micro SD card real time data logger
CAT III 1000V, true rms, L/C
MULTIMETER
Model : DM-9962SD
Your purchase of this
MULTIMETER marks a
step forward for you
into the field of
precision measurement.
Although this
MULTIMETER is a
complex and delicate
instrument, its durable
structure will allow
many years of use if
proper operating
techniques are
developed. Please read
the following
instructions carefully
and always keep this
manual within easy
reach.

Caution :
* Risk of electric shock !
Caution :
* Do not apply the overload
voltage, current to the input
terminal !
* Remove test leads before open
the battery cover !
* Cleaning - Only use the dry
cloth to clean the plastic case !
Environment Conditions
* Installation Categories III-1000V.
* Pollution Degree 2.
* Altitude up to 2000 meters.
* Indoor use.
* Relative humidity 80% max.

FEATURES.......................................................................................
SPECIFICATIONS............................................................................
2-1 General Specifications................................................................
2-2 Electrical Specifications..............................................................
FRONT PANEL DESCRIPTION.......................................................
6
4. PRECAUTIONS & PREPARATIONS FOR
MEASUREMENT................................................................................
MEASURING PROCEDURE..........................................................
5-1 Symbols & Units of Display ......................................................
5-2 DC voltage, AC voltage true rms Measurement ......................................
5-3 Resistance Measurement..........................................................
5-4 DC current, AC current true rms Measurement ..................................
5-5 Continuity Check .......................................................................
5-6 Diode Test .................................................................................
5-7 Capacitance Measurement..........................................................
5-8 Inductance Measurement..........................................................
5-9 L/C Calibration......................................................................
14
5-10 Frequency Measurement...................................................
14
5-11 Temperature Measurement................................................ 15
5-12 Relative Measurement................................................................
5-13 Data Hold Operation................................................................
5-14 LCD Backlight ON/OFF........................................................
5-15 Max/Min. value record..............................................................
5-16 RS232 Computer Interface.....................................................
6-1 Preparation before execute datalogger function............................................
18
6-2 Auto Datalogger................................................................... 18
6-3 Manual Datalogger................................................................
6-4 To check the time and sampling time information.................
6-5 Micro SD Card Data structure..........................................................

Saving data from the Micro SD card to the computer....................................................
ADVANCED SETTING....................................................................
8-1 Micro SD memory card Format.....................................................
8-2 Set sampling time..................................................................
8-3 Set Auto power OFF management..................................................................
8-4 Set beeper sound ON/OFF..................................................................
8-5 Set Decimal point of Micro SD card setting..................................................................
8-6 Set clock time(Year/Month/Date,Hour/Minute/Second)..............
26
8-7 Set Change the Temp. ℃,
℉
unit...........................................
MAINTENANCE...............................................................................
9-1 Replacement of Battery..............................................................
9-2 Cleaning......................................................................................
9-3 Replacement of Fuse..................................................................
OPTIONAL ACCESSORIES & ADAPTERS ...............................
THE ADDRESS OF AFTER SERVICE CENTER........................

1. FEATURES
* Real time Datalogger, save the into the micro SD memory
card and can be downloaded to the Excel, extra software
is no need.
* Real time Datalogger, it Built-in Clock (year/month/date/
hour/minute/second ), sampling time set from 2 seconds
to 3600 seconds.
* Manual datalogger is available ( set the sampling time to be
0 seconds).
* Meet IEC 1010 CAT III 1000 V safety requirement.
* LCD with green light backlight, easy reading.
* 6000 counts A/D, high resolution.
* Accepts micro SD memory cards of up to 32 GB capacity.
* Multi function measurement. DCV, ACV, DCA, ACA,
Resistance, Capacitance, Inductance, Frequency,
Temperature, Diode, Continuity beeper.
* Max. & Min. measurement value with recall.
Auto range with manual range selection.
* Temperature measurement possibility.
* V/A/Hz button, when execute the ACV, ACA function also
can measure the frequency of signal.
* Both 10 A, mA, uA current are build fuse for safety
consideration.
10 M ohm impedance for voltage circuit.
Operates from 9 V ( 006P, MIN1604 ) batteries.
Built-in overload protection for most ranges.
Photo couple RS 232 computer serial interface.
* Uses durable, long-lasting components, enclosed in
strong, light weight ABS-plastic housing.
* Full line optional adapters : Clamp adapter,
Pressure adapter, Sound level adapter,
Anemometer adapter, Light adapter, EMF adapter.
1

2. SPECIFICATIONS
2-1 General Specifications
Display 60 mm x 39 mm large LCD display
Measurement DCV, ACV, DCA, ACA, Resistance,Diode,
Continuity beeper,Capacitance,Inductance,
Datalogger
Auto
2 seconds to 3600 seconds
Sampling Time
Setting range
Manual
Push the data logger button
once will save data one time.
@ Set the sampling time to 0 seconds.
Data error no.
≦
0.1% no. of total saved data typically.
micro SD card 4 GB to 32 GB
Capacity
A/D counts no. 6000 counts.
Range selection Auto range with manual range selecting.
Special function
Relative measurement, Data hold,
Data hold To freeze the display reading on the LCD display.
Power On/Off Auto power off or manual power off.
management @ Details please refer page 7
Memory recall
Records Maximum & Minimum readings with recall.
Relative measurement To offset the measurement value.
V/A/Hz button When execute the voltage or current function
also can measure the frequency of signal.
Data output RS 232 PC serial interface, photo couple.
Polarity
Automatic Switching, " - " indicates negative polarity.
Zero adjustment Automatic.
Sampling time Approx. 0.5 to 1 second.
of display
Operating 0 ℃ to 50 ℃ (32 ℉ to 122 ℉),
Temp. & humidity Max. 80% RH.
Power supply 9 V ( 006P, MIN1604 )
DC 9V adapter input
*AC/DC Power adapter is optional.
Power Normal operation (w/o Micro SD card save data).
consumption Approx. DC 5.5 mA
When Micro SD card save the data
Approx. DC 25 mA.

Dimension 190 x 88 x 40 mm ( 7.5 x 3.5 x 1.6 inch ).
Weight 387 g/0.85 LB ( w.o battery ).
Accessories Red and Black Test Leads
Included ( CAT III 1KV Test Leads )...................... 1 Set
600 mA Spare Fuse................................ 1 PC
Instruction Manual.....................................
1 PC
Optional Full line adapters :
accessories ACA/DCA current adapter,
Pressure adapter,
Light adapter, EMF adapter,
Sound level adapter,
High voltage probe.
AC to DC 9V Adaptre……………………..
Type K Temperature probe.....................
Soft carrying case....................................
RS232 cable to D-Sub 9 connector..............................................
RS232 cable to USB connector..............................................
Micro SD card…………………………………..
Data Acquisition software……………………
2-2 Electrical Specifications (23±5
DC Voltage
Range 600.0 mV /6 V/60 V/600 V /1000 V
Resolution 0.1 mV /0.001V /0.01V /0.1V/1 V
Accuracy ± ( 0.5% + 2d ) 600 mV
± ( 0.8% + 1d ) 6 V, 60 V, 600 V, 1000 V
Input impedance 10 M ohm.
Over load 600 mV range ± 380 DCV, 380 ACV
protection other ranges ±1000 DCV, 1000 ACV
AC Voltage ( True RMS )
Range 600.0 mV /6 V/60 V/600 V /1000 V
Resolution 0.1 mV /0.001V /0.01V /0.1V/1 V
Accuracy ± ( 1% + 2d )
* Spec. are tested under 50/60 Hz.
Input impedance 10 M ohm.
Over load 600 mV range ± 380 DCV, 380 ACV
protection other ranges ±1000 DCV, 1000 ACV
3

DC Current, AC Current ( True RMS )
DC Current, AC Current ( True RMS )DC Current, AC Current ( True RMS )
DC Current, AC Current ( True RMS )
Range 10 A/6 A/600 mA/60 mA/6000 uA/600 uA
0.01 A/0.001 A/0.1 mA/0.01 mA/1 uA/0.1 uA
Accuracy DCA ACA
600 uA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
6000 uA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
60 mA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
600 mA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
6 A ± ( 1.5 % + 5d ) ± ( 1.5 % + 5d )
10 A ± ( 1.5 % + 2d ) ± ( 1.5 % + 2d )
* ACA spec. are tested under 50/60 Hz.
Over load 10A range : 10A / 600V fuse.
protection uA, mA range : 600 mA / 600V fuse.
Diode ( Forward voltage, VF )
Diode ( Forward voltage, VF )Diode ( Forward voltage, VF )
Diode ( Forward voltage, VF )
Range 2.9 V DC.
Accuracy ±( 0.5% + 2d )
Capacitance
CapacitanceCapacitance
Capacitance
Auto Range 600 pF/6 nF/60 nF/600 nF/6 uF/60 uF/100 uF/600 uF
Resolution
0.1 pF/0.001 nF/0.01nF/0.1 nF/0.001 uF/0.01 uF/0.1 uF/1uF
600 pF ±( 2.5% + 5d )
Accuracy 6 nF/60 nF/600 nF/6 uF/60 uF ±( 1.5% + 5d )
100uF/600 uF ±( 5% + 5d )
Test frequency
(Auto frequency configuration)
Remark Discharge capacitor before testing.
Over load protection
±30 DCV, 30 ACV.
Inductance
InductanceInductance
Inductance
Auto Range 600 uH/6 mH/60 mH/600 mH/6 H/60 H/100 H
Resolution 0.1uH/0.001mH/0.01 mH/0.1 mH/0.001 H/0.01 H/0.1 H
Test frequency
100Hz /1KHz /10KHz (Auto frequency configuration)
Accuracy 6 mH/60 mH/600 mH/6 H ±( 1.5% + 5d )
60 H/100 H ±( 2.5% + 5d )
Over load protection
±30 DCV, 30 ACV.
4

Frequency
FrequencyFrequency
Frequency
Range 600 Hz/6 KHz/60 KHz/600 KHz/6 MHz/20 MHz
Resolution
0.1 Hz/0.001 KHz/0.01 KHz/0.1 KHz/0.001 MHz/0.01 MHz
Accuracy ±( 0.5% + 2d )
Sensitivity Min. 1 V rms, Max. 5 V rms.
600 Ω/6 KΩ/60 KΩ/600 KΩ/6 MΩ/60 MΩ
0.1Ω/0.001 KΩ/0.01 KΩ/0.1 KΩ/0.001 MΩ/0.01 MΩ
6K/60K/600K/6 M : ±( 1.5 % + 2d )
Beeper will sound if measured resistance less than 20 ohm.
During the operation can memorize the maximum and the
minimum measurement value.
Temperature
Auto Range Resolution Accuracy
200 ℃ to 1000
℃
1
℃
-148.0 ℉ to 391.9
℉
0.1
℉
±( 1% + 2 ℉ )
392 ℉ to 1832
℉
1
℉
Temp. probe The temperature probe ( TP-11 ) is the
optional accessory.
Remark :
* Spec. tested under the environment RF Field Strength
less than 3 V/M & frequency less than the 30 MHz only.
5

3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display 3-10 Function rotary switch
3-2 MAX/MIN ( ▼ ) button
3-11 Temp./ohm/V/Cap. input terminal
3-3 TIME (SET) button 3-12 COM input terminal
3-4 REL(Backlight) button 3-13 mA/uA input terminal
3-5 HOLD ( ▲ ) button
3-14 10A input terminal
3-6 RANGE button 3-15 Battery compartment/Cover
3-7 V/A/Hz button 3-16 RS232 terminal
3-8 ENTER (LOG) button 3-17 DC 9V Power Adapter Input Socket
3-9 AC/DC button 3-18 micro SD card socket

4. PRECAUTIONS & PREPARATIONS FOR
Ensure that the DC 9 V ( 006P, MIN1604 ) batteries are connected
with the right polarity and placed in the battery compartment
correctly.
2) Place the Red & Black Test Leads into the proper input
terminal before making measurement.
Remove either of the test leads from the circuit when
changing the measurement range.
Except operate the " Data Hold " function, it should
cancel the " Data Hold " function, otherwise the display
reading will freeze permanently.
5) Do not exceed the maximum rated voltage and current to
the input terminal.
6) Always switching the " Function Rotary Switch " to the
" Off " position when the instrument is not operation.
Remove the battery if the instrument is not to be used in
For safety consideration, when change the new test
leads, it should use the replace test leads that already
approval of " CATIII-1000 V " at least.
Power On/Off management :
When not use the meter, should rotate the " Function rotary
switch " ( 3-10, Fig. 1 ) to the " OFF " position.
During the measurement, after 10 minutes the meter
will auto power off. If intend to power on again, it
should rotate the " Function switch " to " OFF "
position then set to the new desiring function position.
Disabling auto power off ( not auto power off )
Press the " TIME button " ( 3-3, Fig. 1 ) into
the setting mode,the Poff set " no "
auto power off function can be lifted.

5-1 Symbols & units of display
Symbols Descriptions
Units
Appears when selecting " Automatic range " mode.
Appears when selecting " Manual range " mode.
DC
Appears when selecting DC mode.
( DC voltage or DC current )
AC
Appears when selecting AC mode.
( AC voltage or AC current )
HOLD
Appears when the " Data hold " function is operated.
Appears when the " Relative " function is operated.
Appears when " Max and Min. value record "
Battery voltage is already under the low condition.
Appears when the " Continuity beeper " is operated.
mV, V
Units for" voltage " measurements.
Units for " resistance" measurements.
uA,mA,A
Units for " Current " measurement.
uH,mH,H
Units for " Inductance " measurement.
pF,nF,uF
Units for " Capacitance " measurement.
Units for " Frequency " measurement.
Appears when the " Diode " function is operated.
–
Appears when measuring a DCV or DCA value
is negative.
Units for " Temperature " measurement.
Over range indicator
AdP
Notice must be inserted 9V ADAPTER
8

5-2 DC Voltage, AC voltage true rms Measurement
1) Connect BLACK test lead into " COM " terminal ( 3-12,
Fig. 1 ).
2) Connect RED test lead into " V " terminal ( 3-11, Fig. 1 ).
3) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to
the " V " position.
4) Push the " AC/DC button " ( 3-9, Fig. 1 ) to select the
" ACV " or " DCV " measurement,
5) When LCD show the " AUTO " marker, the meter is
under the " auto range " mode. Meter will select the
suitable measurement range automatically.
6) Under the operation of " auto range " mode, push the
" Range button " ( 3-6 Fig. 1 ) once will execute the
" Manual Range " mode and hold the range, the LCD will
show the " MANU " marker.
Under the manual range operation, push the " Range
button " ( 3-6 Fig. 1 ) > 2 seconds, will return to auto
range operation.
During the measurement, if push the " V/A/Hz button " ( 3-7
Fig. 1 ) once, until the LCD show the " Hz " marker and the
display will show the frequency value of the measurement signal.
5-3 Resistance Measurement
1) Connect BLACK test lead into " COM " terminal ( 3-12,
Fig. 1 ).
2) Connect RED test lead into " Ω " terminal ( 3-11, Fig. 1 ).
3) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to
the " Ω " position.
4) When LCD show the " AUTO " marker, the meter is
under the " auto range " mode. Meter will select the
suitable measurement range automatically.
9

5) Under the operation of " auto range " mode, push the
" Range button " ( 3-6 Fig. 1 ) will execute the " Manual
Range " mode and hold the range, the LCD will show
the " MANU " marker.
Under the manual range operation, push the " Range
button " ( 3-6 Fig. 1 ) > 2 seconds, will return to auto
range operation.
5-4 DC Current, AC Current true rms Measurement
A : 10A range, 6A range.
mA : 600 mA range, 60 mA range.
uA : 6000 uA range, 600 uA range.
1) Connect BLACK test lead into " COM " terminal ( 3-12, Fig. 1 ).
2) For the " mA, uA " measurement, connect RED
test lead into " mA uA " terminal ( 3-13, Fig. 1 ).
For the " 10 A " current measurement, connect RED
test lead into " A " terminal ( 3-14, Fig. 1 ).
Open the circuit in which current is to be
measured. Now securely connect test leads in series
with the load in which the current is be measured.
3) For the " uA " measurement ( 600 uA, 6000 uA ), select
the " Function rotary switch " ( 3-10, Fig. 1 ) to " uA " position.
For the " mA " measurement ( 60 mA, 600 mA ), select
the " Function rotary switch " ( 3-10, Fig. 1 ) to " mA " position.
For the " 10 A " measurement ( 6A,10A ), select the " Function rotary
switch " ( 3-10, Fig. 1 ) to " A " position.
4) Push the " AC/DC button " ( 3-9, Fig. 1 ) to select the
" ACA " or " DCA " measurement,
5) When LCD show the " AUTO " marker, the meter is
under the " auto range " mode. Meter will select the
suitable measurement range automatically.
10

6) Under the operation of " auto range " mode, push the
" Range button " ( 3-6 Fig. 1 ) once will execute the
" Manual Range " mode and hold the range, the LCD will
show the " MANU " marker.
Under the manual range operation, push the " Range
button " ( 3-6 Fig. 1 ) > 2 seconds, will return to auto
range operation.
Remark :
During the measurement, if push the " V/A/Hz button " ( 3-7
Fig. 1 ) once, until the LCD show the " Hz " marker and the
display will show the frequency value of the measurement signal.
5-5 Continuity Check
1) Connect BLACK test lead into " COM " terminal.
2) Connect RED test lead into " Ω " terminal.
3) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to
the " " position.
4) The LCD display will show the " " marker.
5) when the resistance value is less than 20 ohm, the beeper
sound will be generated.
5-6 Diode Test
1) Connect BLACK test lead into " COM " terminal.
2) Connect RED test lead into " V " terminal.
3) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to
the " " position.
The LCD display will show the " " marker.
4) a. When connected with polarity as shown in Fig. 2, a
forward current flow is established and the approx.
Diode Forward Voltage (VF) value in volt will appears
on the display reading. If the diode under test is
defective, " 0.000 " or near " 0.000 " value ( short circuit )
" – – – – " ( open circuit ) will be displayed.
11

Fig. 2
Fig.2
b. When connected as shown in Fig. 3, a reverse check on
the diode is made. If the diode under test is good, "––––"
will be displayed. If the diode under test is defective,
" 0.000 " or other numbers will be displayed. Proper diode
testing should include both steps a. and b. above.
Fig. 3
Fig.3
12

5-7 Capacitance Measurement
1) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to
the " " position.
2) Connect the tested capacitor to " Input terminals " directly.
* If the measured capacity existing the polarity, then
should connect the " + " polarity of the measured
capacitor to the " " terminal ( 3-11, Fig. 1 ),
connect the " - " polarity of the measured
capacitor to the " COM " terminal ( 3-12, Fig. 1 ),
* Full discharge the measured capacitor before the
make the measurement.
3) When LCD show the " AUTO " marker, the meter is
under the " auto range " mode. Meter will select the
suitable measurement range automatically.
4) Capacitance measurement mode, no manual functions,
" Range button " ( 3-6 Fig. 1 ) is invalid.
5-8 Inductance Measurement
1) Connect BLACK test lead into " COM " terminal ( 3-12,
Connect RED test lead into " " terminal ( 3-11, Fig. 1 ).
3) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to
the " " position.
4) When LCD show the " AUTO " marker, the meter is
under the " auto range " mode. Meter will select the
suitable measurement range automatically.
5) Inductance measurement mode, no manual functions,
" Range button " ( 3-6 Fig. 1 ) is invalid.
13

5-9 L/C Calibration
In order to improve the accuracy of a high / low impedance,
Recommend doing open / short calibration
Before measurement.
Open / short calibration procedure:
1)
Simultaneously push " AC / DC " and " REL " button more
to start open / short calibration,The display will show "OPEn",
than 2 seconds, then push "ENTER" key once,LCD will display 30,
and the countdown to zero,
2)
The display will show " Short ", set the input of a short circuit,
then push the " ENTER " key once,LCD will display 30, and the
countdown to zero, then the calibration data is saved to the
EEPROM, complete the calibration.
5-10 Frequency Measurement
1) Connect BLACK test lead into " COM " terminal ( 3-12, Fig. 1 ).
2) Connect RED test lead into " Hz " terminal ( 3-11, Fig. 1 ).
3) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to the
" Hz " position, LCD will show the " Hz " marker.
4) When LCD show the " AUTO " marker, the meter is
under the " auto range " mode. Meter will select the
suitable measurement range automatically.
5) Under the operation of " auto range " mode, push the
" Range button " ( 3-6 Fig. 1 ) once will execute the
" Manual Range " mode and hold the range, the LCD will
show the " MANU " marker.
Under the manual range operation, push the " Range
button " ( 3-6 Fig. 1 ) > 2 seconds, will return to auto
range operation.
14

5-11 Temperature Measurement
1) Plug in the optional " Type K Temperature probe, TP-11 "
into the input terminals, " TEMP input terminal " ( 3-11, Fig. 1 )
and the " COM input terminal " ( 3-12, Fig. 1 )
2) Select the " Function rotary switch " ( 3-10, Fig. 1 ) to
the " TEMP. " position.
3) Temperature measurement mode, no manual functions,
" Range button " ( 3-6 Fig. 1 ) is invalid.
5-12 Relative Measurement
1) During the measurement of ACV, ACA, DCV, DCA,
ohm, Capacitance, Inductance, Frequency and Temperature,
the circuit will memorize the last measured values if push the
" REL. button " ( 3-4, Fig. 1 ) at once, then LCD will
show zero value & a " REL " indicator.
2) The input measured values will deduct last measured
values automatically, then show those new value on the
display.
3) It will cancel the Relative Measurement function if push
the" REL. button" at once again, at same time the " REL ."
marker will disappear.
5-13 Data Hold Operation
1) During the measurement, pushing the " Hold button "
( 3-5, Fig. 1 ) once a while will freeze the measured value
& the LCD will indicate " HOLD " symbol.
2) Push the " Hold Button " again to cancel the data hold
function.
5-14 LCD Backlight ON/OFF
During the measurement, press the
" REL Button " ( 3-2, Fig. 1 )> 2 seconds will turn ON the
" LCD Backlight ".
Press the " REL Button " once again > 2 will turn OFF the
" LCD Backlight " .
15

5-15 Max and Min. value record
1) Application : To record the maximum and the minimum
reading value during the measurement.
2) Push the " MAX/MIN button " ( 3-2, Fig. 1 ) at once ,
the display will show the " REC " markers
Start recording meter the " MAX " and " MIN " value.
3) Push the " MAX/MIN button " ( 3-2, Fig. 1 ) once again
the display will show the " MAX " maker along with
the maximum measured value.
Push the " MAX/MIN button " ( 3-2, Fig. 1 ) once again
the display will show the " MIN " maker along with
the minimum measured value.
4) If intend to cancel the " MAX/MIN Record function" just
push the " MAX/MIN button " ( 3-2, Fig. 1 ) > 2 seconds
the " REC " marker will disappear.
5-16 RS232 computer interface
1) Connect the optional RS232 cable (UPCB-06) of
RS232 terminal (3-16, Fig. 1)
2) The meter will be the serial data via
" RS232 Terminal " is transferred to the computer.
The data output is a 16 digit stream which can be
utilized for user's specific application.
The 16 digits data stream will be displayed in the
following format
︰
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
16

Each digit indicates the following status :
D15 Start Word (0X02)
D14 4
D13 When send the upper display data = 1
D12, D11 Annunciator for Display
DC mV = 18 AC mV = 49 uH = 68
DC V = 34 AC V = 50 mH = 41
DC uA = 35 AC uA = 51 H = 42
DC mA = 37 AC mA = 53 Hz = 31
DC A = 36 AC A = 52
KHz = 33
Ω = 38
pF = A4
MHz = 67
KΩ = 39 uF = 44
℃
= 01
MΩ = 40 nF = 43
℉
= 02
DIODE = 46
D10 Polarity
0 = Positive 1 = Negative
D9 Decimal Point(DP), position from right to the
left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1 Display reading, D1 = LSD, D8 = MSD
For example :
If the display reading is 1234, then D8 to
D1 is : 00001234
D0 End Word (0x0D)
RS232 FORMAT : 9600, N, 8, 1
Baud rate 9600
Parity No parity
Data bit no. 8 Data bits
Stop bit 1 Stop bit
17

6. DATALOGGER
6. DATALOGGER6. DATALOGGER
6. DATALOGGER
6-1 Preparation before execute datalogger function
6-1 Preparation before execute datalogger function6-1 Preparation before execute datalogger function
6-1 Preparation before execute datalogger function
a. Insert the micro SD card
Prepare a " micro SD memory card "( 4 GB to 32 GB, optional )
Open the Micro SD card cover(3-18, Fig. 1), follow the instructions
to the micro SD card inserted .
* It recommend use memory card capacity is
* It recommend use memory card capacity is * It recommend use memory card capacity is
* It recommend use memory card capacity is
≧≧≧≧
4 GB.
4 GB. 4 GB.
4 GB.
* When the instrument power use AC to DC Adapter,
* When the instrument power use AC to DC Adapter, * When the instrument power use AC to DC Adapter,
* When the instrument power use AC to DC Adapter,
Micro SD Card data logger will be enable.
Micro SD Card data logger will be enable. Micro SD Card data logger will be enable.
Micro SD Card data logger will be enable.
If micro SD card just the first time use into the meter, it recommend to
make the " micro SD card Format " at first,please refer chapter 8-1, page 23.
* If the micro SD card exist the trouble during format by the meter,
* If the micro SD card exist the trouble during format by the meter,* If the micro SD card exist the trouble during format by the meter,
* If the micro SD card exist the trouble during format by the meter,
use the computer to reformat again can fix the problem.
use the computer to reformat again can fix the problem. use the computer to reformat again can fix the problem.
use the computer to reformat again can fix the problem.
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 8-6, page 26.
d. Decimal format setting
The numerical data structure of micro SD card
is default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
6-2 Auto Datalogger ( Set sampling time
6-2 Auto Datalogger ( Set sampling time 6-2 Auto Datalogger ( Set sampling time
6-2 Auto Datalogger ( Set sampling time
≧≧≧≧
2 seconds )
2 seconds ) 2 seconds )
2 seconds )
a. Start the datalogger
Press the " LOG Button ( 3-8, Fig. 1 ) > 1.5 seconds
continuously, the LCD will show the text of " LOGGER "
indicator and flashing per sampling time, at the same
time the measuring data along the time information will
be saved into the memory circuit.

How to set the sampling time, refer to Chapter 8-2 ,
How to set the beeper sound is enable, refer to
b. Pause the datalogger
During execute the Datalogger function, if press the
" LOG Button " ( 3-8, Fig. 1 ) once will pause the
Datalogger function ( stop to save the measuring data
into the memory circuit temporally ). In the same time
the " LOGGER " symbol will stop flashing.
If press the " LOG Button " ( 3-8, Fig. 1 ) once again
will execute the Datalogger again, the " LOGGER "
c. Finish the Datalogger
During execute the Datalogger function, press the
" LOG Button ( 3-8, Fig. 1 ) > 1.5 seconds continuously
again will finish the Datalogger function, the " LOGGER "
text will be disappeared and finish the Datalogger function.
6-3 Manual Datalogger ( Set sampling time = 0 seconds )
a. Set sampling time is to 0 seconds
Press the " LOG Button ( 3-8, Fig. 1 ) > 1.5 seconds, the LCD
will show the indicator " LOGGER " and " " symbol
then press the " LOG Button " ( 3-8, Fig. 1 ) once , the
" " and " Position no. " symbol will flashing once
and Beeper will sound once, at the same time the measuring
data along the time information will be saved into the memory
circuit .
19

During execute the Manual Datalogger, it can use the "
▲▲▲▲
Button " ( 3-5, Fig. 1) or "
▼▼▼▼
Button " ( 3-2, Fig. 1 ) to
set the measuring position ( 1 to 99, for example room 1
to room 99 ) to identify the measurement location , the
Display will show Px ( x = 1 to 99 ).
b. Finish the Datalogger
During execute the Datalogger function, press the " LOG
Button " ( 3-8, Fig. 1 ) > 1.5 seconds continuously again will
finish the Datalooger function, the " LOGGER " indication
will be disappeared and finish the Datalogger.
6-4 To check the time and sampling time information
During the normal measurement ( not execute the
Datalogger ), If press " Time " ( 3-3, Fig. 1 ) )
once , the LCD display will present the time
information of Year/Month/Date , Hour/Minute/Second
and the Sampling time information .
6-5 Micro SD Card Data structure
1) When the first time, the Micro SD card is used into the meter,
the Micro SD card will generate a route :
2) If the first time to execute the Datalogger,
under the route DMA01\, will generate a new
file name DMA01001.XLS.
After exist the Datalogger, then execute again,
the data will save to the DMA01001.XLS until
Data column reach to 30,000 columns, then
will generate a new file, for example DMA01002.XLS
20

3) Under the folder DMA01\, if the total files more
than 99 files, will generate anew route, such as
DMA02\ ........
4) The file's route structure :
DMA01\
DMA01001.XLS
DMA01002.XLS
.....................
DMA01099.XLS
DMA02\
DMA02001.XLS
DMA02002.XLS
.....................
DMA02099.XLS
DMAXX\
.....................
.....................
Remark :
XX : Max. value is 10.
7. Saving data from the Micro SD card
to the computer ( EXCEL software )
1) After execute the Data Logger function, take away the
SD card out from the " Micro SD card socket " ( 3-18, Fig. 1 ).
2) Plug in the Micro SD card into the Computer's Micro SD card slot
( if your computer build in this installation ) or
insert the Micro SD card into the " Micro SD card adapter ". then
connect the " Micro SD card adapter " into the computer.
3) Power ON the computer and run the " EXCEL software ".
Download the saving data file ( for example the file
name : DMA01001.XLS, DMA01002.XLS ) from the SD
card to the computer. The saving data will present into
the EXCEL software screen ( for example as following
EXCEL data screens ) , then user can use those EXCEL
data to make the further Data or Graphic analysis
usefully.
21

EXCEL data screen ( for example )
EXCEL graphic screen ( for example, graphic )
22
50
70
90
110
130
150
170
190
210
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27
A
C
V
Graphic
Ch1_Value

Under do not execute the Datalogger function,
press the " SET Button " ( 3-3, Fig. 1 ) continuously at
least two seconds will enter the " Advanced Setting " mode.
then press the " SET Button " ( 3-3, Fig. 1 ) once a while
in sequence to select the six main function, the
display will show :
Micro SD memory card Format
SP- t.............................
Set sampling time ( 0 to 3600 seconds )
PoFF..........................
SET Auto power OFF management
bEEP.......................
dEC.........................
Set Micro SD card Decimal character
dAtE.......................
Set clock time ( Year/Month/Date, Hour/Minute/
Second )
t-CF……… Set Temperature unit ℃ or
℉
Remark :
In the implementation of the "Advanced Settings" function,
If within 10 seconds without work, will exit
" Advanced Settings " function, the LCD will return
Normal screen.
8-1 Micro SD memory card Format
When the display show " Sd F "
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button " (
3-2, Fig. 1 ) to select the upper text to "YES " or
" no ".
YES - Intend to format the Micro SD memory card
no - Not execute the Micro SD memory card format

2) If select the upper to " YES ", press the " Enter Button
" ( 3-8, Fig. 1 ) once again, the Display will show text
" Ent " to confirm again, if make sure to do the
Micro SD memory card format, then press " Enter Button "
once will format the Micro SD memory clear all the existing
data that already saving into the Micro SD card.
8-2 Set sampling time (Second )
When the display show " SP - t "
1)
Use the "▲ Button" (3-5, Fig. 1) or "▼ button" (3-2, Fig. 1)
to adjust the value ( 0 → to Manual Datalogger or
2 ~ 3600 seconds → to Auto Datalogger ), the adjusted value
will be flashed. When the desired value is set, press the
" Enter key " (3-8, Fig. 1) once, save the default sample value.
2) " sp-t " setting is complete, then display will go to
the " Auto power Off " settings screen (Chapter 8-3).
8-3 Set Auto power OFF management
When the display show " PoFF "
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-2, Fig. 1 ) to select the upper text to "YES " or
"no ".
YES - Auto Power Off management will enable.
- Auto Power Off management will disable.
2) After select the upper text to " YES " or " no ", press the
" Enter Button " ( 3-8, Fig. 1 ) will save the setting
function with default.
24

8-4 Set beeper sound ON/OFF
When the display show " bEEP "
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-2, Fig. 1 ) to select the upper text to " YES " or
" no ".
YES - Meter's beep sound will be ON with default.
no - Meter's beep sound will be OFF with default.
2) After select the upper text to " YES " or " no ", press the
" Enter Button " ( 3-8, Fig. 1 ) will save the setting
function with default.
8-5 Set Decimal point of Micro SD card setting
When the display show " dEC "
The numerical data structure of Micro SD card is default used
the " . " as the decimal, for example "20.6" "1000.53" .
But in certain countries ( Europe ...) is used the " , " as
the decimal point, for example " 20,6 " "1000,53".
Under such situation, it should change the Decimal
character at first.
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-2, Fig. 1 ) to select the upper text to " USA " or
" Euro ".
USA - Use " . " as the Decimal point with default.
Euro - Use " , " as the Decimal point with default.

2) After select the upper text to " USA " or " Euro ",
press the " Enter Button " ( 3-8, Fig. 1 ) will save the
setting function with default.
8-6 Set clock time ( Year/Month/Date,Hour/Minute/ Second )
When the upper display show " dAtE "
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-2, Fig. 1 ) to adjust the value ( Setting start from
Year value ). After the desired value is set, press the
" Enter Button " ( 3-8, Fig. 1 ) once will going to
next value adjustment ( for example, first setting
value is Year then next to adjust Month, Date, Hour,
Minute, Second value ).
Remark :
The adjusted value will be flashed.
2) After set all the time value ( Year, Month, Date, Hour,
Minute, Second ), press the " SET Button " ( 3-3, Fig.
1 ) once will save the time value, then the screen will
jump to " Temp. ℃, ℉ unit " setting screen .
Remark :
After the time value is setting, the internal clock will
run precisely even Power off if the battery is under
normal condition ( No low battery power ).
8-7 Set Change the Temp.
℃℃℃℃, ℉℉℉℉
unit
a. Use " ▲ button " ( 3-5, Fig. 1 ) to select " ℃ " or " ℉ ".
b. After select the desiring text ( ℃ or ℉ ), press the
" Enter button " ( 3-8, Fig. 1 ) to save the data with
default.
26

9. MAINTENANCE
9. MAINTENANCE9. MAINTENANCE
9. MAINTENANCE
9-1 Battery replacement
9-1 Battery replacement9-1 Battery replacement
9-1 Battery replacement
Caution :
Caution :Caution :
Caution :
Remove test leads before
Remove test leads beforeRemove test leads before
Remove test leads before
opening the battery cover !
opening the battery cover !opening the battery cover !
opening the battery cover !
1) When the LCD display showing the mark of " ", it is
necessary to replace the battery, However in-spec.
measurement may still be made for several hours after "
Low battery indicator " appears before the instrument
become inaccurate.
2) Open the screw of " Battery Cover " ( 3-15, Fig. 1 ) by
loosing the screws, then
move the battery.
3) Replace with 9 V batteries ( 006P, MIN1604 ) and
reinstate the cover.
9-2 Cleaning
Cleaning - Only use
the dry cloth to clean
the plastic case !
27

9-3 Replacement of Fuse
Caution :
When make the replacement,
should change the right spec
fuse.
a. Fuse A -
Rating :600 mA, Size : 5 mm dia. x 20 mm
To be protected the circuit from overload current at
" 600 uA, 6000 uA, 60 mA, 600 mA " range.
b. Fuse B -
Rating : 10 A, Size : 5.2 mm dia. x 20 mm
To be protected the circuit from overload current at
" 10 A " range.
2) When the uA, mA current range can not operation,
please check if the Fuse A is broken or not:
When the 10 A current range can not operation,
please check if the Fuse B is broken or not:
3) When replace the fuse should take the test leads from the
measuring circuit and power off the meter.
4) Take the screws away from the down case, loose the
housing case, the fuses are install on the fuse socket on
the PCB.
5) For safety consideration, when replace the fuse according
the spec. ( should use the approval fuse ) and reinstall the
cover.
6) Make sure the housing case is secured with the screw
after replace the fuse.
28

10. OPTIONAL ACCESSORIES & ADAPTERS
Memory card micro SD memory card ( 8 GB )
RS232 cable * Computer interface cable.
UPCB-06 * Used to connect the meter to
the computer ( COM port ).
USB cable * Computer interface cable.
USB-11 * Used to connect the meter to
the computer ( USB port ).
Item Model
Carrying Case CA-05A
Light Adapter LX-02
EMF Adapter EMF-824
Pressure Adapter PS-403
Anemometer Adapter AM-402
Tachometer Adapter TA-601
Sound Adapter SL-406
High Voltage Probe HV-40
29

11. THE ADDRESS OF AFTER SERVICE
CENTER
30 180525-DM-9962SD