Lightolier MD4GPSVA2SB4SB User Manual

Page 1
MD*4
TM
2' x 4' Recessed or Surface MD4GPSVA2SB4SB
Page 1 of 4 Multi-Function Patient Lighting
Features
• Four lighting functions with intregral ambient, reading, examination and nurses night light.
• Precise light control for glare free patient environment.
• Installs as a one-piece unit.
• Recessed or surface mounted to ceiling which eliminates damage to fixture from hospital equipment attached to bed.
• 6-7/8" deep.
All twin tube T5 lamps are accessible for relamping without tools.
• All ballasts accessible thru louver compartment.
• Optional low voltage electronic sequential switch mounted in fixture for ambient and reading light controlled with pillow speaker (optional by others).
• Optional 50 watt or LED night light.
• Meets NYC Code requirements. Chicago Plenum requirements available.
Mounting Methods Dimensions
S
surface
Ceiling
1
6
Style 1 shown
8
3
7
5
6-7/8"
(175mm)
Wall
exposed T-grid ceiling
G
Secure fixture (4 places)
Wall Angle or T-bar (by others)
Wall
48"
Acoustical Tile (by others)
Now available in 3 styles
(Reflectors always semi-specular)
Traditional with specular louver (PS option)
(For multiple patient room)
White louver and horizontal surfaces to blend with ceiling (FF option)
(Reflectors always semi-specular)
All components white (AW option)
(For single patient room)
48" (1219mm)
6
2
1. Reading light with low iridescence anodized semi-specular aluminum reflector.
2. Extruded aluminum lamp baffle.
3. Ambient light with low iridescence anodized secular aluminum reflector.
4. Examination light with specular 36 cell aluminum louver with virgin acrylic overlay (illustrated).
5. Twin tube T5 lamps (6).
6. Clear acrylic lens.
7. Prismatic virgin acrylic overlay.
8. Ballast compartment accessible by opening louver and reflector.
2" x 3" access plate with (2) 7/8"
dia. K.O.s (Ambient/Reading)
4"
23-3/4"
12-3/4"
2-1/32"
19-11/16" 19-11/16"
Mounting holes (8)
11-7/8"
47-3/4"
4
2" x 3" access plate with (2) 7/8" dia. K.O.s (Examination)
11-7/8"
4-5/16"
Tabs for wire supports
(4 places)
14"
4-7/8"
2-1/32"
Job Information Type:
Job Name:
Cat. No.:
Lamp(s):
Volts/Ballast:
6/07
Lightolier a Genlyte company www.lightolier.com
Technical Information: (978) 657-7600 • Fax (978) 658-0595 631 Airport Road, Fall River, MA 02720 • (508) 679-8131 • Fax (508) 674-4710 We reserve the right to change details of design, materials and finish. © 2007 a Genlyte company
Section 8/Folio R40-11 Rev. B
Page 2
MD*4
TM
2' x 4' Recessed or Surface MD4GPSVA2SB4SB
Page 2 of 4 Multi-Function Patient Lighting
Ambient Light
Indirect illumination from wall directly behind patient provides glare-free lighting.. No direct glare to patient. (Light is directed towards wall.) No direct glare to patient's visitor. No direct glare to adjacent patients. (3 footcandles maximum on wall behind adjacent patients bed.)
AMBIENT LIGHT
ON
LIGHTED
ZONE
SHIELDED
ZONE
45 110 140 110 45 14 4 2 1 1 1 1
44 83 95 83 44 19 7 3 2 2 1 1
30 45 50 45 30 16 8 4 3 3 2 1
16 20 22 20 16 11 7 4 3 3 2 1
79 974322 11
3 3221 1
1211
ADJACENT
MD*4
OFF
Reading Light
Eliminates reflected glare from reading material (over shoulder illumination). No direct glare to patient or visitors. (7 footcandles maximum at head of bed. 15 footcandles maximum at foot of bed.) Controls light directly onto patient's bed. No reading light spill-over to adjacent bed (2 foot/candles maximum at head of adjacent patient's bed.)
SHIELDED
ZONE
LIGHTED
ZONE
Illumination at the reading task ranges 50-60 footcandles.
READING LIGHT
3"
6"
222222111111
3 3 3 3 3222
6 7 7 7 6 4332211
33 38 39 38 33 23 16 12 9 5 3 2
30 38 38 38 30 23 15 12 8 6 3 2
14 15 15 15 14 11 9 7 5 4 2 2
555555442211
Illumination at the reading task range 50-60 footcandles
ON
BED
FOOTCANDLES ON BEDS
WALL
1111
BED
Examination Light
Excellent color rendition. (Biax Type lamp.)
LIGHTED
ZONE
Wired for independent switching (1 or 2 circuit). Optional louver or lens frame. Concealed hinging and latching.
F
3"
6"
OOTCANDLES ON BED (with Specular Louver)
12"
Totals of ambient, reading and examination lights
ADJACENT
MD*4
OFF
12"
.
3"
↓ ↑
12"
6"
70 82 82 102 100 70 48
(11) (12) (15) (45) (43) (20 (10)
75 82 86 107 102 72 48
(11) (12) (15) (45) (43) (20) (10)
70 82 82 102 100 70 48
(11) (12) (15) (45) (43) (20) (10)
BED
Numbers in parenthesis ( ) are total footcandle level of ambient and reading lights only.
12"
Page 3
MD*4
TM
2' x 4' Recessed or Surface MD4GPSVA2SB4SB
Page 3 of 4 Multi-Function Patient Lighting
Nurse's Night Light (Optional NL or LD)
Lights night stand between beds. Field adjustable for either right or left side of patient's bed. No glar
e to any patient.
Wired for independent switching.
CAN BE SET TO
LEFT OR RIGHT
SIDE OF BED
Options
120V medium base narrow spot, 50W reflector lamp
Specular anodized aluminum mirror, adjusts to left or right side of bed
Nurse's night light.. Suffix: NL (Lamp included)
Ballast Compartment accessible by opening Louver and Reflector
Night Light
Night Light: 120V, 3 watt, cool white LED (50,000 hours life) (supplied with fixture). Suffix: LD
Sequential Switch. Suffix: SW
MD*4’s optional low voltage (1-3 volts) sequential switch mounted in the fixture is listed in com­pliance with UL 544 for medical and dental equipment. Two red wires coming from the fixture can be wired to a wall receptacle (by others) for use with that manufacturer's pillow speaker. The pil­low speaker is used to control the television, call the nurse and control the lighting. To control the lighting it should be specified to have a normally open contact switch.
Pressing this switch on the pillow speaker the first time, turns the ambient light on; pressed a second time, turns the reading light on and the ambient light off; pressed a third time, turns both the ambient and reading lights on; pressed a fourth time turns both lights off. Pressing the button again repeats the cycle.
Should the patient's bed be rolled away from the wall, the pillow speaker will simply pull free from the wall receptacle avoiding any damage.
TAMPER RESISTANT/GERM SHIELD Germ Shield: Ideal for psychiatric wards and
germ free environments. Fits lay-in grid ceiling systems with 15/16" maximum ceiling tee face. The heavy duty welded extruded aluminum frame with clear 1/8" thick polycarbonate lens is fixtur
e to ceiling system sealed with a heavy duty
gasket. Request Folio R40-15 REV. A.
Pillow speaker (by others)
Tamper Resistant/Germ Shield
Page 4
MD*4
TM
2' x 4' Recessed or Surface MD4GPSVA2SB4SB
Page 4 of 4 Multi-Function Patient Lighting
ORDERING INFORMATION
Explanation of Catalog Number. Example: MD4GPSVA2SB4SB120SW
MD
MD*4
Recessed or
Surface,
Multi-function
Patient Light
4
Fixture
Size:
2' x 4'
G
Ceiling
Type:
Same Unit
May Be Used
For: Grid (lay-in)
T-bar or Surface
Mounted
Louver/
Lens Frame:
AW = All Components
White (For single patient room)
FF = White Louver
and Horizontal surfaces to blend with ceiling
PS = Traditional with
Specular Louver (For multiple patient room)
Louver
Overlay/Lens:
VA = Virgin Acrylic
.095" Nominal Thickness Overlay (standard)
Options/Accessories
Access Plates: Top wiring access plate is shipped with fixture as standard.
When access plates are required in advance for wiring convenience, specify separately. Catalog Number: ACP. Drywall Kit or Plaster Ceiling: Order Catalog Number: FK92X4.
Request Folio OA30-10 Rev. A.
Electronic Sequential Switch (Low Voltage): Switch factory mounted in fixture for ambient and reading light convenience (controlled by pillow switch, by others). Suffix: SW. Nurses Night Light: 120 volt medium base, narrow spot, 50W reflector lamp with specular anodized mirror, field adjustable to right or left side of patient's bed (Lamp included). Suffix: NL.
Night Light: 120 volt, 3 watt, cool white LED (50,000 hours life). Suffix: LD. Fusing: Internal fast-blow fusing. Suffix: GLR.
Internal slow-blow fusing. Suffix: GMF. Radio Interference Filter: ballast; Suffix: B. Electrical/Wiring Options: Consult factory.
To order one RIF per fixture; Suffix: R. To order one RIF per
2
Lamp
Quantity:
(Ambient/
Reading) 2 Lamps
(Total)
Ballast Type:
(Ambient Reading)
1-2 Lamp Elec. TT5 SB* IB**
(40 Watt)
1-2 Lamp Elec. TT5
(40 Watt)
LOL Dimming TT5 PB
**Instant Start Standard
**Program Start
NOTE:
For TT5 UNV, you must use IB Ballast
<20THD <10THD
PR**
Lamp
Quantity 3 = 3 Lamp 4 = 4 Lamp
Ballast Type:
(Examination)
1-3 Lamp Elec. TT5 X3* I3** 2-2 Lamp Elec. TT5 SB* IB**
2-2 Lamp Elec. TT5 PR**
att)*
(40 W
LOL Dimming TT5 PB*
**Instant Start Standard
**Program Start
NOTE:
For TT5 UNV, you must use IB Ballast
<20THD <10THD
Voltage:
120 or 277
Options:
Add appropriate
Suffix to Catalog
No. i.e. (GLR)
Specifications:
Materials: Chassis parts are die-formed 20 gauge steel. Louver (examination light) – Pre-anodized aluminum Coilzak®or equal.
Reading light – Extruded aluminum lamp baffle with clear extruded acrylic
shields and is removable without tools.
Ambient light – Clear acrylic lens that rotates open for easy relamping. Reflectors (Ambient and Reading Lights): Low iridescence specular pre-
anodized aluminum standard. Finish: Cavity (Examination Light) – Powder coated flat white or black polyester enamel, 89% reflectivity. Five stage rust prevention undercoating process. Chassis exterior – post painted white baked polyester powder. Louver (examination light) PS = low iridescence specular pre-anodized aluminum. Comes with a virgin acrylic overlay FF = flat white baked polyester powder louver. AW = flat white baked polyester powder louver. VA = .095" nominal thickness prismatic virgin acrylic overlay.
Specifications (continued):
Electrical: Rapid start HPF, SB (Biax type) thermally protected class "P" ballast
C.B.M. certified by E.T.L. If K.O. is within 3" of ballast, use wire suitable for at least 90°. Labels: I.B.E.W. UL and ULc Listed. This product may have a mercury containing lamp. Manage in accord with Disposal Laws. See: www.lamprecycle.org
Job Information Type:
Lightolier a Genlyte company www.lightolier.com
Technical Information: (978) 657-7600 • Fax (978) 658-0595 631 Airport Road, Fall River, MA 02720 • (508) 679-8131 • Fax (508) 674-4710 We reserve the right to change details of design, materials and finish. © 2007 a Genlyte company
Section 8/Folio R40-11 Rev. B
Loading...