USA __ FRANCE ________________ NORWAY________________
1345 Monroe NW, Suite 255A 30, rue Paul ClaudelTravbaneveien 1
Grand Rapids, MI 49505 91000 Evry, FranceN-7044 Trondheim, Norway
Phone:(+1) 888 863-8766 Phone: (+33) 1 6078 8899Phone:(+47) 7382 8500
Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839Fax:(+47) 7391 7009
E-mail: support.us@metron-biomed.com E-mail: info@metron.fr E-mail: support@metron.no
Disclaimer
METRON provides this publication as is without warranty of any kind, either express or implied,
including but not limited to the implied warranties of merchantability or fitness for any particular pur pose. Further, METRON reserves the right to revise this publication and to make changes from time
to time to the content hereof, without obligation to METRON or its local representatives to notify any
person of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or im plied warranties in certain transactions; therefore, this statement may not apply to you.
Limited Warranty
METRON warrants that the QA-ES Electrosurgical Analyzer will substantially conform to published specifications and to the documentation, provided that it is used for the purpose for which it was
designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair
any defective system, if the fault is due to a manufacturing defect. In no event will METRON or its local representatives be liable for direct, indirect, special, incidental, or consequential damages arising
out of the use of or inability to use the QA-ES Electrosurgical Analyzer, even if advised of the possibility of such damages. METRON or its local representatives are not responsible for any costs, loss of
profits, loss of data, or claims by third parties due to use of, or inability to use the QA-ES Electrosurgical Analyzer. Neither METRON nor its local representatives will accept, nor be bound by any other
form of guarantee concerning the QA-ES Electrosurgical Analyzer other than this guarantee. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; there fore, this statement may not apply to you.
4. ESU Tests with QA-ES......................................................................................................................1
5. Control and Calibration....................................................................................................................1
6. Component Functions and Parts......................................................................................................1
Appendix A – Diagrams ........................................................................................................................A-1
Appendix B – Error Report Form........................................................................................................B-1
Appendix C – Improvement Suggestion Form....................................................................................C-1
iii
This page intentionally left blank.
iv
Manual Revision Record
This record page is for recording revisions to your QA-ES User & Service Manual that have been
published by METRON AS or its authorized representatives. We recommend that only the man agement or facility representative authorized to process changes and revisions to publications:
•make the pen changes or insert the revised pages;
•ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as
superseded and placed in a superseded document file, and;
•enter the information below reflecting that the revisions have been entered.
Rev NoDate EnteredReasonSignature of Person Entering Change
1.30-14-30-01General Update
v
This page intentionally left blank.
vi
1.1QA-ES Description
1. Introduction
This chapter describes the METRON QA-ES Electrosurgical Analyzer, including its features and specifications.
The METRON QA-ES Electrosurgical Analyzer (QA-ES) is a precision instrument designed to perform tests on high-frequency electro surgical units (ESU) in accordance with national and international
standards, and is designed to be used by trained service technicians.
Tests include:
•automatic power distribution measurement;
•crest factor measurement;
•RF leak measurement, and;
•return electrode monitor (REM) test
Testing is accomplished by measuring the ESU output against test
loads that are set and adjusted in the QA-ES. The QA-ES can automatically execute a power distribution test with a load resistance
ranging from 10 ohms to a maximum of 5200 ohms. The automatic
measuring of the QA-ES, comprising crest factor measurements
with a bandwidth of 10 MHz, ensures that the test result is reliable
and reproducible.
Test results, shown in the QA-ES’s LCD display, can be printed out
directly, or transferred to a PC via the PRO-Soft QA-ES test auto mation software. PRO-Soft lets you design test protocols, remotely
control the QA-ES, and store the test results.
1.2 QA-ES Specifications
Generator Output:RF LEAKAGE: From active electrode or neu-
tral plate with an open or closed load circuit.
Mode Of Operation:Manual or user-programmable. Can be re-
motely controlled with PC utilizing accessory
PRO-Soft QA-ES software and RS-232 communication cable connection.
Measurements:True RMS value of applied waveform.
RMS Bandwidth:30 Hz to 10 MHz (+3 dB).
Low Frequency Filter:100 Hz filter to avoid low frequency distur-
bance and/or interference.
Current:20 mA to 2200 mA.
Current Accuracy:20 - 2200 mA ± 2% of reading.
Load Resistance:10 - 2500 ohms in steps of 25 ohms (@ dc).
1
2600 - 5200 ohms in steps of 100 ohms (@
dc).
Additional Fixed Load:200 ohms, 500 watt maximum.
Crest Factor:The higher of the two peak-measurements is
Range:1.4 - 16 (V peak voltage / V RMS).
Foot Switch Output:The output triggers the measurement after a
Peak To Peak Voltage:0 to 10 kV (closed load only) ± 10%. Mea-
Oscilloscope Output:5 V/A uncalibrated, 100 mA RF current mini-
Isolation:10 kV isolation between measurement device
used for calculation.
programmed delay time, defined as the time
period from the activation of the foot switch to
the beginning of data processing. The delay
time is 200 ms - 4000 ms.
surement is taken between the active and
dispersive electrodes with closed load only.
mum input.
and enclosure.
1.3 General Information
Temperature Requirements:
+15°C to +35°C when operating
0°C to +50°C in storage
Display:
TypeLCD graphic display
Alphanumeric format8 lines, 40 characters
Graphics mode:240 x 64 point matrix
Display control:5 F-keys, enter, cancel and an encoder
Data Input/ Output (2):Parallel printer port (1); Bi-directional RS
-232C (1) for Computer control
Power Source:From 115 VAC to 230 VAC, 48/66 MHz.
Mechanical Specifications:
Housing Metal case
Height13.2 cm / 4.48 in.
Width34.2 cm / 11.61 in.
Length39.5 cm / 13.41 in.
Weight9.8 kg / 21.6 lbs.
Standard Accessories:
QA-ES Electrosurgical Analyzer(P.N. 14010)
Power Cord(P.N. 14300)
QA-ES User and Service Manual(P.N. 14025)
Additional Accessories:
E-Input Measuring Cable - Black(P.N. 11451)
2
E-Input Measuring Cable - Red(P.N. 11452)
Alligator Clamp - Black(P.N. 11461)
Alligator Clamp - Red(P.N. 11462)
Carrying case (P.N. 14100)
PRO-Soft QA-ES software (P.N. 12200)
PRO-Soft QA-ES DEMO (P.N. 14201)
PRO-Soft QA-ES User Manual(P.N. 14225)
Storage:
Store in the carrying case in dry surroundings within the temperature range specified. There are no other storage requirements.
Periodic Inspection:
The unit should be calibrated every 12 months.
3
This page intentionally left blank
4
2.1Receipt, Inspection and Return
2. Installation
This chapter explains unpacking, receipt inspection and claims, and
the general procedures for QA-ES setup.
1.Inspect the outer box for damage.
2.Carefully unpack all items from the box and check to see that
you have the following items:
QA-ES Electrosurgical Analyzer (PN 14010)
E-Input Measuring Cable -- Black (PN 11411-B)
E-Input Measuring Cable -- Red (PN 11411-R)
Alligator Clamp -- Black (PN 11412-B)
Alligator Clamp -- Red (PN 11412-R)
Power Cord (No PN)
QA-ES User and Service Manual (PN 14025)
3.If you note physical damage, or if the unit fails to function according to specification, inform the supplier immediately.
When METRON AS or the company’s representative, is in formed, measures will be taken to either repair the unit or dispatch a replacement. The customer will not have to wait for a
claim to be investigated by the supplier. The customer should
place a new purchase order to ensure delivery.
2.2 Setup
4.When returning an instrument to METRON AS, or the company
representative, fill out the address label, describe what is wrong
with the instrument, and provide the model and serial numbers.
If possible, use the original packaging material for return shipping. Otherwise, repack the unit using:
a reinforced cardboard box, strong enough to carry the
weight of the unit.
at least 5 cm of shock-absorbing material around the unit.
nonabrasive dust-free material for the other parts.
Repack the unit in a manner to ensure that it cannot shift in the
box during shipment.
METRON’s product warranty is on page ii of this manual. The
warranty does not cover freight charges. C.O.D. will not be accepted without authorization from METRON A.S or its representative.
1.Equipment connection is as shown in the typical setup below
(for Power Distribution Test).
1
CAUTION
The QA-ES uses a 3-wire
power cord and plug for
the user's safety. Use this
power cord in conjunction
with a properly grounded
electrical outlet to avoid
electrical shock.
2.If you are using an oscilloscope, attach the BNC cable to the
Scope Output connector, located on the front of the QA-ES.
3.If PRO-Soft QA-ES is being used, attach an RS-232 (null modem/data transfer configured) cable to the 9-pin D-sub outlet
port located at the rear of the QA-ES. Do not attach the printer
cable to the QA-ES. See below. However, if you are not using
PRO-Soft QA-ES, and are sending directly to a printer for printouts, attach the printer cable to the 25-pin outlet port.
2.3 PRO-Soft QA-ES
2
PRO-Soft QA-ES is a front-end test automation and presentation
tool for METRON's ESU Performance Analyzer. It allows you to
conduct the same tests, but by remote control via an IBM-compatible PC/XT with MS Windows (Version 3.1 or later). Additionally,
the program has features to enhance your QA-ES’s performance.
Each of the QA-ES tests can be run independently from PRO-Soft in
the “Manual” test mode. Results are shown on the PC screen during
testing, and the user is prompted to set the tested equipment accord ingly. At the conclusion of tests, the user may print a report, store
the test and results on disk, or both. Combinations of tests can be
created and stored as “Test Sequences.” The program maintains a library of these sequences. In this way you can store and retrieve se-
NOTE
PRO-Soft QA-ES has its
own user manual, which
contains all the information
concerning the program. If
you order a demonstration
version of the program you
also receive the manual.
quences that are appropriate for each ESU being tested at your facil ity.
Sequences can then be used independently, or can be attached to a
checklist, written procedure, and equipment data in the form of a
test “Protocol.” The equipment data can be entered manually into
the protocol, or it may be retrieved by PRO-Soft from a database
program, or other equipment files. Protocols can be created easily
for each ESU in your inventory, and stored for use. Test protocols
with results can be printed, or stored on disk, and the results of testing can be sent back to the equipment database to close a work order
and update the service history.
3
This page intentionally left blank
4
3.1 Control Switches
and Connections
Front Panel
3. Operating QA-ES
This chapter explains the operating controls, switches and menus
of the QA-ES, and details how to use them in ESU testing.
1. Power SwitchTurns the power on and off.
2. EncoderSets values according to the specified range and
3. EnterAdmits newly specified information.
4. CancelCancels a new value and returns to previously cho-
5. LCD DisplayShows messages, test results and function menus.
6. Function KeysFl - F5 are used to select the functions shown on the
7. RF-DetectIndicates when the ESU is activated.
8. RemoteIndicates that REMOTE CONTR. (F4) has been
9. Scope Output
Connector
choose between different operations/ measurement
ranges.
sen value.
bottom line of the LCD display, i.e., for selecting the
function that is directly above the key.
pressed.
BNC-cable connector for attenuator signal in real
time. (When oscilloscope output is desired.)
1
Right Side Panel
Rear Panel
10. Terminal
RED and BLACK
11. Terminal
BLUE and BLUE
12. Terminal
GREEN and GREEN
Connection for the electrode outputs of the VAR.
LOAD ESU. Active electrode to the red terminal,
and neutral electrode to the black terminal.
Additional fixed load resistance of 200 ohms
FIXED LOAD 500 watt for serial connection during leakage test.
Foot switch output switch can be used to trigger
the ESU.
13. RS-232 Serial Port9-pin D-sub
14. Printer Outlet Port14-25 pin D-sub
15. Main Connector3-pin SCHUCO-plug
16. Voltage Selector115 VAC/230 VAC
17. FusesT 200 mA @230 VAC / T 400 mA @ 11 5 VAC
2
3.2 QA-ES Menu and
Function Keys
3.3 LCD Display Menu/
Messages (Overview)
The QA-ES uses a display, programmable function keys and a setting regulator to provide flexibility and control over the operations.
The upper part of the screen displays messages, status and results.
The menu bar is at the bottom of the display. The function keys are
numbered from Fl to F5. A function is selected by pressing the key
located directly under the Menu Item displayed in the menu bar.
3
3.4LCD Display
Menu/ Messages (Detail)
1. Startup Screen. The following screen will be displayed for 2
seconds after the QA-ES has been switched on.
2.Main Menu
a. First Menu Bar (Page 1)
b. Second Menu Bar (Page 2)
3.SHOW CHOICES (F1). This function is activated when you
see an asterisk (*) in the status field under ‘Mode.’ Choose a
test function by pressing UP (F2) or DOWN (F3). (The encoder can also be used for choosing a test function) Press
ENTER (F5) to save it under Mode in the STATUS field. Press
CANCEL (F4) to undo.
4.KNOB PARAM. (F2). With this function, you can choose between ‘Mode,’ ‘Load’ and ‘Delay’ in the STATUS field. (*
marks the active item). If you choose ‘Load’, use the encoder to
set the load from 10 ohms to 5200 ohms in steps of:
25 ohms from 50 ohms to 2500 ohms.
100 ohms from 2500 ohms to 5200 ohms.
Save the selected load in ‘Mode’ under the STATUS field by
pressing ENTER (F5). Press CANCEL (F4) to undo. If you
choose ‘Delay’, use the setting regulator to set the delay from
200 ms to 4000 ms in steps of:
50 ms from 200 ms to 1000 ms.
4
100 ms from 1000 ms to 4000 ms.
Save the chosen delay in ‘Delay’ under the STATUS field by
pressing ENTER (F5). Press CANCEL (F4) to undo.
5.START (F3). When you press on START, the test procedure
will begin, and the text in the field ‘Oper.’ will change from
‘Ready’ to ‘Measuring’. If the unit is set to the position for a
REM test, this text will change from ‘Ready’ to ‘lncr'. res.’
Press STOP (F3) to stop the test procedure.
6.SETUP (F4). Here you can set the power distribution level for
start, stop and step in ohms.
Choose the ‘Start load’ by using KNOB PARAM (F4). (see
stars). Use the encoder to set the level. Save the level by pressing ENTER (F5). Press CANCEL (F4) to undo. Go to ‘End
load’ and ‘Step Size’ and repeat the same procedure.
Pwr. distr. Start load is the first load to be used during the
measurements; it can be set from 10 ohms to 2100 ohms, with
steps of 25 ohms starting at 25 ohms onwards.
Pwr. distr. End load is the last load used in the measurements;
R can be set from 525 ohms to 5200 ohms, with steps of 25
ohms from 525 to 2500 ohms and step of 100 ohms from 2500
ohms to 5200 ohms
Pwr, distr. Step Size is the load set with steps of 25, 50, 100,
and 200 ohms
Press QUIT MENU (F5) to return to the main menu.
7.PRINT HEADER (F3). Writes a heading for a new test protocol.
8.REMOTE CONTR. (F4). Enables you to control the QA-ES
through a PC. Required software: PRO-Soft QA-ES.
5
3.5 Printout
Press PRINT HEADER (F3) before printing out a page if you want
it to have a new heading. The QA-ES automatically prints out the
test results via the printer output after every measurement. See ex-
ample below.
3.6 Foot Switch Output
3.7 Main Switch On/Off
6
A Foot Switch Output is activated by use of relays (K11), and locat ed on the right side of the unit. This is used to trigger the foot switch
input on the ESU being tested.
The QA-ES has to be turned off for at least 5 seconds before turning
it on again to allow the reset circuit to unload.
4.1 Power Distribution
4. ESU Tests with QA-ES
This chapter explains the tests that can be conducted by the QA-ES
on an ESU, as well as the features available with the PRO-Soft QAES software accessory.
This test checks the power provided by the ESU over a range of load
resistances. The QA-ES allows you to specify a range of loads, over
which you test the ESU output power to see if it is within the specified limits.
Per IEC 601-2-2 the power output cannot be reduced by more than
10W, or 5% of the minimum power output level. Per ANSI/AAMI
HF18-1993 the power output must be within 20% of the ESU manufacturer’s specifications.
Load Resistance Range
Equipment
Monopolar100 - 1000 ohms50 - 2000 ohms
Bipolar10 - 500 ohms10 - 1000 ohms
Test setup for ESU power distribution test:
IECANSI/AAMI
4.2 HF Current Leakage
This test checks to see whether or not the active and dispersive leak age currents are within acceptable limits. There are four test setups
to accomplish this testing.
Per IEC 601-2-2 and ANSI/AAMI HF18-1993 the ESU shall be operated at the maximum output setting in each operating mode. The
limits for the acceptable leakage currents depend upon the test configuration.
Test ConfigurationLimits of Acceptable Leakage Current
Measured on electrodesThe leakage current should not exceed 150 mA
1
Test ConfigurationLimits of Acceptable Leakage Current
BipolarThe leakage current should not exceed 1% of the
Measured at equipment
terminals
maximum bipolar rated power output.
The leakage current should not exceed 100 mA.
1.Grounded HF Equipment: Measurements of the HF current leakage. The ESU is grounded. The test load is 200 ohms
and the ESU must be operating at maximum power. The current
leakage measured directly at the instrument's terminals must not
exceed 100 mA.
Test setup in compliance with IEC 601.2.2, sec. 19.101a, test 1,
fig. 102 and sec. 19.102. (Adopted by ANSI/AAMI HF18-
1993)
Test setup in compliance with IEC 601.2.2, sec. 19.101a, test 2,
fig. 103 and sec. 19.102. (Adopted by ANSI/AAMI HF18-1993)
2.HF Isolated Equipment: Measurements of the HF current
leakage from the active and neutral electrodes. The test load
is 200 ohm and the ESU must be operating at maximum power.
The current leakage measured directly at the instrument's termi nals must not exceed 100 mA.
Active electrode test setup in compliance with IEC 601.2.2, sec.
19.101b, fig, 104 and sec. 19.102. (Adopted by ANSI/AAMI
HF18-1993)
2
4.3 REM Alarm
Neutral electrode test setup in compliance with IEC 601.2.2,
sec. 19.101b, fig, 104 and sec. 19.102. (Adopted by
ANSI/AAMI HF18-1993)
This test ensures that the ESU will sound an alarm if the resistance
between the two neutral electrodes exceeds your specified limit. The
program directs the OA-ES to gradually increase the resistance. At a
certain value, the ESU should sound an alarm. Test setup for ESU
REM alarm test.
3
This page intentionally left blank
4
5.1 Required Test
Equipment
5. Control and Calibration
This chapter explains the QA-ES maintenance procedures, including
testing and calibration.
ESU, 200W in 75 ohms with 500 kHz
Signal generator, 3 MHz, harmonics better than -40 dBc with
500 kHz, 0 dBm
Digital multimeter
Digital multimeter, HP 34401 or equivalent
RMS / Peak Voltmeter, RHODE & SCHWARY URE 3 or
equivalent
Oscilloscope, 10 MHz
Electrosurgical probe for QA-ES
Computer (PC) with RS-232C interface
Printer with parallel interface
Short circuit SMB terminal, female
5.2 Preparation
The QA-ES should be
switched on for a minimum
of 15 minutes before the
test starts to ensure stable
working temperatures.
5.3 Function Testing
Measurements should
be performed only on
main supply, and not
both 230V and 115V.
Test cables
WARNING!
HIGH VOLTAGES ARE CAPABLE
OF CAUSING DEATH!
USE EXTREME CAUTION WHEN PERFORMING TESTS AND CALIBRATION. USE ONLY INSULATED TOOLS WHEN THE UNIT IS
PLUGGED IN, AND THE CASE HOUSING IS OFF.
Before performing testing and calibration, you must dismantle the
housing. This is accomplished by removing the side plates fastened
with Velcro straps. Then, remove the bottom plate by loosening the
four screws. Afterwards, move the bottom plate of the isolation box
by loosening the 16 screws holding it in place. Adjust the voltage
switch at the rear of the QA-ES so that is corresponds to the main
voltage (115 or 230V).
1.Power Supply. Connect the multimeter in series on one of the
mains supply leads to measure the current consumption. Turn
1
5.4 Calibration
on the QA-ES. Don't activate any functions. Measure the current. Required value:
230±10%: 80mA -10/+20mA
115±10%: 160mA -20/+40mA
2.User Interface and Display. Check that the display and the
user interface are working normally.
3.Cooling Fan. Check that the fan increases speed when you
press on Start (F3) in the main menu and that the speed decreases about 10 seconds after you press STOP (F3).
4.Serial and Parallel Interface. QA-ES can be connected to a
printer and a PC. Check that booth the serial and parallel inter faces work.
5.Foot Switch. Connect a multimeter to the foot switch output
on the QA-ES (green). Check that the relay for the foot switch
is connected when you press START (F3).
6.Fixed Load. Measure the value of the fixed load with a multimeter (blue). Required value: 200 ohms ±10 ohms.
7.HousingIsolation. Use a multimeter to check the isolation between the housing and the measuring inputs. It is important to
check all of the terminals on the housing: ‘Var. Load', 'Fixed
Load' and 'Foot Switch’.
1.Offset Voltage. Short-circuit the RF input on the sampling unit
(J4). All the measurements and reference values in this section
refer to the sampling unit. Measure the voltage at each of the
following points and adjust them to the following values with
the potentiometers specified between brackets. The voltage
should be set as close to 0V as possible.
2.Load Resistance. Set the QA-ES in calibration mode by hold-
It is important to calibrate the resistance of
the measuring leads on
the multimeter itself be-fore starting the test.
ing the function key F1 while switching the instrument ON.
Continue to press on the key until the main menu appears. Connect a multimeter (hp 34401A) to the 'Var. Load' input. The resistances are calibrated by turning the small wheel until the display shows a value as close as possible to the value read. To
calibrate the next resistance, press on UP (F1) or DOWN (F2).
3.Measuring Resistance. Set the QA-ES in calibration mode.
Before mounting it, the measuring resistance R1 on the Load
Board should be measured with hp 34401A. The measured value should be noted near R1 on the Load Board. Use this value
for R1 in the following equation and for ‘Meas. Reas.’ in the
calibration menu by using the wheel. Press QUIT (F5) to save
calibrated values.
2
Meass
R
R
Re
*
=
+
1 100
1 100
4.Measuring Device
Clin
kdB
=
20
10
gaingain
QQ
Clin
=•
Measuring The Effect. Set the QA-ES in test mode by holding
F2 while switching the power on. Connect the signal generator
to the RF input on the sampling unit. The signal applied is a
pure sine of 500 kHz. URE 3 is connected in parallel with the
sampling unit to measure the applied signal. Press START (F3)
and read the measured effect in dB. Vary the level of the ap plied signal from -16 dBm to +10 dBm and find the average
value for the difference between the value read on the QA-ES
and the applied value measured with URE 3. Use the term below to calculate a linear percentage value from the difference
value in dB.
KdB: Average value for the difference between the value read
on QA-ES and the applied value measured with URE 3.
Clin: Linear value of kdB.
The following term will lead to a new value for Q_gain.
Turn the wheel and press ENTER to specify a new value for
Q_gain. Repeat the test until the average value between the
QA-ES and URE 3 is smaller than 0.1 dB. This corresponds to
an error of 2.3%.
Peak Detector. Once the measuring device used in measuring
the effect has been calibrated, you can calibrate the peak detector. Use the same test setup as for calibrating the effect. Measure with a signal level of 200 mVp and 1 VP applied from the
generator. The applied amplitude is measured with URE 3. The
difference between the value read on the QA-ES and URE 3
should not exceed 10% otherwise the P_offset and N_offset in
the calibration menu must be changed until this requirement is
met. If the positive peak value read is too small P_offset should
be reduced. N_offset should be reduced if the negative peak
value is too low. Press QUIT (F5) to save calibrated values.
5.Measuring Accuracy of the total system. Connect the coaxial
cable from the Load Board to the RF-input (J4) on the Sampling
Unit. Put the ground plate of the isolating box back into place.
Connect a high voltage probe in the port for variable load resistance on the QA-ES. The probe's coaxial cable should be connected to URE 3. The banana adapters of the probe are connected with the high voltage generator. The following mathematical
formula presents the applied effect in dBm as a curve of the effect on URE 3.
3
dBcoeff
RR
R
=•
+
20
12
2
log
R1: Resistance in the high voltage probe. Specified on the
mURE
PP
dBcoeff
−+
#
mqa es
PP
dBmdBmdB
[
[][].
−<
−
017
qa es
I
mA
−
>
50[]
mqa es
m
PP
P
−
<
−
4%
probe. Nominal value 47 kohms.
R2: Measuring resistance in the high voltage probe. Speci-
fied on the probe. Nominal value 37.5 ohms.
The measuring accuracy of the QA-ES should be checked for
the following resistance values and the corresponding signals
applied.
RAppliedsignalmeasured on URE3
[ohms][W] [dBm]
75
300
1500
Calculate theaveragevalueforthe difference between QA-ES
and URE 3 and correct the value for Q_gain in the calibration
menu by following the instructions given before under item 4 in
this section.
Requirement:
NOTE
Remember to set the reference impedance in
URE 3 to the same value
as the load resistance
used for measuring.
200∀50W150∀50W100∀50W
52.0 - 54.0
50.0 - 53.0
47.0 - 52.0
Remember that the
scope output is not calibrated!
In linear form this corresponds to:
6.Scope Output. Connect the high voltage generator to 'Var.
Load' and the oscilloscope to 'Scope Output'. Set the QA-ES to
continuous mode and start measuring by pressing on Start. Activate the high voltage generator and measure the peak-to-peak
voltage on the oscilloscope. Check that this value equals the
current flow read on the QA-ES.
Nominal value:4.5 ∀ 0.4 V/A
4
6.1 Processor Board
6. Component Functions and Parts
This chapter provides a detailed description of the functions of the
main components of the QA-ES, as well as a parts list for cross-ref erence. Reference is made to the component location and schematic
diagrams to assist servicing personnel. These diagrams are foldouts,
and are located in Appendix B.
(Refer to QA-ES Processor Board Component Location Diagram
and Schematic Diagrams 1 and 2) The Processor Board is installed
on the inside of the front plate. It comprises the:
power supply;
microprocessor system;
display;
function keys;
interface towards Sampling Unit, and;
RS 232 port and printer port
The Processor Board activates and controls the sampling procedure
in the QA-ES. The data is routed back from the Sampling Unit to
the microprocessor, where the results will be calculated before being
displayed.
1.Power Supply
(See QA-ES Processor Board Schematic Diagram 2). The Processor Board receives 12 VAC from the Transformer located at
the rear of the QA-ES via J4. The voltage is then converted with
diode D9 and filtered through C1. Voltage regulator U8 supplies the circuits with +5 V. The voltage is adjusted with resistors Rl0 and R11. Schottky diode D4 protects the 5V power
supply to the printer interface to avoid any power flow between
the printer and the Processor Board while the QA-ES is
switched off. Capacitive switch regulator U9 generates -12 V
from +12 V. V-comparator U10 monitors the +5V voltage and
sends a reset signal to the microprocessor when the voltage
drops under 4.75 V.
Transistor Q4 controls the voltage to the compressor at the rear
of the QA-ES. The compressor is connected to J5. The microprocessor issues a digital signal at PWMA, depending on the
ventilation speed desired. Transistor Q2 controls the basic power in Q4. When Q4 is turned off (low ventilation speed), the
compressor receives power via transistors R8 and R17, leading
to a voltage drop.
2.Microprocessor System
1
(See QA-ES Processor Board Schematic Diagram 1). The microprocessor system is in PCMCIA card format (U1). It comprises a MC68HC16 microprocessor with I/O, 256K Flash ROM
and 64K RAM. It can be reprogrammed or replaced when upgrading the software. The processor's I/O includes an asynchronous serial port (ACIA), synchronize serial port, parallel I/O,
A/D converter and PWM output. In addition, 8 bit of the data
bus are accessible, as well as chip-select lines for direct access
to the external I/O.
The processor is timed with a 16.67 MHz timer frequency, con trolled by an internal crystal in the component. To check the
timer frequency, a 1024 Hz square signal is applied to one of
the PWM outputs. The signal can be measured at pin TP 1.
Ul3 is an EEPROM connected to the processor component via a
parallel I/O. Ul3 saves the calibration parameters for the Sampling Unit. The parameters can be stored independently of the
processor component so that the QA-ES does not have to be recalibrated when upgrading the software.
3.Function Keys
(See QA-ES Processor Board Schematic Diagram 1). The QAES is operated by touch keys and a universal digital encoder.
The encoder, or 'knob', will be given different functions depending upon the parameters that are to be changed. The encoder is connected to two touch keys with unchangeable functions, 'ENTER' and 'CANCEL'. The QA-ES is operated from 5
'soft-keys' linked to menus shown on the display.
Latch U2 buffers data from touch keys SW1-SW7. If you press
one of the keys, power will flow into the base at Q1 via RP1,
thus controlling an interrupter input on the processor unit. When
the interrupter input is activated the processor reads latch U2
latch to find out which key was pressed. The same interrupter
input is activated via diodes D1 and D2 when operating the encoder. The encoder is read via I/O ports on the processor com ponent.
4.Display
(See QA-ES Processor Board Schematic Diagram 1). The QAES is equipped with a 240 x 64 dot graphic display with a builtin character generator. When in character mode, the display
shows 40 x 8 characters. The display is controlled by the processor unit via the data bus. The display's contrasting voltage is
regulated with voltage regulator U12. Voltage (and contrast) are
controlled with the potentiometer R15. This is the only point of
adjustment at the processor board.
The display offers EL background light to make it easier for the
user to read. U11 is an oscillator generating an operating voltage of about 90 VAC for the EL component.
5.Interface Towards Sampling Unit
2
(See QA-ES Processor Board Schematic Diagram 1). All data
transferred between the Sampling Unit and the Processor Unit is
in serial format to simplify the optical interface on the sampling
board. Communication is controlled by the synchronous serial
interface on the processor unit. The signals are transmitted via
pin board base J1.
6.Serial Port
(See QA-ES Processor Board Schematic Diagram 2). The serial
port is adapted to a 9-pin RS-232C format. The port is set to
9600 baud, 8 data bits, 1 stop bit and no parity. RS-232 driver
U6 drives the data signals. The handshake is software-oriented.
The command responses are returned via the D-sub terminal.
7.Printer Output
(See QA-ES Processor Board Schematic Diagram 2). The Processor Board’s printer output has a standard 25-pin D-sub contact for Centronix interface. The output is built around 3 HCMOS circuits; U3, U4 and U5. The circuits are connected to the
data bus and I/O ports of the processor unit. U3 is a latch for the
8 parallel data lines. U4 is the driver for the outgoing commands, while U5 acts as a buffer for incoming commands. RP3
comprises pull-up resistances for the input lines. All signals to
the printer output are filtered to reduce high frequency radiation.
6.2 Sampling Unit
The QA-ES Sampling Unit is placed inside the internal protection
box of the instrument. The card is fixed vertically on the right side
of the protection box. The unit comprises:
power supply;
measuring device, and;
interfaces towards the Processor Board and the Load Board.
The Sampling Unit converts the applied RF signal to a low frequency signal proportional to the mean square of the RF signal. The peak
value of the applied signal is also measured. The measuring values
are sampled with a 12 bits A/D converter and the data is transmitted
to the Processor Board for calculation and presentation on the display.
1.Power Supply
(See QA-ES Processor Board Schematic Diagram 1). The Sampling Unit receives -10 VDC and +10 VDC from the Load
Board via J1). Voltage regulators U17 and U19 provide, respectively, +6V and -6V for the RMS DC converter. Voltage regulators U18 and U20 provide the peak detector with +9V and -9V.
Voltage regulator U16 supplies the opto-coupler in the interface
towards the Processor Board with + 5V.
2.Measuring Device
3
The measuring device consists of two blocks:
•a RMS DC converter for current measuring, and;
•a peak detector for measuring the peak voltage in the sig-
nal.
The peak detector is divided into two actions for measuring the
positive and negative peak values.
3.Input Filtration
At the measuring system input, the incoming signal is filtered
down to a lower level through a filter based on L1, L2, C5, C6
and C7. The 3dB frequency is set to 10 MHz. The filter can be
found on diagram 1 of 3.
4.RMS DC Converter
(See QA-ES Processor Board Schematic Diagram 1). The RMS
DC converter is based upon an analog multiplicator from analog
devices AD834, U15. U15 and U25 form a circuit that calculates the mean square of the applied signal. A filter, consisting
of C8, C9. C50, C51, R35 and R36, determines the constant
time factor for the integration, which will correspond to the am plitude of the outgoing signal at U25. Further calculations of
RMS values and current are carried out in the software on the
Processor Board.
Two amplification steps following the mean square circuit ensure optimum dynamics in the measuring system. Both steps
U13 and U14 provide 20 dB amplification. By measuring the
signal before U13, and after U13 and U14, you can determine
which signal level makes the best use of the A/D converter dynamics.
5.Peak Detector
(See QA-ES Processor Board Schematic Diagram 3). The peak
detector is divided into two sections: one for detecting the positive peak voltage, and the other for detecting the negative peak
voltage. The detector is based on a transistor diode connection.
The positive peak voltage is detected by transistors Q1 and Q4,
to which are attached various components. The same goes for
the negative peak voltage with Q2 and Q5. The positive peak
voltage is built up over the C27 capacitor, whereas the negative
peak voltage is built up over the C28 capacitor. U23 drains
small amounts of current from C27, and provides the AID converter with correct polarization and low impedance for the positive peak detector. U22 does the same for the negative peak detector. The peak detector ran be reset by short-circuiting C27
and C28 via octol analog switch U24 (See QA-ES Processor Board Schematic Diagram 2).
6.A/D Converter
4
The measuring device uses a 12-bit A/D converter from linear,
Ul. This circuit measures both RMS values, and positive and
negative peak voltage. Each of the signals is multiplexed into
the A/D converter via the U24 switch.
The sampling speed depends on the Processor Board's reading
speed. Schottky diode D7 protects the A/D converter from incoming negative signals.
7.Interface Towards Load Board and Processor Board
The measuring device and load resistances in the QA-ES are
isolated from the processor unit and the housing by a galvanic
shield. This protects the user if a fault in the QA-ES produces
ground currents during the measurements. This interface is
based upon opto-couplers and various mains transformers for
the user interface and the measuring device in the QA-ES.
Data is transferred digitally between the Sampling Unit and the
Processor Board in serial form via J3 to simplify the optical interface. Opto-couplers U5, U6, U7 and U8 are used in transmitting signals from the processor system to the Sampling Unit.
The Opto-coupler U9 transfers the measurement from the Sampling Unit back to the processor system on the Processor Board.
The relays used in choosing load values on the Load Board are
also controlled via terminal J3 on the Sampling Unit. The commands are then transmitted from the Sampling Unit to the Load
Board via terminal J2.
8.High Voltage Protection
6.3 Load Board
Diodes D8-D15 are a protection device against high voltage signals entering the Sampling Unit.
(Refer to QA-ES Load Board Component and Schematic Diagrams).
The Load Board is located inside the internal protection box in the
QA-ES. The card is fixed vertically inside the box. The unit includes a power supply, load resistances with corresponding control
relays, and interface towards the processor system Sampling Unit, a
measurement resistor with attenuators and measuring Transformer
for the scope output.
The Load Board forms the load for the ESU being tested. The load
can vary from 10 ohms to 5200 ohms, or be fixed to 200 ohms. A
high voltage relay inside the Load Board (K1) connects the load to
or from the ESU.
1.Power Supply
The Load Board is supplied with 2 x 9 VAC from the Trafo in
the front part of the protection box via J14. The voltage is converted with diode D9, and filtered through Cl and C2. U1 is a
voltage regulator, providing the relay drivers +5 V. U1 also
supplies Sampling Unit opto-couplers U10, U11 and U12 with 5
5
V. Cl and C2 provide relay drivers U2 and U3 with +10 VDC
and -10 VDC. These voltages are used for controlling the relays
on the Load Board.
2.Load Resistances
There are two load resistances on the Load Board, seen from
the ESU being tested. The first has a fixed load of 200 ohms,
and is used to measure the current leakage. The other is a vari able load resistance, which varies in steps of 25 ohms, from 10
ohms to 5200 ohms.
3.Fixed Load Resistance
The fixed load resistance is based on the two 100 ohms non-inductive resistances (R17 and R18) connected in series. The total
load can be 350 W continuously.
4.Variable Load Resistance
The variable load resistance is also based on a non-inductive resistance (R3 - R16). In addition, relays (K2 - K10) have been
used to offer the choice between different combinations of resistances. You can therefore obtain any value between 10 and
5200 ohms.
5.Measuring Resistance and Attenuator
The measuring signal transmitted to the Sampling Unit is
drained over a 2.0 ohms resistance (R1) in series with the vari able load resistance (R3 - R16). Before the signal is transmitted
to the measuring device, it passes through a variable attenuator,
based on a 10 dB attenuator (R20 - R22) and a 20 dB attenuator
(R26 - R30). By connecting these attenuators in series, you can
obtain an attenuation of 30 dB.
6
6.Interface Towards Sampling Unit
When the signals pass through an optical barrier between the
processor system and the rest of the electronics, the Sampling
Unit's F1 terminal controls the relays, choosing the load and attenuator values on the Load Board.
7.Scope Output
The scope output is based on a measuring Transformer L1,
mounted on the Load Board. The measuring Transformer is ter minated with a 50 ohms resistance R31.
8.Foot Switch
The output foot switch is based on a relay (K11), and is used in
triggering the ESU being tested.
Screw recessed head pozDIN 7985 M3x104
Screw recessed head pozDIN 7985 M4x304
Nut M36
Nut M44
Safety nut M41
Spring washer M44
Washer M4 12mmø DIN 90211
Spacer for fixing the front plate5
Contact plate1
RF protection box1
Front plate w/5 screws for circuit board1
Rear plate1
Front foilv1
58-755-21
SMB coax conn.Straight plug crimp1
SMB coax conn.Straight PCB recept.1J13
Printer board contact6-polt hannPhoe1J1, J2
Printer board clip3-polt Phoe1J14
Safety holder Farn 1344772F1, F2
Safety fuse600mAT Farn 1502022(F1), (F2)
High voltage cable35 cm 10-/m @ 100 m2
Coax. cable RG17440 cm Farn 1253261
Coax terminatorDouble leg right ang.1
Plastic SpacerENISO 3x2538
Plastic SpacerENISO 3x206
Nylon ScrewM4x07x12 mm36
Screw pinDIN 553 M4x22 mm14
Dekk-kappe todelt30
Bracket for power resistor30
1L1
11
This page intentionally left blank
12
Appendix A - GLOSSARY OF TERMS USED
Bipolar ElectrosurgeryElectrosurgery where current flows between two bipolar electrodes that are posi -
tioned around tissue to create a surgical effect. Current passes from one electrode, through the desired tissue, to the other electrode, thus completing the circuit without entering any other part of the patient's body. Neutral plates are not
employed in the bipolar technique. Both electrodes are generally of the same
size.
Bipolar OutputAn isolated electrosurgical output where current flows between two bipolar elec-
trodes that are positioned around tissue to create a surgical effect in that tissue
(usually desiccation).
BlendA waveform that combines features of cut and coag waveforms; current that cuts
with varying degrees of hemostasis.
Crest FactorThe amount of heat generated is relative to the mean power value. The crest
factor depends on the load resistance and is defined as the ratio of peak value to
effective value. A sine wave has a crest factor of 1.4 and provides the cleanest
form of cutting.
Diathermy, also Surgical
Diathermy; Electrosurgery
ESUElectrosurgical Unit. This is a term which is inclusive of both the electrosurgical
Cut (Cut Mode, Pure Cut)A low voltage, continuous waveform optimized for electrosurgical cutting.
Isolated OutputThe output of an electrosurgical generator that is not referenced to earth ground.
Monopolar ElectrosurgeryA type of electrosurgery involving a small (active) electrode and a large neutral
Monopolar OutputA grounded or isolated output on an electrosurgical generator that directs current
Resistance (Impedance)Resistance to the flow of alternating current, including simple direct current re -
A surgical technique used to cut or coagulate cellular tissue. To avoid muscle
contractions, only high frequency currents and voltages of more than 100 kHz
are used. The electric current directs the heat into the tissue. The patient is connected to two electrodes, allowing the current to flow through the body. The ac tive electrode will generate a large amount of heat, due to the high current densi ty and the small surface of the electrode.
generator and its connecting cables.
(neutral plate) electrode. The small surface of the active electrode provides very
good results in coagulating and cutting. The neutral plate of modern units is
split, thus controlling the circuit, including the contact between electrodes and
patient.
through the patient to a patient return electrode.
sistance and the resistance produced by capacitance or inductance. The resistance of a material is its tendency, measured in ohms, to oppose the flow of
electric current or, viewed another way, the material's tendency not to conduct
the current.