LifeSource UA-851 User Manual

0 (0)

Multi-Function Automatic

Blood Pressure Monitor

Bilingual Instruction Guide - Model UA-851

English Español

important information

Please read this important information before using your monitor.

Please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this device should not replace regular medical examinations.

Consult your physician if you have any doubt about your readings. Should a mechanical problem occur, contact LifeSource® at 1-888-726-9966.

Do not attempt to service, calibrate, or repair this monitor.

Because your UA-851 monitor contains delicate, high-precision parts, avoid exposing it to extremes in temperature or humidity or to direct sunlight, shock and dust. LifeSource® guarantees the accuracy of this monitor only when it is stored and used within the temperature and humidity ranges noted on page E-21.

Clean the monitor and cuff with a dry, soft cloth or a cloth dampened with water and a mild detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean monitor or cuff.

Remove and replace batteries if monitor is not used for more than six months. Alkaline batteries recommended.

PRECAUTIONS

The UA-851 is designed to be used at home, by those who are eighteen (12) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate.

IMPORTANT SAFETY INSTRUCTiONS - SAVE THESE INSTRUCTIONS

DANGER - TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK,

CAREFULLY FOLLOW THESE INSTRUCTIONS.

For connection to a supply not in the U.S.A., use an attachment plug adapter of the proper configuration for the power outlet.

The power unit (AC Adapter) is intended to be correctly oriented in a vertical or floor mount position.

Fast. Easy. Accurate.

Congratulations on purchasing a state-of-the-art LifeSource® blood pressure monitor—one of the most technologically advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen.

Physicians agree that daily self-monitoring of blood pressure is an important step individuals can take to maintain cardiovascular health and prevent the serious consequences associated with undetected and untreated hypertension.

A&D Medical has been manufacturing high quality healthcare products for over 20 years. We are committed to providing you and your family with monitoring devices that provide the utmost in accuracy and convenience.

LifeSource - Designed for Life.

 

 

TABLE OF CONTENTS

 

 

 

 

 

 

 

 

 

?

WHAT DISPLAY SYMBOLS MEAN . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . E-2

MONITOR COMPONENTS . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . E-3

HOW UA-851 WORKS . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . E-4

before you start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4 SELECTING THE CORRECT CUFF SIZE . . . . . . . . . . . . . . . . . . . . . . . . . . E-5

TAKING YOUR BLOOD PRESSURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6 WHAT IS AN Irregular HeartBeat . . . . . . . . . . . . . . . . . . . . . . . . . E-11 About cuff inflation METER . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-11 About Average Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-12 About Pressure Rating Indicator™ . . . . . . . . . . . . . . . . . . . . . . . E-12 About Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-13

USING THE MONITOR WITH the optional AC ADAPTER . . . . . . . . . . . . E-14

ABOUT BLOOD PRESSURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-15

TROUBLESHOOTING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-20 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-21

CONTACT INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-22

SPANISH INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1

LifeSource Health Line (Toll-Free): 1-888-726-9966

E-1

?WHAT DISPLAY SYMBOLS MEAN

Display/

 

Condition/Cause

Recommended

Symbol

 

Action

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The symbol shown while measure-

Measurement is in progress,

 

 

 

 

 

 

ment is in progress. It blinks while

 

 

 

 

 

 

remain as still as possible.

 

 

 

 

 

 

detecting the pulse.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

An irregular heartbeat or body

 

 

 

 

 

 

 

movement may have occurred.

Take measurement again and

 

 

 

 

 

 

Refer to page E-11 for more infor-

 

 

 

 

 

 

consult with your physician.

 

 

 

 

 

 

mation on irregular heartbeats.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Previous measurement stored in

 

 

 

 

 

 

 

MEMORY.

 

 

 

 

 

 

 

The battery power indicator during

Replace all batteries with new

 

 

 

 

 

 

 

 

 

 

 

 

measurement.

 

Battery Full

 

 

 

 

 

 

 

 

ones when the symbol blinks.

 

 

 

 

 

 

The batteries are low when it

 

 

 

 

 

 

Alkaline batteries recommended.

 

 

 

 

 

 

blinks.

 

Battery Low

 

 

 

 

 

 

 

 

Unstable blood pressure due to

Try the measurement again.

 

 

 

 

 

 

excessive body movement.

Remain very still during the

 

 

 

 

 

 

 

measurement.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The systolic and diastolic values are

Fasten the cuff correctly, and try

 

 

 

 

 

 

within 10 mmHg of each other.

 

 

 

 

 

 

the measurement again.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The pressure value did not increase

Check for air leaks along the

 

 

 

 

 

 

during inflation.

tube and around the air socket.

 

 

 

 

 

 

The cuff is not fastened correctly.

Refasten the cuff and retake mea-

 

 

 

 

 

 

 

surement.

 

 

 

 

 

 

There is an air leak in the cuff or

Make sure tube is properly con-

 

 

 

 

 

 

monitor.

nected to cuff and monitor.

 

 

 

 

 

 

 

Try the measurement again.

 

 

 

 

 

 

The pulse is not detected correctly.

Remain very still during the

PULSE DISPLAY ERROR

 

 

measurement.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cuff Inflation Meter

Measurement is in progress,

 

 

 

 

 

 

remain as still as possible.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure Rating Indicator™

Refer to the section “About

 

 

 

 

 

 

Pressure Rating Indicator” for

 

 

 

 

 

 

 

further explanation.

 

 

 

 

 

 

 

 

E-2 ?

LifeSource Health Line (Toll-Free): 1-888-726-9966

LifeSource UA-851 User Manual

MONITOR COMPONENTS

Display

Air Connector Plug

FRONT

Arm Cuff

Air Hose

Air Socket

START Button

BACK

Battery Compartment

Battery Cover

AC Adapter Jack

Optional AC

 

 

Adapter

LifeSource Health Line (Toll-Free): 1-888-726-9966

E-3

? HOW UA-851 WORKS

LifeSource® blood pressure monitors are easy to use, accurate and digitally display full measurement readouts. Our technology is based on the “oscillometric method” – a noninvasive blood pressure determination. The term “oscillation” refers to any measure of vibrations caused by the arterial pulse. Our monitor examines the pulsatile pressure generated by the arterial wall as it expands and contracts against the cuff with each heartbeat.

The cuff is inflated until the artery is fully blocked. The inflation speed is maximized and the pressure level is optimized by the device. The monitor takes measurements while the cuff is both inflating and deflating. This results in a faster measurement providing greater comfort to the user.

before you start

must install 4 type AA (1.5 volt) batteries (included), or use the optional AC Adapter (see page E-14 for using the Monitor with AC Adapter) and attach the cuff to the monitor before using it. To install batteries (or replace them if the “Low Battery” symbol appears on display), proceed as follows:

1. Remove battery compartment cover by gently pushing down on arrow and sliding cover forward.

2.Put in bottom row of batteries first. Place the batteries in compartment with positive (+) and negative (–) terminals matching those indicated in the compartment. Be sure batteries make contact with compartment terminals.

3.Replace cover by sliding it into the compartment and gently pressing into place.

Step 2

Step 1

Step 3

NOTE: Rechargeable batteries will not work with this monitor.

E-4 ?!

LifeSource Health Line (Toll-Free): 1-888-726-9966

SELECTING THE CORRECT CUFF SIZE

Using the correct cuff size is important for an accurate reading. A cuff that too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small will produce a reading that is higher than the correct blood pressure. With your arm hanging at the side of your body, measure the circumference of your upper arm at the midpoint between the shoulder and elbow.

ARM SIZE

RECOMMENDED CUFF

REPLACEMENT CUFF MODEL #

SIZE

 

 

 

 

 

9.4" - 14.2" (24 - 36cm)

Medium Cuff

UA-280

 

 

 

14.2" - 17.7" (36-45cm)

Large Cuff

UA-281

 

 

 

NOTE: The LifeSource small cuff is not compatible with the UA-851 monitor.

Cuff Size Indicator On The Cuff

Our cuff has an indicator which tells you whether you are using the correct cuff size or not. Place the cuff on your arm (see Page E-6 to learn how to put the cuff on correctly) and if the Index Mark points within the Proper Fit Range, you are using the correct cuff size. If the Index points outside of the Proper Fit Range, contact A&D Medical at 1-888-726-9966 (in Canada, call Auto Control Medical at 1-800-461-0991) for more information on where to

obtain a cuff replacement.

NOTE: If the cuff size indicator falls on a borderline between two size categories, pick either one and make sure to use the same cuff size for each measurement.

USE SMALL CUFF

PROPER FIT RANGE

USE LARGE CUFF

 

INDEX

 

LifeSource Health Line (Toll-Free): 1-888-726-9966

!

E-5

TAKING YOUR bLOOD PRESSURE

Tips for Blood Pressure Monitoring:

Relax for about 5 minutes before measurement.

Do not smoke or ingest caffeine at least 30 minutes prior to measurement.

Remove constricting clothing and place cuff on bare arm.

Unless your physician recommends otherwise, use left arm to measure pressure.

Do not talk during measurement.

Do not cross legs and keep your feet flat on floor during measurement.

Now you are ready. Follow these simple steps:

1. Sit comfortably with your left arm resting on a flat surface so that the center of your upper arm is at the same height as your heart.

2. Lay left arm on the table, palm up and thread cuff end through metal loop, smooth size against arm. Then position the tube off-center toward the inner side of arm in line with the little finger.

3. Secure cuff around arm. The cuff should be snug but not too tight. You should be able to insert two fingers between the cuff and your arm.

4. Confirm that the Index Mark

points within the proper fit range.

E-6

LifeSource Health Line (Toll-Free): 1-888-726-9966