Mettler Toledo L-2XLS+, L-100XLS+, L-200XLS+, L-300XLS+, L-1000XLS+ Operating Instructions Manual

...
Operating Instructions
Pipet-Lite™ XLS+
Manual Pipettes with RFID
2
• Single channel models with LTS or universal-fit shafts
• Multichannel models with LTS
• Adjustable spacer multichannel models with LTS
• RFID enabled, all models
Table of Contents
1.1 Introduction ........................................................................................................3
1.2 Setting Volume ....................................................................................................4
1.3 Tip Selection and Mounting ...................................................................................6
1.4 Tip Immersion Depth ............................................................................................6
1.5 Operation ...........................................................................................................7
1.6 Tip Ejector Arm Removal ......................................................................................8
1.7 Storage ..............................................................................................................8
1.8 Care and Maintenance .........................................................................................8
1.9 Autoclaving ......................................................................................................10
1.10 Upgrading your Pipette to XLS+ ...........................................................................10
1.11 Troubleshooting ................................................................................................11
1.12 Service, Calibration, and Repair ..........................................................................11
1.13 Replacement Parts ............................................................................................12
2.1 RFID (Radio-Frequency Identication) .................................................................14
3.1 Pipet-Lite XLS+ Multichannel Pipettes...................................................................14
4.1 Pipet-Lite XLS Adjustable Spacing Pipettes ............................................................16
5.1 Specications – All Models ............................................................ inside back cover
Contents of box
• Pipette model as ordered
• Conformance Certicate / Warranty Card
If any item is missing please call 800-4-RAININ (800-472-4646) in the US, or contact your
local MT ofce or distributor.
Rainin, Pipet-Lite, XLS, XLS+, LTS, LiteTouch, Hang-Ups are trademarks of Rainin Instrument, LLC. Pipet-Lite pipettes are manufactured under U.S. Patent Nos. 5,614,153; 5,700,959; and 5,849,248. For use under U.S. patents 6,168,761 B1; 6,171,553 B1; and D426,643.
©2017 Mettler-Toledo Rainin, LLC. No part of this manual may be copied without the express written permission of Mettler-Toledo Rainin, LLC.
Pipet-Lite XLS and XLS+
Manual Pipettes with RFID
Pipet-Lite XLS
3
1.1 Introduction
Rainin’s Pipet-Lite XLS and XLS+ family is continually
being upgraded with new features to further improve ergonomics while maintaining high accuracy and perfor­mance. The pipette is available in several versions, out­lined below; this manual will describe all models.
Single-channel 11 models for volumes from 0.1 µL to 20 mL
Multichannel
12 models for volumes from 1 µL to 1200 µL, including 8 and 12 channels
Adjustable-spacer
5 models for volumes from 5 µL to 1200 µL including 6 and 8 channels, with nozzle spacing adjustable from 9-19 or 9-14 mm.
Spring forces are reduced even further over previous
models. With a contoured handle with ngerhook, the
pipette is designed for maximum comfort and enhanced performance.
New enhancements for Pipet-Lite XLS+ single channel
include low stiction* seals for incredibly smooth opera­tion and a tough high-density polymer tip ejector that is corrosion free and easy to release in one smooth action.
Pipet-Lite XLS+ multichannels are enhanced with low
stiction seals and are up to 35% lighter than the previous model with even lower plunger forces.
In models with LTS™ shafts (L-model single channels,
all multichannels and adjustable spacers), the patented LTS LiteTouch™ Tip Ejection System reduces tip ejection forces by up to 85% in the single-channel L models, and provides absolutely consistent sample pickup across all channels in multichannel models, as well as reducing tip ejection force.
Finally, all models contain an RFID tag (radio-frequency
identication) for facilitating calibration management
when using the optional SmartStand and EDPAM soft­ware.
*stiction = static friction
Figure 1: Pipet-Lite XLS+
PLUNGER BUTTON VOLUME SET
VOLUME LOCK
COUPLING NUT
DISPOSABLE TIP
SHAFT
RFID IN FINGER-HOOK
VOLUME DISPLAY
RUGGED HIGH-IMPACT
LOW THERMAL
CONDUCTIVITY POLYMER
TIP EJECTOR
TIP EJECTOR BUTTON
4
Single-Channel
1.2 Setting Volume
1. Turn the volume lock lever to the “unlock” position so the volume setting mechanism is un-
locked and free to turn. See left image in Figure 2.
UNLOCK LOCK
2. With the mechanism unlocked, orient the pipette so you are looking at the volume indicator, then rotate the plunger button to change volume – counter-clockwise to increase, and clock-
wise to decrease volume.
Read the volume indicator from the top down:
2–20 µL: Black digits indicates µL. Red digits – tenths, hundredths of µL.
100–300µl: All digits black – µL. 1000–5000 µL: red digit – mL. Black digits– tenths, hundedths of mL. 10 mL and 20 mL: Red digit – mL. Black digits – tenths of mL.
3. To eliminate errors due to mechanical backlash: when setting the desired volume, rst turn the knob 1⁄3 turn above the desired volume. Then turn the knob slowly clockwise until the desired
volume is displayed. Always dial down to the desired volume.
4. Turn the volume lock to the “lock” position (Fig. 2 above) to prevent changes to the volume setting. Some volumes for the 200 µL model are shown in Figure 4.
Figure 2: Volume Lock Control
Figure 3: Volume Indicator by Pipette Model
Figure 4: Example volumes
5
Volume ranges and increments for each model are shown below:
Single-Channel
Volume Adjustable Range µL Recommended Range µL Increment µL 2 µL 0 to 2 0.1 to 2 0.002
10 µL 0 to 10 0.5 to 10 0.02 20 µL 0 to 20 2 to 20 0.02 100 µL 0 to 100 10 to 100 0.2 200 µL 0 to 200 20 to 200 0.2 300 µL 0 to 300 20 to 300 0.5 1000 µL 0 to 1000 100 to 1000 2.0 2000 µL 0 to 2000 200 to 2000 2.0 5000 µL 0 to 5000 500 to 5000 5.0 10 mL 0 to 10 mL 1 mL to 10 mL 20.0 20 mL 0 to 20 mL 2 mL to 20 mL 20.0
Multichannel and Adjustable-Spacers
Volume Adjustable Range µL Recommended Range µL Increment µL
10 µL 0 to 10 0.5 to 10 0.02 20 µL 0 to 20 2 to 20 0.02 50 µL 0 to 50 5 to 50 0.05 100 µL 0 to 100 10 to 100 0.2 200 µL 0 to 200 20 to 200 0.2 300 µL 0 to 300 20 to 300 0.5 1200 µL 0 to 1200 100 to 1200 2.0
Filter
The 5000 µL, 10 mL, and 20 mL pipettes use a lter in the end of the shaft to help prevent
liquid entering the shaft and contaminating the piston, should the plunger snap up during
aspiration. Using a lter is recommended when pipetting large volumes. Replace the lter if it
gets wet.
The 5000 µL and 10 mL pipettes use the same lter, as shown in the diagram above: 5000 µL: small diameter into the shaft. 10 mL: large diameter into the shaft. Filter part numbers are 17001944 (pack of 100) and 17001945 (pack of 1000).
The lter for the 20 mL model is a cylinder. Part numbers: 17001951 (pack of 100), 17001952 (pack of 500).
Figure 5: Filter orientation
6
Single-Channel
1.3 Tip Selection and Mounting
Rainin pipettes and tips are designed together as a pipetting system, both in LTS and traditio-
nal versions. All Rainin tips are BioClean and totally inert, so you can be assured of the best pipetting results.
Rainin pipettes are calibrated with Rainin tips, and performance to published specications
can only be guaranteed when Rainin tips are used.
To mount a tip, press the pipette shaft into the end of the tip with light force. With both LTS and
traditional versions, the tip will seal properly on the shaft with minimal force — do not use more force than is required.
1.4 Tip Immersion Depth
The recommended depth for tip insertion into the sample for each model is shown below.
Nominal Volume Volume Range Immersion Depth
2 µL 0.1 - 2 µL 1-2 mm 10 µL 0.5 - 10 µL 1-2 mm 20 µL 2 - 20 µL 2 - 3 mm 100 µL 10 - 100 µL 2 - 3 mm 200 µL 20 - 200 µL 3 - 6 mm 300 µL 20 - 300 µL 3 - 6 mm 1000 µL 100 - 1000 µL 3 - 6 mm 2000 µL 200 - 2000 µL 3 - 6 mm 5000 µL 500 - 5000 µL 6 - 10 mm 10 mL 1 mL - 10 mL 6 - 10 mm 20 mL 2 mL - 20 mL 6 - 10 mm
Tip immersion depth is critical and should not be exceeded, or the volume measured may be
inaccurate, possibly out of specication. The tip angle is also important – the pipette should
always be used in a position within 20 degrees of vertical. See Figure 6 below.
TIP
ANGLE
Figure 6: Tip Immersion Depth and Tip Angle