Maquet Servo User Manual

x
User's Manual
SERVO-air Ventilator System v2.1

TABLE OF CONTENTS

| TABLE OF CONTENTS |
1 2 3 4 5 6 7 8 9 10
5|Introduction 15|System Overview 37|Operation overview 45|Displaying and saving data 55|Ventilation modes 83|Ventilatory settings and functions
109|Alarm handling 123|Service & Settings 131|Technical data 159|Index
SERVO-airVentilator System v2.1, User's Manual Infologic 1.39
3
| TABLE OF CONTENTS |
4
SERVO-airVentilator System v2.1, User's Manual
Infologic 1.39

1 Introduction Table of contents

| Introduction |
6|Device description1.1
8|Safety guidelines1.2 14|Version and configurations1.3
1 |
SERVO-airVentilator System v2.1, User's Manual
5
| Introduction |
| 1

1.1 Device description

1.1.1 Device components

The ventilator system consists of:
1. a user interface for setting ventilation modes, displaying data and indicating alarms
2. a patient unit for mixing gases and controlling gas delivery
3. a patient circuit for delivering and exchanging gases
1

1.1.2 Intended use

The SERVO-air ventilator system is:
intended for respiratory support, monitoring
and treatment of pediatric and adult patients
to be used only by healthcare providers
to be used only in professional healthcare
facilities and for transport within these facilities
2
6
3
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |

1.1.3 User's Manual

This manual summarizes the functions and safety features of the ventilator system. It is not all-inclusive and should not be seen as a substitute for training.

1.1.4 Cleaning & Maintenance

Please refer to the SERVO-air Cleaning and Maintenance User's Manual.

1.1.5 Servicing Guidelines

CAUTIONS:
Regular service: The ventilator system
must be serviced at regular intervals by Maquet authorized personnel who have received specialized training.
Complete service records: All service
performed on the ventilator system must be recorded in a service log in accordance with hospital procedures and local and national regulations.

1.1.6 Disclaimers

Non-professional servicing:
Maquet has no responsibility for the safe operation of the ventilator system if installation, service or repairs are performed by persons other than Maquet authorized personnel.
Service contract: It is strongly
recommended that all service on the ventilator system should be performed as part of a service contract with Maquet.
Note: If the ventilator system is to be a part of another system it requires an evaluation of the requirements of the IEC 60601-1 standard.
SERVO-airVentilator System v2.1, User's Manual
7
| Introduction |
| 1

1.2 Safety guidelines

Follow these safety guidelines. Additional warnings appear in context throughout this document.
Information is highlighted with Warning, Caution, Important or Note, where:
WARNING! Indicates critical information about a potential serious outcome to the patient or the user.
CAUTION: Indicates instructions that must be followed in order to ensure the proper operation of the equipment.
Important: Indicates information intended as help to operate the equipment or its connected devices easily and conveniently.
Note: Indicates information requiring special attention.

1.2.1 General

WARNINGS!
The ventilator system may be operated
only by authorized personnel who are properly trained in its use. It must be operated according to the instructions in this User´s Manual.
After unpacking, perform a routine
cleaning and a pre-use check.
Always perform a pre-use check before
connecting the ventilator system to a patient.
Secure all tubing and cables to avoid the
risk of unwanted movement of the equipment.
If any of the following occurs, discontinue
use of the ventilator system and contact a service technician:
- unfamiliar pop-up windows on the
screen
- unfamiliar sounds
- any unfamiliar or unexplained event
- alarms that cannot be resolved
Make sure that a resuscitator is readily
available.
The air inlet must not be occluded.
Positive pressure ventilation can be
associated with the following adverse events: barotrauma, hypoventilation, hyperventilation or circulatory impairment.
Ventilation must be started manually
when a patient is connected to the ventilator system.
Keep the ventilator system upright during
use.
8
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |
Do not cover the ventilator system in any
way, since the functioning of the equipment may be adversely affected.
Do not modify or remove any original
parts.
The ventilator system must not be used
during radiotherapy, since this may cause system malfunction.
The ventilator system must not be used
in a hyperbaric chamber.
The ventilator system must be kept away
from magnetic resonance imaging (MRI) equipment.
The ventilator system must not be used
with helium or any gas mixture containing helium.
When the ventilator system is used with
MCare Remote Services, use only network equipment that is safe and complies with the relevant electrical and EMC standards such as IEC 60950.
Note:The network cable is excluded
from this requirement.
Only accessories, supplies, and auxiliary
equipment recommended by Maquet should be used with the ventilator system. Use of any other accessories, spare parts or auxiliary equipment may cause degraded system performance and safety.
Use only the Maquet approved active
humidifiers. Use of non-approved active humidifiers may result in higher gas temperatures and increase resistance in filters.
During humidification, carefully monitor
the airway pressure. Increased airway pressure could result from a clogged
filter. Replace the filter if the expiratory resistance increases or according to the instructions for the filter, whichever comes first.
Service, repair and installation may only
be performed by Maquet authorized personnel.
The ambient sound needs to be taken
into consideration when setting the alarm sound level.
Always disconnect the patient from the
ventilator system when performing operations that involve risk for the patient, such as replacing the O2 cell, dismantling etc.
CAUTIONS:
Never leave the patient unattended when
connected to the ventilator system.
Before use, make sure the system
version displayed under SYSTEM STATUS/General corresponds to the system version described in the User's Manual.
Maquet has no responsibility for the safe
operation of the ventilator system if the requirements specified in Intended use on page 6 are not followed.
When lifting or moving the ventilator
system or parts of the system, follow established ergonomic guidelines, ask for assistance, and take appropriate safety precautions. The weight is specified on the ventilator system.
The air inlet filter must be in place when
the system is running.
SERVO-airVentilator System v2.1, User's Manual
9
| Introduction |
| 1
The expiratory channel and expired gas
from the exhaust port may be contaminated.
If a scavenging system (i.e. gas
evacuation) is connected to the ventilator system, it must conform to ISO 80601-2-13 guidelines for subatmospheric pressure and induced flow.
During operation any water traps must
be checked regularly and if necessary emptied.
All technical documentation is available
for use by Maquet authorized personnel.
When using the MCare Remote Services
function, instal the network cable so that there is no risk of anyone tripping over it.
The ventilator system must not be used
with any anaesthetic agent.
Important:
Portable and mobile radio frequency (RF)
communications equipment can affect medical electrical equipment.
The ventilator system must be installed
and put into service according to the EMC declaration.
Securely attach all cables, etc, to
minimize the risk of unintentional disconnection.
While the ventilator system is in use, the
wheels of the mobile cart must be locked and the mobile cart must be in a horizontal position.
When the ventilator system is connected
to a patient:
- Do not lift or disconnect the expiratory
cassette.
- Continuously monitor the settings and
measurements displayed on the screen.
Always use a heat and moisture
exchanger (HME) or an active humidifier to prevent dehydration of lung tissue.
If a heated patient circuit is not used in
the system, a water trap must be used on the expiratory tube to avoid condensation in the system when an active humidifier is used. During operation the water traps must be checked regularly and if necessary emptied.
Check that the cooling fan intakes are
not covered. Do not place the ventilator system on soft surfaces.
The air inlet filters must be checked
regularly and replaced if necessary.
Use an inspiratory filter when ventilating
a highly infectious patient.
All excess fluids must be disposed of
according to hospital routines.
The emergency air intake must not be
blocked.
Do not disconnect the expiratory
cassette while the ventilator system is in operation; if necessary, disconnect the cassette while in STANDBY.
10
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |
Notes:
Do not simultaneously touch the patient
and any accessible connector contacts.
Do not solely rely on the use of an
external monitor to determine the status of the patient and the ventilator system.
Make sure that the ventilator system is
firmly mounted on the mobile cart.
Make sure that cables and patient circuit
is not obstructed or squeezed due to improper mounting.
Extra care should be taken when
handling tubes, connectors and other components of the patient circuit. The use of a support arm to relieve the patient from the weight of the tubing system is recommended.
Contact a Maquet representative
regarding decommissioning of the equipment.
Turn off the active humidifier during
nebulization.
Expiratory filter connection is mandatory
during nebulization.
A water trap is recommended if a single
heated patient circuit is used.
Thermoshell, expiratory cassette must
be used when using heated expiratory tubing or Expiratory heater Servo Duo Guard.

1.2.2 Power supply

WARNINGS!
The power cord must be connected only
to an AC mains power outlet with protective earth to avoid the risk of electric shock.
The power supply cord must be plugged
directly into the mains power outlet without the use of any multiple socket outlets. If a multiple socket outlet is used together with other products, total leakage current might be exceeded in the event of a fault in the protective earth.
CAUTIONS:
Do NOT use antistatic or electrically
conductive tubing with this system.
Avoid contact with external electrical
connector pins.
SERVO-airVentilator System v2.1, User's Manual
11
| Introduction |
| 1
Important: In case of total loss of power during ventilation, an alarm will sound for 2 minutes. When power is restored, the ventilator system will start in the same state and with the same settings as before the power loss.
Note: When the system is connected to an external power supply, all connected battery modules are being recharged. This does not affect ventilation.
Battery
WARNINGS!
Do not use sharp tools when extracting
the batteries.
To guarantee reliable battery backup,
make sure a battery is in place in slot 2 at all times during ventilation.
Dispose of batteries according to local
regulations and not with ordinary waste.
CAUTION:
The battery modules must be charged before first use.
Do not expose the batteries to water, fire or excessive heat.
Do not crush, disassemble, puncture or short circuit the connector terminals.
One battery can be added to an available slot during operation.
Hold onto the battery strap when inserting a battery in the ventilator system.
Important:
If a battery status message is displayed
on the screen, always go to SYSTEM STATUS/Batteries for detailed information.
Check battery in SYST EM
STATUS/Batteries window to ensure safe battery operation. Always charge the battery before use.
Always replace batteries when the
ventilator system provides notification of imminent expiration or of diminished operating capacity.
When not in use, the ventilator system
should always be connected to the mains power to ensure fully charged batteries.
12
Refer to section Battery status on page 23.
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |

1.2.3 Fire hazard

WARNINGS!
Keep all sources of ignition away from
the ventilator system and the oxygen hoses.
Do not use a ventilator system with worn
or frayed gas supply hoses or hoses that have been contaminated by combustible materials such as grease or oil.
Oxygen-enriched gas is extremely
flammable: if a burning odor is detected, disconnect the oxygen supply and mains power and remove the batteries.
Make sure that both the mains power
outlet and the power supply connector are accessible.

1.2.4 Gases

1.2.5 Auxiliary equipment

CAUTION:
Measurements of numerical values that have been processed by auxiliary equipment:
may be inaccurate if equipment not
authorized by Maquet is used
should be disregarded if they conflict
with information on the ventilator screen
must not be used as a substitute for
therapeutic or diagnostic decisions.
Accessories, supplies, and auxiliary equipment used with the ventilator system must be recommended by Maquet.
WARNING! The ventilator system must not be used with helium or any gas mixture containing helium.
Refer to section Ventilator system on page 134.
SERVO-airVentilator System v2.1, User's Manual
13
| Introduction |
| 1

1.3 Version and configurations

This manual applies to version 2.1 of the SERVO-air ventilator system.

1.3.1 Configurations

The ventilator system can be used in both invasive and non invasive ventilation. The configurations includes adult and pediatric. Refer to section System on page 132.

1.3.2 Available modes and functions

Patient categoryModes/Functions
SIMV
(PC) + PS
(PRVC) + PS
(VC) + PS
X
O
X
AdultPediatric
XXPC
OOPRVC
XXVC
OOBi-Vent/APRV
XXPS/CPAP OOVS OOAutomode
X O
X
OONIV PC OONIV PS
XXNebulizer OOAlarm output connection
14
O = option— = not applicableX = standard
SERVO-airVentilator System v2.1, User's Manual

2 System Overview Table of contents

| System Overview |
16|Ventilator2.1 18|Patient unit2.2 22|Batteries2.3 24|Patient circuit2.4 29|User interface2.5 36|Transport2.6
2 |
SERVO-airVentilator System v2.1, User's Manual
15
| System Overview |
| 2

2.1 Ventilator

The ventilator consists of a user interface and a patient unit.
Air is supplied from ambient air by an internal turbine and O2 may be supplied by a medical pipeline system or by gas cylinder.

2.1.1 Mounting on mobile cart

Lock the wheels.
Release the locking clamp on the mobile
cart.
Stand directly in front of the mobile cart
when mounting the ventilator system.
Tilt the ventilator system to fit the two front
clamps in position on the mobile cart.
Press down the rear end of the ventilator to
fit the rear clamp in position.
Lock the ventilator system to the mobile
cart with the locking clamp.
Ensure that the patient unit is firmly fixed
to the mobile cart via the clamps and locking clamp.
16
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |
2
3
1
1. Patient unit
2. User interface
3. Expiratory inlet
4. Inspiratory outlet
5. Emergency air intake
6. Air inlet
7. Battery compartment
8. Patient circuit
9. AC mains power
10. O2 supply
11. Wheel lock
11
4
6
5
7
8
9
10
Note: Lock the wheels whenever the ventilator system is standing still.
SERVO-airVentilator System v2.1, User's Manual
17
| System Overview |
| 2

2.2 Patient unit

1
2
3
4
5
6 8 9 10
7
14 13 12 11
15
16
18
17
20
21
19
1. On/Off switch
The switch must be pulled downwards before it can be switched.
2. Expiratory outlet
3. Power indicators
4. RS-232 connectors
5. Equipotentiality terminal.
6. AC mains power source connector with
fuse
7. Alarm output connection
8. External +12V DC inlet
9. Fuse for external DC power
10. Ethernet connection
18
11. Battery compartments
12. USB ports
13. Gas inlet for O
14. Gas inlet for air including air inlet filter
15. Inspiratory outlet
16. Emergency air intake
17. Nebulizer connector
18. Expiratory inlet
19. Cooling fan with filter (on both sides)
20. Expiratory cassette
21. Expiratory inlet with moisture trap
2
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |
Important: No other external devices than
a USB memory stick may be connected to the USB ports. Only one memory stick can be used at the same time.

2.2.1 Symbols on patient unit

DescriptionSymbol
CE label—indicates compliance with the requirements of the Medical Device Directive 93/42/EEC
CSA label—Indicates compliance with Canadian and US standards
UDI Label - Unique Device Identification. Refer to section UDI label on page 156.
Rx
ONLY
In USA, Federal law restricts this device to sale by or on the order of a physician.
Type B—indicates classification according to IEC 60601-1
Type BF applied part — indicates classification according to IEC 60601-1
Potential equalization terminal Note: The potential equalization
terminal is designed for the connection of a potential equalization conductor according to DIN 42 801 and IEC 60601-1. The function of the potential equalization terminal is to equalize potentials between the ventilator system and other medical devices that can be touched simultaneously. The potential equalization terminal must not be used for a protective earth connection.
SERVO-airVentilator System v2.1, User's Manual
-
+
12V
-
+
Nebulizer connector
RS-232/Serial port - connector for data communication
Ingress protection, IP21
Fuse (specification) External 12V DC input
Battery
Expiratory gas flow from the patient
Inpiratory gas flow to patient
19
| System Overview |
| 2
DescriptionSymbol
Mains power On/Off Mains connected, batteries charging Gas exhaust port—exhaust gas flow
from ventilator system Network connection
USB connection
Alarm output connection
Special waste
Note: This product contains electronic and electrical components. Discard disposable, replaced and left-over parts in accordance with appropriate industrial and environmental standards.
Caution
Consult instructions for use
DescriptionSymbol
MR Unsafe - keep away from magnetic resonance imaging (MRI) equipment.
20
Consult accompanying documentation
Locked
Unlocked
Serial number
Order number
Weight of patient unit with user interface and ventilator including its safe working load.
Use of ON/OFF switch The switch must be pulled
downwards before it can be switched.
Manufacturer The symbol is accompanied by
manufacturer address and manufacturing date.
SERVO-airVentilator System v2.1, User's Manual

2.2.2 Gas flow through the patient unit

| System Overview |
2 |
8
6
7
9
11
10
1. Air inlet with air inlet filter.
2. Turbine module for ambient air.
3. The check valve prevents the gas to flow
backwards.
4. Gas inlet for O2.
5. The gas module for O2 regulates the O
gas flow.
6. The flow meter measures the gas flow.
7. The pressure of the gas delivered to the
patient is measured by the inspiratory pressure transducer. The transducer is protected by a bacterial filter.
2
5
3 1
2
4
12 13
14
10. The expiratory inlet in the expiratory cassette contains a moisture trap.
11. The ultrasonic transducers in the expiratory cassette measure the outgoing gas flow.
12. The expiratory pressure transducer in the expiratory cassette measures the pressure in the patient circuit. The transducer is protected by a bacterial filter.
13. The expiratory valve in the expiratory cassette regulates the pressure in the patient circuit.
14. Exhaust port.
8. The O2 cell measures the oxygen concentration. The O2 cell is protected by a bacterial/viral filter.
9. The inspiratory channel delivers gas to the patient circuit inspiratory tubing and contains a safety valve.
SERVO-airVentilator System v2.1, User's Manual
Note: The expiratory cassette can be exchanged between different ventilator systems. Always perform a pre-use check after exchanging an expiratory cassette.
21
| System Overview |
| 2

2.3 Batteries

2.3.1 Charging battery modules

Important: The battery modules are
delivered in shipping-mode and must be charged before use.
To charge the battery modules, insert the battery modules in the ventilator system. The ventilator must be connected to mains. The battery modules are charged automatically.
The batteries can also be charged with the External battery charger, SERVO-air.

2.3.2 Handling battery modules

The battery compartment is divided into two slots, 1 and 2.
The ventilator system uses the battery in slot 1 first. Make sure that the battery in slot 2 is in place as a backup at all times during ventilation.
Refer to section Battery on page 12.
Remove the battery from the ventilator
system.
To insert a battery module:
Hold onto the battery strap when inserting
a battery in the ventilator system. When inserting a battery module in slot 1, check that the battery strap for the battery in slot 2 is not folded into the battery compartment.
Ensure that the battery is fully inserted so
that the release button returns to a completely closed position.
1
2
1
2
The battery module in slot 1 may be exchanged during ventilation.
To remove a battery module:
Press the release button to the left and pull
the battery strap.
22
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |

2.3.3 Battery status

Important:
If Replace battery is displayed, the
battery is unreliable, regardless of the operating time displayed under Batteries. In this situation, replace the battery even when the STATUS window indicates that significant operating time remains.
At least one battery module must always
be installed.
Detailed information about batteries is accessed via SYSTEM STATUS/Batteries. There is also an indication in the status bar showing the power supply currently being used by the ventilator system.
If the ventilator system is running on battery power, the active battery in the battery symbol turns yellow and the mains power symbol disappears. The estimated remaining battery time in minutes is always displayed, regardless of the power supply in use.
The following information is displayed under Batteries in SYSTEM STATUS for each connected battery module:
BATTERY CAPACITY – usable backup time
in minutes An estimated backup time is shown in Standby. This estimate may differ from the actual usable backup time during running. Usable backup time depends on set mode and selected ventilation settings.
Note: The presented usable backup time is the sum of the estimated operation time displayed for each battery module minus 20 minutes.
Slot number
Serial number
Remaining operation time in minutes for
each battery
Notification – may be displayed close to the
remaining operation time in minutes.
Remaining battery life
The battery symbol also functions as a shortcut to the window otherwise accessed via SYSTEM STATUS/Batteries.
SERVO-airVentilator System v2.1, User's Manual
23
| System Overview |
| 2

2.4 Patient circuit

Notes:
Turn off the active humidifier during
nebulization.
Expiratory filter connection is mandatory
during nebulization.
A water trap is recommended if a single
heated patient circuit is used.
Thermoshell, expiratory cassette must
be used when using heated expiratory tubing or Expiratory heater Servo Duo Guard.
To ensure that the inspiratory gas
temperature is below 43°C the patient circuit inspiratory tube must be at least
1.2 m to let the gas cool down.
Refer to System Flow Chart, SERVO-air for information regarding patient circuit configurations to be used with the ventilator system.
24
SERVO-airVentilator System v2.1, User's Manual
15 - 22 mm Ø patient circuit, reusable
5
2
4
2
3
| System Overview |
2 |
1
4
1. Inspiratory patient tube
2. Y piece
3. Heat and moisture exchanger (HME)
4. Angled adapter for endotracheal tube
7
6
5. Aerogen Pro
6. Expiratory patient tube
7. Expiratory filter
SERVO-airVentilator System v2.1, User's Manual
25
| System Overview |
| 2
22 mm Ø patient circuit, dual heat disposable
8
6
7
5
6
7
1. Inspiratory patient tube
2. Water autofill
3. Humidification chamber
4. Active humidifier
5. Cuff with temperature port
6. Y piece
2
1
1
3
4
9
10
12
11
7. Angled adapter for endotracheal tube
8. Aerogen Solo
9. Expiratory patient tube
10. Expiratory filter
11. Expiratory heater Servo Duo Guard
12. Thermoshell, expiratory cassette
26
SERVO-airVentilator System v2.1, User's Manual
10 mm Ø patient circuit, single heat reusable
| System Overview |
2 |
9
6
8
5
6
7
1. Inspiratory patient tube
2. Water autofill
3. Humidification chamber
4. Active humidifier
5. Cuff with temperature port
6. Y piece
10
11
2
1
1
3
4
10
7. Angled adapter for endotracheal tube
8. Pressure line connection port
9. Aerogen Pro
10. Expiratory patient tube
11. Water trap
12. Expiratory filter
12
SERVO-airVentilator System v2.1, User's Manual
27
| System Overview |
| 2
10 mm Ø patient circuit, dual heat disposable
11
5
10
9
6
7
8
1. Inspiratory patient tube
2. Water autofill
3. Humidification chamber
4. Active humidifier
5. Cuff with temperature port
6. Extension tube for incubator use
7. Angled Y piece
8. Pressure line connection port
2
1
1
3
4
5
13
15
12
9. Angled adapter for endotracheal tube
10. Y piece
11. Aerogen Solo
12. Expiratory patient tube
13. Expiratory filter
14. Expiratory heater Servo Duo Guard
15. Thermoshell, expiratory cassette
14
28
SERVO-airVentilator System v2.1, User's Manual

2.5 User interface

1

2.5.1 User interface adjustment

2
2. Loudspeaker1. Alarm indicator

2.5.2 Interactive areas

| System Overview |
2 |
The user interface can be adjusted into different positions.
1
2
3
4
5
The user interface is completely touch based and is divided into the following areas:
1. Status bar
2. Quick menu/extended menu
3. Display area
4. Direct access bar/media bar
5. Numerical values
SERVO-airVentilator System v2.1, User's Manual
29
| System Overview |
| 2
Important:
Do not use sharp tools on the screen.
Fluid on the screen can disturb touch
functionality.

2.5.3 Navigating

To navigate the user interface, adjust settings and get support:
Tap (the touchpad changes color when the
navigating is registered).
Tap and hold
Scroll vertically or horizontally
Drag and drop

2.5.4 User support

The user is supported by the following:
Alarm management
Safety scales
Dynamic images
Information texts
Pre-use check instructions
Shortcuts
Prompts
Note:
The following colors are used for settings:
Red — not recommended
Yellow — use with caution
Green — normal
Alarm management
1
2
30
3
4
1. Alarm list
2. Number of active alarms
3. Alarm management checklist
4. Alarm history Refer to chapter Alarm handling on page 109.
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |
Safety scales
5
35
1 3
1. Slide bar
2. Increase/decrease setting
3. Full settings range
4. Accept
5. Cancel
Refer to section Safety scales on page 41.
Dynamic images
2
4
Information texts
1
2
1. Information text is available.
2. Indication that more information is available by scrolling vertically in the middle of the information window.
Shortcuts
Some frequently used functions can be accessed via a shortcut. There are shortcuts to the following windows:
Alarm limits
Patient data
Battery status
Leakage compensation deactivation
Circuit compensation deactivation
Nebulization
The dynamic image illustrates the effects of the changes made to selected ventilation settings.
SERVO-airVentilator System v2.1, User's Manual
Refer to section Symbols on user interface on page 32.
31
| System Overview |
| 2
Prompts
Prompts indicate that input is required.

2.5.5 Symbols on user interface

DescriptionSymbol
Extended menu show/extended menu hide
Start ventilation
Stop ventilation/Standby
Do not push the user interface as the ventilator system may tip over.
Alarm limits/Alarm limits shortcut
Audio pause
Audio paused
Audio off
Audio pause - all alarms, active and inactive are pre-silenced.
Alarm on
32
Alarm off
Check alarms
Alarm sound level
Alarm history
Message
Number of messages
Adult/patient data shortcut
Pediatric/patient data shortcut
AC mains power Missing battery
Unknown battery (not a Maquet battery)
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |
DescriptionSymbol
Charging battery
1
2
Total battery capacity, active battery
xx min
and battery status shortcut External 12V DC power
DescriptionSymbol
Screen layout
Disconnect Service & Settings Biomed
Backup on
Backup off
Circuit compensation on
Circuit compensation off/Circuit
Service
Remote services
Exit
compensation deactivation shortcut Invasive ventilation adult
Nebulization period/Nebulization shortcut
Continuous nebulization/Nebulization
Invasive ventilation pediatric
Non invasive ventilation adult
100
O2 BOOST
shortcut O2 boost locked to 100 %
Pneumatic trigger, pressure/flow
Non invasive ventilation pediatric
Organize
Two overlay loops
Reference loop
Panel locked
Panel unlocked
Loop grid on
Progress
Loop grid off
Full settings range
Compensation
Normal settings range
Configuration
Maneuvers
Recorder Recording waveforms 30 seconds
Media
Modes Patient data
System status
Trends & logs
Views
SERVO-airVentilator System v2.1, User's Manual
Recording waveforms in progress
Camera for taking screenshots
Value not within range Uncertain value Test failed (red)
Test not performed (yellow)
33
| System Overview |
| 2
DescriptionSymbol
Test passed (green)
Accept
Decrease
Increase
Information text Cancel (red)
Close (green) Switch between main/backup modes
Additional values & settings hide/Additional values & settings show
Additional information hide/Additional information show
2.5.6 Symbols on accessories and
packaging
DescriptionSymbol
CE label—indicates compliance with the requirements of the Medical Device Directive 93/42/EEC
Order number Number to identify the production
batch
QTY
Quantity Outer diameter in mm Inner diameter in mm
Indicates the inner diameter of the endotracheal tube
Circumference/lengthFr/cm Use by date
Do not re-use. Single use only.
Rx
ONLY
Do not use if packaging is damaged
Consult accompanying documentation
Keep away from sunlight
Manufacturer The symbol is accompanied by
manufacturer address and manufacturing date.
Manufacturing date
In USA, Federal law restricts this device to sale by or on the order of a physician.
Recyclable material. Recycling must be performed in accordance with appropriate industrial and environmental standards.
34
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |
DescriptionSymbol
Special waste to be disposed of in accordance with appropriate
DescriptionSymbol
The support arm must be folded
during transport. industrial and environmental standards
Pb
Gas cylinder
Locked
Defibrillation proof Type CF applied part - indicates classification according to IEC 60601-1
Type BF applied part — indicates classification according to IEC 60601-1
Type B—indicates classification according to IEC 60601-1
Humidity limitation
+55oC
Temperature limitation
o
-10
C
Fragile — handle with care
Unlocked
Consult instructions for use
Keep away from water
This way up — indicates correct upright position of the transport package
Atmospheric pressure limitation
Do not expose to heat or fire.
Do not expose to mechanical force.
Do not dismantle, open or shred.
SERVO-airVentilator System v2.1, User's Manual
35
| System Overview |
| 2

2.6 Transport

2.6.1 Before intrahospital transport

Before transporting the ventilator system with or without a patient connected, follow facility guidelines and:
Be sure that the patient unit is securely
attached and locked.
Be sure that all accessories such as battery
modules, gas cylinder, and humidifier are securely attached and locked.
Be sure that the gas cylinders are
connected and have sufficient gas.
Be sure that the straps are firmly wrapped
across the center of the gas cylinders so that the cylinders do not move during transport.
Be sure that the battery in slot 2 is fully
charged.
Inspect the resuscitator.
Inspect the mobile cart for damage.
Be sure that the support arm is folded
before transport.

2.6.2 During intrahospital transport

While transporting the ventilator system with or without a patient connected, follow facility guidelines and:
Use the handles on the patient unit.
Transport the bed and the ventilator system
slowly, and watch the patient connection carefully to see that no pulling or other movement occurs.
If triggering problems occur during
intrahospital transport because of extreme vibrations, Pressure Control mode is recommended or to set the trigger so that it is less sensitive.
Be careful not to tip the mobile cart when
crossing an obstacle like a threshold.
On arrival, connect the ventilator system to
mains power and lock the brakes.
CAUTION: Do not load the ventilator equipment asymmetrically on the ventilator system to avoid instability.
36
SERVO-airVentilator System v2.1, User's Manual

3 Operation overview Table of contents

| Operation overview |
38|Workflow summary3.1 38|Pre-use check3.2 39|Patient circuit test3.3 40|Select patient category3.4 40|Select ventilation type3.5 41|Set ventilation mode3.6 42|Set alarm limits3.7 43|Enter patient data3.8 44|Start ventilation3.9 44|Stop ventilation3.10
3 |
SERVO-airVentilator System v2.1, User's Manual
37
| Operation overview |
| 3

3.1 Workflow summary

Turn on the ventilator system, prepare the
patient circuit to be used and perform a pre-use check.
Select patient category.
Select invasive or non invasive ventilation.
Set the ventilation mode.
Check, and if necessary, adjust the alarm
limits.
Enter data for the new patient, including
height, weight, and gender (optional).
Start ventilation and connect the ventilator
system to the patient.
Adjust alarm limits if necessary.
Stop ventilation when desired.

3.2 Pre-use check

The pre-use check contains a number of tests that the ventilator system automatically performs. Refer to section Pre-use check tests on page 140.
Each test starts automatically when the previous test is completed.
To ensure correct system functionality, optimal performance and patient safety, a pre-use check must be performed.
WARNINGS!
Always perform a pre-use check before
connecting the ventilator system to a patient.
Do not connect the ventilator system to
a patient while a malfunction persists.
38
Important:
When the pre-use check is completed,
all sources of alarm signals and alarm conditions have been verified and the alarm system operates correctly.
The volume of the patient circuit used
during pre-use check should be the same as during ventilation. If the patient circuit is changed after the pre-use check is completed, perform a new pre-use check or a patient circuit test.
SERVO-airVentilator System v2.1, User's Manual
| Operation overview |
3 |

3.2.1 Start pre-use check

Connect the ventilator system to a mains
power outlet.
Connect the ventilator system to O2 gas
supply.
Turn the ventilator system on.
Refer to section Patient unit on page 18.
Ta p PRE-USE CHECK in STANDBY.
Tap Ye s in the PRE-USE CHECK window
to start, and follow on-screen instructions.

3.2.2 Complete pre-use check

A symbol and a color marking appear on screen for each pre-use check test, as appropriate: Passed, Failed and Not performed.
Tap OK to confirm or tap Redo test to
restart a pre-use check test.
The ventilator system returns to STANDBY when the pre-use check is completed.

3.3 Patient circuit test

In Standby, the patient circuit test may be performed separately from the pre-use check.
CAUTION: The patient circuit test must be performed with a complete patient circuit, including all accessories (e.g. active humidifier filled with water, filter and nebulizer), that is to be used with the patient.
The patient circuit test measures resistance and compliance in the patient circuit. If the patient circuit is changed and no new patient circuit test is performed, the ventilator will compensate incorrectly based on the measurements of the previous patient circuit. If the correct circuit is not tested, the following risks may arise:
In volume-based modes, the volume
delivered to the patient will be incorrect.
In pressure-based modes, the volume
measured will be incorrect.
Notes:
The status of the two latest pre-use
checks and patient circuit tests is displayed under SYSTEM STATUS/General.
The status of the latest pre-use check
and patient circuit test is also displayed in STANDBY.
SERVO-airVentilator System v2.1, User's Manual
Tap PATIENT CIRCUIT TEST and follow on-screen instructions.
The result from the patient circuit test is displayed in PATIENT CIRCUIT TEST in
STANDBY. Detailed result are displayed in the SYSTEM STATUS/General window.
Important: The patient circuit test does not replace the pre-use check.
39
| Operation overview |
| 3

3.4 Select patient category

Tap patient category in STANDBY. All
available patient categories appear.
Select the appropriate patient category.
The patient data shortcut in the status bar changes accordingly.
Important: Always check the alarm settings after changing the patient category.
Notes:
Changing the patient category affects
the following settings:
- default values for alarm limits
- allowed ranges for alarm limits
- default values for ventilatory settings
- allowed ranges for ventilatory settings
- pressure and flow regulation
- scaling

3.4.1 Change patient category

To change the patient category during ventilation:
Tap the patient data shortcut in the status
bar or tap PATIENT DATA in the quick menu.
Select the appropriate patient category.
Follow on-screen instructions.

3.5 Select ventilation type

Select invasive or non invasive ventilation
in STANDBY.
Note: The default values may have been changed by a previous user.
The default values may have been
changed by a previous user.
40
SERVO-airVentilator System v2.1, User's Manual
| Operation overview |
3 |

3.6 Set ventilation mode

Tap MODES in STANDBY to open the
MODES window.
Select mode.
Note: Tap and hold the tile to access more information about the selected mode.
When a ventilation mode has been selected,
all parameters can be set in the mode settings window.
Tap a parameter to adjust its value.
Tap Accept to confirm, or Cancel to cancel
the settings.

3.6.1 Settings

Important: If one or several settings in the
mode settings window are highlighted in yellow, this indicates that it/they should be considered for adjustment, as the values entered there may have been carried over from the previous mode.

3.6.2 Safety scales

5
35
1 3
1. Slide the bar to the right or left to increase or decrease the settings. The bar displays the safety scale, that is the range that is safe to use for most patients.
2. Tap to incrementally increase or decrease the setting. Tap and hold to rapidly increase or decrease the setting.
3. Tap on full settings range to extend the safety scale setting range.
4. Confirm the setting by tapping Accept.
5. Exit settings without changing by tapping cancel.
2
4
Refer to section Ventilatory settings on page 142.
SERVO-airVentilator System v2.1, User's Manual
41
| Operation overview |
| 3

3.7 Set alarm limits

Tap ALARM LIMITS in the quick menu.
The limits are set in the alarm limit bars in
the ALARM LIMITS window.
Tap the upper or lower value in the selected
alarm limit bar.
A scale appears, tap plus or minus or slide
the bar to set the value.
Confirm each setting by tapping Accept. Tap Autoset all alarms, if desired, to get alarm
limit proposals for the following modes:
-VC
-PC
-PRVC
42
Important: Before accepting Autoset all alarms values, make sure they are
appropriate for the patient. If not, enter settings manually.
SERVO-airVentilator System v2.1, User's Manual
| Operation overview |
3 |
To activate the new alarm limits tap Accept.
Note: Autoset all alarms is not available in supported or NIV modes or in STANDBY because the ventilator system requires patient values in order to propose alarm limits.

3.7.1 Set alarm sound level

The ambient sound needs to be taken into consideration when setting the alarm sound level.
Tap ALARM LIMITS in the quick menu.
Tap alarm sound level.

3.8 Enter patient data

Tap the patient data shortcut in the status
bar or tap PATIENT DATA in the quick menu.
Tap in the selected input field to open a
keyboard or keypad.
Tap Accept to confirm new data.
Enter/edit the following data:
- Patient category
- Name
- Date of birth / ID
- Gender
- Height
- Weight
Note: If gender, height and weight have been entered, predicted body weight will be automatically displayed.
10
Tap the sound level bar to set appropriate
alarm sound level.
Tap Accept.
Refer to section Predicted body weight (PBW) on page 91.
Tap Done when entry is complete.
SERVO-airVentilator System v2.1, User's Manual
43
| Operation overview |
| 3

3.9 Start ventilation

Tap START VENTILATION in STANDBY or START in the quick menu to start ventilation.
WARNING! Ventilation must be started manually when a patient is connected to the ventilator system.

3.10 Stop ventilation

To disconnect and stop ventilation:
Physically disconnect the patient from the
ventilator system.
Tap STANDBY in the quick menu.
Tap and hold STOP VENTILATION to stop
ventilation.
44
SERVO-airVentilator System v2.1, User's Manual

4 Displaying and saving data Table of contents

| Displaying and saving data |
46|Views4.1 49|Displaying waveforms4.2 50|Displaying numerical values4.3 50|Trends and logs4.4 51|Saving and exporting data4.5 53|Ventilator configuration4.6 54|System status4.7
4 |
SERVO-airVentilator System v2.1, User's Manual
45
| Displaying and saving data |
| 4

4.1 Views

The ventilator system offers different views to suit different needs. They are accessed via the quick menu during ventilation.

4.1.1 Basic view

Note: All non invasive ventilation modes
start in the BASIC view.
It is possible to adjust the layout by tapping either SCREEN LAYOUT in the extended menu or VIEWS/Screen layout in the quick menu.
This makes it possible to show or hide the volume waveform.
Refer to section Adapting the waveform display on page 49.
It is also possible to adjust the scaling, sweep speed, appearance of the waveforms and user interface brightness in the SCREEN LAYOUT window.
The view consists of:
two or three waveforms – pressure and flow
waveforms are always present, together with the volume waveform, if desired
a single column of numerical values
46
SERVO-airVentilator System v2.1, User's Manual
| Displaying and saving data |
4 |

4.1.2 Advanced view

The view consists of:
two to three waveforms – pressure and flow
waveforms are always present, together with the volume waveform, if desired
two columns of numerical values
Note: All invasive ventilation modes start in the ADVANCED view.
It is possible to adjust the layout by tapping either SCREEN LAYOUT in the extended menu or VIEWS/Screen layout in the quick menu.
Refer to section Adapting the waveform display on page 49.

4.1.3 Loops view

This view provides a graphical representation of the relationship between pressure-flow, pressure-volume and volume-flow.
The view consists of:
up to three loops – pressure-flow,
pressure-volume and volume-flow
two to three waveforms – pressure and flow
waveforms are always present, together with the volume waveform, if desired
two columns of numerical values
Note: The LOOPS view is not available in non invasive modes.
SERVO-airVentilator System v2.1, User's Manual
It is possible to adjust the layout by tapping either SCREEN LAYOUT in the extended menu or VIEWS/Screen layout in the quick menu.
It is also possible to adjust the scaling, sweep speed, appearance of the waveforms and user interface brightness in the SCREEN LAYOUT window.
Refer to section Adapting the waveform display on page 49.
47
| Displaying and saving data |
| 4
In addition, all loops can be shown or hidden. The maximum number of loops displayed is three, in which case no more than three waveforms may be displayed. The loops may also be displayed with or without a loop grid by tapping Loop grid.
To store a reference loop or see two overlaid loops simultaneously:
1. Tap the reference loop symbol. A reference loop will then be displayed together with a time stamp.
2. Tap the two overlay loops symbol to display the two previous loops.

4.1.4 Family view

4.1.5 Distance view

The view is designed for optimal readability from a distance. Information displayed includes numerical values and waveforms.
There are six large tiles displaying:
five enlarged numerical values
the pressure and flow waveforms
The screen layout cannot be adjusted.
The view has a neutral background image and may be used during family visits to hide the standard user interface.
Displayed information is minimized to:
one column of numerical values
the direct access bar
alarms and messages in the status bar
a dynamic representation (moving bubbles)
showing that ventilation is in progress.
The screen layout cannot be adjusted.
Tap anywhere on the screen for rapid access to the most recently used view.
48
SERVO-airVentilator System v2.1, User's Manual
| Displaying and saving data |
4 |

4.2 Displaying waveforms

The ventilator system can display a minimum of two waveforms and a maximum of three, depending on the view selected.
Pressure and flow waveforms are always mandatory except in the FAMILY view.
The waveforms displayed on the user interface are:
pressure – (yellow)
flow – (green)
volume – depending on view selected and
layout adjustments (blue)

4.2.1 Adapting the waveform display

It is possible to adapt the waveform display via the scaling function.
Tap and hold a displayed waveform.
In the window that opens, adjust the scaling
function manually or use Auto to scale automatically.
To close the window, tap anywhere on the
screen other than the opened window.
Alternatively:
Adjust the layout by tapping either SCREEN
LAYOUT in the extended menu orVIEWS/Screen layout in the quick menu.
Tap the tile shown in the figure directly to
the left of each waveform name.
It is also possible to show or hide non-mandatory waveforms in the SCREEN LAYOUT window.
The sweep speed can be adjusted by tapping Sweep speed and selecting either of three speeds.
In addition, there is a choice under Appearance between filled and unfilled waveforms.
SERVO-airVentilator System v2.1, User's Manual
49
| Displaying and saving data |
| 4

4.3 Displaying numerical values

During ventilation, numerical values (measured or calculated) are displayed on the right side of the screen.
20
10
25
Alarm limits (if applicable) are displayed in
small digits for each numerical value.
Values that are off the scale are replaced
by three asterisks.
Values that are uncertain are indicated by
a single asterisk.
Depending on the view, either one or two columns of numerical values are displayed.
To access additional values, tap the arrow at the right edge of the screen to display all numerical values.

4.4 Trends and logs

4.4.1 Trends

Trend values are stored every 60 seconds and retained for a maximum of 72 hours. Stored events and system changes are also displayed here.
To view trends:
In the extended menu, tap TRENDS &
LOGS /TRENDS.
To adjust the time resolution, tap the
number of hours displayed.
The time valid for the cursor position is
displayed. If events have been stored, their number is displayed in the circle shown in the figure and an explanation appears to the left of this circle.
If a recording is saved at a time
corresponding to the cursor position, a recorder is displayed. To view the recording, tap this recorder.
50
Tap Organize trends to place the trends in
the desired order by dragging and dropping the different trended values presented.
Refer to section Trends on page 147.
SERVO-airVentilator System v2.1, User's Manual
| Displaying and saving data |
4 |

4.4.2 Logs

To view the event log:
In the extended menu, tap TRENDS &
LOGS /LOGS.
Scroll among all the events listed.
The LOGS window offers a search function.
Tap the text field to open the keyboard and enter a search word. To display only log items that contain the search word entered, tap Filter. Tap again to deactivate the filter.
Use the backspace arrow to delete the
search word.
Each event includes the event time and date. The event log is cleared when a new patient is admitted.

4.5 Saving and exporting data

Data can be saved in a number of ways:
as screenshots
as recordings
as files for export including event log, trends
and both the above.
The screenshots and recordings are stored under MEDIA.
Data can later be exported to a USB memory stick.

4.5.1 Saving screenshots

To save a screenshot, tap the camera in the status bar.
Refer to section Event log on page 148.
SERVO-airVentilator System v2.1, User's Manual
The screenshot will be stamped with the date and time it was taken and saved under the Saved screens tab in the MEDIA window.
There is space for 40 screenshots under this tab. When the space is full, the next screenshot taken will erase the oldest one.
51
| Displaying and saving data |
| 4

4.5.2 Viewing saved screens

To view screenshots, tap MEDIA in Saved screens. Choose the relevant screenshot
displayed at the bottom of the window. If there are more than ten screens saved, scroll to the right to find more.

4.5.3 Recording waveforms

To make a recording, tap the recorder (not available in Standby) in the status bar.
A 30 second long recording will be made starting 15 seconds before, and lasting until 15 seconds after the time the recording was initiated. A blue progress bar will be displayed under the recorder while the recording is being made.

4.5.4 Viewing recordings

To view recordings, tap TRENDS & LOGS/MEDIA /Recordings in the extended
menu. Choose the relevant recording displayed at the bottom of the window. If more than ten recordings have been saved, scroll to the right to find more.
The cursor (pale green) is positioned on the dotted line indicating the middle point of the recording. It is activated by moving it or by pressing the arrows to the right of the recorder seen above the dotted line. The values at the cursor position are displayed in digits to the right of the waveform name in the recording window.
When viewing a recording, it is also possible to view the settings by tapping Settings at the bottom left of the window. This will open a list of the actual parameter settings in use at the time the recording was initiated.
The recording will be stamped with the date and time that it was initiated and will be saved under the Recordings tab in the MEDIA window. All settings applying at the time the recording is initiated will also be saved.
There is space for 40 recordings under this tab. When the space is full, the next recording made will erase the oldest one.
52
SERVO-airVentilator System v2.1, User's Manual
| Displaying and saving data |
4 |

4.5.5 Exporting and deleting data

To export or delete screenshots or recordings, tap TRENDS & LOGS/MEDIA /Export & Delete in the extended menu.
Alternatively, tap the Export & delete tab.
Both screenshots and recordings can be selected for export or deletion.
The additional choice of exporting files that include other material is also given at the bottom right of this window. To use this function, tap Export files.

4.6 Ventilator configuration

The ventilator system will always start up with the stored configuration settings.
To view the stored configuration settings, tap CONFIGURATION in the extended menu:
The following configurations can be viewed:
Alarms
General
Units
Startup configuration
The alarms configuration can be viewed for each of the patient categories. The other configurations do not vary with patient category.
Note: No editing can be done under CONFIGURATION.
Refer to chapter Service & Settings on page 123.
The following data will be exported to a USB memory stick:
Event log
Trends
Saved screens & recordings
Important: Only one USB memory stick may be connected to the USB ports at the same time.
SERVO-airVentilator System v2.1, User's Manual
53
| Displaying and saving data |
| 4

4.7 System status

To view the current status of the ventilator system:
Tap SYSTEM STATUS in the quick menu
in Standby.
Tap SYSTEM STATUS in the extended
menu during ventilation.
The SYSTEM STATUS window that opens contains:
1. Installed options
2. General
3. Batteries
4. Expiratory cassette
5. Pre-use check
6. O2 cell
7. Patient circuit
8. Turbine
54
SERVO-airVentilator System v2.1, User's Manual

5 Ventilation modes Table of contents

| Ventilation modes |
56|Introduction5.1 57|Pressure Control (PC)5.2 59|Pressure Regulated Volume Control (PRVC)5.3 61|Volume Control (VC)5.4 63|Bi-Vent/APRV5.5 65|Pressure Support (PS)/CPAP5.6 68|Volume Support (VS)5.7 71|Automode5.8 74|SIMV5.9 78|Non Invasive Ventilation (NIV)5.10 80|NIV Pressure Control (NIV PC)5.11 81|NIV Pressure Support (NIV PS)5.12
5 |
SERVO-airVentilator System v2.1, User's Manual
55
| Ventilation modes |
| 5

5.1 Introduction

5.1.1 Safety guidelines

Not all safety guidelines apply to all modes.
WARNINGS!
Autotriggering should be avoided. Do not
set the trigger level too low.
The following warning apply to non
invasive ventilation (NIV) modes only:
- Avoid high inspiratory pressure as it
may lead to gastric overdistension and risk of aspiration. It may also cause excessive leakage.
Be sure to set alarm limits as appropriate for each mode, especially those for:
- expired minute volume
- apnea time
- airway pressure
Important:
To protect the patient’s lungs from
excessive pressure it is important to set the upper pressure limit to a suitable value.
It is important to avoid leakage so as to
ensure the proper functioning of modes such as:
-PRVC
-VS
- Automode PRVC VS
- SIMV (PRVC) + PS
The circuit compensation function should
be used – it is important to make sure that the compressible volume of the patient circuit is not changed after the pre-use check/patient circuit test has been performed (e.g. filling an active humidifier with water or connecting a filter after the test has been performed).
56
Note:
The ventilator system is delivered preset with the following configuration options:
Ventilatory settings are based on either
minute volume or tidal volume.
Ventilatory settings are based on either
I:E ratio or inspiration time.
SERVO-airVentilator System v2.1, User's Manual

5.2 Pressure Control (PC)

| Ventilation modes |
5 |
Pressure Control (PC):
delivers a constant pressure over a preset
inspiratory time and at a preset respiratory rate
delivers the inspiration with a decelerating
flow
changes in lung or thorax resistance or
compliance will affect the volume delivered
1234
567
The following parameters are set:
1. Oxygen concentration (%)
2. PEEP (cmH2O)
3. Respiratory rate (b/min)
4. PC above PEEP (cmH2O)
The delivered volume is dependent on the pressure above PEEP, lung compliance and resistance in the patient circuit and airways. This means that the tidal volume can vary.
The flow during inspiration is decelerating. The patient can trigger extra breaths. As the delivered tidal volume can vary, it is very important to set alarm limits for the minute volume to adequate levels.
PC ventilation is often preferred when there is leakage in the patient circuit, e.g. due to an uncuffed endotracheal tube, or in situations where the maximum airway pressure must be controlled.
If a patient tries to exhale during inspiration, pressure increases. When it increases 3 cmH2O above the set inspiratory pressure level, the active expiratory valve opens and regulates the pressure down to the set inspiratory pressure level. If the pressure increases to the set upper pressure limit, e.g. if the patient is coughing, the expiratory valve opens and the ventilator system switches to expiration.
5. I:E ratio or Inspiration time (s)
6. Inspiratory rise time (% or s)
7. Trigger
Refer to section Settings on page 41.
SERVO-airVentilator System v2.1, User's Manual
57
| Ventilation modes |
| 5

5.2.1 Pressure Control in detail

The circles in the figure indicate patient triggering.
PC ensures that the preset inspiratory
pressure level is constant throughout inspiration. Breaths are delivered in accordance with the preset respiratory rate, inspiration time and inspiratory pressure level, resulting in a decelerating flow.
The preset pressure level is controlled by
the ventilator system. The resulting volume depends on the set pressure level, the inspiration time and the mechanical properties of the patient's lungs during each breath.
Inspiration starts in accordance with the
preset respiratory rate or when the patient triggers.
Expiration starts:
After the termination of the preset
inspiration time.
If the upper pressure limit is exceeded.
58
SERVO-airVentilator System v2.1, User's Manual

5.3 Pressure Regulated Volume Control (PRVC)

| Ventilation modes |
5 |
Pressure Regulated Volume Control (PRVC):
combines the advantages of Volume
Control and Pressure Control by delivering a preset tidal volume with a decelerating inspiratory flow at a preset respiratory rate
maintains the lowest possible constant
pressure throughout inspiration
the inspiratory pressure of a breath will
never exceed 5 cmH2O below the upper pressure limit
The ventilator system can be configured so that either tidal volume or minute volume is set.
1234
567
The flow during inspiration is decelerating. The patient can trigger extra breaths.
5 cmH2O
The first breath is a volume controlled test breath with the pause time set to 10 %. The measured pause pressure of this breath is then used as the pressure level for the following breath.
Following the initial breath, the ventilator system calculates and continuously regulates the pressure needed to deliver the preset tidal volume.
An alarm is activated if the set target volume cannot be delivered due to the fact that the pressure required to deliver it is higher than 5 cmH2O below the set upper pressure limit.
The following parameters are set:
1. Oxygen concentration (%)
2. PEEP (cmH2O)
3. Respiratory rate (b/min)
4. Tidal volume (ml) or minute volume (l/min)
5. I:E ratio or Inspiration time (s)
6. Inspiratory rise time (% or s)
7. Trigger
Refer to section Settings on page 41.
The ventilator system delivers a preset tidal volume. The pressure is automatically regulated to deliver this volume but limited to 5 cmH2O below the set upper pressure limit.
SERVO-airVentilator System v2.1, User's Manual
Refer to section Leakage compensation on page 94.
59
| Ventilation modes |
| 5

5.3.1 PRVC in detail

The circles in the figure indicate patient triggering.
PRVC ensures a set target minute
ventilation to the patient. The target volume is based on settings for tidal volume, respiratory rate and inspiration time.
The inspiratory pressure level is constant
during each breath, but automatically adapts in small increments on a breath-by-breath basis to match the mechanical properties of the patient's lungs, thus ensuring delivery of the target volume.
Inspiration starts in accordance with the
preset respiratory rate or when the patient triggers.
Expiration starts:
After the termination of the preset
inspiration time.
If the upper pressure limit is exceeded.
60
SERVO-airVentilator System v2.1, User's Manual

5.4 Volume Control (VC)

| Ventilation modes |
5 |
Volume Control (VC):
delivers a preset tidal or minute volume over
a preset inspiratory time and at a preset respiratory rate, regardless of changes in lung or thorax resistance or compliance
maintains a constant flow with varying peak
pressure
The ventilator system can be configured so that either tidal volume or minute volume is set.
1234
5 6 7 8
The airway pressure is dependent on the tidal volume, the inspiration time and the resistance and compliance of the respiratory system. The set tidal volume will always be delivered. An increase in resistance and decrease in compliance will lead to an increased airway pressure. The delivered pressure can vary, so in order to protect the patient's lungs from excessive pressure, it is very important to set the upper pressure limit to a suitable value.
Patients may trigger extra breaths if they can overcome the set trigger level.
Flow adaptation
Patient inspiratory efforts can also result in a higher inspiratory flow and tidal volume than were preset. This is because the ventilator system enables the patient to modify both flow rate and timing.
The following parameters are set:
1. Oxygen concentration (%)
2. PEEP (cmH2O)
3. Respiratory rate (b/min)
4. Tidal volume (ml) or minute volume (l/min)
5. I:E ratio or Inspiration time (s)
6. Pause time (% or s)
7. Inspiratory rise time (% or s)
8. Trigger
Refer to section Settings on page 41.
SERVO-airVentilator System v2.1, User's Manual
Thus, if the patient demands a higher flow than the calculated constant flow, the system will sense any sudden pressure drop of > 3 cmH2O and temporarily enables PS to deliver a higher flow adapted to patient demand.
61
| Ventilation modes |
| 5

5.4.1 Volume Control in detail

The circles in the figure indicate patient triggering.
VC ensures a preset tidal volume during a
preset inspiratory time at a preset respiratory rate.
The inspiratory flow is constant or linearly
decelerating and depends on the ventilatory settings.
Inspiration starts in accordance with the
preset respiratory rate or when the patient triggers.
If the patient makes an inspiratory effort
during the inspiratory period, when flow adaption is enabled and flow pattern is 100%, the ventilator system will switch to PS to satisfy the patient´s flow demand, as shown in the second breath in the figure.
Expiration starts:
When the preset tidal volume is delivered
and after the preset pause time.
When the flow returns to the set value after
the preset tidal volume is delivered and after the preset pause time (flow adaptation). The patient is however always guaranteed an expiration time corresponding to at least 20 % of the total breath.
If the upper pressure limit is exceeded.
62
SERVO-airVentilator System v2.1, User's Manual

5.5 Bi-Vent/APRV

| Ventilation modes |
5 |
Bi-Vent:
is a time-cycled, pressure-limited mode that
allows spontaneous breathing throughout the entire ventilatory cycle
has two time-cycled pressure levels and
switches between these two levels. The patient can breathe spontaneously at both these levels and it is possible to give Pressure Support at both levels.
Airway Pressure Release Ventilation (APRV):
is a time-cycled, pressure-limited mode that
allows spontaneous breathing throughout the entire ventilatory cycle
alternates between two levels of positive
airway pressure, with the main time on the high level and a brief expiratory release to facilitate ventilation
differs from Bi-Vent in that it uses an inverse
I:E ratio
1234
5 6 7
9 10
8
The following parameters are set:
1. Oxygen concentration (%)
2. Pressure at the lower pressure level (PEEP)
3. Pressure at the higher pressure level (Phigh) (cmH2O)
4. Time at the higher pressure level (Thigh) (s)
5. Time at the lower pressure level (TPEEP) (s)
6. PS above Phigh (cmH2O)
7. PS above PEEP (cmH2O)
8. End inspiration (%)
9. Inspiratory rise time (s)
10. Trigger
Refer to section Settings on page 41.
SERVO-airVentilator System v2.1, User's Manual
63
| Ventilation modes |
| 5

5.5.1 Bi-Vent/APRV in detail

The circles in the figure indicate patient triggering.
1. Bi-Vent/APRV cycle = Thigh + TPEEP
2. PEEP
3. Phigh
4. PS above PEEP
5. PS above Phigh
Each Bi-Vent/APRV cycle is regarded as autonomous and therefore most of the measured values are updated every cycle, i.e. minute volume, respiratory rate, mean pressure and end expiratory pressure. Associated alarms are also handled for every cycle.
Bi-Vent/APRV allows for spontaneous breathing/PS ventilation at two different pressure levels. These basic levels are individually set, as well as the time in seconds at each level. The ventilator system always tries to synchronize with the patient's breathing. The main difference between Bi-Vent and APRV is the inverse I:E ratio in APRV.
Since Bi-Vent/APRV is basically a controlled
At extreme settings, the update of measured values and alarms will show a mandatory frequency dependence even in the face of preserved spontaneous breathing.
As a result of switching between two different pressure levels, the tidal volumes may vary significantly between different breaths. This may also be the case for etCO2 (end tidal CO2) concentration.
mode of ventilation, apnea alarm and backup ventilation are not available. It is also very important to set the lower and upper alarm limit for expired minute volume.
1
5
3
4
2
64
SERVO-airVentilator System v2.1, User's Manual

5.6 Pressure Support (PS)/CPAP

Pressure Support (PS)/CPAP:
is initiated by the patient, who controls the
respiratory rate and tidal volume
delivers ventilator support using the preset
pressure level and with a decelerating flow
provides backup (PC) ventilation in case of
apnea
Continuous Positive Airway Pressure (CPAP):
is initiated by the patient and works exactly
like PS except that the Pressure Support level is set to zero
maintains positive pressure in the airways
at all times
is effectively a spontaneous breathing mode
with continuous positive pressure to keep the airways open
PS is thus a patient-initiated breathing mode in which the ventilator system supports the patient with a set constant pressure.
The ventilator system provides this constant preset pressure when activated by patient effort. The patient determines the frequency and duration of the breaths, which have a decelerating flow pattern. Duration of inspiration can be adjusted with the help of the End inspiration setting.
| Ventilation modes |
1234
5 6
8 9 10
7
The following parameters are set:
1. Oxygen concentration (%)
2. PEEP (cmH2O)
3. PS above PEEP (cmH2O) (PS level)
4. End inspiration (%)
5. Inspiratory rise time (s)
6. Trigger
7. Apnea time (s)
8. Backup respiratory rate (b/min)
9. Backup PC above PEEP (cmH2O)
10. Backup I:E or Ti (s)
Refer to section Settings on page 41.
The higher the preset inspiratory pressure level from the ventilator system, the more gas flows into the patient. As the patient becomes more active, the PS level may be gradually reduced.
5 |
CPAP may be seen as a special case of PS in which the inspiratory pressure level is set to zero and is used when the patient is breathing spontaneously.
SERVO-airVentilator System v2.1, User's Manual
Always set the apnea time that is appropriate to the individual patient situation. If the apnea alarm limit is reached, the ventilator system will automatically switch to backup ventilation (PC).
The alarm should alert staff to take action by either returning to a supported mode or changing to a controlled mode of ventilation.
65
| Ventilation modes |
| 5
It is also very important to set the lower and upper alarm limits for expired minute volume.
The inspiratory rise time should be set to a comfortable value for the patient. In PS, the inspiratory rise time should normally be increased.
The End inspiration setting is important to patient comfort and ventilator synchronization with the patient. If the patient's expiratory resistance is high, the End inspiration setting should be raised to guarantee enough time for expiration.
It is important to monitor how this affects the tidal volume.
66
SERVO-airVentilator System v2.1, User's Manual

5.6.1 PS/CPAP in detail

The circles in the figure indicate patient triggering.
| Ventilation modes |
5 |
PS ensures that a preset inspiratory
pressure level is constantly maintained in response to patient effort.
The preset pressure level is controlled by
the ventilator system, while the patient determines the respiratory rate and inspiration time.
Inspiration starts when the patient triggers
a breath and gas flows into the lungs at a constant pressure. Since the pressure provided by the ventilator system is constant, the flow will decrease until the level set for End inspiration is reached.
For CPAP, inspiration starts upon patient
effort.
Expiration starts:
When the inspiratory flow decreases below
a preset fraction of the peak inspiratory flow (End inspiration)
If the upper pressure limit is exceeded
If the maximum time for inspiration is
exceeded Refer to section Functions in ventilation modes on page 146.
If the flow drops to a range between 25 %
of peak flow and the lower limit for End inspiration, and remains within the range for a period longer than 50% of the time elapsing between inspiration start and the point when the range was entered.
SERVO-airVentilator System v2.1, User's Manual
67
| Ventilation modes |
| 5

5.7 Volume Support (VS)

Volume Support (VS):
is initiated by the patient, who controls the
respiratory rate
delivers ventilator support with a variable
peak pressure and decelerating flow to guarantee the preset tidal volume
the inspiratory pressure of a breath will
never exceed 5 cmH pressure limit
provides backup (PRVC) ventilation in case
of apnea
A patient-adapted constant inspiratory support is supplied when activated by patient effort. The resulting volume is continuously monitored and the constant inspiratory pressure automatically adjusts to the required level. The patient determines the frequency and duration of the breaths which have a decelerating flow pattern.
O below the upper
2
1234
56
89710
The following parameters are set:
1. Oxygen concentration (%)
2. PEEP (cmH2O)
3. Tidal volume (ml)
4. End inspiration (%)
5. Inspiratory rise time (s)
6. Trigger
7. Apnea time (s)
8. Backup respiratory rate (b/min)
9. Backup tidal volume (ml)
10. Backup I:E or Ti (s)
68
Refer to section Settings on page 41.
SERVO-airVentilator System v2.1, User's Manual
| Ventilation modes |
5 |
If patient activity increases, the inspiratory support will decrease provided that the set tidal volume is maintained. If the patient breathes below the set tidal volume, the inspiratory support will increase.
5 cmH2O
The initial breath provides support with 5 cmH2O. Following this breath, the ventilator system calculates and continuously regulates the pressure needed to deliver the preset tidal volume.
An alarm is activated if the set target volume cannot be delivered due to the fact that the pressure required to deliver it is higher than 5 cmH2O below the set upper pressure limit.
In this mode it is also important to set the apnea time that is appropriate to the individual patient's situation. If this time is reached, the ventilator system will automatically switch to backup ventilation.
Refer to section Leakage compensation on page 94.
SERVO-airVentilator System v2.1, User's Manual
69
| Ventilation modes |
| 5

5.7.1 Volume Support in detail

The circles in the figure indicate patient triggering.
VS ensures delivery of a set target tidal
volume following patient effort by providing inspiratory pressure support that is adapted to the patient.
The inspiratory pressure level is constant
during each breath, but alters in small increments, on a breath-by-breath basis, to match the patient´s breathing ability and the mechanical properties of the lungs.
Inspiration starts when the patient triggers.
Expiration starts:
When the inspiratory flow decreases below
a preset fraction of the peak inspiratory flow (End inspiration)
If the upper pressure limit is exceeded
If the maximum time for inspiration is
exceeded Refer to section Functions in ventilation modes on page 146.
If the flow drops to a range between 25 %
of peak flow and the lower limit for End inspiration, and remains within the range for a period longer than 50% of the time elapsing between inspiration start and the point when the range was entered.
70
SERVO-airVentilator System v2.1, User's Manual
| Ventilation modes |
5 |

5.8 Automode

In Automode, the ventilator system shifts automatically between controlled and supported ventilation, allowing better patient-ventilator interaction. When the patient is making a breathing effort, the ventilator system immediately switches to a supported mode of ventilation. If the patient is not making any breathing effort, the ventilator system will return to the controlled mode and deliver controlled breaths.
The parameters for each Automode combination are adjusted in the settings window and are basically the same as those for the relevant controlled or supported mode. Refer to section Settings on page 41.
Automode PC PS:
is an interactive mode automatically
switching between the controlled mode PC and supported mode PS based on patient triggering
delivers controlled breaths in the absence
of patient breathing effort, switching to supported breaths when a breathing effort is detected
serves as an aid to starting the weaning
period
adapts to the patient’s breathing capacity
Automode PRVC VS:
is an interactive mode automatically
switching between the controlled mode PRVC and supported mode VS based on patient triggering
delivers controlled breaths in the absence
of patient breathing effort, switching to supported breaths when a breathing effort is detected
serves as an aid to starting the weaning
period
adapts to the patient’s breathing capacity
In this combination, the first supported breath delivered to the patient has the same pressure level as the preceding PRVC breath.
Automode VC ⇄ VS:
is an interactive mode automatically
switching between the controlled mode VC and supported mode VS based on patient triggering
delivers controlled breaths in the absence
of patient breathing effort, switching to supported breaths when a breathing effort is detected
serves as an aid to starting the weaning
period
adapts to the patient’s breathing capacity
In this combination, the ventilator system uses the plateau pressure in the VC breath as a reference pressure for the first VS breath.
SERVO-airVentilator System v2.1, User's Manual
71
| Ventilation modes |
| 5

5.8.1 Automode in detail

1. The ventilator system starts in PC, PRVC or VC mode. If the patient triggers a breath, the ventilator system will turn to the relevant supported mode to encourage the patient's respiratory drive.
2. If the patient is breathing adequately:
In VS, the ventilator system adjusts the
a.
inspiratory pressure level on a breath-by-breath basis to ensure delivery of the preset target volume.
b. In PS, the ventilator system ensures
that the preset inspiratory pressure level is maintained throughout inspiration.
3. The ventilator system initially adapts with an increasing apnea time.This means that for the spontaneously triggering patient, the apnea time increases successively until the level set in the settings window for the maximal apnea time parameter is reached after 10 consecutive spontaneously triggered breaths. Refer to section Settings on page 41.
4. Exceeding the maximal apnea time setting without a sufficient patient effort will cause the following:
In VS, a PRVC or VC breath will be
a.
delivered according to the selected Automode functionality.
b. In PS, a PC breath will be delivered.
5. The rings in the figures indicate patient triggering.
72
SERVO-airVentilator System v2.1, User's Manual
PC PS
12b3 4b
PRVC ⇄ VS
| Ventilation modes |
5 |
12a 4a
3
In the Automode PRVC VS combination, the first supported breath delivered to the patient has the same pressure level as the preceding PRVC breath.
VC ⇄ VS
12a 4a
3
In the Automode VC VS combination, the ventilator system uses the plateau pressure in the VC breath as a reference pressure for the first VS breath.
SERVO-airVentilator System v2.1, User's Manual
73
| Ventilation modes |
| 5

5.9 SIMV

SIMV stands for Synchronized Intermittent Mandatory Ventilation. In SIMV modes, mandatory controlled ventilation breaths are delivered with a preset SIMV rate. The patient can breathe spontaneously with PS between the mandatory breaths.
The parameters for each SIMV combination are adjusted in the settings window and are basically the same as those for the relevant controlled or supported mode. Refer to section Settings on page 41.
SIMV (PC) + PS:
delivers mandatory controlled breaths using
a preset respiratory rate and a preset pressure
delivers inspiratory support (PS) during
spontaneous breaths taken between the mandatory breaths
SIMV (PRVC) + PS:
delivers mandatory controlled breaths using
a preset respiratory rate and a preset volume
delivers inspiratory support (PS) during
spontaneous breaths taken between the mandatory breaths
SIMV (VC) + PS:
delivers mandatory controlled breaths using
a preset respiratory rate and a preset volume
delivers inspiratory support (PS) during
spontaneous breaths taken between the mandatory breaths
In SIMV modes, the mandatory breath is defined by the basic settings.
74
SERVO-airVentilator System v2.1, User's Manual
Settings
| Ventilation modes |
SIMV (VC)+ PSSIMV (PRVC)+PSSIMV (PC) + PS
XPC above PEEP
5 |
volume
Breath cycle time
1
X
1
X
Pause time
1
Only when the ventilator system is configured for setting the I:E ratio.
XXTidal volume /Minute
XXXSIMV rate
1
X
XXXI:E ratio / Inspiration time
XXXInsp. rise time
1
X
SERVO-airVentilator System v2.1, User's Manual
75
| Ventilation modes |
| 5
Note: In the minute volume configuration, the tidal volume is determined by dividing the minute volume by the SIMV rate.
The breath cycle time is the length of the mandatory breath in seconds and is the same as the duration of an SIMV period.
In SIMV, the very first breath is always a mandatory one.
If the patient triggers a breath during the SIMV period, the breath delivered is a mandatory one. If the patient fails to trigger a breath within the first 90 % of the SIMV period, a mandatory breath is delivered.
Note: If the ventilator system is configured for setting the inspiration time, an I:E ratio of 1:2 will be used to estimate the breath cycle time.

5.9.1 SIMV in detail

This combination of controlled and
supported ventilation allows for preset mandatory breaths that are synchronized with the patient's breathing.
If there is no trigger attempt within a time
window equal to 90 % of the set breath cycle time, a mandatory breath is delivered (the breath cycle time is the total time for one mandatory breath).
The mandatory breath is defined by the
basic settings (mode of ventilation, breath cycle time, respiratory pattern and volumes/pressures).
The spontaneous/supported breaths are
defined by the setting for PS.
The spontaneous/Pressure Support breaths are defined by setting the level for Pressure Support above PEEP (PS above PEEP).
76
SERVO-airVentilator System v2.1, User's Manual

5.9.2 SIMV (PC) + PS

The circles in the figure indicate patient triggering.

5.9.3 SIMV (PRVC) + PS

| Ventilation modes |
90%
90%
5 |
The circles in the figure indicate patient triggering.

5.9.4 SIMV (VC) + PS

The circles in the figure indicate patient triggering.
SERVO-airVentilator System v2.1, User's Manual
90%
77
| Ventilation modes |
| 5

5.10 Non Invasive Ventilation (NIV)

5.10.1 Safety guidelines

WARNINGS!
Avoid high inspiratory pressure as it may
lead to gastric overdistension and risk of aspiration. It may also cause excessive leakage.
The dead space will increase in NIV when
using a mask or helmet.
NIV is not intended to be used on
intubated or tracheotomized patients.
In non invasive ventilation, the measured
expired volume may be different from the actual volume exhaled by the patient due to leakage around the mask.
If nasal prongs are used, make sure that
they are applied so that air can flow freely through both prongs.
CAUTIONS:
It is not recommended to use a nebulizer
during NIV as the nebulized drug might come in contact with the patient's eyes in case of leakage.
Mask/prongs leakage might affect
nebulizer efficiency.
Important:
The mask/prongs must be properly
applied in order to minimize leakage.
When selecting the mask/prongs, it is
essential to consider such things as proper size and accurate adaptation to the patient.
CO2 rebreathing will increase during NIV
and use of a face mask/prongs.

5.10.2 Introduction

NIV refers to ventilation when the patient is not intubated or tracheotomized. It involves the use of a patient interface such as:
nasal mask
nasal prongs
face mask
total face mask
endotracheal tube positioned above the
vocal cords
NIV helmet
In NIV, the ventilator system adapts to variations in leakage to maintain the required pressure and PEEP level. Excessive leakage will result in a high priority alarm. Ventilation will resume automatically if the leakage decreases. It can also be started manually by tapping Resume ventilation in the LEAKAGE DETECTED window that opens to inform about leakage.
Note: In NIV, flow and volume curves and the following measured values are compensated for leakage: VTi, VTe, MVi, MVe.
Refer to chapter Alarm handling on page 109.
78
SERVO-airVentilator System v2.1, User's Manual
| Ventilation modes |
5 |
In all NIV modes, there is an automatic detection of patient connection and disconnection – the NIV disconnection functionality. This ensures that ventilation starts in a comfortable manner when the patient interface is applied to the patient's face. It ensures that ventilation stops when the interface is removed, avoiding high air flows and alarms.
It is possible to configure the NIV disconnection functionality. The setting made here will ensure a constant disconnection flow while ventilation is paused (at a high flow or a low flow setting). It is also possible to disable the NIV disconnection functionality, which may result in high air flows and alarms.
Refer to chapter Service & Settings on page 123.

5.10.3 Using a NIV helmet

WARNING!
The helmet application shall not be used with volume controlled modes.
The helmet application must only be used with pressure supported ventilation in NIV.
Important:
When using NIV PS, the filling of the
helmet must be initiated by tapping
START VENTILATION or Resume ventilation on the screen. This must also
be done after disconnection.
To secure a proper patient triggering
function, the PEEP level should never be set below 3 cmH2O. When helmets with a safety valve are used, it is recommended to set a PEEP level of at least 5 cmH2O.
Alarms related to volume are not reliable.
To avoid nuisance alarms the alarm limits must be properly set.
It is essential to set pressure alarms
adequately.
It is possible to use a helmet for non invasive ventilation of patients between 10 and 250 kg. Only use the adult patient category when using a helmet.
For instructions for the helmet application, refer to the manufacturer's instructions for use.
There are a few points to remember in order to use a helmet safely:
Do not rely on flow and volume parameters.
The volume in the helmet may cause delays
in signals and patient triggering.
Make sure that the helmet used eliminates
CO2 re-breathing.
High pressure levels may affect the patient's
ears and the flow may affect the patient's eyes.
Patients may perceive the helmet
application as noisy. A Servo Duo Guard filter used on the inspiratory side will reduce the noise level. The noise level may vary between different helmets.
Do not use humidified ventilation gas as this
will cause condensation on the helmet walls.
Do not use nebulizers.
SERVO-airVentilator System v2.1, User's Manual
79
| Ventilation modes |
| 5

5.11 NIV Pressure Control (NIV PC)

NIV Pressure Control (NIV PC):
delivers a constant pressure over a preset
inspiratory time and at a preset respiratory rate
delivers the inspiration with a decelerating
flow
changes in lung or thorax resistance or
compliance will affect the volume delivered
is leakage compensated
1234
56
The following parameters are set:
1. Oxygen concentration (%)
2. PEEP (cmH2O)
Differences compared with invasive PC:
When START VENTILATION is tapped, a
waiting position dialog is displayed. All patient related alarms are turned off for 2 minutes. In this position, ventilation will start if the ventilator system detects patient activity.
The trigger cannot be manually set in NIV.
Detection of pressure below PEEP or
expiratory volume decrease will start a new breath.
3. Respiratory rate (b/min)
4. PC above PEEP (cmH2O)
5. I:E ratio or Inspiration time (s)
6. Inspiratory rise time (% or s) Refer to section Settings on page 41.
80
SERVO-airVentilator System v2.1, User's Manual

5.12 NIV Pressure Support (NIV PS)

| Ventilation modes |
5 |
NIV Pressure Support (NIV PS):
is initiated by the patient, who controls the
respiratory rate and tidal volume
delivers ventilator support using the preset
pressure level and with a decelerating flow
has a fixed trigger
provides backup (PC) ventilation in case of
apnea
is leakage compensated
1234
5
7869
Differences compared with invasive PS:
When START VENTILATION is tapped, a
waiting position dialog is displayed. All patient related alarms are turned off for 2 minutes. In this position, ventilation will start if the ventilator system detects patient activity.
The ventilator system will not lock in backup
ventilation. There is no limit on the number of times the ventilator system can switch between supported mode and backup.
The trigger cannot be manually set in NIV.
The following parameters are set:
1. Oxygen concentration (%)
2. PEEP (cmH2O)
3. PS above PEEP (cmH2O) (PS level)
4. End inspiration (%)
5. Inspiratory rise time (% or s)
6. Apnea time (s)
7. Backup respiratory rate (b/min)
8. Backup PC above PEEP (cmH2O)
9. Backup I:E or Ti (s)
Refer to section Settings on page 41.
SERVO-airVentilator System v2.1, User's Manual
81
| Ventilation modes |
| 5
82
SERVO-airVentilator System v2.1, User's Manual

6 Ventilatory settings and functions Table of contents

| Ventilatory settings and functions |
84|O2 concentration6.1 84|Tidal volume/Minute volume6.2 84|Pressure level6.3 85|I:E ratio/Inspiration time6.4 86|End inspiration6.5 87|Inspiratory rise time6.6 88|Trigger6.7 90|Apnea time6.8 90|SIMV breath cycle time6.9 91|Predicted body weight (PBW)6.10 91|Maneuvers6.11 94|Compensation functions6.12 95|Disconnection/Suction6.13 97|Previous mode6.14
98|Apnea management6.15 101|Nebulization6.16 107|Adjust the O2 cell6.17
6 |
SERVO-airVentilator System v2.1, User's Manual
83
| Ventilatory settings and functions |
| 6

6.1 O2 concentration

The O2 concentration delivered to the patient is set in the mode settings window and is monitored by the ventilator system with upper and lower alarm limits. Refer to section Alarm limits on page 145.
The alarm is delayed 40 seconds after changing the O2 concentration.

6.2 Tidal volume/Minute volume

Depending on the ventilator configuration, the inspiratory volume can be set as either:
tidal volume or
minute volume
Whichever of these is set, the other will be displayed in the lower right information area of the mode settings window.

6.3 Pressure level

PC above PEEP is the set inspiratory pressure level for each mandatory breath in:
PC
SIMV (PC) + PS
Automode PC PS
backup ventilation in PS
PS above PEEP is the set inspiratory pressure support level for triggered breaths in:
PS
all SIMV modes
Automode PC PS
84
Bi-Vent
SERVO-airVentilator System v2.1, User's Manual
| Ventilatory settings and functions |
6 |

6.4 I:E ratio/Inspiration time

The setting of ventilatory settings can be configured in two different ways, based on:
I:E ratio or
inspiration time, in seconds, to better meet
the requirements for pediatric care.

6.4.1 I:E ratio

The I:E ratio expresses the relation between the inspiration phase and the expiration phase. Spontaneous breathing has an I:E ratio of around 1:1.5.
Note that increasing the inspiration time may raise mean airway pressure and improve oxygenation but may also cause hyperinflation. Reversed I:E ratios (e.g. 1.5:1 or 2:1) will further lengthen inspiratory time and shorten expiration, which may be helpful if the lungs are very stiff, but requires low respiratory rates to avoid gas trapping.
1 2
1:2
1. Inspiration
2. Expiration An inverse I:E ratio is also used in
Bi-Vent/APRV mode. Refer to section Bi-Vent/APRV on page 63.
A prolonged expiration time (e.g. 1:3) may be used for weaning and in case of obstructive lung disease, but a short inspiration time may also lower the tidal volume and lead to inadequate ventilation.
SERVO-airVentilator System v2.1, User's Manual
When the ventilator system is configured for setting of I:E, the unit for pause time and inspiratory rise time automatically switches to percent. The corresponding inspiration time for each I:E is displayed in the lower right information area of the mode settings window.
85
| Ventilatory settings and functions |
| 6

6.4.2 Inspiration time

The setting makes it possible to set the inspiration time (Ti) to a fixed time in seconds.
With this configuration, the unit for inspiratory rise time and pause time automatically switches to seconds. The corresponding I:E ratio for each inspiration time setting is displayed in the lower right information area of the mode settings window.
As the inspiration time is explicitly set, a change in the respiratory rate, for example, will affect the I:E ratio. As a safety precaution, there will therefore be an indication when the resulting I:E ratio passes 1:1 in either direction.
Note: When the inspiration time is directly set, the breath cycle time parameter is not displayed when an SIMV mode is selected, since there is no need to set it.

6.5 End inspiration

End inspiration:
is the point at which inspiration changes to
expiration in supported ventilation
if set too low, inspiration will be longer,
which may cause pulmonary hyperinflation and increased work of breathing
if set too high, inspiration will be shorter,
which may mean that the patient receives insufficient tidal volume
100%
50
86
SERVO-airVentilator System v2.1, User's Manual
| Ventilatory settings and functions |
6 |

6.6 Inspiratory rise time

The inspiratory rise time (Tinsp.rise):
is the time taken to reach peak inspiratory
flow or pressure at the start of each breath
is expressed in seconds or as a percentage
of the respiratory cycle time depending on how the ventilator is configured
0.10
An increased inspiratory rise time will affect the rate of flow/pressure increase and can be evaluated by the shape of the flow and pressure waveforms.
Inspiratory rise time set in seconds is applicable in:
PS
VS
Bi-Vent/APRV
Note: When the ventilator system is configured for setting of inspiration time rather than I:E ratio, the unit for inspiratory rise time automatically switches to seconds for all ventilation modes.
In supported modes, the inspiratory rise time should normally be increased from the default setting to enhance patient comfort.
Inspiratory rise time set as a percentage is applicable in:
VC
PC
PRVC
all SIMV modes
all Automode modes
SERVO-airVentilator System v2.1, User's Manual
87
| Ventilatory settings and functions |
| 6

6.7 Trigger

6.7.1 Pneumatic trigger

WARNING! If the flow trigger level is very
low (too far to the left on the scale), an autotriggering condition may be reached. This condition can also be reached if there is leakage in the patient circuit, e.g. if an uncuffed endotracheal tube is used. Triggering will then be initiated by the ventilator system and not by the patient. This should always be avoided by increasing the patient effort required to trigger the ventilator system, moving further to the right on the scale.
The ventilator system has a pneumatic trigger (flow or pressure based) functionality.
The pneumatic trigger setting:
determines the level of patient effort needed
to trigger the ventilator to inspiration
may be set as either flow or pressure
triggering, where flow triggering allows the patient to breathe with less effort
should generally be set so that it requires
minimal patient effort without causing autotriggering
During expiration, the ventilator system continuously delivers a gas flow (bias flow), which is measured in the expiratory channel.
3
1
4
0.40
2
1. Flow
2. Less effort
3. Trigger setting
4. Pressure
5. More effort
When triggering is based on flow, to the left on the scale, the ventilator system senses deviations in the bias flow delivered during expiration. These deviations are caused by the inspiratory efforts of the patient.
5
Refer to section Ventilatory settings on page 142.
88
SERVO-airVentilator System v2.1, User's Manual
The further to the left on the scale, the less effort the patient has to make. At the far left of the scale, there is a risk of autotriggering, and the scale and value are therefore marked in red. The trigger setting is marked with a ring in the dynamic image.
When triggering is based on pressure, to the right on the scale, the ventilator system senses deviations in the pressure below PEEP created by the patient. The pressure below PEEP required to initiate a breath is displayed when the setting is made.
The further to the right on the scale, the greater the patient effort required to trigger.
The trigger scale has different colors based on the setting. Green indicates a normal setting for pneumatic triggering. Red indicates that the setting is not recommended, e.g. when the risk of autotriggering may increase. Yellow is used as a warning color.
| Ventilatory settings and functions |
Patient triggering (flow or pressure) is indicated by a symbol in the status bar.
The pressure or flow curve will also be highlighted in white depending on which type of trigger is used.
6 |
SERVO-airVentilator System v2.1, User's Manual
89
| Ventilatory settings and functions |
| 6

6.8 Apnea time

Apnea time is the time without a patient breathing effort that the ventilator system will allow to elapse in supported ventilation before the No patient effort alarm is activated and the ventilator system switches to the backup mode.
Refer to section Apnea management on page 98.

6.8.1 Maximal apnea time

The maximal apnea time:
is the maximum time without a patient
breathing effort that the ventilator will allow to elapse in supported ventilation before switching to controlled ventilation.
In Automode, the apnea time becomes longer as spontaneous breathing becomes more regular. It is therefore set, in Automode only, as Max. apnea time.

6.9 SIMV breath cycle time

The SIMV breath cycle time:
is the duration of the total respiratory cycle
of the mandatory breath in SIMV (inspiration + pause + expiration)
only applies if the inspiratory time is set
using the I:E ratio
together with a spontaneous period, makes
up one full SIMV cycle
The breath cycle time is sometimes referred to as an SIMV period.
Note: The breath cycle time parameter is not displayed when an SIMV mode is selected and inspiration time is configured.
90
SERVO-airVentilator System v2.1, User's Manual

6.10 Predicted body weight (PBW)

| Ventilatory settings and functions |
6 |
Notes:
For adult patients outside the height
range 130-200 cm, PBW is the same as the patient weight (BW).
For pediatric patients, PBW is the same
as the patient weight (BW).
In mechanical ventilation, predicted body weight can be used to help reduce the risk that differences in body weight will affect the estimated ventilation needs of different patients.
The ventilator monitors the ratio of tidal volume to PBW (VT/PBW) in ml/kg.
In the adult patient category PBW is calculated according to the Devine Formula and requires that height and gender are entered.
The tidal volume setting will, when tapped, present a calculated value for VT/PBW if the necessary patient data has been entered under PATIENT DATA. The value will also be presented in the numerical values and trended under TRENDS & LOGS/TRENDS

6.11 Maneuvers

The following four functions can be accessed under MANEUVERS (only available during ventilation) in the quick menu.
1
2
3
4
1. Manual breath
2. Static measurements
3. Nebulization
4. O2 boost level

6.11.1 Manual breath

When MANUAL BREATH is tapped, the ventilator system will initiate a new breath cycle according to the current ventilator settings.
SERVO-airVentilator System v2.1, User's Manual
91
| Ventilatory settings and functions |
| 6

6.11.2 Static measurements

The inspiratory and expiratory hold can be used to perform certain measurements:
PEEPtot: set PEEP + intrinsic PEEP
Cstatic: static compliance, a measure of the
elastic properties of the respiratory system. A decrease in compliance implies stiffer lungs.
E: elastance, has different compliances and
resistances in the lungs and an increase in elastance implies stiffer lungs.
Ri: inspiratory resistance
Re: expiratory resistance
Tc: time constant, calculated as Cstatic x Re.
Some lung units have decreased compliance, and some have increased resistance, or both. Differences in Re and Cstatic affect the speed at which the lung units are filled and emptied. An expiration time of three time constants is recommended to avoid auto PEEP.
Pplat: pressure during end inspiratory pause
Inspiratory hold
This function is activated by pressing INSPIRATORY HOLD for a maximum of 30 seconds. The inspiratory and expiratory valves close after inspiration. This function can provide an exact measurement of the end inspiratory lung pressure. It can be used to pause ventilation during X-ray or to determine the plateau pressure (Pplat), or, together with the expiratory hold, to calculate static compliance.
Expiratory hold
Expiratory and inspiratory valves are closed after the expiration phase is completed for as long as EXPIRATORY HOLD is pressed, but only up to a maximum of 30 seconds. Expiratory hold provides an exact measurement of the end expiratory pause pressure. It can be used to determine total PEEP and, together with inspiratory hold, static compliance (Cstatic). The dynamic pressure is displayed on the PEEP numerical value.
92

6.11.3 Nebulization

Refer to section Nebulization on page 101.
SERVO-airVentilator System v2.1, User's Manual
| Ventilatory settings and functions |
6 |

6.11.4 O2 boost level

1
3
2
1. O2 boost level
2. O2 boost function
3. O2 concentration setting (O2 conc.)
By tapping O2 BOOST LEVEL, it is possible to change the desired level for the O2 boost function. It is possible to lock the O2 boost level to 100 %. It is also possible to set it to 0 %, in which case the O2 boost function will no longer be active and will be replaced by three asterisks.

6.11.5 O2 boost function

To use the O2 boost function, tap and hold O boost at the bottom left corner of the screen.
When tapped, O2 boost delivers the oxygen setting displayed here for a period of 1 minute. The O2 boost function can be interrupted by tapping the red cancel symbol in the O2 boost timer window anytime during the 1 minute interval.
2
The value entered under O2 boost (%) level specifies the number of percentage units that will be added to the value set for the O
2
concentration. For example: if the current O2 concentration is 40 % and the O2 boost level is 30 %, the O2 boost function will, when tapped, deliver 70 % O2.
The O2 boost function figure displayed will change accordingly. Since the minimum O
2
concentration is 21 %, the O2 boost (%) level scale goes from 0 to 79 %.
Refer to section Ventilatory settings on page 142 and to to section Edit Temporary O
2
increase (%) on page 127.
SERVO-airVentilator System v2.1, User's Manual
93
| Ventilatory settings and functions |
| 6

6.12 Compensation functions

6.12.1 Leakage compensation

Leakage compensation is automatic for all patient categories in non invasive modes.
The function is designed to help maintain PEEP throughout the breath and is activated by default.
Leakage compensation may also affect important ventilatory parameters, such as patient triggering and the termination of inspiration.
Leakage is measured and presented in percent.

6.12.2 Circuit compensation

Part of the volume of each inspiration will not reach the patient because of gas compression in the ventilator and expansion of the tubing. All components in the patient circuit affect such losses.
When circuit compensation is activated, the delivered and measured volume and flow values are automatically compensated for these losses, as indicated by the symbols on the affected values.
The patient circuit test must be passed in order to activate circuit compensation.
94
To deactivate or reactivate, tap COMPENSATION in the extended menu/CIRCUIT COMPENSATION. Follow on-screen instructions.
Important: When monitoring VT/PBW, circuit compensation must be activated.
SERVO-airVentilator System v2.1, User's Manual
| Ventilatory settings and functions |
6 |

6.13 Disconnection/Suction

Important: If a closed-suction system is used, DISCONNECTION/SUCTION, as well as the inspiratory and expiratory hold functions, should not be used.
Tap DISCONNECTION/SUCTION in the quick menu during ventilation to open the DISCONNECTION/SUCTION window.
The window always opens in Preparation.
DISCONNECTION/SUCTION enables automatic inhibition of the ventilator system during a tracheal suction procedure or when briefly pausing ventilation in invasive modes. The ventilator system is prevented from cycling without activating alarms.
Refer to section Ventilatory settings on page 142 and to to section Edit Temporary O increase (%) on page 127.
2

6.13.1 Suctioning procedures

Open suctioning
Important:
Alarms are turned off during the Patient
disconnected phase for a maximum of
60 seconds. If the patient has not been reconnected within 60 seconds, alarms are activated.
The minimum PEEP level during
disconnection is 3 cmH2O. The ventilator system will adjust to the minimum level if the PEEP level is below 3 cmH2O in order to detect disconnection of the patient.
For open suctioning procedures, there are three phases following Preparation:
Pre-oxygenation
Patient disconnected
Post-oxygenation
Notes:
DISCONNECTION/SUCTION is not
available in NIV modes or when Manual breath is activated.
During the Patient disconnected phase
of a suctioning procedure, the nebulizer is temporarily paused.
When only one gas is connected, an
elevated oxygen level cannot be set during the preparation phase. In this case, the post-oxygenation phase will be skipped.
SERVO-airVentilator System v2.1, User's Manual
Adjust the O2 concentration, if desired, then tap Accept.
95
| Ventilatory settings and functions |
| 6
Note: Tapping Cancel will close the DISCONNECTION/SUCTION window.
Pre-oxygenation
Pre-oxygenation of the patient begins automatically after Accept is tapped during preparation.
The Patient circuit disconnected alarm is turned off. The maximum duration of the pre-oxygenation phase is 120 seconds. After this, the system automatically returns to ventilation using the previous oxygen setting. The same thing happens if Cancel is tapped.
Patient disconnected
The system automatically enters the Patient disconnected phase when the patient is disconnected during the pre-oxygenation phase.
During the Patient disconnected phase, the following alarms are turned off for up to 60 seconds:
- apnea
- minute volume
- respiratory rate
- PEEP When the patient is reconnected, the
system automatically enters the post-oxygenation phase and then resumes ventilation. It is also possible to restart ventilation manually by tapping START VENTILATION.
Post-oxygenation
After reconnection, the ventilator system will deliver the same oxygen concentration as in the pre-oxygenation phase for 60 seconds.
96
After 60 seconds the system automatically returns to ventilation using the previous oxygen concentration setting.
SERVO-airVentilator System v2.1, User's Manual
| Ventilatory settings and functions |
6 |
Closed suctioning
When using a closed-suction system, DISCONNECTION/SUCTION should not be used. The O2 boost function should be used instead for oxygenation purposes. Consider pre-silencing the alarms before suctioning.
Use one of the pressure-based modes listed here. Adjust settings to levels suitable for the patient and follow hospital guidelines for closed suctioning.
PC
PS
Bi-Vent/APRV
SIMV (PC) + PS

6.14 Previous mode

When MODES is tapped in the quick menu during operation, the current mode tile is always highlighted and the previous mode tile is marked PREVIOUS, together with the date and time it was used.
Note: If the previous mode was non invasive and the current mode is invasive, or vice versa, it is necessary to go to Standby and choose the relevant ventilation type to find the previous mode.
To recall the previous ventilation mode used:
Tap the tile marked with an arrow in the
MODES window.
A dialog will open asking Do you want to
keep the previous settings for the mode?
Tap one of the two choices Yes or No as
appropriate.
- If Yes is tapped, the mode settings
window will open with the previous settings intact.
SERVO-airVentilator System v2.1, User's Manual
Important: If one or several settings in the mode settings window are highlighted in yellow, this indicates that it/they should be considered for adjustment, as the values entered there may have been carried over from the previous mode.
- If No is tapped, the mode settings
window will open with default settings, which may then be adjusted.
97
| Ventilatory settings and functions |
| 6
Notes:
The previous mode function is not
available:
- after a pre-use check
- after changing the patient category
- after admitting a new patient
- after using the same ventilation mode
for more than 24 hours
- after restarting the system.
When the previous mode function is
activated during backup ventilation, the ventilator system returns to the mode that was active before the supported mode was initiated.
A recall of previous settings is only
possible after a change of ventilation mode.

6.15 Apnea management

6.15.1 Apnea time

Apnea time is the time without a patient breathing effort that the ventilator system will allow to elapse in supported ventilation before the No patient effort alarm is activated and the ventilator system switches to the backup mode.
The relevant backup mode is highlighted in white in the heading on the screen and the alarm No patient effort is displayed.
If the patient triggers a breath, the ventilator system automatically switches back to supported ventilation and the No patient effort alarm disappears.
Apnea time is available in all supported modes and in all SIMV modes. Set the apnea time that is appropriate for each patient in the mode settings window.
98
Note: In SIMV modes, there is no backup ventilation and the apnea time only controls the No patient effort alarm. The apnea time is therefore set in the ALARM LIMITS window.
Refer to chapter Alarm handling on page 109 and to section Alarm limits on page 145.
SERVO-airVentilator System v2.1, User's Manual
| Ventilatory settings and functions |
6 |

6.15.2 Backup ventilation

For invasive modes, backup ventilation entails a switch in case of apnea:
from VS to PRVC
from PS/CPAP to PC.
For non invasive modes, the switch is from NIV PS to NIV PC.
When the relevant backup mode is activated while ventilating in a supported mode, the name of the mode is highlighted in white in the mode heading and the backup parameters in the direct access bar are shown as active.
The following parameters are set under the backup mode heading in the mode settings window:
PC above PEEP (cmH2O) for PS backup.
The minimum backup pressure level is 5 cmH2O.
Tidal volume (ml) for VS backup.
Respiratory rate (b/min)
I:E or Ti (s) (depending on configuration)
Backup ventilation trends
The number of switches to backup ventilation per minute is trended under TRENDS & LOGS/TRENDS.
The percentage time spent in backup ventilation per minute is also trended.
No consistent patient effort
This alarm occurs in invasive ventilation only.
If the patient fulfils the criteria for the No consistent patient effort alarm, the ventilator system will lock in backup ventilation.
A dialog Backup ventilation active - review
ventilation settings or continue in supported mode. is displayed on the screen. A choice
must be made or this dialog will remain open and the ventilator system will remain in backup ventilation.
Tap Review ventilation settings in the dialog window to return to the mode settings window.
Refer to section Settings on page 41.
SERVO-airVentilator System v2.1, User's Manual
Tap Cancel to close the mode settings
window without changes being applied. Ventilation will continue as before, i.e. in backup ventilation.
Tap Accept to accept the settings and
continue in the supported mode with a reset apnea time.
Alternatively, tap Continue in supported mode in the dialog window to return to the supported mode. The apnea time will be reset.
Refer to chapter Alarm handling on page 109 and to section Alarm limits on page 145.
99
| Ventilatory settings and functions |
| 6

6.15.3 Deactivating backup ventilation

It is possible to deactivate backup ventilation for invasive PS/CPAP and VS. If backup ventilation is deactivated, the No patient effort alarm will be activated at the end of the apnea time but no backup ventilation will start.
To allow deactivation of backup ventilation:
Tap SERVICE & SETTINGS in the extended
menu in STANDBY.
Tap BIOMED and enter the code, then tap
CONFIGURATION/STARTUP CONFIGURATION/Deactivation of backup function.
Change from Not allowed to Allowed.
If this choice is made, Deactivate backup ventilation is displayed at the top right of the mode settings window during ventilation.
To deactivate backup ventilation:
Tap Deactivate backup ventilation in the
mode settings window.
A confirmation dialog Do you really want to
deactivate backup ventilation? is displayed. Confirm by pressing Yes.
Tap Accept in the mode settings window.
Backup ventilation off is displayed after the
mode name in the heading when ventilation then begins.
The backup function is automatically re-activated if:
a change is made to a controlled mode of
ventilation
the ventilator system is switched to Standby
the system is turned off.
100
Note: Backup ventilation remains inactive if a change of mode is made between PS/CPAP and VS.
SERVO-airVentilator System v2.1, User's Manual
Loading...