5|Introduction
17|System Overview
37|Power supply
43|Operation overview
65|Monitor and record
77|Ventilation, modes and functions
153|NAVA
173|Alarms
183|Optional Accessories
203|System messages
215|Start-up configuration
219|Technical data
249|Definitions
253|Appendix • User interface
265|Certificates
267|Index
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
Infologic 1.73
3
| TABLE OF CONTENTS |
4
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
Infologic 1.73
1 INTRODUCTION
TABLE OF CONTENTS
| Introduction |
6|Device description1.1
9|Warning, Caution, Important and Note1.2
14|Version and Configurations1.3
1 |
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
5
| Introduction |
| 1
1.1 DEVICE DESCRIPTION
This section provides general information
about the SERVO-i Ventilator System along
with guidelines for appropriate use.
1.1.1 DEVICE DIAGRAM
1
3
SVX-128a
1.1.2 DEVICE COMPONENTS
2
1.1.3 INTENDED USE
The SERVO-i Ventilator System is intended
for treating and monitoring patients ranging
from neonates to adults with respiratory failure
or respiratory insufficiency.
NAVA
The added indications for use of the NAVA or
NIV NAVA software are that the electrical
signal from the brain to the diaphragm is
intact, and that there is no contraindication for
insertion/exchange of nasogastric tube.
1.1.4 INTENDED USER
The SERVO-i Ventilator System should be
used only by those who:
are professional healthcare providers, and
have received training in the use of this
system, and
have experience with ventilation treatment.
The SERVO-i Ventilator System consists of
the following components:
3.Patient Breathing System—for delivering
and exchanging gases
6
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| Introduction |
1 |
1.1.5 INTENDED USE ENVIRONMENT
The SERVO-i Ventilator System should be
used only:
in hospitals
in facilities whose primary purpose is to
provide healthcare
during transport of a patient within hospitals
or healthcare facilities
during MR examinations of patients if the
conditions in the SERVO-i MR Environment
declaration (order no. 66 71 670) are met
and an agreement with MAQUET is signed.
1.1.6 CLEANING AND MAINTENANCE
Please refer to the SERVO-i/s Cleaning and
Maintenance User's Manual.
1.1.7 SERVICING GUIDELINES
CAUTIONS:
Regular Service: The SERVO-i Ventilator
System must be serviced at regular
intervals by MAQUET authorized
personnel who have received specialized
training.
Complete Service Records: All service
performed on the SERVO-i Ventilator
System must be recorded in a service
log in accordance with hospital
procedures and local and national
regulations.
Service Contract: We strongly
recommend that all service on the
SERVO-i Ventilator System should be
performed as part of a service contract
with MAQUET.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
7
| Introduction |
| 1
1.1.8 DISCLAIMERS
Improper Use Environment
MAQUET has no responsibility for the safe
operation of SERVO-i Ventilator System if
the Intended Use Environment requirements
specified in this document are not followed.
Nonprofessional Servicing
MAQUET has no responsibility for the safe
operation of the SERVO-i Ventilator System
if installation, service or repairs are
performed by persons other than MAQUET
authorized personnel.
8
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| Introduction |
1 |
1.2 WARNING, CAUTION,
IMPORTANT AND NOTE
Follow these safety guidelines. Additional
warnings appear in context throughout this
document.
Information is highlighted with Warning,
Caution, Important or Note, where:
WARNING! Indicates critical information
about a potential serious outcome to the
patient or the user.
CAUTION: Indicates instructions that must
be followed in order to ensure the proper
operation of the equipment.
Important: Indicates information intended
to help you operate the equipment or its
connected devices easily and conveniently.
Note: Indicates information requiring special
attention.
1.2.1 GENERAL
This manual summarizes the functions and
safety features of the SERVO-i Ventilator
System. It is not all-inclusive and should not
be construed as a substitute for training.
WARNINGS!
Always perform a Pre-use check before
connecting the ventilator to a patient.
If any of the following occurs, discontinue
use of the ventilator and contact a
service technician:
- Unfamiliar pop-up windows on the
screen
- Unresolvable alarms
- Unfamiliar sounds
- Any unfamiliar or unexplained event
Keep the ventilator upright during use.
Make sure that ventilation is started when
a patient is connected to the ventilator.
When the ventilator is in Standby, a
flashing message, Patient not ventilated,
is displayed as a reminder directly above
the word Standby.
When the ventilator is connected to a
patient:
- Do not lift or disconnect the expiratory
cassette.
- Continuously monitor the settings and
measurements displayed on the
screen.
- Make sure a resuscitator is readily
available.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
The SERVO-i Ventilator System must be
operated only by authorized personnel
who are well trained in its use. It must be
operated according to the instructions in
this User´s Manual.
9
| Introduction |
| 1
Do not modify or remove any original
parts.
Do not cover the ventilator in any way,
since the functioning of the equipment
may be adversely affected.
When the ventilator is used for MCare
Remote Service, use only network
equipment that is safe and in compliance
with the relevant electrical and EMC
standards such as IEC-60950.
Note: The network cable is excluded
from this requirement.
Always disconnect the network cable
before starting ventilation when the
ventilator is used for MCare Remote
Service.
Positive pressure ventilation can be
associated with the following adverse
events: barotrauma, hypoventilation,
hyperventilation or circulatory
impairment.
The SERVO-i Ventilator System should
not be used in MR environments unless
the requirements described in the
SERVO-i MR Environment Declaration
(order no. 66 71 670) are met and an
agreement with MAQUET is signed.
The SERVO-i Ventilator System is not
intended to be used during radiotherapy,
since this may cause system malfunction.
The SERVO-i Ventilator System must not
be used in a hyperbaric chamber unless
it is equipped for hyperbaric oxygenation.
The SERVO-i Ventilator System must not
be used with helium without the Heliox
option.
Only accessories, supplies, and auxiliary
equipment recommended by MAQUET
should be used with the ventilator
system. Use of any other accessories,
spare parts or auxiliary equipment may
cause degraded system performance
and safety.
The power supply cord must be plugged
directly into the mains power outlet
without the use of any multiple socket
outlets. If a multiple socket outlet is used
together with other products, total
leakage current might be exceeded at
earth fault.
CAUTIONS:
In USA, Federal law restricts this device
to sale by or on the order of a physician.
The expiratory channel and expired gas
from the exhaust port may be
contaminated.
Refer to the Installation instructions to
assemble the system or options to obtain
a proper mechanical assembly.
Service, repair and installation must be
performed by MAQUET authorized
personnel only.
When lifting or moving the ventilator
system or parts of the system, follow
established ergonomic guidelines, ask
for assistance, and take appropriate
safety precautions.
Before use, make sure the system
version displayed under Status
corresponds to the system version
described in the User´s Manual.
10
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| Introduction |
1 |
Extra care should be taken when
handling tubes, connectors and other
parts of the patient circuit. The use of a
support arm to relieve the patient from
the weight of the tubing system is
recommended.
When using the MCare Remote Service
function, install the network cable so that
there is no risk of anyone tripping over
it.
Do not leave the patient unattended
when connected to the ventilator.
MAQUET has no responsibility for the
safe operation of the SERVO-i Ventilator
System if the Intended Use requirements
specified in this document are not
followed.
Contact a MAQUET representative
regarding decommissioning of the
equipment.
Disconnect the mains power cable from
the outlet to isolate the ventilator from
mains power.
Do not touch accessible connector
contacts and the patient simultaneously.
Important:
Always use a heat and moisture
exchanger (HME) or equipment to
prevent dehydration of lung tissue.
While the SERVO-i Ventilator System is
in use, the wheels of the carrier shall be
locked and the carrier shall be in a
horizontal position.
Securely attach all cables, etc, to
minimize the risk of unintentional
disconnection.
Extra care should be taken when
handling the gas trolley, as it may tip
over.
All excess fluids must be disposed
according to hospital routines.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
11
| Introduction |
| 1
1.2.2 POWER SUPPLY
WARNINGS!
The power cord must be connected only
to a properly grounded AC electrical
outlet to avoid the risk of electrical shock.
To guarantee reliable battery backup,
two fully charged battery modules must
be installed at all times.
CAUTIONS:
Do NOT use antistatic or electrically
conductive tubing with this system.
Avoid contact with external electrical
connector pins.
Unused module compartments should
always contain an empty module to
protect the electrical connector pins from
spillage and dust.
1.2.3 FIRE HAZARD
WARNINGS!
Keep the system and its gas hoses clear
of all ignition sources.
Do not use the system with worn or
frayed hoses or hoses that have been
contaminated by combustible materials
such as grease or oil.
Oxygen-enriched gas is extremely
flammable: if you detect a burning odor,
disconnect the oxygen supply, mains
power and remove the batteries.
Make sure that both the mains power
outlet and the power supply connector
are accessible.
12
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| Introduction |
1 |
1.2.4 GASES
CAUTION: The system is not intended to
be used with any anesthetic agent.
Important:
Supplied gases shall meet the
requirements for medical grade gases
according to applicable standards.
Maximum levels:
Air
H2O < 7 g/m
Oil < 0.5 mg/m
3
3
Chlorine: must not be detectable
Oxygen
H2O < 20 mg/m
Oil < 0.3 mg/m
3
3
Note: For devices with serial numbers
below 70000 that are updated to version
7.0, the maximum inlet gas pressure will be
reduced (see the Technical data chapter on
page 222).
1.2.5 AUXILIARY EQUIPMENT
CAUTIONS:
Accessories, supplies, and auxiliary
equipment used with the ventilator
should:
- be recommended by MAQUET
- meet EN/IEC 60601-1standards
- meet IEC standards as a whole system
If a scavenging system (i.e. gas
evacuation) is connected to the ventilator,
it must conform to ISO8835-3 guidelines
for subatmospheric pressure and
1
induced flow.
Measurements of parameter values that
have been processed by auxiliary
equipment:
- may be inaccurate if equipment not
authorized by MAQUET is used
- should be discounted if they conflict
with information on the ventilator
screen
- must not substitute for therapeutic or
diagnostic decisions
Note: Applied parts, i.e. equipment making
physical contact with the patient, comprise
nebulizer patient unit and cable, Y Sensor,
CO2 sensor, Edi Catheter and cable and the
Ventilator Breathing System described in
System Flow Chart, Ventilation, Patient
Connection, part no. 66 92 522.
1. If the compressed air is generated by a liquid ring compressor there is a potential risk of chlorine in the supplied
air.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
13
| Introduction |
| 1
1.3 VERSION AND CONFIGURATIONS
This manual applies to version 7.0 of the SERVO-i Ventilator System, which can be delivered
in three configurations: SERVO-i Infant; SERVO-i Adult and SERVO-i Universal.
1.3.1 CONFIGURATIONS
The weight ranges served by each configuration of the SERVO-i Ventilator System.
The SERVO-i Ventilator System can be used in both invasive and non invasive ventilation.
Configuration
ventilation
Universal
NIV = Non Invasive Ventilation
Adult
Weight range
Non invasive ventilationInvasive
Infant
NIV NAVA InfantNIV PC + PS
NIV Nasal
CPAP
0.5 kg – 10 kg0.5 kg – 30 kg3 kg – 30 kgNot Applicable0.5 kg – 30 kgSERVO-i Infant
Not ApplicableNot ApplicableNot Applicable10 kg – 250 kg10 kg – 250 kgSERVO-i Adult
0.5 kg – 10 kg0.5 kg – 30 kg3 kg – 30 kg10 kg – 250 kg0.5 kg – 250 kgSERVO-i
14
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| Introduction |
1.3.2 CONFIGURATION LABELS
NAVA readySERVO-i Universal
NAVASERVO-i Adult
SERVO-i Infant
1.3.3 AVAILABLE OPTIONS
The SERVO-i Ventilator System Configurations - lists the available functions and indicates if
included or optional with each configuration
1 |
Infant
Standard
Configuration
OptionsAdult
Not applicableUniversal
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
15
| Introduction |
| 1
NIV NAVA
NAVA
Nasal CPAP
Non Invasive
Ventilation (NIV)
Y Sensor Measuring
CO2 Analyzer
Nebulizer
Heliox
Stress Index
Open Lung Tool
®
(OLT)
Automode
®
Bi-Vent
VS (Volume Support)
PRVC (incl. SIMV
(PRVC) + PS)
VC (incl. SIMV (VC) +
PS)
PC (incl. SIMV (PC) +
PS)
PS/CPAP
16
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
2 SYSTEM OVERVIEW
TABLE OF CONTENTS
2.2
symbols
| System Overview |
18|Ventilator2.1
19|User Interface - Connections, labels and
22|Navigating the User Interface2.3
28|Patient Unit - Connections, labels and symbols2.4
34|Transport and storage2.5
2 |
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
17
| System Overview |
| 2
2.1 VENTILATOR
The User Interface is used to control ventilator
settings. Settings may be adjusted using
touchpads on the screen or a rotary dial.
Breathing parameters are continuously
measured and controlled. A difference
between the actual measured value of a
parameter and the preset or calculated value
results in the adjustment of gas delivery to
achieve the target value.
The system has two gas modules, one for air
and one for O2. Gases may be supplied by a
medical pipeline system, a compressor, or by
gas tanks.
Ensure that the ventilator is in its locked
position on the cart or holder used, to prevent
unintentional movements.
CAUTION: Lock the wheels if the ventilator
is not to be used for transportation.
6
10
7
10
8
9
1.Air and O2 supply
2.Power cable
3.User Interface
4.Patient Unit
5.Expiratory inlet
6.Servo Duo Guard, viral/bacterial filter
7.Inspiratory outlet
8.Patient circuit
9.Module compartment
10. Emergency air intake
18
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| System Overview |
2.2 USER INTERFACE - CONNECTIONS, LABELS AND SYMBOLS
The User Interface includes:
a screen with active touchpads
fixed keys
rotary dials
2.2.1 USER INTERFACE DIAGRAM
2 |
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
19
| System Overview |
| 2
2.2.2 USER INTERFACE COMPONENTS
Refer to the User Interface Diagram for the location of the following numbered components:
1.Current Patient category
2.Current mode of ventilation
3.Automode On/Off (option)
4.Admit patient/Entered patient data and
admission date
5.Nebulizer On/Off (option)
6.System status parameters
7.Fixed keys
8.Main Rotary Dial—used to select a menu
touchpad or parameter box, to adjust
values, and to confirm settings
9.Special Function Keys—used to start
special ventilatory functions
10. Direct Access Knobs—used for immediate
adjustment of breathing parameters
11. AC Power indicator (green)
12. Standby indicator (yellow)—when the
ventilator is in Standby, a flashing
message, Patient not ventilated, is
displayed on the screen directly above the
word Standby.
Power indicator—green indicates AC
power connected
Battery—indicates ventilator is using
battery power, with estimated
minutes remaining
12 V - indicates that external 12V DC
is connected.
ON/OFF switch
Trigger indication—appears in the
message/alarm field when the patient
triggers a breath
Indicates that the system is
compensated for Heliox.
Infant patient category
DescriptionSymbol
Volume Control with decelerating
flow
Note: The patient unit symbols are
described later in this chapter.
Adult patient category
Volume Control with flow adaptation
Volume Control without flow
adaptation
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
21
| System Overview |
| 2
2.3 NAVIGATING THE USER
INTERFACE
The following subsections provide general
procedures for working with the user interface.
More detailed procedures for specific tasks
are found in later chapters and in the
Appendix.
2.3.1 TOUCH SCREEN
To adjust ventilator settings:
1.Open the desired window by pressing one
of the touchpads on the screen or one of
the fixed keys.
2.Activate the desired touchpad by pressing
it.
The touchpad is now highlighted in white
with a blue frame and it is possible to set
a new value.
3.Turn the Main Rotary Dial to the desired
value.
4.Confirm the settings by pressing the
parameter touchpad or by pressing the
Main Rotary Dial.
5.The touchpad turns grey again indicating
that the new setting has been entered.
6.Press Accept to activate the new settings,
or Cancel to start over.
Important: Different ways of interacting
with the screen will affect its lifetime. Never
use any sharp or pointed objects, such as
ballpoint pens, on the screen.
22
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| System Overview |
2 |
2.3.2 MAIN ROTARY DIAL
SVX-6021_XX
To use an alternative method for adjusting
ventilator settings once the desired menu is
activated:
1.Turn the Main Rotary Dial until the desired
menu touchpad is marked with a blue
frame.
4.Confirm the setting by pressing the Main
Rotary Dial.
The parameter touchpad turns grey again
indicating that a new setting has been
entered.
5.Touch Accept to activate your settings,
or Cancel to start over.
Note: When the defined safety limits for a
given parameter have been reached, the
Main Rotary Dial becomes inoperative for
2 seconds to indicate that a limit has been
reached.
2.Press the Main Rotary Dial to confirm.
The menu touchpad is highlighted in white
with a blue frame, indicating that a new
value can be entered.
3.Turn the Main Rotary Dial to the desired
value or line.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
23
| System Overview |
| 2
2.3.3 FIXED KEYS
There are two groups of fixed keys on the user
interface screen:
The keys in group 1 activate user interface
functions such as Save and access various
screens such as Menu.
The keys in group 2 start special ventilatory
functions
Important: The special ventilatory
functions require continuous supervision.
2.3.4 DIRECT ACCESS KNOBS
Using Direct Access Knobs
To adjust a breathing parameter directly:
1.Turn the Direct Access Knob
corresponding to the parameter you wish
to change until the desired value is
displayed on the screen.
WARNING! When you adjust a breathing
parameter using a Direct Access Knob, the
parameter will change immediately starting
with the next breath; no additional
confirmation is required.
The four dials along the bottom of the User
Interface screen are the Direct Access Knobs.
They permit direct control of four breathing
parameters, which are automatically selected
depending on ventilation mode.
24
The Main Rotary Dial and Direct Access Knobs
become inoperative for 2 seconds when the
user reaches a defined safety limit for the
parameter being adjusted.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| System Overview |
2 |
Direct Access Knobs - Safety
The four Direct Access Knob parameters are
displayed at the bottom of the screen with
color-coded bars that indicate whether the
parameter values are within
generally-recognized safety limits.
Note: When the defined safety limits for a
given parameter have been reached, the
Direct Access Knob becomes inoperative
for 2 seconds to indicate that a safety limit
has been reached.
The figure above shows the following
components.
1.A Direct Access Knob
2.A yellow bar indicating the corresponding
parameter value is outside safety limits; a
text message is displayed accompanied
by an audible signal.
3.A red bar indicating the corresponding
parameter value is significantly outside
safety limits; a text message is displayed
accompanied by an audible signal.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
25
| System Overview |
| 2
2.3.5 MENU KEY
To access the user interface windows:
1.Press the fixed key Menu.
Touchpads leading to the user interface
windows appear.
2.If the touchpad shows a sheet icon, press
the touchpad to open a user interface
window, OR
3.If the touchpad shows an arrow icon,
press the touchpad to display the
submenu.
Press any of the following touchpads.
4.Alarm
5.Review
6.Options
7.Compensate
8.Copy (to Ventilation Record Card)
9.Biomed
10. Panel lock
11. Change patient category (option)
For more information see Appendix • User
interface on page 253.
26
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| System Overview |
2 |
2.3.6 STATUS TOUCHPAD
The Status touchpad indicates the power
supply currently being used by the ventilator
(AC power, battery power, or external 12V DC
power). If the ventilator is running on battery
power, the estimated remaining battery time
in minutes is shown.
To access the status window:
1.Press the Status touchpad.
Touchpads leading to status windows appear.
Press any of the following touchpads.
2.General
3.O2 cell / O2 Sensor
4.Expiratory cassette
5.Batteries
6.Modules - CO2 module (if fitted), Y Sensor
measuring (if fitted), Edi module (if fitted)
7.Installed Options
8.Pre-use check
9.Patient Circuit
CAUTION: When using an external 12V DC
supply, battery modules must be installed
to ensure proper operation.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
27
| System Overview |
| 2
2.4 PATIENT UNIT - CONNECTIONS, LABELS AND SYMBOLS
The patient unit consists of the following components:
gas supplies and their connectors
power supplies and their connectors
connectors for accessories
2.4.1 PATIENT UNIT DIAGRAM
28
0123
-
+
12V
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
| System Overview |
2.4.2 PATIENT UNIT COMPONENTS
Refer to the Patient Unit Diagram for the location of the following numbered components:
B
A
21. Cover, inspiratory channel
1.Handle
2.Gas inlet for air
3.Gas inlet for O
4.Air / Luft
5.O
2
6.Model number
2
22. Expiratory inlet
23. Battery lock
24. Module compartment
25. Nebulizer connector
26. Inspiratory outlet
27. Emergency air intake
7.Serial number
8.Manufacturing information
9.Equipotentiality terminal
10. Label
11. AC power supply connector with fuse
12. Cooling fan with filter
2 |
13. Alarm output connection
14. RS-232 connector
15. External +12V DC inlet
16. Fuse for external DC power supply
17. Optional connector
18. User interface connector
19. RS-232 connector
20. Expiratory outlet
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
29
| System Overview |
| 2
2.4.3 PATIENT UNIT SYMBOLS
& 86
E 01
ExplanationSymbol
CE label—indicates compliance with
the requirements of the Medical
Device Directive 93/42/EEC
CSA label—Indicates compliance
with Canadian and US standards
PCT label - indicates compliance with
Russian standards
Class I equipment, Type B—indicates
classification according to IEC
60601-1/EN 60601-1
Equipotentiality terminal
Note: The equipotentiality terminal
is designed for the connection of a
potential equalization conductor
according to DIN 42 801 and
EN/IEC 60601-1. The function of the
equipotentiality terminal is to equalize
potentials between the system and
other medical electrical devices that
can be touched simultaneously. The
equipotentiality terminal must not be
used for a protective earth
connection.
Nebulizer Connector
-
+
12V
ExplanationSymbol
External 12V DC input
Note: This symbol may be different
depending on Patient Unit version.
Expiratory label—gas flow from
patient.
Inspiratory label—gas flow to patient.
Gas exhaust port label—exhaust gas
flow from ventilator
Note: This port should not be
connected to a spirometer because
the volume through the exhaust port
is not equal to the expired volume
from the patient.
Alarm output connection—external
alarm output communication
Special waste. This product
contains electronic and electrical
components. Discard disposable,
replaced and left-over parts in
accordance with appropriate
industrial and environmental
standards.
Caution
30
RS 232 / Serial port—connector for
data communication.
Note: This symbol may be different
depending on Patient Unit version.
User Interface connector
Note: This symbol may be different
depending on Patient Unit version.
Optional connector / Expansion
Note: This symbol may be different
depending on Patient Unit version.
10A Fuse for external DC power
supply.
Consult Instructions for use
-
Batteries
+
Weight - Patient unit and User
interface.
SERVO-i VENTILATOR SYSTEM V7.0, User´s Manual
Loading...
+ 243 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.