Trilingual Instruction Guide – MODEL UA-767
Blood Pressure Monitor
WHAT DISPLAY SYMBOLS MEAN . . . . . . . . . . . . . . . . . . . . 3
MONITOR COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . 4
HOW 767 WORKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
BEFORE YOU START . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
SELECTING THE CORRECT CUFF . . . . . . . . . . . . . . . . . . . . . . 6
TAKING YOUR BLOOD PRESSURE . . . . . . . . . . . . . . . . 7 — 10
CHECK/ADJUST EXHAUST VELOCITY . . . . . . . . . . . . . 11 — 12
ABOUT BLOOD PRESSURE . . . . . . . . . . . . . . . . . . . . 13 — 15
BLOOD PRESSURE RECORD . . . . . . . . . . . . . . . . . . . . . . . . 16
IMPORTANT INFORMATION . . . . . . . . . . . . . . . . . . . . . . . 17
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
CONTACT INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . 19
TABLE OF CONTENTS
✓
?
❤
Congratulations. You have purchased a LifeSource
state-of-the-art blood pressure monitor, one of the most
technologically advanced yet easy to use products
available in the marketplace today. This LifeSource
monitor is designed to make your daily regimen
useful, and convenient.
Physicians agree that daily self-monitoring of blood
pressure is an important way individuals can contribute
to maintaining their cardiovascular health and preventing
the serious consequences of hypertension left
undiagnosed and untreated.
LifeSource has been manufacturing quality healthcare
products for more than 20 years. Rest assured that we
are committed to providing you and your family with
monitoring devices specifically designed for high accuracy
and ease of use. LifeSource—your source for a lifetime
of health.
1
2
LifeSource Health Line (Toll-Free): 1-888-726-9966 LifeSource Health Line (Toll-Free): 1-888-726-9966
AIR SOCKET
MEASUREMENT
IN PROGRESS
LOW BATTERY
DISPLAY
SYSTOLIC DISPLAY
DIASTOLIC
DISPLAY
STARTBUTTON
AIR CONNECTOR
CUFF
BATTERYCOMPARTMENT
ADJUSTMENT SCREW
BLACK PROTECTION CAP
ON ADJUSTMENT SCREW
PRESSURE
SWITCH
PULSE DISPLAY
MONITOR COMPONENTS
TUBE
Display
Symbol
Recommended
Action
Remain still and refrain
from talking; measurement in progress.
Replace all batteries
with new ones.
Check for air leakage;
make sure cuff tube is
properly connected to
monitor and exhaust
rate is in increments of
2 mmHg. to 5 mmHg
Try measurement again
and remain very still.
Make sure cuff tube is
properly connected to
monitor, cuff is securely
fastened and exhaust rate
is between 2 mmHg and
5 mmHg. Make sure air
connector is inserted
properly into tube
and monitor.
Condition/
Cause
Appears during measurement; blinks and beeps
when pulse is detected.
Battery voltage is
too low for monitor
to work properly.
Appears when systolic
and diastolic measurements are within 10
mmHg of each other.
Appears if pressure
value is unstable due
to movement during
reading.
Appears when pressure
value did not increase
during cuff inflation:
1) cuff is not fastened
2) air connector is
not connected properly.
MEASUREMENT IN
PROGRESS
LOW BATTERY
ERROR 2
ERROR 1
ERROR 3
MONITEUR
TUYAU DU
BRASSARD
WHAT DISPLAY SYMBOLS MEAN
?
3
4
?
?
LifeSource Health Line (Toll-Free): 1-888-726-9966 LifeSource Health Line (Toll-Free): 1-888-726-9966
?
START
TM
UA-767
DIGITAL
BLOOD
PRESSURE
MONITOR
SYSTOLIC
mmHg
DIASTOLIC
mmHg
PUL / min
240
210
180
150
Using the correct cuff size is important for an accurate reading.
A cuff that is too large will produce a reading that is lower than
the correct blood pressure; a cuff that is too small yields a
measurement that is higher. With your arm hanging at the side
of your body, measure the circumference of your upper arm
at the midpoint between shoulder and elbow.
To attach cuff, insert air connector at end of cuff tube into
air socket on left side of monitor. Should you need to change
the cuff size, the preset exhaust velocity rate may have to
be adjusted (see page 11).
ARM SIZE
RECOMMENDED
CUFF SIZE
REPLACEMENT
CUFF MODEL#
5.1” - 7.9” (13-20cm)
7.5” - 12.2” (19-31cm)
11.8” - 17.7” (30-45cm)
Small
Medium
Large
UA-279
UA-280
UA-281
SELECTING THE CORRECT CUFF
TUBE
You must install 4 type AA (1.5 volt) batteries and attach the
cuff to the monitor before using it. To install batteries (or replace
them if the “Low Battery” symbol appears on display), proceed
as follows:
1. Remove battery compartment cover
by gently pushing down on arrow and
sliding cover forward.
2. Put in top row of batteries first. Place
the batteries in compartment with
positive (+) and negative (–) terminals
matching those indicated in the
compartment. Be sure batteries make
contact with compartment terminals.
3. Replace cover by sliding it into
the compartment and gently pressing
into place.
BEFORE YOU START
LifeSource Blood Pressure Monitors are easy to use, accurate
and digitally display full measurement readouts. Our technology
is based on the “oscillometric method” — a noninvasive blood
pressure determination. The term “oscillation” refers to any
measure of vibrations caused by the arterial pulse. The cuff is first
inflated until the artery is fully blocked. Then, the monitor takes
measurements while the cuff deflates. Our monitors examine the
pulsatile pressure generated by the arterial wall as it expands and
contracts against the cuff with each heartbeat.
HOW 767 WORKS
?
5
6
?
LifeSource Health Line (Toll-Free): 1-888-726-9966LifeSource Health Line (Toll-Free): 1-888-726-9966
AIR CONNECTOR
MONITOR