Masterflex 77240-00, 77240-10 Instruction Manual

OPERATING MANUAL:
Compact Pumps
pages 1-8
MANUEL DE L’UTILISATEUR:
Entrainements de Pompes Compactes
pages F1-F8
BEDIENUNGSANLEITUNG:
seiten D1-D8
MANUAL DE OPERACIÓN:
Bombas Compactas
páginas E1-E8
MANUALE DI ISTRUZIONI:
Pompe Compatte
pagine T1-T8
77240-00 77240-10
Model No. Nº de modèle Modellnummern Modelo No. Modello No.
Cole-Parmer Instrument Co.
625 E. Bunker Court, Vernon Hills, Illinois U.S.A. 60061-1844 • 800-MASTERFLEX (627-8373) • (847) 247-2929 (fax) www.masterflex.com • e-mail: techinfo@coleparmer.com
Barnant Company
28W092 Commercial Ave., Barrington, Illinois U.S.A. 60010-2392 • (847) 381-7050 • (847) 381-7053 (fax) www.barnant.com • e-mail: barnant@barnant.com
A-1299-5092
Edition 01
TABLE OF CONTENTS
Title Page
SAFETY PRECAUTIONS........................................................2
GENERAL DESCRIPTION......................................................3
SETUP AND DRIVE OPERATION.......................................4-6
MAINTENANCE ......................................................................6
SPECIFICATIONS...................................................................7
WARRANTY ............................................................................7
PRODUCT RETURN...............................................................7
TECHNICAL ASSISTANCE.....................................................7
SAFETY PRECAUTIONS
WARNING: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures
to protect operator and equipment.
CAUTION: Tubing for use with the MASTERFLEX®L/S®compact pump are Masterflex®two-stop tube sets.
See Appendix A for specifics. Use of tubing other than that specified will result in poor pumping performance and/or pump system damage and voiding of applicable warranty.
WARNING: PRODUCT USE LIMITATION
These products are not designed for, nor intended for use in patient-connected applications, including, but not limited to, medical and dental use and, accordingly, have not been submitted for FDA approval.
2
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
3
GENERAL DESCRIPTION
The MASTERFLEX®L/S®Compact Pump System,
Figure 1,
is enclosed in
an epoxy-painted metal housing. The Pump System can accommodate
one or two tubes simultaneously (depending on model) at controlled speeds as low as 20 rpm. Tubing is held in place by stops bonded to the tubing and slots in the Tubing Retainer brackets. For a list of tube sets for use with the Pump System, refer to Appendix A.
The latching Occlusion Bed allows quick loading or changing of tubing. The Pump Head Rotor contains three rollers for optimum flow and positive flow checking. All units operate from an External DC Power Supply and are supplied with a Universal Power Supply (see Input Specification for operating voltage/frequency). This power supply provides a DC output for connection to the Pump Drive.
The single-turn, adjustable Speed Control, Figure 2, provides variable flow operation. The green PWR On indi­cator lights whenever the pump is oper­ating. The Power On/Direction Switch turns power on when either clockwise or counterclockwise pump rotor direction is selected. The MAX Button is used for priming and purging and operates the pump at maximum speed when depressed.
Occlusion Bed Latch
Universal Power Supply
Figure 1. Masterflex
®
L/S®Compact Pump System
Figure 2. Pump — Front Panel
Tubing Retainer Bracket
Tubing Retainer Bracket
PWR On Indicator
MAX: Press and hold to prime or purge pump
Speed Control: Provides variable flow adjustment
Power On/Direction Switch
Occlusion Bed Latch: Push to right to release occlusion bed for loading or changing tubing
4
The rear panel, Figure 3, contains a DC power input jack for connection of primary power and a 4-terminal barrier strip for connection of remote start/stop and for a DC remote supply.
SETUP
Use only MASTERFLEX®L/S®tubing sets to ensure optimum performance. Use of other tubing may void applicable warranties.
Selecting Tubing Size
Appendix A provides a list of tubing sets which will work efficiently with the MASTERFLEX
®
L/S®Compact Pump System. This list includes tubing diameters, flow rates in mL/min and the maximum flow rates per revolution at 20 rpm and 100 rpm. Tubing is listed by part number. For best results, when using two tubes (two channel models only), select tubes that provide flow rates within a 10:1 range. (Tube diameter ratios should be within approximately 3.5:1.) Silicone tubing should not be used with TYGON
®
or PHARMED®.
Installing Tubing in Pump Head
WARNING: Be sure the Pump Drive is turned off
before proceeding. Loose clothing could be caught in the rollers.
1. Place the Power On/Direction Switch,
Figure 2
, in the
center (Off) position.
2. Push the Occlusion Bed Latch, Figure 4, to the right to
release the Occlusion Bed from the rotor.
3. Insert the tube set into the slot on one side of the tubing
retainer bracket with the tubing stop on top of the bracket (opposite side from the rotor).
4. Wrap the tubing around the rollers and insert the tubing
into the slot on the opposite side of the bracket, stretching the tubing slightly so that the second tubing stop is also on top of the bracket.
5. For two channel modes, the tubing must be in the same
slots on both sides of the bracket (i.e. both in front slot or both in rear slot). If two tubes are being installed, repeat 3 and 4 above with the second tube set.
6. Lift the occlusion bed up against the tubing and push the
occlusion bed latch to the left until it locks in place in the cutout on the occlusion bed.
Figure 3. Pump — Rear Panel
DC Power Input,
2.5 mm Coaxial Receptacle
Shorting Bar
Remote
DC
Input
Remote
Start/Stop
Input
Figure 4. Occlusion Bed in Unlatched Tube Loading Position
Occlusion Bed Latch
Occlusion Bed in
Unlatched (Tube
Loading) Position
Rear Tube
Retainer
Slots
Front Tube
Retainer
Slots
5
Connect the External Power Supply to the applicable input voltage source and the output of the External Power Supply to the DC input connector on the pump unit.
NOTE: The Power Supply output connection is center positive (+).
Connecting Primary Power
Terminals 1 and 2 on the rear panel Barrier Terminal Strip, Figure 3, provide a means for connecting a remote DC power source. The positive (+) terminal is terminal 1. The negative (–) terminal is terminal 2.
NOTE: Input voltage must not exceed 15V DC or equipment may be damaged. A minimum of 11.0V DC is required for proper operation.
Remote Start/Stop Connection
Terminals 3 and 4 on the rear panel Terminal Strip, Figure 3, are used for remote start/stop operation. Pump direction and speed are not remotely controllable. In non-remote operation, these terminals are connected together by a Shorting Bar.
For remote control by switch closure, remove the Shorting Bar and connect the two terminals of the remote control switch to terminals 3 and 4. A closure of the remote control switch contacts will start the Pump System. Opening the contact will stop the Pump System.
Remote DC Power Connection
6
OPERATION
This section describes the procedures for obtaining desired performance. Flow rate is determined by the drive speed and the tubing size.
WARNING: Tubing breakage may result in fluid
being sprayed from pump. Use appropriate measures to protect operator and equipment.
Turning Pump System On and Selecting Direction of Operation
The Pump System can be set to operate in either a clockwise or a counterclockwise direction. The same control used to select direction also turns power on or off. Select direction of pump operation desired. PWR indicator should light.
Setting Pump Speed
Pump speed is controlled by the variable Speed Control. Turning the control clockwise increases the speed. Tube life is decreased with increased operating speed.
Priming Pump System
The MAX Push Button is used for priming and purging the Pump System. When the MAX button is depressed, the pump operates at maximum speed in the selected direction until the button is released.
MAINTENANCE
Cleaning: Keep the drive enclosure clean with mild detergents. Never immerse nor use excessive fluid.
Motor/Brush Replacement: The motor used in this inexpensive drive is not serviceable or repairable and has brushes which cannot be replaced. The motor is rated for a continuous duty life of 1500 operating hours minimum. Operation at other than continuous duty cycles will result in longer motor life. If motor failure for whatever reason occurs, the motor must be replaced.
Replacement Parts
Item Part No.
Rubber feet (4 ea.) A-1390-0001 Shorting bar A-4402 Power supply adapter 77200-07 Speed knob B-1083-0060 Speed knob cap A-4623 Speed knob skirt A-4621
7
SPECIFICATIONS
Output:
Operating Speed: 20 to 100 rpm Number of Tubing Channels:
Model 77240-00 1 Model 77240-10 2 Direction of rotation: Clockwise and Counterclockwise
Input:
Operating 115V AC nominal, 50/60 Hz Voltage/Frequency: (90–130VAC) @500 mA AC
230V AC nominal, 50/60 Hz (190–260VAC) @300 mA AC
13.5V DC nominal (11.0 to 15.0V DC) @ 2.4 A DC
Remote Start/Stop: Contact closure connection at
terminal strip contacts 3 and 4
Construction:
Dimensions (L x W x H): 6.50 in x 5.25 in x 4.50 in
(16.5 cm x 13.4 cm x 11.4 cm) Weight: 3.3 pounds (1.5 kg) Color: Grey Material: Painted Steel
Enclosure Rating: IP22 per IEC529
Environment:
Operating Temperature: 32°F to 104°F (0°C to 40°C) Storage Temperature: –49°F to 149°F (–45°C to 65°C) Humidity (Non-cond): 10% to 90% Pollution Degree: Pollution Degree 2
(Indoor usage — lab, office)
Compliance (For CE Mark):
EN61326 (EMC Directive) Converter is UL listed and CSA approved.
WARRANTY
Use only MASTERFLEX®L/S®tubing links to ensure optimum performance. Use of other tubing may void applicable warranties.
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge, if it is not due to misuse or abuse on your part, as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the Warranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss. Any damages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
8
APPENDIX A
CAUTION: Tubing for use with the MASTERFLEX®L/S®Compact Pump Systems are Masterflex®L/S
®
two-stop tube sets. Use of tubing other than that specified will result in poor pumping performance and/or pump system damage and voiding of applicable warranty.
Tubing
Pump Model Material
Size
77240-00 77240-10 Silicone PharMed
®
Tygon
®
Viton
®
L/S® 13 X X 06421-13 06431-13 06416-13 06428-13 L/S® 14 X X 06421-14 06431-14 06416-14 06428-14 L/S® 16 X N/A 06421-16 06431-16 06416-16 06428-16 L/S® 25 X N/A 06421-25 06431-25 06416-25 06428-25
Tubing Tubing ID
Nominal flow rate per channel in ml/min
Size in (mm)
ml/rev 20 RPM 100 RPM
L/S® 13 0.03 (0.76) 0.04 0.8 4.0 L/S® 14 0.06 (1.5) 0.14 2.8 14 L/S® 16 0.12 (3.0) 0.54 11 54 L/S® 25 0.19 (4.8) 1.1 21 105