SULFIDE TEST KIT
OCTA-SLIDE 2, 0.2-20.0 ppm
CODE 4456-01
QUANTITY CONTENTS CODE
30 mL *Sulfi de Reagent A *4458-G
15 mL *Sulfi de Reagent B *4459-E
60 mL Sulfi de Reagent C 4460-H
2 Test Tubes, 2.5-5-10mL, plastic, w/caps 0106
1 Sulfi de Octa-Slide 2 Bar, 0.2-20.0 ppm 4457-01
1 Octa-Slide 2 Viewer 1101
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com. Search for the four digit reagent code number listed on the reagent label, in the contents list or in the test procedures. Omit any letter that follows or precedes the four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel: (US, 1-800-255-3924) (International, call collect, 813-248-0585).
To order individual reagents or test kit components, use the specifi ed code number.
Note: Collect sample with a minimum of aeration and analyze promptly:
Warning! This set contains chemicals that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision.
USE OF THE OCTA-SLIDE 2 VIEWER
 e Octa-Slide 2 Viewer should be held so non-direct light enters through the back of the Viewer. Viewer. Insert the reacted sample into of the Viewer. Match the color of the to the color standards.
Insert the Octa-Slide 2
Bar into the
the top
reaction
EPA ACCEPTED PROCEDURE
A known concentration of sulfi de in the range of 0.2-20.0 ppm S= may be prepared as a check standard by referring to APHA Standard Methods 4500–S2–D, 17th Ed.,
1989. Run this test on the check standard. If the test result is incorrect, discard the old reagents and order the reagent refi ll package.
PROCEDURE
Note: Collect sample with a minimum of aeration and analyze promptly:
1
Place Sulfi de Octa-Slide 2 Bar (4457-01) into the Octa-Slide 2 Viewer (1101).
45
7
10
Cap and gently invert to mix.
Wait 1 minute.
01:00
Insert test tube into Octa-Slide 2 Viewer (1101).
23
Fill a test tube (0106) to the 5 mL line with the sample water.
Add 3 drops of *Sulfi de Reagent B
(4459).
89
11
Use the pipet (0354) to add 1.0 mL of Sulfi de Reagent C (4460).
Match sample color to a color standard.
Record as ppm Sulfi de.
Add 15 drops of *Sulfi de Reagent A
(4458).
6
Cap and mix.
Cap and mix.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
64456-01 11/19