All trademarks used in this publication are duly marked and the sole property of their
respective owners. No attempt at trademark or copyright infringement is intended or
implied.
Marathon Sensors makes no warranties express or implied beyond the written warranty
presented at initial purchase. Marathon Sensors Inc. is not responsible for any product,
process, damage or injury incurred while using this equipment. Marathon Sensors makes
no representations or warranties with respect to the contents hereof and specifically
disclaims any warranties of merchantability or fitness for any particular application or
purpose.
Table of Contents
GENERAL DESCRIPTION............................................................................................................................ 2
Process Type.............................................................................................................................................5
Carbon Process Factor.............................................................................................................................5
Dew Point Process Factor........................................................................................................................ 5
Function Field......................................................................................................................................... 10
Data Field...............................................................................................................................................10
Error Check Field (CRC)........................................................................................................................10
Please specify the following parameters when ordering a transmitter; process type, process
range (%, ppm), thermocouple type, temperature scale F/C, analog output 1 process and
scale, analog output 2 process and scale.
Typical Oxygen Transmitter Calibration
(F840030)
Calibration
Function
Measured Value or
Input
Output / Units
Cold JunctionRoom Temp°F
Thermocouple
min
Thermocouple
max
800°F (B type)
standard t/c type
3000°F (B type)
standard t/c type
°F
°F
Millivolt0.0 mVMillivolts
Millivolt2000 mVMillivolts
Analog 1 Zero0% O24.0 mA +/- 0.1
Analog 1 Span20.9% O220.0 mA +/- 0.1
Analog 2 Zero800°F +/- 5°4.0 mA +/- 0.1
Analog 2 Span3000°F +/- 5°20.0 mA +/- 0.1
Typical Carbon Transmitter Calibration
(F840031)
Calibration
Function
Measured Value or
Input
Output / Units
Cold JunctionRoom Temp°F
Thermocouple
Min
Thermocouple
Max
Analog 1 Zero0% Carbon4.0 mA +/- 0.1
Analog 1 Span2.55% Carbon20.0 mA +/- 0.1
Analog 2 ZeroMUST BE
4.0 mA +/- 0.1
SPECIFIED
Analog 2 SpanMUST BE
20.0 mA +/- 0.1
SPECIFIED
Page 1 of 23
11/14/2006Rev. 14
General Description
The Oxymit™ Transmitter has been designed to work as an analog or digital interface for
any zirconia based oxygen probe used to track dew point, carbon potential, or oxygen. The
transmitter connects to the temperature and millivolts outputs of an oxygen probe and can
produce analog outputs proportional to the selected process value.
The features available are:
Isolated inputs for thermocouple and probe millivolt
24 bit Sigma-Delta ADC for inputs.
SerialEEPROM to store setup and calibration values.
Two isolated self-powered 4-20mA outputs for process value and temperature.
The transmitter makes a carbon or oxygen probe an intelligent stand alone sensor. The
transmitter is located near the probe, preferably mounted in an enclosure. The transmitter
mounts onto a DIN rail and requires a 24VDC power supply. It measures the probe
temperature and millivolts. At the time of order the transmitter can be configured to
calculate percent carbon, dewpoint, or percent oxygen from these inputs. The results of
any of these calculations are made available via two4-20mA loop outputs. Typically one
first loop is set up for the process value the second loop transmits probe temperature.
T/C INPUT
mV INPUT
+24V
24V
COM
12
11
6
5
8
7
5V_A
44M
EVENT INPUT
Power
Supplies
22M
A/D
CONV.
5V_A
5V_B
+15V
-15V
+15V
-15V
EEPROM
Process
Controller
3
4
Figure 1 BLOCK DIAGRAM
5V_A
5V_A
DISPLAY
CONN.
5V_A
10
5V_B
A
RS485
B
ISOLATED
RTX+
9
RTX-
ISOLATED
-15V
+15V
1
ANALOG
OUT 1
4-20mA
2
D/A
C
CC
ISOLATED
-15V
+15V
14
ANALOG
OUT 2
4-20mA
13
D/A
D
DD
Page 2 of 23
11/14/2006Rev. 14
Safety Summary
All cautions and instructions that appear in this manual must be complied with to prevent
personnel injury or damage to the Probe Transmitter or connected equipment. The
specified limits of this equipment must not be exceeded. If these limits are exceeded or if
this instrument is used in a manner not intended by Marathon Sensors Inc., damage to this
instrument or connected devices could occur.
Do not connect this device directly to AC motors, valves, or other actuators. All AC alarm
functions must be connected through an interposing DC coil relay with a maximum coil
load of 0.5 amps DC. The Probe Transmitter is not rated to act as a safety device. It
should not be used to provide interlocking safety functions for any temperature or process
functions. Alarm capabilities are provided for probe test and input faults only and are not
to be considered or used as safety contacts in any application.
Connections
The Probe Transmitter has four removable terminal blocks grouped with four terminals
each. Each terminal is a wire clamp type with a standard slot screw. Each clamp can
accommodate AWG 24 to 12 flexible stranded wire. Maximum torque on the terminal
screws should not exceed 0.8 Nm.
The figure below shows the arrangement of the terminals.
1234
-+EVT EVT
AO1COM NO
LOWER
5678
-+-+
TCMV
UPPER
UPPER
910 11 12
-+-+
RS48524VDC
LOWER
13 14 15 16
-+N/C N/C
AO2
Figure 2 Terminal Layout
Page 3 of 23
11/14/2006Rev. 14
The next figure shows a schematic representation of the Probe Transmitter and typical
connections required in the field.
Figure 3 Schematic Connections
Grounding and Shielding
To minimize the pick-up of electrical noise, the low voltage DC connections and the sensor
input wiring should be routed away from high-current power cables. Where it is
impractical to do this, use shielded cables with the shield grounded at the Probe Transmitter
enclosure ground as show above.
Parameter Selections
The following tables list the parameters available in the Probe Transmitter. Default values
are also listed. The default values are loaded if a reset is force in the device. Changes to
these parameters must be specified at the time of order.
Process Parameters
The following table shows the process selections and other parameters that effect the
process value.
Page 4 of 23
11/14/2006Rev. 14
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.