07519-05 Pump Head with 07519-70 Large Cartridge
Tête de pompe 07519-05 à cartouche épaisse 07519-70
Pumpenkopf 07519-05 mit großer Schlauchkassette 07519-70
Cabeza de bomba 07519-05 con cartucho grande 07519-70
Testa pompante 07519-05 con cartuccia grande 07519-70
Model Nos.
Modèles n°
Modellnummern
Números de modelo
Modelli n°
07519-05
07519-06
BOMBAS DE CARTUCHOS
MANUALE DI ISTRUZIONI
SISTEMI CON TESTA
POMPANTE A CARTUCCE
Cartridge Small
Cartouche mince
Kleine Schlauchkassette
Cartucho pequeños
Cartucce piccole
C-FLEX — Reg TM Consolidated Polymer Technologies, Inc.
C-FLEX, PHARMED, NORPRENE, TYGON—Reg TM Saint-Gobain Performance Plastics Corp.
SANTOPRENE — Reg TM AES.
VITON — Reg TM E.I. duPont DeNemours & Co.
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
1
SAFETY PRECAUTIONS
WARNINGS:Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to pro
tect operator and equipment.
Turn off drive beforeremoving orinstalling Cartridges. Safetyguards areprovided tominimize risk
of fingers getting caught between the roller mechanism and the base of the module. However, be
safe—Keep your fingers away from these areas.
CAUTIONS: Excessive occlusion can cause high pump temperaturesand early tubing and roller failure. Do not
set occlusion tighter than what is recommended in Table 4.
Use only tubing/cartridge combinations defined by Tables 1, 2, and 3. Use of other combinations
could cause malfunction or damage to the pump.
Not using the BOOT will damage the pump.
Explanation of Symbols
CAUTION:Risk of Danger. Consult Operator’s manual for nature of hazard and corrective actions.
CAUTION:Risk of crushing. Keep fingers away from rotor while pump is in operation. Stop pump before load
ing or unloading tubing.
-
-
WARNING: PRODUCT USE LIMITATION
These products are not designed for, nor intended for use in, patient-connected applications, including, but not
limited to, medical and dental use and, accordingly, have not been submitted for FDA approval.
2
INTRODUCTION
The instructions in this manual are task-oriented for easy reference. You can go directly to a particular section and
quickly find the answers. Appendix A and B list MASTERFLEX
®
L/S®pump drive types which can be used with this
system.
Pump Heads
MODELTYPE
Model 07519-05
Model 07519-06
3-Roller, 4 or 8 Channel
4-Roller, 4 or 8 Channel
Cartridges
MODELTYPE
Model 07519-70
Model 07519-80
These Cartridge Pumps are designed to provide up to 8 simultaneously driven pump channels and incorporate the fol
lowing features.
Simultaneous dispensing into 8 separate containers with one set-up and pump cycle.
·
The four-roller design provides reduced pulsation.
·
(1) Application Data
Large
Small
-
The Cartridge Pump System isdesigned for multi-channeluse where closeocclusion control of avariety of differentflows
is required. Tubing found in Table 3 may be used to deliver from .074 to 1700 mL/min.
The Pump Heads contain either three or four rollers. Both models can operatewith a variety oftubing and with eitherthe
small or the large Cartridges.
®
The Pump Head quickly mounts to most existing MASTERFLEX
L/S®Pump Drive types (see Appendix A and B).
3
INSTALLATION
Tool Required: Allen Key (provided).
Mount the Pump Head to a MASTERFLEX
1. Install the plastic Bootover the tang at the end of pump shaft,(Figure 1). This Boot prevents metal-to-metal contact,
and thus reduces noise. This Boot should be inspected and replaced when worn or at 500 hour intervals.
CAUTION:Not using the boot will damage the pump.
NOTE: The boot with closed ends (A-3480) will fit most drives.Use of the bootwith open ends (A-3514) may be required
on some drives. Both boots are supplied with the pump.
2. Connect the Pump to Drive by aligning the tang on Pump Head, (including installed boot) with the slot in the motor
Drive shaft.
3. Attach Pump Head to drive unit with four screws (provided) using the Allen Key (also provided).
WARNING: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to pro
tect operator and equipment.
4. If the Pump Head requires support, install the appropriate length support screw with plastic foot and locknut (three
provided). Level the Pump Head withthe drive unitby adjusting thefoot. Tighten thelocknut against the bottomof the
pump body.
®
L/S®Pump Drive as follows:
-
FIGURE 1PUMP MOUNTING
A-3480
FIGURE 2PLASTIC BOOTS
4
A-3514
SETUP
Two Cartridge sizes (0.72 in and 0.36 in thick) (see Figure 3) accommodate a broad range of tubing sizes as noted in
Tables 1 and 2.
FIGURE 3SMALL AND LARGE CARTRIDGES
(1) Select Cartridges
NOTE: These Pump Heads are designed to work with the Model 07519-70 and 07519-80 Cartridges. Do not use Model
07519-75 or 07519-85 Cartridges, which are designed for different Pump Heads.
Depending on which Pump Head is used, up to four large or eight small tubing Cartridges can be teamed up with the
Pump Head. But there is a limit depending on available drive powerand pump load. For full tubing flow rates at 0 psi, the
table in Appendix Bshows cartridge capacitiesaccording totubing sizeand drive Type(see AppendixA fordescription of
Drive Types). Cartridge sizes can be intermixed as long as the total load does not exceed the drive capability.
The small Cartridge is designedto operate with MASTERFLEX
accommodate specially designedMicrobore TubingSets (.9 mm wall — seeTables 1 and 3).The small Cartridge can be
used with either the three- or the four-roller Pump Head.
The large Cartridge is designed to operate withMASTERFLEX
®
L/S
17, with either the three- or four-roller Pump Head (see Tables 2 and 3).
®
precision tubing sizes L/S®13 and L/S®14, and will also
®
L/S®precision tubing sizes L/S®14, L/S®16, L/S®25 and
Use only MASTERFLEX Precision Tubing and Microbore Tubing Sets with MASTERFLEX Pumps to ensure
optimum performance. Use of other tubing may void applicable warranties.
(2) Load Cartridges
Easy, snap-in mounting and quick-release removal come with a Cartridge system—no tools required. Since the Car
tridges function uniformly in either direction, Cartridge orientation is not important. However, you may wish to alternate
Cartridges so that the Adjustment Knob on each is easier to turn when several small Cartridges are mounted close to
gether.
WARNING: Turn offdrive beforeremoving or installingCartridges. Safetyguards areprovided tominimize risk
of fingers getting caught between the roller mechanism and the base of the module. However, be
safe—Keep your fingers away from these areas.
CAUTION:Use only tubing/cartridge combinations defined by Tables 1, 2, and 3. Use of other combinations
could cause malfunction or damage to the pump.
With the drive stopped, follow this procedure to install Cartridges.
1. Turn the Occlusion Adjustment Knob (see Figure 4) counterclockwise to maximumopen. This step isnot necessary
when the occlusion is setat the factory setting or if the occlusion has already been setappropriately during previous
pumping.
2. Set the Tubing Retainer according to tubing size (see Figure 6A)when using the largeCartridge. No retainer adjust
ment is required when using the small Cartridge. (Note: The retainer button located on the Occlusion Adjustment
Knob-side of the large Cartridge isa loose component and may fall out. To reassemble, push it back into placewhile
pressing in the button.)
-
-
-
5
FIGURE 4CARTRIDGE ASSEMBLY
3. Select tubingand place in Cartridge (see Figure 6B).If using a Microbore TubingSet (see Figure 5), placethe Collar
up against the Tubing Retainer (see Figure 6C) on both sides of the Cartridge.
4. Attach the Hook Leg of the Cartridge onto one of the Support Rods while holding the tubing between the Cartridge
legs and against the Tubing Retainer (see Figure 6D).
5. Swing the Cartridge with the tubing over the rotor and push down on the Tab (see Figure 4) until the Latch snaps
closed. (See Figure 6E.) Note: Depress the Cartridge as required in the location indicated in Figure 6E to facilitate
latching and unlatching.
6. Lightly pull the tubing at the outlet and inlet to remove the slack when using continuous tubing. No slack removal is
required when using a Microbore Tubing Set.
7. Adjust the occlusion setting (see OPERATION section). For a nominal setting with MASTERFLEX tubing, turn adjustment screw until inside edge of the wedge is aligned midway between #3 and #4 on the label.
FIGURE 5MICROBORE TUBING SET
6A
6B
6D
FIGURE 6TUBING AND CARTRIDGE LOADING
6
6C
6E
(3) Cartridge Removal
Follow this procedure to unload the Cartridge.
1. Turn the Drive OFF.
2. Push down on latch side of Cartridge (see Figure 6E) and pull Latch Leg outward to unhook from rod. The Latch
Release (see Figure 4) can be depressed to aid in this process as desired.
3. Pivot Cartridge upwards and unhook Hook Leg from rod and remove Cartridge.
(4) Bi-Directional Pumping
Fluids can be pumped into and back out of containers byreversing the drive. Ondrives that cannot be reversed, use lon
ger tubing lengths and simply reverse the loaded Cartridge.
WARNING: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to pro
tect operator and equipment.
(5) Partial Bank Pumping
The Pump can be operated with either a partial or a full bank of Car
tridges. There is an elastic, VITON
®
fluoroelastomer O-ring on each
of the two metal rods of the pump (see Figure 7) that holds the Car
-
tridges in place on the pump. These O-rings should be adjusted as
follows. If the pump is partially loaded with Cartridges, slide the
O-rings along the rods, up against the front of the outermost Car
tridge, to keep it in place(note position A, for example, in Figure 7). If
the pump is fully loaded with Cartridges, slide the O-rings to the front
of the pump as far as possible (note position B in Figure 7).
FIGURE 7O-RING POSITION
-
-
7
OPERATION
This section describes the procedures for obtaining the desired performance.
(1) Select Tubing and Pump Heads
Use Tables 1, 2, and 3to select thetubing, Cartridge, andPump Head toprovide the required singlechannel flow rate.
CAUTION:Use only tubing/cartridge combinations defined by Tables 1, 2, and 3. Use of other combinations
could cause malfunction or damage to the pump.
TABLE 1SMALL CARTRIDGE FLOW RATES—07519-80
®
MICROBORE TUBING (mm ID)MASTERFLEX
Pump
Head
3-Roller
3-Roller
4-Roller
4-Roller
Drive
rpmUnits
6
mL/min
600
mL/min
1
mL/min
100
mL/min
6
mL/min
600
mL/min
1
mL/min
100
mL/min
0.891.422.062.79L/S®13L/S®14
FLOW RATES
.44
44
.074
7.4
.44
44
.074
7.4
1.0
100
.17
17
1.0
100
.17
17
2.2
220
.37
37
2.2
220
.37
37
3.8
380
.63
63
3.8
380
.63
63
.30
30
.050
5.0
.28
28
.047
4.7
TABLE 2LARGE CARTRIDGE FLOW RATES—07519-70
Pump
Head
3-Roller
3-Roller
4-Roller
4-Roller
Drive
rpmUnits
6
mL/min
600
mL/min
1
mL/min
100
mL/min
6
mL/min
600
mL/min
1
mL/min
100
mL/min
L/S®14L/S®16L/S®25L/S®17
1.3
130
.21
21
1.2
120
.20
20
MASTERFLEX
4.6
460
.76
76
3.8
380
.63
63
NOTE: The Pumps operate at the same speed as the drive.
NOTE: Improper tubing stretch could result in reduced flow rate (see LOAD CARTRIDGE section).
®
L/S®TUBING
FLOW RATES
10
1000
1.7
170
8.3
830
1.4
140
TABLE 3COMPATIBLE TUBING
*MICROBORE TUBING SET
Tubing Material
Soft
Silicone (platinum)XXX
Silicone (peroxide)XX
TYGON®SiliconeXX
C-FLEX®(50 A)XX
Stiff
PHARMED®BPT (65)X
TYGON LFLX
NORPRENE®food (A 60 F)X
TYGON lab (R-3603)X
NORPRENE (A 60 G)X
TYGON fuel (F-4040-A)X
PVCX
VITONX
SANTOPRENE
*Use only specially designed MICROBORE TUBING SETS with Cartridges (Figure 5).
®
(mm ID)
.89, 1.42, 2.06, 2.79L/S®13, 14, 16, 25L/S®17
X
MASTERFLEX
®
L/S®TUBING
L/S®TUBING
1.3
130
.21
21
1.2
120
.20
20
17
1700
2.8
280
14
1400
2.3
230
8
(2) Select Pump Speed
Select fixed speed drive or adjust speed of variable speed drive to pro
vide desired nominal flow ratewithin the rpmspeed range shownfor the
pump model selected.
(3) Adjust Occlusion Settings on Cartridges
CAUTION:Excessive occlusion can cause high pump tempera
-
tures and early tubing and roller failure. Do not set
occlusion tighter than what is recommended in
Table 4.
NOTE: With largeand small Cartridges,the scale provides nominal oc
clusion for MASTERFLEX
®
L/S®precision tubing at the #3–#4 setting.
-
With a multi-channel Cartridge system, flow rate, pressure sensitivity
and tubing life can be fine-tuned—even while the pump is running. No
lost time, and withoperating speeds up to 600 rpm,you get higher max
imum flows with the same size tubing.
Figure 8 (large and small Cartridge) shows the wedges adjusted at #3
FIGURE 8 OCCLUSION INDICATIONS
on the occlusion setpoint scale for the MASTERFLEX tubing.
The Occlusion Adjustment Knob (see Figure 8) controls a screw mechanism to move the occlusion wedges. Once a
Cartridge is adjustedto a desired setting fora particularapplication, there is no need to reset.Or, youcan record settings
and quickly reproduce them at a later time.
(a) Normal Occlusion Settings
The following table shows the appropriate occlusion settings to provide satisfactory performance for various tubing
sizes. (For optimum occlusion settings, see following section.)
TABLE 4RECOMMENDED OCCLUSION SETTINGS
TubingLarge Cartridge ScaleSmall Cartridge Scale
Microbore Tubing Set
MASTERFLEX®L/S®Precision
N/A#1 – #2
#3 – #4#3 – #4
(b) Occlusion Setting Procedure:
1. Select the recommended occlusion value from the table.
2. Turn the Occlusion AdjustmentKnob to align the inside edge of the white Wedge with thescale number. (Clockwise
rotation increases the occlusion.)
(c) Optimized Occlusion Settings
Some applications require additional fine-tuning of the occlusion setting to vary the flow rate for a particular tubing, to
reduce flow variations caused by changes in system pressure, or to increase tubing life.
1. Adjust the Occlusion Wedges as described in the preceding table.
2. Refine this setting, depending on your objectives:
NOTE:To Maximize Tubing Life
While running the pump, simply reduce the adjusted occlusion setting by turning the Occlusion Adjustment Knob coun
terclockwise to move the wedges to a higher setpoint scale number (toward 5). Continue turning the knob counterclock
wise until the flow drop-off or the pressure sensitivity becomes unacceptable, then turn knob slightly clockwise.
-
-
NOTE: To Reduce or Eliminate Flow Drop-Off Rate with Pressure Fluctuations
While running the pump alternate the backpressure between the expected high and low values and increase the
adjusted occlusion setting.Turn the OcclusionAdjustment Knob clockwise to move the wedges toa lower setpoint scale
number (toward 1) until the flow drop-off is minimized.
NOTE: To Fine-Adjust the Flow Rate
Increase or decrease the occlusion value to vary the flow.
9
MAINTENANCE
No maintenance required. Wipe Pump witha clean clothand a mild detergent.Never immerse oruse excessive fluid.
0° to 40°C (32°F to 104°F) Operating Temperature Range:
Storage Temperature Range:
Noise level: <70 dBA @ 1 meter
Compliance: EN809 (EU Machine Directive)
CARTRIDGES
Dimensions:
40°C to 110°C (
Fiberglass-reinforced polysulfone housing; stainless steel,Construction and
anodized aluminum, sealed stainless steel ball bearingsChemical Resistance:
and Buna N safety guard.
Durable polycarbonate frame with filled nylon and acetalConstruction:
components. Anodized aluminum adjustment knob
on stainless steel screw, VITON O-ring.
4 in Tall x 5 in W x 0.72 in ThickLarge
(102 mm x 127 mm x 18.3 mm)
4 in Tall x 5 in W x 0.36 in ThickSmall
(102 mm x 127 mm x 9.1 mm)
40°F to 230°F)
10
WARRANTY
Use only MASTERFLEX Precision Tubing and Microbore TubingSets with MASTERFLEX Pumpsto ensureoptimum performance. Use of other tubing may void applicable warranties.
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair or
adjustment is necessarywithin the warrantyperiod, the problem will be corrected at no charge if it is not due to misuse or
abuse on your part as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from
product misuse or abuse, may be invoiced to you.
The warranty period for this product is two (2) years from the date of purchase.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for
the return. For your protection, pack the product carefully and insure it against possible damage loss. Any damages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
US & Canada only
Toll Free 1-800-MASTERFLEX | 1-800-637-3739
Outside US & Canada
1-847-549-7600 | 1-847-381-7050
11
*EN809 manufactured by:
Cole-Parmer Instrument Company
28W092 Commercial Avenue, Barrington, IL 60010
techinfo@masterflex.com | www.masterflex.com
Printed in U.S.A.
APPENDIX A
MASTERFLEX
®
L/S®PUMP DRIVE TYPES
FOR USE WITH CARTRIDGE PUMP SYSTEMS
Cartridge
Size
Small
Large
Max
rpm
600
120
6001/2090II
1001/10360III
Power
(hp)
1/10
1/20
Torque
oz-in
180
180
Drive
Type
I
Note: Check Max rpm and Power of the Drive or use the Torque rating to determine the Drive Type.
C-FLEX, PHARMED, NORPRENE et TYGON sont des marques déposées de Saint-Gobain Performance Plastics Corp.
SANTOPRENE est une marque déposée de AES.
VITON est une marque déposée de E.I. duPont DeNemours & Co.
Les marques accompagnées du symbole ® qui sont mentionnées dans la présente
publication sont déposées aux États-Unis et dans d’autres pays.
13
MESURES DE SÉCURITÉ
AVERTISSEMENTS : La rupture d’un tube peut entraîner une pulvérisation de liquide refoulé par la pompe. Prendre des
mesures appropriées pour protéger l’opérateur et les appareils.
Mettre l’entraînement hors tension avant de retirerou de mettre enplace des cartouches. Desdisposi
tifs de protection sont prévus pour minimiser les risques de pincement des doigts entre le mécanisme
à rouleaux et la base du module. Toutefois, être prudent : ne pas approcher les doigts de ces endroits.
CONSEILS DE
PRUDENCE :
Une occlusion excessive peut entraîner une surchauffe de la pompe, ainsi qu’une défaillance préma
turée des tubes et des rouleaux. Ne pas dépasser l’occlusion recommandée au Tableau n°4.
Utiliser uniquement les combinaisons tubes/cartouches définies aux Tableaux 1, 2 et 3. L’utilisation
d’autres combinaisons risque d’entraîner un mauvais fonctionnement de la pompe ou
d’endommager cette dernière.
L’utilisation de la pompe sans TAMPON endommagera celle-ci.
Explication des symboles
ATTENTION :Risque dedanger. Consulter lemanuel de l’opérateur pour vérifier la nature des risque et prendre les
mesures correctives.
ATTENTION :Risque d’écrasement. Eloigner les doigts du rotor lorsque la pompe fonctionne. Arrêter la pompe
avant le chargement ou le déchargement du tubage.
-
-
AVERTISSEMENT : LIMITES D’UTILISATION DES PRODUITS
Ces produits ne sontpas conçuspour, nidestinés à,être utilisés dans des applications avec patients, y comprisentre autresles
applications médicales et dentaires, et n’ont par conséquent pas été soumis à l’agrément de la FDA.
14
INTRODUCTION
Les instructions données dans la présente notice sont orientées vers les tâches, ce qui en facilite la consultation. Il est possible de
consulter directement unesection particulière et d’obtenir rapidement les réponsesaux questions que l’on se pose.Les Annexes A et
B indiquent les types d’entraînements de pompes MASTERFLEX
®
L/S®compatibles avec ce système.
Têtes de pompes
MODÈLETYPE
Modèle 07519-05
Modèle 07519-06
3 rouleaux, 4 ou 8 canaux
4 rouleaux, 4 or 8 canaux
Cartouches
MODÈLETYPE
Modèle 07519-70
Modèle 07519-80
Ces pompes à cartouches sont conçues pour offrir jusqu’à 8 canaux de pompage activés simultanément et présentent les
caractéristiques suivantes.
Distribution simultanée dans 8 récipients distincts avec un seul cycle de préparation et de pompage.
·
La conception à quatre rouleaux permet de réduire la pulsation.
·
(1) Applications
Le système de pompage à cartouches est conçu pour les applications multicanaux dans lesquelles un contrôle fin de l’occlusion à
différents débits est nécessaire. Les tubes indiqués sur le Tableau n° 3 peuvent être utilisés pour obtenir un débit de 0,074 à
1700 ml/min.
(2) Description générale
Chaque système de tête de pompe se compose d’une tête de pompe pour service intensif et de deux tailles de cartouches à tubes
emboîtables dont chacune offre un réglage individuel d’occlusion. La conception à occlusion par « coins » permet une régulation
précise et plus stable du débit.
Les têtes de pompes contiennent trois ou quatre rouleaux. Les deux modèles peuvent fonctionner avec divers tubes et avec les
cartouches minces ou épaisses.
La tête de pompe se monte rapidement sur la plupart des types d’entraînements de pompes MASTERFLEX
Annexes A et B).
Épaisse
Mince
®
L/S®existants (voir les
15
INSTALLATION
TAMPON
TROUS DE
MONTAGE
TROU DE
DÉGAGEMENT
DE TOURNEVIS
PIED
RÉGLABLE
CONTRE-ÉCROU
TIGES DE
MONTAGE DE
CASSETTES
DISPOSITIF DE
PROTECTION
QUEUE (À L’ARRIÈRE DE
LA TÊTE DE POMPE)
VIS DE MONTAGE
DE MOTEUR (4)
Outillage nécessaire : Clé Allen (fournie).
Accoupler la tête de pompe à un entraînement MASTERFLEX
1. Mettre le tampon en plastique en place par-dessus la queue qui se trouve à l’extrémité de l’arbre de la pompe (Figure 1). Ce
tampon empêche le contact métal sur métal, ce qui abaisse le niveau de bruit. Il convient d’examiner ce tampon et de le remplacer
lorsqu’il est usé ou toutes les 500 heures.
ATTENTION :L’utilisation de la pompe sans TAMPON endommagera celle-ci.
REMARQUE : le tampon à côtés fermés (A-3480) s’adapte à la plupart des entraînements. Il peut s’avérer nécessaire d’utiliser le
tampon à côtés ouverts (A-3514) sur certains entraînements. Les deux tampons sont fournis avec la pompe.
2. Accoupler la pompe àl’entraînement enalignant la queue de la tête de pompe (tampon compris) et la fente del’arbre moteur.
3. Fixez la tête de pompe à l’entraînement à l’aide des quatre vis (fournies) en utilisant la Clé Allen.
AVERTISSEMENT :La rupture d’un tube peut entraîner une pulvérisation de liquide refoulé par la pompe. Prendre des
mesures appropriées pour protéger l’opérateur et les appareils.
4. S’il s’avère nécessaire de soutenir la tête de pompe, poser une vis de support de la longueur correcte avec pied en plastique et
contre-écrou (fournis tous les trois). Mettre la tête de pompe à niveau avec l’entraînement en réglant le pied. Serrer le
contre-écrou contre le dessous du corps de la pompe.
®
L/S®en procédant comme suit :
FIGURE 1MONTAGE DE LA POMPE
A-3480
FIGURE 2TAMPONS EN PLASTIQUE
16
A-3514
PRÉPARATION
Les deux tailles de cartouches (0,72 et 0,36 po d’épaisseur) (voir la Figure 3) s’adaptent à une vaste gamme de diamètres de tubes
comme indiqué sur les Tableaux 1 et 2.
FIGURE 3CARTOUCHES MINCES ET ÉPAISSES
(1) Sélection des cartouches
REMARQUE : ces têtes de pompes sont conçues pour fonctionner avec les cartouches 07519-70 et 07519-80. Ne pas utiliser des
cartouches 07519-75 ou 07519-85, car elles sont conçues pour des têtes de pompes différentes.
Suivant la tête de pompe utilisée, il est possible d’associer à cette dernière jusqu’à quatre cartouches épaisses ou huit cartouches
minces. Cette capacité est toutefois limitée par la puissance de l’entraînement et la charge de la pompe. Le tableau de l’Annexe B
indique les capacités en cartouches en fonction du diamètre des tubes et dutype d’entraînement (voir l’Annexe A pourune description
des types d’entraînements) permettant d’obtenir le débit maximum à 0 kPa. Il est possible d’utiliser des cartouches des deux tailles
tant que la charge totale ne dépasse pas la capacité de l’entraînement.
La cartouche minceest conçue pour fonctionner avecles tubes de précision MASTERFLEX
les kits de tubes Microbore (paroi de 0,9 mm — voir les Tableaux 1 et 3) conçus spécialement pour être compatibles avec elle. La
cartouche mince peut être utilisée avec la tête de pompe à trois ou à quatre rouleaux.
La cartouche épaisse est conçue pour fonctionner avec les tubes de précisionMASTERFLEX
avec la tête de pompe à trois ou à quatre rouleaux (voir les Tableaux 2 et 3).
N’utiliser que des kits de tubes de précision MASTERFLEX et de tubes Microbore
avec les pompes MASTERFLEX pour garantir des performances optimales.
L’utilisation d’autres tubes peut annuler les garanties applicables.
®
L/S®13 et L/S®14 et accepteégalement
®
L/S®14, L/S®16, L/S®25 et L/S®17 et
(2) Mise en place des cartouches
Les cartouchess’emboîtent facilement en place et se déboîtent rapidement sans l’aide d’un outil. Dans la mesure où leur fonctionne
ment est identique dans un sens comme dans l’autre, leur orientation est sans importance. Il peut être toutefois préférable d’alterner
leur orientation de façon à pouvoir tourner leur bouton de réglage plus facilement quand plusieurs cartouches minces sont montées
l’une à côté de l’autre.
AVERTISSEMENT :Mettre l’entraînement hors tension avant de retirer ou demettre en place descartouches. Des disposi
tifs de protection sont prévus pour minimiser les risques de pincement des doigts entre le mécanisme
à rouleaux et la base du module. Toutefois, être prudent : ne pas approcher les doigts de ces endroits.
ATTENTION :Utiliser uniquement les combinaisons tubes/cartouches définies aux Tableaux 1, 2 et 3. L’utilisation
d’autres combinaisons risque d’entraîner un mauvais fonctionnement de la pompe ou
d’endommager cette dernière.
L’entraînement étant arrêté, procéder comme suit pour mettre les cartouches en place.
1. Tourner le bouton de réglage d’occlusion (voir la Figure 4) dans le sens antihoraire jusqu’à la position d’ouverture maximum.
Cette opération n’est pas nécessaire lorsque le réglage d’occlusion effectué à l’usine est conservé ou si l’occlusion a déjà été
réglée correctement lors du pompage précédent.
2. Régler le dispositif deretenue detube en fonction du diamètre de ce dernier(voir laFigure 6A) en cas d’utilisation de la cartouche
épaisse. Aucun réglage de ce type n’est nécessaire en cas d’utilisation de la cartouche mince. (Remarque : le bouton de
dispositif de retenue qui se trouve du côté bouton de réglage d’occlusionde la cartouche épaisse n’est pas assujetti et risque de
tomber. Pour le remonter, le réenfoncer en place tout en appuyant dessus.)
17
-
-
Loading...
+ 43 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.