LifeSource DT-605 User Manual

(DT-605)
Jumbo Night Glow Digital Thermometer
Jumbo back-lit display for easy nighttime reading
MEASURING SENSOR
START BUTTON
BATTERY COVER
DISPLAY WINDOW
SAFETY INFORMATION
High, prolonged fever requires medical attention especially for young children. In
case of high readings, please contact your physician.
Do not allow children to take their temperature unattended.Read the enclosed instructions carefully to ensure accurate temperature readings. Do not allow children to walk or run while taking temperature.
The probe could present a hazard.
Never boil the thermometer.Do not open the thermometer except to replace the battery.Keep battery away from children.
SPECIAL PRODUCT FEATURES
Reads temperature in Fahrenheit or Celsius.Beeps when peak temperature is reached.“Nite Glow” feature for easy nighttime reading.Thermometer is water resistant, but not 100% water proof. Do not
submerge unit in any liquid.
Convenient memory feature: automatic recall of last temperature reading.Easy-to-read Jumbo display.Long-life, replaceable battery (approximately 1,500 temperatures).Low Battery Indicator “”.Auto-Off in approximately 10 minutes.Oral, rectal or underarm use.Includes 5 free probe covers.
HOW TO CARE FOR THIS PRODUCT
Use of probe covers is recommended to help avoid spreading germs, and
to keep the thermometer clean. The thermometer should be cleaned after every use even when probe covers are used.
Clean the thermometer using isopropyl alcohol or soap and water. Do
not submerge unit in water. Do not boil the thermometer.
Store the unit in the protective case while not in use.Do not store or keep in direct sunlight or in high temperatures.Do not drop the thermometer, as this may damage the unit.Do not attempt to disassemble the unit except to replace the battery.
WHAT IS A “NORMAL” TEMPERATURE?
Temperature readings vary among individuals. A normal temperature refers to each individual’s average temperature while they are well. Normal aver­age temperatures range from 97°F/36.1°C to 100°F/37.7°C when taken orally. Oral, underarm, and rectal temperatures vary by approximately 1°F (0.5°C). A rectal temperature is considered your real or “core” body tem­perature. A rectal temperature is 1°F/0.5°C higher than an oral reading and an underarm temperature is 1°F/0.5°C lower than an oral reading.
HOW TO TAKE TEMPERATURE
1.Place a probe cover carefully over the tip (see figure on right). Select oral, underarm or rectal as your measurement site (see following instructions).
2.Firmly push the START button. The thermometer will beep and the display will read: “188.8°FM”. This is a function check indicating the unit is working properly. If a previous reading is stored in memory, the thermometer will display the last temperature for three seconds. The night-glow feature automatically lights up when the thermometer is turned on.
3. The °F or °C symbol will flash when the unit is ready to measure temperature. “ ” indicates a low battery. Replace batter y (see “Changing the Battery” section) for accurate measurement. An “ERR” message in the window indicates your unit is not functioning proper­ly. Ensure that the batteries are installed correctly and review instruc­tions for proper operation.
4.When peak temperature is reached, the °F or °C symbol will stop flashing, and the thermometer will beep repeatedly. It is not necessary to hear the beep to ensure the peak temperature has been reached. When the °F or °C symbol stops flashing and the display reads a consistent temperature, the peak temperature has been reached.
5.Read and record the temperature and time for reference. The reading will automatically be stored in the memory until another reading is taken or the batteries are removed from the thermometer.
6.Push the START button to tur n the unit off. If you do not turn the unit off, it will shut off automatically in approximately 10 minutes.
ORAL USE
Do not drink hot or cold fluids, exercise, smoke or
perform other activities that raise or lower temperature. The mouth should remain closed up to 5 minutes before attempting a reading.
Place the probe tip well under the tongue in locations
as indicated by the “” (see figure on right). Instruct children not to bite down on the thermometer.
Approximate measuring time is 30 seconds. Opening of mouth or
improper placement of the probe tip can result in longer read time.
UNDERARM (AXILLARY) USE
From a medical point of view, this method will
always give an inaccurate reading and should not be used if a precise measurement is required.
Wipe underarm with a dry towel.Place probe tip in armpit and bring arm down against
the body. The probe should be pointing upward parallel to the arm.
Approximate measuring time is 30 seconds or longer depending on
placement of the probe and closure of the arm against the body.
RECTAL USE
Commonly used for babies, young children or when it is difficult to
take an oral or underarm temperature.
After placing the probe cover on the thermometer, lubricate with a
water-soluble lubricant.
Gently insert the probe tip no more than 1/2inch or 1.3cm into the
rectum. Do not force the tip into the rectum if resistance is encountered.
Dispose of the used probe cover and wash the thermometer as recom-
mended.
Approximate measuring time is 30 seconds.
SPECIFICATIONS
RANGE:
32.0°C - 43.9°C
89.6°F - 109.9° F
ACCURACY:
± 0.1°C between 35.0° and 40.0°C, ± 0.2°C under 35.0° or over 40°C ± 0.2°F between 96.0° and 104.2°F, ± 0.4°F under 96.0° or over 104.2°F At standard room temperature between 18.0-28.0° C (64.4-82.4° F)
DISPLAY:
Liquid Crystal display 3.5 digits
BATTERY:
One 1.55V button size battery (SR41)
POWER CONSUMPTION:
Measurement mode: 0.1 mA Off: .001 mA Beeper: 0.15 mA Backlight: 5 mA
BATTERY LIFE:
Approximately 600 measurements
DIMENSIONS:
5.7” x 0.8” x 0.4” (14.5 cm x 2 cm x 1 cm)
WEIGHT:
Approximately 10 grams including battery
FAST READ:
Provides readings in approximately 30 seconds
BEEPER:
Beeps approximately 10 seconds after peak temperature is reached.
PRECAUTIONS
Do not drop unit or subject to shock.Do not soak the thermometer in water.Do not boil the probe. Clean the probe with isopropyl alcohol.Do not use solvents or thinners on the unit.Do not store the unit in direct sunlight or at high temperature.Do not attempt to disassemble the unit except to replace the battery.
WARRANTY
A&D Medical warrants to the first retail purchaser that this digital ther­mometer is free from defects in material and workmanship for the lifetime of the original owner. This warranty does not cover batteries or damage resulting from improper handling or use. To obtain replacement product, please return the product, freight and insurance pre-paid, together with proof of purchase and applicable shipping costs to A&DMedical at the address below. Canadian residents can return the product to Auto Control Medical. Call us toll-free at 1-888-726-9966 for shipping costs and further information (in Canada, call 1-800-463-5414). At A&D Medical’s option, the product will be replaced with a new or refurbished product of equal value. THE WARRANTY AND REMEDY PROVIDED ABOVE ARE EXCLU­SIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES. Some states may have laws which amend this warranty.
CHANGING THE BATTERY
When the “ ” appears in the lower right of the display, the battery is exhausted and needs replacing. Make sure you use the same type of battery and proceed as follows: Remove battery cover. Use a small screwdriver or a straight pin to remove battery. Insert new equivalent battery type (1.55V, SR41). Ensure battery is correctly positioned with the “+” facing up. Carefully replace the battery cover. Properly dispose of the battery, keeping it from small children and heat.
SWITCHING BETWEEN FAHRENHEIT & CELSIUS*
1.Press and release the START button.
2.Wait for display to read --- °C / --- °F (approx. 5 seconds) .
3.Before °C/°F begins flashing, press the START button for three seconds.
4.Unit will switch from °C and °F.
5.When °C/°F symbol begins to flash, the thermometer is ready to take your temperature.
NOTE: The last reading is erased from memory when measurement
mode is switched.
TO RECALL THE LAST TEMPERATURE
1.Press the START button while tur ning on the unit. The last measured temperature will be shown in the display with the letter “M” shown under the degree sign for approximately 3 seconds.
2.The thermometer will be in “ready-to-measure” mode and the recalled temperature will be erased. Note: If the device is switched off before completion of temperature, the thermometer automatically goes to “ready-to-measure” mode.
CLEANING AND DISINFECTING
After each use, clean the thermometer using isopropyl alcohol (70%) or soap and water. The thermometer is water resistant, but not 100% water­proof. Do not submerse the thermometer in any liquid.
LifeSource is a trademark of A&D Medical. ©2003 A&D Medical. All rights reserved.
Specifications subject to change without notice. Made in China
In United States: In Canada: A&D Medical Auto Control Medical, Inc. A division of A&D Engineering 6695 Millcreek Drive, Unit 5 1555 McCandless Drive Mississauga, Ontario Milpitas, CA 95035 L5N 5R8 Toll-free LifeSource Health Line 1-888-726-9966 1-800-463-5414 www.LifeSourceOnline.com www.autocontrol.com
U.S. Patent #6,419,388
Loading...