Packaging & Delivery ····································································5
General Precautions ······································································ 5
Safety Precautions ········································································· 5
Limits of Liability ···········································································5
Specifications ·················································································6
Statistical And Technical Definitions
Related To Product Specifications·················································· 7
Contents and Accessories·······························································8
CE Compliance ··············································································9
CHEMICAL TESTING
An Introduction to Colorimetric Analysis ···································· 10
Reagent Blank ···············································································11
Colorimeter Tubes ·········································································11
Meter Care ·····················································································11
Sample Dilution Techniques & Volumetric Measurements ·········· 12
Interferences ··················································································12
The Keypad ···················································································18
The Display & the Menus ····························································· 19
Looping Menus···············································································21
TESTING
Testing Menu ·················································································22
Sequences of Tests ·········································································23
General Testing Procedures ···························································24
Testing With the LaMotte Pre-Programmed Tests ························ 25
EDITING MENU
Edit a Sequence ············································································· 26
Adding or Deleting Tests ·······························································27
Creating a Soil Test Sequence ························································ 30
Calibrating LaMotte Pre-Programmed Tests·································· 34
Setting the Clock············································································37
Turning the Data Logger On and Off············································· 38
Factory Setup ·················································································· 39
Setting the Power Saver Function··················································39
Replacing the Battery ·····································································40
MAINTENANCE
Cleaning ························································································41
Meter Care ·····················································································41
Meter Disposal················································································41
Multiple Test Procedure··························································· 49
Single Test Procedure·······························································49
Experienced packaging personnel at LaMotte Company assure adequate
protection against normal hazards encountered in transportation of shipments.
After the product leaves the manufacturer, all responsibility for its safe delivery
is assured by the transportation company. Damage claims must be filed
immediately with the transportation company to receive compensation for
damaged goods.
Should it be necessary to return the instrument for repair or servicing, pack
instrument carefully in suitable container with adequate packing material. A
return authorization number must be obtained from LaMotte Company by
calling 1-800-344-3100. Attach a letter with the authorization number to the
shipping carton which describes the kind of trouble experienced. This valuable
information will enable the service department to make the required repairs
more efficiently.
GENERAL PRECAUTIONS
n
Before attempting to set up or operate this instrument it is important to read
the instruction manual. Failure to do so could result in personal injury or
damage to the equipment.
The SMART 2 Colorimeter should not be stored or used in a wet or corrosive
environment. Care should be taken to prevent water or reagent chemicals from
wet colorimeter tubes from entering the colorimeter chamber.
NEVER PUT WET TUBES IN COLORIMETER.
SAFETY PRECAUTIONS
n
Read labels on all reagent containers. Some labels include precautionary
notices and first aid information. Certain reagents are considered hazardous
and are designated with a * in the instruction manual. To view or print a
Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or our
web site. To obtain a printed copy, contact us by e-mail, phone or FAX.
Additional information for all LaMotte reagents is available in the United
States, Canada, Puerto Rico, and the US Virgin Islands from Chem-Tel by
calling 1-800-255-3924. For other areas, call 813-248-0585 collect to contact
Chem-Tel’s International access number. Each reagent can be identified by the
four digit number listed on the upper left corner of the reagent label, in the
contents list and in the test procedures.
Keep equipment and reagent chemicals out of the reach of young children.
Protect Yourself and Equipment: Use Proper Analytical Techniques
LIMITS OF LIABILITY
n
Under no circumstances shall LaMotte Company be liable for loss of life,
property, profits, or other damages incurred through the use or misuse of its
products.
SMART2 SOIL MANUAL 1.075
SPECIFICATIONS
n
n
INSTRUMENT TYPE: Colorimeter
ReadoutGraphical 4 line, 16 character per line LCD
Wavelengths430nm, 520 nm, 570 nm, 620 nm
Wavelength Accuracy±2nm
Readable ResolutionDetermined by reagent system
Wavelength Bandwidth10 nm typical
Photometric Range-2to+2A
Photometric Precision± 0.001A
Sample ChamberAccepts 25 mm diameter flat-bottomed test tubes
Light Sources4 LEDs
Detectors4 silicon photodiodes with integrated interference
filters
ModesAbsorbance, pre-programmed tests
Pre-Programmed TestsYES, with automatic wavelength selection
RS232 Port8 pin mini-DIN, 9600b, 8, 1, n
Power RequirementsBattery Operation: 9voltalkaline
Line Operation: 110/220V AC;
50/60 Hz with adapter, 6V 500 mA DC
Dimensions (LxWxH)8.5 x 16.2 x 16.7 cm, 3.4 x 6.4 x 2.6 inches
Weight312 g, 11 oz (meter only)
Data Logger350 test results stored for download to a PC
6SMART2 SOIL MANUAL 1.07
STATISTICAL AND TECHNICAL DEFINITIONS RELATED
n
TO PRODUCT SPECIFICATIONS
Method Detection Limit (MDL): “The method detection limit (MDL) is
defined as the minimum concentration of a substance that can be measured
and reported with 99% confidence that the analyte concentration is greater
than zero and is determined from analysis of a sample in a given matrix
containing the analyte.”
almost all cases when dealing with a limit of detection or limit of
determination, the primary purpose of determining that limit is to stay away
from it.’”
2
1. CFR 40, part 136, appendix B
2. Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and
L Vanatta, American Laboratory, Sept 2003, P. 31.
Precision: Precision is the numerical agreement between two or more
measurements.
3
The precision can be reported as a range for a measurement
(difference between the min and max). It can also be reported as the standard
deviation or the relative standard deviation. It is a measure of how close
together the measurements are, not how close they are to the correct or true
value. The precision can be very good and the accuracy very bad. This is a useful
measure of the performance of a test method.
3. Skoog, D.A., W e st, D. M., Fundamental of Analytical Chemistry, 2
Holt Rinehart and Winston, Inc, 1969, p. 26.
Accuracy: Accuracy is the nearness of a measurement to the accepted or true
4
value.
The accuracy can be expressed as a range, about the true value, in
which a measurement occurs (i.e. ±0.5 ppm). It can also be expressed as the %
recovery of a know amount of analyte in a determination of the analyte (i.e.
103.5 %). This is a useful measure and what most customers are interested in when
they want to know about the performance of a test method.
4. Skoog D.A., West D. M., Fundamental of Analytical Chemistry, 2
Holt Rinehart and Winston, Inc, 1969, p. 26.
Resolution: Resolution is the smallest discernible difference between any two
measurements that can be made.
places are displayed. (i.e. 0.01). For titrations and various comparators it is the
smallest interval the device is calibrated or marked to (i.e. 1 drop = 10 ppm,
0.2 ppm for a DRT, or ±half a unit difference for an octaslide or color chart).
Note that the resolution many change with concentration or range. In some
cases the resolution may be less than the smallest interval, if it is possible to
make a reading that falls between calibration marks. This is often done with
various comparators. One caveat is, that resolution has very little relationship to
accuracy or precision. The resolution will always be less than the accuracy or
precision but it is not a statistical measure of how well a method of analysis works.
The resolution can be very very good and the accuracy and precision can be very, very
bad! This is not a useful measure of the performance of a test method.
5. Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and
L Vanatta, American Laboratory, Sept 2003, P. 34.
1
Note that, “As Dr. William Horwitz once stated, ‘In
nd
ed.,
nd
ed.,
5
For meters this is usually how many decimal
SMART2 SOIL MANUAL 1.077
Sensitivity: Sensitivity is the resolution based on how this term is used in
LaMotte catalogs. This term is not listed in any of the references. Sometimes it
is used for detection limit. It is a confusing term and should be avoided.
Repeatability: Repeatability is the within-run precision.
6
A run is a single
data set, from set up to clean up. Generally, one run occurs on one day.
However, for meter calibrations, a single calibration is considered a single run
or data set, even though it may take 2 or 3 days.
6. Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s T extbook of
Quantitative Chemical Analysis, 5
th
ed., Longman Scientific & Technical,
1989, p. 130.
Reproducibility: Reproducibility is the between-run precision.
7. Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s T extbook of
Quantitative Chemical Analysis, 5
th
ed., Longman Scientific & Technical,
7
1989, p. 130.
CONTENTS AND ACCESSORIES
n
n
CONTENTS
SMART 2 Colorimeter
Test Tubes, with Caps
Sample Cell Holder , Universal
Power Supply, 110/220V
Battery Charger
SMART 2 Colorimeter Quick Start Guide
SMART 2 Soil Manual
n
ACCESSORIES
Small Field Carrying CaseCode 1919-GCS150
Large Field Carrying CaseCode 1919-BCS440
SMARTLink 2 Program & Interface Cable (3.5 disk)Code 1912-3
SMARTLink 2 Program & Interface Cable (CD)Code 1912-CD
8SMART2 SOIL MANUAL 1.07
CE COMPLIANCE
n
The SMART 2 Colorimeter has earned the European CE Mark of Compliance
for electromagnetic compatibility and safety.
DECLARA TION OF CONFORMITY
Standards to which
Conformity Declared:
Manufacturer's Name:
Manufacturer's Address:
Type of Equipment:
Model Name:
Year of Manufacture:
Testing Performed By:
I, the undersigned, hereby declare that the equipment specified above
Most test substances in water or soil extract are colorless and undetectable to
the human eye. T o test for their presence we must find a way to “see” them.
The SMART 2 Colorimeter can be used to measure any test substance that is
itself colored or can be reacted to produce a color. In fact a simple definition of
colorimetry is “the measurement of color” and a colorimetric method is “any
technique used to evaluate an unknown color in reference to known colors”. In
a colorimetric chemical test the intensity of the color from the reaction must
be proportional to the concentration of the substance being tested. Some
reactions have limitations or variances inherent to them that may give
misleading results. Many such interferences are discussed with each particular
test instruction. In the most basic colorimetric method the reacted test sample
is visually compared to a known color standard. However, accurate and
reproducible results are limited by the eyesight of the analyst, inconsistencies
in the light sources, and the fading of color standards.
To avoid these sources of error, a colorimeter can be used to photoelectrically
measure the amount of colored light absorbed by a colored sample in reference
to a colorless sample (blank).
White light is made up of many different colors or wavelengths of light. A
colored sample typically absorbs only one color or one band of wavelengths
from the white light. Only a small difference would be measured between white
light before it passes through a colored sample versus after it passes through a
colored sample. The reason for this is that the one color absorbed by the
sample is only a small portion of the total amount of light passing through the
sample. However, if we could select only that one color or band of wavelengths
of light to which the test sample is most sensitive, we would see a large
difference between the light before it passes through the sample and after it
passes through the sample.
The SMART 2 Colorimeter passes one of four colored light beams through one
of four optical filters which transmits only one particular color or band of
wavelengths of light to the photodectector where it is measured. The difference
in the amount of colored light transmitted by a colored sample is a
measurement of the amount of colored light absorbed by the sample. In most
colorimetric tests the amount of colored light absorbed is directly proportional
to the concentration of the test factor producing the color and the path length
through the sample. However, for some tests the amount of colored light
absorbed is inversely proportional to the concentration.
The choice of the correct wavelength for testing is important. It is interesting
to note that the wavelength that gives the most sensitivity (lower detection
limit) for a test factor is the complementary color of the test sample. For
example the Nitrate-Nitrogen test produces a pink color proportional to the
nitrate concentration in the sample (the greater the nitrate concentration, the
darker the pink color). A wavelength in the green region should be selected to
analyze this sample since a pinkish-red solution absorbs mostly green light.
10SMART2 SOIL MANUAL 1.07
REAGENT BLANK
n
Some tests will provide greater accuracy if a reagent blank is determined to
compensate for any color or turbidity resulting from the reagents themselves. A
reagent blank is performed by running the test procedure on 10 mL of
demineralized water. Use sample water to SCAN BLANK. Insert the reagent
blank in the colorimeter chamber and select SCAN SAMPLE.Noteresultof
reagent blank. Perform the tests on the sample water as described. Subtract
results of reagent blank from all subsequent test results. NOTE: Some tests
require a reagent blank to be used to SCAN BLANK.
COLORIMETER TUBES
n
Colorimeter tubes which have been scratched through excessive use should be
discarded and replaced with new ones. Dirty tubes should be cleaned on both
the inside and outside. Fingerprints on the exterior of the tubes can cause
excessive light scattering and result in errors. Handle the tubes carefully,
making sure the bottom half of the tube is not handled.
LaMotte Company makes every effort to provide high quality colorimeter
tubes. However, wall thicknesses and diameter of tubes may still vary slightly.
This may lead to slight variations in results (e.g. if a tube is turned while in the
sample chamber, the reading will likely change slightly). To eliminate this error
put the tubes into the sample chamber with the same orientation every time.
The tubes that are included with the colorimeter have an index mark to
facilitate this. If possible, use the same tube to SCAN BLANK and SCANSAMPLE.
METER CARE
n
The optical system of the SMART2 must be kept clean and dry for optimal
performance. Dry the colorimeter tubes before placing them in the chamber to
avoid introducing moisture. For best results store the instrument in an area that
is dry and free from aggressive chemical vapors.
SMART2 SOIL MANUAL 1.0711
SAMPLE DILUTION TECHNIQUES
n
& VOLUMETRIC MEASUREMENTS
If a test result using the SMART 2 Colorimeter gives an OVERRANGE message
then the sample concentration could be over range or under range. If it is over
range, the sample must be diluted. Then the test should be repeated on the
diluted sample to obtain a reading which is in the concentration range for the
test. (Note: This is not true for colorimetric determination of pH.)
Example:
Measure 5 mL of the water sample into a graduated cylinder. Add
demineralized water until the cylinder is filled to the 10 mL line. The sample
has been diluted by one-half, and the dilution factor is therefore 2. Perform the
test procedure, then multiply the resulting concentration by 2 to obtain the
test result.
The following table gives quick reference guidelines on dilutions of various
proportions. All dilutions are based on a 10 mL volume, so several dilutions
will require small volumes of the water sample. Graduated pipets should be
used for all dilutions.
Size of Sample
10 mL0 mL1
5mL5mL2
2.5 mL7.5 mL4
1mL9mL10
0.5 mL9.5 mL20
Deionized Water to Bring
Volume to 10 mLMultiplication Factor
If the above glassware is not available, dilutions can be made with the
colorimeter tube. Fill the tube to the 10 mL line with the sample then transfer
it to another container. Add 10 mL volumes of demineralized water to the
container and mix. T r ansfer back 10 mL of the diluted sample to the tube and
follow the test procedure. Continue diluting and testing until a reading, which
is in the concentration range for the test, is obtained. Be sure to multiply the
concentration found by the dilution factor (the number of total 10 mL volumes
used).
Example:
10 mL of sample is diluted with three 10 mL volumes of demineralized water;
the dilution factor is four.
INTERFERENCES
n
LaMotte reagent systems are designed to minimize most common interferences.
Each individual test instruction discusses interferences unique to that test. Be
aware of possible interferences in the soil extract being tested.
The reagent systems also contain buffers to adjust the soil extract to the ideal
pH for the reaction. It is possible that the buffer capacity of the soil extract may
exceed the buffer capacity of the reagent system and the ideal pH will not be
obtained. If this is suspected, measure the pH of a reacted distilled water
12SMART2 SOIL MANUAL 1.07
reagent blank using a pH meter. This is the ideal pH for the test. Measure the
pH of a reacted soil extract using the pH meter. If the pH is significantly
different from the ideal value, the pH of the sample should be adjusted before
testing.
Interferences due to high concentration of the substance being tested, can be
overcome by sample dilution (see page 11).
CALIBRATION
n
As with all pre-calibrated meters, it is highly recommended, even if not
required by regulations, that the user periodically verify the performance of the
meter by running standards with a predetermined concentration. Results
outside of specification are an indication that the meter needs to be adjusted.
This can be done following the user calibration described on page 34. If the
user calibration fails to properly adjust the meter then the meter should be
returned to LaMotte Company for recalibration. (See page 5).
SMART2 SOIL MANUAL 1.0713
OPERA TION OF THE
SMART 2 COLORIMETER
OVERVIEW
n
The SMART 2 Colorimeter is a portable, microprocessor controlled, direct
reading colorimeter. It has a graphical 4 line, 16 character liquid crystal display
for graphical, alphabetical and numerical messages. The operation is controlled
with the keypad through menu driven software in response to selections shown
on the display.
The test library consists of 100 LaMotte tests (not all 100 may be available at
present) and 10 “User Tests”. The LaMotte tests are precalibrated for LaMotte
reagent systems. The colorimeter displays the results of these tests directly in
units of concentration. The 10 “User Tests” may be used to enter additional
calibrations. All of these tests may be arranged in any of 3 sequences. These
sequences can be modified a limitless number of times to meet changing testing
needs.
The optics feature 4 different colored LEDs. Each LED has a corresponding
silicon photodiode with an integrated interference filter. The interference
filters select a narrow band of light from the corresponding LED for the
colorimetric measurements. The microprocessor automatically selects the
correct LED/photodiode combination for a test.
A RS-232 serial port on the back of the colorimeter, and optional software,
allows the SMAR T 2 to be interfaced with an IBM compatible personal
computer for real time data acquisition and data storage. This port also allows
an interface with a RS-232 serial printer.
Due to its portability, alternate power sources, and rugged construction, the
SMART 2 Colorimeter is ideal for lab and field use.
POWER SOURCE
n
The SMART 2 Colorimeter uses a 6V 500 mA AC adapter. Please refer to the
Parts List for the code number for the correct adapter.
USE OF ANY AC ADAPTER OTHER THAN THE ONE SPECIFIED
FOR USE WITH THE SMART 2 COLORIMETER MAY DAMAGE
THE METER AND WILL VOID THE WARRANTY. Do not use the
adapter sold with the original SMART Colorimeter.
To use the adapter, slide the connector pin from the AC adapter into the small
hole on the left side of the meter. Plug the AC adapter into an appropriate wall
socket or power source.
14SMART2 SOIL MANUAL 1/07
COMPONENTS
n
Figure 1 shows a diagram of the SMART 2 Colorimeter and its components.
T op View
SMART2
••••••••••••••••••
ENTER
OFFEXIT
Lid
*
ON
Side Views
Bottom View
Serial
Number
Battery
Compartment
RS232
Serial Port
AC
Adapter Socket
SMART2 SOIL MANUAL 1.0715
QUICK START
n
Some quick instructions to get into testing.
Press the ON buttontoturnontheSMART2.The
LaMotte logo screen will appear for about 2 seconds
and then the Start screen appears. Press the
Q/ENTER button to start testing.
The MAIN MENU will appear. Press Q/ENTER to
select TESTING MENU.
Press Q/ENTER to select All Tests.
Press t or s to move the * to the desired test.
VER 1.0
Smart 2
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
ALL TESTS
* 001 Alk - UDV
002 Aluminum
003 Ammonia-NLF
Press Q/ENTER to select test.
Insert blank, press Q/ENTER to scan blank.
Continued...
16SMART2 SOIL MANUAL 1.07
ALL TESTS
* 015 Chlorine
016 Cl F-UDV
017 Cl Liq-DPD
015 Chlorine
* Scan Blank
The screen will display Blank Done for about 1
second.
015 Chlorine
Blank Done
* Scan Blank
Insert the reacted sample. Press Q/ENTER to scan
sample. The SMAR T 2 will scan the sample and
display the concentration.
After recording test result, scroll with t or s and
make another selection with Q/ENTER.PressEXIT to escape to previous menus.
015 Chlorine
* Scan Sample
015 Chlorine
1.28 ppm
* Scan Sample
SMART2 SOIL MANUAL 1.0717
GENERAL OPERATING PROCEDURES
The operation of the SMART 2 Colorimeter is controlled by a microprocessor.
The microprocessor is programmed with menu driven software. A menu is a list
of choices. This allows a selection of various tasks for the colorimeter to
perform, such as, scan blank, scan sample, and edit test sequences. The keypad
is used to make menu selections which are viewed in the display. There are
three selections accessible from the MAIN MENU: Testing Menu, EditingMenu and PC Link.
THE KEYP AD
n
The keypad has 6 buttons which are used to perform specific tasks.
ON
tThis button will cause the display to scroll down through a list of
sThis button will cause the display to scroll up in a list of menu
ENTER
Q
EXIT
OFF
SAMPLE HOLDERS
n
The sample chamber is designed for 25 mm round tubes. Additional sample
holders for 16 mm COD tubes and for 1 cm square UDV cuvettes are available
for the SMAR T 2 Colorimeter.
This button is used to turn the colorimeter on.
menu choices. It will move through a list viewed in the display. It
will auto scroll when held down.
choices. It will move through a list viewed in the display. It will
auto scroll when held down.
This button is used to select the menu choice adjacent to the “*”in
amenuviewedinthedisplay.
This button is an exit or escape button. When pressed, the display
will exit from the current menu and go to the previous menu.
This button turns the colorimeter off.
18SMART2 SOIL MANUAL 1.07
THE DISPLAY & THE MENUS
n
The display allows menu selections to be viewed and chosen. These choices
instruct the colorimeter to perform specific tasks. The menus are viewed in the
display using two general formats which are followed from one menu to the
next. Each menu is a list of choices or selections.
There are four lines in the display. The top line in each menu is a title or
pertinent instruction. The top line does not change unless a new menu is
selected. The second and third lines are used in two ways. One way is to display
menu choices. The second way takes advantage of the graphical capabilities of
the display. Both lines are used to display important messages, such as test
results, in a large, easy to read format. The fourth line is used for menu choices.
DISPLAY
TESTING MENU
* FIRST CHOICE
SECOND CHOICE
ANOTHER
AND ANOTHER
AND SO ON
TITLE or INSTRUCTION
MENU CHOICE WINDOW
Think of the menu choices as a vertical list in the display which moves up or
down each time an arrow button is pressed. This list or menu is viewed through
a window, the menu choice window, in the display. The menu choice window
is the lower 2 or 3 lines of the display. Pushing the arrow buttons brings
another portion of the menu into menu choice window. This is referred to as
scrolling through the menu.
TESTING MENU
* FIRST CHOICESECOND CHOICEANOTHER
SECOND CHOICE* ANOTHERAND ANOTHER
ANOTHERAND ANOTHER* AND SO ON
AND ANOTHERAND SO ONLAST CHOICE
AND SO ONLAST CHOICE
LAST CHOICE
t
TESTING MENU
t
TESTING MENU
An asterisk, “*”, will start in the far left position of the top line in the menu
choice window. As the menu is scrolled through, different choices appear next
to the “*”. The “*” in the display corresponds with the Q/ENTER button.
Pushing the Q/ENTER button selects the menu choice which is adjacent to
the “*” in the menu choice window.
SMART2 SOIL MANUAL 1.0719
The second general format of the display takes advantage of the graphics
capabilities of the display. The top line of the display is still a title line. The
middle two lines of the display are used to display important messages, results
or graphics in a large, easy to read format. The menus work in the same way as
described previously but only one line of the menu is visible at the bottom of
the display.
TESTING MENU
TESTING MENU
t
TESTING MENU
t
Result or MessageResult or MessageResult or Message
* ANOTHER* AND ANOTHER*AND SO ON
AND ANOTHERAND SO ONLAST CHOICE
AND SO ONLAST CHOICE
LAST CHOICE
As described previously, the EXIT button allows an exit or escape from the
current menu and a return to the previous menu. This allows a rapid exit from
an inner menu to the main menu by repeatedly pushing the EXIT button.
Pushing OFF at any time will turn the colorimeter off.
20SMART2 SOIL MANUAL 1.07
LOOPING MENUS
n
Long menus, such as All Tests, incorporate a looping feature which allow
the user to quickly reach the last choice in the menu from the first choice. In a
looping menu the last choices in the menu are above the first choice and
scrolling upward moves through the menu in reverse order. Scrolling downward
moves through the menu from first choice to last but the menu starts over
following the last choice. So all menu choices can be reached by scrolling in
either direction. The diagrams below demonstrate a looping menu.
AND SO ONAND ANOTHERANOTHER
: : :AND SO ONAND ANOTHER
: : :: : :AND SON ON
THIRD TO LAST: : :: : :
SECOND TO LASTTHIRD TO LAST: : :
LAST CHOICESECOND TO LASTTHIRD TO LAST
TESTING MENU
* FIRST CHOICE* LAST CHOICE*SECOND TO LAST
SECOND CHOICEFIRST CHOICELAST CHOICE
ANOTHERSECOND CHOICEFIRST CHOICE
AND ANOTHERANOTHERSECOND CHOICE
AND SO ONAND ANOTHERANOTHER
: : :AND SO ONAND ANOTHER
: : :: : :AND SO ON
LAST CHOICE: : :: : :
TESTING MENU
s
TESTING MENU
s
SMART2 SOIL MANUAL 1.0721
TESTING
TESTING MENU
n
The Testing Menu is used to run all LaMotte pre-programmed tests, USER
TESTS and Absorbance test at one of four wavelengths. Testing from any of
three sequences can also be done.
Press ON to turn on the SMART 2
Colorimeter. The LaMotte logo will appear for
about 2 seconds and the the Start screen
appears. Press Q/ENTER to begin testing.
The MAIN MENU will appear. Press Q/ENTER
to select Testing Menu.
Scroll with t or s and make a selection with
Q/ENTER. All Tests lists all the available
tests. The three sequences have selected tests
and Absorbance lists %T/ABS tests.
VER 1.0
Smart 2
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
Sequence 3
Absorbance
22SMART2 SOIL MANUAL 1.07
SEQUENCES OF TESTS
n
SEQUENCE 1, SEQUENCE 2,andSEQUENCE 3 are alterable sequences. They
may be edited using the Editing Menu. Any of the LaMotte
pre-programmed tests or User Tests may be placed in these sequences in
whatever testing order that is preferred. Some examples of typical sequences
are given below.
SEQUENCE 1SEQUENCE 2SEQUENCE 3
* 015 Chlorine* 002 Aluminum* 003 Ammonia-N LF
079 Phosphate H035 Cyanide032 Cu-DDC
009 Bromine-LR041 Fluoride064 Nitrate-N L
076 pH TB053 Iron Phen067 Nitrite-N L
061 Moly-HR055 Manganese L074 pH CPR
086 Silica Hi064 Nitrate-N L078 Phosphate L
045 Hydrazine067 Nitrite-N L085 Silica Lo
032 Cu-DDC077 Phenol
051 Iron Bipyr078 Phosphate L
090 Sulfide-LR
These alterable sequences allow a series of tests to be setup that are run
frequently. The order of the individual tests in the sequence is determined by
the user. After running a test, use t to scroll to the next test and press
Q/ENTER to select the next test in the sequence. Continue this pattern until
the entire sequence has been completed.
All Tests is a fixed sequence containing the LaMotte pre-programmed tests,
User Tests, and Absorbance tests.
Modification of the alterable sequences is accomplished through the EditingMenu. This menu is explained in greater detail in EDITING MENU (p. 25).
Pressing the EXIT button while in a sequence menu will escape back to the
Testing Menu.
Pressing the OFF button at any time will turn the colorimeter off.
SMART2 SOIL MANUAL 1.0723
GENERAL TESTING PROCEDURES
n
The following are some step by step examples of how to run tests from the
Testing Menu. These test procedures are designed to be used with LaMotte
SMART Reagent Systems.
TESTING WITH THE LaMOTTE
n
PRE-PROGRAMMED TESTS
Press ON to turn on the SMART 2 Colorimeter.
The LaMotte logo will appear for about 2 seconds
and then the Start screen appears. Press Q/ENTER
to start testing.
The MAIN MENU will appear. Press Q/ENTER to
select Testing Menu.
Press Q/ENTER to select All Tests.
Press t to scroll to 002 Aluminum.
VER 1.0
Smart2
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
ALL TESTS
* 001 Alk - UDV
002 Aluminum
003 Ammonia-N LF
Press Q/ENTER to select 002 Aluminum.
Continued...
24SMART2 SOIL MANUAL 1.07
ALL TESTS
* 002 Aluminum
003 Ammonia-N LF
004 Ammonia-N LS
The SMART 2 Colorimeter is ready to scan at the
correct wavelength. Place the blank in the sample
chamber, close the lid and press Q/ENTER to scan
blank.
NOTE: Do not keep the button depressed.
002 Aluminum
* Scan Blank
The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.
Place the reacted sample in the chamber, close the
lid and press Q/ENTER to scan sample. The
colorimeter will scan the sample and the results
screen will appear.
Record test result. To repeat the test, press
Q/ENTER to scan the sample again. The last blank
scanned is used to zero the colorimeter for repeated
scans. A different blank can be used by pressing s to
scroll back to Scan Blank and then scanning
another blank. Scroll with the t or s buttons and
make another selection with the Q/ENTER button.
The%TorAbsorbanceofthelasttestcanbeviewed
by choosing %T/Abs.PressEXIT to escape to
previous menus.
NOTE: The menus loop in this screen so either the
s or t buttons will lead to the menu selection
needed.
002 Aluminum
Blank Done
* Scan Blank
002 Aluminum
* Scan Sample
002 Aluminum
0.09 ppm
* Scan Blank
Next Test
Previous Test
%/Abs
Scan Blank
SMART2 SOIL MANUAL 1.0725
EDITING MENU
The EDITING MENU allows the user to edit sequences, edit user tests, set the
clock, edit the logging function, and set the power saving function.
EDIT A SEQUENCE
n
The EDIT SEQUENCE menu allows three alterable test sequences (SEQUENCE
1, SEQUENCE 2,andSEQUENCE 3) to be edited.
Press ON to turn on the SMART 2 Colorimeter.
The LaMotte logo will appear for about 2 seconds
and then the Start screen appears. Press Q/ENTER
to start testing.
The Main Menu will appear. Press t to scroll to
Editing Menu.
Press Q/ENTER to select Editing Menu.
The Editing Menu appears. Press Q/ENTER to
select Editing Sequence.
VER 1.0
Smart2
*START
MAIN MENU
Testing Menu
* Editing Menu
PC Link
MAIN MENU
*Editing Menu
PC Link
EDITING MENU
* Edit Sequence
Edit User Test
Set Clock
The Edit Sequence menu appears. PressQ/ENTER to scroll to select Edit Sequence 1.
Continued...
26SMART2 SOIL MANUAL 1.07
EDIT SEQUENCE
*Edit Sequence 1
Edit Sequence 2
Edit Sequence 3
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.