Maquet Servo User Manual

x
User's Manual
SERVO-air Ventilator System v2.1

TABLE OF CONTENTS

| TABLE OF CONTENTS |
1 2 3 4 5 6 7 8 9 10
5|Introduction 15|System Overview 37|Operation overview 45|Displaying and saving data 55|Ventilation modes 83|Ventilatory settings and functions
109|Alarm handling 123|Service & Settings 131|Technical data 159|Index
SERVO-airVentilator System v2.1, User's Manual Infologic 1.39
3
| TABLE OF CONTENTS |
4
SERVO-airVentilator System v2.1, User's Manual
Infologic 1.39

1 Introduction Table of contents

| Introduction |
6|Device description1.1
8|Safety guidelines1.2 14|Version and configurations1.3
1 |
SERVO-airVentilator System v2.1, User's Manual
5
| Introduction |
| 1

1.1 Device description

1.1.1 Device components

The ventilator system consists of:
1. a user interface for setting ventilation modes, displaying data and indicating alarms
2. a patient unit for mixing gases and controlling gas delivery
3. a patient circuit for delivering and exchanging gases
1

1.1.2 Intended use

The SERVO-air ventilator system is:
intended for respiratory support, monitoring
and treatment of pediatric and adult patients
to be used only by healthcare providers
to be used only in professional healthcare
facilities and for transport within these facilities
2
6
3
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |

1.1.3 User's Manual

This manual summarizes the functions and safety features of the ventilator system. It is not all-inclusive and should not be seen as a substitute for training.

1.1.4 Cleaning & Maintenance

Please refer to the SERVO-air Cleaning and Maintenance User's Manual.

1.1.5 Servicing Guidelines

CAUTIONS:
Regular service: The ventilator system
must be serviced at regular intervals by Maquet authorized personnel who have received specialized training.
Complete service records: All service
performed on the ventilator system must be recorded in a service log in accordance with hospital procedures and local and national regulations.

1.1.6 Disclaimers

Non-professional servicing:
Maquet has no responsibility for the safe operation of the ventilator system if installation, service or repairs are performed by persons other than Maquet authorized personnel.
Service contract: It is strongly
recommended that all service on the ventilator system should be performed as part of a service contract with Maquet.
Note: If the ventilator system is to be a part of another system it requires an evaluation of the requirements of the IEC 60601-1 standard.
SERVO-airVentilator System v2.1, User's Manual
7
| Introduction |
| 1

1.2 Safety guidelines

Follow these safety guidelines. Additional warnings appear in context throughout this document.
Information is highlighted with Warning, Caution, Important or Note, where:
WARNING! Indicates critical information about a potential serious outcome to the patient or the user.
CAUTION: Indicates instructions that must be followed in order to ensure the proper operation of the equipment.
Important: Indicates information intended as help to operate the equipment or its connected devices easily and conveniently.
Note: Indicates information requiring special attention.

1.2.1 General

WARNINGS!
The ventilator system may be operated
only by authorized personnel who are properly trained in its use. It must be operated according to the instructions in this User´s Manual.
After unpacking, perform a routine
cleaning and a pre-use check.
Always perform a pre-use check before
connecting the ventilator system to a patient.
Secure all tubing and cables to avoid the
risk of unwanted movement of the equipment.
If any of the following occurs, discontinue
use of the ventilator system and contact a service technician:
- unfamiliar pop-up windows on the
screen
- unfamiliar sounds
- any unfamiliar or unexplained event
- alarms that cannot be resolved
Make sure that a resuscitator is readily
available.
The air inlet must not be occluded.
Positive pressure ventilation can be
associated with the following adverse events: barotrauma, hypoventilation, hyperventilation or circulatory impairment.
Ventilation must be started manually
when a patient is connected to the ventilator system.
Keep the ventilator system upright during
use.
8
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |
Do not cover the ventilator system in any
way, since the functioning of the equipment may be adversely affected.
Do not modify or remove any original
parts.
The ventilator system must not be used
during radiotherapy, since this may cause system malfunction.
The ventilator system must not be used
in a hyperbaric chamber.
The ventilator system must be kept away
from magnetic resonance imaging (MRI) equipment.
The ventilator system must not be used
with helium or any gas mixture containing helium.
When the ventilator system is used with
MCare Remote Services, use only network equipment that is safe and complies with the relevant electrical and EMC standards such as IEC 60950.
Note:The network cable is excluded
from this requirement.
Only accessories, supplies, and auxiliary
equipment recommended by Maquet should be used with the ventilator system. Use of any other accessories, spare parts or auxiliary equipment may cause degraded system performance and safety.
Use only the Maquet approved active
humidifiers. Use of non-approved active humidifiers may result in higher gas temperatures and increase resistance in filters.
During humidification, carefully monitor
the airway pressure. Increased airway pressure could result from a clogged
filter. Replace the filter if the expiratory resistance increases or according to the instructions for the filter, whichever comes first.
Service, repair and installation may only
be performed by Maquet authorized personnel.
The ambient sound needs to be taken
into consideration when setting the alarm sound level.
Always disconnect the patient from the
ventilator system when performing operations that involve risk for the patient, such as replacing the O2 cell, dismantling etc.
CAUTIONS:
Never leave the patient unattended when
connected to the ventilator system.
Before use, make sure the system
version displayed under SYSTEM STATUS/General corresponds to the system version described in the User's Manual.
Maquet has no responsibility for the safe
operation of the ventilator system if the requirements specified in Intended use on page 6 are not followed.
When lifting or moving the ventilator
system or parts of the system, follow established ergonomic guidelines, ask for assistance, and take appropriate safety precautions. The weight is specified on the ventilator system.
The air inlet filter must be in place when
the system is running.
SERVO-airVentilator System v2.1, User's Manual
9
| Introduction |
| 1
The expiratory channel and expired gas
from the exhaust port may be contaminated.
If a scavenging system (i.e. gas
evacuation) is connected to the ventilator system, it must conform to ISO 80601-2-13 guidelines for subatmospheric pressure and induced flow.
During operation any water traps must
be checked regularly and if necessary emptied.
All technical documentation is available
for use by Maquet authorized personnel.
When using the MCare Remote Services
function, instal the network cable so that there is no risk of anyone tripping over it.
The ventilator system must not be used
with any anaesthetic agent.
Important:
Portable and mobile radio frequency (RF)
communications equipment can affect medical electrical equipment.
The ventilator system must be installed
and put into service according to the EMC declaration.
Securely attach all cables, etc, to
minimize the risk of unintentional disconnection.
While the ventilator system is in use, the
wheels of the mobile cart must be locked and the mobile cart must be in a horizontal position.
When the ventilator system is connected
to a patient:
- Do not lift or disconnect the expiratory
cassette.
- Continuously monitor the settings and
measurements displayed on the screen.
Always use a heat and moisture
exchanger (HME) or an active humidifier to prevent dehydration of lung tissue.
If a heated patient circuit is not used in
the system, a water trap must be used on the expiratory tube to avoid condensation in the system when an active humidifier is used. During operation the water traps must be checked regularly and if necessary emptied.
Check that the cooling fan intakes are
not covered. Do not place the ventilator system on soft surfaces.
The air inlet filters must be checked
regularly and replaced if necessary.
Use an inspiratory filter when ventilating
a highly infectious patient.
All excess fluids must be disposed of
according to hospital routines.
The emergency air intake must not be
blocked.
Do not disconnect the expiratory
cassette while the ventilator system is in operation; if necessary, disconnect the cassette while in STANDBY.
10
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |
Notes:
Do not simultaneously touch the patient
and any accessible connector contacts.
Do not solely rely on the use of an
external monitor to determine the status of the patient and the ventilator system.
Make sure that the ventilator system is
firmly mounted on the mobile cart.
Make sure that cables and patient circuit
is not obstructed or squeezed due to improper mounting.
Extra care should be taken when
handling tubes, connectors and other components of the patient circuit. The use of a support arm to relieve the patient from the weight of the tubing system is recommended.
Contact a Maquet representative
regarding decommissioning of the equipment.
Turn off the active humidifier during
nebulization.
Expiratory filter connection is mandatory
during nebulization.
A water trap is recommended if a single
heated patient circuit is used.
Thermoshell, expiratory cassette must
be used when using heated expiratory tubing or Expiratory heater Servo Duo Guard.

1.2.2 Power supply

WARNINGS!
The power cord must be connected only
to an AC mains power outlet with protective earth to avoid the risk of electric shock.
The power supply cord must be plugged
directly into the mains power outlet without the use of any multiple socket outlets. If a multiple socket outlet is used together with other products, total leakage current might be exceeded in the event of a fault in the protective earth.
CAUTIONS:
Do NOT use antistatic or electrically
conductive tubing with this system.
Avoid contact with external electrical
connector pins.
SERVO-airVentilator System v2.1, User's Manual
11
| Introduction |
| 1
Important: In case of total loss of power during ventilation, an alarm will sound for 2 minutes. When power is restored, the ventilator system will start in the same state and with the same settings as before the power loss.
Note: When the system is connected to an external power supply, all connected battery modules are being recharged. This does not affect ventilation.
Battery
WARNINGS!
Do not use sharp tools when extracting
the batteries.
To guarantee reliable battery backup,
make sure a battery is in place in slot 2 at all times during ventilation.
Dispose of batteries according to local
regulations and not with ordinary waste.
CAUTION:
The battery modules must be charged before first use.
Do not expose the batteries to water, fire or excessive heat.
Do not crush, disassemble, puncture or short circuit the connector terminals.
One battery can be added to an available slot during operation.
Hold onto the battery strap when inserting a battery in the ventilator system.
Important:
If a battery status message is displayed
on the screen, always go to SYSTEM STATUS/Batteries for detailed information.
Check battery in SYST EM
STATUS/Batteries window to ensure safe battery operation. Always charge the battery before use.
Always replace batteries when the
ventilator system provides notification of imminent expiration or of diminished operating capacity.
When not in use, the ventilator system
should always be connected to the mains power to ensure fully charged batteries.
12
Refer to section Battery status on page 23.
SERVO-airVentilator System v2.1, User's Manual
| Introduction |
1 |

1.2.3 Fire hazard

WARNINGS!
Keep all sources of ignition away from
the ventilator system and the oxygen hoses.
Do not use a ventilator system with worn
or frayed gas supply hoses or hoses that have been contaminated by combustible materials such as grease or oil.
Oxygen-enriched gas is extremely
flammable: if a burning odor is detected, disconnect the oxygen supply and mains power and remove the batteries.
Make sure that both the mains power
outlet and the power supply connector are accessible.

1.2.4 Gases

1.2.5 Auxiliary equipment

CAUTION:
Measurements of numerical values that have been processed by auxiliary equipment:
may be inaccurate if equipment not
authorized by Maquet is used
should be disregarded if they conflict
with information on the ventilator screen
must not be used as a substitute for
therapeutic or diagnostic decisions.
Accessories, supplies, and auxiliary equipment used with the ventilator system must be recommended by Maquet.
WARNING! The ventilator system must not be used with helium or any gas mixture containing helium.
Refer to section Ventilator system on page 134.
SERVO-airVentilator System v2.1, User's Manual
13
| Introduction |
| 1

1.3 Version and configurations

This manual applies to version 2.1 of the SERVO-air ventilator system.

1.3.1 Configurations

The ventilator system can be used in both invasive and non invasive ventilation. The configurations includes adult and pediatric. Refer to section System on page 132.

1.3.2 Available modes and functions

Patient categoryModes/Functions
SIMV
(PC) + PS
(PRVC) + PS
(VC) + PS
X
O
X
AdultPediatric
XXPC
OOPRVC
XXVC
OOBi-Vent/APRV
XXPS/CPAP OOVS OOAutomode
X O
X
OONIV PC OONIV PS
XXNebulizer OOAlarm output connection
14
O = option— = not applicableX = standard
SERVO-airVentilator System v2.1, User's Manual

2 System Overview Table of contents

| System Overview |
16|Ventilator2.1 18|Patient unit2.2 22|Batteries2.3 24|Patient circuit2.4 29|User interface2.5 36|Transport2.6
2 |
SERVO-airVentilator System v2.1, User's Manual
15
| System Overview |
| 2

2.1 Ventilator

The ventilator consists of a user interface and a patient unit.
Air is supplied from ambient air by an internal turbine and O2 may be supplied by a medical pipeline system or by gas cylinder.

2.1.1 Mounting on mobile cart

Lock the wheels.
Release the locking clamp on the mobile
cart.
Stand directly in front of the mobile cart
when mounting the ventilator system.
Tilt the ventilator system to fit the two front
clamps in position on the mobile cart.
Press down the rear end of the ventilator to
fit the rear clamp in position.
Lock the ventilator system to the mobile
cart with the locking clamp.
Ensure that the patient unit is firmly fixed
to the mobile cart via the clamps and locking clamp.
16
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |
2
3
1
1. Patient unit
2. User interface
3. Expiratory inlet
4. Inspiratory outlet
5. Emergency air intake
6. Air inlet
7. Battery compartment
8. Patient circuit
9. AC mains power
10. O2 supply
11. Wheel lock
11
4
6
5
7
8
9
10
Note: Lock the wheels whenever the ventilator system is standing still.
SERVO-airVentilator System v2.1, User's Manual
17
| System Overview |
| 2

2.2 Patient unit

1
2
3
4
5
6 8 9 10
7
14 13 12 11
15
16
18
17
20
21
19
1. On/Off switch
The switch must be pulled downwards before it can be switched.
2. Expiratory outlet
3. Power indicators
4. RS-232 connectors
5. Equipotentiality terminal.
6. AC mains power source connector with
fuse
7. Alarm output connection
8. External +12V DC inlet
9. Fuse for external DC power
10. Ethernet connection
18
11. Battery compartments
12. USB ports
13. Gas inlet for O
14. Gas inlet for air including air inlet filter
15. Inspiratory outlet
16. Emergency air intake
17. Nebulizer connector
18. Expiratory inlet
19. Cooling fan with filter (on both sides)
20. Expiratory cassette
21. Expiratory inlet with moisture trap
2
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |
Important: No other external devices than
a USB memory stick may be connected to the USB ports. Only one memory stick can be used at the same time.

2.2.1 Symbols on patient unit

DescriptionSymbol
CE label—indicates compliance with the requirements of the Medical Device Directive 93/42/EEC
CSA label—Indicates compliance with Canadian and US standards
UDI Label - Unique Device Identification. Refer to section UDI label on page 156.
Rx
ONLY
In USA, Federal law restricts this device to sale by or on the order of a physician.
Type B—indicates classification according to IEC 60601-1
Type BF applied part — indicates classification according to IEC 60601-1
Potential equalization terminal Note: The potential equalization
terminal is designed for the connection of a potential equalization conductor according to DIN 42 801 and IEC 60601-1. The function of the potential equalization terminal is to equalize potentials between the ventilator system and other medical devices that can be touched simultaneously. The potential equalization terminal must not be used for a protective earth connection.
SERVO-airVentilator System v2.1, User's Manual
-
+
12V
-
+
Nebulizer connector
RS-232/Serial port - connector for data communication
Ingress protection, IP21
Fuse (specification) External 12V DC input
Battery
Expiratory gas flow from the patient
Inpiratory gas flow to patient
19
| System Overview |
| 2
DescriptionSymbol
Mains power On/Off Mains connected, batteries charging Gas exhaust port—exhaust gas flow
from ventilator system Network connection
USB connection
Alarm output connection
Special waste
Note: This product contains electronic and electrical components. Discard disposable, replaced and left-over parts in accordance with appropriate industrial and environmental standards.
Caution
Consult instructions for use
DescriptionSymbol
MR Unsafe - keep away from magnetic resonance imaging (MRI) equipment.
20
Consult accompanying documentation
Locked
Unlocked
Serial number
Order number
Weight of patient unit with user interface and ventilator including its safe working load.
Use of ON/OFF switch The switch must be pulled
downwards before it can be switched.
Manufacturer The symbol is accompanied by
manufacturer address and manufacturing date.
SERVO-airVentilator System v2.1, User's Manual

2.2.2 Gas flow through the patient unit

| System Overview |
2 |
8
6
7
9
11
10
1. Air inlet with air inlet filter.
2. Turbine module for ambient air.
3. The check valve prevents the gas to flow
backwards.
4. Gas inlet for O2.
5. The gas module for O2 regulates the O
gas flow.
6. The flow meter measures the gas flow.
7. The pressure of the gas delivered to the
patient is measured by the inspiratory pressure transducer. The transducer is protected by a bacterial filter.
2
5
3 1
2
4
12 13
14
10. The expiratory inlet in the expiratory cassette contains a moisture trap.
11. The ultrasonic transducers in the expiratory cassette measure the outgoing gas flow.
12. The expiratory pressure transducer in the expiratory cassette measures the pressure in the patient circuit. The transducer is protected by a bacterial filter.
13. The expiratory valve in the expiratory cassette regulates the pressure in the patient circuit.
14. Exhaust port.
8. The O2 cell measures the oxygen concentration. The O2 cell is protected by a bacterial/viral filter.
9. The inspiratory channel delivers gas to the patient circuit inspiratory tubing and contains a safety valve.
SERVO-airVentilator System v2.1, User's Manual
Note: The expiratory cassette can be exchanged between different ventilator systems. Always perform a pre-use check after exchanging an expiratory cassette.
21
| System Overview |
| 2

2.3 Batteries

2.3.1 Charging battery modules

Important: The battery modules are
delivered in shipping-mode and must be charged before use.
To charge the battery modules, insert the battery modules in the ventilator system. The ventilator must be connected to mains. The battery modules are charged automatically.
The batteries can also be charged with the External battery charger, SERVO-air.

2.3.2 Handling battery modules

The battery compartment is divided into two slots, 1 and 2.
The ventilator system uses the battery in slot 1 first. Make sure that the battery in slot 2 is in place as a backup at all times during ventilation.
Refer to section Battery on page 12.
Remove the battery from the ventilator
system.
To insert a battery module:
Hold onto the battery strap when inserting
a battery in the ventilator system. When inserting a battery module in slot 1, check that the battery strap for the battery in slot 2 is not folded into the battery compartment.
Ensure that the battery is fully inserted so
that the release button returns to a completely closed position.
1
2
1
2
The battery module in slot 1 may be exchanged during ventilation.
To remove a battery module:
Press the release button to the left and pull
the battery strap.
22
SERVO-airVentilator System v2.1, User's Manual
| System Overview |
2 |

2.3.3 Battery status

Important:
If Replace battery is displayed, the
battery is unreliable, regardless of the operating time displayed under Batteries. In this situation, replace the battery even when the STATUS window indicates that significant operating time remains.
At least one battery module must always
be installed.
Detailed information about batteries is accessed via SYSTEM STATUS/Batteries. There is also an indication in the status bar showing the power supply currently being used by the ventilator system.
If the ventilator system is running on battery power, the active battery in the battery symbol turns yellow and the mains power symbol disappears. The estimated remaining battery time in minutes is always displayed, regardless of the power supply in use.
The following information is displayed under Batteries in SYSTEM STATUS for each connected battery module:
BATTERY CAPACITY – usable backup time
in minutes An estimated backup time is shown in Standby. This estimate may differ from the actual usable backup time during running. Usable backup time depends on set mode and selected ventilation settings.
Note: The presented usable backup time is the sum of the estimated operation time displayed for each battery module minus 20 minutes.
Slot number
Serial number
Remaining operation time in minutes for
each battery
Notification – may be displayed close to the
remaining operation time in minutes.
Remaining battery life
The battery symbol also functions as a shortcut to the window otherwise accessed via SYSTEM STATUS/Batteries.
SERVO-airVentilator System v2.1, User's Manual
23
| System Overview |
| 2

2.4 Patient circuit

Notes:
Turn off the active humidifier during
nebulization.
Expiratory filter connection is mandatory
during nebulization.
A water trap is recommended if a single
heated patient circuit is used.
Thermoshell, expiratory cassette must
be used when using heated expiratory tubing or Expiratory heater Servo Duo Guard.
To ensure that the inspiratory gas
temperature is below 43°C the patient circuit inspiratory tube must be at least
1.2 m to let the gas cool down.
Refer to System Flow Chart, SERVO-air for information regarding patient circuit configurations to be used with the ventilator system.
24
SERVO-airVentilator System v2.1, User's Manual
15 - 22 mm Ø patient circuit, reusable
5
2
4
2
3
| System Overview |
2 |
1
4
1. Inspiratory patient tube
2. Y piece
3. Heat and moisture exchanger (HME)
4. Angled adapter for endotracheal tube
7
6
5. Aerogen Pro
6. Expiratory patient tube
7. Expiratory filter
SERVO-airVentilator System v2.1, User's Manual
25
| System Overview |
| 2
22 mm Ø patient circuit, dual heat disposable
8
6
7
5
6
7
1. Inspiratory patient tube
2. Water autofill
3. Humidification chamber
4. Active humidifier
5. Cuff with temperature port
6. Y piece
2
1
1
3
4
9
10
12
11
7. Angled adapter for endotracheal tube
8. Aerogen Solo
9. Expiratory patient tube
10. Expiratory filter
11. Expiratory heater Servo Duo Guard
12. Thermoshell, expiratory cassette
26
SERVO-airVentilator System v2.1, User's Manual
10 mm Ø patient circuit, single heat reusable
| System Overview |
2 |
9
6
8
5
6
7
1. Inspiratory patient tube
2. Water autofill
3. Humidification chamber
4. Active humidifier
5. Cuff with temperature port
6. Y piece
10
11
2
1
1
3
4
10
7. Angled adapter for endotracheal tube
8. Pressure line connection port
9. Aerogen Pro
10. Expiratory patient tube
11. Water trap
12. Expiratory filter
12
SERVO-airVentilator System v2.1, User's Manual
27
| System Overview |
| 2
10 mm Ø patient circuit, dual heat disposable
11
5
10
9
6
7
8
1. Inspiratory patient tube
2. Water autofill
3. Humidification chamber
4. Active humidifier
5. Cuff with temperature port
6. Extension tube for incubator use
7. Angled Y piece
8. Pressure line connection port
2
1
1
3
4
5
13
15
12
9. Angled adapter for endotracheal tube
10. Y piece
11. Aerogen Solo
12. Expiratory patient tube
13. Expiratory filter
14. Expiratory heater Servo Duo Guard
15. Thermoshell, expiratory cassette
14
28
SERVO-airVentilator System v2.1, User's Manual

2.5 User interface

1

2.5.1 User interface adjustment

2
2. Loudspeaker1. Alarm indicator

2.5.2 Interactive areas

| System Overview |
2 |
The user interface can be adjusted into different positions.
1
2
3
4
5
The user interface is completely touch based and is divided into the following areas:
1. Status bar
2. Quick menu/extended menu
3. Display area
4. Direct access bar/media bar
5. Numerical values
SERVO-airVentilator System v2.1, User's Manual
29
| System Overview |
| 2
Important:
Do not use sharp tools on the screen.
Fluid on the screen can disturb touch
functionality.

2.5.3 Navigating

To navigate the user interface, adjust settings and get support:
Tap (the touchpad changes color when the
navigating is registered).
Tap and hold
Scroll vertically or horizontally
Drag and drop

2.5.4 User support

The user is supported by the following:
Alarm management
Safety scales
Dynamic images
Information texts
Pre-use check instructions
Shortcuts
Prompts
Note:
The following colors are used for settings:
Red — not recommended
Yellow — use with caution
Green — normal
Alarm management
1
2
30
3
4
1. Alarm list
2. Number of active alarms
3. Alarm management checklist
4. Alarm history Refer to chapter Alarm handling on page 109.
SERVO-airVentilator System v2.1, User's Manual
Loading...
+ 139 hidden pages