Masterflex 77110-07, 77110-00 Instruction Manual

OPERATING MANUAL:
B/T®RAPID - LOAD FIXED SPEED PERISTALTIC PUMPS AND DRIVE
MASTERFLEX®B/T®Pump 77110-00
Model No.
77110-00 77110-07
A-1299-5124
Edition 02
Cole-Parmer
1-800-MASTERFLEX (627-8373) (U.S. and Canada only) • 11 (847) 549-7600 (Outside U.S.) (847) 549-7600 (Local) • www.masterflex.com • techinfo@coleparmer.com
Thermo Fisher Scientific
1-800-637-3739 (U.S. and Canada only) • 11 (847) 381-7050 (Outside U.S.) (847) 381-7050 (Local) • www.thermo.com • bar.barnant@thermofisher.com
2
TABLE OF CONTENTS
Title Page
SAFETY PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
APPLICATION DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
GENERAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
INSTALLATION AND SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
TUBING TYPES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
INSTALLING THE PUMP TUBING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
REPLACING ROLLERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
PRODUCT RETURN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
TECHNICAL ASSISTANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
B/T, PERFECTPOSITION – Reg TM Thermo Fisher C-FLEX – Reg TM Consolidated Polymer Technologies, Inc. NORPRENE, PHARMAPURE, PHARMED and TYGON – Reg TM Saint-Gobain Performance Plastics Corp.
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
SAFETY PRECAUTIONS
DANGER: High voltages exist and are accessible. Do not remove cover of Drive. Use extreme caution
when servicing internal components.
CAUTIONS: Risk of electric shock – this pump is supplied with a grounding conductor and grounding-type
attachment plug. To reduce risk of electric shock, be certain that it is connected only to a properly grounded, grounding-type receptacle.
Electrical connections and grounding (earthing) must conform to local wiring codes.
WARNINGS: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to
protect operator and equipment.
Power must be removed from pump before removing or installing tubing. Fingers or loose clothing could get caught in drive mechanism. Do not operate this pump without cover or interlock door properly closed and latched. Rotating parts can cause serious injury.
To reduce risk of injury, do not pump materials hotter than 150 degrees Fahrenheit, (65.5°C).
Explanation of Symbols
CAUTION: Risk of Danger. Consult Operator’s manual for nature of hazard and corrective actions.
CAUTION: Risk of crushing. Keep fingers away from rotor while pump is in operation. Stop pump before
loading or unloading tubing.
CAUTION: Hot Surface. Do not touch.
CAUTION: Risk of electric shock. Consult Operator’s manual for nature of hazard and corrective actions.
WARNING: PRODUCT USE LIMITATION
This product is not designed for, nor intended for use in, patient-connected applications, including, but not limited to, medical and dental use, and, accordingly, has not been submitted for FDA approval. If drive is used in a manner not specified in this manual the protection provided by the equipment may be impaired.
3
4
INTRODUCTION
This manual provides information for installing, operating and servicing the following models of MASTERFLEX®B/T
®
RAPID-LOAD peristaltic pumps.
MODEL TYPE
77110-00 Fixed Speed w/TEFC, 115V AC @ 60 Hz
77110-07 Fixed Speed w/TEFC, 230V AC @ 50 Hz
The unique design of these peristaltic pumps provides a greatly simplified means for rapid loading and changing of tubing. In addition, different sizes and formulations of tubing can be used. Unlike many older type pumps where tubing had to be carefully measured and tubing connectors were mandated, this pump provides for the use of a continuous piece of tubing to ensure sanitary connections. Fluid contacts only the tubing and fittings. These pumps are designed to handle a wide range of viscosities and can pump up to 10 GPM.
APPLICATION DATA
The gentle peristaltic action of these pumps is ideal for pumping highly viscous and shear-sensitive liquids. These pumps are also ideally suited for use where sterile conditions and purity are required. Toxic and hazardous fluids can be pumped with the proper selection of MASTERFLEX
PERFECTPOSITION
B/T tubing since the fluid contacts only
the tubing and not the pump.
WARNING: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to
protect operator and equipment.
GENERAL DESCRIPTION
Due to its unique design, different MASTERFLEX
PERFECTPOSITION
B/T tubing sizes can be accommodated by
this RAPID-LOAD peristaltic pump.
For an indirect estimate of flow rate, a reflective element attached to the rotor has been provided for use with an optical tachometer. Point tachometer beam through front cover window and target the reflective element.
To obtain flow rate in mL/min., multiply tachometer rpm reading times the nominal flow per revolution value provided in TABLE 1 (Note: 3,785 mL = 1 U.S. liquid gallon).
If the pump is to be operated with back-pressure on the discharge line above 5 PSIG, hose clamps or screw-type band clamps are recommended on the tubing connections on the discharge side of the pump. Silicone or C-FLEX
®
tubings, because of their highly elastic natures, can expand very quickly if back-pressure is present and could create leakage at the fittings if not securely retained. It is this same elastic nature, however, that makes them such excellent materials for this peristaltic type pump and give it its long life. NORPRENE®tubing yields longer life especially under pressure. Do not attempt to use other materials in lieu of these, as pump performance could be severely compromised with possible damage to the pump. Refer to TABLE 1 for tubing recommendations.