IBM is a trademark of the International Business Machine
Corporation.
MS-DOS and Windows are trademarks of the Microsoft
Corporation.
Other product names used in this document may also be
trademarks and are only used for identification purpose.
Before using the instrument for the first time in a hazardous location:
Observe all the safety instructions in these operating instructions.
1.2Essential safety notes for the MLA 900
b
Observe the following safety notes when assembling and operating the instrument,
during maintenance and repair, and when exchanging parts – for your own protection.
The most important safety rules are:
VORSICHT: Risk of wrong measurements
b
Before using the MLA 900: carry out a performance test (→S. 21, §4.3) –
always to be done outside of the hazardous locations.
WARNUNG: Hazard when potential equalisation is missing:
b
Before the measuring probe is lowered into the liquid container: Safely
connect the ground clamp of the display unit with the liquid container
(potential equalisation).
WARNUNG: Hazard when enclosure is open
b
Never open the instrument enclosure as long as it is in a hazardous
location.
WARNUNG: Hazard by unsafe instrument
b
If there is any reason to assume that the instrument is no longer safe: shut
down the instrument and protected it against unauthorised use.
The MLA 900 determines the electrical conductivity and the temperature of petroleum
products such as occur, for example, in aviation fuels and in other petroleum products.
During the filling process, pumping or filtering, these liquids may become electrostatically
charged. If, in addition, there is an inflammable gas mixture present in the ambient
atmosphere, there is a danger that the gas mixture will be ignited by a discharge spark,
which means that it may explode. To evaluate this risk, the electrical conductivity of the
liquid is measured (see German standard DIN 51 412-T02-79, »Determining electrical
conductivity, field procedure«).
The conductivity is stated in the physical unit »pS/m« (picosiemens per meter). This unit
corresponds to »c.u.« (conductivity unit), which is still commonly used in the petroleum
industry:
Ambient conditions
The MLA 900 is designed as a measuring instrument for outside use.
b
Use the MLA 900 at an ambient temperature of –20 °C … +60 °C (–4 … +140 °F).
Outside this temperature range it can no longer be guaranteed that the instrument will
function correctly.
pS
1
1c.u.=
m
● At temperatures above +60 °C (140 °F) there is a danger that caustic liquid
from the battery may escape and damage the electronics. Apart from this,
the LC displays may be damaged (irreversible blackening).
● At temperatures of below –20 °C (–4 °F), the LC displays in the display unit
may freeze, resulting in the LC display enclosure bursting.
● Low temperatures affect the flexibility of the probe cable.
b
Avoid bending the probe cables in low temperatures.
Otherwise the probe cable could brake due to the sheath becoming brittle.
2.1.2Application restrictions
The MLA 900 may only be used to measure the conductivity and temperature of petroleum
products, such as petrol, kerosene and oils/greases. Other liquids such as acids or
solvents containing chlorinated hydrocarbons, for example, can damage the measuring
probe.
● German standard DIN 51 412-T02-79 (»Determining electrical conductivity, field
procedure«)
● ASTM 02624 »Standard Test Methods for Electrical Conductivity of Aviation and
Distaillate Fuels«
The technical design of the MLA 900 corresponds to:
● European standard EN 60079-0:2006
● European standard EN 60079-11:2007
● European standard EN 60079-26:2007
● European Council Directrive 94/9/EC (ATEX guideline)
● German »Ordinance on electrical appliances in hazardous locations« (»Verordnung
über elektrische Anlagen in explosionsgefährdeten Räumen«, ElexV) of 13 December
1996.
2.2Official approval
2.2.1Definitions of terms
● An explosive mixture is a mixture of gases, vapours, mists or dusts in which a reaction
continues independently after ignition.
● An explosive atmosphere contains explosive mixtures of gases, vapours, mists or dusts
with air, including usual admixtures (for example, humidity) under atmospheric
conditions. Atmospheric conditions are here defined as total pressures of 0.8 to 1.1 bar
and gas mixture temperatures of from –20 °C to +60 °C.
A dangerous explosive atmosphere occurs when in the event of an ignition injury may
be caused to persons as a direct or indirect result of the explosion.
● Hazardous locations are those areas in which a dangerous explosive atmosphere may
occur due to local and operational conditions.
Hazardous locations are divided into zones on the basis the probability (in terms of time
and place) of the presence of dangerous explosive atmospheres:
– Zone 0 includes areas in which a dangerous explosive atmosphere is present
»constantly or for longer periods«; for example, in tank facilities.
– Zone 1 covers areas in which a dangerous explosive atmosphere »occasionally«
occurs; for example, in the vicinity of refuelling nozzles.
The MLA 900 is approved for use in hazardous locations in which inflammable gases,
vapours or mists may occur (EEx ia IIB T6). The MLA 900 measuring probe is approved for
use in Zone 0, the display unit for use in Zone 1.
Prescribed operating conditions
The following applies for the use of the MLA 900 conductivity meter in the area of
application as defined in the »Ordinance on electrical appliances in hazardous locations
(ElexV)«:
WARNUNG: Hazards in explosions-hazardous locations
● The MLA 900 is intended exclusively for mobile use at different locations.
● The measuring probe of the MLA 900 is approved for use in containers for
inflammable liquids (Hazardous Location Zone 0).
● The display unit of the MLA 900 may only be used in Hazardous Location
Zone 1.
● The measuring probe may only be used in such liquids against which the
→
materials of which it is made offer adequate chemical resistance (
§2.1.2).
● Damaged measuring probes may not be used in Zone 0.
● Before the measuring probe is taken to Zone 0, all plug and threaded cable
connections must be checked.
● The measuring probe may only be used with the appropriate display unit.
● Before the measuring probe is lowered into the liquid container (tank), the
ground clamp of the display unit must be connected to the container to
equalize the electric potentials.
● Tight bending may damage the cable (cable breaks).
● Damaged cables must not be used in hazardous locations.
b
Never buckle the probe cable and do not bend the cables too far.
b
At temperatures below 0 °C (32 °F), treat the probe cable with particular
care – because the cable sheath may become brittle at these temperatures.
Avoid tight bending (risk of breakage).
Bild 1MLA 900 supply schedule
ground clamp
● Keep the transport box even if you do not wish to use it when operating the
● List of spare parts
display unit
probe cable
2m
MLA 900. It may be useful as a shipping packaging.
The MLA 900 consists of four instrument components (measuring probe, display unit,
ground clamp, probe cables). – Note: These components will only conform to the technical
safety regulations for the petroleum industry when used as an assembled unit.
Measuring probe and display unit are a matching pair
Display unit and measuring probe have been individually adapted in the factory. If a
different probe is connected, the measuring accuracy cannot be guaranteed.
b
Only connect a measuring probe to a display unit with the identical serial number on it.
Probe cables
b
Connect the display unit to the measuring probe with one of the probe cables supplied.
(Choose the cable length which is the best for the application.)
You can also connect the probe cables to each other to have a 12 m length of
cable. Greater cable lengths can be obtained by using further cable sections
→
(order designations
The maximum allowable total cable length is 24 m.
WARNUNG: Risk by wrong equipment
b
Use only use cables of the type supplied to connect the display unit and the
measuring probe.
It is not permitted to operate the instrument with other cables in hazardous
locations.
Ground clamp
b
Make sure that the ground clamp is firmly attached to the display unit (threaded
connections, cable lugs).
Always use the carrying handle to carry the instrument.
⊗ Do not use the cover flap or the cable brackets to carry the instrument.
Handling during measurement
b
During measurements, hold the display unit safely in your hand.
If you put the display unit down while measuring: make sure that it is standing
securely – because it might be pulled down by the weight of the measuring
probe, or even fall into the liquid container.
WARNUNG: Risks by wrong use
b
In hazardous locations the display unit may only be operated within
»Zone 1«.
b
Before the measuring probe is introduced into the liquid container (tank):
connect the ground clamp of the display unit to the container.
4.1.2Switching on and off
b
Switching on: Open the cover flap of the display unit.
➝ The MLA 900 automatically starts up when the light sensor is exposed to light.
b
Switching off: Close the cover flap.
➝ When the flap is closed, the MLA 900 automatically shuts off.
4.1.3Measuring value displays
left-hand display:right-hand display:
temperature of the measuring probe/
the liquid
Display range:
b
Watch out for fault and limit value indications (→S. 19, §4.2).
-199
The MLA 900 is approved for measurement in hazardous locations only in a
temperature range of
the MLA 900 fulfils the stated specifications.
Always make a performance test before operating the MLA 900.
b
Perform this performance test outside hazardous locations.
Check condition of instrument
1 Make sure that
– the cylinder of the measuring probe is tightly screwed on
– the measuring probe is sufficiently clean and dry
– display unit and measuring probe are correctly connected to each other (inspect
cables and plug connectors).
Check basic condition
2 Open the cover flap of the display unit in order to switch on the MLA 900. Let the
measuring probe hang free in the air.
➝ The conductivity value measured should then read
b
If a value above 2pS/m is displayed: carefully clean the measuring probe (→S. 27,
§5.1).
b
If a value below
Check measuring function
3 Hold the surface of the measuring probe with the company logo close to the red disc on
the display unit.
➝ The MLA 900 should display
b
If the value is not between
-2
pS/m is displayed: check the battery (→S. 29, §5.2).
1000
990
pS/m (±10 pS/m).
and
1010
-2…2
pS/m: check the battery (→S. 29, §5.2).
pS/m.
A re-calibration (a new adaptation of measuring probe and electronics) is
usually not required. If it is however necessary, this work must be made at the
manufacturer's factory, for safety reasons. Please observe the transport
Hold the display unit in your hand or use the transport case for holding the display unit (do
not put the display instrument down!).
1 Connect the ground clamp: Attach the ground clamp to a metallic, unpainted, rust-free
and grease-free point on the liquid container. Make sure that there is a good metallic
connection between the liquid container and the display unit.
WARNUNG: Explosion risk
b
Do not lower the measuring probe into the liquid container before the
ground clamp is safely connected to the container.
2 Immerse the measuring probe: Take the measuring probe and the probe cables out of
the bracket on the display unit and lower the measuring probe, hanging from the cable,
carefully into the liquid.
3 Measure: Open the cover flap of the display unit and read the measuring values. Note:
So that the measuring probe won’t get lost
in the liquid container.
Otherwise wrong measurements and/or
faults could occur (
→
S. 19, §4.2).
Contamination results in wrong measuring
→
results (cleaning
S. 27, §5.1).
Otherwise safety is not guaranteed.
b
Make sure that the measuring probe is completely filled with liquid. Air
bubbles would distort the measuring result.
b
If possible, measure the conductivity just when the desired immersion
depth is reached, or move the measuring probe constantly while
1
measuring.
b
Wait for the final temperature reading until the temperature display
remains approximately constant.
b
Please consider: there can be different temperatures at different
2
depths.
1
If the measuring probe is resting in the liquid, the conductivity measuring value will slowly change, due to
unavoidable electrochemical effects (ionic migration, polarisation, surface effects).
2
This takes about 30 seconds after the immersion; you can speed-up this by constantly moving the measuring
probe.
Finishing the measurement
1 Switch off the MLA 900: Close the cover flap of the display unit.
2 Remove the measuring probe: Pull in the measuring probe carefully out of the liquid
container (recommendation: wipe the probe cable when pulling out and dry with a soft
cloth). Stow the cables and measuring probe in the brackets of the display unit.
3 Stow away the ground clamp: Detach the ground clamp and attach it to the underside
If a fault or damage to an instrument component is detected, the MLA 900 must
not be taken into a hazardous location.
If the MLA 900 is faulty or damaged:
b
Mark the instrument clearly as faulty (for example, using a sticker).
b
Make sure that the MLA 900 is no longer used in hazardous locations.
WARNUNG: Risk caused by damaged measuring probe
If the cylinder of the measuring probe is deformed (for example, partly dented),
the conductivity measuring values will be distorted. This fault may not be detected
→
during a performance test (
b
Check the mechanical condition of the measuring probe.
b
If the MLA 900 is faulty of damaged: send both the display unit and the measuring
probe to the manufacturer's factory for repair (detailed information
5316999 permanent magnetdisc on display unit; used for performance test
4039227 sealseal for display unit enclosure
4038445 transport case
Display unit and measuring probe have been individually adapted in the
factory. If you connect a different probe, the measuring accuracy cannot be
guaranteed.
b
Only connect the measuring probe to the display unit with the identical
serial number on it.
If only the measuring probe or the display unit has become unusable or has
been lost, you can send the remaining part to the manufacturer's factory and
have this made back into a complete MLA 900.
ground clamp (with cable), 2 m cable,
10 m cable; without transport case
Please observe these instructions if the MLA 900 is to be taken out of service for more than
3 months:
b
Remove the battery (→S. 29, §5.2).
b
Keep all the components stored in a dry place at room temperature (15 … 20 °C/
59 … 68 °F).
b
Observe the admissible storage and transport temperature (–20 °C … +60 °C).
WARNUNG: Damage risks caused by wrong storage
● At lower temperatures the LC displays in the display unit may freeze,
resulting in the enclosure of the LC display bursting.
● Lower temperatures affect the flexibility of the probe cables. At low
temperatures avoid bending the probe cables too far as the cable sheath
may become brittle and break.
● At higher temperatures there is a risk that caustic fluid may escape from the
battery, damaging the electronics. In addition, the LC displays may be
damaged (irreversible blackening).
6.2Short-distance transports
b
Wind-up both the probe cable and the ground cable onto the cable brackets of the
display unit.
b
Attach the measuring probe and the ground clamp in the supports on the display unit –
do not detach the cable connections.
b
Use the hand-grip to carry the instrument.
6.3Correct shipping
If the MLA 900 is to be transported over long distances:
b
Secure instrument: Wind the probe cable and the cable with the ground clamp carefully
onto the cable bracket of the display unit. Attach the measuring probe and the ground
clamp in the brackets of the display unit.
b
Protect the display unit: Protect the display unit against condensation, humidity and
splashing
b
Packing: Stow the MLA 900 in transport case supplied.
b
Observe the admissible storage and transport temperature (→§6.1).
b
If the instrument is to be sent for repair: observe the notes in §6.4 (→S. 33).
Always send both the measuring probe and the display unit together for repair
(explanation
b
Please attach the following notes:
– A detailed, clear description of the problem (single words are fine, but merely stating
that »the instrument does not work« is of little help).
– A short description of the operating conditions.
– The name of the our representative who is informed about the problem or with whom
you have arranged transport to the workshop.
– The contact person in your company who can answer any questions that may arise.
This will help to bring your instrument quickly back to perfect operation. Please add the
information even if your matter has already been discussed with our customer service
or a representative.
→
S. 13, §3.2).
If only the measuring probe or the display unit has become unusable or
been lost, you can send the remaining part to the manufacturer's factory
and have these made back into a complete MLA 900.