Please contact Mettler-Toledo before returning a defective device. Ship the cleaned device to the address
you have been given. If the device has been in contact with process fluids, it must be decontaminated/
disinfected before shipment.
In that case, please attach a corresponding certificate, for the health and safety of our service personnel.
Disposal
Please observe the applicable local or national regulations concerning the disposal of
“waste electrical and electronic equipment”.
Registered trademarks
The following registered trademarks are used in this instruction manual
without further marking
For modules which are not listed here, refer to corresponding module
instruction manual.
ModuleSoftware version
pH 2700 (X)2.x
pH 2700i (X)1.x
Cond 7700(X)2.x
Cond Ind 7700(X)2.x
O2 4700(X)2.x
O2 4700(X) ppb2.x
O2 4700i(X)1.x
O2 4700i(X) ppb1.x
Out 700(X) 1.x
PID 700(X)1.x
PA 700(X)1.x
EC 700(X)1.x
Query actual device/module software
When the analyzer is in measuring mode: press menu key, open Diagnostics
menu.
MenuDisplayDevice description
0.003 mS/cm
22.7 °C
Device description
Module FRONT M 700-011
Operating panel M 700
Hardware: 2, Software: 6.0
Serial number: 0000815
Provides information about all modules installed: Module type and function, serial number, hardware and
software version and device options.
Select the different modules (FRONT,
Module
FRONT
Return
BASE
Options
BASE, slots 1 - 3) using the arrow
keys.
6
M 700(X)
Modular concept and instruction manuals
Instruction manuals for basic unit, measuring module, additional functions.
The M 700(X) is an expandable modular process analysis system.
The basic unit (M 700(X) FRONT and M 700(X) BASE) provides three slots
which can be equipped by the user with any combination of measuring or
communication modules. The software capabilities can be expanded by additional functions (options). Additional functions must be ordered separately.
They are supplied with a device-specific TAN for function release.
M 700(X) modular process analysis system
n
o
i
t
Additional functions
p
O
Activation via device-
e
l
b
a
r
it
e
f
t
o
r
specific TAN
Measuring modules
• pH / ORP / Temp
•02/Temp
• Noncontacting conductivity / Temp
• Contacting conductivity / Temp
SmartMedia card
Data recording
3 module slots
for free combination of
measuring and
communication modules
Communication modules
• Out 700 (additional switching and
current outputs)
• PID 700 (analog and digital controller)
•PA 700 (Profibus PA)
• The instruction manual for the M 700(X) describes how to install,
commission and operate the basic unit.
• The instruction manual for the measuring or communication module
describes all functions required for commissioning and working with
the respective measuring or communication module.
• Additional functions are supplied with a function description.
7
System overview
M 700(X) modular process analysis system:
Measuring modules and communication modules
pH 2700(X)
module
pH, ORP
MΩ, kΩ
I 3
mA
Out 700(X)
module
Interface
Glass
2 EL / 4 EL
Reference
ORP
RTD
RTD
Temp
µS, mS/cm
% by wt
Temp
µS, mS/cm
% by wt
Interface
Cond 7700(X)
module
Interface
Cond Ind
7700(X)
module
I 4
IV 1
IV 2
mA
mA
mA
Electronic
contacts
PID 700(X)
module
Interface
Electronic
contacts
Electrodeless sensor
O2, %
µg/l, ppb
mg/l, ppm
O2 sensor
8
Temp
Temp
Interface
O2 4700(X)
O2 4700(X) ppb
module
Interface
PA 700(X)
module
PROFIBUS-PA
COMPA
Interface
M 700(X)
System overview
M 700(X) modular process analysis system:
Basic unit and control module for retractable housings
Module
slot
Module
slot
Module
slot
EC 400(X)
SMARTMEDIA
OK 1
VariPower
M 700 (X)
OK 2
Contacts
K1, K2, K3
I 1
I 2
Alarm
K4
EC 700(X)
mA
mA
0(4) ... 20 mA
pH
0(4) ... 20 mA
Temp
Probe
RTD
EC 400(X)
pH, ORP
MΩ, kΩ
Temp
Interface
9
Latest product information
Additional functions
The M 700(X) is an expandable modular process analysis system.
For latest product information, please refer to:
www.mtpro.com
Additional functions
A current version of the respective function description is available as pdf for
download.
10
M 700(X)
How to order an additional function
Device-specific TAN (transaction number)
Additional functions expand the device capabilities.
The additional functions are device-specific. When ordering an additional
function, you therefore have to specify the serial number and hardware
version of your FRONT module in addition to the respective order number.
The manufacturer then supplies a TAN (transaction number) to release the
additional function.
Serial number M 700 FRONT
Menu
DisplaySerial number M 700 FRONT
0.002 mS/cm
23.7 °C
Menu selection
Menu selection
Call up diagnostics.
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys,
Select:
[enter]
confirm with enter.
Return to meas
Diagnostics
Message list
Logbook
Device description
M 700 FRONT
M 700 BASE
Module Cond Ind 7700
Return
Device description
Module FRONT 700-011
Operating panel M 700
Hardware: 2, Software: 6.0
Serial number: 0000815
Lingua
0.003 mS/cm
23.7 °C
Set favorite
0.003 mS/cm
22.7 °C
Diagnostics
Select Device description using arrow
keys, confirm with enter.
Device description
Please specify this
serial number
and the hardware version when
ordering an additional function.
Module
FRONT
Return
BASE
11
Intended use
The M 700(X) modular process analysis system is preferably used to measure
and process electrochemical quantities in liquids. It has a modular design and
consists of the BASE power supply unit, the FRONT door and different measuring and communication modules.
The M 700X is intended for operation in locations subject to explosion hazards which require equipment of Group II, device category 2(1), gas/dust.
The M 700(X) is a flexible measuring system for continuous measurements in
the field of liquid analysis. Thanks to its modular design, the M 700(X) can
easily be adapted to your measuring task. Flexible use of plug-in modules
allows combined measurements as well as later expansions or modifications.
The measured variables depend on the measuring modules installed.
Communication modules are available for further processing of the output
signals. The rugged enclosure (IP 65) can be wall or pipe mounted or fixed
into a control panel. The M 700 version with hygienic, polished stainless steel
enclosure allows application in the field of biotechnology, food processing,
and in the pharmaceutical industry. The M 700 version with coated steel
enclosure – extremely corrosion resistant – has been developed for application in the chemical industry, environmental engineering, water and wastewater treatment, and for application in power plants .
Caution!
Never expose the display to direct sun light! Only operate the display within
the temperature range of 0 °C up to 50 °C max.
Package contents
• M 700(X) basic unit (FRONT and BASE)
• Wall-mount kit
• Test Certificate
• This instruction manual
• EC Declaration of Conformity
• EC-Type-Examination Certificate (M 700X)
Modules as ordered (each in a separate package with Instruction Manual,
Test Certificate, EC Declaration of Conformity):
12
M 700(X)
Safety information
Application in hazardous locations
M 700X modular process analysis system
The M 700X modular process analysis system is intended for operation in
specific environments and specific fields of application. These are listed in the
instruction manual as specifications for environment, installation and commissioning, intended use (= application), assembly and dismantling, and
maintenance.
Observe the influences of humidity, ambient temperature, chemicals and corrosion. If the specifications in the instruction manual are not sufficient for
assessing the safety of operation, e.g. because your specific applications are
not described, please contact the manufacturer to make sure that the application is possible and safe.
Prerequisite to safe use of the equipment is the observance of the specified
ambient conditions and temperature ranges.
When using the M 700X modular process analysis system, the stipulations
for electrical installations in hazardous areas (EN 60079-14) must be
observed. When installing the device outside the range of applicability of the
94/9/EC directive, the appropriate standards and regulations in the country
of use must be observed.
The module has been developed and manufactured in compliance with the
applicable European guidelines and standards. Compliance with the
European Harmonized Standards for use in hazardous locations is confirmed
by the EC-Type-Examination Certificate. Compliance with the European
guidelines and standards is confirmed by the EC Declaration of Conformity.
The EC Declaration of Conformity and the EC-Type-Examination Certificate
are included in the instruction manual.
There is no particular direct hazard caused by the operation of the device in
the specified environment.
13
Safety information
During operation, the M 700(X)
modular process analysis system
may be opened briefly to replace
the SmartMedia card. The mains terminal cover must be opened only
when the unit is de-energized.
Installation:
The power supply must be disconnectable using a 2-pole disconnect
switch installed near the M700(X).
This switch must meet the requirements of EN 60947-1 and EN 60947-3,
be marked as disconnecting device for M 700(X) and be easily accessible by
the user.
Information on commissioning/start-up
Caution!
• Before commissioning it must be proved that the device may be
connected with other equipment.
• Commissioning must only be performed by authorized and trained personnel!
• The combination of hazardous-area and safe-area modules
(mixed configuration) is not permitted.
Whenever it is likely that protection has been impaired, the device shall be
made inoperative and secured against unintended operation. The protection is
likely to be impaired if, for example:
• the device shows visible damage
• the device fails to perform the intended measurements
• after prolonged storage at temperatures above 70 °C
• after severe transport stresses
Before recommissioning the device, a professional routine test in accordance
with EN 61010-1 must be performed. This test should be carried out by the
manufacturer.
14
M 700(X)
Conformity with FDA 21 CFR Part 11
In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11,
Electronic Records; Electronic Signatures” the US American health agency
FDA (Food and Drug Administration) regulates the production and processing
of electronic documents for pharmaceutical development and production.
This results in requirements for measuring devices used for corresponding
applications. The following features ensure that the M 700(X) modular
process analysis system meets the demands of FDA 21 CFR Part 11:
Electronic Signature
Access to the device functions is regulated and limited by individually
adjustable codes – “Passcodes”. This prevents unauthorized modification of
device settings or manipulation of the measurement results. Appropriate use
of these passcodes makes them suitable as electronic signature.
Audit Trail Log
Every change of device settings can be automatically recorded and documented in the Audit Trail Log on the SmartMedia card. The recording can be
encoded.
M 700 SBasic unit, stainless steel enclosure52121174
M 700 CBasic unit, coated steel enclosure52121171
pH 2700Module: pH52121182
pH 2700iModule: pH, ISM52121161
Cond 7700Module: Conductivity52121184
Cond Ind 7700Module: Electrodeless cond.52121186
O
4700Module: Oxygen (standard)52121188
2
O
4700 ppbModule: Oxygen (traces)52121190
2
O
4700iModule: Oxygen (standard, ISM)52121163
2
O
4700i ppbModule: Oxygen (traces, ISM)52121165
2
Out 700Module: Output expansion52121177
PA 700Module: Profibus PA52121210
PID 700Module: PID controller52121179
M 700X S/VPWBasic unit, stainless steel enclosure52121175
with VariPower power supply unit
M 700X S/24VBasic unit, stainless steel enclosure52121176
with 24 V power supply unit
M 700X C/VPWBasic unit, coated steel enclosure52121172
with VariPower power supply unit, AC
M 700X C/24VBasic unit, standard enclosure52121173
with 24 V power supply unit, AC/DC
pH 2700XModule: pH52121183
pH 2700XModule: pH52121162
Cond 7700XModule: Conductivity52121185
Cond Ind 7700XModule: Electrodeless cond.52121187
O
4700XModule: Oxygen (standard)52121189
2
O
4700X ppbModule: Oxygen (traces)52121191
2
O
4700iXModule: Oxygen (standard, ISM)52121164
2
O
4700iX ppbModule: Oxygen (traces, ISM)52121166
2
Out 700XModule: Output expansion52121178
PA 700XModule: Profibus PA52121181
PID 700XModule: PID controller52121180
KI recorderTAN SW 700-00152121198
Buffer sets to be entered (pH)TAN SW 700-00252121199
Adaptive cal timer (pH)TAN SW 700-00352121200
ServiceScope (pH)TAN SW 700-00452121201
Tolerance band recorder (pH)TAN SW 700-00552121202
Current characteristic definableTAN SW 700-00652121203
TC ultrapure water (Cond)TAN SW 700-00852121204
Concentration determination (Cond) TAN SW 700-00952121205
High CO
compensation (O2)TAN SW 700-01152121250
2
ISFET for module pH 2700i(X)TAN SW 700-01252121274
5 loadable parameter setsTAN SW 700-10252121192
Measurement recorderTAN SW 700-10352121193
Extended logbookTAN SW 700-10452121194
Software updateTAN SW 700-10652121195
AuditTrail to FDA 21 CFR Part 11AuditTrail card/TAN, SW 700-10752121196
Message list
Meas. point
description
Logbook
Device description
6
FRONT
BASE
Module 1
Module 2
Module 3
Legend:
(1) Pressing the menu key accesses menu selection
(2) Pressing the meas key returns to measurement
(3) Menu groups are selected using the arrow keys
(4) Press enter to confirm, enter passcode
(5) Further menu items are displayed
(6) Selected functions of the Diagnostics menu can be
recalled via softkey even when in measuring mode (Pg 34)
19
Short description: M 700 FRONT
M 700
View into the open device (M 700 FRONT)
Slot for SmartMedia card
• Data recording
The SmartMedia card expands the
measurement recorder capacity to
> 50000 records.
• Exchange of parameter sets
5 parameter sets can be stored
on the SmartMedia card, 2 of them
can be loaded to the M 700
and switched by remote control.
Configurations can be transferred
from one M 700 to the other.
• Function expansions
are possible with additional software
modules which are released using
transaction numbers (TAN).
• Software updates
Terminal plates
of “hidden” modules
Each module comes with an adhesive
label containing the contact assignments.
This label should be sticked to the inner
side of the front (as shown).
Then, the terminal assignments remain
visible even if further modules are
inserted.
Replacing the front module
Pull off power cord and ground wire.
To separate the M 700 FRONT from
the M 700 BASE, turn the retaining screws
of the pivot hinge by 90°.
The circumferential sealing
guarantees IP 65 protection and allows spray
cleaning / disinfection.
Caution! Keep clean!
20
M 700(X)
Short description: M 700 BASE
M 700
View into the open device (M 700 BASE, 3 function modules installed)
Module equipment
Module identification: Plug & Play
Up to 3 modules can be combined as
desired. Several input and communication
modules are available
(Overview on Page 16)
M 700 BASE
2 current outputs (free assignment of
process variable) and 4 relay contacts,
2 digital inputs.
VariPower broad-range power supply
unit, 20 ... 265 V AC/DC, suitable for all
public mains supplies in the world.
Power supply units, IS version:
100 ... 230 V AC or
24 V AC/DC
Warning!
Do not touch the terminal compartment, there may be
dangerous contact voltages!
Important note concerning SmartMedia card
The SmartMedia card may be inserted or replaced with the power supply
switched on. Before a memory card is removed, it must be “closed” in the
maintenance menu. When closing the device, make sure that the sealing is
properly seated and clean.
21
Connection of power supply
Module M 700 C / M 700 S (non-IS)
Connection of power supply (BASE M 700 C / M 700 S, non-IS)
The M 700(X) comes in three different versions.
The terminal plates and wirings are shown on the following pages.
1. BASE M 700 C / M 700 S (standard version, non-IS)
VariPower broad-range power supply unit, 24 (-15 %) ... 230 (+15%) V AC/DC
2. BASE M 700X C/VPW / M 700X S/VPW (IS version)
VariPower broad-range power supply unit
3. BASE M 700X C/24V / M 700X S/24V (IS version)
24 V power supply unit
22
M 700(X)
1.BASE M 700 C / M 700 S (non-IS)
Standard version. Not suitable for hazardous-area applications!
Information on installation
Caution!
• Installation may only be carried out by trained and authorized
experts in accordance with this instruction manual and as per
applicable local and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the insulation.
• All parameters must be set by the system administrator
prior to commissioning.
Connection of power supply
With the VariPower broad-range power supply unit, the analyzer can
be operated with a power supply of 24 (-15 %) to 230 (+15 %)
V AC/DC making it suitable for all public mains supplies in the
world.
The terminals are suitable for single wires and flexible leads up to
2
2.5 mm
Terminal plate BASE M 700
Standard version. Not suitable for hazardous-area applications!
Connection of power supply. Contact assignment of inputs/outputs.
(AWG 14).
23
2.BASE M 700X C/VPW / M 700X S/VPW
IS version with VariPower power supply unit
Information on installation
When using the M 700X modular process analysis system, the stipulations for electrical installations in hazardous areas (EN 60079-14) must
be observed. When installing the device outside the range of applicability of the 94/9/EC directive, the appropriate standards and regulations in the country of use must be observed.
Caution!
• Installation may only be carried out by trained and authorized experts
in accordance with this instruction manual and as per applicable local
and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the insulation.
• All parameters must be set by the system administrator prior to
commissioning.
Connection of power supply
With the VariPower broad-range power supply unit, the analyzer can
be operated with a power supply of 100 to 230 V AC (-15 %, +10 %)
(EEx em IIC).
The terminals are suitable for single wires and flexible leads up to
2
2.5 mm
Terminal plate BASE M 700X C/VPW / M 700X S/VPW
(IS version with VariPower power supply unit)
Connection of power supply. Contact assignment of inputs/outputs.
(AWG 14).
24
M 700(X)
3.BASE M 700X C/24V / M 700X C/24V
IS version with 24 V power supply unit
Information on installation
When using the M 700X modular process analysis system, the stipulations for electrical installations in hazardous areas (EN 60079-14) must
be observed. When installing the device outside the range of applicability of the 94/9/EC directive, the appropriate standards and regulations in the country of use must be observed.
Caution!
• Installation may only be carried out by trained and authorized
experts in accordance with this instruction manual and as per
applicable local and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the insulation.
• All parameters must be set by the system administrator
prior to commissioning.
Connection of power supply
With the power supply unit, the analyzer can be operated with a
power supply of 24 V AC (-15 %, +10%) or 24 V DC (-15 %, +20%).
The terminals are suitable for single wires and flexible leads up to
2
2.5 mm
Terminal plate BASE M 700X (C/S)/24V
(IS version with 24 V power supply unit)
Connection of power supply. Contact assignment of inputs/outputs.
(AWG 14).
25
Hazardous-area connection M 700X
24 V
GND
24 V
GND
Logic input
Logic input
Safe area
A
A
B
B
B
Hazardous
area
31
33
30
61
63
65
60
M 700X
OK1
OK1
OK1 ... OK2
K1
K2
K3
K1 ... K3
B
0/4 ... 20 mA
0/4 ... 20 mA
Mains connection: laid in Ex „e”
C
C
71
72
51
52
53
54
L1
PE
L2
K4 (Alarm)
I1
I2
Ex „i”
N
L
Ex „e”
26
M 700(X)
Hazardous-area control modules (examples)
DesignationModelManufacturer
Valve control moduleKFD2-SL-Ex 1.48****Pepperl + FuchsA
Valve control moduleMK 72-S17-Ex0/24VDCTURCK
Switch amplifierKF**-SR2-Ex1.W.**Pepperl + FuchsB
Switch amplifierMK1-22Ex0-R/**TURCK
Loop-powered isolatorIsoTrans®36A7KnickC
27