Lutron Electronics DM-9962SD Operation Manual

micro SD card real time data logger CAT III 1000V, true rms, L/C
MULTIMETER
Model : DM-9962SD
Your purchase of this MULTIMETER marks a step forward for you into the field of precision measurement. Although this MULTIMETER is a complex and delicate instrument, its durable structure will allow many years of use if proper operating techniques are developed. Please read the following instructions carefully and always keep this manual within easy reach.
OPERATION MANUAL
OPERATION MANUAL
Caution Symbol
Caution :
* Risk of electric shock !
Caution :
* Do not apply the overload
voltage, current to the input terminal !
* Remove test leads before open
the battery cover !
* Cleaning - Only use the dry
cloth to clean the plastic case !
Environment Conditions
* Installation Categories III-1000V. * Pollution Degree 2.
* Altitude up to 2000 meters. * Indoor use. * Relative humidity 80% max.
TABLE OF CONTENTS
1.
FEATURES.......................................................................................
1
2.
SPECIFICATIONS............................................................................
2
2-1 General Specifications................................................................
2
2-2 Electrical Specifications..............................................................
3
3.
FRONT PANEL DESCRIPTION.......................................................
6
4. PRECAUTIONS & PREPARATIONS FOR
MEASUREMENT................................................................................
7
5.
MEASURING PROCEDURE..........................................................
8
5-1 Symbols & Units of Display ......................................................
8
5-2 DC voltage, AC voltage true rms Measurement ......................................
9
5-3 Resistance Measurement..........................................................
9
5-4 DC current, AC current true rms Measurement ..................................
10
5-5 Continuity Check .......................................................................
11
5-6 Diode Test .................................................................................
11
5-7 Capacitance Measurement..........................................................
13
5-8 Inductance Measurement..........................................................
13
5-9 L/C Calibration......................................................................
14
5-10 Frequency Measurement...................................................
14
5-11 Temperature Measurement................................................ 15
5-12 Relative Measurement................................................................
15
5-13 Data Hold Operation................................................................
15
5-14 LCD Backlight ON/OFF........................................................
15
5-15 Max/Min. value record..............................................................
16
5-16 RS232 Computer Interface.....................................................
16
6. DATALOGGER
6-1 Preparation before execute datalogger function............................................
18
6-2 Auto Datalogger................................................................... 18
6-3 Manual Datalogger................................................................
19
6-4 To check the time and sampling time information.................
20
6-5 Micro SD Card Data structure..........................................................
20
7.
Saving data from the Micro SD card to the computer....................................................
21
8.
ADVANCED SETTING....................................................................
23
8-1 Micro SD memory card Format.....................................................
23
8-2 Set sampling time..................................................................
24
8-3 Set Auto power OFF management..................................................................
24
8-4 Set beeper sound ON/OFF..................................................................
25
8-5 Set Decimal point of Micro SD card setting..................................................................
25
8-6 Set clock time(Year/Month/Date,Hour/Minute/Second)..............
26
8-7 Set Change the Temp. ,
unit...........................................
26
9 .
MAINTENANCE...............................................................................
27
9-1 Replacement of Battery..............................................................
27
9-2 Cleaning......................................................................................
27
9-3 Replacement of Fuse..................................................................
28
10
OPTIONAL ACCESSORIES & ADAPTERS ...............................
29
11
THE ADDRESS OF AFTER SERVICE CENTER........................
30
1. FEATURES
* Real time Datalogger, save the into the micro SD memory card and can be downloaded to the Excel, extra software
is no need. * Real time Datalogger, it Built-in Clock (year/month/date/ hour/minute/second ), sampling time set from 2 seconds to 3600 seconds. * Manual datalogger is available ( set the sampling time to be
0 seconds). * Meet IEC 1010 CAT III 1000 V safety requirement. * LCD with green light backlight, easy reading. * 6000 counts A/D, high resolution. * Accepts micro SD memory cards of up to 32 GB capacity. * Multi function measurement. DCV, ACV, DCA, ACA,
Resistance, Capacitance, Inductance, Frequency,
Temperature, Diode, Continuity beeper. * Max. & Min. measurement value with recall.
*
Relative, Data hold.
*
Auto range with manual range selection.
* Temperature measurement possibility. * V/A/Hz button, when execute the ACV, ACA function also
can measure the frequency of signal. * Both 10 A, mA, uA current are build fuse for safety
consideration.
*
10 M ohm impedance for voltage circuit.
*
Operates from 9 V ( 006P, MIN1604 ) batteries.
*
Built-in overload protection for most ranges.
*
Photo couple RS 232 computer serial interface.
* Uses durable, long-lasting components, enclosed in
strong, light weight ABS-plastic housing. * Full line optional adapters : Clamp adapter,
Pressure adapter, Sound level adapter,
Anemometer adapter, Light adapter, EMF adapter.
1
2. SPECIFICATIONS
2-1 General Specifications
Display 60 mm x 39 mm large LCD display Measurement DCV, ACV, DCA, ACA, Resistance,Diode,
Continuity beeper,Capacitance,Inductance,
Frequency, Temperature.
Datalogger
Auto
2 seconds to 3600 seconds Sampling Time Setting range
Manual
Push the data logger button
once will save data one time.
@ Set the sampling time to 0 seconds.
Data error no.
0.1% no. of total saved data typically.
micro SD card 4 GB to 32 GB Capacity A/D counts no. 6000 counts. Range selection Auto range with manual range selecting. Special function
Relative measurement, Data hold,
Data hold To freeze the display reading on the LCD display. Power On/Off Auto power off or manual power off. management @ Details please refer page 7 Memory recall
Records Maximum & Minimum readings with recall.
Relative measurement To offset the measurement value. V/A/Hz button When execute the voltage or current function
also can measure the frequency of signal. Data output RS 232 PC serial interface, photo couple. Polarity
Automatic Switching, " - " indicates negative polarity.
Zero adjustment Automatic. Sampling time Approx. 0.5 to 1 second. of display Operating 0 to 50 (32 to 122 ), Temp. & humidity Max. 80% RH. Power supply 9 V ( 006P, MIN1604 )
DC 9V adapter input
*AC/DC Power adapter is optional. Power Normal operation (w/o Micro SD card save data). consumption Approx. DC 5.5 mA
When Micro SD card save the data
Approx. DC 25 mA.
2
2
Dimension 190 x 88 x 40 mm ( 7.5 x 3.5 x 1.6 inch ). Weight 387 g/0.85 LB ( w.o battery ). Accessories Red and Black Test Leads
Included ( CAT III 1KV Test Leads )...................... 1 Set
600 mA Spare Fuse................................ 1 PC
Instruction Manual.....................................
1 PC Optional Full line adapters : accessories ACA/DCA current adapter,
Pressure adapter, Light adapter, EMF adapter, Sound level adapter, High voltage probe.
AC to DC 9V Adaptre……………………..
AP-9VA
Holster…………………………………….
HS-03
Type K Temperature probe.....................
TP-11
Soft carrying case....................................
CA-05A
RS232 cable to D-Sub 9 connector..............................................
UPCB-06
RS232 cable to USB connector..............................................
USB-11
Micro SD card……………………………..
AC-MICROSD
Data Acquisition software………………
SW-U801-WIN SW-E802
2-2 Electrical Specifications (23±5
)
DC Voltage
Range 600.0 mV /6 V/60 V/600 V /1000 V Resolution 0.1 mV /0.001V /0.01V /0.1V/1 V Accuracy ± ( 0.5% + 2d ) 600 mV
± ( 0.8% + 1d ) 6 V, 60 V, 600 V, 1000 V Input impedance 10 M ohm. Over load 600 mV range ± 380 DCV, 380 ACV protection other ranges ±1000 DCV, 1000 ACV
AC Voltage ( True RMS )
Range 600.0 mV /6 V/60 V/600 V /1000 V Resolution 0.1 mV /0.001V /0.01V /0.1V/1 V Accuracy ± ( 1% + 2d )
* Spec. are tested under 50/60 Hz.
Input impedance 10 M ohm. Over load 600 mV range ± 380 DCV, 380 ACV protection other ranges ±1000 DCV, 1000 ACV
3
DC Current, AC Current ( True RMS )
DC Current, AC Current ( True RMS )DC Current, AC Current ( True RMS )
DC Current, AC Current ( True RMS )
Range 10 A/6 A/600 mA/60 mA/6000 uA/600 uA
Resolution
0.01 A/0.001 A/0.1 mA/0.01 mA/1 uA/0.1 uA
Accuracy DCA ACA
600 uA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
6000 uA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
60 mA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
600 mA ± ( 0.5 % + 2d ) ± ( 1 % + 7d )
6 A ± ( 1.5 % + 5d ) ± ( 1.5 % + 5d )
10 A ± ( 1.5 % + 2d ) ± ( 1.5 % + 2d )
* ACA spec. are tested under 50/60 Hz.
Over load 10A range : 10A / 600V fuse. protection uA, mA range : 600 mA / 600V fuse.
Diode ( Forward voltage, VF )
Diode ( Forward voltage, VF )Diode ( Forward voltage, VF )
Diode ( Forward voltage, VF )
Range 2.9 V DC. Accuracy ±( 0.5% + 2d )
Capacitance
CapacitanceCapacitance
Capacitance
Auto Range 600 pF/6 nF/60 nF/600 nF/6 uF/60 uF/100 uF/600 uF Resolution
0.1 pF/0.001 nF/0.01nF/0.1 nF/0.001 uF/0.01 uF/0.1 uF/1uF
600 pF ±( 2.5% + 5d ) Accuracy 6 nF/60 nF/600 nF/6 uF/60 uF ±( 1.5% + 5d )
100uF/600 uF ±( 5% + 5d )
Test frequency
100Hz /1KHz /10KHz
(Auto frequency configuration)
Remark Discharge capacitor before testing.
Over load protection
±30 DCV, 30 ACV.
Inductance
InductanceInductance
Inductance
Auto Range 600 uH/6 mH/60 mH/600 mH/6 H/60 H/100 H Resolution 0.1uH/0.001mH/0.01 mH/0.1 mH/0.001 H/0.01 H/0.1 H
Test frequency
100Hz /1KHz /10KHz (Auto frequency configuration)
600 uH
±( 5% + 5d )
Accuracy 6 mH/60 mH/600 mH/6 H ±( 1.5% + 5d )
60 H/100 H ±( 2.5% + 5d )
Over load protection
±30 DCV, 30 ACV.
4
Frequency
FrequencyFrequency
Frequency
Range 600 Hz/6 KHz/60 KHz/600 KHz/6 MHz/20 MHz Resolution
0.1 Hz/0.001 KHz/0.01 KHz/0.1 KHz/0.001 MHz/0.01 MHz
Accuracy ±( 0.5% + 2d ) Sensitivity Min. 1 V rms, Max. 5 V rms.
OHMS
Range
600 /6 K/60 K/600 K/6 M/60 M
Resolution
0.1/0.001 K/0.01 K/0.1 K/0.001 M/0.01 M
Accuracy
600 ohm : ±( 1 % + 2d )
6K/60K/600K/6 M : ±( 1.5 % + 2d )
60 M : ±( 3 % + 5d )
Over load
±350 DCV, 350 ACV.
protection
Continuity Beeper
Beeper will sound if measured resistance less than 20 ohm.
Max. & Min. Measurement
During the operation can memorize the maximum and the
minimum measurement value.
Temperature
Auto Range Resolution Accuracy
-100.0
to 199.9
0.1
±( 1% + 1
)
200 to 1000
1
-148.0 to 391.9
0.1
±( 1% + 2 )
392 to 1832
1
Temp. probe The temperature probe ( TP-11 ) is the
optional accessory.
Remark : * Spec. tested under the environment RF Field Strength
less than 3 V/M & frequency less than the 30 MHz only.
5
3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display 3-10 Function rotary switch 3-2 MAX/MIN ( ) button
3-11 Temp./ohm/V/Cap. input terminal
3-3 TIME (SET) button 3-12 COM input terminal 3-4 REL(Backlight) button 3-13 mA/uA input terminal
3-5 HOLD ( ) button
3-14 10A input terminal
3-6 RANGE button 3-15 Battery compartment/Cover 3-7 V/A/Hz button 3-16 RS232 terminal 3-8 ENTER (LOG) button 3-17 DC 9V Power Adapter Input Socket 3-9 AC/DC button 3-18 micro SD card socket
6
4. PRECAUTIONS & PREPARATIONS FOR
MEASUREMENT
1)
Ensure that the DC 9 V ( 006P, MIN1604 ) batteries are connected
with the right polarity and placed in the battery compartment correctly.
2) Place the Red & Black Test Leads into the proper input
terminal before making measurement.
3)
Remove either of the test leads from the circuit when
changing the measurement range.
4)
Except operate the " Data Hold " function, it should
cancel the " Data Hold " function, otherwise the display
reading will freeze permanently.
5) Do not exceed the maximum rated voltage and current to
the input terminal.
6) Always switching the " Function Rotary Switch " to the
" Off " position when the instrument is not operation.
7)
Remove the battery if the instrument is not to be used in
a long period of time.
8)
For safety consideration, when change the new test
leads, it should use the replace test leads that already
approval of " CATIII-1000 V " at least.
9)
Power On/Off management :
a.
When not use the meter, should rotate the " Function rotary
switch " ( 3-10, Fig. 1 ) to the " OFF " position.
b.
During the measurement, after 10 minutes the meter
will auto power off. If intend to power on again, it
should rotate the " Function switch " to " OFF "
position then set to the new desiring function position.
c.
Disabling auto power off ( not auto power off )
Press the " TIME button " ( 3-3, Fig. 1 ) into
the setting mode,the Poff set " no "
auto power off function can be lifted.
7