IBDV · NDV · IBV · REO
Quick Guide
This Quick Guide is not intended to replace the xMAP® Flock Monitor (IBDV, NDV, IBV, REO) Assay Package Insert (document number 89-6000000-070 rev. 9). Please refer to the Package Insert for complete background, instructions, warnings and procedure precautions.
Sample, Reagent and Plate Preparation
Before setting up the Flock Monitor assay plate, use the following procedures to prepare fresh reagents and sample
dilutions. Allow all components to equilibrate to room temperature, and prepare aliquots and dilutions immediately
prior to performing the assay.
SAMPLES
CONTROLS
MICROSPHERE
MIX
SA-PE
CONJUGATE
DETECTION
ANTIBODY
• Dilute all samples 1:500 using Sample DILUENT. You will need 50 µL of each diluted
sample per well.
• IMPORTANT - Vortex samples before use and after dilution.
• Dilute CONTROLS labeled (-) and (+) 1:500 using Sample DILUENT. You will need 50 µL
of each diluted control per well.
• IMPORTANT - Vortex controls before use and after dilution.
• Calculate the volume of microspheres you will need with the following formula:
Required microsphere volume = (0.05 mL) x (number of wells) + (10%)
• Aliquot the required microsphere volume into another tube or reagent trough, shield
from light, and set aside for use in assay setup.
• Calculate how much conjugate you will need with the following formula:
Required SA-PE conjugate volume = (0.10 mL) x (number of wells*) + (10%)
• Aliquot this amount into a reagent trough, then cover with foil to protect from light.
• Promptly return SA-PE conjugate to 2-8°C storage.
• Calculate how much conjugate you will need with the following formula:
Required conjugate volume = (0.10 mL) x (number of wells*) + (10%)
• Aliquot this amount into a reagent trough, then cover with foil to protect from light.
• Promptly return detection antibody to 2-8°C storage.
WASH
BUFFER
MAGNETIC
SEPERATION
DEVICE
*Consult the product insert for more information
• WASH Buffer is provided ready-to-use in nal dilution.
• May be a LifeSep 96F hand held magnet for maual washing (see http://www.luminexcorp.
com/support/Magnetic%20Microspheres/Washing_Method_whitepaper.pdf for more details)
• May be a Bio Tek ELx405 or Tecan Hydroex washer equipped with a washer
manufacturer provided magnet for automated washing
89-30000-00-261 Revision A - Page 1 of 2
Before You Start
1234 56789 10 11 12
A
B
C
D
E
F
G
H
Prepare reagents, samples and controls as described on page 1 of this Quick Guide. All steps below are performed
at room temperature, taking care to protect MICROSPHERES, SA-PE CONJUGATE, and DETECTION ANTIBODY
from light by wrapping in foil during incubation.
Allow liquid components to equilibrate to room
temperature for 30 to 60 minutes*
Warm up and calibrate Luminex instrument*
Vortex working MICROSPHERES 5 seconds
Sonicate 5 seconds
Add 50 μl MICROSPHERES to each well
SET UP
Vortex diluted CONTROLS and diluted SAMPLES
Add 50 μl CONTROLS or SAMPLE to
corresponding wells
Cover plate with opaque lid or foil
Shake plate at 800 rpm for 60 minutes
INCUBATEINCUBATE
WASH STEP - May be done with manual or
automatic washer technique
Allow plate to sit on magnet for 60 seconds
Remove liquid from wells
WASH
Add 100 μl WASH to each well
Repeat WASH STEP one more time (total of two
washes)
Remove Final WASH from Wells
Plate Layout
Use this template for setting up your assay plate and
software batch on the Luminex 200 / FlexMAP 3D System.
Shake plate at 800 rpm for 30 minutes
INCUBATE
WASH STEP - May be done with manual or automatic
washer technique
Allow plate to sit on magnet for 60 seconds
Remove liquid from wells
WASH
Add 100 μl WASH to each well
Repeat WASH STEP one more time (total of two
washes)
Resuspend MICROSPHERES in 100μl WASH with
a pipette or plate shaker
Read assay results on a Luminex analyzer
READ
Add100 μl DETECTION ANTIBODY to each well
Cover plate with opaque lid or foil
DETECTION ANTIBODY
Shake plate at 800 rpm for 30 minutes
WASH STEP - May be done with manual or
automatic washer technique
Allow plate to sit on magnet for 60 seconds
Remove liquid from wells
WASH
Add 100 μl WASH to each well
Repeat WASH STEP one more time (total of two
washes)
89-30000-00-261 Revision A - Page 2 of 2
UNITED STATES
Luminex Corporation
Austin, Texas
Tel: +1.512.219.8020
Fax: +1.512.219.5195
Remove nal WASH from wells
Add 100 μl SA-PE CONJUGATE to each well
Cover plate with opaque lid or foil
CONJUGATE
*Consult the product insert for more information
CANADA
Luminex Molecular Diagnostics
Toronto, Ontario
Tel: +1.416.593.4323
Fa x: + 1.4 16. 593 .10 66
EUROPE
Luminex B.V.
Oosterhout,The Netherlands
Tel: +31.162.408333
Fax: +31.162.408337
www.luminexcorp.com
-
-
+
+
Protect microspheres, conjugate, and detection antibody
from prolonged light exposure throughout entire procedure.
CHINA
Luminex Shanghai Trading Co..
Shanghai, China
Tel: +21.616.50809
Fax:+21.616.50811
info@luminexcorp.com
USA: 1-877-8-Biovet (824-6838)
Canada & Intl: 1-888-8-Biovet (824-6838)
JAPAN
Luminex Japan Corporation, Ltd.
Tokyo, Japan
Tel: +81.3.5545.7440
Fax:+81.3.5545.0451
AUSTRALIA
BSD Robotics, A Luminex Company
Brisbane, Queensland
Tel: +61.7.3273.0273
Fax:+61.7.3273.0274