LEADER FS 3030 Instruction Manual

Page 1
0802220
FS 3030
CINELITE for the LV 5700A
(S.O.No.0802220)
Instruction Manual
LEADER ELECTRONICS CORP.
Page 2
0802220
1. FEATURES....................................................................................................................... 1-1
2. SPECIFICATIONS ............................................................................................................ 2-1
2.1 %DISPLAY Mode .................................................................................................... 2-1
2.2 f_Stop Mode ............................................................................................................ 2-1
2.3 Level Warning Display When Using the CINELITE Function .................................. 2-1
2.4 General Specifications............................................................................................. 2-1
3. CINELITE INSTALLATION ............................................................................................... 3-1
3.1 Before Installation.................................................................................................... 3-1
3.2 Installation Procedure.............................................................................................. 3-1
4. OPERATING PROCEDURE............................................................................................. 4-1
4.1 Picture Display Menu .............................................................................................. 4-1
4.2 f_Stop/%DISPLAY Function Selection .................................................................... 4-2
4.3 f_Stop Value Display Function................................................................................. 4-3
4.3.1 f_Stop Setup Menu ...................................................................................... 4-3
4.3.2 Cursor Movement ........................................................................................ 4-5
4.3.3 USER_CAL Setup Menu ............................................................................. 4-6
4.3.4 CAL_Table Setup and USER Mode f_Stop Measurement........................... 4-7
4.3.5 CAL_Table Setting Rules............................................................................. 4-8
4.4 %DISPLAY Displaying Function.............................................................................. 4-8
4.4.1 %DISPLAY Setup Menu .............................................................................. 4-9
4.4.2 Cursor Movement ........................................................................................ 4-10
Page 3
0802220
The FS 3030 CINELITE for the LV 5700A is an exclusive software application for the LV 5700A.
For information on how to use the LV 5700A, see the LV 5700A Instruction Manual.
1. FEATURES
Displays the luminance of each component of the video signal (luminance component) in terms
of a camera aperture (f_Stop), luminance level (percentage level), or RGB level.
(1) You can select aperture (f_Stop), luminance level percentage (% Display), or RGB level for
displaying the luminance information. For RGB level, you can select RGB percentage or
RGB tone display.
(2) Displays the f_Stop value according to the CAL Table setting (GAMMA correction 0.45
(ITU-R BT709) or a user-defined correction table USER1 to USER3).
(3) Measures and displays luminance information at any location of the video using the X and
Y cursors.
(4) Displays the measured values in real time when shooting a video.
(5) Displays luminance information at the video signal reference point and at the comparison
point as well as the difference in the luminance levels between the two points (diff).
1-1
Page 4
0802220
2. SPECIFICATIONS
2.1 %DISPLAY Mode
(1) Luminance Level Display
Displays the luminance level at the intersection of the X and Y cursors as a percentage.
0%: Y = 64 (040h) 100%: Y=940 (3ACh)
(2) RGB % Display
Displays the RGB level at the intersection of the X and Y cursors as a percentage.
(3) RGB Tone Display
Displays the RGB level at the intersection of the X and Y cursors using 256 tone
levels.
R = (R% × 255)/100 (R=0 when R%<0, R=255 when R%>100) G = (G% × 255)/100 (G=0 when G%<0, G=255 when G%>100) B = (B% × 255)/100 (B=0 when B%<0, B=255 when B%>100)
2.2 f_Stop Mode
Displays the f_Stop value after performing GAMMA correction with respect to the 18 %
REF_SET reference level. (1) 0.45: Luminance Correction Table for Standard TV Cameras
Displays the f_Stop value after ITU-R BT709 correction If the reference level is set to REF18%
1) When level % 0 F_Stop = Not measurable
2) In case of 0 < level % < 8.1% f_Stop = (Log((LEVEL%/100)/4.5)/Log2) - REF18%
3) When level % 8.1 % f_Stop = (Log((((LEVEL%/100)+0.099)/1.099)^(1/0.45))/Log2) - REF18%
(2) USER1, USER2, USER3: The camera luminance correction table can be stored to a
preset.
Linearly interpolates between each corrected values
2.3 Level Warning Display When Using the CINELITE Function
When using the ftop, level %, RGB %, or RGB tone display, the measured values are displayed in yellow as a warning if the level % value at the X and Y cursor position is as
follows
(1) Level % ≤ 0 %
(2) Level % ≥ 80 %
2.4 General Specifications
Accessories Instruction Manual ....................... 1
CINELITE label............................1
2-1
Page 5
0802220
3. CINELITE INSTALLATION
You can use the CINELITE features by installing the software from the CF card. Follow the procedure below to install the software.
3.1 Before Installation
(1) The software is exclusive to the LV 5700A. Do not connect the CF Card to another
device.
(2) Copying the contents of the CF card without LEADER’s consent is prohibited. Note
that even if you copy the contents to another CF card, you will not be able to install CINELITE from that card.
(3) Once you install CINELITE to one LV 5700A, you will not be able to install it to another
LV 5700A using the same CF card.
(4) Follow the precautions below when handling the CF card.
• Do not touch the connector.
• Avoid exposing the CF card to heat, moisture, or direct sunlight.
• Do not apply shock to it.
• Do not remove it while it is being accessed.
3.2 Installation Procedure
(1) Check that the LV 5700A firmware version is 8.3 or later.
On the LV 5700A startup screen, check that the firmware version is 8.3 or later. You cannot install CINELITE if the firmware version is 8.2 or earlier. The startup screen appears after you turn on the power.
(2) Turn the LV 5700A off.
(3) Connect the CF card to the connector on the rear panel, and turn the power on.
When the power is turned on, the CINELITE features are automatically installed. When the measurement screen appears, the installation is complete. Do not remove the CF card while the LV 5700A is starting up.
(4) Check that the word “CINELITE” is displayed on the startup screen.
To confirm the installation, turn the power off, and then turn it back on. Check that the word “CINELITE” appears below the model name on the startup screen. Then, remove the CF card.
(5) Affix the CINELITE label to the front panel.
The figure below indicates where you can affix the label.
3-1
CINELITE label
Page 6
0802220
4. OPERATING PROCEDURE
4.1 Picture Display Menu
The picture display menu is shown below. The section enclosed in broken lines contains the menu items that have been added to the standard model.
PIC F1
F2
F4
F5
MARKER
LINE SELECT
f_Stop DISPLAY
%DISPLAY
F1 ( ON / OFF )
F2 ( ON / OFF )
F3
F4
F5
F1
F2
F7
F1
F2 ( ON / OFF )
F3
F4
F5
F6 F1 ( LINE / SAMPLE )
F7
F1
F2
F3
F4
F7
4:3
SAFE ACTION
SAFE TITLE
CENTER
up menu
LINE SELECT
FIELD
up menu
F.D
REF_POS DISPLAY
REF_POS SET
18% REF_SET
GAMMA
CAL SET
up menu
F.D
REF_POS DISPLAY
REF_POS SET
%/RGB SELECT
up menu
( ON / OFF )
( ON / OFF )
( ON / OFF )
( FIELD1 / FIELD2 / FRAME )
( LINE / SAMPLE )
( 0.45 / USER1 / USER2 / USER3 )
F.D
Table
F2
AllClear
1 Data
F3
Clear
CAL
F5
SET
F6
CAL_F
up
F7
menu
( LINE / SAMPLE )
( ON / OFF )
( LEVEL% / RGB% / RGB 255 )
( 22.0 / 16.0 / 11.0 / 8.0 / 5.6 / 4.0 / 2.8 / 2.0 )
4-1
Page 7
0802220
4.2 f_Stop/%DISPLAY Function Selection
(1) Picture Screen
Press the PICTURE key on the panel to display the picture screen shown below.
(2) F4 f_Stop DISPLAY
The screen switches to f_Stop display mode.
1) LINE SELECT is forced to ON.
2) X and Y cursors as well as the f_Stop value are displayed.
(3) F5 %DISPLAY
The screen switches to %Display (luminance signal level) display mode.
1) LINE SELECT is forced to ON.
2) X and Y cursors as well as the %Display (VIDEO luminance level) value are displayed.
3) The setting of F4 %/RGB is switched to LEVEL%, RGB% or RGB255.
4-2
Page 8
0802220
4.3 f_Stop Value Display Function
[ PICTURE → F4 f_Stop DISPLAY ] Displays the f_Stop value at the cross-point of LINE_No (Y cursor) and SAMPLE_No (X cursor). Displays the Y (luminance) level (=d_Lev) of the video signal indicated by LINE_No and SAMPLE_No as well as the calculated value of the Y (luminance) level (=d_Lev) by taking the reference value (d_REF) when the F4 18% REF_SET is pressed to be 0. This value indicates the aperture of the camera.
(1) If F•5 GAMMA=0.45
Displays the f_Stop value after ITU-R BT709 correction. If the reference level is set to REF18% (d_REF).
1) When LEVEL % 0
f_Stop = Not measureable
2) In case of 0 < LEVEL % < 8.1%
f_Stop = (Log((LEVEL%/100)/4.5)/Log2) - d_REF
3) When LEVEL% ≥ 8.1%
f_Stop = (Log((((LEVEL%/100)+0.099)/1.099)^(1/0.45))/Log2) - d_REF
(2) If F5 GAMMA=USER
Displays the value that is obtained by linearly approximating the values between the table values using the CAL table that you set using the F6 CAL_SET function menu.
When the luminance level (LEVEL%) at the intersection of the X and Y cursors is 0% or less or 80% or more, the f_Stop value display is displayed in yellow. Others are displayed in white.
4.3.1 f_Stop Setup Menu
Press PICTURE → F•4 f_Stop DISPLAY to display the LINE_No (Y) and SAMPLE_No (X) cursors as well as the LINE_No and SAMPLE_No at the lower left of the screen indicating the X and Y cursor positions. The f_Stop value is displayed near the cross-point of the X and Y cursors.
Note: Because the X and Y cursors indicate the positions represented by LINE_No and
SAMPLE_No, the cursors and luminance level are not displayed for values in the blanking period.
4-3
Page 9
0802220
LINE
SAMPLE
FD
ON
OFF
REF_POS
DISPLAY
REF_POS
SET
18%
REF_SET
0.45 USER1 USER2 USER3
GAMMA
CAL SET
up
menu
f_Stop setup menu
(1) F1 FD LINE_No and SAMPLE_No Selection
Turn the function dial (FD) to change the LINE_No (Y-axis cursor) or SAMPLE_No (X-axis cursor) position. If F1 FD is set to LINE, the LINE_No (horizontal cursor) position is changed; if F1 FD is set to SAMPLE, the SAMPLE_No (vertical cursor) position is changed.
LINE: You can change the line number using the FD. The Y-axis cursor
moves to the position indicated by LINE_No.
SAMPLE: You can change the sample number of the specified line using the F•D.
The X-axis cursor moves to the position indicated by SAMPLE_No.
(2) F•2 REF_POS DISPLAY Reference Position Measurement ON/OFF Setting
Turns ON/OFF the REF_POS (reference position) measurement and display.
REF_POS DISPLAY=OFF: Disables the f_Stop value display at the REF_POS
(reference position). Only the f_Stop value at the cursor position is displayed.
REF_POS DISPLAY=ON: Enables the f_Stop value display at the REF_POS
(reference position). A crosshair cursor is displayed at the REF_POS (reference position). f_Stop REF measurement value and Diff (f_Stop-REF) are displayed at the top section of the function display.
(3) F3 REF_POS SET Reference Position Setting
Move the X and Y cursors to the reference level position using the FD. Then, press F3 REF_POS SET to set the reference position to the position at the X and Y cursors, and fix the crosshair cursor at that position. From this point, the f_Stop values at the crosshair cursor position specified by FD_KNOB and at the REF_POS (reference position) are measured and displayed.
(4) F4 18% REF_SET Reference Level Setting of the f_Stop Value
Move the X and Y cursors to the reference level position using the FD. Then, press F4 18% REF_SET to set the illuminance at the position specified by the X and Y cursors to the reference (0.0). From this point, the f_Stop value indicates the relative value with respect to the reference illuminance.
4-4
Page 10
0802220
(5) F5 GAMMA f_Stop Value Calculation Method Selection
Selects the method used to calculate the f_Stop value at the position specified by the X and Y cursors.
0.45: Determines the illuminance using the calculating equation of the reference. Displays the f_Stop value after ITU-R BT709 correction. If the reference level is set to REF18%.
1) When LEVEL % 0 f_Stop = Not measureable
2) In case of 0 < LEVEL % < 8.1% f_Stop = (Log((LEVEL%/100)/4.5)/Log2) - REF18%
3) When LEVEL% ≥ 8.1% f_Stop = (Log((((LEVEL%/100)+0.099)/1.099)^(1/0.45))/Log2) - REF18%
USER1: Calculates the illuminance using the user-defined USER1_CAL_Table.
USER2: Calculates the illuminance using the user-defined USER2_CAL_Table.
USER3: Calculates the illuminance using the user-defined USER3_CAL_Table.
(6) F•6 CAL SET (Setting of CAL Table)
F6 CAL_SET appears only if GAMMA is set to a user-defined setting (USER1, USER2, or USER3). F6 CAL_SET is used to set the calibration table.
4.3.2 Cursor Movement
Use the (1) V_POS, (2) H_POS, and (3) F•D controls to move the cursors.
(1) V_POS Control
Varies the vertical cursor (LINE_No). Turn the control clockwise to move the cursor up (decrease the Line Select Number). (This is opposite to the operation of F•D_KNOB.) Press the V_POS control to move the vertical cursor to the center of the picture display.
(2) H_POS Control
Varies the horizontal cursor (SAMPLE_No). Press the H_POS control to move the horizontal cursor to the center of the picture display.
(3) FD Control
Press F1 FD or FD_KNOB to switch the cursor that is moved when you operate the FD control between LINE (vertical cursor (LINE_No)) 1) and SAMPLE (horizontal cursor (SAMPLE_No)) 2). By pressing FD_KNOB to switch the function, you can carry out all cursor operations using only F.D_KNOB.
4-5
Page 11
0802220
4.3.3 USER_CAL Setup Menu
Press PICTURE → F•4 f_Stop DISPLAY → F•5 GAMMA to select USER1, USER2, or USER3. Then, press F•6 CAL_SET to display the USER_CAL setup screen. LINE_No (Y cursor) and SAMPLE_No (X cursor) as well as the f_Stop value at the position specified by the X and Y cursors and the video level (Lv) at the cursor position are displayed. The current defined calibration table is displayed at the lower left.
Note: Because the X and Y cursors indicate the positions represented by LINE_No and
SAMPLE_No, the cursors and f_Stop value are not displayed for values in the blanking period.
LINE
SAMPLE
FD
AllClear
Table
1 Data
Clear
CAL SET
CAL_F
22.0
up
menu
USER_CAL setup menu
(1) F1 F.D (LINE_No and SAMPLE_No Selection)
To change the line number (Y-axis cursor) or the sample number (X-axis cursor), turn the function dial (FD). If F1 F.D is set to LINE, the line number (horizontal cursor) will change. If it is set to SAMPLE, the sample number (vertical cursor) will change.
LINE: Turn the function dial to control the line number. The Y-axis moves to
the position indicated by LINE_No.
SAMPLE: Turn the function dial to control the sample number.
The X-axis moves to the position indicated by SAMPLE_No.
4-6
Page 12
0802220
(2) F2 Table All Clear: CAL Table Clear
Clears the entire calibration table. To set a new table, be sure to press F2 Table All Clear to initialize the table. (Correct measured values will not be displayed if unnecessary data is present in the table.)
(3) F3 1 Data Clear: Clears one line of table data specified by F6 CAL_F from the
calibration table.
(4) F5 CAL_SET: Assigns the video level at the cursor position to the table
position specified by F6 CAL_F.
(5) F6 CAL_F: Press F6 CAL_F to select F6. Then, turn the FD control to
change the CAL_F value among 2.0, 2.8, 4.0, 5.6, 8.0, 11.0,
16.0, and 22.0. This value indicates the position in the CAL table that you are setting.
(6) F7 up_menu: Returns to the f_Stop setup menu.
4.3.4 CAL_Table Setup and USER Mode f_Stop Measurement
Press PICTURE → F•4 f_Stop DISPLAY → F•5 GAMMA to select USER1, USER2, or USER3. Then, press F6 CAL_SET to display the function menu as explained in section
3.3.3. The CAL_Table shown below appears at the lower left of the screen.
(1) [USERx]: Displays the selected table number.
(2) REF=x.x: Reference f_Stop Value
Displays the F value when F4 18% REF_SET was pressed in the f_Stop setup menu. Example: In the display above, let us assume that the f_Stop value for
[5.6] F=4.0 and Lev=416 is set to 0.0. The displayed values are as follows depending on the video level at the position specified by the X and Y cursors.
f_Stop = -3.0 for Lv = 152 f_Stop = -2.0 for Lv = 240 f_Stop = -1.0 for Lv = 328 f_Stop = 0.0 for Lv = 416 f_Stop = 1.0 for Lv = 504 f_Stop = 2.0 for Lv = 592
4-7
Page 13
0802220
(3) CAL_F: CAL_Table Setup Position
(4) F (0.0 to 7.0) Calculates the user-defined table’s F value relative to the 18%
5) Lev Displays the video level Y (luminance level).
4.3.5 CAL_Table Setting Rules
In the user-defined measurement, the levels set in the CAL_Table are interpolated using linear approximation. Set the values in the table according to the rules given below. Correct measurements may not be displayed depending on how the table is set. (1) Be sure to set the Lev values in the table consecutively.
Setup Example
[22.0] 0.0 0 [16.0] 1.0 0 [11.0] 2.0 64 [ 8.0] 3.0 102 [ 5.6] 4.0 240 [ 4.0] 5.0 372 [ 2.8] 6.0 504
[ 2.0] 7.0 0 The measurements are valid for Y levels 0 to 504 in the setup above. Do not create unspecified data in the 8.0 to 4.0 range.
(2) The Lev settings in the table must increase monotonically.
In (1) above, Lev(11.0) < Lev(8.0) < Lev(5.6) < Lev(4.0) < Lev(2.8).
The position specified by the F6 CAL_F is assigned the Lv value that is the current cursor position.
REF_SET value. Displays the difference between the F value when you pressed F•4 18% REF_SET in the f_Stop setup menu and the F value that is determined from the Lv value at the cursor position.
4.4 %DISPLAY Displaying Function
[ PICTURE → F5 %DISPLAY ]
Displays the luminance level as a percentage at the cross-point of LINE_No (Y cursor) and SAMPLE_No (X cursor).
(1) If F2 REF_POS_DISPLAY = OFF
Displays only the luminance level as a percentage at the cross-point of LINE_No (Y cursor) and SAMPLE_No (X cursor).
(2) If F2 REF_POS_DISPLAY = ON
Displays the luminance level at the reference position (Ref. Level) and the luminance difference (Diff) between the cursor position and the reference position in addition to the luminance level display at the cross-point of LINE_No (Y cursor) and SAMPLE_No (X cursor).
(3) If F4 %RGB SELECT = LEVEL %
Displays the luminance level at the intersection of the cursor and at the REF_POS position.
4-8
Page 14
0802220
(4) If F4 %RGB SELECT = RGB %
Displays the RGB % level at the intersection of the cursor and at the REF_POS position.
(5) If F4 %RGB SELECT = RGB 255
Displays the RGB 256 tone levels at the intersection of the cursor and at the REF_POS position.
4.4.1 %DISPLAY Setup Menu
Press PICTURE → F•5 %DISPLAY to display the LINE_No (Y) and SAMPLE_No (X) cursors as well as the LINE_No and SAMPLE_No at the lower left of the screen indicating the X and Y cursor positions. The luminance level is displayed near the cross-point of the X and Y cursors.
Note: Because the X and Y cursors indicate the positions represented by LINE_No and
SAMPLE_No, the cursors and luminance level are not displayed for values in the blanking period.
LINE
SAMPLE
FD
(1) F1 F•D (LINE_No and SAMPLE_No Selection)
ON
OFF
REF_POS
DISPLAY
LEVEL%
RGB%
RGB 255
REF_POS
SET
%/RGB
SELECT
%Display setup menu
up
menu
To change the line number (Y-axis cursor) or the sample number (X-axis cursor), turn the function dial (FD). If F1 FD is set to LINE, the line number (horizontal cursor) will change. If it is set to SAMPLE, the sample number (vertical cursor) will change.
4-9
Page 15
0802220
LINE: Turn the function dial to control the line number. The Y-axis moves to
the position indicated by LINE_No.
SAMPLE: Turn the function dial to control the sample number.
The X-axis moves to the position indicated by SAMPLE_No.
(2) F2 REF_POS DISPLAY (Reference Position Measurement ON/OFF Setting)
Turns on/off the REF_POS (reference position) measurement and display.
REF_POS DISPLAY=OFF: Disables the reference luminance level display at
REF_POS (reference position). Only the luminance level at the cursor position is displayed.
REF_POS DISPLAY=ON: Enables the reference luminance level display at
REF_POS (reference position). A crosshair cursor is displayed at REF_POS (reference position). The REF.Level (reference luminance level) and Diff (luminance level - reference luminance level) are displayed at the top section of the function display.
(3) F3 REF_POS SET (Reference Position Setting)
Move the X and Y cursors to the reference level position using the function dial. Then, press F3 REF_POS_SET to set the reference position at the intersection of
X and Y cursors and to fix the crosshair cursor at that position. From this point, the LV 5750 measures and displays the luminance level at the crosshair cursor position that is specified by the function dial and the level at REF_POS (reference position).
(4) F4 %/RGB SELECT (Display Mode Setting)
Sets the measured values that are displayed at the intersection of the X and Y
cursors and at the REF_POS_DISPLAY position to:
1) LEVEL%: Luminance level Y
2) RGB%: R, G, and B percentage levels
3) RGB 255: R, G, B levels using 256 tone levels
4.4.2 Cursor Movement
Use the (1) V_POS, (2) H_POS, and (3) F•D controls to move the cursors.
(1) V_POS Control
Varies the vertical cursor (LINE_No). Turn the control clockwise to move the cursor up (decrease the line select number). (This is opposite to the operation of F•D_KNOB.) Press the V_POS control to move the vertical cursor to the center of the Picture display.
(2) H_POS Control
Varies the horizontal cursor (SAMPLE_No). Press the H_POS control to move the horizontal cursor to the center of the Picture display.
4-10
Page 16
0802220
(3) FD Control
Press F1 FD or FD_KNOB to switch the cursor that is moved when you operate the FD control between LINE (vertical cursor (LINE_No)) 1) and SAMPLE (horizontal cursor (SAMPLE_No)) 2).
By pressing FD_KNOB to switch the function, you can carry out all cursor operations using only F.D_KNOB.
4-11
May 22, 2008 08050.001K TS 1