General Specifications - Mechanical Specifications
Supply
Index
Firmware update :
You may use PC software provided on the CD_ROM and consult Internet site
www.chauvin-arnoux.com
Chapter VII
page 97
Chapter VIII
page 98
page 99
2-channel Oscilloscopes I - 3
General Instructions
General Instructions
Introduction
Precautions and
safety measures
You have just purchased a 2-channel oscilloscope.
Congratulations on your choice and thank you for your trust in the quality of our
products.
This range of oscilloscopes offers the following models:
OX 6062-M
OX 6062-C
OX 6152-C
OX 6202-C
The OX 6062 and OX 6152 models integer also a multimeter mode.
Caution : Measurement ground is connected to earth ground.
It complies with safety standard NF EN 61010-1 (2001), class1, relative to
electronic measuring instruments.
For optimum service, read this manual carefully and comply with the operating
precautions.
Non-compliance with the warnings and/or operating instructions might damage
the instrument and/or its components and could prove dangerous for the user.
•
This instrument has been designed for use:
- indoors
- in an environment with pollution level 2,
- at an altitude of less than 2000 m,
- at a temperature between 0°C and 40°C
- with relative humidity of less than 80% up to 31°C.
monochrom 2-channel 60 MHz 1 GHz sample
color 2-channel 60 MHz 1 GHz sample
color 2-channel 150 MHz 1 GHz sample
color 2-channel 200 MHz 1 GHz sample
•
The safety of any system integrating the apparatus concerns the
responsibility of the assembler of the system.
•
It can be used for measurements on circuits 300 V, CAT II in relation to
earth and can be powered by a 90 to 264 V mains network.
Definition of
measurement
categories
before use
during use
CAT I
CAT II
• Comply with environmental and storage conditions.
• Power supply : it must be connected to the mains network
(from 90 to 264 VAC). Make sure that it is in good working conditions.
• Read carefully all the notes preceded by the symbol .
• The instrument power supply is equipped with an electronic protection
system which is reset automatically when the fault is eliminated.
• As a safety measure, use only suitable cords and accessories supplied with
the instrument or approved by the manufacturer.
Measurement category I is for measurements performed on circuits
not directly connected to mains.
Example : circuits not derived from mains and specially protected
mains-derived circuits.
Measurement category II is for measurements performed on circuits
connected directly to the low-voltage installation.
Example : power supply for domestic appliances and portable tools
I - 4 2-channel Oscilloscopes
General Instructions
General instructions (cont'd)
Symbols used on
the instrument
Warranty
Warning: Risk of danger.
Refer to the operating manual to find out the nature of the potential
hazards and the action necessary to avoid such hazards.
Earth
This equipment is guaranteed for 2 years against any defect in materials or
workmanship, in accordance with the general terms and conditions of sale, .
During this period, the equipment can only be repaired by the manufacturer.
The manufacturer reserves the right to carry out repair or replacement of all or
part of the equipment. If the equipment is returned to the manufacturer, initial
transport costs shall be borne by the customer.
The warranty does not apply following:
•
unsuitable use of the equipment or use with other incompatible equipment
•
modification of the equipment without explicit authorization from the
manufacturer’s technical services
•
repair carried out by a person not certified by the manufacturer
•
adaptation to a specific application, not provided for in the definition of the
equipment or by the operating manual
•
an impact, a fall or a flooding.
According to WEEE directive 2002/96/EC
Maintenance and
metrological
verification
Unpacking Repacking
Cleaning
Before the equipment is opened, it must be disconnected from the mains supply
and the measurement circuits, and the operator must not become charged with
any static electricity. This could cause the destruction of internal parts.
Any adjustment, servicing or repair of the unit under power must be undertaken
only by qualified personnel, after reading the instructions in this manual.
A qualified person is a person who is familiar with the installation, its
construction, its use and the hazards that exist. They are authorized to activate
and deactivate the installation and equipment, in compliance with the safety
instructions.
Information and contact details : contact your nearest distributor.
All the equipment was verified mechanically and electrically before shipping.
When you receive it, carry out a quick check to detect any damage that may
have occurred during transport. If necessary, contact our sales department
immediately and register any legal reservations with the carrier.
In the event of reshipping, it is preferable to use the original package. Indicate
the reasons for the return as clearly as possible in a note attached to the
equipment.
•
Turn the instrument off.
•
Clean it with a damp cloth and soap.
•
Never use abrasive products or solvents.
•
Allow to dry before any further use.
2-channel Oscilloscopes I - 5
Description of the instrument
OX 6152
Oscilloscope
PT100 or TCK
Description of the instrument
This manual describes the operation of anOX 6152oscilloscope.
Presentation
The adjustment of the 2 ways is accessible by the opposite keys.
This instrument is part of our range of portable oscilloscopes.
Its special feature is that it groups 2 instruments into one (OX 6062 and
OX 6152 only) :
FFT
function
OX 6062
OX 6152
•
a digital oscilloscope for laboratory use, intended for the analysis of the
Digital
Multimeter
OX 6062
OX 6202
•
DC voltage measurement
•
AC voltage measurement
(RMS)
•
clamps
•
•
•
•
•
Current measurement for
Resistance measurement
Condenser measurement
Diode test
Continuity test
Temperature measurement
signals encountered in electronics
•
OX 6062 and OX 6152 only : an 4,000-count multimeter.
Caution : “measurement ground is connected to earth ground".
The instrument works with a constant acquisition depth of 2500 counts.
Memory management is organized using a "Windows®"-type file system.
A large colour (or monochrome OX 6062-M) LCD screen is used to view the
signals applied, along with all the settings.
The main control functions are directly accessible using the keys on the front
panel and can be modified using a touch-sensitive pad with the stylus
supplied.
A graphical interface similar to a PC's is used to:
•
select the advanced functions by means of drop-down menus and the
touch-sensitive pad
•
act directly on the objects (curves, cursors, etc.) displayed on the screen.
This means that the settings can be modified.
This instrument is completed by RS232, ETHERNET and CENTRONICS
communication interfaces (option).
Starting up
•
The oscilloscopes are designed to operate on a power source delivering
90 to 264 V (ACrms) at 50 to 60 Hz.
• Take the power cord which is in the trunk (p. 9). The power supply can
remain in the trunk.
• Connect the power cord to the mains.
• Press the instrument's ON switch: it lights on and then a clock is displayed
on the screen during the start-up sequence.
The message "Instrument start-up" is displayed.
The oscilloscope is then ready for use.
By default, the "Advanced" mode is not active
II - 6 2-channel Oscilloscopes
(see p. 58).
Description of the instrument
Description of the instrument (cont'd)
OX 6062
OX 6152
OX 6202
Front Panel
2-channel Oscilloscopes II - 7
Description of the instrument
Description of the instrument (cont'd)
OX 6062
OX 6152
OX 6202
Rear panel
Marking
II - 8 2-channel Oscilloscopes
Description of the instrument
Calibration
Description of the instrument (cont'd)
OX 6062
OX 6152
OX 6202
Interface
Marking
RS232/Ethernet
Connector
2-channel Oscilloscopes II - 9
signal
Extern
Trigger
Description of the instrument
OX 6062
Description of the instrument (cont'd)
Terminal board
OX 6152
OX 6202
OX 6062
OX 6152
marking
OX 6202
marking
Stylus
Magnetized base
12
12
1212
II - 10 2-channel Oscilloscopes
Description of the instrument
Description of the instrument (cont'd)
Front (description)
1 On/Standby/Off key
1 touch-sensitive pad
and stylus
The main functions of the instrument are accessible on the front panel and can
be modified using the touch-sensitive pad (with its stylus) or the menu bar.
•
power on by a short press
•
switch the instrument to standby (yellow LED flashing inside the key) by one
short press. A second press on the key reactivates the instrument.
•
power off by a long press (> 3 s):
- the recording time of the files and configuration is < 15 s.
long press on this key.
If the instrument is not equipped with a battery, never disconnect the
instrument from the mains while the message "System shut down : Please
wait before switch off power" is displayed on the screen. Otherwise, the
current file and all the files previously saved will be lost.
These can be used for:
selection of menus,
validation of functions,
movement of symbols appearing on the LCD screen.
•
The menus at the top of the screen and the submenus selected by the
pointer open and are validated with the stylus.
•
The menus in the curve display area,
the command area
the status area
can be opened with the stylus.
•
The stylus can move the symbols displayed in:
1.
the main display area:
trigger position
position of cursors
reference of the traces displayed
2.
the bargraph:
trigger position
position of cursors
position of zoomed area in the acquisition memory
Place the pointer on the symbol to be moved and keep the stylus pressing
down while you move it to the required position.
•
It is possible to use the stylus to zoom in the display area: drag to create a
rectangle.
4 "operating mode"
keys
28 keys only active
when pressed
You can select the operating mode of the instrument by pressing one of these
4 keys:
"oscilloscope"
"multimeter" (*) (not available on OX 6202)
"harmonic analyser" (not available)
"recorder" (not available)
(*) Caution : measurement ground is connected to earth ground
Shortcut access to the most common functions: see chapter on "The Keys" for
the "Oscilloscope", "Multimeter".
2-channel Oscilloscopes II - 11
Description of the instrument (cont'd)
Description of the instrument
Network
General principles of
the ETHERNET
network
Addressing
ETHERNET physical
addresses
IP addresses
Configuration of the "Network" interface (ETHERNET).
This interface uses the same connector (RS232/ETHERNET), located on the
right-hand side of the instrument, and requires a specific ETHERNET / RJ 45
cable.
ETHERNET and TCP/IP (Transmission Protocol/Internet Protocol) are used
to communicate on a company's network.
Each piece of equipment under TCP/IP has a physical address
(ETHERNET) and an Internet address (IP).
A physical or ETHERNET address, stored in ROM or PAL, identifies each
item of equipment on the network. The physical address enables the
equipment to determine the source of data "packet" transmission.
The physical address is a number coded over 6 bytes represented in
hexadecimal form. Hardware manufacturers procure physical addresses and
allocate them incrementally when the product is manufactured. The physical
addresses cannot be modified.
An IP address is coded over 4 bytes, displayed in decimal format.
( Example: 132.147.250.10). Each field may be coded between 0 and 255
and is separated by a decimal point.
Unlike the physical address, the IP address can be modified by the user.
You must ensure that the IP address is unique on your network. If an address
is duplicated, network operation becomes random.
The IP address is made up of two parts:
•
the network identifier (Network ID) identifying a given physical network
•
the host identifier (Host ID) identifying a specific item of equipment on the
same network.
There are 5 addressing classes. Only classes A, B and C are used to identify
the equipment.
See below:
Class A
0XXXXXXX XXXXXXXX XXXXXXXX XXXXXXXX
Network ID Host ID
Class B
10XXXXXX XXXXXXXX XXXXXXXX XXXXXXXX
Network ID Host ID
Class C 010XXXXX XXXXXXXX XXXXXXXX XXXXXXXX
Network ID Host ID
II - 12 2-channel Oscilloscopes
Description of the instrument
Description of the instrument (cont'd)
SUBNET mask
and GATEWAY
DHCP Protocol
FTP protocol
If the result of the operation ' ET LOGIQUE ' between IP address of the
recipient of the message and the value of subnet mask is different from the
address of the recipient of the message, this message is sent to the
gateway which will be given the responsibility to forward it to destination.
The programming of the mask and the address of the gateway is possible
on the instrument, in the Advanced mode.
This protocol is used to automatically assign an IP address to the
instrument when it connects up to the network.
A DHCP (Dynamic Host Configuration protocol) server must be accessible
on this network (contact your network administrator to make sure that this
server is present).
FTP (File Transfer Protocol) is used in the oscilloscope for fast file
transfers to or from a PC.
To use it, open the browser on the PC and, in the URL field, type the IP
address of the instrument, preceded by "ftp:"
Example: ftp://192.168.3.1
The oscilloscope is an FTP server.
See §. Applications p. 87 and 82.
HTTP protocol
LPD protocol
With this protocol, the instrument can act as a WEB server and you can
access the most frequent settings and view traces on your PC using your
browser (EXPLORER, NETSCAPE, …)
To use it, open the browser on the PC and, in the URL field, type the IP
address of the instrument, preceded by "http:"
Example: http://192.168.3.1
See §. Applications p. 84.
To be able to display the traces, you must install on your PC the Java Virtual
Machine JVM SUN 1.4.1 (or higher). This JVM can be dowloaded from the site
http://java.sun.com
This protocol (Line Printer Daemon) is used by most of the printers
connected to an ETHERNET network, but also by the printing server units
which handle conversion between ETHERNET and CENTRONICS
(
Example: Jet Admin) and UNIX and LINUX workstations.
An LPD server can also be installed on a PC (available as an option with
WINDOWS 2000 or XP).
In all cases, the instrument is an LPD client which has to be configured to
indicate to it the IP address of the LPD server (the workstation PC or
directly the printer) and the logical name of the printer managed by the
server.
See §. Applications p. 83.
2-channel Oscilloscopes II - 13
Oscilloscope Mode
Oscilloscope Mode - The Keys
The Keys
5 "UTILITY" keys or
key pad
By pressing this key, you can select the "oscilloscope" mode.
Direct access to LCD contrast adjustment.
When this key is pressed, the display mode switches from normal to "full screen" display (and vice versa).
The screen is organized in such a way as to leave an optimum trace plotting
surface area: removal of: the menu bar,
the parameters of the traces of the time base,
the bargraph.
Only the permanent settings and the measurements will remain.
The controls on the front panel remain active.
Triggers a hardcopy in accordance with the configuration chosen in the "Util"
and "Hardcopy" menus.
A second press before the end of the process will interrupt the current printout.
If printing is impossible, a "Printing error" message will be sent.
1 "AUTOSET" key
Selective
"AUTOSET"
+
The "
printing is in progress.
The first press will freeze the traces on the screen. They will be displayed in a
lighter colour as a reference to be compared with another acquisition.
A second press will erase them: they will then be lost.
•
Traces will be saved only through the "Memory Trace Save" menu.
•
The reference memories will be accompanied by their reference number.
Automatic optimum adjustment by Autoset on the channels where the signal is
applied.
This concerns the following parameters: the coupling, the vertical sensitivity,
the time base, the gradient, the framing and the trigger.
The lowest frequency signal is used as the trigger source.
If no trace is detected on the inputs, the autoset will be aborted.
When pressed at the same time as a
corresponding channel as the trigger source. It initiates an autoset which will
take this selection into account. Channel CHx then becomes active for
adjustment using the keys:
" symbol is displayed in front of the settings display zone when
CHx key (CH1 or CH4), this defines the
III - 14 2-channel Oscilloscopes
Oscilloscope Mode - The Keys
Oscilloscope Mode (cont’d)
4 "Trigger" keys
Sets the trigger level to the average value of the signal (50%) without
modifying the trigger coupling.
When pressed in combination with a CHx key, this activates the same the
same function, after first selecting the corresponding channel as the trigger
source.
selects the trigger gradient (up or down ) by successive presses.
The gradient is indicated in the status area.
Successive presses can be used select one of the following acquisition modes:
Single shot (Mono) = SINGLE
Triggered Trig
Automatic (Auto) = REFRESH
•
"SINGLE" mode:
A single acquisition is authorized, triggered by pressing the RUN HOLD key.
For any further acquisition, the trigger circuit must be reset by pressing
the RUN HOLD key.
•
"TRIGGERED" mode:
The screen's content is only refreshed when there is a trigger event
linked to the signals present on the oscilloscope's inputs (CH1, CH4,
EXT).
If there is no trigger event linked to the signals present on the inputs (or
if there is no signal on the inputs), the trace is not refreshed.
•
"AUTOMATIC" mode:
The screen's content is refreshed even if the trigger level is not detected
on the signals present on the inputs.
When there is a trigger event, screen refreshing is managed as in the
"Triggered" mode.
• allows acquisition to be started and stopped in "TRIGGERED" and
"AUTOMATIC" modes.
• resets the trigger circuit in "SINGLE" mode.
Acquisition is initiated according to the conditions defined by the acquisition
mode (SGLE REFR key).
The acquisition status is indicated in the status area:
RUN = started
STOP = stopped
PRETRIG = acquisition
2-channel Oscilloscopes III - 15
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The Keys
3 "MEASURE" keys
activates or deactivates the display of the window for the 19 automatic
measurements on the reference trace.
When pressed in combination with a CHx key, it displays the
measurements concerning the corresponding channel.
By means of successive presses, this selects one of the displayed traces as
the reference trace for the automatic and manual measurements.
It appears in the "Measure" menu Reference.
Activates or deactivates the cursor display for manual measurements.
The cursors can be moved directly on the touch -sensitive pad using the
stylus.
•
The "dt" measurements (time difference between the two cursors)
and "dv" measurements (voltage difference between the 2 cursors) are
indicated in the status area.
•
The absolute value of the cursor selected is indicated in the "Current
Settings” area.
3 "HORIZONTAL" keys
or key pads
Adjustment of the time base coefficient (T/DIV).
After a Zoom, the "Z-Pos." setting modifies the position of the screen in
the acquisition memory.
Activates or deactivates the "Zoom" function.
By default, the zoom is performed around the samples located in the
middle of the screen.
A zone can be zoomed by tracing a rectangle around the area to be
enlarged using the stylus on the touch-sensitive pad. The sensitivity, time
base and horizontal and vertical alignment values are recalculated
automatically.
III - 16 2-channel Oscilloscopes
Oscilloscope Mode - The Keys
Oscilloscope Mode (cont'd)
Definition of terms
5 "VERTICAL" keys
or key pads
used
Validated channel: Display enabled, trace displayed after RUN
Displayed channel: Channel validated, trace present on the screen
Selected channel: The parameters of this channel can be set with the keys:
and
Before pressing one of the
following keys :
The channel concerned is
not displayed.
The channel concerned is
displayed, but not selected.
The channel concerned is
displayed and selected.
Step 1 Step 2 Step 3
Press
After pressing one of the
preceding keys:
The channel is displayed and
selected.
On CH1 and CH4, the vertical
sensitivity and the vertical
position are assigned to the
channel selected.
The channel is selected.
The channel is not selected
any longer.
A long press on one of the keys CHx causes a verticalautoset:
•
This modifies the sensitivity and vertical positioning of the channel in
question.
•
It optimizes the display on the screen by activating and selecting the
channel.
The channel is displayed and selected.
2-channel Oscilloscopes III - 17
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The Keys
5 "VERTICAL" keys
or key pads
(cont’d)
Activates or deactivates horizontal splitting of the display zone.
When activated, the "Full Trace" function is indicated by:
- the presence of a continuous horizontal line in the middle of the display
area
- horizontal splitting of the graticule.
After activation of the function:
- traces 1 and 3 are assigned to the upper part of the display,
- traces 2 and 4 are assigned to the lower part in order to prevent overlays.
The traces can then be moved vertically in the two zones.
This function can also be used in "full screen" mode.
Successive presses allow selection of the input coupling (AC, DC or GND)
for the last channel selected.
The coupling is indicated in the channel parameters area:
AC :
DC :
GND :
Adjustment of the vertical sensitivity of the last channel selected:
increases the vertical sensitivity
reduces the vertical sensitivity
Adjustment of the vertical position of the last channel selected:
moves it upwards
moves it downwards
III - 18 2-channel Oscilloscopes
Oscilloscope Mode - Display
4. Menu bar
Oscilloscope Mode (cont'd)
Display
Display
Composition
Cursor1
The oscilloscope display is divided into 4 functional zones.
Direct access to
main settings
3. Display area
2. Control area
1. Status area
Display and
adjustment of
current value
2-channel Oscilloscopes III - 19
Oscilloscope Mode (cont'd)
Trig 1 Pretrig
Position and movement of
manual cursors
T
1
2
Oscilloscope Mode - Display
1. Status area
Bargraph
Cursor measurements
Movement towards left of
screen in acquisition
Permanent settings
memory
Three types of general information appear in this area:
•
The bargraph representing the screen position, the trigger and the
cursors in the acquisition memory.
•
The instrument permanent settings.
•
The measurements, when the cursors are present on the screen.
(1)dt=110,0µs, div=100.0µV
Representation and
movement of the screen in
acquisition memory
Permanent settings
Each element in the bargraph can be moved with the stylus.
This zone refers to the trigger status (mode, edge, source, current status).
Position and movement of
time trigger
Movement towards right of
screen in acquisition
memory
Example: AUTO 1 STOP
When the stylus is placed on this information, the "Trigger Parameters"
menu can be opened by pressing twice.
Cursor
measurements
This zone refers either to:
•
the horizontal (dt) and vertical (dv) differences between 2 cursors in the
case of manual measurements
Example:
•
phase measurement in the case of manual phase measurement (Ph).
Example: (1) Ph = 200.0°
•
the automatic measurements selected using the "Automatic
Measurements" or "Phase measurement" menus
Example: (2) F = 1.0000 kHz, Vpp = 7,500 V
(1) dt = 110.0 µs, div = 100.0 µV
III - 20 2-channel Oscilloscopes
Oscilloscope Mode - Display
Display of ZOOM mode
Oscilloscope Mode (cont'd)
2. Control area
The parameters displayed in this area are:
•
The parameters of each channel and trace: display, sensitivity, coupling,
The time base value, the presence of a Zoom and a change in the signal
representation domain (FFT).
•
Active adjustment of the last selected element:
trigger level
trigger time position
channel offset value
X & Y position of cursor
• Time display, if no measurement has been selected.
Display of the trace parameters (in the
colour of the trace):
validity, coupling, bandwidth limitation,
sensitivity
Display of ZOOM mode
or
Display of the math function parameters
(in the colour of the trace):
validity, value of one division
Display of ZOOM mode
or
Display of memories (M): validity
Value of time base coefficient (s/div) in
oscilloscope mode or of the frequency
(Hz/div) in FFT mode
Contrast
•
The channels and functions can be validated using the stylus or the keys.
•
The "" symbol indicates whether a channel or function is selected, or
whether FFT mode is selected.
•
The settings of the time base (or the frequency) and the value of the active
parameter can be modified using the UP/DOWN button next to the display of
the current value.
•
After modification of the time base, the corresponding sampling frequency
is indicated in the settings area.
•
Double-tapping with the stylus on the parameters of a channel or on the
value of the time base, directly opens the associated menus:
- Sensitivity/Coupling and Vertical Scale, for the channels
- Vertical scale for the functions
-
Source, trigger mode and RUN/STOP, for the time base.
Change of signal representation domain
(FFT selection)
Indication and adjustment of last setting
selected
2-channel Oscilloscopes III - 21
Oscilloscope Mode (cont'd)
The grouped "Source" and "Trigger Mode" menus can be opened by a
double press with the stylus on the time base area.
RUN/STOP starts and stops acquisition from this menu. The acquisition
status is indicated in the status area on the screen.
The symbol "" indicates the source and trigger mode selected.
Oscilloscope Mode - Display
3. Display area
The graphic elements displayed associated with the traces in this area are:
•
a trigger time position indicator
•
a trigger level indicator
•
a trace number identifier
•
a vertical position indicator for the reference level of each trace
•
cursor position indicators linked to the trace for the automatic
automatic measurements
•
position indicators regarding the cursors linked or not to the trace
for manual measurements
•
selection of a zoom zone
III - 22 2-channel Oscilloscopes
Oscilloscope Mode - Display
9
8
1
3
4
5
2
1
2
6
6
Oscilloscope Mode (cont'd)
Display elements
12
>
10
ϕϕϕϕ
v
v
7
11
Definition of display
Refs. Elements selectable using the touch-sensitive pad
1
Tracedisplayed
2
3
4
5
6
7
8
9
10
11
Trace displayed
Vertical position indicator of the reference level of the trace
displayed and identification of the trace number
Indication of Trigger time position
Division of graticule
Position indicator of the cursors for the first automatic
measurement
Manual measurement cursor position indicator
Phase measurement cursor position indicator
Trigger level position indicator
Selection of a zoom zone
Indicator of trigger time position overshoot outside the display
window
Indicator of trigger level position overshoot outside the display
window.
12
Indicator of channel level overshoot outside the display
window.
2-channel Oscilloscopes III - 23
Oscilloscope Mode (cont'd)
Oscilloscope Mode - Display
Menu accessible
from display area
By double-tapping with the stylus in the
display area, the menu concerning the
display can be opened directly.
The "Full Screen" and "Zoom Out"
options are directly accessible
(see §. Display Menu).
It is possible to use the stylus to zoom in the display area by pulling a
rectangle.
After zooming in on part of the screen, the sensitivities of the traces and
the time base are recalculated.
•
The symbol "Z" appears in the signal and time base parameters
display.
•
The zoomed section is represented in the bargraph.
•
The "Zoom Out" menu (see §. Display Menu) can be used to return to
the original display, the zoom key de-activates the horizontal zoom.
•
The value of the horizontal zoom is adjusted to assign a calibrated
value to the horizontal scale (zoom factor: x 5 max.)
•
If the vertical selection of the zoom is greater than 6 divisions, no
vertical zoom is performed (zoom factor: x 16 max.).
All the symbols present in the display area:
- trigger indicators,
- trace position indicator,
- manual cursor position indicator,
- etc.
can be moved using the stylus.
The new modified symbol value is indicated in the current settings display
area.
The same applies to the selection of the
automatic and manual measurement
reference signal (see §. Measure Menu).
Calibration of the
touch-sensitive
screen
III - 24 2-channel Oscilloscopes
To optimize selection of the different elements present in the display area
using the stylus, calibration of the touch-sensitive screen may prove
necessary.
Select the "Touch Screen Calibration" option proposed in the display area
menu or in the Util menu.
Oscilloscope Mode - Display
Oscilloscope Mode (cont'd)
Calibration of the
touch-sensitive
screen (cont'd)
Corner of the touch screen
Follow the instructions on the screen
Use the stylus to point at the centre of the 4 patterns displayed on the
screen.
Validation of the input is indicated by modification of the pattern.
The pointing order is not important.
Once the 4 inputs have been recorded, validate the calibration with OK.
The touch-sensitive screen is calibrated and the display returns to normal
mode.
4. Menu bar
All the oscilloscope functions can be accessed via the main menus.
2-channel Oscilloscopes III - 25
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Vertical" Menu
The "Vert" Menu
(∗)
(∗)
15MHz
(∗) Functions only accessible in
"Advanced" mode.
See §. Description, page 58.
III - 26 2-channel Oscilloscopes
Oscilloscope Mode - The "Vertical" Menu
Oscilloscope Mode (cont'd)
Display
ch1 ch4
Sensitivity / Coupling
Channel Sensitivity
Coupling
Opens the "Trace display" menu for validating or devalidating the traces.
The "" symbol in front of a trace indicates that it has been validated.
The traces can be validated or devalidated from the control area by using the
stylus.
Modify the parameters of channels ch1 and ch4 independently, as well as the
vertical scale of the trace selected.
Modifies the parameters of the selected channel.
Modification of the channel's sensitivity using the stylus on the scrollbar,
adjustable by sequence: from 2.5 mV to 200 V/div.
The sensitivity is indicated in the channel parameter display area. It takes into
consideration the parameters of the "Vertical scale" menu.
Modification of AC - DC - GND coupling
AC: blocks the DC component of the input signal and attenuates the
signals below 10 Hz
DC: transmits AC and DC components of the input signal
GND: internally, the instrument links the input of the channel selected to a
0 V reference level
The "" symbol indicates the coupling selected. Coupling is indicated in the
modified channel's parameter display area.
bw limit
Vertical scale
Coefficient
Limits to 15 MHz the bandwidth of the channel and its trigger circuit to reduce
display noise and false triggering.
The bandwidth of each channel can be limited to 5 kHz, 1.5 MHz or
15 MHz. The bandwidth limit of a channel is indicated in the control area by
following symbols :
This menu can also be called up by double-pointing with the stylus on the
required channel parameter display area.
Defines the vertical scale of the channel selected on the basis of the current
settings. Readings of the direct measurements of the value analyzed and its
unit are provided.
Assignment of a multiplication coefficient to the selected channel's sensitivity.
This can be modified with the stylus, using the table of usable numbers, after
selecting the "Coefficient" zone.
The
2-channel Oscilloscopes III - 27
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Vertical" Menu
Measurement unit
math2 math3
Function definition
Init
Modification of the selected channel's vertical scale unit.
The modification is performed by means of the stylus, using the table of
usable characters after selecting the “measure unit” zone.
The key deletes the character preceding the cursor in this area.
The vertical scale unit will be indicated in the modified channel's parameter
display.
Reinitializes the multiplication coefficient to 1.00 and returns to the V
measurement unit.
Validation of the selections by "OK". Exit from the menu without modification
by "Cancel".
This menu can also be called up by double-pointing with the stylus on the
required channel's parameter display area (ch1 or ch4).
For the math2 math3 traces, definition of a mathematical function and the
vertical scale.
Menus present only in Advanced mode (see §. "Util" Menu).
Defines the mathematical function to be assigned to the selected trace by
means of a vertical keyboard, linking traces ch1and ch4, as well as M1 and
M4 if necessary.
Functions
8 predefined mathematical functions can be linked to the traces:
Validation of the selections by "OK". Exit from the menu without modification by
"Cancel".
If … then …
... the dynamic calculation of the
vertical scale is impossible
... the dynamic calculation of the vertical
scale is possible
Particular cases : Value of the measuring unit
... a message indicates that the
measuring unit on this function will
be vertical division (div).
... it takes into account of the
sensitivities of the channel sources.
Ch1 + Ch4 Sensitivity and measuring unit used on channel 1
Ch1 - Ch4 Sensitivity and measuring unit used on channel 1
In each cases, the measuring unit can be re-defined and a coefficient can be
applied to the measurement results (refer to §. Vertical scale).
Amplitude
Multiplying or dividing coefficients make it possible to optimize the
representation of the result :
Vertical scale math3 = 1.00 div
Vpp math3 = 12.00 div
Vhigh math3 = 6.00 div
There is a high and low overshoot, so division by 2 is necessary to
optimize display of the result.
- math3 = (ch1 + ch4) / 2 (Amplitude : / 2)
Vertical scale math3 = 2.00 div
Vpp math3 = 6.00 div = 12.00 V
Vhigh math3 = 3.00 div = 6.00 V
Division by two adjusts the addition to the dynamics of the screen.
Vamp (math3) = 6 vertical divisions
The measuring unit and the sensitivity of ch1 are used during the
display of measurements.
You can then open the menu “Vertical Scale" of math3 (see §.
of the menu from:
math2, math3,
p. 32) to assign a coefficient to the
Opening
result and to modify the measuring unit.
2-channel Oscilloscopes III - 29
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Vertical" Menu
Example 2
Multiplication
of two traces
As for the addition of traces, there is an even more significant high and low
overshoot. The overshoot is due to the increase in dynamics of the numbers
after a multiplication; to avoid this and standardize the result, the vertical
scale adapts. With an amplitude x 1, the product is included in the dynamics
of the screen. Moreover, the sensitivity of the math function is:
sensitivity
sensitivity
= sensitivity
math
= sensitivity
math3
x sensitivity
ch1
x sensitivity
ch1
ch4
ch4
x 4
x 4
= 5 x 5 x 4
= 100 VV
Vhigh
= Vhigh
math3
x Vhigh
ch1
ch4
= 10 x 10
=100 VV
(= 1 div)
Vamp(Ch1) = 4 vertical divisions
Vamp(Ch4) = 4 vertical divisions
math3 = (Ch1 * Ch4) (amplitude : x 1)
The result of the multiplication is translated into divisions on the screen.
III - 30 2-channel Oscilloscopes
Oscilloscope Mode - The "Vertical" Menu
Oscilloscope Mode (cont'd)
You can then open the menu “Vertical Scale" of math3 (see §.
the menu from: math2, math3
and to modify the measuring unit.
Opening of
, p. 32) to assign a coefficient to the result
Example 3
Division of two
traces
In our example
•
Then select math3 as the reference for the automatic and manual
(amplitude : x 1)
:
measurements (see §. "MEASURE" Menu).
•
Then display the table of 19 measurements made on the math3 trace
math3 (see §. "MEASURE" Menu).
The measurements displayed are the exact result of the multiplication of
The ch1 and ch4 traces are displayed on 4 vertical divisions.
math2 = ch1 / ch4
The sensitivity of the math2 function is : sens
math2
= sens
/ sens
ch1
ch4
= 5 / 5
= 1 V / V
Sensitivity
Vamp
math2
= 1 V/V
math2
= 1 V/V
2-channel Oscilloscopes III - 31
Oscilloscope Mode - The "Vertical" Menu
Oscilloscope Mode (cont'd)
Function definition
(cont’d)
Vertical scale Definition of the vertical scale for the selected trace
Opening of the menu
math2, math3
from:
Measurement unit
RESET
Coefficient Modifies the value of a division (div) in the selected trace.
cancels the selected function and resets the amplitude in x 1.
Calling this menu from math2 and math3 is identical to calling ch1 to ch4.
This can be modified with the stylus, using the table of usable numbers, after
selecting the "Coefficient" zone.
The key deletes the character preceding the cursor in this area.
The predefined values (x1, x10, x100, x1000) correspond to standard probe
coefficients and can be assigned directly.
The value of a division will be entered into the display of the modified trace
parameters.
Modification of the unit of the vertical scale (div) for the selected MATH
function.
This can be modified with the stylus, using the table of usable numbers, after
selecting the measurement unit zone.
The key deletes the character preceding the cursor in this area.
The "" key can be used to switch between upper-case and lower-case
characters.
III - 322-channel Oscilloscopes
Mode Oscilloscope - Le Menu « Déclenchement »
Mode Oscilloscope (suite)
Le Menu « Décl »
160ns
160ns
Oscilloscopes de table, 2 voies III - 33
Mode Oscilloscope (suite)
Mode Oscilloscope - Le Menu « Déclenchement »
Définition
Cette gamme d’oscilloscopes de table est pourvue de « déclenchements
évolués ».
L’onglet « Principal » permet de choisir et de paramétrer la source de
déclenchement principal.
La validation du choix de déclenchement se fait en quittant le menu.
Si … alors …
… l’utilisateur quitte depuis l’onglet
« Principal »,
… l’utilisateur quitte depuis l’onglet
« Pulse »,
etc. etc.
• Il n’existe qu’un seul Holdoff, bien qu’il soit programmable depuis les onglets
« Principal » et « TV ».
Le Holdoff s’applique à la source de déclenchement principal.
• Chaque source de déclenchement possède ses propres attributs :
Couplage, Niveau, Front, Rej Bruit.
… il se trouve en déclenchement
« Principal ».
… il se trouve en déclenchement
« Pulse ».
Le choix de
la voie de
déclenchement
peut être aussi
appelée en
double pointant
avec le stylet
dans la zone
d’affichage de la
base de temps.
Les paramètres de déclenchement
peuvent être aussi appelés en double
pointant avec le stylet dans la zone
d’affichage des paramètres de
déclenchement.
III - 34 Oscilloscopes de table, 2 voies
Mode Oscilloscope - Le Menu « Déclenchement »
Mode Oscilloscope (suite)
Paramètres
Principal
Source
Couplage
Front
Niveau
Sélection des « Paramètres de Déclenchement ».
Déclenchement sur front
Sélection d’une voie comme source de déclenchement :
CH1, CH4, Secteur, Externe, Externe /5.
Sélection du filtre de la source principale de déclenchement :
AC Couplage alternatif (10 Hz à 200 MHz) :
bloque la composante continue du signal.
DC Couplage continu (0 à 200 MHz) :
laisse passer tout le signal.
LF Reject Réjection des fréquences du signal source < 10 kHz :
facilite l’observation des signaux présentant une
composante continue ou une basse fréquence indésirable.
HF Reject Réjection des fréquences du signal source > 10 kHz : facilite l’observation des signaux présentant du bruit haute
fréquence.
Sélection de la pente de déclenchement :
+ pente de déclenchement ascendante
- pente de déclenchement descendante
La pente de déclenchement sélectionnée est reportée dans la zone d’état.
454mV Réglage du niveau de déclenchement avec le stylet en agissant sur
l’ascenseur.
Le niveau de déclenchement est reporté dans la zone d’affichage de la valeur
en cours, après modification. Il peut être ajusté finement.
Rejet bruit
Holdoff
Exemple
Non Hystérésis de ≈ ± 0,5 div.
Oui introduit un hystérésis de ≈ ± 1,5 div.
160ns permet :
• d’inhiber le déclenchement pendant une durée prédéfinie
• de stabiliser le déclenchement sur des trains d’impulsions.
Un double pointage sur ce champ fait apparaître à l’écran un clavier numérique
virtuel pour une saisie directe de la valeur.
Signal injecté sur CH1: un train de 3 impulsions à la fréquence de 20 kHz de
6 Vcc avec une composante continue de 500 mV, séparées de 500 µs.
Oscilloscopes de table, 2 voies III - 35
Mode Oscilloscope (suite)
Le déclenchement est réglé avec la voie 1 comme source, un niveau à 2,00 V,
sur un front montant.
Le Holdoff stabilise le signal en inhibant le déclenchement pendant 108 µs. Le
couplage DC du trigger laisse passer la totalité de signal.
Dans cet exemple, le signal n’étant pas bruité, l’option réjection de bruit n’est
pas nécessaire. Le couplage DC de Ch1 fait apparaître la composante continue
du signal.
Mode Oscilloscope - Le Menu « Déclenchement »
Pulse
Exemple
Sélection du déclenchement sur largeur d’impulsions.
<déclenche sur une impulsion, si sa durée est inférieure à la consigne
(le déclenchement effectif survient sur le front de fin de l’impulsion).
>déclenche sur une impulsion, si sa durée est supérieure à la consigne
(le déclenchement a lieu à l’issue de la durée définie, sans attendre la fin
de l’impulsion).
25.0µs Réglage avec le stylet sur l’ascenseur de la consigne de temps
Un double pointage sur ce champ fait apparaître à l’écran un clavier numérique
virtuel pour une saisie directe de la valeur.
Le choix du front ou dans l’onglet « Principal » définit les limites de
l’analyse :
• front définit une impulsion entre et
• front définit une impulsion entre et
III - 36 Oscilloscopes de table, 2 voies
Mode Oscilloscope - Le Menu « Déclenchement »
Mode Oscilloscope (suite)
TV
Standard
Polarité
Holdoff
Ligne
Déclenchement sur un signal TV. Ce menu n’est applicable qu’à l’entrée CH1.
Déclenchement sur un numéro de ligne spécifique. Le déclenchement part du
front avant du top de synchronisation ligne.
• 625 lignes (SECAM) ou
• 525 lignes (PAL)
+ Vidéo directe
- Vidéo inverse
Réglage avec le stylet en agissant sur l’ascenseur, inhibition du déclenchement
pendant une durée prédéfinie.
Un double pointage sur ce champ fait apparaître à l’écran un clavier numérique
virtuel pour une saisie directe de la valeur.
25Réglage du n° avec le stylet en agissant sur l’ascenseur
Un double pointage sur ce champ fait apparaître à l’écran un clavier numérique
virtuel pour une saisie directe de la valeur.
Les symboles « » et « » indique les paramètres sélectionnés. Validation
des sélections par « OK ».
Oscilloscopes de table, 2 voies III - 37
Mode Oscilloscope (suite)
Mode Oscilloscope - Le Menu « Déclenchement »
Mode déclenché
Mode
automatique
Mode monocoup
Acquisitions et rafraîchissement de l’écran à chaque événement de
déclenchement.
Acquisitions et rafraîchissement automatique de l’écran, même en l’absence
d’événement de déclenchement.
Traces visibles, même en l’absence d’événement de déclenchement.
Acquisition du signal et rafraîchissement de l’écran, au premier déclenchement
survenu après le réarmement du trigger par un appui sur la touche ci-contre (ou
par le menu de la base de temps).
•
Le symbole « » indique le mode de déclenchement sélectionné.
•
Le mode de déclenchement sélectionné est reporté dans la zone d’état
(Décl, Auto, Mono).
•
L’état d’acquisition est indiqué dans la zone d’état : PRETRIG, RUN,
STOP, POSTRIG, PRET, …
Cette sélection peut aussi être appelée en double pointant avec le stylet, dans
la zone d’affichage de la base de temps.
Mode ROLL
Si la base de temps est réglée à une valeur > 50 ms/div, l’activation du mode
MONOCOUP entraîne l’enclenchement du mode ROLL (défilement de la trace
de la droite vers la gauche de l’écran
En l’absence de déclenchement, la trace peut être figée en appuyant à
nouveau sur la touche RUN HOLD.
).
III - 38 Oscilloscopes de table, 2 voies
Oscilloscope Mode - The "Horizontal" Menu
Oscilloscope Mode (cont'd)
The "Horiz" Menu
(∗) Function accessible only in
"Advanced" mode.
See §. Description, page 58.
(∗)
(∗)
Repetitive Signal
Min/Max Acquisition
Increase in the time definition of a trace for a periodic signal. If this option
is valid, the signal can be averaged (see below).
Example: measurement on a microprocessor bus.
• For time bases of less than 100 µs/div. (without mode zoom
active), the signal displayed is reconstituted on the basis of
several acquisitions. The time resolution can reach 40 ps.
• If the signal is not repetitive, do not use this option. The
time resolution will then be ± 1 ns.
Menu present only in "Advanced" mode (see §. "Util" Menu).
The "" symbol indicates that the "Repetitive Signal" option has been
selected.
Sampling of the signal, even at slow time-base speeds. The display only
takes extreme value samples (min. and max.) into consideration.
This mode:
•
detects wrong representation due to under-sampling
•
Whatever time base is used, short-term events (Glitch, ≥ 2 ns) are
displayed.
displays short-term events (Glitch, ≥ 2 ns).
The "" symbol indicates that the "Min/Max Acquisition" mode is active.
2-channel Oscilloscopes III - 39
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Horizontal" Menu
Averaging
No averaging
Average rate 2
Average rate 4
Average rate 16
Average rate 64
FFT
(Fast Fourier
Transform)
Selection of a coefficient to calculate an average for the displayed samples.
For instance, this is a way of attenuating random noise observed in a signal.
For the averaging coefficient to be taken into account for representation of the
signal, the "Repetitive signal" option must be selected.
The calculation is performed using the following formula:
Pixel N = Sample*1/Average rate + Pixel
with:
Sample Value of new sample acquired at abscissa t
Pixel N Ordinate of pixel with abscissa t on the screen, at moment N
Pixel N-1 Ordinate of pixel with abscissa t on screen, at moment N-1
The "" symbol indicates the averaging coefficient selected.
(1-1/Average rate)
N-1
The Fast Fourier Transform (FFT) is used for calculating the discrete
representation of a signal in a frequency domain from its discrete
representation in the time domain.
FFT can be used in the following applications:
• measurement of the different harmonics and the distortion of a signal,
• analysis of a pulse response,
• search for noise source in logic circuits.
Description
The FFT is calculated over 2500 points.
The Fast Fourier Transform is selected by the FFT icon in the control area.
When the trace is zoomed, the FFT applies to the zoomed part of the trace.
The Fast Fourier Transform is calculated using the equation:
N
1
−
2
X (k) =
with: x (n): a sample in the time domain
X (k): a sample in the frequency domain
N: resolution of the FFT
n: time index
k: frequency index
The displayed trace represents the amplitude in V or dB of the various signal
frequency components according to the scale selected (linear or logarithmic).
The DC component of the signal is removed by software.
12
N
x nj
*( )*exp −
∑
N
n
=−
2
nk
π
for k∈ [0 (N – 1) ]
N
III - 40 2-channel Oscilloscopes
Oscilloscope Mode - The "Horizontal" Menu
scan
ning coefficient
2
Oscilloscope Mode (cont'd)
Horizontal unit
10 kHz square signal on CH2 and
5 Vpp
FFT with a Hanning window and a log
scale
FFT with a rectangular window and a
linear scale
This is indicated instead of the time base and is calculated according to the
scanning coefficient:
Unit (in Hz/div.) =
12.5
Vertical unit
Two possibilities are offered by the sub-menus:
a) Linear scale: by selecting the FFT menu and then the linear scale
• in V/div. =
unit of the signal in its time representation (V/div.)
b) Logarithmic scale: by selecting the FFT menu and then log scale (logarithmic)
• in dB/div. = by assigning 0 dB to a signal with an effective
amplitude of 1 in the time representation
The vertical position indicator of the representation is at -40 dB.
Graphic
representation
2-channel Oscilloscopes III - 41
The FFT representation indicates symmetry in relation to the frequency
origin; only positive frequencies are displayed.
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Horizontal" Menu
Rectangular
Hamming
Hanning
Blackman
The sub-menus select a type of window.
Before calculating the FFT, the oscilloscope weights the signal to be
analyzed by means of a window acting as a band-pass filter. The choice
of window type is essential to distinguish between the various lines of a
signal and to make accurate measurements.
Time representation
of signal to be
analyzed
Weighting window
Weighted signal
Frequency
representation of
signal calculated by
FFT
III - 42 2-channel Oscilloscopes
Oscilloscope Mode - The "Horizontal" Menu
Oscilloscope Mode (cont'd)
The finite duration of the study interval results in a convolution in the signal
frequency domain with a function sinx/x.
This convolution modifies the graphic representation of the FFT because
of the lateral lobes characteristic of the sinx/x function (unless the study
interval contains an whole number of periods).
Four types of window selections are available: the menus appear directly
on selection of the FFT menu.
Effects of under-sampling on frequency representation:
If the sampling frequency is not correctly adjusted (less than or twice the
maximum frequency of the signal to be measured), the high-frequency
components will be under-sampled and appear in the graphic
representation of the FFT by symmetry (aliasing).
•
The "Autoset" function is active. This prevents the phenomenon above
and adapts the horizontal scale: the representation is more legible.
•
The "Zoom" function is active.
The "" symbol in front of one of the options indicates the function
selected.
2-channel Oscilloscopes III - 43
Oscilloscope Mode - The "Display" Menu
Oscilloscope Mode (cont'd)
The "Display" Menu
(∗) Function only accessible
in "Advanced" mode.
See §. Description, page 58.
(∗)
Grid
Display modes
Envelope
Zoom off
Vector
Display / Removal of graticule
There are two display modes available: vector and envelope.
A vector is plotted on the basis of each sample.
The minimum and maximum observed on each horizontal position of the
screen are displayed. This mode is used, for example, to view a time or
amplitude variation or a modulation.
The “” symbol indicates the active display mode.
Returns to the original screen size after zooming in on part of the screen.
•
This function is inactive unless the screen is in zoom mode.
•
If the Zoom mode is active, the letter "Z" is displayed in the trace
and time-base parameter display area.
This menu can also be called up by double-pointing with the stylus inside
the trace display area.
III - 44 2-channel Oscilloscopes
Oscilloscope Mode - The "Display" Menu
Oscilloscope Mode (cont'd)
Full screen
Oscilloscope
XY
Switches from the normal display mode to the “full screen” display mode
and vice versa.
The display is organized so as to leave the biggest surface area possible
for trace plotting: only the permanent settings and the automatic or
manual measurements remain
• This function has the same effect as the key.
• The “” symbol indicates that the full screen mode is active.
This function can also be called up by double-pointing with the stylus on
the trace display area.
The settings defined on the front panel remain active.
The following sub-menus can be used to switch from oscilloscope mode
to XY mode.
The “” symbol indicates the active mode.
This is the basic operating mode.
The “XY source” menu is used for assigning the desired traces to the
X axis (horizontal) and Y axis (vertical).
Validation of the selections by "OK". Exit from the menu without modification
by "Cancel".
• Each axis is graduated into 8 divisions.
Example
• The selected traces are identified by a figure corresponding to their
axis.
• The “” symbol indicates the trace selected for each axis.
Two sinusoidal signals assigned to the X and Y axis with an offset of π/2
are then represented by a circle.
2-channel Oscilloscopes III - 45
Oscilloscope Mode - The "Measure" Menu
)
Oscilloscope Mode (cont'd)
The "Measure" Menu
(∗
Reference
Automatic
measurements
Trace 1
Trace 2
Trace 3
Trace 4
Selection of the trace on which the automatic or manual measurements
will be performed.
Only the active traces can be selected. Inactive traces are displayed in a
lighter colour.
The "" symbol indicates the reference trace.
Opens the "Automatic measurements" menu window.
The measurements are performed and refreshed on the selected
reference trace. All the measurements that can be performed on this
trace are displayed.
(- . - -) is displayed for measurements that cannot be performed.
(∗) Function only accessible
in "Advanced" mode.
See §. Description, page 58.
III - 46 2-channel Oscilloscopes
The window can be validated and closed by pointing to OK with the
stylus.
The selected measurement(s) will be displayed in the status area.
Oscilloscope Mode - The "Measure" Menu
Oscilloscope Mode (cont'd)
• It is possible to select two permanent measurements.
• The "" symbol indicates the measurement(s) that will be indicated
in the status area.
• Activation of the automatic measurements causes two markers (+) to
appear on the curve, if at least one period is visible on the screen.
• The display order corresponds to the chronological order of the
selection and the markers are assigned to the first measurement
selected.
Automatic measurements in the status area can be deleted by means of
this menu, by erasing the selected measurements (no "" symbol in the
automatic measurements table).
Reference memory
difference
Example
19 automatic
Activation of the "Reference memory difference" option is a way of
calculating the deviations, for all the automatic measurements, between
the selected trace and the memorized reference trace (see §. Memory
Menu).
Calculation performed and displayed on one of the 19 automatic
measurements:
Vpp (Ref. memory difference) = Vpp (Trace 1) – Vpp (Trace 1 Ref 1)
The calculation is performed in the same way for all the measurements.
•
This option is only active if a reference trace is present.
It must correspond to the trace on which you wish to perform
automatic measurements
( Example: Trace 1 and Trace 1 Ref. 1).
•
Condition: the reference trace must have the same characteristics as
the associated trace (sensitivity and time base)
2-channel Oscilloscopes III - 47
Oscilloscope Mode - The "Measure" Menu
Oscilloscope Mode (cont'd)
Measurement
conditions
•
The measurements are performed on the displayed part of1.31968(e)1.31968( )-4.77687(d)1.3196870l.32232109196870l.32232109196870l.32232109196870l.322327432ETQ0
Automatic measurement of a trace's phase compared with a reference trace
(See §. Reference Measurement).
This menu selects the trace on which phase measurements are to be
performed.
To deactivate phase measurement, deselect the selected phase
measurement using the same menu.
•
The "" symbol indicates the trace selected for phase measurement.
•
Activation of the phase measurement, if it is possible, will cause display of 3
markers:
2 markers for the reference trace period 1 marker indicated as ϕ on the trace for which the phase
measurements will be performed.
These 3 markers are positioned automatically; they cannot be moved.
•
The phase measurement (in °) of the trace selected compared with the
reference trace is indicated in the measurement display status area
Example: (1)Ph (2) = 180.0°
•
If the measurement cannot be performed, " - . - -" is displayed.
Cursor measurements on the reference signal
The measurement cursors (1 and 2) are displayed as soon as the menu is
activated.
The two measurements made are:
dt (time difference between the two cursors),
dv (voltage difference between the two cursors).
The measurements performed and the displayed cursors are linked to the
selected reference trace (see §. Reference Measurement).
•
The "" symbol indicates that the manual measurements (dt, dv) are active.
•
the measurement cursors can be moved directly with the stylus.
They can also be moved with the stylus by selecting the 1 (cursor 1)
or 2 (cursor 2) in the bargraph.
•
If the free cursor option is not active (see §. "Unattached Cursors"
Measurement), the cursors will remain linked to the reference trace during
movements. If the option is active, the cursors can be moved
anywhere on the screen.
•
The dt and dv measurements in relation to the selected reference
are indicated in the measurement display status area.
Example: (1)dt = 500.0 µs, dv = 1.000 V
Manual phase
measurement
2-channel Oscilloscopes III - 49
Phase measurements using 3 cursors:
- Use cursors 1 and 2 to indicate the period of the reference signal.
- Use the ϕ cursor to measure the phase.
•
The "" symbol indicates that manual phase measurement is active.
•
When this menu is active, the 3 cursors are present if at least one
signal is active.
•
The cursor marked ϕ can be moved freely, even if the "Unattached
cursors" menu is not active.
•
The phase measurement (in °) between the cursors is indicated in the
measurement display status area.
Example: (1)Ph = 120.0 °
Oscilloscope Mode - The "Measure" Menu
Oscilloscope Mode (cont'd)
Unattached manual
cursors
Used for linking or not linking the manual measurement cursors (1 and
2) to the reference trace.
When the "Unattached cursors" menu is selected, cursors 1 and 2 can
be moved freely over the whole screen.
• The "
• To deactivate this menu, deselect it by pointing with the stylus.
In the case of "Automatic measurements" and manual measurement
activation:
"
symbol indicates that the "Unattached cursors" menu is active.
If …
… the manual cursors and
the automatic markers are
displayed together,
… the portion defined
between the manual cursors
is too small [in this case, the
fixed markers (+) will not be
displayed],
then …
… the automatic measurements
are performed on the portion of the
trace defined between the manual
cursors.
… the automatic measurements
selected are impossible, so "-.--" is
indicated in the measurement
display area.
Deselect the automatic measurements to validate the manual
measurements (dt, dv).
Storage of the selected trace in its reference memory
( E.g.: Trace 1 in Ref. 1).
The 4 traces have their reference memory.
A reference memory is volatile, which means that it is lost when the
instrument is switched off.
•
For optimum use, the reference trace must have the same
characteristics as the associated trace (sensitivity and time base).
•
A trace can only be saved in its reference memory if it is present
on the screen
•
The memorized traces are displayed in a light colour, accompanied by
their reference number.
•
The "" symbol in the menu means that the corresponding trace has been
saved in the reference memory and that it is present on the screen.
•
A reference trace cannot be moved.
•
A reference memory can be deactivated by deselecting it in the menu.
2-channel Oscilloscopes III - 51
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Memory" Menu
Trace
Save .TRC
Save.TXT
Saving (to the non-volatile memory) or recall of a trace or a reference
memory. The back-up can be saved in two formats: ".TRC" or ".TXT".
The "File copy" menu is adapted to the type of format selected.
Saving of the files for subsequent recall on the oscilloscope screen
The back-up files will take the suffix .TRC; they can be recalled in the
"Trace Recall" menu.
Saving of files for export to another application
The saved files have the suffix .TXT; they cannot be recalled by the
"Trace Recall" menu for screen display. However, they can be exported
in a standard format for use in other software (spreadsheet - e.g.
Microsoft EXCEL) using the menu "Util Files Export".
The selection made opens a "File Copy" menu.
∗ Then in the "Source" drop-down menu, select the trace or the
reference memory to be recorded.
The trace or reference memory to be saved will appear in grey. The stylus
is used for selection.
•
Only traces and reference memories present on the screen are
indicated in the "Source" list (selectable).
•
If all the traces and reference memories are present on the screen,
the right scroll bar in the menu can be used to move around inside
the list.
∗ A default backup file name is proposed above the keyboard. It can be
modified using the virtual keyboard with the stylus.
The key deletes the character preceding the cursor in this area.
∗ Once the name has been written, the key records it by entering it
into the destination menu and closes the menu. The backup file takes
the extension .TRC (internal format) or .TXT (text format), depending
on the previous selection.
You can exit from this menu without saving by pointing with the stylus
on the icon in the top right-hand corner of the window.
• The file name is limited to a maximum of 15 characters + the suffix. If
this rule is not respected, the message 'File name too long'
is displayed. .
• As soon as the pointer passes over a destination file, the name is
displayed with its recording date
its recording time
its size.
• If the name already exists or is incompatible, an error message
‘Impossible! File already exists' will be displayed.
III - 52 2-channel Oscilloscopes
Oscilloscope Mode - The "Memory" Menu
The name of the file selected for recall appears in grey. The selection will be
Oscilloscope Mode (cont'd)
Recall .TRC
When selected, this opens a "File Copy" menu.
In the "Source" list, the .TRC files recorded previously (via the menu "Trace
Save .TRC") are displayed.
The name of the file selected for recall appears in grey. The stylus is used for
selection.
∗
After selecting the file to be recalled, the "destination" list indicates on which
trace you want to recall it. The destination trace selected (1 to 4) by pointing
with the stylus appears in grey. It is also indicated at the bottom of the screen.
∗
Once the trace to be recalled and its destination have been selected, the
key is used to execute the operation and close the menu.
You can exit from this menu without recalling by pointing with the stylus on the
icon in the top right-hand corner.
•
If the destination trace selected is already present on the screen, it
will be replaced by the recalled trace.
•
When a trace is recalled, "Mx" is displayed in the destination trace
parameters.
•
In this menu, the virtual keyboard cannot be used.
Configuration
Save
Recall
Saving or recall of an instrument configuration.
When selected, this opens the "File copy" menu.
∗
In the "Source" menu, there is a file called "Configuration".
It contains the parameters of the instrument's configuration when this menu
is opened.
∗
A back-up file name is proposed above the Qwerty keyboard.
The key deletes the character preceding the cursor in this area.
∗
When the source file name has been entered, the key is used to record
the configuration by transferring it into the destination menu, before closing
the menu (save file: .CFG suffix).
You can exit from this menu without saving by pointing with the stylus on the
icon in the top right-hand corner of the window.
•
The file name is limited to a maximum of 15 characters + the suffix. For a
source file, the name is accompanied by the recording date and time and
the file's size, as soon as the pointer passes over its name.
•
If the name already exists or is incompatible, an error message will appear.
When selected, this opens the "File copy" menu.
∗
In the "Source" list, the .CFG files recorded previously (via the menu
"Configuration Save") are displayed.
made using the left mouse button. The scrollbar on the right is used for
moving through the list.
∗
With the source file selected, the key is used to perform the recall.
∗
You can exit from this menu without recalling by pointing with the stylus on
the top right-hand corner of the window.
•
In this menu, the virtual keyboard cannot be used.
•
Use the "default config" file to restore the factory configuration.
2-channel Oscilloscopes III - 53
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Utilities" Menu
The "Util" Menu
III - 54 2-channel Oscilloscopes
Oscilloscope Mode - The "Utilities" Menu
Oscilloscope Mode (cont'd)
Files
File type
Selection of the "File Management" menu.
It contains the files which have been:
•
recorded since the instrument was first used
•
created since the last startup. These files will only be saved definitively when
the instrument is switched off using the key shown opposite.
If there is a mains power cut during saving of the configuration, the files
in the file manager will be lost.
The configuration back-up files (.CFG),
for traces (.TRC),
for sampling (.TXT),
for printing (.PRN, .PCL, .EPS, .BMP, .GIF)
are accompanied by the date and time when they were saved and their size.
The selected file appears in grey. The stylus is used to select the file. The
scrollbar on the right is used for moving through the list.
Selection of the file type required, using the corresponding filter:
.CFG .TRC .PRN .PCL .EPS .BMP .GIF .REC .TXT
The stylus is used for selection.
" . ∗∗∗∗ " can be used to select all the file types.
The storage capacity of the file manager is 1 Mbyte.
Erase*.*
I/O port config.
Stop bits
Protocol
Open
Erase
Export
RS 232
Speed
Format
Parity
causes restoration of the file selected by the “File Copy” menu.
deletes the selected file.
deletes all the files the extension of which has been selected in “File type”.
transmits the file over the active communication interface
(RS232/CENTRONICS), network.
You can exit from this menu by pointing with the mouse on the icon in the top
right-hand corner of the window.
Selection of the "RS232" or "Network" menus.
This menu can be used to configure the "serial" remote programming interface
or the "network" interface (ETHERNET).
This interface uses the connector (RS232/ETHERNET) on the right-hand side
of the instrument. It requires the use of the RS232 / SUBD9 cable (HX0042).
selects the transmission speed : 300 to 115,200 Baud.
selects the word length: 7 or 8 bits.
selects the type of parity: even, odd or no parity (none).
selects the number of stop bits (1 or 2 stop bits).
selects the serial link management mode.
Hard Hardware: the protocol is provided by the RTS and CTS lines
of the RS232 link.
Soft Software: use of the XON and XOFF characters to synchronize transmission and reception of the messages (reduced "3-wire" link)
None No protocol checking
The "" symbol indicates the selected option.
•
•
The option can be modified using the stylus.
2-channel Oscilloscopes III - 55
ted. In this case, the "LPD Server" software needs to be installed on the
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Utilities" Menu
Network
supplied by a DHCP
server
Physical address
IP address
Configuration of the ETHERNET parameters
If this box is notched, the instrument makes a request towards a DHCP server
to obtain automatically
• an address IP,
• the value of the subnet-mask,
• the address of the gateway,
• possibly the IP address of a network printer (according to the
configuration of the DHCP server).
corresponds to the address of the oscilloscope on the ETHERNET network.
This address cannot be modified (it is specific to the instrument).
Example: 00-01-02-03-04-63
corresponds to the address of the oscilloscope on the ETHERNET network.
This address can be input automatically or manually with the keyboard, after
selecting the zone to be modified.
The key can be used to delete the value preceding the cursor in the zone to
be modified.
An IP address can be assigned automatically by a DHCP server, if the server
is accessible, by ticking the box "provided by a DHCP server".
Example: 132.147.200.74
Printer or LPD server:
IP address
Name
corresponds to the IP address of the printer, or of a PC to which the printer is
connec
PC.
This address must be input manually with the keyboard, after selecting the
zone to be modified.
The key can be used to delete the value preceding the cursor in the zone to
be modified.
Example: 132.147.240.1
Name of the printer as it appears in the printing server (or PC). If the printer is
connected directly to the network, do not enter anything here.
This key is accessible in the "Advanced" mode only.
It gives access to the manual programming of the Subnet mask and to the
programming of the IP address of the gateway.
Validation of the selections by "OK". Exit from the menu without modification
by "Cancel".
III - 56 2-channel Oscilloscopes
Oscilloscope Mode - The "Utilities" Menu
". Exit from the menu without modification
Oscilloscope Mode (cont'd)
Hardcopy
The printing format, the type of printer and the communication port are chosen
in this menu.
The printer type or selected format will appear in reverse video. The stylus is
used for selection. The scrollbar on the right is used for moving through the list
of types or the printer languages.
Configuration
Date/time
Option
Port
Choice of colour or black/white printing.
Selection of the interface used for print data transfer: RS232C, CENTRONICS
(using the adapter provided as an option) or in a file.
•
If the RS232C interface is selected, the parameters (speed, format, parity,
stop bit, protocol) must be configured in the "Config I/O Ports" menu. Check
that the configuration matches the configuration of the peripheral
device connected to the instrument.
•
If the "Network" option is selected, the parameters (IP address, printer) must
be configured in the « Config I/O Ports Network" menu).
•
The "File" option is a way of recording the hardcopy in a file.
".bmp" and ".gif" image formats can be used directly in the Windows
applications (word processing, presentations, etc.). As soon as the printout is
activated, the "File copy" menu opens and you have to enter the name of the
file generated (see "Trace Save" menu).
A hard copy of the screen can be printed by pressing this key. The hard copy is
produced using the parameters defined in the "Hardcopy" menu.
Updating of the date (day, month, year) and the time (hour, minute, second).
You can select the required parameter by using the stylus and the scrollbars
located on either side of the parameters to be adjusted.
The clock starts when the menu is closed.
Language
Screen s
2-channel Oscilloscopes III - 57
aver
Standby
Selection of the language in which the menus are written.
Options available: French, English, German, Italian, Spanish, etc…
Sets the screen to standby after a defined period of time to minimize the
consumption of the equipment and screen ageing.
4 options are available: 15 min, 30 min, 1 hour, no standby mode.
The screen can be reactivated by pressing any key on the front panel.
Shutdown of the instrument after a predefined period, in order to limit its
energy consumption.
In this case, the equipment configuration is saved before the shutdown.
4 options are available: 30 min, 1 hour, 4 hour, 24 hour.
The equipment is reactivated using the key shown opposite.
• The "
• The option can be modified using the stylus.
Validation of the selections by "OK
by "Cancel".
" symbol indicates the option selected.
n
Oscilloscope Mode (cont'd)
Oscilloscope Mode - The "Utilities" Menu
System Info
Next calibration date
Touch-screen
calibration
"Advanced" Mode
Menus present i
"Advanced" mode only
Display of data concerning the operation of the instrument since it was first
used.
Periodic equipment calibration is necessary to guarantee the specifications
announced.
For verification of the instrument, see §. Maintenance.
You can exit from this menu by pointing with the stylus on the icon in the top
right-hand corner.
See pages 24 and 25.
When this "Advanced" mode is not active, the advanced equipment
functions no longer appear in the menus.
math2, math3 accessible via the "Vert" menu
Averaging, repetitive signal accessible via the "Horiz" menu
XY accessible via the "Display" menu
Unattached manual cursors accessible via the "Measure" menu
•
The "" symbol indicates that the "Advanced" mode is active.
•
The stylus can be used to modify this.
•
By default, the "Advanced" mode is not active.
•
In "non-advanced" mode, the instrument's configuration is not
saved when it is shut down using the key opposite and the default
factory configuration is loaded at start-up.
III - 58 2-channel Oscilloscopes
Oscilloscope Mode - The « Help » Menu
regarding the key pressed.
Oscilloscope Mode (cont'd)
The « ? » Menu
OX6102 Vx.00/x00
Help
About
When selected with the stylus, this opens the "Help" menu.
The online help concerns the instrument's keyboard.
Use the and keys of the window to scroll through the description
of the keys on the front panel.
Whenever a keyboard key is pressed, online help will be displayed
The functions associated with the keys will not be activated.
The key name is indicated above the explanation.
You can exit from this menu by pointing with the stylus in the top right-
hand corner of the window.
Provides information on:
- the name of the instrument
- the version of the software
- the version of the hardware
You can exit from this menu by choosing OK.
2-channel Oscilloscopes III - 59
Multimeter Mode - The Keys
Multimeter Mode
The Keys
4 "UTILITY"keys or
key pads
1 "AUTOSET" key
Press the key opposite to select the "Multimeter" mode, OX 6062 and
OX 6152 only.
Direct access to LCD contrast adjustment.
No action.
Triggers a hardcopy in accordance with the configuration chosen in the "Util"
and "Hardcopy" menus.
A second press before the end of the process will interrupt the current printout.
If printing is impossible, a "Printing error" message will be sent.
The " " symbol is displayed in front of the settings display zone when
printing is in progress.
No action.
No action.
(OX 6062 and OX 6152 only)
Selective
"AUTOSET"
+
4 "TRIGGER" keys
No action.
No action.
No action.
No action.
To lock / unlock the display of the measurements. Plotting of the traces never
stops.
IV - 60 2-channel Oscilloscopes
Multimeter Mode - The Keys
Multimeter Mode (cont'd)
3 "MEASURE" keys
3 "HORIZONTAL" keys
or key pads
5 "VERTICAL" keys
or key pads
No action.
For changing the reference trace to which the cursor refers (successive presses).
No action.
Duration of the recording in the display window: 5’, 15’, 30’, 1h, 6h, 12h, 24h,
week, month.
No action.
No action: zoom not possible in this mode.
The instrument contains as many independent multimeters as there are
channels in "Oscilloscope" mode (2 or 4).
Same function as in "Oscilloscope" mode.
A long press enables or disables the autorange on the channel concerned.
The channel is displayed and selected.
No action.
If a channel is activated and selected, this key can be used to change the input
coupling of the channel. By successive presses, the coupling can be changed
from AC to DC to AC+DC.
The coupling is indicated in the multimeter window of the channel concerned.
When selected for ohmmeter, continuity or capacimeter measurements,
component testing or temperature measurements on channel 1, the key has
no effect, as the input coupling in these functions cannot be adjusted.
Manual modification of the measurement range
If the autorange mode is activated, the manual adjustment has no effect: the
autorange will automatically reset the instrument to the measurement range
best suited to the input signal.
No action.
2-channel Oscilloscopes IV - 61
BarGraph
Multimeter Mode (cont'd)
Multimeter Mode - Display
Display
Display
Composition
The multimeter display is divided into 6 functional areas:
1. Menu bar
1. Menu bar
2. Channel 1 Multimeter
3. Cursor
value
5. Graphic window
channel 1
4. BarGraph
2. Channel 4 Multimeter 6. Current
setting
The tool bar gives access to the different menus of the "Multimeter"
function.
4.
channel 4
IV - 62 2-channel Oscilloscopes
Multimeter Mode - Display
Multimeter Mode (cont'd)
2. Channel (x)
multimeter)
Coupling
Autorange
Symbols
Channel
There is a display area reserved for each of the instrument's channels. In each
of these, the following information is indicated:
CH1 or CH4
The input coupling selected (see §. Vert. Menu) is indicated in this field.
For the Ohmmeter, Capacimeter, Continuity, Component Test and
Temperature Measurement modes, the coupling is not shown.
The symbol « » shows the selected bandwidth limit.
Indicates whether range changing is automatic.
A symbol is displayed correspon0 0 1 176.04 486.44 Tm[(34.15818(i)-)-2.99246(h)-1.1409d644091( 3aR7 15.a314091(d).a3147W11.14091(s)-5)-1.14091(r)3.263110910085( )5.13
2-channel Oscilloscopes IV - 63
ast
point
Multimeter Mode (cont'd)
Multimeter Mode - Display
4. Bargraph
5. Graphic window
These graphs show the min. and max. values measured on the channels in
the range during the observation period.
The bargraph is shown with the colour of the channel.
The zero level of the bargraph and the scale are adapted to suit the type of
measurement and the range.
If the range is changed, the bargraph is reinitialized and the curve showing the
evolution of the measurement is erased.
This window shows the evolution of the measurements as a function of time.
The most recent measurement points are those on the right-hand side of the
screen.
Depth of the
display
window
The depth of the window, which represents the observation period, can be
programmed using the keys opposite.
In this way, you can choose to view the measurements with a depth of:
5’, 15’, 30’, 1h, 6h, 12h, 24h, week, month.
acquired
Cursor
L
6. Current settings
When ROLL mode is activated (see §. HORIZ Menu), the measurements are
updated constantly, with the most recent results taking the place of the older
results, which disappear.
The window has the same function as the oscilloscope window: it serves to
indicate and adjust the value of the parameter selected.
IV - 64 2-channel Oscilloscopes
Multimeter Mode - The Menus
Multimeter Mode (cont'd)
The Menus
Presentation
• Screen display when measurements are possible on all the channels:
Example: CH1 is configured as an Ohmmeter, so CH4 is necessarily
set to amplitude measurement (AC, DC or AC + DC)
The screen display is identical if CH1 is configured for amplitude
measurement, as an ohmmeter or for temperature measurement
(PT100).
• Screen display when measurements are possible on CH1 only:
Example: CH1 is configured for a Component Test
The display is identical when CH1 is configured for capacitance
or continuity measurement.
Measurements on channel 4 are impossible.
2-channel Oscilloscopes IV - 65
Multimeter Mode - The "Vertical" Menu
For the ranges available for each type of measurement, refer to the technical
The Range / Coupling menu is displayed in light grey when the range is not
Multimeter Mode (cont'd)
The "Vert" Menu
ch1 ch4
Range/Coupling
Range
Modification of:
•
the parameters of channels ch1 and ch4, independently
•
the vertical scale of the selected trace
Modifies the parameters of the selected channel.
Choice of the measurement range. The unit displayed depends on:
• the type of measurement selected: amplitude (available on all
channels), ohmmeter, continuity, capacimeter, PT100 temperature
probe (available only on channel 1, see §. Measure Menu),
specifications of the "Multimeter" function.
modifiable (single range).
IV - 66 2-channel Oscilloscopes
Multimeter Mode - The "Vertical" Menu
Multimeter Mode (cont'd)
Autorange
Bandwidth limit
Coupling
When this option is selected, the measurement range changes
automatically.
The "" symbol shows that it is active.
If the option is disabled, the range can be modified manually, using the
keys opposite or the "Range" menu, depending on the type of
measurement being performed.
Modification of the AC, DC or AC+DC coupling for amplitude
measurement.
AC: AC voltage measurement
DC: DC voltage measurement
AC + DC: Measurement of AC voltage with a DC component
For AC and AC+DC measurements, the menu "Display Frequency" can
be used to display the frequency of the signal in the secondary display
field.
The "" symbol indicates the type of coupling selected.
The coupling is updated in the modified channel parameter display zone.
If the channel measures an AC or AC+DC voltage (see Coupling), it is
possible to filter the signal with a low-pass analogical filter whose cut-off
frequency is 5 kHz.
Vertical scale
Id. “Oscilloscope” mode.
2-channel Oscilloscopes IV - 67
Multimeter Mode (cont'd)
Multimeter Mode - The "Trigger", "Horizontal" and "Display" Menus
The "Trig" Menu
The "Horiz" Menu
Roll
The "Display" Menu
No trigger parameterization is possible in this mode.
•
If this mode is activated (presence of "" symbol), the measurement history curve is constructed continuously. The oldest points disappear on the
left-hand side of the screen, while the most recent ones appear on the right.
•
If this mode is deactivated, the point display stops as soon as the
first point acquired reaches the left-hand edge of the window. However,
the measurements continue and are still refreshed in the area 'Multimeter
Channel’x'.
Frequency
Statistics
Relative mode
This menu selects the information that you want to display in the secondary
measurement field of each 'Multimeter Channel’x' area.
If this line is ticked and in the case of AC amplitude measurement, the
frequency of the signal measured (if possible and consistent) will be displayed.
Display of the Min. and Max. values of the measurements performed.
The value displayed is the difference between the value measured and the
value displayed when this option was selected.
The "" symbol indicates the secondary function selected.
IV - 68 2-channel Oscilloscopes
Multimeter Mode - The "Measure" Menu
ch1: PT100
Multimeter Mode (cont'd)
The "Measure" Menu
ch1: Amplitude
ch1: Ohmmeter
ch1: Continuity
ch1: Capacimeter
Reference
The reference is used to select the measurement trace on which the cursor is
positioned. The cursor value is therefore relative to the measurement on this
channel.
It is only possible to choose the reference on activated channels: channels
that are not activated are displayed in a lighter colour in the sub-menu.
The "
Channel CH1 is used as a voltmeter and therefore measures the amplitude
of the signal present on the input of this channel.
Channel CH1 is used as an ohmmeter and therefore measures the
resistance of the dipole wired to the input.
Channel CH1 is used as a continuity tester: there is a beep if the input
resistance is less than ≈ 30 Ohms.
In this mode, measurements are impossible on the ch4 channel.
Channel CH1 is used as a capacitance meter and therefore measures the
capacitance of the dipole wired to the channel's input.
In this mode, measurements are impossible on the ch4 channel.
" symbol indicates the reference selected.
ch1: Component Test
2-channel Oscilloscopes IV - 69
Channel CH1 is used as a diode tester. This mode measures the threshold of
the diode wired to the input.
In this mode, measurements are impossible on the ch4 channel.
These configurations measure a temperature using a 100 Ω (PT100) resistive
sensor.
The range of maximum temperatures is - 200°C… + 850 °C and depends on
the sensor used.
Multimeter Mode (cont'd)
Multimeter Mode - The "Memory", "Utilities" and "Help" Menus
The "Memory" Menu
Trace (.TXT)
Configuration
The"Util" Menu
Configuration
Screen saver
Setup
In "Multimeter" mode, it is only possible to save a trace in non-volatile memory
in .TXT format.
Files saved with the suffix .TXT can be exported onto a PC (see §. Util Menu
Files) for processing with other software (spreadsheet, etc.).
This function is identical to the one in "Oscilloscope" mode.
This function is identical to the one in "Oscilloscope" mode, excepted :
•
If the recording time is over 5 minutes, the screen saver will never be
activated.
•
If the recording time is equal to 5 minutes, the screen saver and settings
operate as in “Oscilloscope” mode, p. 57.
Standby
The "?" Menu
IV - 70 2-channel Oscilloscopes
•
If the recording time is over 5 minutes, standby will never be activated.
•
If the recording time is equal to 5 minutes, standby and settings operate as in
“Oscilloscope” mode, p. 57.
This menu is identical to the one in "Oscilloscope" mode, p. 59.
Applications
Applications
Display of the
calibration signal
•
Using the key shown opposite, select the "Oscilloscope" mode.
• Connect a 1/10-ratio probe to input CH1.
A message indicating the characteristics of the probe is displayed briefly,
confirming that it has been detected.
• Validate the signal:
Vert Menu Display Trace 1
or by pressing the CH1 key
or on the display of the CH1 trace parameters.
• Adjust the CH1 sensitivity:
Vert Menu CH1 Sensitivity/coupling: 500 mV/div. (1/10 probe)
or using the keys opposite.
• Adjust the CH1 coupling:
Vert Menu CH1 Sensitivity/coupling AC
or by pressing the AC/DC GND key.
• Adjust the scan speed:
using the scrollbar in the time base window: 500 µs/div
or the keys opposite.
• Select the trigger parameters:
Trig. Menu Parameter Principal Source: CH1
Coupling: AC
Front: + (or using the key opposite).
• Adjust the trigger mode:
Trig. Menu Automatic mode
or using the SGLE REFR key.
• Use the RUN HOLD key to start acquisition ("RUN" mode)
Probe compensation
If necessary:
• Modify the trigger level using the stylus:
- by moving the symbol T (Trigger) on the screen. The value of the trigger
level is indicated in the bottom right-hand corner of the screen.
- or through the trigger parameters menu:
Trig. Menu Parameter Principal Level
• Modify the vertical positioning of the curve:
- by using the stylus to move symbol 1, on the left of the screen.
- or by using the keys opposite.
The key opposite is used to make these adjustments automatically.
Adjust the low-frequency compensation of the probe so that the signal plateau
Measure Manual measurements (dt, dv) (see §. Measure Menu).
∗
Two measurement cursors (1 and 2) are displayed as soon as the
menu has been activated.
∗
The 2 measurements indicated under the trace display are
dt (interval between the 2 cursors as a function of the time base) and
dv (voltage between the 2 cursors as a function of the vert. sensitivity).
Example: (1)dt : 2,150 ms, dv = 250.0 mV
•
The two measurement cursors (1 and 2 can be moved directly on the
screen by means of the stylus.
In the same way, they can also be moved horizontally by means of
the stylus, by selecting the 1 (cursor 1) or the 2 (cursor 2) in the
bargraph of the status zone.
•
If the unattached cursors option is not activated (see §. Measure Menu Unattached cursors), the cursors remain linked to the trace
during the moves.
•
If the unattached cursors option is active, the cursors can be moved
anywhere on the screen.
•
Initially, there must be 2 out-of-phase signals on 2 channels (ch1 and ch2).
Automatic phase
measurement
Reminder
•
Select the reference trace in relation to which you want to perform
the phase measurements via the menu:
Measure Reference Trace 1 to Trace 4 (see §. Reference).
Example: Reference Measurement Trace 1.
•
Select automatic phase measurement via the menu:
Measure Phase measurements (see §. Phase measurement).
Example: Phase Measurement Phase Trace 2.
∗
The 2 cursors (+) of the automatic measurements are displayed on
the reference trace. A cursor "ϕ" is displayed on the trace concerned
by the phase measurements.
∗
The phase measurement (in °) is indicated under the display of
the curves.
Example: (1)Ph (2) = 180.0°
•
The 3 cursors are fixed; they cannot be moved.
•
If it is not possible to perform the measurement , "-.- -" appears.
2-channel Oscilloscopes V - 73
Manual measurement
Applications (cont.)
•
Select manual phase measurement via the menu:
of phase
Measure Manual phase measurement (see §. Measure Menu).
∗
the reference trace. A "ϕ" cursor in relation to which the phase
is measured is displayed.
∗
curve(s).
Applications
The 2 cursors (1 and 2) of the manual measurements are displayed on
The phase measurement (in °) is indicated under the display of the
Example: (1)Ph = 150.0°
Reminder
Display of a TV video
signal
•
The 3 measurement cursors are present if at least one trace is present
on the screen.
•
The 3 measurement cursors can be moved directly on the screen
by using the stylus. They can also be moved using the stylus by
selecting 1 (cursor 1) or 2 (cursor 2) in the bargraph in the
status zone.
•
If the unattached cursors option is not activated (see §. Unattached
cursors), the cursors (1 and 2) will remain linked to the trace when moved.
If the option is active, these cursors can be moved as required on the
screen.
In all cases, the symbol "ϕ" can be moved freely.
This example illustrates the TV synchronization functions.
- The parameters in the TV menu (Trigger Parameters menu) for
displaying a TV signal only apply to the CH1 input.
- it is recommended to use a 75 Ω adapter for observing a video signal.
• Inject on CH1 a composite TV signal with the following characteristics:
- 625 lines
- positive modulation
- vertical grey scale stripes
• In the Trigger Parameters menu, select the tab:
Trig. Menu Parameter TV:
• Set the standard number of lines: 625 lines
- polarity: +
- line: 25 (for a video signal)
- front: + (or using the key opposite).
• Adjust the CH1 coupling:
Vert Menu CH1 Sensitivity/coupling DC
or by pressing the AC/DC GND key.
• Adjust the CH1 sensitivity:
Vert Menu CH1 Sensitivity/coupling 50 Mv/div.
or using the keys opposite.
• Adjust the scan speed:
using the scrollbar in the time base window: 20 µs/div.
or using the keys opposite.
V - 74 2-channel Oscilloscopes
Applications
Applications (cont.)
• Select the trigger mode:
Trig. Menu Automatic mode
or using the SGLE REFR key.
• Use the RUN HOLD key to start acquisition (RUN mode)
or use the time base menu.
Reminder
Use the manual cursors to check the duration of a line
The acquisition status (Ready, RUN, STOP) is indicated on the right, under
the display of the trace, in the trigger status display zone.
• Optimize the time base speed to observe several complete TV lines.
Example of a video signal
• Display the manual cursors:
Menu Measure Manual measurements (dt, dv)
or using the key opposite.
The measurements between the 2 cursors are indicated under the curve
• To move the cursors freely, select:
Measure Menu Unattached cursors.
• Use the stylus to position cursors 1 and 2 on the start and end of the
signal, respectively.
display.
Example: dt = 64.00 µs = duration of a line
2-channel Oscilloscopes V - 75
Applications (cont.)
Applications
Examination of a
specific TV line
For more detailed examination of a video line signal, the TV trigger menu can
be used to select a line number.
• In the Trigger Parameters menu, select the tab:
Trig. Menu Parameter TV:
• Set the standard number of lines: 625 lines
polarity: +
line: 1
front: + (or using the key opposite).
• Adjust the ch1 sensitivity:
Vert Menu CH1 Sensitivity/coupling 100 mV/div
or using the keys opposite.
• Modify the scan speed:
using the scrollbar in the time base window: 20 µs/div
or using the keys opposite.
Example of video line 1
V - 76 2-channel Oscilloscopes
Applications
Applications (cont.)
Display of slow
phenomena “ROLL”
mode
This example examines the analysis of slow phenomena for time bases
ranging from 200 ms to 200 s. The samples are displayed constantly, without
waiting for the Trigger (ROLL mode).
Examination of slow phenomena over a significant period of time
•
Select "Oscilloscope" mode (key opposite).
•
Inject on the CH1 input a sinusoidal signal at 1 Hz and 1 Vrms.
•
Adjust the scan speed: using the scrollbar in the time base window:
500 ms/div
or using the keys opposite.
•
Validate signal CH 1:
Vert Menu Display Trace 1
or by pressing the CH1 key
or on the display of the CH1 trace parameters.
•
Adjust the CH1 sensitivity:
Vert Menu CH1 Sensitivity/coupling 500 mV/div (1/10 probe)
or using the keys opposite.
•
Adjust the CH1 coupling:
Vert Menu CH1 Sensitivity/coupling DC
or by pressing the AC/DC GND key.
Examination of the
signal
•
Select the trigger parameters:
Trig. Menu Parameter Principal Source CH1
Coupling: AC
Front: + or using the key opposite.
•
Adjust the trigger mode:
Trig. Menu Single mode
or using the SGLE REFR key.
Use the stylus to move the Trigger level symbol up or down in the display
area:
Trigger level is < the signal level the oscilloscope stops data acquisition
once it has filled the acquisition memory (STOP mode).
- Trigger level > Signal level data acquisition no longer stops and the signal
is analyzed constantly.
The trigger level may be set precisely, using the trigger parameters menu:
Trig. Menu Parameter Principal Level.
•
Start acquisition using the RUN/HOLD key (RUN mode).
This function is recommended
for studying low-frequency signals.
2-channel Oscilloscopes V - 77
Applications (cont.)
Applications
Min / Max
Acquisition
Detection of a wrong representation due to under-sampling
Examination of an amplitude-modulate signal
• Using the key shown opposite, select the "Oscilloscope" mode.
• Inject on the CH1 input an amplitude-modulated sinusoidal signal at a.
frequency of 15 Hz.
• Inject on the CH4 input a sinusoidal signal at 300 Hz and 3 VDC. for synchronization of the signal on CH1.
• Adjust the scan speed:
using the scrollbar in the time base window: 1 ms/div
or using the keys opposite.
• Validate the CH1 signal: Vert Menu Display Trace 1
or using the key opposite.or on the display of the CH1 trace parameters.
• Adjust the CH1 sensitivity:
Vert Menu CH1 Sensitivity/coupling: 200 mV/div or using the keys opposite.
• Ditto for CH4 with a sensitivity of: 500 mV/div
• Adjust the trigger parameters:
Trig. Menu Parameter Principal:
Source: CH4
Coupling: AC
Front: + or using the key opposite.
• Adjust the trigger mode:
Trig. Menu Auto Mode:
or using the SGLE REFR key.
• Start acquisition using the RUN/HOLD key (RUN mode).
• Use the FULL TRACE key to display the 2 traces separately.
Examination of the signals
V - 78 2-channel Oscilloscopes
Applications
Applications (cont.)
The observation of the amplitude-modulated CH1 trace cannot be used (incorrect
display).
• Validate the MIN / MAX mode:
Horiz MIN/MAX Acquisition menu, to view the amplitude modulation of the
CH1 signal.
Examination of the signals
2-channel Oscilloscopes V - 79
Applications (cont.)
Applications
Measurement in
multimeter mode,
OX 6062 and
OX 6152 only
Measurement in
Multimeter Mode
• Press the key opposite to activate the "Multimeter" mode.
Caution : measurement ground is connected to earth ground.
Ohmmeter mode (Ω) is indicated in the display of the parameters for
CH1.
• Validate the measurements on CH1 by pressing the key opposite (- X - disappears).
The resistance measured is not known:
• Select the "Autorange" mode.
"Channel 1 Parameters" menu under Vert CH1 Range/Coupling
or by a long press on the CH1 key.
The autorange mode (auto) is indicated in the display of the parameters
for CH1.
In this case, the instrument constantly seeks the most suitable
measurement range.
The resistance measured is known:
• Select the appropriate range:
"Channel 1 Parameters" menu under Vert CH1 Range/Coupling
or using the keys opposite.
See the general characteristics for the ranges available.
V - 80 2-channel Oscilloscopes
• Select the "Statistics" mode.
Menu: Display Statistics, to find out the minimum and
maximum values when the measurement variations are analyzed.
The MIN and MAX measurements are indicated in the display of the
parameters for CH1.
• Use the keys opposite to select the duration of the period (5’, 15’, 30’, 1 h, 6 h, 12 h, 24 h, week, month) for analysis of the
measurement variations.
The duration selected is indicated in the top left-hand corner of the
graphic window.
• Deactivate "Roll" mode:
Menu: Horiz Roll to stop the measurements at the end of the
analysis period.
Applications
Applications (cont.)
measurement in
Multimeter Mode
Example of
The graphic window records the changes in the measurements during the
analysis period (5 minutes).
The bargraph shows the amplitude of the variations.
The minimum value (49.9 ohms) and the maximum value (550.5 Ohms)
measured are indicated in the CH1 parameter display.
The cursor linked to the trace indicates the measurement a specific point on
the trace, along with the time of the event.
The value measured which is displayed in the CH1 parameter display
remains active after the analysis period has ended.
2-channel Oscilloscopes V - 81
Applications (cont.)
Applications
ETHERNET
network application
examples
a) File transfer from a
PC via the network
The files in the oscilloscope's "File Management" menu (see §. "Util" Menu)
can be downloaded onto a PC (or uploaded from a PC) via an ETHERNET
network.
IP Address = 132.147.200.74
Oscilloscope
IP Address = 132.147.240.1
PC
ETHERNET
network
•
Use a suitable ETHERNET cable to link the oscilloscope to the network.
•
Open the oscilloscope's "Network" menu.
•
Enter the IP address manually or automatically using the icon
a DHCP server" (if the server is accessible).
•
Then validate the information by choosing OK.
"provided by
•
Example:
132.147.
200.74
•
Use a PC connected to this network.
•
In your browser, type in the URL zone: ftp://132.147.200.74
A list of the files is then displayed.
•
You can use your browser to:
- copy files (PC Scope or Scope PC),
- delete files,
- rename files.
The SX-METRO software (option) simplifies file transfer via the
ETHERNET network.
V - 82 2-channel Oscilloscopes
Applications
Applications (cont.)
b) Hard copy of the
screen on a network
printer
Screen copying can be initiated on a network printer.
ETHERNET
•
•
•
•
•
IP Address = 132.147.200.74
network
Use a suitable ETHERNET cable to link the oscilloscope to the network.
Open the oscilloscope's "Network" menu.
Enter the IP address manually or automatically using the icon
by a DHCP server" (if the server is accessible).
Example:132.147.200.74
Enter the IP address of the network printer using the table of usable
numbers after selecting the zone to be modified.
Example:132.147.200.74
Specify the name of the required printer ( Example: LaserJet 4)
Oscilloscope
IP Address = 132.147.200.10
Printer
"provided
To find out the
contact the network administrator responsible for your IT installation.
•
Then validate the information by choosing OK.
•
Open the "Hardcopy" menu (See §. “Util” Menu) of the oscilloscope.
•
According to the printer connected to the network, select the print format or
type of printer.
•
Validate the colour or black/white option.
•
Validate the Network option of the port menu.
•
Configure the oscilloscope so that it displays the screen as you wish to print it.
Start the required print operation by pressing the key opposite.
IP address of the server or the name of the printer,
2-channel Oscilloscopes V - 83
Applications (cont.)
Applications
WEB Server
Connection to WEB
SERVER
Minimum PC Configuration: Pentium II, 200 MHz, 64 Mb. RAM.
Screen resolution: > 1152 x 864 pixels
Install JVM SUN (minimum version
J2RE 1.4.2) from site //java.sun.com
Recommended browsers: Internet Explorer 6.0 or Netscape 6.0
Screens obtained on PC logged on to same network as the instrument.
In the examples that follow, the IP address of the oscilloscope is: 14.3.250.46
(programmed in the UTIL menu Config I/O Port Network).
Connection is done from a PC by using INTERNET EXPLORER.
To reach the Web-server, register in the bar of address:
'http://IP Address’ of the instrument.
The IP address of the instrument is defined in the following menu:
UTIL Config I/O Port Network. See p. 56.
IP Address of
instrument:
see p. 56.
V - 84 2-channel Oscilloscopes
Applications
One click on
title triggers
Applications (cont.)
"OSCILLOSCOPE"
mode
IP Address of
instrument:
see p. 56
window title
triggers
refreshing
Window
adjustments
integrated
after a
click on
"Remote
Control
".
"FFT" mode
One click
on window
refreshing
Window
adjustments
integrated
after a
click on
"Remote
Control".
.
2-channel Oscilloscopes V - 85
virtual disk of
used in the
Applications (cont.)
"MULTIMETER"
mode,
OX 6062 and
OX 6152 only
Applications
One click on
window title
triggers
refreshing
Window
adjustments
integrated
after a
click on
"Remote
Control".
.
"UTILITIES"
File name
request
(hardcopy
or save
trace)
Display of
FTP
window,
indicating
files on
instrument.
See screen
below.
The selection button
determines the action:
- launching of screen copy or
- save trace.
V - 86 2-channel Oscilloscopes
Technical Specifications
3 dB on range 2,5 mV/div.
Technical Specifications
« Oscilloscope » Function
Vertical deviation
Characteristics OX 6062 OX 6152 OX 6202
Number of channels
Vertical ranges
BW at - 3dB on all vertical ranges
from 5 mV to 100 V/div.
Max. input voltage
Input Type
Vertical offset dynamic
Input coupling
Bandwidth limit
400 VDC, 300 Vrms, 450 Vpk (DC + peak AC at 1 kHz) without 1/10 probe
AC: 10 Hz to 60 MHz
DC: 0 to 60 MHz
GND: reference
Only the values assigned with a tolerance or limits are guaranteed
values (after half an hour warming up). The given specifications
apply in the reference area. Values without a tolerance are
for information only.
2 channels: CH1 & CH4
Variation in steps (no continuous variable coefficient)
60 MHz 150 MHz 200 MHz
60 MHz 100 MHz 150 MHz BW at -
Measured on load 50 Ω with a signal of amplitude 6 div.
for ranges 2,5 mV/div to 5 V/div. 1
derating -20 dB/decade from 100 kHz to 150 MHz
metallic BNC connected to ground
± 10 divisions on all ranges
AC: 10 Hz to 150 MHz
DC: 0 to 150 MHz
GND: reference
15 MHz, 1.5 MHz, 5 kHz
2.5 mV to 100 V/div.
AC: 10 Hz to 200 MHz
DC: 0 to 200 MHz
GND: reference
Rise time2
Cross-talk between channels
Peak-to-peak gain accuracy2
Vertical resolution of the display
DC vertical measurement accuracy2
with offset and
with averaging of 16
AC vertical measurement accuracy
without offset in 1 kHz and
with averaging of 16
Accuracy2 of vertical offset
with averaging of 16
Resolution of the measurements
Vertical ZOOM function on acquired
or saved curve
Electrical safety (accessories not)
Input impedance
Display modes
≈ 5.8 ns ≈ 2.34 ns
DC at 60 MHz > 40 dB
2
DC at 150 MHz ≥ 40 dB
Same sensitivity on both channels
± 2 % (with averaging of 4) at 1 kHz
± 0.4 % of full scale (without ZOOM)
0.1 % in ZOOM mode (10 bits)
± [2,2 % (reading) + 11 % (sensitivity) + 250 µV]
applies to : Vmin, Vmax, Vlow, Vhigh, Vavg, curs(1), curs(2)
± [2 % (reading) + 10 % (sensitivity) + 250 µV]
applies to : Vamp, Vrms, Overshoot +, Overshoot -
± [0.2 % (reading) + 10 % (sensitivity) + 250 µV]
10 bits
ZOOM factors: 16 max.
300 V, CAT II, cl. 1
1 MΩ ± 1 % approx. 15 pF 1 MΩ ± 1 % approx.12 pF
ch1, ch4
≈ 1.75 ns
DC at 200 MHz ≥ 40 dB
1
Beyond 5 V/div. range the bandwidth is guaranteed by the respect of the rising time.
2
Specification given for 5 mV/div. to 100 V/div. ranges. Reduced specification for 2.5 mV/div. range.
VI - 88 2-channel Oscilloscopes
Technical Specifications
Technical Specifications (cont'd)
« Oscilloscope » Function
Horizontal deflection (time base)
Characteristics OX 6062 OX 6152 OX 6202
Time base ranges
Time base accuracy
Sampling rate
Time measurement accuracyHorizontal ZOOM
XY ModePhase error
Bandwidth in X and Y
Representation
Fast Fourier Transform
35 ranges, from 1 ns to 200 s/div.
± 0.1 %
1 GS/sec. in real time
50 GS/sec. with repetitive signal
The horizontal ZOOM factors available range from x 1 to x 10 acc. to seq. 1-2-5
(in ZOOM mode, the same sequence of time base ranges is used as in normal
The smallest time increment between two successive XY points is given
XY mode representation therefore depends on the selected time-base range.
temporal or frequential (FFT)
•
calculation on the traces present in the screen area
•
dynamic refreshment as a function of the signal observed in RUN mode
•
windowing: rectangle, Hamming, Hanning, Blackman
•
scales: logarithmic or linear
•
automatic adjustment with autoset function
± [(0.02 div.) x (time/div.) + 0.01 x reading + 1 ns]
N.B.: The oscilloscope has a memory capacity
for recording of 2,500 counts per channel.
Display at screen is 500 counts
for 10 divisions.
The bandwidth in X and Y is identical
In XY mode, at each instant t:
by the real acquisition frequency of the oscilloscope.
60 MHz
150 MHz
mode).
< 3°
200 MHz
2-channel Oscilloscopes VI - 89
Technical Specifications
Technical Specifications (cont'd)
« Oscilloscope » Function
Trigger circuit
Characteristics OX 6062 OX 6152 OX 6202
Trigger sources
Trigger mode
Trigger sensitivity
Trigger coupling
without bandwidth limit
Trigger gradient
Noise rejection
Trigger level
Variation range
Trigger type
Pre-Trigger
CH1, CH4, EXT, line
Automatic
Triggered
Single shot
Auto Level 50 %
DC : 0.6 div. from DC at 10 MHz
1.5 div. from 10 MHz at 60 MHz
AC : 10 Hz at 60 MHz
HF Reject : 0 at 10 kHz
BF Reject : 10 kHz at 60 MHz
on pulse width < t > t from 20 ns to 20 s
- Selection of line number and polarity, with 525 lines (PAL) and 625 lines (SECAM),
DC : 0.6 div. from DC at 10 MHz
1.5 div. from 10 MHz at 150 MHz
AC : 10 Hz at 150 MHz
HF Reject : 0 at 10 kHz
BF Reject : 10 kHz at 150 MHz
Falling or rising
≈ ± 1,5 div.
± 8 div.
on edge
TV on CH1 only:
even or odd line field
- TV trigger sensitivity: > 1 div.
Adjustable from 0 to 100 %
DC : 0.6 div. from DC at 10 MHz
1.5 div. from 10 MHz at 200 MHz
AC : 10 Hz at 200 MHz
HF Reject : 0 at 10 kHz
BF Reject : 10 kHz at 200 MHz
HOLDOFF
TRIG-EXT
Trigger sensitivity
Trigger level
Variation range
Input impedance Safety (access. not)
EXT: 100 mV p. to p. from DC to 1 MHz
150 mV p. to p. from 1 MHz to 100 MHz
EXT / 5 : 500 mV p. to p. from DC to 1 MHz
750 mV p. to p. from 1 MHz to 100 MHz
EXT : ± 800 mV
EXT / 5 : ± 4 V
Adjustable from 160 ns to 30 sec.
1 MΩ± 1%
300 V Cat. II - Class 1
Trigger in mode LINE is carried out by collecting the ambient electric field at 50 Hz
(or 60 Hz).
For a correct operation, the ambient electric field should not be disturbed by signals of raised
level, close to 50-60 Hz frequency.
VI - 90 2-channel Oscilloscopes
Technical Specifications
Technical Specifications (cont'd)
« Oscilloscope » Function
Acquisition chain
Characteristics
ADC Resolution
Maximum sampling rate
Sampling modes:
10 bits
1 GS/sec. in real time
Specifications Comments
1 converter per channel
2-channel Oscilloscopes VI - 91
Technical Specifications
Technical Specifications (cont'd)
« Oscilloscope » Function
Processing of measurements
Mathematical functions
Automatic
measurements
Resolution of the measurements
Measurements by cursors or automatic
measurements
DC vertical measurement accuracy 3
with offset and averaging of 16
Equation editor (functions on the channels or simulated)
Addition, subtraction, multiplication, division and complex
functions between channels.
Time measurements
up time
down time
positive pulse
negative pulse
cyclic ratio
period
frequency
phase
counting
summation
10 bits / display on 4 digits
± [2,2 % (reading) + 11 % (sensitivity) + 250 µV]
applies to : Vmin., Vmax., Vlow, Vhigh, Vavg., curs(1), curs(2)
Level measurements
DC voltage
RMS voltage
peak-to-peak voltage
amplitude
max. voltage
min voltage
high plateau
low plateau
overshoot
AC vertical measurement accuracy 3
without offset in 1 kHz and averaging of 16
Vertical offset accuracy 3
with averaging of 16
Accuracy of 2-cursor time measurements
The cursors are attached to the trace, but they can be detached
to perform a measurement between channels (offset, delay ...)
In XY mode, the cursors are not attached to the trace.
1/4 VGA, i.e. 320 pixels horizontally x 240 pixels vertically
Delay can be selected in the Tool Menu Configuration
15 ‘, 30’, 1h or none
Complete memory: 2500
500 counts out of the 2500 of the whole memory
Acquired points, interpolated points, averaging
The acquired points are attached by a segment.
Min. and max. on each horizontal screen position are displayed.
Range of factors: none, 2, 4, 16, 64
Graticule
Indications on screen
Triggering
Traces
Miscellaneous
1/10 probe calibration signal
Autoset
Search time
Frequency range
Amplitude range
Cyclic ratio limits
Complete Borders
Trigger level position (with overshoot indicator)
Position of the Trigger point on the bargraph and on the top edge of the
screen (with overshoot indicators)
Trace identifiers, activation of the traces
Position, Sensitivity
Ground reference
High and low overshoot indicators if traces are off screen
Form: rectangular
Amplitude: ≈ 0 - 3 V
Frequency: ≈ 1 kHz
< 5 s
> 30 Hz
15 mVpp to 400 Vpp
20 to 80 %
All the versions of this instrument are equipped with the clock and calendar functions.
Only the values assigned with a tolerance or limits are
guaranteed values (after half an hour warming up). The
percentage of measuring accuracies refers to the measured
value. Values without a tolerance are for information only.
Resistance measurement On Channel 1
Ranges (end of scale)Ohmmeter
2-channel Oscilloscopes VI - 95
Technical Specifications
Technical Specifications (cont'd)
« Multimeter » Function
40 Hz to 198 kHz on a square and sinus signal
Accuracy : 0.1 %
Multimeter operating modes
Selection of modes
Relative mode
Monitoring (statistics)
Frequency
The modes
Digital display rate
Measurement log
HOLD
Display
In numeric form
Graphic trace
RUN
via menu, with touch-sensitive screen
MAX and MIN values for all measurements
Possible display of frequency in AC mode
Relative
Monitoring
Frequency are exclusive
2 measurements per second
Measurement display = f (time)
default window of 5 min (adjustable from 5 min to 31 days)
Initiation of the measurements
Freezing of the measurement
Principal measurement large-size display
Secondary measurement small-size display
The touch-sensitive screen allows you to select the secondary measurement via
a menu.
History of the measurements over time
Objective: Presentation of the measurements as an amplitude histogram.
Warning !
Error Messages
VI - 96 2-channel Oscilloscopes
If one of those codes (or the addition of several codes) is present when getting
started : a default has been detected.
In this case, contact your closest distributor (See §. Maintenance).
Autotest : Error n° 0001 : Micro Problem
Autotest : Error n° 0002 : Flash Problem
Autotest : Error n° 0004 : RAM Problem
Autotest : Error n° 0008 : FPGA Problem
Autotest : Error n° 0010 : Numerisation problem on channel 1
Autotest : Error n° 0080 : Numerisation problem on channel 4
Autotest : Error n° 0100 : Analog problem on channel 1
Autotest : Error n° 0800 : Analog problem on channel 4
Autotest : Error n° 1000 : Problem on ETHERNET link
General Specifications - Mechanical Specifications
General specifications
•
Environment
Power supply
Reference temperature 18°C to 28°C
•
Operating temperature 0°C to 40°C
•
Storage temperature -20°C to +60°C
•
Utilization indoors
•
Altitude < 2000 m
•
Relative humidity < 80 % up to 31°C
•
External power supply
Mains voltage 100 V to 240 V ± 10 %
Frequency from 47 to 63 Hz
Consumption < 20 A
Screen saver
(automatic stand-by) adjustable by menu : 15’, 30’, 1h, none
Automatic stop adjustable by menu : 30’, 1h, 4h, 24h
Safety
EMC
As per IEC 61010-1 (2001):
•
Isolation class 1, earth grounded
•
Pollution level 2
•
"Measurement” input overvoltage category CAT II, 300 V
This instrument conforms the EMC NF EN 61326-1, 07/97+A1, 10/98 norm :
• Emission A class instrument A single spectral line corresponding to the
clock frequency of the processor (60 to
75 MHz) might exceed the requirement
by less than 10 dBµV/m.
• Immunity Influence magnitude:
2 div. in the presence of a 3 V/m
electromagnetic field (class B)
3 div. in the presence of a 10 V/m
electromagnetic field. (class A)
Mechanical Specifications
•
Casing
Dimensions 215 mm x 190 mm x 225 mm
•
Weight 1,2 kg
Packaging
2-channel Oscilloscopes VII - 97
•
Dimensions 345 mm x 275 mm x 200 mm
Supply
Supply
Accessories
delivered with the
instrument
delivered optionally
• User’s manual on CD-ROM
• Remote programming on CD-ROM
• 2 1/10 probes, according to model..........................(OX 6062) HX0210
(OX 6152, OX 6202) HX0220
• Serial link RS232.....................................................................HX0042