1 Before use Before use Before useBefore useefore usefore useore usere usee use useusesee ................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................3
5 Set-ups and preparations Set-ups and preparations Set-ups and preparationsSet-ups and preparationset-ups and preparationst-ups and preparations-ups and preparationsups and preparationsps and preparationss and preparations and preparationsand preparationsnd preparationsd preparations preparationspreparationsreparationseparationsparationsarationsrationsationstionsionsonsnss..........................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................119199
7 Operating your Servo Operating your Servo Operating your ServoOperating your Servoperating your Servoerating your Servorating your Servoating your Servoting your Servoing your Servong your Servog your Servo your Servoyour Servoour Servour Servor Servo ServoServoervorvovoo-ii ....................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................157577
11 Technical data1 Technical data Technical data Technical dataTechnical dataechnical datachnical datahnical datanical dataical datacal dataal datal data datadataatataa ..............................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................241411
12 Abbreviations and definitions2 Abbreviations and definitions Abbreviations and definitions Abbreviations and definitionsAbbreviations and definitionsbbreviations and definitionsbreviations and definitionsreviations and definitionseviations and definitionsviations and definitionsiations and definitionsations and definitionstions and definitionsions and definitionsons and definitionsns and definitionss and definitions and definitionsand definitionsnd definitionsd definitions definitionsdefinitionsefinitionsfinitionsinitionsnitionsitionstionsionsonsnss ......................................................................................................................................................................................................................................................................................................................................................................................................................255555
13 Appendix: User Interface3 Appendix: User Interface Appendix: User Interface Appendix: User InterfaceAppendix: User Interfaceppendix: User Interfacependix: User Interfaceendix: User Interfacendix: User Interfacedix: User Interfaceix: User Interfacex: User Interface: User Interface User InterfaceUser Interfaceser Interfaceer Interfacer Interface InterfaceInterfacenterfaceterfaceerfacerfacefaceacecee......................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................259599
14 Index4 Index Index IndexIndexndexdexexx...................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................273733
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
1
Servo… User´s manual
s
2
InfantAdultUniversalOption
US edition
Order No: 66 00 261
1.. Before useefore usefore useore usere usee use useusesee
Warnings, Cautions and Important in this manualarnings, Cautions and Important in this manualrnings, Cautions and Important in this manualnings, Cautions and Important in this manualings, Cautions and Important in this manualngs, Cautions and Important in this manualgs, Cautions and Important in this manuals, Cautions and Important in this manual, Cautions and Important in this manual Cautions and Important in this manualCautions and Important in this manualautions and Important in this manualutions and Important in this manualtions and Important in this manualions and Important in this manualons and Important in this manualns and Important in this manuals and Important in this manual and Important in this manualand Important in this manualnd Important in this manuald Important in this manual Important in this manualImportant in this manualmportant in this manualportant in this manualortant in this manualrtant in this manualtant in this manualant in this manualnt in this manualt in this manual in this manualin this manualn this manual this manualthis manualhis manualis manuals manual manualmanualanualnualualall . 44
Support material related to the Servoupport material related to the Servopport material related to the Servoport material related to the Servoort material related to the Servort material related to the Servot material related to the Servo material related to the Servomaterial related to the Servoaterial related to the Servoterial related to the Servoerial related to the Servorial related to the Servoial related to the Servoal related to the Servol related to the Servo related to the Servorelated to the Servoelated to the Servolated to the Servoated to the Servoted to the Servoed to the Servod to the Servo to the Servoto the Servoo the Servo the Servothe Servohe Servoe Servo ServoServoervorvovoo-ii systemsystemystemstemtememm. . . . .. . .. 77
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
3
s
Before useefore usefore useore usere usee use useusesee
1
Welcome as a user of the Servo-i Ventilator he Servo-i Ventilator
system! We hope that you will be very We hope that you will be very
satisfied with your new system. For the latest fied with your new system. For the latest
information about it, call your local MAQUET bout it, call your local MAQUET ET
representative. Before use, please read the he
general information below.
The Servo-i Ventilator System consists of a vo-i Ventilator System consists of a
Patient Unit where gases are mixed and Unit where gases are mixed and
administered, and a User Interface where the
settings are made and ventilation is ventilation is
monitored.
The ventilator delivers controlled or
supported breaths to the patient, with either breaths to the patient, with either
constant flow or constant pressure, using a
set oxygen concentration. The entire Servo-i ygen concentration. The entire Servo-i
system includes a wide range of optional ludes a wide range of optional
accessories, e.g. Mobile Cart, breathing bile Cart, breathing
systems, compressors, Battery modules, pressors, Battery modules,
humidifiers and equipment for nebulization, quipment for nebulization,
measurement and Y-piece
COO
2
measurement.
Unit
Intended userntended usertended userended usernded userded usered userd user userusersererr
Servo-i is a ventilator system with advanced dvanced
functionality. It may be used only by
professional health care providers who have rofessional health care providers who have
sufficient experience in ventilator treatment.ment.
Intended populationntended populationtended populationended populationnded populationded populationed populationd population populationpopulationopulationpulationulationlationationtioniononn
The Servo-i Ventilator System can be can be
delivered in three configurations:
• Servo-i Infant range 0,5 - 30kg0,5 - 30kg
NIV (PC+PS) Infant range 3 - 30kg) Infant range 3 - 30kg
NIV Nasal CPAP range 0.5 - 10kgCPAP range 0.5 - 10kg
• Servo-i Adult lt range 10 - 250kg
• Servo-i Universal range 0.5 - 250kg
NIV (PC+PS) Infant range 3 - 30kge 3 - 30kg
NIV Nasal CPAP range 0.5 - 10kgrange 0.5 - 10kg
Note: Servo-i Universal covers both Basic both Basic
and Extended edition.
Warnings, Cautions and arnings, Cautions and rnings, Cautions and nings, Cautions and ings, Cautions and ngs, Cautions and gs, Cautions and s, Cautions and , Cautions and Cautions and Cautions and autions and utions and tions and ions and ons and ns and s and and and nd d
Important in this manualmportant in this manualportant in this manualortant in this manualrtant in this manualtant in this manualant in this manualnt in this manualt in this manual in this manualin this manualn this manual this manualthis manualhis manualis manuals manual manualmanualanualnualualall
WARNING! ARNING! Indicates critical information
about a potential serious outcome to the potential serious outcome to the
patient or the user.
Caution: Indicates instructions that must be must be
followed in order to ensure the proper
operation of the equipment.quipment.
Important:mportant: Indicates information intended to
help you operate the equipment or its
connected devices easily and conveniently.
Intended usentended usetended useended usended useded useed used use useusesee
The Servo-i Ventilator System is intended for he Servo-i Ventilator System is intended for
treatment and monitoring of patients in the g of patients in the
range of neonates, infants and adults with
respiratory failure or respiratory insufficiency. ciency.
Servo-i is a ventilator system to be used only
by health care providers in hospitals or health health care providers in hospitals or health
care facilities and for in-hospital transport.
Note: The Servo-i Ventilator System is not
intended to be used with any anesthetic with any anesthetic
agents.
4
InfantAdultUniversalOption
Servo… User´s manual
US edition
Order No: 66 00 261
Before useefore usefore useore usere usee use useusesee
1
Symbolsymbolsmbolsbolsolslss
User Interfacece
Audio off Silence alarm or confirm
alarm.
Audio pause Silence alarm or
confirm alarm.
Reserved for future use.
Save To save a recording or to copy py
screen.
Attention Consult accompanying
documents.
Standby/Start ventilation Set
standby mode or start ventilation.
Yellow lamp indicating Standby g Standby
mode.
Mains indicatorcator
Green lampp indicating mains
connected.
Battery Symbol indicating battery ry
power supply.The estimated
remaining time with current power ning time with current power
consumption is indicated in minutes.
ON/OFF switch
Patient Unit
CE label The device complies with he device complies with
the requirements of the Medical f the Medical
Device Directive 93/42/EEC.42/EEC.
CSA label The device complies plies
with the Canadian standards.
C US
Class I equipment, Type Buipment, Type B The
device classification according to on according to
according to IEC 60601-1/EN 6060-1.N 6060-1.
Equipotentiality terminalality terminal
Nebulizer
connector for nebulizer.
RS 232 / Serial port232 / Serial port
connector for data communicationmmunication
Note: The symbol has two different different
labels depending on panel version.
User Interface connector / Panelser Interface connector / Panel
Note: The symbol has two different ymbol has two different
labels depending on panel version.g on panel version.
connector for optional equipment.
Note: The symbol has two different
labels depending on panel version.pending on panel version.
Trigger indicationgger indication The indication The indication
appears in the message/alarm field field
when the patient triggers a breath.
The NIV symbol appears in the NIV symbol appears in the Mode
pad field during Non Invasive d field during Non Invasive
Ventilation.
Servo… User´s manual
US edition
Order No: 66 00 261
NIV symbolmbol
10A0A
fuse for external DC power supply.C power supply.
InfantAdultUniversalOptions
5
s
Before useefore usefore useore usere usee use useusesee
1
12V DC / Ext. bat 12Vat 12V
external +12V DC inlet.12V DC inlet.
Note: The symbol has two different bol has two different
labels depending on panel version.
Caution: When external +12 V DC When external +12 V DC C
is used, at least one installed
Battery module is required to ensure
proper operation.
Expiratory labelpiratory label
Gas flow from patient.
Inspiratory label
Gas flow to patient.
Gas exhaust port labelust port label
Exhaust gas flow from ventilator.haust gas flow from ventilator.
Note: Should not be connected to a d not be connected to a
spirometer, as the volume through gh
the exhaust port is not equal to the
expired volume from the patient.me from the patient.
Single use
Special waste
This product contains electronic product contains electronic
and electrical components.
Discard disposable, replaced and
left-over parts in accordance with ver parts in accordance with
appropriate industrial and
environmental standards.
Recycling
Worn-out batteries must be
recycled or disposed of properly posed of properly
in accordance with appropriate
industrial and environmental mental
standards.
Hazardous wastes waste (infectious) The
device contains parts which must vice contains parts which must
not be disposed of with ordinary disposed of with ordinary
waste.
Alarm output connection nnection
option
External alarm output
communication.
In this manualmanual
Adult Information valid for the Adult d for the Adult
configuration
Infant Information valid for the Infant
configuration
Universal (Basic and Extended rsal (Basic and Extended
editions) Information valid for the
Universal configuration.
Options
6
InfantAdultUniversalOption
Servo… User´s manual
US edition
Order No: 66 00 261
Warnings, cautions and importantarnings, cautions and importantrnings, cautions and importantnings, cautions and importantings, cautions and importantngs, cautions and importantgs, cautions and importants, cautions and important, cautions and important cautions and importantcautions and importantautions and importantutions and importanttions and importantions and importantons and importantns and importants and important and importantand importantnd importantd important importantimportantmportantportantortantrtanttantantntt
1
Support material related to upport material related to pport material related to port material related to ort material related to rt material related to t material related to material related to material related to aterial related to terial related to erial related to rial related to ial related to al related to l related to related to related to elated to lated to ated to ted to ed to d to to to o
the Servohe Servoe Servo ServoServoervorvovoo-i systemsystemystemstemtememm
Wall cleaning diagram
Configuration Card
User‘s manual
This concept comprises components concept comprises components
intended to cover the needs of a clinical user.
It is divided into different components
according to use to facilitate accessibility of g to use to facilitate accessibility of
information. If you have any comments or ve any comments or
suggestions regarding this information
material, please let us know.please let us know.
This User´s manual covers functionality and manual covers functionality and
use but should not be regarded as all garded as all
inclusive within the very complex field of mplex field of
ventilatory treatment. Clinical judgements or judgements or
settings are therefore not described in this
manual. Authorized, medically competent Authorized, medically competent
health care providers with good knowledge viders with good knowledge
of Servo-i Ventilator System have the Ventilator System have the
responsibility to determine the clinical bility to determine the clinical
judgement and settings based on the needs settings based on the needs
of the patient.
Read the User´s manual carefully before use ´s manual carefully before use
and follow the instructions.he instructions.
SVX-129_EN
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
7
s
Warnings, cautions and importantarnings, cautions and importantrnings, cautions and importantnings, cautions and importantings, cautions and importantngs, cautions and importantgs, cautions and importants, cautions and important, cautions and important cautions and importantcautions and importantautions and importantutions and importanttions and importantions and importantons and importantns and importants and important and importantand importantnd importantd important importantimportantmportantportantortantrtanttantantntt
1
This User´s manualer´s manual
The information in this User´s manual is valid valid
for Servo-i Ventilator System 3.0 unless 3.0 unless
stated otherwise.
Here you will find the information needed to find the information needed to
use the Servo-i system safely.
It is divided into five main sections: ctions:
• Before use (mandatory information)
• Description
• Operation
• Maintenance
• Miscellaneous
Recommended use
The main document, for every-day use.
Text shown on the User Interface is presented xt shown on the User Interface is presented
in these instructions in a uctions in a special typefacel typeface.
Brief instructionsctions
Overviews and step-by-step instructions for y-step instructions for
the set-ups. These instructions you will find
in the drawer above the ventilator, when awer above the ventilator, when
positioned on the Mobile Cart.Cart.
Recommended usese
These documents are intended to be used as
a guide for the experienced user.guide for the experienced user.
Wall diagramgram
Overviews and step-by-step instructions for verviews and step-by-step instructions for
cleaning, to be posted on a wall.leaning, to be posted on a wall.
Recommended use
Checklist for the experienced user.klist for the experienced user.
Ventilator - Information entilator - Information ntilator - Information tilator - Information ilator - Information lator - Information ator - Information tor - Information or - Information r - Information - Information - Information Information Information nformation formation ormation rmation mation ation tion ion on n
materialaterialterialerialrialialall
Caution: The Servo-i Ventilator System may Servo-i Ventilator System may
have different software versions. ware versions. Before use,
make sure the software version shown on the
screen under the der the Status / General General menu
corresponds to the version number on the
User´s manual.anual. Refer to page 259.
Trademarkrademarkademarkdemarkemarkmarkarkrkk
Trademark ™demark ™ is written only when a product/product/
method name appears for the first time in this
manual.
8
InfantAdultUniversalOption
Servo… User´s manual
US edition
Order No: 66 00 261
Warnings, cautions and importantarnings, cautions and importantrnings, cautions and importantnings, cautions and importantings, cautions and importantngs, cautions and importantgs, cautions and importants, cautions and important, cautions and important cautions and importantcautions and importantautions and importantutions and importanttions and importantions and importantons and importantns and importants and important and importantand importantnd importantd important importantimportantmportantportantortantrtanttantantntt
1
General warningseneral warningsneral warningseral warningsral warningsal warningsl warnings warningswarningsarningsrningsningsingsngsgss
• The Servo-i Ventilator System must be ust be
operated only by authorized personnel
who are well trained in its use. It must be must be
operated according to the instructions in
this User´s manual.User´s manual.
• After unpacking, perform a Routine g, perform a Routine
cleaning and a Pre-use check.se check.
• To provide adequate patient safety, set the nt safety, set the
alarm limits to relevant values.
• To avoid electrical shock hazard, connect ctrical shock hazard, connect
the power cord to a mains outlet equipped utlet equipped
with a protective ground.
• Should any unfamiliar events occur, such ccur, such
as irrelevant pop-up windows on the
screen, unfamiliar sounds, alarms from the miliar sounds, alarms from the
Patient Unit or technical high priority gh priority
alarms, the ventilator should immediately
be checked and, if applicable, replaced.
• Only accessories and auxiliary equipment ories and auxiliary equipment
that meet current IEC standards (e.g. IEC current IEC standards (e.g. IEC
60601-1-1) may be connected to the 1-1-1) may be connected to the
Servo-i Ventilator System. If external vo-i Ventilator System. If external
equipment such as computers, monitors, quipment such as computers, monitors,
humidifiers or printers are connected, the
total system must comply with IEC 60601-ystem must comply with IEC 60601-606011-1.
• The ventilator must only be used in an y be used in an
upright position.
• When a Servo Ultra Nebulizer is used, ulizer is used,
always consult the drug manufacturer manufacturer
regarding the appropriateness of
ultrasonic nebulization for certain bulization for certain
medication.
• All personnel should be aware of the risk sonnel should be aware of the risk
of parts being infected when g infected when
disassembling and cleaning the ventilator.r.
• Service mode may only be used when no when no
patient is connected to the ventilator.
• Positive pressure ventilation can be be
associated with the following adverse
events: barotrauma, hypoventilation, ma, hypoventilation,
hyperventilation or circulatory impairment.
• The Servo-i Ventilator System is verified Servo-i Ventilator System is verified
against and complies with IEC 60601-1-2 gainst and complies with IEC 60601-1-2 C 60601-1-2
regarding electromagnetic compatibility. It magnetic compatibility. It
is the responsibility of the user to take user to take
necessary measures to ensure that the
clinical environment is compatible with the ble with the
limits specified in IEC 60601-1-2. C 60601-1-2.
Exceeding of these limits may impair the these limits may impair the
performance and safety of the system.
Such measures should include, but are not uch measures should include, but are not
limited to:
– Normal precautions with regard to precautions with regard to
relative humidity and conductive midity and conductive
characteristics of clothing in order to
minimize the build-up of electrostatic ze the build-up of electrostatic
charges.
– Avoiding the use of radio-emitting
devices, such as cellular phones and ces, such as cellular phones and
high-frequency apparatus in close tus in close
proximity to the system.
• The Servo-i Ventilator System is not Servo-i Ventilator System is not
intended to be used in MR environment used in MR environment
during MR examinations. This may cause This may cause
deactivation of the system functions and
may result in permanent damage to the permanent damage to the
Servo-i Ventilator System.ystem.
• The Servo-i Ventilator System is not
intended to be used with any anesthetic ded to be used with any anesthetic
agent. To avoid risk of fire, flammable To avoid risk of fire, flammable
agents such as ether and cyclopropane uch as ether and cyclopropane
must not under any circumstances be
used with this device.with this device.
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
9
s
Warnings, cautions and importantarnings, cautions and importantrnings, cautions and importantnings, cautions and importantings, cautions and importantngs, cautions and importantgs, cautions and importants, cautions and important, cautions and important cautions and importantcautions and importantautions and importantutions and importanttions and importantions and importantons and importantns and importants and important and importantand importantnd importantd important importantimportantmportantportantortantrtanttantantntt
1
• To avoid fire hazard, keep all sources of zard, keep all sources of
ignition away from the y from the Servo-i Ventilator
System and oxygen hoses. Do not use xygen hoses. Do not use
oxygen hoses that are worn, frayed, or d, or
contaminated by combustible materials
such as grease or oils. Textiles, oils, and Textiles, oils, and
other combustibles are easily ignited and y ignited and
burn with great intensity in air enriched
with oxygen. Immediately disconnect the xygen. Immediately disconnect the
ventilator from the oxygen supply, facility from the oxygen supply, facility
power, and backup sources if there is a wer, and backup sources if there is a
smell of burning.mell of burning.
General cautionseneral cautionsneral cautionseral cautionsral cautionsal cautionsl cautions cautionscautionsautionsutionstionsionsonsnss
• As a general rule always avoid contact void contact
with external electrical connector pins. It is
recommended to have the module ule
compartment filled up with empty modules s
to protect from spillage and dust.
• Federal law in the USA restricts this device w in the USA restricts this device
to sale by or on the order of a physician (or ale by or on the order of a physician (or
a properly licensed practitioner).
• The Servo-i Ventilator System must be ust be
serviced at regular intervals by specially
trained personnel. The intervals are stated The intervals are stated
in the chapter Regular maintenance. Any maintenance. Any
maintenance must be noted in a log book k
for that purpose in accordance with
national regulations.gulations.
• MAQUET has no responsibility for the safe T has no responsibility for the safe
operation of the equipment if service or ment if service or
repair is done by a non-professional or by
persons who are not employed by or who are not employed by or
authorized by MAQUET. We recommend MAQUET. We recommend mmend
that service is done as part of a service
contract with MAQUET.MAQUET.
• MAQUET has no responsibility for the safe esponsibility for the safe
operation of the equipment if the
equipment is used for anything other than d for anything other than
its intended use, as specified in this User´s User´s
manual.
• A resuscitator should always be readily ways be readily
accessible for extra safety.
• When connected to a patient, the system hen connected to a patient, the system
must never be left unattended. be left unattended.
• The nebulizer module is inoperative when perative when
the ventilator is running on batteries, to
reduce the power consumption.ce the power consumption.
• The Expiratory cassette must not be lifted atory cassette must not be lifted
up when the ventilator is in operation. This when the ventilator is in operation. This
may, however, be done when in Standby be done when in Standby
mode.
• Always use heat and moisture exchanger lways use heat and moisture exchanger ger
(HME) or equipment to prevent t to prevent
dehydration of lung tissue.
• Refer to the Installation instructions to
assemble the system or options to obtain ptions to obtain
a proper mechanical assembly.
• When lifting or moving the ventilator
system or parts of the system, follow parts of the system, follow
established ergonomic guidelines, ask for gonomic guidelines, ask for
assistance, and take appropriate safety y
precautions.
• Antistatic or electrically conductive
breathing tubing should not be used with
this lung ventilator system.m.
• Any scavenging system (Gas evac)
connected must comply to ISO8835-3 onnected must comply to ISO8835-3 8835-3
with regard to subatmospheric pressure mospheric pressure
and induced flow. Otherwise ventilator Otherwise ventilator
functions and patient safety may be patient safety may be
degraded.
• It is not recommended to use the Servo commended to use the Servo
Evac 180 in the Nasal CPAP mode.80 in the Nasal CPAP mode.ode.
• Values measured at the signal outputs of f
the Servo-i Ventilator System and which which
have been processed in auxiliary y
equipment must not be used as a
substitute for therapeutic or diagnostic herapeutic or diagnostic
decisions. Such decisions can be made made
only by staff with medical expertise, pertise,
according to established and accepted
practice. If auxiliary equipment that has xiliary equipment that has
not been delivered by MAQUET with the delivered by MAQUET with the T with the
system is used, MAQUET denies all QUET denies all
responsibility for the accuracy of signal ty for the accuracy of signal
processing.
10
Servo… User´s manual
InfantAdultUniversalOption
US edition
Order No: 66 00 261
Warnings, cautions and importantarnings, cautions and importantrnings, cautions and importantnings, cautions and importantings, cautions and importantngs, cautions and importantgs, cautions and importants, cautions and important, cautions and important cautions and importantcautions and importantautions and importantutions and importanttions and importantions and importantons and importantns and importants and important and importantand importantnd importantd important importantimportantmportantportantortantrtanttantantntt
1
• If there should be any deviation between d be any deviation between
information shown on the User Interface of User Interface of
the ventilator and that shown by the y the
auxiliary equipment, the ventilator
parameters shown on the User Interface wn on the User Interface
shall be considered the primary source for mary source for
information. When combining the g the Servo-i
Ventilator System with accessories and with accessories and
auxiliary equipment other than those pment other than those
recommended by MAQUET, it is the AQUET, it is the
responsibility of the user to ensure the bility of the user to ensure the
integrity of system performance and ce and
safety. In order to maintain electrical
system safety, i.e. such that compliance
with IEC 60601-1-1 is fulfilled, only nly
accessories and auxiliary equipment that hat
meet current IEC standards (e.g. IEC
60601-1, IEC 950) may be connected to connected to
signal inputs and outputs of the Servo-i -i
Ventilator System.
• Only original parts from MAQUET must be parts from MAQUET must be T must be
used in the system.
• Only accessories, supplies or auxiliary ccessories, supplies or auxiliary
equipment recommended by MAQUET by MAQUET
should be used with the ventilator system ld be used with the ventilator system
(“Products and accessories” catalog and
“Spare parts list”). Use of any other ). Use of any other
accessories, spare parts or auxiliary
equipment may cause degraded system quipment may cause degraded system
performance and safety. ormance and safety.
• The displayed information about set and layed information about set and
corresponding measured parameters, g measured parameters,
shall continously be compared by the be compared by the
operator.
Important:
•This symbol on the unit means n the unit means
Attention, consult accompanying
documents.
Note: The are two versions of this symbol ymbol
depending on System version.
• The gases supplied must be free from
water, oil and particles:
Air ................... H
O < 7 g/m
2
........................ Oil < 0.5 mg/m
Oxygen ........... H2O < 20 mg/m20 mg/m
3
3
3
• The environmental declaration is part of declaration is part of
the service manual.
• The Servo-i Ventilator system does not does not
contain any latex.
• Data on pressures can be given in cmHgiven in cmH
where:
1 kPa ~ 10 cmHPa ~ 10 cmH
100 kPa = 1bar ~1atm ~1kgf/cm/cm
O
2
2
(kp/cm2)
100 kPa ~15 psi.
• All disposable parts must be discarded
according to hospital routine and in an hospital routine and in an
environmentally safe way.way.
• Do not expose the Expiratory cassette cassette
compartment to excessive amounts of
fluid, e.g. during cleaning and disinfection,
as this may influence ventilator his may influence ventilator
functionality.
• Do not use sharp tools on the screen.p tools on the screen.
• It is recommended that at least two
batteries always is used in the ventilator for ys is used in the ventilator for
backup.p.
• It is recommended that at least two
batteries are used for ventilation during used for ventilation during
transport.
2
O,
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
11
s
Warnings, cautions and importantarnings, cautions and importantrnings, cautions and importantnings, cautions and importantings, cautions and importantngs, cautions and importantgs, cautions and importants, cautions and important, cautions and important cautions and importantcautions and importantautions and importantutions and importanttions and importantions and importantons and importantns and importants and important and importantand importantnd importantd important importantimportantmportantportantortantrtanttantantntt
1
• Documentation for Servo-i Ventilator
System stem consists of:
–User´s manual
– Brief instructions
– Wall diagramdiagram
– Installation instructions
– Service manual
– Products and accessories, catalog, catalog
– Spare parts list
Context-related warnings ontext-related warnings ntext-related warnings text-related warnings ext-related warnings xt-related warnings t-related warnings -related warnings related warnings elated warnings lated warnings ated warnings ted warnings ed warnings d warnings warnings warnings arnings rnings nings ings ngs gs s
Note: General warnings are not listed here here
even though they are repeated inside the
manual.
Note: Context-related Cautions and
"Important" are not listed here, but are but are
written in the relevant context inside the
manual.
Operationperationerationrationationtioniononn
• Always disconnect the ventilator if any ventilator if any
operation which may involve risk for the k for the
patient will be done, e.g. replacement of ment of
cell, dismantling etc. (page 211, 211,
O
2
page 225).
• If the trigger sensitivity is set too high, a y is set too high, a
self-triggering (auto-triggering) condition
may be reached. This condition can also
be reached if there is leakage in the kage in the
breathing system, e.g. if an uncuffed d
endotracheal tube is used. Triggering will
then be initiated by the system and not by y the system and not by
the patient.This should always be avoided ways be avoided
by decreasing the trigger sensitivity
(page 23). This is also important during page 23). This is also important during during
transport as the movement of the body y
and the breathing system may lead to
false triggering.
• When you turn a Direct Access Knob, Knob,
ventilation will change accordingly from y from
the next breath without additional
confirmation (For further information see
page 166).166).
• If any malfunctions are detected during the detected during the
start-up procedure, please refer to
Chapter, Troubleshooting (page 225).Troubleshooting (page 225).5).
• If a malfunction persists, the ventilator ventilator
may not be connected to the patient.
• A Pre-use check must always be done be done
before connecting the ventilator to a
patient (page 145).45).
• To protect the patient against high airway way
pressures, the upper pressure limit must
always be set to the relevant value so as to
provide adequate patient safety (page de adequate patient safety (page
165). 5).
Caution: If airway pressure rises 6 cmHH
above the set upper pressure limit the
safety valve opens. The safety valve also fety valve opens. The safety valve also
opens if system pressure exceeds 117 xceeds 117
cmH
O.
2
• To provide adequate patient safety, always rovide adequate patient safety, always
set the alarm limits at relevant values he alarm limits at relevant values
(page 165).).
O
2
± 7
12
Servo… User´s manual
InfantAdultUniversalOption
US edition
Order No: 66 00 261
Warnings, cautions and importantarnings, cautions and importantrnings, cautions and importantnings, cautions and importantings, cautions and importantngs, cautions and importantgs, cautions and importants, cautions and important, cautions and important cautions and importantcautions and importantautions and importantutions and importanttions and importantions and importantons and importantns and importants and important and importantand importantnd importantd important importantimportantmportantportantortantrtanttantantntt
• Servo Humidifier/HME must be ME must be
disconnected during nebulization ation
otherwise the humidifier may be blocked cked
(page 128).
• The heated humidifier must be switched ted humidifier must be switched
off during nebulization. Otherwise the zation. Otherwise the
particle size may be affected (page 128).y be affected (page 128).
• During nebulization a filter must be
connected to the expiratory inlet of the d to the expiratory inlet of the
ventilator. Always carefully monitor the s carefully monitor the
airway pressure during nebulization. during nebulization.
Increased airway pressure could be pressure could be
caused by a clogged filter. The filter should he filter should
be replaced if the expiratory resistance
increases or every 24 hours when the very 24 hours when the
nebulizer is being used.g used.
• When a Servo Ultra Nebulizer is used, Ultra Nebulizer is used,
always consult the drug manufacturer onsult the drug manufacturer
regarding the appropriateness of
ultrasonic nebulization for certain bulization for certain
medications (page 128, 187).ge 128, 187).
• The nebulizer must not be used without must not be used without
buffer liquid (sterile water). Otherwise the ater). Otherwise the
ultrasonic generator crystal may break may break
(page 129, 187).7).
• To avoid explosion hazards, flammable flammable
agents such as ether and cyclopropane propane
must not be used with this device. Only ce. Only
agents which comply with the ply with the
requirements on non-flammable agents in ble agents in
the IEC standard “Particular requirements ular requirements
for electrical safety of anaesthetic
machines” are suitable.
• For adult/pediatric patients, never fill the
medication cup with more than 10 ml
(page 129).
• For neonatal patients, never fill the ver fill the
medication cup with more than 4 ml 4 ml
(page 129).
• If the patient unit of the nebulizer is tilted, nt unit of the nebulizer is tilted,
the drug can flow into the patient´s lungs patient´s lungs
or the ventilator.
• The nebulizer must not be left unattended must not be left unattended
when connected to a patient.
• Continuously check that the buffer liquid quid
level is between MIN. and MAX. during AX. during
nebulization (page 187).page 187).
• During nebulization: Continuously check on: Continuously check
that moisture is generated in the d in the
medication cup (page 187).7).
• When the ventilator is running on batteries batteries
the nebulizer module is inoperative, to
reduce the power consumption ce the power consumption
(page 187).87).
• For information about the stand alone stand alone
Aeroneb Professional Nebulizer System, System,
refer to separate manual.
Cleaningleaningeaninganingningingngg
• All personnel should be aware of the risk hould be aware of the risk
of parts being infected when g infected when
disassembling and cleaning the ventilator r
(page 191).
• After removing the Expiratory cassette, do ng the Expiratory cassette, do
not pour any fluid into the Expiratory ur any fluid into the Expiratory
cassette compartment (page 196).mpartment (page 196).
Replacement of Oeplacement of Oplacement of Olacement of Oacement of Ocement of Oement of Oment of Oent of Ont of Ot of O of Oof Of O OO2 cellcellelllll
The sealed unit of the O2 cell, contains a led unit of the O2 cell, contains a
caustic liquid which may cause severe burns which may cause severe burns
to the skin and eyes. In case of contact, kin and eyes. In case of contact,
immediately flush continuously with water for with water for
at least 15 minutes and seek medical cal
attention especially if the eyes are affected
(page 214)14)
Servo… User´s manual
US edition
Order No: 66 00 261
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
15
s
Modes of ventilationodes of ventilationdes of ventilationes of ventilations of ventilation of ventilationof ventilationf ventilation ventilationventilationentilationntilationtilationilationlationationtioniononn
The Servo-i Ventilator System is designed for lator System is designed for
safe and effective treatment. It can be set for be set for
continuous adaptation to the patient´s
prevailing condition or for manually manually
controlled operations. The servo systems for vo systems for
pressure, flow and timing operate in all g operate in all
modes of ventilation (set time in control
modes and patient-related timing in support
modes).).
Important:
• To show all available installed ventilation
modes, please refer to "Setting ventilation please refer to "Setting ventilation
mode" on page 164 in this manual.64 in this manual.
• In all pressure controlled modes, it is d modes, it is
important to set alarm limits to adequate
levels.
• For information about default values and bout default values and
parameter settings refer to page 249.249.
Applicationon
The Servo-i ventilator system also contains ystem also contains
tools to assist the user in application of lung
recruitment methodologies.dologies.
3. spontaneous breathing/CPAPbreathing/CPAP
4-7. It also allows for combined ventilatory It also allows for combined ventilatory
control or support. Spontaneous breathing upport. Spontaneous breathing
efforts are sensed during controlled controlled
ventilation, e.g. Volume Control. Mandatory Mandatory
ventilation can be used during supported/pported/
spontaneous breathing, e.g. the enhanced
SIMV functionality.
8. The Automode functionality continuously he Automode functionality continuously
adapts to the patient´s breathing capability. pts to the patient´s breathing capability.
When required, all ventilation is provided for quired, all ventilation is provided for
mandatorily. When the patient is able to When the patient is able to
initiate a breath, the ventilator supports and
monitors the patient´s breathing capability
and controls ventilation only if required.d controls ventilation only if required.
16
Servo… User´s manual
US edition
InfantAdultUniversalOption
Order No: 66 00 261
Modes of ventilationodes of ventilationdes of ventilationes of ventilations of ventilation of ventilationof ventilationf ventilation ventilationventilationentilationntilationtilationilationlationationtioniononn
Ventilation can be managed and
administered with a focus on:h a focus on:
A. pressure and volume
B. pressure
C. flow/volume.
Extra flow and extra breathsra flow and extra breaths
In flow/volume- oriented modes of d modes of
ventilation, additional on-demand flow can
be triggered during inspiration. Additional uring inspiration. Additional
breaths can always be triggered between the ways be triggered between the
ordinary breaths if the set trigger criteria are
met.
Timingming
In controlled ventilation modes, timing is
related to preset values. In supported
ventilation modes, timing is related to patient
triggering and Inspiratory cycle-off setting. ff setting.
Pressure and volume in focusd volume in focus
In the pressure- he pressure- and flow- oriented modes, a
constant inspiratory Tidal Volume is y Tidal Volume is
maintained. The inspiratory pressure level is ratory pressure level is
constant during each breath. (PRVC, Volume breath. (PRVC, Volume
Support.)
Pressure in focus
In the pressure-oriented modes, a constant
preset pressure level is maintained during vel is maintained during
inspiration. (Pressure Control, Pressure Control, Pressure
Support)
Flow/volume in focusw/volume in focus
In the flow/volume oriented modes a w/volume oriented modes a
constant inspiratory volume is maintained. y volume is maintained.
The inspiratory flow is constant during each flow is constant during each
breath (Volume Control).Volume Control).
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
17
s
Modes of ventilationodes of ventilationdes of ventilationes of ventilations of ventilation of ventilationof ventilationf ventilation ventilationventilationentilationntilationtilationilationlationationtioniononn
2
Basic functionality - An overview asic functionality - An overview sic functionality - An overview ic functionality - An overview c functionality - An overview functionality - An overview functionality - An overview unctionality - An overview nctionality - An overview ctionality - An overview tionality - An overview ionality - An overview onality - An overview nality - An overview ality - An overview lity - An overview ity - An overview ty - An overview y - An overview - An overview - An overview An overview An overview n overview overview overview verview erview rview view iew ew w
18
Servo… User´s manual
US edition
InfantAdultUniversalOption
Order No: 66 00 261
1.(PRVC) Pressure Regulated ressure Regulated
Volume Control
Breaths are delivered mandatorily to assure mandatorily to assure
preset volumes, with a constant inspiratory constant inspiratory
pressure continuously adapting to the
patient´s condition. The flow pattern is flow pattern is
decelerating.
2. Volume Control trol
Breaths are delivered mandatorily with a vered mandatorily with a
constant flow to assure preset volumes. preset volumes.
3. Volume Support pport
A patient-adapted constant inspiratory
support is supplied when activated by when activated by
patient effort. The resulting volume is The resulting volume is
continuously monitored and the constant and the constant
inspiratory pressure automatically adjusts to
the required level. The patient determines quired level. The patient determines
frequency and duration of the breaths which cy and duration of the breaths which
show a decelerating flow pattern.g flow pattern.
When sufficient inspiratory volumes are y volumes are
achieved, spontaneous breathing without breathing without
ventilator support is allowed for in Volume for in Volume
Support.
5. Pressure Control ontrol
Breaths are delivered mandatorily at a preset s are delivered mandatorily at a preset
pressure level, causing a decelerating flow flow
pattern.
6. Pressure Support Support
Inspiration is supported by a constant preset
pressure when activated by patient effort. hen activated by patient effort.
The patient determines frequency and mines frequency and
duration of the breaths, which show a , which show a
decelerating flow pattern. Inspiratory breath breath
duration can be influenced by adjusting the
Inspiratory cycle-off criteria.piratory cycle-off criteria.
7. Spontaneous breathing/CPAPs breathing/CPAPCPAP
True spontaneous breathing (CPAP) occurs g (CPAP) occurs
when the inspiratory pressure level is set to the inspiratory pressure level is set to
zero in Pressure Support.
8. Nasal CPAP
Spontaneous breathing on a set pressure breathing on a set pressure
level.
Modes of ventilationodes of ventilationdes of ventilationes of ventilations of ventilation of ventilationof ventilationf ventilation ventilationventilationentilationntilationtilationilationlationationtioniononn
2
Servo… User´s manual
US edition
Order No: 66 00 261
InfantAdultUniversalOptions
19
s
Modes of ventilationodes of ventilationdes of ventilationes of ventilations of ventilation of ventilationof ventilationf ventilation ventilationventilationentilationntilationtilationilationlationationtioniononn
2
Combined modes - An ombined modes - An mbined modes - An bined modes - An ined modes - An ned modes - An ed modes - An d modes - An modes - An modes - An odes - An des - An es - An s - An - An - An An An n
overviewverviewerviewrviewviewieweww
Automodeutomode
The ventilator continuously adapts to the
patient's breathing capability and allows the breathing capability and allows the
patient to better interact with the ventilator. ventilator.
The ventilator automatically shifts between between
controlled ventilation, supported ventilation
and spontaneous ventilation. Each h
controlled ventilation mode has a
corresponding support mode.
Volume Control<----> Volume Support
PRVCRVC<----> Volume Support
Pressure Control <----> Pressure Support
When the patient is making a breathing
effort, the ventilator immediately switches to ort, the ventilator immediately switches to
a support mode of ventilation. If the patient is pport mode of ventilation. If the patient is
not making any breathing effort, the y breathing effort, the
ventilator will return to the controlled mode d mode
and deliver controlled breaths.
The ventilator provides mandatory breaths y breaths
which are synchronized with the patient´s
spontaneous efforts at a preset rate. The us efforts at a preset rate. The
mandatory breaths can be Volume Control, breaths can be Volume Control,
Pressure Control or PRVC breaths.
Bi-Vent
Bi-Vent is pressure controlled breathing, g,
giving the patient the opportunity of
unrestricted spontaneous breathing. Two breathing. Two
pressure levels are set together with the
individually set duration of each level. dividually set duration of each level.
Spontaneous efforts can be assisted by pontaneous efforts can be assisted by
pressure support.
The graphic display of flow, pressure and ure and
volume is visualized in wave forms. Modes of
ventilation directly affect flow, pressure and ctly affect flow, pressure and
volume patterns.
Volume Contrololume Controllume Controlume Controlme Controle Control ControlControlontrolntroltrolrololl
Pressure-Time waveform. Points ure-Time waveform. Points orm. Points
and regions of interestegions of interest
X. Inspiration time
Y. Pause time
Z. Expiration timeExpiration time
1. Start of Inspiration
2. Peak inspiratory pressure
3. Early inspiratory pause pressure
4. End inspiratory pause pressured inspiratory pause pressure
5. Early expiratory pressure
6. End expiratory pressure
2
4
3
1
12
5
7
8
14
13
6
t
11
t
10
9
15
16
t
Flow-Time waveform. Points and w-Time waveform. Points and Points and
regions of interestf interest
X. Inspiration timeX. Inspiration time
Y. Pause time
Z: Expiration timeExpiration time
This determines the level of patient effort to
trigger the ventilator to inspiration.gger the ventilator to inspiration.
Trigger sensitivity can be set in flow flow
triggering (Trigg. Flow) or pressure triggering
(Trigg. PressurePressure). Normally flow triggering is y flow triggering is
preferable as this enables the patient to
breath with less effort.
The sensitivity is set as high as possible vity is set as high as possible
without self-triggering. This ensures that ng. This ensures that
triggering is patient initiated and avoids autocycling by the ventilator.ycling by the ventilator.
Pressure triggering can be set in the range
-20 to 0 cmH20 to 0 cmH
level, white area on the bar).
O (in reference to set PEEP EEP
2
When the trigger sensitivity is set above 0 gger sensitivity is set above 0
(green and red area on the bar), flow flow
triggering is set, i.e. the amount of the bias
flow that the patient has to inhale to trigg a
new breath. The sensitivity can be set from The sensitivity can be set from
100% of the bias flow (left), to 0% of the bias 00% of the bias flow (left), to 0% of the bias
flow (right). For information about the ght). For information about the
different colors of the bar refer to page 167.lors of the bar refer to page 167.
Important: In
NIV it is not possible to set
trigger sensitivity.
The ventilator continuously delivers a gas he ventilator continuously delivers a gas
flow during expiration, which is measured in xpiration, which is measured in
the expiratory channel.y channel.
1. At a Trigger sensitivityer sensitivity level above zero
(0), the ventilator senses deviations in
the bias flow caused by inspiratory flow caused by inspiratory
efforts of the patient. The more to the he patient. The more to the
right on the scale, the more sensitive is
the trigger function.unction.
2. Weak inspiratory effort.
3. Very weak inspiratory effort.
For further information see page 167.urther information see page 167.
WARNING! NING! If the trigger sensitivity is set too vity is set too
high, a self triggering (auto-triggering)
condition may be reached. This condition dition may be reached. This condition
can also be reached if there is leakage in the kage in the
breathing system, e.g. if an uncuffed d
endotracheal tube is used. Triggering will
then be initiated by the system and not by the y the system and not by the
patient.This should always be avoided by ways be avoided by
decreasing the trigger sensitivity.
Stronger patient effortger patient effort
Trigg. Pressure
-2
SVX-142_EN
1. At a Trigger sensitivity level below zero zero
(0), the ventilator senses negative
pressures created by the patient. created by the patient.
Required preset negative pressure to gative pressure to
initiate a breath is shown numerically. cally.
The more to the left on the scale, the
more effort is required to trigger.d to trigger.
2. Stronger patient effort.
For further information see page 167.mation see page 167.
WARNING! The trigger sensitivity bar has vity bar has
different colors based on the setting. A green A green
bar indicates a normal setting for the flow w
triggering. The risk of self-triggering
increases when the bar is red. A white bar d. A white bar
indicates that pressure triggering is required.quired.
Time to peak inspiratory flow or pressure at nspiratory flow or pressure at
the start of each breath as a percentage of ge of
the respiratory cycle time or in seconds.
Increased rise time will affect the rate of flow/will affect the rate of flow/
pressure increase and can be evaluated by valuated by
the shape of the flow and pressure
waveforms.ms.
Inspiratory rise time (%) is applicable in ble in
Pressure Control, Volume Control, PRVC,
SIMV-Volume Control, SIMV-Pressure
Control, SIMV-PRVC. Setting can be in the RVC. Setting can be in the
range 0-20% of the respiratory cycle time.0-20% of the respiratory cycle time.me.
Inspiratory rise time set in seconds is
applicable in Pressure Support, Volume ure Support, Volume
Support and Bi-Vent. For adults the range is For adults the range is
0-0.4 seconds and for infants the range is 0-for infants the range is 0-
0.2 seconds.
Note: When the ventilator is configured for
setting of Inspiration time, the unit for piration time, the unit for
Inspiratory rise time then automatically
switches to seconds for all ventilation ds for all ventilation
modes.
Inspiratory Cycle-off is the point at which
inspiration changes to expiration in ges to expiration in
spontaneous and supported modes of
ventilation. A decrease of the inspiratory flow A decrease of the inspiratory flow
to a preset level causes the ventilator to vel causes the ventilator to
switch to expiration. This preset level is
measured as a percentage of the maximum
flow during inspiration. The range of w during inspiration. The range of
Inspiratory cycle-off is 1 - 70%. - 70%.
Note: In NIV the range is 10-70%.
Normally in supported modes the Inspiratory he Inspiratory
rise time should be increased from the from the
default setting and so give more comfort to
the patient.
Servo… User´s manual
US edition
Order No: 66 00 261
This is the length of the breath i.e. the total
cycle time of the mandatory breath in SIMV ycle time of the mandatory breath in SIMV MV
(inspiration, pause plus expiration). This is his is
set in seconds within the range:
Infants: 0,5 -15 seconds in half second steps.15 seconds in half second steps.
Adults: 1-15 seconds in one second steps.
Note: The soft key Breath cycle time is not
shown when an SIMV mode is selected and wn when an SIMV mode is selected and cted and
inspiration time is configured. Refer to
heading I:E ratio / Inspiration times.
Trigger Timeout is the maximum allowed
apnea time in Automode before controlled pnea time in Automode before controlled
ventilation is activated. It is applicable in:s activated. It is applicable in:
Automode:
Volume Control
PRVC
Pressure Controlssure Control
The settings are within the ranges:are within the ranges:
• Infant: 3-7 seconds
• Adult: 7-12 seconds
Initially the ventilator adapts with a dynamic he ventilator adapts with a dynamic
Trigger Timeout limit. This means that for the
spontaneously triggering patient the timeout
increases successively during the first ten creases successively during the first ten
breaths.
<--->Volume Support
<--->Volume Support
<--->Pressure Support
PEEPEEPEPP
PEEP
SVX-646_EN
Positive End Expiratory Pressure (PEEP) can ) can
be set in the range of 0 - 50 cmHH
Positive End Expiratory Pressure is piratory Pressure is
maintained in the alveoli and may prevent the
collapse of the airways.
Note: In NIV the range is 2-20 V the range is 2-20 cmH
O. A
2
2
O.
26
Servo… User´s manual
US edition
InfantAdultUniversalOption
Order No: 66 00 261
Settingsettingsttingstingsingsngsgss
2
I:E ratio / Inspiration time:E ratio / Inspiration timeE ratio / Inspiration time ratio / Inspiration timeratio / Inspiration timeatio / Inspiration timetio / Inspiration timeio / Inspiration timeo / Inspiration time / Inspiration time/ Inspiration time Inspiration timeInspiration timenspiration timespiration timepiration timeiration timeration timeation timetion timeion timeon timen time timetimeimemee
The setting of breathing parameters in parameters in
Servo-i can be configured in two different ured in two different
ways, based on:
• I:E ratio (independent of changes of e.g. pendent of changes of e.g.
the breathing frequency) or,y) or,
• Inspiration time in seconds (independent
of changes of e.g. the breathing hanges of e.g. the breathing
frequency), to better meet the meet the
requirements for infant care.
When the ventilator is configured for setting gured for setting
of Inspiration time, the unit for Pause time
and Insp. rise time then automatically cally
switches to seconds. The resulting I:E ratio :E ratio
for each setting is shown in the upper right
information area of the ventilation mode mation area of the ventilation mode
window.
As the inspiration time is explicitly set, a xplicitly set, a
change of for example the Respiratory Rate Respiratory Rate
will affect the I:E ratio. As a safety precaution, As a safety precaution,
it will therefore be indicated when the be indicated when the
resulting I:E ratio passes 1:1 in either passes 1:1 in either
direction.
Note: The soft key Breath cycle time is not y Breath cycle time is not
shown when an SIMV mode is selected, MV mode is selected,
since there is no need to set Breath cycle ycle
time when Inspiration time is directly set.
Note: The configuration is done by a service y a service
technician with a service card.
PC (Pressure Control level) above PEEP is C (Pressure Control level) above PEEP is
the set inspiratory pressure level for each he set inspiratory pressure level for each
mandatory breath in Pressure Control and breath in Pressure Control and
SIMV (PC) + PS, and also for Apnea back-up MV (PC) + PS, and also for Apnea back-up for Apnea back-up
in Pressure Support.Support.
PS (Pressure Support level) above PEEP is
the set inspiratory pressure support level for he set inspiratory pressure support level for
triggered breaths in Pressure Support, SIMV breaths in Pressure Support, SIMV V
modes and Bi-Vent.
The setting range for the gas mixer is 21% Ofor the gas mixer is 21% OO2
to 100% O
automatically set at approximately 6% Outomatically set at approximately 6% OO
above or below the set concentration value.
There is also an absolute minimum alarm minimum alarm
limit of 18% O
operating settings.
Respiratory rate is the number of controlled umber of controlled
mandatory breaths per minute in controlled inute in controlled
modes excluding SIMV. The respiratory rate V. The respiratory rate
is also used for calculation of tidal volume if n of tidal volume if
the ventilator is configured for Minute volume Minute volume
setting. SIMV rate is the number of controlled he number of controlled
mandatory breaths in SIMV modes.SIMV modes.
Servo… User´s manual
US edition
Order No: 66 00 261
1. Time when previous mode was when previous mode was
inactivated.
2. Press the pad Show previous mode to
recall the previous accepted ventilation
mode.
3. Activate the previous used ventilation ntilation
mode settings by pressing the Accept
pad.
Note:
• The previous ventilation mode function is
not available after a Pre-use check, ble after a Pre-use check,
changing of patient category, admitting a patient category, admitting a
new patient, use of the same ventilation f the same ventilation
mode for more than 24 hours or after startup (cold start) of the system.p (cold start) of the system.
• In backup ventilation, the ventilator shows n backup ventilation, the ventilator shows
the settings for the supported mode when gs for the supported mode when
previous mode is activated.ctivated.
• A recall of previous settings is only y
possible after a change of ventilation
mode.
28
InfantAdultUniversalOption
Servo… User´s manual
US edition
Order No: 66 00 261
Loading...
+ 253 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.