1. Important Safeguards
WARNING: To avoid fire, electrocution, or serious burns:
• This is a single-user product. Use by more than one person may present a health risk and voids the warranty.
• Always unplug electrical product immediately after use except for when charging.
• Do not place or store product where it can fall or be pulled into a tub or sink.
• The breast pump and accessories are not heat-resistant: keep away from heated surfaces or open flames.
• Do not use near flammable materials.
• The breast pump should never be left unattended when plugged into a power source.
• Never operate an electrical device if it has a damaged cord or plug, is not working properly, or if it has been dropped or damaged. If damage is found, immediately discontinue use and call Medela Customer Service at 1-800-435-8316.
• Do not use an electrical outlet device that has been exposed to water or other liquids including:
- Do not use while bathing or showering.
- Never place or drop into water or other liquids.
- Do not run water over breast pump.
- If a device has been exposed to water or other liquids, do not touch, unplug the device from electrical outlet, turn off and contact manufacturer.
T
The defined signal words identify all instructions that are important to safety. Failure to observe these instructions can lead to injury or damage to the breast pump or yourself. When used in conjunction with the following words, the defined signal words stand for:
WARNING Can lead to serious injury or death.
CAUTION Can lead to minor injury.
NOTE Can lead to material damage.
TIP Useful or important information that is not related to safety.
When using electrical products, especially when children are present, basic safety precautions should always be followed.
READ ALL INSTRUCTIONS BEFORE USING THIS PRODUCT.
Page 3
Questions? Visit www.medela.com or call us at 1-800-435-8316.
WARNING: To avoid health risk and reduce the risk of injury:
• This device cannot be serviced or repaired. Do not repair yourself. Do not modify the device.
• Never use a damaged device. Replace damaged or worn parts.
• Use only the power adaptor that comes with the breast pump.
• Use the breast pump only for its intended use as described in this manual.
• Do not use the breast pump while sleeping or overly drowsy.
• Do not use Sonata breast pump while operating a moving vehicle.
• Inspect all appropriate pump components before each use.
• If tubing becomes moldy, discontinue use and replace tubing.
• Pumping can induce labor. Do not pump until after giving birth. If you become pregnant while breastfeeding or breast pumping, consult with a licensed healthcare professional before continuing.
• If infected with Hepatitis B, Hepatitis C, or Human Immunodeficiency Virus (HIV), pumping breast milk will not reduce or remove the risk of transmitting the virus to your baby through your breast milk.
• Do not microwave or boil breast milk. Microwaving can cause severe burns to baby’s mouth from hot spots that develop in the milk during microwaving. (Microwaving can also change the composition of breast milk.)
• Clean and sanitize all parts that come into contact with your breast and breast milk prior to first use.
• Wash all parts that come into contact with your breast and breast milk after every use.
• Only use Medela recommended parts with your SonataTM Breast Pump.
• Close supervision is necessary when the breast pump or accessories are used in the vicinity of children.
Page 4
CAUTION: Can lead to minor injury:
• Portable and mobile radio frequency communications equipment can affect the breast pump.
• Make sure the voltage of the power adaptor is compatible with the power source. See Section 23 for technical specifications.
• Wash hands thoroughly with soap and water before touching breast pump, kit and breasts, and avoid touching the inside of bottles or lids.
• Separate and wash all parts that are exposed to breast milk immediately after use. This will help remove breast milk residue and prevent growth of bacteria.
• Always inspect breast shields, connectors, valves, membranes, bottles, lids, and tubing prior to use for cleanliness. Contact Medela Customer Service if cleaning does not resolve the issue.
• Only use drinking-quality tap or bottled water for cleaning your breast pump and parts.
• Do not store wet or damp parts as mold may develop.
• Do not run pump with wet tubing. Doing so may damage the breast pump.
• If you are experiencing discomfort at the base of the nipple due to rubbing of your breast tissue against the breast shield tunnel, use of a lubricant such as Tender CareTM Lanolin may be beneficial. For assistance with correct breast shield sizing and comfort please visit www.MedelaBreastshields.com or see a lactation consultant/breastfeeding specialist.
• If pumping is uncomfortable or causing pain, turn the unit off, break the seal between the breast and the breast shield with your finger and remove the breast shield from your breast.
• Contact your health care professional or breastfeeding specialist if you can express only minimal or no milk or if expression is painful. See Section 7 & 8 for more information.
• While some discomfort may be felt when first using a breast pump, using a breast pump should not cause pain. For assistance with correct breast shield sizing and comfort please visit www.MedelaBreastshields.com or see a lactation consultant / breastfeeding specialist.
• Do not try to express with vacuum that is too high and uncomfortable (painful). The pain, along with potential breast and nipple trauma, may decrease milk output.
• Make sure tubing is not kinked or pinched while pumping.
• Do not hold the pump kit by the bottle. This can lead to blockage of the milk ducts and engorgement. See Section 11 for more information.
• The breast pump uses Bluetooth® technology. Please follow your air carrier rules for the use of Portable Electronic Devices when flying with your breast pump.
• Using a breast pump on an aircraft is not recommended. The cabin pressure may affect the breast pumps performance.
Note
• Plastic bottles and parts become brittle when frozen and may break when dropped.
• Bottles and parts may become damaged if mishandled (e.g., dropped, over-tightened, or knocked over).
• Take appropriate care in handling bottles and components.
• Do not use the breast milk if bottles or components become damaged.
1. Important Safeguards (cont.)
Page 5
Questions? Visit www.medela.com or call us at 1-800-435-8316.
PLEASE SAVE THESE INSTRUCTIONS.
THIS IS A SINGLE-USER PRODUCT.
Indications for use
The SonataTM breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breast. The Sonata Breast Pump is a single user device.
Tip
It is best to wait until your breast milk routine is established (approximately 4 weeks) before expressing breast milk, unless otherwise advised by your healthcare professional.
If you have medical reasons or other needs for exclusively pumping, it is recommended that you use a hospital-grade (multi-user) breast pump such as the Medela Symphony® Breast Pump. To learn more, visit www.medelabreastfeedingus.com.
Contraindications for use
Product Description
This breast pump is a personal-use electric breast pump that includes 2-Phase Expression® technology and is capable of single and double pumping.
The operating life of this breast pump is defined to be approximately three 15-minute sessions per day, for one year. The operating life for the breast pump kit is 6 months.
There are no known contraindications for use with this product.
T
Page 6
Page 7
Questions? Visit www.medela.com or call us at 1-800-435-8316.
1. Important Safeguards ................................................................................................2-5
2. Product Description ...................................................................................................8-9
3. Getting Started ............................................................................................................. 10
4. Cleaning Overview ....................................................................................................... 11
5. Cleaning Instructions .............................................................................................12-13
6. Breast Pump & Tubing Care................................................................................... 14-15
6A Cleaning the breast pump ......................................................................................... 14
6B Tubing care ............................................................................................................... 15
7. Breastfeeding Information .......................................................................................... 16
8. Breast Shield Sizing ..................................................................................................... 17
9. Assembling Your Breast Pump .............................................................................. 18-19
10. Powering Your Breast Pump..................................................................................20-22
10A Charging your breast pump ..................................................................................... 20
10B Battery care & status codes ...............................................................................20-21
10C Power adaptor care................................................................................................. 22
10D Traveling outside the U.S. ........................................................................................ 22
11. Operating Your Breast Pump ................................................................................. 23-26
11A Glossary, controls & icons........................................................................................ 23
11B Single pumping ....................................................................................................... 24
11C Double pumping...................................................................................................... 24
11D Pumping steps ................................................................................................... 25-26
12. Finding Your Maximum Comfort VacuumTM ............................................................... 27
13. Features ................................................................................................................... 28-33
Breast Pump Connectivity ..........................................................................................28-29
Selecting a rhythm .......................................................................................................... 30
Setting the timer ............................................................................................................. 31
Pausing your breast pump .............................................................................................. 32
Silencing your breast pump chimes ................................................................................ 33
14. Storing Breast Milk ...................................................................................................... 34
15. Preparing & Feeding Breast Milk ............................................................................... 35
16. Troubleshooting ...................................................................................................... 36-41
Air leak alert .................................................................................................................... 36
Battery drained ............................................................................................................... 37
Battery broken ................................................................................................................ 38
Bluetooth broken ............................................................................................................ 39
General alert ................................................................................................................... 40
Additional troubleshooting............................................................................................... 41
17. Long-term Storage ...................................................................................................... 42
18. Disposal ........................................................................................................................ 43
19. Supplemental Information .......................................................................................... 44
20. Warranty ....................................................................................................................... 45
21. Meaning of Symbols .................................................................................................... 46
22. EMC/Technical Description ...................................................................................47-53
23. Technical Specifications ............................................................................................. 54
Table of Contents
Page 8
AUTHENTIC
medela spare parts
Quality & performance
you can trust
2x PersonalFit™ Breast Shields (24 mm) Applied Part
2x PersonalFit™ Breast Shields (27 mm) Applied Part
For other breast shield sizes, see Section 8
Replacement 87073 (24 mm) Replacement 87274 (27 mm)
4x Bottles
Item 6007135
4x Lids
Item 8107183
2x Bottle stands
Item 8107322
2x Membranes
Item 8207086 Replacement 68055
2x Connector caps
Item 8007282
2x Valves
Item 6 007151
2x Connector bodies
Item 101033253
2. Product Description
2x Breast shield assemblies
Replacement 68054
Your breast pump system includes
Medela reserves the right to substitute any component or accessory with a replacement of equivalent performance.
(1x) Power adaptor
Item 9007068 Replacement 68050
1x Ice pack
It em 8117010 Replacement 87092
1x Cooler bag
Item 3007304 Replacement 67068 (replacement includes ice pack & 4 bottles with lids)
1x Bag
Item 3007396 Replacement 68052
Page 9
Questions? Visit www.medela.com or call us at 1-800-435-8316.
Rhythm
1x Tubing
Item 8007278 Replacement 68053
Speed/Vacuum level
Rhythm selection
Decrease vacuum
Programmable timer
Start/Stop
Stimulation phase
Expression phase
One-touch letdown
Increase vacuum
Silence control
Status icon light
sonata
Display Icons
Controls
Air leak alert
Battery status
External power connected
Chimes “off”
Timer “on”
Bluetooth® connectivity
sonata
Tubing port
Power adaptor port
Power button
Handle
1x Breast pump
Short Tubing
(connects to breast shield assemblies)
Tubing Holder
Long Tubing
(connects to breast pump)
Page 10
10
3. Getting Started
Note
It is important for you to do the following before using your breast pump for the first time:
1. Plug the Sonata into an external power source. Charge the battery for 12 hours. (See Section 10, Powering Your Pump).
2. Separate all parts that are exposed to the breast or breast milk prior to cleaning.
3. Clean - see Section 5.
Use your thumb to separate the connector cap and body.
Pull down on valve to remove from connector body. Do not rotate.
Parts to clean
Separate all parts that are exposed to
the breast and breast milk prior to cleaning.
Page 11
Questions? Visit www.medela.com or call us at 1-800-435-8316.
11
When to Wash
Breast Shields
Breast Pump Kit
Bottles & Lids
Before 1st use
3 wash
3 sanitize
3 wash
3 sanitize
3 wash
3 sanitize
After each use
3 wash 3 wash 3 wash
Once per day
3 sanitize 3 sanitize 3 sanitize
4. Cleaning Overview
sonata
Washing and sanitizing are two different activities. They must be done separately to protect you, your baby and the performance of your breast pump.
Wash – To clean the surfaces of the parts by physically removing contamination.
Sanitize – To kill living organisms, such as bacteria or viruses, that may be present on the
surfaces of the parts.
Detailed cleaning instructions on pages 12-15:
• Wash – Section 5
• Sanitize – Section 5
• Breast Pump and Tubing Care – Section 6
When to Wash Tubing
Breast Pump
Carry Bag
As needed
3 See tubing care -
Section 6
3 wipe with clean,
damp cloth
3 wipe with clean,
damp cloth
Page 12
12
Supplies needed:
Mild dish soap or Medela Quick Clean™ Breastmilk Removal Soap
• Clean dish towel or soft brush
• Clean sink or bowl
• Drinking-quality water
• Tongs
• Clean pot for boiling water
Parts to wash or sanitize:
Breast shields
Breast milk bottles
Lids
Connector bodies
Valves
Membranes
Connector caps
5. Cleaning Instructions
1. Disassemble
Separate parts for washing.
• If using the dishwasher, parts may become discolored. This will not impact part function.
• Tubing should only be washed if dirty or milk is present. See Section 6 for more details.
Tip
T
Note
• Take care not to damage parts of the breast pump kit during cleaning.
• Store the dry breast pump kit in a clean bag/container until next use.
• Wash hands thoroughly.
• If you notice a white residue on your breast pump parts after boiling, your water may have a high mineral content. Remove residue by wiping parts with a clean cloth and allow to air dry.
• Distilled water is recommended when boiling parts to prevent substantial mineral build-up over time, which may compromise your parts.
• Separate and wash all parts that are exposed to breast milk immediately after use. This will help to remove breast milk residue and prevent growth of bacteria.
• Only use drinking-quality tap or bottled water for cleaning.
• When cleaning the valves do not use small objects such as brushes. Be careful not to puncture.
Caution
Page 13
Questions? Visit www.medela.com or call us at 1-800-435-8316.
13
2. Wash
Before first use and after each use
OR
quick clean
micro-steam bag
STEAM CLEAN YOUR
ACCESSORIES AFTER EVERY USE
steam vent / pour spout
thoroughly disinfects most breastpump accessories, bottles, nipples, paciers, and cups
steam cleaning kills 99.9% of most harmful bacteria and germs
works in approximately 3 minutes
20 uses per bag
This bag will hold:
-2 breastshields and valve membranes or
-2 5oz (150ml) breastmilk bottles and caps, or
-2 - 8oz (250ml) breastmilk bottles and caps
quick clean
micro-steam bag
quick clean
micro-steam bag
3. Sanitize
Before first use and once per day
OR
Step 1 – Fill pot with water to cover all
parts. Bring to a boil. Step 2 – Place parts in boiling water for
10 minutes. Step 3 – Allow parts to air dry. Store
dry parts in a cool place when not in use. Do not store wet or damp parts.
Step 1 – Use Medela's Quick Clean Micro-Steam™ bags. Sold separately (Follow instructions as provided on bag)
Step 2 – Place parts on a clean surface and/or towel. Allow parts to air dry. Store dry parts in a cool place when not in use. Do not store wet or damp parts.
Microwave
StovetopSink
Dishwasher
Step 1 – Rinse all separated parts that
touch the breast and breast milk in cool water to remove breast milk residue.
Step 2 – Soak in warm soapy water for 5 minutes, then wash and rinse.
Step 3 – Allow parts to air dry. Store dry parts in a cool place when not in use. Do not store wet or damp parts.
Step 1 – Wash all separated parts by using the top rack of your dishwasher.
Step 2 – Place parts on a clean surface and/or towel. Allow parts to air dry. Store dry parts in a cool place when not in use. Do not store wet or damp parts.
10
minutes
5
minutes
Page 14
14
sonata
1
6. Breast Pump & Tubing Care
Warning
Do not immerse the
breast pump in water; do not run water over the breast pump.
• When cleaning the breast pump display, only use a damp, soft cloth with mild dish soap. Other cleaners or abrasive cloths may degrade the display performance.
Wipe pump unit with a clean, damp cloth.
6A. Cleaning the breast pump
Supplies needed:
Mild dish soap
• Clean dish towel
• Clean sink or bowl
• Drinking-quality water
Parts needed:
Tubing
• Breast pump
sonata
Tip
T
• Tubing should only be washed if dirty or milk is present. Follow instructions in Section 6B.
• There is no need to clean tubing if condensation inside the tube is from previous washings or atmospheric conditions. Follow drying instructions in Section 6B.
Page 15
Questions? Visit www.medela.com or call us at 1-800-435-8316.
15
sonata
2
4 3
Caution
• Do not store wet or damp parts as mold may develop.
• If tubing becomes moldy, discontinue use and replace tubing. To find replacement parts visit www.ShopMedela.com.
• Do not use breast pump with wet tubing;
doing so will cause damage to your breast pump.
6B. Tubing care
To properly clean your tubing
Washing tubing
• Rinse tubing in cool water.
• Wash tubing in warm, soapy water.
• Rinse tubing with clear water.
• Shake out water droplets.
• Hang to air dry.
• Tubing does not need to be washed before first use.
• Inspect tubing after each pumping session.
Turn off breast pump and unplug from power source.
To remove tubing from the breast pump, grab the tube adaptor and pull directly from the tubing port. Do not wiggle or pull tubing at an angle.
Remove tubing from breast shield.
1
Page 16
16
Common pumping questions & answers
How often should you breast pump?
A breast pump is a replacement for when you are separated from your baby. It is important to pump when the baby would normally be breastfeeding. For example, a working mother may pump 2-3 times during an 8-hour work day.
How long should your pumping session last?
Pumping times can vary from mother to mother. Pumping sessions usually last from 15 to 30 minutes.
How should your breasts feel before and after pumping?
Before pumping, your breasts will have a firm, heavy feeling. After pumping, your breasts should feel soft with no firm areas. Firm areas could indicate that the breast is not completely drained.
7. Breastfeeding Information
Page 17
Questions? Visit www.medela.com or call us at 1-800-435-8316.
17
8. Breast Shield Sizing
Caution
• While some discomfort may be felt when first using a breast pump, using a breast pump should not cause pain. If you are unsure about breast shield sizing, please contact a health care professional or breastfeeding specialist who can help you get a proper fit.
If you are experiencing discomfort at the base of the nipple due to rubbing of your breast
tissue against the breast shield tunnel, use of a lubricant such as Tender CareTM Lanolin may be beneficial.
1. Medela’s PersonalFit™ sizing
Pumping should not hurt. For maximum comfort and pumping efficiency, Medela offers five breast shield sizes.
Visit MedelaBreastshields.com to view the full guide to help you determine your optimal size based on your nipple diameter.
2. Test your breast shield size
1. Start with the 24 mm that came with your pump, or the size determined by measuring.
2. Center nipple and gently hold breast shield against your breast.
3. Adjust for Maximum Comfort Vacuum™, to achieve optimum suction level.
4. Refer to images while pumping in expression.
3. Should you try a new size?
• Does your nipple rub sides of tunnel, to the point of causing discomfort?
• Do you see excessive areola being pulled into tunnel?
• Do you see any redness?
• Is your nipple or areola turning white?
• Do you feel unexpressed milk after pumping?
If you answered “YES” to any of these questions, consider trying a new size by following the measuring instructions above.
If you are still unsure if you selected the correct size, see a lactation consultant, breastfeeding specialist, or visit MedelaBreastshields.com for assistance in choosing the right size breast shield.
36
mm
27
mm
30
mm
24
mm
21
mm
Too Small
Nipple
Areola
Correct Fit
Nipple
Areola
Nipple
Areola
Too Large
Nipple
Areola
Nipple
Areola
Nipple
Areola
• Nipple rubs along side of tunnel
• Try a larger size
• Nipple is centered and moves freely
• Nipple and excessive areola are pulled into the tunnel
• Try a smaller size
Page 18
18
Tip
T
Tip
T
• Only use Authentic Medela Spare Parts. See Section 2 for details.
• Check breast pump kit components for wear or damage before use. Replace if necessary.
• Always inspect all parts prior to use for cleanliness. If dirty, see Section 4.
• To prevent damage to the breast pump all components must be completely dry before use.
Parts Needed:
Tubing
• Breast pump
Breast shields
• Breast milk bottles
• Lids
• Connector bodies
• Valves
• Membranes
• Connector caps
Using the two legs as guides, insert the connector body onto the valve.
Insert short tubing into the connector cap until snug.
9. Assembling Your Breast Pump
Wash hands thoroughly with soap and water before touching breast pump, breast pump kit and breasts. Avoid touching the inside of containers and lids.
4
7
1
Page 19
Questions? Visit www.medela.com or call us at 1-800-435-8316.
19
sonata
• Carefully insert the (dark yellow) membrane into the top of the connector body.
• The membrane must be secure for correct assembly.
Press the connector cap onto the connector body. It will snap closed.
Fasten the bottle onto the complete breast shield assembly.
Insert the long tubing end into the breast pump as far as it will go.
Push breast shield onto the connector body.
Accurately assembled system.
6 5
sonata
8
Snap
3 2
Page 20
20
Power AdaptorRechargeable Battery
10B. Battery Care
10A. There are Two Ways to Power Your Breast Pump for Your Pumping Session:
• Plug power cord into the power adaptor port on the back of the breast pump. Plug power cord into power source.
• Proceed to Section 11, Operating Your Breast Pump to begin your pumping session.
Your breast pump contains a Lithium ION rechargeable battery.
10. Powering Your Breast Pump
2
sonata
1
• Before first use, charge the breast pump battery for 12 hours with the power adaptor. The battery icon will appear fully charged on the display when charging is complete.
• Proceed to Section 11, Operating Your Breast Pump, to begin your pumping session.
• Before using on battery power for the first time, fully charge your breast pump for 12 hours.
• You can use your breast pump while charging the battery. The battery will charge as soon as the breast pump is connected to a power source. It will not charge while the pump is running.
• Keep your pump in a cool place.
• Keep your battery charged.
• Fully charge the battery before storing your pump for an extended period of time.
• Recharge the battery before it completely drains. This is better for the lifecycle of the battery.
• If your breast pump has been stored in a hot location, it may not run on battery power right away. To resume normal battery function, allow the breast pump to cool for one hour. During this time, you can use your breast pump when plugged into external wall power.
Tip
T
OR
Page 21
Questions? Visit www.medela.com or call us at 1-800-435-8316.
21
Battery Charge Status Codes
External power source connected
Battery charge status
Low battery, needs charging
Battery drained. See Troubleshooting (Section16).
Battery broken. See Troubleshooting (Section16) and Call Medela Customer Service.
(flashing)
(flashing)
bAtt
Page 22
22
10D. Traveling Outside of the U.S.
When travelling internationally, we recommend purchasing a Universal Power Plug Adaptor (not included). Please consult with the country you are visiting to find out what type of adaptor will work best. This breast pump's A/C adaptor is two-pronged and does not require a grounded outlet.This breast pump has a Lithium ION Battery which may be restricted from the country you are visiting. Please consult with the country you are visiting to find out if there are any restrictions that pertain to travelling with Lithium ION.
For information about traveling with your breast pump and expressed breast milk, visit www.TSA.gov.
10C. Power Adaptor Care
How to care for your adaptor:
Use only the power adaptor that comes with the breast pump.
• Make sure the voltage of the power adaptor is compatible with the power source.
Warning
Correct!
Correct!
Do not wrap cord of power adaptor around the plug body.
Do not unplug power adaptor by pulling on the cord.
10. Powering Your Breast Pump (cont.)
Never operate an electrical device if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water. If damage is found, immediately discontinue use of power adaptor and call Medela Customer Service at 1-800-435-8316.
Page 23
Questions? Visit www.medela.com or call us at 1-800-435-8316.
23
11. Operating Your Breast Pump
11A. Glossary of Terms
2-Phase Expression
®
Through research, Medela found that babies nurse in 2 Phases – Stimulation and Expression. This research is the basis for the technology in all Medela breast pumps.
Stimulation Phase Fast sucking/pumping rhythm to stimulate milk flow.
Expression Phase
Slower sucking/pumping rhythm for gentle and efficient milk removal after milk has started flowing.
Maximum Comfort Vacuum
Highest vacuum setting where pumping still feels comfortable. Different for every mother.
Pumping Rhythms
This breast pump includes two clinically-tested pumping rhythms. It offers you options to choose the most comfortable and effective for you.
Responsive Pump Technology
Responsive Pump Technology automatically adjusts to the body and environment and makes sessions more consistent and productive.
Letdown When milk begins to flow from the breast.
Rhythm
Speed/Vacuum level
Rhythm selection
Decrease vacuum
Programmable timer
Start/Stop
Stimulation phase
Expression phase
One-touch letdown
Increase vacuum
Silence control
Status icon light
sonata
Display Icons
Controls
Air leak alert
Battery status
External power connected
Chimes “off”
Timer “on”
Bluetooth® connectivity
Page 24
24
11. Operating Your Breast Pump (cont.)
Tip
T
Place the breast shields on your breasts so that your nipples are properly centered in the tunnels.
1
11C. Double Pumping
Hold the breast shield and connector onto your breast with your thumb and index finger. Support your breast with the palm of your hand.
Refer to Section 8 for breast shield sizing information.
11B. Single Pumping
1
Insert the unused tubing end into the tubing holder.
2
• Wash hands thoroughly with soap and water before touching breast pump, kit and breasts. Avoid touching the inside of containers or lids.
• Always inspect all parts prior to use for cleanliness.
Page 25
Questions? Visit www.medela.com or call us at 1-800-435-8316.
25
• Make sure tubing is not kinked or pinched while pumping.
• Do not hold the breast pump kit by the
bottle. This can lead to blockage of the milk ducts and engorgement.
• Do not try to express with vacuum that
is too high and uncomfortable (painful). The pain, along with potential breast and nipple trauma, may decrease milk output. The comfortable level is based on each individual.
• Contact your healthcare professional
or breastfeeding specialist if you can express only minimal or no milk or if expression is painful.
Caution
11D. Pumping Steps
Plug your assembled double-breast pump kit into the tubing port on the front of the breast pump.
1
Turn on the breast pump by pressing the power button on the back.
Start breast pump vacuum by pressing the start/stop button.
sonata
3
sonata
2
Page 26
26
sonata
sonata
When your pumping session is over, stop vacuum by pressing the start/ stop button.
sonata
5
6
To turn off your breast pump, press the power button on the back.
Your breast pump will begin in Stimulation Phase. Adjust the pumping level using the controls to find a comfortable level.
Stimulation Phase – a fast sucking/ pumping rhythm to stimulate milk flow.
After two minutes, your pump will automatically transition to Expression Phase. Adjust the pumping level using the controls to find your Maximum Comfort Vacuum
(see Section 12).
If your milk begins to flow before two minutes, press the let-down button to manually transition to Expression Phase.
Expression Phase - a slower sucking/pumping rhythm for gentle and efficient milk removal after milk has started flowing.
4A 4B
11. Operating Your Breast Pump (cont.)
Tip
The breast pump will automatically
turn off after five minutes of inactivity.
• Do not forget to record milk output in
your MyMedela app.
T
Page 27
Questions? Visit www.medela.com or call us at 1-800-435-8316.
27
sonata
Once you are pumping in the Expression Phase, increase the speed/vacuum until pumping feels slightly uncomfortable (not painful), then decrease slightly.
Tip
• Maximum Comfort Vacuum is the highest vacuum setting where pumping still feels comfortable. This is different for every mother.
• Stimulation should be at a comfortable vacuum level.
• Reassess your Maximum Comfort Vacuum throughout your pumping experience. It can change throughout each stage of lactation.
T
12. Find Your Maximum Comfort Vacuum
TM
Page 28
28
Breast Pump Connectivity
sonata
By tracking your sessions in our MyMedela® app, you can stay on-target with daily highlights and custom content. Having visibility to your pumping history gives you the opportunity to plan efficiently each day and track changes in your body like an increase or decrease in milk output.
By connecting your breast pump to your mobile device you can automatically transfer your pumping data (session length, phases & levels) and manually input the amount of milk expressed into your MyMedela pumping log.
13. Features
MyMedela
How do I connect (pair) my Sonata Breast Pump with my mobile device?
Step 1. Download and install MyMedela to your mobile device. Available for free for
iOS and Android.
Step 2. Open MyMedela on your mobile device. Follow set-up and pairing instructions. Step 3. When correctly paired, the will appear on your breast pump display.
Page 29
Questions? Visit www.medela.com or call us at 1-800-435-8316.
29
Tip
Tip
T
T
How do I remove my Sonata breast pump from my mobile device?
Step 1. Navigate through MyMedela to remove your Sonata Breast Pump.
• After initial set-up, the Bluetooth icon will automatically appear on your breast pump, when powered on, if your paired device is near and MyMedela is open.
• You can pair multiple mobile devices with Sonata.
• Only one mobile device at a time can interact with Sonata.
• Sonata has the ability to store your last 30 pump sessions in its memory. If the pump is not connected prior to session 31 your history will be overwritten with new session data. When connected, MyMedela will log an infinite number of historical sessions.
• Removing Sonata from your mobile device will not delete information stored in MyMedela.
• If you re-connect your breast pump and mobile device the last 30 records stored within your pump will transfer.
To learn more about MyMedela, including download instructions, please visit: www.mymedelaapp.com.
Page 30
30
sonata
Selecting a Rhythm
Two clinically tested pumping rhythms are included:
1. Signature - This rhythm is based on Medela's successful Symphony® Breast Pump used in hospitals.
2. Lifestyle - An alternate rhythm for when you want to change your pumping routine. It includes additional benefits of longer battery, and quieter pump operation when you need it.
Note
Your breast pump will default to Signature rhythm. To change to the alternate Lifestyle rhythm, push the pattern button.
13. Features (cont.)
Page 31
Questions? Visit www.medela.com or call us at 1-800-435-8316.
31
sonata
Tip
If you want to continue the same session, press start/stop once and vacuum will
continue in same level, phase and rhythm. The timer will count up from your previously entered pumping time.
T
Note
• Your breast pump will make a sound and the display will flash to let you know that your session has 1 minute remaining.
• Your breast pump will make a sound and flash to alert you that your session has ended.
The default timer on your breast pump will start at zero and count up. Follow the steps below if you want to count down from a set time.
To set your count-down timer:
1. Before you begin pumping, press the timer button. The timer icon will appear on the display and the buttons will be illuminated.
2. Use the buttons to adjust time in 1 minute increments (up to 30 minutes).
3. Press start to begin your pumping session.
4. Once you’ve reached your goal time, the vacuum will pause and the display will show your
total pumping duration.
Setting the Session Timer
Page 32
32
sonata
Your breast pump has the ability to briefly pause during your pumping session. This allows you to readjust yourself or your breast pump pieces, or attend to your surroundings.
To pause your breast pump:
1. Press the start/stop button. Vacuum will stop and time on the display will flash. Start/stop icon on display will flash.
2. When ready to resume your session, press start/stop button.
Note
• If pumping is not resumed within two minutes, the breast pump will end your session. To resume pumping, you will need to start a new session.
• If you need to pause for more than two minutes, we recommend restarting in Stimulation Phase.
Pausing Your Breast Pump
13. Features (cont.)
Page 33
Questions? Visit www.medela.com or call us at 1-800-435-8316.
33
sonata
Silencing Your Breast Pump Chimes
Your breast pump is equipped to provide sound notifications, such as button presses, during your pumping sessions.
To silence your breast pump chimes:
1. Push the Silence control button.
2. Your breast pump will remember your setting for future sessions. To reactivate the
sound feature, push the button again.
Note
When the breast pump chimes are turned off, the sound icon will appear on the display screen.
Page 34
34
Room
Temperature
Cooler with
Ice Pack
Refrigerator Freezer
Freshly Expressed Breast Milk
Storage Guidelines
(For Healthy Term Babies)
4–6 hours
at 66–78 °F
(19–26 °C)
24 hours
at 59 °F
(15 °C)
3–8 days
at 39 °F or lower
(4 °C)
6–12
months
0–-4 °F
(-18–-20 °C)
References: www.BreastmilkGuidelines.com
• Your Medela cooler bag requires one Medela contoured ice pack. Up to four 5 oz. bottles of breast milk will stay cool for up to 10 hours in a room temperature environment inside your pump bag. Up to four 5oz bottles of breast milk will stay cool for up to 8 hours in a room temperature environment, inside the cooler, stored outside of the pump bag.
• As soon as you are finished pumping, place the breast milk bottle or breast milk storage bag in the cooler bag with the ice pack.
• Transfer breast milk bottles or breast milk storage bags to the refrigerator or freezer once you are home.
14. Storing Breast Milk
Storing Breast Milk in a Medela Cooler Bag
Page 35
Questions? Visit www.medela.com or call us at 1-800-435-8316.
35
Caution
Preparing Breast Milk
• Do not thaw frozen breast milk in a microwave or in a pan of boiling water.
• Do not microwave breast milk. Microwaving can cause severe burns to baby’s mouth from
hot spots that develop in the milk during microwaving. (Microwaving can also change the composition of breast milk.)
• Thaw breast milk overnight in the refrigerator. Thawed breast milk is safe in the
refrigerator for 24hours. Do not refreeze thawed breast milk.
• Quickly thaw breast milk by holding the bottle or bag under warm running water.
• Place the sealed bottle or bag in a bowl of warm water for 20-minutes to bring it to
body temperature.
• If you are adding expressed breast milk to a container of already frozen breast milk,
make sure to add a lesser amount than the already frozen amount.
Feeding Breast Milk
It is recommended that breastfeeding is well established prior to bottle feeding your baby.
• Always inspect the bottle, nipple and other components immediately before and after each
use. If nipple appears cracked or torn, discontinue use immediately.
• To prevent possible choking hazard, test strength of nipple by pulling on bulb
portion of the nipple.
• Do not attempt to enlarge the nipple hole.
• Infants must not be bottle fed without adult supervision.
• Nipple should not be used as a pacifier.
For additional breast milk collection & storage information, please visit www.BreastmilkGuidelines.com.
15. Preparing & Feeding Breast Milk
Page 36
36
sonata
Note
There is no need to turn off your breast pump. The alert icon will disappear once your breast pump detects that the error has been fixed. In order to do this, the breast pump must be running.
16. Troubleshooting
Air Leak Alert
There is an air leak in your system and vacuum is not operating at optimal performance. During this time, the breast pump will continue to run, but an alert icon will appear on the display and a chime will sound every 30 seconds until the correction has been made.
• Verify that the kit connections are properly assembled.
• Make sure the breast shields are secure to your breasts.
• Check that the tubing connection between the breast pump and kit is secure.
• If you still have issues after following these steps, please call
Medela Customer Service at 1-800-435-8316 for further assistance.
Page 37
Questions? Visit www.medela.com or call us at 1-800-435-8316.
37
sonata
16. Troubleshooting (cont.)
Battery Drained
If bAtt appears on the screen with an empty battery cell, the battery is too drained to start a vacuum.
To recover:
• Plug your breast pump into an external power source.
• Fully charge the battery for 12 hours.
Page 38
38
sonata
Battery Broken
If the battery icon is empty and flashing, your battery is broken. The battery cannot be serviced or replaced. Please call Customer Service at 1-800-435-8316 for replacement options.
To continue pumping:
Plug your breast pump into an external power source to use.
Note
If your battery is broken, your breast pump will no longer be able to charge or run on battery power.
16. Troubleshooting (cont.)
Page 39
Questions? Visit www.medela.com or call us at 1-800-435-8316.
39
sonata
Bluetooth® Broken
If ERR flashes on your breast pump for three seconds it means the Bluetooth connection is broken and the breast pump or mobile device are unable to complete data transmission. This will not impact the ability to pump with Sonata, only the ability to transfer data to the MyMedela® app. Please call Customer Service at 1-800-435-8316 for assistance.
Page 40
40
sonata
General Alert
Pump is not operable in this condition. Please contact Medela Customer Service at 1-800-435-8316 for assistance.
16. Troubleshooting (cont.)
Page 41
Questions? Visit www.medela.com or call us at 1-800-435-8316.
41
If your breast pump is not responding as expected
• To reset your device, plug Sonata into external power and press and hold the power button (on the back) for up to 10 seconds or until the pump display flashes. If there is no change contact Medela Customer Service at 1-800-435-8316.
If breast pump does not power on
• Check that the breast pump is plugged into a working electric outlet.
• If there is no power, display does not light up, and there is no error alert appearing on screen, contact Medela Customer Service at 1-800-435-8316.
If breast pump does not run after pushing the start/ stop button
• Make sure you've powered on the breast pump.
• Make sure breast pump is plugged in and/ or charged.
• For more information, refer to General Alert (Section 16).
If breast pump exterior gets wet
• Unplug from power source and power off and discontinue use.
• Dry off the exterior of the breast pump.
If breast pump is submerged in water
• Unplug from power source, power off and discontinue use.
• Contact Medela Customer Service at 1-800-435-8316.
If there is low or no suction
• Check that all of the breast pump and kit connections are secure.
• Disassemble the double-pumping kit and check that all pieces are clean and not damaged.
• Assemble the pump set as directed in Assembling your Breast Pump (Section 9).
• While pumping, make sure the breast shields form a complete seal around the breast.
• When single pumping, check that the unused end is correctly stored in the tubing holder.
• If suction does not improve after following these steps, contact Medela Customer Service at 1-800-435-8316.
If there is a breast milk overflow
Turn off the breast pump and unplug the power cord from the electrical outlet.
• Use a damp (not wet) cloth to wipe the breast pump.
• If there is milk in the tubing: Remove tubing from the pump unit as well as breast shield assembly. Clean as directed in Section 6.
• When all components are dry, reassemble.
No “let-down" or breast milk expression
• Ensure your breast pump kit is assembled correctly and suction is available.
• Relax and take a 10-15 minute break if let-down is not achieved after two consecutive pumping sessions.
• Consult with a health care or lactation professional if expression does not occur.
Data cannot be sent • The Bluetooth® function on the mobile device is turned off. Turn on the
Bluetooth function and try sending data again.
• Pairing between the two devices has not been completed. Perform pairing instructions again. Refer to “How do I pair my Sonata Breast Pump with my mobile device”? in Section 13.
• The mobile device is too far away from the breast pump. Move the device and breast pump closer together.
• The application on the mobile device is not ready. Check the application.
• If the Err symbol appears after checking the application, contact Customer Service at 1-800-435-8316.
If you have not resolved the problem with your breast pump or you have further questions, please contact Medela Customer Service at 1-800-435-8316 or fill out the Contact Us form at www.medelabreastfeedingus.com.
Page 42
42
17. Long-term Storage
Your breast pump contains a rechargeable battery. To preserve your breast pump, when not in use for an extended amount of time, it will automatically transition into storage mode.
To remove from storage mode, plug your breast pump into an external power source and fully charge the battery for up to 12 hours.
sonata
Before storing your breast pump for an
extended amount of time remember to:
Clean breast pump kit
(Section 5)
Clean breast pump
(Section 6)
Make sure breast pump battery is fully-charged
3
sonata
Page 43
Questions? Visit www.medela.com or call us at 1-800-435-8316.
43
Disposal
The symbol on the product or its packaging indicates that this product must not be disposed of with your other household waste. Instead, it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment. The separate collection and recycling of your waste equipment of your waste equipment at the time of disposal will help conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment. for more information about where you can drop off your waste for recycling, please contact your local authority, or where you purchased your product.
Recycle your breast pump
Through the Medela Recycles program, Medela LLC. provides customers the option to recycle their breast pump after they have finished their breast milk feeding journey. For information on how to recycle your pump, visit www.MedelaRecycles.com.
This program is currently only available to U.S. residents. We only accept the return of electric breast pumps. All other breast pump kit components (connectors, breast shields, tubing, bottles, valves, membranes, and cooling elements) can be disposed of in your waste management recycling container at your home. Cooler bags and other soft-good items are not recyclable.
18. Disposal
Page 44
44
Supplemental Information
Following are some common breastfeeding-related conditions. If you experience any of these symptoms, contact a healthcare professional or lactation professional.
Symptom Potential Causes
Engorgement Breasts are hard and uncomfortable, possibly
with reddened areas.
Milk not draining properly from breast. Milk can drain back into the tissue causing areas to swell and become tender.
Blocked/Plugged Ducts
Area of breast will look reddened and may be tender to touch.
Milk not being drained from a specific duct. The area becomes "clogged" up and milk is then prevented from flowing.
Mastitis Often follows engorgement but can just occur
suddenly. The initial symptoms are similar to flu – fatigue, headache and muscle aches, fever and localized breast tenderness. It can be in one or both breasts and needs immediate treatment.
A bacterial infection in the breast tissue which is often accompanied by cracked nipples.
A Healthcare Professional or Lactation Professional can give you guidance in breastfeeding your baby and using a breast pump. Their guidance is valuable in terms of the lifetime of health benefits for both you and your baby. A listing of lactation professionals are available at www.medelalocator.com or calling 1-800-TELL-YOU (1-800-835-5968).
19. Supplemental Information
Page 45
Questions? Visit www.medela.com or call us at 1-800-435-8316.
45
20. Warranty
This product is warranted by Medela LLC. to the original retail purchaser to be free from defects in material and workmanship for the period of 1 year for pump mechanism (90 days for other parts and accessories) from the date of purchase. Warranty can only be claimed in the country of purchase. In the event of defect, at Medela's option, Medela will replace this product, without charge for such replacement parts. Purchaser shall bear all expense for returning this product to Medela. This warranty does not apply to any product used commercially or which has been subjected to misuse, abuse or alteration.
ANY AND ALL IMPLIED WARRANTIES, INCLUDING THE WARRANTY OF MERCHANTABILITY, ARE LIMITED TO A DURATION OF 3 YEARS FROM DATE OF PURCHASE. SOME AREAS DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE LIMITATIONS MAY NOT APPLY TO YOU. THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS AND YOU MAY ALSO HAVE OTHER RIGHTS WHICH VARY FROM STATE TO STATE OR IN YOUR COUNTRY.
Before you make a claim under this warranty, it may save time and expense to call Medela Customer Service (toll free) at 1-800-435-8316. You may also call this number for additional information concerning this warranty. If you wish to make a claim under this warranty, you must return this product to Medela with a return authorization number received from Medela Customer Service, prepaid, together with your dated bill of sale or other proof of purchase and a brief statement of the problem to the following address:
Medela LLC – Returns, Door 4501 1101 Corporate Dr. McHenry, IL 60050 ATTENTION: RETURNS
Contact Customer Service for a return authorization number. Returns are not accepted without an authorization number.
We recommend recording your serial number here when you first open your pump for easy reference when contacting Medela Customer Service.
Your serial number can be found on the sticker on the bottom of the breast pump.
SERIAL NUMBER:
Page 46
46
21. Meaning of Symbols
On/off
Session duration
Refer to Operating Instructions for important safety information, failure to follow instructions could place operator at risk. (when shown on the product this symbol is blue)
This symbol indicates the manufacturer.
This symbol indicates do not dispose the device together with unsorted municipal waste (in accordance with local regulations).
This symbol indicates compliance with international requirements for protection from electric shock. (Type BF applied parts.)
This symbol indicates the protection against entrance of solid foreign objects and against harmful effects due to the entrance of water.
This symbol indicates the date of manufacture (four digits for the year and two digits for the month).
IP21
Glossary of Symbols
This symbol indicates Class II.
This symbol indicates that the power adaptor is for indoor use only.
Classified by UL according to the standards ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012), CAN/CSA-C22.2 No. 60601-1 (2014), IEC 60601-1-6:2010 + AMD1(2013), IEC 62366:2007 + AMD1(2014) and IEC 60601-1-11:2015
This symbol indicates alternating current.
This symbol indicates direct current.
Symbol showing that the manufacturer has taken all necessary measures to ensure that the product complies with the applicable safety legislation for distribution within the European Community. This symbol only applies to the power adaptor.
Friwo Gerätebau GmbH Trademark.
Warning sign (when shown on product this symbol is yellow)
UL Recognized Component.
Keep Dry.
This device contains Bluetooth® wireless technology. “The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc., and any use of such marks by Medela LLC. is under license. Other trademarks and trade names are those of their respective owners”.
Page 47
Questions? Visit www.medela.com or call us at 1-800-435-8316.
47
22. EMC / Technical Description
The breast pump needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the instructions for use. Portable and mobile RF communications can affect the breast pump.
Note
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Page 48
48
22. EMC / Technical Description (cont.)
Guidance and manufacturer’s declaration – electromagnetic emissions
This breast pump is intended for use in the electromagnetic environment specified below. The customer or the user of the breast pump should assure that it is used in such an environment.
Emission tests Compliance Electromagnetic environment – guidance
RF Emissions CIS PR 11
Group 1
The breast pump uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CIS PR 11
Class B
The breast pump is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3
Pst < 1.0
Warning – This breast pump should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, this breast pump should be observed to verify normal operation in the configuration in which it will be used.
Warning
Page 49
Questions? Visit www.medela.com or call us at 1-800-435-8316.
49
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge ( ESD ) IEC 61000-4-2
+/- 2kV, +/- 4kV, +/- 6 kV, +/- 8kV contact discharge
+/- 2kV, +/-4kV, +/- 6 kV, +/8kV, +/- 15 air discharge
+/- 2kV, +/-4kV, +/- 6 kV, +/- 8kV contact discharge
+/- 2kV, +/-4kV, +/- 6 kV, +/8kV, +/­15 air discharge
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4
± 2 kV 100 kHz repetition frequency
± 2 kV 100 kHz repetition frequency
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
± 0,5 kV , ± 1 kV Line-to-line
± 0,5 kV , ± 1 kV Line-to-line
Mains power quality should be that of a typical commercial or hospital environment.
This breast pump is intended for use in the electromagnetic environment specified below. The customer or the user of the breast pump should assure that it is used in such an environment. The Sonata Breast Pump has no essential performance but was tested for immunity to electromagnetic disturbances and passed using the following criteria:
1. No visible change in the operation of the breast pump.
2. The breast pump changes settings but returns automatically to previous settings.
3. The breast pump changes settings but can return to previous settings by intervention of the user.
4. The breast pump changes settings but can return to previous settings by executing the INITIATION SEQUENCE.
INITIATION SEQUENCE: (The following sequence can be used to reset the device in the event of an EM Disturbance)
1. Unplug the power cord from the back of the breast pump.
2. Unplug the power adaptor from the wall outlet.
3. Attach a breast pump kit to the pump port (if none is attached).
4. Press and hold the power button for at least 30 seconds. (Note: The breast pump may cycle on/off several times. After 30 seconds the breast pump should remain off.)
5. Plug the power adaptor back into the wall outlet.
6. Plug the power cord into the back of the breast pump.
7. Press the on/off switch to power on the breast pump.
8. Press the ‘PLAY’ button to begin pumping.
Page 50
50
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
0 % UT ; 0,5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0 % UT ; 0,5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
Mains power quality should be that of a typical commercial or hospital environment. If the user of the breast pump requires continued operation during power mains interruptions, it is recommended that the breast pump be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
30 A/m, 50 or 60 Hz 30 A/m, 50/60 Hz
It may be necessary to position the breast pump further from sources of power frequency magnetic fields or to install magnetic shielding. The power frequency magnetic field should be measured in the intended installation location to assure that it is sufficiently low.
NOTE UT is the a.c. mains voltage prior to application of the test level.
22. EMC / Technical Description (cont.)
Page 51
Questions? Visit www.medela.com or call us at 1-800-435-8316.
51
Guidance and manufacturer’s declaration – electromagnetic immunity
This breast pump is intended for use in the electromagnetic environment specified below. The customer or the user of the breast pump should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of the breast pump, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommend separation distance
d = 1,2P d = 1,2√P 80 MHz - 800 MHz d = 2,3√P 800 MHz - 2.7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol.
10 V/m (minimum)
6V3 V 0,15 MHz – 80 MHz 6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80% AM at 1 kHz
10 V/m (minimum) 80 MHz – 2,7 GHz 80 % AM at 1 kHz
Radiated RF IEC 61000-4-3
Conducted RF IEC 61000-4-6
Page 52
52
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the breast pump is used exceeds the applicable RF compliance level above, the breast pump should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the breast pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies. Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
22. EMC / Technical Description (cont.)
Page 53
Questions? Visit www.medela.com or call us at 1-800-435-8316.
53
Recommended separation distances between portable and mobile RF communications equipment and the breast pump
This electric breast pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the electric breast pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the electric breast pump as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800
MHz
d = 1.2√P
800 MHz to 2,7 GHz
d = 2.3√P
0,01 0,06 0,04 0,07
0,1 0,18 0,11 0,22
1 0,58 0,35 0,22
10 1,84 1,11 2,21
100 5,83 3,50 7,0 0
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the breast pump including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Page 54
54
Vacuum Range
-50....-250 m mHg
54 … 120 cpm
Operation Temperature
Transport / Storage Temperature
Battery Capacity and Type
7.4 V, 2200 mAh Li-Ion
Power In / Power Out
Operation Humidity
Size 160 x 160 x 120
Weight
1.1 kg
Bluetooth RF Output Power 0dBm (class 3) Frequency band 2379–2496 MHz Frequency-hopping spread spectrum per Bluetooth® 4.0 Core Specification.
Transport / Storage Humidity
Ambient Pressure kPa
In Out
100-240V~ 50/60 Hz
12 V
1.5 A
°C
+40
+5
°C
+70
-25
93
15
93
15
106
83
23. Technical Specifications
Technical Specifications
Page 55
Questions? Visit www.medela.com or call us at 1-800-435-8316.
55
Page 56
Medela wordmark and logo, Symphony, 2-Phase Expression, MyMedela and Easy Expression are registered in the U.S. Patent and Trademark Office.
Sonata, Responsive Pump Technology, Quick Clean, Micro-Steam, Maximum Comfort Vacuum, and PersonalFit are trademarks of Medela.
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Medela is under license.
iOS is a trademark of Cisco in the U.S. and other countries. Medela is not affiliated with Cisco. Android is a trademark of Google Inc. Medela is not affiliated with Google.
1548311 C 0217 © 2017 Medela LLC 1908462
Manufactured and Assembled by Medela LLC 1101 Corporate Drive McHenry, IL 60050, USA Ph: (800) 435-8316 / (815) 363-1166 Email: customer.service@medela.com
www.medela.com
Loading...