Masterflex 77122-00, 77122-10, 77122-20, 77122-02, 77122-12 Instruction Manual

...
The MASTERFLEX C/L Pump Systems are small peristaltic pumps with integral drive motors and are intended for use with a series of tubing sizes that provide flow rates in the range of
2.0 µL/min to 43 mL/min.
OOPPEERRAATTIINNGG MMAANNUUAALL::
PPUUMMPP SSYYSSTTEEMM
Model No.
7777112222--0000 7777112222--1100 7777112222--2200
7777112222--0022 7777112222--1122 7777112222--2222
Edition 02
Cole-Parmer
1-800-MASTERFLEX (627-8373) (U.S. and Canada only) • 11 (847) 549-7600 (Outside U.S.) (847) 549-7600 (Local) • www.masterflex.com • techinfo@coleparmer.com
Thermo Fisher Scientific
1-800-637-3739 (U.S. and Canada only) • 11 (847) 381-7050 (Outside U.S.) (847) 381-7050 (Local) • www.thermo.com • bar.barnant@thermofisher.com
TTAABBLLEE OOFF CCOONNTTEENNTTSS
Title Page
SAFETY PRECAUTIONS ........................................................2
INTRODUCTION......................................................................3
GENERAL DESCRIPTION ...................................................4-5
SETUP...................................................................................5-6
Selecting Tubing Size .......................................................5
Installing Tubing in Pump Head........................................5
Installing Panel Mount Mounting Brackets.......................6
Installing Rubber Feet.......................................................6
Connecting Primary Power...............................................6
Backup Battery Connection..............................................6
Remote Start/Stop Connection.........................................6
OPERATION ............................................................................7
Turning Pump System On and Selecting Direction of
Operation ..........................................................................7
Setting Pump Speed.........................................................7
Priming Pump System ......................................................7
MAINTENANCE.......................................................................7
Cleaning ............................................................................7
Replacement Parts and Accessories................................7
SPECIFICATIONS ...................................................................8
WARRANTY .............................................................................9
PRODUCT RETURN................................................................9
TECHNICAL ASSISTANCE ....................................................9
APPENDIX A
Available Microbore Autoanalysis Tubing.......................10
Tubing Flow Rates ..........................................................11
SSAAFFEETTYY PPRREECCAAUUTTIIOONNSS
WARNING: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures
to protect operator and equipment.
CAUTIONS: Tubing for use with the MASTERFLEX C/L Pump Systems is Microbore Autoanalysis Tubing. See
Appendix A for specifics. Use of tubing other than that specified will result in poor pumping perfor­mance and/or pump system damage and voiding of applicable warranty.
Do not close the occlusion bed by pulling up on the locking tabs. Damage to the tabs may occur.
WWAARRNNIINNGG:: PPRROODDUUCCTT UUSSEE LLIIMMIITTAATTIIOONN
These products are not designed for, nor intended for use in patient connected applications; including, but not limited to, medical and dental use, and accordingly have not been submitted for FDA approval.
C-FLEX — Reg TM Consolidated Polymer Technologies, Inc. NORYL — Reg TM General Electric Company PHARMED, TYGON — Reg TM Saint-Gobain Performance Plastics Corp. VITON — Reg TM E.I. duPont DeNemours & Co.
2
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
IINNTTRROODDUUCCTTIIOONN
The MASTERFLEX C/L Pump System is designed to pump fluid through Microbore tubing by means of peristaltic action at very low flow rates. It is ideal for sanitizers, reagent dis-
pensing, analyzers, printing systems, controlled feeding and non-human infusion procedures.
3
GGEENNEERRAALL DDEESSCCRRIIPPTTIIOONN
The MASTERFLEX C/L Pump System, Figure 1, is enclosed in a plastic 1/4 DIN Case.
Rubber Feet, supplied with the unit, can be attached to the bottom of the Case for oper­ating on a bench or other flat surface and for stability when stacking multiple units. The enclosure allows panel mounting using optional Mounting Brackets (see Accessory Section for Part No. and page 6 for mount­ing instructions).
The Pump System accommodates one tube at controlled speeds as low as 1.7 rpm. Tubing is held in place by spring-loaded Tubing Retainers. For a list of tubing sizes for use with the Pump System, refer to Appendix A.
Bezel
1/4 DIN Case
Mounting
Bracket
(not Supplied)
Universal
Power Supply
The latching Occlusion Bed allows quick loading or changing of tubing. The Pump Head Rotor contains four rollers for minimum pulsation. All units operate from an External DC Power Supply. The AC mod­els are supplied with a Universal Power Supply, which provides a DC output for connection to the Pump Drive.
The single-turn, adjustable Speed Control, Figure 2, provides variable flow - operation. The green PWR On indicator lights whenever the pump is operating. The Power On/Direction Switch turns power on when either clockwise or counterclockwise pump rotor direction is selected. The MAX Button is used for priming and purging and operates the pump at maximum speed while depressed.
Figure 1. MASTERFLEX®C/L®Pump System
Door
Finger Recesses
Figure 2. Pump — Front Panel
PWR On Indicator
MAX: Press and hold to prime or purge pump
Speed Control: Provides variable flow adjustment
Power On/Direction Switch
4
Rollers
Locking
Tab
Occlusion
Bed
Locking Ta b
Recess - Push Here
to Close Bed
Tubing
Retainers
The rear panel, Figure 3, contains a DC power input jack for
Locking
Tab
Occlusion
Bed
Bed Stop
Recess - Push Here
to Close Bed
Bed Stop
Locking
Tab
connection of primary power and a 4-terminal barrier strip for connection of remote start/stop and for a DC backup supply.
SSEETTUUPP
Use only Microbore Autoanalysis tubing with MASTERFLEX C/L pumps to ensure optimum performance. Use of other tubing may void applicable warranties.
DC Power Input,
2.5 mm Coaxial Receptacle
Backup
DC
Input
SSeelleeccttiinngg TTuubbiinngg SSiizzee
Appendix A provides a list of tubing sizes which will work efficiently with the MASTERFLEX C/L Pump System. This list includes tubing diameters, flow rates in µL/min and the maximum flow rates at 10 rpm, 80 rpm and 300 rpm. Tubing is listed by part number.
IInnssttaalllliinngg TTuubbiinngg iinn PPuummpp HHeeaadd
WARNING: Be sure the Pump Drive is turned off
before proceeding. Loose clothing could be caught in the rollers.
1. Place the Power On/Direction Switch, Figure 2, in the center (Off) position.
2. Open the Door, Figure 2, by grasping the finger recesses on either side and pulling away from the pump.
3. Release the Occlusion Bed, Figure 4, by squeezing the Locking Tabs, Figure 4, together and sliding the bed down towards the door hinge.
4. Insert the selected tubing into one of the Tubing Retainers, Figure 4, by moving the retainer toward the center of the pump and placing the tubing in the v-notch of the retainer. Release the retainer so that the tubing is secured in place.
5. Wrap the tubing around the Rollers, Figure 4, and then secure it in the other tubing retainer as in step 4. The tubing should be stretched slightly around the rollers so that there is no excess tubing between the rollers and the retainers.
Remote
Start/Stop
Input
Shorting Bar
Shorting Bar
Figure 3. Pump — Rear Panel
Figure 4. Occlusion Bed in Unlatched Tube Loading Position
6. Close the occlusion bed by pushing up on the recessed area at the bottom, Figure 5, until the locking tabs snap into place against the Bed Stops, Figure 5, on the housing. Verify that there is no excess tubing between the rollers and the retainers - any excess can be removed by pulling slightly on the tubing outside of the pump.
CAUTION: Do not close the occlusion bed by
pulling up on the locking tabs. Damage to the tabs may occur.
7. Close the door. The locking tabs must be on the outside of the door and the door fully closed for proper operation.
Figure 5. Closing Occlusion Bed
5
IInnssttaalllliinngg PPaanneell MMoouunntt MMoouunnttiinngg BBrraacckkeettss
The optional Mounting Brackets are used to attach the 1/4 DIN Case to a Mounting Rack. Use Mounting Brackets Part No. 77120-03, (2/set).
1. Be sure the panel bezel is on the pump, then slide the pump system through the front of the mounting rack.
2. Place a Mounting Bracket, Figure 6, between the grooved bracket retainers on one side of the Case and slide the Mounting Bracket back to lock into the retainer grooves.
3. Tighten the bracket screw against the back of the rack to hold the Pump System in place.
4. Repeat steps 2 and 3 for the second Mounting Bracket on the opposite side.
IInnssttaalllliinngg RRuubbbbeerr FFeeeett
Four rubber feet are provided for operating the Pump System on a bench or other flat surface and for stability when stack­ing multiple units. The feet should be installed at the four cor­ners on the bottom of the unit. Remove the protective paper from the adhesive surface of each foot and press the foot firmly on the bottom surface about one-eighth inch in from and parallel to the outer edges.
CCoonnnneeccttiinngg PPrriimmaarryy PPoowweerr
Depending on the Pump System model, primary power may be 100-240V AC or 12V DC. The DC models can be con­nected to any DC* supply and are not supplied with a Universal Power Supply unit. Connect the External Power Supply to the applicable input voltage source and the output of the External Power Supply to the DC input connector on the pump unit.
*Refer to Specifications for current supply voltage.
NOTE: The Power Supply output connection is center positive (+).
Mounting Bracket
Bracket
Retainers
Bezel
Figure 6. Mounting Bracket Installation
BBaacckkuupp BBaatttteerryy CCoonnnneeccttiioonn
Terminals 1 and 2 on the rear panel Barrier Terminal Strip, Figure 3, provide a means for connecting a backup DC power source. The positive (+) terminal is terminal 1. The negative (–) terminal is terminal 2. Connection to these terminals would usually be made only for emergency type operation in con­junction with an External Power Supply, or a stand-alone operation for the DC models.
NOTE: Input voltage must not exceed 15V DC or equipment may be damaged. A minimum of 11.0V DC is required for proper operation.
RReemmoottee SSttaarrtt//SSttoopp CCoonnnneeccttiioonn
Terminals 3 and 4 on the rear panel Terminal Strip, Figure 3, are used for remote start/stop operation. Pump direction and speed are not remotely controllable. In non-remote operation, these terminals are connected together by a Shorting Bar.
For remote control by switch closure, remove the Shorting Bar and connect the two terminals of the remote control switch to terminals 3 and 4. A closure of the remote control switch contacts will start the Pump System. Opening the contact will stop the Pump System.
6
OOPPEERRAATTIIOONN
This section describes the procedures for obtaining desired performance. Flow rate is determined by the drive speed and the tubing size.
WARNING: Tubing breakage may result in fluid being
sprayed from pump. Use appropriate measures to protect operator and equipment.
TTuurrnniinngg PPuummpp SSyysstteemm OOnn aanndd SSeelleeccttiinngg DDiirreeccttiioonn ooff OOppeerraattiioonn
The Pump System can be set to operate in either a clockwise or a counterclockwise direction. The same control used to select direction also turns power on or off. Select direction of pump operation desired. PWR indicator should light.
MMAAIINNTTEENNAANNCCEE
SSeettttiinngg PPuummpp SSppeeeedd
Pump speed is controlled by the variable Speed Control. Turning the control clockwise increases the speed. Tube life is decreased with increased operating speed.
PPrriimmiinngg PPuummpp SSyysstteemm
The MAX Push Button is used for priming and purging the Pump System. When MAX is depressed, the pump operates at maximum speed in the selected direction until the button is released. Approximately two feet of tubing length can be filled or emptied per minute using the MAX Push Button on the 10 rpm units.
CClleeaanniinngg
Clean exterior surfaces of case, control panel and pump rollers using dry or damp cloth. Never immerse nor use excessive fluid.
RReeppllaacceemmeenntt PPaarrttss aanndd AAcccceessssoorriieess
Item Part No.
Rubber Feet (4 ea.) A-1390-0001 Shorting Bar A-4402 Door 108229
Accessory Part No.
Mounting Brackets (2/set) 77120-03 Universal Power Supply 100-240V AC 77200-07
7
SSPPEECCIIFFIICCAATTIIOONNSS
OOuuttppuutt::
Operating Speed:
Models 77122-00 1.7 to 10 rpm and 77122-02
Models 77122-10 13 to 80 rpm and 77122-12
Models 77122-20 50 to 300 rpm and 77122-22
Maximum No. of Tubes: 1
Direction of rotation: Clockwise and
Counterclockwise
IInnppuutt::
Supply Voltage/Frequency:
Model 77122-00, -10, -22 115V AC nominal, 50/60 Hz
(90–130V AC) @ 400 mA AC 230V AC nominal, 50/60 Hz (190–260V AC) @ 250 mA max.
Model 77122-02, -12, -22 12.0V DC nominal,
(11.0–16.0V DC) @ 2.4 A DC
Power Input: 11.0–16.0V DC to terminals
1 and 2 of terminal strip or External Power Supply
Installation Category:
77122-00, -10, -20 Category II per IEC664
(Local level—appliances, portable equipment, etc.)
77122-02, -12, -22 Category I per IEC664
(Signal Level)
Remote Start/Stop: Contact Closure connection
at terminal strip contacts 3 and 4
CCoonnssttrruuccttiioonn::
Dimensions (L x W x H): 7.0 in x 3.5 in x 3.5 in
(17.8 cm x 8.9 cm x 8.9 cm) Weight: 1.5 pounds (0.68 kg) Color: Black
Material:
Pump Head: PPS, polyester
and stainless steel Case: NORYL
Enclosure Rating: IP22 per IEC529
®
EEnnvviirroonnmmeenntt::
Operating Temperature: 32°F to 105°F (0°C to 40°C) Storage Temperature: –49°F to 149°F (–45°C to 65°C) Humidity (Non-cond): 10% to 90% Altitude: Less than 6600 ft (2000 m) Pollution Degree: Pollution Degree 2 per IEC664
(Indoor usage — lab, office)
CCoommpplliiaannccee ((FFoorr CCEE MMaarrkk))::
EN61326-1/A2: 2001 (EMC Directive) Converter is UL listed and CSA approved. Regulatory agency specifications not applicable to the balance of the unit due to low voltage.
8
WWAARRRRAANNTTYY
Use only Microbore Autoanalysis tubing with MASTERFLEX C/L pumps to ensure optimum
performance. Use of other tubing may void
applicable warranties.
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part, as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is one (1) year from date of purchase.
PPRROODDUUCCTT RREETTUURRNN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss. Any damages resulting from improper packaging are your responsibility.
TTEECCHHNNIICCAALL AASSSSIISSTTAANNCCEE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
9
AAPPPPEENNDDIIXX AA
CAUTION: Tubing for use with the MASTERFLEX C/L Pump Systems is Microbore Autoanalysis Tubing. Use of
tubing other than that specified will result in poor pumping performance and/or pump system damage and voiding of applicable warranty.
Available Microbore Autoanalysis Tubing
Tubing
Suffix.
10 0.007 (0.19)
12 0.010 (0.25)
14 0.015 (0.38)
16 0.017 (0.44)
18 0.020 (0.51)
22 0.025 (0.64)
24 0.030 (0.76)
26 0.035 (0.89)
28 0.040 (1.02)
30 0.045 (1.14)
32 0.051 (1.30)
34 0.056 (1.42)
36 0.060 (1.52)
38 0.065 (1.65)
40 0.073 (1.85)
42 0.081 (2.06)
44 0.090 (2.29)
46 0.100 (2.54)
48 0.110 (2.79)
Tubing
ID in (mm)
PharMed®
BPT
95709 07625 95612 95609 95718 96429 95712 07632
Silicone
peroxide
Silicone
platinum
Tygon®
lab
C-FLEX®
Tygon®
LFL
Solvent/
hydrocarbon
Viton®
Note: Dark Gray indicates availability.
10
Tubing Flow Rates
See the following table for Tubing size versus ID and flow rates. Flow rates are for water pumped at room temperature and 0 psi. Flow rates for silicone tubing are approximately 25%–50% less than listed.
Flow rate is determined by drive, speed, tubing size and material.
Tubing Tubing Flow Rate
Suffix. ID 10 rpm 80 rpm 300 rpm
in (mm) (µL/min) (µL/min) (µL/min)
-10 0.007 (0.19) 13.0 100 385
-12 0.010 (0.25) 22.5 180 670
-14 0.015 (0.38) 50.5 400 1510
-16 0.017 (0.44) 67.0 535 2000
-18 0.020 (0.51) 87.0 700 2600
-22 0.025 (0.64) 135 1050 4050
-24 0.030 (0.76) 185 1450 5550
-26 0.035 (0.89) 245 1950 7400
-28 0.040 (1.02) 315 2500 9400
-30 0.045 (1.14) 385 3100 11500
-32 0.051 (1.30) 485 3850 14500
-34 0.056 (1.42) 565 4500 17000
-36 0.060 (1.52) 635 5100 19000
-38 0.065 (1.65) 730 5850 22000
-40 0.073 (1.85) 885 7100 26500
-42 0.081 (2.06) 1050 8500 32000
-44 0.090 (2.29) 1250 9350 36000
-46 0.100 (2.54) 1450 10200 40500
-48 0.110 (2.79) 1650 11000 43000
All flow rates based on pumping water @ 0 psig 70°F (21°C).
11
1-800-MASTERFLEX (627-8373) (U.S. and Canada only)
11 (847) 549-7600 (Outside U.S.)
(847) 549-7600 (Local)
www.masterflex.com
techinfo@coleparmer.com
Thermo Fisher Scientific
1-800-637-3739 (U.S. and Canada only)
11 (847) 381-7050 (Outside U.S.)
(847) 381-7050 (Local)
www.thermo.com
bar.barnant@thermofisher.com
Printed in U.S.A.