LEADER LV 58SER07 Instruction Manual

Page 1
LV 58SER07
EYE PATTERN
INSTRUCTION MANUAL
Page 2

TABLE OF CONTENTS

1. INTRODUCTION ........................................................................................................ 1
1.1 Scope of Warranty.....................................................................................................................1
1.2 Operating Precautions...............................................................................................................1
1.2.1 Input Connectors................................................................................................................1
1.2.2 Output Connectors.............................................................................................................1
1.2.4 Included Cables .................................................................................................................2
1.3 Notations Used in This Manual .................................................................................................2
2. SPECIFICATIONS ...................................................................................................... 3
2.1 General......................................................................................................................................3
2.2 Features.....................................................................................................................................3
2.3 Specifications.............................................................................................................................4
2.3.1 Supported Formats ............................................................................................................4
2.3.2 Input Connectors................................................................................................................4
2.3.3 Output Connectors.............................................................................................................4
2.3.4 Installation Position and Unit to Combine With..................................................................4
2.3.5 Eye-Pattern Display ...........................................................................................................5
2.3.6 Automatic Eye-Pattern Measurement................................................................................5
2.3.7 Jitter Display.......................................................................................................................6
2.3.8 Automatic Jitter Measurement ...........................................................................................6
2.3.9 Error Detection ...................................................................................................................7
2.3.10 General Specifications .......................................................................................................7
3. NAMES AND FUNCTIONS OF PARTS ...................................................................... 8
4. MEASUREMENT BASICS.......................................................................................... 9
4.1 Unit Installation..........................................................................................................................9
4.2 Applying the SDI Signal.............................................................................................................9
4.3 Measurement Procedure.........................................................................................................10
4.4 Multi Screen Display................................................................................................................ 11
5. EYE-PATTERN AND JITTER DISPLAYS ................................................................. 12
5.1 Setting the Waveform Display Position...................................................................................12
5.2 Adjusting the Waveform Intensity............................................................................................12
5.3 Adjusting the Scale Intensity...................................................................................................12
5.4 Selecting the Display Mode.....................................................................................................13
5.5 Selecting Which Link to Display ..............................................................................................13
5.6 Event Log.................................................................................................................................14
6. EYE-PATTERN DISPLAY......................................................................................... 15
6.1 Turning Automatic Measurement On and Off..........................................................................15
Page 3
6.2 Adjusting the Gain ...................................................................................................................18
6.3 Selecting the Sweep Time.......................................................................................................18
6.4 Selecting the Filter...................................................................................................................19
6.5 Configuring Cursor Settings ....................................................................................................20
6.5.1 Turning Cursors On and Off.............................................................................................20
6.5.2 Selecting the Cursors.......................................................................................................21
6.5.3 Selecting the X-Axis Measurement Unit ..........................................................................22
6.5.4 Selecting the Y-Axis Measurement Unit...........................................................................22
6.5.5 Moving Cursors ................................................................................................................23
7. JITTER DISPLAY ..................................................................................................... 24
7.1 Configuring the Peak Hold Settings ........................................................................................25
7.1.1 Turning the Peak Hold On and Off...................................................................................25
7.1.2 Clearing the Peak Hold....................................................................................................26
7.2 Selecting the Gain ...................................................................................................................26
7.3 Selecting the Sweep Time.......................................................................................................28
7.4 Selecting the Filter...................................................................................................................28
7.5 Configuring Cursor Settings ....................................................................................................28
7.5.1 Turning Cursors On and Off.............................................................................................29
7.5.2 Selecting the Cursors.......................................................................................................29
7.5.3 Selecting the X-Axis Measurement Unit ..........................................................................29
7.5.4 Selecting the Y-Axis Measurement Unit...........................................................................30
7.5.5 Moving Cursors ................................................................................................................30
8. UNIT SETUP ............................................................................................................ 31
8.1 Operating Tab Menus ..............................................................................................................31
8.2 Configuring 3G-SDI Error Settings..........................................................................................32
8.3 Configuring HD-SDI Error Settings .........................................................................................34
8.4 Configuring SD-SDI Error Settings..........................................................................................36
9. REFERENCE MATERIAL......................................................................................... 38
9.1 Menu Tree................................................................................................................................38
9.2 Change History of the Software ..............................................................................................39
Index
Page 4
1. INTRODUCTION
Thank you for purchasing this LEADER instrument. To use this instrument safely, read this instruction manual thoroughly, and make sure that you know how to use the instrument properly.
If some point about the operation of this instrument is still unclear after you have read this instruction manual, refer to the contact information on the back cover of the manual to contact LEADER, or contact your local LEADER agent. After you have finished reading this manual, keep it in a convenient place so that you can refer to it when necessary.

1.1 Scope of Warranty

This LEADER instrument has been manufactured under the strictest quality control guidelines.
LEADER shall not be obligated to furnish the following free services during the warranty period.

1. INTRODUCTION

1. Repair of malfunction or damages resulting from fire, natural calamity, or improper voltage applied by the user.
2. Repair of an instrument that has been improperly repaired, adjusted, or modified by personnel other than a factory-trained LEADER representative.
3. Repair of malfunctions or damages resulting from improper use.
4. Repair of malfunctions caused by devices other than this instrument.
5. Repair of malfunctions or damages without the presentation of a proof of purchase or receipt bill for the instrument.

1.2 Operating Precautions

1.2.1 Input Connectors

The maximum signal voltage that can be received through the input connectors is indicated below. Do not apply excessive voltage to the connectors. Doing so may damage the instrument or lead to injury.
Table 1-1 Maximum allowable input voltage
Input Connector Maximum Allowable Input Voltage
INPUT ±2 V (DC + peak AC)

1.2.2 Output Connectors

The output connectors are designed to be connected to the LV 58SER06 input connectors. Connecting the output connectors to a different instrument or shorting them may damage the unit or the instrument that is connected to the unit.
1
Page 5

1.2.3 Warming Up

To ensure more accurate measurements, turn on the instrument approximately 30 minutes before you intend to use it to allow its internal temperature to stabilize.

1.2.4 Included Cables

The included cables are designed to connect this unit to an LV 58SER06. Do not use them for any other purpose.

1.3 Notations Used in This Manual

The key and other operations explained in this manual apply to the LV 5800, but you can also perform similar operations on the LV 7800.
In this manual, abbreviations are used for procedures as shown below. Colons are followed by a list of options or a range of settings. The default setting is underlined.
Procedure
EYE → F・2 MODE: EYE / JITTER
1. INTRODUCTION
2
Page 6
2. SPECIFICATIONS

2.1 General

Installing this unit in an LV 5800 or LV 7800 and using it together with an LV 58SER06 enables the measurement and observation of the eye pattern and jitter of SDI signals. It enables the measurement and observation of the physical characteristics of not only 3G-SDI signals but also HD-SDI and SD-SDI signals.

2.2 Features

Support for 3G-SDI, HD-SDI, and SD-SDI Signals
Combining this unit with an LV 58SER06 enables the LV 5800 or the LV 7800 to display eye patterns and jitter and perform automatic measurements for not only 3G-SDI signals (level A and B), but also for HD-SDI and SD-SDI signals.
Two Switchable SDI Inputs
The signal that is displayed on the screen can be switched from the front panel between the two input connectors, which support 3G-SDI, HD-SDI, and SD-SDI signals.

2. SPECIFICATIONS

Eye-Pattern Display
Measurements of 3G-SDI signals have low noise and wide bandwidth characteristics thanks to the use of a new kind of circuit. The eye-pattern displays of timing jitter and alignment jitter can be displayed through the changing of the filter.
Jitter Display
Because a phase detection method is used, accurate jitter measurements can be performed even on degraded signals for which eye patterns would not be useful. Also, V rate and H rate sweep displays synchronized to the video signal are useful for analyzing jitter that originates in digital video data.
Simultaneous Eye-Pattern and Jitter Display
The eye-pattern and jitter waveforms of the selected SDI signal can be displayed simultaneously on the multi screen display.(*1)
Filter Settings
The measurement of the timing jitter and alignment jitter of an SDI signal can be performed through the switching of filters in the eye-pattern and jitter displays.
Automatic Measurement
An eye pattern’s amplitude, rise time, fall time, timing jitter, current jitter(*2), overshoot of the rising edge, and overshoot of the falling edge can be measured automatically.
Error Detection
An eye pattern’s amplitude, rise time, fall time, time difference between rise and fall times, timing jitter, current jitter, overshoot of the rising edge, and overshoot of the falling edge can be monitored, and when errors (values passing their thresholds) are detected, they can be displayed and logged to the event log screen(*3). The threshold values can be changed.
3
Page 7
*1 Signals applied to different input connectors cannot be displayed simultaneously. *2 The current jitter is the jitter value when the currently selected filter is applied. *3 The event log display is enabled when one of the display areas is set to EYE mode. Error detection
cannot be performed in the background.
* The LV 58SER07 is used in combination with a single LV 58SER06. It cannot be used in combination
with the LV 58SER01A or the LV 58SER04. In addition, multiple LV 58SER07s cannot be installed in the same instrument, nor can the LV 58SER07 be installed in the same instrument with the LV 58SER02.

2.3 Specifications

2.3.1 Supported Formats

Standard
3G-SDI SMPTE 424M HD-SDI SMPTE 292M SD-SDI SMPTE 259M
Bit rate
3G-SDI 2.970 Gbps or 2.970/1.001 Gbps HD-SDI 1.485 Gbps or 1.485/1.001 Gbps SD-SDI 270Mbps
2. SPECIFICATIONS

2.3.2 Input Connectors

Function Input of SDI signals for eye-pattern and jitter display Input Connectors 2 BNC connectors
1 channel selectable from channels A and B (When in HD dual link mode, 1 link selectable from
links A and B) Input Impedance 75 Ω Input Return Loss 15 dB or greater (5 MHz to 1.485 GHz)
10 dB or greater (1.485 to 2.97 GHz) Maximum Input Voltage ±2V (DC + peak AC) Supported Signals 3G-SDI / HD-SDI / SD-SDI

2.3.3 Output Connectors

Function Dedicated output connectors for connecting to the
LV 58SER06 INPUT connectors Output Connectors 2 BNC connectors Output Impedance 75 Ω

2.3.4 Installation Position and Unit to Combine With

Installation Position
LV 5800 Input slot (UNIT 1 to 4) LV 7800 Input slot (UNIT 1) or I/O slot (UNIT 3)
Unit to Combine With LV 58SER06
4
Page 8

2.3.5 Eye-Pattern Display

Display Displays the input SDI waveform before equalizing Method Equivalent time sampling Frequency Response 7 GHZ -3dB (converted from the rise time) Amplitude Accuracy 800 mV ± 5 % (for 800 mV input) Time Axis
2 UI Display
3G-SDI 50 ps/div HD-SDI 100 ps/div SD-SDI 550 ps/div
4 UI Display
3G-SDI 100 ps/div HD-SDI 200 ps/div SD-SDI 1100 ps/div
16 UI Display
3G-SDI 400 ps/div HD-SDI 800 ps/div
SD-SDI 4400 ps/div Time Axis Accuracy ±3 % Jitter Filter
10 Hz HPF 10 Hz 100 Hz HPF 100 Hz 1 kHz HPF 1 kHz 100 kHz HPF 100 kHz TIMING HPF 10 Hz ALIGNMENT
3G-SDI, HD-SDI HPF 100 kHz
SD-SDI HPF 1 kHz Cursor Measurement Amplitude measurement using Y cursors
2. SPECIFICATIONS
Time measurement using X cursors Rise time and fall time measurement using the Tr and Tf cursors

2.3.6 Automatic Eye-Pattern Measurement

Automatic Measurement On or off Measurable Items Eye pattern’s amplitude
Rise time (the time for the signal to rise from 20 to 80 % of its amplitude) Fall time (the time for the signal to fall from 80 to 20 % of its amplitude) Timing jitter Current jitter Overshoot of the rising edge Overshoot of the falling edge
5
Page 9

2.3.7 Jitter Display

Display Displays the jitter component of an SDI signal Method Phase detection method Gain ×8 / ×2 / ×1 Measurement Range
×8 0.00 to 1.20 UI ×2 1.20 to 4.80 UI ×1 4.80 to 9.60 UI
Frequency Response
SD-SDI 600 kHz or more (with 0.2 UI input) HD-SDI 2 MHz or more (with 0.2 UI input)
3G-SDI 2 MHz or more (with 0.3 UI input) Time Axis 1H / 2H / 1V / 2V Time Axis Accuracy ±3 % Jitter Filter
10 Hz HPF 10 Hz
100 Hz HPF 100 Hz
1 kHz HPF 1 kHz
100 kHz HPF 100 kHz
TIMING HPF 10 Hz
ALIGNMENT
3G-SDI, HD-SDI HPF 100 kHz SD-SDI HPF 1 kHz
Cursor Measurement Jitter value measurement through the use of cursors
2. SPECIFICATIONS

2.3.8 Automatic Jitter Measurement

Automatic Measurement Displays the jitter value in seconds (sec) and unit
Measurable Items Timing jitter, current jitter Measurement Range 0.00 to 9.60 UI (differs depending on the gain) Accuracy Input jitter frequency: 1 kHz. Filter setting: 10 Hz,
0 UI < automatic measured value 1 UI
1 UI < automatic measured value 7 UI
intervals (UI)
within measurement range
±10 % + 0.05 UI
±10 %
6
Page 10

2.3.9 Error Detection

Error Detection On or off per item Error Threshold Settings Settable separately for 3G-SDI, HD-SDI, and
Event Log Events can only be logged when eye patterns or
Threshold Values 100 % of the values in the SMPTE standard
Eye-Pattern Amplitude
Upper Limit 80 to 140 % (640 to 1120 mV) Lower Limit 40 to 100 % (320 to 800 mV)
Rise Time
3G-SDI 40 to 140 % (54.0 to 189.0 ps) HD-SDI 40 to 140 % (108.0 to 378.0 ps) SD-SDI 40 to 140 % (0.60 to 2.10 ns)
Fall Time
3G-SDI 40 to 140 % (54.0 to 189.0 ps) HD-SDI 40 to 140 % (108.0 to 378.0 ps) SD-SDI 40 to 140 % (0.60 to 2.10 ns)
Difference between the Rise and Fall Times
3G-SDI 40 to 140 % (20 to 70 ps) HD-SDI 40 to 140 % (40 to 140 ps) SD-SDI 40 to 140 % (0.20 to 0.70 ns)
Timing Jitter
3G-SDI 10 to 200 % (0.20 to 4.00 UI, 67.4 to 1348.0 ps) HD-SDI 10 to 200 % (0.10 to 2.00 UI, 67.4 to 1348.0 ps) SD-SDI 10 to 200 % (0.02 to 0.40 UI, 0.07 to 1.48 ns)
Current Jitter
3G-SDI 10 to 200 % (0.03 to 0.60 UI, 10.1 to 202.5 ps) HD-SDI 10 to 200 % (0.02 to 0.40 UI, 13.5 to 270.0 ps) SD-SDI 10 to 200 % (0.02 to 0.40 UI, 0.07 to 1.48 ns)
Overshoot of the rising edge
Overshoot of the falling edge
2. SPECIFICATIONS
SD-SDI
jitter are being displayed.
0 to 200 % (0.0 to 20.0 %)
0 to 200 % (0.0 to 20.0 %)

2.3.10 General Specifications

Environmental Conditions Same as the LV 5800/7800 Power Consumption 23 W max. from the LV 5800/7800 Weight 0.5 kg
Accessories Cables (designed for connecting to an LV 58SER06).................2
Instruction manual ....................................................................... 1
Precautions about installing an LV 58SER07 in an LV 7800 ...... 1
7
Page 11

3. NAMES AND FUNCTIONS OF PARTS

3. NAMES AND FUNCTIONS OF PARTS
Figure 3-1 Rear panel
1 INPUT
SDI signal input connectors. The signals that are applied to INPUT A and INPUT B cannot be displayed simultaneously. Switch between these signals to show the signal that you want to display.
2 OUTPUT
Use the included cables to connect these outputs to the input connectors on an LV 58SER06.
The output connectors are designed to be connected to the LV 58SER06 input connectors. Connecting the output connectors to a different instrument may damage the unit or the instrument that is connected to the unit.
8
Page 12

4. MEASUREMENT BASICS

4. MEASUREMENT BASICS

4.1 Unit Installation

Install this unit in an LV 5800 or LV 7800. To perform measurements, an LV 58SER06 must also be installed together with this unit.
• Precautions
You can only install one LV 58SER07 in an LV 5800 or LV 7800. You cannot install
multiple LV 58SER07s.
You cannot install an LV 58SER07 and an LV 58SER02 in the same instrument.
• When Installing in an LV 5800
This unit is an input unit. Refer to the LV 5800 instruction manual, and install the unit in a slot from 1 to 4.
• When Installing in an LV 7800
Installation of units in the LV 7800 is performed as a factory option. You cannot install or uninstall units. Contact your local LEADER agent.
The LV58SER07 can be installed in UNIT 1 or UNIT 3. If the LV 58SER07 is installed in UNIT 1, the LV 58SER06 must be installed in UNIT 2. If the LV 58SER07 is installed in UNIT 3, the LV 58SER06 must be installed in UNIT 4.

4.2 Applying the SDI Signal

• Input Format
This unit supports 3G-SDI, HD-SDI (including dual-link), and SD-SDI signals. Apply signals that this unit supports to its input connectors.
• Input Pattern
Use a color bar signal to measure the amplitude of an eye pattern and jitter values.
• Cables
Because measured values are influenced greatly by the cable that is used, we recommend that you use a high-quality, low-loss 5C-FB or Belden 1694A cable. Before connecting the cables, check that the cable connectors are not dirty, deformed, or otherwise damaged.
Especially if you are measuring the amplitude of eye patterns, rise time, fall time, overshoot of the rising edge, or overshoot of the falling edge of relay instruments, use one of the cables described above (1 meter in length).
Static electricity build-up in the cables can cause damage to the input circuit. Discharge any built-up static electricity before you connect the cables.
• Connectors
This unit’s input connectors are terminated internally at 75 Ω, so there is no need to connect terminators to them. Connect cables with a characteristic impedance of 75 Ω to the input connectors.
9
Page 13

4.3 Measurement Procedure

To measure eye patterns, follow the procedure below.
4. MEASUREMENT BASICS
Figure 4-1 Measurement procedure
1. Set the Input Mode on the LV 58SER06’s GENERAL SETUP tab.
If you want to measure a 3G-SDI signal or HD dual link signal, you have to set Input
Mode to Single Input. For details, see the LV 58SER06 instruction manual.
Figure 4-2 GENERAL SETUP (LV 58SER06)
2. Apply the SDI signal to the LV 58SER07’s INPUT A or INPUT B connector.
The signals that are applied to INPUT A and INPUT B cannot be displayed
simultaneously. Switch between these signals to show the signal that you want to
display.
3. Use the included cables to connect the LV 58SER07 OUTPUT connectors to the LV 58SER06 INPUT connectors.
As shown in the figure above, connect OUTPUT A to INPUT A and OUTPUT B to INPUT B. Connect one LV 58SER07 to one LV 58SER06. Do not connect one LV 58SER07 to multiple LV 58SER06s.
4. Select the unit number of the slot that the LV 58SER06 (which is connected to the LV 58SER07) is installed in.
If you select the unit number of the slot that the LV 58SER07 is installed in, the message “This is an option board” will be displayed, and you will not be able to perform measurements.
10
Page 14
5. Press EYE.
The eye pattern will be displayed.
6. Select the input channel.
When you are applying a single-link signal, use the A and B keys to select the input channel. When you are applying a dual-link signal, use F•6 LINK SELECT to select the input channel.

4.4 Multi Screen Display

This unit has two display modes: the eye-pattern display and the jitter display. Press F•2 MODE to switch between these modes.
See section 5.4, “Selecting the Display Mode”
In addition, you can simultaneously display the eye-pattern and jitter waveforms each on their own screen on the multi screen display. (You cannot display eye-pattern or jitter waveforms in multiple areas.)
4. MEASUREMENT BASICS
Figure 4-3 Multi screen display
• Simultaneous display method
If you select EYE for two areas, an eye-pattern display will automatically be assigned to one area and a jitter display will automatically be assigned to the other. If you have already assigned an eye-pattern display to one area, a jitter display will be assigned to the second area that you select EYE for. If you have already assigned a jitter display to one area, an eye-pattern display will be assigned to the second area that you select EYE for.
• Precautions
To perform this simultaneous display, both displays have to use the same input signal. Select the same unit number and the same channel (A or B) for both of the screens.
Even if you set Input Mode on the LV 58SER06’s GENERAL SETUP tab to Simultaneous Input, you cannot simultaneously display the input from different channels on the eye-pattern display and jitter display.
11
Page 15

5. EYE-PATTERN AND JITTER DISPLAYS

5. EYE-PATTERN AND JITTER DISPLAYS
This chapter explains the settings that are common to the eye-pattern display and jitter display.

5.1 Setting the Waveform Display Position

You can use the V POS and H POS knobs on the front panel to adjust the waveform display position.
LV 5800 LV 7800
Figure 5-1 V POS and H POS knobs
• V POS knob
Adjusts the vertical position of the waveform. Pressing the knob returns the waveform to its default position.
• H POS knob
Adjusts the horizontal position of the waveform. Pressing the knob returns the waveform to its default position.

5.2 Adjusting the Waveform Intensity

To adjust the intensity of the eye-pattern waveform and jitter waveform, follow the procedure below. Turn the function dial (F•D) to perform the adjustment. Press the function dial (F•D) to return the setting to its default value (32).
Procedure
EYE → F・1 INTEN → F・1 EYE INTEN: -128 to 32 to 127

5.3 Adjusting the Scale Intensity

To adjust the scale intensity, follow the procedure below. Turn the function dial (F•D) to perform the adjustment. Press the function dial (F•D) to return the setting to its default value (4).
Procedure
EYE → F・1 INTEN → F・2 SCALE INTEN: -8 to 4 to 7
12
Page 16
5. EYE-PATTERN AND JITTER DISPLAYS

5.4 Selecting the Display Mode

To select the display mode, follow the procedure below. You can simultaneously display the eye-pattern and jitter waveforms each on their own screen on the multi screen display.
See section 4.4, “Multi Screen Display.”
Procedure
EYE → F・2 MODE: EYE / JITTER
Settings
EYE: The eye pattern display is shown. JITTER: The jitter waveform is shown.
MODE = EYE MODE = JITTER
Figure 5-2 Selecting the display mode

5.5 Selecting Which Link to Display

When an HD dual link signal is being applied, you cannot display the link A and link B signals simultaneously. To select which waveform is displayed, follow the procedure below.
See section 4.3, “Measurement Procedure.”
Procedure
EYE → F・6 LINK SELECT: LINK A / LINK B
Settings
LINK A: The link A waveform is shown. LINK B: The link B waveform is shown.
13
Page 17

5.6 Event Log

If one of the errors whose detection has been set to ON (as explained in chapter 8, “UNIT SETUP”) occurs, the error is displayed on the LV 58SER06 event log screen. The names of the events that may be generated by this unit are shown below. For details about the event log, see the LV 58SER06 instruction manual.
* You can also use other units that have an event logging feature (such as the LV 58SER01A) to check
the event log.
* The event log display is enabled when one of the display areas is set to EYE mode. Error detection
cannot be performed in the background.
Table 5-1 Events
Event Name Description
EYE_3G_AMP 3G-SDI Amplitude Error
EYE_3G_TR 3G-SDI Risetime Error
EYE_3G_TF 3G-SDI Falltime Error
EYE_3G_TR_TF 3G-SDI Deltatime Error
EYE_3G_T_JIT 3G-SDI Timing Jitter Error
EYE_3G_A_JIT 3G-SDI Current Jitter Error
EYE_3G_OR 3G-SDI OverShoot Rising Error
EYE_3G_OF 3G-SDI OverShoot Falling Error
EYE_HD_AMP HD-SDI Amplitude Error
EYE_HD_TR HD-SDI Risetime Error
EYE_HD_TF HD-SDI Falltime Error
EYE_HD_TR_TF HD-SDI Deltatime Error
EYE_HD_T_JIT HD-SDI Timing Jitter Error
EYE_HD_A_JIT HD-SDI Current Jitter Error
EYE_HD_OR HD-SDI OverShoot Rising Error
EYE_HD_OF HD-SDI OverShoot Falling Error
EYE_SD_AMP SD-SDI Amplitude Error
EYE_SD_TR SD-SDI Risetime Error
EYE_SD_TF SD-SDI Falltime Error
EYE_SD_TR_TF SD-SDI Deltatime Error
EYE_SD_T_JIT SD-SDI Timing Jitter Error
EYE_SD_A_JIT SD-SDI Current Jitter Error
EYE_SD_OR SD-SDI OverShoot Rising Error
EYE_SD_OF SD-SDI OverShoot Falling Error
5. EYE-PATTERN AND JITTER DISPLAYS
14
Page 18
6. EYE-PATTERN DISPLAY
This chapter explains the eye-pattern display. To show the eye-pattern display, set F•2 MODE to EYE.

6. EYE-PATTERN DISPLAY

Figure 6-1 Eye-pattern display screen
• Setting the eye-pattern display
To configure the eye-pattern-display settings, press F•4 EYE SETUP. F•4 EYE SETUP appears when F•2 MODE is set to EYE.
SYS → F4 EYE SETUP →
Figure 6-2 EYE SETUP menu

6.1 Turning Automatic Measurement On and Off

To automatically measure the amplitudes of eye patterns, jitter values, and other values, follow the procedure below. When you set F•1 AUTO MEASURE to ON, the measured values are displayed in the bottom part of the screen. Measured values are normally displayed in white, but they are displayed in red if they exceed the values that you have specified in unit setup.
See chapter 8, “UNIT SETUP.”
Procedure
EYE → F・4 EYE SETUP → F・1 AUTO MEASURE: ON / OFF
15
Page 19
6. EYE-PATTERN DISPLAY
AUTO MEASURE = ON
Figure 6-3 Automatic measurement
• Measurement items
The items that can be automatically measured are shown below.
Table 6-1 Measurement items
Symbol Display Indication Description
a Amp Eye-pattern amplitude
b Tr Rise time
c Tf Fall time (not shown in the following figure)
d T.J Timing jitter
e C.J Current jitter (jitter value when the currently selected filter is applied)
f Or Overshoot of the rising edge
g Of Overshoot of the falling edge
Figure 6-4 Explanation of measurement items
16
Page 20
6. EYE-PATTERN DISPLAY
• Unit interval
This unit uses unit intervals (UI) as jitter measurement units. One cycle of the eye pattern is 1 UI. The time that corresponds to 1 UI varies depending on the input signal, as shown below.
Table 6-2 Time that corresponds to 1 UI
Input Signal Bit Rate Time That Corresponds to 1 UI
2.970/1.001 Gbps 337.0 ps3G-SDI
2.970 Gbps 336.7 ps
1.485/1.001 Gbps 674.1 psHD-SDI
1.485 Gbps 673.4 ps
SD-SDI 270 Mbps 3.7 ns
Figure 6-5 Unit interval
• Precautions
When the LV 58SER07 is performing automatic measurements of rise time, fall time, overshoot of the rising edge, or overshoot of the falling edge, it may take some time for the measured values to stabilize.
The LV 58SER07 uses a histogram to automatically measure eye-pattern amplitude, rise time, fall time, overshoot of the rising edge, and overshoot of the falling edge. The LV 58SER07 algorithms are optimized for reducing noise.
When the LV 58SER07 is performing automatic measurements of timing jitter or current jitter, the values that were measured in jitter display mode are displayed. The LV 58SER07 uses a phase modulator to perform these measurements.
If there is a large amount of waveform distortion—such as jitter, noise, or overshoot—present in the input signal, the difference between the automatically measured values and the cursor-measured values may become large.
17
Page 21

6.2 Adjusting the Gain

To adjust the eye-pattern gain, follow the procedure below.
Procedure
EYE → F・4 EYE SETUP → F・2 GAIN VARIABLE: CAL / VARIABLE
Settings
CAL: The eye pattern is shown without gain. VARIABLE: The eye pattern is shown with the specified gain (×0.50 to ×2.00). The gain
value appears in the upper right of the screen. Turn the function dial (F•D) to adjust the gain. Press the function dial (F•D) to return the setting to its default value (×1.00).
GAIN VAR = ×0.50 GAIN VAR = ×2.00
6. EYE-PATTERN DISPLAY
Figure 6-6 Adjusting the gain

6.3 Selecting the Sweep Time

To select the eye-pattern sweep time, follow the procedure below.
Procedure
EYE → F・4 EYE SETUP → F・3 SWEEP:
Settings
2UI: Two cycles of the eye pattern are shown. 4UI: Four cycles of the eye pattern are shown. 16UI: Sixteen cycles of the eye pattern are shown.
SWEEP = 2UI SWEEP = 4UI
2UI / 4UI / 16UI
18
Page 22
SWEEP = 16UI

6.4 Selecting the Filter

To select the filter that is used during jitter measurement, follow the procedure below. The selected filter is indicated in the lower right of the display. If you change this setting, the filter that you selected for jitter display mode also changes.
See section 7.4, “Selecting the Filter.”
6. EYE-PATTERN DISPLAY
Figure 6-7 Selecting the sweep time
Procedure
EYE → F・4 EYE SETUP → F・4 FILTER: 100kHz / 1kHz / 100Hz / 10Hz / TIMING /
ALIGNMENT
Settings
100kHz: Jitter at 100 kHz or higher is measured. 1kHz: Jitter at 1 kHz or higher is measured. 100Hz: Jitter at 100 Hz or higher is measured. 10Hz: Jitter at 10 Hz or higher is measured. TIMING: Timing jitter is measured. Jitter at 10 Hz or higher is measured. ALIGNMENT: Alignment jitter is measured. When the input signal is 3G-SDI or HD-SDI,
jitter at 100 kHz or higher is measured. When the input signal is SD-SDI, jitter at 1 kHz or higher is measured.
FILTER = TIMING (3G-SDI signal) FILTER = ALIGNMENT (3G-SDI signal)
19
Page 23
FILTER = TIMING (SD-SDI signal) FILTER = ALIGNMENT (SD-SDI signal)
Figure 6-8 Selecting the filter

6.5 Configuring Cursor Settings

You can use cursors to measure time and amplitude. The measured values that you obtain by using the cursors and the automatically measured values may be different. See the precautions in section 6.1, “Turning Automatic Measurement On and Off.”
To configure cursor settings, press F•5 CURSOR on the EYE SETUP menu.
6. EYE-PATTERN DISPLAY
SYS → F4 EYE SETUP → F5 CURSOR →
Figure 6-9 CURSOR menu

6.5.1 Turning Cursors On and Off

To turn cursors on and off, follow the procedure shown below. When you set F•1 CURSOR to ON, the REF cursor is displayed in blue, and the DELTA cursor is displayed in green. The value of DELTA-REF appears as a measured value in the upper part of the screen.
Procedure
EYE → F・4 EYE SETUP → F・5 CURSOR → F・1 CURSOR: ON / OFF
CURSOR = ON
Figure 6-10 Cursor display
20
Page 24

6.5.2 Selecting the Cursors

The X-axis and Y-axis cursors are displayed at the same time, but you can only use the function dial (F•D) to move one set of cursors at a time. To select the cursors that you want to move, follow the procedure below.
Procedure
EYE → F・4 EYE SETUP → F・5 CURSOR → F・2 XY SEL: X / Y / Tr,Tf
If you select Tr,Tf, you can measure the rise time (Tr) and fall time (Tf) by following the procedure below.
1. Set F•2 XY SEL to Tr,Tf.
This selects the Y-axis cursors.
2. Use the function dial (F•D) to align the cursors with the amplitude of the eye pattern.
6. EYE-PATTERN DISPLAY
Figure 6-11 Tr,Tf measurement (1)
3. Press F•5 REF SET.
The Y-axis cursors move to the 20 % and 80 % positions of the amplitude, and then F•2 XY SEL is automatically set to X.
Figure 6-12 Tr,Tf measurement (2)
21
Page 25
6. EYE-PATTERN DISPLAY
4. Align the X-axis cursors with the intersections of the Y-axis cursors and the eye pattern.
You can now measure the rise time and fall time. The measured value is displayed next to X in the upper part of the screen.
Figure 6-13 Tr,Tf measurement (3)

6.5.3 Selecting the X-Axis Measurement Unit

To select the units used in X-axis cursor measurement, follow the procedure below. This setting appears when F•2 XY SEL is set to X.
Procedure
EYE → F・4 EYE SETUP → F・5 CURSOR → F・3 X UNIT: sec / Hz / UIp-p
Settings
sec: The measurement unit is seconds. Hz: The measurement unit is hertz (frequency). The length of one period is the
time between two cursors.
UIp-p: The measurement unit is UIp-p. One UIp-p is one cycle of the eye pattern.

6.5.4 Selecting the Y-Axis Measurement Unit

To select the units used in Y-axis cursor measurement, follow the procedure below. This setting appears when F•2 XY SEL is set to Y.
Procedure
EYE → F・4 EYE SETUP → F・5 CURSOR → F・3 Y UNIT: V / %
Settings
V: The measurement unit is volts. %: The measurement unit is percentage. If you select %, F•5 REF SET appears.
100 % is set to the amplitude at the time when you press F•5 REF SET.
22
Page 26

6.5.5 Moving Cursors

To move a cursor, follow the procedure shown below to select a cursor. Then, move the cursor by turning the function dial (F•D). Triangles appear on both ends of the selected cursor.
You can also select a cursor by pressing the function dial (F•D). Each time you press the function dial (F•D), the selected cursor switches from REF, to DELTA, to TRACK, and then back to REF.
Procedure
EYE → F・4 EYE SETUP → F・5 CURSOR → F・4 FD VAR: REF / DELTA / TRACK
Settings
REF: The REF cursor (blue) is selected. DELTA: The DELTA cursor (green) is selected. TRACK: The REF cursor and DELTA cursor are both selected.
6. EYE-PATTERN DISPLAY
23
Page 27
7. JITTER DISPLAY
This chapter explains the jitter display. To show the jitter display, set F•2 MODE to JITTER.

7. JITTER DISPLAY

Figure 7-1 Jitter display screen
• Explanation of the jitter display
In the jitter display mode, the jitter component is extracted from the input signal and plotted on a graph in which time is assigned to the Y axis. In this mode, the jitter can be measured accurately even when the eye pattern is not displayed or when jitter of 1 UI or greater is present. Also, because the length of the horizontal axis is equal to one line or field (frame) of the video signal, you can observe jitter that is synchronized with lines and fields (frames).
• Automatic measurement on the jitter display
Timing jitter (T.J) and current jitter (C.J) are automatically measured and displayed on the jitter display screen.
SMPTE defines two methods of measuring jitter. One method uses an eye pattern, and the other method uses a phase modulator. LV 58SER07 automatic measurement uses the latter method. These two methods can produce different measured values.
The method of measuring jitter from an eye pattern waveform can only be used when the jitter value is small, because in principle, measurement can only be carried out when the eyes in the pattern are open. Another shortcoming is that measurement is prone to errors, because it can be difficult to distinguish between jitter and waveform distortions such as noise and sag.
On the other hand, the phase modulator measurement method does not suffer from the shortcomings described above, and so is more suitable for making accurate measurements. In automatic measurement, the LV 58SER07 averages measured values so that they do not fluctuate minutely even when the jitter value does. The LV 58SER07 uses the optimum number of values for averaging.
24
Page 28
7. JITTER DISPLAY
Measured values are normally displayed in white, but they are displayed in red if they exceed the values that you have specified in unit setup.
See chapter 8, “UNIT SETUP.”
• Jitter display setup
To configure the jitter-display settings, press F•4 JITTER SETUP. This setting appears when F•2 MODE is set to JITTER.
SYS → F4 JITTER SETUP →
Figure 7-2 JITTER SETUP menu

7.1 Configuring the Peak Hold Settings

To configure the peak hold settings, press F•1 PEAK HOLD on the JITTER SETUP menu.
SYS → F4 JITTER SETUP → F1 PEAK HOLD →
Figure 7-3 PEAK HOLD menu

7.1.1 Turning the Peak Hold On and Off

To measure the peak values of the timing jitter (T.J) and the current jitter (C.J), follow the procedure below. When you set F•1 PEAK HOLD to ON, the peak values are displayed in the lower part of the screen next to “PEAK.” The peak values are retained until you press F•2 CLEAR. If a peak value exceeds a limit, “OVER” is displayed.
For details about the “OVER” indication, see section 7.2, “Selecting the Gain.”
Procedure
EYE → F・4 JITTER SETUP → F・1 PEAK HOLD → F・1 PEAK HOLD: ON / OFF
PEAK HOLD = ON
Figure 7-4 Peak hold display
25
Page 29

7.1.2 Clearing the Peak Hold

To clear the peak values, follow the procedure below.
Procedure
EYE → F・4 JITTER SETUP → F・1 PEAK HOLD → F・2 CLEAR

7.2 Selecting the Gain

To select the jitter-waveform gain, follow the procedure below.
Procedure
EYE → F・4 JITTER SETUP → F・2 GAIN MAG:×1 / ×2 / ×8
Settings
×1: The jitter waveform is shown without gain. ×2: The jitter waveform is shown with ×2 gain. ×8: The jitter waveform is shown with ×8 gain.
GAIN MAG = ×1 GAIN MAG = ×2
7. JITTER DISPLAY
GAIN MAG = ×8
Figure 7-5 Selecting the gain
26
Page 30
7. JITTER DISPLAY
• Measurement range
The measurement range and the various display conditions for each gain setting are shown below. Select an appropriate gain that matches the measured value that you want to obtain.
Table 7-1 Measurement range
F2 GAIN MAG Measurement Range *UNDER RANGE* Appears OVER Appears
×1 4.80 to 9.60 UI 0.60 UI or less 10.01 UI or more
×2 1.20 to 4.80 UI 0.60 UI or less 5.21 UI or more
×8 0.00 to 1.20 UI Does not appear 1.31 UI or more
• UNDER RANGE display
When F•2 GAIN MAG is set to ×1 or ×2 and the measured jitter value is 0.60 UI or less, the measured value becomes yellow, and “*UNDER RANGE*” appears at the bottom left of the screen. If this happens, set F•2 GAIN MAG to ×8.
Figure 7-6 UNDER RANGE display
• OVER display
If the measured jitter value exceeds its limit, “OVER” will appear in red as the measured value at the bottom left of the screen. If this happens, set F•2 GAIN MAG to ×8, ×2, and then ×1, in that order.
Figure 7-7 OVER display
27
Page 31

7.3 Selecting the Sweep Time

To select the sweep time, follow the procedure below.
Procedure
EYE → F・4 JITTER SETUP → F・3 SWEEP: 1H / 2H / 1V / 2V
Settings
1H: The jitter from a period of one line is shown. 2H: The jitter from a period of two lines is shown. 1V: When the input signal is interlace or segmented frame, the jitter from a period of
one field is shown. When the input signal is progressive, the jitter from a period of one frame is shown.
2V: When the input signal is interlace or segmented frame, the jitter from a period of
one frame is shown. When the input signal is progressive, the jitter from a period of two frames is shown.

7.4 Selecting the Filter

7. JITTER DISPLAY
To select the filter that is used during jitter measurement, follow the procedure below. The selected filter is indicated in the lower right of the display. If you change this setting, the filter that you selected for eye-pattern-display mode also changes.
See section 6.4, “Selecting the Filter.”
Procedure
EYE → F・4 JITTER SETUP → F・4 FILTER: 100kHz / 1kHz / 100Hz / 10Hz / TIMING /
ALIGNMENT
Settings
100kHz: Jitter at 100 kHz or higher is measured. 1kHz: Jitter at 1 kHz or higher is measured. 100Hz: Jitter at 100 Hz or higher is measured. 10Hz: Jitter at 10 Hz or higher is measured. TIMING: Timing jitter is measured. Jitter at 10 Hz or higher is measured. ALIGNMENT: Alignment jitter is measured. When the input signal is 3G-SDI or HD-SDI,
jitter at 100 kHz or higher is measured. When the input signal is SD-SDI, jitter at 1 kHz or higher is measured.

7.5 Configuring Cursor Settings

You can use cursors to measure time and amplitude. To configure cursor settings, press F•5 CURSOR on the JITTER SETUP menu.
SYS → F4 JITTER SETUP → F5 CURSOR →
Figure 7-8 CURSOR menu
28
Page 32

7.5.1 Turning Cursors On and Off

To turn cursors on and off, follow the procedure shown below. When you set F•1 CURSOR to ON, the REF cursor is displayed in blue, and the DELTA cursor is displayed in green. The value of DELTA-REF appears as a measured value in the lower right of the screen.
Procedure
EYE → F・4 JITTER SETUP → F・5 CURSOR → F・1 CURSOR: ON / OFF
CURSOR = ON
7. JITTER DISPLAY
Figure 7-9 Cursor display

7.5.2 Selecting the Cursors

The X-axis and Y-axis cursors are displayed at the same time, but you can only use the function dial (F•D) to move one set of cursors at a time. To select the cursors that you want to move, follow the procedure below.
Procedure
EYE → F・4 JITTER SETUP → F・5 CURSOR → F・2 XY SEL: X / Y

7.5.3 Selecting the X-Axis Measurement Unit

To select the units used in X-axis cursor measurement, follow the procedure below. This setting appears when F•2 XY SEL is set to X.
Procedure
EYE → F・4 JITTER SETUP → F・5 CURSOR → F・3 X UNIT: sec / Hz
Settings
sec: The measurement unit is seconds. Hz: The measurement unit is hertz (frequency). The length of one period is the
time between two cursors.
29
Page 33
7. JITTER DISPLAY

7.5.4 Selecting the Y-Axis Measurement Unit

To select the units used in Y-axis cursor measurement, follow the procedure below. This setting appears when F•2 XY SEL is set to Y.
Procedure
EYE → F・4 JITTER SETUP → F・5 CURSOR → F・3 Y UNIT: sec / UIp-p
Settings
sec: The measurement unit is seconds. UIp-p: The measurement unit is UIp-p. One UIp-p is one cycle of the eye pattern.

7.5.5 Moving Cursors

To move a cursor, follow the procedure shown below to select a cursor. Then, move the cursor by turning the function dial (F•D). Triangles appear on both ends of the selected cursor.
You can also select a cursor by pressing the function dial (F•D). Each time you press the function dial (F•D), the selected cursor switches from REF, to DELTA, to TRACK, and then back to REF.
Procedure
EYE → F・4 JITTER SETUP → F・5 CURSOR → F・4 FD VAR: REF / DELTA / TRACK
Settings
REF: The REF cursor (blue) is selected. DELTA: The DELTA cursor (green) is selected. TRACK: The REF cursor and DELTA cursor are both selected.
30
Page 34
8. UNIT SETUP
To configure error detection settings, use the UNIT SETUP menu in the system settings. The LV 58SER07 has two input channels (channels A and B), but the settings that you configure on the UNIT SETUP menu are applied to both channels. You cannot configure settings for each channel individually.
To display the UNIT SETUP menu, press F•1 UNIT SETUP on the system menu. Next, press the function key (F•1 to F•4) that corresponds to the unit number that the LV 58SER07 is installed in.
SYS → F1 UNIT SETUP →
Figure 8-1 UNIT SETUP menu

8.1 Operating Tab Menus

8. UNIT SETUP

You can usually use the function menu to change the settings, but the unit setup settings appear on a tab menu. The list below explains how to perform different operations in the tab menu.
To Move the Cursor
Turn the function dial (F•D).
To Change the Tab
Press F•2 PREV or F•3 NEXT. Even if you switch to another tab, the settings are not confirmed until you press F•1 COMPLETE.
To Select a Check Box
Press the function dial (F•D).
To Enter a Value
When you push the function dial (F•D), the color of the cursor changes from blue to green, and you can set the value. Turn the function dial (F•D) to set the value. To confirm the value that you have set, press the function dial (F•D) again.
To Apply the Tab Settings
Press F•1 COMPLETE. All the tab settings are applied, and the previous menu is displayed.
To Cancel the Tab Settings
Press F•7 CANCEL. All the tab settings are cancelled, and the previous menu is displayed.
31
Page 35
8. UNIT SETUP

8.2 Configuring 3G-SDI Error Settings

Use the 3G-SDI ERR SETUP tab to configure error detection settings for 3G-SDI signals.
If one of the errors whose detection has been set to ON occurs, the corresponding measured value is displayed in red on the eye-pattern and jitter display screens. The error is also displayed on the event log screen. The detection for all errors is initially set to OFF.
You can set the threshold values when you set the error detection to ON. Measured values given in SMPTE 424M are used as 100 %.
Figure 8-2 3G-SDI ERR SETUP tab
A configuration example showing threshold values that correspond to SMPTE 424M is given below.
Table 8-1 3G-SDI ERR SETUP configuration example
Item Value Corresponding Value
Upper 110 % 880 mVAmplitude Error
Lower 90 % 720 mV
Risetime Error Max 100 % 135.0 ps
Falltime Error Max 100 % 135.0 ps
Deltatime Error(Tr-Tf) Max 100 % 50 ps
Timing Jitter Error Max 100 % 2.00 UI (674.0 ps)
Current Jitter Error Max 100 % 0.30 UI (101.2 ps)
OverShoot Rising Error Max 100 % 10.0 %
OverShoot Falling Error Max 100 % 10.0 %
• Amplitude Error
Turns the eye pattern’s amplitude error detection on and off. When you select ON, you can set the upper and lower limits. You cannot set Lower to a value that is greater than Upper, even if the value is within the selectable range.
32
Page 36
8. UNIT SETUP
Upper: 80 to 140 % (640 to 1120 mV) Lower: 40 to 100 % (320 to 800 mV)
• Risetime Error
Turns the eye pattern’s rise time (the time for the signal to rise from 20 to 80 % of its amplitude) error detection on and off. When you select ON, you can set the upper limit.
Max: 40 to 140 % (54.0 to 189.0 ps)
• Falltime Error
Turns the eye pattern’s fall time (the time for the signal to fall from 80 to 20 % of its amplitude) error detection on and off. When you select ON, you can set the upper limit.
Max: 40 to 140 % (54.0 to 189.0 ps)
• Deltatime Error
Turns the eye pattern’s time difference (between the rise and fall times) error detection on and off. When the measured values exceed the specified value, Tr and Tf are displayed in red. When you select ON, you can set the upper limit.
Max: 40 to 140 % (20 to 70 ps)
• Timing Jitter Error
Turns the eye pattern and jitter waveform’s timing jitter error detection on and off. When you select ON, you can set the upper limit.
Max: 10 to 200 % (0.20 to 4.00 UI, 67.4 to 1348.0 ps)
• Current Jitter Error
Turns the eye pattern and jitter waveform’s current jitter error detection on and off. When you select ON, you can set the upper limit.
Max: 10 to 200 % (0.03 to 0.60 UI, 10.1 to 202.5 ps)
• OverShoot Rising Error
Turns the overshoot of the rising edge error detection on and off. When you select ON, you can set the upper limit.
Max: 0 to 200 % (0.0 to 20.0 %)
• OverShoot Falling Error
Turns the overshoot of the falling edge error detection on and off. When you select ON, you can set the upper limit.
Max: 0 to 200 % (0.0 to 20.0 %)
33
Page 37
8. UNIT SETUP

8.3 Configuring HD-SDI Error Settings

Use the HD-SDI ERR SETUP tab to configure error detection settings for HD-SDI signals.
If one of the errors whose detection has been set to ON occurs, the corresponding measured value is displayed in red on the eye-pattern and jitter display screens. The error is also displayed on the event log screen. The detection for all errors is initially set to OFF.
You can set the threshold values when you set the error detection to ON. Measured values given in SMPTE 292M are used as 100 %.
Figure 8-3 HD-SDI ERR SETUP tab
A configuration example showing threshold values that correspond to SMPTE 292M is given below.
Table 8-2 HD-SDI ERR SETUP configuration example
Item Value Corresponding Value
Upper 110 % 880 mVAmplitude Error
Lower 90 % 720 mV
Risetime Error Max 100 % 270.0 ps
Falltime Error Max 100 % 270.0 ps
Deltatime Error(Tr-Tf) Max 100 % 100 ps
Timing Jitter Error Max 100 % 1.00 UI (674.0 ps)
Current Jitter Error Max 100 % 0.20 UI (135.0 ps)
OverShoot Rising Error Max 100 % 10.0 %
OverShoot Falling Error Max 100 % 10.0 %
• Amplitude Error
Turns the eye pattern’s amplitude error detection on and off. When you select ON, you can set the upper and lower limits. You cannot set Lower to a value that is greater than Upper, even if the value is within the selectable range.
34
Page 38
8. UNIT SETUP
Upper: 80 to 140 % (640 to 1120 mV) Lower: 40 to 100 % (320 to 800 mV)
• Risetime Error
Turns the eye pattern’s rise time (the time for the signal to rise from 20 to 80 % of its amplitude) error detection on and off. When you select ON, you can set the upper limit.
Max: 40 to 140 % (108.0 to 378.0 ps)
• Falltime Error
Turns the eye pattern’s fall time (the time for the signal to fall from 80 to 20 % of its amplitude) error detection on and off. When you select ON, you can set the upper limit.
Max: 40 to 140 % (108.0 to 378.0 ps)
• Deltatime Error
Turns the eye pattern’s time difference (between the rise and fall times) error detection on and off. When the measured values exceed the specified value, Tr and Tf are displayed in red. When you select ON, you can set the upper limit.
Max: 40 to 140 % (40 to 140 ps)
• Timing Jitter Error
Turns the eye pattern and jitter waveform’s timing jitter error detection on and off. When you select ON, you can set the upper limit.
Max: 10 to 200 % (0.10 to 2.00 UI, 67.4 to 1348.0 ps)
• Current Jitter Error
Turns the eye pattern and jitter waveform’s current jitter error detection on and off. When you select ON, you can set the upper limit.
Max: 10 to 200 % (0.02 to 0.40 UI, 13.5 to 270.0 ps)
• OverShoot Rising Error
Turns the overshoot of the rising edge error detection on and off. When you select ON, you can set the upper limit.
Max: 0 to 200 % (0.0 to 20.0 %)
• OverShoot Falling Error
Turns the overshoot of the falling edge error detection on and off. When you select ON, you can set the upper limit.
Max: 0 to 200 % (0.0 to 20.0 %)
35
Page 39
8. UNIT SETUP

8.4 Configuring SD-SDI Error Settings

Use the SD-SDI ERR SETUP tab to configure error detection settings for SD-SDI signals.
If one of the errors whose detection has been set to ON occurs, the corresponding measured value is displayed in red on the eye-pattern and jitter display screens. The error is also displayed on the event log screen. The detection for all errors is initially set to OFF.
You can set the threshold values when you set the error detection to ON. Measured values given in SMPTE 259M are used as 100 %.
Figure 8-4 SD-SDI ERR SETUP tab
A configuration example showing threshold values that correspond to SMPTE 259M is given below.
Table 8-3 SD-SDI ERR SETUP configuration example
Item Value Corresponding Value
Upper 110 % 880 mVAmplitude Error
Lower 90 % 720 mV
Risetime Error Max 100 % 1.50 ns
Falltime Error Max 100 % 1.50 ns
Deltatime Error(Tr-Tf) Max 100 % 0.50 ns
Timing Jitter Error Max 100 % 0.20 UI (0.74 ns)
Current Jitter Error Max 100 % 0.20 UI (0.74 ns)
OverShoot Rising Error Max 100 % 10.0 %
OverShoot Falling Error Max 100 % 10.0 %
• Amplitude Error
Turns the eye pattern’s amplitude error detection on and off. When you select ON, you can set the upper and lower limits. You cannot set Lower to a value that is greater than Upper, even if the value is within the selectable range.
36
Page 40
8. UNIT SETUP
Upper: 80 to 140 % (640 to 1120 mV) Lower: 40 to 100 % (320 to 800 mV)
• Risetime Error
Turns the eye pattern’s rise time (the time for the signal to rise from 20 to 80 % of its amplitude) error detection on and off. When you select ON, you can set the upper limit.
Max: 40 to 140 % (0.60 to 2.10 ns)
• Falltime Error
Turns the eye pattern’s fall time (the time for the signal to fall from 80 to 20 % of its amplitude) error detection on and off. When you select ON, you can set the upper limit.
Max: 40 to 140 % (0.60 to 2.10 ns)
• Deltatime Error
Turns the eye pattern’s time difference (between the rise and fall times) error detection on and off. When the measured values exceed the specified value, Tr and Tf are displayed in red. When you select ON, you can set the upper limit.
Max: 40 to 140 % (0.20 to 0.70 ns)
• Timing Jitter Error
Turns the eye pattern and jitter waveform’s timing jitter error detection on and off. When you select ON, you can set the upper limit.
Max: 10 to 200 % (0.02 to 0.40 UI, 0.07 to 1.48 ns)
• Current Jitter Error
Turns the eye pattern and jitter waveform’s current jitter error detection on and off. When you select ON, you can set the upper limit.
Max: 10 to 200 % (0.02 to 0.40 UI, 0.07 to 1.48 ns)
• OverShoot Rising Error
Turns the overshoot of the rising edge error detection on and off. When you select ON, you can set the upper limit.
Max: 0 to 200 % (0.0 to 20.0 %)
• OverShoot Falling Error
Turns the overshoot of the falling edge error detection on and off. When you select ON, you can set the upper limit.
Max: 0 to 200 % (0.0 to 20.0 %)
37
Page 41
9. REFERENCE MATERIAL

9.1 Menu Tree

The menu tree for the EYE key is shown below. The default values are underlined.

9. REFERENCE MATERIAL

EYE
F1 F1
INTEN
F2
MODE ( EYE / JITTER )
1
EYE
F4
SETUP
5
JITTER
F4
SETUP
F2
F7
F1
F2
F3
F4
F5
F7
F2
F3
EYE INTEN
SCALE INTEN
up menu
AUTO MEASURE
GAIN VARIABLE
SWEEP
FILTER
CURSOR
up menu
PEAK HOLD
GAIN MAG
SWEEP
( -128 - 32 - 127 )
( -8 - 4 - 7 )
( ON / OFF )
( CAL / VARIABLE )
( 2UI / 4UI / 16UI )
( 100kHz / 1kHz / 100Hz / 10Hz / TIMING / ALIGNMENT )
F1
CURSOR ( ON / OFF )
F2
XY SEL
2
F3
X UNIT
3
F3
Y UNIT ( V / % )
F4
FD VAR
4
F5
REF SET
up
F7
menu
PEAK
F1F1
HOLD
F2
CLEAR
up
F7
menu
( X1 / X2 / X8 )
( 1H / 2H / 1V / 2V )
( X / Y / Tr,Tf )
( sec / Hz / UIp-p )
( REF / DELTA / TRACK )
( ON / OFF )
F4
FILTER
F5 F1
CURSOR
up
F7
menu
6
LINK
F6
SELECT
( LINK A / LINK B )
( 100kHz / 1kHz / 100Hz / 10Hz / TIMING / ALIGNMENT )
CURSOR
F2
XY SEL
2
F3
X UNIT
3
F3
Y UNIT
F4
up
F7
menu
( ON / OFF )
( X / Y )
( sec / Hz )
( sec / UIp-p )
( REF / DELTA / TRACK )FD VAR
Figure 9-1 EYE menu
38
Page 42
9. REFERENCE MATERIAL
1 Appears when MODE is set to EYE. 2 Appears when XY SEL is set to X. 3 Appears when XY SEL is set to Y. 4 Appears when XY SEL is set to Tr,Tf or Y UNIT is set to %. 5 Appears when MODE is set to JITTER. 6 Appears when the input signal is HD dual link.

9.2 Change History of the Software

This manual was written for the following firmware versions:
• Ver 6.8 (LV 5800)
• Ver 1.4 (LV 7800)
To check the version number, press SYS and then F•5 SYSTEM INFORMATION.
Ver 6.6 (LV 5800) and Ver 1.4 (LV 7800)
• LV 58SER07 support has been added.
39
Page 43

Index

3G-SDI ERR SETUP ................................................. 32
AUTO MEASURE...................................................... 15
CLEAR....................................................................... 26
CURSOR ............................................................. 20, 29
EYE INTEN................................................................ 12
FD VAR................................................................ 23, 30
FILTER ................................................................ 19, 28
LINK SELECT ........................................................... 13
MODE........................................................................ 13
PEAK HOLD.............................................................. 25
REF SET ................................................................... 22
SCALE INTEN........................................................... 12
SD-SDI ERR SETUP................................................. 36
SWEEP ............................................................... 18, 28
GAIN MAG................................................................. 26
GAIN VARIABLE........................................................ 18
H POS ....................................................................... 12
HD-SDI ERR SETUP................................................. 34
INTEN........................................................................ 12
UNIT SETUP ............................................................. 31
V POS ....................................................................... 12
X UNIT................................................................. 22, 29
XY SEL................................................................ 21, 29
Y UNIT................................................................. 22, 30
Page 44
Following information is for Chinese RoHS only
所含有毒有害物质信息
部件号码:
此标志适用于在中国销售的电子信息产品,依据2006年2月28日公布的 《电子信息产品污染控制管理办法》以及SJ/T11364-2006《电子信息产品污染 控制标识要求》,表示该产品在使用完结后可再利用。数字表示的是环境保护使 用期限,只要遵守与本产品有关的安全和使用上的注意事项,从制造日算起在数 字所表示的年限内,产品不会产生环境污染和对人体、财产的影响。 产品适当使用后报废的方法请遵从电子信息产品的回收、再利用相关法令。 详细请咨询各级政府主管部门。
部件名称
Parts 六价铬 多溴联苯 多溴二苯醚
(Pb) (Hg) (Cd) (Cr( Ⅵ)) (PBB) (PBDE)
实装基板 ×
主体部 × ○ 包装材
LV 58SER07
产品中有毒有害物质或元素的名称及含量
有毒有害物质或元素 Hazardous Substances in each Part
备注)
○:表示该有毒有害物质在该部件所有均质材料中的含量均在SJ/T11363-2006 规定的限量要求以下。
×:表示该有毒有害物质或元素至少在该部件的某一均质材料中的含量超出SJ/T11363-2006
标准规定的限量要求。
Ver.1
Page 45
LEADER ELECTRONICS CORP.
2-6-33 Tsunashima-Higashi, Kohoku-ku, Yokohama 223-8505, Japan
PHONE:81-45-541-2123 FAX:81-45-541-2823 http://www.leader.co.jp
Sep. 16, 2010 Ver.1 (FW Ver.6.8)
Loading...