LeCroy warrants this oscilloscope accessory for normal use and operation within specification for a period of one year from
the date of shipment. Spare parts, replacement parts and repairs are warranted for 90 days.
In exercising its warranty, LeCroy, at its option, will either repair or replace any assembly returned within its warranty period
to the Customer Service Department or an authorized service center. However, this will be done only if the product is determined by LeCroy’s examination to be defective due to workmanship or materials, and the defect is not caused by misuse,
neglect, accident, abnormal conditions of operation, or damage resulting from attempted repair or modifications by a nonauthorized service facility.
The customer will be responsible for the transportation and insurance charges for the return of products to the service facility. LeCroy will return all products under warranty with transportation charges prepaid.
This warranty replaces all other warranties, expressed or implied, including but not limited to any implied warranty of merchantability, fitness or adequacy for any particular purposes or use. LeCroy shall not be liable for any special, incidental, or
consequential damages, whether in contract or otherwise.
Copyright 2008 by LeCroy Corporation. All rights reserved.
LeCroy, ActiveDSO, JitterTrack, WavePro, WaveMaster, WaveSurfer, WaveLink, WaveExpert, WaveJet, and Waverunner
are registered trademarks of LeCroy Corporation. Other product or brand names are trademarks or requested trademarks
of their respective holders. Information in this publication supersedes all earlier versions. Specifications subject to change
without notice.
HFP1000-OM-E Rev F
Manufactured under a
tered Quality Management System
Visit www.lecroy.com to view the certificate.
n ISO 90
00 Regis-
This electronic product is subject to disposal and recycling regulations that
vary by country and region. Many countries prohibit the disposal of waste electronic equipment in standard waste
receptacles.
For more information about proper disposal and recycling of your LeCroy
product, please visit www.lecroy.com/
recycle.
Page 3
DECLARATION OF CONFORMITY
according to ISO/IEC Guide 22 and EN 45014:1998
Manufacturer’s Name:
Manufacturer’s Address:
herewith declare that
Product(s) Name: Active Voltage Probe
Model Number(s): HFP 3500, HFP2500, HFP1500, HFP1000
including all their options are in conformity with the provisions of the following EC directive(s), including
the latest amendments, and with national legislation implementing these directives:
and that conformity with Council Directive 73/23/EEC is based on
EN 61010-2-031:2002 Safety requirements for electrical equipment for measurement
control and lab
and that conformity with Council Directive 89/336/EEC is based on
EN 61326/A3:2003 EMC requirements for electrical equipment for measurement
control and laboratory use
Emissions EN 55011/A2:2002 Conducted and Radiated Emissions
Immunity EN 61000-4-2/A2:2001 Electrostatic Discharge
LeCroy Corporation
700 Chestnut Ridge Road
Chestnut Ridge, NY 10977
USA
73/23/EEC Low Voltage Directive
89/336/EEC EMC Directive
oratory use
Part 031: Particular requirements for
for electrical test and measurement
EN 61000-4-3/A1:2003 RF Radiated Electromagnetic Field
EN 61000-4-4:2004 Electrical Fast Transient/Burst
EN 61000-4-5/A1:2001 Surge
EN 61000-4-6/A1:2001 RF Conducted Electromagnetic Field
hand-held probe assemblies
By: David C. Graef European Contact:
Vice President & Chief Technology Officer Your local LeCroy Sales Office or
Place: LeCroy Corporation LeCroy Europe GmbH
700 Chestnut Ridge Road Waldhofer Str 104
Chestnut Ridge, NY 10977 D-69123 Heidelberg
USA Germany
Date: December 18, 2007 Tel: (49) 6221 82700
Fax: (49) 6221 834655
Warning: This is a Class A product. In a domestic environment this product may cause
radio interference, in which case the user may be required to take adequ ate measu res.
HFP1000 Test Record .......................................................................................... A–2
viHFP1000-OM-E Rev E
Page 7
Safety Information
1 Safety Information
OPERATOR SAFETY
To avoid personal injury and to prevent damage to the probe or any products connected to it,
review the following safety precautions. To avoid potential hazards, use the probe only as specified.
Warning
For indoor use only. To avoid fire or personal injury, comply with the following:
Connect probe to the measurement instrument before connecting test leads to a voltage source.
Do not disconnect probe from instrument while test leads are connected to a voltage source.
Use of the probe and/or the oscilloscope it is connected to in a manner other than that specified
may impair the protection mechanisms.
Do not apply a potential to any input that exceeds the maximum rating of that input. (Section 9,
Specifications.)
When measuring high frequency signals, be sure to comply with the voltage vs. frequency derating
curve. (Section 9, Specifications.)
Do not use in wet or explosive atmospheres.
Do not use the probe if any part is damaged. All maintenance should be referred to qualified service personnel.
CONVENTIONS USED IN THIS MANUAL
The following conventions may appear in this manual:
Note
A Note contains information relating to the use of the product.
CAUTION
A Caution contains information that should be followed to avoid possible damage
to the instrument or the device under test.
HFP1000-OM-E Rev E1–1
Page 8
HFP1000 High Frequency Probe
WARNING
A Warning alerts you to a potential hazard. Failure to adhere to the statement in a
WARNING message could result in personal injury.
The following symbols may appear on the product:
CAUTION: Refer to
accompanying documents
This refers you to additional information contained in this manual. The corresponding information
in the manual is similarly denoted.
CAUTION: Risk of electric
shock
This is a reminder that high voltage may be present and that appropriate caution should be taken.
This is the symbol for earth ground.
###
1–2HFP1000-OM-E Rev E
Page 9
Overview
2 Overview
PRODUCT DESCRIPTION
The 1 GHz HFP1000 is a small, high frequency active probe designed to meet today’s increasing
demand for measurements on a variety of test points.
With low input capacitance and high input resistance, circuit loading is minimized.
The HFP1000 can be used with LeCroy’s WavePro™, Waverunner™ and LC series oscilloscope
with firmware version 8.7.0 or higher.
When the probe is used with any of these oscilloscopes, an AutoColor ID feature automatically illuminates the probe head in the default trace color of the channel to which the probe is connected,
eliminating the need for color bands or other markers.
With the ProBus interface, the HFP1000 becomes an integral part of the oscilloscope. The probe
can be controlled from the oscilloscope’s front panel. The oscilloscope provides power to the
probe, so there is no need for a separate power supply or batteries.
When used in combination with an ADPPS power supply, the HFP1000 probe can be used with
non-LeCroy instruments. (See Section 4 for detailed operation with an ADPPS adapter).
Note:
The AutoColor ID feature will be disabled when the probe is used with an ADPPS
adapter or a LeCroy oscilloscope with firmware version lower than 8.7.0.
KEY BENEFITS
• High frequency performance
• Low input capacitance
• Wide dynamic range
• ProBus interface
• AutoColor feature matches the probe color to the oscilloscope’s default trace color
• Five interchangeable tips for probing a variety of test points (two standard, three
optional accessories)
• Replaceable probe tip socket
• Hands free probing with optional FreeHand Probe Holder
HFP1000-OM-E Rev E2–1
Page 10
HFP1000 High Frequency Probe
STANDARD ACCESSORIES
The HFP1000 probe is shipped with the following standard accessories:
Item:Quantity:
Straight Tip4
Sharp Tip4
Clip, 0.8 mm2
Square Pin Ground Spring1
Offset Pin2
Short Right Angle Lead1
Long Right Angle Lead1
Short Single lead1
Long Single Lead1
Replaceable Cartridge1
Instruction Manual1
Certificate of calibration1
OPTIONAL ACCESSORIES
The following items are available as optional accessories for the HFP1000 probe:
IC Lead Tip
SMD Discrete Tip
Bent Sharp Tip
Micro Clip
Ground Spring with Hook
FreeHand Probe Holder
High Frequency Cartridge
Soft Accessory Case
For part number information for standard and optional accessories refer to Section 6, Care and
Maintenance, Replaceable Parts List.
###
2–2HFP1000-OM-E Rev E
Page 11
Features and Accessories
3 Features and Accessories
The HFP1000 probe is provided with numerous features and accessories to make probing and
connecting to different test points easier than ever.
PROBE HEAD
The small, low mass probe head is designed for ease of
use and high performance.
Ground
Socket
Input socket
AUTOCOLOR ID
AutoColor ID LED
DESCRIPTION OF STANDARD AND OPTIONAL ACCESSORIES
The probe tip socket fits easily onto 0.025 inch square
pins for direct access to test points. Several different
adapters are available which connect directly in the
probe socket.
The probe tip socket has a removable tip cartridge for
easy replacement in case the probe socket gets damaged.
The ground socket will accept several different ground
leads to provide a short ground path for high frequency
performance.
The AutoColor ID consists of an LED inside the probe
head which illuminates the probe body in the default
trace color of the channel to which the probe is connected.
The AutoColor ID will only function when the probe is
connected to a LeCroy oscilloscope supplied with the
ProBus interface and firmware version 8.7.0 or higher.
The colors are correct when factory default color
scheme 1 is selected.
The following Tip and Clip accessories can be pushed
into the probe tip socket, ground socket or any other
socketed lead or adapter.
HFP1000-OM-E Rev E3–1
Page 12
HFP1000 High Frequency Probe
A. Tips
Straight Tip
The straight tip is rugged and designed for general
probing. Fits in either probe socket.
PACC-PT001, package of 4.
Sharp Tip
Rugged, titanium tip designed to connect to the smallest vias and small test points. Fits in either probe
socket.
PACC-PT002, package of 4.
IC Lead Tip
Covered in insulation on all sides (except for a small
edge), this tip was designed to prevent shorting neighboring IC leads. The gold part of the tip is not insulated
and should touch the IC lead to be tested. It is one-sizefits-all and will work with any IC lead pitch. Fits in either
probe socket.
The IC Lead Tip is an optional accessory for the
HFP1000.
PACC-PT003, package of 4.
SMD Discrete Tip
The crescent shape of this tip is designed to fit tightly on
capacitors, resistors, transistors and other surface
mount components with discrete leads. Fits in either
probe socket.
The SMD Discrete Tip is an optional accessory for the
HFP1000.
PACC-PT004, package of 4.
3–2HFP1000-OM-E Rev E
Page 13
Bent Sharp Tip
B. Clips
Micro Clip (0.5 mm)
Features and Accessories
Made out of titanium, this tip is ideal for situations that
require the user to hold the probe parallel to the circuit
board under test. Also gives the user more control when
holding the probe like a pencil. Fits in either probe
socket.
The Bent Sharp Tip is an optional accessory for the
HFP1000.
PACC-PT005, package of 4.
A pincher like tip designed to hold onto fine pitch leads
and small components, commonly found in SMD ICs.
Fits in either probe socket, or can be used with a lead.
The Micro Clip is an optional accessory for the
HFP1000.
PACC-CL001, package of 4.
Clip (0.8 mm)
A pincher like tip designed to hold onto larger wires and
components than possible with the Micro Clip, including
through-hole mounted components.
This clip cannot be connected directly into either of the
probe head sockets; it must be connected to a lead.
PK006-4, package of 2.
Ground Spring with Hook
A flexible spring connected to a square pin that fits into
either of the probe head sockets. Designed to be used
as a ground lead, there is a hook on the end of the
spring so that it can probe general circuits.
The Ground Spring with Hook is an optional accessory
for the HFP1000.
PACC-LD001, package of 4.
HFP1000-OM-E Rev E3–3
Page 14
HFP1000 High Frequency Probe
Square Pin Ground Spring
A flexible spring connected to a square pin that fits into
either of the probe sockets. Designed to be attached to
a square pin on the circuit under test.
PACC-LD002, package of 4.
Offset Pin
The offset pin is designed to be attached to either socket
of the probe head. The offset pin is the highest quality
grounding solution and is recommended in high frequency applications.
405400003, package of 1.
C. Leads
While longer leads provide greater flexibility when connecting the probe to a circuit, the added
inductance may degrade the fidelity of high frequency signals. See Section 4 for additional information.
3–4HFP1000-OM-E Rev E
Page 15
Short and Long Right Angle Lead
Short and Long Single Lead
D. Probe Holder
FreeHand Probe Holder
Features and Accessories
This lead has a socket on one end and a bent square
pin on the other to connect to the input or ground socket
of the probe body, and may be used for general purpose probing.
PACC-LD003 (short), PACC-LD004 (long), packages
of 4.
This lead can be used for either ground or input lead.
It has a socket on one end and a square pin on the
other and may be used for general purpose probing.
PACC-LD005 (short), PACC-LD006 (long), packages
of 4.
FreeHand lets you focus on the oscilloscope screen
instead of on maintaining contact to multiple test points.
It allows the user to concentrate on what is really important – the waveform.
It is designed to keep most of the weight on the probe
tip and will prevent lost contact when a bump to the
table shakes the circuit under test.
Additionally, the HFP probe can be mounted horizontally or vertically in the FreeHand, giving added mea-
surement flexibility.
The FreeHand probe holder is an optional accessory for
the HFP1000.
PACC-MS001, package of 1.
HFP1000-OM-E Rev E3–5
Page 16
HFP1000 High Frequency Probe
To use the FreeHand probe holder
d
c
Installing probe into FreeHand
E. Cartridges
Replaceable Cartridge
1. Slide the probe cable through the slot on the bottom
of the translucent holder section.
2. Slide probe backwards in the probe holder.
If the input tip socket gets damaged, you don’t have to
replace the entire probe, because the HFP series active
probe has a removable tip socket cartridge.
PACC-MS002, package of 1.
High Frequency Cartridge
By having a fixed tip rather than a socket, the High Frequency cartridge is able to increase signal fidelity at
higher frequencies.
The High Frequency Cartridge is an optional accessory
for the HFP1000.
PACC-MS003, package of 1.
3–6HFP1000-OM-E Rev E
Page 17
Removal and Installation of the Replaceable Cartridge
To remove old cartridge:
1. Slide the cable strain relief over the cable away from
the probe body.
2. To release the latch, lift the part closest to the strain
relief away from the probe body and slide the cartridge toward the strain relief.
Removing old cartridge
To install a new cartridge:
1. Slide the new cartridge onto the probe body until the
latch engages.
2. Slide the cable strain relief forward to cover the back
end of the probe body.
Installing new cartridge
Features and Accessories
Note:
The cable strain relief is polarized and fits over the probe body in
one direction only.
###
HFP1000-OM-E Rev E3–7
Page 18
HFP1000 High Frequency Probe
BLANK PAGE
3–8HFP1000-OM-E Rev E
Page 19
Operation
4 Operation
HANDLING THE PROBE
Exercise care when handling and storing the probe. Always handle the probe by the probe body or
compensation box. Avoid putting excessive strain or exposing the probe cable to sharp bends.
CONNECTING THE PROBE TO THE TEST INSTRUMENT
The HFP1000 probe has been designed for use with LeCroy’s WavePro™, Waverunner™ and LC
oscilloscopes equipped with the ProBus interface. When you attach the probe output connector to
the oscilloscope’s input connector, the oscilloscope will recognize the probe, provide proper termination and activate the probe control functions in the user interface.
CONNECTING THE PROBE TO THE TEST CIRCUIT
To maintain the high performance capability of the probe in measurement applications, care must
be exercised in connecting the probe to the test circuit. Increasing the parasitic capacitance or
inductance in the input paths may introduce a “ring” or slow the rise time of fast signals. Input
leads which form a large loop area will pick up any radiated electromagnetic field which passes
through the loop and may induce noise into the probe input.
Using one of the available accessories makes the HFP1000 probe with its small profile and low
mass head ideally suited for applications in dense circuitry.
OPERATION WITH A LECROY OSCILLOSCOPE
When the HFP1000 probe is connected to any LeCroy oscilloscope, the displayed scale factor and
measurement values will be automatically adjusted.
Control through the oscilloscope’s interface can be found in the ’Coupling’ menu of the channel to
which the probe is connected.
Turning the Volts/Div knob will control the oscilloscope’s scale factor to give full available dynamic
range up to 2 V/div.
USE WITH NON-LECROY INSTRUMENTS
The HFP1000 can be used with other instruments with the optional ADPPS power supply. The output of the ADPPS must be terminated into 50 Ω.
The AutoColor ID feature will be disabled when the probe is used with an ADPPS adapter.
USE WITH OLDER LECROY INSTRUMENTS
When used with LeCroy instruments with software version lower than 8.7.0, the oscilloscope will
provide the correct scale factor but no AutoColor ID.
With V/div settings greater than 2 V/div, it is possible to display clipped waveforms on screen.
###
HFP1000-OM-E Rev E4–1
Page 20
HFP1000 High Frequency Probe
BLANK PAGE
4–2HFP1000-OM-E Rev E
Page 21
High Frequency Measurements
5 High Frequency Measurements
INPUT LOADING
When you touch a probe to the circuit under test, the probe will affect your measurement because
of the probe’s input impedance introduced into the circuit. All probes present resistive, capacitive
and inductive loading.
INDUCTIVE LOADING (LEAD LENGTH)
A significant element in this circuit is the inductance shown in the input ground leads of the oscilloscope probe.
Figure 5-1. Probe Input Equivalent Circuit
The ground lead is the primary return path for the current resulting from the input voltage acting on
the probe’s input impedance. The ground lead and input lead inductances act with the probe’s
input capacitance to form series L-C network. The impedance of a series LC network will drop dramatically at its resonant frequency. This is the cause of the "ring" we often see after the leading
edge of pulses in measured waveforms. This effect is referred to as ground lead corruption.
Because it is impossible to eliminate either the L or C from this circuit, the method to improve
waveform fidelity is to raise the resonant frequency beyond the bandwidth of interest in the measurement.
The resonant frequency of a simple LC circuit can be represented by:
1
F
Resonance
The resonant frequency of a series LC circuit can be raised by decreasing the inductance, capacitance or both.
Since the input capacitance is already very low and cannot be reduced, you can only try to reduce
the inductance. This can be accomplished by using the shortest possible input lead as well as the
shortest possible ground lead.
For example, to obtain the shortest possible ground lead when measuring IC related signals,
attach a small piece of copper clad material to the top of the IC package and connect this to the
HFP1000-OM-E Rev E5–1
------------------
=
2π LC
Page 22
HFP1000 High Frequency Probe
package grounding wires. Using the shortest ground lead and input lead available makes probing
signals on the package easier and makes for the shortest lead length for the best signal fidelity.
To illustrate how dramatic this effect is, we will work a simple example.
Assuming an input capacitance of 0.7 pF and a total lead length (input and ground) of 2 inches
(inductance of ≈ 25 nH/inch) such a setup may cause ringing with a resonant frequency (f
This frequency is well within the passband of the probe and will therefore show up as part of the
measured signal at faster time/div settings.
To determine how fast a waveform to be measured can be without causing ringing on a probe like,
this divide the BW (ringing frequency) of the probe into 0.35:
0.35
t
rise
Any input signal with a rise time faster than 0.4 ns can cause ringing.
CAPACITIVE LOADING
Capacitive loading is usually the most troublesome of the three loading effects.
It can affect the rise time, bandwidth and delay time measurements.
At higher frequencies the capacitive loading can affect the amplitude as well as the waveshape of
the measured signal by introducing an exponential response to the waveform.
For a simple RC network the time constant of this exponential response is:
----------
t
rise
where C
resistance.
For a setup where C
0.385 ns, which will correspond to a bandwidth of 909 MHz, assuming no inductive loads.
5–2HFP1000-OM-E Rev E
is the combined probe and circuit capacitance and R
total
= 0.7 pF and a source resistance is 250 Ω, the measured rise time will be
t
(t
=2.2 x 0.7 X 10
rise
-12
x 250 Ω = 0.385 ns)
2.2C
×R
total
×=
total
is combined circuit and probe
total
Page 23
High Frequency Measurements
(parallel combination of 250 Ω and 100 kΩ is still 250 Ω)
Input inductance
250 Ω
0.7 pF100 kΩ
Source
Figure 5-2. Probe input equivalent circuit
To illustrate the effect of capacitive loading at higher frequencies:
At a frequency of 851 MHz the reactance of the 0.7 pF capacitance is 267 Ω, and at 1.0 GHz the
reactance has been lowered to 227 Ω
If, at a given frequency, the source impedance is large with respect to the input impedance, a measurable reduction in the output signal amplitude may occur.
V
out
V
Z
--------------------------------------
Z
probe
+
probeZsource
Vin×=
where: Z
Z
As an example:
At 851 MHz, where the probe input impedance has reduced to 267 Ω, and a source resistance of
250 Ω the probe output amplitude is reduced to:
HFP1000-OM-E Rev E5–3
is the probe’s input impedance and
probe
is the source impedance
source
V
out
267
-----------------------267250+
###
×==
0.52V
in
Page 24
HFP1000 High Frequency Probe
BLANK PAGE
5–4HFP1000-OM-E Rev E
Page 25
Care and Maintenance
6 Care and Maintenance
CLEANING
The exterior of the probe and cable should be cleaned only using a soft cloth lightly moistened with
water or isopropyl alcohol. The use of abrasive agents, strong detergents, or other solvents may
damage the probe. Always ensure that the input leads are free of debris.
Note
The probe case is not sealed and should never be immersed in any fluid.
CALIBRATION INTERVAL
The recommended calibration interval is one year. (Performance Verification and Adjustment Procedures are included in this manual.)
SERVICE STRATEGY
The HFP1000 probe utilizes fine pitch surface mount devices. It is therefore impractical to attempt
to repair in the field. Defective probes must be returned to a LeCroy service facility for diagnosis
and exchange. A defective probe under warranty will be replaced with a factory refurbished probe.
A probe that is not under warranty can be exchanged for a factory refurbished probe for a modest
fee. You must return the defective probe in order to receive credit for the probe core.
RETURNING A DEFECTIVE PROBE
The procedure for returning a defective probe is as follows:
Contact your local LeCroy sales representative to find out where to return the product. All returned
products should be identified by model number and serial number. Provide your name and contact
number and if possible describe the defect or failure. In case of products returned to the factory, a
Return Authorization Number (RAN) should be used. The RAN can be established by contacting
your nearest LeCroy office, or the New York Customer Care Center.
Return shipment should be made prepaid. LeCroy cannot accept COD or Collect Return shipments. We recommend air-freighting. It is important that the RAN be clearly shown on the outside
of the shipping package for prompt redirection to the appropriate department.
1. Contact your local LeCroy sales or service representative to obtain a Return Authorization
Number.
2. Remove all accessories from the probe. Do not include the manual.
3. Pack the probe in its case, surrounded by the original packing material (or equivalent) and
box.
4. Label the case with a tag containing
• The RAN
HFP1000-OM-E Rev E6–1
Page 26
HFP1000 High Frequency Probe
• Name and address of the owner
• Probe model and serial number
• Description of failure
5. Package the probe case in a cardboard shipping box with adequate padding to avoid damage
in transit.
6. Mark the outside of the box with the shipping address given to you by the LeCroy representative; be sure to add the following:
• ATTN: <RAN assigned by the LeCroy representative>
• FRAGILE
7. Insure the item for the replacement cost of the probe.
8. Ship the package to the appropriate address.
REPLACEMENT PARTS
The probe connection accessories and other common parts can be ordered through the regional
customer care centers. Refer to list below for LeCroy part numbers. Defective probes can be
replaced on an exchange basis. The replacement exchange probe will have been factory repaired,
inspected and calibrated to the same standards as a new product. In order to obtain an exchange
probe, you must return the defective probe. The returned probe should be sent back to the
regional customer care center without any accessories, manual or case.
6–2HFP1000-OM-E Rev E
Page 27
HFP1000
Care and Maintenance
PACC-MS002
PACC-MS003 (HF)
Drawing Not To Scale
Figure 6-1. Replaceable Parts Selected by Input and Ground Accessory
(Some accessories may be optional and not included with the probe)
PACC-PT001
PACC-PT002
PACC-PT003
PACC-PT004
PACC-PT005
PACC-CL001
PACC-LD005 (short)
PACC-LD006 (long)
Table 6-1. Replaceable Parts List
ItemLeCroy P/N
405400003
PACC-LD001
PACC-LD002
PACC-LD003 (short)
PACC-LD004 (long)
PK006-4
Replacement
Quantity
Straight TipPACC-PT0014
Sharp TipPACC-PT0024
IC Lead TipPACC-PT0034
SMD Discrete TipPACC-PT0044
Bent Sharp TipPACC-PT0054
Micro ClipPACC-CL0014
ClipPK006-42
Ground Spring with HookPACC-LD0014
HFP1000-OM-E Rev E6–3
Page 28
HFP1000 High Frequency Probe
ItemLeCroy P/N
Square Pin Ground SpringPACC-LD0024
Offset Pin4054000031
Short Right Angle LeadPACC-LD0034
Long Right Angle LeadPACC-LD0044
Short Single LeadPACC-LD0054
Long Single LeadPACC-LD0064
FreeHand Probe HolderPACC-MS0011
Replaceable CartridgePACC-MS0021
Low C CartridgePACC-MS0031
Soft Accessory caseSAC-011
Instruction ManualHFP1500-OM-E1
Replacement
Quantity
###
6–4HFP1000-OM-E Rev E
Page 29
Performance Verification
7 Performance Verification
This procedure can be used to verify the warranted characteristics of the HFP1000 High Frequency Probe.
The recommended calibration interval for the model HFP1000 is one year. The complete performance verification procedure should be performed as the first step of annual calibration. Test
results can be recorded on a photocopy of the Test Record provided in Appendix A at the end of
the manual.
Performance verification can be completed without removing the probe covers or exposing the
user to hazardous voltages. Adjustment should only be attempted if a parameter measured in the
Performance Verification Procedure is outside the specification limits.
Note:
Adjustment should only be performed by qualified personnel.
This procedure tests the following specifications:
• Output Zero Voltage
• LF Attenuation Accuracy
TEST EQUIPMENT REQUIRED
Table 7-1 lists the test equipment and accessories (or their equivalents) that are required for performance verification of the HFP1000 Probe.
This procedure has been developed to minimize the number of calibrated test instruments
required.
Only the parameters listed in boldface in the "Minimum requirements" column must be calibrated
to the accuracy indicated.
Because the input and output connectors types may vary on different brands and models of test
instruments, additional adapters or cables may be required.
BNC to Mini-grabberBNC Male to Mini-grabber Cable, 36" Pomona 5187-C-36
PRELIMINARY PROCEDURE
1. Connect the HFP1000 probe to the female end of the ProBus Extension Cable. Connect the
male end of the ProBus Extension Cable to channel 1 of the oscilloscope.
2. Turn the oscilloscope on and allow at least 30 minutes warm-up time for the HFP1000 and test
equipment before performing the Verification Procedure.
3. Turn on the other test equipment and allow these to warm up for the time recommended by the
manufacturer.
4. While the instruments are reaching operating temperature, make a photocopy of the Performance Verification Test Record (located in Appendix A), and fill in the necessary data.
5. Select the channel to which the probe is connected. Set the oscilloscope scale factor to 20
mV/div.
6. Disconnect the ProBus Extender Cable from the oscilloscope. Verify that the scale factor
changes from 20 mV/div to 2 mV/div.
7. Re-connect the ProBus extender Cable to the oscilloscope.
The warranted characteristics of the HFP1000 are valid at any temperature within the Environmental Characteristics listed in the Specifications. However, some of the other test equipment used to
verify the performance may have environmental limitations required to meet the accuracy needed
for the procedure. Make sure that the ambient conditions meet the requirements of all the test
instruments used in his procedure.
Note
The correct operation of the HFP1000 controls requires software version 8.7.0 or
higher. The software version in the test oscilloscope can be verified by pushing
SCOPE STATUS, then selecting the System menu option.
Contact your local LeCroy representative if the software in your oscilloscope
requires updating.
7–2HFP1000-OM-E Rev E
Page 31
Performance Verification
FUNCTIONAL CHECK
The functional check will verify the basic operation of the probe functions.
It is recommended the Functional Check be performed prior to the Performance Verification Procedure.
1. Return to the factory default settings by:
a. Pressing the oscilloscope’s front panel PANELS button.
b. From the Menu buttons press FROM DEFAULT SETUP.
3. Select Channel 1 and enter the Coupling menu.
4. Verify that Probe sensed (HFP1000) is displayed on the right hand menu.
5. If the trace colors have been reassigned or you are unsure, restore the default colors by pressing the following menus: DISPLAY, More Display Setup, Color Scheme and in the Color
Scheme menu press 1.
6. Verify that the probe head LED shows basically the same color as the channel 1 trace color.
7. Disconnect the probe from channel 1 and connect respectively to channel 2, 3 and 4.
8. Verify that in each case the LED color corresponds to the trace color of the channel to which
the probe is connected.
PROCEDURE
A. Output Zero Voltage
1. Connect one end of a BNC cable to the female BNC connector on the probe end of the ProBus
extender cable. Connect the precision 50 Ω terminator to the other end of the BNC cable.
2. Connect the banana plugs of the Precision terminator to the input of the DMM. Make sure that
the plug corresponding to the BNC shield (marked "Ground") is connected to the LO or COM-MON input of the DMM. Refer to figure 7-1 for setup information.
HFP1000-OM-E Rev E7–3
Page 32
HFP1000 High Frequency Probe
LeCroy
Oscilloscope
ProBus Extender Cable
HFP1000 Probe
HFP2500
DMM
Precision 50Ω
Terminator
Figure 7-1. Output Zero Voltage Test Setup
3. Set the OFFSET on the oscilloscope to zero, as indicated by on-screen display.
4. Set the DMM to read DC Volt on the most sensitive range.
5. Record the voltage measured on the DMM to 10 µV resolution as ’Output Zero Voltage’ in the
Test record.
6. Check that the voltage indicated by the DMM is between ±800 µV.
7. Disconnect the DMM from the precision 50 Ω terminator. Leave the remaining setup in place
for the next step.
7–4HFP1000-OM-E Rev E
Page 33
Performance Verification
B. LF Attenuation Accuracy
1. Connect the BNC end of the BNC to mini-grabber cable to a female end of the BNC tee
adapter. (Refer to figure 7-2)
Function Generator
LeCroy Oscilloscope
ProBus Extender Cable
DMM
HFP1000 Probe
HFP2500HFP2500
Precision 50Ω
Terminator
Figure 7-2. LF Attenuation Accuracy Test Setup
2. Carefully insert the Straight Tips (supplied in accessory kit) into the sockets of the probe head.
Attach the red lead of the mini-grabber to the signal input and the black lead to the ground
input of the probe head.
3. Connect the BNC tee to the output of the function generator. (Use a 50 Ω termination if the
function generator requires such a load).
4. Attach a BNC cable to the unused female port of the BNC tee and connect a dual banana plug
adapter to the other end of the cable and plug the dual banana plug adapter into the DMM
input. Make sure the side of the banana plug adapter corresponding to the BNC shield
(marked "GROUND") is connected to the LOW or COMMON input of the DMM.
5. Set the DMM to read AC volt and set the range to measure 5.0 Vrms.
6. Set the mode of the function generator to sine wave, the frequency to 70 Hz and the output
amplitude to 5 Vrms ±10 mV as measured on the DMM.
HFP1000-OM-E Rev E7–5
Page 34
HFP1000 High Frequency Probe
7. Record the output voltage to 1 mV resolution as ’Generator Output Voltage’ in the Test Record.
Be careful not to alter the output amplitude after the reading is recorded.
8. Divide the reading recorded in step B-7 by 10 and record the result with 100 µV resolution as
’Expected Output Voltage, top range" in the Test Record.
9. Remove the banana plug adapter, connected to the function generator, from the DMM and
connect the precision 50 Ω terminator to the DMM, making sure that the banana plug side
marked ’GROUND" is connected to the LOW or COMMON input of the DMM.
10. After the DMM reading has stabilized, record the reading to 100 µV resolution as ’Measured
Output Voltage, top range’ in the Test Record.
11. Calculate the error by dividing the measured top range output voltage recorded in step B10 by
the expected top range output voltage recorded in step B-8. Subtract 1 from this ratio and multiply by 100% to get the error in percent.
Measured Output Voltage
⎛⎞
Error
12. Record the calculated error to two decimal places (±0.xx%) as ’Gain Error, top range’ in the
test record.
14. Disconnect the precision 50 Ω terminator from the DMM.
15. Connect the banana plug adapter connected via a BNC cable to the BNC tee at the function
generator to the DMM. Verify that the side of the plug marked ’Ground’ is connected to the
LOW or COMMON input of the DMM.
16. Adjust the sine wave generator output amplitude to approximately 2.5 Vrms as measured on
the DMM.
17. Record the reading to 1 mV resolution as ’Generator Output Voltage, mid range’ in the Test
Record. Be careful not to alter the output amplitude after the reading is recorded.
18. Divide the reading recorded in step B-17 by 10.
19. Record the result to 100 µV resolution as ’Expected Output Voltage, mid range’ in the test
record.
20. Remove the banana plug adapter from the DMM and connect the precision 50 Ω terminator to
the DMM, making sure that the banana plug side marked ’GROUND" is connected to the LOW
or COMMON input of the DMM.
21. After the DMM has stabilized, record the reading to 100 µV resolution as ’Measured Output
Voltage, mid range’ in the Test record.
22. Calculate the error by dividing the measured mid range output voltage recorded in step B-21
7–6HFP1000-OM-E Rev E
Page 35
Performance Verification
by the expected mid range output voltage recorded in step B-19. Subtract 1 from this ratio and
multiply by 100% to get the error in percent.
Measured Output Voltage
⎛⎞
Error
23. Record the calculated error to two decimal places (±0.xx %) as ’Gain Error, mid range’ in the
Test record.
24. Verify that the mid range gain error is less than ±1.0%
This completes the Performance Verification of the HFP1000. Complete and file the Test Record,
as required to support your internal calibration procedure.
Apply suitable calibration label to the HFP1000 housing as required.
You can use this procedure to adjust the HFP1000 probe to meet the warranted specifications.
This procedure should only be performed if the probe fails to meet the Performance verification
tests for Output Zero.
Gain which affects LF attenuation accuracy cannot be adjusted during routine calibration. Probes
which fail LF attenuation accuracy during performance verification must be returned to the factory
for rework.
If the probe cannot be adjusted to meet the Performance verification limits, repair may be necessary.
To assure probe accuracy, check the calibration of the HFP1000 every 1000 hours or once a year
if used infrequently. Before calibration, thoroughly clean and inspect the probe as outlined in the
Care and Maintenance section.
To assure the probe will meet the published specifications over the entire temperature range,
adjustment must be performed in a controlled ambient environment with temperature of
23 °C ±5 °C.
Caution
The adjustment procedure will require removal of the probe control circuitry cover.
This cover is part of the ESD protection system of the HFP1000. To protect the
probe, you should perform the entire procedure on a static dissipating work surface. Wear an antistatic wrist strap and follow standard static control procedures.
TEST EQUIPMENT REQUIRED
Table 8-1 lists the test equipment and accessories (or their equivalents) that are required for complete calibration of the HFP1000 Probe. Specifications given for the test equipment are the minimum necessary for accurate calibration. All test equipment is assumed to be correctly calibrated
and operating within the specification listed. Detailed operating instructions for the test equipment
are not given in this procedure. refer to the test equipment manual if more information is needed
If alternate test equipment is substituted, control settings or calibration equipment setups may
need to be altered.
1. Remove the two screws that secure the plastic cover on the cable end of the ProBus interface
housing.
2. Gently pull on the probe cable to slide the circuit board assembly from the metal housing.
3. Connect the HFP1000 probe to the female end of the ProBus extension cable, being careful to
line up all six pins of the probe connector. Connect the male end of the ProBus extension
cable to channel 1 of the oscilloscope.
4. Apply power to the oscilloscope and test equipment.
5. Allow at least 30 minutes warm-up time for the HFP1000 and test equipment before starting
the calibration procedure.
PROCEDURE
A. Adjust Output Zero
1. Connect one end of a BNC cable to the probe end of the ProBus extension cable. Connect the
Precision 50 Ω Terminator to the other end of the BNC cable.
2. Connect the banana plugs of the precision 50 Ω terminator to the input of the DMM. Make sure
the plug corresponding to the BNC shield (marked ’Ground’) is connected to the LO or COM-
MON input of the DMM. Refer to figure 8-1 for setup information.
8–2HFP1000-OM-E Rev E
Page 39
Adjustment Procedure
LeCroy
Oscilloscope
ProBus Extender Cable
HFP1000 Probe
HFP2500
DMM
Precision 50Ω
Terminator
Figure 8-1. Output Zero Voltage Adjustment Setup
3. Select the channel to which the probe and ProBus extender is connected. Set OFFSET on the
oscilloscope to zero as indicated on the on-screen display.
4. Set the DMM to read DC Volt on the most sensitive range.
5. Verify that the probe inputs are not connected to any signal.
6. Adjust OUTPUT ZERO on the board until the DMM reads 0 V ±100 µV. Refer to figure 8-2 for
adjustment location.
OUTPUT ZEROOUTPUT ZERO
Figure 8-2a. Adjustment Location S/N 100 and higher.
HFP1000-OM-E Rev E8–3
Page 40
HFP1000 High Frequency Probe
Output Zero
Figure 8-2b. Adjustment Location S/N below 1000.
7. Disconnect the probe from the ProBus extender and re-install the circuit board into the probe
case, being careful to align the ProBus interface connector with the opening on the other end
of the probe.
B. Verify Calibration
Repeat the Performance Verification procedure to ensure compliance with the warranted specifications.
Apply a calibration sticker, if required, in accordance with your quality control procedures.
###
8–4HFP1000-OM-E Rev E
Page 41
Specifications
9 Specifications
NOMINAL CHARACTERISTICS
Nominal characteristics describe parameters and attributes that are guaranteed by design, but do
not have associated tolerances.
Input Dynamic range±8 V
Maximum Input Voltage40 V
Intended Output Load50 Ω
Attenuation÷10
Output ConnectorProBus
InterfaceProBus
Oscilloscope Full Compatibility LeCroy oscilloscope with firm-
1
Subject to input voltage vs. frequency derating. See figure 9-1.
WARRANTED CHARACTERISTICS
1
ware version 8.7.0 or higher
Warranted characteristics are parameters with guaranteed performance. Unless otherwise noted,
tests are provided in the Performance Verification Procedure for all warranted specifications.
Low Frequency Attenuation±1% plus uncertainty of 50 Ω
Accuracytermination
Output Zero≤ 8 mV referred to the input
TYPICAL CHARACTERISTICS
Typical characteristics are parameters with no guaranteed performance. Tests for typical characteristics are not provided in the Performance Verification Procedure.
Output Zero≤ 4 mV, referred to the input
Bandwidth (Probe only)1 GHz
Input Capacitance0.7 pF
DC Input Resistance100 kΩ
ENVIRONMENTAL CHARACTERISTICS
Temperature, operating0 °C to 50 °C
Temperature, non-operating-40 °C to 71 °C
HFP1000-OM-E Rev E9–1
Page 42
HFP1000 High Frequency Probe
Relative Humidity80% max up to 31 °C, decreas-
Altitudeup to 2000 m
PHYSICAL CHARACTERISTICS
Probe Head Size:
Length61 mm (2.4 in)
Width7.3 mm (0.29 in)
Height13.1 mm (0.52 in)
Cable Length1.3 m (51.1 in)
Weight:
Probe only100 g (3.5 oz.)
Shipping0.85 kg (1.87 labs)
Input SocketsSignal and ground sockets are
ing linearly to 45% max at 50 °C
compatible with 0.635 mm
(0.025 in) square pins, and
0.91 mm (0.036 in) maximum
diameter round pins
100
Maximum
1 GHz
Non-destruct
Range
10 GHz
Volts
Pk-Pk
40
16
10
1
1 MHz
10 MHz
Maximum
Operating
Range
100 MHz
Frequency
Figure 9-1. Input Voltage vs. Frequency
9–2HFP1000-OM-E Rev E
Page 43
COMPLIANCE AND CERTIFICATIONS
CE Declaration of Conformity
Low Voltage Directive: EN 61010-2-031:2002
Safety requirements for electrical equipment for measurement control and laboratory use.
Part 031:
Particular requirements for hand-held probe assemblies
for electrical measurement and test.
EMC Directive:EN 61326/A3:2003
EMC requirements for electrical equipment for measurement control and laboratory use.
Electromagnetic Emissions: EN 55011/A2:2002 Class A Radiated Emissions.
Electromagnetic Immunity: EN 61000-4-2/A2:2001Electrostatic Discharge.
(Air/Contact Discharge: 8 kV/4 kV)
Specifications
EN 61000-4-3/A1:2003RF Radiated Electromagnetic
Field (80 MHz to 1 GHz; 3 V/m)
HFP1000-OM-E Rev E9–3
Page 44
HFP1000 High Frequency Probe
Toxic or Hazardous Substances and Elements
)
Polybrominated
Biphenyls
(PBB)
Part NameLead
(Pb)
PCBAsXOXXXX
Mechanical HardwareOOXOOO
Sheet MetalOOXOOO
Plastic PartsOOOOXX
Protective CaseOOOOXX
Cable AssembliesXOXOXX
Probe TipsXOXOXX
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit
requirement specified in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the homogenous materials used for this part is
above the limit requirement specified in SJ/T11363-2006.
Mercury
(Hg)
Cadmium
(Cd)
Hexavalent
Chromium
6+
(Cr
Polybrominated
Diphenyl Ethers
(PBDE)
EFUP (Environmental Friendly Use Period) Use Conditions: refer to the environmental conditions stated in the Specifications section of this manual.
###
9–4HFP1000-OM-E Rev E
Page 45
Appendix A
Appendix A
PERFORMANCE VERIFICATION TEST RECORD
This record can be used to record the results of measurements made during the performance verification of the HFP1000 High Frequency Probe.
Photocopy this page and record the results on the copy. File the completed record as required by
applicable internal quality procedures.
The section in the test record corresponds to the parameters tested in the performance verification
procedure. The numbers preceding the individual data records correspond to the steps in the procedure that require the recording of data. Results to be recorded in the column labeled "Test
Result" are the actual specification limit check. The test limits are included in all of these steps.
Other measurements and the results of intermediate calculations that support the limit check are to
be recorded in the column labeled "Intermediate Results".
Permission is granted to reproduce these pages for the purpose of recording test results.
Probe Model: HFP1000
Serial Number:_______________________________________
Asset or Tracking Number:_______________________________________
The function generator used in this Performance Verification Procedure is used for making relative measurements. The
output of the generator is measured with a DMM or oscilloscope in this procedure. Thus, the generator is not required to
be calibrated.
HFP1000-OM-E Rev EA–1
1
CALIBRATION
DUE DATE
N/A
Page 46
HFP1000 High Frequency Probe
HFP1000 TEST RECORD
StepDescriptionIntermediate dataTest Result
Output Zero Voltage
A-5 Output Zero Voltage (Test limit ≤± 800 µV)_____________ V
LF Attenuation Accuracy
B-7 Generator Output Voltage________________ V
B-8 Expected Output Voltage, top range________________ V
B-10 Measured Output Voltage, top range________________ V
B-12 Gain Error, top range (Test limit ≤± 1.0%)_____________ %
B-17 Generator Output Voltage________________ V
B-19 Expected Output Voltage, mid range________________ V
B-21 Measured Output Voltage, mid range________________ V
B-23 Gain Error, mid range (Test limit ≤± 1.0%)_____________ %
###