9|Important
17|System overview
47|Startup and system checkout
65|System functionality
119|Breathing system
149|AFGO (Additional Fresh Gas Outlet, Option)
155|Membrane buttons
183|Alarms and patient safety
209|Ventilation modes
233|System shutdown
235|Routine cleaning and maintenance
265|Technical specifications
291|EMC Declaration
299|Definitions
301|Certificates
303|Index
309|Log sheet
FLOW-i 4.2, User's Manual
Infologic 1.67
3
| Table of contents |
4
FLOW-i 4.2, User's Manual
Infologic 1.67
1 Introduction
Table of contents
| Introduction |
6|Indications for use1.1
6|Intended use1.2
7|Using this manual1.3
1 |
FLOW-i 4.2, User's Manual
5
| Introduction |
| 1
1.1 Indications for use
The indication for FLOW-i Anesthesia System is administering inhalation anesthesia while
controlling the entire ventilation of patients with no ability to breathe, as well as in supporting
patients with a limited ability to breathe. The system is intended for use on neonatal to adult
patient populations. The system is intended for use in hospital environments, except MRI
environment, by healthcare professionals trained in inhalation anesthesia administration.
1.2 Intended use
The system is intended for use in administrating anesthesia while controlling the entire ventilation
of patients with no ability to breathe, as well as in supporting patients with a limited ability to
breathe.
The system is intended for use by healthcare professionals, trained in the administration of
anesthesia.
The system is intended for use on neonate to adult patient populations.
The system is intended for use in hospital environments, except the MRI environment.
When not in operation, the system is designed for in-hospital transport.
6
FLOW-i 4.2, User's Manual
| Introduction |
1 |
1.3 Using this manual
This manual is divided into the following
chapters:
• Chapters 1 and 2: Important information
before use, along with a basic introduction
to the system and User's Manual.
• Chapter 3: Overview of the functional parts
of the system.
• Chapters 4 through 11: Understanding and
working with the system.
• Chapter 12: Routine cleaning and
maintenance procedures.
• Chapter 13: The technical specification of
the system.
• Chapter 14: EMC Declaration
• Chapter 15: Terminology and definitions
• Chapter 16: Certificates
• Chapter 17: Index
Important information is highlighted with
Warning or Caution, where:
WARNING! Indicates critical information
about a potential serious outcome to the
patient or the user.
CAUTION: Indicates instructions that must
be followed in order to ensure the proper
operation of the equipment.
• Chapter 18: Log sheets
References made to optional accessories
within the FLOW-i User's Manual are
accompanied by the text 'option' in brackets.
FLOW-i 4.2, User's Manual
7
| Introduction |
| 1
8
FLOW-i 4.2, User's Manual
2 Important
Table of contents
| Important |
10|General information2.1
11|Connection2.2
12|Operation2.3
13|Installation and service2.4
14|Accessories and auxiliary equipment2.5
2 |
FLOW-i 4.2, User's Manual
9
| Important |
| 2
2.1 General information
• The following applies throughout this User's
Manual:
- 'FLOW-i', 'anesthesia system' and
'system' represent FLOW-i anesthesia
system version 4.2.
• This manual applies to the FLOW-i
anesthesia system models C20, C30 and
C40, and optional equipment that can be
fitted onto these.
• Only authorized personnel who are well
trained in its use should operate the
anesthesia system. It must be operated
according to the instructions in this User's
Manual.
• Gas volumes, flows and leakages
associated with the breathing system are
stated in the technical specifications and
adhere to BTPS reference conditions. (Body
temperature, ambient pressure, Saturated).
• All gas concentration readings are normally
referenced to dry gas conditions, ambient
room temperature and atmospheric
pressure (ATPD).
• The condition for measured inlet gas
pressures and flows is STPD (Standard
Temperature and Pressure Dry); 20° C,
standard pressure at 101.3 kPa and 0 %
relative humidity (dry).
• All data on pressures are given in either
cmH2O or bar, where
1 cmH2O = 1 hPa = 1 mbar
1 bar = 15 psi = 1 atm = 1 kgf/cm2 (kp/cm2)
• The anesthesia system is not made with
natural latex.
• Applied parts, i.e. equipment making
physical contact with the patient, comprise
gases (including agents) and the patient
mask.
• If the mains power supply is interrupted,
the internal battery will provide temporary
power to the system (approx. 90 minutes
when fully charged).
• The fresh gas/gas supply outlets are not
affected by switch to battery power.
• Malfunction of the central gas supply can
potentially cause one or several of the
devices connected to the system to stop
their operation simultaneously.
• When the system is in use, a backup gas
supply shall always be available.
• If the central gas supply is interrupted, the
backup gas cylinders O2/N2O or O2/Air
(option) will provide gas to the system.
• In case of automatic ventilation failure,
switch to manual ventilation. In case of
manual ventilation failure, switch to
emergency ventilation.
• In case of a total power (mains power and
battery) or other electronic failures, the
built-in emergency ventilation system can
be used.
• The system maintains its performance when
tilted up to two degrees.
• In case of a complete system failure,
immediate access to alternative means of
ventilation, e.g. a manually powered
resuscitator, must be ensured to avoid
possible patient injury.
• MAQUET takes full responsibility for
compliance of the CE mark requirements
for the MAQUET CO2 cartridge produced
by Molecular Products Ltd (MPL). MAQUET
takes full responsibility for the supply of the
instructions for use in accordance with the
legislation relevant for the intended use for
this product.
10
FLOW-i 4.2, User's Manual
| Important |
2 |
WARNING! To avoid the risk of electric
shock, this equipment must only be
connected to a supply mains with protective
earth.
CAUTION: Federal law restricts this device
to sale by or on the order of a physician.
2.2 Connection
• A full System checkout procedure must be
performed at least once a day.
• The system must never be left unattended
while connected to a patient.
• Electronic accessories and auxiliary
equipment other than the vaporizers must
not be connected or disconnected during
operation or when the system is plugged
into a mains power outlet. Such connection
or disconnection may interfere with the
functioning of the system.
• Supplied gases shall meet the requirements
for medical grade gases according to
applicable standards.
• The backup gas supply should only be
turned ON (valves open) when the backup
gas supply is in use, or during System
checkout.
FLOW-i 4.2, User's Manual
11
| Important |
| 2
2.3 Operation
• The system shall always be used in
combination with other vital signs
monitoring devices and/or professional
human judgements of patient condition.
• To protect the patient from high airway
pressure, an upper pressure limit must
always be set. See Chapter 10 for details.
• To protect the patient, an alarm limit must
always be set for low expired minute
volume. See Chapter 10 for details.
• Only anesthetic agents recommended by
MAQUET are suitable for use.
• Anesthetic agent bottles without keying may
not be used with the system, nor is it
allowed to tamper with the keying of
anesthetic agent bottles.
• Ensure that the operating room is properly
ventilated.
• The system is not intended for use during
interhospital transportation.
• The system is not intended for use in an MR
environment.
• Antistatic or electrically conductive
breathing tubes should not be used. If such
breathing tubes are used in combination
with high frequency electric surgery
equipment, burns may occur.
• To allow for mains power disconnection,
make sure that the power cable connected
to the mains power supply remains visible
and fully accessible during patient treatment
and not obstructed in any way by ME
(Medical Electrical) equipment.
• If the integrity of the protective earth
conductor or the protective earthing system
in the installation is in doubt, unplug the
mains power cable and use battery power.
12
• The equipotentiality terminal is designed for
the connection of a potential equalization
conductor according to DIN 42 801 and
IEC 60601-1. The function of the
equipotentiality terminal is to equalize
potentials between the system and other
medical electrical devices that can be
touched simultaneously. The
equipotentiality terminal must not be used
for a protective earth connection.
• The anesthesia system has been designed
and tested to comply with requirements
specified in electromagnetic compatibility
standard IEC 60601-1-2. It is the
responsibility of the user to take necessary
measures to ensure that the EMC
environment in which the workstation will
be used is compatible with the requirements
of IEC 60601-1-2 and that the installation
is carried out according to the EMC
information. See Chapter 16 for details on
EMC environments. If limits are exceeded,
the accuracy and safety of the system may
be impaired. Proactive measures include,
but are not limited to, avoiding the use of
portable and mobile radio-emitting devices,
such as cellular phones and high frequency
apparatus, in the proximity of the system.
• Full performance is reached after a 15
minute warm-up. In case of an emergency,
the system can be used immediately.
• FLOW-i shall be connected to a
centre-tapped single phase supply circuit
when connected to a 240 Vac
(L1-L2-N-GND) supply in the United States.
• The breathing system can handle negative
pressures down to -200 cmH2O, but is not
designed to withstand pressures below that.
• The operator must not touch the patient
and any of the following parts
simultaneously:
Accessible contacts of connectors-
FLOW-i 4.2, User's Manual
| Important |
2 |
- Contacts of fuse holders that are
accessible during replacement of the
fuse.
- Contacts of lamp holders that are
accessible after removal of the lamp.
- Parts inside access covers that can be
opened without the use of a tool, e.g.
patient cassette lid connector.
WARNINGS!
• In case of system failure, the lack of
immediate access to appropriate
alternative means of ventilation can result
in patient injury.
• Gas inlets and outlets shall not be
covered or in any other way be
obstructed.
• FLOW-i is not designed to be resistant
to direct exposure to high ionizing
radiation. Such exposure may result in
memory erasure and/or interruption of
ventilation.
2.4 Installation and service
• Installation, service and maintenance of the
system must be performed by personnel
trained and authorized by MAQUET.
• Instructions for installation, service and
maintenance, i.e. a service manual, is
available for personnel trained and
authorized by MAQUET.
• Only original spare parts from MAQUET
must be used in the system.
FLOW-i 4.2, User's Manual
13
| Important |
| 2
2.5 Accessories and auxiliary
equipment
• External equipment intended for connection
to signal input, signal output or other
connectors shall comply with relevant IEC
standards (e.g. IEC 60950 for IT equipment
and the IEC 60601 series for medical
electrical equipment). In addition, all such
combinations – systems – shall comply with
the standard IEC 60601-1 'Safety
requirements for medical electrical
systems'. Equipment not complying with
IEC 60601-1 shall be kept outside the
patient environment, as defined in the
standard.
• Any person who connects external
equipment to signal input, signal output or
other connectors has formed a system and
is therefore responsible for ensuring that
the whole system complies with the
requirements of IEC 60601-1. If in doubt,
contact a qualified medical technician or
your local representative.
• For optional equipment and accessories,
refer to the user documentation supplied
by the manufacturer.
• Use of an anesthesia gas scavenging
system is compulsory and must comply
with ISO 7396-2 (wall connection) and ISO
80601-2-13 (tubings).
• Values measured at the signal outputs of
the anesthesia system, which have been
processed in auxiliary equipment, must not
be used as a substitute for therapeutic or
diagnostic decisions. Such decisions can
be made only by staff with medical
expertise, and according to established and
accepted practice.
• If there should be any deviation between
information shown on the system and that
shown by auxiliary equipment, the
parameters shown on the system shall be
considered the primary source of
information.
• When electrical equipment is connected to
the auxiliary power outlet or
communications interface, a medical
electrical system (ME system) is effectively
created, potentially reducing the level of
safety. This could result in previously
unidentified risks to patient, users or third
parties. It is the responsibility of the user to
ensure that the connected equipment is
compatible with the requirements of IEC
60601-1.
• Connecting auxiliary equipment to the
auxiliary power outlet can potentially
increase leakage currents to values above
the allowable limits.
• External monitors or similar devices
connected to the VGA port of the system,
must be powered via a medical grade
isolation transformer. No other use is
allowed.
• The responsible organization should
identify, analyze, evaluate and control these
risks.
• Subsequent changes to the medical
electrical system could introduce new risks
and require additional analysis.
• Changes to the medical electrical system
include configuration changes, connection
of additional items, disconnection of items,
update or upgrade of connected equipment.
14
FLOW-i 4.2, User's Manual
| Important |
2 |
• The use of O2 and Air gas outlets may,
depending on central gas supply pressure
and ventilation settings, affect ventilation
performance.
• Reprocessing parts labelled 'Single-use'
will degrade biocompatibility and
cleanliness.
WARNINGS!
• The use of other accessories,
transducers and cables other than those
specified by MAQUET may result in
increased emissions or decreased
immunity (EMC) of the system.
• No other electrical equipment other than
those described in this user's manual
may be placed on, or in the immediate
vicinity of the system.
• The patient monitor power connector
must only be used for patient monitors
or equipment mounted on the top shelf.
CAUTIONS:
• Only MAQUET recommended
accessories, supplies and auxiliary
equipment must be connected to or used
in conjunction with the system. Use of
other unvalidated accessories, supplies
and auxiliary equipment may impair the
performance and safety of the system.
• To prevent the system from tilting, follow
the restrictions for patient monitors and
auxiliary equipment specified in Chapter
15, page 268. Maximum torque on the
side rails by the table is 20 Nm.
• Equipment placed on shafts or rails must
not restrict opening of the Emergency
Ventilation cover.
The anesthesia system is designed to enable the operator to work with the basic parts of the
system in the most suitable way for each procedure. As the system is mounted on wheels,
and the control panel is mounted on a rotatable arm, it can be easily moved into an
ergonomically-suitable position.
There are a number of different system models available. For more information on the different
models, see page 40.
3.1 System parts
The system comprises the following basic parts:
1
7
2
3
5
220-240V~
400 VA
T 1.6 A
6
250 V
T 4 A
250 V
5.Emergency ventilation system1.Control panel
2.Patient monitor (option)6.Vaporizer unit
7.3.Additional arm 'A' (option) and/or
Breathing system
Gas backup holder 'B' (option)
1
4.External connections
A
B
4
1. C30 can only be equipped with the additional arm or the extra backup gas holder.
18
FLOW-i 4.2, User's Manual
3.2 Control panel
01-01 13 00
| System overview |
3 |
312
3
2
1
The control panel includes:
1.Screen with active touch pads
2.Rotary knob
3.Membrane buttons
FLOW-i 4.2, User's Manual
19
| System overview |
| 3
3.2.1 Areas of the screen
10
9
8
7
6
111213
521
43
The screen is divided up into a number of different areas:
The touch screen and rotary knob allow the
operator to control the primary functions of
the anesthesia system, where the patient's
condition is monitored through measured
values and waveform displays.
Measured values and waveforms are displayed
on the screen in the following color groups:
Messages are displayed in the Alarm message
area or System message area. The following
color scheme is adopted:
Color for alarms and system messages
High priority alarms
Medium priority alarms
Low priority alarms
Technical alarms, i.e.
alarms with the prefix
TEXXX, where XXX is an
integer.
System messages
Black text on red
background
Black text on yellow
background
Black text on blue
background
Black text on red,
yellow or blue
background, dependent
on priority
(high/medium/low)
White text on black
background
A detailed description of alarms and patient
safety is found in Chapter 10.
The colors for O2 and N2O in the above table
may vary due to country-specific standards.
FLOW-i 4.2, User's Manual
21
| System overview |
| 3
3.2.2 Navigating the screen
There are several ways of navigating the
screen and setting values.
Using the touch screen
1.Press the required touch pad. The touch
pad becomes active, which is indicated
by a blue highlight.
2.Turn the rotary knob to the required value.
3.Press the touch pad to confirm setting.
An activated touch pad is only active for 20
seconds. The system will prompt the user to
enter a value when 10 seconds have passed.
If no value is entered and confirmed within the
following 10 seconds, the touch pad setting
returns to its previous setting and is
deactivated.
Using the rotary knob
1.Turn the Rotary knob to move between
the touch pads on the screen. The
selected touch pad is indicated by a blue
frame.
2.On required touch pad, press the rotary
knob to activate the touch pad. This will
highlight the touch pad in blue.
3.Turn the Rotary knob to the required
value.
4.Press the Rotary knob to confirm setting.
22
FLOW-i 4.2, User's Manual
Active screen
| System overview |
3 |
5
6
4
8
3
2
Pressing any of the areas labelled with a
number in the illustration, displays a window
according to the table below.
7.Mode indicator area
9.Tab area (displayed window depends on
selected tab)
7
9
1
Dialog/window producedScreen area
Multiple windows, see Chapter 51.Gas settings/ Ventilation settings
Patient category, see Chapter 5, page 75.2.MAC value
Alarm profile window, see Chapter 10, page 187.3.Gas measurement area
Alarm profile window, see Chapter 10, page 187.4.Ventilation measurement area
Patient category, see Chapter 5, page 75.5.Patient category symbol
Alarm profile window, see Chapter 10, page 187.6.Alarm functions
Ventilation mode selection, see Chapter 5, page
87.
Waveforms and scales, see Chapter 9, page 163.8.Waveform area
- Waveforms and scales, see Chapter 9, page 163.
- Screen layout, see Chapter 9, page 163.
FLOW-i 4.2, User's Manual
23
| System overview |
| 3
3.2.3 Touch pad settings
A range bar is located at the bottom of all
ventilation and gas touch pads, where a
numeric value can be specified by the user. If
the entered value deviates too far from the
norm given the current ventilation mode and
other parameter settings, the bar will change
color according to the table below.
DescriptionTouch pad setting
A black range bar
indicates normal
20
0120
cmH20
50
0120
cmH20
75
0120
cmH20
parameter setting.
A yellow range bar
indicates that the
parameter setting is
high (or low, depending
on the parameter).
A red range bar
indicates that the
parameter setting is
very high (or very low,
depending on the
parameter).
3.2.4 Membrane buttons
1
3
5
7
10
2
4
6
8
9
11
The change in color of the range bar is
accompanied by a system message that
remains displayed on the screen until the
trigger condition is relieved.
24
1.Audio pause
2.Alarm profile
3.Start case
4.End case
5.Save screen
6.Trends
7.Start/Stop timer
8.Reset timer
9.Home
10. Screen layout
11. Menu
See Chapter 9 for more information.
FLOW-i 4.2, User's Manual
| System overview |
3 |
3.3 Patient monitor (option)
The system can be connected to a selection
of different patient monitors. For full details,
contact your local MAQUET representative.
The patient monitor must comply with
IEC 60601-1 ed. 3.
During mains power failure, the backup battery
in the anesthesia system will not power the
patient monitor.
3.3.1 Panel interchangeability
If desired, the placement of the control panel
and the patient monitor can be switched so
that the patient monitor is placed on the
display arm. This procedure shall only be
performed by a service technician trained and
authorized by MAQUET. Contact your local
MAQUET supplier for more information.
3.4 Breathing system
4
1
2
13
12
11
10
9
8
7
The breathing system comprises the following:
1.Patient cassette
2.O2 flush
3.MAN/AUTO ventilation switch
4.APL valve
5.Volume reflector
6.AGS (Anesthetic Gas Scavenging) flow
indicator
7.Patient tubing
8.Manual breathing bag with tubing
9.Y-piece
10. AFGO - Additional Fresh Gas Outlet
(option)
11. Water trap and sampling line
12. CO2absorber
13. Auxiliary O2 and suction module (option)
53
6
FLOW-i 4.2, User's Manual
MAQUET recommends that a bacterial/viral
filter is always connected to the expiratory
connection on the patient cassette. This will
minimize the risk of cross-contamination.
25
| System overview |
| 3
3.5 Vaporizer unit
The vaporizer unit holds one to two vaporizers,
where vaporizers can be selected for the
following agents:
• Isoflurane
• Sevoflurane
• Desflurane
3.6 Emergency ventilation system
O
2
10
8
mbar / cmH2O
6
4
2
30
I/min
Emergency ventilation
O
I/min
2
In case of a total power (i.e. mains power and
battery) or system failure, this system will allow
the patient to be manually ventilated. The
emergency ventilation system comprises:
1.Activation switch
2.O2 gas supply and flowmeter, graded up
to 10 l/min.
3.Mechanical APL
O
2
10
8
6
4
2
I/min
I/min
O
2
Not for use with AFGO
APL
80
SP
mbar / cmH2O
APL
60
26
13
2
WARNING! If the emergency ventilation
system is activated while the anesthesia
system is in operation, the anesthesia
system will be shutdown.
FLOW-i 4.2, User's Manual
3.7 External connections
| System overview |
3 |
2
O
2
Air
1
N2O
3
The following external connections exist on
the system:
1.Power supply and fuses
2.Gas connections
3.Input/Output ports
FLOW-i 4.2, User's Manual
27
| System overview |
| 3
3.7.1 Power supply and fuses
1
2
3
4
5
6
7
9
8
10
1.Auxiliary power outlets fuse 2A (option)
2.Auxiliary power outlets (option)
3.Auxiliary power outlets fuse 1A (option)
4.Auxiliary power outlets fuse 1A (option)
5.Patient monitor fuse (option)
6.Isolation transformer fuse (option)
7.Mains power inlet fuses
8.Lift fuses (C30 model only)
9.Mains power inlet
10. Equipotential terminal (earth)
The patient monitor outlet and the auxiliary
power outlets do not have a battery backup.
Auxiliary power supply outlet connections vary
depending on country specific standards.
In addition to the individual fuses, the
maximum current delivered through the
system is regulated by a shared fuse.
In case of mains power failure, a battery
backup will power the system for a limited
time.
28
FLOW-i 4.2, User's Manual
| System overview |
3 |
WARNING!
Restrictions apply to the use of auxiliary
power outlets:
• Outlets must not be used to power life
support equipment unless the life support
equipment itself is equipped with battery
backup.
• A multiple socket extension cord must
not be connected to any of the outlets.
CAUTION: The auxiliary power outlets shall
only be used for supplying power to
equipment intended to form part of the
medical electric system.
For more details on the power supply and
status, see Chapter 10, page 203.
3.7.2 Gas connections
1
2
3
7
4
5
6
FLOW-i 4.2, User's Manual
1.O2 outlet (option)
2.O2 inlet
3.Air outlet (option)
4.Air inlet
5.N2O inlet
6.AGS outlet (country-specific connections)
7.Gas cylinder inlets for O2, Air and N2O
(option)
If equipment with high gas consumption are
connected to the gas outlets, the central gas
supply pressure should be above 3 bar (300
kPa, 44 PSI). Make sure that the central gas
supply is sufficient for the extra equipment.
29
| System overview |
| 3
The intended use for the Auxiliary O2 outlet is
to provide oxygen for patient therapy.
The intended use for the Auxiliary AIR outlet
is to provide drive gas for small ejector driven
suction devices.
Maximum allowed air flow from each of the
O
and Air outlets is 60 l/min.
2
The central N2O gas supply (wall supply) is
connected to the system via a pressure
regulator attached to the N2O gas inlet.
3.7.3 Input/Output ports
1
234
1.RS232 serial data communication ports
2.Control panel connection
3.VGA connector
4.USB port
5.Network connection
6.Cable restrainer
The intended use of the USB port is only for
USB memory flash drives. Restricted items
include, but are not limited to, external hard
drives, USB hubs and any equipment using
the USB port primarily as a power source.
6
5
Clinical situations that require high gas
consumption might affect the supply gas
pressure.
If the gas supply is low, i.e. close to 2.5 bar
(250 kPa, 36 PSI), the system's ability to
deliver gas according to all possible settings
might be compromised by additional gas
hoses and connections.
30
FLOW-i 4.2, User's Manual
Loading...
+ 287 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.