Metrix DOX3104, DOX3304 User Manual

Page 1
Digital Storage Oscilloscope SPO
«Sensitive Phosphor Oscilloscopes»
DOX3104
WaveGenerator - Serial Bus Decode
DOX3304
WaveGenerator - Serial Bus Decode
8-Channel Logic Analyzer
User’s Manual
Pôle Test et Mesure de CHAUVIN-ARNOUX
Analog Inputs
Digital Inputs
Arbitrary Generator
Interactive Menu buttons
LCD TFT 8″ 800 x 480
USB
Cal. 3V 1kHz
Control Front Panel ON/OFF
Page 2
Content
2 4-Channel SPO Digital Oscilloscopes
Content
General Instructions
Introduction .............................................................................. 6
Precautions and safety measures ............................................. 6
Symbols on instrument ............................................................. 7
Guarantee, Repair, Servicing .................................................... 7
Instrument Description
Front panel ............................................................................... 8
Rear panel ................................................................................ 9
User display interface ............................................................. 10
Menu and Control buttons ...................................................... 11
Front panel Inputs and Ouputs................................................ 13
Rear panel Connectors .......................................................... 13
Universal knob ........................................................................ 13
Getting Started
Function Checking .................................................................. 14
Probe ...................................................................................... 15
Probe compensation .............................................................. 16
Function Description
AUTO Config/Default Config .................................................................................................... 17
Auto Setup .............................................................................. 17
Default Configuration ............................................................. 18
I - Vertical System .................................................................................................................... 19
Channels CH1 - CH2 - CH3 - CH4 ............................................. 19
Channel configuration ............................................................ 20
Input coupling ........................................................................ 21
Bandwidth Limit ...................................................................... 21
Vertical Sensitivity .................................................................. 22
Probe Factor ........................................................................... 23
Invert function ........................................................................ 23
Input impedance ..................................................................... 24
Vertical Unit ............................................................................ 25
Vertical knobs ............................................................................................................... 26
Vertical Position knob ............................................................. 26
Vertical Sensitivity knob ......................................................... 26
Storing and Displaying Reference waveforms ............................................................... 27
REF Menu ............................................................................... 27
Mathematical Funstions ................................................................................................ 28
MATH Menu ............................................................................ 28
Mathematical operations ..................................................... 28
Fast Fourier Transform FFT .................................................... 29
FFT Menu ............................................................................ 29
How to use the FFT ............................................................. 29
Displaying the FFT spectrum .............................................. 30
Selecting the FFT window ................................................... 31
FFT : Vertical and Horizontal scales .................................. 32
Vertical and Horizontal Position .......................................... 32
Cursors Measurements ....................................................... 32
Amplitude .......................................................................... 32
Frequency .......................................................................... 33
FFT of « Cal 3V 1kHz » signal ............................................. 34
Page 3
Content
4-Channel SPO Digital Oscilloscopes 3
II - HORIZONTAL System ......................................................................................................... 35
Horizontal menu ..................................................................... 35
HORIZONTAL Pad ................................................................... 36
Horizontal Position ................................................................ 37
Time Base Coefficient S/div ................................................... 37
Horizontal Zoom .................................................................... 37
III - TRIGGER System ............................................................................................................... 38
Trigger Menu .......................................................................... 38
Tigger Source ......................................................................... 39
Trigger Type ........................................................................... 40
Edge .................................................................................. 40
Pulse ................................................................................. 41
Video ................................................................................ 43
Slope .................................................................................. 44
Window ............................................................................... 46
Interval ............................................................................... 47
DropOut .............................................................................. 48
Runt .................................................................................... 49
Pattern ................................................................................ 50
Serial 1-2 ............................................................................ 53
I2C ................................................................................... 53
SPI ................................................................................... 54
UART/RS232 .................................................................... 55
CAN ................................................................................. 56
LIN ................................................................................... 57
Trigger Coupling .................................................................... 58
Horizontal Position ................................................................. 58
Slope&Level ........................................................................... 58
Holdoff .................................................................................... 59
IV - ACQUISITION System ........................................................................................................ 61
Acquisition menu .................................................................... 61
Acquisition Configuration ................................................ 62
Acquisition modes .................................................................. 63
Normal ................................................................................ 63
Peak detect ........................................................................ 63
Average ............................................................................. 64
High Res ............................................................................ 64
Sequence ............................................................................ 65
Memory Depth setting ......................................................... 67
Time Base ............................................................................... 68
Under-Sampling - Aliasing ...................................................... 68
Sequence Mode ...................................................................... 70
Run/Stop, Single, Auto, Normal ............................................... 71
Page 4
Content
4 4-Channel SPO Digital Oscilloscopes
Content (cont’d)
V - Display System .................................................................................................................. 72
Display ................................................................................... 72
SPO : Color ON - Color Off ...................................................... 73
X-Y Format .............................................................................. 75
VI - Measurement System ........................................................................................................ 77
Scale measurements ............................................................... 78
Cursors measurement ............................................................ 77
Manual Cursors .......................... 77
Track Mode .......................... 79
Automatic measurements ....................................................... 81
Automatic Measurements menu .......................... 81
Voltage Measurements .......................... 81
Time Measurements .......................... 81
Delay Measurements ......................... 81
All Measurements .......................... 82
Measurement Types .......................... 83
VII - SAVE/RECALL System ..................................................................................................... 87
Save/Recall ........................................................................... 87
Saving Setup files in Internal
and External memory ......................... 89
Recall a Setup file ......................... 90
Rename a Setup file .......................... 91
Delete » a Setup file .......................... 91
Save/Recall Waveforms ......................................................... 92
Save/Recall Picture ............................................................... 95
Save .CSV file ....................................................................... 96
Recall Default Setup .............................................................. 98
VIII - UTILITY System ............................................................................................................. 101
Utility Menu ........................................................................... 101
Utility page 1/3 ................................................................... 102
Utility page 2/3 ................................................................... 103
Utility page 3/3 ................................................................... 105
System Status ...................................................................... 107
Language ............................................................................. 107
Print ...................................................................................... 108
Do Self Cal ........................................................................... 111
Do Self Test ......................................................................... 111
Screen Test ....................................................................... 111
Keyboard Test ................................................................... 112
LED Test .................................................................................... 112
Firmware Update .................................................................. 113
Pass/Fail .............................................................................. 113
History ................................................................................. 119
Options .......................................................................................... 122
Date/Time ............................................................................. 124
I/O ......................................................................................... 125
Page 5
Content
4-Channel SPO Digital Oscilloscopes 5
IX - Arbitrary Waveform Generator ............................................................................................. 126
Creating an arbitrary waveform with EasyWave ..................... 126
Auto-Calibrating the arbitrary waveform generator ............... 129
X - DECODE option for Decoding Serial Bus .............................................................................. 130
I2C serial bus ....................................................................... 130
Setting I2C Trigger .............................................................. 131
Decoding I2C ...................................................................... 134
Reading I2C decoding ......................................................... 135
SPI serial bus ....................................................................... 136
Setting SPI Trigger .............................................................. 138
Decoding SPI ...................................................................... 139
Reading SPI decoding ......................................................... 140
UART/RS232 serial bus ........................................................ 141
Setting UART/RS232 Trigger ............................................... 142
Decoding UART/RS232 ....................................................... 144
Reading UART/RS232 decoding .......................................... 145
CAN serial bus .................................................................... 146
Setting CAN Trigger ............................................................ 151
Decoding CAN .................................................................... 157
Reading CAN decoding ....................................................... 158
LIN serial bus ....................................................................... 157
Setting LIN Trigger ............................................................. 158
Decoding LIN ...................................................................... 159
Reading LIN decoding ......................................................... 160
XI - DIGITAL Option Logic Analyzer…………………………………………………………………………160
XII - Power Analysis Option.............…………………………………………………………………………167
XIII - Remote Control .......................…………………………………………………………………………172
Messages ...................................................................................................................................... 178
Help ..................................................................................... 178
Messages ............................................................................. 179
Diagnostic ............................................................................ 180
Technical Specifications ................................................................................................................ 181
Appendix ...................................................................................................................................... 188
Page 6
General Instructions
6 4-Channel SPO Digital Oscilloscopes
General Instructions
Introduction
You have just acquired a 4-channel SPO Digital oscilloscope:
DOX3104, 4-channel, 100 MHz, 2 GSPS, 28Mpts, Arbitrary Waveform Generator,
Serial Bus Decode or
DOX3304, 4-channel, 300 MHz, 2 GSPS, 28 Mpts, Arbitrary Waveform Generator,
Serial Bus Decode, 8-channel Logic Analyzer. This 2-4-channel oscilloscope provides a set of powerful features for a wide range of
applications such as production, education, maintenance, service, research and development.
Congratulations for your choice and thank you for your trust in the quality of our products.
This instrument conforms to safety standard NF EN 61010-1, single insulation, and relative to electronic measurement instruments. This is a class 1 device which must be connected to the protective earth by its power cord.
To obtain optimum service, read these instructions with care and comply with the precautions for use.
Failure to comply with these warnings and/or user instructions is liable to cause damage to equipment to the equipment and/or components. This could be dangerous to the user.
Précautions and safety measures
This instrument has been designed for use:
- indoors,
- in a pollution degree 2 environment,
- at an altitude of less than 2000 m,
- at a temperature in the range of 0°C to 40°C
- at a relative humidity below 80 % up to 31°C.
It can be used for measurements on 300V CAT I circuits and can be powered by a
300V CAT II network.
Définition of
installation
categories
Overvoltage Category l : Is applicable to instruments and equipment which are not intended to be connected to the mains supply. Because the available energy is very limited this category is normally not marked on the equipment. CAT I is for connection to circuits in which measures are taken to limit transient over-voltages to an appropriately low level. Examples: Protected electronic circuits. Overvoltage Category ll : is for equipment intended to be supplied from building wiring. It applies to both equipments plug-connected or permanently connected.
E.g.:Measurements on the network circuits of household appliances, portables tools and other similar appliances.
Overvoltage Category lll : is for equipment intended to be incorporated into the building wiring. Such equipment includes socket outlets, fuse panels, and some mains installation control equipment. E.g.: Measurements on distribution panels (including secondary
meters) circuits breakers, cabling including cables, busbars, junction boxes, disconnecting switches, power outlets in the fixed installation, and industrial appliances and other equipment, such as motors permanently connected to the fixed installation.
La catégorie de surtension lV : is for equipment installed at or near the origin of electrical supply to a building, between the building entrance and the main distribution board. Such equipment may include electricity tariff meters and primary overcurrent protection devices. E.g.: Measurements on systems installed before the main fuse or the
circuit breaker of the building’s installation.
before use
Comply with environment and storage conditions.
during use
Read carefully all the notes preceded by the symbol . Connect the instrument to an outlet with a protective ground pin. Be sure not to obstruct the aeration points. As a safety measure, use only suitable cords and accessories supplied with the
instrument or type approved by the manufacturer.
When the instrument is connected to the measurement circuits, never touch an
unused terminal.
Page 7
General Instructions
4-Channel SPO Digital Oscilloscopes 7
General Instructions (cont’d)
Symbols on instrument
Warning: Risk of danger. Refer to the operating manual to find out the nature of the potential hazards and the action necessary to avoid such hazards.
Selective sorting of waste for recycling electric and electronic materials. In accordance with the WEEE 2002/96/CE directive: must not be treated as household waste.
Earth symbol
On/Off switch
LAN symbol
USB symbol
Fuse symbol
European Conformity
Hazardous voltage
Alternative Current
Guarantee
This equipment is guaranteed for 3 years against any material defect or manufacturing faults, in conformity with the general conditions of sale. During this period, only the manufacturer may repair the equipment and it reserves the rigth to carry out repair or replacement of all or part of the equipment. If the equipment is returned to the manufacturer, forward transport is at the expense of the customer. The guarantee does not apply in the event of :
Improper use of the equipment or by association with incompatible equipment Modification of the equipment without the explicit authorization of the manufacturer
technical services
Equipment repaired by a person not authorized by the manufacturer Equipment used in a particular application not suitable for the device or not described
in the user manual
Shock, fall or flooding.
Repair
For repairs under or out off warranty, please contact our nearest « CHAUVIN­ARNOUX » sales office or our “ MANUMESURE” regional technical center , which will establish a feedback file and will communicate the procedure to be followed.
Contact information on our website : http://www.chauvin-arnoux.com or by phone at the following numbers: 02 31 64 51 43 (MANUMESURE Technical Center) 01 44 85 44 85 (Chauvin Arnoux). For repairs outside metropolitan France, return the instrument to our “Chauvin-Arnoux” local agency or distributor.
Servicing
Turn the instrument off. Clean it with a damp cloth and soap. Never use abrasive products or solvents. Dry it before any further use.
Page 8
Instrument Description
8 4-Channel SPO Digital Oscilloscopes
Instrument Description
Front Panel
No.
Description
No.
Description
1
Universal Knob
12
Trigger Control
2
Horizontal Controls
13
Common Functions Keys
3
Auto Setup
14
Vertical Controls
4
« Default » Configuration
15
Six menu buttons
5
Run/Stop
16
«Up » Menu button
6
Generator set button « Wave Gen »
17
On/Off button
7
« DECODE » set button
18
Analog Inputs BNCs
8
« Digital » set button
19
Logic Analyzer Inputs
9
« Math » set button
20
« USB Host » connector
10
« Ref» set button
21
Probe Adjust Output
11
Vertical Controls for (Decode/Digital/Math/Ref
22
Arbitrary Waveform Generator Output
17 18 19 20 21 22
1 2 3 4 5 6 7
7 8
11
9
10
14
Page 9
Instrument Description
4-Channel SPO Digital Oscilloscopes 9
Instrument Description (cont’d)
Rear Panel
1. Handle
To transport the oscilloscope turn the handle uprigth.
2. EXT TRIG
External trig input BNC “EXT TRIG.
3. « PASS/FAIL » output or trigger output « TRIG OUT » :
The oscilloscope delivers either a square signal whose frequency is the number of waveforms acquired per second (Aux Output=Trig Out) or the number of “Fail” tests (Aux Output = Pass/Fail).
4. LAN RJ45 connector for a remote control of the oscilloscope (VXI11).
5. USB Device USB connector for « PictBridge » printer (printer) or PC control (USBTMC).
6. Kensington Lock The Kensington cable is not supplied with the unit
7. Power cord connector The supply voltage range is 100-240 Vac, 45-440 Hz. Use the power cord supplied with the unit.
8. Fuse Slow blow fuse 5x20mm 250V, 1.25A.
Page 10
Instrument Description
10 4-Channel SPO Digital Oscilloscopes
Instrument Description (cont’d)
User display interface
1.
Trigger status:
Arm: The oscilloscope is acquiring pre-trigger data. All triggers are ignored in this state. Ready: All pre-trigger data has been acquired and the oscilloscope is ready to accept a trigger event.
Trig’d: The oscilloscope has detected a trigger event and is acquiring the post-trigger data.
Stop: The oscilloscope has stopped acquiring waveform data. Auto: In trigger « Auto » mode, the oscilloscope refreshes the trace even in absence of a trigger event.
2.
Time Base coefficient : S/div.
3.
Indicates the horizontal trigger position relative to the screen center, use the horizontal Position knob to adjust it.
4.
Zoom” Time Base coefficient : S/div
5.
Awg : High-Z Arbitrary waveform Generator output impedance: High-Z or 50
6.
USB flash plugged-in
7.
Indicates whether the back USB option is set to : Computer or Printer
“back USB ” option is set to Computer
“back USB” option is set to Printer
8.
Shows the Channel vertical position symbol
9.
Shows the CHi active Menu (CH 1)
10.
Input Coupling symbol
11.
Indicates whether the bandwidth limit is « 20MHz » or « Full Bandwidth».
12.
Vertical Scale Adjust : « Coarse » or « Fine »
13.
Probe Factor Selection : .1X .2X .5X 10X 20X 50X 100X 200X 500X 1000X 2000X 5000X 10000X
14.
Input impedance : 1M - 50
15.
Next page Menu CHi (CANAL i)
16.
Display for each of the active channels: Channel number, Probe factor, Input Impedance, « B » for the 20MHz bandwidth limit if enabled, Input
Coupling, Vertical Sensitivity V/div
17.
Indicates the trigger type, the trigger source, the trigger slope, the trigger coupling and the trigger level in Volt.
18.
Indicates the sample rate and the current memory depth.
19.
Indicates the frequency of the trigger source signal (hardware counter)
1 2 3 4 5 6 7
8
9
19 18
17
16
20
10 11 12 13 14 15
Page 11
Instrument Description
4-Channel SPO Digital Oscilloscopes 11
20.
Indicates the date and time
Instrument Description (cont’d)
Menu and Control buttons and knobs
Channel
buttons
CH1, CH2,
CH3, CH4
Press the CH1 (or CH2 or CH3 or CH4) button to enable (ON) or disable (OFF) the channel and to open the CHi configuration menu. When the channel is “On” the button is lit.
Run/Stop
Run : The traces are refreshed continuously, Stop: acquisition is Stopped.
Note: In « Stop », use the S/div knob to zoom or compress the trace.
Auto Setup
Automatically sets the oscilloscope controls to produce a suitable display of the input signals.
Wave Gen
Press this button to enable the « Arbitrary Waveform Generator »
Horizontal Control
Trigger Control
Common Menu buttons
Vertical Control
Advanced
Functions
Arbitrary Generator
Submenu
buttons
Page 12
Instrument Description
12 4-Channel SPO Digital Oscilloscopes
Default
Press this button to reset the oscilloscope settings to the « Default » configuration
Decode
Press this button to enable and configure the serial bus « Decode » option
Math
Press this button to enable and configure the « MATH » function.
Digital
Press this button to enable and configure the 8 channel « Digital » bus option
Ref
Press this button to open the REF WAVE menu. Use this menu to Save or Recall reference waveforms (REFA REFB REFC REFD) in the internal memory. When a reference waveform is displayed the “Ref” button is lit.
Horiz
Press this button to open the horizontal menu to select the horizontal format : YT, XY or Roll.
Setup TRIG
Press this button to open the TRIGGER menu to select the trigger type : Edge, Slope, Pulse, Video, Window, Interval, DropOut, Runt, Pattern, Serial1, Serial 2 and to configure it.
50%
Push the « Level » knob to obtain a stable trace. The oscilloscope automatically sets the trigger level to be halfway between the minimum and maximum voltage level of the trigger source signal. This is particularly useful when the trigger source is a non displayed signal like EXT TRIG.
« Auto »
Press this button to enable the Auto mode. In this mode the traces are refreshed in the presence or absence of a trigger event.
Normal
Press this button to enable the Normal mode. In this mode the traces are refreshed only in the presence of a trigger event.
Single
Press this button to enable the « Single » mode trigger. In this mode the oscilloscope performs a single acquisition and then « STOP ».
Utility
Press this button to open the « Utility » menu. The utility menu allows to configure the oscilloscope options such as: Sound, Language, Printer, Date/Hour etc. The Utility menu allows to display the « System Status » and to update the firmware.
Display
Press this button to open the « Display » menu. The display menu allows to configure the display Type, Color, Grid, Persistence, Intensity, Brightness, etc..
Cursors
Press this button to open the « CURSORS » menu. Use the « Universal » knob to position the active cursor. The sliders are displayed after leaving the « CURSORS » menu (exept if mode = Off) but can not be adjusted.
Help
Press this button to open the access to the internal Help.
Print
Press this button to open the « Print » menu. The back USB connector must be set on « Printer ».
Measure
Press this button to open the « MEASURE » menu.
Acquire
Press this button to open the « Acquire » menu. The Acquire menu allows to configure the sampling mode : Normal, Peak Detect, Average, High Resolution.
Save / Recall
Press this button to open the « Save/Recall » menu. This menu allows to Save (or Recall) “Setups” in the internal or external memory and “Pictures”, “Waveforms”, “CSV files” in external memory.
Soft Keys
Six soft keys and the « Up » key provide access to submenus.
Page 13
Instrument Description
4-Channel SPO Digital Oscilloscopes 13
Instrument Description (cont’d)
Input BNCs , Logic Analyzer Probe Connector, USB host, Probe adjust Output, Arbitrary Waveform Generator Output BNC
Channel input BNC : CH1, CH2, CH3, CH4
4 BNCs connectors for input signals.
Logic Analyzer
8-channel logic analyzer input connector
USB A
USB host connector
Cal Probe Adjust
Output
The « Cal » Output allows compensation of 1/10 attenuating probes.
Arbitrary Waveform
Generator Output
Arbitrary Waveform Generator BNC output
Rear Panel Connectors
2 BNCs
USB
Device
LAN
Main
Supply Socket
1 Aux Output BNC :« PASS/FAIL OUT » or « TRIG OUT » according to setting
1 « EXT TRIG » BNC: External Trig Input
USB type B Connector : to connect the oscilloscope to a PC or to a « Pictbridge » printer
RJ45 Connector : to connect the oscilloscope to Ethernet LAN
Main Supply Input Connector
Universal Knob
You can use it with many functions, such as adjusting the holdoff time, moving cursors, setting the pulse width, setting the video line, adjusting X and Y masks when using the pass/fail function etc. The “Universal” knob allows to adjust the storage location of setups, waveforms, pictures when saving/recalling and to select menu options. When the “Universal” button is active the “Select LED” is lit.
Input BNCs
Logic Analyzer
USB A
Cal
AWG out
On/Off
Page 14
Getting started
14 4-Channel SPO Digital Oscilloscopes
Getting started
Verification of instrument operation
To check the operation of the oscilloscope, perform the following steps:
Steps
1. Turn on the oscilloscope. Press the « Default » button of the « RUN CONTROL » pad to load the
« default » configuration of the oscilloscope. Note : The default value for the attenuation compensation coefficient of the probe is 1X.
2. Switch the attenuation of the probe to the « 1X » position, plug it on CH1 and turn clockwise to block. Connect the probe tip and the ground alligator clip to the terminals Cal 3V 1kHz and Ground of the probe adjust output.
3. Press the "Auto Setup" button to display the 3Vpp, 1kHz Cal square wave.
4. Press the "CH1" button to disable CH1. Then press the "CH2" button to enable CH2 and repeat steps 2 and 3 to display the Cal signal on CH2. Repeat the previous steps with the channels CH3 and CH4.
Cal Output : 3V 1kHz
GND
CH1 channel
« Default Setup »
Page 15
Getting started
4-Channel SPO Digital Oscilloscopes 15
Getting started (cont’d)
Oscilloscope Probe
A guard around the probe body protects fingers from electric shocks.
Before performing a measurement : Plug the probe in the “BNC” input of the oscilloscope and then connect the
alligator clip of the probe to the ground reference potential of the circuit under test.
Probe
Attenuation
The probes can have different attenuation factors that affect the vertical scale of the oscilloscope. Press the corresponding channel button (CH 1 or CH2 or CH3 or CH4), and select the factor that matches the attenuation factor of your probe.
Be sure that the attenuation switch on the probe matches the probe factor in the oscilloscope. The probes supplied with the oscilloscope have two attenuation coefficients : X1 et X10.
Note:
● To avoid electric shock when using the probe, keep fingers
behind the probe guard.
● To avoid electric shock when using the probe, do not touch the metal parts of the head of the probe when it is connected to a voltage source. Connect the probe to the oscilloscope input and then connect the probe ground lead to the ground reference of the “circuit under test “before performing any measurement.
Note : The default value of the probe factor is 1X.
Note : When the attenuation switch is in the X1 position, the probe limits the bandwidth of the oscilloscope to about 6MHz (depending on the characteristics of the probe ). To use the full bandwidth of the oscilloscope, be sure that the position of the probe switch is set to X10.
Page 16
Getting started
16 4-Channel SPO Digital Oscilloscopes
Getting Started (cont’d)
Probe Adjust
Warning : The DOX3000 series have a switchable input impedance: 1MΩ - 50Ω. When using a « 1/10 » probe of 10MΩ input impedance (= 9MΩ for the
probe + 1MΩ oscilloscope input) it is imperative to set the oscilloscope input impedance to 1M. If we visualize the « Cal 3V 1kHz » signal with a 1/10
probe and the oscilloscope input impedance is set to 50, we will see on the screen a near zero amplitude signal because instead of having an attenuation of 1/10, we have an attenuation of 1/180000 (50/9000050).
Manual method of compensating the frequency response of a « 1/10 » probe connected to one channel of the oscilloscope.
Operation Steps
1. Open the CH1 menu and set the Probe attenuation factor to 10X and the oscilloscope input impedance to 1M. Set the probe switch to X10 and connect the probe to CH1. If you use the probe hook-tip, insure a proper connection by firmly inserting the tip onto the probe.
2. Attach the probe tip to the Cal 3V 1kHz output and the reference lead to the GND connector. Set CH1 « On » (button lit) and then press the "Auto Setup" button.
3. Check the shape of the displayed waveform :
Over Compensated Under Compensated Correctly Compensated
4. If necessary, adjust the probe variable capacitor with the supplied screwdriver to get a « Correctly Compensated » signal. Repeat when you connect the probe to a different channel or when using another probe.
Note : For more information consult the manual of the probes supplied with the oscilloscope.
Page 17
Functional Description
4-Channel SPO Digital Oscilloscopes 17
Functional Description
AUTO SETUP
Auto Setup
« Auto Setup » button on « RUN CONTROL” block
DOX3000 digital oscilloscopes have an « Auto Setup » function that configures automatically the device to produce a display adapted to the signals at the channel inputs.
After launching an “Auto Setup”, select, if necessary, the type of signal displayed among the four proposed : Several periods, One period, Rising or Falling edge at the center of the screen.
Auto Setup
Function
Option
Description
Multi-cycle
Autoset and displays several cycle signal.
Single-cycle
Autoset and displays single cycle signal.
Rising edge
Autoset and displays the Rising edge.
Falling edge
Autoset and displays the Falling edge.
Undo config.
The oscilloscope recalls the previous Setup.
The “Auto Setup” sets the trigger source and displays the channels according to the following criteria :
If multiple channels have an input signal, channel with the lowest frequency
signal has the priority.
If no channel is « On » and no signal is found, no channels are displayed.
If one or more channels are « On » and no signal is found, only the « On »
channels remains active.
Page 18
Functional Description
18 4-Channel SPO Digital Oscilloscopes
Functional Description
AUTO SETUP (cont’d)
« Auto setup » function
Function
Value
Sampling Mode
unchanged
Display Format
Y-T
Display Type
Vectors
Input Coupling
AC Bandwidth limit
Unchanged
V/div
Set
V/div adjust
Coarse (séquence 1 2 5)
Invert Signal
unchanged
Horizontal Position
Centered
S/div
Set
Trigger Type
Edge
Trigger Source
Auto detection of the signals on the input channels and selection of the proper trigger source.
Trigger Slope
Rising
Trigger Mode
Auto
Trigger Coupling
DC
Holdoff
unchanged
Trigger Level
Adjusted to 50%
« Default » Setup
The oscilloscope is set up for normal operation when it is shipped from the factory. This is the « Default Setup ». To recall this setup, press the « Default » button. Settings may change for options, buttons and controls when you press the « Default » button, refer to appendix B.
The « Default » setup does not change the following settings :
● Language option
● Saved reference waveform files
● Saved setup files
● Display settings
Calibration data
Page 19
Functional Description
4-Channel SPO Digital Oscilloscopes 19
Functional Description
I - VERTICAL System
Vertical knobs and buttons allow to display waveforms and to modify sensitivity and vertical position.
Channels: CH1 CH2 CH3 CH4
page 1
Option
Value
Description
Coupling
DC AC GND
DC: passes the AC and DC components of the input signal. AC: blocks the DC component of the input signal and attenuates AC components below 10Hz. GND: disconnects the input signal.
Bandwidth limit
ON OFF
Limits the channel bandwidth to 20 MHz and reduces high frequency noise. Full bandwidth
Volts/Div
Coarse Fine
Selects the resolution of the Volts/div knob : « Coarse » : defines a 1-2-5 sequence. « Fine » : small steps continuous adjustement.
Probe
.1x, .2x, .5x,1x , 5x, 10x, 20x, 50x,100x, 200x, 500x, 1000x, 2000x, 5000x,10000x
Set to match the attenuation of the probe you are using to ensure correct vertical readouts.
Input Impedance
1M 50Ω
Selects the channel input impedance. « 1M » : for use with 10MΩ input impedance «1/10» probes. « 50Ω » : for use with 50Ω cable and 50Ω output impedance Generator to preserve impedance matching which is necessary when working with high frequency signals.
Next Page
Page 1/2
To access page 2/2
page 2
Option
Values
Description
Unit
V A
Vertical unit : Volts or Ampères.
Invert
ON OFF
To invert the channel. To disable inversion.
Next Page
Page 2/2
To access page 1/2.
CH1 CH2 CH3 CH4
buttons to open vertical menu and set channels
“on”
CH1 CH2 CH3 CH4
« Volt/div » knobs
CH1 CH2 CH3 CH4
Vertical « Position » knobs
Page 20
Functional Description
20 4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL System (cont’d)
Setting up: CH1,CH2, CH3, CH4 channels
Press the CH1 (CH2, CH3 or CH4) button to open the corresponding channel
menu.
Selecting the
channel Input
Coupling
● Pressing CH1 “Coupling”→ “AC”, Set to AC the CH1 input coupling:
The DC component of input signal is blocked.
● Pressing CH1”→ “Coupling”→ “DC”, Set to DC the CH1 input coupling:
Both DC and AC components of input signal pass.
● Pressing CH1”→ “Coupling”→ “GND”, Set the input coupling to GROUND :
The input signal is disconnected.
The CH1 input signal is a positive square wave displayed with DC input coupling:
Proceed similarly to select the input coupling of CH2 - CH3 - CH4
Bandwidth
Limit
● Pressing CH1”→“BW Limit”→ “20Mto enable the channel Bandwidth Limit”.
The sinusoïdal signals (F> 20MHz ) are attenuated.
● Pressing “CH1”→ “BW Limit”→ “Off” disable the channel Bandwidth Limit :
The channel has « full » bandwidth.
DC Input
Coupling
Positive square wave
« 20M » channel BW limited to 20MHz
Page 21
Functional Description
4-Channel SPO Digital Oscilloscopes 21
Injecting a <1ns” rise time “1MHz” square wave on channels CH1 and CH2 . The picture below shows a 1MHz square wave displayed with the bandwidth limit
"20M" on CH2 and with the “full” bandwidth on CH1:
The measured rise time is 1.8ns on CH1 and 17.20ns on CH2, the presence of the
« bandwidth limit » on CH2 leads to a slower rise time.
Page 22
Functional Description
22 4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL System (cont’d)
Adjusting
vertical
sensitivity V/div
Vertical scale adjusting has « Coarse » and « Fine » mode, vertical sensitivity range is 2mV/div to 10V/div.
For example, for CH1:
● Pressing CH1”→ “Volts/Div”→ “Coarse”  sets “V/div” adjust to coarse default value.
Vertical sensitivity can be adjusted in a 1-2-5 sequence from 2mV/div, 5mV/div,
10mV/div to 10V/div.
● Pressing CH1”→“Volts/Div”→ Finesets “V/div” adjust to “fine” This setting allows a continuous adjustement of the vertical sensitivity between coarse steps.
Note : Pressing the vertical sensitivity button also toggles the setting from
« Coarse » to « Fine ».
For example : Use the « Fine » vertical sensitivity adjust to determine the channel bandwidth when the bandwidth limit is « On »:
1° Set the bandwidth limit to 20MHz on CH2 and “Full” bandwidth on CH1 2° Inject on CH1 and CH2 the same 300mVpp, 1kHz” sinusoïdal signal 3° Adjust the CH1 vertical sensitivity to 50mV/div (coarse) and CH2 (Fine) to
35mV/div (35mV= 0.7x50mV).
Fine adjust sensitivity
Bandwidth Limit « 20M » on CH2
21.6MHz
50mV/div
35mV/div
Page 23
Functional Description
4-Channel SPO Digital Oscilloscopes 23
4° Increase the frequency of the generator to obtain a « 6 divisions amplitude » signal on both channels, the frequency measured by the oscilloscope (21.6MHz in our example) is the CH2 cutoff frequency with the bandwidth limit to 20M.
Probe
attenuation
factor
In the CHi menu, select the probe « factor » that matches the attenuation coefficient of the probe used.
To compensate an attenuation coefficient of 1/10, the probe factor must be set to 10X to integrate the attenuation coefficient of the probe in the vertical sensitivity.
For example if you are using a « 1/100 » probe, the probe factor must be set to 100X :
Press “CH1”→Probe” → and select the “100X” factor
Then the vertical sensitivity range will cover : 200mV/div to 1kV/div
Probe Factor “100X”
to compensate a “1/100” probe
Maximum sensitivity 1kV/div
Page 24
Functional Description
24 4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL System (cont’d)
Inverting
Waveforms
For example, on CH1:
● Press CH1”→ ″Page Suiv (Next Page) page 2/2 →“Invert”→“On”:
We show below, the same positive square wave on CH1 with invert « On » and on CH2 with invert « Off »
Input
Impedance
Press “CH1”→ «Impedance » and select the channel input impedance
« 1MΩ » or « 50Ω ».
1M input impedance is necessary when using 1/10 probes:
The 1MΩ” impedance allows to compensate the « 1/10 » probes by adjusting its variable capacitor :« 9MΩ//Var. Cap ».
The « 1M » impedance is particularly well suited to « low frequency » and “high
voltage” signals.
● « 50Ω » impedance : for use with 50Ω cable and 50Ω output impedance Generators
to preserve impedance matching .
“50Ω » impedance is particularly well suited to « High Frequency » and « Low voltage » signals.
● In the example below we show a fast rising edge (<1ns) square wave on CH1 with
an impedance of 50 and 1MΩ 1° CH1 input impedance « 50Ω » :
Page 25
Functional Description
4-Channel SPO Digital Oscilloscopes 25
The rise time = 1.5ns and the amplitude = 286mV 2° CH1 input impedance « 1MΩ » :
The rise time = 2.2ns and the amplitude = 576mV (twice that obtain with « 50Ω ») Conclusion : The « 50Ω » input impedance is best suited to observe fast rising edge
waves provided by 50Ω output impedance generators.
Vertical Unit
V ou A
● Press CH1”→ “Next Page page2/2”→ “Unit” → and select the vertical unit : Volt or Ampere.
Page 26
Functional Description
26 4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL System (cont’d)
2. Vertical Knobs
Vertical Position knobs
1. Use the vertical "Position" knobs to move traces up and down the screen. The vertical position variation range is :
+ /- 1V for vertical sensitivities from 2mV/div to 100mV/div +/- 10V for vertical sensitivities from 102mV/div to 1V/div +/- 100V for vertical sensitivities from 1.02V/div to 10V/div
2. While you adjust the trace vertical position the ″Volts Pos=″ value is displayed on the screen.
3. Push the vertical “Position” knob to set the vertical position to zero (= screen center).
Vertical
Sensitivity
Volts/div
1. Use the “Volt/div” knob to adjust the vertical sensitivity. When you turn clockwise or counter-clockwise the sensitivity increases or decreases.
2. Push the “Volt/div knob, to toogle between Coarse and Fine adjust. ″Coarse″ Vertical sensitivity can be adjusted in a 1-2-5 sequence from
2mV/div, 5mV/div, 10mV/div to 10V/div (12 steps).
Fine a continuous adjustement of the vertical sensitivity from 2mV/div to
10V/div.
3. Save and Recall
Reference Waveform
The REF menu allows to save (or recall) reference waveforms in four internal memorie locations : REFA - REFB - REFC and REFD.
Press the « Ref » button to open the REF menu :
Page 27
Functional Description
4-Channel SPO Digital Oscilloscopes 27
REF Menu
Option
Value
Description
Source
CH1 CH2 CH3 CH4 Math
Select the waveform to save
REFA REFB REFC REFD
Select the REF memory location.
Save
Save the waveform in the selected memory location
REFA/REFB/ REFC/REFD
On Off
Recall and Display the selected REF. Clear the displayed REF waveform.
Operation Steps
1. Press the "Refbutton to open the REF menu.
2. Press the "Source” button to select the reference source : CH1 or CH2 or
CH3 or CH4 or Math.
3. Use the “Position” and “V/div” knobs to set the vertical position and the
sensitivity to the proper values.
4. Press the REF A (or REF B or REFC or REFD) button to select the REF
memory location.
5. Press the « Save » button to save the REF waveform.
6. Select REF A (ou REF B ou REFC ou REFD) Display ON to display the REF
waveform
The reference trace is dispayed in blue color.
Page 28
Functional Description
28 4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL System / MATH Functions
Menu MATH
The « MATH » menu displays the results of mathematicals operations: +, -, *, /, FFT, d/dt, dt and on channels CH1, CH2, CH3, CH4 and REFA, REFB, REFC, REFD waveforms.
Press the « Math » button to open the MATH menu.
MATH Menu
Fonction
Value
Description
Operation
+, -, *, FFT, d/dt, dt and
Select the mathematical function.
Source A Source B
CH1, CH2, CH3, CH4, REFA, REFB, REFC, REFD
Select the Math source Invert
On Off
Invert the MATH waveform. “Invert is disable.
Vertical Position
Use the vertical « Position » knob to adjust the MATH waveform vertical position.
Vertical Scale
Use the « Variable » knob to adjust the MATH vertical scale.
Mathematical
operations
Opération
Value
Description
+
CHi+CHj
CHi added to CHj.
-
CHi-CHj
CHj subtracted from CHi.
*
CHi*CHj
CHi multiplied by CHj.
/
CHi/CHj
CHi divided by CHj.
FFT
Fast Fourier Transform.
d/dt
Derivative
dt
Integral
Square Root
Example of
Math function:
CH1+CH2
Page 29
Functional Description
4-Channel SPO Digital Oscilloscopes 29
Functional Description
I - VERTICAL System / MATH Funtion (Cont’d)
« Fast Fourier Transform »
The Fast Fourier Transform « FFT » process mathematically converts à time domain signal into its components in the frequency domain.
Two measurements are available on the FFT spectrum : Magnitude (in Vrms or dBVrms) and Frequency (in Hz).
Menu FFT
FFT Option
Setting
Description
Operation
FFT
Source
CH1, CH2, CH3, CH4, REFA, REFB, REFC, REFD
Select the FFT source channel.
Window
Hanning, Hamming Rectangulaire Blackman
Select the FFT window type. FFT ZOOM
1X, 2X 5X,10X
Changes the horizontal magnification of FFT display.
Vertical « Scale »
Vrms, dBVrms
FFT vertical scale
FFT « Display »
Split Full Screen
Displays the FFT spectrum on half screen Displays the FFT spectrum on full screen
How to use FFT
To use the FFT Math function, select the time domain signal source Y(t):
● Press « Auto Setup» button to display the Y(t) signal and select Multi cycles.
Push the vertical “Position” knob to reset the vertical position to zero divisions
(screen center).
Use the horizontal Positionknob to position the Y(t) signal portion to be
analyzed with the FFT, in the center of the screen.
The oscilloscope calculates the FFT spectrum using the center 1024 points of
the time-domain waveform.
● Turn the V/divknob to ensure that the entire waveform remains on the
screen.
● Turn the S/divknob to provide the resolution you want in the FFT spectrum.
Operation Sequence
To display the FFT correctly:
1. Push the "Math" button.
2. Set the "Opération" option to « FFT ».
3. Press the "Source" button and select the channel "CH1" or "CH2" or "CH3"
or "CH4" or "REFA" or "REFB" or "REFC" or "REFD".
4. To comply with the Nyquist-Shanon theorem turn the "S/div" knob to adjust the
sampling rate (displayed on top rigth of the screen) at least twice the frequency of the input signal.
5. To determine the appropriate value for the “S/div coefficient, compare the FFT
horizontal scale to the frequency displayed by the hardware counter, on the top rigth of the screen.
Page 30
Functional Description
30 4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL System / MATH Function (cont’d)
2) Displaying the
FFT Spectrum
Press the « Math » button to display the MATH menu. Use the options to select: the Source channel, the Window, and the FFT Zoom Factor. You can display only one FFT spectrum at a time. You can select Full Screen" or "Split" in the Display option to display the waveform and its FFT spectrum on full screen, or the waveform and the FFT on half screen at a time.
With « Split » display and « Cursors » Off, you can display simultaneously: The entire Waveform, a Zoom portion and the FFT spectrum.
FFT Menu: Source - Window - FFT Zoom - Scale - Display
We inject on CH1 a sine wave: frequency = 21.7MHz and amplitude=896mVpp We will use the « Hanning » window to calculate the FFT, to determine the
frequency and the amplitude of the fundamental :
Using the “X1” manual cursor we can determine the frequency (X1=21.8MHz) of the fundamental
Page 31
Functional Description
4-Channel SPO Digital Oscilloscopes 31
And using the Y1 manual cursor we can determine the amplitude (Vrms)
Y1=316mVrms (896mV/2)x0,707 = 316,7mVrms)
3) Selecting the
FFT window
The FFT window reduces spectral leakage in the FFT spectrum. The FFT assumes that the Y(t) waveform repeats indefinitely. With an integer number of cycles, the Y(t) waveform starts and ends at the same amplitude and there are no discontinuities in the signal shape. A non-integer number of cycles in the Y(t) waveform causes the signal start and end to be at different amplitudes. The transition between the start and end points causes discontinuities in the signal that introduce high frequency transients.
Window
Speciality
Adapted to
Rectangular
Best frequency resolution, worst magnitude resolution. This is essentially the same as no window.
Symmetric transients or bursts. Equal amplitude sine waves with fixed frequency. Broadband ramdon noise with relatively slowly varying spectrum.
Hanning Hamming
Better frequency resolution and poorer magnitude accuracy than Rectangular. Hamming has slightly better frequency resolution than Hanning.
Sine, periodic, and narrow­band random noise. Asymmetric transient or bursts.
Blackman
Best magnitude and worst frequency resolution.
Single frequency waveforms, to find higher order harmonics
Page 32
Functional Description
32 4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL System/ MATH Function (cont’d)
FFT:Vertical and Horizontal Scale, Vertical and Horizontal Position
You can magnify and use cursors to take measurements on the FFT spectrum. The oscilloscope includes an "FFT Zoom" option to magnify horizontally, press this button
to select : "1X", "2X", "5X", or "10X". The FFT spectrum can be moved horizontally using the horizontal « Position » knob. The FFT spectrum can be moved vertically using the Math « Vertical » knob. The Math « Variable » knob allows to adjust the FFT vertical scale.
5) Measuring the FFT
spectrum with Cursors
Two types of measurements are possible on the FFT spectrum : Amplitude : in dBVrms or Vrms and Frequency : in Hz
Amplitude
1. Input a sine wave to CH1 and press the "Auto Setup" button then select the “Multi cycles” option.
2. Press the "Math" button to open the "MATH" menu.
3. Press the "Opération" button and select "FFT"
4. Press the "Source" button and select "CH1"
5. Press the CH1button to open the CH1 menu.
6. Turn the "S/div" knob to adjust the sample rate (at least the double than the input signal frequency).
7. If the FFT display is on full screen, press CH1 button again to remove CH1 waveform from display.
8. Press the "Cursors" button to open the CURSORS menu.
9. Press the cursor Mode button and select Manual”.
10. Select the « Yi » cursor type
11. Press the "Source" button and select "MATH"
12. Select the "Y1" cursor, and use the "Universal" knob to move Y1 to the highest amplitude component of the FFT spectrum.
13. Select the "Y2" cursor, and use the "Universal" knob to move “Y2” to the lowest amplitude component of FFT spectrum.
14. The amplitude (∆V) displays on the top left of the screen.
In the following example we will consider a composite periodic signal : fundamental frequency 5MHz and 20MHz - 35.6MHz spectral components
The FFT displays the fundamental (F=5MHz Amplitude =2.66 Vrms) and two spectral components at 20MHz and 35.6MHz of Amplitude=0.4Vrms
Using Yi horizontal cursors to measure the amplitude of spectral components
Page 33
Functional Description
4-Channel SPO Digital Oscilloscopes 33
Functional Description
I - MATH Function (cont’d)
Measure
frequency with
« Xi » Vertical
Cursors
1. Press the « Cursors » button
2. Press cursor Modebutton and select Manual”.
3. Select "Xi" vertical cursor type
4. Press the "Source" button and select "MATH".
5. Select the "X1" cursor, and use the "Universal" knob to move “X1” to the highest amplitude component of the FFT spectrum.
6. The value X1=5.00MHzdisplayed on the top left of the screen is the frequency of the fundamental of the CH1 input signal.
Use vertical cursors to measure the frequency of the spectral components
Fundamental frequency X1=5MHz and the harmonic X2=35.6MHz
Frequency of the two harmonics: X1=20MHz and X2=35.6MHz
Page 34
Functional Description
34 4-Channel SPO Digital Oscilloscopes
FFT of « Cal 3V 1kHz »
1° Enter the « CAL 3V 1kHz » signal on CH4 with a 1/1 probe 2° Channel CH4 coupling = DC 3° Select 1.4Mpts memory depth and a 500ms/div Time Base 4° Select the FFT Math function with a « Blackman » window (highest
amplitude resolution) and FFT Zoom = « X10 » so that the horizontal scale is 1kHz/div.
5° Set the FFT vertical scale to Vrms 6° Use the manual cursors to measure the amplitude of the fundamental
(1kHz) and the first harmonic (3kHz) :
The fundamental (1kHz) amplitude is Y1=1.12Vrms (close to 1.06Vrms =
1.5V /1.414) Note : The « Cal 3V 1kHz » output provides a positive square wave, if the
input coupling is « DC » we observe a DC component in the FFT spectrum. Set the input coupling to « AC » to block the DC component of the « CAL 3V
1kHz » signal to obtain a FFT spectrum without DC component :
Warning : The presence of a DC component can hinder the observation of low-frequency components of the signal.
With FFT math function, the use of AC input coupling is recommended.
Page 35
Functional Description
4-Channel SPO Digital Oscilloscopes 35
Functional Description
II - HORIZONTAL System
HORIZONTAL Menu
The HORIZONTAL pad contains two buttons (Horiz and Zoom) and two knobs (S/div and Horizontal Position).
HORIZONTAL
Menu
« Horiz » button of the HORIZONTAL pad
Option
Setting
Description
Format
YT
XY
ROLL
«YT» : Displays the voltage (on the vertical axis) versus time (on the horizontal axis).
«XY» : To display in « XY » mode the two channel couples [CH1(X), CH2(Y)] and [CH3(X), CH4(Y)]
Time Base coefficient >100ms. The traces scroll from the right to the left of the screen as in a recorder, the “Roll” format is suitable for slow signals.
In « ROLL » there is no trigger or horizontal position control. Warning : When working with very low frequency signals
(<10Hz) the DC input coupling is not recommended.
ROLL mode
example
Note : The hardware frequency counter does not work for frequencies <10Hz
S/div knob
Horiz and Zoom menu buttons
Horizontal Positon knob
Page 36
Functional Description
36 4-Channel SPO Digital Oscilloscopes
Functional Description
II - HORIZONTAL System (cont’d)
HORIZONTAL pad
The HORIZONTAL pad knobs allow to adjust the Time Base coefficient « S/div » and the traces horizontal Position.
The two buttons allow to open the HORIZONTAL menu (“Horiz”) and to activate the “Zoom” function.
Horizontal
"POSITION" knob
1. Adjusts the trace horizontal position (the trigger position relative to the center of the screen). The time resolution of this command depends on the selected Time Base coefficient.
2. To reset to zero the horizontal position, push the Position knob.
S/div knob
1. Adjusts the Time Base coefficient. When the acquisition is stopped ( with RUN/STOP” button or in “SINGLE” mode), turn the « S/div » knob to expand or compress the waveform.
2. « S/div » knob adjusts the coefficient of the « Main » or « Zoom » Time Bases. In Zoom mode, changing the Time Base coefficient (S/div) modifies the width of the « Zoom » window.
Horizontal Zoom
Use the horizontal Zoom” function to define the portion of the waveform to be observed in details. The slowest « zoom » time base coefficient may not be lower than the “main” time base coefficient. Use the horizontal « Position » knob to move the « zoom window » position and the « S/div » knob to magnify or compress the “zoom window”.
"M" stands for « Main » time base, "Z" stands for “Zoom” time base. The vertical « blue » arrows indicate the horizontal position of the trigger
event in the Main time base and in the “Zoom” window. The horizontal « yellow » arrow indicate the vertical position of the trigger
level. Note: The display of the « Trigger Position » symbol in the “zoomed window” is
only possible if the position of the “zoomed window” includes the trigger event. If the trigger event is outside the “zoomed window” the horizontal blue arrow
indicates the direction in which to move the “zoomed window” to display the
trigger.
Page 37
Functional Description
4-Channel SPO Digital Oscilloscopes 37
Operation Steps
To display in details a portion of the waveform:
1. Press the "Horiz" button to open the HORIZONTAL menu.
2. Turn the "S/div" knob to adjust the main timebase.
3. Press the « Zoom » button to activate the horizontal Zoom.
4. Turn the « S/div » knob to adjust the “Zoom” time base.
5. Turn the horizontal "Position" knob to move the zoom window.
In the example below, we display in « Zoom » mode, a detail (a 0.24µs wide window with a resolution of 0.5ns situated at 235µs from the trig event) of a signal acquired at 2GSPS sample rate with a memory depth of 28Mpoints. The oscilloscope has recorded the evolution of the signal during a lapse of time of 14ms with an elementary time step of 0.5ns : The arrows indicates the delay value (Delay=235µs) relative to the trigger event and the signal value for this delay :
We can « Stop » the trace refresh (« Run/Stop » button) to observe it in detail. For example we can determine the value that the signal has 74ns before the trigger event by moving the zoomed window (Horizontal « Position » knob): The red arrows indicate the value of the signal at the trigger event and 74ns before :
Page 38
Functional Description
38 4-Channel SPO Digital Oscilloscopes
Functional Description
III - Trigger System
TRIGGER Menu
The DOX3000 oscilloscope series have a digital trigger system that has the following advantages:
Precise Trigger - Low « Jitter » - High Sensitivity Precise Trigger timing < 1ns - Configurable noise reject High stability with temperature
A hardware counter allows to display the frequency of the trigger source signal in the top right of the screen
The « Setup » button gives access to the « Trigger » setting menu.
The trigger symbol indicates when the trigger event has occurred. The trigger event divides the acquisition memory in two parts:
- The « Pre-Trig Buffer » before the trigger event and
- The « Post-Trig Buffer » after the trigger event
The DOX3000 oscilloscope series have eleven trigger modes: Edge, Slope, Pulse, Video (HDTV), Window, Interval, Dropout, Runt, Pattern, Serial 1, Serial 2 (Serial bus: I2C, UART, SPI, CAN, LIN)
TRIGGER pad
 « Setup » :Press the « Setup » button to open the TRIGGER menu.  “Level” knob: Use the LEVEL knob to adjust the trigger level. Push the
"Level" knob to set the trigger level to 50% of the peak to peak amplitude of the signal. This function is very useful in “Single” or “Normaltrigger modes.
“Auto” button : The waveform is refreshed even in the absence of a trigger
event.
« Normal » button: in « Normal » mode the trace is refreshed only if a
trigger event occur.
«Level» knob push to force level to 50%
« Setup » button for Trigger setting
« Single » button
« Auto » button
« Normal » button
Page 39
Functional Description
4-Channel SPO Digital Oscilloscopes 39
"Single button: Press the "Single" button to activate the SINGLE mode. In
this mode only one acquisition is allowed at a time.
Pre-trig/Post-trig/Trig-Delay: Datas before and after the trigger event . If
the trigger symbol is at the center of the screen, the portion of the traces corresponding to the first 7 divisions represent the pre-trig and the 7 following divisions the post-trig. This function is very useful because you can observe the samples that occur before and after the trigger event. Everithing that is to the right of the trigger event corresponds to the post-trig.
Trigger SOURCE
In « Edge » mode : the trigger source may be the signal to the following inputs : CH1, CH2, CH3, CH4, EXT, EXT/5 or AC line.
For the other trigger modes (Slope, Pulse, Vidéo, Window, Interval, DropOut, Runt, Pattern and Serial 1 & 2) the 4 possible trigger sources are: CH1, CH2, CH3 and CH4.
Page 40
Functional Description
40 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d)
1. Edge
Trigger « Edge »
Option
Settings
Description
Type
Edge
The rising or falling edge of the signal source is used to trigger.
Source
CH1 CH2 CH3 CH4
Note : The trigger source is active even if the channel is off.
EXT
The EXT TRIG source signal is not displayed. The EXT source is the signal to the “EXT TRIG” input BNC in the back of the oscilloscope. The trigger level adjustement range is: ­1,6V to +1,6V.
EXT/5
Same as EXT option, but the signal is attenuated by a factor of 5. The trigger level range is multiplied by 5 : -8V to +8V.
AC line
The trigger source signal is the AC line. The trigger coupling is set to « DC » and the trigger level to 0V.
Slope
Trigger on the rising edge of the source signal.
Trigger on the falling edge of the source signal. Trigger on the rising and falling edges of the source signal.
Holdoff
Adjustable Fixed
Use the « Universal » knob to adjust the « holdoff ». The « holdoff » is fixed to the latest set value
Coupling
DC
All components of the signal source passes
AC
Blocks the DC component and attenuates the AC signal below 5.8Hz.
Low Pass
Filter
HF reject
Attenuates « high frequency » components above
1.27MHz.
High Pass
Filter
LF reject
Blocks the DC component and attenuates the « low
frequency » components below 2.08MHz.
Noise Reject
On Off
Enable or Disable the Noise Reject
Operation Steps
1. Set the trigger type
1) Press the “Setup » button to open the TRIGGER menu.
2) Press the "Type" button and select "Edge".
2. Set the trigger Source
Press the "Source" button and select the trigger source: "CH1", "CH2", "CH3", "CH4", "EXT",
"EXT/5" or "AC Line".
3. Set the trigger slope
Press the « Slope » button and select the trigger slope “ ”, “ ” or “ ”.
4. Set the Holdoff
Press the « Holdoff close » button « Holdoff time » and use the “Universal” knob to adjust
the « Holdoff » time.
5. Set the Trigger coupling
a. Press the “Set Up” button to open the trigger settings menu. b. Press the “Coupling” button and select the coupling type: “DC”, “AC”, “HF Reject” ou “LF
Reject”.
Page 41
Functional Description
4-Channel SPO Digital Oscilloscopes 41
Functional Description
III - TRIGGER System (cont’d)
2. «Pulse»
Pulse
Trigger page 1
Use the Pulse type to trigger on a particular “pulse width” with respect to the entire signal.
Option
Settings
Description
Type
Pulse
Select the trigger pulse type and width.
Source
CH1, CH2, CH3, CH4
Select the trigger source.
Polarity
Select the pulse polarity : Positive or Negative
Limite Range
< (Pulse width less than pulse width setting)
> (Pulse width larger than pulse width setting)
<> (Pulse width outside the range of set values)
>< (Pulse width in the range of set values)
Select the condition that the pulse width must meet to obtain a trigger event.
Set the width
2,0ns to 4,2s
Use the « Universal » knob to set the pulse width.
Next page 1/2
Press this button open page 2/2
Page 42
Functional Description
42 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d)
Pulse Trigger
page 2
Option
Settings
Description
Type
Pulse
Coupling
DC AC LF Reject HF Reject
Select the trigger coupling.
Noise Reject
On
Off
Noise Reject enable
Noise Reject disable
Next Page 2/2
Press this button to open the page 1/2
Operation Steps
1. Select the type
1) Press the « Setup » button to open the TRIGGER” menu.
2) Press the "Type" button and select "Pulse"
2. Set the condition
Polarity : ”, ”,
Press the Limit Range to select the condition
“< ”, “ > ”, “< >” or “> <”
3. Set the pulse width
Use the "Universal" knob to set the pulse width limit.
Page 43
Functional Description
4-Channel SPO Digital Oscilloscopes 43
Functional Description
III - TRIGGER System (cont’d)
«VIDEO»Trigger
To trigger on « Lines » or « Frames » of standard video signals.
Video Trigger
Option
Settings
Description
Type
Video
To trig on video signals: NTSC, PAL/SECAM, HDTV and custom.
Source
CH1, CH2, CH3, CH4
Select the trigger source.
Standard
NTSC - PAL/SECAM 720p/50 - 720p/60 1080p/50 - 1080p/60 1080i/50 - 1080i/60 Custom
Select the video standard
Sync
Any line Line Number Field
Choose the appropriate video trigger:
Any line - Line Number - Field 1 - 2
Operation
Steps
1. Set the type
1) Press the Setup button to open the TRIGGER menu.
2) Press the "Type" button and select "Video"
2. Set the Sync
1) Press the Sync” button to select “Any line , or Line Number (select), Field 1 or 2.
2) If you choose « Select », use the “Universal” knob to set the line number.
3. Set the Standard
1) Press the “Standard” button and select “PAL/SECAM” or “NTSC” or ....
Page 44
Functional Description
44 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d)
« Slope »
Trigger on a positive or negative slope whose duration is specified.
« Slope » Trigger
page 1
Option
Settings
Description
Type
Slope
Trigger on a positive or negative slope of specific duration.
Source
CH1, CH2 CH3, CH4
Select the trigger source.
Slope
Positive slope
Negative slope
Limit Range
< , > < > > <
Select the trigger condition on slope duration.
< Less than > Greater than < > Outside the limited range
> < Inside the limited range
Duration
Use the "Universal" knob to adjust the slope duration in the range: 2 ns to
4.20s.
Next Page
1/2
Page 45
Functional Description
4-Channel SPO Digital Oscilloscopes 45
Functional Description
III - TRIGGER System (cont’d)
Slope Trigger
page 2
Option
Settings
Description
Type
Slope
Level
Lower Upper
Select the level (L1 or L2) to adjust with the trigger "Level" knob. You can adjust the two levels 'L1' & "L 2" that define the slope.
Coupling
DC - AC LF Reject HF Reject
Select the trigger coupling
Noise Reject
On Off
NextPage
Page 2/2
Press this button to open page 1/2
Operation Steps
1. Enter the signal on CH1 or CH2 or CH3 or CH4.
2. Press the Auto button.
3. Press the Setupbutton to open the Trigger Menu”.
4. Press the Typebutton and select "Slope".
5. Press the Sourcebutton and select “CH1” or “CH2” or CH3 or CH4.
6. Press the Limit Rangebutton and select the condition: <”, “>”, “<>”, “><”.
7. Press the Timebutton and use the “Universal” knob to adjust the slope duration.
8. Press the Next Pagebutton to open page 2/2.
9. Press the Lower Upper button to select the level to adjust L1 or L2.
10. Use the Level knob to adjust L1 or L2.
Page 46
Functional Description
46 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d)
« Window » trigger
When the « window » trigger type is enabled, the oscilloscope triggers when the signal exits, the window defined by the two levels L1 and L2, from the top or bottom level.
Window Trigger
page 1
Option
Settings
Description
Type
Window
The window is defined by two voltage levels L1 and L2.
Source
CH1, CH2, CH3, CH4
Select the trigger source CH1 or CH2 or CH3 or CH4
Window type
Absolute Relative
Absolute: the two levels are set independently Relative: we adjust the delta on either side of a
central value.
Lower Upper Center Delta
The trigger window is define by the lower and upper value or by the delta relative to a central value.
Coupling
DC AC LF Reject HF Reject
Trigger source coupling
Noise Reject
ON Off
Enable (On) or Disable(Off) the noise reject
Page 47
Functional Description
4-Channel SPO Digital Oscilloscopes 47
Functional Description
III - TRIGGER System (cont’d)
Interval
Trigger
page 1
Option
Settings
Description
Type
Interval
In this mode the trigger event occur on the second edge, when the time interval between two consecutive edges is : < or > or <> or >< to the set value.
Source
CH1, CH2, CH3, CH4
Select the trigger source CH1 or CH2 or CH3 or CH4
Slope
Select the slope (Rising or Falling) that starts the interval
« Limit Range »
< , >
< > , > <
Trigger Condition:
< Less than > Greater than < > Outside the window > < Inside the window
Duration
Range : 2ns à 4.20s
Next Page
1/2
To open page 2/2
« Interval »
Trigger page 2
Option
Settings
Description
Type
Interval
Coupling
DC - AC - LF Reject - HF Reject
Trigger source coupling.
Noise Reject
On Off
Enable or Disable noise reject
Next page
2/2
open page 1/2
Page 48
Functional Description
48 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (Cont’d)
« DropOut »
trigger page 1
Settings
Description
Type
DropOut
A trigger event occurs if the signal disappears for a time longer than the specified « DropOut » duration.
Source
CH1, CH2 CH3, CH4
Select the trigger source CH1 or CH2 or CH3 or CH4
Slope
Select the edge (state) that reset the dropout duration.
Over Time type
Edge State
Duration
« DropOut » duration range : 2ns to 4.20s
Next Page
1/2
open page 2/2
« DropOut »
page 2
Option
Settings
Description
Type
DropOut
Coupling
DC AC LF Reject HF Reject
Trigger source coupling.
Noise reject
On Off
Enable or disable noise reject.
Next Page
2/2
Open page 1/2
Page 49
Functional Description
4-Channel SPO Digital Oscilloscopes 49
Functional Description
III - TRIGGER System (cont’d)
« Runt »
Trigger
page 1
Option
Settings
Description
Type
Runt
A trigger event occurs if the pulse (negative or positive) crosses the first level (L1) but not the second level (L2) of the window, before recrossing the (L1) level in a given time.
Source
CH1, CH2, CH3, CH4
Select the trigger source CH1 or CH2 or CH3 or CH4
Polarity
Select the pulse polarity : Positive or Negative
Limit Range
< , >
< > , > <
Select the limits for « Runt »
< Less than > Greater than < > Outside the window > < Inside the window
Duration
« Runt » duration range : 2ns - 4.20s
Next Page
1/2
Open page 2/2
« Runt »
Trigger
page 2
Option
Settings
Description
Type
Runt
Time
Use the "Universal" knob to adjust the Runt duration in the range : 2ns to 10s.
Lower - Upper
Use the Level knob to adjust the Lower and
Upper levels 'L1' and "L2".
Coupling
DC AC LF Rej HF Rej
Trigger source coupling.
Noise Reject
On Off
Enable or disable the « Noise Reject ».
Next Page
2/2
Open page 1/2
Page 50
Functional Description
50 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d)
« Pattern »
Trigger
without the
« Digital » option
page 1
Option
Settings
Description
Type
Pattern
Triggers when the «pattern» condition goes from false to true.
We can also adjust the time duration of the pattern “true” condition.
Note : Input coupling must be : DC.
Inputs CHi
CH1 CH2 CH3 CH4
Select the Pattern inputs and the active level :
Low High Invalid
Next Page
Page 1/2
Open page 2/2
Trigger type
« Pattern »
Without the
Digital option
page 2
Option
Settings
Description
Type
Pattern
Logic
AND, OR NAND, NOR
Fonction
Limit Range
< > < > > <
« True » pattern durantion condition limit
Duration
Use the "Universal" knob to adjust the duration of the pattern « true » condition in the range: 2 ns to 4.20s.
HoldOff Time
Setting range : 100ns to 1.50s.
« Next page »
2/2
open page 1/2
Page 51
Functional Description
4-Channel SPO Digital Oscilloscopes 51
« Pattern » with « Digital » option
page 1/4
Option
Settings
Description
Type
Pattern
Triggering when the «pattern» condition goes from false to true.
We can also adjust the time duration of the pattern “true” condition.
Note : Input coupling must be : DC.
Inputs CHi
CH1 CH2 CH3 CH4
Select the Pattern inputs and the active level :
Low High Invalid
Next Page
Page 1/2
Open page 2/2
We inject on CH4 the « Cal 3V 1kHz » signal and on the 8bit digital bus square waves of frequency 68Hz.
Page 52
Functional Description
52 4-Channel SPO Digital Oscilloscopes
« Pattern » with « Digital » option
page 2/4
Option
Settings
Description
Type
Pattern
Digital inputs : select the logic type to set automatically the threshold :
TTL - CMOS - LVCMOS3.3 - LVCMOS2.5 or Custom
Di Inputs
D0 D1 D2 D3
Select the pattern inputs and the active level: Low High Invalid
Next Page
Page 2/4
Open page 3/4
Pattern with the
Digital option
page 3/4
Option
Valeurs
Instruction
Type
Pattern
Digital inputs : select the logic type to set automatically the threshold :
TTL - CMOS - LVCMOS3.3 - LVCMOS2.5 or Custom
Di Inputs
D4 D5 D6 D7
Select the pattern inputs and the active level: Low High Invalid
Next Page
Page 3/4
Open page 4/4
Pattern with the
Digital option
page 4/4
Option
Settings
Description
Type
Pattern
Logic
AND, OR NAND, NOR
Function
Limit Range
< , > < > , > <
True pattern duration limit
Durée
Use the "Universal" knob to adjust the « true » pattern duration within the range : 2 ns à 4.20s.
HoldOff
Setting range : 100ns to 1.50s.
Next page
4/4
Open page 1/4
More details  § XI - DIGITAL, option.
Page 53
Functional Description
4-Channel SPO Digital Oscilloscopes 53
Functional Description
III - TRIGGER System (cont’d) ()
« Serial1-2 »
Trigger
Option
Settings
Description
Serial1
I2C
I2C is the default setting on Serial 1.
Condition
Start - Stop - Restart ­No Ack
Condition
EEPROM
Limit Range
= , > , <
Data1
0xXX
Condition
7 Addr & Data
Addr, Data 1, Data 2
0xXX, 0xXX, 0xXX
R/W bit
Write - Read - Dont Care
Condition
10 Addr & Data
Addr, Data1, Data2
0xXX , 0xXX , 0xXX
R/W bit
Write - Read - Dont Care
Condition
Data Length
Address
7bit 10bit Byte Length
1 to 12 adjustable
Page 54
Functional Description
54 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d) ()
Trigger
« Serial 2 »
Option
Settings
Description
Serial 2
SPI
SPI is the default setting on Serial 2
Trigger Setting
SPI Trigger Setting
Trigger type
MOSI MISO
Data Length
4 to 96 ajustable
Bit Roll
0 to 95
Bit Value
0 1 X
All Same
0 1 X
Bit Order
MSB LSB
()
Further informations : § DECODE : Decode bus SPI/I2C/UART/LIN/CAN
Page 55
Functional Description
4-Channel SPO Digital Oscilloscopes 55
Functional Description
III - TRIGGER System (cont’d)
Trigger
« Serial1-2 »
Option
Settings
Description
Serial1
UART/RS232
Trigger on UART/RS232 serial bus.
Trigger Setting
Source Type
RX TX
Condition
Start - Stop - Data - Parity error
Bus Configure
Baud
600-1200-2400-4800-9600-19200-38400-57600­115200-Custom
Data Length
5 - 6 - 7 - 8
Parity Check
None - Odd - Even
Stop Bit
1 - 1.5 - 2
Idle Level
Low - High
Page 56
Functional Description
56 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d)
Trigger
« Serial 1-2 »
Option
Settings
Description
Serial 1
CAN
Trigger on CAN serial Bus
Trigger Setting
CAN TRIG SET
Start - Remote - ID - ID+DATA - Error
Bus Configure
Baud
5kb/s-10kb/s-20kb/s-50kb/s-100kb/s-125kb/s-250kb/s­500kb/s-800kb/s-1Mb/s-Custom
Page 57
Functional Description
4-Channel SPO Digital Oscilloscopes 57
Functional Description
III - TRIGGER System (cont’d)
Trigger
« Serial1-2 »
Option
Settings
Description
Serial1
LIN
Trigger on LIN serial Bus.
Trigger Setting
Condition
Break - ID - ID+DATA - Data Error
Bus Configure
Bit Rate
600-1200-2400-4800-9600-19200-Custom
Page 58
Functional Description
58 4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER System (cont’d)
Trigger COUPLING
Use the appropriate coupling to the trigger source signal.
To select the trigger coupling, first press the "Setup" button to open the trigger menu and select the trigger type : "Edge", "Pulse", Video", "Slope", "Window", "Interval", "DropOut" or "Runt" then press the “Coupling” submenu and select the coupling of the trigger source: CH1 or CH2 or CH3 or CH4.
Four types of coupling are available : AC - DC - LF Reject - HF Reject
Horizontal POSITION
The horizontal « Position » knob allows to adjust the interval of time between the trigger event and the screen center. The horizontal "Position" knob allows to adjust the display portion of the waveform before or after the trigger event . The
horizontal position reference is the center of the screen, the ″Position (Delay :) is positive on the left side of the screen and negative on the right side.
Trigger Slope & Trigger Level
The "Slope" and "Level" control define the trigger point. The "slope" polarity (Edge trigger type only) determines which signal edge (Rising or
Falling or both) causes the trigger. The Level knob controls the vertical position (voltage) of trigger threshold.
Falling edge
Rising edge
T
r
i
g g
e r l
e v
e l c
a n
b
e a
d j
u
s
t
e d
v
e
r
t
i
c
a l
l
y
Note :
Use the SINGLE shot mode to capture a single signal (not
repetitive).
The trigger source coupling does not affect the bandwidth
of the displayed channel.
Page 59
Functional Description
4-Channel SPO Digital Oscilloscopes 59
Functional Description
III - TRIGGER System (cont’d)
HOLDOFF
Triggering Inhibition time
Use the « Holdoff » to obtain a stable display in the case of complex signals such as pulse bursts. The HoldOff represents the time interval between the detection of a valid trigger event and the moment when the oscilloscope is ready to detect a new trigger event. The oscilloscope will not trigger during the Holdoff time, because the trigger circuit is inhibited. With "pulse bursts", set the Holdoff equal to the burst length, so that the oscilloscope will trigger on the first pulse of the burst masking the other pulses.
Holdoff time
Trigger position
Trigger level
Operation
Steps
To set the Hold-Off :
1. Press the Setupbutton to open the TRIGGER Menu.
2. Press the Typebutton and select the trigger type.
3. Press the Holdoff button and use the ″Universal knob to adjust the Holdoff.
4.
Set the HoldOff time to obtain a stable waveform.
Note : With periodic complex waveforms use the « Holdoff » to stabilize the
display.
Example : We display on CH3 with the option Color « On » a periodic signal (generated by the DOX3304 AWG) having two rising edges separated by 392µs (measured with the manual cursors ΔT= X2-X1= 390µs - (-2µs) = 392µs).
Color : is « On » (Color grading: Red corresponds to frequent points and Blue to
rare points) and the “Edge” trigger type.
If the holdoff is less than 392µs we observe double triggers instability : The
most frequent triggering is on the first rising edge (Red trace) and rarely on the second rising edge (Blue trace).
If the holdoff is greater than 392µs the trigger event is always on the same
edge the oscilloscope displays only a red trace.
Page 60
Functional Description
60 4-Channel SPO Digital Oscilloscopes
Holdoff =
355µs<392µs
the oscilloscope
displays two
waveforms
resulting from a
double triggering
(Blue and Red
waveforms)
Holdoff = 393µs
> 392µs the
oscilloscope
display only the
red waveform
This holdoff
value prevents
the double
triggering
Page 61
Functional Description
4-Channel SPO Digital Oscilloscopes 61
Functional Description
IV - ACQUISITION System
Acquisition menu
« ACQUIRE »
When acquiring an analog signal, the oscilloscope converts it into digital format and displays a waveform. The acquisition mode defines how the signal is digitized. The time base setting multiplied by the number of horizontal divisions of the screen determines the duration of the recording in seconds (S/div x 14div), and the elementary step of acquisition is given by the sampling period. The Memory Depth, the Sample Rate and the Waveform Recorded Length are related by the following formula:
Memory Depth (Nb Recorded Samples) = Sampling rate ( Samples/sec) x Waveform Recorded Length (in sec)
The oscilloscope adjusts the current memory depth according to the the selected time base coefficient : For example we set the maximum memory depth to 14Mpts (Menu «Acquire» Sub-menu « Memory Depth ») and the channel CH1 “On” (all other channels are Off), the oscilloscope will adjust the “Current” memory depth according to the selected Time Base coefficient:
S/div
Sample Rate
Memory depth (Samples)
1ns
2GSPS
28 Samples
2ns
2GSPS
56 Sa
5ns
2GSPS
140 Sa
10ns
2GSPS
280 Sa
20ns
2GSPS
560 Sa
50ns
2GSPS
1.4k Sa
100ns
2GSPS
2.8k Sa
200ns
2GSPS
5.6k Sa
500ns
2GSPS
14k Sa
1µs
2GSPS
28k Sa
2µs
2GSPS
56k Sa
5µs
2GSPS
140k Sa
10µs
2GSPS
280k Sa
20µs
2GSPS
560k Sa
50µs
2GSPS
1.40M Sa
100µs
2GSPS
2.80M Sa
200µs
2GSPS
5.60M Sa
500µs
2GSPS
14.0M Sa
1ms
2GSPS
28.0M Sa
2ms
1GSPS
28.0M Sa
5ms
0.4GSPS
28.0M Sa
10ms
200MSPS
28.0M Sa
20ms
100MSPS
28.0M Sa
50ms
40MSPS
28.0M Sa
100ms
20MSPS
28.0M Sa
200ms
10MSPS
28.0M Sa
500ms
4MSPS
28.0M Sa
1s
2MSPS
28.0M Sa
2s
1MSPS
28.0M Sa
5s
400KSPS
28.0M Sa
10s
200KSPS
28.0M Sa
20s
100KSPS
28.0M Sa
50s
40KSPS
28.0M Sa
Page 62
Functional Description
62 4-Channel SPO Digital Oscilloscopes
Functional Description
IV - ACQUISITION System (cont‟d)
« Acquire »
Press the "Acquire" button to open the acquisition menu.
Acquisition
Settings
Option
Settings
Description
Acquisition
Normal
This mode is used for an accurate sampling and display of most signals.
Peak Detect
This mode is used to display noise and to reduce aliasing.
High Res
To increase the vertical resolution and to reduce random noise.
Average
To reduce the random noise on displayed waveforms. Average: 4, 16, 32, 64,128, 256, 512, 1024
Sinx/x
Sinx
X
Sinusoidal Interpolation "Sinx" Linear Interpolation "x"
Sequence
Acq Mode
Off Normal Single
Display
Mode
Overlay Waterfall
Frame Set
Setting range : 1 to 80000 (The max value depends on the time base coefficient S/div and the number of active channels)
Mem Depth
7k - 14k ­70k - 140k ­700k - 1.4M
- 7M - 14M
Select the maximum Memory Depth value
Page 63
Functional Description
4-Channel SPO Digital Oscilloscopes 63
Functional Description
IV - ACQUISITION System (cont‟d)
Sampling Modes
The oscilloscope has the following sampling modes: Normal - Peak Detect - Average - High Resolution
Normal
The oscilloscope samples the signal at regular intervals to build the digital waveform. In general this mode is a true representation of the signal.
Advantage
It is the most common sampling mode. This mode reduces the random noise.
Disadvantage
The "Normal" mode does not capture fast signal changes that may occur between to successive samples. An aliasing phenomenon can result and the short pulses may not be detected. In the presence of short pulses (glitch) you must use the Peak Detect mode.
« Peak Detect »
The « Peak Detect » mode capture the minimum and maximum values of the signal (Oversampled waveform) between two successive samples and displays two Waveforms, one with the Low values and the other with the High values.
Advantage
In this mode, the oscilloscope can capture fine pulses (glitch) which have not been detected by the "Normal" mode.
Disadvantage
The trace thickness (noise) will be higher in this mode.
Page 64
Functional Description
64 4-Channel SPO Digital Oscilloscopes
Functional Description
IV - ACQUISITION System (cont‟d)
Average Mode
In this mode, the oscilloscope performs Multiple Waveforms acquisitions, average them and displays the resulting 1 Waveform.
Advantage
Use this mode to reduce the random noise.
A noisy sine wave acquire in « Normal » and « Average » (128) modes
High Res mode
High Res
The High Res mode uses « on the fly » over-sampling and averaging. Each displayed sample point represents the average value of the samples acquired in the elementary time interval. The «High Res» mode increases the effective vertical resolution of the oscilloscope. The High Res mode reduces the signal random noise.
The High Resolution mode can be used even with SINGLE mode because the averaging takes place at each point displayed and not on successive waveforms acquisitions as in the « Average » mode.
Page 65
Functional Description
4-Channel SPO Digital Oscilloscopes 65
Functional Description
IV - ACQUISITION System (cont‟d)
Sequence Mode
In the Sequence mode, the recording memory is segmented. The oscilloscope captures and records a waveform (Frame) at each trigger event. When the acquisition sequence is completed, the oscilloscope displays the first 20 acquired waveform segments. In this mode, the waveform capture speed can reach up to 300000 waveforms/sec. When a trigger event occurs, the oscilloscope acquires and records a waveform (segment) and then waits for a new trigger event. The oscilloscope blind time” is minimized allowing the capture of certain signal details, that were not visible with other acquisition modes.
For a 14Mpts maximum memory depth and with all four channels activated, the table below shows the maximum number of Frames according to the « Time Base » coefficient :
Time Base Coefficient
Sampling
Frequency
Current memory
Maximum
Frame Set
1ns/div
1.00GSPS
14pts
80 000
2ns/div
1.00GSPS
28pts
80 000
5ns/div
1.00GSPS
70pts
80 000
10ns/div
1.00GSPS
140pts
80 000
20ns/div
1.00GSPS
280pts
80 000
50ns/div
1.00GSPS
700pts
80 000
100ns/div
1.00GSPS
1.4kpts
63 157
200ns/div
1.00GSPS
2.8kpts
33 519
500ns/div
1.00GSPS
7kpts
16 042
1µs/div
1.00GSPS
14kpts
8 108
2µs/div
1.00GSPS
28kpts
4 067
5µs/div
1.00GSPS
70kpts
1 630
10µs/div
1.00GSPS
140kpts
815
20µs/div
1.00GSPS
280kpts
408
50µs/div
1.00GSPS
700kpts
163
100µs/div
1.00GSPS
1.4Mpts
81
200µs/div
500MSPS
1.4Mpts
74
500µs/div
200MSPS
1.4Mpts
74
1ms/div
100MSPS
1.4Mpts
74
2ms/div
50MSPS
1.4Mpts
74
5ms/div
20MSPS
1.4Mpts
74
10ms/div
10MSPS
1.4Mpts
74
20ms/div
5MSPS
1.4Mpts
74
50ms/div
2MSPS
1.4Mpts
74
100ms/div
1MSPS
1.4Mpts
74
200ms/div
500kSPS
1.4Mpts
74
500ms/div
200kSPS
1.4Mpts
74
1s/div
100kSPS
1.4Mpts
74
2s/div
50kSPS
1.4Mpts
74
5s/div
20kSPS
1.4Mpts
74
10s/div
10kSPS
1.4Mpts
74
20s/div
5kSPS
1.4Mpts
74
50s/div
2kSPS
1.4Mpts
74
Due to display limitations, the oscilloscope can not display more than 20 waveforms at a time (Overlay» or « Waterfall » display modes) .
Page 66
Functional Description
66 4-Channel SPO Digital Oscilloscopes
Functional Description
IV - ACQUISITION System (cont‟d)
Advantage
This mode allows you to increase the maximum waveforms recording (in a segmented memory) frequency to 300000 Waveforms/s. It allows us to observe details that rarely occur, because the « blind time » (time without signal acquisitions) is minimized.
Disadvantage
The oscilloscope displays the acquired waveforms once the acquisition sequence is completed.
The oscilloscope can not display more than 20 segments (Frames) at a time.
Waterfall ( 3 segments at a time)
Overlay (3 segments at a time)
Using the « Sequence » mode we were allowed to capture a waveform that occurs
rarely. We used the two viewing modes « Waterfall » and « Overlay »
Page 67
Functional Description
4-Channel SPO Digital Oscilloscopes 67
Functional Description
IV - ACQUISITION System (cont‟d)
Memory Depth
Advantage
A deep memory allows to record the signal evolution over a long period of time, and then analyze it in detail with a very fine elementary pitch. For example with a 28Mpoints memory depth we can record the evolution of a signal for 14ms with an elementary pitch of 0.5ns.
Disadvantage
We have to search in the long recording memory, the waveform portion to observe in details.
Using the x50000 Zoom mode to display simultaneously the overall waveform and a detail
Using x1000000 Zoom coefficient to analyze in detail the detail
Using a 28Mpoints memory depth, we were able to record “14ms” of signal evolution at a sampling rate of 2GSPS (=0.5ns elementary pitch). We then used two Zoom factors : x50000 (to display a 280ns portion of 14ms) and x1000000 (to display a 14ns portion of 14ms) to observe an elementary portion of the recorded waveform in “RUN” or STOP. We can move the Zoom window in the entire memory depth by using the horizontal « Position » knob.
Page 68
Functional Description
68 4-Channel SPO Digital Oscilloscopes
Functional Description
IV - ACQUISITION System (cont‟d)
Time Base
The oscilloscope digitizes the waveform by sampling the input signal at a regular time interval. The Time Base coefficient and the Memory Depth set the sampling frequency.
Use the "S/div" knob to set the “Time Base” coefficient. Open the « Acquire » menu to set the suitable memory depth.
Under-Sampling Aliasing
Example of
«aliasing»
observed with a
160MHz sinusoïdal
signal sampled at
100MSa/s : the
hardware counter
frequency is :
« f=160.66MHz »
the automatic
frequency
measurement is
« Freq=667.8kHz »
Comparing the two
different frequency
values we can
highligth the
presence of
«aliasing»
The aliasing occurs when the oscilloscope sampling rate is not fast enough to precisely reconstruct the digital waveform. The oscilloscope displays a lower frequency waveform, or an unstable waveform.
The real high frequency signal
The displayed low-frequency signal resulting from aliasing
Sample points
Page 69
Functional Description
4-Channel SPO Digital Oscilloscopes 69
Functional Description
IV - ACQUISITION System (cont‟d)
Operation Steps
Example:
a 250MHz sine
wave displayed
using the “Sinx”
sinusoidal
interpolation
To set sampling
Press the Acquirebutton to open the ACQUIRE menu. Press the « Acquisition » button and use the “Universal” knob to select the
acquisition mode : “Normal”, “Peak detect”, “High Res” or “Average”.
To set the « Average »
Press the « Averages » button to select the number of averages: “4”, “16”, “32”, “64”, “128”, “256”, “512” or “1024”.
To set the interpolation Press the Sinx/xbutton to select the interpolation: « Sinx » or « x ». "Sinx" : Sinusoidal interpolation:
Example :
a 250MHz sine wave displayed using the linear
interpolation « X »
"X" Linear interpolation:
Page 70
Functional Description
70 4-Channel SPO Digital Oscilloscopes
Functional Description
IV - ACQUISITION System (cont‟d)
SEQUENCE Mode
In the « Sequence » mode, the oscilloscope is ready to display waveforms, once the acquisitions are completed. In this mode, the maximum number of acquisitions per second can reach 300000. For each trigger event, the oscilloscope fills a segment of the recording memory and continues until the memory is completely filled. Once completed, the oscilloscope displays the acquired segments up to 20 segments at a time.
To use the « Sequence » the Horizontal format must be: YT. To set the Sequence mode : 1° Press the « Acquire » button to open the ACQUIRE menuPress the "Sequence" button to open the SEQUENCE menuPress the « Acq. Mode » button to disable or enable the Acquisition
« Off - On » (Off is the default value). 4° Press the « Display Mode » button to select the display mode : Overlay - Waterfall 5° Press the « Frame set » button and use the « Universal” knob to set the
displayed « Frame set » Press the «Run/Stop» button to start recording the sequence When the acquisition sequence is completed and the oscilloscope
displays the first 20 segments, press the « Run/Stop » button Manually scroll the acquired segments using the « Start », « End »
buttons and the « Universal » knob or Press the « Operation » button to automatically scroll the acquired
segments.
Page 71
Functional Description
4-Channel SPO Digital Oscilloscopes 71
Functional Description
IV - ACQUISITION System (cont‟d)
RUN/STOP SINGLE AUTO NORMAL
Run/Stop
Press the "Run/Stop" button to start (Run/Stop button is lit in yellow) the signal acquisition, press this button again to Stop (Run/Stop button is lit in red).
Single
Press the Single button to Arm the trigger (the oscilloscope is Ready, and the “Run/Stop” button is lit in yellow), then when a trigger event is detected a unique acquisition starts, when the acquisition is completed the oscilloscope Stops (Run/Stop button is lit in red).
In SINGLE mode when you press the "Single" button to start a unique acquisition, the oscilloscope performs the following steps:
1. Acquisition of Pre-Trigger samples, which are represented on the screen by the portion of the waveform to the left of the Trigger point.
2. The oscilloscope acquires the « Pre-Trigger » samples continously and
waits for the trigger event.
3. Detection of the trigger event.
4. After the trigger event, the acquistion continues until the recording memory is filled with the « Post-Trig » samples, which are represented on the screen by the portion of the waveform to the right of the trigger point.
5. The oscilloscope displays the waveform and Stops (Run/Stop button is lit in
red). Press again the « Single » button to start a new single acquisition .
Auto
Press the « Auto » button to enable the Auto mode. In this mode, acquisitions occur in the presence or absence of a trigger event.
Normal
Press the « Normal » button to enable the « Normal » mode. In this mode, acquisitions occur only when a trigger event is detected.
Page 72
Functional Description
72 4-Channel SPO Digital Oscilloscopes
Functional Description
V - DISPLAY System
Display Menu
Press the "Display" button to open the DISPLAY menu.
MENU pad
Display
page 1
Option
Settings
Description
Type
Vectors
Points
Two adjacent points (samples) are connected by a straight segment.
Points (samples) are not connected together .
Color
ON OFF
Allows to switch the SPO display from the “color palette” type to the “intensity grading” type.
If « Color : On » the SPO «color palette» is active. If « Color : Off » the SPO « intensity grading » is active.
Persistence
Off 1 sec 5 sec 10 sec 30 sec Infinite
To set the persistence duration.
Clear Pers
Dump the Persistence
Clear Display
Dump the Display
Next Page
Page1/2
Open page 2/2
Page 73
Functional Description
4-Channel SPO Digital Oscilloscopes 73
Functional Description
V - DISPLAY (cont‟d)
«Display» Menu
page 2
Option
Settings
Description
Grid
Display the grid and the axes. Display the axis. Grid and axis are not displayed.
Intensity
Sets the trace intensity.
Brightness
Sets the Grid intensity.
Transparence
Set the transparence.
Next Page
Page 2/2
Press this button to open page 1/2.
SPO
Two SPO display modes:
Color On : Display with a « Color palette »
Color Off : Display with an « intensity grading »
Color On
The SPO « Color
palette » is « On ».
The point (sample)
color depends on
the occurrence : the
Red color corresponding to the most frequent points
(samples) and the
purple color to rare
points. All channel
waveforms use the
same “palette”.
Color Off
The SPO display
uses an « intensity
grading » (256
levels). Each
channel has its own
color but the
intensity of the
points change
according to their
occurrence.
Rare very frequent
Page 74
Functional Description
74 4-Channel SPO Digital Oscilloscopes
Functional Description
V - DYSPLAY System (cont‟d)
Operation Steps
1. To set the display type
1) Press the "Display" button to open the DISPLAY menu.
2) Press the "Type" button and select "Vectors" or "Points".
2. To set the Color Press the "Colorbutton and set "ON" or "Off" : Off : Each Channel is displayed with its own color and an intensity grading : CH1 yellow, CH2 red, CH3 blue and CH4 green ON : All channels are displayed using the same « color palette » (Red Yellow Green Blue Purple)
3. To set the Persistance Press the "Persistance" button and select "Off", "1 Sec", "5 Sec", "10 Sec", "30 Sec", or "Infinite". This option is used to view, for example, the drift of a signal over time.
4. «Clear» Persistence
Press this button to dump the persistence
5. «Clear» Display
Press this button to dump the Display
6. Next page button
Press this button to open page 2/2
7. To set the Grid Press the “Grid” button and select the grid type ”, “ ” or
8. To set the Intensity Press the Intensity” button and use the “Universal” knob to set the trace Intensity.
9. To set the « Brightness » Press the “Brightness” button and use the “Universal” knob to set the grid Brightness.
10. To set the transparence
Press the “Transparence” button and use the « Universal » knob to set the transparence.
11. Next Page
Press this button to open page 1/2.
Caution in XY mode,
the «Persistence» and the «Color» are «Off» and the “Dots”
display type is
imposed
Page 75
Functional Description
4-Channel SPO Digital Oscilloscopes 75
Functional Description
V - DISPLAY System (cont‟d)
X-Y Format
The XY format is used to analyze the phase difference between signals using the Lissajous figures. In « XY » mode, channel CH1 (CH3) is the X axis and channel CH2 (CH4) the Y axis. The oscilloscope uses an acquisition mode without trigger and displays XY datas in the form of points, each XY point corresponds to a simultaneous acquisition on CH1(X) and CH2(Y) : X= CH1 sample (or CH3), Y= CH2 sample (or CH4).
The "XY" mode accepts sampling frequencies from 20Sa/s to 1GSa/s (1-2-5 sequence).
If we visualize, in XY mode, two sinusoidal signals with the same frequency but phase shifted by an angle θ :
We will have : Sin(θ)=A/B or C/D θ=± arcsin(A/B) or ± arcsin(C/D)
Two sine waves with same frequency but phase shifted by 40° :
Page 76
Functional Description
76 4-Channel SPO Digital Oscilloscopes
Operation Steps
Channel CH1 (or CH3) X axis, the corresponding “V/div” and vertical “Position” knobs allows to adjust the horizontal scale and position.
Channel CH2 (or CH4) Y axis, The corresponding “V/div” and vertical “Position” knobs allows to adjust the vertical scale and position.
Turn the “S/div” knob to set the sampling rate
The following functions are not available in “XY” mode:
- XY between a Real waveform and a Math waveform
- Cursor
- Trigger Control
- Horizontal position knob
- Vector display type
Note : The « Auto setup » function resets automatically to YT format.
Page 77
Functional Description
4-Channel SPO Digital Oscilloscopes 77
Functional Description
VI - MEASURE System
MEASUREMENTS
The oscilloscope displays the waveform, that is to say the variation of the signal amplitude (voltage or current) as a function of time. The oscilloscope displays : the vertical and horizontal scales and the automatic and cursors measurements values.
The oscilloscope has a « hardware » frequency counter on the signal of the active trigger source channel.
Scale measurements
This method allows a quick visual estimate of the amplitude or the period of the signal. For example, you can observe the signal amplitude and roughly determine it is slightly higher than 100 mV. You can estimate the amplitude by simple measurements by counting the number of major and minor divisions and multiplying by the vertical sensitivity. For example, if you count five major divisions between the minimum and maximum values of a signal, and if the vertical sensitivity is 100mV/div, then it is easy to calculate the peak-to-peak amplitude: 5 div x 100 mV/div = 500 mV.
Cursors measurements
Three modes : Manual, Track (cursors attached to the trace) and Off Press the "Cursors" button to open the cursors menu.
1. Manual Cursors
Option
Settings
Description
Mode
Manual
For measurements using manual cursors.
Type
Voltage Time
Use the cursors to measure the signal amplitude : voltage ∆V. Use the cursors to measure the period (or the frequency) of the signal : T et 1/T.
Source
CH1 - CH2 CH3 - CH4 MATH REFA - REFB REFC - REFD
Select the reference source for measurements.
Cur X1
Select X1 cursor and use the "Universal" knob to adjust.
Cur X2
Select X2 cursor and use the "Universal" knob to adjust.
Cur X1-X2
Select X1-X2 cursors pair and use the "Universal" knob to adjust.
Cur Y1
Select Y1 cursor and use the "Universal" knob to adjust.
Cur Y2
Select Y2 cursor and use the "Universal" knob to adjust.
Cur Y1-Y2
Select Y1-Y2 cursors pair and use the "Universal" knob to adjust.
Page 78
Functional Description
78 4-Channel SPO Digital Oscilloscopes
Functional Description
VI - MEASURE Sytem (cont‟d)
Examples :
Use of X and Y
cursors to measure
the period (or frequency) and the amplitude of a sine
wave
When the « Manual Cursors » are active, the display shows two horizontal or vertical parallel cursors to measure voltage (Y cursors) or time (X cursors). Use the "Universal" knob to move the cursors. Select the appropriated reference source for cursors measurements.
Voltage cursor (Y cursors) : The voltage cursors materialized on the screen in the form of horizontal lines. The cursor position is in Volt.
Time Cursor (X cursors) : The time cursors materialized on the screen in the form of vertical lines. The cursor position is in second.
Moving the cursors : Select the cursor (X or Y) and use the "Universal" knob to move it. When you move the cursor the value corresponding to the vertical and(or) horizontal position of the cursor is displayed in the top left of the screen.
Page 79
Functional Description
4-Channel SPO Digital Oscilloscopes 79
Functional Description
VI - MEASURE Sytem (cont‟d)
Operation
Steps
1. Press the "Cursors" button to open the CURSORS menu.
2. Press the “Mode” button and select “Manual”.
3. Press the "Type" button and select : "X1" or "X2" or "X1-X2" or "Y1" or "Y2" or "Y1­Y2".
4. Press the "Source" button and selectet : "CH1", "CH2", "CH3", "CH4", "MATH", " REFA", "REFB", " REFC", " REFD", as the source reference for measurements.
5. Select "X1" and use the "Universal" knob to set X1.
6. Select "X2" and use the "Universal" knob to set X2.
7. Select "Y1" and use the "Universal" knob to set Y1.
8. Select "Y2" and use the "Universal" knob to set Y2.
9. The measurements values are displayed on the top left of the screen:
- The voltage difference (in Volts) between Cursor Y1 and Cursor Y2: ∆V
- The time difference (in Seconds) between Cursor X1 and Cursor X2: ∆T
- The inverse of ∆T : 1/∆T is in Hz
Track Mode
Option
Settings
Description
Cursor Mode
Track
Track mode allows to measure with cursors
attached to the waveform.
Cursor X1
CH1 - CH2 CH3 - CH4
Select the channel attached to cursor X1.
Cursor X2
CH1 - CH2 CH3 - CH4
Select the channel attached to cursor X2.
Cur X1
Select the cursor X1 and use the "Universal" knob to adjust.
Cur X2
Select the cursor X2 and use the "Universal" knob to adjust.
X1-X2
Select the X1-X2 cursors pair and use the "Universal" knob to adjust.
Page 80
Functional Description
80 4-Channel SPO Digital Oscilloscopes
The « Track » mode displays two cross cursors attached to the trace. With the Universal knob, you can only adjust the horizontal position of the selected cross cursor. The oscilloscope displays the ∆V and ∆T values.
Operation
Steps
1. Press the "Cursors" button to open the CURSORS menu.
2. Press the “Mode” button and select "Track".
3. Press the "X1 source” button and select the attached channel.
4. Press the "X2 source” button and select the attached channel.
5. Select “X1” and use the “Universal” knob to move horizontally the X1 cross cursor.
6. Select “X2” and use the “Universal” knob to move horizontally the X2 cross cursor.
7. The measured values are displayed on the top left of the screen:
X1→T: The horizontal position of X1 Cursor (Time Cursor). Y1→V: The vertical position of Y1 Cursor (Voltage Cursor). X2→T: The horizontal position of X2 Cursor (Time Cursor). Y2→V: The vertical position of Y2 Cursor (Voltage Cursor). ∆T : The time interval between X1 and X2 cursors. 1/∆T : The inverse of the time interval between X1 and X2 cursors in Hz. ∆V : The voltage difference between Y1 and Y2 cursors.
Example :
Measuring the
amplitude and
the frequency
of a 1MHz
square wave
Page 81
Functional Description
4-Channel SPO Digital Oscilloscopes 81
Functional Description
VI - MEASURE Sytem (cont‟d)
Automatic Measurements Menu
Press the "Measure" button to open the MEASURE menu :
Menu Pad
Three types of automatic measurements :
Voltage measurements Time measurements Delay measurements
Of the 32 automatic measurements available, 23 of them (14 voltage and 9 time) can be displayed individually with or without their statistics.
In the example below we have choosen to display the frequency :
Measure menu
Press the « Measure » button to open the MEASURE menu :
We can select : * The channel reference for measurements : CH1 or CH2 or CH3 or CH4 * The type of measurement of the 23 available : Vpp - Vmax - Vmin - Vamp ­Vtop - Vbase - Vmean - Mean - Vrms - Crms - FOV - FPRE - ROV - RPRE -
Period - Freq - +Wid - -Wid - Rise Time - Fall Time - Bwid - +Dut - -Dut * Press the « Add » button to display the selected measurement
* Press the « Clear » button to clear the displayed measurements * Press the « Statistics» button to display the statistics on the measurements
displayed. * Press the « All Measure » button to display a table with all the voltage (or
time, or delay) measurements.
Page 82
Functional Description
82 4-Channel SPO Digital Oscilloscopes
Functional Description
VI - MEASURE Sytem (cont‟d)
Statistics
Press the «Statistics» button to open the «Statistics» sub-menu:
We can then view statistics of the displayed measurements, the oscilloscope displays the following values:
Average, Min, Max, Standard Deviation and the Number of measurements.
All Measures
Option
Description
1. SOURCE
The reference channel for measurements : CH1 CH2 CH3 CH4
2.
VOLTAGE
Press this button to display the 14 voltage measurements.
3. TIME
Press this button to display the 9 time measurements.
4. DELAY
SOURCE
Press this button to select the channel pair for the delay measurement :
CH1-CH2, CH1-CH3, CH1-CH4, CH2-CH3, CH2-CH4, CH3­CH4
5. DELAY
Press this button to display the 9 delay measurements.
Example : displaying the 32 automatic measurements available: Voltage, Time, Delay
CH1 is the reference channel for the voltage and time measurements
For the delay measurements the reference is the channel pair : CH1-CH2
Page 83
Functional Description
4-Channel SPO Digital Oscilloscopes 83
Functional Description
VI - MEASURE System (cont‟d)
1. Automatic Voltage
Measurements
Option
Settings
Description
Source
CH1 CH2 CH3 CH4
Select the reference channel for voltage measurements.
Type
Vmax, Vmin, Vpp, Vamp, Vtop, Vbase, Cycle Mean, Mean, Cycle Vrms, Vrms, ROVShoot, FOVShoot, RPREShoot, FPREShoot
Press the "Type" button and use the "Universal" knob to select the voltage measurement type.
ROVShoot = Rise Overshoot FOVShoot = Fall Overshoot RPREShoot= Rise Preshoot FPREShoot= Fall Preshoot
2. Automatic Time Measurements
Option
Settings
Description
Source
CH1 CH2 CH3 CH4 Select the reference channel for
time measurements
Type
Rise Time, Fall Time, Freq, Period, Bwidth, +Width,
-Width, +Duty, -Duty
Press the "Type" button and use the "Universal" knob to select the time measurement type.
3.Automatic Delay
Measurements
Option
Settings
Description
Source
CH1-CH2 CH1-CH3 CH1-CH4 CH2-CH3 CH2-CH4 CH3-CH4
Reference channel pair for Delay measurements.
Type
Phase, FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF
Press the "Delay" button « On » or « Off » to display or not the delay measurements.
4. All measures
Option
Settings
Description
Source
CH1, CH2 CH3, CH4
Select the reference channel for measurements.
Voltage
On Off
Enable « All Voltage measurements ». Disable « All Voltage measurements ».
Time
On Off
Enable « All Time measurements ». Disable « All Time measurements ».
Delay
On Off
Enable « All Delay measurements ». Disable « All Delay measurements ».
Return
Press the « Up » button to return to the MEASURE menu.
Page 84
Functional Description
84 4-Channel SPO Digital Oscilloscopes
Functional Description
VI - MEASURE System (cont‟d)
Description
of the type of
measure
Type of measure
Description
Vmax
Waveform maximum peak voltage.
Vmin
Waveform minimum peak voltage.
Vpp
Absolute difference value between maximum and minimum peak voltage of the entire waveform
Vtop
Upper level most common voltage value of the waveform.
Vbase
Lower level most common voltage value of the waveform
Vamp
Voltage difference between Vtop and Vbase.
Vavg
Arithmetic mean over the first cycle of the waveform.
Mean
Arithmetic mean over the entire waveform.
Crms
The true RMS voltage value over the first cycle of the signal.
Vrms
The true RMS voltage value over the entire waveform.
ROVShoot
« Rise OVer Shoot » : Defined as (Vmax-Vtop)/Vamp after a positive edge.
FOVShoot
« Fall OVer Shoot » : Defined as (Vmin-Vbase)/Vamp after a negative edge.
RPREshoot
« Rise PRE shoot »: Defined as (Vmin-Vbase)/Vamp
before a rising edge.
FPREshoot
« Fall PRE shoot »: Defined as (Vmax-Vtop)/Vamp before a
falling edge.
Rise Time
« Rise Time »: the interval of time between 10% and 90%
of the first rising edge.
Fall Time
« Fall Time »: the interval of time between 90% and 10%
of the first falling edge.
BWid
« Burst Width »: The duration of a pulse burst over the
entire waveform.
+ Wid
+ Width: interval of time between the first rising edge and the next falling edge at 50% of Vamp.
- Wid
-Width: interval of time between the first falling edge and the next rising edge at 50% of Vamp.
+ Duty
+ Duty is the ratio between the positive pulse width and period.
- Duty
- Duty is the ratio between the negative pulse width and period.
Phase
The amount one waveform leads or lags another in time. Expressed in degrees, where 360 degrees comprise one waveform cycle.
FRR
The interval of time between the first rising edge of CHi source and the first rising edge of CHj source
FRF
The interval of time between the first rising edge of CHi source and the first falling edge of CHj source
FFR
The interval of time between the first falling edge of CHi source and the first rising edge of CHj source
FFF
The interval of time between the first falling edge of CHi source and the first falling edge of CHj source
LRR
The interval of time between the first rising edge of CHi source and the last rising edge of CHj source
LRF
The interval of time between the first rising edge of CHi source and the last falling edge of CHj source
LFR
The interval of time between the first falling edge of CHi source and the last rising edge of CHj source
LFF
The interval of time between the first falling edge of CHi source and the last falling edge of CHj source
Page 85
Functional Description
4-Channel SPO Digital Oscilloscopes 85
Functional Description
VI - MEASURE System (cont‟d)
Operation Steps
Voltage measurements
To display a particular voltage measurement :
1. Press the "Measure" button to open the "MEASURE" menu.
2. Press the "Source button to select the channel source for
measurements: "CH1", "CH2", "CH3", "CH4".
3. Press the Type” button to select the measurement type to display.
4. Press the « Add » button to display the selected measurement
5. Press the « Clear » button to clear the displayed measurements
6. Press the « Statistics » button to display the measurement statistics
To display all voltage measurements
7. Press the « All measure » button to open the ALL MEASURE menu.
8. Press the "Voltage" button (On, Off) to display all voltage measurements.
9. Press the "Up" button to return to the « MEASURE » menu.
Note: You can display up to five automatic measurements at a time. Note : To display all the available measurements (Voltage, Time, Delay)
you have to activate them from the « All Measure » submenu.
Example : Displaying 5 automatic measurements, their statistics and the 32 available measurements
Page 86
Functional Description
86 4-Channel SPO Digital Oscilloscopes
Functional Description
VI - MEASURE System (cont‟d)
Operation Steps
Time measurements
To display a particular time measurement :
1. Press the "Measure" button to open the "MEASURE" menu
2. Press the "Source button to select the channel source for
measurements: "CH1", "CH2", "CH3", "CH4".
3. Press the “Type” button to select the measurement type to display.
4. Press the « Add » button to display the selected measurement
5. Press the « Clear » button to clear the displayed measurements
6. Press the « Statistics » button to display the measurement statistics
To display all time measurements
7. Press the « All measure » button to open the ALL MEASURE menu.
8. Press the "Time" button (On, Off) to display all time measurements.
9. Press the "Up" button to return to the « MEASURE » menu.
Page 87
Functional Description
4-Channel SPO Digital Oscilloscopes 87
Functional Description
VII - SAVE/RECALL System
Save/Recall
Press the Save/Recallbutton to open the SAVE/RECALL menu. This menu allows to Save/Recall : Setups, Waveforms, Picture, .CSV files.
4 data types can be saved :
- Setups - Waveforms - Pictures - .CSV files
The «Setups» files can be saved in internal (the oscilloscope) or external (USB memory device plugged in the “USB host” front panel connector) memories. You can save up to 20 Setups in the internal memory.
The Waveforms, the Pictures, and the .CSV files must be saved in external memories.
The setups and the waveforms can be recalled and displayed on the oscilloscope.
The pictures and the .CSV files can not be recalled or displayed on the oscilloscope, this files can be open on a PC.
Page 88
Functional Description
88 4-Channel SPO Digital Oscilloscopes
Functional Description
VII - SAVE/RECALL System
1. SAVE/RECALL
Menu
Pres the « Save/Recall » button to open the ”SAVE/RECALL” menu :
You can save
« Setups » in
internal or external
memory
You can save
Waveforms,
Pictures and .CSV
files only in
external memory
Page 89
Functional Description
4-Channel SPO Digital Oscilloscopes 89
Functional Description
VII - SAVE/RECALL System (cont‟d)
Saving « Setups »
in Internal or
External memory
To save a «Setup» file in the oscilloscope internal memory : 1° Press the « Save/Recall » button to open the SAVE/RECALL menu. 2° Press the « Type » button and select « Setups » 3° Select Save to « Internal » 4° Select a Setup N° location among the 20 available
Press the «Save» button to save the current Setup To save a setup file in an external memory (USB memory device
plugged in the « USB host » front panel connector) : 1° Press the « Save/Recall » button to open the SAVE/RECALL menu
Press the « Type » button and select « Setups » 3° Select Save to « External » 4° Plug the USB device in the « USB host » front panel connector and wait
for the USB icon appears and the following Popup” message : « USB Flash Drive Plugged in ! » is displayed.
Press the « Save » button and set the « save » function, wait for the file directory of the “USB memory device” is displayed.
Note : If the USB device is not plugged in and we press the «Save» button, the following message is displayed : « USB Flash Drive isn‟t connected
Press the « New » button, to enter from the keyboard (shift to keyboard)
the name of the new setup file to save, the maximum width allowed is 8 characters
Example : To enter the name « dox3302s » select one by one the characters (using the Universal knob or the buttons “→” and “←”) and press the
« Enter » button for each selected character.
Press the «Switch to keyboard» button to switch to «Switch to Name» 8° Press the « Enter » button to save the Setup file: “dox3302S”
Page 90
Functional Description
90 4-Channel SPO Digital Oscilloscopes
9° The Setup file « dox3302s » is saved in the USB memory device as
dox3302s.SET
To Recall a Setup file
We can « recall » a setup file from an internal or an external memory.
To Recall a Setup file from an internal or external memory
To Recall a Setup file from the oscilloscope internal memory
Press the « Save/Recall » button to open the SAVE/RECALL menu. 2° Press the « Type » button and select « Setups » 3° Select Save to «Internal»
Select a Setup N° location among the 20 availablePress the « Recall » button to recall the selected Setup file; wait for the message «Read Data success» Note : If the selected « setup N° » corresponds to an empty location the
following message is displayed : « Location Empty ! » To Recall a Setup file from an external memory (USB memory device
plugged in the « USB host » front panel connector) : 1° Press the « Save/Recall » button to open the SAVE/RECALL menu
Press the « Type » button and select « Setups »
Select Save to « External »
4° Plug the USB memory device in the USB host connector and wait for the
USB device icon and the Popup message « USB Flash Drive Plugged
in ! ». 5° Press the « Recall » button and wait for the file directory of the “USB
memory device” is displayed.Use the Universal knob to select the .SET file to recallPress the « Load » button to recall the selected setup file and wait for the
message «Read Data success» .
Page 91
Functional Description
4-Channel SPO Digital Oscilloscopes 91
Functional Description
VII - SAVE/RECALL System (cont‟d)
Rename
We can Rename an existing Setup file
Press the « Save/Recall » button to open the SAVE/RECALL menu
Press the « Type » button and select « Setups »
Select Save to « External »
4° Plug the USB memory device in the USB host connector and wait for the
USB device icon and the Popup message « USB Flash Drive Plugged
in ! ». 5° Press the « Recall » button and wait for the file directory of the “USB
memory device” is displayed
Select the file to Rename 7° Press « Next Page » button Press the « Rename » button Modify the file name using the buttons and the Universal knob: « Switch
to Name », « Universal knob», « Delete », « switch to keyboard », « Universal knob », « Enter », « Switch to Name », « Enter », « Confirm » or « Cancel »
Delete
We can delete a file in a directory:
Press the « Save/Recall » button to open the SAVE/RECALL menu
Press the « Type » button and select « Setups »
Select Save to « External »
Plug the USB memory device in the USB host connector and wait for the
USB device icon and the Popup message « USB Flash Drive Plugged
in ! ». 5° Press the « Recall » button and wait for the file directory of the “USB
memory device” is displayed
Select the file to Delete 7° Press the Delete button Press the « Confirm » or « Cancel » button to delete or not the selected
file.
Page 92
Functional Description
92 4-Channel SPO Digital Oscilloscopes
Functional Description
VII - SAVE/RECALL System (cont‟d)
2. Save/Recall Waveforms
Waveforms are saved only in external memory, therefore before saving or
recalling a waveform plug the USB memory device in the “USB host” front
panel connector of the oscilloscope.
You can save a « waveform » only in an external memory.
The saved waveform can be recalled and displayed on the oscilloscope screen.
To « SAVE or RECALL » a « Waveform » you need an USB memory device.
Option
Settings
Description
Type
Waveform
Menu de « Sauvegarde/Restauration » des Traces
Save
External memory
Save waveforms in an external memory (USB memory device plugged in the USB host connector of the oscilloscope).
Recall
External memory
Recall saved waveforms .DAV files
2.1 Saving / Recalling
Waveforms
Press the Save or Recall buttons to open the submenus « Save » or « Recall » waveforms. Both submenus have the same functionality : 1° You are able to modify a « File » or a « Directory ». 2° You are able to save waveforms in a New file 3° You can Delete a selected file. 4° You are able to Load a waveform file (.DAV) from an USB device 5° You can Rename a file.
Notes : Waveforms are saved in .DAV files When a waveform is saved or recalled the acquisition is stopped.
Page 93
Functional Description
4-Channel SPO Digital Oscilloscopes 93
Press the « Save » or « Load » buttons to open the SAVE or RECALL sub­menus, both submenus have the same functionality :
Option
Settings
Description
Modify
Files Directory
to modify files or directories
New
Press the “New” button to enter the file name using the “Universal” knob or the buttons “→” and “←”.
Delete
To delete the selected file.
Load
To load and display the selected file.
Next Page
1/2
To open page 2/2
Rename
To rename an existing file
Return
To return to the main Save/Recall menu
Next Page
2/2
To open page ½
Page 94
Functional Description
94 4-Channel SPO Digital Oscilloscopes
Functional Description
VII - SAVE/RECALL System (cont‟d)
Save a Picture
The oscilloscope can save a screen copy (« Picture ») in a external memory. The oscilloscope can’t recall a picture file (.BMP) .
Saving a screen
copy « Picture »
To save a screen copy (« Picture ») as a .BMP file
1. Press the "Save/Recall" button to open the SAVE/RECALL menu.
2. Press the "Type" button and select "Picture"
3. Press the "Save" button to open the Save sub-menu.
4. Press the « New » button and enter the picture file name .BMP using the Universal knob or the buttons “→” and “←”.
5. Confirm the Name to save the picture.
Notes : 1° A Picture must be saved in an external memory. 2° A Picture file (.BMP) can’t be open by the oscilloscope, use for example
the “Paint” PC software to open it.
Picture save
and recall menu
Page 95
Functional Description
4-Channel SPO Digital Oscilloscopes 95
Functional Description
VII - SAVE/RECALL System (cont‟d)
2.2 Saving a screen copy
(Picture)
To save a screen copy in an external memory : Open the SAVE/RECALL menu and select the Picture type, press the «Save»
button to open the Save submenu to save a screen copy .BMP.
Option
Settings
Description
Type
Picture
Save a « Picture» file.
Save
Press the « Save » button to open the save menu.
Submenu « Save » « Picture »
Option
Settings
Description
Modify
File Directory
To modify a file or a directory
New
To create a new screen copy (picture) file press the « New » button to open the New file submenu. Enter the file name (.BMP) using the Universal knob or the buttons “→” and “←”.
Delete
Press this button to delete the selected file
Next page
1/2
Press this button to open page 2/2
Rename
Press this button to rename an existing file
Return
To return to the main menu
Next Page
2/2
To open page ½
To enter the picture file name first press the « Switch to Keyboard » button (to “switch to name”) and then press the “Enter” button
Page 96
Functional Description
96 4-Channel SPO Digital Oscilloscopes
Functional Description
VII - SAVE/RECALL System (cont‟d)
Saving CSV files
The CSV files must be saved in external memory. The CSV files can’t be opened in the oscilloscope. The CSV files can be opened in table software EXCEL.
1. Open the "SAVE/RECALL" menu and select the file type: "CSV" .
2. Plug the USB memory device in the «USB host» connector and wait for the USB device icon and the Popup message « USB Flash Drive
Plugged in”.
3. Press the « Para Save » button to set : "On" or "Off".
4. Press the "Save" button to open the « Save .CSV file » submenu.
5. Press the « Modify » button and select: "File" or "Directory"
6. Press the « New » button to enter the file name using the Universal knob or the buttons “→” and “←”.
7. Press the « Delete » button to delete the selected file.
8. Press the « Next Page » button
9. Press the « Rename » button to modify the name of the selected file or directory.
10. Press the « Return » Button to return to the main menu.
Notes: 1° The .CSV files can‟t be opened in the oscilloscope. 2° The .CSV files can be opened in EXCEL Warning : The .CSV files generated by the DOX3000, can exceed the size limit allowed by the version of EXCEL (eg limits : 65536 lines for EXCEL 2003 and 1048576 lines for EXCEL 2007). To be able to open the complete file, set the maximum memory depth of the oscilloscope less than the limit of the EXCEL version (examples : 14kpts for “EXCEL 2003” or 700kpts for EXCEL 2007).
If the file exceeds the limit, the following « PopUp » message appears :
Click « OK » to display the lines allowed by the EXCEL version, starting from line N°1 to the imposed limit value:
Page 97
Functional Description
4-Channel SPO Digital Oscilloscopes 97
Menu Save
Type CSV
Submenu
« Save» « New »
.CSV file
Page 98
Functional Description
98 4-Channel SPO Digital Oscilloscopes
Functional Description
VII - SAVE/RECALL System (cont‟d)
Restoring the «Factory Settings»
(Default Setup)
Press the « Default » button to restore the « Default Setup » :
Display obtained after restoring the « default
setup »:
Channel CH1
Sensitivity: 1V/div,
Probe factor: 1X,
Impedance: 1MΩ,
Coupling: DC, BW
Limit : Full Time Base:
1µs/div
Restoring Waveforms
Waveform files must be saved in external memory (USB memory device). To restore a waveform plug first the USB device in the « USB host »
connector of the oscilloscope front panel
Plug the USB memory device in the «USB host» connector and wait for the USB device icon and the Popup message « USB Flash Drive Plugged
in”.
Press the « Save/Recall » button to open the SAVE/RECALL menuPress the « type » button and select Waveforms 4° Press the « Recall » button and wait for the contents of the « USB memory
device » is displayed on the screenUse the Universal knob to select the waveform file « .DAV » to be displayed
on the screenPress the « Load » button to restore the selected waveform file Wait until the selected waveform is displayed on the screen Notes : 1° When the saved waveform is displayed the acquisition stops. The displayed
values of the time base coefficient, the sampling rate, and the memory depth correspond to the restored waveform. To refresh the display of vertical
Page 99
Functional Description
4-Channel SPO Digital Oscilloscopes 99
parameters of the restored waveform you must press the corresponding “CHi” button.
We can perform automatic or cursors measurements. If we « Run » new acquisitions by pressing the « Run/Stop » button, the
restored waveform is cleared and replaced with the new acquire waveform. Warning : The new acquisitions are done with the settings corresponding to the restored waveform.
In the example below we provide : 1° The initial configuration of the oscilloscope before restoring the saved
waveform.
The configuration corresponding to the restored waveformThe configuration after « Running » new acquisitions
Initial
Configuration :
Channels CH1 and
CH2 actives,
Sensitivity 1V/div,
BdT 1.0µs/div
The restored
waveform on CH3 :
the S/div coefficient
= 20µs/div is
automatically
refreshed but not the
vertical parameters
of the restored
waveform
Page 100
Functional Description
100 4-Channel SPO Digital Oscilloscopes
To refreshed the
vertical parameters of
the restored (CH3)
waveform you must
press a « CHi »
button.
If we « Run » new
acquisitions by
pressing the
« Run/Stop » button,
the restored
waveform is cleared
and replaced with the
new acquire
waveform. The new
acquisitions are done
with the settings
corresponding to the
restored waveform.
(The initial
configuration is lost)
Loading...