Metron QA-ST Service manual

QA-ST
User and Service Manual
QA-ST Safety Tester
North American Version 1.1
Copyright 2000 by METRON. All rights reserved.
METRON:
USA __ FRANCE ________________ NORWAY________________
Disclaimer
METRON provides this publication as is without warranty of any kind, either express or implied, in­cluding but not limited to the implied warranties of merchantability or fitness for any particular purpose. Further, METRON reserves the right to revise this publication and to make changes from time to time to the content hereof, without obligation to METRON or its local representatives to notify any person of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or implied war­ranties in certain transactions; therefore, this statement may not apply to you.
Limited Warranty
METRON warrants that the QA-ST Safety Tester will substantially conform to published specifica­tions and to the documentation, provided that it is used for the purpose for which it was designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair any defective simulator, if the fault is due to a manufacturing defect. In no event will METRON or its local representa­tives be liable for direct, indirect, special, incidental, or consequential damages arising out of the use of or inability to use the QA-ST Safety Tester, even if advised of the possibility of such damages. METRON or its local representatives are not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or inability to use the QA-ST Safety Tester. Neither METRON nor its local representatives will accept, nor be bound by any other form of guarantee concerning the QA­ST Safety Tester other than this guarantee. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.
ii
Table of Contents
1. INTRODUCTION.................................................................................................................1-1
1.1 Features....................................................................................................................................1-1
1.2 Specifications.............................................................................................................................1-1
1.3 General Information...................................................................................................................1-2
2. INSTALLATION..................................................................................................................2-1
2.1 Receipt, Inspection and Return.................................................................................................2-1
3. OPERATING QA-ST...........................................................................................................3-1
4. QA-ST SAFETY TESTS ...................................................................................................4-1
4.1 Linearity Test of Voltmeter (Lead to Lead)...............................................................................4-1
4.2 Mains Supply Voltage................................................................................................................4-2
4.3 Ground Wire Resistance DC.....................................................................................................4-3
4.4 Leakage Current – Internal Case........................................................................................4-4
4.5 Leakage Current – External Case.............................................................................................4-5
4.6 Patient Leakage Current – Lead to Lead..................................................................................4-6
4.7 Patient Leakage Current – Lead to Ground..............................................................................4-7
4.8 Leakage Current – Lead to ISO................................................................................................4-8
5. CALIBRATION....................................................................................................................5-1
5.1 Linearity Test of Voltmeter (Lead to Lead)...............................................................................5-1
5.2 Mains Supply Voltage................................................................................................................5-2
5.3 Ground Wire Resistance DC.....................................................................................................5-3
5.4 Leakage Current – Internal Case........................................................................................5-4
5.5 Leakage Current – External Case.............................................................................................5-5
5.6 Patient Leakage Current – Lead to Lead..................................................................................5-6
5.7 Patient Leakage Current – Lead to Ground..............................................................................5-7
5.8 Leakage Current – Lead to ISO................................................................................................5-8
APPENDIX A – CIRCUIT BOARD SCHEMATIC (NORTH AMERICAN VERSION)..............A-1
APPENDIX B – ERROR REPORT FORM...............................................................................B-1
APPENDIX C - IMPROVEMENT SUGGESTION FORM.........................................................C-1
iii
This page intentionally left blank.
iv
Manual Revision Record
This record page is for recording revisions to your QA-ST User and Service Manual that have been published by METRON AS or it’s authorized representatives. We recommend that only the man­agement or facility representative authorized to process changes and revisions to publications:
make the pen changes or insert the revised pages;
ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as
superseded and placed in a superseded document file, and;
enter the information below reflecting that the revisions have been entered.
Rev No Date Entered Reason Signature of Person Entering Change
v
This page intentionally left blank.
vi

1. Introduction

This chapter describes the METRON QA-ST Safety Tester’s features and specifications.

1 . 1 F e a t u r e s

The QA-ST is designed to quickly check electrical safety on medical and industrial devices. It is palm-sized and very portable. The tests can be performed manually or in step-by-step mode. The results are displayed in a bright 4-digit LCD display.
The QA-ST performs all electrical safety tests necessary to satisfy AAMI ESI, NFPA-99, VDE 701, VDE 751 and HEI 95. It also features differential current measurements obtained by an internal current clamp. This presents a new leakage current measurement technique for convenient testing of medical equipment with a fixed power cord (hardwired to wall, not detachable).
1 . 2 S p e c i f i c a t i o n s
1

1. Line Voltage Measurement Range: 0 - 300 V

Accuracy: ± 1% reading, ± 1 digit
1
For North American QA-ST Version 1.1
Resolution: 1 V L-N
2. Current Capacity
15 Amps @ 100 Volts 15 Amps @ 120 Volts 10 Amps @ 230 Volts
3. Resistance Measurement
DC-Source: Range: 0 - 19.99 Ohms Accuracy: ± 2%reading, ± 1 digit Resolution: 0.01 ohms Current Source: 100 mA DC
AC-Source: Range: 0 - 1999 mOhms Accuracy: ± 2% reading, ± 1 digit Resolution: 1 mOhm Current Source: 200 mA AC
4. Leakage Current
Range: 0 - 1999 µA RMS Accuracy: DC and 25 to 1.0 kHz, ± 2% reading, ± 1 digit
1.0 kHz to 100 kHz, ± 2.5%R, ± 1 digit 100 kHz to 1 MHz, ± 5.0%R, ± 1 digit
Resolution: 1 µA
5. Input Impedance
1000 ohm, AAMI ES1-1993 load or IEC 60601.1 load
6. Power Consumption
110 V (50/60) Hz = 200 mA 115 V (50/60) Hz = 200 mA 240 V (50/60) Hz = 100 mA
2
1 . 3 G e n e r a l
I n f o r m a t i o n
1. Temperature Requirements:
+15/59 to +35/95 °C/F while operating 0 to +50/122 °C/F for storage
2. Display:
Type: LCD Numeric format: 4 characters Display control: Keypad
3. Power: 100V/110V or 230V/240V, 50/60 Hz
4. Housing: Plastic case
5. Physical Dimensions (D x W x H):
191mm x 135mm x 45mm
7.5 in x 5.3 in x 1.8 in
6. Weight: 930 g / 2.0 lbs.
3
1 . 4 Q A
-
S T O r­d e r­i n g I n f o r­m a ti o n
Order no: 11040: QA-ST Electrical Safety Analyzer
(Specify power supply socket.)
Standard Accessories: 11045: User/Service Manual QA-ST
17290: Test Lead (Kelvin Cable)
Various power supply socket including:
European Schuco, French Schuco, UK, Swiss, Australian, US.
Optional Accessories: 11041: Carrying Case
4

2. Installation

This chapter explains unpacking, receipt inspection, claims, and the general procedures for QA-ST setup.
2 . 1 R e c e i p t ,
I n s p e c t i o n
a n d
R e t u r n
1. Inspect the outer box for physical damage.
2. Carefully unpack all items from the box and check to see that you
have the following items:
QA-ST Electrical Safety Analyzer (P.N. 11040)
User/Service Manual QA-ST (P.N. 11045)
Test Lead (Kelvin Cable) (P.N. 17290)
Specified power supply socket
3. If you note physical damage, or if the unit fails to function accord-
ing to specification, inform the supplier immediately. When METRON or the company’s Sales Agent, is informed, measures
1
will be taken to either repair the unit or dispatch a replacement. The customer will not have to wait for a claim to be investigated by the Sales Agent. The Customer should issue a new purchase order to ensure delivery.
4. When returning an instrument to METRON, or the Sales Agent,
fill out the address label, describe what is wrong with the instru­ment, and provide the model and serial numbers. If possible, use the original packaging material for return shipping. Otherwise, repack the unit using:
a reinforced cardboard box, strong enough to carry the
weight of the unit.
at least 5 cm of shock-absorbing material around the unit.
nonabrasive dust-free material for the other parts.
Repack the unit in a manner to ensure that it cannot shift in the
box during shipment.
METRON’s product warranty is on page ii of this manual. The war­ranty does not cover freight charges. C.O.D. will not be accepted without authorization from METRON or its Sales Agent.
This page intentionally left blank.
3

3. Operating the QA-ST

This chapter describes the controls and terminals of the QA-ST Safety Tester.
1
Remove the batteries and dis­connect the AC Adapter / Bat­tery Eliminator if you do not in­tend to use the PS-410 for an extended period of time.