Maquet Alphamaxx User manual

ALPHAMAXX
1133.02BX/FX
Operating instructions
Universal operating table
Englisch
GA113302GB11
1133.02BX
15
1
2
4
3
14
68
5
12
20
13
11
7
10
9
12
14
Mains cable 16
18
Potential equalization
Socket for mains cable
15
17
19
1 Head rest (e.g. 1130.64) 2 Fastening screws for head rest 3 Locking lever for upper back plate 4 Back plate, removable
1133.31X0 Back plate for general surgery 5 Connection point for removable back plates 6 Lower back plate 7 Bellows 8 Leg plates 9 Locking lever for leg plate abduction
10 Unlocking of leg plates 11 Seat plate 12 Infrared receiver and transmitter 13 Override system with status indicators for oper-
ating table
14 Sockets for cable-connected hand control and
foot switch
15 Socket for mains cable 16 Mains cable 17 Potential equalization 18 Cable-connected hand control (optional) 19 IR remote control (optional) 20 Interface for servicing
2
GA113302GB11
1133.02BX
Contents
I. Important safety instructions ....................................................................................................... 4
II. Important information .................................................................................................................. 6
1. Important notes on these operating instructions ....................................................................... 6
2. Using the operating table in accordance with the intended purpose .......................................... 6
3. Intended purpose ..................................................................................................................... 6
4. Fundamental safety aspects ....................................................................................................6
III. General description...................................................................................................................... 7
1. General features ....................................................................................................................... 7
2. Control of functions .................................................................................................................. 8
2.1. Cable-connected hand control – functions ......................................................................... 9
2.2. Override system = emergency operation ......................................................................... 10
2.3 IR remote control (optional) ............................................................................................. 12
2.3.1 IR system code ...................................................................................................... 13
2.3.2 IR charging station ................................................................................................. 14
2.4 IR wall control panel ........................................................................................................14
2.5 Control via foot switch ..................................................................................................... 14
2.6 Special functions............................................................................................................. 14
3. Table top ................................................................................................................................ 15
3.1 General features ............................................................................................................. 15
3.2 Table top components ..................................................................................................... 16
3.3 Attaching the back plate modules ................................................................................... 17
3.4 Removing the back plate modules................................................................................... 18
3.5 Leg plates ....................................................................................................................... 18
3.6 Table top padding ............................................................................................................ 19
3.7 Accessories.................................................................................................................... 19
4. Motor drive for the base .......................................................................................................... 20
IV. Use of the operating table ........................................................................................................... 21
1. General .................................................................................................................................. 21
2. Operation on battery power .................................................................................................... 21
3. Operation on mains power...................................................................................................... 22
4. HF surgery / Use of defibrillators ............................................................................................22
V. Patient positioning ..................................................................................................................... 23
1. General .................................................................................................................................. 23
2. Positioning a patient weighing up to 135 kg (300 lb.) .............................................................. 25
3. Positioning a patient weighing from 135 kg up to 225 kg (300 lb. up to 500 lb.) ...................... 28
4. Positioning a patient weighing from 225 kg up to 450 kg (500 lb. up to 1000 lb.) .................... 31
VI. Care and maintenance .............................................................................................................. 32
1. Cleaning ................................................................................................................................. 32
2. Disinfection ............................................................................................................................ 32
3. Maintenance .......................................................................................................................... 33
4. Malfunctions .......................................................................................................................... 33
5. Environmental protection ........................................................................................................ 34
VII. Technical data ............................................................................................................................ 35
1. Operating table adjustments .................................................................................................. 35
2. Electrical data ........................................................................................................................ 35
VIII.List of accessories ...................................................................................................................... 36
1. Accessories for the head rest interface .................................................................................. 38
2. Accessories for the back plate interface................................................................................. 40
3. Accessories for the legplate interface..................................................................................... 40
4. Accessories for attachment to the side rails........................................................................... 41
5. Special accessories ............................................................................................................... 41
6. Control modules ..................................................................................................................... 42
IX. Spare parts list ............................................................................................................................ 42
GA113302GB11
3
I. Important safety instructions
I. Important safety instructions
Mount only the accessories listed in Chapter VIII.
Make sure that the operating table is set up horizontally and is stable!
Lock the operating table in position prior to use! Press the LOCK key!
Lock the operating table in position prior to transferring the patient! Press the LOCK key!
The patient may only be transferred from the side of the operating table. It is not allowed to transfer the patient from the head or foot end onto the table.
Danger of tipping over!
Danger of injury! If the patient is not secured during transport, adjustments of the operating table or
positioning (especially in Trendelenburg positions), he/she may slip off the operating table. Always secure and observe the patient.
1133.02BX
Danger of burns to the patient!
When using HF surgical equipment, defibrillators and defibrillation monitors, make sure to avoid that the patient comes into contact with the metal parts of the operating table, the table tops or the accessories. Also make sure to avoid that the patient is lying on moist underlays or sheets on the conductive pads.
Be absolutely sure to comply with the manufacturer’s operating instructions.
Danger of damage! Take care to avoid collisions between accessories and the operating table during adjustment procedures.
Consult MAQUET prior to using accessories produced by other manufacturers if they affect the longitudinal geometry of the table top.
There is no explosion protection while recharging the batteries or the operating table is on mains power.
When connecting and removing the mains cable, observe the correct order of the procedure! (see page 22)
The patient has to be positioned and observed so as to avoid hazards to respiration, the nervous system and the circulatory system.
The operating table’s electrical conductivity has to be checked once a year by trained service personnel!
Unplug the operating table before opening!
Electrical devices (such as cellphones, radios, magnetic resonance tomographs) can interfere with the functioning of the operating table when used near the operating table. Please observe the Technical Description for the operating table. It contains information on electromagnetic compatibility (EMC) for the operating table (both radiation and resistance to interference). Adhere to those specifications when using electrical equipment and respond properly in the event of effects on the operating table.
4
GA113302GB11
1133.02BX
I. Important safety instructions
Danger of injury!
Before aligning the leg plates horizontally, check the patient’s position carefully, particularly when using knee crutches.
Danger of crushing!
Do not touch the side rails in the area of the joints when adjusting the operating table top!
Danger of pinching!
Danger of collision!
Danger of pinching!
Do not reach under the table top or between the table top components during adjustment procedures.
Danger of crushing!
When locking the operating table (LOCK), the operator’s feet may be crushed or shorn. When setting the operating table, keep enough distance from the table base.
Danger of damage!
Do not place any objects on the base of the operating table!
Danger of damage!
Do not press or pull on the covers of the joints.
Danger of damage!
Make sure the cable of the hand control does not get jammed when adjusting the operating table.
Î
GA113302GB11
Î
5
II. Important information
1133.02BX
II. Important information
Your hospital has opted for the 1133.02 operating table. Thank you for the confidence you have demonstrated in doing so.
1. Important notes on these operating instructions:
Please read these operating instructions thoroughly and carefully. They were written to make you familiar with the features of the operating table.
Please observe these instructions at all times.
Please keep these instructions close to the equipment for ease of reference.
In these operating instructions we have used the following symbols:
This symbol indicates safety instructions in accordance with the prevailing international standard for medical equipment.
ÎÎ
Î Note
ÎÎ
This symbol is placed before all instructions that ensure the trouble-free operation of the unit.
(X) An „X“ in the order number (e.g. 1133.53XX) stands in place of the different versions.
2. Using the operating table in accordance with the intended purpose
The operating tables distributed by MAQUET are to be used exclusively in the field of human medicine. The user must have been instructed in the proper operation of the table before using it. This familiarization session is confirmed in an acceptance log. The location selected for the table must satisfy the requirements set down in applicable standards and guidelines. Mandatory for utilization is the proper assembly and maintenance of the unit.
3. Intended purpose
The 1133.02 operating table serves to support and position patients for surgical treatment
immediately before, during and after the operative phase
in rooms designated for medical use and which provide the following required features: – equipment to protect against electric shocks – additional potential equalization – electrically conductive floor (corresponding to Group 1 or 2 as per VDE 0100 part 710);
under medical supervision.
Considering patient positioning a max. 450 kg (1000 lb.) patient weight is permissible.
Avoid excentric patient positions, i.e. the patient’s centre of gravity (near the navel) must be as close as possible to the centre of the column.
Due to the design of the table top, the operating table is suitable for all surgical disciplines. The patient positioning options are listed in these Operating instructions. The operating table adjustments are made electrohydraulically using a cable-connected hand control, an IR re­mote control, an IR wall control panel or a foot switch. Side rails serve to attach accessory components in accordance with the manufacturer’s specifications. The radiotranslucent operating table top allows for intraoperative use of X-ray equipment.
ÎÎ
Î Note
ÎÎ
Make sure to comply with the generally applicable provisions relating to hygienics, and to cover the table top of the operating table with sheets before positioning the patient.
4. Basic safety aspects
„CLASSIFIED BY UNDERWRITERS LABORATORIES INC: WITH RESPECT TO ELECTRIC SHOCK; FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 2601-1“
This operating table meets the requirements set forth in the IEC 601-1 international standard, which is equivalent to EN 60601-1. The operating table is classified as Class I equipment of the Medical Device Directive. Whenever maintenance work is required, remember that the MAQUET service is the right address in terms of quality and maintaining guarantee protection. You will certainly understand that MAQUET can be held responsible for the safety technology features of the product only if assembly, enhancements, readjustment, modifications and repairs are made exclusively by our service department or by an authorized agent and if the unit is used in full compliance with the Operating Instructions.
6
GA113302GB11
1133.02BX
III. General description
III. General description
1. General features
In principle, the ALPHAMAXX 1133.02 operating table is suitable for the following surgical procedures:
in the thorax and abdominal areas in the head area in gynaecology in urology in orthopaedics
If required, separate accessories can be attached to the operating table.
Fig. 1: Permissible patient orientation
The table top is subdivided into 6 sections:
Head rest (removable) Upper back plate (removable) Lower back plate Seat plate Split leg plates
The lower back, seat and leg plates are adjusted under motor power. The head rest and the upper back plate are adjusted manually.
X-ray cassettes can be inserted underneath the back plate.
The side rails are used to fix the accessories.
For patients weighing up to 135 kg (300 lb), place the operating table only on its castors (UNLOCK) up to the maximum table top height of 940 mm (37 in)! The warning sign is not seen!
For patients weighing from 135 kg up to 225 kg (300 - 500 lb), place the operating table only on its castors (UNLOCK) when the table top is in its lowest position!
GA113302GB11
NO
YES
Using the UNLOCK function for patient weights exceeding 225 kg (500 lb) is not permissible!
2. Versions
Various versions of the 1133.02 operating table are avail­able:
1133.02B1 Operating table incl. SFC pads and stand­ard side rails
1133.02B3 Operating table incl. SFC pads, standard side rails and motor drive for the base
1133.02F1 Operating table incl. SFC pads and US side rails
1133.02F3 Operating table incl. SFC pads and US side rails and motor drive for the base
The cable-connected hand control, the IR transmitter and charging station for the IR transmitter as well as the head rests, upper back plates and leg plates are to be ordered separately. The order numbers are listed in Chapters VIII and IX.
7
III. General description
2. Control of functions
The electrohydraulic functions of the operating table can be operated using the following control elements:
Cable-connected hand control (see 2.1)
Override system = emergency operation (see 2.2)
or optionally:
IR remote control (see 2.3)
Wall control panel (see 2.4)
Foot switch 1009.79F0 (see 2.5)
The cable-connected hand control and the IR remote control have almost the same function keys. The override system is only used for emergency opera­tion of the operating table. The foot switch functions are: table height, Trendelen­burg/reverse Trendelenburg position and another func­tion (lateral tilt, longitudinal shift for the back plate or leg plate adjustment).
ÎÎ
Î Note:
ÎÎ
For safety reasons only the TRENDELENBURG/ REVERSE TRENDELENBURG functions can be used when the UNLOCK function is activated.
1133.02BX
The sockets 14 for the cable-connected hand control and the foot switch are located on the override control panel.
ÎÎ
Î Note:
ÎÎ
If the functions are controlled simultaneously by more than one operating element, each movement of the operating table is stopped immediately.
Danger of operating table tipping over!
Make sure to remove any potential hindrances before adjusting the operating table!
Danger of damage during adjustments!
Do not place any objects on the base of the op­erating table!
Danger of pinching!
Do not reach under the table top or between the table top components during adjustment proce­dures.
Danger of injury!
If the patient is not secured during transport, adjustments of the operating table or positioning (especially in Trendelenburg positions), he/she may slip off the operating table. Always secure and observe the patient.
Danger of damage!
When not in use, hang up the cable-connected hand control at the side rail of the operating table. Make sure the cable of the hand control does not get jammed when adjusting the operating table.
8
GA113302GB11
1133.02BX
v w
t
r
p
l
j
g
e
c
a
III. General description
2.1 Cable-connected hand control
The cable-connected hand control is connected to the operating table at one of the two sockets 14 at the over­ride control panel (for simultaneous connection of ca­ble-connected hand control and foot switch). The elec­tric motor-powered adjustments of the operating table can be made immediately by pressing completely the corresponding function keys.
a)
Lower
b)
Raise
c) Tilt to the d) Tilt to the e)
Trendelenburg position
f)
Reverse Trendelenburg position
g)
Flex
position
operates the
back plate
achieved to force the surgical field upward
Beach chair
raises the back plate, moves the entire table top into Trendelenburg position and lowers the leg plates. The patient is moved into a seated posi­tion.
The left leg plate can be controlled individually only together with the appropriate function key for the desired movement – downward (r+j) or upward (r+k), under two-button control
Similar to the individual plate, by pressing the two keys for downward movement (s+j) or for upward movement (s+k) (two-button control)
Lock
Press the LOCK key for at least 1 second; the table will be locked automatically.
Unlock
To unlock the operating table, press the UNLOCK key for at least 1 second, then the castors will be extended automatically and the table can be displaced.
Displace the operating table the head end).
foot end). When displacing the operating table under motor power, it must be guided with a hand.
m
u
s
h)
q
j) Slope (both) leg plates
k
h
f
d
b
z
k) Slope (both) leg plates
l) Slope (lower) back plate
m) Slope (lower) back plate
p) Shift the table top toward the q) Shift the table top toward the
r) Adjust the left leg plate individually
s) Adjust the right leg plate individually
t)
u)
v) Motor drive for the base
w) Displace the operating table
table top
table top
left
(view from head end)
right
(view from head end)
reverse Trendelenburg
functions until a flex position is
downward upward
downward upward
foot end head end
for 1133.02B2 (optional)
backward
forward
and
lower
(toward
(toward the
GA113302GB11
9
III. General description
z) 0-position
Resets the lateral tilt, Trendelenburg/reverse Tren­delenburg, back plate and leg plate adjustments to the horizontal position. If the function key z is pressed, the operating table top will first reset the adjustments in the following order – lateral tilt, Tren­delenburg/reverse Trendelenburg, back plate. An acoustic signal will indicate that these adjustments have been reset. Release the function key z and press once again to reset the back plates to the horizontal position; if the leg plates have been ad­justed individually, they will be positioned parallel to one another.
ÎÎ
Î Notes:
ÎÎ
Manual adjustment functions such as those for the upper back plate, head rest and leg plate abduction will not be affected by the IR remote control (and as a consequence neither by the automatic 0-position function).
ÎÎ
Î The transmission power of the IR hand control is
ÎÎ
sufficient to be able to adjust the operating table even from a greater distance. If the operating table does not respond to the command even though the function key is completely depressed, either reaim the hand control or move to a slightly different trans­mission position. In such cases the receiver at the column is probably being blocked by a person’s body.
1133.02BX
2.2. Override system = emergency operation
2.2.1 Override control panel
In case of a malfunction or a defective hand control, you can control the operating table functions via the control panel 13 on the upper column casing = emergency op- eration.
ÎÎ
Î Note:
ÎÎ
Always press the ON key together with the desired
13
function key.
ÎÎ
Î Note:
ÎÎ
If the operating table batteries are discharged, the override system only works if the operating table is connected to the mains.
ÎÎ
Î Note:
ÎÎ
When using the override control panel, the anticollision feature is not activated. Avoid extreme adjustments – the combination of maximum Tren­delenburg/reverse Trendelenburg position and maxi­mum lateral tilt.
10
GA113302GB11
1133.02BX
III. General description
The functions of the override control panel are shown in the following:
A) LED status indicators of operating table
ÎÎ
Î Note:
ÎÎ
The status indicators are only active when the ca­ble-connected hand control or IR remote control is switched on.
Description of the symbols from left to right: z Mains connection: the green LED lights up when
the operating table is connected to the mains.
z Battery charge level: the red LED flashes when the
operating table batteries have to be recharged as soon as possible.
z Lock operating table: the green LED lights up when
the operating table is locked, i.e.
z Extend operating table castors: the red LED lights
up when the operating table is on its castors and can be displaced, e.g.
UNLOCK
LOCK
activated.
activated.
AC
EGILN
DFHK
B
MO
B) ON
switches the motor ON and must be pressed in addition to each function key
C) TREND
Moves table top into Trendelenburg position
D) REV.
Moves table top into reverse Trendelenburg position
E) UP Move table top up
F) DOWN
Move table top down
G) TILT
Lateral tilt to the left
GA113302GB11
H) TILT
Lateral tilt to the right
I) UP
Slope back plate upward
K) DOWN
Slope back plate downward
L) UP
Slope leg plate upward
11
III. General description
v w
t
r
p
l
j
m
1133.02BX
M) DOWN
Slope leg plate downward
N) LOCK
Operating table locked
O) UNLOCK
Operating table on castors, movable
2.3. IR remote control (optional)
u
s
q
k
The operating table can be equipped with an IR remote control, as desired. The required charging station for the IR remote control is not included with the order no. 1133.91A0. The layout of the function keys for the operating table adjustments is the same on the cable-connected hand control and IR remote control apart from the following exceptions: There are no function keys for the motor drive for the base. For safety reasons, you can only activate this function via the cable-connected hand control.
g
e
c
a
h
f
d
b
z
ÎÎ
Î Note:
ÎÎ
When switching on the IR remote control, the LED status indicators on the control panel on the oper­ating table column are active (see chapter on “Over­ride system”).
ÎÎ
Î Note:
ÎÎ
When using the IR remote control or the cable-con­nected hand control, the electronics generally avoid collisions with the floor or parts of the operating table.
12
GA113302GB11
1133.02BX
III. General description
2.3.1 IR system code
An IR remote control is clearly assigned to each oper­ating table through a code. The operating table has the same system code as the associated IR remote con­trol. You can see from the label at the operating table column whether the IR reception has been activated for this operating table. If the system code is crossed out, then the IR reception has been locked for this operating table. The operating table and the IR remote control are activated / locked for IR reception by authorized per­sonnel.
The operating table and the associated IR remote con­trol are marked by adhesive labels bearing the set sys­tem code:
The adhesive labels are affixed to the operating ta­ble column next to the IR receiver.
The adhesive labels are affixed to the back of the IR remote control.
The two dot-shaped adhesive labels bear a number from 0 to 9 or a letter from A to F:
left: 0, 1, 2 ...A, B, ... F
right: 0, 1, 2 ...A, B, ... F
03
Examples of preset IR system code markings:
15 3A C2 DA
ÎÎ
Î Note:
ÎÎ
The order of the identification letters/numbers is sig­nificant, i.e. the IR system code 4 B does not equal B 4.
ÎÎ
Î Note:
ÎÎ
Always leave the IR remote control in the immedi­ate vicinity of the operating table to which it is as­signed, since the operating table can only be moved via this IR remote control. Of course, the operating table can be adjusted via a cable-connected hand control or the override sys­tem at any time, regardless of the IR remote con­trol.
ÎÎ
Î Note:
ÎÎ
The transmission power of the IR hand control is sufficient to be able to adjust the operating table even from a greater distance. If the operating table does not respond to the command even though the function key is completely depressed, either reaim the hand control or move to a slightly different trans­mission position. In such cases the IR sensor at the column is being blocked by a person’s body.
GA113302GB11
13
III. General description
2.3.2 Charging station
Recharging the IR remote control:
Connect the charging station to the mains.
The green LED in the charging station indicates whether the required supply voltage is available or not.
Place the IR remote control in the charging station as follows: – Keypad visible, in the position for normal use,
i.e. the 0-position key x is located at the bottom right
– The base of the hand control is in contact with
the lower limit stop of the holder.
ÎÎ
Î Note:
ÎÎ
The rechargeable battery of the IR remote control has enough capacity to ensure several day’s use of the operating table. Nevertheless, we recommend that you insert the hand control into the holder of the charging station so that the battery is recharged overnight.
1133.02BX
2.4 Control via IR wall panel
The desired adjustment functions for the operating ta­ble can be controlled by means of the function keys (1) at the wall control panel or the separate IR hand control (2).
ÎÎ
Î Note:
ÎÎ
The IR wall control panel can be mounted to the wall at a later date. For further information, please refer to the operating instructions GA115095GBxx.
14
GA113302GB11
1133.02BX
2.5 Control via foot switch 1009.81DX (optional)
14
14
Connect the foot switch to one of the two sockets (14) at the column head.
Three functions can be activated using the following foot switch versions:
1009.81D0 Foot switch for height, Trendelenburg/rev.,
back plate
1009.81D1 Foot switch for height, Trendelenburg/rev.,
lateral tilt
1009.81D2 Foot switch for height, Trendelenburg/rev.,
leg plates
1009.81D3 Foot switch for height, Trendelenburg/rev.,
longit. shift
ÎÎ
Î Note
ÎÎ
Be sure to observe the Operating instructions for the foot switch.
2.6 Special functions
If requested, the MAQUET service personnel can set the NEUROLOCK function at the operating table. This function avoids adjustments of the table position due unintentional actuation of the keys at the hand control.
a+b) The ”Neurolock” function can be unlocked by
pressing simultaneously the ”table top down” key a and the ”table top up” key b on the control module. The operating table can now be adjusted. If no further keys on the control module are actu­ated for approx. 8 sec., then the operating table is locked again. Adjusting the operating table again is only possi­ble after actuating the ”table top down” a and ”ta­ble top up” b function keys simultaneously.
ÎÎ
Î Note:
ÎÎ
The ”table top up” and ”table top down” functions cannot be actuated simultaneously using the foot switch 1009.81X0. Use a second control module to unlock the Neurolock function.
GA113302GB11
15
III. General description
3. Table top
3.1 General features
The 1133.02 has been designed to meet the require­ments of various surgical disciplines. Selecting from a variety of modules lets you adapt the geometry of the table top to accommodate individual patient height. The table top is divided into a maximum of 6 sections, depending on the type of leg plate and upper back plate module selected:
head rest
upper back plate (extension plate)
lower back plate
seat plate
leg plate(s)
The table top is equipped with a motor-powered longitu­dinal shift feature through 215 mm (9 inches) and mo­tor-powered adjustment functions for the back plate and leg plates. Furthermore, the mounting sockets for the leg plates automatically detect whether one-piece com­ponents have been installed, thus increasing reliability and safety in use. Depending on the application, the adjustment ranges for the accessories mounted at the leg plate connec­tion point may be restricted. Please refer to the Operat­ing instructions for the accessory in question. Mounting various back plate modules at the upper/lower back plate connection point allow for positions match­ing the patient perfectly and create a favourable working environment for the surgical team and for the interven­tion. Guide rails used to insert X-ray film cassettes are lo­cated below the back plate. The side rails are intended for the attachment of acces­sories. Various head positioning accessories, used in­stead of a head rest, can be attached at the head end of the back plate modules (see Section V. Patient posi­tioning).
1133.02BX
16
ÎÎ
Î Note:
ÎÎ
According to the internationally valid standard IEC 60601-1, the operating table is designed for safe handling of a patient weight of 450 kg (1000 lb.). When using accessories, the maximum load may be restricted. Please observe the Section VIII. List of accessories.
GA113302GB11
1133.02BX
3
x
fig. 1
III. General description
3.3 Attaching the back plate modules
1) First remove the pad
2) Loosen the eccentric safety levers at the right
and left (3); press the safety push buttons x (fig.1) to do so. If a back plate is already installed, hold it with one hand.
3
fig. 2
3) Insert the back plate into the bars from above,
holding it almost vertically and set it to the re­quested position (fig. 2)
4) When making the adjustment, hold the back plate
with one hand and re-tighten the eccentric levers (3) (fig. 3).
5) Re-place the pad in the correct position.
3
x
3
fig. 3
GA113302GB11
3.4 Removing the back plate modules
To remove the back plate modules, refer to section
3.3 ”Attaching the back plate modules” and pro­ceed in reverse order.
17
fig.5
10
III. General description
3.5 Leg plates
1133.53
To mount the split leg plates, press the push button
10 and insert the leg plate pins into the mounting holes up to the stop (Fig. 5).
They are locked in position automatically as soon
as the button 10 is released (Fig. 5a).
The slope of the back plate is adjusted electro-hy-
draulically by means of the hand control.
9
To abduct the leg plates, open the eccentric lever 9
and set to the desired abduction angle.
Re-tighten the eccentric lever after making the ad-
justment.
A dual-pivot joint is used for abduction of the 1133.53
leg plates so that on the one hand the surgeon has access to the patient’s body from the foot end while at the same time the patient’s legs are given suffi­cient support.
To remove the leg plate, press the push button 10
and pull the leg plate out of the mounting holes.
1133.02BX
fig.5a
10
Take care that the leg plates are not crossed / on top of each other before aligning the leg plates for the horizontal position.
Danger of collision! Danger of injury!
18
GA113302GB11
1133.02BX
III. General description
1133.58BC Single-section leg plate
To attach or remove the single-section leg plate,
actuate the release lever a and insert the leg plate in the bore holes up to the stop or remove it.
The single-section leg plate is suitable for surgical procedures in orthopaedics and vascular surgery where the patient’s legs are to be X-rayed. It is sloped electrohydraulically by means of the control mod­ule.
a
1133.55BC Seat plate extension
To attach or remove the seat plate extension, actu­ate the release lever a and insert the seat plate ex­tension in the bore holes up to the stop or remove it.
The seat plate extension is suitable for surgical pro­cedures in gynaecology and urology.
3.6 Table top padding
The table top will be supplied with table top plates and SFC padding secured with Velcro straps.
3.7 Accessories
Mount only the accessories listed in Chapter VIII.
Use accessory parts from other manufacturers, which can affect the table top geometry, only after consulta­tion with MAQUET!
Accessories which are mounted on the side rails of the operating table provided for this purpose should also been drawn exclusively from the MAQUET line of ac­cessories. MAQUET attaches great importance to care­ful harmonization with the operating tables in our range. This results in maximum safety for the patient and the surgical staff when used in accordance with the instruc­tions. Accessories (to be mounted on the side rails) which are made by other manufacturers must be particularly carefully examined to ensure that they do not represent a hazard to the patient and staff or to the operating ta­ble (due to possible collisions). Particularly accessory units with high weights or long lever arms (such as hand operating tables, retractor systems, etc.) must be ex­pressly permitted by MAQUET.
The downward movement for the seat plate exten­sion is locked for safety reasons.
a
Danger of injury!
Do not use worn out or damaged accessories!
Danger of injury!
Make sure to fasten the table top accessories properly!
GA113302GB11
Danger of damage!
While making adjustments, avoid collisions between the operating table and its accessories!
19
III. General description
4. Motor drive for the base
The table top versions 1133.02B2/F3 are equipped with a motor drive for the base. A table equipped with a mo­tor drive for the base starts moving slowly and slows down softly after releasing the function key (i or k).
i) BACKW.
k) FORW.
To move the operating table, proceed as follows:
Remove the mains cable from the operating table
Extend the operating table castors fully (press UNLOCK key)
Guide the operating table with at least one hand
Press the function key (v or w) and the operating table will move in the selected direction.
Release the function key (v or w) and the operating table will gently brake until it stops.
1133.02BX
Move table backwards in the longitudinal direction
Move table forwards in the longitudinal direction
ÎÎ
Î Note
ÎÎ
If the LEDs of the LOCK and UNLOCK functions are flashing, press the UNLOCK key.
Guide the operating table with at least one hand during displacement.
ÎÎ
Î Note
ÎÎ
When moving the operating table, do not press the override system keys, since this will cause several functions to be activated simultaneously and involves the risk of injuries or collisions with potential hin­drances.
ÎÎ
Î Note
ÎÎ
The operating table can only be moved laterally if you have not activated the BACKW. or FORW. func­tion.
20
GA113302GB11
1133.02BX
IV. Use of the operating table
IV. Use of the operating table
1. General
The operating table’s electrical conductivity has to be checked once a year by trained servicing personnel!
ÎÎ
Î Note :
ÎÎ
In countries where potential equalization is required, the operating table has to be connected to the po­tential equalization system of the operating room. For this purpose, connect the potential equaliza­tion pin of the operating table to one connector of the potential equalization pin of the operating room.
In countries where potential equalization is not used, it is not required to connect the operating table to the potential equalization system.
If no potential equalization cable (5 m long) with connectors at both ends is included in the scope of supply of your operating table and you are required to use PE cable during surgery please order p/n 0200 403 4 at your local MAQUET representative.
2. Operation on battery power
When operating the table with the integral batteries with no connection to the mains supply, the operating table satisfies the requirements of Class M for the anaesthet­ics test. It may be used in areas subject to explosion hazard, Zone M. Battery operation of the operating table is always pref­erable to operation on mains power.
The capacity of the fully charged batteries allow approx. 1 week’s use of the operating table, depending on how often adjustments are made.
If the charge level of the batteries drops below a mini­mum permissible value, this is indicated by the red operating table status indicator LED when the operat­ing table is switched on.
ÎÎ
Î Note
ÎÎ
The operating table switches off automatically when the battery charge level drops below 10%.
GA113302GB11
Recharging the batteries
Discharged batteries will again be 90% recharged after approx. 5 hours of charging and 100% recharged after approx. 10 hours.
The batteries are automatically recharged as soon as the operating table is connected to the mains by means of the mains cable.
21
IV. Use of the operating table
ÎÎ
Î Note
ÎÎ
We recommend to recharge the batteries overnight, but at least every 2 days, even if “low charge level” of the rechargeable batteries has not been indicated yet. Regular recharging of the batteries will not only en­sure sufficient battery power available for the opera­tion programme, it will also serve to prolong the use­ful life of the rechargeable batteries.
1133.02BX
Socket for mains cable
Mains cable
Potential equalization
3. Operation on mains power
As soon as the operating table is connected to the mains via the mains cable, the batteries are recharged auto­matically.
There is no explosion protection while recharg­ing the batteries or the operating table is on mains power.
The separation of the operating table from the mains is made by the mains plug.
Connect the operating table to the mains via the mains cable.
When connecting and removing the mains ca­ble, observe the correct order of the procedure:
When connecting the charging cable, first of
all connect the operating table and then plug the cable into the socket.
When disconnecting the charging cable, first
of all pull the plug out of the socket and then remove the cable from the operating table.
ÎÎ
Î Note
ÎÎ
Damaged cables must be replaced immediately.
When replacing the fuses please note:
1. Unplug the operating table
2. Unscrew the two fuse holders located on the
right and left next to the socket for the main plug with a coin or a large screw driver
3. Insert the appropriate fuse according to the
opposite drawing
4. Reinsert the fuse holders
4. HF surgery / Use of defibrillators
In the field of HF surgery and when using defibrillators, be absolutely sure to observe the following safety in­structions:
Danger of burns!
When using HF surgical equipment, defibrillators and defibrillation monitors, make sure to avoid that the patient comes into contact with the metal parts of the operating table, the table tops or the accessories. Also make sure to avoid that the patient is lying on moist underlays or sheets on the conductive pads. Be absolutely sure to comply with the manufac­turer’s operating instructions.
22
GA113302GB11
1133.02BX
V. Patient positioning
V. Patient positioning
1. General
The 1133.02 modular universal operating table offers the possibility to use various modules as the upper back plate, thus providing the following advantages:
The operating table top which can be used for various purposes, can be adapted to meet the special requirements of diverse surgical disciplines (modules see Chapter VIII).
In accordance with generally accepted hygiene standards, the operating table is to be covered with surgical drapes during use.
Position the patient always in the patient orientation, i.e. the patient’s head is lying above the long end of the operating table base.
Observe the defined patient orientation. Never position the patient’s upper body on the leg plate.
Danger of injury!
Risk of injury due to the operating table tipping over or material breaks when using accessories which are not approved. The patient positions vary, depending on the patient weight (see sketches in Chapter V Pati­ent positioning) and the accessories which may be employed (see Accessories list, Chapter VIII) for the operating table. Observe the specifications for positioning the patient according to his/her weight.
NORMAL
GA113302GB11
Danger of injury!
If the patient is not secured during transport, adjustments of the operating table or positioning (especially in Trendelenburg positions), he/she may slip off the operating table. Always secure and observe the patient.
23
V. Patient positioning
Danger of injury!
If the table top width is not sufficient for an ergonomic position, the patient may be injured. The table width extenders (1001.75A0/ 76A0) can be mounted on the side rails to ensure an ergonomic patient position. Only one infusion stand (1009.01C0), one anaesthesia screen (1002.57A0), one wristlet (1002.24C0) and one arm protector (1002.25A0) may be mounted to the table width extenders (1001.75A0/76A0). When positioning the patient, be sure to observe that the patient is lying above the centre of the column in the lateral direction (see sketch).
Danger of injury!
It is not permissible to mount another width ex­tender to a table width extender (1001.75A0 / 76A0).
1133.02BX
Danger of injury!
The nursing staff must prevent risks to breath­ing, nerves and circulation by careful positioning and observation of the patient.
Danger of injury!
Watch the patient permanently while adjusting the knee crutches.
Danger of operating table tipping over!
Lock the operating table before transferring the patient! Press the LOCK key!
ÎÎ
Î Note:
ÎÎ
After standstills, actuate the table functions to check proper functioning before positioning the patient. Press the function keys table top up/down, Trende­lenburg/reverse Trendelenburg, lateral tilt left/right, back plate up/down, leg plates up/down, for 2 to 3 seconds each. Finally, press the Level key (z) to align the operating table horizontally.
The functions of the cable-connected hand con­trol are described on page 9.
24
ÎÎ
Î Note:
ÎÎ
Depending on the patient weight, the speed of the motor-powered adjustment functions will change. The speed of the downward movements will slightly increase. The adjustment speed will be reduced during up­ward movements.
GA113302GB11
1133.02BX
V. Patient positioning
2. Positioning a patient weighing up to 135 kg (300 lb)
2.1 Approved accessories
Danger of injury!
The mounted accessories may not exceed the maximum permissible length specified in the sketch. For further accessories, please refer to the accessories list, see Chapter VIII.
2.2 Operating table unlocked (UNLOCK)
Danger of tipping the operating table over!
If the patient is in the wrong position, the operating table may tip over when the castors are deployed (UNLOCK). Position the patient like Fig. V-1 shows and set the operating table to a height of less than 940 mm before deploying the operating table castors (UNLOCK). The position below 940 mm is reached when the warning sign on the column is no longer seen.
NO
YES
GA113302GB11
25
2.2 Operating table unlocked (UNLOCK)
V. Patient positioning
1133.02BX
Fig. V-1 Maximum patient weight 135 kg (300 lb), UNLOCK
26
GA113302GB11
1133.02BX
2.3 Operating table locked LOCK
V. Patient positioning
Fig. V-2 Maximum patient weight 135 kg (300 lb), LOCK
GA113302GB11
27
V. Patient positioning
3. Positioning a patient weighing 135 kg up to 225 kg (300 lb - 500 lb)
3.1 Approved accessories
Danger of injury!
The mounted accessories may not exceed the maximum permissible length specified in the sketch. For further accessories, please refer to the accessories list, see Chapter VIII.
To extend the table top at the head end by 600 mm (23.6 in) , it is recommended to use the upper back segment 1133.31xx in conjunction with the head rest 1130.67xx.
Accessories mounted on the side rails
1001.44D0 Arm posturing device 450 mm long
1001.75A0 Table width extenders 100 x 300 mm
1001.76A0 Table width extenders 100 x 200 mm
1002.25A0 Arm protector
1001.85B0 Foot plate
1001.86A0/B0 Divided foot plate
3.2. Operating table unlocked (UNLOCK)
1133.02BX
Danger of tipping the operating table over!
If the patient is in the wrong position, the operating table may tip over when the castors are deployed (UNLOCK). Position the patient like Fig. V-3 shows and set the operating table to its lowest position before deploying the operating table castors (UNLOCK).
28
GA113302GB11
1133.02BX
3.2. Operating table unlocked (UNLOCK)
V. Patient positioning
Fig. V-3 Maximum patient weight 225 kg (500 lb), UNLOCK
GA113302GB11
29
3.3 Operating table locked (LOCK)
V. Patient positioning
1133.02BX
Fig. V-4 Maximum patient weight 225 kg (500 lb), LOCK
30
GA113302GB11
1133.02BX
4. Positioning a patient weighing 225 kg up to 450 kg (500 lb - 1000 lb)
4.1 Approved accessories
Danger of injury!
The mounted accessories may not exceed the maximum permissible length specified in the sketch. For further accessories, please refer to the accessories list, see Chapter VIII.
To extend the table top at the head end by 380 mm (15 in) , it is recommended to use the head plate adapter
1133.36A0 in conjunction with the head plate 1130.64xx.
The approved leg plates are: 1133.53xx abductable leg plate, max. length 670mm as well as the 1133.58xx one-section leg plate.
Accessories mounted on the side rails
1001.44D0 Arm posturing device 450 mm long
1001.75A0 Table width extenders 100 x 300 mm
1001.76A0 Table width extenders 100 x 200 mm
1002.25A0 Arm protector
1001.85B0 Foot plate
1001.86A0/B0 Divided foot plate
4.2 Operating table locked (LOCK)
Danger of tipping the OR table over!
Risk of injury due to the operating table tipping over when deploying the operating table castors (UNLOCK). The operating must always be locked (LOCK) for patient weights exceeding 225 kg.
V. Patient positioning
Danger of injury!
Risk of injury due to the patient slipping off the table during unrestricted use of adjustment functions. Therefore, restrict the adjustment functions as follows for patient weights exceeding 225 kg:
Longitudinal shift is not permissible
maximum Trendelenburg/reverse Trendelenburg position ± 10°
maximum lateral tilt ± 5°
maximum slope of the back plate ± 10°
maximum slope of the leg plate ± 10° Although the operating table would be able to perform the adjustments, these restrictions should be ob­served in the interest of patient safety.
Fig. V-5 Maximum patient weight 450 kg (1000 lb), LOCK
GA113302GB11
31
VI. Care and maintenance
VI. Care and maintenance
Cover the operating table top with sterile sheets when it is not in use.
1. Cleaning
Use a mild alkaline general detergent (soap suds) con­taining tensides and phosphates as its active cleaning agents for cleaning the operating table and its accesso­ries. If the surfaces are extremely dirty, a concentrated de­tergent should be used. This will make it necessary to wipe down the surfaces using a cloth wet with clear water.
ÎÎ
Î Note
ÎÎ
Only use as much water as is absolutely neces­sary and immediately remove any excess water with a dry cloth.
ÎÎ
Î Note
ÎÎ
Do not use pressurized jets of water or detergent. High-pressure cleaning can result in the ingress of fluid through the gaps between joints and cause corrosion.
1133.02BX
2. Disinfection
Use a commercial aldehyde-based surface disinfectant in an aqueous solution for manual disinfection of the operating table top. Be sure to use agents which are included in the list published by the German DGHM*. Refer to this list for further information on the individual components of the disinfectant.
ÎÎ
Î Note
ÎÎ
The disinfectant must not contain the substances specified below:
Chlorine or compounds which could liberate chlo-
rine, since they will damage metal surfaces
• Alcohol or alcohol-containing compounds
Hand disinfectants contain alcohol!
They will damage plastic parts such as table top pads or control levers.
• They will form explosive mixtures.
ÎÎ
Î Note
ÎÎ
Never spray the disinfectant directly into the joints between the sections of the operating table.
* Deutsche Gesellschaft für Hygiene und Mikrobiologie DGHM
c/o Institut für Hygiene und Mikrobiologie Universität Würzburg Josef-Schneider-Str. 2 97080 Würzburg http://www.dghm.org
32
GA113302GB11
1133.02BX
VI. Care and maintenance
3. Maintenance The operating table will have to be serviced once
a year by the MAQUET factory service department or by service technicians authorized by MAQUET to do so.
Please adhere to prescribed maintenance intervals as otherwise guarantee coverage will lapse.
In so doing you also make a contribution to patient and staff safety and extend the useful life of your costly equip­ment.
We recommend concluding a maintenance contract with MAQUET.
Outside Germany please contact your local representa­tive in all matters relating to service.
4. Malfunctions
If the operating table is defective, we recommend pro­ceeding as follows in order to place the operating table on its castors and remove it from the operating room:
1. Use the UNLOCK function on the Override panel: press the ON key and the UNLOCK key simultaneously until the table is standing on its castors. If required:
2. Use the mains cable to connect the operating table to the mains supply and repeat step 1. If required:
2
3. Use the ratchet:
Remove the sealing cap at the base using a wide
bladed screw driver or a coin.
Insert the ratchet (ref. no. 3113.3649, included in
the delivery) into the opening.
Turn the ratchet (2) clockwise until the castors
are extended (visual check). When the castors are fully extended, it will be very difficult or im­possible to turn the ratchet (2).
Screw in the sealing cap without tightening.
Move the defective operating table to the desired
location and park it on its castors (secure it, if required)
Inform the service department.
Attention:
If the castors have been extended using the hand crank, the UNLOCK key must not be actuated, oth­erwise the base will be lowered and cannot be placed on its castors again.
GA113302GB11
If the operating table malfunctions, please contact the MAQUET agency. If you can describe the symptoms and have the system’s serial number ready, they will be able to react and solve the problem more quickly.
ÎÎ
Î Note
ÎÎ
In case of malfunctions, do not attempt to repair the unit yourself and do not try to force the unit.
33
VI. Care and maintenance
5.3. Protection of the environment
Packaging materials:
The materials used for packaging MAQUET products are environmentally friendly. They are made of untreated wood, cardboard, recyclable plastics or other reusable materials. If you wish, MAQUET can also dispose of them.
MAQUET products:
MAQUET will take back any used MAQUET products, or component parts thereof, which are no longer usable and reuse the components in a way which is not dam­aging to the environment. Please ring your local MAQUET service centre for more details.
Plastic components:
Large plastic components have a symbol to indicate what type of plastic they are made of in order to make recycling easier. Padding can be disposed of through the hospital waste system.
(Rechargeable) Batteries:
can be disposed of through the local waste system. You can also ask your MAQUET agency for informa­tion.
1133.02BX
34
GA113302GB11
1133.02BX
VII. Technical data
1. Operating table adjustments
Length without head rest: 1900 mm (74.8 inches) Length with head rest: 2196 mm (86.5 inches) Width: 530 mm (20.9 inches) Width across side rails: max. 580 mm (22.8 inches) Height (without pads): 594 - 1056 mm (23.4 – 41.6 inches) Weight: approx. 328 kg (723 lb.) Trendelenburg position: max. 30° Reverse Trendelenburg position: max. 30° Lateral tilt left / right: max. 20° Lower back plate up: max. 80° Lower back plate down: max. 40° Upper back plate up: 90° (manual adjustment) Upper back plate down : 45° (manual adjustment) Leg plate up: 10° Leg plate down: 90° Longitudinal shift: 230 mm (9 inches)
Flex position: max. 40°
Reverse Trend. position: 20° Back plate down 20°
VII. Technical data
Beach chair:
Max. patient weight: 450 kg (1000 lb)
2. Electrical data
Nominal voltage: AC100/110-115/127
Frequency: 50/60 Hz Power consumption: max. 400 VA Class: II (Double Insulated) Applied part
Duty cycle: int. 10 min. ON / 20 min. OFF
Trendelenburg position: 25° Back plate up 80° Leg plate down 70°
200/220/230-240V
Type B The enclosure leakage current meets the requirements of the patient leakage current for CF conditions ac­cording to IEC 60601-1 IP X4* IPS ** SELV*** DC 24V
For patients weights from 225 kg up to 450 kg (500-1000 lb): 2 min. ON / 120 min. OFF
Explosion protection: AP for operation on battery power
Rechargeable batteries: Type A512C / 15SR
* IPX 4 = protection against ingress of liquids ** IPS = internal power source *** SELV = safety extra low voltage
GA113302GB11
When connected to the mains, the operating table is not approved for areas subject to explosion hazards (AP-M)
2 rechargeable batteries 12 V / 15 Ah
35
VIII. List of accessories
VIII. List of accessories
1133.02BX
Fig.: VIII-1 Accessories for patients weighing up to 135 kg (300 lb). For further accessories, please refer to the accessories list, see Chapter VIII.
36
GA113302GB11
1133.02BX
VIII. List of accessories
Fig.: VIII-2 Accessories for patients weighing up to 225 kg (500 lb). For further accessories, please refer to the accessories list, see Chapter VIII.
GA113302GB11
37
VIII. List of accessories
1133.02BX
Fig.: VIII-3 Accessories for patients weighing up to 450 kg (1000 lb) For further accessories, please refer to the accessories list, see Chapter VIII.
1. Accessories for the head rest interface
1.1 for patients weighing up to 135 kg (300lb)
1150.71X0 Head rest approved only in conjunction with the 1150.37X0 carbon-fibre back plate for neurosurgery
1130.54B0 Connecting bracket
1005.2900 Basic unit
1005.25A0 Connecting piece +
1002.71A0 Horseshoe-shaped head rest
1002.73A0 Head rest, flat
1002.7000 Horseshoe-shaped head rest, split +
1004.7500 Wrist support
1002.72A0 Horseshoe-shaped head rest, split +
1004.7500 Wrist support
1005.3400 Clamp adaptor for skull clamp +
1005.3500 Skull clamp
In addition, the accessories listed under sections 1.2 and 1.3 are approved.
38
GA113302GB11
1133.02BX
1.2 for patients weighing up to 225 kg (500lb)
1130.52X0 Head rest, simple version
1130.53X0 Head rest with double articulation
1130.54B0 Connecting bracket
1002.07A0 Head plate adapter
1002.71A0 Horseshoe-shaped head rest
1002.73A0 Head rest, flat
1002.6200 Head rest, flat
1002.7000 Horseshoe-shaped head rest, split+
1002.72A0 Horseshoe-shaped head rest, split +
1003.4700 Head support plate
1002.65A0 Connection fixture
1002.71A0 Horseshoe-shaped head rest
1002.73A0 Head rest, flat
1002.6200 Head rest, flat
1002.7000 Horseshoe-shaped head rest, split +
1002.72A0 Horseshoe-shaped head rest, split +
1003.4700 Head support plate +
1000.5700 Head ring, PUR or
1005.3600 Special clamp for skull clamp +
1005.5800 Supporting arm
1002.5800 Adaptor +
1130.64X0 Head rest
1130.67X0 Head rest, folding with gas strut, washable
1130.68B0 Head rest long, folding with gas strut, SFC pad, with side rails, Europe
1130.69X0 Head rest with double articulation
1002.74A0 Motorized head plate adjusting unit
1002.71A0 Horseshoe-shaped head rest
1002.73A0 Head rest, flat
1002.6200 Head rest, flat
1002.7000 Horseshoe-shaped head rest, split +
1004.7500 Wrist support
1002.72A0 Horseshoe-shaped head rest, split +
1004.7500 Wrist support
1003.4700 Head support plate
VIII. List of accessories
1004.7500 Wrist support
1004.7500 Wrist support
1004.7500 Wrist support
1004.7500 Wrist support
1000.5900 Gel head ring
1005.3500 Skull clamp
1002.71A0 Horseshoe-shaped head rest
1002.73A0 Head rest, flat
1002.7000 Horseshoe-shaped head rest, split +
1004.7500 Wrist support
1002.72A0 Horseshoe-shaped head rest, split +
1004.7500 Wrist support
1.3 for patients weighing up to 450 kg (1000lb)
1130.67X0 Head rest, folding with gas strut, washable
GA113302GB11
39
VIII. List of accessories
2. Accessories for back plate interface
2.1 for patients weighing up to 135 kg (300lb)
1150.37A0 Carbon-fibre back plate
In addition, the accessories listed under section 2.2 are approved.
2.2 for patients weighing up to 225 kg (500lb)
1133.31B0 Back plate for general surgery, SFC padding, Europe
1133.32B0 Short back plate for general surgery, SFC padding, Europe
1133.33B0 Back plate for neurosurgery, SFC padding, Europe
1150.34F0 Back plate for shoulder operations, SFC padding
1150.35F0 Back plate for ophthalmology, SFC padding
1150.36A0 Head plate adapter
2.3 for patients weighing up to 450 kg (1000lb)
1150.36A0 Head plate adapter
1133.02BX
3. Accessories for legplate interface
3.1 for patients weighing up to 135 kg (300lb)
1130.56D0 Rectal positioning device
1003.3300 Knee supports, for fastening on side rails with clamps
1130.63X0 Dual-joint leg plate with SFC padding
1130.65X0 Transfer board
1130.70A0 Maquetmatic leg plate with SFC padding
1132.65A0 Transfer board
1133.55BC Seat plate extension, coded, SFC padding
1133.53BC Leg plate, with abduction, coded, SFC padding
1133.58BC One-piece leg plate, coded, SFC padding
1130.65X0 Transfer board
1132.65A0 Transfer board
1419.01HC Extension device, coded
4831.001T Rectal positioning device
In addition, the accessories listed under section 3.2 are approved.
3.2 for patients weighing up to 225 kg (500lb)
1133.53BC Leg plate, with abduction, coded, SFC padding
1133.54BC Dual-joint leg plate, manually adjustable via 2 joints, coded, SFC padding
1133.58BC One-piece leg plate, coded, SFC padding
1133.67BC Carbon-fibre leg plate, abductable, coded, SFC padding
3.3 for patients weighing up to 450 kg (1000lb)
1133.53BC Leg plate, with abduction, coded, SFC padding
1133.58BC One-piece leg plate, coded, SFC padding
40
GA113302GB11
1133.02BX
4. Accessories to be attached to the side rails
1001.62A0 Hand operating table, Europa
9092.3214 Padding
1001.28C0 Hand support
1001.44C0 Arm posturing device 300 mm
1001.44D0 Arm posturing device 450 mm
1001.45C0 Upper arm posturing plate
1001.4700 Leg restraint cuff
1001.56A0 Body strap, washable
1001.6000 Arm board with ball-and-socket joint adjustment
1001.85B0 Foot plate
1001.92C0 Axilla support, washable
1002.09C0 Shoulder supports
1002.19C0 Fixture for body supports
1002.11A0 Back-buttocks support
1002.11B0 Pubis-sacrum-sternum support
1002.11C0 Lateral support
1002.11D0 Supporting roll
1002.21C0 Lateral support
1002.24C0 Wristlet
1003.22C0 Standard clamp
1002.57A0 Anaesthesia screen, L-shaped
1001.4600 Arm restraint cuff
1002.59A0 Pivot-mounted elongation tubes for anaesthesia screen
1005.86A0 Direct placement stirrups
1003.23C0 Radial setting clamp or
1003.61A0 Radial setting clamp
1002.2000 Thigh holder
1002.2300 Leg holder
1002.57A0 Anaesthesia screen, L-shaped
1001.4600 Arm restraint cuff
1002.59A0 Pivot-mounted elongation tubes for anaesthesia screen
1003.1000 Iliac support
1004.2400 Arm supports
1005.2300 Crossbar attachment
1005.2900 Basic unit
1005.8700 Leg holder
1001.65A0 Goepel knee crutches
1009.02C0 Meniscus positioning device, radiotranslucent
1003.32C0 Mayo instrument table
1003.3300 Knee supports for rectal positioning
1004.23C0 Elbow support
1004.8000 Countertraction post for humerus
1004.91A0 Side rail extension
1004.84A0 Knee positioning device, motor-powered
1004.84B0 Knee positioning device, manual
VIII. List of accessories
1005.3400 Clamp adapter +
1005.3500 Skull clamp +
1005.3200 Pins for children
1004.9800 Humerus positioning device
1001.4800 Weinberger hand traction device
1009.01C0 Infusion stand
5. Special accessories
1000.5600 Pad
1000.5700 Head ring, PUR foam
1000.5900 Gel head ring
1000.6800 Pad, segments (for percutaneous kidney operations, spinal surgery)
1003.7400 PUR cushion for kidney operations
GA113302GB11
41
IX. Spare parts list
6. Control modules
1009.81D0 Foot switch for height, Trendelenburg/rev., back plate
1009.81D1 Foot switch for height, Trendelenburg/rev., lateral tilt
1009.81D2 Foot switch for height, Trendelenburg/rev., leg plates
1009.81D3 Foot switch for height, Trendelenburg/rev., longit. shift
1133.90A0 Cable-connected hand control
1133.91A0 IR transmitter
3 110 26E9 Charging station EU + US + Japan
1150.95X0 Wall panel
1133.02BX
IX. Spare parts list
You can order replacement parts under the numbers specified. For further details, please refer to the spare parts lists.
1. Operating table
Head rest pad 1130.53B0 9092.0104 Pad for upper back plate 1133.31B0 9092.2424 Pad for lower back plate and seat plate1133.02BX 9092.2383
Leg plate pad 1133.53BC (2 pcs. always required) 9092.0204
Mains cable
for 1133 BX (230V) 3113.3659 for 1133.02 FX (110V) 3113.3669
Rechargeable battery (2 pcs. are always required) 0227.0154
2. Accessories
• Hand control with coiled cable 1133.90A0
• IR transmitter
IR transmitter without charging station 1133.91A0 Amplified IR transmitter incl. charging station (230V) 1133.91L0 System codes numbers and letters 9801.6234 Cover for leg plate 1132.65A0 9082.1223 Control line with two plugs for foot switc 0200.4273
Use only spare parts approved by MAQUET.
42
GA113302GB11
1133.02BX
GA113302GB11
43
All notes and technical data contained in this documentation reflect the status at the publication date. Since we constantly strive to improve all MAQUET products, however, we reserve the right to adopt modifications at any time, without giving prior notice. There might therefore be minor differences between the illustrations in this manual and the product’s current design. This will. however, not impair the validity and correctness of this manual as far as the technical features herein are concerned.
MAQUET GmbH & Co. KG Surgical Tables Division Kehler Straße 31 D-76437 Rastatt / Germany
Phone: + 49 7222 932-0 Fax: + 49 7222 932-838
e-mail: info.sales@maquet.de Internet: http://www.maquet.com Service Hotline: ++ 49 7222 / 932-745
Subject to modifications due to technical development.
GA113302GB11 07 / 2005
Loading...