Please contact Mettler-Toledo before returning a defective device. Ship the cleaned device to the address
you have been given. If the device has been in contact with process fluids, it must be decontaminated/
disinfected before shipment.
In that case, please attach a corresponding certificate, for the health and safety of our service personnel.
Disposal
Please observe the applicable local or national regulations concerning the disposal of
“waste electrical and electronic equipment”.
Registered trademarks
The following registered trademarks are used in this instruction manual
without further marking
For modules which are not listed here, refer to corresponding module
instruction manual.
ModuleSoftware version
pH 2700 (X)2.x
pH 2700i (X)1.x
Cond 7700(X)2.x
Cond Ind 7700(X)2.x
O2 4700(X)2.x
O2 4700(X) ppb2.x
O2 4700i(X)1.x
O2 4700i(X) ppb1.x
Out 700(X) 1.x
PID 700(X)1.x
PA 700(X)1.x
EC 700(X)1.x
Query actual device/module software
When the analyzer is in measuring mode: press menu key, open Diagnostics
menu.
MenuDisplayDevice description
0.003 mS/cm
22.7 °C
Device description
Module FRONT M 700-011
Operating panel M 700
Hardware: 2, Software: 6.0
Serial number: 0000815
Provides information about all modules installed: Module type and function, serial number, hardware and
software version and device options.
Select the different modules (FRONT,
Module
FRONT
Return
BASE
Options
BASE, slots 1 - 3) using the arrow
keys.
6
M 700(X)
Modular concept and instruction manuals
Instruction manuals for basic unit, measuring module, additional functions.
The M 700(X) is an expandable modular process analysis system.
The basic unit (M 700(X) FRONT and M 700(X) BASE) provides three slots
which can be equipped by the user with any combination of measuring or
communication modules. The software capabilities can be expanded by additional functions (options). Additional functions must be ordered separately.
They are supplied with a device-specific TAN for function release.
M 700(X) modular process analysis system
n
o
i
t
Additional functions
p
O
Activation via device-
e
l
b
a
r
it
e
f
t
o
r
specific TAN
Measuring modules
• pH / ORP / Temp
•02/Temp
• Noncontacting conductivity / Temp
• Contacting conductivity / Temp
SmartMedia card
Data recording
3 module slots
for free combination of
measuring and
communication modules
Communication modules
• Out 700 (additional switching and
current outputs)
• PID 700 (analog and digital controller)
•PA 700 (Profibus PA)
• The instruction manual for the M 700(X) describes how to install,
commission and operate the basic unit.
• The instruction manual for the measuring or communication module
describes all functions required for commissioning and working with
the respective measuring or communication module.
• Additional functions are supplied with a function description.
7
System overview
M 700(X) modular process analysis system:
Measuring modules and communication modules
pH 2700(X)
module
pH, ORP
MΩ, kΩ
I 3
mA
Out 700(X)
module
Interface
Glass
2 EL / 4 EL
Reference
ORP
RTD
RTD
Temp
µS, mS/cm
% by wt
Temp
µS, mS/cm
% by wt
Interface
Cond 7700(X)
module
Interface
Cond Ind
7700(X)
module
I 4
IV 1
IV 2
mA
mA
mA
Electronic
contacts
PID 700(X)
module
Interface
Electronic
contacts
Electrodeless sensor
O2, %
µg/l, ppb
mg/l, ppm
O2 sensor
8
Temp
Temp
Interface
O2 4700(X)
O2 4700(X) ppb
module
Interface
PA 700(X)
module
PROFIBUS-PA
COMPA
Interface
M 700(X)
System overview
M 700(X) modular process analysis system:
Basic unit and control module for retractable housings
Module
slot
Module
slot
Module
slot
EC 400(X)
SMARTMEDIA
OK 1
VariPower
M 700 (X)
OK 2
Contacts
K1, K2, K3
I 1
I 2
Alarm
K4
EC 700(X)
mA
mA
0(4) ... 20 mA
pH
0(4) ... 20 mA
Temp
Probe
RTD
EC 400(X)
pH, ORP
MΩ, kΩ
Temp
Interface
9
Latest product information
Additional functions
The M 700(X) is an expandable modular process analysis system.
For latest product information, please refer to:
www.mtpro.com
Additional functions
A current version of the respective function description is available as pdf for
download.
10
M 700(X)
How to order an additional function
Device-specific TAN (transaction number)
Additional functions expand the device capabilities.
The additional functions are device-specific. When ordering an additional
function, you therefore have to specify the serial number and hardware
version of your FRONT module in addition to the respective order number.
The manufacturer then supplies a TAN (transaction number) to release the
additional function.
Serial number M 700 FRONT
Menu
DisplaySerial number M 700 FRONT
0.002 mS/cm
23.7 °C
Menu selection
Menu selection
Call up diagnostics.
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys,
Select:
[enter]
confirm with enter.
Return to meas
Diagnostics
Message list
Logbook
Device description
M 700 FRONT
M 700 BASE
Module Cond Ind 7700
Return
Device description
Module FRONT 700-011
Operating panel M 700
Hardware: 2, Software: 6.0
Serial number: 0000815
Lingua
0.003 mS/cm
23.7 °C
Set favorite
0.003 mS/cm
22.7 °C
Diagnostics
Select Device description using arrow
keys, confirm with enter.
Device description
Please specify this
serial number
and the hardware version when
ordering an additional function.
Module
FRONT
Return
BASE
11
Intended use
The M 700(X) modular process analysis system is preferably used to measure
and process electrochemical quantities in liquids. It has a modular design and
consists of the BASE power supply unit, the FRONT door and different measuring and communication modules.
The M 700X is intended for operation in locations subject to explosion hazards which require equipment of Group II, device category 2(1), gas/dust.
The M 700(X) is a flexible measuring system for continuous measurements in
the field of liquid analysis. Thanks to its modular design, the M 700(X) can
easily be adapted to your measuring task. Flexible use of plug-in modules
allows combined measurements as well as later expansions or modifications.
The measured variables depend on the measuring modules installed.
Communication modules are available for further processing of the output
signals. The rugged enclosure (IP 65) can be wall or pipe mounted or fixed
into a control panel. The M 700 version with hygienic, polished stainless steel
enclosure allows application in the field of biotechnology, food processing,
and in the pharmaceutical industry. The M 700 version with coated steel
enclosure – extremely corrosion resistant – has been developed for application in the chemical industry, environmental engineering, water and wastewater treatment, and for application in power plants .
Caution!
Never expose the display to direct sun light! Only operate the display within
the temperature range of 0 °C up to 50 °C max.
Package contents
• M 700(X) basic unit (FRONT and BASE)
• Wall-mount kit
• Test Certificate
• This instruction manual
• EC Declaration of Conformity
• EC-Type-Examination Certificate (M 700X)
Modules as ordered (each in a separate package with Instruction Manual,
Test Certificate, EC Declaration of Conformity):
12
M 700(X)
Safety information
Application in hazardous locations
M 700X modular process analysis system
The M 700X modular process analysis system is intended for operation in
specific environments and specific fields of application. These are listed in the
instruction manual as specifications for environment, installation and commissioning, intended use (= application), assembly and dismantling, and
maintenance.
Observe the influences of humidity, ambient temperature, chemicals and corrosion. If the specifications in the instruction manual are not sufficient for
assessing the safety of operation, e.g. because your specific applications are
not described, please contact the manufacturer to make sure that the application is possible and safe.
Prerequisite to safe use of the equipment is the observance of the specified
ambient conditions and temperature ranges.
When using the M 700X modular process analysis system, the stipulations
for electrical installations in hazardous areas (EN 60079-14) must be
observed. When installing the device outside the range of applicability of the
94/9/EC directive, the appropriate standards and regulations in the country
of use must be observed.
The module has been developed and manufactured in compliance with the
applicable European guidelines and standards. Compliance with the
European Harmonized Standards for use in hazardous locations is confirmed
by the EC-Type-Examination Certificate. Compliance with the European
guidelines and standards is confirmed by the EC Declaration of Conformity.
The EC Declaration of Conformity and the EC-Type-Examination Certificate
are included in the instruction manual.
There is no particular direct hazard caused by the operation of the device in
the specified environment.
13
Safety information
During operation, the M 700(X)
modular process analysis system
may be opened briefly to replace
the SmartMedia card. The mains terminal cover must be opened only
when the unit is de-energized.
Installation:
The power supply must be disconnectable using a 2-pole disconnect
switch installed near the M700(X).
This switch must meet the requirements of EN 60947-1 and EN 60947-3,
be marked as disconnecting device for M 700(X) and be easily accessible by
the user.
Information on commissioning/start-up
Caution!
• Before commissioning it must be proved that the device may be
connected with other equipment.
• Commissioning must only be performed by authorized and trained personnel!
• The combination of hazardous-area and safe-area modules
(mixed configuration) is not permitted.
Whenever it is likely that protection has been impaired, the device shall be
made inoperative and secured against unintended operation. The protection is
likely to be impaired if, for example:
• the device shows visible damage
• the device fails to perform the intended measurements
• after prolonged storage at temperatures above 70 °C
• after severe transport stresses
Before recommissioning the device, a professional routine test in accordance
with EN 61010-1 must be performed. This test should be carried out by the
manufacturer.
14
M 700(X)
Conformity with FDA 21 CFR Part 11
In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11,
Electronic Records; Electronic Signatures” the US American health agency
FDA (Food and Drug Administration) regulates the production and processing
of electronic documents for pharmaceutical development and production.
This results in requirements for measuring devices used for corresponding
applications. The following features ensure that the M 700(X) modular
process analysis system meets the demands of FDA 21 CFR Part 11:
Electronic Signature
Access to the device functions is regulated and limited by individually
adjustable codes – “Passcodes”. This prevents unauthorized modification of
device settings or manipulation of the measurement results. Appropriate use
of these passcodes makes them suitable as electronic signature.
Audit Trail Log
Every change of device settings can be automatically recorded and documented in the Audit Trail Log on the SmartMedia card. The recording can be
encoded.
M 700 SBasic unit, stainless steel enclosure52121174
M 700 CBasic unit, coated steel enclosure52121171
pH 2700Module: pH52121182
pH 2700iModule: pH, ISM52121161
Cond 7700Module: Conductivity52121184
Cond Ind 7700Module: Electrodeless cond.52121186
O
4700Module: Oxygen (standard)52121188
2
O
4700 ppbModule: Oxygen (traces)52121190
2
O
4700iModule: Oxygen (standard, ISM)52121163
2
O
4700i ppbModule: Oxygen (traces, ISM)52121165
2
Out 700Module: Output expansion52121177
PA 700Module: Profibus PA52121210
PID 700Module: PID controller52121179
M 700X S/VPWBasic unit, stainless steel enclosure52121175
with VariPower power supply unit
M 700X S/24VBasic unit, stainless steel enclosure52121176
with 24 V power supply unit
M 700X C/VPWBasic unit, coated steel enclosure52121172
with VariPower power supply unit, AC
M 700X C/24VBasic unit, standard enclosure52121173
with 24 V power supply unit, AC/DC
pH 2700XModule: pH52121183
pH 2700XModule: pH52121162
Cond 7700XModule: Conductivity52121185
Cond Ind 7700XModule: Electrodeless cond.52121187
O
4700XModule: Oxygen (standard)52121189
2
O
4700X ppbModule: Oxygen (traces)52121191
2
O
4700iXModule: Oxygen (standard, ISM)52121164
2
O
4700iX ppbModule: Oxygen (traces, ISM)52121166
2
Out 700XModule: Output expansion52121178
PA 700XModule: Profibus PA52121181
PID 700XModule: PID controller52121180
KI recorderTAN SW 700-00152121198
Buffer sets to be entered (pH)TAN SW 700-00252121199
Adaptive cal timer (pH)TAN SW 700-00352121200
ServiceScope (pH)TAN SW 700-00452121201
Tolerance band recorder (pH)TAN SW 700-00552121202
Current characteristic definableTAN SW 700-00652121203
TC ultrapure water (Cond)TAN SW 700-00852121204
Concentration determination (Cond) TAN SW 700-00952121205
High CO
compensation (O2)TAN SW 700-01152121250
2
ISFET for module pH 2700i(X)TAN SW 700-01252121274
5 loadable parameter setsTAN SW 700-10252121192
Measurement recorderTAN SW 700-10352121193
Extended logbookTAN SW 700-10452121194
Software updateTAN SW 700-10652121195
AuditTrail to FDA 21 CFR Part 11AuditTrail card/TAN, SW 700-10752121196
Message list
Meas. point
description
Logbook
Device description
6
FRONT
BASE
Module 1
Module 2
Module 3
Legend:
(1) Pressing the menu key accesses menu selection
(2) Pressing the meas key returns to measurement
(3) Menu groups are selected using the arrow keys
(4) Press enter to confirm, enter passcode
(5) Further menu items are displayed
(6) Selected functions of the Diagnostics menu can be
recalled via softkey even when in measuring mode (Pg 34)
19
Short description: M 700 FRONT
M 700
View into the open device (M 700 FRONT)
Slot for SmartMedia card
• Data recording
The SmartMedia card expands the
measurement recorder capacity to
> 50000 records.
• Exchange of parameter sets
5 parameter sets can be stored
on the SmartMedia card, 2 of them
can be loaded to the M 700
and switched by remote control.
Configurations can be transferred
from one M 700 to the other.
• Function expansions
are possible with additional software
modules which are released using
transaction numbers (TAN).
• Software updates
Terminal plates
of “hidden” modules
Each module comes with an adhesive
label containing the contact assignments.
This label should be sticked to the inner
side of the front (as shown).
Then, the terminal assignments remain
visible even if further modules are
inserted.
Replacing the front module
Pull off power cord and ground wire.
To separate the M 700 FRONT from
the M 700 BASE, turn the retaining screws
of the pivot hinge by 90°.
The circumferential sealing
guarantees IP 65 protection and allows spray
cleaning / disinfection.
Caution! Keep clean!
20
M 700(X)
Short description: M 700 BASE
M 700
View into the open device (M 700 BASE, 3 function modules installed)
Module equipment
Module identification: Plug & Play
Up to 3 modules can be combined as
desired. Several input and communication
modules are available
(Overview on Page 16)
M 700 BASE
2 current outputs (free assignment of
process variable) and 4 relay contacts,
2 digital inputs.
VariPower broad-range power supply
unit, 20 ... 265 V AC/DC, suitable for all
public mains supplies in the world.
Power supply units, IS version:
100 ... 230 V AC or
24 V AC/DC
Warning!
Do not touch the terminal compartment, there may be
dangerous contact voltages!
Important note concerning SmartMedia card
The SmartMedia card may be inserted or replaced with the power supply
switched on. Before a memory card is removed, it must be “closed” in the
maintenance menu. When closing the device, make sure that the sealing is
properly seated and clean.
21
Connection of power supply
Module M 700 C / M 700 S (non-IS)
Connection of power supply (BASE M 700 C / M 700 S, non-IS)
The M 700(X) comes in three different versions.
The terminal plates and wirings are shown on the following pages.
1. BASE M 700 C / M 700 S (standard version, non-IS)
VariPower broad-range power supply unit, 24 (-15 %) ... 230 (+15%) V AC/DC
2. BASE M 700X C/VPW / M 700X S/VPW (IS version)
VariPower broad-range power supply unit
3. BASE M 700X C/24V / M 700X S/24V (IS version)
24 V power supply unit
22
M 700(X)
1.BASE M 700 C / M 700 S (non-IS)
Standard version. Not suitable for hazardous-area applications!
Information on installation
Caution!
• Installation may only be carried out by trained and authorized
experts in accordance with this instruction manual and as per
applicable local and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the insulation.
• All parameters must be set by the system administrator
prior to commissioning.
Connection of power supply
With the VariPower broad-range power supply unit, the analyzer can
be operated with a power supply of 24 (-15 %) to 230 (+15 %)
V AC/DC making it suitable for all public mains supplies in the
world.
The terminals are suitable for single wires and flexible leads up to
2
2.5 mm
Terminal plate BASE M 700
Standard version. Not suitable for hazardous-area applications!
Connection of power supply. Contact assignment of inputs/outputs.
(AWG 14).
23
2.BASE M 700X C/VPW / M 700X S/VPW
IS version with VariPower power supply unit
Information on installation
When using the M 700X modular process analysis system, the stipulations for electrical installations in hazardous areas (EN 60079-14) must
be observed. When installing the device outside the range of applicability of the 94/9/EC directive, the appropriate standards and regulations in the country of use must be observed.
Caution!
• Installation may only be carried out by trained and authorized experts
in accordance with this instruction manual and as per applicable local
and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the insulation.
• All parameters must be set by the system administrator prior to
commissioning.
Connection of power supply
With the VariPower broad-range power supply unit, the analyzer can
be operated with a power supply of 100 to 230 V AC (-15 %, +10 %)
(EEx em IIC).
The terminals are suitable for single wires and flexible leads up to
2
2.5 mm
Terminal plate BASE M 700X C/VPW / M 700X S/VPW
(IS version with VariPower power supply unit)
Connection of power supply. Contact assignment of inputs/outputs.
(AWG 14).
24
M 700(X)
3.BASE M 700X C/24V / M 700X C/24V
IS version with 24 V power supply unit
Information on installation
When using the M 700X modular process analysis system, the stipulations for electrical installations in hazardous areas (EN 60079-14) must
be observed. When installing the device outside the range of applicability of the 94/9/EC directive, the appropriate standards and regulations in the country of use must be observed.
Caution!
• Installation may only be carried out by trained and authorized
experts in accordance with this instruction manual and as per
applicable local and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the insulation.
• All parameters must be set by the system administrator
prior to commissioning.
Connection of power supply
With the power supply unit, the analyzer can be operated with a
power supply of 24 V AC (-15 %, +10%) or 24 V DC (-15 %, +20%).
The terminals are suitable for single wires and flexible leads up to
2
2.5 mm
Terminal plate BASE M 700X (C/S)/24V
(IS version with 24 V power supply unit)
Connection of power supply. Contact assignment of inputs/outputs.
(AWG 14).
25
Hazardous-area connection M 700X
24 V
GND
24 V
GND
Logic input
Logic input
Safe area
A
A
B
B
B
Hazardous
area
31
33
30
61
63
65
60
M 700X
OK1
OK1
OK1 ... OK2
K1
K2
K3
K1 ... K3
B
0/4 ... 20 mA
0/4 ... 20 mA
Mains connection: laid in Ex „e”
C
C
71
72
51
52
53
54
L1
PE
L2
K4 (Alarm)
I1
I2
Ex „i”
N
L
Ex „e”
26
M 700(X)
Hazardous-area control modules (examples)
DesignationModelManufacturer
Valve control moduleKFD2-SL-Ex 1.48****Pepperl + FuchsA
Valve control moduleMK 72-S17-Ex0/24VDCTURCK
Switch amplifierKF**-SR2-Ex1.W.**Pepperl + FuchsB
Switch amplifierMK1-22Ex0-R/**TURCK
Loop-powered isolatorIsoTrans®36A7KnickC
27
Panel mounting
Dimension drawings
159,5
Front view
45,5
163
72
147,5
39,5
213
Side view
144
70,5
28
M6 internal (4x)
Rear view
141
72
189
194
r15 ± 3
Control panel cutout
Contour
of
enclosure
Control
panel cutout
Contour of
enclosure
front
M 700(X)
Wall mounting, pipe mounting
Dimension drawings
Ø 30 ... 65
72
Ø 6.5 (4x)
72
150
Wall mounting
Ø 30 ... 65 mm
for vertical or horizontal mounting
Legend:
(1) Pressing the menu key accesses menu selection
(2) Pressing the meas key returns to measurement
(3) Menu groups are selected using the arrow keys
(4) Press enter to confirm, enter passcode
(5) Further menu items are displayed
(6) Selected functions of the Diagnostics menu can be
recalled via softkey even when in measuring mode (Pg 34)
30
M 700(X)
man
Menu selection
M 700 FRONT
After switching on, the M 700 performs an internal test routine and
automatically detects the number and
type of modules installed. Then,
the M 700 goes to measuring mode
(Pg 34).
• Configure measurement display (7)
Pg 34
• Secondary displays/softkeys (8)
Pg 36
Menu selection
(1) Pressing the menu key accesses menu selection
(2) Pressing the meas key returns to measurement
1
Menu selection
01.02.03
2
7.00
25.6 °C
1
7
pH
8
3
4
7.00 pH
25.6 °C
7.00
13.20
24.3°C
(Measuring mode)
Select the desired menu group using the arrow keys (3).
Press enter (4) to confirm your choice.
An overview of the menu structure is given on Pg 30.
pH
%Air
1014 mbar
2
Select:
Return to meas
(Menu selection)
[enter]
Lingua
31
Mode indicators in the display
Icons
The plain-text user interface is supplemented by icons which provide
information on the operating status:
HOLD
Function check is active
SmartMedia card
located in M 700 FRONT
Mode indicator
Parameter setting
Module BASE (Administrator)
Output current I1
Output current I2
Contact K4
Contact K3
Contact K2
Contact K1
Return
(NAMUR Failure)
(NAMUR maintenance)
(NAMUR function check)
(Limit)
0.003 mS/cm
19.0 °C
Block
Module identification
After a menu has been called up, the currently measured values of the respective modules
are displayed here. (number of module slot,
no module designation)
Menu level (Administrator level)
Current selection
is displayed in reverse video.
(Parameters represented in gray cannot be
edited, they have been blocked at the
Administrator level.)
Safety of operation
To ensure increased safety of operation, the
M 700 provides three operating levels:
• Administrator level
Access to all device parameters.
Settings can be blocked against access
from the Operator level.
• Operator level
Access to all functions which have been
released at the Administrator level.
Blocked functions are displayed in gray and
cannot be edited.
•Viewing level
Display of all settings.
No editing possible.
32
M 700(X)
How to enter numbers and text
M 700 FRONT: Operation
Select the position using the left/right keys,
then edit the number or letter using the up/down keys.
Confirm with enter.
Example:
Entering a tag number (point of measurement)
Pressing meas (1) returns the M 700 to the measuring mode from any
function. (Pressing meas in measuring mode, successively displays the
activated special functions such as measurement recorder or KI recorder).
All process variables coming from the modules can be displayed. The table
on the next page describes how to configure the measurement display.
Measurement display
Typical measurement display
, pH modules)
(O
2
7.00
13.20
24.3°C
1
pH
%Air
1014 mbars
Secondary displays
Additional values, also date and
time, can be displayed depending
on the modules installed. They
are selected using the softkeys
(Pg 36).
Softkeys
The softkeys allow selection of
values for the secondary displays.
In addition, Diagnostics functions
which are set as “Favorites” can
be called up.
(Pg 37)
If required, you can also change
the parameter set via softkey
(Pg 37)
Furthermore, the softkeys include
self-explaining, context-sensitive
functions, e.g. with measurement
or KI recorder activated.
34
M 700(X)
MenuDisplayConfigure measurement display
7.00 pH
25.6 °C
Menu selection
Select:
Return to meas
Parameter setting (Administrator)
System control
Module FRONT M 700-011
Module BASE M 700-021
Module O2 4700
Module PH 2700
Module Cond Ind 7700
Return
Module FRONT (Administrator)
Languages
Measurement display
Measurement recorder
[enter]
16.9 %Air
25.6 °C
16.8 %Air
25.6 °C
English
Configure meas. display
Press menu key to select menu.
Select parameter setting using arrow
keys, confirm with enter. Select:
“Administrator level”: Passcode 1989
(For passcodes, see Pg 55)
Lingua
Parameter setting:
Select “M 700 FRONT”
M 700 FRONT:
Select “Measurement display”
Return
16.9 %Air
16.9 %Air
25.6 °C
Measurement display (Administrator)
Main display
1st primary value
2nd primary value
Viewing angle
Abort
Measurement display (Administrator)
Main display
1st primary value
2nd primary value
Viewing angle
Abort
1 primary value
2 primary values
OK
16.9 %Air
25.6 °C
2
mg/l
mg/l
ppm
mbar
pH
values
K
%Air
%O
°C
Measurement display:
Set the number of primary values
(large display) to be displayed
Select process variable(s) to be
displayed and confirm with enter.
Pressing the meas key returns to
measurement.
35
Softkey function (function control)
Select menu: Parameter setting / System control / Function control matrix
In measuring mode you can use the
softkeys (1) to control functions.
The functions are assigned in the
function contr
ol matrix (Fig.)
(Parameter setting/System control).
Softkeys which have not been assigned to
a certain function are automatically used
for selecting the secondary displays.
Secondary display (2)
Display of additional values in the measuring mode when the respective softkey is
pressed. Always active.
You can choose one of the process
variables supplied by the modules (and
Calculation Blocks) or the date or time.
Favorites menu
Selected Diagnostics functions can be
called up directly from the measuring
mode using a softkey. The following table
(Pg 37) explains how to select favorites.
2
1
7.00 pH
25.6 °C
Function control matrix (Administrator)
ParSet KI rec.Fav
Input OK2
Left softkey
Right softkey
Profibus DO 2
Return
Example:
“Parameter set” to be selected with
“Left softkey”
–
–
Connect
Further functions which can be controlled
via softkey:
• Parameter set
• KI recorder
• EasyClean 400
36
To select a softkey function:
Select desired function using arrow
keys, press “Connect” softkey
and confirm with enter.
Deselect function:
Press “Disconnect” softkey,
confirm with enter.
M 700(X)
MenuDisplaySelect favorites
Favorites menu
Diagnostics functions can be called
7.00
11.00
24.3°C
Menu selection
Select:
Return to meas
Diagnostics
Message list
Return
Meas. point description
Logbook
Device description
Module FRONT
Module BASE
pH
pH
Favor ites menu
7.00 pH
25.6 °C
[enter]
Lingua
16.8 %Air
25.0 °C
Set favorite
up directly from the measuring mode
using a softkey.
The “Favorites” are selected in the
Diagnostics menu.
Select favorites
Press menu key to select menu.
Select diagnostics using arrow keys,
confirm with enter.
Set/delete favorite:
“Set favorite” allows activation of
the selected diagnostic function
directly from the measuring mode via
softkey. The respective function is
marked with a heart icon.
(See Softkey usage, Pg 36)
7.00
11.00
24.3°C
Note:
When one of the softkeys has been assigned to the “Favorites menu” function, diagnostic functions which have been set as “Favorite” can be directly
called up from the measuring mode.
pH
pH
Favor ites menu
Pressing the meas key returns to
measurement. When the softkey has
been assigned to “Favorites”,
“Favorites menu” is read in the
secondary display
(see Softkey usage, Pg 36).
37
Diagnostics functions: Overview
Selected diagnostics functions for quality management
Diagnostics functions (quality management, ISO 9000)
To meet the quality management requirements to ISO 9000, the M 700
provides comprehensive diagnostics and safety functions such as Sensocheck
sensor monitoring and CalCheck monitoring of calibration ranges, a logbook
for time- and date-stamped recording of function activations, warning and
failure messages. Further features are:
7 pH25°C
5 pH
20°C
2-channel measurement recorder
Can be called up directly from the measuring mode.
Allows detailed evaluation of events by placing the
cursor on measured values of interest.
S/cm
14/06/02 11:32, Status: Measurement
0.003mS/cm, 23.7°C
Sensor network diagram
1 - Slope
2 - Zero point
3 - Ref impedance
4 - Glass impedance
5 - Response time
6 - Cal timer
7 - Calcheck
Return
ServiceScope
Fail
Noise level
Return
1min/pixel
Info
7.0pH
24.1°C
Info
7.00 pH
0.2 °C
10x
°C
Sensor network diagram
(pH 2700, O24700 modules)
Graphical representation of the sensor parameters in a
network diagram – with slope, zero, reference impedance, glass impedance, response time, calibration timer,
deviation from calibration range (Calcheck).
ServiceScope
(pH module)
Displays the noise levels over the time. Allows distinction
of individual disturbances, periodic and broadband disturbances, which is helpful for troubleshooting.
An error message is generated if the noise level exceeds
the failure limit.
t
38
M 700(X)
Diagnostics functions: Sensoface
Graphical indication of sensor condition
Sensocheck must have been activated during parameter setting
Sensocheck - Sensor monitoring
Module Sensocheck function
4700:Monitoring membrane/electrolyte
O
7.00
pH
2
Cond 7700:Information on sensor condition
11.00
pH
24.3°C
Cond Ind 7700: Information on sensor condition
pH 2700:Automatic monitoring of glass and
reference electrode
The “smileys” provide information on wear and required maintenance of the
sensor (“friendly” – “neutral” – “sad”).
MenuDisplayActivate Sensocheck
7.00 pH
25.6°C
Menu selection
Open menu selection
Select parameter setting
Enter passcode (Administrator)
Select:
Return to meas
[enter]
Lingua
Parameter setting (Administrator)
System control
M 700 FRONT 700-011
M 700 BASE 700-021
Module O2 4700
Module pH 2700
Return
Sensocheck Ref el (Administrator)
Nominal
Min
Max
Message
Abort
7.00 pH
25.6°C
7.00 pH
25.6°C
005.0 kOhm
003.1 kOhm
100.0 kOhm
Off
Failure
Maint. request
OK
Select module
(“pH 2700” or “O2 4700”)
Confirm with enter.
Select “Sensor data”.
Confirm with enter. Then select
“Sensocheck Ref el”
(Fig.)
Assign function and confirm with
enter.
39
Diagnostics functions
General status information of the measuring system
Select menu: Diagnostics
Menu
DisplayDiagnostics functions
0.002 mS/cm
23.7 °C
Menu selection
Call up diagnostics
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys,
confirm with enter.
Select:
Return to meas
Diagnostics
Message list
Logbook
Device description
Module FRONT 700-011
Module BASE 700-021
Module Cond 7700
Return
Message list
Fail Current IV1 load
[enter]
Lingua
0.003 mS/cm
23.7 °C
Set favorite
0.003 mS/cm
23.7 °C
1 messg.
The “Diagnostics” menu gives an
overview of all functions available.
Functions which have been set as
“Favorite” can be directly accessed
from the measuring mode
(see Pg 37).
Message list
Shows the currently activated warning or failure messages in plain text.
Return
Logbook
06/11/02 15:17 Measurement active
06/11/02 15:00 Diagnostics active
06/11/02 10:05 Measurement active
06/11/02 09:50 Parameter setting active
06/11/02 09:42 Wrong passcode
06/10/02 14:33 Measurement active
06/10/02 14:11 Parameter setting active
Return
40
0.003 mS/cm
23.7 °C
Logbook
Shows the last 50 events with date
and time, e.g. calibrations, warning
and failure messages, power failure
etc. This permits quality management
documentation to ISO 9000.
Extended logbook:
SmartMedia card (SW 700-104)
M 700(X)
MenuDisplayDiagnostics functions
Device description
Module FRONT 700-011
Operating panel M 700
Hardware: 2, Software: 6.0
Serial number: 0000815
Module
Module FRONT 700-011
Module diagnostics
Display test
Keypad test
Keypad test
FRONT
Return
Return
BASE
0.003 mS/cm
22.7 °C
0.003 mS/cm
22.7 °C
0.003 mS/cm
22.7 °C
Device description
Provides information about all
modules installed: Module type and
function, serial number, hardware
and software version and device
options (Example: FRONT).
M 700 FRONT
The module the display and keypad
control. Test possibilities:
• Module diagnostics
• Display test
• Keypad test
Example:
M 700 FRONT, keypad test.
Return (2x)
Module BASE 700-021
Module diagnostics
Input/output status
Return
Input/output status
Current load I1
Current load I2
Contact
Input OK1
Input OK2
0.003 mS/cm
22.7 °C
Set favorite
0.003 mS/cm
22.7 °C
ok
ok
K1
K3
Inactive
Inactive
Correct functioning of each key can
be checked by pressing it down.
M 700 BASE
The module generates the standard
output signals. Test possibilities:
From the measuring mode:
Press menu key to select menu.
Select parameter setting using arrow
keys, confirm with enter.
Administrator level:
Access to all functions, also passcode
setting.
Releasing or blocking a function for
access from the Operator level.
Module FRONT (Administrator)
Languages
Measurement display
Measurement recorder
KI recorder
Return
Module FRONT
Languages
Measurement display
Measurement recorder
KI recorder
Return
English
Release
English
25.6 °C
0.003 mS/cm
25.6 °C
0.003 mS/cm
Functions which can be blocked for
the Operator level are marked with
the “lock” symbol.
The functions are released or
blocked using the softkey.
Operator level:
Access to all functions which have
been released at the Administrator
level. Blocked functions are displayed
in gray and cannot be edited (Fig.).
Viewing level
Display of all settings.
No editing possible!
42
M 700(X)
Parameter setting: Lock functions
Administrator level: Enable / lock functions for Operator level
Note: Function check active
MenuDisplay
Parameter setting (Administrator)
System control
Module FRONT
Module BASE
Module Cond Ind 7700
Module PID 700
Module Out 700
M 700 BASE (Administrator)
Output current I2
Contact K4 (NAMUR Failure)
Contact K3 (NAMUR maintenance)
Contact K2 (NAMUR function check)
Contact K1 (Limit)
Inputs OK1, OK2
Return
0.003 mS/cm
25.6 °C
7.00 pH
19.2°C
Block
Administrator level:
Enable / lock functions
Example: Blocking access to the configuration of relay contact K1 (M 700 BASE) from
the Operator level
Call up parameter setting
Select Administrator level.
Enter passcode (1989).
Select “M 700 BASE” with arrow
keys, confirm with enter.
Select “Contact K1” with arrow keys,
“Block” with softkey.
M 700 BASE (Administrator)
Output current I2
Contact K4 (NAMUR Failure)
Contact K3 (NAMUR maintenance)
Contact K2 (NAMUR function check)
Contact K1 (Limit)
Inputs OK1, OK2
Return
M 700 BASE
Output current I2
Contact K4 (NAMUR Failure)
Contact K3 (NAMUR maintenance)
Contact K2 (NAMUR function check)
Contact K1 (Limit)
Inputs OK1, OK2
Return
7.00 pH
19.2°C
Release
7.00 pH
19.2°C
Now, the “Contact K1” line is marked
with the “lock” icon. This function
cannot be accessed from the
Operator level any more. The softkey
function changes to “Release”.
Call up parameter setting
Select Operator level
, passcode
(1246), select “M 700 BASE”.
The locked “Contact K1” function is
displayed in gray and marked with the
lock icon.
43
Function control matrix, time/date
Select menu: Parameter setting/System control
Note: Function check active
MenuDisplay
Parameter setting (Administrator)
System control
Module FRONT 700X-015
Module BASE 700X-025/VPW
Module Cond Ind 7700X
Module PID 700X
Module Out 700X
Return
System control (Administrator)
Parameter sets
Function control matrix
Time/date
Meas. point description
Release of options
Factory setting
Return
0.003 mS/cm
25.6 °C
0.003 mS/cm
25.6 °C
Function control matrix,
time/date
Call up parameter setting
Select Administrator level.
Enter passcode (1989).
Select system control using arrow
keys, confirm with enter.
Submenus of system control:
• Parameter sets
• Function control matrix
•Time/date
• Meas. point description (Point of measurement)
• Release of options
• Factory setting
• Passcode entry
• Software update ... more, depending on Options.
7.00 pH
25.6 °C
Function control matrix (Administrator)
ParSet KI rec.Fav
Input OK2
Left softkey
Right softkey
Profibus DO 2
Return
Softkey usage (Administrator)
Date format
Time
Date
Weekday
Abort
DD,MM,YY
DD/MM/YY
MM/DD/YY
YY-MM-DD
Connect
EC 400
––
0.003 mS/cm
25.6 °C
OK
Function control matrix
Clear assignment of function
(parameter sets, KI recorder, Favorites
menu control) to control element
––
(optocoupler, softkey or Profibus).
Time/date
Selection of date format,
adjustment of time and date
44
M 700(X)
Point of measurement, passcodes, release of options
Select menu: Parameter setting/System control
Note: Function check active
Menu
Display
0.003 mS/cm
25.6 °C
Point of measurement (Administrator)
Enter 0...9A...Z-+/
using [ ] [ ]
Meas. point
Note
Abort
Passcode entry (Administrator)
cal
maint
opl
Passcode entry (Administrator)
cal
maint
opl
adm
Calibration
Calibration
Change passcode
Maintenance
Operator level
Administrator level
Return
M402.......................
MAINT...07/01/02...
OK
0.003 mS/cm
25.6 °C
Off
On
If you lose your
adm passcode, system
access will be locked!
Close
On
On
On
Off
Off
OK
0.003 mS/cm
25.6 °C
Off
1147
Info
Point of measurement, passcode entry, release of options
Point of measurement
You can enter a tag number or notes
(e.g. date of last maintenance).
Passcode entry
When this menu is opened, the analyzer displays a warning (Fig.).
Passcodes (factory settings):
Calibration(cal)1147
Maintenance(maint) 2958
Operator level(opl)1246
Administrator level(adm)1989
Caution
If you lose the Administrator passcode, system access is locked!
Release of options (Administrator)
SW 700-001 KI recorder
SW 700-002 Buffer sets
SW 700-003 Adaptive timer
SW 700-004 ServiceScope
SW 700-005 Cal tolerance band
SW 700-006 Current charact.
When you have purchased an option
to be released via TAN:
• Parameter setting, Administrator
• System control
• Select “Release of options”
Set option to “active”. Enter the
TAN at the prompt. The option is
available after the TAN has been
entered.
45
Inserting the SmartMedia card
To release an additional function via TAN, see Pg 45, Release of options
Inserting the SmartMedia card
Please note when inserting the SmartMedia card:
The SmartMedia card may be inserted or replaced with the power supply
switched on. Protect against electrostatic discharge! When closing the
device, make sure that the sealing is properly seated and clean.
Warning!
Do not touch the terminal compartment, there may be dangerous contact voltages!
To open the analyzer
• Loosen the 4 front screws
• Open the M 700 FRONT at its right side
(pivot hinge inside at the left)
• The slit for inserting the SmartMedia card is
located at the inner side of the M 700 FRONT
To insert the SmartMedia card
• Remove SmartMedia card from its package
without touching the contact surface
• Insert card in the slit at the inner side of the
M 700 FRONT
Inserting the
SmartMedia card:
The label must be
facing you.
To remove the SmartMedia card
• “Close memory card” (Maintenance menu)
To avoid data loss, please call up the
Maintenance menu.
“Close memory card” to terminate software
access to the SmartMedia card.
Now the card can be taken out.
46
M 700(X)
SmartMedia card: Application
Use as memory card in combination with additional functions.
Additional functions must be ordered separately (release via TAN).
Types of SmartMedia cards
The SmartMedia cards are delivered from the manufacturer as
• Softwar
e update (SW 700-106, see Pg 52)
• Memory card (additional function SW 700-1xx, see Pg 56)
Using commercially available SmartMedia cards
Commercially available SmartMedia cards can be used as memor
(Softwar
e updates are supplied by the manufacturer and are device-specific).
y card
The following types of cards are supported: 8 MB, 16 MB, 32 MB, 64 MB
and 128 MB storage capacity. Externally produced files, such as from a
digital camera, are tolerated. Long file names can be read. The M 700
generates file names in the 8.3 format (8 characters file name, 3 characters
program-specific file name extension).
Formatting a commercial SmartMedia card
Some devices (e.g. digital cameras, scanners) cause a formatting of the
SmartMedia card which does not correspond to the SSFDC specification or
the SmartMedia Interface Library (SMIL). The manufacturer therefore recommends to format a commercial SmartMedia card as M 700 memory card
prior to first use.
MenuDisplayFormatting a SmartMedia card
Memory card (Administrator)
Record logbook
Record recorder
Decimal separator
Card full
Format card
On
On
Point
Record stop
0.003 mS/cm
20.4 °C
Off
Off
Comma
Format card
• Insert SmartMedia card (Pg 46)
• Open menu selection
• Parameter setting, Admin. level
• Enter passcode
• System control: Memory card
Return
(The “Memory card” function is
only available with the SmartMedia
Card inserted!)
• Format card
47
System control: Parameter sets A, B
Parameter setting/System control/Parameter sets
Parameter sets A, B
2 complete parameter sets (A, B) can be stored in the analyzer.
An icon in the measurement display shows which parameter set is active:
or
Select “Parameter setting/System control/function control matrix”to select
the control element for switching between the parameter sets (optocoupler,
softkey or PROFIBUS). The currently activated set can be signaled by a relay
contact (see Pg 69).
Menu Display Parameter sets
7.00 pH
25.6 °C
Function control matrix (Administrator)
ParSet KI recFav
Input OK2
Left softkey
Right softkey
Profibus DO 2
Return
System control (Administrator)
Memory card
Copy configuration
Parameter sets
Function control matrix
Time/date
Point of measurement
Return
Parameter sets ( Administrator)
For switchover see
“Function control matrix”
––
Connect
7.00 pH
25,6 °C
7.00 pH
25,6 °C
Select control element for switching between the parameter sets
• Call up Menu selection
• Select Parameter setting,
–
–
• Enter passcode
––
Administrator level
• System control
• Select “Function control matrix”
Parameter sets A, B
• Call up Menu selection
• Select Parameter setting,
Administrator level
• Enter passcode
• System control
• Select “Parameter sets”,
confirm with enter.
• Save parameter set
The active parameter set A over-
Parameter sets
Save parameter set
Load parameter set
Return
48
A, B (internal)
writes the internal parameter set B
• Load parameter set
Parameter set B is loaded
M 700(X)
SmartMedia card: Parameter sets
Parameter setting/System control/Parameter sets
Note: Additional function SW 700-102 required.
Menu
Display
Parameter sets (Administrator)
Select with softkey
Settings in menu
Softkey usage
Parameter sets
Save parameter set A
Load parameter set A
Abort
Parameter sets (Administrator)
Select with softkey
Settings in menu
Softkey usage
Parameter sets
Save parameter set A
Load parameter set A
Return
A,B (internal)
1,2 (card)
1,2,3 (card)
1,2,3,4 (card)
1,2,3,4,5 (card)
0.003 mS/cm
25.1 °C
OK
0.003 mS/cm
25.0 °C
1,2,3,4 (card)
Load parameter set from
SmartMedia card
Loading a parameter set from
SmartMedia card
• Call up parameter setting
• System control
• Call up “Parameter sets” (Fig)
The M 700(X) provides 2 complete
parameter sets (A, B).
5 parameter sets can be stored on
the SmartMedia card. One of those
can be saved as parameter set A to
the analyzer:
0.003 mS/cm
24.9 °C
Load parameter set A (Administrator)
NOTE: Old parameter set
will be overwritten!
Load from
Abort
0.003
25.0
Time 11:49
B (internal)
1(card)
2(card)
3(card)
4(card)
5(card)
mS/cm
°C
Favor ites menu
SmartMedia card
Set 1
Set A
Set 2Set 3
Set B
(internal only)
• Select parameter set to be loaded
Activated parameter set is displayed
in measuring mode.
Note: Remote switching between A
and B is possible via the OK2 input.
Select with softkey
Settings in menu
Softkey usage
Parameter sets
Save parameter set A
Load parameter set A
Abort
Parameter sets (Administrator)
Select with softkey
Settings in menu
Softkey usage
Parameter sets
Save parameter set A
Load parameter set A
Return
0.003 mS/cm
25.1 °C
A,B (internal)
1,2 (card)
1,2,3 (card)
1,2,3,4 (card)
1,2,3,4,5 (card)
OK
0.002 mS/cm
24.2 °C
1,2,3,4 (card)
Saving parameter set on
SmartMedia card
For parameter settings, refer to
www.mtpro.com.
To save a parameter set on
SmartMedia card
• Call up parameter setting
• System control
• Call up “Parameter sets” (Fig)
The M 700(X) provides 2 complete
parameter sets (A, B).
Up to 5 parameter sets can be
loaded to the SmartMedia card.
To do so, a parameter set (1, 2, 3, 4,
or 5) of the SmartMedia card is overwritten by the M 700(X) parameter
set A.
0.002 mS/cm
24.2 °C
Save parameter set A (Administrator)
NOTE: Old parameter set
will be overwritten!
Save to
Abort
B (internal)
1(card)
2(card)
3(card)
4(card)
5(card)
• Select parameter set
on SmartMedia card
Set A
Set 1
Set 2Set 3
Set B
(internal only)
SmartMedia card
Copying parameter sets (for similarly equipped M 700)
Parameter sets can be generated with the M700 and then be copied from
the PC to further SmartMedia cards, thus being available for further M 700
units.
50
M 700(X)
SmartMedia card: Memory card
Parameter setting/System control/Memory card
Menu DisplayUsing the memory card
To use the memory card
• Insert SmartMedia card
• Open menu selection
• Parameter setting, Admin. level
• Enter passcode
• System control: Memory card
Caution! “Close” memory card before
removing it (Maintenance menu)
0.003 mS/cm
25.6 °C
System control (Administrator)
Memory card
Copy configuration
Parameter sets
Function control matrix
Time/date
Point of measurement
Return
0.003 mS/cm
25.6 °C
Memory card (Administrator)
Record logbook
Record recorder
Decimal separator
Card full
Copy configuration
Format card
Return
Copy configuration (Administrator)
Identical module equipment
required for transfer.
On
On
Point
Record stop
Off
Off
Comma
0.003 mS/cm
25.6 °C
With SmartMedia card inserted, the
display shown on the left appears
(The “Memory card” line is displayed
only if there really is a memory card
in the slot).
• Select “Memory card”,
confirm with enter.
The menu is self-explanatory.
Behavior when the memory
card is full:
Continuous recording (as with a
flight recorder) or Stop (card replacement).
Copy configuration
(see Pg 52)
• Save: Saving all data on
the memory card
Configuration
Return
Save
Load
Execute
• Load: Overwriting all device data
with the data from the memory
card
Copy configuration from one M 700 to further M 700 units
Prerequisite:
The devices must have the same hardware.The module equipment must be
absolutely identical – the same modules in the same slots!
Options:
All required options must be enabled on the “master device”, this is not
required for the“slave devices”.
It is the parameters of the options that are transferred, not the option itself.
When an option will be enabled on a “slave device” at a later point in time,
the parameters of this option will be preset corresponding to the “master
device”.
SmartMedia card: Software update
Parameter setting/System control/Release of options
Additional functions must be ordered separately (release via TAN).
Additional function SW 700-106: Software update
For a software update (additional function SW 700-106), the manufacturer
supplies a specially formatted SmartMedia card. The M 700 replaces its own
firmware (operating program) by the new version (“Update”).
Caution!
During a software update the analyzer is not operable!
After a software update you should check the configuration.
This icon indicates that a SmartMedia card is inserted in the slot.
UP
The update card allows loading of new software into the device as well
as storing of current device software on the card.
A memory card can be generated by formatting an update card (irreversible!). Formatting erases the update.
terminates software access to the
SmartMedia card. Must be executed
before removing the card from the
SmartMedia card slot to prevent data
loss.
56
M 700(X)
Factory setting, logbook
Parameter setting/System control/Logbook
Note: Function check active
Menu
Display
Factory setting (Administrator)
The factory setting erases
all your set parameters!
Set factory setting
Return
Logbook (Administrator)
Log failure
Log warning
Erase logbook
Yes
Yes
0.003 mS/cm
24.8 °C
Yes
No
0.003 mS/cm
24.8 °C
No
No
Factory setting, logbook
Factory setting
When this menu is opened, the
analyzer displays a warning (Fig.).
• For factory settings, see module
description
(Free download available at:
www.mtpro.com)
Logbook
Select which messages are to be
logged in the logbook. The last 50
events are recorded with date and
time.
Return
Logbook
12/13/03 09:50 Measurement active
12/13/03 09:36 Parameter setting active
12/12/03 17:52 Measurement active
12/12/03 17:44 Parameter setting active
12/12/03 17:40 Wrong passcode
12/12/03 17:04 Measurement active
12/12/03 16:53 Diagnostics active
Return
0.003 mS/cm
24.8 °C
This permits quality management
documentation to ISO 9000.
The logbook can be called up from
the diagnostics menu (Fig.).
Additional function SW 700-104:
Extended logbook for recording data
on SmartMedia card (TAN).
57
Language, measurement display, viewing angle
Select menu: Parameter setting/M 700 FRONT
Note: Function check active
Menu
Display
0.003 mS/cm
24.0 °C
M 700 FRONT (Administrator)
Languages
Measurement display
Measurement recorder
KI recorder
Abort
Measurement display (Administrator)
Main display
1st primary value
2nd primary value
Viewing angle
Abort
German
English
Français
Italiano
Svenska
Español
OK
0.003 mS/cm
23.9 °C
1 primary value
2 primary values
OK
Language, measurement display, viewing angle
Language setting
• Call up parameter setting
• Select M 700 FRONT
• Call up “Language”
Measurement display
• Call up parameter setting
• Select M 700 FRONT
• Call up “Measurement display”
• Select number and type
of values to be displayed
M 700 FRONT (Administrator)
Main display
1st primary value
2nd primary value
Viewing angle
Return
2primary values
0.003 mS/cm
24.0 °C
S/cm
°C
Viewing angle
• Call up parameter setting
• Select M 700 FRONT
• Call up “Measurement display”
• Adjust display to local light
conditions
• Confirm with enter
58
M 700(X)
Current outputs, contacts, OK inputs
Select menu: Parameter setting/M 700 BASE
Note: Function check active
Menu
Display
M 700 BASE (Administrator)
Output current I2
Contact K4 (NAMUR Failure)
Contact K3 (NAMUR Maintenance)
Contact K2 (NAMUR Function check)
Contact K1 (Limit)
Inputs OK1, OK2
Return
Output current I1 (Administrator)
Measured variable
Characteristic
Output
Start
End
Behavior during messages
Abort
Output current I1 (Administrator)
Measured variable
Characteristic
Output
Start
End
Output filter
Abort
Block
7.00 pH
19.0 °C
OK
7.00 pH
19.0 °C
OK
7.00 pH
19.2°C
Off
pH
°C
rH
S/cm
°C
g/kg
°C
Linear
Trilinear
Function
Table
Parameter setting M 700 BASE
Configure current output
• Call up parameter setting
•Enter passcode
• Call up parameter setting
• Select M 700 BASE
• Enter passcode
•Select “Output current I1”
• Select M 700 BASE
• Select “Output current ...”
• Select measured variable
• Select Characteristic, e.g. “Linear”:
The measured variable is represented
by a linear output current curve.
The desired range of the measured
variable is specified by the values for
“Start” and “End”.
Assignment of measured values: Start (4 mA) and end (20 mA)
Example 1: Range pH 0 - 14 Example 2: Range pH 5 - 7
Output current
20
4
[mA]
0
14
[pH]
Advantage: Higher resolution in range of interest
[mA]
20
4
5
7
[pH]
59
Current outputs: Current output curves
Select menu: Parameter setting/M 700 BASE
• Linear characteristic
The measured variable is represented by a linear output current curve.
Output current
Y
[%]
100
0
Start
End
• Trilinear characteristic
Two additional corner points must be entered:
Output current
Y
[%]
100
2nd corner Y
1st corner Y
0
Start
1st corner X
End
X
Measured variable
X
Measured variable
• Note: Bilinear characteristic
For a bilinear characteristic, identical parameters are entered
for the two corner points (1st corner, 2nd corner).
60
2nd corner X
M 700(X)
• Function characteristic
Nonlinear output current characteristic: allows measurements over several
decades, e.g. for measuring very low values with a high resolution
and high values with a low resolution.
Required: Entering a value for 50 % output current.
Output current
Y
[%]
100
50
0
Equation
Output current (4 to 20 mA) = 16 mA + 4 mA
E + I - 2 * X50%
K =
I:Initial value at 4 mA
X50%: 50% value at 12 mA (output current range 4 to 20 mA)
E:End value at 20 mA
M: Measured value
Entry for “50 % point”:
e.g.:
10 % of measured value
X50% - I
X
Measured variable
End
(1+K)x
1+Kx
M - I
x =
E - I
Logarithmic output curve over one decade:
I:10 % of maximum value
X50%: 31.6 % of maximum value
E:Maximum value
Logarithmic output curve over two decades
I:1 % of maximum value
X50%: 10 % of maximum value
E:Maximum value
61
Output filter
Time constant.
Time constant of output filter
To smoothen the current output, a low-pass filter with adjustable time
constant can be switched on. When there is a jump at the input (100 %),
the output level is at 63 % after the time constant has been reached.
The time constant can be set from 0 to 120 sec. If the time constant is set
to 0 s, the current output follows the input.
Note:
The filter only acts on the current output and the current value of the
secondary display, not on the measurement display, the limit values, or the
controller!
Measured variable0 (4) - 20 mA
Time constant 0 to 120 sec
62
M 700(X)
NAMUR signals: Current outputs
Behavior during messages: Function check, 22 mA signal
Behavior during messages
Behavior during messages (Admin.)
Function check
22mA message
Abort
Output current
[mA]
22
4
Curr.meas. value
Last usable value
Fixed 22 mA
7.00 pH
19.0 °C
OK
“Current meas.”
Function
check
Depending on the parameter setting
(“Messages”), the current outputs switch to:
• Currently measured value
• Last measured value (HOLD function)
• Fixed value (22 mA)
In the case of a fault a 22 mA signal can be
generated for the selected process variable
(1st primary value).
“Last usable value”
Function
check
“Fixed 22 mA”
Function
check
63
NAMUR signals: Relay contacts
Failure, maintenance request, function check
As delivered, the floating relay outputs of the M 700 BASE are assigned to
the NAMUR signals:
• Failure: Contact K4, normally closed contact (signaling current failure)
• Maintenance request: Contact K3, normally open contact
• Function check: Contact K2, normally open contact
NAMUR signals: Factory setting of contacts
• Select parameter setting
• Select Administrator level
• Select M 700 BASE (Fig.)
You can define a delay time for “Maintenance request”
and “Failure”, resp.
If an alarm message is released, the contact will only be
activated after expiry of this delay time.
M 700 BASE (Administrator)
Output current I1
Output current I2
Contact K4
Contact K3
Contact K2
Contact K1
Return
(NAMUR Failure)
(NAMUR maintenance)
(NAMUR function check)
(Limit)
0.003 mS/cm
19.0 °C
Block
Failure is active
when a value has exceeded (or fallen below, resp.) a preset ”Failure Limit Hi”
or ”Failure Limit Lo”, when the measured value is out of range or in the
event of other failure messages. That means that the equipment no longer
operates properly or that process parameters have reached a critical value.
Failure is disabled during “Function check”.
Maintenance request is active
when a value has exceeded (or fallen below, resp.) a preset “Warning Limit
Hi” or “Warning Limit Lo”, or when other warning messages have been activated. That means that the equipment is still operating properly but should
be serviced, or that process parameters have reached a value requiring intervention.
Warning is disabled during “Function check”.
Function check is active:
• during calibration
• during maintenance (current source, meas. point maintenance)
• during parameter setting at the Operator level and the Administrator level
• during an automatic rinsing cycle.
64
M 700(X)
Relay contacts: Protective wiring
Protective wiring of relay contacts
Relay contacts are subjected to electrical erosion. Especially with inductive
and capacitive loads, the service life of the contacts will be reduced.
For suppression of sparks and arcing, components such as RC combinations,
nonlinear resistors, series resistors and diodes should be used.
Warning!
Make sure that the maximum ratings of the relay contacts are
not exceeded even during switching!
Information concerning relay contacts
As delivered, the relay contacts are suitable for low signal currents (down to
approx. 1mA). If currents above approx. 100 mA are switched, the gold
plating is destroyed during the switching process. After that, the contacts
will not reliably switch low currents.
65
Relay contacts
Parameter setting/M 700 BASE/Relay contacts
Menu
DisplaySetting the relay contacts
Contact K1 (Administrator)
Usage
Measured variabl
Limit value
Hysteresis
Effective direction
Contact type
M 700 BASE
Alarm (K4)
Return
7.00 pH
19.2°C
NAMUR maintenance
NAMUR function check
Limit value
Rinse contact
Parameter set 2 active
USP output
The M 700 BASE provides 4 relay contacts
(max. AC/DC rating 30 V / 3 A each).
Contact K4 is provided for failure message.
The switching behavior (normally open or
normally closed), as well as a switch-on or
switch-off delay can be defined.
Configure current output
Relay contacts, usage
• Call up parameter setting
• Call up parameter setting
•Enter passcode
• Enter passcode
• Select M 700 BASE
• Select M 700 BASE
•Select “Output current I1”
• Select “Contact ...”
• “Usage” (Fig.)
K1
Default settings of the user-definable relay
contacts of the M 700 BASE:
K3: NAMUR maintenance request
K2: NAMUR function check
K2
K1: Limit value
The contact assignment K1 - K3 is user defined
K3
(“Usage”):
• NAMUR maintenance request
• NAMUR function check
• Limit value
• Rinse contact
• Parameter set 2 active
• USP output (COND 7700 module only)
Contact assignment: See terminal plate M 700 BASE
66
M 700(X)
Limit value, hysteresis, contact type
Parameter setting/M 700 BASE/Relay contacts/Usage
MenuDisplayUsage as limit value
Contact K1 (Administrator)
Usage
Measured variabl
Limit value
Hysteresis
Effective direction
Contact type
Return
NAMUR maintenance
NAMUR function check
Limit value
Rinse contact
Parameter set 2 active
USP output
Limit value
Effective direction min
Signal
Hysteresis +
Limit value
7.00 pH
19.2°C
Configure current output
Relay output: Limit value
• Call up parameter setting
• Call up parameter setting
•Enter passcode
• Enter passcode
• Select M 700 BASE
• Select M 700 BASE
•Select “Output current I1”
• Select “Contact ...”
• “Usage: Limit value” (Fig.)
Limit value
Effective direction max
Limit value
Hysteresis -
Signal
Contact
1
Contact
0
1
0
Hysteresis
Tolerance band around the limit value, within which the contact is not actuated. Serves to obtain appropriate switching behavior at the output and
suppress slight fluctuations of the measured variable (Fig.)
Contact type
Specifies whether the active contact is closed (N/O) or open (N/C).
67
OK1, OK2 inputs: Specify level
Parameter setting/M 700 BASE/Inputs OK1, OK2
Note: Function check active
MenuDisplay
Inputs OK1, OK2 (Administrator)
For OK2 usage see
“Function control matrix”
OK1 usage
OK2 usage
Return
Inputs OK1, OK2 (Administrator)
For OK2 usage see
“Function control matrix”
OK1 usage
Input OK1
Input OK2
Return
7.00 pH
19.2°C
Off
Function check
7.00 pH
19.2°C
Function check
active 10 ... 30 V
active < 2 V
OK
Setting the OK inputs
Configure current output
OK1 usage
• Call up parameter setting
• Call up parameter setting
•Enter passcode
• Enter passcode
• Select M 700 BASE
• Select M 700 BASE
•Select “Output current I1”
• Select “Inputs OK1/OK2”
• Select “OK1 usage”
Configure current output
OK1/OK2 switching level
• Call up parameter setting
• Call up parameter setting
•Enter passcode
• Enter passcode
• Select M 700 BASE
• Select M 700 BASE
•Select “Output current I1”
• Select “Inputs OK1/OK2”
• Specify active switching level
The M 700 BASE provides 2 digital inputs
(OK1, OK2). The following functions (depending on the parameter setting) can be started
via a control signal:
• OK1: “Off” or “Function check”
• OK2: For selection, see Pg 36, System control
menu/Function control matrix.
(“Off”, “Parameter set A/B”,
“Start KI recorder”)
The switching level for the control signal
must be specified:
(active 10...30 V or active > 2 V).
M 700 BASE
OK1
OK1
OK2
OK2
OK
Contact assignment:
See terminal plate
M 700 BASE
68
M 700(X)
Switching parameter sets via OK2
Parameter setting / System control / Function control matrix
Note: Function check active
Parameter sets
2 complete parameter sets (A, B) can be stored in the M 700.
You can switch between the parameter sets using the OK2 input.
The currently activated set can be signaled by a relay contact . An icon in the
measurement display shows which parameter set is active: or
MenuDisplay
Function control matrix (Administrator)
Input OK2
Left softkey
Right softkey
Profibus DO 2
Return
Contact K3 (Administrator)
Usage
Contact type
ON delay
OFF delay
Abort
0.003 mS/cm
24.8 °C
ParSet KI rec.Fav
––
Connect
0.003 mS/cm
24.8 °C
NAMUR maintenance
NAMUR function check
Limit value
Rinse contact
Parameter set B active
USP output
OK
Parameter sets
Select parameter set (A, B)
via input OK2
• Call up parameter setting
–
–
––
• System control
• Function control matrix
• Select “OK2”
• Connect “Parameter set A/B”
Signal active parameter set
via relay contact
• Call up parameter setting
• M 700 BASE
• Select contact
• Usage: “Parameter set ...”.
Note
The selection has no effect when working on SmartMedia card
with SW 700-102.
69
Calculation Blocks
Select menu: Parameter setting/System control/Calculation Blocks
Calculation of new variables from measured variables
Calculation Blocks
Two measuring modules with all their measured values serve as input for the calculation block. In addition, the general device status (NAMUR signals) is taken into
account. The following variables are calculated from the existing values:
• Ratio
• Pass (passage)
• Reject (rejection)
• Measured-value difference
• Deviation
• pH value calculation by means of dual conductivity measurement
These output variables are then available in the system and can be assigned
to the outputs (current, limit values, display ...)
All current outputs can be set to output the new process variables formed by
the Calculation Blocks.
Measurement display
All new process variables can be displayed as primary or as secondary value.
Controller
Controller functions are not supported.
72
M 700(X)
Calculation formulas
Module combinations, Calculation Block, process variables
Measured
variable
Difference
(selectable in menu)
Ratio
(selectable in menu)
Passage
Rejection
Calculation formulaRangeSpan
DIFF = A - B
or
Meas. variableMeas. variable
DIFF = B - A
A
0.00 ... 19.990.10
RATIO =
B
PASS =• 100
B
0.00 ... 199.910 %
A
REJECT=
(
1-
B
100 %
)
-199.9 ... 199.910 %
A
Deviation
DEVIAT=
B
-1)100 %
(
-199.9 ... 199.910 %
A
pH value calculation by means of dual conductivity measurement
See instruction manual for Cond 7700 module. Principle:
COND measuring
point 1
Cation
exchanger
COND measuring
point 2
H2O
Module 1:
Cond 7700
Module 2:
Cond 7700
COND1- 1/3 COND2
c(NaOH) =
243
Calculation Block
pH = 11+log[c(NaOH)]
pH
c (NaOH) [mmol/l]
Secondary display
only
73
Configuring a Calculation Block
Select menu: Parameter setting/System control/Select Calculation Block
Setting the process variable to be calculated
MenuDisplay
Parameter setting (Administrator)
System control
Module FRONT 700-011
Module BASE 700-021
Module pH 2700
Module pH 2700
+
Calc pH/pH
Return
+ Calc pH/pH (Administrator)
Difference pH
Difference ORP
Difference °C
Messages
Return
+ Messages (Administrator)
Messages pH -diff
Messages ORP- diff
Messages Temp- diff
Off
Off
Off
7.20 pH
25.6 °C
7.20 pH
25.6 °C
–
–
–
7.20 pH
25.6 °C
Configure Calculation Block
Select Calculation Block
• Call up parameter setting
• System control
• Select module
• Depending on modules installed,
the possible combinations
–
–
–
for Calculation Blocks
are offered
Messages
You can activate messages for the
selected variables.
Return
+ Messages pH diff (Administrator)
Monitoring
Failure Limit Lo
Warning Limit Lo
Warning Limit Hi
Failure Limit Hi
Return
74
7.20 pH
25.6 °C
Variable limits
-01.00 pH
-00.50 pH
+00.50 pH
+01.00 pH
Variables which have been set as
“Off” cannot be processed further.
The measured values which shall
release a message are set using the
arrow keys
(left/right: select position,
up/down: edit number).
Confirm with enter.
M 700(X)
USP function
Monitoring of ultrapure water in the pharmaceutical industry
(To configure: Select Parameter setting COND 7700 module)
USP function, define switching output
When a COND 7700 module is installed, one of the floating relay outputs of the BASE module (K1, K2 or K3) can
be assigned to the USP function.
• Select parameter setting:
• Administrator level (function check active!)
• M 700 BASE
• Define contact “Usage” (Fig.)
Contact K1 (Administrator)
Usage
Measured varia
Limit value
Hysteresis
Effective direc
Contact type
Abort
NAMUR maintenance
NAMUR function check
Limit value
Rinse contact
Parameter set 2 active
USP output
0.003 mS/cm
18.8 °C
OK
According to the “USP” directive (U.S. Pharmacopeia), Appendix 5, Section
645 “Water Conductivity” the conductivity of pharmaceutical waters can be
monitored online.
To do so, the conductivity is measured without temperature compensation and
compared with limit values. The water is usable without further test steps
when the conductivity is below the USP limit.
Reduced limit:
The USP limit can be reduced down to 10 % (Parameter setting).
From the measuring mode:
Press menu key to select menu.
Select maintenance using arrow keys,
confirm with enter.
Then select M 700 BASE
Lingua
Current source
For testing purposes, the output
current can be manually specified
(range 0 ... 22 mA).
Close memory card
terminates software access to the
SmartMedia card. Must be executed
before removing the card from the
SmartMedia card slot to prevent
76
data loss.
M 700(X)
Specifications
Specifications
–––––––––––––––––––––
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Display*LC graphic display, white backlighting
Resolution240 x 160 pixels
LanguagesGerman, English, French, Italian, Spanish, Swedish
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
KeypadNAMUR keypad, individual keys, no double assignments
‚
[meas] [menu] [
] [ƒ] [„] [‘] [enter] [softkey 1]
[softkey 2], NAMUR LEDs red and green.
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
LogbookRecording of function activations, appearance and
disappearance of warning and failure messages
with date and time
Storage capacityApprox. 50 entries, without SmartMedia-Card read on display,
recording on SmartMedia card
Extended logbook> 50 000 entries, depending on free memory of
SmartMedia card
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Measurement recorder2-channel measured value recorder with marking of
events (failure, maintenance request, function check, limit
values)
Recording mediumSmartMedia card
Recording capacity> 50 000 entries, depending on free memory of
SmartMedia card
RecordingProcess variables and span selectable
Recording method–– Snapshot
–– Min/Max value
–– Average
Time base–– 10 s to 10 h/ pixel
Zoom function–– 10fold zoom in the event of
high rate of change
–––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
KI recorderAdaptive representation of process flow with
monitoring and signaling of critical process parameters
* Caution! Never expose the display to direct, strong sun light!
Only operate the display within the temperature range of
0 °C up to 50 °C max.
77
Specifications
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Device self-testTest of RAM, FLASH, EEPROM, display, and keypad,
Record for QM documentation to ISO 9000
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
ClockReal-time clock with date
Power reserveApprox. 1 year (lithium battery)
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Data retentionParameters and factory settings> 10 years (EEPROM)
in case of power failureLogbook, statistics, records> 1 year (lithium battery)
Measurement recorder: SmartMedia card
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Module slots3
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Power supply (M 700 C/S)24 (–15 %) to 230 (+15 %) V AC/DC appr. 10 VA/10 W
Overvoltage categoryII
Protection class I
Pollution degree2 (EN 61010-1)
Wire cross-section2.5 mm
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
2
Power supply (M 700 (C/S)/VPW)100 (–15 %) ... 230 (+10 %) V AC < 15 VA, 48 ... 62 Hz
EEx em IIC
or
Power supply (M 700 (C/S)/24V)24 V AC/DC
EEx em IICAC 24 V (- 15 %, + 10 %) < 15 VA, 48 ... 62 Hz
DC 24 V (- 15 %, + 20 %) < 8 VA
Overvoltage categoryII
Protection class I
Pollution degree2 (EN 61010-1)
Wire cross-section2.5 mm
2
Ground wire connection2.5 mm2, M4 screw (EN 61010-1, 6..5.1.2.)
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Sensor monitorDirect display of measured values from sensor for validation
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Protection againstProtective connection according to EN 61010-1, 6.5.1
electrical shock
78
M 700(X)
Specifications
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Input OK 1Galv. separated (OPTO coupler)
EEx ib IICVi ≤ 30 V, floating, galvanic isolation up to 60 V
FunctionSwitches device to HOLD mode
(function check)
Switching voltage0 ... 2 V AC/DC inactive10 ... 30 V AC/DC active
(invertible)
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Input OK 2Galv. separated (OPTO coupler)
EEx ib IICVi ≤ 30 V, floating, galvanic isolation up to 60 V
FunctionSTART / STOP KI recorder
Switch-over to second parameter set
Switching voltage0 ... 2 V AC/DC inactive10 ... 30 V AC/DC active
(invertible)
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Current output I10/4... 20 mA (22 mA), max. 10 V,
EEx ib IICgalvanic isolation up to 60 V
(galvanically connected with output I2)
Load monitoringError message if load is exceeded
Overrange
Measurement error
*
**
22 mA in the case of messages
< 0.25 % current value + 0.05 mA
Current source0,00 ... 22.00 mA
––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Current output I20/4 ... 20 mA (22 mA), max. 10 V,
EEx ib IICgalvanic isolation up to 60 V
(galvanically connected with output I1)
Load monitoringError message if load is exceeded
Overrange
Measurement error
*
**
22 mA in the case of messages
< 0.25 % current value + 0.05 mA
Current source0,00 ... 22.00 mA
––––––––––––––––––––––
l
Switching contacts
*)
–––––––––––––––––––––––––––––––––––––––––––––––––
l
4 relay contacts K1 to K4, floating
EEx ib IICgalvanic isolation up to 60 V
K1, K2, K3 connected on one side
LoadabilityDC: < 30 V / < 500 mA, < 10 W
Usage
*
K1 - K3, user definable for NAMUR maintenance request/
function check, limit values, parameter set 2
active, rinsing contact, USP contact,
K4 permanently set as alarm contact (NAMUR failure)
79
Specifications
General data
––––––––––––––––––––––
l
Explosion protectionII 2 G D EEx em ib IIC T4 KEMA 04 ATEX 2056
––––––––––––––––––––––
l
EMCNAMUR NE 21 and
Emitted interferenceClass B
Immunity to interferenceIndustry
––––––––––––––––––––––
l
Lightning protectionEN 61000-4-5, Installation Class 2
––––––––––––––––––––––
l
Nominal operatingAmbient temperature –20 ... +55 °C (IS: max. +50 °C)
conditionsRelative humidity 10 to 95 % not condensing
––––––––––––––––––––––
l
Transport/–20 ... +70 °C
Storage temperature
––––––––––––––––––––––
l
EnclosureM 700 C:Steel, coated
Assembly–– Wall mounting
DimensionsSee dimension drawing.
Ingress protectionIP 65 / NEMA 4 X
Cable glands5 times M20 x 1.5
Te rminalsSingle wires and flexible leads up to 2.5 mm
WeightApprox. 3.2 kg plus approx. 150 g per module
–––––––––––––––––––––––––––––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
EN 61326 VDE 0843 Part 20 /01.98
EN 61326/A1 VDE 0843 Part 20/A1 /05.99
–––––––––––––––––––––––––––––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
Power supply 24 (–15 %) to 230 (+15%) V AC/DC
Frequency AC 45 ... 65 Hz
–––––––––––––––––––––––––––––––––––––––––––––––––
l
–––––––––––––––––––––––––––––––––––––––––––––––––
l
M 700 S:Stainless steel, polished, 1.4305
–– Post/pipe mounting
–– Panel mounting
–– Sealed against panel
2
(AWG 14)
*
User-defined
**
To IEC 746 Part 1, at nominal operating conditions
80
M 700(X)
Glossary
Glossary
Alarm limit
For each process variable, you can define high and low warning and failure
limits (NAMUR states: maintenance request, failure). The alarm can be activated individually for each variable.
When an alarm limit is exceeded, an error message appears and the corresponding NAMUR contact is activated.
Calibration/adjustment passcode
Protects access to calibration. Can be set or disabled at the Administrator
level.
Cell factor
Mechanical characteristic of electrodeless (toroidal) conductivity sensors.
Cleaning
User-defined time during which the cleaning contact is closed during a rinsing cycle.
Controlled variable
User-defined variable that acts on the controller.
Diagnostics menu
Display of all relevant information on the device status.
Failure
Alarm message and NAMUR contact. Failure means that the equipment no
longer operates properly or that a process parameter has reached a critical
value. Failure is disabled during “function check”.
Feed time alarm
Monitors the time during which the controller output is at 100 %.
81
Glossary
Glossary
Function check
NAMUR contact. Always active when the unit does not output the
configured measured value.
GLP/GMP
Good Laboratory Practice / Good Manufacturing Practice:
Rules for performance and documentation of measurements.
Interval
The interval extends from the start of one rinsing cycle to the start of the
next rinsing cycle, user defined.
Isothermal potential
The isothermal intersection point is the point of intersection between two
calibration lines at two different temperatures. The potential difference
between the electrode zero point and this intersection point is the isothermal
potential “V
Limit contacts
Are controlled by a user-definable process variable. The limit contact is activated if the measured value falls below or exceeds an alarm limit, depending
on the user-defined effective direction.
Logbook
The logbook shows the last 50 events with date and time, e.g. calibrations,
warning and failure messages, power failure etc. etc. This permits quality
management documentation to ISO 9000. Longer recordings are possible
with the additional function “Extended logbook”.
iso
”.
Main display
Large measured-value display in the measuring mode. You can select which
process variable is to be displayed.
82
M 700(X)
Glossary
Glossary
Maintenance menu
The Maintenance menu provides all functions for sensor maintenance and
signal outputs.
Maintenance passcode
Protects access to Maintenance. Can be set or disabled at the Administrator
level.
Measuring mode
When no menu function is activated, the unit is in measuring mode. The
selected measured value is displayed. Pressing the meas key always returns
you to the measuring mode.
Menu structure
The analyzer provides a very clear menu structure. Menu selection is called
up by pressing the menu key. Four basic functions can be accessed:
Calibration (cal), Maintenance (maint), Parameter setting (par), and
Diagnostics (diag). From each of these functions, the individual module
blocks (system control, M 700 FRONT (display functions), M 700 BASE
(signal outputs) can be accessed, as well as all added measuring and
communication modules. See Pg 30.
Message list
The message list shows the number of currently activated messages and the
individual warning or failure messages in plain text.
NAMUR
German committee for measurement and control standards in the chemical
industry
NAMUR contacts
”Funcion check”, ”maintenance request” and ”failure”.
Indicate status of measured variable and measuring system.
83
Glossary
Glossary
Operator level
Menu level of the Parameter Setting menu. You can edit the device settings
that have been enabled at the Administrator level.
Operator passcode
Protects access to the Operator level. Can be set or disabled at the
Administrator level.
Parameter Setting menu
The Parameter Setting menu provides 3 access levels:
Viewing, Operator, and Administrator level.
Passcode protection
Access to the Calibration, Maintenance, Operator and Administrator levels is
protected by passcodes.
The passcodes can be defined or disabled at the Administrator level.
Point of measurement (Tag number)
Can be defined to identify the unit and can be displayed in the Diagnostics
menu.
Reference temperature
With temperature compensation activated, the measured value is calculated
to the value at the reference temperature (usually 20 oder 25 °C) using the
temperature coefficient.
Second rinsing
User-defined time during which the “Rinsing” contact is closed at the end of
the rinsing cycle.
84
M 700(X)
Glossary
Glossary
Secondary displays
Two small displays located below the main display in measuring mode. The
process variables to be displayed can be selected using the softkeys underneath.
Sensor coding
Here, internal settings for electrodeless sensors are encoded
(Cond Ind module).
Slope
The slope of an electrode is the voltage change per pH unit.
For an ideal pH electrode, it lies at -59.2 mV/pH (25 °C).
Viewing level
Menu level of the Parameter Setting menu. Display of all device settings,
however no editing possible.
Zero point
The zero point is the voltage value delivered by an electrode at 25 °C and
pH = 7.00. For an ideal pH electrode, it lies at 0 mV.
In practice, the real zero point is slightly different.