Mettler Toledo M400 Operation Manual

Operation Manual Multiparameter Transmitter M400

Transmitter Multiparameter M400 52 121 378
Transmitter Multiparameter M400 2
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 3
Operation Manual Multiparameter Transmitter M400
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 4

Content

1 Introduction ___________________________________________________________________________________________ 7
Safety instructions ______________________________________________________________________________________ 7
2.1 Definition of equipment and documentation symbols and designations ________________________________________
2.2 Correct disposal of the unit _________________________________________________________________________
Unit overview __________________________________________________________________________________________ 9
3.1 Overview 1/2DIN __________________________________________________________________________________
3.2 Control / Navigation Keys ___________________________________________________________________________
3.2.1 Menu Structure ___________________________________________________________________________
3.2.2 Navigation keys __________________________________________________________________________
3.2.2.1 Navigating the menu tree _________________________________________________________
3.2.2.2 Escape _______________________________________________________________________
3.2.2.3 Enter _________________________________________________________________________
3.2.2.4 Menu _________________________________________________________________________
3.2.2.5 Calibration mode _______________________________________________________________
3.2.2.6 Info mode _____________________________________________________________________
3.2.3 Navigation of data entry fields _______________________________________________________________
3.2.4 Entry of data values, selection of data entry options ______________________________________________
3.2.5 Navigation with
3.2.6 ”Save changes” dialog _____________________________________________________________________
3.2.7 Security Passwords _______________________________________________________________________
3.3 Display ________________________________________________________________________________________
4 Installation instruction __________________________________________________________________________________
4.1 Unpacking and inspection of equipment _______________________________________________________________
4.1.1 Panel cutout dimensional information – 1/2DIN models ___________________________________________
4.1.2 Installation procedure _____________________________________________________________________
4.1.3 1/2DIN version – Assembly _________________________________________________________________
4.1.4 1/2DIN version – Dimension drawings ________________________________________________________
4.1.5 1/2DIN version – Pipe mounting _____________________________________________________________
4.2 Connection of power supply ________________________________________________________________________
4.2.1 Housing (wall mount) _____________________________________________________________________
4.3 Connector PIN definition ___________________________________________________________________________
4.3.1 TB1 and TB2 ____________________________________________________________________________
4.3.2 TB3 – Conventional (analog) Conductivity Sensors ______________________________________________
4.3.3 TB3 – Conventional (analog) pH / ORP Sensors _________________________________________________
4.3.4 TB3 – Conventional (analog) Dissolved Oxygen _________________________________________________
4.3.5 TB4 – ISM (digital) Sensors, pH and Dissolved Oxygen ___________________________________________
4.3.6 TB4 – Optical Oxygen Sensors ______________________________________________________________
4.4 Connection of Sensor – pH / ORP ____________________________________________________________________
4.4.1 Connection of ISM Sensor, pH and Dissolved Oxygen _____________________________________________
4.4.2 AK9 Cable Assignment ____________________________________________________________________
4.4.3 Connecting the Sensor to the VP Cable ________________________________________________________
4.4.4 VP Cable Assignment ______________________________________________________________________
4.4.5 Typical Wiring (Using TB3 / TB4) _____________________________________________________________
4.4.5.1 Example 1 ____________________________________________________________________
4.4.5.2 Example 2 ____________________________________________________________________
4.4.5.3 Example 3 ____________________________________________________________________
4.4.5.4 Example 4 ____________________________________________________________________
4.5 Connection of Sensor – Dissolved Oxygen _____________________________________________________________
4.5.1 Connecting the Sensor to the VP Cable ________________________________________________________
4.5.2 Typical Wiring using TB3 ___________________________________________________________________
4.6 Connection of Optical Dissolved Oxygen ______________________________________________________________
4.6.1 Connecting the Optical Sensor to a Cable ______________________________________________________
4.6.2 Optical DO Sensor Cable Assignment using TB4 _________________________________________________
5 Placing transmitter in, or out, of service ___________________________________________________________________
5.1 Placing transmitter in service _______________________________________________________________________
5.2 Placing transmitter out of service ____________________________________________________________________
6 Quick Setup __________________________________________________________________________________________
in Display ________________________________________________________________ 11
10 10 10 10 11 11 11 11 11 11 11
12 12 12
13
13 13 14 14 15 15 16 16 17 17 17 18 18 19 19 20 20 20 21 22 23 23 24 25 26 27 27 28 29 29 29
30
30 30
31
7 8
9
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 5
7 Sensor Calibration _____________________________________________________________________________________ 32
7.1 Enter Calibration Mode ____________________________________________________________________________
7.2 Conductivity Calibration ___________________________________________________________________________
7.2.1 One-point Sensor Calibration _______________________________________________________________
7.2.2 Two-point Sensor Calibration (4-electrode sensors only) __________________________________________
7.3 Oxygen Calibration _______________________________________________________________________________
7.3.1 One-Point Sensor Calibration ________________________________________________________________
7.3.2 Two Point Sensor Calibration (only available for Optical Sensors) ___________________________________
7.3.3 Process Calibration _______________________________________________________________________
7.4 pH Calibration ___________________________________________________________________________________
7.4.1 One point calibration _____________________________________________________________________
7.4.2 Two point calibration _____________________________________________________________________
7.4.3 Process calibration _______________________________________________________________________
7.4.4 mV calibration __________________________________________________________________________
7.5 Sensor Temperature Calibration _____________________________________________________________________
7.5.1 One-Point Sensor Temperature Calibration ______________________________________________________
7.5.2 Two – Point Sensor Temperature Calibration ____________________________________________________
7.6 Edit Sensor Calibration Constants (only for analog sensor) ________________________________________________
7.7 Sensor Verification
________________________________________________________________________________ 38
8 Configuration _________________________________________________________________________________________
8.2 Measurement ___________________________________________________________________________________
8.2.1 Channel Setup ___________________________________________________________________________
8.2.2 Temperature Source (Not used with ISM Sensors) ________________________________________________
8.2.3 pH / O
Parameter related settings ____________________________________________________________ 40
2
8.2.3.1 Conductivity Temperature Compensation _____________________________________________
8.2.3.2 pH Parameters _________________________________________________________________
8.2.3.3 Dissolved Oxygen Parameters _____________________________________________________
8.2.4 Set Averaging ____________________________________________________________________________
8.3 Analog Outputs __________________________________________________________________________________
8.4 Setpoints ______________________________________________________________________________________
8.5 Alarm / Clean ___________________________________________________________________________________
8.5.1 Alarm __________________________________________________________________________________
8.5.2 Clean __________________________________________________________________________________
8.6 ISM Set up (only available if an ISM sensor is connected) _________________________________________________
8.6.1 Sensor Monitoring ________________________________________________________________________
8.6.2 CIP Cycle Limit ___________________________________________________________________________
8.6.3 SIP Cycle Limit ___________________________________________________________________________
8.6.4 Autoclaving Cycle Limit ____________________________________________________________________
8.6.5 Reset ISM Counter / Timer ___________________________________________________________________
8.7 Display ________________________________________________________________________________________
8.7.1 Measurement ____________________________________________________________________________
8.7.2 Resolution ______________________________________________________________________________
8.7.3 Backlight _______________________________________________________________________________
8.7.4 Name __________________________________________________________________________________
8.7.5 ISM Sensor Monitoring (available when ISM sensor connected) _____________________________________
8.8 Hold Analog Outputs ______________________________________________________________________________
9 System ______________________________________________________________________________________________
9.1 Set Language ___________________________________________________________________________________
9.2 USB ___________________________________________________________________________________________
9.3 Passwords _____________________________________________________________________________________
9.3.1 Changing Passwords ______________________________________________________________________
9.3.2 Configuring Menu Access for Operator _________________________________________________________
9.4 Set / Clear Lockout ________________________________________________________________________________
9.5 Reset __________________________________________________________________________________________
9.5.1 Reset System ____________________________________________________________________________
9.5.2 Reset Meter Calibration ____________________________________________________________________
9.5.3 Reset Analog Calibration ___________________________________________________________________
9.6 Set Date & Time _________________________________________________________________________________
10 PID Setup ____________________________________________________________________________________________
10.1 Enter PID Setup __________________________________________________________________________________
10.2 PID Auto / Manual ________________________________________________________________________________
10.3 Mode __________________________________________________________________________________________
10.3.1 PID Mode _______________________________________________________________________________
32 32 33 33 33 33 34 34 35 35 36 36 36 37 37 37 38
39
39 39 40
40 41 42 42 43 44 45 45 46 46 46 47 48 48 49 49 49 50 50 50 51 51
52
52 52 53 53 53 54 54 54 54 55 55
56
57 57 58 58
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 6
10.4 Tune Parameters _________________________________________________________________________________ 59
10.4.1 PID Assignment & Tuning __________________________________________________________________
10.4.2 Setpoint & Deadband ______________________________________________________________________
10.4.3 Proportional Limits ________________________________________________________________________
10.4.4 Corner Points ____________________________________________________________________________
10.5 PID Display _____________________________________________________________________________________
11 Service ______________________________________________________________________________________________
11.1 Diagnostics _____________________________________________________________________________________
11.1.1 Model / Software Revision ___________________________________________________________________
11.1.2 Digital Input _____________________________________________________________________________
11.1.3 Display ________________________________________________________________________________
11.1.4 Keypad ________________________________________________________________________________
11.1.5 Memory ________________________________________________________________________________
11.1.6 Set Relay _______________________________________________________________________________
11.1.7 Read Relays _____________________________________________________________________________
11.1.8 Set Analog Outputs _______________________________________________________________________
11.1.9 Read Analog Outputs ______________________________________________________________________
11.2 Calibrate _______________________________________________________________________________________
11.2.1 Calibrate Meter (only for channel A) __________________________________________________________
11.2.1.1 Temperature ___________________________________________________________________
11.2.1.2 Current _______________________________________________________________________
11.2.1.3 Voltage _______________________________________________________________________
11.2.1.4 Rg Diagnostic __________________________________________________________________
11.2.1.5 Rr Diagnostics _________________________________________________________________
11.2.2 Calibrate Analog _________________________________________________________________________
11.2.3 Calibrate Unlock __________________________________________________________________________
11.3 Tech Service ____________________________________________________________________________________
12 Info _________________________________________________________________________________________________
12.1 Messages ______________________________________________________________________________________
12.2 Calibration Data _________________________________________________________________________________
12.3 Model / Software Revision __________________________________________________________________________
12.4 ISM Sensor Info (available when ISM sensor connected) __________________________________________________
12.5 ISM Sensor Diagnostics (available when ISM sensor connected) ____________________________________________
13 Maintenance __________________________________________________________________________________________
13.1 Front Panel Cleaning ______________________________________________________________________________
14 Troubleshooting _______________________________________________________________________________________
14.1 Changing the Fuse _______________________________________________________________________________
14.2 Warning- and Alarm list pH ________________________________________________________________________
14.3 Warning- and Alarm list O
14.4 Warning- and Alarm list Optical O
________________________________________________________________________ 72
2
__________________________________________________________________ 73
2
15 Accessories and Spare Parts _____________________________________________________________________________
16 Specifications _________________________________________________________________________________________
16.1 General specifications ____________________________________________________________________________
16.2 Electrical specifications ____________________________________________________________________________
16.3 Mechanical specifications __________________________________________________________________________
16.4 Environmental specifications _______________________________________________________________________
17
Default tables _________________________________________________________________________________________ 77
17.1 General ________________________________________________________________________________________
17.2 M400 _________________________________________________________________________________________
17.3 Error messages __________________________________________________________________________________
18 Warranty _____________________________________________________________________________________________
19 Buffer tables __________________________________________________________________________________________
19.1 Mettler-9 _______________________________________________________________________________________
19.2 Mettler-10 ______________________________________________________________________________________
19.3 NIST Technical Buffers _____________________________________________________________________________
19.4 NIST standard buffers (DIN and JIS 19266: 2000–01) ___________________________________________________
19.5 Hach buffers ___________________________________________________________________________________
19.6 Ciba (94) buffers ________________________________________________________________________________
19.7 Merck Titrisole, Riedel-de-Haën Fixanale ______________________________________________________________
19.8 WTW buffers ____________________________________________________________________________________
59 60 60 60 60
61
61 61 61 62 62 62 62 63 63 63 64 64 64 65 65 66 66 67 67 67
68
68 68 69 69 69
71
71
71
72 72
74
75
75 76 76 76
77 78 80
81
82
82 82 83 83 84 84 85 85
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 7

1 Introduction

Statement of Intended Use – The M400 Multiparameter transmitter is a single- channel online process instrument for measuring various properties of fluids. These include Conductivity, Dissolved Oxygen and pH / ORP. The M400 is not suitable for O2 gas-phase measurement applications. The M400 is available in three different levels. The level indicates the amount of measurement parameters which can be covered. The parameters are indicated on the label on the back of the system.
The M400 is a unique mixed mode transmitter who can handle conventional sensors (analog) or ISM sensors (digital).
M400 parameter fit guide Parameter Type 1 Type 2 Type 3
Analog ISM Analog ISM Analog ISM pH / ORP • Conductivity •* •* •* Oxygen ppm / ppb / traces • / – / – • / – / – • / • / – • / • / – Oxygen optical ppm / ppb • / – • / •
* Q1 / 2009
A large four line backlit Liquid Crystal Display conveys measuring data and setup information. The menu structure allows the operator to modify all operational parameters by using keys on the front panel. A menu-lockout feature, with password protection, is available to prevent the unauthorized use of the meter. The M400 Multiparameter transmitter can be configured to use its four analog and / or six relay outputs for process control.
The M400 Multiparameter transmitter is equipped with a USB communication interface. This interface provides real-time data output and complete instrument configuration capabilities for central monitoring via Personal Computer (PC).

2 Safety instructions

This manual includes safety information with the following designations and formats.
2.1 Definition of equipment and documentation symbols and designations
WARNING: POTENTIAL FOR PERSONAL INJURY.
CAUTION: possible instrument damage or malfunction.
NOTE: Important operating information.
C
On the transmitter or in this manual text indicates: Caution and / or other possible hazard including risk of electric shock (refer to accompanying documents)
The following is a list of general safety instructions and warnings. Failure to adhere to these instructions can result in damage to the equipment and / or personal injury to the operator.
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 8
– The M400 Transmitter should be installed and operated only by personnel familiar with
the transmitter and who are qualified for such work.
– The M400 Transmitter must only be operated under the specified operating conditions
(see section 16). – Repair of the M400 Transmitter must be performed by authorized, trained personnel only. – With the exception of routine maintenance, cleaning procedures or fuse replacement, as
described in this manual, the M400 Transmitter must not be tampered with or altered in
any manner. – Mettler-Toledo accepts no responsibility for damage caused by unauthorized modifications
to the transmitter. – Follow all warnings, cautions, and instructions indicated on and supplied with this product. – Install equipment as specified in this instruction manual. Follow appropriate local and
national codes. – Protective covers must be in place at all times during normal operation. – If this equipment is used in a manner not specified by the manufacturer, the protection
provided by it against hazards may be impaired.
WARNINGS:
Installation of cable connections and servicing of this product require access to shock hazard voltage levels. Main power and relay contacts wired to separate power source must be disconnected before servicing. Switch or circuit breaker shall be in close proximity to the equipment and within easy reach of the OPERATOR; it shall be marked as the disconnecting device for the equipment. Main power must employ a switch or circuit breaker as the disconnecting device for the equipment. Electrical installation must be in accordance with the National Electrical Code and / or any other applicable national or local codes.
NOTE: RELAY CONTROL ACTION: the M400 transmitter relays will always de-energize on loss of
C
NOTE: PROCESS UPSETS: Because process and safety conditions may depend on consistent
C
NOTE: This is a 4-wire-product with an active 4–20 mA analog output.
C
power, equivalent to normal state, regardless of relay state setting for powered operation. Configure any control system using these relays with fail-safe logic accordingly.
operation of this transmitter, provide appropriate means to maintain operation during sensor cleaning, replacement or sensor or instrument calibration.
Please do not supply to Pin1–Pin6 of TB2.

2.2 Correct disposal of the unit

When the transmitter is finally removed from service, observe all local environmental regulations for proper disposal.
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 9
5.90
[150]
5.90
[150]
Enter
Info
CalMenu
M300
1
2
3
ESC
4
5
6
7
8
9

3 Unit overview

The M400 models are available in 1/2DIN case size. The M400 models provide an integral IP65 housing for wall- or pipemount.

3.1 Overview 1/2DIN

1 – Hard Polycarbonate case 2 – Five Tactile-Feedback Navigation Keys 3 – Four-line LCD Display 4 – Power Supply Terminals 5 – USB Interface Port 6 – Relay Output Terminals 7 – Analog Output / Digital Input Terminals 8 – Sensor Input Terminals (analog TB, digital TB) 9 – List of parameters to be measured with this unit
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 10
Measurement
Mode M400
CalMenu Info
Quick Setup Configure System PID Setup Service Messages
ISM Sensor
Info*
Measurement
Analog Outputs
Set Points
Alarm/Clean
ISM Setup*
Set Language
USB
Passwords
Set/Clear Lockout
Reset
PID A/M
Tune Parameters
Mode
Calibrate
Tech Service
Diagnostics
Display
Hold Outputs
Set Date & Time
Channel Select
Output
Set Points
Model/Software
Revision
PID Display Setup
Calibration Data
ISM
Diagnostics*
* Only available in combination with ISM sensors
Menu Cal
Info
Enter
ESC

3.2 Control / Navigation Keys

3.2.1 Menu Structure

Below is the structure of the M400 menu tree:

3.2.2 Navigation keys

3.2.2.1 Navigating the menu tree
Enter the desired main Menu branch with the or keys. Use the and keys to
C
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
NOTE: In order to back up one menu page, without escaping to the measurement mode,
navigate through the selected Menu branch.
move the cursor under the UP Arrow character () at the bottom right of the display screen and press [Enter].
Transmitter Multiparameter M400 11
3.2.2.2 Escape
Press the and key simultaneously (escape) to return to the Measurement mode.
3.2.2.3 Enter
Use the key to confirm action or selections.
3.2.2.4 Menu
Press the key to access the main Menu.
3.2.2.5 Calibration mode
Press the key to enter Calibration Mode.
3.2.2.6 Info mode
Press the key to enter Info Mode
3.2.3 Navigation of data entry fields
Use the key to navigate forward or the key to navigate backwards within the changeable data entry fields of the display.

3.2.4 Entry of data values, selection of data entry options

Use the key to increase or the key to decrease a digit. Use the same keys to navigate within a selection of values or options of a data entry field.
NOTE: Some screens require configuring multiple values via the same data field (ex: configuring
C
multiple setpoints). Be sure to use the or key to return to the primary field and the or
key to toggle between all configuration options before entering to the next display screen.

3.2.5 Navigation with in Display

If a is displayed on the bottom right hand corner of the display, you can use the or the
key to navigate to it. If you click [ENTER] you will navigate backwards through the menu
(go back one screen). This can be a very useful option to move back up the menu tree without having to exit into the measuring mode and re-enter the menu.
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 12

3.2.6 ”Save changes” dialog

Three options are possible for the ”Save changes” dialog: Yes & Exit (Save changes and exit to measuring mode), ”Yes & ” (Save changes and go back one screen) and ”No & Exit” (Don’t save changes and exit to measuring mode). The ”Yes & ” option is very useful if you want to continue configuring without having to re-enter the menu.

3.2.7 Security Passwords

The M400 transmitter allows a security lock-out of various menus. If the security lock-out feature of the transmitter has been enabled, a security password must be entered to allow access to the menu. See section 9.3 for more information.

3.3 Display

NOTE: In the event of an alarm or other error condition the M400 Transmitter will display a
C
flashing in the upper right corner of the display. This symbol will remain until the condition that caused it has been cleared.
NOTE: During calibrations (Channel A), clean, Digital In with Analog Output / Relay / USB in Hold
C
NOTE: Channel A (A is shown on the left side of the display) indicates that a conventional
C
state, a flashing ”H” (Hold) will appear in the upper left corner of the display. During calibration on Channel B, a flashing ”H” (Hold) will appear in the second line. Change to B and flash. This symbol will remain for 20 sec., after end of calibration. This symbol will remain for 20 seconds until after the calibration or clean is completed. This symbol will also disappear when Digital In is deactivated.
Sensor is connected to the transmitter. Channel B (B is shown on the left side of the display) indicates, that an ISM Sensor is connected
to the transmitter. The M400 is a single input channel transmitter, and only one sensor can be connected at the
same time.
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 13
5.39
+0.02
–0.00
[137
+0.5
]
–0.0
5.39
+0.02
–0.00
[137
+0.5
]
–0.0
PANEL CUT-OUT

4 Installation instruction

4.1 Unpacking and inspection of equipment

Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Do not discard the box.
If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Mettler-Toledo immediately

4.1.1 Panel cutout dimensional information – 1/2DIN models

1/2DIN Model transmitters are designed with an integral rear cover for stand-alone wall mount installation. The unit may also be wall mounted using the integral rear cover. See installation instructions in Section 4.1.2.
Below are cut-out dimensions required by the 1/2DIN models when mounted within a flat panel or on a flat enclosure door. This surface must be flat and smooth. Textured or rough surfaces are not recommended and may limit the effectiveness of the gasket seal provided.
Optional hardware accessories are available that allow for panel- or pipe-mount. Refer to Section 15 for ordering information.
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 14
1
2
3

4.1.2 Installation procedure

For Wall Mount: – Remove rear cover from front housing. – Start by unscrewing the four screws located on the face of the transmitter, in each corner.
This allows the front cover to swing away from the rear housing. – Remove the hinge-pin by squeezing the pin from each end.
This allows the front housing to be removed from the rear housing – Drill out wall-mount breakthroughs in the rear housing. – Mount rear housing to wall using appropriate mounting hardware for wall surface. Be sure
it is level and securely fastened and the installation adheres to any and all clearance
dimensions required for transmitter service and maintenance. – Insert two black protective covers (supplied with the M400 transmitter) over the fixing
hardware and into the space on the inside back cover, as shown in the drawing below.
This is necessary to maintain unit integrity. – Replace the front housing to the rear housing. The unit is ready to be wired.
For Pipe Mount: – Use only manufacturer-supplied components for pipe-mounting the M400 transmitter and
install per the supplied instructions. See section 15 for ordering information.

4.1.3 1/2DIN version – Assembly

1 3 Pg 13.5 cable glands 2 2 plastic plugs 3 4 screws
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 15
84 mm
3.31"
42 mm
1.65"
136 mm/5.35"
15 0
mm
/
5. 9
"
6.5 mm
0.256"
116 mm/4.57"
150 mm/5.9"
29 mm
1.14"
29 mm
1.14"
90 mm/3.54"
80 mm/3.15"
75 mm/2.95"
35 mm/
1.38"
6 mm/
0.236"
90 mm/3.54"
138 mm ± 0.5 mm
5.43" ± 0.02"
138 mm ± 0.5 mm
5.43" ± 0.02"
40 ... �60 mm1.57... �2.36"

4.1.4 1/2DIN version – Dimension drawings

4.1.5 1/2DIN version – Pipe mounting

© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 16
1
2

4.2 Connection of power supply

All connections to the transmitter are made on the rear panel of all models.
Be sure power to all wires is turned off before proceeding with the installation. High voltage may be present on the input power wires and relay wires.
A two-terminal connector on the rear panel of all M400 models is provided for power connection. All M400 models are designed to operate from a 20–30 VDC or a 100 to 240 VAC power source. Refer to specifications for power requirements and ratings and size power wiring accordingly (AWG 14, wire cross-section 2.5 mm2).
The terminal block for power connections is labeled ”Power” on the rear panel of the transmitter. One terminal is labeled – N for the Neutral wire and the other + L for the Line (or Load) wire. The terminals are suitable for single wires and flexible leads up to 2.5 mm2 (AWG 14). There is no earth ground terminal on the transmitter. For this reason the internal power wiring within the transmitter is double insulated and the product label designates this using the h symbol.

4.2.1 Housing (wall mount)

1 Connection of power supply 2 Terminal for sensors
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 17
TB2 for1⁄2DIN
1 AO1+ 2 AO1–/AO2– 3 AO2+ 4 AO3+ 5 AO3–/AO4– 6 AO4+ 7 DI1+ 8 DI1–/DI2– 9 DI2+
1 NO1 2 COM1 3 NC1 4 NO2 5 COM2 6 NC2 7 COM5
8 NC5
9 COM6 10 NO6 11 NO3 12 COM3 13 NO4 14 COM4
TB2 TB3
TB4
TB1
1 14
1 9 1 9 1 9
TB1 for1⁄2 DIN
4.3 Connector PIN definition

4.3.1 TB1 and TB2

Power connections are labeled – N for Neutral and + L for Line, for 100 to 240 VAC or 20–30 VDC.
C
NO = normally open (contact open if un-actuated). AO = Analog Output NC = normally closed (contact closed if un-actuated). DI = Digital Input
NOTE: This is a 4-wire-product with an active 4–20 mA analog output.
Please do not supply to Pin1–Pin6 of TB2.

4.3.2 TB3 – Conventional (analog) Conductivity Sensors

Pin no. Sensor wire Color Function
1 white Cnd inner 1 2 white / blue Cnd outer 1 3 blue Cnd inner 2 4 black Cnd outer 2 / Shield 5 not used 6 bare shield RTD ret / GND 7 red RTD sense 8 green RTD 9 +5V
Transparent not used.
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 18

4.3.3 TB3 – Conventional (analog) pH / ORP Sensors

pH / ORP sensors use 52 300 1XX series VP cables, or 10 001 XX02 series AS9 cables (ORP only).
Pin no. Sensor wire Color Function
1 Coax inner / transparent Glass 2 not used 3* Coax shield / red Reference 4* green / yellow, blue Solution GND / Shield 5 not used 6 white RTD ret / GND 7 RTD sense 8 green RTD 9 +5V
grey (no connection)
Take care that AS9 cable and AK9 cable have the same configuration. So, if you want AS9 cable with InPro 2000 and AK9 cable with InPro 3030 connect to TB3, do it as DPAS sensor. Pin no. 1 – Sensing (electrode). Pin no. 3 – Reference (Install jumper 3 to 4).
NOTE: * Install Jumper 3 to 4 when used without Solution Ground.
C
4.3.4 TB3 – Conventional (analog) Dissolved Oxygen
These sensors use 52 300 1XX series VP cables.
Pin no. Sensor wire Color Function
1* not used 2 Coax Shield / red Anode 3* not used 4* green / yellow Shield / GND 5 Coax Inner / transparent Cathode 6 white, grey Temperature, Guard 7 not used 8 green Temperature 9 +5V
Blue wire not used.
NOTE: * Install Jumper (supplied) 1 to 3 to 4 when using InPro 6900 (ppb measurement).
C
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 19
4.3.5 TB4 – ISM (digital) Sensors, pH and Dissolved Oxygen
The wiring of the digital 9 terminal connectors is:
Pin no. Sensor wire Color Function
1 24 VDC 2 GND (24 VDC) 3 Cable core 1-Wire 4 Shield GND (5 VDC) 5 No connection 6 GND (5 VDC) 7 RS485– 8 RS485+ 9 5 VDC
– ISM digital sensors can only be connected on TB4. – Analog sensors can only be connected on TB3.
4.3.6 TB4 – Optical Oxygen Sensors
Pin no. Sensor wire Color Function
1 brown 24 VDC 2 black GND (24 VDC) 3 1-Wire 4 grey and yellow GND (5 VDC) 5 No connection 6 GND (5 VDC) 7 blue RS485– 8 white RS485+ 9 5 VDC
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 20
A
B

4.4 Connection of Sensor – pH / ORP

4.4.1 Connection of ISM Sensor, pH and Dissolved Oxygen

NOTE: Connect the Sensor and screw the plug head clockwise (hand tight).
C

4.4.2 AK9 Cable Assignment

A: 1-wire data (transparent) B: Ground/shield
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 21
A B C D E F S
A
F
C
B
E
D

4.4.3 Connecting the Sensor to the VP Cable

NOTE: Cable lengths > 20 m can worsen the response during pH measurement. Be sure to
C
observe the sensor instruction manual.
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 22
Glass electrode
Ref. electrode
Solution ground
T3
T1
T2
Outer shield
A transparent
B red
D blue
C gray
E white
F green
S green/yellow
C = 220 nF

4.4.4 VP Cable Assignment

© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
T1 / T2 = Temperature probe for 2-wire connection T3 = Additional connection for temperature probe (3-wire connection)
Transmitter Multiparameter M400 23
Cable
Combination pH electrode
Temperature probe
yellowgreen
green
red
transparent
white
Jumper
1
2
3
4
5
6
7
8
9

4.4.5 Typical Wiring (Using TB3 / TB4)

4.4.5.1 Example 1
pH measurement without Solution Ground
NOTE: Jumper Terminals 3 and 4.
C
Wire colors only valid for connection with VP cable; blue and grey not connected. 1 – Glass 2 – Not used 3 – Reference 4 – Shield / GND 5 – Not used 6 – Solution GND / RTD ret 7 – Not used 8 – RTD 9 – Not used
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 24
1
2
3
4
5
6
7
8
9
yellowgreen
green
red
transparent
white
Combination pH electrode with RTD and SG
Temperature probe
blue
Cable
4.4.5.2 Example 2
pH measurement with Solution Ground
NOTE: Wire colors only valid for connection with VP cable, grey not connected.
C
1 – Glass 2 – Not used 3 – Reference 4 – Shield / Solution GND 5 – Not used 6 – GND / RTD ret 7 – Not used 8 – RTD 9 – Not used
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 25
Cable
ORP electrode
Temperature probe
Reference electrode
Sensing electrode
1
2
3
4
5
6
7
8
9
Jumper
4.4.5.3 Example 3
ORP (redox) measurement (temperature optional)
NOTE: Jumper Terminal 3 and 4
C
1 – Platinum 2 – Not used 3 – Reference 4 – Shield / GND 5 – Not used 6 – RTD ret 7 – Not used 8 – RTD 9 – Not used
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 26
1
2
3
5
6
7
8
9
yellowgreen
green
red
transparent
white
Combination pH electrode with RTD and SG
Temperature probe
blue
Cable
Not connected
4
Jumper
4.4.5.4 Example 4
ORP measurement with pH Solution ground electrode (e.g. InPro 3250SG, InPro 4800SG).
NOTE: Jumper Terminal 3 and 4
C
1 – Platinum 2 – Not used 3 – Reference 4 – Shield / GND 5 – Not used 6 – RTD ret 7 – Not used 8 – RTD 9 – Not used
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Transmitter Multiparameter M400 27
A B C D E F S
A
F
C
B
E
D
4.5 Connection of Sensor – Dissolved Oxygen

4.5.1 Connecting the Sensor to the VP Cable

NOTE: Be sure to observe the sensor instruction manual.
C
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400 Printed in Switzerland 52 121 378
Loading...
+ 61 hidden pages