Firmware update : You may use PC software provided on the CD-ROM and consult Internet site
www.chauvin-arnoux.com. The software update involves a rebootstrrapping of the filesystem : all
Two-channel digital oscilloscopes, 60 MHz or 100 MHz I - 3
the saved files are erased and lost.
General instructions
General Instructions
Introduction
Precautions and
safety measures
definition of
measurement
categories
You have just acquired a two-channel digital oscilloscope with “SPO”
analog persistence display, 60 MHz or 100 MHz, ETHERNET.
It can also features a « harmonic analyser » (option) mode and a
« recorder » mode (option).
Congratulations for your choice and thank you for your trust in the quality of
our products.
This instrument conforms to safety standard NF EN 61010-1 (2001), single
insulation, relative to electronic measurement instruments.
To obtain optimum service, read these instructions with care and comply with
the precautions for use.
Failure to comply with these warnings and/or user instructions is liable to
cause damage to the equipment. This could be dangerous to the user.
•
This instrument has been designed for use:
- indoors,
- in a pollution degree 2 environment,
- at an altitude of less than 2000 m,
- at a temperature included between 0°C and 40°C
- with relative humidity of less than 80 % at up to 31°C.
•
It can be used for measurements on circuits at 300 V CAT II, relative to ground
and can be supplied by a 240 V CAT II network.
CAT I : Measurement category I is for measurements performed on circuits not
directly connected to mains.
E.g.: protected electronic circuits
CAT II : Measurement category II is for measurements performed on circuits
directly connected to the low voltage installation.
E.g.: power supply to domestic appliances and portable tools.
CAT III : Measurement category III is for measurements performed in the building
installation
E.g.: machine or industrial apparatus power supply.
CAT IV : Measurement category IV is for measurements performed at the source
of the low-voltage installation.
E.g.: energy inputs
before use
during
use
I - 4 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
• Comply with environment storage conditions.
• Make sure that the three-wire phase/neutral/ground power supply cord
supplied with the unit is in suitable condition. It conforms to standard NF
EN 61010-1 (2001) and must be connected to the instrument on the one
hand, and to the network on the other (variation from 100 to 240 V
• Read carefully all the notes preceded by the symbol .
• Connect the instrument to an outlet with a ground pin.
• The instrument power supply has automatically reset electric protection
operating after the fault has been eliminated.
• Be sure not to obstruct the aeration points.
• As a safety measure, use only suitable cords and accessories supplied with
the instrument or type approved by the manufacturer.
• When the instrument is connected to the measurement circuits, never
touch an unused terminal.
AC
).
General Instructions
General Instructions (cont’d)
Symbols
used
Guarantee
Warning: Risk of danger.
Refer to the operating manual to find out the nature of the potential hazards and
the action necessary to avoid such hazards.
Earth
This equipment is guaranteed against any material defect or manufacturing
faults, in conformity with the general conditions of sale.
During this period, the equipment may only be repaired by the manufacturer.
He reserves the right to carry out repair or replacement of all or part of the
equipment.
If the equipment is returned to the manufacturer, forward transport is at the
expense of the customer.
The guarantee does not apply in the event of:
•
unsuitable use of the equipment or by association with incompatible
equipment
•
modification of the equipment without the explicit authorization of the
manufacturer technical services
•
operation by a person not approved by the manufacturer
•
adaptation to a specific application not provided for in the equipment
definition or in the operating instructions
•
impact, fall or flooding.
According to WEEE directive 2002/96/EC
Maintenance and
metrological
checking
Unpacking and
repacking
Servicing
Before the equipment is opened, it must be disconnected from the network
supply and the measurement circuits. The operator must not become charged
with any static electricity. This could cause the destruction of internal parts.
Any adjustment, maintenance or repair of the energized equipment shall
only be undertaken by qualified personnel, after referring to the instructions
given in this document.
A qualified person is a person who is familiar with the installation, construction,
use and the hazards that exist. This person is authorized to start up and shut
down the installation and equipment in conformity with the safety rules.
Return your instrument to your distributor for any work to be done within or
outside the guarantee.
All the equipment has been checked mechanically and electrically before
shipping.
On reception, carry out a quick check to detect any damage caused by
transport. If necessary, contact our commercial department immediately and
make all legal reservations with the carrier.
In the event of reshipping, it is preferable to use the original package. Indicate
as clearly as possible, by a note attached to the equipment, the reasons for
the return.
•
Turn the instrument off.
•
Clean it with a damp cloth and soap.
•
Never use abrasive products or solvents.
•
Allow to dry before any further use.
Two-channel digital oscilloscopes, 60 MHz or 100 MHz I - 5
Description of instrument
Smart
Harmonic
analyser
Description of instrument
Presentation
Persistence
Oscilloscope
mtx 3x52
Recorder
This instrument has the particularity of grouping together 3 units in 1 :
•
a Digital Oscilloscope incorporating the
FFT(Fast Fourier Transform) function and
SPO(Analog Persistence Display)
intended to analyze the signals appearing in the electronics and
electrotechnical fields
•
a Harmonic Analyser (option) mode for breaking down four signals
simultaneously while representing their fundamental and their first 31
harmonics
•
a Recorder mode (option) intended for the capture of single or slow signals.
The instrument works at a constant acquisition depth of 50,000 points.
Memory management is organized from a system of files “Windows ®” like.
A large-size colour LCD screen displays the applied signals, together with all
the adjustment parameters.
The main control functions are accessible directly from the front panel.
The adjustment parameters can be modified using the thumbwheel.
A graphic interface is similar to that of the PC and is used for:
•
using the mouse to select the functions proposed by the pull-down menus
•
acting directly on objects (traces, cursors, etc.) displayed on the screen.
The adjustment parameters can thus be modified by a variety of means.
This instrument is completed by RS232/CENTRONICS, ETHERNET and USB
to RS232 interfaces as standard.
Overall view
II - 6 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Description of instrument
VERTICAL
HORIZONTAL
ACQUISITION
Description of instrument (cont’d)
Front panel
(illustration)
SPO
TRIGGER
Measurement
terminal block
CH1 signal input CH2 signalInput EXT input Calibrator output
Two-channel digital oscilloscopes, 60 MHz or 100 MHz II - 7
Description of instrument
Description of instrument (cont’d)
Rear panel
USB connector
USB to RS232
interface
ETHERNET RJ45
connector
SUBD 25- pin female
connector
for RS232/CENTRONICS
interfaces
6-pin MiniDin
connector
PS/2 mouse
Mains plug
II - 8 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Description of instrument
Description of instrument (cont’d)
Front panel
(
description)
1 startup / standby
1 mouse / 2 keys
button
The main features of the equipment can be obtained from the front panel.
They can also be modified directly by the mouse or using the menu toolbar.
activates:
•
startup (green LED) when pressed briefly.
•
Setting the oscilloscope to standby (red LED) by a long press (> 3 s).
The files and the configuration are saved.
If a menu is open when the key is pressed, the backup is performed,
but no message is displayed.
connected to the back of the oscilloscope (PS/2 connector) used for:
selection of menus,
function validation,
movement of symbols appearing on LCD screen.
•
The menus appearing at the top of the screen and the submenus
selected by the mouse pointer open and are validated by the left key.
•
The menus in the trace display zone
in the control zone
in the status zone
are opened with the right mouse key.
1 rotary control
button
•
The mouse can be used for moving:
symbols appearing in the main display zone:
trigger position, cursor position, display trace reference
symbols appearing in the bargraph:
trigger position, zoomed area cursor position
Position the mouse pointer on the symbol to be moved, holding the left
key of the mouse down during the movement to the desired position.
•
Zooming in the display zone is possible using the mouse:
hold the left key down when defining the zone with the pointer (cannot
be activated in “SPO” analog persistence display
•
The outer wheel of this encoder is used for incrementing or
decrementing the selected setting (by rotation).
•
The LED comes on when adjustment is possible using the wheel. After
20 seconds without action on the wheel, the LED goes out and the
function is no longer active.
•
While the LED is on, pressing the central part of the encoder
(TOGGLE key) will toggle the adjustment of the main function to the
secondary function of the key.
The symbol appears on the keys concerned (except for the key
opposite which has no secondary function.).
21 fleeting action
keys
Two-channel digital oscilloscopes, 60 MHz or 100 MHz II - 9
giving direct access to the more basic functions.
Description of instrument
Description of instrument (cont’d)
Preparation for use
Instructions before
startup
Error messages
Mains voltage
Fuse
Before initial startup:
•
Connect the mouse to the PS/2 connector at the back of the unit.
•
Check the power supply cord which will be connected to the back of the unit
and to the grounded mains is in good condition.
•
When on, the LED shown opposite verifies that the mains voltage is applied
to the oscilloscope.
Refer to §. Technical Specifications p. 139.
The oscilloscope power supply is designed for:
•
a mains voltage from 90 VAC to 264 VAC (use nominal range 100 VAC
and 240 VAC)
•
a frequency included between 47 Hz and 63 Hz.
The unit is protected by a fuse:
•
Fuse range: T, 2,5 A, 250 V, 5 x 20 mm
This protection fuse must only be replaced by an identical model.
Startup
LCD contrast
LCD brightness
Replacement may only be performed by qualified personnel.
Contact your closest distributor.
The pushbutton starts up the oscilloscope (the LED becomes green after
loading).
A long press (> 3 seconds) switches the equipment to standby (the LED
becomes red).
When the unit is powered up, the last memorized configuration is restored.
At initial startup, the default configuration parameters are applied.
Never disconnect the instrument from the network while the message
"System shutdown. Please wait before switch off power !" is displayed
on the screen, otherwise the current file and all the files saved
beforehand will be lost.
This key is used for adjusting the LCD contrast. Adjustment is by the
thumbwheel as long as the LED associated with it is on.
The « TOGGLE » key on the thumbwheel is used for switching from LCD
contrast adjustment to brightness adjustment (and vice versa).
II - 10 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Description of instrument
Description of instrument (cont’d)
ETHERNET
network
General principles
of the ETHERNET
network
Addressing
ETHERNET physical
addresses
IP addresses
ETHERNET and TCP/IP (Transmission Protocol/Internet Protocol) are used to
communicate on a company's network.
Each piece of equipment under TCP/IP has a physical address (ETHERNET)
and an Internet address (IP).
A physical or ETHERNET address, stored in ROM, identifies each item of
equipment on the network. The physical address enables the equipment to
determine the source of data "packet" transmission.
The physical address is a number coded over 6 bytes represented in
hexadecimal form. Hardware manufacturers procure physical addresses and
allocate them incrementally when the product is manufactured. The physical
addresses cannot be modified.
An IP address is coded over 4 bytes, displayed in decimal format.
( Example: 132.147.250.10).
Each field may be coded between 0 and 255 and is separated by a decimal
point.
Unlike the physical address, the IP address can be modified by the user.
You must ensure that the IP address is unique on your network. If an address
is duplicated, network operation becomes random.
The IP address is made up of two parts:
•
the network identifier (Network ID) identifying a given physical network
•
the host identifier (Host ID) identifying a specific item of equipment on the
same network.
There are 5 addressing classes. Only classes A, B and C are used to identify
the equipment. See below:
Class A
0XXXXXXX XXXXXXXX XXXXXXXX XXXXXXXX Network ID Host ID
Class B
10XXXXXX XXXXXXXX XXXXXXXX XXXXXXXX
Network ID Host ID
Class C
010XXXXX XXXXXXXX XXXXXXXX XXXXXXXX
Network ID Host ID
Two-channel digital oscilloscopes, 60 MHz or 100 MHz II - 11
Description of instrument
Description of instrument (cont’d)
SUBNET mask and
GATEWAY
DHCP Protocol
FTP protocol
If the result of the operation ' ET LOGIQUE' between IP address of the
recipient of the message and the value of subnet mask is different from the
address of the recipient of the message, this message is sent to the gateway
which will be given the responsibility to forward it to destination.
The programming of the mask and the address of the gateway is possible on
the instrument, in the Advanced mode.
This protocol is used to automatically assign an IP address to the instrument
when it connects up to the network.
A DHCP (Dynamic Host Configuration protocol) server must be accessible on
this network (contact your network administrator to make sure that this server
is present).
FTP (File Transfer Protocol) is used in the oscilloscope for fast file transfers to
or from a PC.
To use it, open the browser on the PC and, in the URL field, type the IP
address of the instrument, preceded by "ftp:"
Example: ftp://192.168.3.1
The oscilloscope is an FTP server.
See §. Applications p. 126, 132.
HTTP protocol
LPD protocol
With this protocol, the instrument can act as a WEB server and you can access
the most frequent settings and view traces on your PC using your browser
(EXPLORER, NETSCAPE, …)
To use it, open the browser on the PC and, in the URL field, type the IP
address of the instrument, preceded by "http:"
Example: http://192.168.3.1
See §. Applications p. 128.
To be able to display the traces, you must install on your PC the Java Virtual
Machine JVM SUN 1.4.1 (or higher). This JVM can be downloaded from the
site http://java.sun.com
This protocol (Line Printer Daemon) is used by most of the printers connected
to an ETHERNET network, but also by the printing server units which handle
conversion between ETHERNET and CENTRONICS
( Example: Jet Admin) and UNIX and LINUX workstations.
An LPD server can also be installed on a PC (available as an option with
WINDOWS 2000 or XP).
In all cases, the instrument is an LPD client which has to be configured to
indicate to it the IP address of the LPD server (the workstation PC or directly
the printer) and the logical name of the printer managed by the server.
See §. Applications p. 127.
II - 12 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Oscilloscope Mode - Keys
Selective«AUTOSET»
Oscilloscope Mode
Keys
4 «UTILITY» keys
Attention
Pressing this key will configure the instrument in the following modes:
« oscilloscope »,
« harmonic analyser » (option),
« recorder » (option).
used for access to the LCD contrast adjustment using the thumbwheel.
The LED combined with the thumbwheel comes on adjustment is accessible.
The thumbwheel TOGGLE key is used for switching the key assignment from
contrast adjustment to LCD brightness.
Pressing causes switchover from normal display mode to "full screen" display
mode (and vice versa).
The screen is organized in such a way as to leave an optimum trace plotting
surface area: deletion of menu bar,
parameters of traces in time base,
bargraph.
Only permanent adjustments and measurements will remain.
In «SPOOscilloscope » mode, this key has no action : when pressing this key the
following message is displayed : « Not possible in this mode ».
launches a hardcopy depending on the configuration produced in the « Util » and
« Hardcopy » menus.
A second press before the process end will interrupt current printing.
If printing is impossible, a « Printing error » message will be sent.
Attention
1 «AUTOSET» key
with
1 HELP key
The « » symbol is displayed in front of the display zone of the last selected
element when printing is underway.
The first press will freeze the traces on the screen. They will be displayed in plain
language as a reference to be compared with further acquisition.
A second press will erase them: the latter will then be lost.
• Traces will be saved only in the « Memory Trace Save » menu.
• The reference memory will be accompanied by a reference Nr.
In «SPOOscilloscope » mode, this key has no action : when pressing this key the
following message is displayed : « Not possible in this mode ».
used on channels to which a signal is applied to obtain optimum automatic
adjustment (General Autoset) of coupling, vertical sensitivity, time base, slope,
framing and trigger.
The lowest frequency signal is used as a triggering source.
If no trace is detected at the inputs, the autoset will be aborted.
Simultaneously pressing with the CHx key (ch1 or ch2) will assign the
corresponding channel as trigger source, initiating an autoset which will take this
selection into consideration.
The CHx channel becomes active for adjustments by means of AC/DC/GND key.
activating or deactivating help on the keys.
Whenever a keyboard key is pressed, on-line help will be displayed for the
depressed key (except for the key ????). The functions associated with the keys will
not be started up.
On-line help can also be deactivated with the mouse (icon at top right).
The keyboard then resumes normal operation.
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 13
Oscilloscope Mode - Keys
Oscilloscope Mode (cont’d)
2 «ACQUISITION»
keys
by successive pressing, select one of the following acquisition modes:
Single mode Single
Trigger mode Trig’d
Automatic mode Auto
« SINGLE » mode:
A single acquisition is triggered by pressing the RUN STOP key.
For any further acquisition, the triggering circuit must be reset by pressing the
RUN STOP key.
«TRIGGER » mode:
The screen content is only refreshed in the presence of a triggering event
related to signals appearing at the oscilloscope (ch1, ch2, ch3, ch4 or mains).
If there is no triggering event linked with the signals appearing at the inputs (or
if there is no signal at the inputs), the trace is not refreshed.
«AUTOMATIC » mode:
The screen content is refreshed even if the triggering level is not detected on
the signals appearing at the inputs.
In the presence of a triggering event, screen refreshing is managed as in the
« Triggered » mode.
2 «TRIGGER» keys
• allows starting or stopping of acquisition in « TRIGGER » and
« AUTOMATIC » mode.
• resets the triggering circuit in the «Single» mode.
• A long press forces the trigger (Force TRIG).
Acquisition is initiated according to the conditions defined by the
acquisition mode (SGLE REFR key).
Acquisition status is indicated in the status zone:
RUN = started READY = wait STOP = stopped
PRETRIG = before trigger POSTRIG = after trigger
selects the trigger slope (positive or negative ) by successive
pressing.
The slope is indicated in the status zone.
sets the trigger level to the average value of the signal (50%) without
modifying the trigger coupling.
The thumbwheel is assigned to adjusting the trigger level.
Combined pressing with the CHx key launches the same function, but
previously sets the corresponding channel as triggering source.
No functions are associated with the TOGGLE key of the thumbwheel.
Attention
III - 14 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
In «SPOOscilloscope » mode, this key has no action : when pressing this key the
following message is displayed : « Not possible in this mode ».
Oscilloscope Mode - Keys
Oscilloscope Mode (cont’d)
3 « MEASUREMENT »
keys
activates or deactivates the display of the window for the 19 automatic
measurements of the reference trace.
Combined pressing with the CHx key displays the measurements on the
corresponding channel.
When the automatic measurement window is active, the left mouse button
is used for selecting at most 2 measurements that will appear in the status
zone at the bottom of the screen.
used for selecting (successive pressing) among the displayed traces, the
reference trace for the automatic and manual measurements.
Appears in the « Measurement » Reference menu.
activates or deactivates the cursor displays for manual measurements.
The LED combined with the thumbwheel comes on: the latter allows cursor
1 to be moved horizontally over the screen.
The thumbwheel TOGGLE key is used for moving from the cursor 1 to
cursor 2 horizontal movements and vice versa.
•
The dt measurements made (time difference between the two cursors)
and the dv measurement (voltage deviation between the two cursors) are
reported in the status zone.
•
The selected cursor position is entered into the active adjustment zone.
2 «HORIZONTAL»
keys
adjusts the time base coefficient (T/div.) by thumbwheel or the horizontal
position (H-Pos.) by the thumbwheel TOGGLE key.
The LED associated with the wheel lights up the selected adjustment is
possible with this device.
The H-Pos. adjustment modifies the horizontal (time-related) position of the
trigger point.
activates or deactivates the « Zoom » function.
The LED associated with the wheel lights up: the thumbwheel is assigned to
the horizontal zoom coefficient adjustment.
Pressing the thumbwheel TOGGLE key makes it possible to change from
the horizontal zoom coefficient setting to Z-Pos. horizontal movement in the
zoomed zone.
A zone can be zoomed by tracing a rectangle around the zone to be
enlarged using the left mouse button. The sensitivity, time base and
horizontal and vertical alignment values are recalculated automatically.
If no zones to be zoomed are selected with the mouse, a simple horizontal
zoom by default will be performed with respect to the screen center.
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 15
Oscilloscope Mode - Keys
Oscilloscope Mode (cont’d)
Definition :
5 «VERTICAL» keys
Validated channel = Display enable (trace displayed after RUN)
Displayed channel = Validated channel and trace on screen
Selected channel = Parameter settings enabled for this channel via the
opposite key.
Before pressing one of the
four keys:
The channel concerned is not
displayed.
The channel concerned is
displayed, but not selected.
The channel concerned is
displayed and selected.
Press
on
After pressing one of the 4
keys:
The channel is displayed and is
selected. The thumbwheel is
assigned to sensitivity adjustment.
Double pressing one of those keys devalidates (erases) the concerned
signal.
In « SPO mode », the math functions are not authorized. In this case, the
MATH {3 , 4} keys validate or devalidate the M {3 , 4} memory channels.
Pressing one of these 2 keys for a long time generates a verticalautoset:
•
This modifies the sensitivity and vertical positioning of the wheel in
question.
•
It optimizes the display on the screen by activating and selecting the
channel.
Each press gives access, through the thumbweel, to the adjustment of the
sensitivity (V/div.) of the last selected channel.
By pressing the TOGGLE, you switch from the sensitivity adjustment to the
verticalposition adjustment (V-Pos.).
When the LED of the thumbwheel lights on means that you may adjust with
the thumbwheel.
used for selecting, by successive pressing, input coupling « AC », « DC »
or « GND » for the last channel selected.
Coupling is indicated in the channel parameters zone:
AC :, DC : ground :
III - 16 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Oscilloscope Mode - Display
4. Menu bar
uency of appearance (red corresponds to
Oscilloscope Mode (cont’d)
Display
in oscilloscope
mode
in SPO Persistence
mode
See chapter IV, p. 67.
View serial bus Data in envelope mode.
In this mode, the absence of accumulation of
successive acquisitions and multi-colour display
does not show infrequent signals. All points
displayed at the same colour and intensity.
Displays a warbled triangular signal
The modulation amplitude in Envelope
mode may be seen on the screen. All
points displayed at the same colour and
intensity.
Composition
Displays serial bus Data.
In "Persistence SPO" display, successive
acquisitions are accumulated and the multicoloured view visually represents the
frequency with which points (traces) appear,
infrequent signals in "blue", and the more
frequent signals in "red".
In "Persistence SPO" view, a warbled
triangular signal is displayed: the modulation
amplitude can be viewed on the screen. The
colour of traces varies according to their
freq
high appearance frequencies and blue to
low).
The oscilloscope display is divided into 4 functional zones.
3. Display zone
zone
2. Control
Direct access
to current
adjustment
1. Status zone
Display and
adjustment of
current value
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 17
Oscilloscope Mode - Display
manual cursors
Movement to left of screen in
T
Oscilloscope Mode (cont’d)
1. Status zone
Bargraph
Cursor Measurements
acquisition memory
Three types of general information appear in this zone:
•
the bargraph represents the screen position, that of the trigger and the
cursors in the acquisition memory
•
permanent instrument settings
•
measurements when the cursors are on the screen or the trigger type
(1)dt = 500µs, dv =-6,400V Trig 1 Pretrig
Each bargraph element can be moved via the mouse left button.
1
Position and movement of
Representation and
movement of screen in
acquisition memory
Position and movement
of time trigger
2
adjustments
Movement of screen to right
in acquisition memory
Permanent
Permanent settings
Cursor measurements
This zone refers to the triggering status (mode, front, source, current
status).
Example : AUTO 1 STOP
When the cursor of the mouse is placed over this information, the right
mouse button will open the "Trigger parameters" menu.
This zone refers either to:
•
horizontal (dt) and vertical (dv) difference between 2 cursors in the
case of manual measurements
(
Example : (1)dt = 500,0 µs, dv = -6,400 V)
•
phase measurement in the case of manual phase measurement (Ph)
(
Example : (1)Ph = 130.0°)
•
automatic measurements selected by the "Automatic measurements"
or "Phase measurement" menu.
(
Example : (1)F = 1.000kHz, Vpp = 15.00 V)
III - 18 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Oscilloscope Mode - Display
Indication and adjustment of last selected
Display of ZOOM mode (Z
)
Display of the parameters
Oscilloscope Mode (cont’d)
2. Control zone
The parameters displayed in this zone are:
•
The parameters of each channel and trace: display,
The time base value and the loading of the signal representation domain
•
The indication and the active adjustment of the last selected element:
- trigger level, trigger type
- trigger time position
- channel offset value
- X & Y position of cursor
- time display, if no setting has been selected
- symbol of current printing ….
Display of trace parameters (in trace
color):
validity, coupling, bandwidth limit,
sensitivity
Display of ZOOM mode (Z)
Display of function parameters (in trace
color):
validity, value of division
of a memory : sensitivity,
coupling, bandwidth limit
Value of time base coefficient (s/div) in
oscilloscope mode or frequency (Hz/div)
in FFT mode
Change of signal representation domain
(FFT selection)
Contrast
25.0%
element or otherwise time display and
trigger type
•
In SPO mode, the FFT is not available. The SPO logo is displayed.
•
The mouse left key validates the channels and the functions.
•
The «
» symbol indicates whether a channel or function has been
selected, or whether the FFT mode has been selected.
•
The adjustments of the time base (or frequency) and the value of the
active parameter can be made with the UP/DOWN button alongside the
current value display using the mouse key button.
•
After a change to the time base, the corresponding sampling frequency is
entered into the adjustment zone.
•
When the mouse is placed over the parameters of a channel or time base
value, the right mouse button opens the associated menus directly.
- Sensitivity/Coupling and Vertical scale, for channels and functions
- Source, trigger mode and RUN/STOP, for the time base
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 19
Oscilloscope Mode - Display
Oscilloscope Mode (cont’d)
The « Source » and « Trigger mode » menus are grouped together and
can be opened using the right mouse button by placing the pointer over
the time base zone.
3. Display zone
RUN/STOP is used for starting and stopping acquisition from this menu.
The acquisition status is indicated in the screen status.
•
The symbol « » indicates the source and selected trigger mode
•
The trigger source selectable from this menu is limited to the channels
(ch1 or ch2).
The displayed graphic elements associated with the traces in this zone
are:
•
a trigger time position indicator
•
a trigger level indicator
•
a vertical position indicator for the reference level of each trace
•
cursor position indicators linked with the curve for automatic
measurements
•
cursor position indicators linked or not linked with the curve for manual
measurements
•
the selection of a zoom zone
III - 20 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Oscilloscope Mode - Display
7 9
8
11
10
1
3
4
5
2
1
>
v
ϕϕϕϕ
2
6
6
v
12
Oscilloscope Mode (cont’d)
Display elements
Definition of
display
Items Elements
1 Trace displayed
2 Indication of displayed trace reference level vertical position
Traceaffichée
3 Indication of time-related position of trigger
4 Graticule division
5 Automatic measurement cursor position indicator
6 Manual measurement cursor position indicator
7 Phase measurement cursor position indicator
8 Trigger level position indicator
9 Zoom zone selection (not available in SPO)
10 Time position output indicator of trigger outside displayed
11 Trigger level position output indicator outside displayed
window (in SPO the trigger is limited to the view zone)
window.
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 21
Oscilloscope Mode - Display
Oscilloscope Mode (cont’d)
Menu accessible
from display zone
How to zoom
Attention
When the mouse pointer is placed in the display zone,
the right key gives direct access to a display menu.
The « Full Screen » and « Zoom out » options are directly
accessible (see the Display menu). The same applies to
the selection of the automatic and manual measurement
reference signal (see Measurement menu).
The « » symbol indicates that the display is in the « Full Screen » or
“Full Trace” mode
(if present) and gives the reference trace for automatic and manual
measurements.
Zooming in the display zone is possible using a mouse, holding the left
mouse key down when the zone is selected by the pointer.
The zoom function is not available in SPO.
After zooming in to part of the screen, the sensitivities of the traces and
the time base are recalculated.
•
The « z » symbol appears in the parameter display of the signals and
the time base.
•
The zoomed section is represented in the bargraph.
How to move the
symbols
•
The « Zoom Out » mode (see Display menu) returns to the original
display.
•
The horizontal zoom value is adjusted to assign a calibrated value to
the horizontal scale (zoom factor: x 100 max.)
If the zoom vertical selection is greater than 6 divisions, no vertical zoom
will be performed.
All the symbols appearing in the display zone:
- trigger indicators,
- trace position indicator,
- manual cursor position indicator,
- etc …
can be moved using the left button of the mouse.
The new modified symbol value is entered into the current adjustment
display zone.
III - 22 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Oscilloscope Mode - Display
Oscilloscope Mode (Cont’d)
Example of an
expanded area
Expanded area (in blue)
Normal mode screen
Expanded mode screen
4. Menu bar
All the oscilloscope functions can be accessed by the main menus.
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 23
Oscilloscope Mode - The « Vertical » Menu
Oscilloscope Mode (cont’d)
The « VERT » Menu
15MHz
(∗)
Only available in advanced mode
if the SPO mode is not selected.
See §. Description, page 65.
III - 24 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Oscilloscope Mode - The « Vertical » Menu
Modifies the sensitivity of the channel using the scroll bar and the left mouse
Oscilloscope Mode (cont’d)
Display
ch1 ch2
Sensitivity/Coupling
Channel Sensitivity
Coupling
opens the « Trace display » menu for validating or devalidating the traces.
Validation of selections by « OK ». Exit from menu without modification by
« Cancel ».
The «
» symbol in front of a trace indicates its validation.
The traces can be validated or devalidated from the command zone using
the left mouse button.
modifies independently the ch1 and ch2 parameters and modifies the
selected trace vertical scale.
modifies the selected channel parameters.
button, adjustable by sequence: from 2.5 mV to 100 V/div.
Sensitivity is entered into the channel parameter display zone. It takes the
« Vertical scale » menu parameters into consideration.
Modification of the AC - DC - GND coupling
AC : blocks the DC component of the input signal and attenuates the signals
lower than 10 Hz.
DC : transmits the DC and AC components of the input signal.
GND : the equipment connects internally the selected channel input to a
0 V reference level.
The « » symbol indicates the selected coupling. Coupling is reported in
the modified channel parameter display zone.
BW limit
This menu can also be called by clicking with the right mouse key in the
Vertical scale
Coefficient assigns a multiplication coefficient to the selected channel sensitivity.
Limits the bandwidth of the channel and its trigger circuit to reduce display
noise and false triggering.
The bandwidth of each channel can be limited to 5 kHz, 1.5 MHz or
15 MHz. The bandwidth limit of a channel is indicated in the control area by
following symbols :
Validation of selections by « OK ». Exit from menu without modification by
« Cancel ».
desired channel parameter display zone.
defines the vertical scale of the channel selected from the current
adjustments. Reading and direct measurements of the analyzed
magnitude and its unit are obtained.
The modification is made using the mouse with the table of usable
numbers, having selected the « Coefficient » zone.
The key deletes the previous value of the cursor in this zone.
15 MHz
1.5 MHz
5 kHz
The predefined values (x1, x10, x100, x1000) correspond to standard
probe coefficients and can be assigned directly.
The sensitivity value indicated in the channel parameter display will be
modified according to this coefficient.
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 25
Oscilloscope Mode - The « Vertical » Menu
∗)
Oscilloscope Mode (cont’d)
Measure unit
math1 math2
math3 math4
Function definition
modifies the vertical scale unit of the selected channel.
The modification is carried out by the mouse using the table of characters that
can be used after selecting the “measure unit” zone.
The key is used for deleting the character leading the cursor in this zone.
The "Padlock " key can be used to switch between upper-case and lowercase characters.
The vertical scale unit will be entered into the modified channel parameter
display.
re-initializes the multiplication coefficient at 1.00 and returns to a measure unit
Init
in Volt.
Validation of selections by « OK ». Exit from menu without modification by
« Cancel ».
This menu can be called by dialogue with the right mouse button in the display
of the desired channel parameters (ch1 or ch2).
defines for each trace a mathematical function and the vertical scale.
Menus present only in Advanced mode (see menu « Util », p. 61).
defines the mathematical function to be assigned to the selected trace.
The function is defined using the usable characters table, and associating the
ch1 and ch2 traces.
•
The mathematical function can be defined on 2 lines.
•
mathx cannot be used in the definition of a function.
Functions
The key deletes the character preceding the cursor in the window.
8 predefined mathematical functions can be linked to the traces :
divh(
divv(
step(
sin(
(
divh(1) is equivalent to 5000 samples (counts) = 1 horizontal div.
Validation of the selections by "OK". Exit from the menu without modification by
"Cancel".
If … then …
... the dynamic calculation of the
vertical scale is impossible
... the dynamic calculation of the
vertical scale is possible
CHx + CHy Sensitivity and measuring unit used on CHx
CHx - CHy Sensitivity and measuring unit used on CHx
In each cases, the measuring unit can be re-defined and a coefficient can be applied
to the measurement results (refer to §. Vertical scale).
("horizontal division")
("vertical division")
("step") using "t" (∗)
("sine")
t = abscissa of the sample in the 50,000-sample acquisition memory.
Note: The results of the automatic Vhigh and Vpp by math 3 must be
multiplied by two to be correct.
III - 28 Two-channel digital oscilloscopes, 60 MHz or 100 MHz
Oscilloscope Mode - The « Vertical » Menu
Oscilloscope Mode (cont’d)
For immediate interpretation of the results, configure the "Vertical scale"
menu of mathx (see §. Opening from math1, math2, math3, math4 p. 35).
In our example:
• The sum of ch1 + ch2 is the sum of two values in volts, so the
result is expressed in volts.
• The sum of ch1 + ch2 must be divided by 2, so the coefficient of
math3 can be replaced with 2 to obtain the automatic measurement
results of math3 immediately.
• Then select math3 as the reference for the automatic and manual
measurements (see "MEASUREMENT" menu, p. 53).
• Then display the table of 18 measurements made on the math3 trace
(see "MEASUREMENT" menu, p. 53).
The measurements displayed are the exact result of the addition of the
two traces ch1 + ch2 in the correct unit (Volts).
Vhigh
Vertical scale math3 = 2.00 V
Vpp math3 =12.00 V
Vhigh math3 = 6.00 V
Two-channel digital oscilloscopes, 60 MHz or 100 MHz III - 29
As for the addition of traces, we can observe a much more significant high
and low overshoot.
Vhigh math3 = ch1 x ch2 = 1 vertical division x 1 vertical division
= 32000 LSB x 32000 LSB = 1024 106 LSB
> (4 vertical divisions = 128000 LSBs)
The function divv (vertical division) is necessary to optimise the display.
- math3 = (ch1*ch2)/divv(1)
Divv(1) can be used to divide by 32,000 (1 vertical division = 32,000 LSBs):
the result of the multiplication is translated into divisions on the screen.
Note: If Vpp of ch1 and ch2 had been 8 vertical divisions, the multiplication
would have had to be divided by divv(4).
When mathematical functions associated with traces are used, the
dynamics of the result obtained must be verified.
Correction of the result of the operations by mathematical functions (divv(),
divvh(), / …) is recommended to optimise the screen display.
III - 30 Two-channel digital oscilloscopes, 60 MHz or 100 MHz