• Servo Ventilator 900 C must be operated only by authorized
personnel who are well trained in its use.
It must be operated according to the instructions in this
Operating Manual.
• After unpacking, the ventilator must be checked and, if
necessary, calibrated.
• All data on pressures for Servo Ventilator 900 C are given in
cm H2O.
1 kPa (kilopascal) ≈ 10 cm H
100 kPa = 1 bar ≈ 1 atm ≈ 1 kgf/cm
100 kPa ≈ 15 psi
• Responsibility for the safe functioning of the equipment
reverts to the owner or user in all cases in which service or
repair has been done by a non-professional or by persons
who are not employed or authorized by MAQUET, and when
the equipment is used for other than its intended purpose.
•A full technical description – including circuit diagrams, parts
list and service data – is contained in the service
documentation, copies of which are held by your supplier.
O
2
2
(kp/cm2)
Connection
• When connected to a patient, the ventilator must never be
left unattended.
•A check on functions must be done before a patient is
connected to the ventilator.
• When anaesthetic gas is metered via a flow meter on the low
pressure inlet, compressed air must not be connected at the
same time.
• To avoid explosion hazards, flammable agents such as ether
and cyclopropane must not be used in this machine. Only
agents which comply with the requirements on nonflammable agents in the IEC standard “Particular
requirements for electrical safety of anaesthetic machines”
are suitable in this machine.
• As this machine is not suitable for use with flammable agents
such as ether and cyclopropane, the use of antistatic
breathing tubes and face masks is not necessary.
The use of antistatic or electrically conductive breathing
tubes when using high frequency electric surgery equipment,
may cause burns and is therefore not recommended in any
application of this machine.
• Never connect or disconnect auxiliary equipment to the outlet
on the rear of the ventilator when the ventilator is connected
to mains.
• All gases must fulfill the specifications for medical grade gas.
The gases supplied must be dry and free from oil and dust.
AirH2O< 5 g/m
• The APNEA ALARM is not intended to and will not monitor
for disconnections.
• The APNEA ALARM is not functional in VOL. CONTR., VOL.
CONTR. + SIGH, PRESS. CONTR. or MAN.
• In the case of a power failure, manual ventilation using a
Servo Ventilator 900 C is possible only with the help of
power supply from external battery. A resuscitator should
always be available, however, as an extra safety measure.
• The SV 900 C is certified, with regard to safety, to be
compatible with electromagnetic environments complying
with IEC 601-1-2. It is the responsibility of the user to take
necessary measures in order to ascertain that the specified
limits are not exceeded as this may impair the safety of the
ventilator.
Such measures should include, but are not limited to:
– normal precautions with regard to relative humidity and
conductive characteristics of clothing in order to minimize the
build-up of electrostatic charges.
– avoiding the use of radio emitting devices in close
proximity of the ventilator, such as high-frequency surgery
apparatus or cordless (mobile) telephones, resulting in a field
level exceeding 3 V/m (IEC 601-1-2).
Magnetic fields of MR equipment having flux densities above
20 mT may cause deactivation of the ventilator functions and
may result in permanent damage to the Servo Ventilator.
• To protect the patient against high pressures, the WORKING
PRESSURE and UPPER PRESS. LIMIT must always be set
at suitable values.
• Do not forget to set the manual ventilation valve to position
AUT after completed manual ventilation. Otherwise the
patient may be hypoventilated without any alarm from the
ventilator. (Not applicable to manual ventilation accessory
with motor).
• When mains supply is switched off or in the case of a mains
power failure, the inspiratory and expiratory valves will
automatically open. This may also occur in the case of an
internal electronic failure. Thus, if the WORKING
PRESSURE is set too high and the gas supply through the
gas supply unit continues, this may result in increased airway
pressure.
• When excess gas is being scavenged, the scavenging
system must meet the following requirements:
– At the point at which the scavenging system is
connected to the ventilator, the sub-atmospheric pressure
must not exceed 0.5 cm H
the breathing system greater than 0.5 l/min.
– With continuous air flows of 30 l/min and 90 l/min for
not less than 5 seconds at the inlet of the anaesthetic
gas scavenging system, the resistance in the system
shall not exceed 0.25 cm H
The Servo Evac 180 basic evacuation unit meets these
requirements.
• If the ventilator is equipped with electronic gas supply unit,
the following applies:
When mains supply is switched off or in the case of power
failure, the gas supply is automatically blocked.
O or cause an induced flow from
2
O and 2.5 cm H2O, respectively.
2
Cleaning
• The ventilator must not be gas sterilized.
• The flow transducers must not be cleaned in a dish washing
machine, by ultrasonic methods or by using agents that
contain aldehydes.
• Agents used for cleaning must have a pH value between
4–8.5.
• Complete cleaning should be done after every 1000 hours
of operation or, at the latest, after every six months.
Service
• The Servo Ventilator 900 C must be serviced at regular
intervals by specially trained personnel. Any maintenance
must be noted in the log book provided for that purpose,
in accordance with national regulations. We recommend that
service is done as a part of a service contract with MAQUET.
• The 1000 hours overhaul shall be done after every 1000
operating hours or, at the latest, every six months. In
addition, the ventilator shall undergo a technical safety check
twice a year, at six months intervals, according to national
regulations.
• Service and repairs on the ventilator may be done only
by MAQUET authorized personnel.
• Only original parts from MAQUET must be used in the
ventilator.
Equipment combinations
• Only MAQUET-approved accessories and auxiliary
equipment may be connected to the ventilator.
• In order to maintain system safety and integrity only
accessories complying with IEC 601-1, or the safety of which
has been verified in another way must be
connected to the signal outputs on the rear of the ventilator.
For details on connections and allowed
voltages, please see Circuit Diagram.
Product information program
This Operating Manual is a part of a comprehensive
information program for Servo Ventilator 900 C.
The program is planned to contain the following:
TrainingAdvisorySlide Series“I am breathingThe Patient’s
InstructionsBookletincludingthrough aABSee
for InstructorsTextbookventilator”Cards
Film andand Poster
Booklet
Front PanelPanel BlockTrainee’s setVideo programs
Flip-chartVideo guide
Video news
®
Servo Ventilator 900 C-Front
Operating instructions
Servo Ventilator 900 C is simple to operate.
Operating instructions are found:
!
On the ventilator in the form of a panel and
cleaning instructions on the lid of the
pneumatic unit.
“Normal” settings are indicated in green on
the front panel. Settings indicated in red
should be used with caution, since these
settings may involve a certain risk for the
patient.
@
In the Brief Operating Instructions in the drawer
under the ventilator.
A log sheet is available with the Brief
Operating Instructions. After certain routines,
e. g. cleaning etc., the person responsible for
the work should complete and sign the log
sheet. The log sheet can then be filed.
#
On a separate wall poster with cleaning
instructions.
$
In this Operating Manual.
The inside of the cover shows a picture of
the ventilator. This picture can be used as a
foldout when reading the manual.
The following information is found in the
corresponding chapters:
The pneumatic unit comprises the gas
conduction system, pressure and flow
transducers and control valves.
The control of flow and pressure is done by a
feed-back system. The transducers continually
measure the flow and pressures. The
information is compared with the panel settings
and a difference between the actual and the
preset values results in correction signals to
the control valves.
For detailed description, see chapter “Basic
principles” in the Training Instructions.
Low pressure
High pressure
Electronic unit
The electronic unit contains a number of plugin PC-boards with the circuits for regulation,
alarms and monitoring.
The unit effects the electronic control of the
pneumatic unit.
Servicing of the ventilator is facilitated by the
spare parts exchange system. The faulty parts
are replaced by factory trimmed exchange
parts.
Service on the electronic unit must be done
by MAQUET, or by MAQUET authorized
personnel only.
Pneumatic unit
1
3
!
Gas connection. The upper inlet, not visible in
the picture, is used for low pressure gas and
the lower inlet is used for high pressure gas.
For anaesthesia, a gas supply unit with three
inlets is available.
@
The gas supply is regulated by a valve so that a
constant pressure is maintained in the bellows.
#
The O2 cell measures the O2 concentration in
the gas.
$
The gas flows through a bacteria filter.
%
The respiratory gases are mixed and stored at
a constant pressure in the bellows.
^
The working pressure is set with an adjustment
screw.
&
A manometer shows the working pressure.
*
A safety valve opens if the bellows is overfilled,
or if the pressure exceeds approx. 120 cm H
O.
2
(
The flow transducer on the inspiration side
measures the gas flow to the patient.
BL
The inspiration valve regulates the inspiratory
gas flow. It is closed during the pause and
expiratory parts of the respiratory cycle.
BM
The pressure transducer on the inspiration side
measures the airway pressure.
BN
The flow transducer on the expiration side
measures the gas flow from the patient.
To prevent condensation, the transducer is
heated to approx. 60°C.
BO
The pressure transducer on the expiration side
measures the airway pressure.
BP
The expiration valve is closed during
inspiration. During the expiratory phase it is
either fully open or regulating the PEEP level.
BQ
The flap valve prevents a back flow of exhaled
gases. It is also necessary for the trigg function.
1
Rear
4
Operating time meter
Equipotential socket
On/Off switch
Fuse holders
Outputs for
auxiliary
equipment
Control
terminal
Recorder
output
Rear of
Servo Ventilator 900 C
To the right on the electronic unit are the mains
inlet socket, fuse holders, On/Off switch,
voltage rating, equipotential socket and
operating time meter.
To the left on the electronic unit are three
15-pole sockets for auxiliary equipment.
The 37-pole socket is used for recording of
pressure and flow curves.
There are labels with the serial number of the
apparatus on both the pneumatic unit and the
electronic unit.
Mains inlet socket
Control terminal
Under the secured lid, there is a connector for
external control of ventilation.
For details see separate Operating Manual.
Chapter 2
Control panel and
ventilation modes
In this chapter:
Panel design
and symbols ...................2:2
Panel functions...............2:3
Working pressure...........2:3
Mode selection ...............2:3
Respiratory pattern ........2:4
I:E ratio............................2:5
Expired minute volume..2:6
Special functions............2:7
Airway pressure .............2:8
SIMV ................................ 2:10
O
alarm........................... 2:11
2
Parameter selection ....... 2:11
2
121
Controlled ventilation .... 2:12
Supported ventilation .... 2:13
Spontaneous ventilation. 2:15
Manual ventilation .......... 2:16
Panel design and symbols
2
For easy operation, the panel is divided into
fields. Each field is described in this chapter.
“Normal” settings are indicated in green on
the front panel. Settings indicated in red should
be used with caution since these settings may
involve a certain risk for the patient.
2
A number of knobs and the EXPIRED
MINUTE VOLUME meter have dual scales.
The blue scale is always to be used when the
low range is selected.
Power on is indicated by a green lamp.
A yellow lamp is either an indication of a
certain setting or a reminder to set an alarm
limit.
A red lamp always indicates an alarm.
! Working Pressure
@ Mode Selection
# Respiratory Pattern
$ Expired Minute Volume
% Special Functions
^ Airway Pressure
& SIMV
* O
Concentration Alarm
2
( Monitoring
Panel functions
Working pressure
The working pressure is set with an adjustment
screw and is read on the manometer
WORKING PRESSURE.
The working pressure must always be set at
a value somewhat higher than the highest
airway pressure.
Mode selection
The Servo Ventilator 900 C can be operated in
8 different modes, which are selected by means
of the mode selector.
The modes are described in special sections
in this chapter.
The inspiratory minute volume is set with the
knob PRESET INSP. MIN.VOL. l/min. The
knob has a locking button on the underside.
A suitable initial value can be calculated by,
for instance, using a Radford nomogram. The
minute volume can be readjusted after
indication from a CO
auxiliary equipment for the Servo Ventilator.
The setting range is 0.4-40 l/min.
The tidal volume =
PRESET INSP. MIN. VOL. l/min
BREATHS/min
The tidal volume can be read on the digital
display.
At extremely high flows it may be necessary
to increase the working pressure in order to
obtain the desired minute volume.
Flow pattern switch
During VOL. CONTR. and VOL. CONTR. +
SIGH, a constant or an accelerating inspiratory
flow can be selected. Constant flow gives a
lower peak pressure and is the pattern
normally used.
A decelerating flow will be obtained in
PRESS. CONTR.
Breaths per minute
The respiratory rate is steplessly adjustable
within the range of 5 to 120 BREATHS/min.
Analyzer 930, which is
2
Inspiration time %
There are 6 fixed inspiration times, 20, 25, 33,
50, 67 and 80% of the breathing cycle.
Pause time %
There are 5 fixed pause times, 0, 5, 10, 20 and
30% of the breathing cycle.
To the left is a conversion table for INSP. TIME
% and PAUSE TIME % to I:E ratio for those
who are more used to working with the latter.
On the basis of an I:E ratio, i.e. the relation of
inspiration time to expiration time, the pause is
allocated to inspiration time.
For example, insp. time 25% and pause time
10% = I:E ratio 35/65 = approx. 1:1.9.
If the ventilator has been set so that insp. time
+ pause time exceeds 80%, then expiration
automatically begins when 20% of the
respiratory cycle remains (safety function).
The pause time is then reduced.
Example 1:
Preset insp. time = 67% ➞ 67% inspiration
Preset pause time = 20% ➞ 13% pause
20% expiration
Example 2:
Preset insp. time = 80% ➞ 80% inspiration
Preset pause time = 30% ➞ 0% pause
20% expiration
The settings are independent of changes in the
patient’s compliance and resistance.
2
5
Inspiration time settings of 33%, 67% or
80% should be avoided in volumecontrolled respiration with accelerating flow
patterns. With these settings patient
triggering may cause the temporary partial
blocking of the inspiration valve which,
however, would be indicated by a minute
volume alarm. Should the valve be blocked,
the situation can be remedied immediately
by switching over to a constant flow.
Inspiratory flow
In mode VOL. CONTR. (constant flow) or SIMV the
inspiratory flow is calculated by:
PRESET INSP. MIN. VOL. l/min
INSP./TIME %
Example:
PRESET INSP. MIN. VOL. l/min= 8 l/min
INSP. TIME %= 25%
Results in an inspiratory flow:= 8 × 100/25
× PRESET INSP. MIN. VOL = inspiratory flow (l/min)
1.5
1.25
= 1.5
= 1.25
= 32 l/min
2
Panel functions
6
Expired minute volume
EXPIRED MINUTE VOLUME is indicated on a
meter which has dual scales: 0-40 l/min and
0-4 l/min.
The scale 0-4 l/min is intended for use when
greater accuracy, e. g. for infants, is needed.
The knobs for LOWER ALARM LIMIT and
UPPER ALARM LIMIT also have the same
dual scales.
The end stop positions for the knobs do not
coincide with the end positions on the scale.
This is correct and no attempt should be made
to alter the range.
The desired scale is selected with the switch
INFANTS/ADULTS.
The alarm limits for expired minute volume
must always be set. If either of the knobs has
been set in end position (out of scale), the
yellow lamp SET MIN. VOL. ALARM flashes.
The alarm limits are normally set at approx.
20% below and above the selected minute
volume. The upper limit may be set at approx.
30% above the selected minute volume in the
mode VOL. CONTR. + SIGH.
The APNEA ALARM is activated with audible
signals and flashing light if the time between
any two consecutive breaths, spontaneous or
mandatory or a combination of the two, is
greater than approximately 15 seconds
(4 breaths per minute or less). The APNEA
ALARM is not intended to and will not monitor
for disconnections. The APNEA ALARM is
operative in CPAP, PRESS. SUPPORT, SIMV
and SIMV + PRESS. SUPPORT.
Failure of gas supply to the ventilator will
result in visual and audible GAS SUPPLY
ALARM signals. The GAS SUPPLY ALARM is
not operative at respiratory rates over 80
breaths/min., and INSP. TIME % 20 or 25.
Panel functions
INSP.EXP. GAS
PAUSEPAUSE CHANGE
HOLDHOLD
Special functions
Under the small hood below the UPPER
ALARM LIMIT knob are three pushbuttons for
special functions.
Inspiratory pause hold
When the pushbutton INSP. PAUSE HOLD is
depressed, the valves close after inspiration
and the pause is prolonged as long as the
pushbutton is depressed.
This provides an exact measurement of the
end inspiratory lung pressure. It may also give
time for an equilibration of the blood gas and
alveolar pressures.
Expiratory pause hold
The valves are closed, after the expiration, as
long as the pushbutton EXP. PAUSE HOLD is
depressed, i.e. a prolonged expiratory pause.
This provides an exact measurement of the
end expiratory lung pressure.
2
7
EXP. PAUSE HOLD must not be used in
SIMV or SIMV + PRESS. SUPPORT mode.
Two inspirations may occur at the same
time. However, the UPPER PRESS. LIMIT
is still functional.
Gas Change
The pushbutton GAS CHANGE is used, for
instance, when it is desirable to rapidly alter the
gas mixture to the patient. The concentration
setting on the vaporizer/gas mixer/flowmeters
must be altered first.
The inspiration and expiration valves open so
that the pressure in the patient circuit is max.
20 cm H2O.
During the gas changing time, the EXPIRED
MINUTE VOLUME meter is zeroed
automatically.
2
Panel functions
8
Airway Pressure
Airway pressure meter
The meter continually indicates the airway
pressure reading.
Upper pressure limit
The knob UPPER PRESS. LIMIT is used to set
the upper limit for airway pressure. When this
pressure limit has been reached, inspiration is
immediately discontinued and there is a
change over to the expiratory phase. Visual
and audible alarms are activated.
Since the inspiration is discontinued, the
complete tidal volume will not be delivered and,
consequently, the LOWER ALARM LIMIT for
EXPIRED MINUTE VOLUME may be
activated.
Upper pressure limit is connected in all
modes.
There are safety catches on the knob at 80
and 100 cm H
PEEP
PEEP is the setting of a Positive End
Expiratory Pressure, e.g. used to counteract
the formation of atelectasis.
The PEEP range is from 0 to 50 cm H2O.
There are safety catches on the PEEP knob at
0 and at 20 cm H
In case of a leakage, e.g. if a tracheal tube
without cuff is being used, the PEEP-setting
may cause self-triggering. To avoid this, the
TRIG. SENSITIVITY should be set at a
somewhat lower value.
The PEEP knob is prepared for NEEP
(Negative End Expiratory Pressure). In order
to use NEEP, it is necessary to connect
auxiliary equipment to the expiration outlet.
The NEEP range is from 0 to -10 cm H
O.
2
O.
2
O.
2
Panel functions
Trig. sensitivity
The knob TRIG. SENSITIVITY is used to set
the value of negative pressure that the patient
must produce in order to trigger a breath.
The TRIG. SENSITIVITY is relative to PEEP.
For instance, PEEP +10 and TRIG.
SENSITIVITY -2 cm H2O mean that the patient
must produce a pressure of -2 cm H
to the setting to trigger a breath.
This corresponds to a pressure of +8 cm H
relative to atmospheric pressure in the patient’s
airways.
The airway pressure, measured on the
expiration side, is compared with the preset
TRIG. SENSITIVITY + PEEP. If the airway
pressure drops below TRIG. SENSITIVITY +
PEEP, a breath is triggered.
If it is desirable for the patient to be able to
easily trigger the ventilator, the knob is set at
green value.
If it is undesirable for the patient to be able to
trigger the ventilator, the knob is set at
-20 cm H
Trig. sensitivity is connected in all modes
except MAN. For further details, see
description on different modes.
O.
2
O relative
2
2
2
9
O
Inspiratory pressure level
The knob INSP. PRESS. LEVEL is used to set
the constant inspiratory pressure, relative to
PEEP, when any of the following modes of
ventilation are used: PRESS. CONTR.,
PRESS. SUPPORT, SIMV + PRESS.
SUPPORT.
A safety catch is located at 30 cm H2O.
The total inspiratory pressure level also
includes PEEP. This is why there is a
marking under the PEEP knob leading to
the INSP. PRESS. LEVEL knob. At a PEEP
of +5 cm H
LEVEL of +15 cm H
pressure level would be +20 cm H
relative to atmospheric pressure.
O and an INSP. PRESS.
2
O, the total inspiratory
2
O,
2
2
10
Panel functions
SIMV
SIMV (Synchronized Intermittent Mandatory
Ventilation) means that the patient gets
mandatory breaths controlled by the ventilator,
and that he also has the possibility of breathing
spontaneously.
The mandatory breaths are synchronized
with the breaths of the patient so that he need
not breathe against the ventilator.
SIMV-rate (SIMV BREATHS/min) is the rate
of the mandatory breaths per minute. The rate
is set between 0.4 and 4 breaths per minute on
the LOW RATE scale, and between 4 and 40
breaths/min on the HIGH RATE scale. The
desired rate is selected by means of a switch.
There are two indicating lamps, one for each
rate.
The tidal volume and the frequency can be
read on the digital display.
Minute volume, respiratory rate, inspiration
time and pause time must be set for the
SIMV mode. These functions determine the
respiratory pattern of the mandatory
breaths. The BREATHS/min must always
be set higher than the SIMV BREATHS/min
to allow time for the spontaneous period.
The SIMV cycle is divided into SIMV period and
spontaneous period. See page 2:14.
Panel functions
O2 alarm
Alarm limits for O2 concentration are set with the
knobs UPPER ALARM LIMIT and LOWER
ALARM LIMIT. The SET O
alarm limits have not been set.
The end stop positions for the knobs do not
coincide with the end positions on the scale.
This is correct and no attempt should be made to
alter the range.
The alarm limits are set as follows:
• Set O2 concentration on the gas mixer/
flowmeters.
• Set the parameter selector at O
and read the value.
• Set the alarm limits at approx. 6% by volume
(3 scale divisions) below and above the O
concentration reading on the digital display.
Visual and audible alarms are activated when
any of the alarm limits is exceeded.
If the O
the ventilator, the O
If not, the LOWER ALARM LIMIT will activate
alarm.
If no O2 cell is mounted, neither digital displays
nor alarm is activated when the parameter
selector is set at O2 CONC. %.
cell is expended but still mounted in
2
2
ALARM flashes if the
2
CONC. %
2
2
cell has to be disconnected.
2
11
Parameter selection
A digital display of the O2 concentration is given
when the selector is set in position O
Other displayable parameters are:
INSPIRED TIDAL VOLUME ml—the volume
provided by the ventilator at each breath.
EXPIRED TIDAL VOLUME ml—the volume
produced by the patient at each breath (with no
leakage in the patient circuit).
EXPIRED MINUTE VOLUME l/min—(also
indicated on the EXPIRED MINUTE VOLUME
meter).
PEAK PRESSURE cm H2O—the pressure at
the end of the inspiration phase.
PAUSE PRESSURE cm H
the end of the pause. This pressure normally
corresponds to the maximal alveolar pressure in
the lungs.
MEAN AIRWAY PRESSURE cm H2O—the
mean value in the patient circuit, continually
measured.
BREATHS/min—the respiratory rate of the
patient. In controlled ventilation, this value
equals the set respiratory rate plus eventual
patient triggered breaths. In the SIMV mode, it is
the sum of the spontaneous and the mandatory
breaths. In the CPAP and PRESSURE
SUPPORT modes, the value represents the
spontaneous breaths/min of the patient.
O—the pressure at
2
CONC. %.
2
Controlled ventilation
2
12
O
2
Insp.Exp.
l/s cm H
Flow Pressure
Flow Pressure
Flow Pressure
WORKING PRESSURE and UPPER
PRESS. LIMIT must always be set at
adequate values. This is extremely
important when treating infants.
Pause
Sigh
Volume controlled ventilation
Volume controlled ventilation (VOL. CONTR.)
ensures that the patient receives a certain preset
tidal volume.
Servo Ventilator 900 C delivers a specific tidal
volume at a specific rate during a specific time
(preset values).
If the patient makes inspiratory efforts so that
the airway pressure falls below the triggering
level, a preset tidal volume will be delivered
earlier and the expired minute volume will
increase.
A constant or an accelerating flow can be
selected.
A patient trigg initiates a breath with the preset
values.
Volume controlled ventilation
+ sigh
In this mode of operation, the Servo Ventilator
900 C delivers a sigh every hundredth breath.
At sigh, double tidal volume is delivered since
the flow is constant and the inspiration time is
doubled.
The first sigh will occur at the second
inspiration after setting the mode selector to
VOL. CONTR. + SIGH.
The UPPER ALARM LIMIT for EXPIRED
MINUTE VOLUME may have to be set
somewhat higher in this mode.
Pressure controlled
ventilation
In the mode PRESS. CONTR., gas is delivered
at a constant pressure during the set inspiration
time.
The pressure is set with the knob INSP.
PRESS. LEVEL.
The flow is decelerating in this mode.
The set inspiratory pressure, respiratory rate,
and inspiration time determine the volume the
patient will receive.
Servo Ventilator 900 C works with a constant
pressure during the entire inspiration.
The minute volume can be read on the
EXPIRED MINUTE VOLUME meter.
Tidal volume can be read on the digital display
with the parameter selector in position INSP.
TIDAL VOLUME or EXP. TIDAL VOLUME.
In case of leakage in the system, e.g. due to
the use of a tracheal tube without cuff, a pause
time of 0% is recommended to ensure that the
measurement of the expired minute volume will
be as correct as possible.
A patient trigg initiates a breath with the preset
values.
Supported ventilation
Trig.
FlowPressure
Pressure supported
ventilation
PRESS. SUPPORT is a spontaneous breathing
mode in which the patient must trigger breaths.
Some applications for PRESS. SUPPORT
are: weaning, patients suffering from asthma
or, in post-operative use, when the patient’s
own breathing efforts are insufficient.
When the patient triggers the ventilator, an
inspiration pressure support is given at a preset
constant pressure.
The ventilator regulates the pressure during
inspiration so that it corresponds to preset
INSP. PRESS. LEVEL + PEEP.
The pressure during expiration drops to
0 cm H
occurs when the flow decreases to 25% of the
peak inspiratory flow.
change from inspiration to expiration, for
instance in the case of a leakage.
has closed the inspiratory valve and the airway
pressure has increased to +3 cm H
the preset INSP. PRESS. LEVEL + PEEP or
after 80% of the preset respiratory cycle.
Because of this safety function the respiratory
rate should always be set with the knob
BREATHS/min.
O, or alternatively to PEEP-level.
2
The change from inspiration to expiration
There are also safety functions to control the
Expiration will then start when the step motor
O above
2
2
13
2
14
Supported ventilation
SIMV cycle 10 sSIMV cycle 10 s
SIMV period 4 sSpontaneous period 6 sSpontaneous period 6 sSIMV period 4 s
FlowPressure
!
A patient trig. during the SIMV period initiates a
mandatory breath. After that, spontaneous
breathing.
SIMV
SIMV (Synchronized Intermittent Mandatory
Ventilation) means that a preset number of
breaths are ventilator controlled, mandatory
breaths (SlMV BREATHS/min). The patient
may breathe spontaneously between those
breaths.
Breathing pattern and tidal volume for the
mandatory breaths are selected with the knobs
PRESET INSP. MIN. VOL., BREATHS/min,
INSP. TIME % and PAUSE TIME %.
The time in seconds for one SIMV cycle is
calculated from the formula
60
preset SIMV freq.
The SIMV cycle is divided into SIMV period
and spontaneous period.
The spontaneous period consists of
spontaneous breathing time if the patient has
triggered a mandatory breath during the
previous SIMV period.
If the patient has insufficient spontaneous
breathing, the maximum time between two
mandatory breaths is just over one SIMV cycle,
(cp 1 to 2 in the picture above).
@
During the SIMV period no breath has been
triggered. The next period begins with a
mandatory breath. After that, spontaneous
breathing.
To ensure adequate ventilation, apnea alarm is
activated if the time between two breaths
exceeds approx. 15 sec., and the minute volume
alarm is activated if any of the preset alarm limits
is exceeded.
Either of these alarms is a command to take
the necessary action (e.g. clinical judgement of
the patient, increased SIMV frequency, altered
trig. sensitivity).
The SIMV period is approx. equal to one
respiratory cycle.
The spontaneous period is calculated as SIMV
cycle-SIMV period.
Example:
BREATHS/min 15 ➞ SIMV period =
SIMV BREATHS/min 6 ➞SIMV cycle =
Spontaneous period ➞10-4 = 6 s
A patient trig during the SIMV period initiates a
breath with the preset values. A patient trig
during the spontaneous period opens the
inspiratory valve, and the patient can breathe
through the ventilator.
60 = 4 s
15
60 = 10 s
6
Supported and spontaneous ventilation
SIMV cycle 10 s
SIMV
period 4 s
Flow Pressure
These settings determine
the SIMV period.
This setting determines the
SIMV cycle.
SIMV cycle
SIMV periodSpontaneous period
Spontaneous
period 6 s
SIMV + pressure support
The main difference between this mode and the
SIMV mode is that the spontaneous, triggered
breaths are pressure supported.
The level of support is set with the control
INSP. PRESS. LEVEL. The level also includes
PEEP.
See also description of pressure support on
page 2:13.
INSP. PRESS. LEVEL
2
15
0 or PEEP-level
TRIG. SENSITIVITY
Flow Pressure
Mandatory breathPressure supported
spontaneous breaths
TRIG. SENSITIVITY
Flow Pressure
PEEP
Care should be taken in setting the preset
BREATHS/min, so that the spontaneous
pressure-supported breaths are not
prematurely time-cycled rather than flow-
cycled to expiration.
Spontaneous ventilation
CPAP
In this mode, CPAP (Continuous Positive Airway
Pressure), the patient breathes spontaneously
through the ventilator at an elevated pressure
level. The positive pressure is set with the
PEEP control.
The TRIG. SENSITIVITY control must be set
at a position which allows the patient to trigger
the ventilator. Upon triggering, the inspiration
valve opens, and the patient can inspire through
the ventilator and control the tidal volume and
respiratory rate.
To avoid self-triggering due to leakage, the
trig. level should be set at a low value.
If CPAP (PEEP) = 0, it will be possible to
check the patient’s ability to breathe
spontaneously without help from the ventilator.
The minute volume, tidal volume and
respiratory rate can be monitored.
2
16
Manual ventilation
T
U
A
N
A
M
Electronic unit
MAN
AUT
Mode selector
If the breathing bag becomes overfilled, the
expiratory valve in the ventilator may go to a
locked position.
To unlock the valve:
• Set UPPER PRESS. LIMIT below 20 cm
H
O.
2
• Squeeze the breathing bag and check that
the pressure in the bag is lowered.
• Set UPPER PRESS. LIMIT back to the
lowest possible value for normal operation.
If the breathing bag becomes overfilled with
the manual ventilation valve in posotition AUT:
• Remove the breathing bag.
• Exchange the manual ventilation valve.
Motor/
valve unit
Manual ventilation
The Manual Ventilation Accessory with Motor
consists of two units:
• Electronic unit
• Motor/valve unit
With this accessory switching between
manual ventilation and other modes is done with
the mode selector.
The valve is automatically set to the correct
position by a motor. The switch-over time is
about 5 seconds. An alarm with audible signals
and red light is activated if the positions of the
valve and mode selector are not the same.
In position
pressure of about 4 cm H
valve closes.
When the breathing bag is squeezed, the gas
flows to the patient via a non-return valve.
Always set UPPER PRESS. LIMIT at the
lowest possible value for normal operation.
The APNEA alarm is inoperative during
manual ventilation.
In the case of a power failure, manual
ventilation is not possible unless an
external power pack is connected.
PEEP cannot be used in manual
ventilation mode.
Set the mode selector to VOL. CONTR.
immediately before connecting the
ventilator to a patient. The reason is to zero
the gas flow reading.
At spontaneous breathing during manual
ventilation the following readings will be
incorrect.
– the displayed values for BREATHS/min.,
EXP. TIDAL VOLUME and EXP.
MINUTE VOLUME.
– the alarm monitoring instrument
EXPIRED MINUTE VOLUME.
The accuracy of the EXPIRED MINUTE
VOLUME meter reading decreases. After
about 5 minutes, the accuracy is about
±2 l/min. To get a correct meter reading,
the breathing bag should be squeezed at
least every 5 minutes.
For spontaneous breathing, the
PRESSURE SUPPORT mode is
recommeded. The INSP. PRESS. LEVEL
should be set at zero or at a low value. This
enables continuous monitoring of pressure
and volume.
MAN., the breathing bag is filled to a
O, then the inspiratory
2
Chapter 3
Patient safety
In this chapter:
Protective devices......... 3:2
Alarm schedule.............. 3:4
3
131
Protective devices
3
Alarms
There are a number of alarms on the ventilator
which protect the patient by alerting ward
personnel to any malfunctions. Alarms are
given with audible signals and flashing red
lights. An audible signal only is given as alarm
for power failure and certain technical
malfunctions.
2
Some audible alarms can be switched off for
a period of approximately 2 minutes. For details
see the alarm schedule on page 3:4.
In addition there are panel indicators with
fixed or flashing yellow lights.
Examples are given in chapter 8,
Troubleshooting, regarding actions for various
alarms.
The patient is protected against high
pressure by an electronic limiter, which is
controlled by the knob UPPER PRESS. LIMIT.
If the value set with this knob is too high or if
the limiting function should fail, the pressure is
limited to the set WORKING PRESSURE.
The WORKING PRESSURE is prevented
from exceeding the set value by a safety valve
and cannot be set at values exceeding
120 cm H
As an extra patient safety protection against
too high an airway pressure, there is a HIGH
PRESSURE PROTECTION circuit incorporated
when either of the modes CPAP, PRESS.
CONTR. or PRESS. SUPPORT is selected.
This protection circuit functions when the
airway pressure in the inspiration channel for
some reason, e.g. coughing, exceeds
30 cm H
level.
O.
2
O above INSP. PRESS. LEVEL/PEEP
2
Important!
The SV 900 is not designed to withstand
severe negative pressures, which, if applied
to the system may damage the internal
pressure transducers rendering the unit
inoperable.
Transducer pressure range
SV 900 up to and including SIN 188499 are
factory equipped with pressure transducers
that will tolerate pressures down to -100 cm
H20(mbar), whereas units from S/N 188500
are factory equipped with pressure
transducers that will tolerate pressures
ranging from -500 cm H20(mbar) to +500
cm H20(mbar).
Earlier units may have been retrofitted with
Transducer Upgrade EM12102/l featuring the
same pressure tolerance as units with S/N
≥188500. Uppgraded units are identified with
a label !Upgrade EMO12/02/l’ attached to
the inside of the pneumatic unit lid.
If closed system suctioning is applied, the
following must be considered:
A suctioning flow exceeding the flow
delivered by the ventilator will result in a
negative pressure affecting the patients
airways as well as the ventilators breathing
system.
Do not use the “Pause hold” function during
the procedure.
All audible alarms, with the exception of
upper pressure limit for airway pressure and
alarm for mains power failure, are switched
off for a period of 2 minutes when the button
2 min is depressed, while the red light
continues flashing. This means that no new
audible alarm will be possible during
approximately 2 minutes.
To switch off the power failure alarm, the
button must be depressed until the audible
alarm stops.
The APNEA ALARM is not intended to
and will not monitor for disconnections.
Loading...
+ 75 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.