Product Troubles and Solutions
Product Troubles and Solutions
Trouble
Cause
Solution
The unit can not be
turned on, abnormal
display or no use
of pressing keys
Battery is powered off
Take out adaptor and batteries,after 1 min,
put in batteries and connect adaptor again
Battery is powered off or
not installed well
Charge it in time or check if two
polars of batteries matched right
Laser output
power reduced
Laser port is dirty
Insufficient battery power
Clean laser port with alcohol
Charge batteries
External electrostatic or
electromagnetic interference
Take out batteries and
put in again, turn it on
Power is still low after
above measures
Change Laser Output Wire (laser output
power will be reduced after long-time use )
Low-frequency
pulse stimulation
is weak or there is
no stimulation
The low-frequency wire plug not
inserted electrode or Bandangle
plug well
Insufficient battery power
Charge batteries
Electrodes or Conductive
Rubber Mat not stick skin well
Stick electrodes or Conductive
Rubber Mat on skin w ell
Electrodes fold
Separate the electrodes and stick again
There is something w rong with
electrodes or Conductive Rubber
Mat
Change electrodes or
Conductive Rubber Mat
Low-frequency pulse
intensity is too low
Adjust low-frequency
pulse intensity
Skin
rubefaction
Skin is too dry with bad
conductivity
Moisture skin with some clean
water before sticking electrodes
Low-frequency pulse
treatment time is too long
Reduce low-frequency treatment time
Electrodes or Conductive Rubber
Mat not stick well on the skin
Electrodes is broken
(consumable )
Low-frequency pulse
intensity is too high
Lower low-frequency pulse intensity
(Higher intensity not means better effect)
Stick electrodes or Conductive
Rubber Mat well on the skin
Electrodes
sticking side is dirty
or Conductive Rubber
Mat
Clean it according to Maintenance
Change electrodes of the same size
(improper size probably cause skin burn)
If there is any trouble with the unit, turn it off and check it according to
, if it is confirmed unsolved, please
contact distributor for check and repair.
Please do not assemble it by yourself, which will lead to unsafety.
Product Troubles and Solutions
The Bandangle plug not inserted
socket of Rubber Mat well
Reinsert Bandangle plug well
Reinsert low-frequency
wire plug well