Maquet Alphamaxx Quick user manual

Brief operating instructions
ALPHAMAXX
1133.12B1/B3/F1/F3
Control of functions:
The electrohydraulically driven operating table functions can be controlled by
cable-connected hand control or
optional infrared hand control
optional foot switch
BACKW.
Motor drive for the base backward
LOCK
Operating table locked
LEFT
Left leg plate up or down, depending on the key which is pressed simultaneously LEG UP/DOWN
L.FOOT
Table top – longitudinal shift towards foot end
FORW.
Motor drive for the base forward
UNLOCK
Table top unlocked; can be displaced
RIGHT
Right leg plate up or down, depending on the key which is pressed simultaneously – LEG UP/DOWN
L.HEAD
Table top – longitudinal shift towards head end
BACK DN.
Back plate down
LEG DN.
Leg plate down
FLEX
(Kidney / gall bladder eleva­tion) controls the ”back plate” and ”inclination” func­tions alternatively until the flex position is achieved.
TREND
Table top – Trendelenburg
TILT LEFT
Lateral tilt left
DOWN
Table top down
LEVEL
Table top – horizontal alignment
BACK UP
Back plate up
LEG UP
Leg plate up
BEACH CHAIR
controls the ”back plate” and ”inclination” functions alter­natively until the reverse flex position is achieved, then the leg plates are lowered
REV. TREND
Reverse Trendelenburg po­sition
TILT RIGHT
Lateral tilt right
UP
Table top up
GK113312EN02
The back plate, seat plate and leg plate adjustments are motor-driven. The head rest is adjusted manually. The NORMAL patient orientation is preferable to the REVERSE patient orientation
.
Fig. 1: The preferred patient orientation
Precautionary measures
This operating table may not be used for a patient weight exceeding 450 kg (1000 lb.).
Surgical procedures are only permissible on the locked table in the patient orientation shown above.
Displacing (Unlock) the operating table with a patient weight between 135 kg (300 lb.) and 225 kg (500 lb.) is only permissible when the table is in its lowest position.
Accessories which themselves are quite heavy or which have long lever arms (retractor systems and the like) are particularly liable to cause damage to the side rails.
The user must ensure that the operating table is set up at a flat, level and solid location.
Some adjustment functions can cause collisions between accessories and the operating table and damages to the accessories.
ÎÎ
Î Note:
ÎÎ
Do not use the override control panel to set up extreme combinations of Trendelenburg/reverse Trende­lenburg and lateral tilt adjustments.
In case of malfunctions or if the hand control is defective, the most important operating table functions can be controlled via the control panel at the upper column casing (override system).
For more detailed information please refer to the Operating instructions GA113312ENxx
Phone: ++ 49(7222) 932-0, Fax: ++ 49(7222) 932-838, Service-Hotline: ++ 49(7222) 932-745,
MAQUET GmbH & Co.KG, Kehler Straße 31 D-76437 Rastatt,
e-mail: info.sales@maquet.de, Internet: http://www.maquet.com