Metron QA-45 Service manual

QA-45
User & Service Manual
QA-45 Defibrillator and
Transutaneous Pacemaker Analyzer
P/N 13060
Copyright 2001 by METRON. All rights reserved.
METRON:
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1 Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500 Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009 E-mail: support.us@metron-biomed.com E-mail: info@metron.fr E-mail: support@metron.no
Disclaimer
METRON provides this publication as is without warranty of any kind, either express or implied, in­cluding but not limited to the implied warranties of merchantability or fitness for any particular purpose. Further, METRON reserves the right to revise this publication and to make changes from time to time to the content hereof, without obligation to METRON or its local representatives to notify any person of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.
Limited Warranty
METRON warrants that the QA-45 Defibrillator/Transcutaneous Pacemaker Analyzer will substan­tially conform to published specifications and to the documentation, provided that it is used for the pur­pose for which it was designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair any defective system, if the fault is due to a manufacturing defect. In no event will METRON or its local representatives be liable for direct, indirect, special, incidental, or consequential damages arising out of the use of or inability to use the QA-45 Defibrillator/Transcutaneous Pacemaker Analyzer, even if advised of the possibility of such damages. METRON or its local representatives are not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or in­ability to use the QA-45 Defibrillator/Transcutaneous Pacemaker Analyzer. Neither METRON nor its lo­cal representatives will accept, nor be bound by any other form of guarantee concerning the QA-45 De­fibrillator/Transcutaneous Pacemaker Analyzer other than this guarantee. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.
ii
Table of Contents
1. Introduction........................................................................................................................................1
1.1 QA-45 Features ...........................................................................................................................1
1.2 Defibrillator Analyzer Specifications..........................................................................................2
1.3 Transcutaneous PPacemaker Analyzer Specifications.............................................................5
1.4 General Information......................................................................................................................6
2. Installation..........................................................................................................................................1
2.1 Receipt, Inspection and Return...................................................................................................1
2.2 Setup..............................................................................................................................................1
2.3 Power ............................................................................................................................................2
2.4 Internal Paddles.............................................................................................................................4
2.5 Special Contacts............................................................................................................................4
2.6 PRO-Soft QA-40M/45..................................................................................................................4
3. Operating QA-45................................................................................................................................1
3.1 Control Switches and Connections.............................................................................................1
3.2 QA-45 Menu and Function Keys.................................................................................................3
3.3 Menu and Messages: Defibrillator Mode....................................................................................4
3.4 Menu and Messages: Transcutaneous Pacemaker Mode.........................................................6
3.5 Test Result Printouts.....................................................................................................................8
4. Defibrillator Mode Testing...............................................................................................................1
4.1 Introduction..................................................................................................................................1
4.2 Test Preparation............................................................................................................................1
4.3 Energy Test ..................................................................................................................................3
4.4 Cardioversion Test........................................................................................................................4
4.5 Maximum Energy Charging Time Test.......................................................................................6
4.6 Shock Advisory Algorithm Test.................................................................................................7
5. Transcutaneous Pacemaker Mode Testing.....................................................................................1
5.1 Introduction..................................................................................................................................1
5.2 Testing Preparation ......................................................................................................................1
5.3 Demand Sensitivity Test..............................................................................................................3
5.4 Refractory Period Test..................................................................................................................5
6. Control and Calibration....................................................................................................................1
6.1 Required Test Equipment............................................................................................................1
6.2 Preparation....................................................................................................................................1
6.3 References.....................................................................................................................................1
6.4 Test................................................................................................................................................1
7. Component Functions and Parts......................................................................................................1
7.1 Theory of Operation .....................................................................................................................1
7.2 Processor Board.............................................................................................................................1
7.3 Sensor Board.................................................................................................................................3
7.4 ECG Signal Distribu- tion Board..................................................................................................4
7.5 Pacer Unit......................................................................................................................................5
7.6 Component Parts ..........................................................................................................................6
Appendix A – Diagrams ........................................................................................................................A-1
Appendix B – Error Report Form........................................................................................................B-1
Appendix C – Improvement Suggestion Form....................................................................................C-1
iii
Manual Revision Record
This record page is for recording revisions to your QA-45 User & Service Manual that have been published by METRON AS or its authorized representatives. We recommend that only the manage­ment or facility representative authorized to process changes and revisions to publications:
make the pen changes or insert the revised pages;
ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as su-
perseded and placed in a superseded document file, and;
enter the information below reflecting that the revisions have been entered.
Rev No Date Entered Reason Signature of Person Entering Change
2.60-1 4-30-01 General update
iv
This page intentionally left blank.
v

1.1 QA-45 Features

1. Introduction

This chapter describes the Metron’s QA-45 Defibrillator / Transcutaneous Pacemaker Analyzer features and specifications.
The QA-45 Analyzer is a precision instrument for testing defibrilla­tors and transcutaneous pacemakers, and is designed to be used by trained service technicians.
The defibrillator function of the QA-45 measures the energy output, and ensures that the defibrillator complies with specified require­ments. QA-45 has a built-in load resistance of 50 ohm, which roughly corresponds to the impedance of the human body. The de­fibrillator pads are placed on the QA-45 contact plates. Thus, the de­fibrillator is connected through the load resistance. When the defib­rillator is discharged, QA-45 calculates and displays the energy de­livered.
In the pacer function the QA-45 tests all types of transthoracic pace­makers. The testing is menu driven, and simple to operate. QA-45 measures and displays a pulse’s amplitude, rate, energy and width. It also conducts demand sensitivity tests, measuring and displaying re­fractory periods, and immunity tests, which determine the pacemak­er’s susceptibility to 50/60 Hz interference.
1

1.2 Defibrillator Analyzer Specifications

1. Energy Output Measurement
High Range
Voltage <5000 volts
Maximum current 120 amperes
Maximum energy 1000 Joules
Accuracy
Trigger level 100 volts
Playback amplitude 1 mV/1000 V Lead I
Test pulse 100 + 4 Joules
Low Range
Voltage <1000 volts
Maximum current 24 amperes
Maximum energy 50 Joules
± 2 % of reading for >100 Joules ± 2 Joule of reading for <100 Joules
Accuracy
Trigger level 20 volts
Playback amplitude 1 mV/200 V Lead I
Test pulse Approx. 4 Joules
Load Resistance
Display Resolution 0.1 Joules
Measurement Time Window Absolute Max. Peak Voltage Pulse Width 100 ms
Cardioversion
Oscilloscope Output
High measure range 1000:1 amplitude-attenuated
Low measure range 200:1 amplitude-attenuated
Waveform Storage And Playback
Discharge can be viewed via ECG outputs and paddles.
Output: 200:1 Time Base expansion.
Sync Time Measurements
Timing window Starts - 40 ms at each R-wave peak.
Test waveforms All waveform simulations available.
Delay time accuracy
Charge Time Measurement
From 0.1 seconds to 99.9 seconds.
± 2% of reading for >20 Joules ± 2 Joule of reading for <20 Joules
50 ohms ± 1%, non-inductive (<1 µH)
100 ms
6000 volts
Measured time delay ± 2 ms
± 1 ms
2
2. ECG Wave
ECG General
Lead configuration 12-lead simulation. RL, RA, LA, LL, V1-6
Output impedance Limb leads 1000 ohms to RL
V Leads 1000 ohms to RL
All other signals are in relative proportion to Lead amplitude as follows: The amplitudes are shown for a Lead I amplitude by 1 mV:
Lead I 1.0 mV (LA - RA)
Lead II 1.5 mV (LL - RA)
Lead III 0.5 mV (LL - LA)
V Lead 1.5 mV (V - 1/3 (LL+LA+RA))
High Level Output (ECG Jack)
1/4" standard phone-jack with an amplitude of 1V/mV of low level Lead II signal
Defibrillator Contact Plates
Same amplitude as Lead I low level ECG.
1 mV between contact surfaces.
Playback
200 to 1 time-base expansion of defibrillator pulse by playback to ECG Leads
Manual ECG Performance Test
DC Pulse 4 seconds 1.0 mV
Square wave 2 Hz 1.0 mV p-p biphasic
Triangular wave 2 Hz 1.0 mV
Sine 0.1, 0.2, 0.5, 10, 40, 50, 60, and 100 Hz
Amplitude 0.5, 1.0, 1.5, 2.0 mV (Lead II)
Accuracy
± 5 % (Lead II 1.0 mV)
ECG Performance Test
Gain/Damping 2 Hz square wave
Frequency Response
Low Frequency 4 second DC pulse
Band Pass 10 Hz sine
Monitor -3dB point: 40 Hz sine
Power Line Notch Filter Linearity 2 Hz triangle wave
Normal Sinus
Rates 30, 60, 80, 120, 180, 240 and 300 BPM.
Accuracy
Amplitudes 0.5, 1.0, 1.5 and 2.0 mV (Lead II)
Accuracy
Automatic ECG Rate Test
Arrhythmia Selections
vfib Ventricular Fibrillation
afib Atrial Fibrillation
50 Hz sine
±1% of selection
±5 % (Lead II 1.0 mV)
3
blk II Second degree A-V block
RBBB Right Bundle Branch Block
PAC Premature Atrial Contraction
PVC_E Early PVC
PVC_STD PVC
PVCRonT R on T PVC
mfPVC Multifocal PVC
bigeminy Bigeminy
run5PVC Bigeminy Run of 5 PVCs
vtach Ventricular Tachycardia
Shock Advisory Test Algorithms
ASYS Asystole
SVTa_90 Supraventricular Tachycardia
PVT_140
PVT_ 160
MVT_140
MVT_160
CVF Course Ventricular Fibrillation
FVF Fine Ventricular Fibrillation
4
1.3 Transcutaneous PPace-
maker Analyzer Specifications
1. TEST LOAD RANGE
2. PULSE MEASUREMENTS
3. DEMAND SENSITIVITY TEST
50 to 2300 ohms in step of:
50 ohm up to 200 ohms
100 ohm from 200 up to 2300 ohms
Accuracy
Oscilloscope Output
50 - 150 ohm 10.24:1 amplitude attenuation
200 - 500 ohm 41:1 amplitude attenuation
600 - 2300 ohm 164:1 amplitude attenuation
Amplitude 4 to 300 mA (100 ohm load)
Accuracy
Max. Amplitude 300 mA all loads
Rate 30 to 800 ppm
Accuracy
Pulse width 0.6 to 80 ms
Accuracy
Waveforms Square(SQR), Triangle(TRI), and Havemine
ECG output Amplitude 0 - 4 mV
Pacer input (Load depended)
Amplitude (50 ohm) 0 10 mV
Resolution (50 ohm)
Amplitude: (500 ohm) 0 - 100 mV Resolution: (500 ohm) 1 mV
Defib. Pads Amplitude 0 10 mV
Resolution 0.1 mV
Waveform width 10, 25, 40, 100 and 200 ms
Pacer rate 30 to 120 ppm
50 - 1300 ohm ±1% 1400 - 2300 ohm ±1.5 %
±5 % or ±0.5 mA
±1% or 2 ppm
±1% or ±0.3 ms
(SSQ)
Resolution 40 µV
40 µV
Immunity Test
50/60 Hz Interference Signal
ECG output 0 - 4 mV peak in steps of 0.4 mV
Pacer input (Load dependent)
0 - 10 mV peak in steps of 1 mV (50 ohm)
0 - 100 mV peak in steps of 10 mV (500
ohm)
Defibrillator pads 0 - 10 mV peak in steps of 1 mV
5
4. Refractory Period Measurement
20 to 500 ms (both Pacing and Sensing) Accuracy: ±2 ms

1.4 General Information

Temperature Requirements
+15°C to +35°C when operating 0°C to +50°C in storage
Display
Type LCD graphic display
Alphanumeric format 6 lines, 40 characters
Data Input/ Output (2) Parallel printer port (1); Bi-directional RS
Power
Mechanical Specifications
Housing High impact plastic case
Height 9.8 cm 3.9 in.
Width 24.8 cm 9.8 in.
Depth 28.0 cm 11.0 in.
Weight 2.06 kg (with battery) 4.5 lbs
Recommended Printer HP DeskJet 500C / 550C and Canon BJ
-232C (1) for Computer control
2 x 9 volt alkaline Battery Duracell MN1604 (or equivalent) for 20 -25 opera­tional hours, or 240 VAC (Battery Elimina­tor), 115 VAC for US.
-10SX.
Standard Accessories
110 V or 220 V AC Adapter (P/N 17021)
Internal paddle-contact adapter (P/N 13403)
Ground contact adapter (P/N 13404)
Snap-to-banana adapters (10 pk) (P/N 17023)
User and Service Manual QA-45 (P/N 13060)
Additional Accessories
Defib. paddle adapter
(specify defibrillator type) (P/N 13410)
Pacemaker external load cable
(specify type pacemaker type} (P/N 13415)
Carrying case (P/N 13422)
Carrying case, ext. printer (P/N 10500)
PRO-Soft QA-40M/45 software (P/N 13600)
PRO-Soft QA-40M/45 DEMO (P/N 13601)
User Manual PRO-Soft QA-40M/45 (P/N 13605)
Storage
Store in the carrying case in dry surroundings within the temperature range specified, without battery. There are no other storage requirements.
6
Periodic Inspection
The unit should be calibrated every 12 months.
7
2.1 Receipt, Inspec-
tion and Return

2. Installation

This chapter explains unpacking, receipt inspection and claims, and the general procedures for QA-45 setup.
1. Inspect the outer box for damage.
2. Carefully unpack all items from the box and check to see that you have the following items:
QA-45 Defibrillator/Transcutaneous Pacemaker Analyzer
(PN 17020)
110 V or 220 V AC Adapter (P/N 17021)
Internal paddle-contact adapter (P/N 13403)
Ground contact adapter (P/N 13404)
10 pack, Snap-to-banana adapter (P/N 17023)
QA-45 User and Service manual (P/N 13060)
3. If you note physical damage, or if the unit fails to function ac­cording to specification, inform the supplier immediately. When METRON AS or the company’s representative, is in­formed, measures will be taken to either repair the unit or dis­patch a replacement. The customer will not have to wait for a claim to be investigated by the supplier. The customer should place a new purchase order to ensure delivery.

2.2 Setup

4. When returning an instrument to METRON AS, or the company representative, fill out the address label, describe what is wrong with the instrument, and provide the model and serial numbers. If possible, use the original packaging material for return ship­ping. Otherwise, repack the unit using:
a reinforced cardboard box, strong enough to carry the
weight of the unit.
at least 5 cm of shock-absorbing material around the unit.
nonabrasive dust-free material for the other parts.
Repack the unit in a manner to ensure that it cannot shift in the
box during shipment.
METRON’s product warranty is on page ii of this manual. The
warranty does not cover freight charges. C.O.D. will not be ac­cepted without authorization from METRON A.S or its repre­sentative.
1. Equipment connection is as shown in the typical setup below.
1
2. If PRO-Soft QA-40M/45 is being used, attach an RS-232 (null modem/data transfer configured) cable to the 9-pin D-sub outlet port located at the rear of the QA-45. Do not attach the printer cable to the QA-45. See below. However, if you are not using PRO-Soft QA-40M/45, and are sending directly to a printer for printouts, attach the printer cable to the 25-pin outlet port.
Some RS-232C cables are missing the connection be­tween the seventh and the eighth wires in the cable. The cable may still be called NULL-modem, but it will not work with the QA-45. Refer to the PRO-Soft QA-40M/45 Users Manual for more informa­tion.
NOTE

2.3 Power

1. Main On/Off Switch. QA-45 should remain off for at least 5 seconds before switching on again, in order to allow the test cir­cuits to discharge fully.
2. Low Battery Power. If battery power falls below 6.9 volts
(± 0.3 volts), the display will show 'Change battery, and reset
system'. This means that the battery should either be replaced or the instrument should be connected to a battery eliminator. The
2
Do not use mercury, air or car­bon-zinc batteries.
NOTE
main switch has to be switched off and then on again in order to use the instrument.
3. Changing Batteries. Open the compartments in the base of the instrument, replace the old batteries with new ones, and close the compartment covers. Use 9 volt alkaline batteries (Duracell MN1604 or similar).
3
4. Battery Eliminator
Remove the batteries and dis­connect the AC Adapter if you do not intend to use the QA-45 for an extended period of time.
NOTE

2.4 Internal Paddles

2.5 Special Contacts

2.6 PRO-Soft QA-40M/45

METRON’s AC Adapter plug-in power supply transformer al-
lows you to use the QA-45 anywhere a standard electrical outlet is available. To attach the AC Adapter insert the adapter’s small connector into the micro jack labeled “Batt. Elim. 9V DC” on the right rear of the unit. Plug the large connector into the near­est standard electrical outlet.
To be able to test defibrillators with internal paddles, an internal paddle adapter has to be used. These contacts have a banana plug that is attached to the standard paddle contact, and which is pro­tected by a plastic insulation washer.
Certain defibrillators (automatic models and those with pacer op­tions) have special contacts that are fastened to the electrodes at­tached to the patient. Metron AS has special adapters to suit the ma­jority of these defibrillators. These are available as accessories. They are more or less the same as the internal pad adapter except that they have a special adapter on the top, which matches the con­tact on the defibrillator.
Defibrillator paddle adapter (specify defibrillator type): (P/N 13410)
Pacemaker external load cable (specify type pacemaker type): (P/N 13415)
PRO-Soft QA-40M/45 is a front-end test automation and presenta­tion tool for METRON's QA-40M/45 Defibrillator/Transcutaneous Pacemaker Analyzer. It allows you to conduct the same tests, but by remote control via an IBM-compatible PC/XT with MS Windows (Version 3.1 or later). Additionally, the program has additional fea­tures to enhance your defibrillator and pacemaker maintenance.
Each of the QA-40M/45 tests can be run independently from PRO­Soft in the “Manual” test mode. Results are shown on the PC screen during testing, and the user is prompted to set the tested equipment accordingly. At the conclusion of tests, the user may print a report, store the test and results on disk, or both. Combinations of tests can be created and stored as “Test Sequences.” The program maintains a library of these sequences. In this way you can store and retrieve se­quences that are appropriate for each kind of equipment being tested at your facility.
PRO-Soft QA-40M/45 has its own user manual, which con­tains all the information con­cerning the program. If you or­der a demonstration version of the program you also receive the manual.
NO TE
4
Sequences can then be used independently, or can be attached to a checklist, written procedure, and equipment data in the form of a test “Protocol.” The equipment data can be entered manually into the protocol, or it may be retrieved by PRO-Soft from database soft­ware or other equipment files. Protocols can be created easily for each defibrillator or transcutaneous pacemaker in your inventory, and stored for use. Test protocols with results can be printed, or
stored on disk, and the results of testing can be sent back to the equipment database to close a work order and update the service his­tory.
5

3.1 Control Switches and Connections

Top Panel

3. Operating QA-45

This chapter explains the operating controls, switches and menus of the QA-45, details how to use them in testing , and provides general information on printouts and operator maintenance.
1. Power Switch Turns the power on and off.
2. Mode Switch Switches between PACE and Low / High ranges of defibrilla-
3 LCD Display Shows messages, test results and function menus.
4 Function Keys Fl - F5 are used to select the functions shown on the bottom
5. Contact Surfaces The defibrillator’s paddles are placed on these so that the
tor energy.
line of the LCD display, i.e., for selecting the function that is directly above the key.
discharged energy passes through the instrument in defib. mode and that the pacer signal passes through the instru­ment with a fixed 50 ohm load in the PACE mode.
1
6. Low Level ECG Connectors
7. Pacer Input Connectors
10 color-coded 4 mm safety terminals with snap-to-banana adapters.
The pacer output cables are connected to these so that the pacer signal passes through the instrument with a variable load selectable from 50 to 2300 ohms.
2
Rear Panel

3.2 QA-45 Menu and Function Keys

8. High Level ECG Jack 1/4” standard phone-jack for amplitude of 1 V/mV of low
9. Oscilloscope Output BNC-contact for attenuated signal in real time.
10. RS-232 Serial Port 9-pin D-sub
11. Printer Outlet Port 14-25 pin D-sub
12. Location of Batteries 2 compartments in the base of the instrument can be
13. Battery Eliminator Socket
level Lead 1 signal.
opened to replace the batteries.
Battery contact for connecting 9V 100 mA battery elimina­tor.
The QA-45 uses display and programmable function keys to provide flexibility and control over the operations. The upper part of the screen displays messages, status and results. The menu bar is at the bottom of the display. The function keys are numbered from Fl to F5.
A function is selected by pressing the key located directly under the Menu Item displayed in the menu bar. A menu unit is written in cap­ital letters.
The menu comprises three pages. The next pages of the menu are selected by pressing more-2, more-3 or more-1.
3

3.3 Menu and Messages: Defibrillator Mode

1. Startup Screen. The following screen will be displayed for 2 seconds after the QA-45 has been switched on.
2. Main Menu
a. Main Menu Bar (Page 1) - Mode switch in Low or High
position.
b. Second Menu Bar (Page 2)
c. Third Menu Bar (Page 3)
3. ECG WAVE (F1)
Choose desired wave by pressing UP (F2) or DOWN (F3). Save
this under ‘Wave” in the STATUS field by pressing SELECT (F4). Press CANCEL (F5) to cancel selection.
4
4. ADV. ALG. (Advisory Algorithms) (F2).
These ECG algorithms are meant to test the analysis and prompt-
ing feature of automatic and semi-automatic defibrillators. Choose desired selection by pressing UP (F2) or DOWN (F3). Save this under ‘Wave” in the STATUS field by pressing SELECT (F4). Press CANCEL (F5) to cancel selection.
5. CHARGE TIME (F3). Used to test the battery and charging capacitor in the defibrillator. It changes the text ‘Delay’ to ‘Chrg T’ in the RESULT field in the main menu.
6. PRINT HEADER (F4). Automatically writes a heading for the new test protocol.
7. WAVE AMPL. (Wave Amplitude) (F1).
Choose desired amplitude by pressing UP (F2) or DOWN (F3).
Save this under ‘Ampl” in the STATUS field by pressing SELECT (F4). Press CANCEL (F5) to cancel selection.
8. PLAY PULSE (F2) enables playback of the last discharge.
9. PERF. WAVE (Performance ECG) (F3).
Choose desired wave by pressing UP (F2) or DOWN (F3). Save
this under ‘Wave” in the STATUS field by pressing SELECT (F4). Press CANCEL (F5) to cancel selection.
10. SYSTEM TEST (F1) .
5
Note
QA-40M has an internally gener­ated test pulse. The control pulse is set at 1.2 Joules in the Low range and 28.5 Joules in the High range. The test pulse is not a cali­bration pulse, and should not be used as an indication of the gen­eral accuracy of the instrument. The test pulse is a good control for testing functions.
3.4 Menu and Messages: Transcutaneous
Pacemaker Mode
Choose a test variant by pressing UP (F2) or DOWN (F3) or
TEST PULSE (F1). Press CANCEL (F5) to cancel selection.
For ‘ECG0’, ‘ECG+’ and ‘ECG-’ see Chapter 6, Control and Calibration. For ‘A/D-read’, see paragraph 7.3.7, page 7-5. Memory’ is for factory testing. Also, see paragraph 4.3.5, page 4-3.
11. REMOTE CONTR. (Remote Control) (F4) enables communi­cation with a PC with test automation software. Required soft­ware: PRO-Soft QA-40/45.
1. Startup Screen. The following screen will be displayed for 2 seconds after the QA-45 has been switched on.
2. Main Menu
a. Main Menu Bar (Page 1) - Mode switch in PACE position.
6
b. Second Menu Bar (Page 2)
3. SELECT LOAD (F1)
Choose desired PACER load by pressing UP (F2) or DOWN
(F3) and then SELECT (F4). Press CANCEL (F5) to cancel selection.
4. SELECT NOISE (F2)
Choose desired noise for the immunity test by UP (F2) or
DOWN (F3) and then SELECT (F4). Press CANCEL (F5) to cancel selection.
5. PRINT HEADER (F3). Automatically writes a heading for the new test protocol.
6. PRINT RESULT (F3). Prints the results of measurements.
7. SELECT WAVE (F2)
7

3.5 Test Result Printouts

Choose desired waveform for the sensitivity test by pressing UP
(F2) or DOWN (F3) and then SELECT (F4). Press CANCEL (F5) to cancel selection.
8. SENS. TEST (Sensitivity Test) (F2). Sensitivity is the QRS minimum amplitude (mV) required to cause the pacemaker to operate in the demand mode. This waveform is delayed from the pacer pulse so that it is outside the pacing refractory period. See ‘Sensitivity Measurements’ in Chapter 5.
9. REF. PER TEST (F3). Used to test time interval (ms) if the pacemaker is insensitive to any external inputs, the maximum time interval after the generation of a pacer pulse and maximum time interval after a QRS wave. See ‘Pacing Refractory Period’ and ‘Sensing Refractory Period’ in Chapter 5.
10. REMOTE (Remote Control) (F4) enables communication with a PC with test automation software. Required software: PRO­Soft QA-40/45.
1. Defibrillator Mode. QA-45 automatically prints out the test re­sults, via the printer output, after each discharge generated. Se­lect PRINT HEADER (F4) if you want to print out a page with a new header.
2. Pace Mode. QA-45 prints out the test results, after the measure­ments, when you press PRINT RESULT (F4) in the Main menu.
8
This page intentionally left blank.
9

4.1 Introduction

4. Defibrillator Mode Testing

This chapter describes QA-45 defibrillator mode testing.
The defibrillator function of the QA-45 measures the energy output, and ensures that the defibrillator complies with specified require­ments. QA-45 has a built-in load resistance of 50 ohm, which roughly corresponds to the impedance of the human body. The de­fibrillator pads are placed on the QA-45 contact plates. Thus, the de­fibrillator is connected through the load resistance. When the defib­rillator is discharged, QA-45 will calculate and display the energy delivered.
Defibrillator energy is defined as an integral of the moment of the discharged energy from the defibrillator. The energy is equal to the square of the voltage, divided by the load resistance.
E =  p dt =  V2 / R dt = V2 dt / R
QA-45 measures and records the voltage pulse every 100 µs, 1000
times, for a total time of 100 ms. The squares of the voltages are
then summed, multiplied by 100 µs, and divided by the load resis-
tance, 50 ohms.
1000 1000

4.2 Test Preparation

E =  (V2) dt / R =  (V2) 100 µs / 50 ohms
0 0
The unit for energy is 'joule', which is equal to Ws (Watt second).
1. If checking ECG monitoring, prompting, or triggering from the ECG, connect the low level or high level ECG connectors to the ten 4 mm AHA color-coded safety terminals or standard phone jack, as appropriate.
2. Switch the QA-45 on. The following will be displayed in the LCD display for about two seconds:
3. The following main menu will then appear. It will show LO­CAL, indicating that the testing is not remotely controlled by PRO-Soft QA-40M/45 test automation software.
1
2

4.3 Energy Test

If the maximum voltage for a se­lected range is exceeded, the LCD display will show ‘WARNING! Overload’
APEX (+) pad right plate STERNUM pad left plate
Note
1. Select a suitable energy range using the mode switch.
Use the HIGH range for normal adult testing.
Use the LOW range for low energy testing, where the en-
ergy does not exceed 50 Joule and the peak voltage does not exceed 1200 volts.
2. Securely place the defibrillator paddles on the QA-45 contact plates, and discharge the defibrillator. The APEX (+) pad should be connected to the right-hand plate, and the STERNUM pad to the left plate. This ensures correct signal polarity for the oscilloscope output. A reversal of this configuration will not damage the QA-45, nor will it give incorrect energy readings. However, the polarity of the oscilloscope output will simply be reversed. The discharge from the defibrillator is transferred to the QA-45's load resistance.
3. QA-45 calculates the energy delivered over the load resistance and displays the result in joules under RESULT. See below:
QA-45 also shows the energy measured, the maximum voltage
and the maximum current in the energy wave. Following the discharge from the defibrillator, QA-45 shows a playback of the wave from the ECG output. A new pulse can be generated when the LCD display shows 'LOCAL'.
4. Following a discharge from the defibrillator, the instrument shows a playback of the wave from the ECG output. The dis­play will thus be in playback mode. When this is shown in one line, QA-45 automatically prints out the result.
5. The discharged pulse can be repeated. To do this press more-2 (F5) to advance to page 2 of the main menu.
Press PLAY PULSE (F2). The display will show 'Oper: Play-
back,' and displays the result in joules under RESULT.
3

4.4 Cardioversion Test

Following playback, the apparatus is ready to receive a new dis-
charge from the defibrillator. The display will show 'LOCAL'.
6. When testing automatic defibrillators, it is quite common to have to select 'vfib' from the ECG menu 'ECG WAVE' for the 'ven­tricular fibrillation' wave. Automatic defibrillators typically do not fire without seeing 'v-fib'.
1. Select ECG WAVE (F1) from the main menu.
2. The ECG Wave menu opens. QA-45 includes the following ECG wave selection for cardioversion tests, or for the testing of elec­trocardiograph monitors.
Normal Sine Rates: 30, 60, 80, 120, 180, 240 and 300 BPM. ECG Arrhythmia types as follows:
vfib Ventricular Fibrillation afib Atrial Fibrillation blk II Second degree A-V block RBBB Right Bundle Branch Block PAC Premature Atrial Contraction PVC_E Early PVC PVC_STD PVC PVCRonT R on T PVC mfPVC Multifocal PVC bigeminy Bigeminy run5PVC Bigeminy Run of 5 PVCs vtach Ventricular Tachycardia
4
Select a desired wave by pressing UP (F2) or DOWN (F3).
Save this under ‘Wave” in the STATUS field by pressing SE- LECT (F4). Press CANCEL (F5) to cancel selection.
3. QA-45 includes the following ECG wave amplitude options:
0.5 mV, 1.0 mV, 1.5 mV and 2.0 mV. To change wave ampli­tude select more-2 on the main menu. Select WAVE AMPL. (F1).
5
The Wave Amplitude Menu appears:
APEX (+) pad right plate STERNUM pad left plate

4.5 Maximum Energy Charging Time Test

Select the desired amplitude by pressing UP (F2) or DOWN
(F3). Save this under ‘Ampl” in the STATUS field by pressing SELECT (F4). Press CANCEL (F5) to cancel selection.
4. Set the defibrillator to synchronized cardioversion mode. Dis­charge the defibrillator over the instrument's load resistance.
5. QA-45 measures the time delay in milliseconds (ms) between the top of the 'R' wave and the discharging of the defibrillator pulse. This delay will be shown in the LCD display as: 'Delay: xxx ms'.
QA-45 also shows the energy measured, the maximum voltage
and the maximum current in the energy wave. Following the discharge from the defibrillator, QA-45 shows a playback of the wave from the ECG output. A new pulse can be generated when the LCD display shows 'LOCAL'.
1. The charge time function is used to test the battery and the charging capacitor in the defibrillator.
2. Set the defibrillator to maximum energy.
3. Securely place the defibrillator paddles on the QA-45 contact plates, and discharge the defibrillator. The APEX (+) pad should be connected to the right-hand plate, and the STERNUM pad to the left plate. This ensures correct signal polarity for the oscilloscope output. A reversal of this configuration will not damage the QA-45, nor will it give incorrect energy readings. However, the polarity of the oscilloscope output will simply be reversed. The discharge from the defibrillator is transferred to the QA-45's load resistance.
4. Select CHARGE TIME (F3) from the main menu and the charge button on the defibrillator simultaneously.
6

4.6 Shock Advisory Algorithm Test

When the defibrillator is charged, discharge it through the instru-
ment.
5. Charging time will be shown in the display as ‘Chrg T: xx.x MS’ under RESULT.
1. This tests the analysis and prompting of automatic and semi-au­tomatic defibrillators. A series of arrhythmia is available for analysis by the defibrillator that should then prompt the user to ‘shock’ of ‘no shock,’ in accordance with national and interna­tional guidelines, as shown below:
ASYS No shock SVTa_90 No shock PVT_140 No shock MVT_140 No shock CVF Shock FVF Shock PVT_160 Shock MVT_160 Shock
2. Select ADV. ALG. (F2) from the main menu.
3. The Advisory Algorithms Menu opens.
7
Select the desired rhythm by pressing UP (F2) or DOWN (F3).
Save this under ‘Wave” in the STATUS field by pressing Se- lect. Press CANCEL (F5) to cancel selection. The ECG signal is output through the low-level ECG connectors, high-level ECG connector, and paddle contact plates on the QA-45.
4. Set the defibrillator to analyze the ECG rhythm and operate in the automatic and semi-automatic mode.
5. Records the defibrillator’s response.
8

5. Transcutaneous Pacemaker Mode Testing

5.1 Introduction

5.2 Testing Preparation

This chapter explains QA-45 transcutaneous, or pacer mode testing,
QA-45 tests all types of transthoracic pacemakers. The testing is menu driven, and simple to operate. QA-45 measures and displays a pacer pulse’s amplitude, rate, energy and width. It also conducts de­mand sensitivity tests, measuring and displaying refractory periods, and immunity tests, which determine the pacemaker’s susceptibility to 50/60 Hz interference.
1. Connect the pacer output cables to the pacer input connectors.
2. Switch the mode switch to ‘PACE’ mode.
3. Turn the QA-45 on. The following will be displayed in the LCD display for about two seconds:
4. The following main menu will then appear:
5. Press SELECT LOAD (F1). The following load options will appear:
The load range is 50 to 2300 ohms in steps of 50 ohms up to 200
ohms, and 100 ohms from 200 up to 2300 ohms
1
Select the desired noise form by pressing UP (F2) or DOWN
(F3) and then Select (F4). Press CANCEL (F5) to cancel the selection. After selection the main menu will reappear.
2
6. Select the desired waveform by pressing UP (F2) or DOWN (F3) and then SELECT (F4). Press CANCEL (F5) to cancel the selection. After selection the main menu will reappear.
7. For Immunity Testing Only. The immunity test determines the pacemaker’s susceptibility to 50/60 Hz interference signals. If you desire to test immunity simultaneously with other testing, press SELECT NOISE (F2). The following load options will appear:

5.3 Demand Sensitivity Test

Select the desired noise form by pressing UP (F2) or DOWN
(F3) and then SELECT (F4). Press CANCEL (F5) to cancel
the selection. After selection the main menu will reappear.
1. General. Sensitivity is the minimum QRS amplitude (mV) re­quired to cause the pacemaker to operate in the demand mode. During sensitivity measurement three different waveforms are selectable with widths varying in steps from 10 to 200 ms. This waveform is delayed from the pacer pulse so that it is outside the pacing refractory period. QA-45 then checks whether this wave is sensed or not by the pacemaker.
If it is not sensed, a message 'exceeded' is displayed which
means that the pacemaker needs an amplitude more than 100 mV for sensing at that setting. If the wave is sensed, QA-45 then reduces the amplitude in steps until it reaches the lowest value required for the pacemaker to sense it. (The internal algo­rithm used converges to the lowest value in the least number of cycles.) This lowest value is the sensitivity.
2. Procedure
a. From the main menu press more-2, then SELECT WAVE
(F1).
3
4
b. The following menu will be displayed:
c. Select the desired waveform by pressing UP (F2) or
DOWN (F3) and then Select (F4). Press CANCEL (F5) to cancel the selection. After selection the main menu will reappear.
d. Select SENS. TEST (F2). The following display will ap-
pear:

5.4 Refractory Period Test

e. Upon completion of testing the results will be displayed un-
der RESULT. Press SENS. TEST. CANCEL (F5) to can­cel the test.
1. General. This test is used to test the time interval in millisec­onds (ms) during which the pacemaker is insensitive to any ex­ternal inputs. QA-45 does this by measuring the maximum time interval after the generation of a pacer pulse, and maximum time interval after a QRS wave.
a. Refractory Period. A time interval in milliseconds, during
which a pacemaker is insensitive to any external inputs. If a QRS is detected during this period, the pacemaker ignores it. On the other hand, if a QRS is detected outside the re­fractory interval, then the pacemaker resets its internal
5
timer and the next pacer pulse is generated after a delay of one time period from this QRS wave.
b. Paced Refractory Period. The maximum time interval af-
ter the generation of a pacer pulse during which time the presence of a QRS wave is ignored.
The measurement of paced refractory period takes a few cycles of the pacemaker output. First, QA-45 measures the pacer-to-pacer interval T. Then, it puts out a square wave 40 milliseconds wide, delayed by delay time D, which is more than the pacing refractory period, from the last pacer pulse. The pacemaker senses this square wave. The delay time D is gradually decremented in subsequent cycles until the square waveform is not sensed by the pacemaker. The maximum value of the delay time D, for which the pace maker does not sense the square wave, is the paced refrac­tory period.
c. Sensed Refractory Period. The maximum time interval
after a QRS wave is sensed by the pacemaker during which time the presence of a second QRS wave is ignored.
The sensed refractory period is measured in a similar man­ner, except that QA-45 now generates two square waves in­stead of one. The first square wave is generated at a fixed time delay from a pacer pulse, which is greater than the paced refractory period. The pacemaker always senses this square wave.
The second square wave is generated at a delay D from the first square wave. The initial value of D is selected to be greater than the sensed refractory period. Therefore the first time the pacemaker is on it also senses the second square wave. In subsequent cycles, the delay 'D' is gradually re­duced until the pacemaker is unable to sense the second square wave. The maximum value of D, for which the pacemaker does not sense the second square wave, is the sensed refractory period.
2. Procedure
a. From the main menu press more-2. Press REF. PER.
TEST (F3).
b. The following display will appear while testing:
6
c. Upon completion of testing the results will be displayed un-
der RESULT. Press REF. PER. CANCEL (F5) to cancel the test.
7
This page intentionally left blank.
8

6.1 Required Test Equipment

6.2 Preparation

6. Control and Calibration

This chapter explains the QA-45 maintenance procedures, including testing and calibration.
Digital multimeter, 10 uV resolution, 0.1% accuracy.
Frequency counter
Oscilloscope
Variable VIA power supply
10 V (+0.01 v) power source
Pulse generator: square pulse, 10 ms width, 10V amplitude, 80
pulses per. minutes (for pacer module).
Set the switches on QA-45 to the following positions:
Mode: Low Power: Off
Connect a power supply to the battery eliminator input on QA-45.
Adjust the power supply to 9V (±0.2v) with a power limitation of 200 mA (±50 mA).

6.3 References

6.4 Test

The function keys are numbered 1 to 5, with switch 1 farthest to left.
1. Set power to ON. Wait 3 seconds, and measure the current from
the power supply. Requirement: 68 mA (±5 mA).
2. Adjust P2 on the processor board to obtain the best possible con­trast on the display. The display should show the main menu and the result with 0-data. Press function switch 5 and check that QA-45 changes between various menus. Adjust P103 out of limit.
3. Set power to OFF, and then back to ON after about 1 second. The display should show the software version number for a brief period, before showing the main active display. (By test under production a additional item shall be used.)
4. Measure the operating voltages in QA-45 with the multimeter. The following values are acceptable:
Test point
- +
Level Maximum
Deviation
1
TP8 - TP6 +8.8V TP8 - TP4 -8.0V TP8 - TP2 +2.5V TP8 - TP5 -8V +2 -1V
TP8 - TP3 +5V
±0.2V ±0.3V ±0.075V
±0.1V
5. Connect the frequency counter to TP7 and read the frequency.
Requirement: 2 MHz (±0.002 MHz).
6. Slowly reduce the voltage from the power supply until the in­stant the display gives the message: 'Change battery, and reset system'. Measure the operating voltage with the multimeter.
Requirement: 6.9V (±0.3v). Return the operating voltage to
9.0V. and reset QA-45 by switching off the power for a short period.
7. Select 80 BPM in the ECG Wave Menu. Connect the oscillo­scope to TP1 for signal, and TP8 to ground. Check that QA-45 generates a 80 BPM signal with an amplitude of approximately 250 mV on the R pulse. Connect the oscilloscope to the High Level ECG contact and check that the same signal is present, only with an amplitude of approximately 1V on the R pulse.
8. Set power to OFF. Measure the resistance from the RL output to the RA, LA and LL outputs.
Requirement: 1000 ohm (±30 ohm).
9. Measure the resistance from the RL output to the V2, V3, V4, V5 and V6 outputs.
Requirement: 1000 ohm (±30 ohm).
10. Set power to ON and go into the SYSTEM TEST (F1) of the page 3 menu and choose ECG + by moving the cursor and pressing SELECT (F4). Measure and adjust the voltage be-
tween TP1 and TP8 to 0.490V ±1 mV. Choose ECG 0 from
menu. Measure the voltage between TP1 and TP8 and check that it is between +20 mV and -10 mV.
11. Choose ECG + from the System Test Menu, measure the volt­ages between the ECG outputs, and check that they fall within the limits shown in the following table:
Contact Power limit Nom. Value Upper limit
RL - RA 1.20 mV 1.35 mV 1.50 mV RL - LA 2.40 mV 2.65 mV 2.90 mV RL - LL 3.00 mV 3.35 mV 3.70 mV RL - V1 2.70 mV 3.02 mV 3.30 mV RL - V2 3.30 mV 3.72 mV 4.10 mV RL -V3 4.00 mV 4.49 mV 5.00 mV RL - V4 4.50 mV 5.06 mV 5.60 mV RL- V5 4.00 mV 4.49 mV 5.00 mV RL - V6 3.30 mV 3.72 mV 4.10 mV
12. Measure the voltage at the High ECG output.
Requirement: 2.0V (±0.05v).
2
13. Measure and check that the voltage between the defibrillator
pads is 2 mV ±50 µV. Choose ECG - by moving the cursor and
pressing SELECT (F4). Measure and check that the voltage be-
tween the defibrillator pads is -2 mV ±50 µV.
14. Set the power to Off, and wait 10 seconds. Measure and note the exact value of the resistance between the defibrillator pads.
Requirement: 50 ohm (±0.5 ohm).
Set the power switch to ON while simultaneously holding down
function key 1, until the main menu appears. The main menu will now be quickly replaced by the menu for calibrating resis­tance. Adjust two measurement values by pressing + and -. The values will be stored in QA-45's EEPROM when SAVE & QUIT is pressed.
15. Set the power switch to Off. Remove the covers at J106 and J107. Measure the resistance between TP101 and TP102.
Requirement: 2 MOhm (±2 kOhm).
Repeat the measurement for TP103 and TP104.
16. Replace the covers at J106 and J107. Check that the Mode switch is set to Low. Measure the resistance between TP102 and TP105.
Requirement: 2 kOhm (±2 ohm).
Measure the resistance between TP108 and TP100.
Requirement: 2 kOhm (±2 ohm).
17. Set the Mode switch to High. Measure the resistance between TP102 and TP106.
Requirement 10 kOhm (±10 ohm).
Measure the resistance between TP109 and TP100.
Requirement 10 kOhm (±10 ohm).
18. Measure the resistance between TP102 and TP113.
Requirement: 200 kOhm (±200 ohm).
Measure the resistance between TPl14 and TP100.
Requirement: 200 kOhm (±200 ohm).
3
19. Measure the resistance between TP102 and TP107.
Requirement 10 kOhm (±10 ohm).
Measure the resistance between TP104 and TP110.
Requirement 10 kOhm (±10 ohm).
20. Set power switch to ON. Connect a frequency counter to pin 3 on IC106 and measure the frequency.
Requirement: 1.9 MHz (±200 kHz).
21. Measure the voltage between TPl12 and TP100. Adjust P104
until the voltage is 5V (±0.0005v).
22. Go to the System Test Menu and choose A/D-read by moving the cursor and pressing SELECT (F4). The input voltage at the A/D converter should be displayed on the screen. The value is updated once every second. Adjust P103 until the value is as close to 0 as possible.
23. Connect a 10V (±0.01v) power source between TP107 and
TP110. Check that the Mode switch is set to High. Adjust P102 until the displayed value is 1997.5 mV and 2000 mV. Change the polarity on the power source and re-measure the voltage. Adjust P102 and P103 until the value displayed is between
1997.5 and 2000 mV regardless of the polarity status. Check
that the scope output is +2000 mV (±20 mV) or -2000 mV (±20
mV). Remove the power source and secure P102, P103 and P104. Switch Off the power.
24. Connect a printer to the printer port. Switch on the power. heck that the Mode switch is set to High. Go to the System Test Menu and activate TEST PULSE (F1). Check that QA-45 gives a correct printout, that the energy measurement on the
display and printer is 125 Joules (±20 %), and that the measured
voltage is approx. 2500 V. Set the Mode switch to Low and ac­tivate TEST PULSE (F1). Check that the measured energy
value is approx. 5 Joules (±20 %), and that the measured volt-
age is approx. 500V.
25. Connect a PC to the serial port and try to control QA-45 re­motely. Check that communication functions in both directions.
26. Go to the System Test Menu and choose ECG +. (Last setting for pacer should be default setting: 500 ohm). Measure the volt­ages In pacer module with the multimeter. The following values are acceptable:
Test point
- +
TP1 - TP5 -2.378V TP1 - TP6 +2.378V
Level Maximum
Deviation
±23 mV ±23 mV
4
27. Connect a 10V (±0.01v) power source to the pacer input. (Last
setting for pacer should be default setting: 500 ohm). Measure the voltages on pacer module with the multimeter. The follow­ing values are acceptable:
Test point
- +
TP1 - TP2 +61.04 mV TP1 - TP3 +244.1 mV TP1 - TP4 +0.9766 V
Level Maximum
Deviation
±625 µV ±2.5 mV ±10 mV
28. Go out of the System Test Menu and choose pacer program with the slide switch. Go to the SELECT LOAD (F1) and choose 50 ohm. Connect a ohm meter to the pacer input and measure the resistance for all load settings. The following val­ues are acceptable:
Nominal Min. value Max. value
50 ohm 49.5 ohm 50.5 ohm 100 ohm 99 ohm 101 ohm 150 ohm 148.5 ohm 151.5 ohm 200 ohm 198 ohm 202 ohm 300 ohm 297 ohm 303 ohm 400 ohm 396 ohm 404 ohm 500 ohm 495 ohm 505 ohm 600 ohm 594 ohm 606 ohm 700 ohm 693 ohm 707 ohm 800 ohm 792 ohm 808 ohm 900 ohm 892 ohm 909 ohm
1000 ohm 990 ohm 1010 ohm 1100 ohm 1089 ohm 1111 ohm 1200 ohm 1188 ohm 1212 ohm 1300 ohm 1287 ohm 1313 ohm 1400 ohm 1379 ohm 1421 ohm 1500 ohm 1478 ohm 1522 ohm 1600 ohm 1576 ohm 1624 ohm 1700 ohm 1675 ohm 1725 ohm 1800 ohm .1773 ohm 1827 ohm 1900 ohm 1872 ohm 1928 ohm 2000 ohm 1970 ohm 2030 ohm 2100 ohm 2069 ohm 2131 ohm 2200 ohm 2167 ohm 2233 ohm 2300 ohm 2266 ohm 2335 ohm
Open
(43.78 ohm)
43.649 ohm 43.911 ohm
29. Set the pacer load to 50 ohm. Connect a pulse generator to the pacer input. Check that the measured values: (If the pulse gen­erator not handle 10V amplitude in 50 ohm, you must scale the values).
Parameter Value Max. Deviation
Rate: 80 ppm Width: 10 ms Amplitude: 200 mA Energy: 20.0 mJ
±0.5 ppm ±0.3 ms ±1.5 mA ±2.0 mJ
5
30. Set the pacer load to 500 ohm. Connect a pulse generator to the
pacer input. The following values are acceptable:
Parameter Value Max. Deviation
Rate: 80 ppm Width: 10 ms Amplitude: 20 mA Energy: 2.0 mJ
±0.5 ppm ±0.3 ms ±1 mA ±0.5 mJ
31. Set the pacer load to 1000 ohm. Connect a pulse generator to the pacer input. The following values are acceptable:
Parameter Value Max. Deviation
Rate: 80 ppm Width: 10 ms Amplitude: 10 mA Energy: 1.0 mJ
±0.5 ppm ±0.3 ms ±1 mA ±0.5 mJ
The test is complete!
6

7.1 Theory of Operation

7. Component Functions and Parts

This chapter provides a detailed description of the functions of the main components of the QA-45, as well as a parts list for cross-ref­erence.
The QA-45 is based on 4 circuit boards: a processor board, sensor board, ECG signal distribution board and pacer module. The boards are described in 12 circuit diagrams which are located in Appendix A. Diagrams 1 through 4 describe the processor board; Diagrams 5 through 8 describe the sensor board; Diagrams 9 and 10 describe the ECG distribution board, and Diagrams 11 and 12 describe the pacer module.
The QA-45 Defibrillator/Transcutaneous Pacer Analyzer is a battery powered unit, based on a microprocessor and analogue electronics with precision data acquisition circuits. Using controls on the front panel of the unit, one can analyze defibrillator and pacer pulses, when these are discharged through built-in load resistors. The unit can also generate a number of different ECG, test and stimulus sig­nals. The measurement results appear on an LCD display, and can be output to a printer. A serial port (RS-232C) makes remote control from a PC possible. All measurement and control signals are con­nected via contacts on the top and rear of the unit. To operate the unit, there are 5 soft keys linked with menus on the LCD display.

7.2 Processor Board

(Refer to QA-45 Processor Board Component Location Diagram and Schematic Diagrams 1-3)
1. Printer Output (Schematic Diagram 1). QA-45 has a printer output with a standard 25-pin D-sub contact for Centronics in­terface. The output is based on 3 HCMOS circuits: IC9, IC10
and IC11. All the circuits are connected to I/O-ports on the mi­croprocessor. ICll is a latch for the 8 parallel data lines. IC9 is the driver for outgoing control signals, while IC10 is a buffer for incoming control signals. RP1 contains pull-up resistors for the input lines. All connections to the printer contact are filtered to reduce high frequency radiation.
2. Serial Port (Schematic Diagram 1). The serial port is suitable for 9-pin RS-232C format. The QA-45 is configured as DTE (data terminal equipment) and should be connected to, for ex­ample, a PC with a null-modem cable. IC6 is the driver for the data signals. All handshakes take place via software. The con­trol signals are routed back to the contact.
3. Curve Generator (Schematic Diagram 1).
1
ECG curves are generated by the microprocessor based on data
tables. The processor updates the 8-bit D/A-converter in IC1 (channel A), usually 500 times per second. IC-3 converts cur­rent to voltage, and IC-2 makes the signal bipolar. The ampli­tude is adjusted using P1, and the zero point is adjusted using P4. From pin 1 on IC2 (TP-1) the stimulus signal is led via ca­bles to resistive voltage dividers on the ECG signal distribution card.
IC1 D/A-converter part B sets the reference level for the curve
generator D/A-converter part A. Part B thus determines the am­plitude of the stimulus signal. The DC value for maximum am­plitude is read back by the microprocessor's 10-bit A/D-input El, and can therefore be adjusted precisely.
The other half of IC2 is used for operating the high-level output
for the ECG signal. The circuit has an amplification of 4 times. The signal is filtered for high frequency noise with F18 before it is conducted to contact J8. C3 ensures bandwidth limitation in IC2. D9 is over voltage protection.
4. Power Supply (Schematic Diagram 2).
The unit is powered either from 2 internal 9V batteries or from
an external 9V DC power supply. The batteries are connected in parallel via J6 and J10, while the external power supply can be connected to J7. D1 protects against incorrect polarization. R16 is a PTC fuse that provides high impedance if too much power is drawn. F19 provides high-frequency filtering and protects against voltage transients from the power supply. The supply voltage is conducted to the sensor board through the power switch and back. F23 and F24 ensure reduction of high-fre­quency radiation from the power supply cables.
After the power switch SW108, the supply voltage is conducted
to the power supply circuits. The supply voltage is also used di­rectly as internal +9V. IC5 is a capacitive switch regulator that regulates -9V from the +9V. These voltages will alternate, and may vary from 7V to 12V.
IC4 is a series regulator that supplies the circuits with +5V. The
circuit has an output that resets the micro controller when the +5V voltage falls below 4.75V. D6 and C13 provide the printer output with its own +5V. D6 will block power inflow through the printer cable when the QA-45 is switched off.
D7 and D8 rectify an AC signal from pin 2 IC-5, and build up a
voltage of -18V via C29. From this voltage, IC17 generates bias voltage to the LCD display. The bias voltage is about -7V, and is adjusted using P2 to set the contrast of the display.
5. Microprocessor (Schematic Diagram 3).
The microprocessor IC13 contains the CPU, RAM, A/D-con-
verter, parallel I/O and serial I/O. Y1 functions as a clock and
2

7.3 Sensor Board

time reference for the processor. The frequency on TP7 is the crystal frequency/4 (2 MHz).
U9 is an EPROM that stores the processor's program. The circuit
can store 128 Kbytes of program data. IC16 is a RAM circuit used for storage of measurement data. Together with the LCD display, these circuits are connected directly to the micropro­cessor's bus. IC15 is a PAL that sends chip-select signals to the EPROM, RAM and LCD display. The PAL circuit is in the base, and can be replaced if the program is updated. IC7 is a se­rial EEPROM that stores calibration data. Data is transferred via the processor's I/O ports.
The A/D converter in the processor is used to monitor the battery
voltage. The voltage is divided by R12 and R13, and fed into the A/D converter via port E0. The Vref. voltage from IC12
(Diagram 1) is used as the reference for the A/D (VRH).
(Refer to QA-45 Sensor Board Component Location Diagram and Schematic Diagrams 4-6)
1. Serial Interface (Schematic Diagram 4). Signal transfer be­tween the processor board and the sensor board is in digital se­ries form. IC108 is a serial-to-parallel converter that sends data to the signal M UX ICl14. IC112 reads parallel data from the switches used for operation. The data format is adapted to the microprocessor's SPI interface, and clocked by the SCK clock from the processor. The A/D converter IC104 (Diagram 3) has its own serial interface connected to the SPI lines.
2. Switches For Operation (Schematic Diagram 4).
The front panel on the QA-45 has 5 push buttons and 2 sliding
controls for operation. The push buttons SW101 - SW105 are soft keys to which different functions are assigned by the soft­ware. The switches are read from the processor via ICl12.
SW108 is an On/Off switch that breaks the battery voltage (see
Diagram 2). SW106 (Diagram 6) sets the input attenuation by selecting different resistor values. The switch also has a digital connection (HI_LO and PACED signal) so that the micropro­cessor can read the switch position.
3. Stimulus Output (Schematic Diagram 4). IClll amplifies the stimulus signal from the processor board, and sends this differ­entially to the pads. R141 is connected to the range selector SW106, which provides different amplification for pacer stimuli and defibrillation stimuli.
4. Scope Output (Schematic Diagram 5). Scope output is fed with a signal from IC114, which is an analogue MUX. The signal can be selected digitally from 4 sources; the scope signal from the input amplifier for the defibrillator pads, or 3 versions (lev-
3
els) of the input signal from the pacer module. IC110 buffers the signal, which is fed to J109 via R129.
5. Internal Power Supply Sensor Board (Schematic Diagram 5). IC107 regulates the -9 volt potential down to stable -5V for use in the A/D converter system. Dl14 prevents incorrect polariza­tion and latch-up when the QA-45 is switched on and off. D108, D109 and Dl10 provide over-voltage protection.
6. Measurement Input (Schematic Diagram 6).
R185 and R186 together make up a 50-ohm load resistance for
the discharges from the defibrillator being tested. The voltage over the resistors is conducted via IC101 and IC102 into the A/
D converter IC104. This samples the voltage every 100 µs, and
the measurement values are transferred digitally to the micro­processor.
The measurement input is based on an attenuator that consists
primarily of a differentially connected operational amplifier (IC101). Using SW109, the attenuation can be set as 10, 200 or 1000 times. The output signal from IC101 may fluctuate from +5 to -5 Volt. With an input attenuation of 1000, the measure­ment range becomes +5000 to -5000 Volt (High Range). With an input attenuation of 200, the measurement range becomes +1000 to -1000 Volt (Low Range). With an input attenuation of 10, the measurement range becomes +50 to -50V (pacer mode).
IC102 is used as an addition amplifier. Here, the measurement
signal is added to an offset signal generated by the other half of IC102, set using P103. A test signal can also be added in to test the A/D converter function. P102 adjusts the input attenuation.
7.4 ECG Signal Distribu­tion Board
7. A/D Converter (Schematic Diagram 6). The A/D converter sys­tem consists of IC104, IC105 and, IC106. IC104 is the A/D con­verter itself, while IC105 is the voltage reference, and IC106 is the clock generator.
The analogue signal is fed into the A/D converter on pin 7, and
the measurement result is read serially on pins 15 to 18. The A/ D has a differential input, and can function with 12- or 13-bit resolution. IC106 is a square wave oscillator that generates the clock signal to the analogue part of the A/D converter. The fre­quency of the signal is nominally 1.9 MHz, but may vary be­tween 1.7 and 2.1 MHz.
IC105 is the voltage reference for the A/D converter. P104 can
be used for precise adjustment of the voltage to exactly 5 Volt (TPl12).
(Refer to QA-45 ECG Distribution Board Component Location Dia­gram and Schematic Diagram 7)
4

7.5 Pacer Unit

Here, the stimulus signal (ECG-signal) is fed through 10 attenuators that set up the correct level for the outputs. The output impedance to earth is 500 ohm for all outputs. Filters on the outputs ensure damp­ing of high-frequency components in the signal.
(Refer to QA-45 ECG Pacer Unit Component Location Diagram and Schematic Diagram 8)
The pacer module consists of 3 blocks. These are the resistor group, stimulus output and input amplifier.
1. Resistor Group. The value of the load resistance for the pacer input is set using 6 relays, which are connected over a range of resistors. The relays are controlled by U4, which in turn is con­trolled from the microprocessor via a serial interface. The relay configuration is determined in the software, and the relays are operated in such a way that the load resistance can be varied from 50 to 2300 ohm, with 50/100 ohm resolution. The relays are of the side stable (latching) type, and will therefore use no power from the batteries except when they are operated. Before the relays are set, they are all reset via pin 18 on U4.
2. Stimulus Output. The stimulus signal that is generated on the processor board is fed to U1, which, together with U2, makes up a programmable gain block. The signal is amplified with U3, and fed differentially over the load resistance via R17 and R18. As the load resistance may be varied, and R17 and R18 are con­stant, the amplitude out from U3 must be adapted to the load setting. This is done using the gain block (U1 and u2), which is controlled from the microprocessor via the same serial interface as U4.
3. Input Amplifier. The input signal is measured over the load re­sistor with the differential amplifier U5. The measurement sig­nal is fed to the A/D converter's MUX from the output on U5. U6 contains 2 amplifiers that each have a gain of 4 times (12 dB). The outputs from these amplifiers are also added to the A/ D converter MUX, so that one can choose between 3 different dynamic measurement ranges. The measurement ranges are se­lected according to the load setting.
5

7.6 Component Parts

COMPONENT PART TYPE/VALUE QTY. DIAGRAM REFERENCE
PROCESSOR BOARD
Processor board AR-095 1 Microprocessor MC68HC11G5FN 1 IC13 RAM LH5160HD-10L 1 IC16 EPROM 27C010-120 1 IC14 EEPROM X24C16P 1 IC7 PAL PAL. 22V10Z-25PC 1 IC15 D/A converter MX7528KN 1 IC1 Op.amp LM358P 1 IC17 Op.amp LT1013DN8 2 IC2, IC3 V-ref. LM-385Z-2V5 2 IC12, IC18 Timer LMC555CN 1 IC106 Port 74HC05N 1 IC9 Bus driver 74HC541N 1 IC10 Latch 74HC573N 1 ICll Volt. Regulator LP2951CN 1 IC4 Volt. Converter LT1054CN8 1 IC5 IC base 20 pin DIL 1 IC1 IC base 20 pin SIL 1 IC14 IC base 24 pin 300mil 1 IC15 IC base 84 pin PLCC 1 IC13 LCD display DMF5005N 1 Diode 1N4148 2 D2,D5 Schottky diode 1N5819 2 D4, D6 Schottky diode BAT54 2 D7, D8 Zener diode 13V 1 D1 Transzorbdiode BZW06-6V4B 1 D3 Transzorbdiode BZW06-17B 3 D9 - Dll Crystal 8 MHz 1 Y1 Resistor 1K0 1% 0.5W 1 R8 Resistor 2K2 1% 0.5W 1 R9 Resistor 3K3 1% 0.5W 1 R13 Resistor 10K 1% 0.5W 3 R6, R10, Rll Resistor 18K 1% 0.5W 1 R19 Resistor 22K 1% 0.5W 1 R12 Resistor 30K 1% 0.5W 1 R7 Resistor 40K 1% 0.5W 1 R1 Resistor 47K 1% 0.5W 3 R5, R14, R18 Resistor 80K 1% 0.5W 2 R3, R4 Resistor 10M 1% 0.5W 1 R2 Resistor OR 1 L4 PTC resistor MF-R030 Boums 1 R16 Potentiometer 50K 1-turn 1 P2 Potentiometer 10K 20 turn 1 P1 Resistor pack 5 x 2K2 1 RP3 Resistor pack 8 x 4K7 1 RP4 Resistor pack 8 x 10K 2 RP1, RP2 Cer. capacitor 22pF 50V 2 C1, C2 Multilay. X7R lnF 50V 1 C3 Multilay. X7R 4n7 50V 1 C4 Multilay. X7R 100nF 50V 9 C6 - Cll, C15, C16, C19 Electrolyte Capacitor
Electrolyte Capacitor Electrolyte Capacitor RF coil 1.2mH 2 L2, L3
10µF 16V 100µF 16V 10µF 30V
1 C13 6 C18, C23 - C27 1 C28 - C30
6
COMPONENT PART TYPE/VALUE QTY. DIAGRAM REFERENCE
EMI Filter DSS306-91Y 28 F1-F18, F21, F22, F25 - F31 EMI Filter DSS710-D22 1 F19 EMI Filter DSS306-91F 9 F20, F23, F24, F32 - F34, F37 - F38 Battery contact FemaleS-G9312#0 1 J7 Header 2 pole 2 J6, J10 Phono-base CLIFF PHS2A 1 J8 D-sub 9p Male 90 degrees 1 J2 D-sub 28p Female 90 de-
grees Flat cable terminal 26P 2 J35, Jll Flat cable plug 16P 1 Flat cable plug 20P 1 Box header 20P 1 J9 Connector 36-pol 1 TP101 -TP9 Flat cable 16 leads 280 mm Flat cable 25 leads 200 mm
SENSOR BOARD
Sensor board AR-082 1 Volt. regulator MAX664CPA 1 IC107 A/D converter LTC1290CCN 1 IC104 Op.amp TL1413CN8 3 IC102, ICl10, IClll Op.amp LT1012ACN8 1 IC101 V-ref. LT1027CCN8-5 1 IC105 Timer LMC555CN 1 IC106 Shift register 74HC589AN 1 ICl12 Shift register 74HC595AN 1 IC108 NAND port 74HC00AN 1 ICl13 Multiplexer 74HC4052P 1 ICl14 Diode 1N4148 3 Dlll - Dl13 Zener diode 5V1 4 D104 - D107 Schottky diode 1N5819 1 Dl14 Transzorb diode BZW06-10B 2 D101, D102 Transzorb diode BZW06-17B 3 D108 - Dl10 Resistor 47R 1% 0.5W 1 R184 Resistor 100R 1% 0.5W 1 R192 Resistor 499R 1% 0.5W 3 R131, R138, R139 Resistor 1K0 1% 0.5W 1 R129 Resistor 2K0 0.1% 0.5W 2 R121, R124 Resistor 3K16 0.1% 0.5W 1 R140 Resistor 3K3 1% 0.5W 1 R132 Resistor 3K74 1% 0.5W 1 R130 Resistor 3K9 0.1% 0.5W 16 R1245-R151, R153 - R159, R180,
Resistor 4K7 0.1% 0.5W 1 R141 Resistor 5K36 1% 0.5W 2 R188, R193 Resistor 8K06 1% 0.5W 1 R109, Rl18 Resistor 10K 0.1% 0.5W 4 R122, R123, R125, R126 Resistor 10K 1% 0.5W 5 R128,R136,R137,R143,R190 Resistor 20K 0.1% 0.5W 3 R127, R142, R144 Resistor 20K 1% 0.5W 1 R135 Resistor 40K2 1% 0.5W 1 R134 Resistor 90K9 1% 0.5W 1 R191 Resistor 180K 1% 0.5W 2 R133, R189 Resistor 200K 0.1% 0.5W 2 R120, R187 Resistor 249K 0.1% 0.5W 16 R101 - R108, Rl10 - Rl17 Resistor 10M 1% 0.5W 3 R194 - R196 Resistor 0R 2 L102, L103 Resistor N.C. 1 Rl19 Power resistor 25R 1% 2 R185, R186 Potentiometer 1K 20-turn 1 P102
1 J1
R181
7
COMPONENT PART TYPE/VALUE QTY. DIAGRAM REFERENCE
Potentiometer 10K 20 turn 2 P103, P104 Resistor pack 8 x 10K 1 RP101 Cer. capacitor 220pF 50V 1 C102 Cer. capacitor 470pF 50V 1 C101 Multilay. X7R 10nF 50V 1 Cl13 Multilay. X7R 100nF 50V 11 C104-Cl12, Cl16, C120 Multilay. X7R 470nF 50V 1 C114 Electrolyte Capacitor
Tantal RF coil 1.2mH 1 L101
EMI Filter DSS306-91Y 1 F101 Sliding switch 4P3T 1 SW106 Sliding switch 4PDT 1 SW108 Pressure switch 15.501 5 SW101 - SW105 Switch top 16.300.09 5 KK contact 2 pole pin 2 J102, J109 KK contact 2 pole house 3 Pin for KK 6 BNC base 1 Solder washer for BNC base 1 Cable 120 mm Cable 230 mm Base 16 pin 1 J101 Spacer M3x5 nylon 4 Screw M3x12 nylon 4 Nut M3 nylon 4
ECG SIGNAL DISTR. BOARD
ECG signal distr. circuit board AR-094 1 Resistor 499R 1% 0.5W 10 R2, R3, R5, R7, R9, Rll, R13, R15,
Resistor 47K5 1% 0.5W 1 R14 Resistor 53K6 1% 0.5W 2 R12, R16 Resistor 64K9 1% 0.5W 2 R10, R18 Resistor 72K3 1% 0.5W 1 R6 Resistor 80K6 1% 0.5W 1 R8 Resistor 90K9 1% 0.5W 1 R4 Resistor 180K 1% 0.5W 1 R1 N.C. D1 EMI Filter 2n2 DSS306 10 F1 - F10 KK contact 1 J1 KK contact house 2 J1 Pin for KK 4 J1 Connector 36-pol 2 TP1, TP2 Cable 1L+Sk 120 mm
100µF 25V
10µF 25V
1 C103 4 C115, Cl17-C119
R17, R19
PACER UNIT
Pacer circuit board PCB1 AR-081 1 D/A converter DAC8043FP 1 U1 Op.amp LT1413CN8 2 U2, U3 Op.amp LT1012ACN8 1 U5 Op.amp LT1112CN8 1 U6 Relay driver UCN5842A 1 U4 Resistor 7R5 1% 0.5W 1 R32 Resistor 49R9 0.1% 0.5W 2 Rll, R7 Resistor 100R 0.1% 0.5W 1 R2 Resistor 420R 0.1% 0.5W 2 R3, R30 Resistor 820R 0.1% 0.5W 3 R4, R5, R31 Resistor 820R 1% 0.5W 3 R22, R33, R34 Resistor 1690R 0.1% 0.5W 1 R6
8
COMPONENT PART TYPE/VALUE QTY. DIAGRAM REFERENCE
Resistor 1K52 0.1% 0.5W 2 R21, R22 Resistor 1K8 1% 0.5W 1 R12 Resistor 2K 0.1% 0.5W 1 Rll Resistor 10K 0.1% 0.5W 4 R9, R13, R,?.5, R28 Resistor 10K 1% 0.5W 3 R15, R24, R27 Resistor 20K 0.1% 0.5W 4 R8, R10, R14, R16 Resistor 24K 0.1% 0.5W 2 R17, R18 Resistor 30K 0.1% 0.5W 2 R26, R29 Resistor 249K 0.1% 0.5W 2 R19, R20 Multilay. X7R 100nF 50V 8 C1 - C5, C7, C8, C10 Electrolyte Capacitor
N.C. C6 RF coil 1.2mH 1 L1 Relay -B201 4 K1 - K4 Relay -B101 2 K5, K6 Connector 2x36-pol 1 J1 KK contact 4-pol 1 J4 Connector 36-pol 1 TP1 -TP6
100µF 25V
1 C9
9
This page intentionally left blank.
10
Appendix A - Diagrams
Processor Board Component Location Diagram............................................................................A-2
Schematic Diagram Part 1 (Processor Board)................................................................................A-3
Schematic Diagram Part 2 (Processor Board)...............................................................................A-4
Schematic Diagram Part 3 (Processor Board)................................................................................A-5
Sensor Board Component Location Diagram................................................................................A-6
Schematic Diagram Part 4 (Sensor Board).....................................................................................A-7
Schematic Diagram Part 5 (Sensor Board)....................................................................................A-8
Schematic Diagram Part 6 (Sensor Board)....................................................................................A-9
ECG Signal Distribution Board Component Location Diagram.................................................A-10
Schematic Diagram Part 7 (ECG Signal Distribution Board).....................................................A-11
Pacer Unit Component Location Diagram..................................................................................A-12
Schematic Diagram Part 8 (Pacer Unit)......................................................................................A-13
A-1
A-2
A-3
A-4
A-5
A-6
A-7
A-8
A-9
A-10
A-11
A-12
A-13
USA
1345 Monroe NW, Suite 255A Grand Rapids, MI 49505 Phone: (+1) 888 863-8766 Fax: (+1) 616 454-3350 E-mail: support.us@metron-biomed.com
From: (name)       Phone:       Address:       Fax:      
      E-mail:             Date:      
FRANCE
30, rue Paul Claudel 91000 Evry, France Phone: (+33) 1 6078 8899 Fax: (+33) 1 6078 6839 E-mail: info@metron.fr
NORWAY
Travbaneveien 1 N-7044 Trondheim, Norway Phone: (+47) 7382 8500 Fax: (+47) 7391 7009 E-mail: support@metron.no
Error Report
Product:       Version:       Serial no.:
Description of the situation prior to the error:      
Description of the error:        
(METRON AS internally) Comments:
Received date:       Correction date:       Ref No.       Critical Normal Minor
USA
1345 Monroe NW, Suite 255A Grand Rapids, MI 49505 Phone: (+1) 888 863-8766 Fax: (+1) 616 454-3350 E-mail: support.us@metron-biomed.com
FRANCE
30, rue Paul Claudel 91000 Evry, France Phone: (+33) 1 6078 8899 Fax: (+33) 1 6078 6839 E-mail: info@metron.fr
NORWAY
Travbaneveien 1 N-7044 Trondheim, Norway Phone: (+47) 7382 8500 Fax: (+47) 7391 7009 E-mail: support@metron.no
B-1
From: (name) Address:       Fax:      
      Phone:      
      E-mail:             Date:      
Improvement Suggestion
Product:       Version:      
Description of the suggested improvement:      
(METRON AS internally) Comments:
Received date:       Correction date:       Ref No.       Critical Normal Minor
C-2
.
USA
1345 Monroe NW, Suite 255A Grand Rapids, MI 49505 Phone: (+1) 888 863-8766 Fax: (+1) 616 454-3350 E-mail: support.us@metron-biomed.com
.
FRANCE
30, rue Paul Claudel 91000 Evry, France Phone: (+33) 1 6078 8899 Fax: (+33) 1 6078 6839 E-mail: info@metron.fr
NORWAY
Travbaneveien 1 N-7044 Trondheim, Norway Phone: (+47) 7382 8500 Fax: (+47) 7391 7009 E-mail: support@metron.no
Loading...