Masterflex 07519-60, 07519-50 Instruction Manual

Masterflex 07519-60, 07519-50 Instruction Manual

OPERATING MANUAL

®

CARTRIDGE

PUMP HEAD SYSTEM

Pump Head Models 07519-05 07519-06

Cartridge Models

Small: 07519-60

Large: 07519-50

Multi-channel peristaltic pumps accommodate two sizes of snap-in tubing cartridges which provide balanced, adjustable flow via instantly changeable occlusion setting. The system mates with most existing MASTERFLEX® L/S® pump drives.

Cole-Parmer Instrument Co.

625 East Bunker Court

Vernon Hills, Illinois U.S.A. 60061-1844

800-MASTERFLEX (627-8373)

(847) 247-2929 (Fax) www.masterflex.com

e-mail: techinfo@coleparmer.com

A-1299-0539

Edition 08

®

CONTENTS

Title

Page

SAFETY PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

(1) Application Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

(2) General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 SETUP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

(1) Select Cartridges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

(2) Load Cartridges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

(3) Retainer Adjustment/Cartridge Removal. . . . . . . . . . . . . . . . . . . . 7

(4) Bi-Directional Pumping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

(1) Select Tubing and Pump Heads . . . . . . . . . . . . . . . . . . . . . . . . . 8

(2) Select Pump Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

(3) Adjust Occlusion Settings on Cartridges. . . . . . . . . . . . . . . . . . . . 9 MAINTENANCE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Service Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 PRODUCT RETURN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 TECHNICAL ASSISTANCE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 APPENDIX A MASTERFLEX® L/S® Pump Drive Types . . . . . . . . . . . . A-1 APPENDIX B Cartridge Loading Capacity . . . . . . . . . . . . . . . . . . . . . B-1

C-FLEX-Reg TM Consolidated Polymer Technologies, Inc.

 

PHARMED, NORPRENE, TYGON-Reg TM Norton Co.

These products are covered by one

VITON-Reg TM DuPont Dow Elastomers L.L.C.

or more of the following U.S. and

Trademarks bearing the ® symbol in this publication

corresponding foreign patents:

are registered in the U.S. and in other countries

4,886,431 and 5,257,917

i

SAFETY PRECAUTIONS

WARNINGS: Tubing breakage may result in fluid being sprayed from

pump. Use appropriate measures to protect operator and equipment.

Turn drive off before removing or installing cartridges. Safety guards are provided to minimize risk of fingers getting caught between the roller mechanism and the base of the module. However, be safe—Keep your fingers away from these areas.

CAUTIONS: Excessive occlusion can cause high pump temperatures and early tubing and roller failure. Do not set occlusion tighter than what is recommended in Table #3.

Large Cartridge 07519-50 should not be used with microbore tubing—may result in premature tubing failure.

WARNING: PRODUCT USE LIMITATION

These products are not designed for, nor intended for use in patient connected applications; including, but not limited to medical and dental use, and accordingly have not been submitted for FDA approval.

ii

INTRODUCTION

The instructions in this manual are task-oriented for easy reference. You can go directly to a particular section and quickly find the answers. Appendix A and B list MASTERFLEX® L/S® pump drive types that can be used with this system.

 

Pump Heads

MODEL

 

TYPE

 

 

 

 

Model 07519-05

 

3-Roller, 4 or 8 Channel

Model 07519-06

 

4-Roller, 4 or 8 Channel

 

 

 

 

Cartridges

MODEL

 

TYPE

 

 

 

 

Model 07519-50

 

Large

Model 07519-60

 

Small

 

 

 

These cartridge pumps are designed to provide up to 8 simultaneously driven pump channels and incorporate the following features.

Simultaneous dispensing into an 8 bank set of test tubes with one set-up and pump cycle.

The four-roller design provides reduced pulsation.

(1) Application Data

The Cartridge Pump System is designed for multi-channel use where close occlusion control of a variety of different flows is required. All standard and many special tubing sizes and materials can be used, delivering from .015 mL/min to 2 L/min.

1

(2) General Description

Each Pump Head System is composed of a heavy-duty Pump Head and two sizes of snap-in tubing cartridges, each with individually adjustable occlusion settings. The occlusion “wedge” design provides accurate and more balanced flow control.

The Pump Heads contain either three or four rollers. All models can operate with a variety of tubing and with either the small or the large cartridges.

The Pump Head quickly mounts to most existing MASTERFLEX® L/S® Pump Drive types (see Appendix A and B).

MOUNTING

HOLES

TANG (REAR OF

PUMP HEAD)

MOTOR MOUNTING

BOOT

SCREWS (4)

SAFETY

GUARD

SCREWDRIVER

CLEARANCE HOLE CARTRIDGE MOUNTING

RODS

ADJUSTABLE

FOOT

LOCKNUT

FIGURE 1 PUMP MOUNTING

2

Loading...
+ 13 hidden pages