Metron QA-45, QA-40M User Manual

P R O - S o f t Q A - 4 0 M / 4 5
U s e r M a n u a l
D e f i b r i l l a t o r - / T r a n s c u t a n e o u s
P a c e m a k e r A n a l y z e r
1
Copyright 1997 by METRON. All rights reserved.
METRON:
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1 Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500 Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009 E-mail: metronus@aol.com E-mail: metronfrance@infonie.fr E-mail: support@metron.no
License
By using the enclosed program and/or installing the files from the original disks to your hard drive, you are agreeing to become bound by the terms of this agreement. If you do not agree to the terms of this agreement, return all materials in this package to METRON U.S. within 30 days for a refund.
You have purchased a software license granting you the right to use the software contained herein. METRON U.S. retains ownership of this software on the original disks, as well as any subsequent copies on any media or in any form. METRON U.S. does not sell any rights to the software. METRON U.S. grants you the right to use the software on the amount of computers for which you have purchased a License. If you wish to use this software for more computers than you have Licenses, you must purchase additional Licenses from METRON U.S. Your License to the program contained in the software grants you, the purchaser, permission to use the pro­gram for the purpose to which it was designed. The program files, in whole or in part, and all copyright or other intellectual and proprietary rights therein are and remain the property of METRON U.S..
THIS SOFTWARE IS LICENSED ONLY TO YOU, THE LICENSEE, AND MAY NOT BE TRANS­FERRED TO ANYONE WITHOUT THE WRITTEN CONSENT OF METRON U.S. Any authorized transferee of the software shall be bound by the terms and conditions of this License. In no event may you transfer, assign, rent, lease, sell, or otherwise dispose of the software on a temporary or permanent basis except as expressly pro­vided herein.
This License is governed by the State of Michigan and shall benefit METRON U.S., its successors, and as­signs. Licensee consents to jurisdiction in the state and federal courts located in the State of Michigan.
Limited Warranty/Disclaimer of Liability
METRON U.S. warrants that the software will substantially conform to published specifications and to the documentation, provided that it is used on the computer hardware and with the operating system for which it was designed. METRON U.S. will replace defective media or documentation, or correct substantial software errors at no charge within 90 days of purchase. Contact METRON U.S. for information. If METRON U.S. is unable to cor­rect the problem, the License fee will be refunded to you. These are your sole remedies for any breach of war­ranty.
BECAUSE SOFTWARE IS INHERENTLY COMPLEX AND MAY NOT BE COMPLETELY FREE OF ERRORS, YOU ARE ADVISED TO VERIFY YOUR WORK. In no event will METRON U.S. be liable for direct, indirect, special, incidental, or consequential damages arising out of the use of or inability to use the software or documentation, even if advised of the possibility of such dam­ages. METRON U.S. is not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or inability to use the software.
Trademarks
IBM is a registered trademark, and PC/XT is a trademark of IBM Corporation. Microsoft is a registered trade­mark and Windows is a trademark of Microsoft Corporation.
2
Table of Contents
MANUAL REVISION RECORD..........................................................................................................5
1. ABOUT PRO-SOFT QA-40M/45.......................................................................................................7
1.1 Introduction ..................................................................................................................................7
1.2 PRO-Soft QA-40M/45..................................................................................................................7
1.3 About your Manual.......................................................................................................................8
1.4 Updating PRO-Soft Program Files...............................................................................................8
2. GETTING STARTED..........................................................................................................................9
2.1 System Requirements....................................................................................................................9
2.2 Installing PRO-Soft .....................................................................................................................9
2.3 Initial Startup...............................................................................................................................10
2.4 Option Settings............................................................................................................................11
2.5 Printing Test Results...................................................................................................................14
3. TESTING SETUP...............................................................................................................................15
3.1 Equipment Setup.........................................................................................................................15
3.2 Power ..........................................................................................................................................15
3.3 Internal Paddles...........................................................................................................................16
3.4 Special Contacts..........................................................................................................................16
3.5 Initializing Remoting -Defibrillator Mode.................................................................................16
3.6 Initializing Remoting -Pacemaker Mode (QA-45 only)..........................................................17
4. MANUAL TESTING.........................................................................................................................18
4.1 Defibrillator Discharge (Energy) Test........................................................................................18
4.2 ECG Performance Test...............................................................................................................20
4.3 Shock Advisory Algorithm Test.................................................................................................22
4.4 Pacemaker Accuracy Test (QA-45 Only)...................................................................................23
4.5 Pacemaker Sensitivity Test (QA-45 Only).................................................................................24
4.6 Pacemaker Refractory Period Test (QA-45 Only) ....................................................................25
5. SEQUENCE TESTING.....................................................................................................................28
5.1 Sequence Form Window.............................................................................................................28
5.2 Defibrillator Test Sequence Setup..............................................................................................29
5.3 Pacemaker Test Sequence Setup................................................................................................30
5.4 Running a Sequence....................................................................................................................32
5.5 Sequence File Management........................................................................................................33
6. EMERGENCY SEQUENCE TRAINING.......................................................................................34
6.1 Sequence Setup...........................................................................................................................34
6.2 Running Emergency Sequences 1 - 7.........................................................................................35
6.3 Running Cardio Sequence...........................................................................................................36
7. SELECT CLOSE TO TERMINATE THE TEST AND RETURN TO THE MAIN SCREEN.
...................................................................................................................................................................36
3
8. CHECKLISTS....................................................................................................................................37
8.1 Checklist Form Window.............................................................................................................37
8.2 Checklist Files Management ......................................................................................................37
9. PROTOCOLS.....................................................................................................................................39
9.1 Protocol Form Window...............................................................................................................39
9.2 Protocol Database........................................................................................................................41
9.3 Protocol File Management..........................................................................................................42
9.4 Importing Data ..........................................................................................................................43
10. TROUBLESHOOTING...................................................................................................................45
APPENDIX A: ERROR REPORT FORM, PRO-SOFT QA-40M/45...........................................47
APPENDIX B: SUGGESTION FORM, PRO-SOFT QA-40M/45.................................................49
4

Manual Revision Record

This record page is for recording revisions to your PRO-Soft QA-40M/45 User Manual that have been published by METRON or its authorized representatives. We recommend that only the man­agement or facility representative authorized to process changes and revisions to publications:
make the pen changes or insert the revised pages;
ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as
superseded and placed in a superseded document file, and;
enter the information below reflecting that the revisions have been entered.
Rev No Date Entered Reason Signature of Person Entering Change
0 - Initial Release
5
This page intentionally left blank.
6

1.1 Introduction

1. About PRO-Soft QA-40M/45

The QA-40M and QA-45 Analyzers are precision instruments for testing defibrillators and, additionally with the QA-45, transcuta­neous pacemakers. They are designed to be used by trained techni­cians.
The defibrillator function of the QA-40M and QA-45 measures the energy output, and ensures that the defibrillator complies with speci­fied requirements. QA-40M and QA-45 have built-in load resis­tances of 50 ohms, which roughly correspond to the impedance of the human body. The defibrillator pads are placed on the analyzers’ contact plates. Thus, the defibrillator is connected through the load resistance. When the defibrillator is discharged, the analyzers calcu­late and display the energy delivered.
In the pacer function the QA-45 tests all types of transthoracic pace­makers. The testing is menu driven, and simple to operate. QA-45 measures and displays a pulse’s amplitude, rate, energy and width. It also conducts demand sensitivity tests, measuring and displaying re­fractory periods, and immunity tests, which determine the pacemak­er’s susceptibility to 50/60 Hz interference.
Testing with the QA-40M/45 is usually a matter of setting up the preferred simulation parameters. Using PRO-Soft QA-40M/45 sim­plifies the process by providing an all Windows-based environment through which simulation parameters are created by the user and sent to the QA-40M/45 unit.

1.2 PRO-Soft QA-40M/45

PRO-Soft QA-40M/45 has two testing methods: manual or se­quence. The manual method allows you to set up test parameters quickly, and run one of the following stand alone tests. There are no restrictions on the parameters used.
Defibrillator Discharge Test ECG Performance Test
Shock Advisory Algorithm Test Pacemaker Accuracy Test
Pacemaker Sensitivity Test Pacemaker Refractory Period Test
Test results are entered on the screen during testing. At the conclu­sion of testing the user may print a report, store the tests and results on disk, or both.
Combinations of tests can be created and stored as “sequences.” PRO-Soft QA-40M/45 maintains a library of these sequences. In this way the user can store and retrieve sequences that are appropriate for the kind of equipment being tested. Sequences can then be used independently, or can be attached to a checklist, written procedure, and equipment data in the form of a test “protocol.” The equipment
7

1.3 About your Manual

data can be entered manually into the protocol, or it may be re­trieved by PRO-Soft from the QA-MAP program, or other equip­ment files. Protocols with test results can be printed, or stored on disk. Also, the results of testing can be sent to QA-MAP, or other equipment maintenance management program, to close a work order and update the defibrillator/transcutaneous pacemaker’s service his­tory.
An additional feature of PRO-Soft QA-40M/45 is its ability to make QA-40/45 serve as a training aid device for medical staff personnel who operate defibrillators in the course of their duties. Several ECG waveform emergency sequences are incorporated into the program. These sequences interact with incoming defibrillator pulses, simu­lating various patient cardiovascular conditions.
This manual is designed to assist you in the basic procedures for cre­ating test protocols with PRO-Soft QA-40M/45. It also covers all features resident in the full PRO-Soft program. Familiarity with Mi­crosoft Windows and its features is assumed. If you are unfamiliar with it, we recommend that you use your Microsoft Windows User’s Guide along with this manual.
This manual contains the following standardized text formatting conventions:
This Represents
Bold title case Menu items and control buttons that can be selected
to perform operations. The underline ( _ ) represents the shortcut key. For example, “Select File, Save” in- structs you to press ”Alt+F,” then press “Alt+S.“ The comma (,) between selections indicates that both se­lections are to be made in sequence.
< braces > Text information specified and entered by the user.
CAPITALS File names and paths. For example, “QAMDAT.50”
The menu under discussion
1.4 Updating PRO-Soft
Program Files
When installing upgrades, backup or copy the original program files to ensure that data you desire to retain is not overwritten.
NOTE
8
Occasionally, updated versions of PRO-Soft QA-40M/45 program files are posted on METRON’s Web Site (http://www.metron­biomed. com). These can be downloaded through file transfer proce­dures. Registered users should contact their local METRON repre­sentative for more information regarding updating.

2.1 System Requirements

2. Getting Started

The following are the minimum requirements for installation:
IBM PC/XT-compatible machine, 80486 SX 25 MHZ or
higher.
A 3.5" floppy drive.
An EGA, VGA, 8514, Hercules or compatible display (rec-
ommended: VGA).
8 MB of RAM.
1 MB of unoccupied hard disk space, plus 3 MB of free
space on the drive contained the WINDOWS:\SYSTEM di­rectory.
Microsoft Windows version 3.11 or later version.
Metron QA-40M/45.
RS-232 cable (null modem/data transfer configured).
Printer for reports. (MS-Windows compatible)
2.2 Installing PRO­Soft
1. Start Windows.
2. Insert PRO-Soft QA-40M/45 install disk # 1 into drive A (or B).
3. Select Run from the File Menu in the Windows Program Manager. (In Win 95, click the Start button and then se­lect Run.)
9
If you are installing PRO­Soft QA-40M/45 for the first time, we recommend using the default drive, directory, and program group (C:\QA-40M/45)
4. Type <drive:\setup>. Then, click the OK button. If you are installing PRO-Soft QA-40M/45 from a server or shared directory, type the full path plus setup.
5. When setup starts, the following dialog box will appear. Type in the desired drive and directory for PRO-Soft QA­40M/45, then click the Continue button.
6. After setup is completed, a program group will be cre­ated in the Windows Program Manager called "PRO-Soft QA-40M/45," or another name you provided during in­stallation.

2.3 Initial Startup

7. Remove the last disk and store the install disks in a safe place.
The first time that PRO-Soft QA-40M/45 is run, information has to be entered, and the program configured for further use.
1. To start the program, double-click on the QA-40M/45 icon.
2. The following dialog box will appear. Enter the Estab- lishment <license holder name>, and the Code <alphanu­meric license number> exactly as written on the license agreement supplied:
10
Be careful to enter the name and code exactly as written on the license agreement.
3. When you click the OK button the PRO-Soft QA-40M/45 main application window is displayed:
Protocol Checklist

2.4 Option Settings

Options
General Ctl+G Reports Ctl+R
The Main Application Window contains menus as well as an icon bar. This bar enables the user to advance quickly to the following program operations through pressing the appropriate icon:
Defibrillator Discharge Test
Pacemaker Sensitivity Test
ECG Performance Test
Emergency Sequence Test
I I I I I I
I I I I I I
Sequence Equipment
Information
Pacemaker Accuracy Test
Pacemaker Refractory Period Test
Shock Advisory Algorithm Test
To configure program parameters click the Option Menu. Two fur­ther selections appear: General, which sets program pathing param­eters, and Reports, which sets standardized report headers .
Help
11
1. G eneral. When selected, the following dialog box appears:
In the topmost section of this box there are three tabs: PRO- Soft, QA-40M/45, and QA-MAP. By clicking one of these tabs, its corresponding card will be displayed, containing the settings available for the item described on the tab.
After editing the option settings press OK to save changes, or Cancel to discard them. When a <Path> text box is highlighted the Browse button will also be available and, by pressing this button, a file window will be displayed where you can select the appropriate path.
a. P RO-Soft Tab. This tab provides basic identification of the
operator of the test equipment, as well as pathing directions for the PRO-Soft program in order for it to function. The configuration parameters and their descriptions are:
12
Operator: The person executing the tests.
Load protocol at startup:
Path in the PRO-Soft files: Path to the QA-40M/45 database files.
Path to procedure files:
This option, when checked, automati­cally advances the program and opens to a blank protocol window at startup.
Default path to the files where the test procedures are stored.
The operator, automatic loading, and pathing settings, once made, become the default settings. To change the defaults, enter new operator and pathing settings or, in the case of automatic protocol loading, uncheck the box. Then, save by pressing OK.
b. Q A-40M/45 Tab. This tab is used to enter the serial num-
ber of the QA-40M/45 utilized for testing, and to set the communications port (COM1-4) to be utilized for initiating the QA-40M/45 remote setting at start-up. Once entered, these become the default settings. To change the defaults,
enter a new serial number or enter a new COM Port, and save by pressing OK.
c. Q A-MAP Tab. Most QA-MAP interaction control settings
included in the PRO-Soft options may be changed during the execution of the PRO-Soft QA-MAP Program1. How­ever, the following default settings are only available through the PRO-Soft QA-40M/45 options menu:
Path: Path to the QA-MAP file QAMDAT.MDB.
List equipment codes when retrieving data:
Retrieve from QA-MAP:
- Equipment data
- Sequence
- Procedure
Controls whether the equipment list window or the equipment code text box window opens in the retrieval operation.
Sets the defaults as to the types of informa­tion extracted from QA-MAP.
Once entered, these become the default settings. To change them, enter new setting and save by pressing OK.
2. R eport. When selected, the following dialog box appears:
Report header title: The path and filename of the header to be
Report language: This text box is blank, as default is auto-
Margins in millimeters: Path to the QA-40M/45 database files.
Font name: Report default font.
Font size: Report default font size.
used on all reports.
matically English. If other languages are desired, contact your METRON represen­tative for the appropriate files.
The heading layout of test reports can be customized by editing the file ‘STANDARD.HDR’ in the PRO-Soft di­rectory with any text editor (e.g., MS Windows Notpad). See Appendix A for further guidance.
1
Available separately from your Metron Representative.
13

2.5 Printing Test Results

1. Test results are printed by selecting the Print button on the test result window.
If this is the first test result for a manual test session you will be prompted to fill in a Equipment Information Form (see below).
Each time a test is run thereafter, the printed result is buffered. A complete report will be sent to the printer when the test win­dow is closed.
2. After selecting Print (or after entering the Equipment Informa­tion Form and selecting OK) the following window appears. This is for prepublication review, and is in a WYSIWYG for­mat.
14
Window controls enable you to enlarge the results, and navigate from one page to the next. When review is complete, press the Print button.

3.1 Equipment Setup

3. Testing Setup

Communication between the PRO-Soft QA-40M/45 and the QA-40M/45 is based upon an RS-232C interface, that is standard on most modern personal computers. The operation is essentially a “master-slave” configuration, whereby the computer operates the QA-40M/45.
Attach an RS-232 (null modem/data transfer configured) cable to the 9-pin D-sub outlet port located at the rear of the QA-45. Do not attach the printer cable to the QA-45. See below.
Some RS-232C cables are missing the connection be­tween the seventh and the eighth wires in the cable. The cable may still be called NULL-modem, but it will not work with the QA-40M/45.

3.2 Power

Do not use mercury, air or carbon-zinc batteries.
NOTE
NOTE
1. Main On/Off Switch. QA-45 should remain off for at least 5 seconds before switching on again, in order to al­low the test circuits to discharge fully.
2. Low Battery Power. If battery power falls below 6.9
volts (± 0.3 volts), the display will show 'Change battery,
and reset system'. This means that the battery should ei­ther be replaced or the instrument should be connected to a battery eliminator. The main switch has to be switched off and then on again in order to use the instrument.
3. Changing Batteries. Open the compartments in the base of the instrument, replace the old batteries with new ones, and close the compartment covers. Use 9 volt alkaline batteries (Duracell MN1604 or similar).
15
NOTE
Remove the batteries and disconnect the AC Adapter if you do not intend to use the QA-40M/45 for an ex­tended period of time.

3.3 Internal Paddles

3.4 Special Contacts

4. Battery Eliminator
METRON’s AC Adapter plug-in power supply trans­former allows you to use the QA-40M/45 anywhere a standard electrical outlet is available. To attach the adapter insert its small connector into the micro jack la­beled “Batt. Elim. 9V DC” on the right rear of the unit. Plug the large connector into a standard electrical outlet.
To be able to test defibrillators with internal paddles, an internal paddle adapter has to be used. These contacts have a banana plug that is attached to the standard paddle contact, and which is pro­tected by a plastic insulation washer.
Certain defibrillators (automatic models and those with pacer op­tions) have special contacts that are fastened to the electrodes at­tached to the patient. Metron AS has special accessory adapters to suit the majority of these defibrillators. They are more or less the same as the internal pad adapter except that they have a special adapter on the top, which matches the contact on the defibrillator.
Defibrillator paddle adapter (specify defibrillator type): (P/N 13410)
Pacemaker external load cable (specify type pacemaker type): (P/N 13415)
3.5 Initializing Remoting ­Defibrillator Mode
1. Before initiating remote control, ensure that the:
RS-232C cable is securely connected to the communication
ports on the QA-40M/45 and your computer.
Proper communications port has been assigned.
‘Mode’ Switch is in the ‘HIGH’ or ‘LOW’ position.
High For Normal adult testing
Low For low energy testing, where energy does not exceed 50
joules, and peak voltage does not exceed 1200 volts.
Low or high level ECG connectors are connected to the ten
4 mm AHA color-coded safety terminals, or standard phone jack, as appropriate.
2. With the QA-40M/45 switched ON, advance to the Second Main Menu Bar by pressing more-2 (F5):
16
3.6 Initializing Remoting ­Pacemaker Mode (QA-45 only)
Press REMOTE CONTR. (F4) to initiate remote control.
3. Communication will be established when testing or download­ing of presets start. A rolling ball, located in the lower left cor­ner of the main application window (see below), indicates that communication between the computer and the QA-40M/45 is active.
1. Before initiating remote control, ensure that the:
RS-232C cable is securely connected to the communication
ports on the QA-40M/45 and your computer.
Proper communications port has been assigned.
Mode Switch is in the ‘PACE’ position.
Pacemaker output cables are connected to either the 'Pacer
Input Variable Load' or 'Pacer Input Fixed Load' on QA-45.
Pacemaker is in ‘DEMAND’ mode (if immunity testing).
2. With the QA-45 switched ON, advance to the Second Main Menu Bar by pressing more-2 (F5):
Press REMOTE CONTR. (F4) to initiate remote control.
3. Communication will be established when testing or download­ing of presets start. A rolling ball, located in the lower left cor­ner of the main application window (see below), indicates that communication between the computer and the QA-45 is active.
17

4.1 Defibrillator Discharge (Energy) Test

4. Manual Testing

The Manual method allows you to set up test parameters quickly, and run one of the following stand alone tests:
Defibrillator Discharge Test ECG Performance Test
Shock Advisory Algorithm Test Pacemaker Accuracy Test
Pacemaker Sensitivity Test Pacemaker Refractory Period Test
There are no restrictions on the parameters used. However, only the QA-45 does both defibrillator and transcutaneous pacemaker testing. The QA-40M is restricted to defibrillator testing only. Manual tests can be activated in two ways. One is to select the ap­propriate icon from the icon menu bar in the Main Application Win­dow. (The icons below are those which activate the tests.) The other method is to select QA-45 from the PRO-Soft QA-40M/45 Main Menu, and the Manual Test Menu appears. (See the menu below.)
Defibrillator Discharge Test
ECG Perform. Test
Shock Adv. Algorithm Test
1. General. QA-40M/45 measure the energy output, and ensure that the defibrillator complies with specified requirements. They have a built-in load resistance of 50 ohms, which roughly corresponds to the impedance of the human body. The defibril­lator pads are placed on the QA-40M/45 contact plates, thus en­suring that the defibrillator is connected through the load resis­tance. When the defibrillator is discharged, PRO-Soft calculates and displays the energy delivered. Defibrillator energy is defined as an integral of the moment of the discharged energy from the defibrillator. The energy is equal to the square of the voltage, divided by the load resis­tance.
E = p dt =  V2 / R dt = V2 dt / R
Pacemaker Accuracy Test
Pacemaker Sensitivity Test
Pacemaker Refractory Period Test
QA-40M/45
Defibrillator discharge test ECG performance test Shock advisory algorithm test Pacemaker accuracy test Pacemaker sensitivity test Pacemaker refractory period test
18
QA-40M/45 measures and records the voltage pulse every 100
µs, 1000 times, for a total time of 100 ms. The squares of the voltages are then summed, multiplied by 100 µs, and divided by
the load resistance, 50 ohms.
1000 1000
E =  (V
2
) dt / R =  (V2) 100 µs / 50 ohms
2. Test Procedure.
a. Initiate remoting. (See chapter 3 for more information)
b. Set the defibrillator to ‘SYNCHRONIZED’ mode, if
desired. If set, there must be an ECG wave and ampli­tude to simulate before testing begins. While the test is activated, the waveform and amplitude cannot be changed.
c. Press the Defibrillator Discharge Test icon button or se-
lect QA-45, Defibrillator discharge test from the PRO­Soft QA-40M/45 Main Menu, and the following appears.
d. In the Manual Energy Settings field, select the desired :
ECG Settings PRO-Soft QA-40M/45 includes the following
Amplitude 0.5, 1.0, 1.5 and 2.0 mV.
ECG wave selection for cardioversion tests, or for the testing of ECG monitors:
Normal Sine Rates: 30, 60, 80, 120, 180, 240, 300 BPM.
ECG arrhythmia types are as follows:
Afib Atrial Fibrillation
2^BLKII Second degree A-V block
RBBB Right Bundle Branch Block
PAC Premature Atrial Contraction
PVC Early Early PVC
PVC Std PVC
PVC R on T R on T PVC
MF PVC Multifocal PVC
Bigeminy Bigeminy
Run of 5 PC Bigeminy Run of 5 PVCs
Vtach Ventricular Tachycardia
Vfib Ventricular Fibrillation
e. Click the Run button. PRO-Soft will set the ECG wave and
the amplitude and start waiting for a defibrillation. A de­fibrillation can then be fired until the Stop button is pushed.
19
If the Test Pulse button is activated while PRO-Soft is waiting for a defibrillation, the QA-40M/45 will generate a test pulse. The test pulse result will be shown in the form.
APEX (+) pad right plate STERNUM pad left plate

4.2 ECG Performance Test

f. Securely place the defibrillator paddles on the QA-
40M/45 contact plates, and discharge the defibrillator. The APEX (+) pad should be connected to the right-hand plate, and the STERNUM pad to the left plate. This ensures cor­rect signal polarity for the oscilloscope output. A reversal of this configuration will not damage the QA-40M/45, nor will it give incorrect energy readings. However, the polarity of the oscilloscope output will simply be reversed. The dis­charge from the defibrillator is transferred to the QA­40M/45's load resistance.
g. Following discharge PRO-Soft will display delivered en-
ergy, peak voltage, current, and synchronization delay in the upper right portion of the Test Result field. After ten seconds PRO-Soft also displays the discharge wave form in the lower right portion of the Test Result field.
h. Before changing ECG wave or amplitude, click the Stop
button. Then enter setting changes.
i. Select Close to terminate testing and return to the Main
Screen. To print a report select the Print button.
1. General. The ECG Performance Test simulates the following waves, and is designed for testing ECG monitors. Testing accu-
racy is ± 5 % (Lead II 1.0 mV).
Pulse: 1.0 mV Pulse lasts for 4 seconds. Used for testing
low frequency responses.
Square: 2 Hz 1.0 mV p-p biphasic. Used for testing amplifi-
cation and damping.
Triangle: 2 Hz 1.0 mV. Used for testing linearity.
Sinus: 0.1, 0.2, 0.5, 10, 40, 50, 60, and 100 Hz. Used for
testing frequency response.
Amplitude: 0.5, 1.0, 1.5, 2.0 mV (Lead II)
20
2. Test Procedure.
a. Initiate remoting. (See chapter 3.)
b. Set the defibrillator to ‘SYNCHRONIZED’ mode, if
desired. If set, there must be an ECG wave and ampli­tude to simulate before testing begins. While the test is activated, the waveform and amplitude cannot be changed.
c. Press the ECG Performance Test iicon button or se-
lect QA-45, ECG performance test from the PRO-
APEX (+) pad right plate STERNUM pad left plate
Soft QA-40M/45 Main Menu, and the following ap­pears.
d. Select the desired ECG wave and Amplitude settings.
d. Click on the Run button. PRO-Soft runs the ECG per-
formance wave until the Stop button is selected.
d. Securely place the defibrillator paddles on the QA-
40M/45 contact plates, and discharge the defibrillator. The APEX (+) pad should be connected to the right­hand plate, and the STERNUM pad to the left plate. This ensures correct signal polarity for the oscilloscope output. A reversal of this configuration will not dam­age the QA-40M/45, nor will it give incorrect energy readings. However, the polarity of the oscilloscope out­put will simply be reversed. The discharge from the de­fibrillator is transferred to the QA-40M/45's load resis­tance.
e. ECG performance waves can be run in sequence. By
selecting Autosequence, a mark will appear in the checkbox. This mark tells you that if the Run button is pushed, PRO-Soft will run the chosen ECG for ten sec­onds before jumping to the next one. All ECG waves are performed in sequence.
If you do not want to wait ten seconds for the next wave, the Advance button may be used. By selecting this button the program will stop executing the active ECG wave and start running the next on the list. To select the previous wave in the list click the Back button.
f. Select Close to terminate testing and return to the Main
Screen. To print a report select the Print button.
21
4.3 Shock Advisory Algo­rithm Test
1. General. This tests the analysis and prompting of certain defib­rillators that can advise operators as to whether or not to per­form a shock. PRO-Soft QA-40M/45 can simulate eight specific waveforms for analysis by the defibrillator that should then prompt the user to ‘shock’ or ‘no shock,’ in accordance with na­tional and international guidelines. The waveforms are:
Asystolic: No ECG No shock
SVTa_90: Supraventricular tachycardia, 90 BPM No shock
PVT 140: Monoventricular tachycardia, 140 BPM No shock
MVT 140: Polyventricular tachycardia, 140 BPM No shock
PVT 160: Monoventricular tachycardia, 160 BPM Shock
MVT 160: Polyventricular tachycardia, 160 BPM Shock
CVF: Coarse Ventricular Fibrillation Shock
FVF: Fine Ventricular Fibrillation Shock
2. Test Procedure.
a. Initiate remoting. (See chapter 3 for more information)

b. Press the Shock Advisory Algorithm Test iicon button

or select QA-45, Shock advisory algorithm test from the PRO-Soft QA-40M/45 Main Menu, and the follow­ing test window appears.
APEX (+) pad right plate STERNUM pad left plate
22
c. Securely place the defibrillator paddles on the QA-
40M/45 contact plates, and discharge the defibrillator. The APEX (+) pad should be connected to the right­hand plate, and the STERNUM pad to the left plate. This ensures correct signal polarity for the oscilloscope output. A reversal of this configuration will not dam­age the QA-40M/45, nor will it give incorrect energy readings. However, the polarity of the oscilloscope out-

4.4 Pacemaker Accuracy Test (QA-45 Only)

put will be reversed. The discharge from the defibrilla­tor is transferred to the QA-40M/45's load resistance.
d. Set the defibrillator to ‘SHOCK ADVISORY’ mode.
e. Run each of the curve forms for three or more seconds.
f. Check the appropriate check box if the defibrillator ad-
vised ‘shock’ or ‘not to shock.’
g. Select Close to terminate the test and return to the
Main Screen. To print a report select the Print button.
1. General. This test evaluates the pacer frequency and delivered energy, and determines if noise interferes with the pacing. To conduct it, the operator selects the following parameters:
Load: The load that will be applied between the pads.
The options are Pacer Input Fixed Load, or load ranges of 50 to 2300 ohms. The latter are in steps of 50 ohms up to 200 ohms, and 100 ohms from 200 up to 2300 ohms.
Noise: To simulate disturbances, 50 Hz, 1-10 mV or 60
Hz, 1-10 mV noise can be added
Rate set and limit: This specifies the target rate in ppm, and the ac-
cepted deviance in percent.
Amplitude set: This specifies the target amplitude in mA, and the
accepted deviance in percent.
2. Test Procedure.
a. If testing with noise, set pacemaker to ‘DEMAND’
mode.
b. Press the Pacemaker Accuracy Test iicon button or
select QA-45, Pacemaker accuracy test from the PRO-Soft QA-40M/45 Main Menu, and the test setup window appears.
c. Enter the desired settings.
d. Click on the Run button. The test result window then
appears (see below) displaying the rate, amplitude,
23
pulse width and energy. For the rate and amplitude it also shows the relative deviance (in percent) from the target settings.
e. The result window continues to show the measure-
ments until you click the Close button. To print a re­port select the Print button. If you want to make a new pacemaker accuracy test select the Next button to set test parameters.

4.5 Pacemaker Sensitivity Test (QA-45 Only)

1. General. Sensitivity is the minimum QRS amplitude (mV) re­quired to cause the pacemaker to operate in the ‘DEMAND’ mode. During sensitivity measurement three different wave­forms are selectable, with widths varying in steps from 10 to 200 ms. This waveform is delayed from the pacer pulse so that it is outside the pacing refractory period. PRO-Soft QA-45 then checks whether this wave is sensed or not by the pacemaker. If it is not sensed, a message 'EXCEEDED' is displayed, which means that the pacemaker needs an amplitude of more than 100 mV for sensing at that setting. If the wave is sensed, PRO-Soft then reduces the amplitude in steps until it reaches the lowest value required for the pacemaker to sense it. (The internal algo­rithm used converges to the lowest value in the least number of cycles.) This lowest value is the pace trigger sensitivity. Parameters to this test are:
Load: The load that will be applied between the pads
Trigger wave: The waveform that will be used to trigger the
pacemaker. NOTE: A high pacer rate shortens the length of the test
Limit pads: The maximum allowed threshold
Limit ECG: The maximum allowed threshold
24
2. Test Procedure.
a. Press the Pacemaker Sensitivity Test iicon button or
select QA-45, Pacemaker sensitivity test from the PRO-Soft QA-40M/45 Main Menu, and the test setup window appears.
b. Enter the desired settings.
c. Click on the Run button. The test takes approximately
40 seconds with a pacemaker setting of 120 ppm.
b. The test result window then appears (see below) dis-
playing the Pads and ECG sensitivity in ms, and ampli­tude in mV.

4.6 Pacemaker Refractory Period Test (QA-45 Only)

c. The result window continues to show the measure-
ments until you click the Close button. To print a re­port select the Print button. If you want to make a new pacemaker accuracy test select the Next button to set test parameters.
1. General. This test is used to test the time interval in millisec­onds during which the pacemaker is insensitive to any external
25
inputs. PRO-Soft does this by measuring the maximum time in­terval after the generation of a pacer pulse, and maximum time interval after a QRS wave.
a. Refractory Period. A time interval in milliseconds, during
which a pacemaker is insensitive to any external inputs. If a QRS is detected during this period, the pacemaker ignores it. On the other hand, if a QRS is detected outside the re­fractory interval, then the pacemaker resets its internal timer and the next pacer pulse is generated after a delay of one time period from this QRS wave.
b. Paced Refractory Period. The maximum time interval af-
ter the generation of a pacer pulse during which time the presence of a QRS wave is ignored.
The measurement of paced refractory period takes a few cycles of the pacemaker output. First, PRO-Soft measures the pacer-to-pacer interval T. Then, it puts out a square wave 40 milliseconds wide, delayed by delay time D, which is more than the pacing refractory period, from the last pacer pulse. The pacemaker senses this square wave. The delay time D is gradually decremented in subsequent cycles until the square waveform is not sensed by the pace­maker. The maximum value of the delay time D, for which the pace maker does not sense the square wave, is the paced refractory period.
c. Sensed Refractory Period. The maximum time interval
after a QRS wave is sensed by the pacemaker during which time the presence of a second QRS wave is ignored.
The sensed refractory period is measured in a similar man­ner, except that PRO-Soft now generates two square waves instead of one. The first square wave is generated at a fixed time delay from a pacer pulse, which is greater than the paced refractory period. The pacemaker always senses this square wave.
The second square wave is generated at a delay D from the first square wave. The initial value of D is selected to be greater than the sensed refractory period. Therefore the first time the pacemaker is on it also senses the second square wave. In subsequent cycles, the delay 'D' is gradually re­duced until the pacemaker is unable to sense the second square wave. The maximum value of D, for which the pacemaker does not sense the second square wave, is the sensed refractory period. The purpose of the Pacemaker Re­fractory Period Test is to ensure that the pacer does not give pulse when the heart is in the refractory state (T wave).
26
Parameters to this test are:
Load: The load that will be applied between
Rate set: The rate at which the pacemaker is set.
Paced refractory period limit:
2. Test Procedure.
a. Press the Pacemaker Refractory Period Test iicon
button or select QA-45, Pacemaker refractory period test from the PRO-Soft QA-40M/45 Main Menu, and the test setup window appear.
the pads
NOTE: A high pacer rate shortens the length of the test
Specifies the minimum paced refractory period limit. NOTE: The limits are speci­fied for each pacemaker, and depend on the pulse rate that is set in the pace­maker.
b. Enter the desired settings.
c. Click on the Run button.
d. The test result window then appears (see below) dis-
playing the paced and sensed refractory periods in mil­liseconds.
e. The result window continues to show the measure-
ments until you click the Close button. To print a re­port select the Print button. If you want to make a new pacemaker accuracy test select the Next button to set test parameters.
27
5.1 Sequence Form Win­dow

5. Sequence Testing

Sequence testing enables the operator to perform multiple tests semiautomatically. Instead of manually setting up and running one test at a time, a sequence of several tests can be set up in advance of testing. Once a sequence is created it can be it for later retrieval, and reuse. There are no restrictions on how the sequence is composed.
Form Sequence F2
Checklist F3 Protocol F4
Press the Sequenceiicon button from the icon menu bar, or select Form, Sequence from the PRO-Soft QA- 40M/45 Main Menu. The sequence window appears. The window contains six tabs. The Se­quence info tab contains general information about the sequence,
and has three control buttons with which to activate the tabs. Checking:
1. Defibrillator activates the tabs for:
Defibrillator Discharge Test
ECG Performance Test
2. External transcutaneous pacemaker activates the tabs for:
Pacemaker Discharge (Accuracy) Test
Pacemaker Sensitivity Test
Pacemaker Refractory Period Test
3. Shock advisory algorithm activates the window for the Shock Advisory Algorithm Test. (See Chapter 4)
28
5.2 Defibrillator Test Se­quence Setup
1. General. By clicking one of the defibrillator test tabs, the cor­responding card will be displayed, containing a table for enter­ing the settings to be incorporated in the testing.
2. Defibrillator Discharge Test. This tab provides a table for en­tering configuration parameters as follows:
Set Energy (J) Energy test value in joules
Limit (%) Limit value for delivered energy in percent.
NOTE: PRO-Soft allows to set illegal limit values
(i.e., less than)% or greater than 100%). There­fore, enter only values between 0 -100%.
Max. Delay (ms) If desired, the upper limit value for the delay, in
milliseconds
If an ECG wave with an amplitude is desired for each test click, respectively, the ECG and Ampl fields. When the ECG field is clicked, the following window appears. Only the presets shown are available. Select the desired wave, then press OK. The preset will then be entered in the ECG field in the test ta­ble.
When the Ampl field is clicked, the following window appears. Select the desired amplitude, then press OK. The preset will then be entered in the Ampl field in the test table.
29
5.3 Pacemaker Test Se­quence Setup
3. Defibrillator Charging Test. This tab provides a table for en­tering configuration parameters as follows:
Set Energy (J) Energy test value in joules
Limit (%) Limit value for delivered energy in percent.
NOTE: PRO-Soft allows to set illegal limit values
(i.e., less than)% or greater than 100%). There­fore, enter only values between 0 -100%.
Max. Delay (ms) If desired, the upper limit value for the delay, in
milliseconds
1. General. By clicking one of pacemaker test tabs, the corre­sponding card will be displayed, containing a table for entering the settings to be incorporated in the testing.
2. Pace Discharge (Accuracy) Test. This tab provides a table for entering the following configuration parameters:
Rate (ppm): Target rate in ppm. Amplitude (mA): Target amplitude in mA.
Rate limit (%): Accepted deviance in percent.
Amplitude limit (%): Accepted deviance in percent.
Load: The load that will be applied between the pads.
The options are Pacer Input Fixed Load, or load ranges of 50 to 2300 ohms. The latter are in steps of 50 ohms up to 200 ohms, and 100 ohms from 200 up to 2300 ohms.
Noise: To simulate disturbances, 50 Hz, 1-10 mV or 60
Hz, 1-10 mV noise can be added
When any field is clicked, the Pacemaker Accuracy Test Win­dow appears.
30
Enter either the values into the text boxes, or select from the drop down lists. When the desired values are entered, press Run. The configuration is then transferred to the Pace Dis­charge test tab.
3. Sensitivity Test. This tab provides a table for entering the fol­lowing configuration parameters:
Limit pads (mV): The maximum allowed threshold in mV
Limit ECG (mV): The maximum allowed threshold in mV
Wave: The waveform that will be used to trigger the
Load: The load that will be applied between the pads
pacemaker. NOTE: A high pacer rate shortens the length of the test
When any field is clicked, the Pacemaker sensitivity test pa­rameters window appears.
Enter either the values into the text boxes, or select from the drop down lists. When the desired values are entered, press Run. The configuration is then transferred to the Sensitivity test tab.
4. Refractory Period Test. This tab provides a table for entering the following configuration parameters:
Pacer rate (ppm): The rate at which the pacemaker is set. NOTE:
Limit paced (ms): Minimum paced refractory period limit, in ms.
Limit sensed (ms): Minimum sensed refractory period limit, in ms.
A high pacer rate shortens the length of the test
NOTE: The limits are specified for each pace­maker, and depend on the pulse rate that is set in the pacemaker.
NOTE: The limits are specified for each pace­maker, and depend on the pulse rate that is set in the pacemaker.
When any field is clicked, the following manual pacemaker re­fractory period parameters window appears.
31

5.4 Running a Sequence

Sequence
Add test Remove test
Run test F11
Stop test F12
Enter either the values into the text boxes, or select from the drop down lists. When the desired values are entered, press Run. The configuration is then automatically entered in the se­quence test tab.
After a test sequence has been defined and saved, it can be retrieved and used at any time. To run a sequence test:
1. Load the test sequence desired by pressing the Sequenceiicon button or selecting Form, Sequence, or Form, Protocol, View Sequence from the PRO-Soft QA- 40M/45 Main Menu.
2. Select File, Load sequence. The Browse window opens, from which the desired test sequence can be opened. Either double­click a file to open it, or highlight it and press the Open button.
3. From the Sequence Menu select Sequence, Run test. The se­quence will then start running in the following order:
Defibrillator Discharge Test
Defibrillator Charging Test
Pacemaker Discharge Test (QA-45 Only)
Pacemaker Sensitivity Test (QA-45 Only)
Pacemaker Refractory Period Test (QA-45 Only)
Shock Advisory Algorithm Test (if selected)
4. In the Defibrillator Discharge Test, following discharge, PRO­Soft displays delivered energy, peak voltage, current, and syn­chronization delay for each test automatically. After ten sec­onds PRO-Soft also displays the discharge wave form in the lower portion of the test tab.
32
5. When the Defibrillator Charging Test tab opens press the Charge Timer (clock at the lower left of the tab). PRO-Soft counts five seconds after test start, showing the following win­dow, with a red background, during the countdown:
5
Wait until countdown reaches 0. Then
charge the defibrillator and fire.
When the countdown reaches zero, the window changes to green, and the following message appears:
0
Charge the defibrillator
and fire immediately.
Press charge, and then press discharge until the pulse is given. The test will then display the measured time from the end of countdown until the pulse is given, as well as the resultant en­ergy.
5.5 Sequence File Man­agement
File
New Sequence Ctl+N Load Sequence Ctl+L Save Sequence Ctl+S Save Sequence As Ctl+A Close Ctl+F4 Print Printer Setup Exit
1. N ew Sequence. Creates a blank sequence form. If an existing sequence is in the form when this option is chosen, PRO-Soft QA-40M/45 will prompt the user as to whether or not the cur­rent sequence is to be saved before clearing the form.
2. L oad Sequence. Opens the Browse form (see below) where a previously saved sequence can be retrieved. This is done by highlighting the sequence to be retrieved and pressing Open, or double-clicking the desired sequence. If an existing sequence is in the form when this option is chosen, PRO-Soft QA-40M/45 will prompt the user as to whether or not the current sequence is to be saved before loading the next sequence.
Sequence
Add test Remove test Run test F11 Stop test F12
3. S ave Sequence and Save Sequence as. Saves the current se- quence in the Sequence Database under the current or new file name, as appropriate. By selecting the preferred storage format both the sequence setup and the sequence result can be saved. When saving test sequences the Browse form is opened, and you must specify a sequence name.
4. Adding and Removing Tests. New tests are added by selecting Sequence, Add test, or Sequence, Remove test.
5. P rint. To print a protocol, select File, Print. Then select whether to print the protocol or write it to a file.
33

6.1 Sequence Setup

6. Emergency Sequence Training

A unique feature of PRO-Soft QA-40M/45 is its ability to make QA-40/45 serve as a training aid device for medical staff personnel who operate defibrillators in the course of their duties. Several ECG waveform emergency sequences are incorporated into the program. These sequences interact with incoming defibrillator pulses, simu­lating various patient cardiovascular conditions.
Emergency sequences can be activated in two ways. One is to select the Emergency Sequence Test icon button from the menu bar in the Main Application Window. The other method is to select QA-45, Emergency sequence test from the PRO-Soft QA-40M/45 Main Menu. In doing either, the Emergency Sequence window opens:
34
1. Emergency Sequence. This field has seven emergency se­quences, and one cardioversion sequence from which to select.
SEQUENCE PARAMETERS
Emergency 1 Starts with an 80 BPM normal sinus rhythm for 10
beats, a PVC (R on T), six normal beats, another PVC (R on T) and then goes to a continuous Vfib. Firing the defibrillator at any energy level will con­vert to normal ECG.
Emergency 2 Starts with 10 normal beats, then has 25 beats of
Vtach, and then also goes to a continuous Vfib. Fir­ing the defibrillator at any energy level will convert to normal ECG.
Emergency 3 Starts with Vfib. If a 200 Joule pulse is fired within
two minutes of starting, the waveform converts to normal. Otherwise, the ECG will continue running

6.2 Running Emergency Sequences 1 - 7

SEQUENCE PARAMETERS
Vfib.
Emergency 4 Same as Emergency 3, except that two pulses at
200 Joule are required within two minutes to con­vert to normal ECG.
Emergency 5 Same as Emergency 4, except that two pulses at
200 Joule followed by one pulse at 360 Joule are required within two minutes to convert to normal ECG.
Emergency 6 Same as Emergency 5, except that after converting
to normal and beating normally for three minutes, the ECG shows two PVCs (R on T) and then reen­ters Vfib. The same series of two pulses at 200 J and one at 360 J are required within two minutes to convert to normal. Otherwise, the ECG will continue running Vfib.
Emergency 7 ECG waveform starts at Vtach at 180 BPM for one
minute then goes to Vfib. Any defibrillator pulse within two minutes will convert to normal. Other­wise, the ECG will continue running Vfib.
Cardio This sequence starts at continuous Atrial fibrillation.
The <<patient>> must be defibrillated syn­chronously to be successful.
2. Result. Following discharge PRO-Soft displays delivered en­ergy in joules, peak voltage in volts, current in amps, and syn­chronization delay in milliseconds, in this field.
After a correct defibrillation, the ECG may continue run­ning Vfib for a short time be­fore converting to normal. This is not physiological, and is due to the speed of the communication between the QA-40M/45 and the PC.
NOTE
1. Select one of the emergency sequences. Emergency sequences 3 through 6 require a certain energy of defibrillation, or sequence of energies, to convert to a normal ECG waveform. In these se­quences allowed variations must be specified. To do so, enter the variation in the field labeled Energy Limit (%) (see illus-
tration).
2. Click on the Run button. PRO-Soft will then start executing this sequence.
3. These sequences feature Vfib either immediately, or after run­ning for a while. During execution, PRO-Soft displays text showing the active ECG wave.
4. Whenever the sequence is in a Vfib state, the system waits for a defibrillation. The measured energy, voltage, current and delay will be displayed in the Result field.
5. When the Test pulse button is clicked, QA-40M/45 substitutes the internal test pulse for an actual pulse, and assumes that the pulse is the correct energy for the current sequence.
6. Before changing emergency sequences, or to cardio sequence, click the Stop button.
35
6.3 Running Cardio Se­quence
7. Select Close to terminate the test and return to the Main Screen.
1. Select Cardio. When selected the following additional control buttons appear in the window:
2. Click Run. PRO-Soft will then execute the sequence, starting at continuous Afib. The <<patient>> must be defibrillated syn­chronously to be successful.
3. Fire a real synchronized defibrillator into the load, or click on the Sync button to use a test pulse.
a. If the defibrillator pulse comes within 120 ms of the ECG
<<R>> wave, then the defibrillation is successful and the wave will convert to a normal ECG.
b. If the defibrillator pulse is not synchronized within 120 ms,
or if the Late button for a Test pulse is pushed, the wave­form will show the progress of Vfib. The <<patient>> is then in an emergency condition, and a 200 joule pulse is re­quired to convert the waveform to normal.
4. Before changing to an emergency sequence, click the Stop but­ton.

7. Select Close to terminate the test and return to the Main Screen.

36
8.1 Checklist Form Win­dow
FormSequenceF2
klistF3
ProtocolF4

8. Checklists

The checklist adds manually or visually performed tests and func­tion checks to the protocols (see Chapter 8). Once created, a check­list can be edited, stored and retrieved for future use.
Press the Checklist icon button from the icon menu bar, or select Form, Checklist from the PRO-Soft QA-40M/45 Main Menu and
Chec
the Checklist form appears (see below). It is a table allowing for up to 30 lines, where each line contains a description of what to exam­ine (up to 50 characters). Each line also has two check boxes, 'OK' and 'Failed,' to be used when examining the defibrillator or transcu­taneous pacemaker.
Checklist
Add item Remove item
8.2 Checklist Files Man­agement
File
New Ctl+N Load... Ctl+L Save... Ctl+S Save As... Ctl+A Delete... Print... Ctl+P Printer setup Exit
To enter text, or edit text highlight the field you want to edit with the mouse, and then enter the text. After entering the desired text ei­ther press  to confirm the new or changed text, or  to discard it. Another method is to select Checklist, then either Add item to cre­ate a new checklist item, or Remove item to discard it.
In order to update the check boxes the checklist must be connected to a protocol and edited through this protocol (see the protocols con- nected checklist in the section Protocol).
1. New Checklist. Select File, New. The current checklist window clears and initializes a new checklist form. If you already have an active checklist containing unsaved changes, a dialog box appears. It prompts you as to whether or not you want to save
37
the changes, discard them, or abort the operation and return to the active checklist.
2. Loading and Saving Checklists. To save the active checklist, select File, Save or File, Save As. A Save As window will open (see below), prompting for a name to use for the checklist. After the name has been entered in File name, press OK.
To retrieve an existing checklist, select File, Load. A Browse
window opens (see below). Select among the available check­lists and double click it, or highlight it and press Open.
38
3. Printing Checklists. To print a checklist, select File, Print. Then, select whether to print the checklist, or write it to a file.
9.1 Protocol Form Win­dow

9. Protocols

Protocol contains a description of the defibrillator or transcutaneous pacemaker being tested, and specifies an associated test sequence and checklist. It also includes a procedure file, stating special con­cerns for each defibrillator or pacemaker.
When saved, the results are added to PRO-Soft 40M/45’s protocol database, or to QA-MAP, if installed. This database is an historical record of those protocols for which safety tests have been done with PRO-Soft. Information stored includes the:
Tested defibrillator or pacemaker’s identifying information.
Test sequence used in the test
Date the safety test was executed
Status of the test (OK, Failed)
Protocol filename containing the test results.
Form
Sequence F2 Checklist F3
Protocol F4
Press the Protocol icon button from the icon menu bar, or select Form, Protocol from the PRO-Soft QA-40M/45 Main Menu and the
Protocol form appears (see below).
39
The Protocol application window contains the following fields:
1. E quipment Information Field. This field contains text boxes that are used to describe the defibrillator or transcutaneous pacemaker under test, and are also used in the safety testing re­port to document the test of a particular defibrillator or transcu­taneous pacemaker. The text boxes, each containing space for up to 30 characters, are:
Equipment code: The equipment identifier
Serial number: The equipment serial number
Status: The equipment status
Group: The equipment group
Manufacturer: Manufacturer of the equipment on test
Type: The equipment type
Model: The equipment model
Location: Location of the equipment on test
PRO-Soft can load equipment information already entered in an
ASCII text file or, if you have PRO-Soft QA-MAP installed, from that program. See paragraph 8.4.
2. R emark Field. This field is for inserting remarks on the test. The remark may contain up to 300 characters long, describing special events that may have occurred during the testing. Any testing anomalies peculiar to the equipment under test, or on the testing itself, should be entered in this box.
File
New protocol Ctl+N
Load Ctl+L
Save Ctl+S
Save As Ctl+A
Equipment data
Procedure Load
Import Save Export Save As Print
Printer Setup
Exit
Protocol
View sequence View checklist Run test F11
Stop test F12
3. P rocedure Field. This field is for detailing the testing proce- dure used. For example, it can be a description of how to con­nect the QA-40M/45 to the instrument under test. Procedure is always saved to a separate text file (*.txt). It can be created and edited by typing the information in the procedure text box or, by using any text editor/word processor capable of saving plain text files (e.g., Windows Notepad). When you select an item in File, Procedure, a dialog box appears where you can specify the file you need by its filename. The title of the dialog box shown depends on whether you select Load, Save or Save As. If you have already saved the procedure, Save will store your changes under the filename you specified the first time you saved the file, while Save As will prompt you for a new file­name.
4. Sequence and Checklist Fields. These drop down list boxes contain listings of all sequences and checklists resident in their respective databases. Select which of them PRO-Soft will use to test the equipment described in your protocol. When you have selected a sequence and checklist the menu items Protocol, View sequence and Protocol, View checklist become avail­able. When selecting either of these menu items PRO-Soft will load the sequence or checklist and switch to the respective ap­plication window, where you may edit it before running the test. If you already have an active sequence or checklist, containing
40

9.2 Protocol Database

unsaved changes, a dialog box appears, giving you an opportu­nity to save the changes.
When first opening the protocol application window with a new
protocol these fields will be blank, indicating that no sequence or checklist is connected to the protocol. They have to be cre­ated and connected to the protocol before they will appear.
Protocols are saved and retrieved from the Protocol Database (see below). This database is an historical record of those protocols for which safety tests have been done with PRO-Soft. The information stored includes the:
Tested defibrillator or transcutaneous pacemaker’s equipment
code and serial number
Test sequence used in the test
Date the safety test was executed
Status of the test (OK, Failed)
Protocol filename containing the test results.
To access the protocol database the Protocol format must be open. To do so press the Protocol icon button from the icon menu bar, or select Form, Protocol from the PRO-Soft QA-40M/45 Main Menu. Then, select File, Load. The following dialog box appears:
The Browse window contains two fields.
1. Identification/Test Time/File name. These list boxes contain the protocols already added to the database by equipment serial number, test time and file name.
2. Control. This field manages the database.
a. Open loads the selected protocol into the active protocol
window, allowing you to view the safety test results and
41
other information stored in the protocol. If you already have a protocol open in the active window, and this proto­col contains unsaved changes, PRO-Soft will not load the protocol selected in the database until you have responded to a dialog box prompting you to save the open protocol.
b. Delete removes selected protocol(s) from the library,
prompting for confirmation. Clicking No prevents deletion. If more than one for deletion is selected, PRO-Soft will leave the current equipment and continue to the next one selected, and display the message for this one.
c. Cancel interrupts the entire delete operation. Note that
equipment protocols deleted before selecting Cancel (or No) will not be restored.
d. Update QA-MAP is designed specifically to update
METRON’s QA-MAP. For this to operate QA-MAP must be installed, and the QA-MAP tab in Options, Edit Op- tions must be configured accordingly (see Chapter 2). When this button is clicked, PRO-Soft updates QA-MAP’s Test table with the status of all test results added to the pro­tocol database since the last update operation. Included in the update is the:
Tested defibrillator or pacemaker’s equipment code and
serial number
Test sequence used in the test
9.3 Protocol File Manage­ment
File
New protocol Ctl+N Load Ctl+L Save Ctl+S Save As Ctl+A Equipment data Procedure Import Export Print Printer Setup Exit
 
Date the safety test was executed
Status of the test (OK, Failed)
PRO-Soft program executing the update (i.e., PRO-Soft
QA-40M/45)
1. New Protocol. Select File, New protocol. The system will dis­play an empty Protocol Form window. Fill in the required infor­mation and select the test sequence and checklist to use with the defibrillator or transcutaneous pacemaker being tested.
You can open multiple protocols to allow you to compare, for ex-
ample, this year’s and last year’s test results. To do so select Form, Protocol, and then open another protocol window by se­lecting File, New Protocol. Then, select Window, Tile Hori- zontal, so that both forms appear together. To compare the test results load the protocol corresponding to this year’s testing into one of the protocol windows, and the protocol corresponding the last year’s testing into the other window.
2. Saving Protocols. Protocols are saved to the Protocol Data- base. To save the active protocol, select File, Save or File,
42

9.4 Importing Data

File
New protocol Ctl+N Load Ctl+L Save Ctl+S Save As Ctl+A
Equipment data From QA-MAP
Procedure From ASCII-file Import Export Print Printer Setup Exit
Save As. The Protocol Database window will open, and you will then be prompted for a name to use for the protocol.
3. Retrieving Protocols. Select File, Load, and the Browse win­dow is opened. Select among the available protocols the one that is desired, and double click to load it into the protocol win­dow.
4. Printing Protocols. To print a protocol, select File, Print. Then, select whether to print the protocol or write it to a file.
Equipment files pertaining to defibrillator or transcutaneous pace­makers in the inventory can be created from ASCII text files, or can be imported into PRO-Soft QA-40M/45 from METRON’s QA-MAP program. This appendix details the procedures for these options.
1. Retrieving from QA-MAP. Data on the defibrillator or tran­scutaneous pacemaker to be tested can be imported directly from METRON’s QA-MAP. For this feature to operate QA­MAP must be installed, and the QA-MAP tab in Options, Edit
Options must be configured to reflect Retrieve Appliance Data. (cf. Paragraph 2.2.2).
To import the data select File, Equipment data, From QA-MAP
in the Main Menu.. When the following dialogue box appears, select the appliance code for the defibrillator or transcutaneous pacemaker you wish to test and click on OK. Information stored in QA-MAP states which test sequence to use for each defibrillator or transcutaneous pacemaker.
2. Making Protocol Formats from an ASCII Text File. PRO­Soft can load equipment information from an ASCII text file. ASCII text files can be created on each of the defibrillator or transcutaneous pacemakers for which you intend to produce protocol formats. The ASCII file must follow this pattern :
43
EquipCode = <Equipment Code> SerNo = <Serial Number> Status = <Status> Group = <Group> Manufacturer = <Manufacturer> Model = <Model> Type = <Type> Location = <Location> Sequence Name = <Name of Test Sequence> Procedure = <Procedure Filename>
File
New protocol Ctl+N Load Ctl+L Save Ctl+S Save As Ctl+A
Equipment data From QA-MAP
Procedure From ASCII-file Import Export Print Printer Setup Exit
This file contains information on one defibrillator or transcuta-
neous pacemaker. The text preceding the '=' sign is only for your convenience. The text following the '=' sign is used to con­struct a protocol format. Make sure that the lines appear in the
right order, i.e., appliance code first, .... , procedure text last.
Use a text editor or a word processor that can save ASCII text to create the file (e.g., Notepad). To produce protocol formats from the data in the file, select File, Equipment data, From ASCII-file. Select the file containing the defibrillator or tran­scutaneous pacemaker information and click on OK. The pro­gram will read the file and add one new protocol format to the database library for each defibrillator or transcutaneous pace­maker
44

10. Troubleshooting

Problem Possible Solution
Test sequence grid will not accept data
Check the tests to be made in the 'Tests' frame.
Reports are not printed
Strange characters ap­pear in a report
Ball does not rotate when remoting is attempted, or is set.
Messages:
‘The port is already taken (8005)’
‘Error in Send Command at line 0. Operator valid only when port is open (8018).’
1. Wrong printer driver. Choose the correct printer
when running print report.
2. Do not check the print to file box when running
print report.
Wrong printer driver. Choose the appropriate printer in the Print Manager program.
Communications are not established between the computer and the QA-IDS. Check to ensure that:
1. The QA-IDS is set to “Remote Control.”
2. The correct COM-Port is assigned in the Options Menu.
3. The RS-232 cable has a null modem/data trans-
fer configuration.
45
This page intentionally left blank.
46
APPENDIX A: ERROR REPORT FORM, PRO-SOFT
QA-40M/45
PRO-SOFT QA-40M/45 ERROR REPORT FORM
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1 Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500 Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009 E-mail: metronus@aol.com E-mail: metronfrance@infonie.fr E-mail: support@metron.no
From: (name) Phone: Address: Fax:
Date:
PRO-Soft QA-40M/45 Error Report Product:
Version:
Type
Wrong results Error messages, without reasonProgram stops, no reaction Wrong responses on commands.Other
Description of the situation prior to the error:
Description of the error:
(METRON use internally)
Received date: Comments:
Correction date:
Ref No.
Critical
Minor
Normal
47
This page intentionally left blank.
48
APPENDIX B: SUGGESTION FORM, PRO-SOFT QA-
40M/45
PRO-SOFT QA-40M/45 SUGGESTION FORM
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1 Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500 Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009 E-mail: metronus@aol.com E-mail: metronfrance@infonie.fr E-mail: support@metron.no
From: (name) Phone: Address: Fax:
Date:
PRO-Soft QA-40M/45 Improvement Suggestion Product:
Version:
Type
One window PresentationSeveral windows Options, configuration possibilitiesDocumentation Other
Description of the suggested improvement:
(METRON use internally)
Received date: Comments:
Correction date:
Ref No.
49
This page intentionally left blank.
50
51
Loading...