Masterflex 77120-20, 77120-30, 77120-40, 77120-25, 77120-35 Instruction Manual

...
OPERATING MANU AL
®
PUMP SYSTEM
77120-20 77120-30 77120-40 77120-25 77120-35 77120-45 77120-52 77120-62 77120-70
The MASTERFLEX® C/L® Pump Systems are small peristaltic pumps with inte­gral drive motors and are intended for use with a series of tubing sizes that provide flow rates in the range of 2.2 µL/min to 37 mL/min.
®
Cole-Parmer Instrument Co.
1-800-MASTERFLEX (627-8373) (U.S . and Canada only)
11 (847) 549-7600 (outside U.S.)
(847) 549-7600 (Local)
www.masterflex.com
Barnant Company
1-800-637-3739 (U.S. and Canada only)
11 (847) 381-7050 (outside U.S.)
(847) 381-7050 (Local)
www.barnant.com
A-1299-0711
Edition 04
TABLE OF CONTENTS
TITLE PAGE
SAFETY PRECAUTIONS ......................................................................... 3
INTRODUCTION....................................................................................... 4
GENERAL DESCRIPTION ....................................................................... 4
SETUP ...................................................................................................... 6
Selecting Tubing Size ............................................................................ 6
Installing Tubing in Pump Head............................................................. 6
Installing Panel Mount Mounting Brackets ............................................ 7
Installing Rubber Feet ........................................................................... 7
Connecting Primary Power.................................................................... 8
Backup Battery Connection .................................................................. 8
Remote Start/Stop Connection ............................................................. 8
OPERATION ............................................................................................. 9
Turning Pump System On and Selecting Direction of Operation........... 9
Setting Pump Speed ............................................................................. 9
Priming Pump System .......................................................................... 9
MAINTENANCE ...................................................................................... 10
Cleaning .............................................................................................. 10
Replacement Parts and Accessories .................................................. 10
SPECIFICATIONS .................................................................................. 11
WARRANTY ............................................................................................ 13
PRODUCT RETURN............................................................................... 13
TECHNICAL ASSISTANCE .................................................................... 13
APPENDIX A MASTERFLEX® C/L
Pump System Tubing Flow Rates ..................................................... A-1
®
NORYL — Reg TM General Electric Company PHARMED, TYGON — Reg TM Nor ton Co. VITON — Reg TM DuPont Dow Elastomers, L.L.C. Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
2
SAFETY PRECAUTIONS
WARNING: Tubing breakage may result in fluid being sprayed
from pump. Use appropriate measures to protect operator and equipment.
CAUTION: Tubing for use with the MASTERFLEX
®
C/L® Pump Systems is Microbore Autoanalysis Tubing. See Appendix A for specifics. Use of tubing other than that specified will result in poor pumping perfor­mance and/or pump system damage and voiding of applicable warranty.
WARNING: PRODUCT USE LIMITATION
These products are not designed for, nor intended for use in patient­connected applications; including, but not limited to , medical and dental use, and accordingly have not been submitted for FDA approval.
These products are covered by one or more of the following U.S. and corresponding foreign patents: 5,380,173.
3
INTRODUCTION
The MASTERFLEX® C/L® Pump System is designed to pump fluid through Microbore tubing by means of peristaltic action at very low flow rates. It is ideal for sanitizers, reagent dispensing, analyz ers, printing systems, controlled feeding and non-human infusion procedures.
GENERAL DESCRIPTION
The MASTERFLEX® C/L® Pump System, Figure 1, is enclosed in a plastic 1/4 DIN Case. The enclosure allows panel mounting using optional Mounting Brackets (see Accessory Section). Rub ber Feet, supplied with the unit, can be attached to the bottom of the Case for operating on a bench or other flat surf ace. The Pump System can accommodate two tubes simultaneously for the 6 and 60 rpm systems at controlled speeds as low as 1 rpm. The 200 rpm system accommodates one tube at controlled speeds as low as 35 rpm. Tubing is held in place by spring-clip Tubing Retainers. For a list of tubing sizes f or use with the Pump System, refer to Appendix A. The latching Occlusion Bed allows quick loading or changing of tubing. The Pump Head Rotor contains six rollers for minimum pulsation. All units operate from an Exter nal DC Power Supply. The 115V AC and 230V AC models are supplied with an External Power Supply, which provides a DC output for connection to the Pump Drive.
Figure 1. MASTERFLEX® C/L® Pump System
The single-turn, adjustable Speed Control, Figure 2, provides variable flow operation. The green PWR On indicator lights whenever the pump is operat­ing. The Power On/Direction Switch turns power on when either cloc kwise or counterclockwise pump rotor direction is selected. The MAX Button is used for priming and purging and operates the pump at maximum speed while depressed.
4
Figure 2. Pump — Front Panel
The rear panel, Figure 3, contains a DC power input jack for connection of primary power and a 4-terminal barrier str ip for connection of remote start/ stop and for a DC backup supply.
Figure 3. Pump — Rear Panel
5
SETUP
Use only Microbore Autoanalysis tubing with MASTERFLEX® C/L® pumps
to ensure optimum performance.
Use of other tubing may void applicable warranties.
Selecting Tubing Size
Appendix A provides a list of tubing sizes which will work efficiently with the MASTERFLEX µL/min and the maximum flow rates at 6 rpm, 60 rpm and 200 rpm. T ubing is listed by part number. For best results, when using two tubes, select tubes that provide flow rates within a 10:1 range. (T ube diameter ratios should be within approximately 3.5:1.) Silicone tubing should not be used with TYGON® or NORPRENE®.
Installing Tubing in Pump Head
®
C/L® Pump System. This list includes tubing diameters, flow r ates in
WARNING: Be sure the Pump Drive is turned off before
proceeding. Loose clothing could be caught in the rollers.
1. Place the Power On/Direction Switch, Figure 2, in the center (Off) position.
2. Press down on the Occlusion Bed Latch, Figure 4, to release the Occlusion Bed from the rotor.
3. Gently stretch the tubing and insert the tubing into the spr ing-loaded left rear Retainer.
4. While maintaining the slight amount of stretching, wrap the selected tubing around the lower Rollers and secure the tubing in the right rear spring­loaded Retainer. If a second tubing is to be installed, secure the tubing in the same manner using the front spring-loaded Retainers.
5. Lift up on the outer edge of the Occlusion Bed until it locks in position holding the tubing against the bottom Rollers.
Figure 4. Occlusion Bed in Unlatched Tube Loading Position
6
Installing Panel Mount Mounting Brackets
The optional Mounting Brackets are used to attach the 1/4 DIN Case to a Mount­ing Rack. Use Mounting Brackets Part No. 77120-03, (2/set).
1. Be sure the panel bezel is on the pump, then slide the pump system through the front of the mounting rack.
2. Place a Mounting Bracket, Figure 5, between the groo ved brack et retainers on one side of the Case and slide the Mounting Bracket back to lock into the retainer grooves.
3. Tighten the bracket screw against the back of the rack to hold the Pump System in place.
4. Repeat steps 2 and 3 for the second Mounting Bracket on the opposite side.
Figure 5. Mounting Bracket Installation
Installing Rubber Feet
Four rubber feet are provided for operating the Pump System on a bench or other flat surface. The feet should be installed at the four corners on the bottom of the unit. Remo ve the protectiv e paper from the adhesive surf ace of each foot and press the foot firmly on the bottom surface about one-eighth inch in from and parallel to the outer edges.
7
Connecting Primary Power
Depending on the Pump System model, primary power may be 115V AC, 230V AC, 12V DC or 13.5V DC . The DC models can be connected to any DC* supply and are not supplied with an External Power Supply unit. Connect the External Power Supply to the applicable input voltage source and the output of the External Power Supply to the DC input connector on the pump unit.
*Refer to Specifications for current supply voltage.
NOTE
The Power Supply output connection is center positive (+).
Backup Battery Connection
Terminals 1 and 2 on the rear panel Barrier Terminal Strip, Figure 3, provide a means for connecting a backup DC pow er source. The positive (+) terminal is terminal 1. The negativ e (–) terminal is terminal 2. Connection to these termi­nals would usually be made only for emergency type operation in conjunction with an External Power Supply, or a stand-alone operation f or the DC models.
NOTE
For Models 77120-30, -35, -40, -45, -52 and -62, input voltage must not exceed 16V DC or equipment may be damaged. A minimum of 11.5V DC is required for proper operation.
For Models 77120-20, -25 and -70, input voltage must not exceed 15V DC or equipment may be damaged. A minimum of 11.0V DC is required for proper operation.
Remote Start/Stop Connection
Terminals 3 and 4 on the rear panel Terminal Strip, Figure 3, are used for re­mote start/stop operation. Pump direction and speed are not remotely control­lable. In non-remote operation, these terminals are connected together by a Shorting Bar. For remote control by switch closure, remove the Shorting Bar and connect the two terminals of the remote control switch to terminals 3 and 4. A closure of the remote control switch contacts will start the Pump System. Opening the contact will stop the Pump System.
8
OPERATION
This section describes the procedures for obtaining desired performance. Flow rate is determined by the drive speed and the tubing size.
WARNING: Tubing breakage may result in fluid being
sprayed from pump. Use appropriate measures to protect operator and equipment.
Turning Pump System On and Selecting Direction of Operation
The Pump System can be set to operate in either a clockwise or a counterclock­wise direction. The same control used to select direction also turns power on or off. Select direction of pump operation desired. PWR indicator should light.
Setting Pump Speed
Pump speed is controlled by the variable Speed Control. Turning the control clockwise increases the speed. Tube lif e is decreased with increased oper ating speed.
Priming Pump System
The MAX Push Button is used for priming and purging the Pump System. When MAX is depressed, the pump operates at maximum speed in the selected direc­tion until the button is released. Approximately two feet of tubing length can be filled or emptied per minute using the MAX Push Button on the 6 rpm units.
9
MAINTENANCE
Cleaning
Clean exterior surfaces of case, control panel and pump rollers using dry or damp cloth. Never immerse nor use excessive fluid.
Replacement Parts and Accessories
Item Part No.
Rubber Feet (4 ea.) A-1390-0001 Shorting Bar A-4402
Accessory Part No.
Mounting Brackets (2/set) 77120-03 6 and 60 rpm External Power Supply, 115V AC 77120-01 6 and 60 rpm External Power Supply, 230V AC 77120-06 200 rpm External Power Supply 115V AC 77120-21 200 rpm External Power Supply, 230V AC 77120-26
10
SPECIFICATIONS
Output:
Operating Speed:
Models 77120-30, 77120-35 1 to 6 rpm
and 77120-52
Models 77120-40, 77120-45 10 to 60 rpm
and 77120-62
Models 77120-20, 77120-25 35 to 200 rpm
and 77120-70
Maximum No. of Tubes: 2 for 6 and 60 rpm models Direction of rotation: Clockwise and Counterclockwise
Input:
Supply V oltage/F requency:
Models 77120-30 and 77120-40 115V AC nominal, 50/60 Hz Models 77120-35 and 77120-45 230V AC nominal, 50/60 Hz Models 77120-52 and 77120-62 12V DC nominal (11.5–16.0V DC) Model 77120-20 115V AC nominal, 50/60 Hz Model 77120-25 230V AC nominal, 50/60 Hz Model 77120-70 13.5V DC nominal, (11.0–15.0V DC)
Power Input: For 6 and 60 rpm models
Installation Category:
77120-20, -25, -30, -35, -40, -45 Category II per IEC664
77120-52, -62, -70 Category I per IEC664
Remote Start/Stop: Contact Closure connection
Construction:
Dimensions (L x W x H): 5.3 in x 3.5 in x 3.5 in Weight: 2 pounds (0.91 kg)
Color: Black Material:
Pump Head: PPS, polypropylene and
1 for 200 rpm models
(90–130V AC) @104 mA AC (190–260V AC) @52 mA AC @ 1.0 A DC (90–130V AC) @ 120 mA AC (190–260V AC) @ 60 mA AC @ 2.4 A DC
11.5–16.0V DC to terminals 1 and 2 of terminal strip or External Power Supply For 200 rpm model
11.0–15.0V DC to terminals 1 and 2 of terminal strip or External Power Supply
(Local level—appliances, portable equipment, etc.)
(Signal Level) at terminal strip contacts 3 and 4
(13.5 cm x 8.9 cm x 8.9 cm)
stainless steel
11
Case: NORYL
®
Enclosure Rating: IP22 per IEC529
Environment:
Operating Temperature: 32°F to 105°F (0°C to 40°C) Storage Temperature: –49°F to 149°F (–45°C to 65°C) Humidity (Non-cond): 10% to 90% Altitude: Less than 6600 ft (2000 m)
Pollution Degree: Pollution Degree 2 per IEC664
(Indoor usage — lab, office)
Compliance (For CE Mark): EN61326-1/A1: 1998
(EMC Directive) 115V AC Power Supply AC/DC Converter is UL listed and CSA approved. Regulatory agency specifications not applicable to the balance of the unit due to low voltage.
12
WARRANTY
Use only Microbore Autoanalysis tubing with MASTERFLEX® C/L pumps to ensure optimum performance. Use of other tubing may void applicable warranties.
The Manufacturer warr ants this product to be free from significant deviations from published specifications. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part, as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the Warranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authoriza­tion and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and in­sure it against possible damage or loss. Any damages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you have an y questions about the use of this product, contact the Manuf ac­turer or authorized seller.
®
Cole-Parmer Instrument Co.
625 East Bunker Court
Vernon Hills, Illinois U .S.A. 60061-1844
1-800-MASTERFLEX (627-8373) (U.S. and Canada only)
11 (847) 549-7600 (outside U.S.)
(847) 549-7600 (Local)
FAX (847) 247-2929 (U .S . and Canada only)
11 (847) 549-1700 (Fax outside U .S.)
www.masterflex.com
e-mail: techinfo@coleparmer .com
Barnant Company
28W092 Commercial Ave.
Barrington, Illinois U.S.A. 60010-2392
1-800-637-3739 (U.S. and Canada only)
11 (847) 381-7050 (outside U.S.)
(847) 381-7050 (Local)
11 (847) 381-7053 (Fax outside U .S.)
(847) 381-7053 (Local Fax)
www.barnant.com
e-mail: barnant@barnant.com
13
Printed in U.S.A.
040901
APPENDIX A
CAUTION: Tubing for use with the MASTERFLEX® C/L® Pump Systems
is Microbore Autoanalysis Tubing. Use of tubing other than that specified will result in poor pumping performance and/or pump system damage and voiding of applicable warranty.
Available Microbore Autoanalysis Tubing
TYGON — Catalog prefix number 95609-XX available in sizes 10 through 48. TYGON LFL — Catalog prefix number 96429-XX availab le in sizes -18, -26, -30,
-34, -42 and -48. Silicone (peroxide-cured) — Catalog prefix number 07625-XX av ailab le in sizes
22 through 48. Silicone (platinum-cured) — Catalog prefix number 95612-XX availab le in siz es 22 through 48. PHARMED® — Catalog prefix number 95709-XX availab le in sizes -12, -18, -26,
-32, -36, -40, -44 and -48. VITON® fluoroelastomer — Catalog prefix number 07632-XX available in sizes
-26, -30, -34, -38, -42, -46 and -48. See the following tab le for T ubing size v ersus ID and flow rates. Flow rates are for water pumped at room temperature and 0 psi. Flo w r ates f or silicone tubing are approximately 25%–50% less than listed. Flow rate is determined by drive, speed, tubing size and material.
Tubing Tubing Suffix. ID 200 rpm 60 rpm 6 rpm
in (mm) (µL/min) (µL/min) (µL/min)
-10 0.008 (0.19) 400 128 13
-12 0.010 (0.25) 610 178 17
-14 0.015 (0.38) 1290 378 45
-16 0.017 (0.44) 1673 490 55
-18 0.020 (0.51) 2280 698 70
-22 0.025 (0.64) 3630 1110 105
-24 0.030 (0.76) 5180 1580 150
-26 0.035 (0.89) 7160 2110 205
-28 0.040 (1.02) 8980 2680 270
-30 0.045 (1.14) 10950 3290 335
-32 0.051 (1.30) 14120 4130 420
-34 0.056 (1.42) 16800 4900 495
-36 0.060 (1.52) 18700 5510 555
-38 0.065 (1.65) 21200 6330 630
-40 0.073 (1.85) 24890 7650 750
-42 0.081 (2.06) 28090 8910 875
-44 0.090 (2.29) 31210 10160 1015
-46 0.100 (2.54) 34160 11340 1170
-48 0.110 (2.79) 36720 12340 1330
All flow rates based on pumping water @ 0 psig 70°F (21°C).
Flow Rate
A-1
EU DECLARATION OF CONFORMITY
Name of Apparatus: MASTERFLEX® C/L® Pump System Model Numbers: 77120-20, -25, -30, -35, -40, -45, -52, -62, -70 Description of Apparatus: Variable Speed P eristaltic Pump Motor Drive. Used with
Barnant Company declares that the above models are in conformity to the following harmonized standards and directives:
Applicable Applicable Manufacturer’s Directives Specifications Report Number
89/336/EEC EN61326-1/A1: 1998 TR9420 92/31/EEC 93/68/EEC
pump head and tubing to pump fluids. 77120-20 (200 rpm @115V AC), -25 (200 rpm @ 230V
AC), -30 (6 rpm @115V AC), -35 (6 rpm @ 230V AC),
-40 (60 rpm @ 115V AC), -45 (60 rpm @ 230V AC),
-52 (6 rpm @12V DC), -62 (60 rpm @ 12V DC),
-70 (200 rpm @13.5V DC)
Manufacturer: Barnant Company Division
Cole-Parmer Instrument Company 28W092 Commercial Avenue Barrington, IL 60010-2392 USA T el.: 847-381-7050
Manufacturer’s Signature:
James W . Doll
Vice President, Engineering
23 August, 2000
Date