Leader LBO-516 User Manual

WARNING!
THE SERVICING INSTRUCTIONS CONTAINED IN THIS MANUAL ARE FOR USE BY QUALIFIED PERSONNEL ONLY. TO AVOID ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING OTHER THAN THAT CONTAINED IN THE OPERATING INSTRUCTIONS UNLESS YOU ARE QUALIFIED TO DO SO.
LBO-516 100 MHz
DELAYED TIME BASE OSCIILOSCOPE
TABLE OF CONTENTS
1. GENERALINFORMATION
1-1 INTRODUCTION ................................................................................................................................1
1-2 SPECIFICATIONS ..........…..................................................................................................................................1
2. OPERATING INSTRUCTIONS
2-1 FUNCTION OF CONTROLS, CONNECTORS, AND INDICATORS .......................................................... 3
2-1-1 Display Block ........................................................................................................................................... 3
2-1-2 Vertical Amplifier Block ........................................................................................................................... 4
2-1-3 Sweep and Trigger Blocks ......................................................................................................................... 5
2-1-4 Miscellaneous ........................................................................................................................................... 7
2-2 INITIAL OPERATION ................................................................................................................................ 7
2-2-1 Power Connections and Adjustments ......................................................................................................... 7
2-2-2 Installation ................................................................................................................................................ 8
2-2-3 Preliminary Control Settings and Adjustments ........................................................................................... 8
2-3 BASIC OPERATING PROCEDURES ......................................................................................................... 9
2-3-1 Signal Connections ................................................................................................................................... 9
2-3-2 Single-trace Operation ..............................................................................................................................11
2-3-3 Triggering Alternatives ............................................................................................................................11
2-3-4 Probe Compensation ................................................................................................................................14
2-3-5 Dual-trace Operation ................................................................................................................................15
2-3-6 Additive and Differential Operation ..........................................................................................................16
2-3-7 Triple-trace Operation ..............................................................................................................................17
2-3-8 Four-trace Operation ................................................................................................................................17
2-3-9 Delayed Timebase Operation ....................................................................................................................18
2-3-10 Single-shot Operation .............................................................................................................................20
2-3-11 X-Y Operation .......................................................................................................................................20
2-3-12 Intensity Modulation ..............................................................................................................................20
2-4 MEASUREMENT APPLICATIONS ...........................................................................................................21
2-4-1 Amplitude Measurements .........................................................................................................................21
2-4-2 Differential Measurement Techniques .......................................................................................................22
2-4-3 Time Interval Measurements ....................................................................................................................23
2-4-4 Phase Difference Measurements ...............................................................................................................24
2-4-5 Distortion Comparison .............................................................................................................................26
2-4-6 Frequency Measurements .........................................................................................................................27
2-4-7 Risetime Measurements ............................................................................................................................27
2-4-8 -3dB Bandwidth Measurement ..................................................................................................................28
........................................................................................................................................1
..................................................................................................................................3
1. GENERAL INFORMATION
1-1. INTRODUCTION
The LBO-516, shown in Figure 1-1, is a 100 MHz oscilloscope
with all of the features normally found on a lab-grade scope: high­fidelity pulse response, stable operation, dual timebase with calibrated sweep delay, flexible triggering facilities, and a bright CRT display with illuminated internal graticule. Moreover, it also has a very unusual feature found on few scopes in any price class: it can simultaneously display up to eight traces from three different input signals! In addition to the two vertical-input channels and their differ­ence signal, the signal used to externally trigger the main timebase can also appear on the CRT display. The alternate sweep mode, which allows the main and delayed timebases to simultaneously sweep the CRT, effectively doubles this four-trace display to an eight-trace display.
The comprehensive triggering facilities of the LBO-516 include
several features that ease the problem of triggering on complex signals: a variety of frequency-selective coupling filters, a trigger holdoff-control, and a trigger pickoff that alternates between the two vertical channels.
1-2. SPECIFICATIONS
Specifications for the model LBO-516 oscilloscope are given in
Table 1-1.
Table 1-1
SPECIFICATIONS
Vertical Amplifiers (Ch. 1 & 2)
Bandwidth (-3 dB) DC coupled DC - 100 MHz AC coupled 10 Hz - 100 MHz Rise Time 3.5 KS Deflection Coefficients
Accuracy 5 mV/div to 5 V/div in 10 calibrated
steps, 1-2-5 sequence. Continuously variable between steps. XI0 magnification adds 0.5, 1, and 2 mV/div steps for frequencies below 5 MHz
Input Impedance -+3%; -+5% with Xl0magniflca-tion
1 megohm +-2%, 25 pF +-3 pF
Maximum Input Voltage 400 V (DC plus AC peak)
1
Signal Delay Leading edge displayed. Leading edge displayed. CH-1 only, CH-2 only,
CH-1 & CH-2 displayed alternately, CH-1 & CH-2 chopped (at 250 kHz rate), CH-1 & CH-2 added, CH-1 & CH-2 subtracted, CH-1 & CH-2 & CH-3 displayed alternately, CH-1 & CH-2 & CH-3 chopped, CH-1 & CH-2 & CH-3 & CH-1 + CH-2 alternated, CH-1 & CH-2 & CH-3 & CH-1 + CH-2 chopped, CH-1 & CH-2 & CH-3 & CH-1
- CH-2 alternated, CH-1 & CH-2 & CH-3 & CH-1
- CH-2 chopped.
Common Mode Rejection 20dB at 20MHz CH- 1 Output 25 mV/div into 50 ohms
Horizontal Amplifier (X-Y Mode)
Bandwidth (- 3 dB) DC coupled DC - 3 MHz AC coupled 10Hz - 3MHz Rise Time 120 KS
Phase Shift <3° at 100 kHz Deflection Coefficients 0.5 mV/div to 5 V/div in 13 calibrated
steps, 1-2-5 sequence, continuously variable between steps
Accuracy +-3% for 5 mV/div to 5 V/div,
+-5% for 0.5 mV/div to 2 mV/div Input Impedance 1 megohm -+2%, 25 pF +-3 pF Maximum Input Voltage 400 V (DC plus AC peak)
Time-Base Generators
Display Modes Main timebase (TB) only,
Main TB intensified by delayed
TB,
Delayed timebase,
Main TB alternated with
delayed TB. Main (A) Time Base 0.02 KS/div to 0.5 S/div in 23
calibrated steps, 1-2-5 sequence.
Continuously variable between steps. Delayed (B) Time Base 0.2 PS/div to 50 mS/div in 20
calibrated steps, 1-2-5 sequence. Magnifier XI0 deflection increase at any TB
setting extends sweep speeds of main
and delayed TB's to 2 KS/div. Accuracy +- 3% unmagnified
+- 5% magnified Delay Time Continuously variable multiplier with
1000 divisions. Delayed TB Jitter 1/20,000
Trigger Circuits
Sources CH-l, CH-2, Alternate, Line, External
Modes Auto, Normal, Single-shot Coupling AC, DC, HF reject, TV vertical, TV
horizontal Slope + or­Holdoff Normal, Variable (to greater than
one sweep), B ends A Sensitivity Internal Trigger DC – 10 MHz: 0.4div
External Trigger DC - 10MHz: 100mV
External Trigger Amplifier (Ch. 3)
Bandwidth (-3 dB) DC coupled DC - 100 MHz AC coupled 10Hz- l00 MHz Rise Time 3.5 KS Deflection Coefficients 0.2 V/div and 2 V/div Accuracy +-3% Input Impedance 1 megohm +-2%, 30 pF Maximum Input Voltage 400 V (DC plus AC peak)
Z-Axis Modulation
Level for Blanking Standard TTL high (+ 2 to + 5V) Coupling DC Maximum Input Voltage 50 Vp-p Input Impedance 10k: Bandwidth DC-5 MHz
Calibrator
Output Voltage 500 m Vp-p--+ 2%, positive-going,
Frequency 1 kHz nominal Waveform Fast-rise rectangular wave
CRT Display
Phosphor P31 (P39 optional) Accelerating Potential 20 kV/2kV Graticule Internal 1 cm square divisions, 8 div
Graticule Illumination Continuously variable Trace Adjustments on Rotation, focus, intensity,
Front Panel B intensity
Other Features
"Out-of-Calibration" Indicator Other Indicators Main timebase triggered
Power Requirements
Line Voltage 100/120/200 VAC 220/240 VAC
Line Frequency 50-60 Hz Power Consumption 55W
Physical & Environmental Data
Case Size (WxHxD) 12.3 x 5.8 x 16 inches 305 x 145 x 400 mm Overall Size (WxHxD), 13.75 x 7.25 x 18.5 inches handle folded back 350 x 185 x 470 mm
10 - 100 MHz: 1.5 divs
10 - 100 MHz: 400mV
ground referenced
high, 10 div wide.
Central axis subdivided into 0.2 cm
graduations.
Main timebase
Single-shot ready
Power on
2
S
1
2
3 4 5 6 7 8 9
Weight 20.9 lbs, 9.5 kg Ambient Operating 0-40°C (32-104°F) maximum Temperature 15-35°C (60-95°F) for guaranteed
specs
Vibration Tolerance 2 mmp-p displacement at 12-33 Hz
and 33-35 Hz
Shock Tolerance 30g
Accessories
Supplied Instruction Manual
Two (2) LP- 100X probes Two (2) BNC-to-post adaptors
Optional LP-2017 Probe Pouch
LC-2016 Protective Front Cover LR-2402 Rack Mount Adaptor LH-2015 Hood
Specifications for the model LP-100X scope probe are given in
Table 1-2.
2. OPERATING INSTRUCTION
This section contains the information needed to operate the LBO-516 and utilize it in a variety of basic and advanced measurement procedures. Included are the identification and function of controls, connectors, and indicators, initial startup procedures, basic operating routines, and selected measurement applications.
2-1. FUNCTION OF CONTROLS, CONNECTORS, AND
INDICATORS
Before turning on
controls, connectors, indicators, and other features described in this section. The descriptions given below are keyed to the items called out in Figures 2-1 to 2-4.
2-1-1 Display Block
Refer to Figure 2-1 for reference (1) to (9).
POWER switch Push in to turn instrument power on
POWER lamp Lamp lights when power is on
A INTEN control To adjust the overall brightness of the
B INTEN control Provides adjustment of CRT brightness
FOCUS control To attain maximum trace sharpness.
ROTATION control Provides screwdriver adjustment of
this instrument, familiarize yourself with the
and off
CRT display. Clockwise rotation increases brightness
during INTEN BY B interval and B timebase sweeps
Astigmatism is automatically adjusted.
horizontal trace alignment with regard to the CRT graticule lines
Table 1-2
Attenuation Ratio 10:1 +-2% and 1:1, switch selectable Input Impedance 10X attenuation 10 megohms, 12 pF + 1X attenuation Scope input Z plus < 150 pF Rise Time (10X atten.) 3.5 KS nominal Overshoot & Ringing <10%
(10X atten.) Bandwidth 10X attenuation DC- 100MHz 1X attenuation DC - 6 MHz Maximum Input Voltage 600 V (DC plus AC peak) Ambient Operating Temperature Maximum - 10 to + 55°C For guaranteed
Specifications +5 to +35°C Ambient Humidity Maximum 40 to 90% For guaranteed
Specifications 45 to 85%
CRT Display device having 1 cm square
ILLUM control To adjust graticule illumination. Clockwise rotation increases brightness CAL connector Provides fast-rise waveform of precise
LP-100X SPECIFICATIONS
10%
graticule lines inscribed on the inner CRT surface for parallax-free measurements. Blue filter provides good contrast and pleasing display.
amplitude for probe adjustment and vertical amplifier calibration.
3
2-1-2 Vertical Amplifier Block
10
11 23 22 12
13 14
15
16 17 18 19 20 21
Refer to Figure 2-2 for references (10) to (12) and (14) to (23).
Refer to Figure 2-3 for reference (13).
VOLTS/DIV switches To select the calibrated deflection
factor of the input signals fed to the vertical amplifier.
VARIABLEcontrols Provide continuously variable
adjustment of deflection factor between steps of the VOLTS/DIV switches. Calibrations are accurate only when the VARIABLE controls are detented in
PULL X 10 MAG Pulling these out will increase the (on VARIABLE sensitivity of the associated vertical controls) amplifiers by ten times at reduced
Ground Connector Provides convenient point to attach
CH-1 OUTPUT Provides scaled output of the channel 1 Connector signal suitable for driving a frequency
CH- 1 or X-IN For applying an input signal to vertical connector amplifier channel 1, or the X-axis
CH-2 or Y-IN For applying an input signal to vertical connector amplifier 2, or the Y-axis (vertical)
AC/GND/DC To select the method of coupling the switches input signals to the vertical amplifiers.
Channel 1 Vertical For vertically positioning trace 1 on the POSITION control CRT screen. Clockwise rotation moves
Channel 2 Vertical or For vertically positioning trace 2 on the Y POSITION control CRT screen. Clockwise rotation moves
CH-2 INV switch Push in to invert the polarity of the
X-Y switch Push in to select X-Y operation.
V MODE switches To select the vertical amplifier display
their fully clockwise positions.
bandwidth.
separate ground lead to oscilloscope.
counter or other instrument.
(horizontal) amplifier during X-Y operation.
amplifier during X-Y operation.
AC position connects a capacitor between the input connector and its associated amplifier circuitry to block any DC component in the input signal. GND position connects the amplifier input to ground instead of the input connector, so a ground reference can be established. DC position connects the amplifier inputs directly to the associated input connector, thereby passing alt signal components on to the amplifiers.
the trace up. Inoperative during X-Y operation.
the trace up. Adjusts the Y-axis of the trace during X-Y operation.
Channel 2 signal.
CH-1
mode. the input signal of channel 1 on the CRT when pressed
push-button displays only
Figure 2-2
Vertical Amplifier Block
CH-2
push-button displays only the input signal
of channel 2 on the CRT when pressed.
ALT
push-button displays the input signals of both channels 1 and 2 (or more) on the CRT when pressed. The CRT beam is switched between channels at the end of each sweep to achieve this multi-channel display.
CHOP
push-button displays the input signals of both channels 1 and 2 (or more) on the CRT when pressed. The CRT beam is switched between channels at a 250 kHz rate during the horizontal sweep to achieve this multi-channel display.
ADD
push-button displays a single trace that is the algebraic sum of the input signals of channels 1 and 2 when pressed.
PULL TRIPLE control When pulled, displays traces for CH-1,
CH-2, and CH-3 (trigger), providing ALT or CHOP push-button is also pressed. Rotating this control also vertically positions the CH-3 trace on the CRT screen. This control is not operative if any single-trace display mode is selected.
PULL QUAD control When pulled, displays traces for CH-l,
CH-2, CH-3 (trigger), and algebraic sum of CH-1 and CH-2 signals, providing ALT of CHOP pushbutton is also pressed. This control is not operative if any single-trace display mode is selected.
4
2-1-3 Sweep and Trigger
24 32 25 26 27 28 29 30 31 33
Refer to Figure 2-3 for reference (24) to (42).
ATIME/DIV and To select either the calibrated DELAY TIME sweep of the main (A) time-base switch or the delay time range for delayed­ sweep operation.
B TIME/DIV switch To select the calibrated sweep rate of
A VARIABLE/ Provides continuously variable PULL X 10 MAG adjustment of sweep rate between steps Control of the TIME/DIV switches. TIME/DIV
UNCAL lamp Indicates when the VARIABLE control
DLY TIME MULT To determine the exact starting point control within the A time base delay range at
Horizontal POSITION To adjust the horizontal position of the
control traces displayed on the CRT.
X FINE Position To adjust the horizontal position of the
control CRT traces as described above, but has
HORIZ DISPLAY To select the sweep mode. Switches
Blocks
the delayed (B) time base.
calibrations are accurate only when the A VARIABLE control is detented in its fully clockwise position. Pulling the control out expands the horizontal deflection by 10 times for X-Y operation. The effective time-base sweep rate is also increased by 10 times, making 2 nS per division the highest sweep rate available.
is not detented as described above.
which the B timebase will begin sweeping. The absolute delay time is equal to the sweep time rate (A TIME/DIV) multiplied by the DLY TIME MULT.
Clockwise rotation moves the trace(s) to the right. During X-Y operation, this control must be used for X-axis positioning.
less effect per degree of rotation. This facilitates precise positioning when Xl0 magnification is used.
A push-button sweeps the CRT at the main (A) timebase rate when pressed.
INTEN BY B push-button sweeps the CRT at the main (A) time-base rate when pressed, and the delayed (B) time-base intensifies a section of the trace(s). The location of the intensified section is determined by the DLY TIME
MULT control, and under some circumstances also by the START switch (32). B push-button sweeps the CRT at the rate selected by the B TIMEY DIV switch, after a delay determined by the A TIME/DIV switch and DLY TIME MULT control. The trace displayed over the full CRT graticule width corresponds to the intensified section of trace displayed during INTEN BY B operation. ALT push-button alternately sweeps the CRT at the main (A) time-base and delayed (B) time-base rates when pressed. This results in twice as many traces displayed on the CRT as are dis­played during any of the sweep modes described above.
START switch When pressed in (TRIG'D position),
causes the B sweep to be triggered by the first trigger pulse occurring after the delay time set by the DLY TIME MULT control. In this position, the delay time is adjustable only in whole increments of the time between trigger pulses. When released (AFTER DELAY position), causes the B sweep to start immediately after the delay time set by the DLY TIME MULT control. In this position, the delay time is adjustable with infinite resolution.
A/B TRACE SEP Permits adjustments of the distance
control between corresponding A and B traces
when the ALT sweep mode is selected.
5
SOURCE switch To select the signal used for A or B
34 36 37 35
36
38 39 40
time-base triggering. CH-1 position selects the channel 1 signal for triggering. CH-2 position selects the channel 2 signal for triggering. ALT position selects the triggering mode that allows a stable display of two asynchronous signals on the CRT. Must be used in conjunction with the ALT vertical mode. LINE position (A only) selects a trigger signal derived from the AC power line, permitting the scope to display stabilized line-related components of a signal even though they may be very small compared to other signal components.
0.2 V/DIV position selects the full signal applied to the EXT TRIG IN connector. 2 V/DIV position selects an attenuated sample of the signal applied to the EXT TRIG IN connector.
CH-3 or EXT TRIG For applying an external signal to the
IN connector oscilloscope for triggering the A
timebase and/or displaying the channel 3 trace.
COUPLING switch To select the frequency characteristics
of the coupling to the trigger circuits. DC position selects direct trigger coupling so all components of the trigger signal are applied to the trigger circuit. AC position inserts a large capacitor in the trigger-coupling chain to remove any DC components from the trigger signal. AC signals below 10 Hz are also attenuated, as is the case in all of the trigger coupling modes listed below. HF-REJ position inserts a filter in the trigger-coupling chain that removes signal components higher in frequency than 35 kHz. TV-V (A only) position inserts a shaping filter (TV sync separator) whose low-frequency output (vertical sync pulses) is used for triggering. This trigger mode will also pass and differentiate waveforms in the 1-100 Hz range. TV-H position inserts a shaping filter (TV sync separator) whose high­frequency output (horizontal sync pulses) is used for triggering. This trigger mode will also pass and differentiate waveforms in the 2-500 kHz range.
When triggered B sweep is selected as the horizontal-display mode, and the COUPLING switch is set to any position other than TV-V, the A- and B- time base trigger signals are identical. However, when the COUPLING switch is set to TV-V during triggered B sweep, the TV-V shaping filter is inserted only in the A­time base trigger signal. The trigger signal fed to the B time base will be shaped by the TV-H filter.
SWEEP MODE To select the triggering mode. AUTO
switches push-button allows sweep to flee-run
and display a base-line in the absence of signal when pressed. Automatically switches to triggered sweep mode when signal of 20 Hz or higher is present and other trigger controls are properly set. NORM push-button produces sweep only when signal is present and other triggering controls are properly set. No trace is visible if any trigger requirement is missing.
SINGLE pushbutton disables recurrent sweep operation when pressed. The sweep generator can then be manually reset before each sweep by depressing this switch again. No trace is visible before or after sweep occurs.
READY lamp Indicates when sweep generator is
armed for single-sweep operation. Lamp is extinguished at start of sweep.
SLOPE switch Push-button selects the positive or
negative slope of the trigger signal for initiating sweep. +position n causes triggering on the positive-going edge or slope of the trigger signal. - .position causes triggering on the negative-going edge or slope of the trigger signal.
LEVEL control Selects the amplitude level at which the
sweep is triggered. When rotated clockwise (+ direction), the trigger point moves towards the positive peak of the trigger signal. When rotated counterclockwise (- direction), the trigger point moves towards the negative peak of the trigger signal. Pulling the HOLDOFF knob (concentric to the LEVEL control) selects PRESET level, a trigger point near the zero-crossing point of the trigger waveform.
6
TRIG'D lamp Indicates when the sweep generator is
41 42 46 47 48 43 44 45
being triggered.
HOLDOFF control Allows triggering on certain complex
signals by changing the hold-off (dead) time of the main (A) sweep. This avoids triggering on intermediate trigger points within the repetition cycle of the desired display. NORM is a detented position at full CCW rotation that is used for ordinary signals. B ENDS A is a detented position at full clockwise rotation that increases the A timebase repetition rate to the maximum to improve the apparent brightness of "short" duration (low­duty cycle) pulses in the delayed mode.
2-1-4 Miscellaneous
Refer to Figure 2-4 for references (43) to (48).
Z AXIS INPUT For applying signal to intensity Connector modulate the CRT.
FUSE Receptacle permits quick fuse re-
placement without opening case
Power Connector Permits removal or replacement of AC
power cord.
Voltage Label Indicates the voltage of the oscil-
loscope's primary wiring.
Cord Caddy Provides a quick method of securing
the power cord, and supports the oscilloscope for vertical operation.
Feet Supports the oscilloscope for shelf
mounting.
2-2. INITIAL OPERATION
Before the instrument is operated for the first time, perform the following procedures in the order listed to ensure satisfaction and prevent damage to the instrument.
2-2-1 Power Connections and Adjustments
The instrument is normally shipped wired for a 120-volt power source but can be adapted to operate from power sources with - 10% of the rated values given in Table 2-1. Operation with a voltage less that t0% of the rated value may result in improper performance of the instrument and a voltage more than 10% in excess of the rated value may damage the power supply circuitry.
To check or alter the power-transformer wiring, proceed as follows:
1. Disconnect the power cord from the Power Connector (45).
2. Remove the 14 Phillips-head screws around the periphery of the
top and bottom covers.
3. Remove the covers and carefully turn the LBO-516 upside
down.
7
Loading...
+ 21 hidden pages