ADVANCED/DELUXE ONE STEP AUTO-INFLATION
Blood Pressure Monitor
Trilingual Instruction Guide – MODEL UA-767 Plus
English • Français • E s pañol
THIS MONITOR IS CLINICALLY VALIDATED
✓
®
This LifeSource
Clinical Validation study to determine the performance and accuracy. This
monitor is clinically validated according to the British Hypertension Society
(BHS) standard. The BHS is the most prestigious international blood pressure
organization. The published study was performed by a reviewing committee
consisting of physicians and/or nurses. Among blood pressure manufactur
ers, A&D Medical is proud to have the highest number of Clinically Validated
monitors with a positive recommendation from professionals that are available to
consumers.
The monitor is recommended by the Canadian Hypertension Society.
blood pressure monitor has undergone and passed a
-
IMPORTANT INFORMATION
Please read this important information before using your monitor.
Please remember that only a medical practitioner is qualified to inter-
☞
pret your blood pressure measurements. Use of this device should not
replace regular medical examinations.
Have your physician review your procedure for using this device. He
☞
or she will want to verify blood pressure measurements before prescribing or adjusting medications.
Consult your physician if you have any doubt about your readings.
☞
Should a mechanical problem occur, contact A&D Medical.
Do not attempt to service, calibrate, or repair this monitor.
☞
Because your UA-767 Plus monitor contains delicate, high-preci-
☞
sion parts, avoid exposing it to extremes in temperature or humidity
or to direct sunlight, shock and dust. A&D Medical guarantees the
accuracy of this monitor only when it is stored and used within the
temperature and humidity ranges noted on page E-18.
Clean the monitor and cuff with a dry, soft cloth or a cloth dampened
☞
with water and a mild detergent. Never use alcohol, benzene, thinner
or other harsh chemicals to clean monitor or cuff.
Remove and replace batteries if monitor is not used for more than
☞
six months. Alkaline batteries recommended.
✓
PRECAUTIONS
The UA-767 Plus is designed to be used at home, by those who are eighteen
(18) years and older, to monitor blood pressure (systolic and diastolic) and pulse
rate.
Fast. Easy. Accurate.
Congratulations on purchasing a state-of-the-art LifeSource®
blood pressure monitor—one of the most technologically
advanced monitors available today. Designed for ease of use
and accuracy, this monitor will facilitate your daily blood
pressure regimen.
Physicians agree that daily self-monitoring of blood pressure is
an important step individuals can take to maintain cardiovascular health and prevent the serious consequences associated
with undetected and untreated hypertension.
?
TABLE OF CONTENTS
WHAT DISPLAY SYMBOLS MEAN . . . . . . . . . . . . . . . . . . . . . . . . . E-2
MONITOR COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
HOW UA-767 PLUS WORKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4
BEFORE YOU START . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4
SELECTING THE CORRECT CUFF SIZE. . . . . . . . . . . . . . . . . . . . . . E-5
TAKING YOUR BLOOD PRESSURE. . . . . . . . . . . . . . . . . . . E-6 — E-7
WHAT IS AN IRREGULAR HEARTBEAT . . . . . . . . . . . . . . . . . . . . . . E-8
ABOUT CUFF INFLATION METER . . . . . . . . . . . . . . . . . . . . . . . . . . E-8
ABOUT AVERAGE READING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-9
ABOUT PRESSURE RATING INDICATOR™ . . . . . . . . . . . . . . . . . . . E-9
ABOUT MEMORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-10
USING THE MONITOR WITH AC ADAPTER . . . . . . . . . . . . . . . . .E-11
A&D Medical has been manufacturing high quality healthcare
products for over 20 years. We are committed to providing
you and your family with monitoring devices that provide the
utmost in accuracy and convenience.
LifeSource - Designed for Life.
ABOUT BLOOD PRESSURE . . . . . . . . . . . . . . . . . . . . . . . E-12 — E-16
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-17
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-18
CONTACT INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-19
FRENCH INSTRUCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
SPANISH INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1
E-1
?
p-3
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
p-3
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
WHAT DISPLAY SYMBOLS MEAN
MONITOR COMPONENTS
?
Display
Condition/Cause
Symbol
The symbol shown while measurement is in progress. It blinks
while detecting the pulse.
An irregular heartbeat or body
movement may have occurred.
Refer to page E-8 for more infor
mation on irregular heartbeats.
Previous measurement stored in
MEMORY.
The battery power indicator
Battery Full
Battery Low
PULSE DISPLAY ERROR
during measurement.
The batteries are low when it
blinks.
Unstable blood pressure due to
excessive body movement.
The systolic and diastolic values
are within 10 mmHg of each
other.
The pressure value did not
increase during inflation.
The cuff is not fastened correctly.
There is an air leak in the cuff or
monitor.
The pulse is not detected correctly.
Cuff Inflation Meter
Pressure Rating Indicator™
Recommended
Action
Measurement is in progress,
remain as still as possible.
Take measurement again and
consult with your physician.
-
Replace all batteries with new
ones when the symbol blinks.
Alkaline batteries recommended.
Try the measurement again.
Remain very still during the
measurement.
Fasten the cuff correctly, and try
the measurement again.
Check for air leaks along the
tube and around the
air socket.
Refasten the cuff and retake
measurement.
Make sure tube is properly con
nected to cuff and monitor.
Try the measurement again.
Remain very still during
the measurement.
Measurement is in progress,
remain as still as possible.
Refer to the section “About
Pressure Rating Indicator” for
further explanation.
-
AIR SOCKET
SYSTOLIC PRESSURE
DIASTOLIC PRESSURE
CUFF INFLATION
METER
PRESSURE RATING
INDICATOR
PULSE RATE
AIR HOSE
BATTERY COMPARTMENT
AC ADAPTER JACK
MEMORY BUTTON
START BUTTON
AIR CONNECTOR PLUG
CUFF
BATTERY COVER
E-2
?
LifeSource Health Line (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
?
E-3