Thank you for choosing the Swing Flex / Swing Maxi Flex breast pump. Breast milk is the natural choice for your
baby, but there are reasons and situations where breastfeeding is not possible. For over 50 years, it has been
Medela’s conviction that these mothers and babies can be supported. In close cooperation with breastfeeding
experts, Medela, one of the market leaders, has researched and developed breast pumps that perfectly meet
breastfeeding mother’s needs. For mothers who insist on the best for their babies and themselves. Because you
care.
Medela vous remercie d‘avoir acheté un tire-lait Swing Flex/Swing Maxi Flex. Le lait maternel est ce qu‘il y a de
plus naturel pour votre enfant, mais dans certaines situations, l‘allaitement au sein n‘est pas possible. Depuis
plus de 50 ans, Medela se donne pour mission d‘assister ces mamans et ces bébés en difficulté. En étroite
collaboration avec des experts de l’allaitement, Medela a développé des tire-laits qui répondent de façon
optimale aux besoins des mères allaitantes. Ces appareils s‘adressent aux mères qui ne veulent que le meilleur
pour leur bébé et pour elles-mêmes. Parce que votre bien-être est important pour nous.
Hartelijk dank dat u voor de Swing Flex/Swing Maxi Flex borstkolf hebt gekozen. Moedermelk is een
natuurlijke keuze voor uw baby, maar er zijn redenen en omstandigheden waarin het niet mogelijk is om
borstvoeding te geven. Medela is er al meer dan 50 jaar van overtuigd dat deze moeders en baby‘s
ondersteund kunnen worden. In nauwe samenwerking met borstvoedingsdeskundigen heeft Medela, een van
de marktleiders, borstkolven onderzocht en ontwikkeld die perfect aansluiten bij de behoeften van moeders
die borstvoeding geven. Voor moeders die het beste willen voor hun baby en voor zichzelf. Uit liefde.
Terima kasih telah memilih Pompa ASI Swing Flex / Swing Maxi Flex. ASI adalah pilihan alami bagi bayi Anda,
tetapi situasi dan kondisi kadang membuat menyusui tidak mungkin. Selama lebih dari 50 tahun, Medela yakin
ada yang dapat dilakukan untuk membantu ibu-ibu dan bayi-bayi ini. Medela, salah satu pemimpin pasar,
bekerja sama dengan pakar per-ASI-an, berhasil meneliti dan mengembangkan pompa ASI yang benarbenar memenuhi kebutuhan ibu menyusui. Untuk para ibu yang menginginkan yang terbaik bagi bayi mereka
juga mereka sendiri. Karena Anda peduli.
EN
4–35
FR
36–67
NL
68–99
ID
100–131
Table of Contents
44
1. Intended use / Population – Contraindication5
2. Meaning of symbols6
3. Important safety information9
4. Product description Swing Flex10
5. Product description Swing Maxi Flex11
6. Cleaning12
6.1 Cleaning overview12
6.2 Disassemble
6.3 Wash
6.4 Sanitise
6.5 Cleaning tubing
6.6 Cleaning the motor unit
13
14
14
15
15
7. Preparing for pumping16
7.1 Mains operation16
7.2 Battery operation
7.3 Assembling the pump set
16
18
8. Expression20
8.1 Swing Flex – Preparing for single pumping20
8.2 Swing Maxi Flex – Preparing for single pumping
8.3 Swing Maxi Flex – Preparing for double pumping
8.4 Swing Maxi Flex – Preparing for hands-free expression with
the Easy Expression Bustier
8.5 Pumping
26
24
20
22
9. Storing and thawing breast milk28
9.1 Storage28
9.2 Freezing
9.3 Thawing
2828
10. Troubleshooting 29
11. Warranty / Disposal30
12. Spare parts and Accessories31
13. EMC / Technical description33
14. Technical specifications34
1. Intended use / Population – Contraindication
The intended use and indications of breast pumps
Breast pumps are intended to be used by lactating women in the hospital or home environment for expressing breast milk. Breast pumps are indicated to alleviate the symptoms of
engorgement by expressing milk. In case of mastitis breast pumps support the healing
process by removing breast milk from the affected breast. Breast pumps are indicated to
relieve sore and cracked nipples and to bring out flat or inverted nipples. Also breast
pumps enable mothers to provide breast milk for babies who cannot feed directly at the
breast for a variety of reasons such as latch-on problems, cleft palate or premature babies.
5
Intended patient population/user
Adolescent paediatric or adult lactating women. Many women find it convenient, or even
necessary, to use a breast pump to express and store their breast milk once they have
returned to work, are travelling or are otherwise separated from their baby. Women are
often delighted to learn that a breast pump can be used as a supplement to breastfeeding
and that some pumps are designed to mimic the suckling of a nursing baby.
Contraindications
There are no known contraindications for the Swing Flex / Swing Maxi Flex breast pump.
2. Meaning of symbols
The warning symbol identifies all instructions that are important for safety. Failure to
66
observe these instructions can lead to injury or damage to the breast pump!
When used in conjunction with the following words, the warning symbol stands for:
Warning
Can lead to serious injury or death.
Caution
Can lead to minor injury.
Symbols on the packaging
This symbol indicates that the material is part of a recovery/recycling process.
This symbol indicates a carton package.
This symbol indicates to keep the device away from sunlight.
This symbol indicates to handle the fragile device with care.
This symbol indicates the temperature limitation for operation, transport and storage.
This symbol indicates the humidity limitation for operation, transport and storage.
This symbol indicates the atmospheric pressure limitation for operation,
transport and storage.
This symbol indicates to keep the device dry.
i
i
Note
Can lead to material damage.
Info
Useful or important information that is
not related to safety.
This symbol indicates that the packaging contains products intended to come in contact with
food according to regulation 1935/2004.
This symbol indicates unique GSI Global Trade Item Numbers (GTIN).
This symbol indicates do not dispose the device together with unsorted municipal waste (for
EU only).
This symbol indicates to follow instructions for use.
Symbols on the device
This symbol indicates to follow instructions for use.
This symbol indicates the compliance with the essential requirements of the Council Directive
93/42/EEC of 14 June 1993 concerning medical devices.
This symbol indicates the manufacturer.
This symbol indicates do not dispose the device together with unsorted municipal waste (for
EU only).
This symbol indicates the compliance with additional USA and Canada safety
requirements for medical electrical equipment.
This symbol indicates a type BF applied part.
This symbol indicates manufacturer’s serial number of the device.
This symbol indicates the positioning of battery cell.
This symbol indicates the protection against ingress of solid foreign objects and against
harmful effects due to the ingress of water.
This symbol indicates the date of manufacture (four digits for the year and
two digits for the month).
Symbols on the mains adapter
This symbol indicates that the mains adapter is a class II device.
7
This symbol indicates the compliance with China Compulsory Certification.
This symbol indicates that the mains adapter is for indoor use only.
This symbol indicates the compliance with USA and Canada safety requirements.
This symbol indicates the compliance with Federal Communications Commission requirements.
This symbol indicates polarity of d.c. power connector.
This CE-mark indicates the compliance with the low voltage and electromagnetic compatibility directive.
This symbol indicates to not dispose the device together with unsorted municipal waste (for
EU only). The solid bar symbol indicates that mains adapter is put on the market after 13
August 2005.
88
SGS
This symbol indicates the compliance with energy efficiency requirements.
This symbol indicates alternating current.
This symbol indicates direct current.
This symbol indicates the compliance with energy efficiency requirements.
This symbol indicates that the mains adapter is safety tested.
This symbol indicates a short-circuit proof safety isolating transformer.
170053-12
S
TÜV
SÜD
PSB
IP20
Singapore Consumer Protection Registration.
This symbol indicates that the electronic device incorporating transformer(s) and electronic
circuitry(ies), converts electrical power into single or multiple power outputs.
This symbol indicates the compliance with AUS/NZ requirements (regulatory compliance
mark).
This symbol indicates the compliance with PSE requirements (Japan, Product Safety Electrical
Appliance and Material Safety Law).
The KC mark indicates that the product meets Korean EMC and electric safety requirements.
This symbol indicates the protection against ingress of solid foreign objects and against
harmful effects due to the ingress of water.
3. Important safety information
MAX.
275 h
MAX.
250 h
Failure to follow these instructions/safety information can lead to danger from the device.
Subject to technical changes.
Device
Danger of electrocution! Keep the device
dry! Never immerse in water or other
liquids!
Do not expose the motor unit to direct
sunlight.
9
The Swing Flex / Swing Maxi Flex breast
pump is not heat-resistant: Keep away
from radiators and open flames.
Never use a damaged device. Replace
damaged or worn parts.
The operating life of the Swing Flex
breast pump is 275 hours. The service
life is two years.
The operating life of the Swing Maxi Flex
breast pump is 250 hours. The service
life is two years.
Electrical use
Separation from the mains is only
assured through the disconnection of
the mains adapter from the wall socket
outlet.
Do not reach for any electrical device if it
has fallen into water.
Unplug immediately.
Use
Use the Swing Flex / Swing Maxi Flex
breast pump only for its intended use as
described in this manual.
Do not use the Swing Flex /Swing Maxi
Flex breast pump while bathing or
showering.
Do not use the Swing Flex / Swing Maxi
Flex breast pump while sleeping or
overly drowsy.
This is a single user product. Use by
more than one person may present a
health risk.
Never operate an electrical device if it has a damaged
cord or plug, if it is not working properly, if it has
been dropped or damaged, or dropped into water.
Repairs must be performed only by an
authorised service agency.
Do not repair yourself! No modifications
to the device are permitted.
Keep the mains adapter away from
heated surfaces.
The breast pump should never be left
unattended when plugged into a power
source.
Do not drive whilst hands-free pumping.
Supervision is necessary when the Swing
Flex / Swing Maxi Flex breast pump is used
in the vicinity of children.
Contact your lactation consultant or
breastfeeding specialist if you experience
problems or pain.
Portable and mobile radio frequency
communications equipment can affect the
Swing Flex / Swing Maxi Flex breast pump.
4. Product description Swing Flex
Swing Flex – single electric 2-Phase
1010
breast pump
1x motor unit:
article 600.0975
spare part 099.0015
on/off button
LED
swing
fixing clip
PVC tubing port
reduce
vacuum
breast shields:
1 x 21 mm (article 101035239)
1 x 24 mm (article 101035240)
1x mains adapter
article spare part
Euro 4.8V 201.1055 099.0016
UK 4.8V 201.1056 099.0017
AUS/NZ 4.8V 201.1057 099.0018
CN 4.8V 201.1058 099.0276
US/JP 4.8V 201.1059 101033542
KR 4.8V 101034044 101038305
1x Swing Flex tubing
ar ticl e 201.1079
spare part 1010340 02
increase
vacuum
let-down
button
battery
compartment
1x connector
article 200.9728
power adapter port
1x multi lid
ar ticle 20 0.2714
1x lid
1x membrane
1x bottle
article 811.0021
1x bottle stand
article 101033405
Spare parts and accessories available for order can be found in chapter 12.
5. Product description Swing Maxi Flex
Swing Maxi Flex – double electric 2-Phase
breast pump
1x motor unit:
article 200.4932
spare part 099.0270
on/off button
reduce
vacuum
1x mains adapter
article spare part
Euro 12V 201.1146 099.0271
UK 12V 201.1147 099.0272
AUS/NZ 12V 201.1149 099.0273
CN 12V 101032865 099.0274
US/JP 12V 201.1148 101033543
KR 12V 101034045 101038303
LED
swing maxi
increase
vacuum
let-down
button
fixing clip
PVC tubing port
battery
compartment
power adapter port
breast shields:
2 x 21 mm (article 101035239)
2 x 24 mm (article 101035240)
2x connector
article 200.9728
11
2x multi lid
ar ticle 20 0.2714
2x lid
2x membrane
short tubings
tubing holder
long tubing
1x Swing Maxi Flex
tubing
ar ticl e 201.1087
spare part 1010340 03
2x bottle
article 811.0021
2x bottle stand
article 101033405
6. Cleaning
6.1 Cleaning overview
1212
Washing and sanitising are two different activities. They must be done separately to protect
you, your baby and the performance of your breast pump.
Wash – To clean the surfaces of the parts by physically removing contamination.
Sanitise – To kill living organisms, such as bacteria or viruses, that may be present
on the surfaces of the parts.
When to clean
Before 1st use
After each use
Once per day
When to clean
Breast shield(s)
✔ 6.2 disassemble
✔ 6.3 wash
✔ 6.4 sanitise
✔ 6.3 wash
✔ 6.4 sanitise
Tubing
Connector and
membrane
✔ 6.2 disassemble
✔ 6.3 wash
✔ 6.4 sanitise
✔ 6.2 disassemble
✔ 6.3 wash
✔ 6.2 disassemble
✔ 6.4 sanitise
Bottles & lids
✔ 6.2 disassemble
✔ 6.3 wash
✔ 6.4 sanitise
✔ 6.2 disassemble
✔ 6.3 wash
✔ 6.2 disassemble
✔ 6.4 sanitise
Motor unit
As needed
see chapter 6.5see chapter 6.6
6.2 Disassemble
*
Caution
• Do not spray or pour liquid directly onto
the pump.
• Only use drinking-quality water for
cleaning.
• Take apart and wash all parts that come
in contact with the breast and breast milk
immediately after use to avoid drying up
of milk residues and to prevent growth of
bacteria.
• Do not clean the tubing with Quick Clean
bags in the microwave.
• Unplug the breast pump before cleaning
the motor unit.
i
Note
• The pump set requires no maintenance.
• Take care not to damage parts of the
pump set during cleaning.
• Do not use the breast milk if bottles
or components become damaged.
13
• Remove and dispose the protective cover*
from the breast shield before first use.
• Disassemble the pump set into its
individual parts (breast shield, connector,
membrane, milk bottle).
• Open back cap of the connector by
squeezing both flaps.
• Check all parts for damage.
• Replace damaged or worn parts.
6. Cleaning
1414
6.3 Wash
Before first use and after each use
SinkStovetop
1. Rinse all parts with cold, clear
drinking-quality water (approx.
20 °C).
2. Clean these parts with plenty of
6.4 Sanitise
Before first use and once per day
Cover all parts, except the tubing,
with water and boil at least for
5minutes. Dry after sanitising.
warm, soapy water (approx. 30 °C).
Use a commercially available
washing-up liquid, preferably without
Microwave
artificial fragrances and colouring
(pH neutral).
3. Rinse the parts with cold, clear
drinking-quality water for
10–15seconds (approx. 20 °C).
Dry after washing.
OR
OR
Dishwasher
Place all parts, except for the tubing,
on the top rack or in the cutlery section.
Use a commercially available dishwashing detergent. Dry after washing.
Use Quick Clean* bags in the microwave in accordance with the instructions. Dry after sanitising.
* Accessories available for order:
see chapter 12. For more information,
visit www.medela.com.
Drying
Dry with a clean cloth or leave to dry on a clean cloth. Put the clean parts in a clean
storage bag or a clean environment. Do not store parts in an airtight container/bag if
moist. It is important that all residual moisture dries.
6.5 Cleaning tubing
1
• No visible residues
No cleaning needed.
•With visible residues
1. Rinse tubing in cool water
2. Wash tubing in warm, soapy water
3. Rinse tubing with clear water
4. Shake out water droplets
5. Hang to air dry
Do not sanitise the tubing.
6.6 Cleaning the motor unit
i
Note
• If the individual components of the pump
set are cleaned in a dishwasher, parts
may be discoloured by food pigments.
This has no impact on the function.
• The membrane at the connector prevents
the milk from flowing into the tubing,
however, inspect the tubing after each
pumping session for visible residues. If
there are visible residues in the tubing,
follow the instructions in chapter 6.5.
15
1
To clean the motor unit, wipe the housing
with a clean, damp (not wet) cloth. You may
use drinking-quality water and a basic
washing-up liquid or a non-abrasive
detergent.
2
Additionally you may use an alcohol-based
disinfecting agent.
7. Preparing for pumping
1616
Warning
• Mains operation:
Follow the exact order of the steps.
• Only use the mains adapter supplied with
the Swing Flex / Swing Maxi Flex.
• Make sure the voltage of the mains
adapter is compatible with the power
source.
Info
i
• You can operate the SwingFlex / Swing
Maxi Flex breast pump via battery or
mains.
• Battery operation:
The batteries do not need to be removed
when the mains adapter is used for
pumping.
• Regularly check the condition of the
batteries in the motor unit.
• If Swing Flex / Swing Maxi Flex will remain
unused for a longer period of time,
remove the batteries from the battery
compartment.
• Batteries last for 4 – 6 pumping sessions
(approx. 1 ½ hours pumping time).
7.1 Mains operation
1
Plug the mains adapter cord into
the motor unit.
7.2 Battery operation
1
Open the battery compartment on
the back in the direction of the arrow.
2
Plug the mains adapter into
the mains socket. The motor unit
is now ready for operation.
17
2
Insert four AA Mignon LR6 batteries
for Swing Flex .
Insert six AAA Micro 1.5 V batteries
for Swing Maxi Flex.
3
Close the battery compartment.
The pump is ready for operation
via battery.
7. Preparing for pumping
1818
Caution
• Wash hands thoroughly (at least one
minute) with soap and water before
touching the breast and pump parts. Dry
your hands with a fresh towel or a
single-use paper towel.
• Also, bottles and component parts may
become damaged if mishandled, e.g.
dropped, over-tightened, or knocked
over.
• Take appropriate care in handling bottles
and components.
i
Note
• Use Medela original accessories only.
• Hold the tubing by the connecting piece
while attaching and detaching it.
• All components must be completely dry
before use.
i
Info
7.3 Assembling the pump set
1
Check pump set components for wear or
damage before use and replace if necessary.
Make sure parts are cleaned according to
chapter 6 before assembling the pump set.
Carefully place the membrane into the
connector with the flap into the opening of
the connector.
Make sure that the membrane forms a seal
around the edge of the connector.
4
• Carry out all the steps with care and
assemble the pump set correctly. Otherwise a good vacuum may not be
achieved.
Screw bottle into connector.
19
32
Close the lid of the connector
until it audibly clicks.
56
Insert the tubing into the corresponding
opening in the lid of the connector.
Insert the breast shield onto the
connector.
Insert the other end of the tubing
into the motor unit.
8. Expression
2020
Caution
• Step 2 (chapters 8.1, 8.2, 8.3):
• The product comes with two different
breast shield sizes 21 mm and 24 mm.
• Answer the following questions
to find out whether the available breast
shield sizes fit:
• Does your nipple rub the tunnel sides to
the point of causing discomfort?
• Do you see excessive areola being pulled
into the tunnel?
• Do you see any redness?
• Is your nipple or areola turning white?
• Do you feel unexpressed milk after breast
milk pumping?
If you answered “YES” to any of these
questions, consider trying a new size by
following the measuring instructions at
www.medela.com.
If you are still unsure that you selected the
right breast shield size, contact your
lactation consultant or breastfeeding
specialist.
See brochure “Choosing your Medela
breast shield size”.
8.1 Swing Flex – Preparing for
single pumping
1
The oval breast shield can be rotated (360°)
and placed in the desired position to have
the most comfortable fit for you.
8.2 Swing Maxi Flex – Preparing
for single pumping
1
i
Note
• Step 3 (chapters 8.1, 8.2, 8.3):
Do not hold the pump set by the bottle.
This can lead to blockage of the milk
ducts and engorgement.
Insert the unused short tubing into the
tubing holder.
Place the breast shield on your breast
so that your nipple is properly centred
in the tunnel.
23
32
For more information
on pumping, see
Hold the breast shield onto your breast
with your thumb and index finger. Support
your breast with the palm of your hand.
21
chapter 8.5.
Place the breast shield of one bottle
on the first breast so that the nipple is
prop erly centred in the tunnel.
For more information
on pumping, see
chapter 8.5.
Hold the breast shield onto your breast with
your thumb and index finger. Support your
breast with the palm of your hand.
8. Expression
2222
Caution
• Step 2 (chapter 8.3):
• Wash hands thoroughly (at least one
minute) with soap and water before
touching the breast, the clean pump parts
and clean pump set. Dry your hands with
a fresh towel or a single-use paper towel.
• Do not hold the pump set by the bottle.
This can lead to blockage of the milk ducts
and engorgement.
i
Info
• Double pumping saves time and increases
the energy content of the milk.
8.3 Swing Maxi Flex –
Preparing for double
pumping
1
The oval breast shield can be rotated (360°)
and placed in the desired position to have
the most comfortable fit for you.
4
Switch on the breast pump with
.
23
2
Place the breast shield on your breast so
that your nipple is properly centred in the
tunnel.
5
3
Hold the breast shield onto your breast
with your thumb and index finger.
Support your breast with the palm
of your hand.
1
For more information
Place the second breast shield on your
breast so that your nipple is properly
centred in the tunnel.
on pumping, see
chapter 8.5.
8. Expression
i
2424
Note
• Easy Expression Bustier must be pur-
chased seperately. More information on
www.medela.com
8.4 Swing Maxi Flex –
Preparing for hands-free
expression with the
Easy Expression Bustier
1
Put on the Easy Expression
Bustier and close it using the hook on top.
4
Connect the assembled pump set to the
breast shields.
23
25
Leave the zip open slightly.
5
Start pumping.
Place the breast shield on your breast
under the bustier such that your nipple is
centred in the tunnel of the breast shield.
For more information
on pumping, see
chapter 8.5.
8. Expression
2626
Caution
• Contact your your lactation consultant or
breastfeeding specialist if you can express
only minimal or no milk, or if expression is
painful.
i
Note
• Always disconnect the Swing Flex /Swing
Maxi Flex from the power source after
expression.
• Only fill the bottle to the 150 ml mark.
i
Info
• Swing Flex / Swing Maxi Flex switches off
automatically if it has been running for
30 minutes with no interaction.
• Do not kink the tubing during expression.
The following table shows the different
LED status:
8.5 Pumping
1
Switch on the breast pump with .
The stimulation phase lasts a maximum
æ
of two minutes. The stimulation phase
will then automatically change to the
expression phase.
4
Lights upOn
Does not light upOff
Even, regular flashingStimulation phase
Light constantly onExpression phase
Rapid flashingMin./Max. vacuum
Switch off the breast pump with .
23
27
If your milk starts to flow earlier, press the
button to change to the expression
phase.
Make sure that the milk is flowing into
æ
the bottle properly.
56
Use the bottle stand to prevent the bottle
from tipping over.
The Max. Comfort vacuum can be adjusted
in every phase. Max. Comfort vacuum: Use
to increase the vacuum until it is slightly
uncomfortable. Press
vacuum, until it is comfortable again.
Close the bottle with a lid.
Follow the instructions in chapter 9
æ
“Storing and thawing breast milk”.
to reduce the
Clean in accordance
with chapter 6.
9. Storing and thawing breast milk
9.1 Storage
2828
Storage guidelines for freshly expressed breast milk (for healthy term babies)
Room temperature
16 to 25 °C
(60 to 77 °F)
Up to 4 hours is best
*Up to 6 hours for milk
expressed under very
clean conditions
Refrigerator
4 °C (39 °F)
or colder
Up to 3 days is best
*Up to 5 days for milk
expressed under very
clean conditions
Freezer
–18 °C (0 °F )
or colder
Up to 6 months is best
*Up to 9 months for milk
expressed under very
clean conditions
*Guidelines for expressing milk under very clean conditions:
Before expressing breast milk, mothers should wash their hands with soap and water or an
alcohol-based hand sanitiser. The pump parts, bottles and pumping area must be clean.
Breasts and nipples do not need to be washed before pumping.
• These guidelines for storage and thawing of breast milk are a recommendation. Contact
your lactation consultant or breastfeeding specialist for further information.
• Store breast milk in the coldest spot of the refrigerator (at the back of the glass shelf
above the vegetable compartment).
Breast milk thawed in
the refrigerator
At room temperature:
Up to 2hours
Refrigerator:
Up to 24 hours
Do not refreeze!
9.2 Freezing
• Freeze expressed breast milk in plastic bottles or milk storage bags. Do not fill the bottles
or bags more than 3/4 full to allow space for possible expansion.
• Label the bottles or bags with the date and volume of expression.
• Plastic bottles and component parts become brittle when frozen and may break when
dropped.
• Do not use the breast milk if bottles or components become damaged.
9.3 Thawing
Caution
Do not thaw or warm breast milk in a microwave or a pan of boiling water. This helps
preserve important components and prevents burns.
• To preserve breast milk components, thaw the milk overnight in the refrigerator. Alternatively, hold the bottle or bag under warm water (max. 37 °C / 98.6 °F).
• Gently swirl the bottle or bag to mix any fat that has separated. Avoid shaking or stirring
the milk.
10. Troubleshooting
Problem
Solution
29
If the motor is not
running
If there is low
or no suction
If condensation
appears in
the tubing
The pump
becomes wet
Check presence of mains current.
Check the position of the batteries in the battery compartment.
Verify the membrane is clean and not damaged.
Make sure that the membrane forms a seal around the edge of the
connector and the lid is firmly connected with the connector.
Check all connections.
Check if the ends of the tubing fit securely into the back of the connector
and into the port on the breast pump. During single pumping, insert the
unused tubing into the tubing holder.
There should be no water in the tubing.
Do not kink the tubing during expression.
Ensure the breast shield forms a complete seal around the breast.
All components must be completely dry.
Clean the tubing as described in chapter 6.5. Shake out any water drops.
Unplug the mains adapter from the socket.
Switch off the pump.
Do not place the pump upside down. The control panel must point upwards.
Store the pump in a dry, warm place for 24 hours.
Note:
Do not expose the pump to direct sunlight.
Do not accelerate the drying process by use of heat.
11. Warranty / disposal
Warranty
3030
The “International Warranty” brochure contains detailed warranty information.
Disposal
This is a single user product. Use by more than one person may present a health risk.
The unit is made of metals and plastics and should be disposed of in accordance with the
European directives 2011/65/EU and 2012/19/EU. Additional, local guide-lines must also
be observed. Please take care that you dispose of the unit and its accessories in accordance with your local disposal guidelines.
Loading...
+ 102 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.