Maquet VASOVIEW 7 Training Manual

TRAINING MANUAL
VASOVIEW
7 ENDOSCOPIC VESSEL HARVESTING SYSTEM
VASOVIEW 7 | Training Manual | English | v1
Copyright MAQUET Gmbh
This training manual for endoscopic vessel harvesting (EVH) using the VASOVIEW 7 xB is designed to provide the operator with step-by-step instructions for performing the EVH procedure. Included in this document are preoperative considerations, intra-operative technique tips, postoperative recommendations, and troubleshooting suggestions. This document is not intended to replace reading the product Instructions For Use nor is it intended to replace the independent judgement of the medical professional performing the procedure.
VASOVIEW 7 Training Manual
MAQUET has a strong history pioneering and advancing endoscopic vessel harvesting systems.
The VASOVIEW 7 xB Endoscopic Vessel Harvesting System is the product of multiple generations of learning and experience. MAQUET combines leading-edge advances in product design with a solid history of surgical success to provide the most advanced and effective EVH solutions available. VASOVIEW products are supported by clinical consultants who have both hospital surgical experience and industry-leading training in EVH. The VASOVIEW System provides real options to cardiac surgeons and their surgical team and real benefits to the patients they serve.
VASOVIEW 7 | Training Manual | English | v1 Copyright MAQUET Gmbh
TABLE OF CONTENTS
| MAQUET Medical Systems USA | ENDOSCOPIC VESSEL HARVESTING |
1 Endoscopic Vessel Harvesting Saphenous Vein | 4 Procedure-at-a-Glance 2 Endoscopic Vessel Harvesting Radial Artery | 5 Procedure-at-a-Glance
3 Operating Room Instrumentation and Equipment | 6 4 Equipment Testing | 7 5 Endoscopic Vessel Harvesting Procedure: | 8
Saphenous Vein Patient Positioning and Preparation | 8 Incision Site | 10 Inserting the 7 mm Endoscope With Attached | 11 Dissection Tip CO
Insufflation | 12
2
Endoscopic Dissection | 13 Branch Cauterization and Transection | 16 Optimizing Visualization | 19 Guidelines for Safe and Effective Use: | 20 BiSECTOR™ Bipolar Ligating Forceps Running the Vein | 21 Distal Transection and Ligation | 21 Vein Removal and Preparation | 22 Closing and Dressings | 22 6 Endoscopic Vessel Harvesting Procedure: | 23 Radial Artery Patient Positioning and Preparation | 23 Incision and Tourniquet Deployment | 24 Endoscopic Dissection | 24 Fasciotomy | 26 Branch Cauterization and Transection | 26 Transection and Retrieval of the Radial Artery | 27 Hemostasis and Conduit Preparation | 28 Closing and Dressings | 28
7 Documentation | 29 8 Nursing Considerations | 30 9 Case Setup | 31 10 Postoperative Procedures | 32 12 Troubleshooting | 33 13 Appendix A: Eight Steps for Tying a Ligation Loop | 34
3 |
VASOVIEW 7 | Training Manual | English | v1
Copyright MAQUET Gmbh
| 4
| ENDOSCOPIC VESSEL HARVESTING | MAQUET Medical Systems USA |
ENDOSCOPIC VESSEL HARVESTING
SAPHENOUS VEIN PROCEDURE-AT-A-GLANCE
Incision
Create an incision in the area near the knee, over the greater saphenous vein. Attach the Dissection Tip to the end of the 7 mm endoscope. Insert the 7 mm endoscope into the incision and visualize on the video monitor.
Initiating CO2 Insufflation
Advance the Short Port Blunt Tip Trocar (BTT) into position, and insufflate CO posterior and branch dissection with the Dissection Tip.
gas to create a tunnel. Perform anterior,
2
Cauterizing and Transecting Branches
Insert the VASOVIEW 7 Harvesting Cannula to cauterize and transect branches, using the BiSECTOR™ Bipolar Ligating Forceps.
Enhanced Visualization
For enhanced visualization during vessel branch cauteriza­tion and transection, transfer the CO
Insufflation Port to the Distal Insufflation CO2 Port on the
CO
2
source from the BTT
2
VASOVIEW Harvesting Cannula.
Running the Vessel
Deploy the C-Ring from the VASOVIEW Harvesting Cannula, engage the vein with the C-Ring and run the length of the vein to ensure that it is completely free and no branches remain.
VASOVIEW 7 | Training Manual | English | v1 Copyright MAQUET Gmbh
Transection and Retrieval
Make a stab wound over the distal end of the tunnel, clamp the saphenous vein, externalize it through the stab wound. Transect the vein. Ligate the terminal end of the vein. Retrieve the harvested vein from the original incision at the knee.
| MAQUET Medical Systems USA | ENDOSCOPIC VESSEL HARVESTING |
ENDOSCOPIC VESSEL HARVESTING
RADIAL ARTERY PROCEDURE-AT-A-GLANCE
Incision
Beginning just proximal to the wrist crease, create a longitudinal incision. Attach the Dissection Tip to the end of the 7 mm endoscope. Insert the 7 mm endoscope into the incision and visualize on the video monitor.
Initiating CO2 Insufflation
Insert the Short Port Blunt Tip Trocar (BTT) into position and begin CO dissection of the radial artery pedicle with the Dissection Tip on the 7 mm endoscope.
insufflation. Perform anterior, posterior, and lateral
2
5 |
Fasciotomy
Insert the VASOVIEW 7 Harvesting Cannula. Using the BiSECTOR™ Bipolar Ligating Forceps, release the fascia, cutting it from the distal to the proximal forearm.
Cauterizing and Transecting Branches
Use the C-Ring to protect the radial artery pedicle and expose the branches. Using the BiSECTOR tool, cauterize and transect branches.
Running the Vessel
Deploy the C-Ring from the VASOVIEW 7 Harvesting Cannula, engage the radial artery with the C-Ring and run the length of the artery to ensure that it is completely free and that no bran­ches remain.
Transection and Retrieval
Make a stab incision near the elbow, clamp the radial artery pedicle and externalize it through the stab incision. Transect the radial artery. Ligate the proximal radial artery, and then in retrograde fashion, retrieve the radial artery pedicle from the original incision at the wrist.
VASOVIEW 7 | Training Manual | English | v1
Copyright MAQUET Gmbh
| 6
| ENDOSCOPIC VESSEL HARVESTING | MAQUET Medical Systems USA |
OPERATING ROOM
INSTRUMENTATION AND EQUIPMENT
VASOVIEW 7 xB Endoscopic Vessel Harvesting System
7 mm Extended Length Endoscope
7 mm Endoscope Seal
Harvesting Cannula
Harvesting Cannula Seal
Port for distal insufflation
BiSECTOR™ Tool
Dissection Tip CO2 Insufflation
Syringe
for BTT
port with one-way valve
Short Port Blunt Tip Trocar (BTT)
Additional Items:
Anti-fog Solution Surgical instruments used for open harvest Bipolar Cord Bovie
Syringe for scope wash
Port for scope wash
Video Equipment:
Camera Box Video Monitor Light Source Insufflator Insufflator tubing CO2 Source Fiber Optic Cable with appropriate Scope Adaptor End Camera compatible to Endoscope
VASOVIEW 7 | Training Manual | English | v1 Copyright MAQUET Gmbh
EQUIPMENT TESTING
| MAQUET Medical Systems USA | ENDOSCOPIC VESSEL HARVESTING |
7 |
Prior to starting the EVH procedure, turn on all of the equipment on the video tower (video monitor, camera, light source, and insufflator). Be sure that the electric cords are attached to the back of each piece of equipment. Once all equipment checks are performed, turn the system com- pletely off until the procedure begins. The monitor should show color bars until a camera head is plugged into the camera box. If the color bars are not bal­anced, press the reset button to bring them back to the fac­tory setting. The camera head should be plugged into the camera box to check for a clear picture prior to sterilization (except in the case of terminal sterilization). The light source should be checked to be sure that light is coming from the unit. Do not look directly into the light. If the fiber optic light cable will be sterilized with the camera head, it should be checked for damaged fiber optics. To do this, hold one end toward a direct light source and look into the other end. If it illuminates with no more than one-third of the area grayed, the cable is in good order. If any part of the cable does not illuminate (reflects a blackened area), use a different cable. If the fiber optic cables need to be terminally sterilized, this quality-control procedure should take place prior to sterilization.
The endoscope should be checked in two ways prior to sterilization. First, look through the endoscope eyepiece and be sure that the image is clear. Second, hold the light post up to a direct light source and look directly into the distal end of the endoscope. A complete circle of light should be visible around the distal tip. Turn the CO
gas tank to the open position and check to be
2
sure there is either a full tank of gas or an extra tank avail­able in the room. Be sure that the gas is flowing from the unit when the insufflator is turned on. Set the insufflator at 3 – 5 L/min flow and 10 – 12 mmHg pressure. Be sure that the bipolar foot pedal is connected to the correct port of the electrocautery generator.
VASOVIEW 7 | Training Manual | English | v1
Copyright MAQUET Gmbh
| 8
| ENDOSCOPIC VESSEL HARVESTING | MAQUET Medical Systems USA |
ENDOSCOPIC VESSEL HARVESTING PROCEDURE
SAPHENOUS VEIN
PATIENT POSITIONING AND PREPARATION
Patient Positioning
Place the patient in a supine position. Externally rotate the leg with the knee flexed and with support behind the knee and the thigh. (Figure 1)
Surgical Preparation and Vein Location
Prior to the skin prep, assess the patient to select the incision site around the knee area. Once the intended incision site is identified, mark the area with a surgical marker.
Figure 1
The saphenous vein can usually be found by using anatomical landmarks. For morbidly obese patients, an ultrasonic doppler may be used to help locate the vein. If needed, use the doppler prior to the patient prep and then mark the site with a surgical marker. If appropriate, vein mapping may be performed preoperatively. Note that the saphenous vein in its natural position lies fairly posterior in relationship to the patella. Perform skin prep according to hospital protocol.
VASOVIEW 7 | Training Manual | English | v1 Copyright MAQUET Gmbh
| MAQUET Medical Systems USA | ENDOSCOPIC VESSEL HARVESTING |
9 |
Product Assembly
Attach the 7 mm Extended Length Endoscope to the camera coupler. Attach the light cable to the post on the endoscope. Focus the endoscope on a gauze pad and white balance the video camera as described below. Ensure proper orientation of the camera head. It is important to maintain correct orientation throughout the procedure. Thread the Dissection Tip onto the distal end of the endo­scope until the proximal edge of the Dissection Tip lines up with the indicator etched on the endoscope shaft. (Figures 2 and 3) Insert the endoscope with attached Dissection Tip through the BTT. Ensure that the Endoscope Seal is on the BTT (it comes packaged with this seal attached). (Figure 4)
Figure 2
Proper White Balancing
White balancing is the process by which the camera learns what “white” is. This process sets the full range of color through the camera.
To white balance the camera, attach the camera head and fiber optic cables to the endoscope and hold a piece of white gauze approximately 1.5 inches from the distal end of the endoscope. (Figure 5) Focus the camera and ensure that only white is visible. Press the white-balance button from either the camera head or camera box, depending on the camera system. Hold the camera and endoscope still until the system indicates white balance is completed.
Figure 3
Figure 4
Figure 5
VASOVIEW 7 | Training Manual | English | v1
Copyright MAQUET Gmbh
| 10
| ENDOSCOPIC VESSEL HARVESTING | MAQUET Medical Systems USA |
INCISION SITE
Selecting the Incision Site
There are several factors to consider when selecting the inci­sion site, such as operator preference, number of grafts need­ed, and whether the upper or lower leg is used for the vein harvest.
There are two commonly used incision sites (Figure 6):
Below the knee: Palpate along the tibia until reaching the medial tibial epicondyle. Make a 2 cm incision along the posterior border of the tibial epicondyle. (Figure 7) Above the knee: Make a 2 cm incision in the thigh where the vein lies in the groove between the sartorius and gracilis muscles. (Figure 8)
Making the Incision
Once the site has been identified, make a 2 cm skin incision directly over the vein. A transverse, longitudinal or oblique incision may be used depending on operator preference. Under direct visualization, dissect the subcutaneous tissue to expose the greater saphenous vein. Continue dissection of the anterior surface of the vein in the direction of planned harvesting, creating a small space approximately the length of the distal end of the BTT to ensure easy insertion.
Figure 6
Figure 7 Figure 8
VASOVIEW 7 | Training Manual | English | v1 Copyright MAQUET Gmbh
INSERTING THE 7 MM ENDOSCOPE WITH ATTACHED
DISSECTION TIP
Under direct visualization, insert the Dissection Tip into the space created, and place it onto the anterior surface of the saphenous vein. Note (on the video monitor) the color of the vein and the surrounding yellow fatty tissue. It is important to always identify the vein prior to advancement of the instrument. (Figure 9) Advance the endoscope approximately 4 cm, then slide the BTT through the incision site and into the previously dissected space. (Figure 10)
| MAQUET Medical Systems USA | ENDOSCOPIC VESSEL HARVESTING |
Figure 9
11 |
Figure 10
VASOVIEW 7 | Training Manual | English | v1
Copyright MAQUET Gmbh
| 12
| ENDOSCOPIC VESSEL HARVESTING | MAQUET Medical Systems USA |
CO2 INSUFFLATION
Connect the CO
insufflation tubing to the clear flexible CO2
2
Insufflation Port on the BTT and begin insufflation of CO Ensure the CO
Insufflator is set to the proper flow and
2
pressure settings.
Flow: 3 – 5 L/min
CO
2
Pressure: 10 – 12 mmHg
CO
2
The goal is to delineate a tissue plane for easier dissection and to maintain the tunnel. If desired pressure is not obtained in the tunnel, the slip tip syringe provided can be used to inflate the balloon on the BTT. Inflate the balloon with the minimal amount of air needed to create an adequate seal (0 – 25 cc’s of air). If there is obvious leakage around the incision, use a purse­string suture to ensure a tight seal around the BTT port. Once the CO
insufflation has begun, note the initial
2
“webbing” effect or partial separation of tissues from the saphenous vein. (Figure 11)
.
2
Figure 11
Technique Tips
If partial separation of tissue from saphenous vein is not observed, check that the CO
is flowing, the pressure
2
setting is 10 – 12 mmHg, and the tubing is connected to the BTT and the insufflator. Also, check the CO
tank to ensure
2
that it is in the open position, with an adequate amount of
. (See Troubleshooting Section)
CO
2
VASOVIEW 7 | Training Manual | English | v1 Copyright MAQUET Gmbh
Loading...
+ 28 hidden pages