Maquet SERVO-i User manual V3.1

User’s manual (US Version)
VENTILATOR SYSTEM SERVO-i V3.1
Contents
2 Ventilation..............................................................15
3 Patient safety.........................................................71
4 Device description.................................................81
5 Set-ups and preparations....................................119
6 Pre-use check .....................................................145
7 Operating your Servo-i ........................................157
8 Routine cleaning..................................................191
9 Maintenance........................................................211
10 Troubleshooting.................................................225
11 Technical data ...................................................241
12 Abbreviations and definitions ............................255
13 Appendix: User Interface...................................259
14 Index..................................................................273
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
1
Servo User´s manual
s
2
Infant Adult Universal Option
US edition Order No: 66 00 261
1. Before use
Contents
Brief device description . . . . . . . . . . . . . . . . . . . . . 4
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intended user . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warnings, Cautions and Important in this manual. 4
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Support material related to the Servo-i system . . . 7
General warnings . . . . . . . . . . . . . . . . . . . . . . . . . . 9
General cautions . . . . . . . . . . . . . . . . . . . . . . . . . 10
Context-related warnings. . . . . . . . . . . . . . . . . . . 12
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
3
s
Before use
1
Welcome as a user of the Servo-i Ventilator system! We hope that you will be very satisfied with your new system. For the latest information about it, call your local MAQUET representative. Before use, please read the general information below.
Brief device description
User Interface
Patient Patient breathing system
SVX-128_EN
The Servo-i Ventilator System consists of a Patient Unit where gases are mixed and administered, and a User Interface where the settings are made and ventilation is monitored.
The ventilator delivers controlled or supported breaths to the patient, with either constant flow or constant pressure, using a set oxygen concentration. The entire Servo-i system includes a wide range of optional accessories, e.g. Mobile Cart, breathing systems, compressors, Battery modules, humidifiers and equipment for nebulization, CO
measurement and Y-piece
2
measurement.
Unit
Intended use
The Servo-i Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency.
Note: The Servo-i Ventilator System is not intended to be used with any anesthetic agents.
professional health care providers who have sufficient experience in ventilator treatment.
Intended population
The Servo-i Ventilator System can be delivered in three configurations:
Servo-i Infant range 0,5 - 30kg
NIV (PC+PS) Infant range 3 - 30kg NIV Nasal CPAP range 0.5 - 10kg
Servo-i Adult range 10 - 250kg
Servo-i Universal range 0.5 - 250kg
NIV (PC+PS) Infant range 3 - 30kg NIV Nasal CPAP range 0.5 - 10kg
Note: Servo-i Universal covers both Basic and Extended edition.
Intended Use Environment
The SERVO-i ventilator system should be used:
• in hospitals
• in facilities whose primary purpose is to provide health care
• for in-hospital transport
• for interhospital transport if the conditions stated in the Servo-i Interhospital Transport declaration are fulfilled and an agreement with MAQUET is signed.
• during MR examinations of patients if the conditions in the Servo-i MR Environment declaration are met and an agreement with Maquet signed.
Warnings, Cautions and Important in this manual
WARNING! Indicates critical information
about a potential serious outcome to the patient or the user.
Caution: Indicates instructions that must be followed in order to ensure the proper operation of the equipment.
Intended user
Servo-i is a ventilator system with advanced functionality. It may be used only by
4
Infant Adult Universal Option
Important: Indicates information intended to help you operate the equipment or its connected devices easily and conveniently.
Servo User´s manual US edition Order No: 66 00 261
Before use
1
Symbols
User Interface
Audio off Silence alarm or confirm
alarm.
Alarm off .
Audio pause Silence alarm or
confirm alarm.
Reserved for future use.
Save To save a recording or to copy screen.
Attention Consult accompanying documents.
Standby/Start ventilation Set standby mode or start ventilation.
Yellow lamp indicating Standby mode.
Patient Unit
CE label The device complies with
the requirements of the Medical Device Directive 93/42/EEC.
CSA label The device complies with the Canadian standards.
C US
Class I equipment, Type B The device classification according to according to IEC 60601-1/EN 6060-1.
Equipotentiality terminal
Nebulizer
connector for nebulizer.
RS 232 / Serial port
connector for data communication Note: The symbol has two different
labels depending on panel version.
Mains indicator Green lamp indicating mains
connected.
Battery Symbol indicating battery power supply.The estimated remaining time with current power consumption is indicated in minutes.
Trigger indication The indication appears in the message/alarm field when the patient triggers a breath.
The NIV symbol appears in the Mode pad field during Non Invasive Ventilation.
Servo User´s manual US edition Order No: 66 00 261
ON/OFF switch
NIV symbol
User Interface connector / Panel Note: The symbol has two different
labels depending on panel version.
Optional connector / Expansion
connector for optional equipment. Note: The symbol has two different labels depending on panel version.
10A
fuse for external DC power supply.
Infant Adult Universal Options
5
s
Before use
1
12V DC / Ext. bat 12V
external +12V DC inlet. Note: The symbol has two different labels depending on panel version.
Caution: When external +12 V DC is used, at least one installed Battery module is required to ensure proper operation.
Expiratory label
Gas flow from patient.
Inspiratory label
Gas flow to patient.
Gas exhaust port label
Exhaust gas flow from ventilator. Note: Should not be connected to a
spirometer, as the volume through the exhaust port is not equal to the expired volume from the patient.
Single use
Special waste
This product contains electronic and electrical components. Discard disposable, replaced and left-over parts in accordance with appropriate industrial and environmental standards.
Recycling
Worn-out batteries must be recycled or disposed of properly in accordance with appropriate industrial and environmental standards.
Hazardous waste (infectious) The device contains parts which must not be disposed of with ordinary waste.
Alarm output connection
option
External alarm output communication.
In this manual
Adult Information valid for the Adult
configuration
Infant Information valid for the Infant configuration
Universal (Basic and Extended editions) Information valid for the
Universal configuration.
Options
6
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Warnings, cautions and important
1
Support material related to the Servo-i system
Wall cleaning diagram
Configuration Card
User‘s manual
This concept comprises components intended to cover the needs of a clinical user. It is divided into different components according to use to facilitate accessibility of information. If you have any comments or suggestions regarding this information material, please let us know.
This User´s manual covers functionality and use but should not be regarded as all inclusive within the very complex field of ventilatory treatment. Clinical judgements or settings are therefore not described in this manual. Authorized, medically competent health care providers with good knowledge of Servo-i Ventilator System have the responsibility to determine the clinical judgement and settings based on the needs of the patient.
Read the User´s manual carefully before use and follow the instructions.
SVX-129_EN
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
7
s
Warnings, cautions and important
1
This User´s manual
The information in this User´s manual is valid for Servo-i Ventilator System 3.1 unless stated otherwise.
Here you will find the information needed to use the Servo-i system safely.
It is divided into five main sections:
• Before use (mandatory information)
•Description
• Operation
• Maintenance
• Miscellaneous
Recommended use
The main document, for every-day use.
Text shown on the User Interface is presented in these instructions in a special typeface.
Brief instructions
Overviews and step-by-step instructions for the set-ups. These instructions you will find in the drawer above the ventilator, when positioned on the Mobile Cart.
Recommended use These documents are intended to be used as
a guide for the experienced user.
Wall diagram
Overviews and step-by-step instructions for cleaning, to be posted on a wall.
Recommended use Checklist for the experienced user.
Ventilator - Information material
Caution: The Servo-i Ventilator System may
have different software versions. Before use, make sure the software version shown on the screen under the Status / General menu corresponds to the version number on the User´s manual. Refer to page 259.
Trademark
Trademark ™ is written only when a product/ method name appears for the first time in this manual.
8
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Warnings, cautions and important
1
General warnings
• The Servo-i Ventilator System must be operated only by authorized personnel who are well trained in its use. It must be operated according to the instructions in this User´s manual.
• After unpacking, perform a Routine cleaning and a Pre-use check.
• To provide adequate patient safety, set the alarm limits to relevant values.
• To avoid electrical shock hazard, connect the power cord to a mains outlet equipped with a protective ground.
• Should any unfamiliar events occur, such as irrelevant pop-up windows on the screen, unfamiliar sounds, alarms from the Patient Unit or technical high priority alarms, the ventilator should immediately be checked and, if applicable, replaced.
• Only accessories and auxiliary equipment that meet current IEC standards (e.g. IEC 60601-1-1) may be connected to the Servo-i Ventilator System. If external equipment such as computers, monitors, humidifiers or printers are connected, the total system must comply with IEC 60601­1-1.
• The ventilator must only be used in an upright position.
• When a Servo Ultra Nebulizer is used, always consult the drug manufacturer regarding the appropriateness of ultrasonic nebulization for certain medication.
• All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
• Service mode may only be used when no patient is connected to the ventilator.
• Positive pressure ventilation can be associated with the following adverse events: barotrauma, hypoventilation, hyperventilation or circulatory impairment.
• The Servo-i Ventilator System is verified against and complies with IEC 60601-1-2 regarding electromagnetic compatibility. It is the responsibility of the user to take necessary measures to ensure that the clinical environment is compatible with the limits specified in IEC 60601-1-2. Exceeding of these limits may impair the performance and safety of the system. Such measures should include, but are not limited to:
– Normal precautions with regard to
relative humidity and conductive characteristics of clothing in order to minimize the build-up of electrostatic charges.
– Avoiding the use of radio-emitting
devices, such as cellular phones and high-frequency apparatus in close proximity to the system.
• The SERVO-i Ventilator System may only be used during MR examinations if the conditions in the MR Environment Declaration are met and an agreement with Maquet is signed. Disregard of these conditions may cause deactivation of the system functions and may result in permanent damage to the SERVO-i Ventilator System.
• The Servo-i Ventilator System is not intended to be used with any anesthetic agent. To avoid risk of fire, flammable agents such as ether and cyclopropane must not under any circumstances be used with this device.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
9
s
Warnings, cautions and important
1
• To avoid fire hazard, keep all sources of ignition away from the Servo-i Ventilator System and oxygen hoses. Do not use oxygen hoses that are worn, frayed, or contaminated by combustible materials such as grease or oils. Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen. Immediately disconnect the ventilator from the oxygen supply, facility power, and backup sources if there is a smell of burning.
General cautions
• As a general rule always avoid contact with external electrical connector pins. It is recommended to have the module compartment filled up with empty modules to protect from spillage and dust.
• Federal law in the USA restricts this device to sale by or on the order of a physician (or a properly licensed practitioner).
• The Servo-i Ventilator System must be serviced at regular intervals by specially trained personnel. The intervals are stated in the chapter Regular maintenance. Any maintenance must be noted in a log book for that purpose in accordance with national regulations.
• MAQUET has no responsibility for the safe operation of the equipment if service or repair is done by a non-professional or by persons who are not employed by or authorized by MAQUET. We recommend that service is done as part of a service contract with MAQUET.
• MAQUET has no responsibility for the safe operation of the equipment if the equipment is used for anything other than its intended use, as specified in this User´s manual.
• A resuscitator should always be readily accessible for extra safety.
• When connected to a patient, the system must never be left unattended.
• The nebulizer module is inoperative when the ventilator is running on batteries, to reduce the power consumption.
• The Expiratory cassette must not be lifted up when the ventilator is in operation. This may, however, be done when in Standby mode.
• Always use heat and moisture exchanger (HME) or equipment to prevent dehydration of lung tissue.
• Refer to the Installation instructions to assemble the system or options to obtain a proper mechanical assembly.
• When lifting or moving the ventilator system or parts of the system, follow established ergonomic guidelines, ask for assistance, and take appropriate safety precautions.
• Antistatic or electrically conductive breathing tubing should not be used with this lung ventilator system.
• Any scavenging system (Gas evac) connected must comply to ISO8835-3 with regard to subatmospheric pressure and induced flow. Otherwise ventilator functions and patient safety may be degraded.
• It is not recommended to use the Servo Evac 180 in the Nasal CPAP mode.
• Values measured at the signal outputs of the Servo-i Ventilator System and which have been processed in auxiliary equipment must not be used as a substitute for therapeutic or diagnostic decisions. Such decisions can be made only by staff with medical expertise, according to established and accepted practice. If auxiliary equipment that has not been delivered by MAQUET with the system is used, MAQUET denies all responsibility for the accuracy of signal processing.
10
Servo User´s manual
Infant Adult Universal Option
US edition Order No: 66 00 261
Warnings, cautions and important
1
• If there should be any deviation between information shown on the User Interface of the ventilator and that shown by the auxiliary equipment, the ventilator parameters shown on the User Interface shall be considered the primary source for information. When combining the Servo-i Ventilator System with accessories and auxiliary equipment other than those recommended by MAQUET, it is the responsibility of the user to ensure the integrity of system performance and safety. In order to maintain electrical system safety, i.e. such that compliance with IEC 60601-1-1 is fulfilled, only accessories and auxiliary equipment that meet current IEC standards (e.g. IEC 60601-1, IEC 950) may be connected to signal inputs and outputs of the Servo-i Ventilator System.
• Only original parts from MAQUET must be used in the system.
• Only accessories, supplies or auxiliary equipment recommended by MAQUET should be used with the ventilator system (“Products and accessories” catalog and “Spare parts list”). Use of any other accessories, spare parts or auxiliary equipment may cause degraded system performance and safety.
• The displayed information about set and corresponding measured parameters, shall continously be compared by the operator.
Important:
This symbol on the unit means Attention, consult accompanying documents.
Note: The are two versions of this symbol depending on System version.
• The gases supplied must be free from water, oil, particles and other contaminants:
Air ........ H
O < 7 g/m
2
............. Oil < 0.5 mg/m
............. Chlorine: Must not be detectable
Oxygen H2O < 20 mg/m
3
3
3
• The environmental declaration is part of the service manual.
• The Servo-i Ventilator system does not contain any latex.
• Data on pressures can be given in cmH where: 1 kPa ~ 10 cmH 100 kPa = 1bar ~1atm ~1kgf/cm
O
2
2
(kp/cm2)
100 kPa ~15 psi.
• All disposable parts must be discarded according to hospital routine and in an environmentally safe way.
• Do not expose the Expiratory cassette compartment to excessive amounts of fluid, e.g. during cleaning and disinfection, as this may influence ventilator functionality.
• Do not use sharp tools on the screen.
• It is recommended that at least two batteries always is used in the ventilator for backup.
• It is recommended that at least two batteries are used for ventilation during transport.
2
1
O,
Servo User´s manual US edition Order No: 66 00 261
1. If the compressed air is generated by a liquid ring compressor there is a potential risk of chlorine in the supplied air.
Infant Adult Universal Options
11
s
Warnings, cautions and important
1
• Documentation for Servo-i Ventilator System consists of:
–User´s manual – Brief instructions – Wall diagram – Installation instructions – Service manual – Products and accessories, catalog – Spare parts list
Context-related warnings
Note: General warnings are not listed here
even though they are repeated inside the manual.
Note: Context-related Cautions and "Important" are not listed here, but are written in the relevant context inside the manual.
Operation
• Always disconnect the ventilator if any operation which may involve risk for the patient will be done, e.g. replacement of O
cell, dismantling etc. (page 211,
2
page 225).
• If the trigger sensitivity is set too high, a self-triggering (auto-triggering) condition may be reached. This condition can also be reached if there is leakage in the breathing system, e.g. if an uncuffed endotracheal tube is used. Triggering will then be initiated by the system and not by the patient.This should always be avoided by decreasing the trigger sensitivity (page 23). This is also important during transport as the movement of the body and the breathing system may lead to false triggering.
• When you turn a Direct Access Knob, ventilation will change accordingly from the next breath without additional confirmation (For further information see page 166).
• If any malfunctions are detected during the start-up procedure, please refer to Chapter, Troubleshooting (page 225).
• If a malfunction persists, the ventilator may not be connected to the patient.
• A Pre-use check must always be done before connecting the ventilator to a patient (page 145).
• To protect the patient against high airway pressures, the upper pressure limit must always be set to the relevant value so as to provide adequate patient safety (page
165). Caution: If airway pressure rises 6 cmH
above the set upper pressure limit the safety valve opens. The safety valve also opens if system pressure exceeds 117 cmH
O.
2
• To provide adequate patient safety, always set the alarm limits at relevant values (page 165).
O
2
± 7
12
Servo User´s manual
Infant Adult Universal Option
US edition Order No: 66 00 261
Warnings, cautions and important
1
Nebulization
• Servo Humidifier/HME must be disconnected during nebulization otherwise the humidifier may be blocked (page 128).
• The heated humidifier must be switched off during nebulization. Otherwise the particle size may be affected (page 128).
• During nebulization a filter must be connected to the expiratory inlet of the ventilator. Always carefully monitor the airway pressure during nebulization. Increased airway pressure could be caused by a clogged filter. The filter should be replaced if the expiratory resistance increases or every 24 hours when the nebulizer is being used.
• When a Servo Ultra Nebulizer is used, always consult the drug manufacturer regarding the appropriateness of ultrasonic nebulization for certain medications (page 128, 187).
• The nebulizer must not be used without buffer liquid (sterile water). Otherwise the ultrasonic generator crystal may break (page 129, 187).
• To avoid explosion hazards, flammable agents such as ether and cyclopropane must not be used with this device. Only agents which comply with the requirements on non-flammable agents in the IEC standard “Particular requirements for electrical safety of anaesthetic machines” are suitable.
• For adult/pediatric patients, never fill the medication cup with more than 10 ml (page 129).
• For neonatal patients, never fill the medication cup with more than 4 ml (page 129).
• If the patient unit of the nebulizer is tilted, the drug can flow into the patient´s lungs or the ventilator.
• The nebulizer must not be left unattended when connected to a patient.
• Continuously check that the buffer liquid level is between MIN. and MAX. during nebulization (page 187).
• During nebulization: Continuously check that moisture is generated in the medication cup (page 187).
• When the ventilator is running on batteries the nebulizer module is inoperative, to reduce the power consumption (page 187).
• For information about the stand alone Aeroneb Professional Nebulizer System, refer to separate manual.
Cleaning
• All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator (page 191).
• After removing the Expiratory cassette, do not pour any fluid into the Expiratory cassette compartment (page 196).
Replacement of O2 cell
The sealed unit of the O2 cell, contains a caustic liquid which may cause severe burns to the skin and eyes. In case of contact, immediately flush continuously with water for at least 15 minutes and seek medical attention especially if the eyes are affected (page 214)
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
13
1
s
Notes
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
14
Infant Adult Universal Option
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
Servo User´s manual US edition Order No: 66 00 261
2. Ventilation
Contents
Modes of ventilation. . . . . . . . . . . . . . . . . . . . . . . 16
Important definitions . . . . . . . . . . . . . . . . . . . . . . 21
Trigger sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . 23
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Special functions . . . . . . . . . . . . . . . . . . . . . . . . . 29
Controlled ventilation - PRVC . . . . . . . . . . . . . . . 32
Controlled ventilation - Volume Control . . . . . . . . 35
Controlled ventilation - Pressure Control . . . . . . . 38
Supported ventilation - Volume Support . . . . . . . 40
Supported ventilation - Pressure Support . . . . . . 43
Spontaneous/CPAP . . . . . . . . . . . . . . . . . . . . . . . 46
Automode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
SIMV. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Bi-Vent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Non Invasive Ventilation (NIV) . . . . . . . . . . . . . . 61
NIV - Pressure Control . . . . . . . . . . . . . . . . . . . . . 62
NIV - Pressure Support . . . . . . . . . . . . . . . . . . . . 63
NIV - Nasal CPAP. . . . . . . . . . . . . . . . . . . . . . . . . 64
Open Lung Tool . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Ventilatory parameters, overview . . . . . . . . . . . . . 66
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
15
s
Modes of ventilation
2
Ventilatory management
The Servo-i Ventilator System is designed for safe and effective treatment. It can be set for continuous adaptation to the patient´s prevailing condition or for manually controlled operations. The servo systems for pressure, flow and timing operate in all modes of ventilation (set time in control modes and patient-related timing in support modes).
Important:
• To show all available installed ventilation modes, please refer to "Setting ventilation mode" on page 164 in this manual.
• In all pressure controlled modes, it is important to set alarm limits to adequate levels.
• For information about default values and parameter settings refer to page 249.
Application
The Servo-i ventilator system also contains tools to assist the user in application of lung recruitment methodologies.
Scope - ventilatory needs
The ventilator can be used for true:
1. controlled ventilation
2. supported ventilation, or
3. spontaneous breathing/CPAP 4-7. It also allows for combined ventilatory
control or support. Spontaneous breathing efforts are sensed during controlled ventilation, e.g. Volume Control. Mandatory ventilation can be used during supported/ spontaneous breathing, e.g. the enhanced SIMV functionality.
8. The Automode functionality continuously adapts to the patient´s breathing capability.
When required, all ventilation is provided for mandatorily. When the patient is able to initiate a breath, the ventilator supports and monitors the patient´s breathing capability and controls ventilation only if required.
16
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Modes of ventilation
2
Implementation
Ventilation can be managed and administered with a focus on:
A. pressure and volume B. pressure C. flow/volume.
Extra flow and extra breaths
In flow/volume- oriented modes of ventilation, additional on-demand flow can be triggered during inspiration. Additional breaths can always be triggered between the ordinary breaths if the set trigger criteria are met.
Timing
In controlled ventilation modes, timing is related to preset values. In supported ventilation modes, timing is related to patient triggering and Inspiratory cycle-off setting.
Pressure and volume in focus
In the pressure- and flow- oriented modes, a constant inspiratory Tidal Volume is maintained. The inspiratory pressure level is constant during each breath. (PRVC, Volume Support.)
Pressure in focus
In the pressure-oriented modes, a constant preset pressure level is maintained during inspiration. (Pressure Control, Pressure Support)
Flow/volume in focus
In the flow/volume oriented modes a constant inspiratory volume is maintained. The inspiratory flow is constant during each breath (Volume Control).
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
17
s
Modes of ventilation
2
Basic functionality - An overview
18
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
1.(PRVC) Pressure Regulated Volume Control
Breaths are delivered mandatorily to assure preset volumes, with a constant inspiratory pressure continuously adapting to the patient´s condition. The flow pattern is decelerating.
2. Volume Control
Breaths are delivered mandatorily with a constant flow to assure preset volumes.
3. Volume Support
A patient-adapted constant inspiratory support is supplied when activated by patient effort. The resulting volume is continuously monitored and the constant inspiratory pressure automatically adjusts to the required level. The patient determines frequency and duration of the breaths which show a decelerating flow pattern.
4. Spontaneous breathing (CPAP)
When sufficient inspiratory volumes are achieved, spontaneous breathing without ventilator support is allowed for in Volume Support.
5. Pressure Control
Breaths are delivered mandatorily at a preset pressure level, causing a decelerating flow pattern.
6. Pressure Support
Inspiration is supported by a constant preset pressure when activated by patient effort. The patient determines frequency and duration of the breaths, which show a decelerating flow pattern. Inspiratory breath duration can be influenced by adjusting the Inspiratory cycle-off criteria.
7. Spontaneous breathing/CPAP
True spontaneous breathing (CPAP) occurs when the inspiratory pressure level is set to zero in Pressure Support.
8. Nasal CPAP
Spontaneous breathing on a set pressure level.
Modes of ventilation
2
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
19
s
Modes of ventilation
2
Combined modes - An overview
Automode
The ventilator continuously adapts to the patient's breathing capability and allows the patient to better interact with the ventilator. The ventilator automatically shifts between controlled ventilation, supported ventilation and spontaneous ventilation. Each controlled ventilation mode has a corresponding support mode.
Volume Control <----> Volume Support PRVC <----> Volume Support Pressure Control <----> Pressure Support
When the patient is making a breathing effort, the ventilator immediately switches to a support mode of ventilation. If the patient is not making any breathing effort, the ventilator will return to the controlled mode and deliver controlled breaths.
Synchronized intermittent Mandatory ventilation (SIMV)
The ventilator provides mandatory breaths which are synchronized with the patient´s spontaneous efforts at a preset rate. The mandatory breaths can be Volume Control, Pressure Control or PRVC breaths.
Bi-Vent
Bi-Vent is pressure controlled breathing, giving the patient the opportunity of unrestricted spontaneous breathing. Two pressure levels are set together with the individually set duration of each level. Spontaneous efforts can be assisted by pressure support.
20
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Important definitions
2
x
V
V
ServoS-0046_XX
z
y
I:E
The graphic display of flow, pressure and volume is visualized in wave forms. Modes of ventilation directly affect flow, pressure and volume patterns.
Volume Control
Pressure-Time waveform. Points and regions of interest
X. Inspiration time Y. Pause time Z. Expiration time
1. Start of Inspiration
2. Peak inspiratory pressure
3. Early inspiratory pause pressure
4. End inspiratory pause pressure
5. Early expiratory pressure
6. End expiratory pressure
2
4
3
1
12
5
7
8
14
13
6
t
11
t
10
9
15
16
t
Flow-Time waveform. Points and regions of interest
X. Inspiration time Y. Pause time Z: Expiration time
7. Peak inspiratory flow
8. Zero flow phase
9. Peak expiratory flow
10. Slope decelerating expiratory limb
11. End expiratory flow
Volume-Time waveform. Points and regions of interest
X. Inspiration time Y. Pause time Z. Expiration time
12. Start of inspiration
13. The slope represents current delivery of inspiratory tidal volume
14. End inspiration
15. The slope represents current patient delivery of expiratory tidal volume
16. End expiration
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
21
s
Important definitions
2
P
V
ServoS-0047_XX
X
Z
I:E
Pressure Control
Pressure-Time waveform. Points and regions of interest
X. Inspiration time Z. Expiration time
1. Start of Inspiration
2. Peak inspiratory pressure
3. End expiratory pressure
2
1
4
5
6
9
8
3
t
7
t
10
t
Volume-Time waveform. Points and regions of interest
X. Inspiration time Z.: Expiration time
8. Start of inspiration
9. End inspiration
10. End expiration
Flow-Time waveform. Points and regions of interest
X. Inspiration time Z. Expiration time
4. Peak inspiratory flow
5. End inspiratory flow
6. Peak expiratory flow
7. End expiratory flow
22
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Trigger functionality
Trigg. Flow
5
Trigg. Pressure
-2
T rig g e r s e n s itiv it y
2
SVX-638_EN
This determines the level of patient effort to trigger the ventilator to inspiration.
Trigger sensitivity can be set in flow triggering (Trigg. Flow) or pressure triggering (Trigg. Pressure). Normally flow triggering is preferable as this enables the patient to breath with less effort.
The sensitivity is set as high as possible without self-triggering. This ensures that triggering is patient initiated and avoids auto­cycling by the ventilator.
Pressure triggering can be set in the range
-20 to 0 cmH level, white area on the bar).
O (in reference to set PEEP
2
When the trigger sensitivity is set above 0 (green and red area on the bar), flow triggering is set, i.e. the amount of the bias flow that the patient has to inhale to trigg a new breath. The sensitivity can be set from 100% of the bias flow (left), to 0% of the bias flow (right). For information about the different colors of the bar refer to page 167.
Important: In
NIV it is not possible to set
trigger sensitivity.
The ventilator continuously delivers a gas flow during expiration, which is measured in the expiratory channel.
1. Inspiration.
Bias flow during expiration.
2. Bias flow Infant 0.5 l/min. Bias flow: Adult 2 l/min.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
23
s
Trigger sensitivity
2
Weak patient effort
Trigg. Flow
5
SVX-141_EN
1. At a Trigger sensitivity level above zero
(0), the ventilator senses deviations in the bias flow caused by inspiratory efforts of the patient. The more to the right on the scale, the more sensitive is the trigger function.
2. Weak inspiratory effort.
3. Very weak inspiratory effort. For further information see page 167.
WARNING! If the trigger sensitivity is set too high, a self triggering (auto-triggering) condition may be reached. This condition can also be reached if there is leakage in the breathing system, e.g. if an uncuffed endotracheal tube is used. Triggering will then be initiated by the system and not by the patient.This should always be avoided by decreasing the trigger sensitivity.
Stronger patient effort
Trigg. Pressure
-2
SVX-142_EN
1. At a Trigger sensitivity level below zero (0), the ventilator senses negative pressures created by the patient. Required preset negative pressure to initiate a breath is shown numerically. The more to the left on the scale, the more effort is required to trigger.
2. Stronger patient effort.
For further information see page 167.
WARNING! The trigger sensitivity bar has different colors based on the setting. A green bar indicates a normal setting for the flow triggering. The risk of self-triggering increases when the bar is red. A white bar indicates that pressure triggering is required.
24
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Settings
2
Inspiratory rise time
Insp rise time
P
0
100 %
0
SVX-644_EN
Time to peak inspiratory flow or pressure at the start of each breath as a percentage of the respiratory cycle time or in seconds. Increased rise time will affect the rate of flow/ pressure increase and can be evaluated by the shape of the flow and pressure waveforms.
Inspiratory rise time (%) is applicable in Pressure Control, Volume Control, PRVC, SIMV-Volume Control, SIMV-Pressure Control, SIMV-PRVC. Setting can be in the range 0-20% of the respiratory cycle time.
Inspiratory rise time set in seconds is applicable in Pressure Support, Volume Support and Bi-Vent. For adults the range is 0-0.4 seconds and for infants the range is 0-
0.2 seconds. Note: When the ventilator is configured for setting of Inspiration time, the unit for Inspiratory rise time then automatically switches to seconds for all ventilation modes.
t
t
Inspiratory cycle-off
SVX-205_XX
Inspiratory Cycle-off is the point at which inspiration changes to expiration in spontaneous and supported modes of ventilation. A decrease of the inspiratory flow to a preset level causes the ventilator to switch to expiration. This preset level is measured as a percentage of the maximum flow during inspiration. The range of Inspiratory cycle-off is 1 - 70%.
Note: In NIV the range is 10-70%.
Normally in supported modes the Inspiratory rise time should be increased from the default setting and so give more comfort to the patient.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
25
s
Settings
2
Breath cycle time
This is the length of the breath i.e. the total cycle time of the mandatory breath in SIMV (inspiration, pause plus expiration). This is set in seconds within the range:
Infants: 0,5 -15 seconds in half second steps. Adults: 1-15 seconds in one second steps. Note: The soft key Breath cycle time is not shown when an SIMV mode is selected and inspiration time is configured. Refer to heading I:E ratio / Inspiration times.
Trigger timeout
Trigger Timeout is the maximum allowed apnea time in Automode before controlled ventilation is activated. It is applicable in:
Automode: Volume Control PRVC Pressure Control The settings are within the ranges:
• Infant: 3-15 seconds
• Adult: 7-12 seconds
Initially the ventilator adapts with a dynamic Trigger Timeout limit. This means that for the spontaneously triggering patient the timeout increases successively during the first ten breaths.
<--->Volume Support <--->Volume Support <--->Pressure Support
PEEP
PEEP
SVX-646_EN
Positive End Expiratory Pressure (PEEP) can be set in the range of 0 - 50 cmH Positive End Expiratory Pressure is maintained in the alveoli and may prevent the collapse of the airways.
Note: In NIV the range is 2-20 cmH
O. A
2
2
O.
26
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Settings
2
I:E ratio / Inspiration time
The setting of breathing parameters in Servo-i can be configured in two different ways, based on:
• I:E ratio (independent of changes of e.g. the breathing frequency) or,
• Inspiration time in seconds (independent of changes of e.g. the breathing frequency), to better meet the requirements for infant care.
When the ventilator is configured for setting of Inspiration time, the unit for Pause time and Insp. rise time then automatically switches to seconds. The resulting I:E ratio for each setting is shown in the upper right information area of the ventilation mode window.
As the inspiration time is explicitly set, a change of for example the Respiratory Rate will affect the I:E ratio. As a safety precaution, it will therefore be indicated when the resulting I:E ratio passes 1:1 in either direction.
Note: The soft key Breath cycle time is not shown when an SIMV mode is selected, since there is no need to set Breath cycle time when Inspiration time is directly set.
Note: The configuration is done by a service technician with a service card.
Volume setting
Depending on the ventilator configuration the inspiratory volume can be set as:
– Minute Volume or, – Tidal Volume
Note: The configuration is done by a service technician with a service card.
Controlled / supported pressure level
PC (Pressure Control level) above PEEP is the set inspiratory pressure level for each mandatory breath in Pressure Control and SIMV (PC) + PS, and also for Apnea back-up in Pressure Support.
PS (Pressure Support level) above PEEP is the set inspiratory pressure support level for triggered breaths in Pressure Support, SIMV modes and Bi-Vent.
O2 concentration
The setting range for the gas mixer is 21% O2 to 100% O automatically set at approximately 6% O2 above or below the set concentration value. There is also an absolute minimum alarm limit of 18% O operating settings.
. The alarm limits are
2
which is independent of
2
Respiratory rate / SIMV frequency
Respiratory rate is the number of controlled mandatory breaths per minute in controlled modes excluding SIMV. The respiratory rate is also used for calculation of tidal volume if the ventilator is configured for Minute volume setting. SIMV rate is the number of controlled mandatory breaths in SIMV modes.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
27
s
Settings
2
Previous ventilation mode
1. Time when previous mode was inactivated.
2. Press the pad Show previous mode to recall the previous accepted ventilation mode.
3. Activate the previous used ventilation mode settings by pressing the Accept pad.
Note:
• The previous ventilation mode function is
not available after a Pre-use check, changing of patient category, admitting a new patient, use of the same ventilation mode for more than 24 hours or after start­up (cold start) of the system.
• In backup ventilation, the ventilator shows
the settings for the supported mode when previous mode is activated.
• A recall of previous settings is only
possible after a change of ventilation mode.
28
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Special functions
2
Fixed keys
1. Start breath
2. O
breaths
2
3. Expiratory hold
4. Inspiratory hold can all be chosen by manually pressing the
respective fixed key.
Start breath
The ventilator will initiate a new breath cycle according to the current ventilator settings.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
29
s
Special functions
2
O2 breaths
This function allows 100% oxygen to be given for 1 minute. After this time the oxygen concentration will return to the pre-set value. The oxygen breaths can be interrupted by repressing the O the 1 minute interval.
breaths fixed key during
2
Expiratory hold
Expiratory and inspiratory valves are closed after the expiration phase is completed, for as long as the fixed key is depressed, up to a maximum of 30 seconds. Expiratory hold provides an exact measurement of the end expiratory pause pressure. It can be used for static compliance measuring and to determine the total PEEP. The dynamic pressure is shown on the PEEP numerical value.
30
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Special functions
2
Inspiratory hold
Inspiratory hold is activated by manually pressing the fixed key. The maximum time is 30 seconds. The inspiratory and expiratory valves close after inspiration. This function can provide an exact measurement of the end inspiratory lung pressure. It can be used during x-ray or to determine Plateau pressure, or static compliance calculation.
Back-up ventilation
Pressure support/
CPAP
Apnea
Volume support
SVX-647_EN
Back-up ventilation is available in all support modes (not applicable in Automode and NIV Pressure Support mode).
The Back-up function switches Volume Support to Volume Control, Pressure Support and CPAP to Pressure Control. During Back-up ventilation default settings are used for I:E ratio, Respiratory Rate, and Inspiratory rise time. Apnea alarm can be set in infant mode (5-45 seconds) and in adult mode (15-45 seconds). The Back-up pressure level is adjustable, minimum settable value is 5 cmH
Note: Back-up not applicable in NIV Nasal CPAP.
O.
2
Pressure control
Volume control
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
31
s
Controlled ventilation - PRVC
2
Functional description PRVC
The Pressure Regulated Volume Control (PRVC) mode is a controlled breathing mode.
Servo-i Ventilator can be configured to set Tidal Volume or Minute Volume. The following parameters are set:
1. Tidal Volume (ml) or Minute Volume (l/ min)
2. Respiratory Rate (b/min)
3. PEEP (cmH
4. Oxygen concentration (%)
5. I:E ratio / Insp. time
6. Inspiratory rise time (%/s)
7. Trigg. Flow / Trigg. Pressure
O)
2
The ventilator delivers a pre-set Tidal Volume. The pressure is automatically regulated to deliver the pre-set volume but limited to 5 cmH pressure limit.
The flow during inspiration is decelerating. The patient can trigger extra breaths.
O below the set upper
2
32
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
PRVC in detail
Controlled ventilation - PRVC
2
1
SVX-9006_XX
1. PRVC assures a set target minute ventilation to the patient. The target volume is based upon settings for Tidal Volume, frequency and inspiration time.
2. The inspiratory pressure level is constant during each breath, but automatically adapts in small increments breath-by­breath to match the patient´s lung mechanical properties for target volume delivery.
3. Inspiration starts according to a preset frequency or when the patient triggers.
Expiration starts:
a. After the termination of preset
inspiration time
b. If the upper pressure limit is
exceeded.
2
3
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
33
s
Controlled ventilation - PRVC
2
SVX-697_EN
The first breath of a start sequence is a volume-controlled test breath with Pause time set to 10%. The measured pause pressure of this breath is then used as the pressure level for the following breath. An alarm is activated if the pressure level required to achieve the set target volume cannot be delivered due to a lower setting of the upper pressure limit (- 5 cmH
2
O).
34
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Controlled ventilation - Volume Control
Functional description Volume Control
2
Volume Controlled ventilation ensures that the patient receives a certain pre-set Minute/ Tidal Volume.
Servo-i Ventilator can be configured to set Tidal Volume or Minute Volume. The following parameters are set:
1. Tidal Volume (ml) or the Minute Volume
(l/min)
2. Respiratory Rate (b/min)
3. PEEP (cmH
4. Oxygen concentration (%)
5. I:E ratio / Insp. time
6. Pause time (%/s)
7. Inspiratory rise time (%/s)
8. Trigg. Flow / Trigg. Pressure
O)
2
The airway pressure is dependent on the tidal volume, inspiration time and the resistance and compliance of the respiratory system. The set tidal volume will always be delivered. An increase in the resistance and decrease in compliance will lead to an increased airway pressure. To protect the patient's lungs from excessive pressure, it is very important to set the upper pressure limit to a suitable value.
It is possible for the patient to trigger extra breaths if they can overcome the pre-set trigger sensitivity. It is also possible for the patient, by their own inspiratory efforts, to receive a higher inspiratory flow and Tidal Volume during an inspiration than pre-set. The flow during inspiration is constant. The patient can trigger extra breaths.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
35
s
Controlled ventilation - Volume Control
2
Volume Controlled ventilation has, by tradition, delivered each breath with a constant flow and constant inspiratory and expiratory times, according to the settings. The Servo-i gives the possibility to the patient to modify both flow rate and timing. So, if a pressure drop of 3 cmH during inspiration, the ventilator cycles to
O is detected
2
Pressure Support with a resulting increase in inspiratory flow. When the flow decreases to the calculated target level this flow will be maintained until the set Tidal Volume is delivered.
SVX-652_EN
The waveform illustrations above show some practical consequences of this enhanced functionality.
• the top waveform shows the trace for a normal Volume Controlled breath
• the second waveform shows a situation when inspiration is prematurely interrupted as the set tidal volume has been delivered
• the third waveform shows a situation where the patient maintains a flow rate higher than the calculated target value. The set Tidal Volume has been delivered when calculated target flow is reached and the inspiration is prematurely interrupted
• the bottom waveform, shows a situation where the increased flow rate is maintained into the expiratory period. The patient will receive a higher tidal volume than set due to a higher flow/volume demand than calculated.
36
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Controlled ventilation - Volume Control
Volume Control in detail
2
SVX-9002_XX
1. Volume Control assures a preset tidal volume with constant flow during a preset inspiratory time at a preset frequency.
2. The inspiratory flow is constant and depends on User Interface setting.
3. Inspiration starts according to the preset frequency or when the patient triggers.
4. If the patient makes an inspiratory effort during the inspiratory period, the ventilator will switch to Pressure Support to satisfy the patient´s flow demand.
Expiration starts:
a. When the preset tidal volume is
delivered and after the preset pause time.
b. When the flow returns to the set value
after the preset tidal volume is delivered and after the preset pause time (on-demand support). The patient is however always guaranteed an expiration time corresponding to at least 20% of the total breath.
c. If the upper pressure limit is
exceeded.
1 2
3
4
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
37
s
Controlled ventilation - Pressure Control
2
Functional description Pressure Control
The Pressure Controlled mode is a controlled breathing mode.
The following parameters are set:
1. PC (Pressure Control level) above PEEP (cmH
O)
2
2. Respiratory Rate (b/min)
3. PEEP (cmH2O)
4. Oxygen concentration (%)
5. I:E ratio / Insp. time
6. Inspiratory rise time (%/s)
7. Trigg. Flow / Trigg. Pressure
The delivered volume is dependent upon the pressure above PEEP, lung compliance and resistance in the patient tube system and airways. This means that the Tidal Volume can vary. Pressure Controlled mode is preferred when there is leakage in the breathing system e.g. due to uncuffed endotracheal tube or in situations when the maximum airway pressure must be controlled. The flow during inspiration is decelerating. The patient can trigger extra breaths. If the patient tries to exhale during the inspiration, the expiratory valve will allow exhalation as long as the pressure is more than 3 cmH As the delivered tidal volume can vary it is very important to set alarm limits for Minute Volume to adequate levels.
O above the set pressure level.
2
38
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Controlled ventilation - Pressure Control
Pressure Control in detail
2
1
SVX-9003_XX
1. Pressure Control assures that the preset inspiratory pressure level is maintained constantly during the entire inspiration. Breaths are delivered according to the preset frequency, inspiration time and inspiratory pressure level resulting in a decelerating flow.
2. The preset pressure level is controlled by the ventilator. The resulting volume depends on the set pressure level, inspiration time and the patient´s lung mechanical properties during each breath with a decelerating flow.
3. Inspiration starts according to the preset frequency or when the patient triggers.
Expiration starts:
a. After the termination of preset
inspiration time.
b. If the upper pressure limit is
exceeded.
2 3
Active expiratory valve
SVX-9008_XX
If a patient tries to exhale during the inspiration, pressure increases. When it increases 3 cmH pressure level, the expiratory valve opens and regulates the pressure down to the set inspiratory pressure level.
Upper pressure Limit
SVX-9009_EN
If the pressure increases to the set upper pressure limit e.g. the patient is coughing, the expiratory valve opens and the ventilator switches to expiration.
t
O above the set inspiratory
2
t
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
39
s
Supported ventilation - Volume Support
2
Functional description Volume Support
The Volume Support mode is a patient initiated breathing mode, where the patient will be given support in proportion to their inspiratory effort and the target Tidal Volume.
The following parameters are set:
1. Tidal Volume (ml)
2. PEEP (cmH
3. Oxygen concentration (%)
4. Inspiratory rise time (s)
5. Trigg. Flow / Trigg. Pressure
6. Inspiratory Cycle-off (%)
O)
2
If the patient’s activity increases the inspiratory pressure support will decrease provided the set Tidal Volume is maintained. If the patient breathes below the set Tidal Volume the inspiratory pressure support will increase.
40
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
SVX-657_EN
Supported ventilation - Volume Support
2
The start breath is given with 10 cmH2O support. From that breath the ventilator calculates and continuously regulates the pressure needed to deliver the pre-set Tidal Volume.
During the remaining 3 breaths of the start up sequence the maximum pressure increase is 20 cmH sequence the pressure increases or decreases in steps of maximum 3 cmH
O for each breath. After the start up
2
O.
2
If the delivered Tidal Volume decreases below the set Tidal Volume the pressure support level is increased in steps of maximum 3 cmH is delivered. If the pressure support level
O until preset Tidal Volume
2
causes a larger Tidal Volume than preset, the support pressure is lowered in steps of maximum 3 cmH Volume is delivered.
O until the preset Tidal
2
The maximum time for inspiration is:
• Infant 1.5 seconds
• Adult 2.5 seconds
An alarm is activated if the pressure level required to achieve the set target volume cannot be delivered due to a lower setting of the upper pressure limit (- 5 cmH
2
O).
In this mode it is also important to set the apnea time appropriate to the individual patient situation. If this time is reached then the ventilator will automatically switch to Back-up mode providing controlled ventilation. In all spontaneous modes it is important to set the Minute Volume alarm.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
41
s
Supported ventilation - Volume Support
2
Volume Support in detail
1
SVX-9005_XX
1. Volume Support assures a set target Tidal Volume upon patient effort by an adapted inspiratory pressure support.
2. The inspiratory pressure level is constant during each breath, but alters in small increments, breath-by-breath, to match the patient´s breathing ability and lung mechanical properties.
3. Inspiration with Volume Support starts: When the patient triggers.
Expiration starts:
a. When the inspiratory flow decreases
below a preset fraction of the inspiratory peak flow (Inspiratory cycle-off)
b. If the upper pressure limit is
exceeded.
c. Maximum time for inspiration is
exceeded.
2
3
42
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Supported ventilation - Pressure Support
Functional description Pressure Support
Pressure Support is a patient initiated breathing mode in which the ventilator supports the patient with a set constant pressure.
2
The following parameters are set:
1. PS (Pressure Support level) above PEEP (cmH
O)
2
2. PEEP (cmH
3. Oxygen concentration (%)
4. Inspiratory rise time (s)
5. Trigg. Flow / Trigg. Pressure
6. Inspiratory Cycle-off (%)
7. PC (pressure control level) above PEEP (cmH
2
O).
O)
2
During Pressure Supported ventilation the patient regulates the respiratory rate and the Tidal Volume with support from the ventilator. The higher the pre-set inspiratory pressure level from the ventilator the more gas flows into the patient. As the patient becomes more active the pressure support level may be gradually reduced. It is important to set the Inspiratory rise time to a comfortable value for the patient. In Pressure Support the Inspiratory rise time should normally be increased.
It is also very important to set lower and upper alarm limit for expired Minute Volume.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
43
s
Supported ventilation - Pressure Support
2
SVX-661_XX
Inspiratory Cycle-off is important for the patient’s comfort and ventilator synchronization with the patient. Inspiratory Cycle-off is the point when inspiration switches to expiration. E.g. for a patient with expiratory resistance the inspiratory Cycle­off should be set to a high value to guarantee enough time for expiration.
Note: It is important to monitor the corresponding Tidal Volume levels.
Inspiration: when the patient triggers a breath, gas flows into the lungs at a constant pressure. Since the pressure provided by the ventilator is constant, the flow will decrease until the Inspiratory Cycle-off is reached.
Expiration starts when:
– The inspiratory flow decreases to the
pre-set Inspiratory Cycle-off level. – If the upper pressure limit is exceeded. – If the flow drops to a flow range between
25% of the peak flow and lower limit for
Inspiratory Cycle-off fraction level and
the spent time within this range exceeds
50% of the time spent in between the
start of the inspiration and entering this
range.
The maximum time for inspiration is:
• Infant 1.5 seconds
• Adult 2.5 seconds
44
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Supported ventilation - Pressure Support
Pressure Support in detail
1 2 3
SVX-9004_XX
1. Pressure Support assures that a preset inspiratory pressure level is constantly maintained upon patient effort.
2. The preset pressure level is controlled by the ventilator, while the patient determines frequency and inspiration time.
3. Inspiration starts when the patient triggers.
2
Expiration starts:
a. When the inspiratory flow decreases
below a preset fraction of the inspiratory peak flow (Inspiratory cycle-off)
b. If the upper pressure limit is
exceeded.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
45
s
Spontaneous/CPAP
2
Functional description Spontaneous breathing/CPAP
The mode Continuous Positive Airway Pressure is used when the patient is breathing spontaneously.
The following parameters are set:
1. PS (Pressure Support level) above PEEP (cmH
O)
2
O).
O)
2
2. PEEP (cmH
3. Oxygen concentration (%)
4. Inspiratory rise time (s)
5. Trigg. Flow / Trigg. Pressure
6. Inspiratory Cycle-off (%)
7. PC (pressure control level) above PEEP (cmH
2
A continuous positive pressure is maintained in the airways. Properly set this may prevent collapse of airways. Inspiration starts upon patient effort. Expiration starts as for Pressure Support above. Always set the Apnea time appropriate to the individual patient situation. If the apnea alarm limit is reached the ventilator will automatically switch back to a Back-up mode.
The alarm should alert staff to take action, either to go back to supported mode or change to a controlled mode of ventilation.
It is also very important to set lower and upper alarm limit for expired Minute Volume
The maximum time for inspiration is:
• Infant 1.5 seconds
• Adult 2.5 seconds.
46
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Spontaneous breathing/CPAP in detail
– True spontaneous breathing will occur:
a. In Volume Support when the target
volume is maintained without support (automatically regulated by the ventilator)
b. In Pressure Support when the
inspiratory pressure level is set to zero
c. In Automode when either of the above
defined conditions is met.
– Inspiration starts upon patient effort.
Expiration starts:
a. When the inspiratory flow decreases
below a preset fraction of the inspiratory peak flow (Inspiratory cycle-off)
b. If the upper pressure limit is
exceeded.
Spontaneous/CPAP
2
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
47
s
Automode
2
Automode Functional description
SVX-602_EN
Automode is a ventilator functionality where the ventilator adapts to the patient's varying breathing capacity and automatically shifts between a control mode and a support mode using a fixed combination of ventilation modes. There are three different combinations, depending on the modes installed:
• Volume Control<----> Volume Support
• PRVC <----> Volume Support
• Pressure Control <----> Pressure Support.
Note: Automode is not possible in NIV.
48
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Automode
2
Volume Control<->Volume Support
The ventilator uses the plateau pressure in the Volume Controlled breath as a reference pressure for the first Volume Supported breath.
PRVC <-> Volume Support
Pressure Control<->Pressure Support
In this combination of Automode - Pressure Control and Pressure Support - the Direct Access Knob will regulate the PC above PEEP (Pressure Control level).
The first supported breath delivered to the patient has the same pressure level as the preceding PRVC breath.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
49
s
Automode
2
Automode in detail
1. The ventilator starts in control mode and operates according to the Volume Control, PRVC or Pressure Control mode. If the patient triggers a breath, the ventilator will turn to support mode, to encourage the patient's respiratory drive.
2. If the patient is breathing adequately: a. In Volume Support the ventilator
adjusts the inspiratory pressure level breath-by-breath to assure the preset target volume.
b. In Pressure Support the ventilator
assures that the preset inspiratory pressure level is maintained constantly during the entire inspiration.
3. Exceeding the default or manually set
trigger timeout limit without a sufficient patient effort will cause:
a. In Volume Support; a PRVC or Volume
controlled breath will be delivered according to the selected automode functionality.
b. In Pressure Support; a Pressure
controlled breath will be delivered.
4. The ventilator initially adapts with a
dynamic trigger timeout limit. This means that for the spontaneously triggering patient, the trigger timeout limit increases successively until the set trigger timeout limit is reached.
50
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
PRVC - Volume Support
1
SVX-165_EN
2a
3a
Volume Control - Volume Support
Automode
4
2
1
SVX-222_EN
2a
3a
Pressure Control - Pressure Support
1
SVX-167_EN
3b2b
4
4
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
51
s
SIMV
2
Functional description SIMV
SIMV is a combination mode where the patient receives mandatory breaths synchronized with his breathing efforts and according to the selected SIMV mode. The patient can breath spontaneously with Pressure Support in between the mandatory breaths.
There are three different SIMV modes, depending on the modes installed:
• SIMV (PRVC) + Pressure Support
• SIMV (Volume Control) + Pressure Support
• SIMV (Pressure Control) + Pressure
Support
The mandatory breath
SIMV
(PRVC)+P
S
1
X
2
X
SIMV
(PC) + PS
X
2
X
PC above PEEP
Tidal volume / Minute volume
SIMV rate
Breath cycle time
SIMV
(VC)+ PS
XX
1
X
2
X
The Breath cycle time is the length of the mandatory breath in seconds.
For example: A SIMV rate of 6, a breath cycle time of 3 seconds with an I:E ratio of 1:2 means that the inspiration will take 1 second and the expiration 2 seconds.
SIMV Cycle
10 sec
373
SIMV Period
SVX-9010_EN
Spon. Period
SIMV Period
During the SIMV period, the first triggered breath will be a mandatory breath. If the patient has not triggered a breath within the first 90% of the Breath Cycle time a mandatory breath will be delivered.
Note: If the ventilator is configured for setting of Inspiration time, an I:E ratio of 1:2 will be used to estimate the Breath cycle time.
The spontaneous/pressure supported breaths are defined by setting the Pressure support level above PEEP.
I:E ratio / Inspiration time
Insp. rise time
Pause time
1
Only when the ventilator is configured for
XXX
XXX
2
X
Minute volume setting.
2
Only when the ventilator is configured for
I:E ratio setting. The Mandatory breath is defined by the basic
settings (as shown in the table above): Minute Volume/Tidal Volume (depending on configuration), PC above PEEP, I:E ratio/ Inspiration time (depending on configuration), Pause time, Inspiratory rise time and Breath cycle time.
Note: In the Minute Volume configuration the Tidal Volume is determined by Minute Volume divided by SIMV rate.
52
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
SIMV (PRVC) + Pressure Support
The following parameters are set:
1. Tidal Volume (ml)/Minute Volume (l/min)
2. SIMV rate (b/min)
SIMV
2
3. PEEP (cmH
4. Oxygen concentration (%)
5. I:E ratio / Insp. time
6. Inspiratory rise time (%/s)
7. Breath cycle time (s)
Note: The soft key Breath cycle time is not shown when an SIMV mode is selected and inspiration time is configured. Refer to page 27.
8. Trigg. Flow / Trigg. Pressure
9. Inspiratory Cycle-off (%)
10. PS (Pressure Support level) above PEEP (cmH
2
O)
O)
2
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
53
s
SIMV
2
SIMV (PRVC) + Pressure Support
P
SIMV Breath cycle
3
time
Spont. period
4
SIMV
Breath cycle
time
2
V
SVX-9027_EN
1
SIMV - in detail
1. This combined control and pressure support/spontaneous function allows for preset mandatory breaths synchronized with the patient's breathing.
2. If there is no trigger attempt within a time window equal to 90% of the set Breath cycle time, a mandatory breath is delivered. (The Breath cycle time is the total time for one mandatory breath.)
3. The mandatory breath is defined by the basic settings (mode of ventilation, breath cycle time, respiratory pattern and volumes/pressures).
4. The spontaneous/pressure supported breaths are defined by the setting for Pressure Support.
90%
time
time
54
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
SIMV (Volume Control) + Pressure Support
The following parameters are set:
1. Tidal Volume (ml)/Minute Volume (l/min)
2. SIMV rate (b/min)
SIMV
2
3. PEEP (cmH
4. Oxygen concentration (%)
5. I:E ratio / Insp. time
6. Pause time (%/s)
7. Inspiratory rise time (%/s)
8. Breath cycle time (s)
Note: The soft key Breath cycle time is not shown when an SIMV mode is selected and inspiration time is configured. Refer to page 27.
9. Trigg. Flow / Trigg. Pressure
10. Inspiratory Cycle-off (%)
11. PS (Pressure support) above PEEP (cmH
2
O)
O)
2
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
55
s
SIMV
2
SIMV (Volume Control) + Pressure Support
P
SIMV
Breath cycle time Spont. period
3
4
SIMV
Breath cycle time
2
V
SVX-9011_EN
1
SIMV - in detail
1. This combined control and pressure support/spontaneous function allows for preset mandatory breaths synchronized with the patient's breathing.
2. If there is no trigger attempt within a time window equal to 90% of the set Breath cycle time, a mandatory breath is delivered. (The Breath cycle time is the total time for one mandatory breath.)
3. The mandatory breath is defined by the basic settings (mode of ventilation, breath cycle time, respiratory pattern and volumes/pressures).
4. The spontaneous/pressure supported breaths are defined by the setting for Pressure Support.
90%
time
time
56
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
SIMV (Pressure Control) + Pressure Support
The following parameters are set:
1. PC (Pressure Control level) above PEEP (cmH
O)
2
2. SIMV rate (b/min)
SIMV
2
3. PEEP (cmH2O)
4. Oxygen concentration (%)
5. I:E ratio / Insp. time
6. Inspiratory rise time (%/s)
7. Breath cycle time (s) Note: The soft key Breath cycle time is
not shown when an SIMV mode is selected and inspiration time is configured. Refer to page 27.
8. Trigg. Flow / Trigg. Pressure
9. Inspiratory Cycle-off (%)
10. PS (Pressure Support level) above PEEP (cmH
O)
2
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
57
s
SIMV
2
SIMV (Pressure Control) + Pressure Support
P
SIMV Breath cycle
3
time
Spont. period
4
SIMV
Breath cycle
2
time
V
SVX-9027_EN
1
SIMV - in detail
1. This combined control and pressure support/spontaneous function allows for preset mandatory breaths synchronized with the patient's breathing.
2. If there is no trigger attempt within a time window equal to 90% of the set Breath cycle time, a mandatory breath is delivered. (The Breath cycle time is the total time for one mandatory breath.)
3. The mandatory breath is defined by the basic settings (mode of ventilation, breath cycle time, respiratory pattern and volumes/pressures).
4. The spontaneous/pressure supported breaths are defined by the setting for Pressure Support.
90%
time
time
58
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Bi-Vent Functional description
Bi-Vent
2
Bi-Vent is pressure controlled breathing that allows the patient the opportunity of unrestricted spontaneous breathing. Two pressure levels are set together with the individually set duration of each level. Spontaneous breathing efforts can be. assisted by pressure support
The following parameters are set:
1. Pressure high (P
pressure level (cmH
) for the higher
High
O)
2
2. PEEP for the lower pressure level O)
(cmH
2
3. Oxygen concentration (%)
4. Time at the higher pressure (T
High
) level
(s)
5. Time at the lower pressure (T
PEEP
) level
(s)
6. Inspiratory rise time (s)
7. Trigg. Flow / Trigg. Pressure
8. Inspiratory Cycle-off (%)
9. Pressure Support level above P O)
(cmH
2
High
In the Bi-Vent mode the ventilator uses two shifting pressure levels, with the patient being able to breath spontaneously on both these levels.
Since Bi-Vent is basically a controlled mode of ventilation, apnea alarm and back-up ventilation are not available. It is also very important to set lower and upper alarm limit for expired Minute Volume.
Every Bi-Vent cycle is regarded as autonomous and therefore most of the measured values are updated every Bi-Vent cycle, i.e. minute volumes, respiratory rate, mean pressure and end expiratory pressure. In accordance to this, associated alarms are also handled for every Bi-Vent cycle.
At extreme settings the update of measured values and alarms will show a mandatory frequency dependence even in the face of preserved spontaneous breathing.
As a result of switching between two different pressure levels, the tidal volumes may vary significantly between different breaths. This may also be the case for etCO concentration.
It is not recommended to use Auto scale in Bi-Vent mode, when patient is breathing spontaneous on both levels.
2
10. Pressure Support level above PEEP O)
(cmH
2
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
59
s
Bi-Vent
2
Bi-Vent in detail
1
1
5
3
3
SVX-184_XX
This function allows for spontaneous breathing / pressure supported ventilation at two different pressure levels. These basic levels are individually set, as well as the time in seconds at each level.The ventilator always tries to synchronize with the patient´s breathing.
1. Bi-Vent cycle; T
High
+ T
PEEP
2. PEEP
3. P
High
4. The pressure support level is set
individually: PS above PEEP
5. PS above P
High
2
2
4
60
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Non Invasive Ventilation (NIV)
2
Non Invasive Ventilation
This chapter refers to when the Servo-i is used during Non Invasive Ventilation (NIV). NIV refers to ventilation, where the patient is not intubated or tracheotomized. It is achieved using a nasal mask / prongs, face mask / prongs or full-face mask / prongs.
Note: In NIV, flow and pressure curves and the measured values: VTi, VTe, MVe, MVi are compensated for leakage.
WARNINGS!
• Avoid high inspiratory pressure as it may
lead to gastric overdistension and risk of aspiration. It may also cause excessive leakage.
• The dead space will increase when use of
a mask / prongs.
• NIV is not intended to be used on
intubated patients.
•CO
measurement will be affected by
2
mask / prongs leakage.
Cautions:
Read more about NIV
Intended population page 4 Ventilation modes (NIV): pages 62, 63 Alarm settings: page 73 Preparation: page 160
• Mask / prongs leakage might affect the
nebulizer efficiency.
• It is not recommended to use the nebulizer
during NIV as the nebulized drug might come in contact with the patient eyes in case of leakage.
Important:
• The mask / prongs must be applied in
order to avoid leakage.
• Selection of the mask / prongs must take
into consideration proper size and an accurate adaptation to the patient.
•CO
rebreathing will increase during NIV
2
and use of a face mask / prongs.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
61
s
NIV - Pressure Control
2
Functional description Pressure Control
The Pressure Controlled (NIV) mode is a controlled breathing mode.
SVX-9013_XX
The following parameters are set:
1. PC (Pressure Control level) above PEEP (cmH
O)
2
2. Respiratory Rate (b/min)
3. PEEP (cmH
4. Oxygen concentration (%)
O)
2
Differences from invasive Pressure control mode:
• When the Standby key is pressed a
waiting position dialog is shown. All patient related alarms are turned off during 120 seconds. Press the Start ventilation pad to start the ventilation.
• During NIV the ventilator automatically adapts to the variation of leakage in order to maintain the required pressure and PEEP level. If the leakage is excessive, the ventilator will issue a high priority alarm, deliver a continuous flow and pause breath cycling. Ventilation will resume automatically if the leakage decreases. Ventilation can also be started manually by pressing the Start ventilation pad in the excessive leakage dialog.
• Trigger sensitivity cannot be set in NIV.
Read more about NIV
Intended population page 4 NIV general information: page 61 Ventilation modes (NIV): page 63 Alarm settings: page 73 Preparation: page 160
5. I:E ratio / Insp. time
6. Inspiratory rise time (%/s)
62
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
NIV - Pressure Support
2
Functional description Pressure Support
Pressure Support (NIV) is a patient initiated breathing mode in which the ventilator supports the patient with a set constant pressure.
SVX-9014_XX
The following parameters are set:
1. PS (Pressure Support level) above PEEP (cmH
O)
2
2. PEEP (cmH
3. Oxygen concentration (%)
4. Inspiratory rise time (s)
5. Inspiratory Cycle-off (%)
6. NIV rate (b/min)
7. Backup Ti (s)
O)
2
Differences from invasive Pressure support mode:
• When the Standby key is pressed a
waiting position dialog is shown. All patient related alarms are turned off during 120 seconds. Press the Start ventilation pad to start the ventilation.
• During NIV the ventilator automatically adapts to the variation of leakage in order to maintain the required pressure and PEEP level. If the leakage is excessive, the ventilator will issue a high priority alarm, deliver a continuous flow and pause breath cycling. Ventilation will resume automatically if the leakage decreases. Ventilation can also be started manually by pressing the Start ventilation pad in the excessive leakage dialog.
• During Pressure support the system ensures a minimum Back-up Rate and maintains the set Inspiratory pressure and PEEP level. The Back-up Rate is activated when the spontaneous breathing rate is lower then the Back-up Rate, but the ventilator does not activate a Backup ventilation mode as in Invasive Pressure Support.
• Trigger sensitivity cannot be set in NIV.
Read more about NIV
Intended population page 4 NIV general information: page 61 Ventilation modes (NIV): page 62 Alarm settings: page 73 Preparation: page 160
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
63
s
NIV - Nasal CPAP
2
Functional description Nasal CPAP
The mode Nasal Continuous Positive Airway Pressure is used when the patient is breathing spontaneously.
SVX-9057
The following parameters are set:
1. CPAP (cmH
2. Oxygen concentration (%)
O)
2
Differences from invasive CPAP
• When the Standby key is pressed a
waiting position dialog is shown. All patient related alarms are turned off during 120 seconds. Press the Start ventilation pad to start the ventilation.
• Trigger and cycle-off is automatically adapted to the leakage and cannot be set in Nasal CPAP.
• There is no backup ventilation available in Nasal CPAP.
• The apnea alarm can be turned off in Nasal CPAP
The following functions are not available during Nasal CPAP ventilation:
•Volume curve
• Loops
• Open Lung Tool
• Additional values
• Additional settings
• Inspiratory hold
• Expiratory hold
•CO
Analyzer.
2
SVX-9061
64
WARNING! Patient effort and artifacts affecting patient flow or pressure such as heart beats, movement of patient tubings, intermittent leakage may not always be correctly detected or discriminated. This may affect the accuracy of alarms and measured parameters, therefore, we advise that a ventilator-independent means of monitoring the patient should be in place.
Read more about NIV
Intended population page 4 NIV general information: page 61 Ventilation modes (NIV): page 62 Alarm settings: page 73
Preparation: page 160
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Open Lung Tool
Clinical performance
The Open Lung Tool is a tool for graphically visualizing measured and calculated values for easier interpretation of already available ventilation data. Three simultaneous graphical trends are presented with a fixed set of parameters as a function of a number of collected breaths. The User Interface features an adjustable cursor which helps illustrate the opening and closing airway pressures. This alternative presentation may be used for immediate visualization of the effect of altered settings.
Note: When the Y Sensor Measuring function is active, then the values recorded in the Open Lung Tool are based on values measured at the Y-piece. Note that when this function is disabled or enabled, then the compliance in the patient circuit may cause the values in the Open Lung Tool to change.
Read more about the Open Lung Tool
Operating: page 175
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
Open Lung Tool
SVX-9023_XX
The following parameters are presented:
– In the top window, measured End
Inspiratory Pressure (EIP) and Positive End Expiratory Pressure (PEEP) are simultaneously presented, breath-by­breath.
– In the middle window, measured
Inspiratory tidal volume (VTi) and Expiratory tidal volume (VTe) are simultaneously presented, breath-by­breath.
– Dynamic compliance (C dyn i) is
calculated breath-by-breath and filtered before presentation. (C dyn i = VTi / EIP – PEEP)
– In the lower window measured Tidal CO
elimination (VTCO presented as well, breath-by-breath (CO
– The time parameter on the lower right
screen indicates how long it will take at the current settings for the waveform to fill the axis. Changing the scaling with the zoom in or out function will change the time and number of breaths needed for filling the axis.
– The breaths parameter on the lower right
screen indicates the number of breaths at the current respiratory rate it will take for the waveform to fill the axis.
Analyzer).
2
) is simultaneously
2
2
65
2
s
Ventilatory parameters, overview
2
SVX-202_EN
When a ventilation mode is selected, the only parameters shown are those affecting the actual mode. Below are all the mode-related parameters presented.
1. Respiratory rate (RR) Rate of controlled mandatory breaths or used for calculation of target volume (b/min).
2. Tidal volume (VT) Volume per breath or target volume (ml). Minute volume (Vmin) Volume per minute or target Minute volume (ml/min or l/min). Presentation can be configured to either tidal or minute volume.
3. PC above PEEP Inspiratory pressure level for each breath (cmH
O) in Pressure Control.
2
4. PS above PEEP Inspiratory pressure support level for triggered breaths (cmH in Pressure Support.
O)
2
5. Inspiratory rise time (T inspiratory rise) Time to full inspiratory flow or pressure at the start of each breath, as a percentage of the breath cycle time (%), or in seconds (s).
9. Trigger sensitivity
a) Below zero: Trigger sensitivity is pressure dependant. The pressure below PEEP which the patient must create to initiate an inspiration (cmH
O) is indicated.
2
b) Above zero: Trigger sensitivity is flow dependent. As the dial is advanced to the right (step wise from the green into the red area) the trigger sensitivity increases i.e the inhaled fraction of the bias flow leading to triggering is reduced.
10. PEEP Positive End Expiratory Pressure
(cmH
O).
2
11. Inspiratory cycle-off Fraction of
maximum flow at which inspiration should switch to expiration (%).
6. I:E ratio (I:E) (Inspiration time + Pause time): Expiration time.
7. Inspiration time (T
pressure delivery to the patient (s).
8. Pause time (T
pressure delivery (% or s).
) Time for active flow or
i
) Time for no flow or
pause
66
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Ventilatory parameters, overview
SVX-218_EN
SVX-203_EN
12. Breath cycle time (Breath cycle T) Total cycle time per mandatory breath in SIMV (inspiratory + pause + expiratory). Set in seconds.
2
13. SIMV rate Rate of controlled mandatory breaths (b/min).
14. Trigger timeout The maximum allowed apnea time in Automode, after which the system switches to controlled ventilation (s).
O
concentration (O2 Conc.) O2
2
concentration in inspiratory gas (not shown in the figure).
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
67
s
Ventilatory parameters, overview
2
SVX-204_XX
15. Time high (T Bi-Vent (s).
16. Time PEEP (T Bi-Vent (s).
) Time at P
High
) Time at PEEP level in
PEEP
High
level in
17. Pressure Support above Pressure high (PS above P
support level for breaths triggered during the
T
period in Bi-Vent (cmH2O).
High
) Inspiratory pressure
High
18. Pressure Support above PEEP (PS above PEEP) Inspiratory pressure
support level for breaths triggered during the
T
period in Bi-Vent (cmH2O).
PEEP
(P
19. Pressure high
Expiratory Pressure at the upper level in Bi­Vent (cmH
2
O).
) Positive End
High
20. PEEP Positive End Expiratory Pressure at the lower level in Bi-Vent (cmH
2
O).
68
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Notes
2
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
69
2
s
Notes
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
70
Infant Adult Universal Option
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
Servo User´s manual US edition Order No: 66 00 261
3. Patient safety
Contents
Alarms- General . . . . . . . . . . . . . . . . . . . . . . . . . 72
High priority alarms . . . . . . . . . . . . . . . . . . . . . . . 72
Medium priority alarms . . . . . . . . . . . . . . . . . . . . 72
Low priority alarms. . . . . . . . . . . . . . . . . . . . . . . . 72
Technical alarms. . . . . . . . . . . . . . . . . . . . . . . . . . 72
Alarm output connection option. . . . . . . . . . . . 72
The alarm profile window . . . . . . . . . . . . . . . . . 73
The alarm window . . . . . . . . . . . . . . . . . . . . . . . 74
Alarms - Visual / audible . . . . . . . . . . . . . . . . . . 75
Audio off (Silence / Pre-silence of alarm) . . . . 76
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Audio off of non-latching alarms . . . . . . . . . . . . . 76
Built-in safety precautions . . . . . . . . . . . . . . . . 78
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
71
s
Alarms- General
3
Several measures have been taken to design this system for safe treatment and use.
The alarms are based on three priority levels; High, Medium and Low.
High priority alarms
These alarms are warnings and are indicated by a red background. They are latched, i.e. the visual indication remains even though the alarm condition ceases.The background color switches to yellow if the alarm condition returns to normal. Latched alarms require manual resetting.
Note: NIV alarm Leakage out of range is not latched.
For more information about the high priority alarms see page 226.
Medium priority alarms
These alarms are advisory. They may be reset (cleared) even though the alarm condition remains.
For more information about the medium priority alarms see page 231.
Low priority alarms
These alarms are cautionary and are indicated by a yellow background.
For more information about the low priority alarms see page 234.
Alarm output connection
option
An Alarm output connection option makes it possible to connect the ventilator to an external alarm signal system. High and medium priority alarms are transferred. The alarm output signal is active as long as the audio alarm is active on the ventilator.
Importants:
• It is required that the patient is never left unattended and that external alarm is used only to draw extra attention to a patient.
• The alarm output is a non-guaranteed alarm according to IEC 60601-1-8 and it is recommended that the user establish a pre-use check routine for this application.
Technical alarms
Technical problem identified by a code.
For more information about the high priority alarms see page 239.
Alarm signals
All alarms are visual and audible.
WARNINGS!
• To protect the patient against high airway pressures, the Upper pressure limit must always be set to the relevant value so as to provide adequate patient safety.
• To provide adequate patient safety, always set the alarm limits at relevant values.
72
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
The alarm profile window
3
Alarm profile window
1. Press the Alarm profile key Shows all applicable alarms and settings for
both lower and upper limits. Also used for adjusting current limits and alarm sound level.
Note: Current alarm limits are displayed adjacent to the measured value, in smaller figures to the right of the display. Default values are displayed during power up and when admitting a new patient. Always make sure that values are appropriate for the patient.
Non Invasive Ventilation
SVX-9016_XX
1. Press the Alarm profile key to show the
applicable alarms for Non Invasive Ventilation (NIV).
2. The bell indicates if the alarm is audible active or Audio off (permanently silenced, a crossed bell).
The apnea alarm can be turned off in Nasal CPAP.
3. To turn off the apnea alarm in Nasal CPAP continue turning control wheel after maximum time limit has been reached.
Servo User´s manual US edition Order No: 66 00 261
When the apnea alarm is turned off, a message is displayed in the message area.
Read more about NIV
Ventilation modes page 61 Preparation page 160
Infant Adult Universal Options
73
s
The alarm window
3
Current alarms window
This window can be displayed if more than one alarm is active.
1. Press the bell (s) in the alarm message pad.
2. All alarms are shown in a window. This is dynamic and will be updated if more alarms occur while the window is open. The alarms are listed by priority and 10 alarm messages are displayed at the most.
3. Press the History pad.
4. The last 16 alarm-dependent events are listed chronologically. The most recent event is at the bottom.
Note: For viewing more than the latest 10 alarms, use the Event log to view all logged alarms (refer to page 270).
74
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Alarms - Visual / audible
3
Visual
1. A text message explaining the cause of the alarm flashes in the alarm message area. The alarm with highest priority is displayed first.
2. The corresponding measured value or set value box flashes and an arrow points at the exceeded limit.
A red background color indicates a high priority alarm. A yellow background indicates a medium or low priority alarm.
A high priority alarm which has been active but for which the condition has returned to normal is latched and requires manual resetting. (Latched alarms: The alarm text remains even though the alarm condition ceases.)
Note: NIV alarm Leakage out of range is not latched.
Audible
An active alarm is indicated by a distinct, but soft alarm signal. The sound level can be adjusted, e.g. lowered during the night time. (Set sound level is indicated in the Alarm profile window.)
Technical errors may also be indicated by a signal similar to that a medium priority alarm, generated by a sounding device in the Patient Unit.
Two bells in the alarm message area indicate that more than one alarm is activated.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
75
s
Audio off (Silence / Pre-silence of alarm)
3
General
All alarms except for No battery capacity and technical error alarms can be silenced (Audio pause) for two minutes. New alarms can be activated during this period. In Standby, only the following alarms are applicable:
No battery capacity (when Battery
module is connected)
Limited battery capacity (when Battery
module is connected)
Battery operation (when Battery module
is connected)
– Technical error – Touch screen or knob press time
exceeded – Internal temperature: High – Exp. cassette exchanged – Technical error in Expiratory cassette
Audio off (Silence of alarms)
At any time while the ventilator is operating (either Invasive or Non Invasive Ventilation modes), alarms can be placed into a state of audio off (silence alarms).
Non Invasive Ventilation
When Non Invasive Ventilation is chosen, the following alarms can be placed into a state of audio off (silenced alarms):
– Minute volume – Respiratory rate – PEEP
– End tidal CO – CPAP ( Nasal CPAP)
By pressing the corresponding bell symbol in the alarm profile window the button changes to a crossed bell to indicate Audio Off. It is also possible to configure these alarms individually to be set to the Audio Off state by default.
Note: When an alarm is silenced (Audio off) in the NIV mode, a symbol will appear on the screen, next to the corresponding measured value, saying Audio Off (a bell with negation cross). The Audio Off symbol will remain on the screen until the user reactivates the alarms or returns to the standby mode. If the user then enters Invasive ventilation and after that returns to NIV, the audible alarms will be reset to their default states.
(CO2 Analyzer)
2
Audio off of non-latching alarms
For a very limited number of alarms a single alarm condition can be silenced (Audio off) during the remaining time of the continuing alarm condition when the message Audio off? is shown. These alarms, such as Battery operation and Low Air/O will be re-activated the next time the alarm condition occurs.
supply pressure,
2
76
Servo User´s manual
Infant Adult Universal Option
US edition Order No: 66 00 261
Audio pause (Silence / Pre-silence alarm)
3
Audio pause/Prolong pre-silence period/Clear latched alarm
SVX-5098_EN
1. Press the Audio pause (Silence/Pre­silence alarm) key briefly, for less than two seconds:
• Active alarms are silenced (Audio
pause) for 2 minutes.
• If already silenced, the silent period is
prolonged for two minutes.
• Latched alarms disappear (latched
alarms: the alarm text remains even though the alarm condition ceases).
An alarm silence (Audio pause) symbol and the remaining time are then displayed in the message area.
Pre-silence alarm (Audio pause))
SVX-5099_EN
1. If you press the Audio pause (Silence/ Pre-silence alarm) key for more than two seconds:
•The active alarms are silenced (Audio
pause), i.e. a two minute period is started.
•All other alarms are silenced (Audio
pause) for 2 minutes, except those which cannot be silenced.
–Latched alarms disappear from the
alarm message area.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
77
s
Built-in safety precautions
3
For patient safety your Servo-i Ventilator System also has a range of built-in safety precautions.
Apnea alarm
The apnea alarm is applicable in all supported/spontaneous modes.
Note: When using the knob to adjust a value, the defined safety limits may be unintentionally reached or exceeded. In this case, the knob will become inoperable for 2 seconds to make you aware that the safety limit has been passed. (Note that this is only valid for Servo-i Infant and Servo-i Universal System versions).
Backup ventilation
In case of exceeded apnea in Volume Support or Pressure Support, a safety backup mode is activated with default breathing frequency and set / default values.
High pressures
The safety valve opens if the pressure in the inspiratory channel is too high.
Caution: If airway pressure rises 6 cmH above the set upper pressure limit the safety
valve opens. The safety valve also opens if system pressure exceeds 117
± 7 cmH
O
2
O.
2
NIV rate
During Pressure support (NIV) the system ensures a minimum Back-up Rate and maintains the set Inspiratory pressure and PEEP level. The Back-up Rate is activated when the spontaneous breathing rate is lower then the Back-up Rate.
No gas supply
If the air and O2 pressure is too low the safety valve and the expiratory valve will open. An alarm will be activated simultaneously.
Parameters and alarm limits
The system has default values for parameters and alarm limits. These are valid until you adjust them before/after connection to a patient. You can also enter new default values or use the values previously applied.
Standby position
All settings will be saved when the ventilator is set in standby position. The ventilator can thus be prepared and the CO warmed up for admission in advance.
transducer
2
Gas supply O2/Air
If the O2 or air supply pressure is too low the flow from the missing gas is automatically compensated for. The patient will get preset volumes and pressure with O alarm will be activated.
/air and an
2
Mains failure and battery
In case of a mains failure, the ventilator will automatically switch over to battery operation. The switch is indicated by a medium priority alarm. The remaining battery capacity is displayed in the status menu on top of the screen. In case of a mains failure and no Battery module has been inserted or connected, a high priority alarm is activated. The inspiratory and expiratory valves are opened to allow for breathing through the ventilator. All settings are saved until the ventilator is powered again.
78
Infant Adult Universal Option
Read more about the alarms
Alarm settings: page 165 Troubleshooting: page 225 Auto set values
calculation: page 244 Default values: page 249 Keys and touch pads: pages 268, 269.
Servo User´s manual US edition Order No: 66 00 261
Notes
3
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
79
3
s
Notes
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
80
Infant Adult Universal Option
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
..................................................................
Servo User´s manual US edition Order No: 66 00 261
4. Device description
Contents
The system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
An overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Options / Accessories . . . . . . . . . . . . . . . . . . . . . 86
User Interface
User Interface - General. . . . . . . . . . . . . . . . . . . . 87
User Interface - Connections and labels . . . . . . . 88
Touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Main Rotary Dial. . . . . . . . . . . . . . . . . . . . . . . . . . 90
Fixed keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Adjusting parameter values . . . . . . . . . . . . . . . . . 91
Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Measured value boxes . . . . . . . . . . . . . . . . . . . . . 92
Servo User´s manual US edition Order No: 66 00 261
User Interface - Positioning . . . . . . . . . . . . . . . . . 93
User Interface - Accessories . . . . . . . . . . . . . . . . 94
Patient Unit
Patient Unit - Connections and labels . . . . . . . . . 96
Patient Unit - Expiratory cassette . . . . . . . . . . . . 98
Mobile Cart Servo-i . . . . . . . . . . . . . . . . . . . . . . 99
Holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Servo Ultra Nebulizer. . . . . . . . . . . . . . . . . . . . 102
Analyzer Servo-i . . . . . . . . . . . . . . . . . . . . 104
CO
2
Y Sensor measuring. . . . . . . . . . . . . . . . . . . . . 106
Battery module . . . . . . . . . . . . . . . . . . . . . . . . . 108
Ventilation Record Card. . . . . . . . . . . . . . . . . . 114
Compressor Mini . . . . . . . . . . . . . . . . . . . . . . . 115
System transport and storage . . . . . . . . . . . . 116
Infant Adult Universal Options
81
s
The system
4
The ventilator
All ventilatory settings are made on the User Interface panel. It can either be operated by the touch screen and the Main Rotary Dial or by using the Main Rotary Dial only. Flow and pressure are continuously measured by transducers and controlled by a feedback system in the Patient Unit. The information is compared with the User Interface settings, and a difference between the actual measured value and the preset/calculated values will cause adjusted gas delivery according to the target flow/volume/ pressure. The Servo-i Ventilator System has two gas modules, one for air and one for O Gas can be connected from a medical pipeline system, a compressor or gas tanks. Air can also be supplied by a compressor.
2
6
7
.
8
9
82
1. Air and O
2. Mains supply
3. User Interface
4. Patient Unit
5. Expiratory inlet
6. Servo guard, viral/bacterial filter
7. Inspiratory outlet
8. Patient system
9. Module compartment
Infant Adult Universal Option
supply
2
Servo User´s manual US edition Order No: 66 00 261
An overview
The system
4
6
1. The ventilator can be delivered in three configurations. Your configuration is clearly indicated on the Patient Unit, at start up and in the Brief Instructions.
The Servo-i Infant with
defaults, scales and safety precautions, designed for use with Infant patients. It is standard configurated for
pressure controlled modes of
ventilation.
The Servo-i Adult with defaults, scales and safety precautions, designed for use with adolescent and adult patients. It is standard configurated for volume controlled modes of ventilation.
The Servo-i Universal (Basic or
Extended edition) is an advanced ventilator to be used with infants and adults. Enhanced functionality i.e. a comprehensive array of different ventilation modes, extended Tidal Volume range, allows for advanced ventilatory treatment in both categories.
2. The User Interface, where all settings are made and effects are monitored.
3. The Patient Unit, where gases are administrated, also has slots for Battery modules and future function modules. Battery module allow backup during mains failure and transport.
4. The Servo Ultra Nebulizer is operated from the User Interface.
Note: The Aeroneb Professional Nebulizer System can be used as a stand alone nebulizer system. Refer to separate manual.
5. The mainstream CO and calculations are displayed on the
User Interface.
6. Y Sensor measuring. Allows measurement of pressure and flow right up to the Y-piece.
measurement
2
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
83
s
6
The system
4
In the system
Default values give fast system start-up. User set values tailor the ventilatory management. Signals are fed to the Patient Unit, which executes ventilation managed by the servo control system. The internal design of the Patient Unit allows for true inspiratory and expiratory regulation and measurement. Set, measured and calculated values are presented on-screen breath-by-breath.
1. Patient data can be transferred to a Personal Computer via the Ventilation record card for further processing and storage.
2. Signals are conveyed from the User Interface for controlling of drug nebulization.
Important: Only valid for the built in Servo Ultra Nebulizer.
3. Data communication to a Personal Computer is possible via the Communication Interface Emulator (CIE) and the serial communication port (RS 232C).
84
Infant Adult Universal Option
4. An alarm output connection option makes it possible to connect the ventilator to an external alarm signal system
5. Signals from the CO sensor are conveyed to the User
Interface, where they are calculated and displayed on the screen.
6. Y Sensor measurements are conveyed to the User Interface, where they are calculated and displayed on the screen.
Capnostat
2
Servo User´s manual US edition Order No: 66 00 261
System elements; overview
Configuration
4
Functionality/Configuration
Alarm output connection option Automode, pressure Automode, PRVC Automode, volume Bi-Vent CO
Analyzer
2
NIV (Non Invasive Ventilation) Nasal CPAP Open Lung Tool
Pressure Control Pressure Support PRVC (Pressure Reg. Volume
Control) SIMV (Press. Contr.) +Pressure
Support SIMV (PRVC) + Pressure Support SIMV (Vol. Contr.) + Pressure
Support Suction Support Upgrade to universal (all patient
categories) Volume Control Volume Support Y Sensor measuring
Reference in
this manual
Basic Extended
ttt t
ttt %
ttt %
ttt %
ttt t
ttt t
pages 72, 150 page 48 page 48 page 48 page 59 pages 104,
151
ttt t t - ttpage 64
ttt %
pages 61, 160
pages 65, 175, 176
%t % %
%%% %
ttt %
%t % %
tt% %
t%% %
%%% %
tt
t%% %
ttt %
ttt t
page 38 page 43 page 32
pages 52, 57
pages 52, 53 pages 52, 55
page 170
-
page 35 page 40 pages
106,135, 192, 248
% included (in standard configuration)
t optional
Note: Refer to page 83 for Servo-i configuration definition.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
85
s
Options / Accessories
4
General options / accessories Reference in this manual
Aeroneb Professional Nebulizer System Refer to separate manual Battery module Servo-i page 108 CO
Analyzer Servo-i pages 104, 151, 207, 238 ,247
2
Compressor Mini pages 115, 248 Drawer kit Servo-i page 99 Extra Expiratory cassette. page 196 Fisher & Paykel humidfier MR730/MR850 page 127 Gas cylinder restrainer Servo-i pages 101, 248 Gas trolley Servo-i pages 101, 126, 248 Holder Servo-i pages 100, 124, 248 Humidifier Holder pages 100, 127 IV pole pages 100, 248 Mobile Cart Servo-i pages 97, 123, 248 O
cell / O2 Sensor pages 98, 213
2
Patient tubing (10, 15, 22 mm diameter) pages 120, 121, 122 Servo guard (viral/bacterial filter) page 137 Servo humidifier page 136 Servo Ultra Nebulizer Servo-i pages 102, 128, 205, 246 Shelf base Servo-i page 100 Support Arm 177 pages 100, 125 User Interface panel cover page 95 Y Sensor measuring pages 106,135, 192, 248
86
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
An overview
Mode
Volume control
User Interface - General
4
SVX-5001_EN
The User Interface is ergonomically designed. You can operate it via the touch screen or by means of the Main Rotary Dial. Fixed keys allow immediate access. Direct Access Knobs allow for immediate adjustments. Data can be shown as waveforms and/or as numerics. The measured value boxes are always visible (also while setting the ventilator). Alarm limits are displayed adjacent to the measured value. All functions and necessary information are gathered in the User Interface. Do not use sharp tools on the screen.
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
Read more about the User Interface
Positioning: page 93 Operating: page 159 Alarm settings page 165 Cleaning: page 191 Technical data: page 241 Keys and touch pads: page 259
87
s
User Interface - Connections and labels
4
88
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
User Interface - Connections and labels
It consists of:
1. Patient category.
2. Active mode of ventilation.
3. Automode On/Off.
4. Admit patient/Entered patient data and admission date.
5. Nebulizer On/Off.
6. System status parameters.
7. Fixed keys for immediate access to special windows.
8. The Main Rotary Dial with which you select the desired menu touch pad or parameter box. You can also adjust values. By pressing it, you confirm your settings.
9. Special function keys for immediate ventilatory functions.
10. Direct Access Knobs for immediate adjustments of vital parameters. A built­in 2 seconds safety delay with inactive knobs when a setting reaches predefined safety limits.
11. Mains indicator (green).
12. Standby indicator (yellow).
13. Start/Stop (Standby) ventilation key. In Standby everything is turned on, except for ventilation.
14. On/Off switch (rear side)
15. Slot for Ventilation record card
16. Luminiscens detector: adjusts screen brightness automatically.
17. Informative text messages. A purple symbol indicates patient triggering.
18. Alarm messages.
19. The waveform area which monitors two to four parameters, individually scaled. You can add an pressure/flow loop, with the ordinary waveforms still visible. This waveform area is also used for the trend presentation.
4
20. A section where measured values and set alarm limits are displayed in boxes. You can choose which parameter values to show.
21. Additional settings.
22. Additional measured values.
23. Loudspeaker.
24. Cable reel for the control cable.
25. Slot for Ventilation record card with a cover.
26. Screen rotation locking lever.
27. Locking screw for alternative cart mounting.
28. Panel holder for positioning on the Mobile Cart.
29. Control cable, 2.9 meters between User Interface and Patient Unit.
30. Service connector
31. On/Off switch. In the off position everything is turned off; however the plug-in battery continues to charge when connected to mains (Set to On in the graph).
32. Locking arm to tilt the screen.
Read more about the User interface
Positioning: page 93 Operating: page 159 Alarm settings page 165 Cleaning: page 191 Technical data: page 241 Keys and touch pads: page 259
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
89
s
User Interface - Functionality
4
Touch screen
1. Activate the desired menu touch pad by pressing it.
2. Activate the desired parameter by pressing the touch pad (highlighted white with a blue frame). It is now possible to enter a new value.
3. Turn the Main Rotary Dial to the desired value or line.
4. Confirm your setting by pressing the Dial or the parameter touch pad (turns blue again). To set more parameter values repeat steps 2) - 4).
5. To activate your settings, press Accept.
6. To cancel your settings, press Cancel.
Note: For more information about settings and operating, refer to page 145.
Main Rotary Dial
SVX-6021_XX
1. Turn the Main Rotary Dial until the desired menu touch pad is marked with a blue frame.
2. Press the Dial to confirm.
– The menu touch pad is highlighted in
white with a blue frame.
– Change values by turning the Dial and
confirm the settings by pressing the Dial.
Note: For more information about settings and operating, refer to page 145.
90
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
User Interface - Functionality
4
Fixed keys
There are two kinds of fixed keys:
1. Short-cut to function or screen.
2. Start special ventilatory function, which demands continuous supervision when used.
Press to activate. Note: For more information about settings and operating, refer to page 157.
Adjusting parameter values
SVX-5089_EN
Immediate adjustment
1. Turn the Direct Access Knob to the desired value. When you reach the defined safety limits the knob is inoperative for 2 seconds, to make you aware that you have passed a safety limit.
Servo User´s manual US edition Order No: 66 00 261
Combined adjustments
2. Press Additional settings and adjust values. Confirm your setting
Infant Adult Universal Options
91
s
User Interface - Functionality
4
Waveforms
As default four waveforms are shown simultaneously (If CO2 Analyzer is connected).
1. Each waveform displays one measured parameter against time (x-axis). The displayed variable and scale are indicated on the y-axis.
2. The waveforms are color-coded (default from factory):
– Yellow for pressure – Green for flow – Light blue for volume – Light yellow for CO
The waveform amplitude can be set individually or by the system, using Auto. Sweep speed can also be adjusted. For further information see page 157. The settings are effective from the first breath after adjustment.
The displayed waveforms can be configurated using Waveform configuration. For further information see page 169.
concentration.
2
Measured value boxes
20
15
10
8.5
8.5
6.5
40
6.2
8.5
6.5
30
11
6
The measured value boxes show measured/ calculated values in numerics and the unit being used.
1. Set Lower and Upper alarm limits are also shown.
2. If an alarm limit is exceeded, the box turns red for a high priority alarm (page
72) and yellow for a medium priority alarm (page 72). The exceeded limit is indicated by an arrow.
3. A value out of range is also labelled "***".
4. Additional measured values can be shown in the box.
92
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
User Interface - Positioning
User Interface positioning
The User Interface can be positioned on the Mobile Cart, a table, a shelf or a pipe.
1. Lift the User Interface straight up.
2. Place the panel on a table, shelf or on a pipe and fasten it securely by turning the handle of the locking screw.
Note: Make sure that the User Interface is fastened firmly. When positioned on a pipe the dimension of the pipe must be between 15 - 30 mm.
4
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
93
s
User Interface - Accessories
4
Knob cover
The knob cover protects the Direct Access Knobs against inadvertent activation. Raise the cover to access the Direct Access Knobs.
94
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
User Interface - Accessories
User Interface panel cover
The User Interface panel cover protects the screen from inadvertent activation of settings and mechanical damage during transport. While attached the user can still access the vital settings. Raise the cover to access the screen.
4
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
95
s
Patient Unit - Connections and labels
4
Note: Refer to chapter Before use (page 5) for more information about symbols on the Servo-i Ventilator system.
96
Infant Adult Universal Option
Servo User´s manual US edition Order No: 66 00 261
Patient Unit - Connections and labels
4
1. Handle
2. Gas inlet for air
3. Gas inlet for O
4. Air / Luft
5. O
2
6. Model number
7. Serial number
8. Manufacturing information
9. Equipotentiality terminal, Label
10. Fuse label T 2.5AL
11. Mains power voltage
12. Mains supply connector with fuse
13. Cooling fan with filter
14. Alarm output connection option
15. External +12V DC inlet Caution: When external +12 V DC is
used, at least one installed Battery module is required to ensure proper operation.
16. Fuse for external DC power supply
17. Optional connector
18. User Interface connector
19. RS232 connector
20. Expiratory outlet
21. Cover, inspiratory channel
22. Expiratory inlet
23. Battery lock
24. Module compartment Note: The slots are numbered (1,2,3...)
from top to bottom.
2
Read more about the patient unit
Positioning: pages 123, 124 Cleaning: page 191 Technical data: page 241
25. Nebulizer connector (only for Servo Ultra Nebulizer)
SVX-6076_XX
26. Inspiratory outlet
Servo User´s manual US edition Order No: 66 00 261
Infant Adult Universal Options
97
s
Patient Unit - Expiratory cassette
4
Gas flow through the Patient Unit
1. Gas inlet for O2.
2. Gas inlet for air.
3. The gas flow is regulated by the gas modules for Air and O
.
2
4. The gases are mixed in the inspiratory mixing section.
5. The Oxygen concentration can be measured either with an O
Sensor. The O
cell is protected by a
2
cell or an O2
2
bacterial filter. Note: On the illustration an O
cell is
2
connected.
6. The pressure of the mixed gas delivered to the patient is measured by the Inspiratory pressure transducer. The transducer is protected by a bacterial filter.
7. The inspiratory channel delivers the mixed gas to the patient system´s inspiratory tubing. The inspiratory channel also contains a safety valve.
8. The patient system´s expiratory tubing is connected to the expiratory inlet. The inlet also contains a moisture trap.
9. The gas flow through the expiratory channel is measured by ultrasonic transducers.
10. The expiratory pressure is measured by the expiratory pressure transducer (located inside the ventilator). The transducer is protected by a bacterial filter in the cassette.
11. The pressure (PEEP pressure) in the patient system is regulated by the expiratory valve.
12. Gas from the patient system leaves the ventilator via the expiratory outlet. The outlet contains a non-return valve.
Note: The Expiratory cassette can be exchanged between different Servo-i Ventilator System. Always perform a Pre-use check after exchanging an Expiratory cassette.
98
Servo User´s manual US edition
Infant Adult Universal Option
Order No: 66 00 261
Loading...