Memmert INCO 108, INCO 108 med, INCO 153, INCO 246 med, INCO 153 med Operating Manual

...
Page 1
OPERATING MANUAL
CO2 incubators
INCO 108 INCO 108 med INCO 153 INCO 153 med INCO 246 INCO 246 med
Page 2
anufacturer and
ustomer servic
e
EMMERT GmbH + Co. KG
ostfach 17 2
0
-91107 Schwabac
h
Äußere
Rittersbacherstr . 3
8
-91126 Schwabac
h
German
y
on: +49 (0) 09122 / 925-0 ax: +49 (0) 09122 / 14585
-Mail: sales@memmert.co
m
nternet: www.memmert.com
ustomer servic
e:
on: +49 (0) 09122/925-128
and: +49 (0) 09122/925-126
-Mail: service@memmert.co
m
or service enquiries, please always specify the appliance number on the nameplate (see page
15 ). © 2
0
14 Memmert GmbH + Co. KG
ition 11/2014
Subject to modifications
Page 3
3
A
bout this manual
About this manua
l
urpose and target grou
p
T
his manual describes the setup, function, operation and maintenance of CO2 incubators
o
f the types INCO and INCOmed with a chamber volume of 108, 153 and 246 litres. In this
manual, the term INCO is used for INCO as well as INCOmed for the purpose of simplification.
istinctions are explicitly pointed out.
T
his manual is intended for use by the trained staff of the operator in charge of operating
and/or maintainin
g
the incubator . If you are asked to work on the incubator, you should read
this manual carefully before startin
g
work on the unit. Familiarise yourself with the safety
re
g
ulations. Only perform the work that is described in this manual. If there is something you
d
on't understand, or certain information is missing, ask your superior or get in touch with the
manufacturer . Do not do anythin
g
without authorisation
.
ntents
T
he INCO incubator is available with different fittings: as a basic model, and with six different
additional modules. The technical fittin
g
s and functional range of the basic model and the
individual additional modules are described from pa
g
e 11 .
f specific equipment features or functions are available only with one of the additional mod-
es, this is pointed out in the relevant sections in this manual.
ue to individual fittings, depictions in this manual may be different from the actual appear-
ance.
ther documents that you must read:
f
or service and repair work (see page
) – a separate service manua
l
torage and Forwardin
g
Th
is instruction manual belongs with the incubator and should always be stored so that those who work on the incubator have access to it. It is the responsibility of the operator to ensure that persons who work on or who will work on the incubator are informed as to the wherea­bouts of this instruction manual. We recommend that it is always stored in a protected loca­tion c
l
ose to the incubator . Make sure that the instruction manual is not damaged by heat or
d
amp. If the incubator is sold on or transported and then set up again at a different location,
t
h
is instruction manual must go with it
.
Page 4
C
ontent
ntent
. Safety regulations 6
1.1 Terms and icons used ..........................................................................................................
1.2 Product safety and dangers ................................................................................................
1.3 Requirements of the operating personnel ..........................................................................
.................................................................................................
1.5 Changes and conversions ....................................................................................................
1.6 Behaviour in case of malfunctions and irregularitie
s
1.7 What to do in case of accidents..........................................................................................
2. Design and Function 1
0
2.1 Design ................................................................................................................................1
0
2.2 Function.............................................................................................................................11
2.3 Optiona
l
extras..................................................................................................................11
2.4 Mat
e
rial.............................................................................................................................1
3
2.5 Electrical equipment
1
3
2.6 Connections.......................................................................................................................1
3
14
2.8 EC Declaration of Conformit
y
...........................................................................................15
2.9 Desi
g
nation (nameplate) ...................................................................................................15
2.1
0
Technical data ....................................................................................................................1
6
2.11 Ambient conditions ...........................................................................................................1
8
2.12 Accessories inclu
ded
..........................................................................................................1
8
. Delivery , Transport and Setting Up 1
9
3
.1 Safety regulations ..............................................................................................................1
9
3.2
Transport ............................................................................................................................1
9
3.3 Delivery
..............................................................................................................................
1
9
3
.4 Setup
.................................................................................................................................2
0
4
. Putting into Operation 2
2
4
.1 Checks
................................................................................................................................
4
.2 Connecting ........................................................................................................................
4
.3 Oxygen calibratio
n
4
5. Operation and control 2
5
5.1 Operating personnel.........................................................................................................
5
5.2 Openin
g the d
oor ............................................................................................................
5
.......................................................................................................
5
5.4 Insertin
g
water tray(s
)
...........................................................................................................
..........................................................................................................
5.7 Basic operation ..................................................................................................................
7
5.8 Sett
ing
parameters............................................................................................................
7
5.9 Operatin
g
modes ...............................................................................................................2
8
..............................................................................................2
8
5.11 During operation ...............................................................................................................34
5.12 En
ding
operation ...............................................................................................................37
6. Warning messages and malfunctions 3
8
6.1 Warning messages ............................................................................................................
38
6.2 System/appliance errors ....................................................................................................
39
......................................................................................................................
42
Page 5
C
ontent
7. Advanced functions 4
3
7.1 Printer
43
7.2 Basic appliance settings (Setup)
........................................................................................
43
7.3 Temperature monitorin
g
4
5
7.4 Steri
l
isation chipcar
d
49
7.5 User ID card (optionally available as an accessory) ...........................................................
7.6 Calibration .........................................................................................................................
7.7 Communication interfaces ................................................................................................
5
......................................................................................................................
. Maintenance and Servicing 5
9
.1 Cleaning .............................................................................................................................
.2 Regular maintenanc
e
...................................................................................................................
9. Storage and Disposal 6
1
1
9.2 Disposa
l
.............................................................................................................................
1
ndex
6
2
Page 6
afety regulations
Safety regulations
1.1
Te
rms and icons us
ed
n this manual, certain common terms and icons are used to warn you of dangers or to give
you notes that are important in avoidin
g
injury or damage. Observe and follow these notes
an
d
regulations to avoid accidents and damage. These terms and icons are explained below.
1.1.1 Terms used
"Warning
"
is always used whenever you or somebody else could be injured if you do
ot observe the accompanying safety regulation
.
"Caution
"
is used for information that is important for avoiding damage
.
1.1.2
cons us
ed
Prohibited sign (forbidding an action)
not tilt
applianc
e
o not lift appliance
Warning icons (warning of a danger)
anger of
ele
ctrical shoc
k
xplosive
atmospher
e
anger of frost
-
ite/cold burns
Warning of gas bottles
heat/hot surface
s
Ga
s
Regulation signs (stipulate an action)
isconnect t
he
mains plu
g
Wear gloves
Wear safety
shoe
s
serve
information in
s
eparate manua
l
Two o
r mor
e
persons require
d
Other icons
nformation on
f
irst ai
d
irst Aid:
nse eyes out
mportant or
seful addition-
al informati
on
Page 7
7
afety regulations
roduct safety and
anger
s
ncubators of the INCO type are technically well-developed, manufactured using high-quality
materials and tested for many hours in the factory . They contain the latest technolo
g
y and
c
omply with recognised technical safety regulations. But there are still dangers involved, even
w
h
en the appliance is used as intended. These dangers are described below.
W
arning
!
A
fter removing covers, live parts may be exposed. You may receive
an electric shock if you touch these parts. Disconnect the mains plug
efore removing any covers. Any work inside the unit may only be erformed by qualified electricians.
W
arning
!
W
hen loading the chamber with an unsuitable load, poisonous or explosive vapours or gases may be produced. This could cause the chamber to explode, and people could be badly injured or poisoned. The chamber may only be loaded with materials/test objects which
d
o not form any poisonous or explosive vapours when heated up
see also Chapter 2.7 Intended use on page 14).
W
arning
!
anger of suffocation. In high concentrations, CO2 and N2 can have
a suffocating effect. In normal operation, the incubator gives off
s
mall amounts of
CO
2
and – if equipped with the O2 module – N
2
t
o its environment. You should therefore ensure that the room in
which it is installed is properly ventilated. If a gas bottle is not con-
ected, or is empty, always close the stop valve or pressure reducer
on
the bottle
.
W
arning
!
igh concentrations of CO2 can cause cold burns or frostbite. Avoid
contact w
ith CO
2
gas to the eyes and skin.
W
arning
!
CO
2
gas bottles may burst or explode at high temperature. Do not use naked flames in the vicinity of the gas bottles. Store gas bottles at lower than 5
0 °C
in a well-ventilated location. Prevent water from
enetrating, as well as backflow into the gas bottles. It is essential
that you read the safety notes and regulations of the gas suppliers.
2
and N2are not dangerous substances in terms of the German Ordinance on Hazardous Substances (GefStoffV). Y ou should nevertheless familiarise yourself with the applicable safety re
gul
ations prior to handling such gas bottles
.
Page 8
8
afety regulations
1.3
equirements of the
perating personne
l
T
he incubator may only be operated and maintained by persons who are of legal age, and
who have received instructions for the incubator . P ersonnel who are to be trained
,
instructed
o
r who are undergoing general training may only be active on the incubator under the con
-
tinuous supervision of an experienced person.
T
he incubator may only be transported by persons (fork-lift truck, manual pallet jack), who are
trained for this work and who know the correspondin
g
safety regulations
.
epairs may only be performed by qualified electricians. In this case the regulations in the
s
eparate service manual must be observed.
1.4
esponsibility of the owner
The owner o
f the incubator
is responsible for the flawless condition of the incubator and for the incubator being oper-
ated in accordance with its intended use (see pa
g
e
);
is responsible for ensuring that persons who are to operate or service the incubator are
q
ualified to do this, have received instructions about the incubator and are familiar with
t
his operating manual
;
ust know about the applicable regulations, requirements and work protection regula-
t
ions, and train staff acordingly
;
is responsible for ensuring that unauthorised persons have no access to the incubator
;
is responsible for ensuring that the maintenance plan is adhered to and that maintenance
and repair work is properly carried out (see pa
g
e
);
e
nsures, for example through corresponding instructions and inspections, that the incuba
-
t
or and its surroundings are kept clean and tidy;
is responsible for ensuring that personal protective clothing is worn by operating person-
el, e.g. work clothes, safety shoes, protective gloves.
1.5
anges andconversion
s
o independent conversions or alterations may be made to the incubator . No parts may be
added or inserted which have not been approved by the manufacturer .
ndependent conversions or alterations result in the EC declaration of conformity (see page
15 ) losin
g
its validity , and the incubator may no longer be operated.
T
he manufaturer is not liable for any damage, danger or injuries that result from independent
c
onversions or alterations, or from non-observation of the regulations in this manual
.
1.6
haviour in case of malfunctions and irregularitie
s
T
he incubator may only be used when in a flawless condition. If you as the operator notice
irre
g
ularities, malfunctions or damage, immediately put the incubator out of service (see
apter 1.8) and inform your superiors.
Y
ou can find information on eliminating malfunctions from page 38 .
Page 9
afety regulations
What to do in case of accidents
1.
eep calm. Act resolutely and with consideration. Pay attention to your own
s
afety .
. Switch off the incubator and close the valves on the gas bottle.
3
.
all a doctor.
.
nitiate first aid measures. If available: Call a trained first aid helper .
In case of contact with
CO
to the eyes and skin:
y
inse eyes out with water for at least 15 minutes. With cold burns, rinse with water
fo
r at least 15 minutes. Cover over in a sterile manner . Call a doctor.
Inhaling CO
g
or N
:
igh concentrations can cause suffocation. Symptoms may include a loss of mobility and
c
onsciousness. The victim is not aware of suffocatin
g.
ow concentrations of CO can cause accelerated breathing and headaches
.
A
nyone affected should breathe fresh air , using a breathing device independent of recircu-
latin
g
air . K eep the person warm and calm. Call a doctor. In case of respiratory arrest, use
artificial respiration
.
If gas is escaping:
gpg
eave the room immediately , warn others and ventilate the room. If you re-enter the room,
se an autonomous breathing device (independent of ambient air) if it has not been estab-
ished that the atmosphere is harmless.
1.8 Switching off incubator in an
mer
gency
ush main switch on front side of appliance
. 1 ).
ose the valves on the gas bottle.
Fig. 1 Switch off incubator by
pressing the main switch
g
y
Page 10
1
0
esign and Functio
n
Design and Functio
n
.1 Des
ign
4
1
2
3
14
13
11 12
9
10
8
7
6
5
18
17
16
15
Fig. 2 Design of INCO incubators
Controller/control panel (see page
27
)
2
Set key
3
Push/turn control
y
4
C
onnections on the rear of the oven
(
see page
)
5
Heating ribs (see page
)
6
Pt100 temperature sensors
7
Humidity sensor
p
8
Glass d
oor
y
9
Chambe
r sea
l
Sliding shelf
1Door sea
l
g
2
Ventilator/inlet opening
Heating ribs
4
Adjustable feet
g
g
Nameplate (beneath door, see page
j
)
Door knob (see page
p
p
25
)
Doo
r
Chip card reader
Page 11
1
esign and Functio
n
unct
ion
T
he air in the incubator is heated up by an all-round heater with a large surface (
. 3 , No. 1).
Th
e CO and/or
N
g
as is introduced into the working chamber via a sterile filter . Because CO has a much higher specific weight than air , the gas is let into the working chamber above the interior fan (2). The turbulence-free interior ventilation (3) ensures a uniform distribution of t
he g
ases, creating a homogenous atmosphere
.
n the basic version, humidification takes place by means of water trays (4). To avoid uncon­trolled condensation, dosed fresh air is piped into the interior . If the appliance is equipped with a humidity module, humidification takes place via a hot-air
g
enerator , which allows
water to eva
p
orate at a set rate. The sterile hot air is let into the interior above the fan and is
mixed with the air current (5).
1
1
1
1
1
1
4 4
2
5
3
1
Fig. 3 How INCO incubators wor
k
.3
ptional
xtras
.3.1
asic equipment
lectronic fuzzy-supported PID process controller with pulse width modulation and per-
anent performance adjustments and time-saving self-diagnosis system to quickly locate
e
rrors (see page
39
)
A
ll-round heater with larger surface area, with additional thermal conduction layer (see
. 3
)
apacitive humidity sensor umidity control (fresh air via sterile filter) ensures that setpoint humidity is quickly
eached and guarantees short recovery times, while avoiding condensation formatio
n
omogenous atmosphere and temperature distribution through encapsulated, turbu-
lence-free ventilation system
TERICard for fully automatic sterilisation process control for hot air sterilisation of appli
-
ance, including sensors and fan rotor (see page 49
)
Page 12
1
esign and Functio
n
igitalised electronic CO control with automatic zero position, NDIR measuring system
with self-dia
g
nosis system and acoustic error display , air pressure compensatio
n
anguage settings (see page
)
Alph
anumeric text display
ntegrated week time switch with grup function (e.g. each working day) (see page 31
)
etracting push-turn control for simple operation of appliance (see page
)
T
wo separate Pt100 temperature sensors DIN Class A in a 4-wire circuit for control and
onitorin
g
igital monitoring control for overtemperature, undertemperature and automatic
s
etpoint following (ASF) (see page 48
)
echanical temperature limiter (TB protection class 1, see page 45
)
onitoring relay to switch off heater in case of erro
r
ptical alarm displa
y
A
coustic signal messages if temperature or CO limits are crossed, when door is open and
if
g
as bottle is empty (see page 38
)
alibration of temperature, humidity , CO
and O possible on the device without a sepa-
ate PC (see from page
)
.3.2
ptionally availableadditional modules
C
omfort module:
T
wo gas connections with quick release connectors (see page 1
3
)
A
utomatic switch-over of gas bottles
Hygiene module
yg
lectro-polished, seamless laser-welded chamber
C
ommunication modul
e
ogging option of temperature, CO2 and relative humidity via computer/laptop
nternal log memory with 1024 kB as ring memory for all setpoint and actual values, errors
an
d
settings in real time and with date, logging approx. 3 months at 1 minute storage
interval (see pa
g
e
)
arallel printer port (PCL3-compatible) for printing out log data (see page 13 and page 43
)
o
ptionally USB, Ethernet, RS-232 or RS-485 interface (see from page
5
)
CO
2
modu
le
xtended adjustment range from 0 to 20 %
3
-point calibration (5 %, 10 %, 15 % CO
2
)
O
2
modu
le
ontrol of oxygen concentration through the introduction of nitrogen (
N
2
)
; adjustment
ange 1 % to 20 %
O
2
(not in combination with Comfort or Premium module
)
Premium modu
le
ncludes comfort, hygiene, communication and C
O
2
modules
Humidity module
y
A
ctive microprocessor humidification and dehumidification control (40-97 % rh
)
Page 13
1
3
esign and Functio
n
Materia
l
or the outer housing, MEMMERT uses stainless steel (W .St.No. 1.4016), and for the interior,
s
tainless steel (W .St.No. 1.4301), which stands out through its high stability , optimal hygienic properties and corrosion-resistance towards many (but not all!) chemical compounds (caution for example with chlorine compounds). The chamber load for the appliance must be carefully
checked
with respect to chemical compatibility with the materials mentioned.
A
material resistance table can be requested from the manufacturer .
.5
ectrical equipment
perating voltage: See nameplate (page 15 ), 50/60 Hz
urrent consumption: See nameplate (page 15
)
rotection class 1, i.e. operating insulation with safety earth terminal in accordance with N 6101
0
rotection type IP 20 acc. to EN 60 529
nterference-suppressed acc. to EN 55011 class B
A
ppliance fuse: Fusible link 250 V/15 A quick-blow
T
he temperature controller is protected with a miniature fuse 100 mA (200 mA at 115 V)
or models with a humidity model the CO
controller is protected with a 6.3 A miniature
fuse
.6
nnection
s
3
4
5
6
2
1
H2O
7
N2
In
Fig. 4
Connections on rear of a
ppl
iance
pp
pp
2
Mains lead
3
alternatively Ethernet, RS-232 or RS-485 connection) (details from page
yp
yp
55
)
4
Printer connection (only for models with communication or premium modules) (see page
ypg
4
3
)
5
CO
2
connection to main gas bottle
6
CO
connection to reserve gas bottle (only for models with CO
g
or premium modules)
N
2
connection (quick-release connector, only if equipped with O
2
gy
gy
222
module
)
p
Page 14
14
esign and Functio
n
.6.1
ctrical connecti
on
bserve the country-specific regulations when making connections (e. g. in Germany DIN VDE
100 with residual current device).
T
his appliance is intended for operation on an electrical power system with a system imped-
anc
e Z
at the point of transfer (service line) of a maximum of 0.292 ohm. The operator must
e
nsure that the incubator is operated only on an electrical power system that meets these
requirements. If necessary , you can ask your local ener
g
y supply company what the system
impe
d
ance is.
.6.2
onnection of external appliances
nly appliances may be connected externally (depending on the model, USB, RS 232, RS 485,
thernet, printer) whose interfaces comply with the requirements for safety extra-low voltage
e.g. PC).
.6.3 Gas connect
ion
Th
e oven can be connected with the supplied compressed air hose via a pressure regulator
with
g
as bottle monitor (DIN 8546) to a CO
compressed gas bottle or directly to a central CO
as supply.
or models with the CO or premium modules, two gas bottles with quick release connectors
c
an be connected. If equipped with the O
module, an N
gas bottle can be connected instead
o
f a second C
O
2
g
as bottle (
. 4 ).
Th
e pre-pressure must not exceed 1.2 bar . A value between 0.8 and 1 bar is considered to be
an i
deal
value.
.6.4 Water connect
ion
or models with a humidity module, the incubator can be connected with the supplied hose
to t
h
e also supplied water supply tank.
.7 Inten
ded
us
e
NCO incubators may only be used for incubating cell cultures or similar . Any other use is
improper , an
d
may result in hazards and damage.
The
incubators are n
ot
e
xplosion-proof (they do not comply with workplace health & safety
re
g
ulation VBG 24). The chambers may only be loaded with materials and substances which
c
annot produce any toxic or explosive vapours at temperature ranges up to 50 ºC, and which
t
h
emselves cannot explode, burst or ignite.
T
he incubators may not be used for drying, vaporising and branding paints or similar materi
-
als, the solvents of which could form an explosive mixture when combined with air . If there is any doubt as to the composition of materials, they must not be loaded into the incubator . P o­tentially explosive
g
as-air mixtures must not be produced, either in the interior of the chamber
o
r in the direct vicinity of the appliance.
T
he incubator may not be used for sterilisation purposes. It is not a steriliser with respect to
t
he
Law on Medical Devices.
nly gas bottles with a pressure regulator may be connected to the gas connections of the
o
ven. Introducing other gases or materials than C
O
2
or
N
2
is not permitted.
Page 15
1
5
esign and Functio
n
.7.1
ntended use of the INCOmed
or INCOmed incubators, which are subject to the 93/42/EEC guideline (Council Directive on
the approximation of the laws of the Member States relatin
g
to medical devices), the intended
se is defined as follows:
T
he CO2 incubator INCO med is intended for the creation and maintenance of constant envi­ronmental conditions for application in the field of in vitro fertilisation (IVF), especially for the incubation of oocytes, spermatozoa and zy
g
otes in special culture dishes for IVF application as
well as for
g
ene expression and the biosynthesis of RNA and proteins.
2.8 EC
claration of
onformit
y
Y
ou can download the EC declaration of conformity of the appliance online:
nglish: http://www.memmert.com/en/service/downloads/ce-statement/ German: http://www.memmert.com/de/service/downloads/e
g
-konformitaetserklaerung/
2.9 Designation (nameplate
)
T
he nameplate (
. 5 ) provides information about the appliance model, manufacturer and
technical data. It is attached to the front of the appliance, on the ri
g
ht beneath the door (see
10 ).
Typ: INC 108 F.-Nr.: 0109.0088 230 V
~
4.4 A 50/60 Hz 1000 W
DIN12880-Kl.3.1 Nenntemp.: 50 °C
1 2
3 4
5
6
10
8
9
7
Fig. 5 Nameplat
e
e designation
2
Operating voltage
yp g
3
Applied standard
pgg
gg
4
Protection type
5
CE conformity
yp
6
Address of manufactur
er
7 Disposal note
8
emperature ran
ge
9
Connection / performance values
pg
g
Factory number
p
Page 16
1
6
esign and Functio
n
2.10 Technical data
Model
* See Fig. 6 on page 17 .
108 153 246
hamber width A* [mm]
6
0
48040
hamber height B* [mm
]
4804040
hamber depth C* [mm
]
400
A
ppliance width D* [mm] 71
030790
A
ppliance height E* (varies
d
ue to adjustable feet) [mm]
77
8
2
0
3
8
A
ppliance depth F* (includ-
ing door handle) [mm
]
7
90
hamber volume [litres
]
108 15
3
4
6
Weight [kg] 7
0110
erformance [W
]
1
0001500
ax. number of slidin
g
s
helves half size / full size
-/4
-/6
2 x 6/6
ax. load per sliding shelf
k
g]
15 15 15
ax. load per appliance [kg]
40
40
T
emperatur
e
T
emperature recording by means of Pt100 in a 4-wire circuit
Ad
justment range: Normal mode: 20 °C to 50 °C
Sterilisation mode: 160 °C (4 hours) via STERICard
Ad
justment precision: 0.1 °C
ontrol range: from 8 °C above room temperature to 50 °C Variation (time): max. ±0,1 °C at 37 °C Variation (spatial): max. ±0.3 °C at 37 °C
umidity
Th
e relative humidity in the chamber is measured by a capac-
itive humidity sensor and displayed digitally in percent.
T
he measurement precision of the humidity sensor is 1 % rh
A
djustment range: 88 to 97 % rh (for models with
umidity module 40 to 97 % rh)
Ad
justment precision: 1 % rh
isplay range: 10 to 98 % r
h
Variation (time): max. ±1 % rh
Page 17
17
esign and Functio
n
Model
* See Fig. 6 on page 17 .
108 153 246
Th
e CO content is determined through an NDIR measur-
ing system, controlled constantly by a microprocessor an
d
d
isplayed digitally in percent
A
djustment range: 0 to 10 % (for models with CO
or
premium modules 0 to 20 %)
Ad
justment precision: 0.1 % Variation (time): max. ±0.1 % rh Variation (spatial): max. ±0.3 %
(only if equipped with O
module
)
The O
content is determined with a long-lasting, mainte-
nance-free zirconium dioxide sensor , constantly controlled by a microprocessor and displayed digitally in percent
Ad
justment range: 1 to 20 %
Ad
justment precision: 0.1 % Variation (time): max. ±0.1 % Variation (spatial): max. ±0.3 %
D A75 75
38
38
F
C
202
B
76
E
8
12
Fig. 6 Dimensions of INCO incubators
Page 18
1
8
esign and Functio
n
Ambie
nt conditions
T
he incubator may only be used in enclosed rooms and under the following environmental
cond
itions
:
Amb
ient temperature: 5 ºC to 35 ºC
umidity: max. 80 % not condensing
egree of pollution: 2
A
ltitude of installation max. 3,000 m above sea level
T
he incubator may not be used in areas where there is a risk of explosions. The ambient air
ust not contain any explosive dusts, gases, vapours or gas-air mixtures. The incubator is
ot explosion-proof. eavy dust production or aggressive vapours in the vicinity of the appliance could lead to
sed
imentation in the chamber interior and as a consequence, could result in short circuits
o
r damage to electrical parts. For this reason, sufficient measures should be taken to revent large clouds of dust or aggressive vapours from developing.
.1
Accessories inclu
ded
For incubators with basic fittings:
g
T
wo (INCO 108) or three (INCO 153 and 246) sliding shelves
Gas pressure hos
e
Ź Water tray (one for INCO 108 and 153 incubators, two for INCO 246 incubators) Ź
terilisation chipcar
d
Additionally, for models with humidity module:
y, y
Water supply tank and connection hos
e
Additionally, for models with CO
y,
2
or premium module:
p
econd gas pressure hose with quick-release connecto
r
Additionally, for models with O
y,
2
module:
econd gas pressure hose with quick-release connecto
r
Page 19
1
9
elivery, Transport and Setting Up
.
elivery, Transport andSetting Up
.1 Safety regulations
W
arning!
You may injure your hands or feet when transporting and
nstalling the incubator. You should wear protective gloves
and work
shoes
.
W
arning
!
ecause of the weight of the incubator, you could injure yourself if
you try to lift it on your own.
f possible, only transport the incubator with a fork-lift truck or
anual pallet jack. The incubator may only be moved using a means
of transport by persons who have the required qualification for this
e.g. fork-lift truck licence). This incubator may not be transported
wi
t
h a crane.
f the incubator has to be carried, at least two people are required
for models 108 and 153, and at least four people for model 246.
W
arning
!
The incubator could fall over and seriously injure you. Never tilt the
ncubator and transport it only in an upright position.
.2 Transpor
t
Th
e incubator can be transported in three ways:
with a fork-lift truck; move the forks of the truck entirely under the incubator
o
n the manual pallet jack. y carrying; to do this, at least two people are required for models 108 and 153, and four
eople for model 246. Read the weight information detailed on page 16 .
.3
elivery
Th
e incubator is delivered in cardboard packaging on a cardboard pallet.
.3.1 Unpackin
g
1.
emove cardboard packaging or cut open carefully along an edge
.
.
ift up incubator from pallet and put down on the appliance feet.
.3.2
hecking for completeness and transport dama
ge
eck the delivery note to ensure that the delivery is complete.
heck the inside and outside of the incubator for damage.
f you notice deviations from the delivery note, damage or irregularities, do not put the incu-
bator into operation, but inform the haula
g
e company and the manufacturer
.
.3.3
isposing of packaging materia
l
ispose of the packaging material (cardboard) in accordance with the effective legal disposal
re
g
ulations for cardboard packaging in your country
.
Page 20
2
0
elivery, Transport and Setting Up
.3.4 Storage after delivery
f the incubator is initially to be stored after delivery: Read the storage conditions from page
1 .
.4
etup
T
he incubator can be placed either on the ground or on a table (work surface). When doing
t
h
is, ensure that the appliance is positioned exactly horizontally.
The
installation site must be
level and able to reliably carry the wei
g
ht of the incubator (see page 1
6
)
. Do not place the ap-
pliance on an inflammable surface
.
A
power connection must be available at the installation site in accordance with the connec-
tion data on the nameplate (see pa
g
e 15 ).
The
distance between the wall and the rear of the chamber must be at least 15 cm. The clear-
ance from the ceilin
g
must not be less than 20 cm and the side clearance from the wall not
less than 8 cm (
. 7 ). Sufficient air circulation in the vicinity of the chamber must be guaran
-
teed at all times.
FP

FP
FP

FP

Fig. 7 Minimum clearance from walls and ceiling
Page 21
1
elivery, Transport and Setting Up
.4.1
ase (accessory
)
T
he incubator can be placed on a base (
. 8 ).
.4.2 Stacking frame (accessory
)
T
wo appliances of the same model size can be placed on top of one another . To do this, foot
ali
g
nment provisions must be attached to the lower oven (
. 9 ):
1.
emove the housing cover from the lower oven.
.
nsert drilling template (supplied with the foot alignments) into the overturned lid.
3
.
ark drilling points and drill with a 4.2 mm diameter drill bit.
.
crew the foot alignment provisions to the top of the lid with the screws and nuts sup-
ied.
5.
ut the covers back on.
Fig. 8
Base
Fig. 9
A
ssembly of the foot alignment
provisions w
h
en two incubators are
p
l
aced on top of one another
Page 22
utting into Operatio
n
utting into Operation
4
.1
c
ks
ecking the temperature senso
r
specially strong vibrations during transport could
resu
l
t in the temperature sensors being moved
in t
h
eir holders in the working chamber .
heck the temperature sensor for its correct
positionin
g
and if necessary adjust its position
in the holder (
. 10 ).
.1. heck the door and adjust if
cessar
y
ee page 60 .
.
onnectin
g
.
.1
ower supply
C
aution:
bserve the country-specific regula-
t
ions when making connections (e. g. in Germany DIN VDE 0100 with residual current
d
evice / RCD). Observe the connection and power ratings (see nameplate).
T
he incubator is intended for operation on an electrical power system with a system imped-
anc
e Z
at the point of transfer (service line) of a maximum of 0.292 ohms. The operator
must ensure that the incubator is operated only on an electrical power system that meets these requirements. If necessary , you can ask your local ener
g
y supply company what the
s
ystem impedance is.
onnect power cable (see Fig. 4 on
page
13).
4.2.2
xternal devices
only for models with communication or premium modules
)
nly appliances whose interfaces comply with the requirements for safety extra-low voltge e.g. PC, laptop, printer) may be connected to the connections on the rear of the incubator see
. 4on
page
13 ). Which devices may be connected depends on the chosen model /
module variant
(
mmunication interfaces described in detail from page 55 ).
4.2
.3 Water connect
ion
only for models with humidity module
)
or steam creation, use only
d
e
mineralised water (aqua dem) in accordance with
VDE 0510/DIN EN 50272; re
g
ulations must be strictly adhered to (production conductivity
6FP%DWWHU\ZDWHULQDFFRUGDQFHZLWK9'(LVDYDLODEOHLQODUJHUFKHPLVW·V
sh
ops, super markets, hardware stores and in the wholesale trade. The standard VDE 0510/
IN EN 50272 has to be explicitly specified on the label.
therwise, possible calcification in the steam generators, steam piping and peristaltic pumps
c
ould affect the operability of the device.
T
he water used must have a pH value between 5 and 7 and be chlorine-free.
Fig. 10 Temperature sensor
Page 23
utting into Operatio
n
1.
ill up the supplied water supply tank (canister) with distilled water and place behind/
xt to the incubator.
A
ttach the supplied hose the quick-release connections to the canister and the water sup­ly "H
“ on the rear of the appliance (see
. 4 on
page
13).
.
.4 Gas connect
ion
W
arning
!
anger of suffocation: In high concentrations, CO
and N
can have
a suffocating effect. In normal operation, the incubator gives off
s
mall amounts of CO and – if equipped with the O
module – N
t
o its environment. You should therefore ensure that the room in
which it is installed is properly ventilated.
W
arning
!
igh concentrations of CO
can cause cold burns or frostbite. Avoid
contact w
ith CO
gas to the eyes and skin.
W
arning
!
CO
gas bottles may burst or explode at high temperature. Do not use naked flames in the vicinity of the gas bottles. Store gas bottles at lower than 5
0 °C
in a well-ventilated location. Prevent water from
enetrating, as well as backflow into the gas bottles. It is essential
that you read the safety notes and regulations of the gas suppliers.
For incubators with basic fittings:
g
A
ttach the supplied pressure hose to the as bottle (pressure regulator) and to the
"C
O
2
" connection on the rear of the appli-
ance with two hose clamps (
. 11, see
a
l
so page
).
For incubators with
CO
2
or premium
p
module
:
T
wo gas bottles can be connected by simply
pus
hing th
e supplied pressure hoses onto
t
h
e "C
O
2
In1" and "C
O
2
In2" quick-release
c
onnections on the rear of the appliance
Fig. 12, see also page 13 ).
onnect the main gas bottle to "In1", a re-
s
erve gas bottle can be connected to "In2".
A
ttach the pressure hose to the gas bottles
pressure regulator) with hose clamps.
CO
2
„CO2 In“
Gas connection for incubators with basic
f
ittin
gs
CO2CO
2
„CO2 In1“ „CO
2
In2“
Fig. 12 Gas conn
e
ction for incubators with
CO
o
r Premium modul
es
Page 24
4
utting into Operatio
n
For incubators with
O
2
module:
A
ttach the su
pplied
CO
pressure hose
t
o the C
O
2
as bottle (pressure reducer)
an
d
to the “CO2 In” connection on the ear of the incubator with a hose clamp Fig. 13, see also page 1
3
)
.
A
ttach (push on) the supplied N2 pres-
sure hose to the
N
gas bottle (pressure
educer) and to the “N2 In” connection
o
n the rear of the incubator with a hose
cl
amp
.
4.3
xygen calibratio
n
only for models with O
module
)
erform an Ocalibration before putting
into operation (see pa
g
e 54 ).
CO2N
2
„CO2 In“
„N
2
In“
Fig. 13 Gas conn
e
ction for incubators with O mod-
ul
e
Page 25
Operation and control
.
peration and
ntr
ol
5.1
perating personne
l
T
he incubator may only be operated by persons who are of legal age and have received in-
s
tructions for the incubator . P ersonnel who are to be trained, instructed or who are undergo-
in
g g
eneral training may only be active on the incubator under the continuous supervision of
an experience
d
person.
5.2 Opening the door
T
o open the door , turn handle to the right ( Fig. 14 ).
T
o close, turn door handle to the left.
W
h
en the door is opened, the CO
supply is automati-
c
ally interrupted. If the heated outer door is left open for
any len
g
th of time, condensation may form on the glass
doo
r.
5.3
oading the incubato
r
W
arning
!
Wh
en loading the chamber with an
unsu
itable load
, poisonous or explosive
v
apours or gases may be produced. This
could cause the chamber to explode, and people could be badly in-
j
ured or poisoned. The chamber may only be loaded with materials/
test objects which do not form any toxic or explosive vapours when
eated up, and which cannot ignite. If there is any doubt as to the
composition of materials, they must not be loaded into the incuba-
t
or.
C
aution:
ck the chamber load for chemical compatibility with the materials of the incubator see page 13), since considerable damage could otherwise occur to the chamber load,
th
e appliance or the surroundings
.
The
incubators are notexplosion-proof (they do not comply with workplace health & safety
re
g
ulation VBG 24) and are therefore not suitable for drying, vaporising and branding paints
o
r similar materials, the solvents of which could form an explosive mixture when combined
with air . P otentially explosive
g
as-air mixtures must not be produced, either in the interior of
the chamber or in the direct vicinity of the appliance.
eavy dust production or aggressive vapours in the chamber or in the vicinity of the appli-
ance cou
ld lead
to sedimentation in the chamber interior and as a consequence, could result
in short circuits or dama
g
e to electrical parts. For this reason, sufficient measures should be
taken to prevent lar
g
e clouds of dust or aggressive vapours from developing.
Th
e chamber must not be loaded too tightly , so that proper air circulation in the working
c
hamber is guaranteed. Do not place any of the chamber load on the floor , touching the side
walls or ri
g
ht below the ceiling (heating ribs) of the working chamber . To guarantee an opti-
mal air circulation, push in the slidin
g
shelves so that the gaps between the door , sliding shelf
and rear wall of the chamber are rou
g
hly the same size.
open
close
Fig. 14 Opening and c
l
osing the door
Page 26
Operation and contro
l
5.4
nsertingwater tray(s
)
for appliances with basic fittings
)
ill the water tray with distilled water and push into the lowest slot (
. 15 ).
Fig. 15 Inserting water tray(s
)
INCO model
Number
of
water
trays
Filling
level in
cm for
each tray
approx.
Amount of water in ltr . for
each tray
approx.
1 1.5 to 2.5 1 to 1.5
15
3
1 1.5 to 2.5 1 to 1.5
4
6
1.5 to 2.5 1 to 1.5
5.5 Connect
as supp
ly
1.
heck that the gas bottle(s) is properly connected (see also page 13 ).
.
pen valve(s).
5.6 Switch on applianc
e
The incubator is switched on and off by pressing the main switch/
push-turn control on the
front of the appliance.
witching on:
ress the main switch so that it comes out of the appliance (
. 16 ).
witching off: press the main switch so that it retracts back into the appliance ( Fig. 17 ).
Fig. 16 Switching on incubator
Fig. 17 Switching off incubator
Page 27
7
O
peration and control
5.7 Basic operatio
n
T
he desired parameters are entered on the operating panel of the controller on the front of
the appliance (
. 18 ). Basic settings, as well as those for time and printing, can also be made
ere. In addition, programmed and current parameters are displayed, as well as warning mes-
sages:
PRINT
SETUP
loop
t3
t4
t2
t1
on off
Mo
Tu
We
Th
Fr
Sa Su
3
4
2
1
STERI DEFRO
°C
°C
rh
mb
%
CO
mb
2
IN 1
IN 2
OUT
IN 1
IN 2
OUT
MIN
AUTO
MAX
123465 7 8 109 11 12
13141516
Fig. 18 Operating panel
ime displa
y
Display appliance locked with user-ID
py
py
card (see page
pypp
p
50
)
3
Display appliance is heating up
pg
pg
Sterilisation mode (see page
pypp g
ypp
49
)
5
6
Alarm display
p
p
7
Monitoring temperature display
py
(
see page
45
)
8
Ho
rn
9
Humidity display
g
g
1
Gas bottle 1 active
ypy
y
2 Gas bottle 2 activ
e
CO
displa
y
4
Display appliance is humidifying
py
Operating mode display (see page
pypp yg
28
)
ext display/O
pg
g
display (O
py
p
display only if
pg
pg
equipped with O
py
y
2
2
module)
py
2
A
ll operating functions are selected by turning the push-turn control to the
left or ri
g
ht...
..and adjusted by turning this with the SET key held down.
5.8 Setting
arameters
n general, all setting actions on the operating panel described on the following pages are
ma
d
e in the same way:
1.
Y
ou select the desired parameter with the push-turn control (menu item, e. g.
temperature); then all other parameters
g
o dark and the selected one flashes.
. With the SET key held down, set the desired value (e. g. 37.0
C) with the
pus
h
-turn contro
l.
3
.
elease the SET key , and the set value is saved. The display briefly shows the
s
et value, flashing. The current temperature is displayed and the incubator
egins to heat up to the set temperature.
Settings for other parameters are made in the same way .
set
set
set
set
set
Page 28
2
8
Operation and control
T
he control returns automatically to the main menu if the push-turn key or set key is
not operated for approx. 30 seconds.
Setting the temperature (Quick adjustment):
gp(Qj)
1.
old down the SET key and set the desired temperature setpoint with the push-turn
co
ntrol.
.
elease the SET key
T
he appliance flashes briefly , showing the temperature setpoint. Then the current temperature
appears on t
he d
isplay and the controller begins to move to the set temperature.
5.9 Operating modes
NCO incubators can be operated in three ways (
. 19 ).
ormal mode: The incubator runs in permanent operation at the temperature, humidity
an
d CO
values set on the operating panel. Operation in this mode is described from page
2
9
Week time switch: The incubator runs at the set values only at certain times. Operation in
t
his mode is described from page 31 .
nterface mode with PC/laptop (for models with communication or premium modules, see
f
rom page
5 )
n addition, basic appliance settings can be made (SETUP, see page 43 ) and printouts can be
made if the appliance is equipped with the communication or premium module (PRINT, see
43 ).
PRINT
SETUP
rma
l
o
peratio
n
see Page 29
)
W
eek
tim
e
s
witc
h
see Page 31
)
rint
er
see Page 43
)
asic appliance
s
ettings
see Page 43
)
Fig. 19 Operating mode
s
5.10 Setting the operating mode
1.
old SET key down for approx. three seconds, the selected operating mode
then be
g
ins to flash.
. Select the desired operating mode (normal mode, week time switch, pro-
ramming mode, printer or basic appliance settings/setup by turning contro
l
with SET key held down.
3
.
elease the SET key , and the selected operating mode is saved.
se
t
set
set
Page 29
2
9
O
peration and control
5.10.1
rmal m
ode
T
he appliance runs in this operating mode in permanent operation. The desired setpoints for
o
perating the chamber can be selected. The settings have an immediate effect on the func-
tions of the appliance.
oad incubator (see page
5 ).
.
witch on appliance. To do this, press the push-turn control on the operating panel so that
it comes out of the appliance (see
. 16 on
page
).
3
.
elect the normal operating mode
with the push-turn control:
PRINT
SETUP
. As described above, set the individual parameters with the push-turn control and the set
ey:
Temperature setpoint
pp
Ad
justment range:
to
°C
Temperature monitoring
pg
Ad
justment range:
IN MAX AUT
O
see also page 45
)
°C
MIN
AUTO
MAX
yp
Ad
justment range:
r incubators with basic
f
ittings: 88 to 97 %r
h
or models with humidity
ule: 40 to 97 %r
h
rh
%
igh levels of air humidity in the interior can only be achieved without condensation if the
interior is t
h
orou
ghly h
eated. Please set the humidity only when the temperature in the
interior has been stable for at least 3
0
minutes.
2
setpoint
p
Ad
justment range: 0 bis
0 %
or models with CO2 or premium
mo
dul
es 0 to 20 0 to 20 %
)
CO
2
2
setpoint
p
2
mod-
le
)
Ad
justment range:
Off
to 20
%
t2
t1
on off
Mo
Tu
We
Th
Fr
Sa Su
o2=3.4%
Page 30
30
Operation and contro
l
5.10.2 Settings example normal mo
de
With a 5% COcontent, a 3% oxygen content and a humidity of 96%, the appliance should heat up to 37°C. The monitorin
g
function should respond at 38.5 °C (
. 20 ).
30
40
20 10
50
60
80
40 20
100
Temperature in °C
Humidity rh in %
Time Time Time
CO
2
in %
5
10
15
20
25
Time
O
2
in %
5
10
15
20
25
Monitoring temperature
Fig. 20 Example for normal mode
1.
etting the normal operating mode:
old SET key down for approx. 3 seconds, the current
o
perating mode then begins to flash.
S
ele
ct the
operating mode with the push-turn con-
trol, while the SET key is held down. After you let go of the SET key , the control is in the normal operating mode.
.
etting the temperature setpoint:
old down the SET key and set the desired temperature
s
etpoint of 37.0°C with the push-turn control.
elease the SET key , the appliance will briefly flash, show-
ing the temperature setpoint. Then the current tempera-
ture
appears on the display and the controller begins to move to the set temperature of 37.0 °C.
eating up is indicated by the icon
.
°C
3
.
etting the monitoring temperature:
T
urn the push-turn control to the right until the monitor-
ing temperature and the MIN or MAX icon flashes. Hold
d
own the SET key and with the push-turn control, set the overtemperature limit to 38.5 °C and the undertem­perature limit to 36.0°C. Turn the push-turn control to the right until the monitoring temperature and the AUTO icons flash. Hold down the SET key and set to on with the push-turn contro
l.
T
he tolerance band is set in the SETUP menu (see
44).
°C
MIN
AUTO
MAX
PRINT
SETUP
Page 31
31
O
peration and control
.
etting the humidity setpoint:
igh levels of air humidity in the interior can only be
achieved without condensation if the interior is
th
orou
ghly h
eated. Please set the humidity only when
t
he temperature in the interior has been stable for at
east 30 minutes.
T
urn the push-turn control to the right until the humid-
ity display flashes. Hold down the SET key and set the
d
esired humidity setpoint of . %rh with the push-turn
c
ontrol. After releasing the SET key , the humidity setpoint briefly flashes. The current humidity value appears on the display and the controller begins to move to the set value.
T
he humidification process is indicated by the icon
only for models with humidity module).
rh
%
5.
Setting the
g
setpoint
p
T
urn the push-turn control to the right until the CO
d
is-
play flashes. Hold down the SET key and set the desired
2
setpoint of 5.0 % with the push-turn control. The
appliance flashes briefly , showing the C
O
2
setpoint. The
c
urrent C
O
2
actual value appears on the display and the
c
ontroller begins to move to the set CO2 setpoint.
ayed by the
IN 1
or
IN 2
icon.
CO
2
.
Adjust the
j
2
setpoint
p
only for models with O2 module
)
T
urn the push-turn control to the left until the O2 display
f
lashes. Hold down the set-key and set the desired O
2
s
etpoint of 3.0 % with the push-turn control. Release
the set-key . The appliance flashes briefly, showing the O
s
etpoint. Then, the current O
actual value appears on
the display and the controller begins to move to the set
value.
on
off
Mo
Tu
We
Th
Fr
Sa Su
o2=3.0%
The incubator is now running in permanent operation with the set values.
5.10.3 Week time switch
n this operating mode, the appliance switches on and
o
ff automatically at the times programmed
.
uring the OFF phase of the week time switch, the
appliance is in standby mode. The heatin
g
and cooling functions, along with the CO2 an
d
humidity supply are switched off here and the controller display shows the time, dimmed.
T
he sequence of the week time switch repeats itself each week.
n total, a maximum of 9 time blocks can be programmed, consisting of the switching on and
s
witching off times.
PRINT
SETUP
Page 32
32
Operation and control
Weekday
y
Ad
justment range: Monday to Sunday
Mo
Tu
We
Th
Fr
Sa
Su
yg p
Ad
justment range: Working days Mo-F
r
W
eekend
Sat-Sun
Mo
Tu
We
Th
Fr
Sa
Su
Mo
Tu
We
Th
Fr
Sa
Su
switch on time: ---
Appl
iance not switched on on this day
on
off
Switch on time (on
)
()
Ad
justment range:
0
0:00 to 23:59 hours
on
off
h
Switch off time (off
)
()
ne minute beyond the switch on time up to
24:0
0
on
off
h
sel
ected as in the normal operating mode.
f no settings (temperature setpoint etc.) are made for the ON phase, the controller takes
o
ver the values from the normal operating mode.
or reasons of safety , you should always check that only one switch one time is programmed in the desired time blocks and days. By turnin
g
further to the right, parameters (temperature
s
etpoint etc.) can be selected as in the normal operating mode.
f the controller is in standby mode or the week time switch is in the ON phase, the tempera-
ture setpoint can be directly accessed by briefly pressin
g
the SET key . By turning the control to
t
h
e right, you are returned to temperature monitoring, humidity and CO2 setting. By turning
to the left, you come back to the settin
g
s for the individual time blocks.
Page 33
33
O
peration and control
5.10.4
ettings example week time switc
h
rom Mo-Fr (workdays group), the appliance should switch on at 9.30 and switch off at
19.00. In addition it should work on Saturday from 10.00 to 14.00 (
. 21 ).
Mo
Sa
Fr
Th
We
Tu
Su
9:30h - 19:00h
24:00h
12:00h
01:00h
9:30h - 19:00h
9:30h - 19:00h
9:30h - 19:00h
9:30h - 19:00h
10:0h -
14:00h
Fig. 21 Operation with week time switch (example)
1.
etting the week time switch operating mode
old the SET key down for approx. 3 seconds, the current
o
perating mode then begins to flash. Select the week
time switc
h
operating mode with the push-turn control,
w
hil
e the SET key is held down.
elease the SET key , the control is now in the week time
s
witch operating mode.
PRINT
SETUP
.
witch on Mo-Fr at 09:3
0
T
urning the push-turn control to the left, select the
Mo-Fr on
"
icons (group working days)
.
old down the SET key and set the desired switch-on
time wit
h th
e push-turn switch to 9:
30
on
off
Mo
Tu
We
Th
Fr
Sa Su
3
.
witch off Mo-Fr at 19:
00
S
ele
ct "Mo-Fr
off
" (group working days) with the push-
turn c
o
ntr
ol.
old down the SET key and set the desired switch-off
time wit
h th
e push-turn switch to 19:00
.
on
off
Mo
Tu
We
Th
Fr
Sa Su
Page 34
3
4
Operation and control
.
witch on Sa at 10:
00
With the push-turn control, select "
Sat on
".
old down the SET key and set the desired switch-on
time wit
h th
e push-turn switch to 10:00
.
Mo
Tu
We
Th
Fr
Su
off
Sa
h
on
5.
witch off Sa at 14:
00
With the push-turn control, select "Sat off
old down the SET key and set the desired switch-off
time wit
h th
e push-turn switch to 14:00
.
Mo
Tu
We
Th
Fr
Su
on
Sa
h
off
5.10.5 Operation with
C/laptop (optional
)
f equipped with the communication or premium module, the incubator can optionally be
sed, controlled and programmed with a PC/laptop. It has corresponding communication
interfaces for this purpose (see pa
g
e 13 and page
5 ).
peration is described in a separate manual. It is delivered with the incubator for
the rele
vant m
odel
s.
.11
uring operatio
n
egularly check the water level. If necessary , add distilled water.
Warning messages during operation: See Page 38 .
CO
2
mod
e
n the heating up phase, the CO2 controller is initially deactivated. The CO2 intake is inter­rupted during this period. About 5 minutes after the setpoint temperature has been reached, t
h
e C
O
2
control begins measuring and C
O
2
gas is let in to the chamber via a sterile filter (the
valve on the
g
as bottle must be open). To ensure a homogenous distribution of the CO2 gas in
the interior , the
g
as is piped in above the chamber fan. The setpoint can be adjusted in 0.1 %
s
teps from 0 to 10% (for models with CO2 or premium modules 0 to 20 %)
A
fter a sterilisation procedure, and in cycles every 24 hours, an automatic zero balance
adjustment is carried out. This automatic zero balance adjustment is completed after a
fe
w minutes.
Displays in CO
py
2
mode:
HEAT UP
H
EAT UP is shown during the heating up phase of
th
e C
O
2
sensor . In the C
O
2
display , CO2 is shown.
A
UTOZERO - DO NOT OPEN DOO
R
is displayed during the
zero balance adjustment.
Page 35
35
O
peration and control
CO
2
IN 1
IN 2
OUT
A
fter the setpoint temperature has been reached,
th
e C
O
concentration is displayed in %, depending
o
n the setting.
IN 1
indicates that gas bottle 1
s active.
s displayed if the CO
concentration exc
eed
s the
d
efined setpoint by at least 1 % for more than 3
minutes. If the concentration is hi
g
her , the CO
d
isplay and the
icon flash.
n this case you should open the door for 30 sec.
and wait to see if the controller steadily adjusts
t
o the setpoint. If the error occurs again, contact
ustomer Servic
e.
T
his monitoring function only starts to work
o
nce the CO setpoint has been reached.
C
O2 emp
t
is displayed if gas bottle 1 and/or 2 is/are
e
mpty . In this case, you should connect new gas
bottles (see pa
g
e
).
Th
e pressure in the gas bottles is a constant
a
pprox. 57 bar at 20 °C ambient temperature. It
is not
p
ossible to determine how full the bottle
is t
h
rough the pressure, since the pressure only
d
rops immediately before the bottle is com-
p
l
etely empty .
Th
e C
O
2
supply is automatically interrupted when
th
e outer doors are opened.
door open
is indicated
in t
h
e text display .
on off
2
empty
2
supply is interrupted. In this
c
ase, check if the N2 bottle is correctly connected
and if the valve is open.
f that does not solve the problem, you should con-
nect a new
g
as bottle (see page
).
Automatic switch-over of
CO
2
gas bottles (only for models with comfort or premium
g(y p
modules)
)
T
he automatic switch-over of gas bottles guarantees an uninterrupted supply with CO2 gas
w
h
en two independent supply systems are connected.
Gas bottle 1 is always the main supply bottle. Gas bottle 2 is always the reserve bottle.
peration is only possible with one bottle. In this case it must be connected to IN1 (see also
. 4 on
page
).
Page 36
36
Operation and control
T
o be on the safe side, a freshly filled CO
gas bottle should always be used. So if the gas in
ottle 1 is used up, you connect the opened bottle to IN1 and the newly filled bottle as a
serve bottle to IN2.
T
he hose connection system used by Memmert shuts off automatically if a connection hose is
pulled off. You should still always close the stop valve on the
g
as bottle if a bottle is empty or
n
o
t connected.
IN 1
lights up when gas bottle 1 is active. If gas bottle 1 is empty ,
there
is an automatic switch-over to the reserve bottle.
IN 2
lights up after the switch-over to gas bottle 2 (reserve bottle). Switc
hing
over to the reserve bottle is marked by a short re-
peated acoustic si
g
nal (about 3 seconds long) (the default after
s
witching on is gas bottle 1).
n the following cases, a switch-over is made from reserve bottle 2 back to the main supply
ttle 1:
if the reserve bottle is empty
e
ach time after the appliance is switched on
after every change of the CO
setpoint
Humidity limit control
y
A
humidity limit control prevents the formation of condensation water in the chamber and
at t
h
e same time ensures that the setpoint humidity is quickly reached, with short recovery
t
ime
s.
T
he maximum achievable humidity can be adjusted in the standard model from 88 to 97% rh.
Active humidity control (only for models with humidity module)
y(y y)
Th
e active humidity control guarantees that setpoint humidity is quickly reached, without the
se of water trays.
n the heating up phase, the humidity control is initially deactivated. Approx. 5 minutes after the setpoint temperature is reached, the humidification and dehumidification control starts workin
g
. The setpoint can be set from 40 to 97 % rh. The humidity setpoint can also be ad-
j
usted during the transient state. For humidification, water is let into the chamber via a dosing pump. To avoid the formation of
g
erms, the steam is first heated to approx. 140 °C. Dehumid-
ifcation takes place throu
g
h the supply of dry air via a sterile filter .
f no humidity is required in the chamber , the humidity control can be deactivated by
s
etting it to
OFF
.
Page 37
3
7
O
peration and control
5.12
nding operatio
n
1.
witch off appliance. To do this, press the
ain switch on the operating panel so
th
at it clicks into place in the appliance
see
. 22).
.
lose the valve(s) of the gas bottle(s).
3
.
pen the door (see page
5 ).
.
move the chamber load.
5.
or appliances with the basic model:
emove and empty the water trays. Fill
water trays an
d
insert them only when
th
e appliance is next used.
or incubators with a humidity module,
e
mpty the water tanks if the incubator is
ot used for several days.
Fig. 22
S
witching off incubator
Page 38
38
W
arning messages and malfunctions
. Warning messages andmalfunction
s
.1 Warningmess
ages
A
n intermittent tone is also set off by the warning messages. This can be temporarily
s
witched off by pressing the SET key.
DOOR
OPENif the door of
a
ppliance is opened for
onger than 3 minutes.
emedy: Close the door
Error in the temperature control system (see also Chapter “
py(p
pg
on page
pg
45
):
)
Tb active
- temperature
limiter triggered if the
t
emperature limiter
respon
d
s
emedy: Switch off the appliance
and l
eave to coo
l down. If the
e
rror occurs again after switch
-
ing the appliance back on: Switch off
th
e appliance and Contact the
c
ustomer servic
e
h
i-alarm- overtempera-
ture al
arm limit max ex
-
c
eeded if overtemperature
rotection responds
emedy: Check the setting of
th
e MAX temperature monitor-
in
g
(see Chapter “ Temperature
mon
i
toring ” on page 45
- undertem-
erature alarm limit min
c
rossed if undertempera-
ture
protection responds
emedy: Check the setting of
t
he MIN temperature monitor-
in
g
(see Chapter “ Temperature
mon
i
toring ” on page 45
a
sf alarm- temperature
o
utside tolerance band if
a
utomatic monitorin
g
f
unction responds
emedy: Check the setting of the
A
SF temperature monitoring (see
utomatic temperature
monitor (ASF) ” on page 48 .
Error in humidity system (only for appliances with humidity module)
yy ( y pp y )
if the water supply is
defe
ctiv
e
emedy: Check that the water
s
upply hose is properly con-
necte
d
. Fill the water supply tank
with distilled water if it is empty .
Page 39
39
W
arning messages and malfunctions
f humidity exceeds the
preset setpoint for lon
g
er
t
han 30 minutes.
emedy: Open door for 30 sec.
and wait t
o see
if the controller
s
teadily adjusts to the setpoint.
f the error occurs again, Contact
the
customer service.
Error in
CO
system:
y
if the CO supply is defec-
tive
emedy: Set CO
setpoint value
t
o 0, check stop valve of gas bottle(s) and ensure that they are properly connected; if
g
as
bottle(s) empty , chan
ge g
as bot-
tles,
set CO
setpoint to desired
va
lue
f the COconcentra
-
tion excee
ds the defined
s
etpoint by at least 1 % for
m
ore th
an 3 minutes.
emedy: Open door for 30 sec.
and wait t
o see
if the controller
s
teadily adjusts to the setpoint.
f the error occurs again, Contact
the
customer service.
Error in
O
2
system:
y
on off
Mo
Tu
W
e
Th
Fr
Sa Su
2
empty
2
supply is defective
emedy: check if the N2 bottle is correctly connected and if the valve is open. If that does not
sol
ve the problem, you should
c
onnect a new gas bottle (see
).
.2 System/appliance
rror
s
W
arning
!
A
fter removing covers, live parts may be exposed. You may receive an electric shock if you touch these parts. Malfunctions requiring in­tervention inside the appliance may only be rectified by electricians.
Y
ou must read the separate service manual for the INCO incubator
for
this
.
o not try and rectify the error yourself, but contact an authorised customer service point for
EMMERT appliances or Contact the customer services department directly of the company EMMERT (see page
).
n case of enquiries, please always specify the model and appliance number on the nameplate see page
5 ).
Page 40
40
W
arning messages and malfunctions
Error P ossible causes Remedy
d
isplay although the
incubator is switc
hed
on.
Power supply interrupte
d
heck power supply and fuse/safety
s
witch.
Appliance fuse or miniature fuse or controller fault
y
ntact the customer service and rea
d
the se
rvice manua
l.
Appliance error Mainboard fault
y
Appl
iance cannot be
o
perate
d
Incubator loc
ked
with
ser-ID car
d
Unlock incubator with user-ID card (see
).
Push/turn contro
l
fault
y
ntact the customer service and rea
d
the se
rvice manua
l.
o CO display shown
in the CO modu
le
Switched mode power
s
upply SP 200 faulty
ntact the customer service and rea
d
the se
rvice manua
l.
icon flashes ...
Temp
erature fuse
(TWW , ASF) has re-
s
ponde
d
ncrease temperature difference
etween monitoring and working
t
emperature (see page 45 ).
eplace Pt100 temperature sensor
o
f monitoring controller if necessar
y
see Service manual
)
... and rh empt
y
Water supply tank empty
Set humidity setpoint to OFF, fill up
d
istilled water , then reset humidity
s
etpoint back to desired value.
... and rh empt
y
Humidity setpoint
exceeded
pen door for 30 sec. and wait to see if
sub
sequently the controller steadily ad-
j
usts to the setpoint. If the error occurs
again, Contact the customer service
.
... and
co2 over
2
setpoint is ex-
c
eeded
pen door for 30 sec. and wait to see if
sub
sequently the controller steadily ad-
j
usts to the setpoint. If the error occurs
again, Contact the customer service
.
... anderror Autoze
ro
A
utozero pump
f
aulty
2
contr
olle
r
f
ault
y
ntact the customer service and rea
d
the se
rvice manua
l.
rror display (E..) in
d
isplay
Appliance/system error
ntact the customer service and rea
d
the se
rvice manua
l.
Page 41
1
W
arning messages and malfunctions
Error P ossible causes Remedy
rror message CON
F
is displayed for only 10
s
ec. after switching on
)
hecksum error (error when saving setpoint values
)
T
he error can be rectified just by the
c
ontroller after a setpoint parameter
has again been saved. If the error
should
continue occurring, or cannot be rectified: Contact the customer service and replace controller (see service manual
)
hamber fan without
functi
on
Miniature fuse or mains adapter fault
y
ntact the customer service and rea
d
the se
rvice manua
l.
eating icon
not on
Amb
ient tempera-
t
ure too hi
gh
Setp the appliance up in a cooler room, minimum setpoint temperature = am-
ient temp. + 8 °C
T
emperature in
appliance higher
th
an set setpoint
t
emperatur
e
Wait until appliance has cooled down.
icon is permanently
it up
T
emperature fuse has
responde
d
Switch off appliance and leave to
cool down
eck temperature limiter
Th
e appliance is only operational again
after it has cooled down and the error
as
been el
iminated.
CAL
O
2
ERRO
R
Fault in oxygen calibra
-
t
ion
pen the door for one minute and
s
tart calibration again. If the message
appears gain: Contact the customer
service
.
ERROR O
2
Sensor
xygen sensor fault
y
ntact the customer servic
e
Page 42
W
arning messages and malfunctions
.3
wer failur
e
n case of a power failure, the incubator operates as follows:
In normal and week time switch operating modes
pg
A
fter the power supply has been restored, operation is continued with the parameters set.
T
he time and the duration of the power failure is documented in the
og memory
.
For remote operation
p
f there is a power failure in remote operation, the appliance starts in the normal operating
mode for safety reasons and all setpoint values are chan
g
ed to safe default values (see table).
T
he programme can only be continued from the computer . The time and duration of the
power failure are documented in the lo
g
memory .
Parameters Default value
T
emperatur
e
20 °C
off
(only for models with O
module
)
OFF
Page 43
43
A
dvanced function
s
Advanced function
s
7.1 Printer
only for models with communication or premium modules
)
ncubators with a communication or premium module are equipped with a parallel printer
port, as used in computers. Standard PCL3 compatible inkjet printers with a parallel port (e.
g
.
P DeskJet 5550 or HP DeskJet 9xx) can be connected to the printer interface on the rear of
the appliance (see pa
g
e ).
ake sure that a shi
elded
interface cable is used. The shielding must be connected to the plug
casing.
The co
ntr
oller h
as an internallog memory (see page
). The log data can be printed out in
t
h
is mode via the connected printer .
f a colour printer is connected, the various graphs are printed out in colour .
or a printout, the GLP header is also printed automatically , containing the following details:
ate of printout
eriod of lo
g
onsecutive page numbers
erial numbers and appliance nam
e
Running the printing function:
gp g
Select the PRINT operating mode with the pus
h
-turn control, as described on page 28 . By
turnin
g th
e push-turn control and holding down
the SET key , the followin
g
parameters can be selected and altered one after another , as
d
escribed in Chapter “
asic operation “ on page
7 :
uerying the date of the first print page
uerying the date of the last print page
eaving the print menu and returning to the main menu
Basic appliance settings (Setup
)
Setting options may vary depending on the appliance model) n this operating mode, the basic settings for the
app
l
iance can be made. The time, date, day and
year are set here, alon
g
with the settings for the
orn, the address allocation, the monitorin
g
nits and the calibration.
PRINT
SETUP
PRINT SETUP
Page 44
4
A
dvanced function
s
y turning the push-turn control and holding down the SET key , the following parameters can
be selected and altered one after another , as described in Chapter “
asic operation “ on page
7 :
lock time in 24 hr . format
onversion to summer time is not automatic, ut must be done manually.
at
e
Th
e controller contains a calendar which automatica
lly
accounts for the different lengths of months, and for leap y
ea
rs.
Weekday
Tu
Year
A
djustment range: from 2000 to 210
0
Acoustic signal at programme end
gpg
ENDSOUN
D
Setting:
OFF
or
ON
Acoustic Signal for alarm, e.g. over/undertemperature
g,g/p
ALARM s
o
Setting:
OFF
or
ON
mmunication addres
s
A
djustment range: 0 to 15 (see Chapter
mmunication interfaces “ on page
5
)
A
djustment range: 0.5 bis 5°C (see page 48
)
Language
gg
Setting: GERMAN, English, franCAIS, ESPANOL and italIAN
O
CAL 1-3 rh20 RH9
0
)
customer-side calibra
-
tion of temperature and humidity (see Chapter “
alibration “
on page
5
0
)
Page 45
45
Advanced function
s
yg
(Cal
0
)
only for models with O
module, see page
4
)
on
off
CAL o2 off
p
Save all settings and leave the SETUP mode.
The
realtime clock, which is set in the SETUP, contains the date and clock time. It is used
for lo
gging
purposes in accordance with GLP. Date and clock time are specified on the lo
g
printout. On graphical printouts, the time axis is labeled with the realtime. The clock is battery­buffered and independent of the mains connection. The inte
g
rated Lithium battery of the
type CR 2032 has a lifetime of approx. 10 years.
7.3 Temperature monitorin
g
The
monitoring temperature is measured via a separate Pt100 temperature sensor in the
chamb
er interior . The monitoring unit is used to protect the chamber load and as a protection
for the appliance and surroundin
gs.
T
he appliance has a double overtemperature fuse (mechanical/electronic) in accordance with
IN 12880.
°C
MIN
AUTO
MAX
1
2
3
1
6 5
4
Fig. 23
Visual alarm icon
p
Lit up: TB alarm
Flashing: TWW alarm, ASF alarm
p
p
2
Undertemperature limit
g
g
3
Automatic temperature monitor (ASF, see page
48
)
4
Overtemperature protection (TWW, TWB, see page
46 )5
Response temperature
6
Acoustic alarm icon
pp
7.3.1 Mechanical temperature monitoring:temperature limiter (TB)
T
he incubator is equipped with a mechanical temperature limiter (TB) protection class 1 in ac-
cord
ance with DIN 12880.
f the electronic monitoring unit should fail duiring operation, and the factory preset maxi-
tive measure, switches off the heating permanently . As a warning, the icon
lights up.
Page 46
46
A
dvanced function
s
Error rectification after the TB has been triggered:
gg
witch off appliance and leave to cool down
ontact the customer service and have the error rectified (e.g. replace temperature sen-
s
or).
T
he appliance is only operational again after it has cooled down and the error has been elimi-
nat
ed
.
7.3.2 Electronictemperature monitoring
(
T
WW
)
T
emperature monitoring can be adjusted independently of the operating modes.
Th
e manually set monitoring temperature
MIN
and
MAX
the overtemperature control is monitored by an adjustable over/undertemperature controller (TWW) protection class 3.3 acc. to DIN 1288
0
.
f the manually set monitoring temperature
MAX
is exceeded, the TWW takes over tempera-
ture control and be
g
ins to regulate the monitoring temperature (Fig. 21). The alarm icon
flashes as a warnin
g
Setting MAX Set
temperature
t
°C
Controller error
Emergency operation
Fig. 24 Schematic diagram of how the TWW temperature monitoring function
s
T
he monitoring temperature must always be set sufficiently high above the maximum
wor
king
temperature.
f the acoustic alarm is switched on in the SETUP, the TWW alarm is additionally signalled by an intermittent tone. If the SET key is pressed, the horn can be temporarily switched off until t
he ne
xt alarm event occurs.
Page 47
7
A
dvanced function
s
Setting:
g
vertemperature limit:
p to max. 10 °C above nominal temperature
for details of the nominal temperature, see nameplate)
Sett
ing
:
1.
ct the
MAX
icon with the push-turn control.
.
old down the SET key and set the desired temperature limit
with the push-turn control (e.
g
. 38.°C
)
°C
MIN
AUTO
MAX
Undertemperature limit:
A
djustment range: 10 °C below the minimum temperature of the appliance up to 10 °C above nominal temperature of the appli­ance (for details of the nominal temperature, see nameplate).
The l
ower alarm limit value cannot be set higher than the top
o
ne. If no undertemperature limit is required, set the lowest
t
emperature.
°C
MIN
AUTO
MAX
Page 48
48
A
dvanced function
s
7.3.3 Automatictemperature monitor (ASF)
A
SF is a monitoring device that automatically follows the set temperature setpoint within an
adjustable tolerance band (
. 25 ).
T
he ASF is activated – if switched on – automatically if the actual temperature value reaches 50 % of the set tolerance band of the setpoint (in the example: 37 °C – 1.5 °C) reached for the first time (section A). The activation of the ASF is shown by the bri
g
htly lit
AUTO
icon
.
When the temperature moves outside the set tolerance band around the setpoint (in the
e
xample in
. 25: 37 °C ± 3 °C) – e. g. if the door is opened during operation (section B of
illustration) – the alarm is set off. This is shown by the flashin
g
AUTO
and icons.
f the acoustic alarm is switched on in the SETUP, the ASF alarm is additionally signalled by an intermittent tone. If the SET key is pressed, the horn can be temporarily switched off until the n
e
xt alarm event occurs
.
T
he ASF alarm goes off automatically as soon as 50 % of the set tolerance band of the set-
point (in the example: 37 °C ± 1.5 °C) are reached a
g
ain (section C).
f the temperature setpoint is altered, the ASF is automatically disabled temporarily (in the
e
xample: The setpoint is changed from 37 °C auf 30 °C, section D), until it has reached the
tolerance ran
g
e of the new temperature setpoint (section E).
40 °C
34 °C
ASF active
40 °C
34 °C
ASF alarm
ASF active ASF active
33 °C
27 °C
°C
37 °C
t
ABCDE
AUTO AUTO AUTO
Fig. 25 Schematic diagram of how the ASF temperature monitoring function
s
Switching on the automatic temperature monitor:
gp
Select the AUTO icon with the push-turn control.
old down the SET key and set to on with the push-turn control.
MIN
MAX
AUTO
°C
Switching off the automatic temperature monitor:
gp
Select the AUTO icon with the push-turn control.
old down the SET key and set to
off
with the push-turn control.
T
he tolerance band for the ASF can be set in the SETUP in the
0.5 ... 5 °C (see page 44 ).
MIN
MAX
AUTO
°C
Page 49
49
A
dvanced function
s
Sterilisation
ipcar
d
or decontamination, the incubator i
s
e
quipped with a STERICard as standard
. 26). The STERICard starts an
automatic an
d
predetermined sterilisa-
t
i
on sequence.
W
arning
!
The incubator becomes
v
ery hot inside during
the sterilisation
p
roce-
d
ure. Do no open the
doo
r.
C
aution:
Th
e sterilisation programme is not
intended for sterilisin
g
the chamber
load, but only for sterilisin
g
the inte-
ior of the appliance. The incubator
is not a steri
l
iser with respect to the
aw on Medical Devices.
ake sure that the chamber is empty before the sterilisation procedure is started. Sliding
shel
ves and removed water trays may also be sterilised.
To activate,the STERICard, insert it into the chip card reader (see page
10 ) on the operating
panel of the appliance and select the Start icon
with the push-turn control while the SET
ey is held down.
A
s soon as the sterilisation procedure has been enabled, the automatic sterilisation procedure
egins: amp 1: Heating up to 160 °C amp 2: Hold time 4 hours (after setpoint temperature has been reached) amp 3: Cooling down to 70 °C amp 4 Wait time of 10 minutes
Th
e automatic sterilisation programme cannot be change
d.
A
fter the sterilisation has been completed, STERILISATION OK is shown in the controller display .
or documentation purposes, the sterilisation procedure is logged on the STERICard and can
be read out with a com
p
uter/laptop.
Sterilisation Process Control Card
Sterilisation Parameter: 160 °C / 4:00 h
Unit-ID: ________________________
Fig. 26 STERICard
Page 50
A
dvanced function
s
7.5 User ID card (optionally available as an accessory
)
The de
vice number of the incubator and a
nique user number are stored in
e
ncrypted form on the user ID card (
.
7 ). The user ID card therefore works
onl
y in the appliance with the corre-
s
ponding serial number
.
T
o use it, insert the card into the chip
c
ard reader (see page 10 ) on the appli-
ance operatin
g
panel
.
f the user ID card has been inserted, the
m
e
nu itemID LOCK also appears in the
SETUP menu. If the settin
g
is set to ON
t
h
e appliance is locked against all altera-
tions once t
h
e chip card is removed.
ocking with the user ID card is displayed with the lit up icon on the operating
anel.
very login process with the user ID card is logged in the internal Flash memory of the control-
r.
7.6 Calibrati
on
7.6.1
T
emperature calibratio
n
T
he incubator can be calibrated customer-specifically using three calibration temperatures of
your c
h
oice
:
C
AL.1Temperature calibration at low temperature (to 30°C)
C
AL.2 Temperature calibration at medium temperature (from 30°C to 38°C)
C
AL.3 Temperature calibration at high temperature (over 38°C)
or each selected balance point ( Fig. 28 ) a positive or negative compensation correction value
c
an be set between -2.9 °C and +2. 9 °C
There must be a difference of at least 10 ºC between
e
ach of the individual balance points.
CAL. 1
+0,2 °C
CAL. 3
+0,3 °C
CAL. 2
-0,4 °C
Default
calibration
30°C
20°C
10°C
40°C
Fig. 28
emperature calibration (example)
access authority card
Name:
ID:
_______________ ______ _______________ ______
Fig. 27 User ID Card
Page 51
1
A
dvanced function
s
Setting:
g
et compensation temperature in the SETUP (see page
) and set accompanying com
-
ensation correction value to
°C.
With a reference instrument, measure the deviation in the stationary state in the selected
c
ompensation temperature.
3
.
et the compensation correction value in the SETUP. If the measured reference tempera-
t
ure is too low, the compensation correction value must be set with a negative sign.
4
.
rform a control measurement with the reference instrument
.
5.
epeat the procedure for the other two balance points if necessary .
xample: The temperature deviation in the chamber load at 35 °C should be corrected.
1.
et the compensation temperature CAL.2 in the SETUP to 35.
°C and the accompanyin
g
c
ompensation correction value to. °C:
h
°C
°C
. With a calibrated reference instrument and at a set setpoint temperature of 35 °C in nor
-
al operation, an actual temperature of 34.6 °C is measured.
3
.
et compensation correction value for CAL.2 in the SETUP to - 0.4 °C:
h
°C
°C
4
. The reference instrument after the calibration procedure should now display 35 °C
5. With CAL.1 another compensation temperature below CAL.2 can be programmed in the
s
ame way , and with CAL.3
one lying above this.
f all compensation correction values are set to
°C, the factory calibration settings are
r
estored
.
Page 52
A
dvanced function
s
7.6.2
umidity calibration
T
he incubator can be calibrated for the individual customer by means of two balance points
at 20 an
d
at 90 % relative humidity . F or each selected balance point a positive or negative
c
ompensation correction value can be set between –5 % and +5 %.
rh 20
+3,0 %
rh 90
-2,0 %
60%
40%
0%
20%
80% 100%
Default calibration
Fig. 29 Humidity calibration (example)
Setting:
g
1.
et the desired humidity balance point in the SETUP (see page 44 ) and set accompanyin
g
c
ompensation correction value to0.0 °C.
. With a reference instrument, measure the deviation in the stationary state in the selected
umidity balance point.
3
.
etting the compensation correction value in the SETUP. If the measured reference humid-
ity is too
l
ow, the compensation correction value must be set with a negative sign.
4
.
rform a control measurement with the reference instrument
.
T
he procedure can be performed with humidity balance points of 20% rh and 90% rh. xample: Humidity deviation at 90 % should be corrected.
1.
et humidity balance point in the SETUP to RH 90 and set the accompanying compensation
corre
ction value t
o
0.0
%rh
:
PRINT
SETUP
loop
t3
t4
t2
t1
on off
Mo
Tu
We
Th
Fr
Sa Su
3
4
2
1
STERI DEFRO
°C
°C
rh
mb
%
CO mb
2
IN 1
IN 2
OUT
MIN
AUTO
MAX
. With a calibrated reference instrument, an actual humidity of 88% rh is measured at nor-
al operation, with a set setpoint humidity of 90 %rh
.
3
.
et the compensation correction value in the SETUP for RH 90 to –2.0 %rh
:
PRINT
SETUP
loop
t3
t4
t2
t1
on off
Mo
Tu
We
Th
Fr
Sa Su
3
4
2
1
STERI DEFRO
°C
°C
rh
mb
%
CO mb
2
IN 2
OUT
IN 1
IN 2
OUT
MIN
AUTO
MAX
Page 53
A
dvanced function
s
4
. The reference instrument should display 90.0% rh after the calibration procedure.
With
R
H 20 a further comparison can be programmed at 20 % relative humidity .
f all compensation correction values are set to
% rh, the factory calibration settings
ar
e restored
.
7.6.3
calibrati
on
ustomer-side calibration of appliance on controller by means of three C
O
2
points:
C
O
calibration at 5 % CO conten
t
C
O2 10 C
O
2
calibration at 10 % C
O
2
conten
t
C
O2 15 CO calibration at 15 % C
O
c
ontent (only for models with
C
O
or premium modules
)
or each selected balance point a positive or negative compensation correction value can be
set
.
co2 5
-0,4 %
20%
0%
10%
co2 10 +1,5 %
co2 15
-0,7 %
Default calibration
Fig. 30
CO
2
calibration (example)
Setting
g
5.
et the desired C
O
2
balance point in the SETUP (see page 44 ) and set accompanyin
g
c
ompensation correction value to 0.0 % rh.
. With a reference instrument, measure the deviation in the stationary state in the selected
2
balance point.
et the compensation correction value in the SETUP. If the measured reference CO
content
is too
l
ow, the compensation correction value must be set with a negative sign.
.
rform a control measurement with the reference instrument
.
. The procedure can be performed for the CO2 balance points 5 %, 10 % and 15 %
15 % only for models with CO2or premium modules).
Page 54
4
A
dvanced function
s
xample: C
O
d
eviation in chamber load at 10 % should be corrected:
1.
et the C
O
bal
ance point in the SETUP to CO2 10 and set the accompanying compensation
corre
ction value to.%:
PRINT
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa Su
3
4
2
1
STERI DEFRO
°C
°C
rh
mb
%
CO
mb
2
IN 1
IN 2
OUT
IN 1
IN 2
OUT
MIN
AUTO
MAX
. With a calibrated reference instrument and at a set CO content of 10.0 %, an actual C
O
c
ontent of 11.5 % is measured.
3
.
et the compensation correction value in the SETUP for CO2 1
0
to
1.5 %:
PRINT
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa Su
3
4
2
1
STERI DEFRO
°C
°C
rh
mb
%
CO
mb
2
IN 1
IN 2
OUT
IN 1
IN 2
OUT
MIN
AUTO
MAX
. The reference instrument should display 10 % after the calibration procedure.
Wit
h
CO2 5
and CO2 15 further calibrations can be programmed at 5 % and 15 % (15 % only
f
or models with C
O
2
or premium modules).
f all compensation correction values are set to0.0%, the factory CO2 calibration settings
ar
e restore
d.
7.6.4 Oxygen calibration
only for models with O2 module
)
n the
O
2
calibration, the oxygen proportion in the incubator at normal atmosphere is deter-
min
ed
.
1.
witch off the
O
2
module, if active. To do this, turn the ush-turn control to the left until the O2 display flashes. old down the set-key and turn the push-turn control ntil
O
2
off
is displayed. Release the set-key .
.
et CO setpoint value to 0% (see page 31 ).
3
.
pen the oven door and inner glass door , in case of
f
our-part door, remove glass panes. Keep the doors
o
pen for at least one minute so that a normal atmos
-
here can form in the chamber . Close the doors again
.
4
.
n the setup, select the item O2 calibration with the
ush-turn control
(
c
al 02 of
f
). Hold down the set key
and set to cal 02 on with the push-turn control. Release
th
e set key and leave the setup via EXIT.
ow, the oxygen content in the incubator is measured
f
or about one minute. In the display , cal o2 acti
ve
can be
s
een. Do not switch off the oven while this is taking
ace.
5. When the calibration is finished, the oxy
g
en content
d
etermined is shown in the display:
on off
Mo
Tu
We
Th
Fr
Sa Su
o2 off
off
CAL o2 on
CAL o2 active
o2=20.7%
Page 55
A
dvanced function
s
Communicationinterfaces
only for models with communication or premium modules
)
f the incubator is equipped with the communication or premium module, its log data can be
rea
d
out with a computer/laptop. For this purpose, the incubator has corresponding commu
-
nication interfaces on the rear of the appliance (see page 13 ).
T
o be able to use the RS 232, RS 485 And USB interfaces, the incubator must be assigned a
nique device address in the SETUP, menu item ADDRESS (see page
4), via which the computer
c
an communicate with the oven. The default setting is ADDRess 0 Using this address, the ap-
propriate incubator can be selected from the computer .
7.7.1 Communication Interface
S-232 C in accordance with DIN 12900-1
T
he computer can be connected with a shielded interface
c
able to the 9-pin interface on the rear of the appliance (see
. 31
and page 13). The shielding must be connected to the plug casing.
f the serial port is not used, put on the cover included. f several ovens are to be connected to a computer via RS-232-C
interfaces, an appropriate interface on the computer and a separate
c
able are required for each oven (
. 32 ). The maximum cable
ength is 15 m.
Pin allocation:
1
3
4
not oc
-
c
upie
d
x
D
TxD
not oc
-
c
upie
d
GN
D
not oc
-
c
upie
d
not oc
-
c
upie
d
oc-
c
upie
d
oc-
c
upie
d
RS 232
9-pin serial port
12345
6987
Fig. 31 RS-232-C
interface
Page 56
A
dvanced function
s
Fig. 32 Connecting incubators via the RS 232-C communication interface
7.7.2
us interface
S 485
f requested on the corresponding order , the oven can be fitted in the factory with a RS-485
interface instead of a RS-232-C interface. This enables the networkin
g
of several ovens (up to
16) with one computer via a shared two-wire cable (
. 34 ). The chamber must be given a
nique device address in the SETUP submenu, menu item
ADdRESS
(see page 44 ), via which the
c
omputer communicates with the oven. A maximum of 16 devices can be addressed on the
S-485 bus.
T
e default setting is ADDRess 0 Using this, the appropriate incubator
c
an be selected and programmed from the PC.
T
he computer needs to be equipped with either an RS-485 interface
o
r with an RS 232/RS 485 converter (
. 34). Connecting the wires
is done individually with a shielded cable, dependin
g
on the installa-
tion site. The maximum total cable len
g
th is 150 m.
220 ohm terminating resistor must be connected to the last device.
Pin allocation:
1
3
4
not oc
-
c
upie
d
not oc
-
c
upie
d
not oc
-
c
upie
d
not oc
-
c
upie
d
not oc
-
c
upie
d
not oc
-
c
upie
d
oc-
c
upie
d
RS 485
9-pin serial port
12345
6987
Fig. 33 RS-48
5
interface
g
Page 57
7
Advanced function
s
RS232
RS485
Interface converter
Fig. 34 Connecting incubators via the RS485 bus interface
7.7.3
SB interfac
e
f several chambers are to be connected to a computer via USB interface, an appropriate
interface on the com
p
uter and a separate cable are required for each chamber . The maximum
cable length
is 5 m.
7.7.4
thernet interfac
e
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
Fig. 35
Connecting one or more incubators to a network using an Ethernet interface
(schematic diagram)
g
g
g
Page 58
A
dvanced function
s
or identification purposes, each appliance connected must have its own unique IP address.
ach chamber is delivered by default with the IP address 192.168.100.100. The program
XTADMIN”, which can be found on the CD-ROM provided, can be used to change the IP ad-
dre
ss.
Settin
g
the IP address is described in a separate manual.
7.8 Logmemor
y
only for models with communication or premium modules
)
Th
e controller continually logs all relevant measured values, settings and error messages at
1-minut
e
intervals.
Th
e internal log memory is listed as a ring memory , i. e. the oldest log data are always over-
written automatica
ll
y with new data.
T
he logging function cannot be switched off, but is always active. The measured data are
s
tored in the controller , safe from manipulation. Each dataset is stored with a unique times-
t
amp.
T
he internal log memory has a size of 1024 kB. This corresponds to a storage capacity for
a
b
out three months in permanent operation.
f the power supply is interrupted, the time of the power cut and the return of voltage are
stored
in the contr
olle
r.
Reading in the log memory to the PC via USB interface
ggy
or documentation purposes, log data can be read out via an interface.
T
he log memory of the controller is not modified or deleted by the reading out.
Printing out log memory
gg y
see also Chapter “
“ on page
43
)
or a printout, the GLP header is also printed automatically , and contains the following details:
ate of printout
eriod of lo
g
onsecutive page numbers
erial numbers and appliance nam
e
f the printer is not ready (e. g. ink cartridge or paper tray empty), no log data will be
lost. Multiple printouts can also be made, since the lo
g
memory is not deleted after print
-
n
g.
Page 59
aintenance and Servicing
. Maintenance andServicin
g
.1
eanin
g
W
arning
!
anger of injury. Before any cleaning work, pull out the mains plug.
egular cleaning of the easy-to-clean chamber interior prevents build up of material remains, which over time could impair the appearance and functionality of the stainless steel
chambe
r.
T
o clean the interior , the fan cover can be removed by pulling
it sli
g
htly forwards, after the two fixing screws have been
s
crewed out (
. 36 ). The protective grid of the O
sensor c
an
if the model is equipped with a O
module) be taken off after
the fixin
g
screws have been removed (
. 37).
The me
tal surfaces of the chamber can be cleaned with nor-
mal stainless steel cleanin
g ag
ents. Make sure that no rusty
ob
jects come into contact with the working chamber or with
the stainless steel housin
g
. Rust deposits lead to an infection
o
f the stainless steel.
f rust spots should appear on the surface of the working
chamb
er , due to impurities, immediately clean and polish
the
affected area.
o not clean the operating panel, the plastic input module
and other
p
lastic parts of the chamber with caustic or solvent-
ased cleaning agents
.
.2
egular maintenanc
e
Monthly
y
or models with O module: Perform oxygen calibration (see page
4 ).
Every three months
y
heck that door seals fit tightly , adjust door if necessary (see page
).
n permanent mode: Grease the moving parts of the doors (hinges and lock) with thin
sil
icone grease and check that the hinge screws are not loose.
Annually:
y
Grease the moving parts of the doors (hinges and lock) with thin silicone grease and check
th
at the hinge screws are not loose.
heck that door seals fit tightly , adjust door if necessary (see page
).
Fig. 36 To take off th
e
fan cover, remove two
g
scre
ws
Fig. 37
O
2
sensor
Page 60
aintenance and Servicing
.3 Adjusting doo
r
A
well-closing door is indispensable for incubators.
n Memmert appliances, the tight closing of
t
he d
oor is optimally guaranteed by a chamber
s
eal and a door seal (see also page 10 ). In
permanent operation, it is possi
bl
e that the
flexible seal material will be
g
in to sag. To ensure
t
h
at the door closes exactly despite this, an
adjustment may be necessary (
. 38 ).
T
he top section (1) of the door hinge can be moved sli
g
htly in the direction of the arrow after
the two screws (2) at the to
p
and bottom of the
d
oor have been undone. Afterwards, tighten the
s
crews down again
.
Ad
justing the door:
1. Undo the ma
gg
ot screw (5) (with a jolting
otion, as it is fixed with locking paint).
. Adjust the door by turning the eccentric (3)
wit
h
a screwdriver.
3
. Apply locking paint to the maggot screw
an
d
re-tighten it.
T
he locking plate (
. 39 ) can also be adjusted
in the direction of the arrows after undoin
g
the
s
crew (6).
1. Undo screw (6).
.
ove locking plate (7) in direction of arrow.
3
. Tighten the screws again
.
.4 Repairs andServic
e
W
arning
!
A
fter removing covers, live arts may be exposed. You
ay receive an electric shock f you touch these parts. Disconnect the mains plug before remov­ng any covers. Any work inside the unit may only be performed by
q
ualified electricians.
epairs and service work are described in a separate service manual for the INCO
incu
bato
r.
2
1
3 4
5
Fig. 38
Adjusting door
jg
Mounting bolts
pp
3
Eccentric tappet
Eccentric socket
pp
p
5
Headless screw
7
6
Adjusting locking plat
e
6
Screw
7 Locking plat
e
Page 61
1
torage and Disposal
. Storage and Disposa
l
.1 Stora
ge
T
he incubator may only be used under the following conditions.
ry and in an enclosed, dust-free room
rost-fr
ee
isconnected from the power supply and gas supply
Undo
g
as bottle connections and close valves of gas bottles.
Gas bottles may be stored in closed rooms if these are sufficiently well ventilated.
For appliances with basic fittings:
pp g
emove, empty and clean the water tray(s).
For appliances with the humidity module:
pp y
Undo hose connection of water supply tank; empty water tank.
.2
isposa
l
T
his product is subject to the Directive 2002/96/EC on Waste Electrical lectronic Equipment (WEEE) of the European Parliament and of the EU
ouncil of Ministers. This appliance has been brought to the market
after 13th Au
g
ust 2005 in countries which have already integreated this directive into their national laws. It may not be disposed of in norma
l h
ousehold waste. To dispose, please contact your dealer or
manufacturer . Any appliances that are infected, infectious or contami
-
nated with materials that are a hazard to health are excluded from
eing taken back. Please observe the other regulations in this context.
efore disposing of the appliance, please render the door locking
mechanism unusable, for example, to prevent playin
g
children from
eing locked inside the appliance.
y
T
he appliance may not be left at public or communal recycling or collection points.
Page 62
nd
ex
Index
Accessorie
s 1
8
A
ccidents 9
A
coustic signal 4
4
Add
itional modules 12
Ad
justing door 60
A
dvanced functions 43
Ambie
nt conditions 1
8
Appl
iance error 39
A
SF 4
8
A
utomatic temperature
nitor 4
8
ase 21 asic device settings 43 asic equipment 11
C
alibration 44,
amber load 25 anges
8
cks 2
2
ipcard 49
eaning 5
9
ck time 44 2 calibration 5
3
O2 setpoint 31
mmunication address 4
4
mmunication interfaces
5
ompensation temperature
1 onformity 15 onnecting 2
2
onnection of external appli-
anc
e
s 14
nnections 1
3
nversions
8
rrective values 44,
1,
ustomer service
2
ate 44 eclaration of conformity 1
5
ecommissioning 6
1
elivery 19, esign 10
mensions 1
7
isposal 61
r 25 r seals 59
C Declaration of Conform
-
ty 15
ctrical connection 14 ectrical equipment 13 ectronic temperature moni-
toring
46
mergency
9
nding operation 37 NDSOUND 44 rrors 3
9
thernet interface 57 xplosion protection 14 xternal devices 22 xtras 11
unction 11
Gas connection 14 Gas supp
l
y 26
azards 7
w to set the IP address 5
8
umidity calibration 52,
cons used
6
nten
ded
use 14 nterface cable 43 nterface mode 2
8
nterfaces 55 rregularities
8
anguage 4
4
aptop 34 ength of lead 57 oading 25 og memory 4
2
,
43
ain switch 2
6
aintenance 59 alfunctions
8
,
38
anufacturer
2
aterial 13 emory 5
8
essages 3
8
inimum clearances 2
0
onitoring temperature 45
ameplate 15
rmal m
ode
2
8
,
2
9
O
2 17 2 module 12 2 setpoint 31 perating modes 28 perating mode - settings
2
8
perating panel 27 perating personnel 8,
5 peration 25 peration with PC/laptop 34 perator 25 xygen calibration 45,
4
ackaging material 1
9
arameters 27 C 34
rformance 1
6
wer failure 42 ower supply 22 rinter 4
3
rinting out log memory 5
8
roduct safety 7 t100 temperature sensor
4
5
ush/turn control 2
6
utting into Operation 22
altime clock 45
egular maintenance 5
9
ng memory 5
8
S-232 55 S 485 5
6
Page 63
nd
ex
Safety regulations 6 Serv
i
ce 6
0
Servicing 5
9
,
Setting parameters 27 Settin
g
s example norma
l
ode 3
0
Settings example week time
s
witch 33
Setu
p
19, 2
0
,
43
Stacking frame 2
1
Standard accessories 1
8
Storage after delivery 20 Stora
g
e capacity 58
Switchin
g
off 26
Switchin
g
on 26
System impedance 22
T
Tech
nical data 1
6
T
emperature calibration 50,
1
T
emperature comparison 5
0
T
emperature deviation 51
Temp
erature limiter (TB) 45
Temp
erature monitor 48
T
emperature monitoring 45
T
emperature monitorin
g
TWW) 46
Temp
erature sensor 45
Te
rms used
6
Tole
rance band ASF 44
T
ransport 1
9
T
ransport damage 1
9
T
WW 4
6
Unpacking 1
9
SB interface 57 ser ID card 5
0
W
Warning messages 4
9
Water connection 22 Water tray(s) 26 Wee
kd
ay 44
W
eek
time switch 2
8
,
31
Wei
gh
t 16
What t
o do
in case of ac-
cide
nts 9
Ye
ar 44
Page 64
Memmert GmbH + Co KG | Postfach 1720 | D-91107 Schwabach | Tel. +49 (0) 9122-925-0 | Fax +49 (0) 9122-145-85 | E-Mail: service@memmert.com | www.memmert.com
05.11.2014
INCO/INCO med englisch
D23532
Loading...