LaMotte WAT-MP-DR User Manual

Page 1
pH W aterproof Testr 1
5-0008
HOLD
CON
CAL
ON
OFF
pHTestr 1
WATERPROOF
© 1999 LaMotte Company • Printed in USA • 12/02 • Code 65-0008
Page 2
Specifications
Range: –1.0 to 15.0 pH extended range
Resolution: 0.1 pH
Accuracy: ±0.2 pH
Operating Temperature: 0 to 50 °C, 32 to 122 °C
Battery/Life: Three 1.5V Eveready A76BP/24 hours
or equivalent
Size (meter only): 6.5" X 1.5"
Weight (meter only): 3.25 oz (90 g)
Before First Use
1. 2.
Remove the cap from the bottom of the Testr to expose the electrode bulb and reference junction.
Do not be alarmed if white crystals are present in the cap or electrode assembly . They will dissolve during the pre-soak.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
Soak the electrode in about 1" of electrode storage solution, pH 4 bufferortapwaterfor at least 30 minutes.
If the electrode dries out between uses, recondition it by following this soaking procedure.
pHTestr1
30
CON
HOLD
CAL
ON
OFF
WATERPROOF
1"
2
Page 3
Calibration
Calibration should be done regularly, typically every day that the Testr is used.
1. 2.
Dip1/2"to1"ofthe Turn the Testr on by pressing the ON/OFF button.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
electrode into a pH 7.0
buffer for general purpose
use, into a pH 4.0 buffer
solution if testing acidic
solutions, or a pH 10.0
buffer if testing
alkaline (or basic)
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
solutions. If testing
solutions close to
7.0, use the 7.0
buffer for
1/2"
to 1"
calibration.
3. 4.
Press the CAL button to enter the Calibrate (CA) mode.
The display will flash "CA" and then a pH value close to the buffer value will flash repeatedly .
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
After about 30 seconds, press the HOLD/CON button to confirm the calibration.
"CON" will appear on the display . It will then switch back toapHreadingof the buffer.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
30
5.
Rinse the electrode with deionized water or tap water.
No further calibration is needed if all measurements are expected to
CON
HOLD
CAL
ON
OFF
be between pH 6 and pH 8 for general purpose use. Perform the calibration using pH 4.0 buffer if readings below pH 6 are expected.
pHTestr1
WATERPROOF
Perform the calibration using pH
10.0 buffer if readings above pH 8 are expected.
3
Page 4
pH Testing
1. 2.
Remove the cap from the Testr.
CON
HOLD
CAL
ON
OFF
Press the On/Off button to turn the Testr on.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
3. 4.
Dip1/2"to1"ofthe electrode into the solution.
1/2"
to 1"
5.
Record the pH.
Or press the HOLD/CON button to freeze the reading. Press the HOLD/CON button again to release the reading.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
5.1
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
Stir once with the probe and let the reading stabilize.
6.
Press the ON/OFF button to turn the Testr off.
If a button is not pushed for 8.5 minutes, the meter will automatically shut off.
pHTestr1
WATERPROOF
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
4
Page 5
Maintenance
1. 2.
Rinse the electrode with deionized waterortapwater after each measurement.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
In aggressive chemicals, dirty, or viscous solutions, and solutions with heavy metals or proteins, take readings quickly and rinse the electrode immediately after use.
3. 4.
Periodic soaking in warm pH 4 buffer will help remove any contaminants that will ruin the electrode.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
Keep a small piece of paper or sponge, moistened with tap water or pH 4 buffer, in the Testr cap to keep the probe moist during storage.
5.
Electrode life will be shortened when a Testr is used in applications where the electrode is exposed to materials that will contaminate the electrode reference junction.
5
Page 6
Changing the Batteries
The battery compartment is located under the cap with the lanyard loop.
1. 2.
Firmly grip the cap and the Testr body with dry hands. Slowly twist the cap counter-clockwise. The cap must be tight to ensure a watertight seal so it will require some effort to loosen it.
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
Remove the old batteries by pulling up the cloth tab.
+-+-+
3. 4.
Reposition cloth tab and insert new batteries. Note polarity. Tuck in the cloth tab.
-
+-+-+
Replace the cap. Tighten until the black O-ring is completely hidden under the cap and the lanyard loop is lined up with the face plate.
-
CON
HOLD
CAL
ON
OFF
pHTestr1
WATERPROOF
6
Page 7
Electrode Replacement
Replace the electrode module when the Testr fails to calibrate, gives fluctuating readings in buffers, shows error messages “E2" or ”OR", and the procedures in the maintenance section fail to resolve the problem.
1. 2.
Place the rubber gripper band overthecollaroftheTestr.
Firmly grip the rubber gripper band and the ribbed collar of the Testr with dry hands.
pHTestr1
ON
HOLD
CAL
OFF
WATERPROOF
CON
3. 4.
Slowly twist counter-clockwise to loosen the collar.
pHTestr1
ON
HOLD
CAL
OFF
WATERPROOF
CON
5.
Pull the electrode module straight out from the bottom of the Testr.
pHTestr1
ON
HOLD
CAL
OFF
WATERPROOF
CON
Unscrew the ribbed electrode module collar and remove it. Save the collar and the O-ring.
6.
Align the four tabs on the new electrode module to match the four slots on the Testr.
pHTestr1
ON OFF
WATERPROOF
small o-ring (hidden)
pHTestr1
WATERPROOF
pHTestr1
ON OFF
WATERPROOF
HOLD
CAL
CON
ON
HOLD
CAL
OFF
CON
HOLD
CAL
CON
7
Page 8
Electrode Replacement
Note: For older models with only two slots, break off the two small tabs using needle-nosed pliers.
7. 8.
Put the smaller O-ring all the way into the ribbed collar.
Self-Diagnostic Messages
6.
Gently push the electrode module into the bottom of the Testr to fully seat it in position.
pHTestr1
ON
HOLD
CAL
OFF
WATERPROOF
Push the collar on over the module. Thread it into place by firmly twisting it clockwise until the larger O-ring that is near the faceplate disappears under the ribbed electrode module collar.
pHTestr1
WATERPROOF
CON
ON
HOLD
CAL
OFF
CON
ER1 Batteries are low and need replacing
ER2 Wrong or bad buffer value (out of range) or the electrode is
failing. Check buffers or replace electrode.
OR Over range signal from voltage in a solution, the electrode is
not contacting the solution, or the electrode is failing. Check solution or replace electrode.
Accessories
Replacement probe 5-0009
Buffers (4, 7, 10) 2866, 2881, 2896
Mini buffer tablets (4, 7, 10) 3983, 3984, 3985
8
Loading...