Interfacing
Serial Output
The MC126 can interface with computers, printers
and other RS232 compatible devices. The RS232
Interface Kit (Cat. 51302031) converts data from the
6 pin socket to RS232 format.
Type of communication - uni directional, baud rate
1200, data format - 1 start bit : 8 data bits : 1 stop
bit : no parity.
n Data is sent to the output socket at
measurement endpoint (manual or automatic)
and memory recall.
n To use the output to monitor continually, you
need to select continuous measurement mode
and continuous data transfer mode.
CAUTION: The MC126 does not meet IP67 when the
waterproof output socket seal has been removed.
Maintenance
General
The MC126 requires very little maintenance.
Occasionally wipe the meter with a damp cloth. The
casework is made of ABS/PC which is known to be
affected by some organic solvents, including
toluene, xylene and methyl-ethyl-ketone. It is good
practice to wipe away any spillages as soon as they
occur.
Every six months inspect and lightly grease the
socket seals, and connector and battery ‘O’ rings
with the grease supplied. Replace the ‘O’ rings if
damaged.
CAUTION: To prevent permanent damage care
should be taken to prevent liquid entering the
MC126 when fitting probes or replacing batteries. If
it does, remove the batteries and allow the meter to
dry before using.
Replacing the Batteries
Replace batteries as soon as the low battery
indicator appears. To prevent data loss always turn
the MC126 off using the mode key before replacing
the batteries. See the Installation section for details
on installing batteries. Do not mix old and new
batteries.
NOTE: Check the Program Menu settings after
changing the batteries.
Probe Maintenance
CAUTION: To prevent static damage to the MC126
always disconnect the conductivity probe from the
meter before cleaning or removing the shield.
Refer to the conductivity probe product insert for full
details on maintaining your probe.
Problem Solving
•
+ Err 1 - measurement out of range
Check probe is connected and immersed in sample.
Err 2 - cal 1 out of range
Clean the probe and shield, or replace.
Err 3 - cal 2 out of range
Check correct standard is being used. Check correct
solids factor is being used (for Mettler Toledo
standards always use a factor of 0.5). Clean the
probe and shield, or replace.
Data Entry Errors
Entered solids factor value changes to 0.40 or 1.00
- the MC126 will not accept solids factors outside
this range.
Entered temperature coefficient value changes to 0.0
or 10.0 - the MC126 will not accept temperature
coefficients outside this range.
Display/Controls Inactive
Low battery power or batteries fitted incorrectly remove batteries and replace correctly (with new
batteries if necessary).
Operating Hints
n Make sure the plastic shield is always in place
when the probe is connected to the MC126, and
when measuring.
n When measuring make sure the solution is
above the cell chamber rings and below the vent
hole.
n Make sure the shield is free of bubbles when
measuring. To reduce air bubbles, immerse
probe at an angle and then raise to vertical
position.
n Rinse probe with distilled water between
measurements, and after use.
n Do not use standards after the expiry date.
n For greatest accuracy, standards and sample
should be at the same temperature, preferably
the reference temperature.
Calibrating
1. Select conductivity or TDS mode
2. 1-point calibration
Example
Auto endpoint or
to return to sample measurement, or
continue with step 3 for the second cal
point
3. 2-point calibration - repeat step 2 using standard
auto or manual endpoint
4. Rinse probe
hold in
air
ATC
cal1
0.03
25.0°c
µS
cal1
0.00
µS
ATC
25.0°c
µS
ATC
25.0°c
cal2
1413.
Measuring Samples
1. Select conductivity or TDS mode
2. Measure sample
Example
Auto endpoint or
3. Rinse probe
n If you press mode to select conductivity or TDS, the
MC126 begins measuring immediately; you will not
need to press read to start the sample measurement.
n The decimal point flashes during measurement.
When the probe output has stabilized
the stability indicator appears.
Manual endpoint - press read to endpoint
Auto endpoint - the MC126 automatically endpoints
ATC
25.0°c
µS
1253.
sample
ATC
25.0°c
µS
1253.
Using the Memory
Entering a Reading into Memory
The MC126 can store up to 10 endpointed results.
Press M when the measurement has endpointed.
M 1 is displayed (or M 2 to M 10 if readings have
already been stored).
M indicates the memory is full.
Recalling Memory
You can only recall stored memories if the current
measurement has endpointed.
Press R - the last stored memory is displayed. Press
∨
to scroll through memories. RM 1 to RM 10 indicates
which memory is being displayed.
Clearing last Memory
Press R - the last stored memory is displayed. Press M
to clear this memory. Only the last entered measurement can be cleared, e.g. M 2 cannot be cleared
without first clearing M 3.
Clearing all Memories
Press R then press ∨ until M C is displayed.
Press M to clear the memories (press read to exit
without clearing the memories).
Using Continuous
Measurement Mode
In normal mode, the MC126 will auto-off 10 minutes
after the last key press, whether the unit is measuring
or has endpointed. If you select continuous
measurement mode, the MC126 will not auto-off.
To select continuous measurement mode press and
hold M for 2 seconds.
Using Continuous
Data Transfer Mode
In continuous data transfer mode readings are sent to
the data output approximately every second. If the
measurement endpoints (manually, or using auto
endpoint) data transfer stops.
To select continuous data transfer mode press and hold
R for 2 seconds.
n If you want to use this mode to monitor continually,
you also need to select continuous measurement
mode.
n To maintain continuity of sample readings
calibration data is not output in continuous data
transfer mode.
H2O
H
2
O