LNE/G-MED certies that the quality system developed by MAQUET SA for design, implementation, sales,
installation and after-sales service of surgical lights complies with the requirements of the following interna-
tional standards:
• ISO 9001:2000
• NF EN ISO 13485:2004
Reference standards
The BLUE 30/80 surgical light was designed to comply with the following standards:
• EN ISO 14971:2000 Medical devices - Application of risk management to medical devices
(ISO 14971:2000)
• EN ISO 14971:2000/A1:2003
• EN 60601-1:1990 Medical electrical equipment - Part 1: General requirements for safety
Amendment A1:1993 to EN 60601-1:1990
Amendment A2:1995 to EN 60601-1:1990
Amendment A13:1996 to EN 60601-1:1990
• EN 60601-1-2:2001 Medical electrical equipment - Part 1-2: General requirements for safety Collateral standard: Electromagnetic compatibility - Requirements and tests
• EN 60601-1-4:1996 Medical electrical equipment - Part 1-4: General requirements for safety -
Collateral standard: Programmable electrical medical systems
Amendment A1:1999 to EN 60601-1-4:1996
• EN 60601-1-6:2004 Medical electrical equipment - Part 1-6: General requirements for safety -
Collateral standard: Usability
• EN 60601-2-41:2000 Medical electrical equipment - Part 2-41: Particular requirements for the safety of
surgical luminaires and luminaires for diagnosis
This product has been veried for compliance with the following additional standards: CAN/CSA-C22.2
No. 601.1-M90 (R2005) (includes national differences for Canada), EN 60601-1:1990 + A1:1993 + A2:1995
+ A13:1996, UL 60601-1, rst edition, 2006-04-26 (includes national differences for the USA).
This product only meets FCC/UL requirements when equipped with a MAQUET SA power supply.
CE labelling/intended use
Compliance with the requirements of Directive 93/42/EEC on medical devices has been assessed and
approved by LNE/G-MED. BLUE 30/80 surgical lights belong to Class I as described in Annex IX of Directive
93/42/EEC.
This range includes single or dual ceiling-mounted, single wall-mounted and mobile light units with or without
battery-backed power supply.
Ceiling, wall-mounted or
mobile surgical light
0136201
4
Technical manual
BLUELINE
SYMBOLS USED IN THE MANUAL
SymbolMeaning
F
Mandatory:
may affect patient or user safety
Recommendation:
risk of damage to device or accessories
CE mark: The device complies with the requirements of
European Directive 93/42/EEC on Medical Devices.
This device complies with Canadian and US safety
requirements.
Ceiling, wall-mounted or
mobile surgical light
0136201
5
Technical manual
BLUELINE
1 GENERAL CHARACTERISTICS
(IN ACCORDANCE WITH STANDARD IEC 60 601-2-41)
SpecicationsUnit BLUE 30 BLUE 80
Nominal illumination*
Nominal voltageV DC22.822.8
Maximum powerW4080
Colour temperature
Colour rendering index>85>85
Illumination depthcm (inch)175 (69)140 (67)
Diameter d10
Diameter d50
With one mask%0 %6 %
lx
+/- 10%
K
+/-6%
cm (inch)
+/- 10%
cm (inch)
+/- 10%
35,00090,000
4600 4600
17 (6.7)16 (6.3)
10 (3.9)10 (3.9)
With two masks%62%49 %
At base of tube%100 %100 %
Shadow dilution
With one mask, at base of tube%0 %6 %
With two masks, at base of tube%62 %49 %
Irradiance (Ee)*W/m
* Values measured during certication:
BLUE 30: 38.400 lx / 139 W/m
BLUE 80: 85.500 lx / 305 W/m
2
2
2
140360
Ceiling, wall-mounted or
mobile surgical light
0136201
6
Technical manual
BLUELINE
2 CLEANING / DISINFECTION / STERILISATION
Users must contact their hospital's sanitary specialists. The recommended products and procedures must
be applied. Should there be any doubt concerning the compatibility of active agents to be used, contact the
local Maquet customer service department.
2.1 Cleaning and disinfecting the surgical light
F
2.1.1 General instructions concerning cleaning, disinfection and safety
• Remove the sterilisable handles.
• Clean the system using a cloth soaked in surface detergent, in line with the manufacturer's recom-
• Use a cloth to rinse the unit with clean water and wipe dry.
• Disinfect uniformly using a cloth soaked in a disinfectant, in line with the manufacturer’s recommenda-
• Use a cloth to rinse with clean water in order to remove residues (in particular products containing al-
• Wipe with a dry cloth.
2.1.2 Examples of recommended products
Getinge USA product: TEC-QUAT 256
Anios products: SURFA’SAFE, 0.5% HEXANIOS G + R (ammonium IV, polyhexanide); ANIOSYME P.L.A
Schülke & Mayr products: Antifect Plus (Glyoxal), Terralin (Benzyl-C12-18-alkyldimethyl ammonium, phe-
noxypropan and phenoxypropanol).
2.1.3 Examples of prohibited products
F
Check that the power is switched off and the light is cool before cleaning.
mended dilution, application time and temperature.
tions.
dehydes, quaternary ammonium or surfactants).
Solutions containing glutaraldehyde, phenol, iodine, bleach, alcohol or chloride ions must not be
used.
Fumigation methods are unsuitable for disinfecting the unit and must not be used.
2.2 Cleaning and sterilising the handles
2.2.1 Before cleaning
• Use a soft cloth immediately after use to wipe away soiling from the handle surface.
• Store handles in a place that keeps them moist to make further cleaning easier.
• Take care to store them in such a way that the inside does not get soiled.
2.2.2 Cleaning
• Soak the handles in a detergent solution.
• Soak for 15 minutes to allow the solution to act, then clean by hand with a soft brush and a lint-free
cloth.
• During cleaning, check regularly that the handles are fully clean and that no soiling remains on the
inside or outside.
• If any soiling remains, repeat cleaning or use an ultrasonic cleaning process.
• Rinsing: Rinse thoroughly in clean water to completely eliminate the detergent solution.
• Drying: Wipe with a clean lint-free cloth.
1
Never soak the handles in enzyme-based detergents as they may damage the material used; rinse thoroughly if
these detergents are used.
1
Ceiling, wall-mounted or
mobile surgical light
0136201
7
Technical manual
BLUELINE
2.2.3 Washing and disinfection
Handles may be disinfected by machine (Clean Maquet) and rinsed at a maximum temperature of 93°C.
Typical recommended cycles:
2.2.4 Sterilisation
StepTemperatureTime
Pre-wash18 - 35°C60 sec
Wash46 - 50°C300 sec
Neutralisation41 - 43°C30 sec
Wash 224 - 28°C30 sec
Rinse92 - 93°C600 sec
Dry20 min
After cleaning, the handles must be steam sterilised as set out below:
Countries
USA & Canada Prevacuum
France
Other countries Prevacuum Comply with national regulations
Sterilisation cycle
2
ATNC (Prion)
(Prevacuum)
Temperature [°C] Time [min] Drying [min]
132 - 135 1016
13418
• Check that each handle is clean before continuing the process.
• Wrap the handles with sterilisation wrapper material (double wrapper or equivalent).
Handles may also be placed in paper or plastic sterilisation bags3, for easier identication and reuse.
• Place the handles on steriliser trays with the opening downwards.
• Package with biological and/or chemical indicators for monitoring the sterilisation process, in accord-
ance with applicable regulations.
• Run the sterilisation cycle according to the steriliser manufacturer's instructions.
F
F
• Dispose of sterilisable handles in the same way as other hazardous products in a hospital environ-
To ensure correct sterilisation do not allow any soiling to penetrate inside the handle.
Handles are not guaranteed beyond 350 sterilisation cycles with the above sterilisation param-
eters.
ment.
4
2
This handle is made of a porous material.
3
Possible sterilisation bag suppliers:
Medical Action Industries
SBW Medical
Baxter International
4
For air removal and faster drying.
Ceiling, wall-mounted or
mobile surgical light
0136201
8
Technical manual
BLUELINE
3 GENERAL MAINTENANCE
Isolate the surgical light from all electrical power sources and/or check that it is connected to an
isolating transformer at very low voltage.
If the spring arm is not disabled when dismantling the lighthead, the arm may spring up abruptly
and present a hazard. The rst step is therefore to disable the spring arm.
The light unit becomes hot during operation. Allow it to cool down before any maintenance or
repair work.
3.1 Preventive maintenance
To preserve your surgical light’s original performance and reliability, annual maintenance and inspections
should be performed as follows:
• By a Maquet technician or Maquet-approved distributor during the guarantee period.
• By a Maquet technician or Maquet-approved distributor or by the hospital's technical maintenance de-
partment outside the guarantee period (contact your dealer to arrange the required technical training).
3.2 First level maintenance
3.2.1 Daily inspection (user)
• Check that the bulbs operate correctly.
• Check that the sterilisable handle locks correctly in place, and replace it if not.
• Check the arm position.
• If a backup power supply is installed, check that the light would operate correctly if a power cut were to
occur.
• Check the overall condition of the unit: paintwork in good condition, screws fastened tightly, bezels and
covers in position.
3.2.2 Annual inspection (must be performed by an authorised technician)
• Check that the limit stops are in place and in good condition.
• Check and tighten the brakes.
• Check the bulb holder and replace if necessary.
• Check that the lighthead limit stops are in place and in good condition.
• For the BLUE HOSPITAL, check the battery autonomy and the backup supply changeover system (see
section 5.4).
• Check that the underside is in good condition and replace it if necessary.
Safety items
Check the following points:
• Attachment screws on suspension tube correctly tightened, seals in position.
• Arm(s) correctly mounted.
• All covers and caps installed.
Other checks
• Nominal illumination: see technical data.
• Earth continuity: 0.1 Ohm maximum.
• Suspension tube vertical.
• Balancing system adjusted correctly.
• Sterilisable handle locking mechanism.
• Voltage.
Ceiling, wall-mounted or
mobile surgical light
0136201
9
Technical manual
BLUELINE
Dismantling certain elements may affect operation and safety. For example:
• Servicing the electrical power supply.
• Servicing the suspension arm and balance system.
• Servicing the optical system of lightheads equipped with lters designed to eliminate radiation not
visible to the patient. Surgical lights must never be used without these lters.
Contact the authorised Maquet after-sales service department for this type of inspection.
Ceiling, wall-mounted or
mobile surgical light
0136201
10
Technical manual
BLUELINE
BLUE 30
2
1
Figure 1
3
Figure 3
4
Figure 2
Figure 4
Figure 5
Figure 7
Figure 6
Ceiling, wall-mounted or
mobile surgical light
0136201
11
Technical manual
BLUELINE
3.3 Changing the bulb
BLUE 30 lighthead
F
Figures 1 and 2
• Press the locking tab (1) and raise the casing of the failed bulb (2).
Figure 3
• Remove the bulb holder (3), taking care not to knock the hot bulb against any surfaces.
Figure 4
• Take the failed bulb (4) and remove it from the holder.
• Take the new bulb and remove it from its packaging.
Figure 5
• Insert the bulb pins into the bulb holder as far as they can go.
• Check that the bulb is correctly seated.
F
Figure 6
• Replace the bulb holder (3) and check that the locking tab clicks audibly into place.
Figure 7
• Close the cover and check that the locking tab clicks audibly into place.
F
When changing a blown bulb:
• Switch off the power supply and leave the lighthead to cool for 25 minutes.
• Only use MAQUET bulbs.
• Handle bulbs carefully through a cloth or wearing gloves.
• Never touch bulbs with bare hands. Grease on bulbs can shorten their life.
Do not change bulbs during an operation.
The unit is designed to operate with the cover closed. When conducting maintenance, take precautions to avoid touching surfaces marked as hot with the following symbol:
To avoid failures during operations, we recommend changing the bulbs on a preventive basis
every 600 hours.
If the bulb cannot be inserted fully, do not force it beyond the limit stop.
Check that the cover is correctly in place.
Ceiling, wall-mounted or
mobile surgical light
0136201
12
Technical manual
BLUELINE
BLUE 80
4
Spare bulb
Figure 1
1
Figure 2
1
1
3
Main bulb
2
Figure 3
Figure 5
2
1
Figure 4
5
Figure 6
1
Figure 7
Figure 8
Ceiling, wall-mounted or
mobile surgical light
0136201
13
Technical manual
BLUELINE
BLUE 80 lighthead
F
Figures 1 and 2
• Push the cover in slightly (1) and slide it sideways.
Figures 3 and 4
• First remove the main bulb holder (3), then the spare bulb holder (4), taking care not to knock the hot
Figure 5
• Take the failed bulb (5) and remove it from the holder.
• Take the new bulb and remove it from its packaging.
• Insert the bulb pins into the bulb holder as far as they can go.
F
Figure 6
• Check that the bulb is correctly seated.
Figure 7
• Replace the spare bulb holder rst, then the main bulb holder, checking that the locking tab clicks audi-
Figure 8
• Close the cover (1) and check that it is correctly in place.
When changing a blown bulb:
• Switch off the power supply and leave the lighthead to cool for 20 minutes.
• Only use MAQUET bulbs.
• Handle bulbs carefully through a cloth or wearing gloves.
• Never touch bulbs with bare hands. Grease on bulbs can shorten their life.
Do not change bulbs during an operation.
The unit is designed to operate with the cover closed. When conducting maintenance, take precautions to avoid touching surfaces marked as hot with the following symbol:
To avoid failures during operations, we recommend changing the main and spare bulbs on a preventive basis every 750 hours.
The yellow indicator on the underside of the BLUE 80 lighthead is lit if the spare bulb is activated.
To avoid a total failure of the surgical light, change the main bulb as soon as possible. We recommended changing the spare bulb at the same time.
bulb against any surfaces.
If the bulb cannot be inserted fully, do not force it beyond the limit stop.
bly into place in each case.
Ceiling, wall-mounted or
mobile surgical light
0136201
14
Technical manual
BLUELINE
2
1
1 Bezel on mobile arm of spring arm
2 Bezel on xed arm of spring arm
3 Adjustment nut
3
Figure 33
1
1 Bezel on mobile arm of spring arm
2 Set screw
Figure 34
2
Ceiling, wall-mounted or
mobile surgical light
0136201
15
Technical manual
BLUELINE
3.4 Adjusting the balance
Figure 33
F
• Remove the bezel from the mobile arm of the spring arm (1) and, for the BLUE 80, the bezel from the
• Turn the adjustment nut (3) by no more than half a turn, holding the hexagon socket screw on the op-
• If the lighthead goes down, turn to the right.
• If the lighthead goes up, turn to the left.
• Check that the spring arm remains steady in each position.
• Replace the bezel(s).
3.5 Adjusting the lighthead top limit stop
Figure 34
F
• Carefully remove the bezel from the mobile arm of the spring arm (1).
• Position the lighthead at the maximum desired height.
• Tighten the two socket screws (2) until resistance is felt.
If the spring arm does not remain steady in all positions, the force of the spring in the arm can be
increased or decreased.
xed arm (2).
posite side in position for the BLUE 80.
BLUE 80
BLUE 80 lightheads can be adjusted so that the travel cannot exceed the ceiling height.
The spring arm may be damaged if the socket screws are not tightened evenly.
Tighten the socket screws uniformly.
Ceiling, wall-mounted or
mobile surgical light
0136201
16
Technical manual
BLUELINE
BLUE 30
1
2
3
1 Locking tab
2 Brake screw
3 Retaining screw (do not loosen)
BLUE 80
2
Figure 35
1
Figure 36
1 Brake screw
2 Lighthead fork
Ceiling, wall-mounted or
mobile surgical light
0136201
17
Technical manual
BLUELINE
3.6 Adjusting the lighthead brake
The lighthead brakes are adjusted during installation. Brakes, like all mechanical parts, are subject
to wear. If the lighthead no longer remains steady in all positions the brake must be readjusted.
The light unit becomes hot during operation. Allow it to cool down before adjusting the brake.
BLUE 30 lighthead
Figure 35
• Turn off the BLUE 30 lighthead and allow it to cool down.
• Press the locking tab (1) to remove the bezel.
• Adjust the lighthead braking by tightening or loosening the brake screw (2). The retaining screw (3)
holds the lighthead in place and must not be loosened.
• Close the cover and check that the locking tab clicks audibly into place.
• Check that the cover is correctly in place.
BLUE 80 lighthead
Figure 36
• Turn off the BLUE 80 lighthead and allow it to cool down.
• Adjust the lighthead braking by tightening or loosening the brake screw (1) above the lighthead fork
(2).
Ceiling, wall-mounted or
mobile surgical light
0136201
18
Technical manual
BLUELINE
2
3
1
Figure 4
465
2
7
Figure 5
Ceiling, wall-mounted or
mobile surgical light
0136201
19
Technical manual
BLUELINE
3.7 Removing and installing the BLUE 80 lighthead
If the spring arm is not disabled when removing the lighthead, the arm may spring up abruptly and
present a hazard.
• Lower the spring arm and lock it in a safe position when installing the lighthead.
• Assemble in the correct order: rst install the lighthead then enable the arm.
• Place the arm in the topmost position when removing the lighthead.
• Keep clear of the spring arm when enabling it.
Removing the BLUE 30 lighthead
Figure 4
• Place the spring arm (2) in the top position.
• Remove the two taper-head hex socket screws (1) to disable the spring arm.
Figure 5
F
• Unscrew the hexagon socket screw (5).
• Unscrew the brake screw (7).
• Remove the fork (6) from the spring arm and separate the connectors (4).
The hex socket screw (5) is a safety screw* that must be replaced each time the unit is
reassembled.
Reinstalling the BLUE 30 lighthead
Figure 4
F
• To disable the spring arm, unscrew the two taper-head hex socket screws (1).
Figure 5
• Secure the connectors (4).
F
• Insert the fork (6) in the spring arm (2) such that the hole in the spring arm lines up with the hole in the
• Lock the fork in place with the hex socket safety screw (5).
• Use a at screwdriver to tighten the brake screw (7).
Figure 4
• To enable the spring arm, tilt the lighthead and spring arm (2) upwards until the holes (3) line up with
• Install the two taper-head hex socket screws (1).
* Safety screws are available in packs of 5 (3 690 51 555).
The spring arm must be disabled before installing the BLUE 30 lighthead.
The hex socket screw (5) is a safety screw that must be replaced each time the unit is
reassembled.
fork.
the tappings (3).
Ceiling, wall-mounted or
mobile surgical light
0136201
20
Technical manual
BLUELINE
1
2
4
8
2
5
76
3
Figure 9
Figure 10
Ceiling, wall-mounted or
mobile surgical light
0136201
21
Technical manual
BLUELINE
3.8 Removing and installing the BLUE 80 lighthead
The order of disassembly is important:
If the spring arm is not disabled when disassembling the BLUE 80 lighthead, the arm may spring
up abruptly and present a hazard. The rst step is therefore to disable the spring arm.
Disabling the BLUE 80 spring arm
Figure 9
• Remove the bezel (1) from the mobile side of the spring arm (2).
• Pivot the spring arm down to the oor and insert the locking pin (3) in hole (4).
If the locking pin is not inserted deeply enough in the hole the spring arm may spring up abruptly
and present a hazard.
Check that the spring arm correctly locked in place before releasing it.
Only use the supplied locking pin.
Removing the BLUE 80 lighthead
Figure 10
F
• Unscrew the hexagon socket screw (5).
• Unscrew the brake screw (6).
• Remove the fork (7) from the spring arm and separate the connectors (8).
Reinstalling the BLUE 80 lighthead
Figure 10
• Secure the connectors (8).
F
• Insert the fork in the spring arm such that the hole in the fork lines up with the hole in the ring.
• Lock the fork in place using the shakeproof hex socket screw (5) and the brake screw (6).
Enabling the BLUE 80 spring arm
Figure 9
• Lower the spring arm (2) and lighthead slightly by pressing on it.
• Remove the locking pin (3).
• Install the bezel (1) and check that it is correctly in place.
The hex socket screw (5) is a safety screw that must be replaced each time the unit is reassembled.
The hex socket screw (5) is a safety screw* that must be replaced each time the unit is
reassembled.
* Safety screws are available in packs of 5 (3 690 52 555).
Ceiling, wall-mounted or
mobile surgical light
0136201
22
Technical manual
BLUELINE
3.9 Checking the BLUE 30/80 HOSPITAL
F
To ensure that the surgical light will operate correctly in the event of a power failure it must be
checked before each use.
LED 1
CheckMains
LED 1
Plug the light into
the power outlet
and turn it on
Disconnect the
power outlet (the
light should remain
on)
Testing the autonomy of the batteries in the BLUE 30/80 HOSPITAL light
F
To guarantee that the light will operate correctly in the event of a power cut, the battery autonomy
must be checked at least once a year, as part of preventive maintenance work.
CheckMains
Turn off the light
and charge the
batteries for at
least 14 hours
Turning on the light LED greenLED offLEDs 3 to 8
Disconnect the
power outlet (the
light should remain
on)
BLUE 30:
After 1 hour's battery operation
BLUE 80:
After 4 hours' battery operation
Connect the power
outlet
LED greenLED offLEDs 3 to 8
LED offLED yellowOne of LEDs 3 to
LED 1
LED greenLED offLEDs 3 to 8
LED offLED yellowOne of LEDs 3 to
LED offLED yellowOne of LEDs 3 to
LED greenLED offLEDs 3 to 8
Batteries
LED 2
Batteries
LED 2
Charging
LEDs 3 to 8
scrolling
LED 8 ashesBatteries
8 lit
(battery charge
level)
Charging
LEDs 3 to 8
scrolling
LED 8 ashesBatteries
scrolling
LED 8 ashesBatteries
8 lit
(battery charge
level)
8 lit
(battery charge
level)
scrolling
LED 2
LEDs 3 to 8
Comment
Batteries charging
completely
charged
Operation on
batteries
Comment
Batteries charging
completely
charged
Batteries charging
completely
charged
Operation on
batteries
Operation on
batteries
Batteries charging
Ceiling, wall-mounted or
mobile surgical light
0136201
23
Technical manual
BLUELINE
Anomalies
OperationAnomalyLikely causeCorrective action
On power outlet
(light on)
On batteries (light
on)
LED 1 not lit greenElectronic faultCall Maquet technical
department
LED 2 lit yellowMains fuse missing or
blown
LED 1 ashes redCharging circuit safety
fuse fault
LEDs 3 to 8 not scrolling;
LED 8 not lit
LED 2 not lit yellowElectronic faultCall Maquet technical
None of LEDs 3 to 8 litElectronic faultCall Maquet technical
The light goes out when
the power outlet is
disconnected
LED 4 ashesBatteries dischargedRecharge batteries
Electronic faultCall Maquet technical
Batteries faulty or
incorrectly connected
Charging circuit safety
fuse fault
Electronic fault (charger
board)
Change the fuse
Change the fuse
(quick-blow automotive
type, 7.5 A/32 V)
department
department
department
Check the connections
and change the batteries if
necessary
Change the fuse
(quick-blow automotive
type, 7.5 A/32 V)
Call Maquet technical
department
F
LED 3 lit red
LED 1 lit red
No warning is provided if the batteries are completely discharged or out of order. The fault may be
checked by disconnecting the power outlet. No indicators are lit in this case.
Batteries close to deep
discharge
Recharge batteries
urgently
Ceiling, wall-mounted or
mobile surgical light
0136201
24
Technical manual
BLUELINE
4 MAINTENANCE PROCEDURES
Required tools
• Cross-head screwdriver
• Torx screwdriver
• Allen wrench
4.1 BLUE 30
4.1.1 Removing the BLUE 30 lighthead cover
• Press the tab.
• Lift the cover.
4.1.2 Reinstalling the BLUE 30 lighthead cover
• Proceed in reverse order.
Check that the tab is correctly located.
4.1.3 Replacing the reector subassembly
• Remove the BLUE 30 lighthead cover.
• Disconnect and remove the lamp bracket.
• Unscrew and remove the four screws and wash-
ers.
• Remove the reector subassembly.
The spacers (and lens) may adhere to the
reector slightly.
• Install the new reector subassembly on the
spacers.
• Put the screws in place.
• Tighten the screws.
• Connect and position the lamp bracket.
• Reinstall the BLUE 30 lighthead cover.
4.1.4 Replacing the lens
• Remove the BLUE 30 lighthead cover.
• Remove the reector subassembly.
• Remove the lens.
• Remove the four spacers from the lens.
• Take the new lens and place the four spacers 90°
apart.
• Place this assembly on the base of the BLUE 30
lighthead.
• Reinstall the reector subassembly.
• Reinstall the lamp bracket.
• Reinstall the BLUE 30 lighthead cover.
Ceiling, wall-mounted or
mobile surgical light
0136201
25
Technical manual
BLUELINE
4.1.5 Replacing the spacers
Always change the four spacers at the
same time.
• Remove the BLUE 30 lighthead cover.
• Remove the reector subassembly.
• Remove the lens.
• Remove the four spacers.
• Place the four new spacers 90° apart on the lens.
• Place this assembly on the base of the BLUE 30
lighthead.
• Reinstall the reector subassembly.
• Reinstall the lamp bracket.
• Reinstall the BLUE 30 lighthead cover.
4.1.6 Replacing the underside
• Remove the BLUE 30 lighthead cover.
• Loosen the brake screw.
• Loosen the screw to allow the fork to slide.
• Unscrew the screw.
• Remove the sterilisable handle bracket.
The handle bracket may fall.
• Remove the screw.
• Remove the underside (with plastic washer and clip).
The underside may fall.
• Put the new underside in place.
• Re-attach the side with the clip.
• Install the sterilisable handle bracket.
• Tighten the screw holding the sterilisable handle bracket.
• Reposition the fork in the base of the BLUE 30 lighthead.
• Tighten the limit stop screw.
Check that it is correctly in place.
• Tighten the brake screw.
• Reinstall the BLUE 30 lighthead cover.
Ceiling, wall-mounted or
mobile surgical light
0136201
26
Technical manual
BLUELINE
4.1.7 Replacing the switch
• Remove the BLUE 30 lighthead cover.
• Remove the switch, separating the pivoting head
from the rest of the switch.
• Disconnect the switch from the connectors.
• Connect the new switch.
The wires are colour-coded.
• Separate the pivoting head to put the switch in
place.
• Reinstall the BLUE 30 lighthead cover.
4.1.8 Replacing the fork
• Remove the lighthead from the spring arm.
• Remove the lighthead from the spring arm.
• Remove the BLUE 30 lighthead cover.
• Disconnect the connector connected to the lamp
• Disconnect the connectors from the switch.
• Unscrew the brake screw to release the fork.
• Unscrew the limit stop screw.
• Remove the fork from the base of the BLUE 30
• Slide the wiring through rst.
• Position the fork in the base of the BLUE 30
• Tighten the limit stop screw.
• Tighten and adjust the brake screw.
• Connect the connectors to the switch.
The spring arm may spring up sharply.
bracket.
lighthead along with the wiring.
body.
Check that it is correctly in place.
The wires are colour-coded.
• Connect the connector to the lamp bracket.
• Reinstall the BLUE 30 lighthead cover.
Ceiling, wall-mounted or
mobile surgical light
0136201
27
Technical manual
BLUELINE
4.1.9 Replacing the lamp bracket
• Remove the BLUE 30 lighthead cover.
• Disconnect the lamp bracket.
• Remove the lamp bracket from the reector
subassembly.
• Connect the new lamp bracket.
• Place the lamp bracket on the reector.
• Reinstall the BLUE 30 lighthead cover.
Ceiling, wall-mounted or
mobile surgical light
0136201
28
Technical manual
BLUELINE
4.2 BLUE 80
4.2.1 Removing the BLUE 80 lighthead cover
• Remove the four central screws from the underside.
• Remove the BLUE 80 cover.
4.2.2 Reinstalling the BLUE 80 lighthead cover
• Put the BLUE 80 cover in place.
• Tighten the four screws.
4.2.3 Replacing the reector subassembly
• Remove the BLUE 80 cover.
• Remove the lamp brackets from the reector.
• Unscrew and remove the four screws and washers.
• Remove the reector subassembly.
Avoid touching the inside of the reector.
• Install the new reector subassembly on the spacers.
Ensure that it is correctly oriented relative
to the fork.
• Put the screws in place and tighten them.
• Put the lamp brackets back in position.
• Reinstall the BLUE 80 cover.
4.2.4 Replacing the lens
• Remove the BLUE 80 cover.
• Remove the reector subassembly.
• Remove the lens.
• Remove the four spacers from the lens.
• Take the new lens and put the four spacers in place.
• Place the assembly on the base of the BLUE 80.
• Reinstall the reector subassembly.
• Reinstall the lamp brackets.
• Reinstall the BLUE 80 cover.
Ceiling, wall-mounted or
mobile surgical light
0136201
29
Technical manual
BLUELINE
4.2.5 Replacing the spacers
Always change the four spacers at the
same time.
• Remove the BLUE 80 cover.
• Remove the reector subassembly.
• Remove the lens.
• Remove the four spacers from the lens.
• Put the four new spacers in place.
• Place the assembly on the base of the BLUE 80.
• Reinstall the reector subassembly.
• Reinstall the lamp brackets.
• Reinstall the BLUE 80 cover.
4.2.6 Replacing the underside
• Remove the six visible screws from the underside.
Four of these six screws hold the cover in
place. It may fall after they are removed.
• Remove the two screws from the sterilisable handle bracket.
The underside may fall.
• Put the new underside in place.
• Install the sterilisable handle bracket.
• Tighten the screws holding the sterilisable han-
dle.
• Tighten the six visible screws on the underside.
• Ensure that the two short screws are placed at
the join with the ends of the fork.
• Reinstall the BLUE 80 cover.
4.2.7 Replacing the switch
• Remove the BLUE 80 cover.
• Remove the switch, separating the pivoting head
from the rest of the switch.
• Disconnect the connectors from the switch.
• Connect the new switch.
• Separate the pivoting head to put the switch in
place.
• Reinstall the BLUE 80 cover.
Ceiling, wall-mounted or
mobile surgical light
0136201
30
Technical manual
BLUELINE
4.2.8 Replacing the ap
• Remove the BLUE 80 cover.
• Remove the reector subassembly.
Take care not to let the lens fall.
• Unscrew the two screws holding the ap.
• Check the ap.
• Put the new ap in place.
• Tighten the two screws holding the ap.
• Reinstall the reector subassembly.
• Reinstall the BLUE 80 cover.
4.2.9 Replacing the fork bracket
• Remove the BLUE 80 cover.
• Remove the reector subassembly.
• Remove the ap.
Remove the lens to avoid it falling by accident.
• Remove the underside.
• If both fork brackets are to be replaced, remove
and reinstall one fork at a time.
Left side:
• Unscrew (but do not remove) the screw from the
limit stop.
• Remove the four screws.
• Remove the fork bracket from the base and from
the fork.
• Install the new fork bracket.
Check that the limit stop screw, brake
screw and nut are in place on the new fork
bracket.
• Tighten the four screws holding the fork bracket
to the base.
• Position the fork bracket correctly in the fork.
• Tighten the limit stop screw.
Loading...
+ 82 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.