Warning! This set contains chemicals that
may be harmful if misused. Read cautions on
individual containers carefully. Not to be used
by children except under adult supervision.
1910-MN-V3 06.28.19
1910-MN-V3.indd
PO Box 329 • Chestertown, MD 21620
800-344-3100 • f 410-778-6394
www.lamotte.com
General Precautions .................................................................................................................................. 5
Register Your Meter......................................................................................................................................6
CE Compliance ........................................................................................................................................... 11
IP 67 Certification ...................................................................................................................................... 11
CHEMICAL TESTING
Water Sampling for Chemical Analysis ...............................................................................................12
Meter Care .................................................................................................................................................. 16
Selecting an Appropriate Wavelength ............................................................................................... 16
Standard Additions ................................................................................................................................. 19
The Keypad ................................................................................................................................................ 23
The Display & the Menus ....................................................................................................................... 24
TESTING
Testing Menu ............................................................................................................................................. 26
Test Sequences ........................................................................................................................................ 27
General Testing Procedures .................................................................................................................. 27
Testing With LaMotte Pre-Programmed Tests ................................................................................ 28
Measuring in the Absorbance Mode .................................................................................................... 33
Page 4
EDITING MENU
Editing a Sequence .................................................................................................................................. 36
Adding or Deleting a Test ....................................................................................................................... 38
Edit User Tests .......................................................................................................................................... 42
Naming the Test ....................................................................................................................................... 44
Selecting the Vial and Wavelength ..................................................................................................... 47
Entering a Two Point Calibration ......................................................................................................... 48
Entering a Multiple Point Calibration....................................................................................................51
Selecting the Numerical Format of the Result .................................................................................54
Selecting Units of Concentration ......................................................................................................... 55
Setting the Clock ....................................................................................................................................... 56
Logging Data .............................................................................................................................................. 57
Setting Power Save .................................................................................................................................. 58
Setting Backlight Time ............................................................................................................................. 59
Bluetooth Menu .......................................................................................................................................... 60
Selecting a Language ............................................................................................................................... 67
Looping Menu .............................................................................................................................................69
Bluetooth Connection ............................................................................................................................... 70
Meter Disposal ........................................................................................................................................... 73
Experienced packaging personnel at LaMotte Company assure adequate protection against
normal hazards encountered in transportation of shipments. After the product leaves
the manufacturer, all responsibility for its safe delivery is assured by the transportation
company. Damage claims must be filed immediately with the transportation company to
receive compensation for damaged goods.
Should it be necessary to return the instrument for repair or servicing, pack instrument
carefully in a suitable container with adequate packing material. A return authorization
number must be obtained from LaMotte Company by calling 1-800-344-3100 or emailing
tech@lamotte.com. Attach a letter with the authorization number to the shipping carton
which describes the kind of trouble experienced. This valuable information will enable the
service department to make the required repairs more efficiently.
GENERAL PRECAUTIONS
Before attempting to set up or operate this instrument it is important to read the
instruction manual. Failure to do so could result in personal injury or damage to the
equipment.
The Smart3 Colorimeter should not be stored or used in a wet or corrosive environment.
Care should be taken to prevent water or reagent chemicals from wet colorimeter tubes
from entering the colorimeter chamber.
NEVER PUT WET TUBES IN COLORIMETER.
SAFETY PRECAUTIONS
*WARNING: Reagents marked with an * are considered to be potential health hazards. To
view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the contents
list or in the test procedures. Omit any letter that follows or precedes the four digit code
number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy,
contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel: (US, 1-800255-3924) (International, call collect, 813-248-0585)
To order individual reagents or test kit components, use the specified code number.
Keep equipment and reagent chemicals out of the reach of young children.
Ensure that the protection provided by this equipment is not impaired. Do not intall or use
this equipment in a manner that is not indicated in this manual.
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life, property, profits,
or other damages incurred through the use or misuse of its products.
WARRANTY
LaMotte Company warrants this instrument to be free of defects in parts and workmanship
for 2 years from the date of shipment. If it should become necessary to return the
instrument for service during or beyond the warranty period, contact our Technical
Service Department at 1-800-344-3100 or tech@lamotte.com for a return authorization
SMART3 Colorimeter 09.16 5
Page 6
number or visit www.lamotte.com for troubleshooting help. The sender is responsible for
shipping charges, freight, insurance and proper packaging to prevent damage in transit.
This warranty does not apply to defects resulting from action of the user such as misuse,
improper wiring, operation outside of specification, improper maintenance or repair, or
unauthorized modification. LaMotte Company specifically disclaims any implied warranties
or merchantability or fitness for a specific purpose and will not be liable for any direct,
indirect, incidental or consequential damages. LaMotte Company’s total liability is limited
to repair or replacement of the product. The warranty set forth above is inclusive and no
other warranty, whether written or oral, is expressed or implied.
REGISTER YOUR METER
To register your meter with the LaMotte Service Department, go to www.lamotte.com and
choose SUPPORT on the top navigation bar.
SPECIFICATIONS
INSTRUMENT TYPE: Colorimeter
Readout160 x 100 backlit LCD, 20 x 6 line graphical display
Wavelengths428 nm, 525 nm, 568 nm, 635 nm
Wavelength Accuracy±2% FS
Readable ResolutionDetermined by reagent system
Wavelength Bandwidth10 nm typical
Photometric Range–2 to +2 AU
Photometric Precision± 0.001 AU at 1.0 AU
Photometric Accuracy±0.005 AU at 1.0 AU
Sample ChamberAccepts 25 mm diameter flat-bottomed test tubes, 10 mm
square cuvettes†, 16 mm COD test tubes
Light Sources4 LEDs
Detectors4 silicon photodiodes
ModesPre-programmed tests, absorbance, %T
Pre-Programmed TestsYES, with automatic wavelength selection
Power RequirementsUSB wall adapter, USB computer connection or lithium ion
0-90 % RH, non-condensing
rechargeable battery
6 SMART3 Colorimeter 09.16
Page 7
BatteryCharge Life: Approximately 380 tests with backlight on to
1000 tests with backlight off. (Signal averaging disabled).
Battery Life: Approximately 500 charges.
Electrical RatingRated voltage (5V), Rated power of input current (1.0A) at
mini-USB input port
Data Logger500 test results stored for download to a PC
WaterproofIP67 with USB port plug in place
Certifications
BT RadioFCC ID:
Cert#
BQB
LISTED
CE MarkSafety
EMC
T9J-RN42
6514A-RN42
B014867-SPP and DUN
profiles
EN61610-1:2010
+Corrigendum 1:2011
EN 301 489-17 and
EN 301 489
USEMCFCC CFR 47, Part 15,
subpart B
CanadaEMCCAN ICES-3(A)/NMB-3(A)
AU/NZEMCAS/NZS CISPR 22
Dimensions (LxWxH)3.5 x 7.5 x 2.5 inches, 8.84 x 19.05 x 6.35 cm
Weight13 oz, 362 g (meter only)
†525 nm and 568 nm only
STATISTICAL & TECHNICAL DEFINITIONS RELATED TO
PRODUCT SPECIFICATIONS
Method Detection Limit (MDL): “The method detection limit (MDL) is defined as the
minimum concentration of a substance that can be measured and reported with 99%
confidence that the analyte concentration is greater than zero and is determined from
analysis of a sample in a given matrix containing the analyte.”1 Note that, “As Dr. William
Horwitz once stated, ‘In almost all cases when dealing with a limit of detection or limit of
determination, the primary purpose of determining that limit is to stay away from it.’”
Accuracy: Accuracy is the nearness of a measurement to the accepted or true value.3
The accuracy can be expressed as a range, about the true value, in which a measurement
occurs (i.e. ±0.5 ppm). It can also be expressed as the % recovery of a known amount of
analyte in a determination of the analyte (i.e. 103.5 %).
Resolution: Resolution is the smallest discernible difference between any two
measurements that can be made.4 For meters this is usually how many decimal places are
displayed. (i.e. 0.01). Note that the resolution many change with concentration or range.
In some cases the resolution may be less than the smallest interval, if it is possible to
make a reading that falls between calibration marks. A word of caution, that resolution has
very little relationship to accuracy or precision. The resolution will always be less than the
accuracy or precision but it is not a statistical measure of how well a method of analysis
SMART3 Colorimeter 09.16 7
2
Page 8
works. The resolution can be very, very good and the accuracy and precision can be very
bad! This is not a useful measure of the performance of a test method.
Repeatability: Repeatability is the within-run precision.5 A run is a single data set, from
set up to clean up. Generally, one run occurs on one day. However, for meter calibrations,
a single calibration is considered a single run or data set, even though it may take 2 or 3
days.
Reproducibility: Reproducibility is the between-run precision.
6
Detection Limit (DL): The detection limit (DL) for the 2020we/wi
is defined as the minimum value or concentration that can be determined by the meter,
which is greater than zero, independent of matrix, glassware, and other sample handling
sources of error. It is the detection limit for the optical system of the meter.
1
CFR 40, part 136, appendix B
2
Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and L Vanatta, American
Laboratory, Sept 2003, P. 31.
3
Skoog, D.A., West, D. M., Fundamental of Analytical Chemistry, 2nd ed., Holt Rinehart and
Winston, Inc, 1969, p. 26.
4
Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and L Vanatta, American
Laboratory, Sept 2003, P. 34.
5
Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s Textbook of Quantitative
Chemical Analysis, 5th ed., Longman Scientific & Technical, 1989, p. 130.
6
Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s Textbook of Quantitative
Chemical Analysis, 5th ed., Longman Scientific & Technical, 1989, p. 130
CONTENTS AND ACCESSORIES
Contents
Smart3 Colorimeter
Test Tubes, with CapsCode 0290
COD/UDV AdapterCode 1724
USB Wall AdapterCode 1721
USB CableCode 1720-01
Smart3 Colorimeter Quick Start Guide
Smart3 Colorimeter Manual
Accessories
Test Tubes, with Caps Code 0290-6
Replacement ChamberCode 3-0038
USB CableCode 1720-01
8 SMART3 Colorimeter 09.16
3.17
Page 9
USB Wall AdapterCode 1721
COD/UDV Adapter Code 1724
Bluetooth Mobile PrinterCode 3-0066
Car ChargerCode 5-0132
Small Field Carrying Case (37.5 27.5 x 13.75 cm)Code 1910-GCS150
Large Field Carrying Case (45 x 32.5 x 20 cm)Code 1910-GCS440
WARNING: Only use the USB cable (1720-01) that is supplied with the kit. Make no substitutions.
EPA COMPLIANCE
The Smart3 Colorimeter is an EPA-Accepted instrument. EPA-Accepted means that the
instrument meets the requirements for instrumentation as found in test procedures that
are approved for the National Primary Drinking Water Regulations (NPDWR) or National
Pollutant Discharge Elimination System (NPDES) compliance monitoring programs. EPAAccepted instruments may be used with approved test procedures without additional
approval.
CE COMPLIANCE
The Smart3 Colorimeter has earned the European CE Mark of Compliance for
electromagnetic compatibility and safety. The Declaration of Conformity for the Smart3
colorimeter is available at www.lamotte.com.
This device complies with Part 15 of the FCC rules. Operation is subject to the following
two conditions (1) This device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.
Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is
commercial environment. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interferences in which case the
user will be required to correct the interference at his own expense.
operated in a
IP67 CERTIFICATION
The Smart3 meets IP67 standards for protection against dust and immersion only when
the USB port plug is in place. Documentation is available at www.lamotte.com.
SMART3 Colorimeter 09.16 9
Page 10
CHEMICAL TESTING
WATER SAMPLING FOR CHEMICAL ANALYSIS
Taking Representative Samples
The underlying factor to be considered for any type of water sampling is whether or not the
sample is truly representative of the source. To properly collect a representative sample:
◊ Sample as frequently as possible.
◊ Collect a large sample or at least enough to conduct whatever tests are necessary.
◊ Make a composite sample for the same sampling area.
◊ Handle the sample in such a way as to prevent deterioration or contamination before
the analysis is performed.
◊ Perform analysis for dissolved gases such as dissolved oxygen, carbon dioxide, and
hydrogen sulfide immediately at the site of sampling. Samples for testing these
factors, as well as samples for pH, cannot be stored for later examination.
◊ Make a list of conditions or observations which may affect the sample. Other
considerations for taking representative samples are dependent upon the source of
the sample. Taking samples from surface waters involves different considerations
than taking samples from impounded and sub-surface waters.
Sampling of Open Water Systems
Surface waters, such as those found in streams and rivers, are usually well mixed. The
sample should be taken downstream from any tributary, industrial or sewage pollution
source. For comparison purposes samples may be taken upstream and at the source of the
pollution before mixing.
In ponds, lakes, and reservoirs with restricted flow, it is necessary to collect a number of
samples in a cross section of the body of water, and where possible composite samples
should be made to ensure representative samples.
To collect samples from surface waters, select a suitable plastic container with a tight
fitting screw cap. Rinse the container several times with the sample to be tested, then
immerse the container below the surface until it is filled to overflowing and replace the
cap. If the sample is not to be tested immediately, pour a small part of the sample out
and reseal. This will allow for any expansion. Any condition which might affect the sample
should be listed.
Sub-surface sampling is required to obtain a vertical profile of streams, lakes, ponds, and
reservoirs at specific depths. This type of sampling requires more sophisticated sampling
equipment.
For dissolved oxygen studies, or for tests requiring small sample sizes, a Water Sampler
(LaMotte Code 1060) will serve as a subsurface or in-depth sampler. This weighted device
is lowered to the sampling depth and allowed to rest at this depth for a few minutes. The
water percolates into the sample chamber displacing the air which bubbles to the surface.
When the bubbles cease to rise, the device has flushed itself approximately five times
and it may be raised to the surface for examination. The inner chamber of the sampling
device is lifted out and portions of the water sample are carefully dispensed for subsequent
chemical analysis.
10 SMART3 Colorimeter 09.16
Page 11
A Snap-Plunger Water Sampler (LaMotte Code 1077) is another “in-depth” sampling
device which is designed to collect large samples which can be used for a multitude of
tests. Basically, this collection apparatus is a hollow cylinder with a spring loaded plunger
attached to each end. The device is cocked above the surface of the water and lowered to
the desired depth. A weighted messenger is sent down the calibrated line to trip the closing
mechanism and the plungers seal the sample from mixing with intermediate layers as it is
brought to the surface. A special drain outlet is provided to draw off samples for chemical
analysis.
Sampling of Closed System
To obtain representative samples from confined water systems, such as pipe lines, tanks,
vats, filters, water softeners, evaporators and condensers, different considerations are
required because of chemical changes which occur between the inlet and outlet water.
One must have a basic understanding of the type of chemical changes which occur for the
type of equipment used. Also, consideration should be given to the rate of passage and
retaining time for the process water.
Temperature changes play an important part in deciding exactly what test should be
performed. Process water should be allowed to come to room temperature, 20–25°C,
before conducting any tests.
When drawing off samples from an outlet pipe such as a tap, allow sample to run for
several minutes, rinsing the container several times before taking the final sample. Avoid
splashing and introduction of any contaminating material.
FILTRATION
When testing natural waters that contain significant turbidity due to suspended solids and
algae, filtration is an option. Reagent systems, whether EPA, Standard Methods, LaMotte
or any others, will generally only determine dissolved constituents. Both EPA and Standard
Methods suggest filtration through a 0.45 micron filter membrane, to remove turbidity, for
the determination of dissolved constituents.** To test for total constituents, organically
bound and suspended or colloidal materials, a rigorous high temperature acid digestion is
necessary.
**LaMotte offers a filtering apparatus: syringe assembly (Code 1050) and membrane filters, 0.45 micron,
(Code 1103).
AN INTRODUCTION TO COLORIMETRIC ANALYSIS
Most test substances in water are colorless and undetectable to the human eye. To test
for their presence we must find a way to “see” them. The Smart3 Colorimeter can be
used to measure any test substance that is itself colored or can be reacted to produce
a color. In fact a simple definition of colorimetry is “the measurement of color” and a
colorimetric method is “any technique used to evaluate an unknown color in reference to
known colors”. In a colorimetric chemical test the intensity of the color from the reaction
must be proportional to the concentration of the substance being tested. Some reactions
have limitations or variances inherent to them that may give misleading results. Many
such interferences are discussed with each particular test instruction. In the most basic
colorimetric method the reacted test sample is visually compared to a known color
standard. However, accurate and reproducible results are limited by the eyesight of the
analyst, inconsistencies in the light sources, and the fading of color standards.
SMART3 Colorimeter 09.16 11
Page 12
To avoid these sources of error, a colorimeter can be used to photoelectrically measure the
amount of colored light absorbed by a colored sample in reference to a colorless sample
(blank).
White light is made up of many different colors or wavelengths of light. A colored sample
typically absorbs only one color or one band of wavelengths from the white light. Only a
small difference would be measured between white light before it passes through a colored
sample versus after it passes through a colored sample. The reason for this is that the one
color absorbed by the sample is only a small portion of the total amount of light passing
through the sample. However, if we could select only that one color or band of wavelengths
of light to which the test sample is most sensitive, we would see a large difference between
the light before it passes through the sample and after it passes through the sample.
The Smart3 Colorimeter passes one of four colored light beams through one of four
optical filters which transmits only one particular color or band of wavelengths of light
to the photodectector where it is measured. The difference in the amount of colored
light transmitted by a colored sample is a measurement of the amount of colored light
absorbed by the sample. In most colorimetric tests the amount of colored light absorbed
is directly proportional to the concentration of the test factor producing the color and
the path length through the sample. However, for some tests the amount of colored light
absorbed is inversely proportional to the concentration.
The choice of the correct wavelength for testing is important. It is interesting to note that
the wavelength that gives the most sensitivity (lower detection limit) for a test factor is the
complementary color of the test sample. For example the Nitrate-Nitrogen test produces a
pink color proportional to the nitrate-nitrogen concentration in the sample (the greater the
nitrate-nitrogen concentration, the darker the pink color). A wavelength in the green region
should be selected to analyze this sample since a pinkish-red solution absorbs mostly
green light.
REAGENT BLANK
Some tests will provide greater accuracy if a reagent blank is determined to compensate
for any color or turbidity resulting from the reagents themselves. A reagent blank is
performed by running the test procedure on demineralized or deionized water. Use sample
water to Scan Blank. Insert the reacted reagent blank in the colorimeter chamber and
select Scan Sample. Note result of reagent blank. Perform the tests on the sample water
as described. Subtract results of reagent blank from all subsequent test results. NOTE:
Some tests require a reagent blank to be used to Scan Blank.
COLORIMETER TUBES AND CHAMBER
The colorimeter uses one type of tube (Code 0290) for all test factors. The handling of the
tubes is of utmost importance. Tubes must be clean and free from lint, fingerprints, dried
spills and significant scratches, especially the central zone between the bottom and the
sample line.
Scratches, fingerprints and water droplets on the tube can cause stray light interference
leading to inaccurate results. Tubes that have been scratched in the light zone through
excessive use should be discarded and replaced with new ones.
Tubes should always be washed on the inside and outside with mild detergent prior to use
to remove dirt or fingerprints. The tubes should be allowed to air-dry in an inverted position
to prevent dust from entering the tubes. Dry tubes should be stored with the caps on to
prevent contamination.
12 SMART3 Colorimeter 09.16
Page 13
After a tube has been filled and capped, it should be held by the cap and the outside
surface should be wiped with a clean, lint-free absorbent cloth until it is dry and smudgefree. Handling the tube only by the cap will avoid problems from fingerprints. Always set the
clean tube aside on a clean surface that will not contaminate the tube. It is imperative that
the tubes and light chamber be clean and dry. The outside of the tubes should be dried with
a clean, lint-free cloth or disposable wipe before they are placed in the meter chamber.
Tubes should be emptied and cleaned as soon as possible after reading a sample to
prevent deposition of particulates on the inside of the tubes.
Variability in the geometry of the glassware and technique is the predominate cause of
variability in results. Slight variations in wall thickness and the diameter of the tubes may
lead to slight variations in the test results. To eliminate this error the tubes should be
placed in the chamber with the same orientation each time.
Chambers which have been scratched through excessive use should be discarded and
replaced with a new one.
METER CARE
The optical system of the Smart3 must be kept clean and dry for optimal performance. Dry
the colorimeter tubes before placing them in the chamber to avoid introducing moisture.
For best results store the instrument in a area that is dry and free from aggressive
chemical vapors.
SELECTING AN APPROPRIATE WAVELENGTH
The most appropriate wavelength to use when creating a calibration curve is usually the
one which gives the greatest change from the lowest reacted standard concentration
to the highest reacted standard concentration. However, the absorbance of the highest
reacted standard concentration should never be greater than 2.0 absorbance units. Scan
the lowest and highest reacted standards at different wavelengths using the absorbance
mode to find the wavelength which gives the greatest change in absorbance without
exceeding 2.0 absorbance units. Use this wavelength to create a calibration curve.
Below is a list of suggested wavelengths for the color of the reacted samples. Use these as
a starting point.
Sample ColorWavelength Range
Yellow428
Pink525
Red568
Green and Blue635
CALIBRATION
As with all pre-calibrated meters, it is highly recommended, even if not required by
regulations, that the user periodically verify the performance of the meter by running
standards with a predetermined concentration. Results outside of specification are
an indication that the meter needs to be adjusted. This can be done following the user
calibration described on page 29. If the user calibration fails to properly adjust the meter
then the meter should be returned to LaMotte Company for recalibration. (See page 74).
SMART3 Colorimeter 09.16 13
Page 14
CALIBRATION CURVES
The Smart3 Colorimeter contains tests for the LaMotte reagent systems. The first step
in using a non-LaMotte reagent system with your Smart3 Colorimeter is to create a
calibration curve for the reagent system. To create a calibration curve, prepare standard
solutions of the test factor and use the reagent system to test the standard solutions with
the Smart3 Colorimeter. Select a wavelength for the test as described above.
Plot the results (in ABS or %Transmittance) versus concentration to create a calibration
curve. The calibration curve may then be used to identify the concentration of an unknown
sample by testing the unknown, reading Absorbance or %T, and finding the corresponding
concentration from the curve. The linear range of the reagent system can be determined
and this information can be used to input a User Test into the Smart3 Colorimeter (see Edit
User Tests, page 42)
PROCEDURE
Prepare 5 or 6 standard solutions of the factor being tested. The concentration of these
standards should be evenly distributed throughout the range of the reagent system,
and should include a 0 ppm standard (distilled water). For instance, the solutions could
measure 0, 10%, 30%, 50%, 70%, and 90% of the system’s maximum range.
1. Turn on the Smart3 Colorimeter. Select the appropriate wavelength from the
absorbance mode. Be sure to select the appropriate wavelength for the color produced
by the reagent system.
2. Use the unreacted 0 ppm standard to standardize the colorimeter by using it to scan
blank.
3. Following the individual reagent system instructions, react each standard solution
beginning with 0 ppm. Continue with standards in increasing concentration. Record
the reading and the standard solution concentration on a chart. Readings can be
recorded as percent transmittance (%T) or absorbance (A).
4. Plot results on graph paper or computer using any available plotting program. If
results are as %T versus concentration, semilog graph paper must be used. Plot
the standard solution concentrations on the horizontal, linear axis, and the %T on
the vertical, logarithmic axis. If results are as absorbance versus standard solution
concentration, simple linear graph paper can be used. Plot the standard solution
concentration on the horizontal axis, and the absorbance on the vertical axis.
5. After plotting the results, draw a line, or curve, of best fit through the plotted points.
The best fit may not connect the points. There should be approximately an equal
number of points above the curve as below the curve. Some reagent systems will
produce a straight line, while others produce a curve. Many computer spreadsheet
programs can produce the curve of best fit by regression analysis of the standard
solution data.
NOTE: Only reagent systems which produce a straight line can be used for a User Test.
14 SMART3 Colorimeter 09.16
Page 15
A sample of each type of graph appears below:
PREPARING DILUTE STANDARD SOLUTIONS
Standard solutions should be prepared to create a calibration curve. Standard solutions
can be prepared by diluting a known concentrated standard by specified amounts. A chart
or computer spreadsheet can be created to determine the proper dilutions. Use volumetric
flasks and volumetric pipets for all dilutions.
6. In Column A – Record the maximum concentration of test as determined by the range
and path length.
7. In Column B – Record the percent of the maximum concentration the standard
solution will be.
8. In Column C – Calculate the final concentration of the diluted standard solutions
by multiplying the maximum concentration (In Column A) by the % of maximum
concentration divided by 100. (C = A x B/100).
9. In Column D – Record the final volume of the diluted sample (i.e. volume of volumetric
flask).
10. In Column E – Record the concentration of the original standard.
11. In Column F – Calculate the milliliters of original standard required
(F = (C x D/E)).
A sample chart appears below:
ABC =
Maximum
concentration
of test
% of Maximum
concentration
A x B/
100
Final
concentration
of Diluted
Standard
DEF =
Volume of
Standard
Concentration
of Original
Standard
C x D/E
mL of
Original
Standard
Required
10.0 ppm909.0 ppm100 mL1000 ppm0.90 mL
10.0 ppm707.0 ppm100 mL1000 ppm0.70 mL
10.0 ppm505.0 ppm100 mL1000 ppm0.50 mL
10.0 ppm303.0 ppm100 mL1000 ppm0.30 mL
10.0 ppm101.0 ppm100 mL1000 ppm0.10 mL
10.0 ppm00 ppm100 mL1000 ppm0 mL
SMART3 Colorimeter 09.16 15
Page 16
STANDARD ADDITIONS
A common method to check the accuracy and precision of a test is by standard
additions. In this method a sample is tested to determine the concentration of the test
substance. A second sample is then “spiked” by the addition of a known quantity of the
test substance. The second sample is then tested. The determined concentration of the
spiked sample should equal the concentration of the first plus the amount added with the
spike. The procedure can be repeated with larger and larger “spikes.” If the determined
concentrations do not equal the concentration of the sample plus that added with the
“spike”, then an interference may exist.
For example, a 10.0 mL water sample was determined to contain 0.3 ppm iron. To a second
10.0 mL sample, 0.1 mL of 50 ppm iron standard was added. The concentration of iron
due to the “spike” was (0.10 mL x 50 ppm)/10.0 mL = 0.50 ppm. The concentration of iron
determined in the spiked sample should be 0.3 + 0.5 = 0.8 ppm iron. (Note: any error due to
the increased volume from the “spike” is negligible).
LaMotte offers a line of calibration standards which can be used to generate calibration
curves and perform standard additions.
If a test result using the Smart3 Colorimeter gives an over range message then the the
sample must be diluted. The test should be repeated on the diluted sample to obtain
a reading which is in the concentration range for the test. (Note: This is not true for
colorimetric determination of pH.)
Example:
Measure 5 mL of the water sample into a graduated cylinder. Add demineralized water
until the cylinder is filled to the 10 mL line. The sample has been diluted by one-half, and
the dilution factor is therefore 2. Perform the test procedure, then multiply the resulting
concentration by 2 to obtain the test result.
The following table gives quick reference guidelines on dilutions of various proportions. All
dilutions are based on a 10 mL volume, so several dilutions will require small volumes of
the water sample. Graduated pipets should be used for all dilutions.
Size of SampleDeionized Water to Bring Volume to 10 mLMultiplication Factor
10 mL0 mL1
5 mL5 mL2
2.5 mL7.5 mL4
1 mL9 mL10
0.5 mL9.5 mL20
If volumetric glassware is not available, dilutions can be made with the colorimeter tube.
Fill the tube to the 10 mL line with the sample then transfer it to another container. Add 10
mL volumes of demineralized water to the container and mix. Transfer back 10 mL of the
diluted sample to the tube and follow the test procedure. Continue diluting and testing until
a reading, which is in the concentration range for the test, is obtained. Be sure to multiply
the concentration found by the dilution factor (the number of total 10 mL volumes used).
16 SMART3 Colorimeter 09.16
Page 17
Example:
10 mL of sample is diluted with three 10 mL volumes of demineralized water; the dilution
factor is four.
INTERFERENCES
LaMotte reagent systems are designed to minimize most common interferences. Each
individual test instruction discusses interferences unique to that test. Be aware of possible
interferences in the water being tested.
The reagent systems also contain buffers to adjust the water sample to the ideal pH for the
reaction. It is possible that the buffer capacity of the water sample may exceed the buffer
capacity of the reagent system and the ideal pH will not be obtained. If this is suspected,
measure the pH of a reacted distilled water reagent blank using a pH meter. This is the
ideal pH for the test. Measure the pH of a reacted water sample using the pH meter. If the
pH is significantly different from the ideal value, the pH of the sample should be adjusted
before testing.
Interferences due to high concentration of the substance being tested, can be overcome by
sample dilution (see page 18)
STRAY LIGHT INTERFERENCE
When scanning samples in 16 mm tubes, such as COD, the sample chamber lid can not be
closed. The COD adapter minimizes stray light. To further reduce stray light interference, do
not scan sample in direct sunlight.
OPERATION OF THE SMART3 COLORIMETER
OVERVIEW
The Smart3 is a portable, microprocessor controlled, direct reading colorimeter. It has a
graphical liquid crystal display and 6 button keypad. These allow the user to select options
from the menu driven software, to directly read test results or to review stored results of
previous tests in the data logger. The menus can be displayed in seven different languages.
The test library consists of over 80 LaMotte tests and 25 “User Tests”. The LaMotte tests
are precalibrated for LaMotte reagent systems.The colorimeter displays the result of these
tests directly in units of concentration. The 25 “User Tests” may be used to enter additional
calibrations. All of these tests may be arranged in any of 3 sequences. These sequences
can be modified a limitless number of times to meet changing testing needs.
The optics feature 4 different colored LEDs. Each LED has a corresponding silicon
photoiode with an integrated interference filter. The interference filters select a narrow
band of light from the corresponding LED for the colorimetric measurements. The
microporcessor automatically selects the correct LED/photodiode combination for the
test.
A USB wall adapter, USB computer connection or lithium battery powers the Smart3.
Bluetooth wireless technology allows communication between the colorimeter and the
Bluetooth Mobile Printer (Code 3-0066 only).
SMART3 Colorimeter 09.16 17
Page 18
COMPONENTS
Figure 1 shows a diagram of the Smart3 Colorimeter and its components.
Top View
SMART3
ENTER
Bottom View
Serial
Number
COLORIMETER
EXIT
Lid
Side Views
USB Port
with plug
Figure 1
18 SMART3 Colorimeter 09.16
Page 19
GENERAL OPERATING PROCEDURES
The operation of the Smart3 Colorimeter is controlled by a microprocessor. The
microprocessor is programmed with menu driven software. A menu is a list of choices.
This allows a selection of various tasks for the colorimeter to perform, such as, scan blank,
scan sample, and edit test sequences. The keypad is used to make menu selections which
are viewed in the display. There are two selections accessible from the Main Menu: Testing
Menu and Editing Menu.
THE KEYPAD
The keypad has 6 buttons which are used to perform specific tasks.
This button will scroll up through a list of menu selections or go to the
next character on a character selection line.
ENTER
EXIT
The button is used to select choices in a menu viewed in the display
and to add the selected character to the name and advance one
character.
This button controls the backlight on the display.
This button will scroll down through a list of menu selections or go to
the previous character selection line.
This button exits to the previous menu on the name edit screen, the
name will not be saved.
This button turns the meter on or off.
ENTER
EXI T
SAMPLE HOLDERS
The sample chamber is designed for 25 mm round tubes. An adapter to hold 16 mm COD
tubes and 10 mm square UDV cuvettes is included. 10 cm cuvettes can be scanned only at
525 nm and 568 nm. The light path is blocked at 428 nm and 635 nm with the use of the
adapter. COD tubes can be scanned at all wavelengths.
SMART3 Colorimeter 09.16 19
Page 20
Position the COD/UDV Adapter (Code 1724) so that the
notches in the adapter fit around the posts on the chamber.
Turn the adapter counterclockwise until the arrows are at
the top and bottom of the chamber and the adapter is locked
into place. Turn the adapter clockwise to unlock the adapter
and remove it from the chamber.
THE DISPLAY & THE MENUS
The display allows menu selections to be viewed and selected. These selections instruct the
Smart3 to perform specific tasks. The menus are viewed in the display using two general
formats that are followed from one menu to the next. Each menu is a list of choices or
selections.
The display has a header line at the top and a footer line at the bottom. The header displays
the title of the current menu. The footer line displays the time and the date, the data logger
status, the bluetooth/printer status and the battery status. The menu selection window is
in the middle of the display between the header and the footer.
The menu selection window displays information in two general formats. In the first format
only menu selections are displayed. Up to 4 lines of menu selections may be displayed.
If more selections are available they can be viewed by pressing the arrow buttons
to scroll the other menu selections into the menu selection window. Think of the
menu selections as a vertical list in the display that moves up or down each time an arrow
button is pressed. Some menus in the Smart3 are looping menus. The top
and bottom menu choices are connected in a loop. Scrolling down past the bottom of the
menu will lead to the top of the menu. Scrolling up past the top of the menu will lead to the
bottom of the menu.
HeaderMenu Title
Main Window Selection First Choice
Second Choice
Third Choice
Another
Footer
12:00:00 001/500 P
And Another
And So On
A light bar will indicate the menu choice. As the menu is scrolled through, the light bar will
highlight different menu choices. Pressing the
ENTER
button will select the menu choice
that is indicated by the light bar.
In the second format the menu choice window takes advantage of the graphical
capabilities of the display. Large format graphic information, such as test results or error
messages or the LaMotte logo is displayed. The top two lines of the display are used to
display information in a large, easy to read format. The menus work in the same way as
previously described but two lines of the menu are visible at the bottom of the display.
20 SMART3 Colorimeter 09.16
Page 21
HeaderMenu Title
Message or Result Window
Result or
Message
Another
Main Window Selection And Another
Footer
12:00:00 001/500
And So On
Last Choice
As described previously, the
and a return to the previous menu. This allows a rapid exit from an inner menu to the main
menu by repeatedly pushing the
Smart3 off.
The display may show the following messages:
Battery Status
More choices are available and can be viewed by scrolling up and/or
down through the display.
HeaderIdentifies the current menu and information on units and reagent
systems if applicable.
FooterIn the data logging mode the number of the data point is displayed
and the total number of data points in the memory will be shown. The
footer also shows current time and battery status
EXIT
button allows an exit or escape from the current menu
EXIT
button. Pushing at any time will turn the
SMART3 Colorimeter 09.16 21
Page 22
TESTING
TESTING MENU
The Testing Menu is used to run all LaMotte pre-programmed tests, User Tests and
Absorbance tests at one of four wavelengths. Testing from any of three sequences can also
be done.
1. Press and briefly hold to turn the
meter on. The LaMotte logo screen will
appear for about 3 seconds and the Main
Menu will appear.
2. Press
ENTER
to select Testing Menu.
3. Press or to scroll to desired
option. All Tests contains all of the
available pre-programmed tests. The
three sequences have user selected tests.
Absorbance has %T/ABS tests.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
Testing Menu
All Tests Menu
Sequence 1
Sequence 2
Sequence 3
12:00:00 001/500
Testing Menu
All Tests Menu
Sequence 1
Sequence 2
Sequence 3
12:00:00 001/500
4. Press
ENTER
to select All Tests.
All Tests
002 Alkalinity UDV
005 Aluminum
006 Ammonia-N LRF
007 Ammonia-N LRS
12:00:00 001/500
22 SMART3 Colorimeter 09.16
Page 23
TEST SEQUENCES
Sequence 1, Sequence 2, And Sequence 3 are alterable sequences. They may be edited
using the Editing Menu. Any of the LaMotte pre-programmed tests or User Tests may be
placed in these sequences in whatever testing order that is preferred. There are three intial
sequences in the Smart3 that can be customized by adding or deleting tests:
Sequence 1Sequence 2Sequence 3
025 Chlorine T DPD005 Aluminum006 Ammonia-N LF
080 Phosphate HR025 Chlorine T DPD036 Cu Thiocarbamate
These alterable sequences allow a series of tests to be setup that are run frequently. The
order of the individual tests in the sequence can be determined by the user. After running
a test, press
ENTER
to select the next test in the sequence. Continue this pattern until the
entire sequence has been completed.
All Tests is a fixed sequence containing the LaMotte pre-programmed tests, User Tests,
and Absorbance tests. Modification of the alterable sequences is accomplished through
the Editing Menu. This menu is explained in greater detail in Editing Menu (page 36).
Pressing
EXIT
while in a sequence menu will escape back to the Testing Menu.
Pressing the at any time will turn the colorimeter off.
NOTE: A sequence must contain at least one test.
GENERAL TESTING PROCEDURES
The following is a step-by-step example of how to run tests from the Testing Menu. These
test procedures are designed to be used with LaMotte SMART Reagent Systems.
LaMotte Company continuously updates the list of pre-programmed tests as the
calibrations become available. Call LaMotte Technical Services at 1-800-344-3100 (410778-3100 outside the USA) or email at tech@lamotte.com for a current list of available
calibrations.
SMART3 Colorimeter 09.16 23
Page 24
TESTING WITH LaMOTTE PRE-PROGRAMMED TESTS
1. Press and briefly hold to turn the
meter on. The LaMotte logo screen will
appear for about 3 seconds and the Main Menu will appear.
2. Press
3. Press
ENTER
to select Testing Menu.
ENTER
to select All Tests Menu.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
Testing Menu
All Test Menu
Sequence 1
Sequence 2
Sequence 3
12:00:00 001/500
All Tests
002 Alkalinity UDV
005 Aluminum
006 Ammonia-N LRF
007 Ammonia-N LRS
12:00:00 001/500
4. Press or to scroll to the
desired test.
002 Alkalinity UDV
All Tests
005 Aluminum
006 Ammonia-N LRF
007 Ammonia-N LRS
12:00:00 001/500
5. Press
ENTER
to select the test.
005 Aluminum
Scan Bank
Scan Sample
12:00:00 001/500
24 SMART3 Colorimeter 09.16
Page 25
6. Insert the blank into the chamber. Close
the lid.Press
screen wil display Scan Blank Blank Done
for about 1 second and then return to the
test menu.
ENTER
to scan the blank. The
005 Aluminum
Scan Blank
Scan Sample
12:00:00 001/500
7. Insert the reacted sample into the
chamber. Close the lid. Press
scan the sample. The screen will display
Scan Sample Sample Done for about
1 second. The result will appear on the
screen.
To repeat the test, press
sample again. The last blank scanned is used
by the colorimeter for repeated scans. A
different blank can be used by pressing
or to scroll to Scan Blank and then
scanning another blank. Scroll with or
and make another selection with
. The %T or Absorbance of the last test can be
viewed by scrolling down and choosing %T/
Abs. Press
NOTE: The menus loop in this screen so either
needed.
To use the Print Test function, see the Bluetooth Menu section (page 60).
EXIT
to escape to previous menus.
or will lead to the menu selection
ENTER
ENTER
to
to scan the
ENTER
1.00 ppm
Scan Blank
Scan Sample
12:00:00 001/500
1.00 ppm
Scan Bank
Scan Sample
12:00:00 001/500
005 Aluminum
005 Aluminum
CALIBRATING LaMOTTE PRE-PROGRAMMED TESTS
The LaMotte Pre-Programmed Tests have been pre-calibrated. Recalibration of the preprogrammed tests by the user is not possible. However, a procedure to standardize the
calibration can be performed to obtain the most accurate readings or to meet regulatory
requirements.
The LaMotte Pre-Programmed tests are standardized with one standard solution. To
standardize over the full range of the test, the concentration of the standard should be
chosen from the high end of the range. Alternatively, if samples do not cover the full range
of the test, a standard should be chosen that is close to the concentration of the samples.
The standardization procedure should be followed as often as required by regulations and
laws for compliance monitoring.
In the example below, the Aluminum calibration will be standardized.
Prepare a standard solution to be tested. In this example, 0.30 ppm aluminum.
SMART3 Colorimeter 09.16 25
Page 26
1. Press and briefly hold to turn the
meter on. The LaMotte logo screen will
appear for about 3 seconds and the Main Menu will appear.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
2. Press
3. Press
ENTER
to select Testing Menu.
ENTER
to select All Tests Menu.
4. Press or to scroll to the
desired test factor.
Testing Menu
All Test Menu
Sequence 1
Sequence 2
Sequence 3
12:00:00 001/500
All Tests
002 Alkalinity UDV
005 Aluminum
006 Ammonia-N LRF
007 Ammonia-N LRS
12:00:00 001/500
All Tests
002 Alkalinity UDV
005 Aluminum
006 Ammonia-N LRF
007 Ammonia-N LRS
12:00:00 001/500
5. Press
ENTER
to select the test.
005 Aluminum
Scan Blank
Scan Sample
12:00:00 001/500
26 SMART3 Colorimeter 09.16
Page 27
6. Follow the test procedure in the manual
to test the prepared standard. Insert the
blank into the chamber. Close the lid. Press
ENTER
to scan the blank. The screen will
display Scan Blank Blank Done for about
1 second and then return to the Test Menu.
005 Aluminum
Testing
Scan Blank
Scan Sample
12:00:00 001/500
7. Insert the reacted standard solution into
the chamber. Close the lid. Press
scan the sample. The screen will display
Scan Sample Sample Done for about
1 second. The result will appear on the
screen.
8. The displayed result can now be
standardized. Press or to
scroll to Calibrate.
9. Press
ENTER
to select Calibrate. A
reverse font (light background with dark
characters) will appear to indicate that the
reading can be adjusted.
ENTER
005 Aluminum
to
0.28 ppm
Scan Blank
Scan Sample
12:00:00 001/500
005 Aluminum
0.28 ppm
%T/Abs
Calibrate
12:00:00 001/500
005 Aluminum
0.28 ppm
∧, ∨=Edit, ENTER=Save
∧ +ENTER=Default
12:00:00 001/500
10. Press or to adjust the
value shown to the concentration of the
prepared standard, 0.30 in this example.
NOTE: A maximum adjustment of 25% is
possible.
SMART3 Colorimeter 09.16 27
0.30 ppm
∧, ∨=Edit, ENTER=Save
∧ +ENTER=Default
12:00:00 001/500
005 Aluminum
Page 28
11. Press
ENTER
to save the value.
To leave the Calibration procedure without
saving the adjustment, press
EXIT
.
Testing
Press and
ENTER
at any time to
return to the default value.
The calibration has now been standardized
and can be used for testing. Scroll to Scan Blank and begin testing.
MEASURING IN THE ABSORBANCE MODE
005 Aluminum
0.30 ppm
%T/Abs
Calibrate
12:00:00 001/500
1. Press and briefly hold to turn the
meter on. The LaMotte logo screen will
appear for about 3 seconds and the Main Menu will appear.
2. Press
ENTER
to select Testing Menu.
3. Press or to scroll to
Absorbance.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
Testing Menu
All Test Menu
Sequence 1
Sequence 2
Sequence 3
12:00:00 001/500
Testing Menu
Sequence 1
Sequence 2
Sequence 3
Absorbance
12:00:00 001/500
28 SMART3 Colorimeter 09.16
Page 29
4. Press
ENTER
to select Absorbance.
Absorbance
101 Absorbance 428
102 Absorbance 525
103 Absorbance 568
104 Absorbance 635
12:00:00 001/500
Testing
5. Press or to scroll to desired
wavelength.
6. Press
7. Insert the blank. Close the lid. Press
ENTER
to select the wavelength.
ENTER
to scan the blank. The screen wil display
Scan Blank Blank Done for about 1 second
and return to the Absorbance menu.
Absorbance
101 Absorbance 428
102 Absorbance 525
103 Absorbance 568
104 Absorbance 635
12:00:00 001/500
102 Absorbance 525
Scan Blank
Scan Sample
12:00:00 001/500
102 Absorbance 525
Scan Blank
Scan Sample
12:00:00 001/500
8. Insert the reacted sample. Press
ENTER
to
102 Absorbance 525
scan the sample. The screen will display
Scan Sample Sample Done for about
0.425
1 second. The result will appear on the
screen.
Scan Sample
Next Test
12:00:00 001/500
SMART3 Colorimeter 09.16 29
Page 30
9. To repeat the test, scroll to Scan Sample
and press
again. The last blank scanned is used
by the colorimeter for repeated scans. A
different blank can be used by pressing
Testing
and then scanning another blank.
Scroll with or and make
another selection with
Absorbance of the last test can be viewed
by choosing %T/Abs. Press
escape to previous menus.
NOTE: The menus loop in this screen so
either or will lead to the
menu selection needed.
NOTE: The calibrate function does not work
in the Absorbance mode.
ENTER
to scan the sample
or to scroll to Scan Blank
ENTER
. The %T or
EXIT
to
0.425
Scan Sample
Next Test
12:00:00 001/500
Previous Test
%T/Abs
Calibrate
Print Test
102 Absorbance 525
EDITING MENU
The Editing Menu allows the user to edit sequences, edit user tests, set the clock, edit the
logging function, access factory setting, set the power saving function, set the backlight
time, and select a language.
The default factory settings are:
Date FormatMM-DD-YYYY
LoggingEnabled
Power Save5 minutes
Backlight10 seconds
LanguageEnglish
EDITING A SEQUENCE
The Edit Sequence menu allows three alterable test sequences (Sequence 1, Sequence 2,
Sequence 3) to be edited.
1. Press and briefly hold to turn the
meter on. The LaMotte logo screen will
appear for about 3 seconds and the Main Menu will appear.
30 SMART3 Colorimeter 09.16
Testing Menu
Editing Menu
12:00:00 001/500
Main Menu
Page 31
2. Press or to scroll to the
Editing Menu.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
Testing
3. Press
4. Press
ENTER
to select Editing Menu.
ENTER
to select Edit Sequences.
5. Press or to scroll to the
desired sequence.
Editing Menu
Edit Sequences
Edit User Test
Set Clock
Logging
12:00:00 001/500
Edit Sequences
Edit Sequence 1
Edit Sequence 2
Edit Sequence 3
12:00:00 001/500
Edit Sequences
Edit Sequence 1
Edit Sequence 2
Edit Sequence 3
12:00:00 001/500
6. Press
edited.
ENTER
to select the sequence to be
005 Aluminum
EDIT SEQUENCE 2
025 Chlorine T DPD
056 Iron Phenanthro
060 Manganese LR
12:00:00 001/500
SMART3 Colorimeter 09.16 31
Page 32
ADDING OR DELETING A TEST
There are three ways to alter a sequence: Insert Before, Insert After, and Delete. Insert
Before adds a new test to the sequence before the selected test. Insert After adds a new
test to the sequence after the selected test. Delete is used to remove an existing test from
a sequence.
Testing
ADDING A TEST
Below is a step-by-step example of how to add a test to SEQUENCE 2 starting from the
EDIT SEQUENCE 2 menu.
1. To add a test before or after an existing
test, press or to scroll to the
existing test.
2. Press
ENTER
to select the existing test.
3. Press or to scroll to Insert Before or Insert After.
EDIT SEQUENCE 2
005 Aluminum
025 Chlorine T DPD
056 Iron Phenanthro
060 Manganese LR
12:00:00 001/500
Add or Delete
Insert Before
Insert After
Delete
12:00:00 001/500
Add or Delete
Insert Before
Insert After
Delete
12:00:00 001/500
4. Press
ENTER
to select the option, Insert
Before, in this example. The All Test Menu
will appear.
002 Alkalinity UDV
005 Aluminum
All Tests
006 Ammonia-N LRF
007 Ammonia-N LRS
12:00:00 001/500
32 SMART3 Colorimeter 09.16
Page 33
5. Press or to scroll to the test
that will be added to the sequence. In this
example, 006 Ammonia-N LRS.
All Tests
002 Alkalinity UDV
005 Aluminum
006 Ammonia-N LRF
007 Ammonia-N LRS
12:00:00 001/500
Testing
6. Press
ENTER
to select the test. The
sequence will appear in the Edit Sequence
menu and the new test will be added to the
sequence. All changes in the sequence will
be automatically saved.
7. Press
EXIT
to exit the Edit Sequence
menu and return to the Editing Sequences Menu.
8. Press
ENTER
to select Edit Sequences
to continue editing the sequences or
EXIT
press
twice to return to the Main
Menu.
EDIT SEQUENCE 2
005 Aluminum
025 Chlorine T DPD
006 Ammonia-N LF
056 Iron Phenanthro
12:00:00 001/500
Edit Sequences
Edit Sequence 1
Edit Sequence 2
Edit Sequence 3
12:00:00 001/500
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
SMART3 Colorimeter 09.16 33
Page 34
DELETING A TEST
Below is a step-by-step example of how to delete a test in SEQUENCE 2 starting from the
EDIT SEQUENCE 2 menu.
1. To delete a test, press or to
scroll to the test in the sequence.
Testing
2. Press
ENTER
to select the test.
3. Press or to scroll to
Delete.
EDIT SEQUENCE 2
005 Aluminum
025 Chlorine T DPD
006 Ammonia-N LF
056 Iron Phenathro
12:00:00 001/500
Add or Delete
Insert Before
Insert After
Delete
12:00:00 001/500
Add or Delete
Insert Before
Insert After
Delete
12:00:00 001/500
4. Press
ENTER
to select Delete. The
sequence will appear in the Edit Sequences menu and the selected test
will have been deleted. All changes to the
sequence will automatically have been
saved.
5. Press
EXIT
to exit the Edit Sequence
menu and return to the Edit Sequences Menu.
005 Aluminum
025 Chlorine T DPD
056 Iron Phenanthro
060 Manganese LR
12:00:00 001/500
Edit Sequence 1
Edit Sequences 2
EDIT SEQUENCE 2
Edit Sequences
Edit Sequences 3
12:00:00 001/500
34 SMART3 Colorimeter 09.16
Page 35
6. Press
EXIT
to exit the Edit Sequences
menu and return to the Editing Menu.
Editing Menu
Edit Sequences
Edit User Test
Set Clock
Logging
12:00:00 001/500
Testing
7. Press
ENTER
to select Edit Sequences to
continue editing the sequences or press
EXIT
to return to the Main Menu.
Testing Menu
Editing Menu
12:00:00 001/500
Main Menu
EDIT USER TESTS
If a test other than the LaMotte programmed tests is performed regularly, a calibration for
it may be entered in one of the 25 User Tests. These tests are originally named “User Test
1 - 25”. It will be possible to rename the test, select a wavelength, enter a new calibration,
select the number of decimal places used to display the results, and select the units. A
User Test may be added for a reagent system for which no precalibrated test exists. A
calibration of a LaMotte reagent system may also be entered. The calibration of a User Test
can be changed at any time.
The User Tests have the ability to handle 2 data points. The colorimeter will determine the
absorbance of the standards and calculate a response that will be stored to determine the
concentration of future samples of unknown concentration. These standards should cover
all the concentrations for the range of the test being performed and be scanned beginning
with the low concentration and finishing with the high concentration (for more information
about this, see CALIBRATION CURVES, page 16). Prepare these standards prior to entering
a new calibration.
NOTE: A calibration procedure must be performed before using any of the User Tests.
The User Tests can be placed in any of the alterable sequences using Edit Sequences.
1. Press and briefly hold to turn the
meter on. The LaMotte logo screen will
appear for about 3 seconds and the Main Menu will appear.
SMART3 Colorimeter 09.16 35
Testing Menu
Editing Menu
12:00:00 001/500
Main Menu
Page 36
2. Press or to scroll to the
Editing Menu.
Testing
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
3. Press
ENTER
to select Editing Menu. Press
to scroll to Edit User Test.
4. Press
ENTER
to select Edit User Test.
5. Press or to scroll to the
desired user test.
Editing Menu
Edit Sequences
Edit User Test
Set Clock
Logging
12:00:00 001/500
Edit User Test
105 USER TEST 01
106 USER TEST 02
107 USER TEST 03
108 USER TEST 04
12:00:00 001/500
Edit User Test
108 USER TEST 04
109 USER TEST 05
110 USER TEST 06
111 USER TEST 07
12:00:00 001/500
6. Press
ENTER
to select the User Test.
111 USER TEST 07
Name the Test
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
36 SMART3 Colorimeter 09.16
Page 37
NAMING THE TEST
A User Test can be up to 16 characters long. The menu choices for each character are 26
upper case letters A to Z, 26 lower case letters a to z, ten numerals 0 to 9, a space, a dash
(-) and a decimal point (.). The existing name is displayed on the bottom line of the display.
The character which is to be edited will blink and that character is also displayed in the
center of the display. The character can be changed by using or to scroll to
other characters. Use
pressing
EXIT
ENTER
to select a character. The edited name is saved at any time by
or by pressing
ENTER
after selecting the sixteenth character.
Testing
1. From the User Test menu, press to
scroll to Name the Test.
2. Press
ENTER
to select Name the Test. A
blinking reverse font (dark background
with a light character) will appear to
indicate the character that will be
adjusted. The same character will also
appear in the center of the display.
3. Press or to scroll to the
desired character. In this example, the test
name will be H2O.
The character for a blank space is located
after the lower case z.
111 USER TEST 07
Name the Test
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
Name the Test
P Q R S T U V W X Z
111 USER TEST 07
12:00:00 001/500
Name the Test
C D E F G H I J K L M
111 USER TEST 07
12:00:00 001/500
4. Press
ENTER
to save the character and
Name the Test
move to the next character.
N O P Q R S T U V W X
111 HSER TEST 07
12:00:00 001/500
SMART3 Colorimeter 09.16 37
Page 38
5. Press or to scroll to the
desired character.
Name the Test
X Y Z . 0 1 2 3 4 5 6 7
111 HSER TEST 07
12:00:00 001/500
6. Press
ENTER
to save the character and
move to the next character.
7. Press or to scroll to the
Editing/Set Up
desired character.
8. Press
ENTER
to save the character. Repeat
the procedure until the test name is
complete.
After selecting the last character in
the name: Hold down and press
and release
ENTER
to save all selected
characters and to delete all characters to
the right of the cursor and exit the name
edit screen.
Hold down and press and release
ENTER
to save all remaining characters
and exit the name edit screen.
The meter will return to the Edit Test
manu.
Name the Test
9 A B C D E F G H I J K
111 H2ER TEST 07
12:00:00 001/500
Name the Test
J K L M N O P Q R S T
111 H2OR TEST 07
12:00:00 001/500
111 H2O
Name the Test
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
38 SMART3 Colorimeter 09.16
Page 39
SELECT THE VIAL AND WAVELENGTH
The Smart3 Colorimeter accepts three different vials (the 25 mm 0290 tube, UDVs and
COD tubes) at 4 different wavelengths (428, 525, 560, and 635 nm). The colorimeter uses
different settings for each of the twelve combinations of vial and wavelength.
9. From the User Test menu, press or
to scroll to Select Vial/WL.
10. Press
ENTER
to select Select Vial/WL.
11. Press or to scroll to the
channel with the desired wavelingth and
vial size combination. The screen will
display storing... for about 1 second and
the meter will return to the User Test
menu.
111 H2O
Name the Test
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
Select Vial/WL
Ch1 428nm 25mm
Ch2 525nm 25mm
Ch3 635nm 25mm
Ch4 568nm 25mm
12:00:00 001/500
Select Vial/WL
Ch1 428nm 25mm
Ch2 525nm 25mm
Ch3 635nm 25mm
Ch4 568nm 25mm
12:00:00 001/500
Editing/Set Up
12. Press
ENTER
to select the Vial/WL option.
The screen will display Storing... for about
1 second and the meter will return to the
Edit Test menu.
Name the Test
Select Vial/WL
111 H2O
STD Calibration
Enter Constants
12:00:00 001/500
ENTERING A TWO POINT CALIBRATION
The Smart3 Colorimeter can scan two reacted standards and create a calibration curve. To
create a calibration curve with multiple data points see Entering a Multiple Point Calibration
Curve (pg. 51).
SMART3 Colorimeter 09.16 39
Page 40
1. From the User Test menu, press
or to scroll to STD (Standard)
Calibration.
111 H2O
Name the Test
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
2. Press
ENTER
to select STD Calibration. The
values for the standards will be entered.
The screen will display the Low Standard
screen. A blinking reverse font (dark
background with a light character) will
appear to indicate the character that will
be adjusted. The same character will also
appear in the center of the display.
Editing/Set Up
3. Press or to scroll to the first
character of the low concentration. In this
example, 1.00 ppm.
4. Press
ENTER
to save the character and
move to the next character.
Low Standard
6 7 8 9 . - 0 1 2 3 4 5
0 0.000000
12:00:00 001/500
Low Standard
7 8 9 . - 0 1 2 3 4 5 6
0 .000000
12:00:00 001/500
Low Standard
5 6 7 8 9 . - 0 1 2 3 4
1 . 000000
12:00:00 001/500
5. Press or to scroll to the
Low Standard
desired character.
5 6 7 8 9 . - 0 1 2 3 4
1 . 000000
12:00:00 001/500
40 SMART3 Colorimeter 09.16
Page 41
6. Press
ENTER
to save the character and
move to the next character.
Low Standard
6 7 8 9 . - 0 1 2 3 4 5
1. 0 00000
12:00:00 001/500
7. Press or to scroll to the
desired character.
8. Press
ENTER
to save the character. Repeat
the procedure until the low concentration
value is complete. (the character for
a blank space is located after the 9.)
After the final character is complete the
meter will save the low concentration
value. The screen will display the High Standard screen. A blinking reverse font
(dark background with a light character)
will appear to indicate the character that
will be adjusted. The same character will
appear in the center of the display.
9. Use or and
ENTER
to select
the characters for the high concentration
value. In this example, 7.5 ppm.
Low Standard
6 7 8 9 . - 0 1 2 3 4 5
1. 0 00000
Editing/Set Up
12:00:00 001/500
High Standard
6 7 8 9 . - 0 1 2 3 4 5
0 .0000
12:00:00 001/500
High Standard
6 7 8 9 . - 0 1 2 3 4 5
7.500000
12:00:00 001/500
SMART3 Colorimeter 09.16 41
Page 42
10. After the final character is entered the
meter will save the high concentration
value. The screen will display instructions
for completing the calibration
procedure.
STD Calibration
Insert Blank
<Enter> continue
12:00:00 001/500
11. The standards wil be scanned. Insert the
blank. Press
ENTER
and the Insert Low
Standard screen will appear.
12. Insert the low standard. Press
Editing/Set Up
ENTER
The Insert High Standard screen will be
displayed.
13. Insert the high standard. Press
ENTER
meter will display the constants for the
calibration curve.
.
. The
STD Calibration
Insert Low Standard
<Enter> continue
12:00:00 001/500
STD Calibration
Insert High Standard
<Enter> continue
12:00:00 001/500
K0 = 0.57246
K1 = 7.14547
K2 = 0.00000
K3 = 0.00000
<Enter> Save
12:00:00 001/500
14. Press
ENTER
to save the calibration curve
and return to the Edit User Test Menu.
Name the Test
111 H2O
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
ENTERING A MULTIPLE POINT CALIBRATION
The Smart3 can directly create a 2 point calibration curve. (See Entering a Two Point
Calibration on page 48.) To create a multiple point calibration curve, constants obtained
42 SMART3 Colorimeter 09.16
Page 43
from a linear regression of multiple data points can be entered into the Smart3.
1. Scan reactions of multiple concentrations at the appropriate wavelength in the
absorbance mode on the Smart3.
2. Plot the concentration (y axis) versus absorbance (x axis) in a program capable of
linear regression such as Excel.
3. Enter the constants obtained from the linear regression equation into the Smart3.
For Example:
y = 0.001x3 – 0.017x2 + 0.181x – 0.049
K0 = – 0.049
K1 = 0.181
K2 = – 0.017
K3 = 0.001
OR (Over Range) = 10
1. From the User Test menu, press or
to scroll to Enter Constants.
Name the Test
111 H2O
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
Editing/Set Up
2. Press
ENTER
to select Enter Constants.
K0=0.00000
K1=1.00000
K2=0.00000
K3=0.00000
OR=100.00000
12:00:00 001/500
3. Press
ENTER
to begin entering the values
Enter Constants K0
for the constants. A reverse font (dark
background with a light character) will
appear to indicate the character that will
be adjusted. The same character will also
6 7 8 9 . - 0 1 2 3 4 5
0 .00000
appear in the center of the display.
12:00:00 001/500
SMART3 Colorimeter 09.16 43
Page 44
4. Press or to scroll to the
first character of K0. In this example, –
0.049.
Enter Constants K0
6 7 8 9 . - 0 1 2 3 4 5
0 .00000
12:00:00 001/500
5. Press
ENTER
to save the character and
move to the next character.
6. Press or to scroll to the next
Editing/Set Up
character.
7. Press
ENTER
to save the character and
move to the next character. Press
or to scroll to the next character.
Note: The character for a blank space is
located after the 9.
Enter Constants K0
5 6 7 8 9 . - 0 1 2 3 4
– .00000
12:00:00 001/500
Enter Constants K0
6 7 8 9 . - 0 1 2 3 4 5
– .00000
12:00:00 001/500
Enter Constants K0
5 6 7 8 9 . - 0 1 2 3 4
– 0 0 0000
12:00:00 001/500
8. Press
ENTER
to save the character.
Enter Constants K1
Repeat the procedure until the K0 value
is complete. After the final character is
complete the meter will save the K0 value
and the meter will display K1 screen.
44 SMART3 Colorimeter 09.16
6 7 8 9 . - 0 1 2 3 4 5
0 .00000
12:00:00 001/500
Page 45
9. Use , and
ENTER
to select the
characters for the remaining values: K1,
K2, K3, and over range. In this example, 10
ppm.
Constants OR
6 7 8 9 . - 0 1 2 3 4 5
10.000000
12:00:00 001/500
10. After the final character is entered the
meter will save the constants. The screen
will display Storing... and return to the Edit Test menu.
Name the Test
Select Vial/WL
111 H2O
STD Calibration
Enter Constants
12:00:00 001/500
SELECTING THE NUMERICAL FORMAT OF THE RESULT
To input tests with very different ranges, the number of decimal places displayed for a
result can be selected. A test which ranges from 20 to 1000 ppm should not be displayed
with three decimal places. A test with a range from 0.010 to 0.500 needs three decimal
places (the microprocessor will always calculate the concentration to many more
significant figures than will be displayed). The choice of 0, 1, 2, or 3 decimal places are
available.
1. From the User Test menu, press or
to scroll to Decimal Places.
Select Vial/WL
STD Calibration
Enter Constants
Decimal Places
12:00:00 001/500
111 H2O
Editing/Set Up
2. Press
ENTER
to select Decimal Places.
Decimal Places
None 0
One 0.0
Two 0.00
Three 0.000
12:00:00 001/500
SMART3 Colorimeter 09.16 45
Page 46
3. Press or to scroll to the
desired number of decimal places.
Decimal Places
None 0
One 0.0
Two 0.00
Three 0.000
12:00:00 001/500
4. Press
ENTER
to select the decimal places.
The screen wil display Storing... for about
1 second and the meter will return to the
Edit Test menu.
Select Vial/WL
STD Calibration
111 H2O
Enter Constants
Decimal Places
12:00:00 001/500
Editing/Set Up
SELECTING THE UNITS OF CONCENTRATION
The Smart3 Colorimeter has seven options for units of concentration. They are No Units,
ppm, pH, FAU, ppb, ppt and mgL.
1. From the User Menu, press to scroll
to Select Units.
STD Calibration
111 H2O
Enter Constants
Decimal Places
Select Units
12:00:00 001/500
2. Press
ENTER
to select Select Units.
Select Units
No Units
ppm
pH
FAU
12:00:00 001/500
3. Press or to scroll to the
desired units.
No Units
Select Units
ppm
pH
FAU
12:00:00 001/500
46 SMART3 Colorimeter 09.16
Page 47
4. Press
ENTER
to select the units. The screen
will display Storing... for about 1 second
and the meter will return to the Edit Test
menu.
STD Calibration
Enter Constants
111 H2O
Decimal Places
Select Units
12:00:00 001/500
SETTING THE CLOCK
Setting the clock allows the correct time and date stamp to be stored with each reading in
the data logger.
1. From the Editing Menu, press or
to scroll to Set Clock.
2. Press
ENTER
to select Set Clock. The year
is displayed. Press or to scroll
to the appropriate character. Press
ENTER
to select the character. The month, day,
hour, format hour, minute, second, AM/PM
will be dislpayed. Repeat for each.
3. Press
ENTER
to select the final character.
The time and date will be saved and the
meter will return to the EditingMenu.
Editing Menu
Edit Sequences
Edit User Test
Set Clock
Logging
12:00:00 001/500
Set Time
Year: 2000
12:00:00 001/500
Editing Menu
Edit Sequences
Edit User Test
Set Clock
Logging
12:00:00 001/500
Editing/Set Up
LOGGING DATA
The default setting for the data logger is enabled. The meter will log the last 500 data
points. The counter in the center bottom of the display will show how many data points
have been logged. The display will show 500+ when the data logger has exceeded 500
points and the data points are being overwritten.
SMART3 Colorimeter 09.16 47
Page 48
1. From the Editing Menu, press or
to scroll to Logging.
Editing Menu
Edit Sequences
Edit User Test
Set Clock
Logging
12:00:00 001/500
2. Press
3. Press or to scroll to desired
Editing/Set Up
ENTER
to select Logging.
function.
4. Press
ENTER
. The screen will display
Storing... for about 1 second and return to
the EditingMenu.
Logging
Display Test Log
Logging Enabled
Logging Disabled
Erase Log
12:00:00 001/500
Logging
Display Test Log
Logging Enabled
Logging Disabled
Erase Log
12:00:00 001/500
Editing Menu
Edit Sequences
Edit User Test
Set Clock
Logging
12:00:00 001/500
FACTORY SETUP
The Factory Setup menu is used in manufacturing of the Smart3 Colorimeter. This menu is
not for use by the operator in the field.
SETTING POWER SAVE
The power saving Auto Shutoff feature will turn the meter off when a button has not been
pushed for a set amount of time. The default setting is disabled. To change the setting:
48 SMART3 Colorimeter 09.16
Page 49
1. From the Editing Menu, press or
to scroll to Set Power Save.
Editing Menu
Set Clock
Logging
Factory Setup
Set Power Save
12:00:00 001/500
2. Press
ENTER
to select Set PWR Save.
3. Press or to scroll to desired
function.
4. Press
ENTER
. The screen will display
Storing.... for about 1 second and the
meter will retun to the Editing Menu.
Set PWR Save
Disable
5 Minutes
15 Minutes
30 Minutes
12:00:00 001/500
Set PWR Save
Disable
5 Minutes
15 Minutes
30 Minutes
12:00:00 001/500
Editing Menu
Set Clock
Logging
Factory Setup
Set PWR Save
12:00:00 001/500
Editing/Set Up
SETTING THE BACKLIGHT TIME
The backlight illuminates the display for enhanced viewing. The default setting is 10
seconds. If Button Control is chosen the backlight button on the key pad will act as an on/
off switch and the backlight will remain on or off when the meter is being used. When one
of the other settings – 10, 20 or 30 seconds – is chosen, the display will be illuminated
for the specified amount of time after any button is pressed. NOTE: The backlight feature
uses a significant amount of power. The longer the backlight is on, the more frequently the
battery will have to be charged if the USB/Wall Adapter is not being used.
SMART3 Colorimeter 09.16 49
Page 50
1. From the Editing Menu, press or
to scroll to Backlight Time.
Editing Menu
Logging
Factory Setup
Set Power Save
Set Backlight Time
12:00:00 001/500
2. Press
ENTER
to select Set Backlight
Time.
3. Press or to scroll to desired
Editing/Set Up
option.
4. Press
ENTER
. The screen will display
Storing... for about 1 second and the
meter will return to the Editing Menu.
Set Backlight Time
Button Control
10 seconds
20 seconds
30 seconds
12:00:00 001/500
Set Backlight Time
Button Control
10 seconds
20 seconds
30 seconds
12:00:00 001/500
Editing Menu
Logging
Factory Setup
Set Power Save
Set Backlight Time
12:00:00 001/500
BLUETOOTH MENU
Enabling Bluetooth Power
Bluetooth wireless technology allows communication between the colorimeter and a
Bluetooth Mobile Printer (Code 5-0066 only). Bluetooth Power must be enabled to use the
Bluetooth Mobile Printer (Code 5-0066). When Bluetooth Power is enabled, the Bluetooth
icon will flash next to the battery icon on the bottom line of the display.
If the Bluetooth feature is not being used to connect to the printer, Bluetooth Power should
be disabled to conserve the battery life. The default Bluetooth Power setting is disabled.
50 SMART3 Colorimeter 09.16
Page 51
1. From the Editing Menu, press or
to scroll to Bluetooth Menu.
Editing Menu
Factory Setup
Set Power Save
Set Backlight Time
Bluetooth Menu
12:00:00 001/500
2. Press
ENTER
to select Bluetooth Menu.
3. Press or to scroll to
Bluetooth Power.
4. Press
ENTER
to select Bluetooth
Power.
Bluetooth Menu
Bluetooth Power
Bluetooth Mode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500
Bluetooth Menu
Bluetooth Power
Bluetooth Mode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500
Bluetooth Power
Enable Bluetooth
Disable Bluetooth
12:00:00 001/500
Editing/Set Up
5. Press or to scroll to desired
option.
Enable Bluetooth
Bluetooth Power
Disable Bluetooth
12:00:00 001/500
SMART3 Colorimeter 09.16 51
Page 52
6. Press
ENTER
to select the option.
Bluetooth Power must be enabled to
use the Bluetooth Mobile Printer (Code
5-0066).
The screen will display Storing… for about
Bluetooth Power
Bluetooth Mode
Bluetooth Scan
Set BT Timeout
Bluetooth Menu
1 second and the meter will return to the
Bluetooth Menu.
12:00:00 001/500 P
A flashing P symbol will indicate that
Bluetooth is enabled.
Setting The Bluetooth Mode
A Bluetooth connection is available only for the Bluetooth Mobile Printer (Code 5-0066).
1. From the Editing Menu, press or
Editing/Set Up
2. Press
to scroll to Bluetooth Menu.
ENTER
to select Bluetooth Menu.
3. Press or to scroll to
Bluetooth Mode.
Editing Menu
Factory Setup
Set Power Save
Set Backlight Time
Bluetooth Menu
12:00:00 001/500 P
Bluetooth Menu
Bluetooth Power
Bluetooth Mode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
Bluetooth Menu
Bluetooth Power
Bluetooth Mode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
52 SMART3 Colorimeter 09.16
Page 53
4. Press
Mode.
ENTER
to select Bluetooth
Bluetooth Mode
Mobile Device
BT Printer
12:00:00 001/500 P
5. Press or to scroll to BT Printer.
Mobile Device
Bluetooth Mode
BT Printer
12:00:00 001/500 P
6. Press
ENTER
. The screen will display
Storing… for about 1 second and the
meter will return to the Bluetooth Menu.
Bluetooth Power
Bluetooth Mode
Bluetooth Menu
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
Scan For Printers
The Smart3 can only be used with Bluetooth Mobile Printer (Code 5-0066).
1. Turn on the power to the Bluetooth Mobile
Printer (Code 5-0066).
Editing/Set Up
SMART3 Colorimeter 09.16 53
Page 54
2. From the Editing Menu, press or
to scroll to Bluetooth Menu.
Editing Menu
Factory Setup
Set Power Save
Set Backlight Time
Bluetooth Menu
12:00:00 001/500 P
3. Press
ENTER
to select Bluetooth
Menu.
Editing/Set Up
4. Press or to scroll to
Bluetooth Scan.
5. Press
ENTER
to select Bluetooth Scan.
The meter will begin scanning. A count
down timer will begin counting down from
11 seconds.
If “No BT Available…” is displayed,
Bluetooth is not enabled on the meter. Go
to Bluetooth Power to enable it.
Bluetooth Menu
Bluetooth Power
Bluetooth Mode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
Bluetooth Menu
Bluetooth Power
Bluetooth Mode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
Bluetooth Scan
Scanning (11)...
12:00:00 001/500 P
54 SMART3 Colorimeter 09.16
Page 55
6. The addresses of the available printers
will be displayed. NOTE: The printer
address for the Bluetooth Mobile Printer
(Code 5-0066) can be determined by
turning the power off and then pressing
the Power button and the Paper Feed
button simultaneously to print the printer
specifications. The printer address is
designated as Bluetooth: BT Address.
Select Printer
000D163140A0
Set Backlight Time
12:00:00 001/500 P
Editing/Set Up
7. Scroll to the desired printer and press
ENTER to pair the printer. The screen will
display the BT Address and Working… for
about 1 second. The meter will display
000D163140A0
Connecting
Connect to Printer
a status message when the printer is
connected.
12:00:00 001/500 P
Set BT Timeout
The Bluetooth Timeout can be set to allow the Bluetooth feature to be on to allow data to
be sent to the Bluetooth printer and then turn off after a specified amount of time.
1. From the Editing Menu, press or
to scroll to Bluetooth Menu.
Factory Setup
Set Power Save
Set Backlight Time
Bluetooth Menu
12:00:00 001/500 P
Editing Menu
SMART3 Colorimeter 09.16 55
Page 56
2. Press
ENTER
to select Bluetooth Menu.
Bluetooth Menu
Bluetooth Power
BluetoothMode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
3. Press or to scroll to Set BT
Timeout.
Editing/Set Up
4. Press
ENTER
to select Set BT Timeout.
5. Press or to scroll to desired
option.
Bluetooth Menu
Bluetooth Power
BluetoothMode
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
Set BT Timeout
15 minutes
5 minutes
30 Seconds
Disable
12:00:00 001/500 P
Set BT Timeout
15 minutes
5 minutes
30 Seconds
Disable
12:00:00 001/500 P
6. Press
ENTER
. The screen will display
Storing… for about 1 second and the
meter will return to the Bluetooth Menu.
Bluetooth Power
Bluetooth Mode
Bluetooth Menu
Bluetooth Scan
Set BT Timeout
12:00:00 001/500 P
56 SMART3 Colorimeter 09.16
Page 57
SELECTING A LANGUAGE
There are seven languages available in the Smart3: English, Spanish, French,
Portuguese, Italian, Chinese, and Japanese.
1. From the Editing Menu, press or
to scroll to Select Language.
2. Press
ENTER
to select Select
Language.
3. Press or to scroll to desired
language.
Editing Menu
Set Power Save
Set Backlight Time
Bluetooth Menu
Select Language
12:00:00 001/500
Select Language
English
Spanish
French
Portugese
12:00:00 001/500
Select Language
English
Spanish
French
Portugese
12:00:00 001/500
Editing/Set Up
4. Press
ENTER
. The screen will display
Storing... for about 1 second and the
meter will return to the Editing Menu.
Set Power Save
Set Backlight Time
Editing Menu
Bluetooth Menu
Select Language
12:00:00 001/500
NOTE: If meter unintentionally switches to another language, use the procedure above to
reset the meter to the desired language. For example, to reset the meter to English:
Turn meter on.
1. Press one time. Press
2. Press eight times. Press
3. Press
SMART3 Colorimeter 09.16 57
ENTER
.
ENTER
ENTER
.
.
Page 58
LOOPING MENUS
Long menus, such as All Tests, incorporate a looping feature which allows the user to
quickly reach the last choice in the menu from the first choice. In a looping menu the last
choices in the menu are above the first choice and scrolling upward moves through the
menu in reverse order. Scrolling downward moves through the menu from first choice to
last but the menu starts over following the last choice. So all menu choices can be reached
by scrolling in either direction. The diagrams below demonstrate a looping menu.
The feature called Looping Menu can be turned on and off in the Editing Menu. The default
setting is ON.
: : :: : :
SECOND TO LASTSECOND TO LAST
COMPUTER CONNECTION
OUTPUT
USB
COMPUTER CONNECTION
USB Type A, USB mini B, Order Cable Code 1720-01.
BLUETOOTH CONNECTION
Barriers to wireless signals can reduce the range of wireless devices. The Smart3 will work
best if there are no walls between it and the receiving device.
The Smart3 colorimeter uses Bluetooth wireless technology to allow communication
between the colorimeter and a bluetooth printer (Code 3-0066 only).
58 SMART3 Colorimeter 09.16
Page 59
WATERLINK CONNECT
The meter may be interfaced with any Windows-based 64 bit computer by using the
LaMotte WaterLink Connect program and a USB cable. The program will store test
information and results in a local database, and allow for exporting this data to a comma
separated value (CSV) file. The meter will send the following data: time/date stamp, name
of test, sample value, sample units, meter name, and location. To download WaterLink go
to http://softwarecenter.lamotte.com/. Select “WaterLink Connect Software for Windows
FREE Download”.
BATTERY
BATTERY/AC OPERATION
The Smart3 may be operated on battery power using a USB wall adapter or USB computer
connection. If using the meter as a bench top unit, use the wall adapter if possible to
extend the battery life. The meter will remain on when the USB adapter is used.
To charge the battery with the wall adapter, plug the smaller end of the USB cable
(USB mini B connector) into the meter and the larger end of the USB cable (USB type
A connector) into the wall adapter. Plug the wall adapter into an AC outlet. Reinsert the
rubber USB port plug after charging. To charge the battery from a computer, plug the
smaller end of the USB cable (USB mini B connector) into the meter and the larger end
of the USB cable (USB Type A connector) into a USB port on a computer. The battery icon
will show no bars and flash when the unit first turns on. Then the indicator will indicate the
battery status by showing 0, 1, 2, 3, or 4 bars.
It will take 6 hours to fully charge a low battery. The battery icon will flash when the
battery is charging. The battery icon will show four bars and stop flashing when it is fully
charged. The charging circuit will automatically switch to a float charge when the battery
is fully charged. The charger may remain connected. Some computers will NOT supply
power to their USB ports during standby operation. The wall adapter will charge the unit
continuously.
Storing the meter above ambient room temperature will decrease the battery charge more
quickly than storage at room temperature. If the meter does not turn on, it means that
the battery is at a very low charge. Charging the battery with the wall adapter in this state
may take up to 10 hours. At low temperatures, approaching 0 °C, the battery will charge
more slowly. It will not charge at all below 0 °C. The battery icon will show no bars and
continuously flash if the battery is getting low but the unit will still operate normally. A “Low
Battery” message on the status bar of the display will replace the time when the battery
voltage is too low for proper operation and accuracy may be degraded. A “Shutdown Low
Batt” message on the display will appear for a few seconds before the power is switched
off when the battery is too low to operate the unit. When the battery icon simultaneously
flashes bars 1 and 2 followed by bars 3 and 4, it is an indication that the battery is
damaged and technical support should be contacted.
To extend the battery life:
◊ Shut down the unit with the power switch when not taking measurements or use the
power save option to have the unit automatically turn off after 5 minutes.
◊ Store the unit at a moderate temperature.
Editing/Set Up
SMART3 Colorimeter 09.16 59
Page 60
◊ Fully charge the battery before storing the unit for extended periods of time.
◊ Fully charge the battery at least once per year. Failure to do so may result in a
permanently drained battery.
◊ Limit backlight use. The unit consumes 3X normal power with the backlight on. Set
the backlight time option to 10 seconds, or select “Button Control” and keep the
backlight off.
BATTERY REPLACEMENT
The lithium ion battery used in this unit should last for many years with normal use.
When it no longer powers the unit long enough to meet testing requirements it will
need to be replaced. Lithium ion batteries that are properly charged and stored do not
usually lose all capacity; they just have less capacity after hundreds of charge cycles.
This unit uses a custom battery assembly that is only available from LaMotte Company.
Battery replacement must be performed at a LaMotte authorized repair facility. The
water resistant housing of this meter should not be opened by the user. Contact LaMotte
Company by phone (1-800-344-3100) or email (tech@lamotte.com) for a return
authorization number.
MAINTENANCE
Editing/Set Up
CLEANING
Clean the exterior housing with a damp, lint-free cloth. Do not allow water to enter the light
chamber or any other parts of the meter. To clean the light chamber and optics area, point
a can of compressed air into the light chamber and blow the pressurized air into the light
chamber. Use a cotton swab dampened with Windex® window cleaner to gently swab the
interior of the chamber. Do not use alcohol; it will leave a thin residue over the optics when
dry.
REPAIRS
Should it be necessary to return the meter for repair or servicing, pack the meter carefully
in a suitable container with adequate packing material. A return authorization number
must be obtained from LaMotte Company by calling 800-344-3100 (US only) or 410778-3100, faxing 410-778-6394, or emailing tech@lamotte.com. Often a problem can
be resolved over the phone or by email. If a return of the meter is necessary, attach a
letter with the return authorization number, meter serial number, a brief description of
problem and contact information including phone and FAX numbers to the shipping carton.
This information will enable the service department to make the required repairs more
efficiently.
METER DISPOSAL
Waste Electrical and Electronic Equipment (WEEE)
Natural resources were used in the production of this equipment. This equipment may
contain materials that are hazardous to health and the environment. To avoid harm to
the environment and natural resources, the use of appropriate take-back systems is
60 SMART3 Colorimeter 09.16
Page 61
recommended. The crossed out wheeled bin symbol on the meter encourages the use of
these systems when disposing of this equipment.
Take-back systems will allow the materials to be reused or recycled in a way that
will not harm the environment. For more information on approved collection,
reuse, and recycling systems contact local or regional waste administration or
recycling services.
TROUBLESHOOTING
ERROR MESSAGES
OVER RANGE
If the message OVERRANGE is displayed when scanning a sample, the sample may be over
range or under range. If the sample is over range the sample should be diluted and tested
again (see Sample Dilution Techniques and Volumetric Measurements, page 19).
If overrange is displayed, press
continue testing on diluted samples.
Note: After pressing
cncentration will be displayed. This
concentration is an approximation only.
ENTER
ENTER
to
, the overrange
Overrange
<Enter> continue
Scan Blank
Scan Sample
12:00:00 001/500
CALIBRATION
As with all pre-calibrated meters, it is highly recommended, even if not required by
regulations, that the user periodically verify the performance of the meter by running
standards with a predetermined concentration. Results outside of specification are
an indication that the meter needs to be adjusted. This can be done following the user
calibration described on page 28. If the user calibration fails to properly adjust the meter
then the meter should be returned to LaMotte Company for recalibration. (See page 73).
STRAY LIGHT
The Smart3 Colorimeter should have no problems with stray light. Make sure that the
sample compartment lid is always fully closed, except when testing COD with the adapter.
TROUBLESHOOTING GUIDE
005 Aluminum
Editing/Set Up
PROBLEMREASONSOLUTION
Flashing
“Low Battery”Battery voltage is very low.
“Shut Down Low Batt”
Shut Down
SMART3 Colorimeter 09.16 61
Low battery. Readings are
reliable.
Readings are not reliable.
Battery is too low to operate
the unit.
Charge battery or use USB
wall/computer adapter.
Charge battery or use USB
wall/computer adapter.
Charge battery or use USB
wall/computer adapter.
Page 62
“Overrange”Sample is outside of
acceptable range.
Dilute sample and test
again.
Unusually large
negative or positive
Incorrect standards used to
calibrate meter.
readings when
performing calibration
SOFTWARE TROUBLESHOOTING
PROBLEMREASONSOLUTION
No printer connectedColorimeter has not been
paired via Bluetooth, or pairing
was lost by leaving Bluetooth
range.
Printer does not
appear in the list of
Bluetooth devices
Editing/Set Up
Connection lostBluetooth pairing with the
Printer in not powered on.Power on the printer
printer was lost.
Use fresh 0.0 standard in
clean tube. Reset meter to
factory default settings.
Recalibrate meter.
Re-pair the colorimeter to
the printer.
Re-pair the Printer to the
colorimeter
62 SMART3 Colorimeter 09.16
Page 63
SMART3
Colorimeter
Test
Procedures
1910-Test-V3.indd
Page 64
Page 65
SMART3 COLORIMETER REAGENT SYSTEMS
SMART3 REAGENT SYSTEMS LIST
LaMotte Company continuously updates the list of pre-programmed tests as the
calibrations become available.
Call LaMotte Technical Services at 1-800-344-3100 (410-778-3100 outside the USA)
or email at tech@lamotte.com for a current list of available calibrations and more
information. Some reagent systems are not available for shipment to Canada.
# of
Test Factor (Test #)Range (ppm)MDLTest Method (# of Reagents)
Alkalinity-UDV (002)0–25010Unit Dose Vials (1)100
Aluminum (005) †0.00–0.300.01Eriochrome Cyanine R (4)50
Silica-Low Range (087)0.00–4.000.05Heteropoly Blue (4)50
Silica-High Range (086)0–751Silicomolybdate (3)50
Sulfate-High Range (089) †0–1003Barium Chloride (1)50
Sulfide-Low Range (090)0.00–1.500.06Methylene Blue (3)50
Surfactants (092)0.0-8.00.5Bromphenol Blue (3)100
Tannin (093)0.0–10.00.1Tungsto-molybdophosphoric
Turbidity (095)0–400 FAU3Absorption (0)–
Zinc-Low Range (097) †0.00–3.000.05Zincon (6)50
***Requires COD Adapter Code 5-0087 and Heater Block.
†Available for shipment to Canada
0.00–3.50
mg/L
5-100 mg/L5Molybdovanadate/Digestion (5)25
0.50Ascorbic Acid/Digestion25
Acid (2)
25
25
50
Page 68
Page 69
ALKALINITY–UDV
UNIT DOSE VIALS • CODE 4318-J
QUANTITY CONTENTS CODE
1*Alkalinity Unit Dose Vials, 20 pouches*4318-J
Equipment needed but not supplied:
STANDARD ACCESSORY PACKAGE · CODE 1961
1Package of 3 Vials (empty)0156
1Syringe, 3 mL, plastic1184
1Foil Storage Bag9467
Or:
ADVANCED ACCESSORY PACKAGE · CODE 1962
1Pipettor, 3mL, w/tip30529-BRAND
1Cuvette Rack31695
1Package of 3 Vials (empty)0156
1Foil Storage Bag9467
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Test Procedures
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
APPLICATION:Drinking and surface waters; swimming pool water.
RANGE:0–250 ppm as CaCO
MDL:10 ppm
METHODThe sample is added to a buffered indicator reagent. The
color that develops, ranging from yellow to blue, will indicate
the amount of alkalinity in the sample.
SAMPLE HANDLING &
PRESERVATION:
INTERFERENCES:Quats and poly quats at high concentrations will interfere.
Smart3 Test Procedures 09.16 ALKALINITY, UDV
Samples should be analyzed as soon as possible after
collection. Sample may be refrigerated for 24 hours.
3
Page 70
PROCEDURE
Use COD/UDV adapter
1. Press and hold until colorimeter turns on.
2. Press
3. Select All Tests (or another sequence containing 002 Alkalinity UDV) from Testing
4. Scroll to and select 002 Alkalinity UDV from menu.
5. Rinse a clean vial (0156) with sample water.
6. Use the syringe (1184) to add 3 mL of sample to the vial.
7. Insert the vial into chamber, close lid and select Scan Blank.
8. Remove vial from the colorimeter.
9. Use the syringe (1184) to add 3 mL of sample to an *Alkalinity-UDV vial
10. Wait 90 seconds.
11. Invert vial 3 times to mix.
12. Insert tube into chamber, close lid and select Scan Sample. Record result.
13. Press to turn the colorimeter off or press
NOTES: For best possible results, a reagent blank should be determined to account for
Test Procedures
any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank.
Then follow the above procedure to perform the test on a distilled or deionized water
sample. This test result is the reagent blank. Subtract the reagent blank from all
subsequent test results of unknown samples. It is necessary to determine the reagent
blank only when a new lot number of reagents is obtained.
ENTER
to select Testing Menu.
Menu.
(4318).
NOTE: If powder residue remains in the bottom of the vial after inverting, invert
once more and tap bottom of vial sharply once or twice to dislodge powder.
Mix.
EXIT
to exit to a previous menu
or make another menu selection.
UDVs from opened pouches should be used promptly. Store unused vials from
opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the reagent.
Generally, UDVs stored in the bag should be used within 10 days if the humidity is
less than 50% and within 5 days if humidity is greater than 50%. The Foil Storage Bag
contains a desiccant pack with indicator. When the indicator in the window turns from
blue to pink, the bag should be replaced.
ALKALINITY, UDV Smart3 Test Procedures 09.16
Page 71
ALUMINUM
ERIOCHROME CYANINE R METHOD • CODE 364I-01-SC
QUANTITY CONTENTS CODE
5 g Aluminum Inhibitor Reagent 7865-C
2 x 120 mL*Aluminum Buffer Reagent*7866-J
120 mL Aluminum Indicator Reagent 7867-J
15 mL Aluminum Complexing Reagent 7868-E
1 Spoon, 0.05 g, plastic 0696
2 Pipets, 1.0 mL, plastic 0354
1 Test Tube, glass, 5 mL w/cap 0230
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Aluminum is the third most common element in the earth’s crust, which accounts for
its wide appearance in many water supplies. Aluminum exists in water as soluble salts,
colloidal compounds, and insoluble compounds. In wastewater that has been treated
by alum coagulation it will appear in one or more of the above forms. Properly treated
drinking water should have an aluminum concentration below 0.05 mg/L.
Test Procedures
APPLICATION:Drinking, surface, and saline waters; domestic and
industrial wastewater.
RANGE:0.00–0.30 ppm Aluminum
MDL:0.01 ppm
METHOD:Aluminum ions buffered to a pH of 6.0 react with
Eriochrome Cyanine R dye to produce a pink to red complex
in proportion to the concentration.
SAMPLE HANDLING &
PRESERVATION:
INTERFERENCES:Fluoride and polyphosphate will interfere. Interference from
Smart3 Test Procedures 11.18 ALUMINUM
Collect sample in acid washed glass or plastic bottle.
Analyze as soon as possible.
iron and manganese is eliminated by the addition of an
inhibitor.
Page 72
PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
3. Select All Tests (or another sequence containing 005 Aluminum) from Testing
4. Scroll to and select 005 Aluminum from menu.
5. Rinse a clean colorimeter tube (0290) with sample water. Fill to the 10 mL line
6. Insert tube into colorimeter chamber and select Scan Blank.
7. Rinse a clean test tube (0230) with sample water. Fill to the 5 mL line with
8. Remove tube from colorimeter. Empty sample from tube (0290).
9. Add 5 mL sample from test tube (0230) to empty tube (0290).
10. Use the 0.05 g spoon (0696) to add one measure of Aluminum Inhibitor Reagent
11. Use a 1.0 mL pipet (0354) to add 2 mL of *Aluminum Buffer Reagent (7866). Cap
12. Use a second 1.0 mL pipet (0354) to add 1 mL of Aluminum Indicator
13. At end of 5 minute waiting period, mix, insert tube into chamber, close lid and
14. Press to turn the colorimeter off or press
Test Procedures
NOTE: For the best possible results, a reagent blank should be determined to account
for any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank. Add
5 drops of Aluminum Complexing Reagent (7868). Then follow the above procedure
to perform the test on a distilled or deionized water sample. This test result is the
reagent blank. Subtract the reagent blank from all subsequent test results of unknown
samples. It is necessary to determine the reagent blank only when a new lot number of
reagents is obtained.
ENTER
to select Testing Menu.
Menu.
with sample.
sample.
(7865). Cap and mix to dissolve powder.
and mix.
Reagent (7867). Cap and mix contents. Wait 5 minutes for maximum color
development.
select Scan Sample. Record result.
EXIT
to exit to a previous menu
or make another menu selection.
ALUMINUM Smart3 Test Procedures 11.18
Page 73
AMMONIA NITROGEN - LOW RANGE
SALICYLATE METHOD • CODE 3659-01-SC
QUANTITY CONTENTS CODE
60 mL*Salicylate Ammonia #1*3978-H
10 g*Salicylate #2 Reagent*7457-D
2 x 5 g*Salicylate #3 Reagent Powder*7458-C
1 Spoon, 0.1 g, plastic 0699
1 Spoon, 0.15 g, plastic 0727
1 Pipet, 1.0 mL, plastic 0354
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Ammonia nitrogen is present in various concentrations in many surface and ground
water supplies. Any sudden change in the concentration of ammonia nitrogen in a
water supply is cause for suspicion. A product of microbiological activity, ammonia
nitrogen is sometimes accepted as chemical evidence of pollution when encountered
in natural waters.
Test Procedures
Ammonia is rapidly oxidized in natural water systems by special bacterial groups that
produce nitrite and nitrate. This oxidation requires that dissolved oxygen be available
in the water. Ammonia is an additional source of nitrogen as a nutrient which may
contribute to the expanded growth of undesirable algae and other forms of plant
growth that overload the natural system and cause pollution.
Smart3 Test Procedures 09.16 AMMONIA NITROGEN, Low Range
Page 74
APPLICATION:Low concentrations of ammonia in fresh, brackish and salt
water; fresh and salt water aquariums.
RANGE:0.00 - 1.00 ppm Ammonia-Nitrogen
MDL:0.05 ppm Fresh Water
0.10 ppm Salt Water
METHOD:Salicylate and ammonia react at high pH in the presence of a
chlorine donor and an iron catalyst to form a blue indophenol
dye, the concentration of which is proportional to the
ammonia concentration in the sample.
SAMPLE HANDLE &
PRESERVATION:
Ammonia solutions tend to be unstable and should be
analyzed immediately. Samples may be stored for 24 hours at
4°C or 28 days at –20°C.
INTERFERENCES:There are few interferences in most natural waters. High
concentrations of reducing agents, such as hydrazine,
react with the chlorine donor and can result in negative
interferences. Color and turbidity can also interfere.
Test Procedures
AMMONIA NITROGEN, Low Range Smart3 Test Procedures 09.16
Page 75
PROCEDURE - FRESH WATER
1. Press and hold until colorimeter turns on.
2. Press
3. Select All Tests (or another sequence containing 006 Ammonia-N LRF) from
4. Scroll to and select 006 Ammonia-N LRF from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
6. Insert tube into chamber, close lid and select Scan Blank. (See Note.)
7. Remove tube from colorimeter. Use the 1.0 mL plastic pipet (0354) to add 2.0 mL
8. Use the 0.15 g spoon (0727) to add two measures of *Salicylate #2 Reagent
9. At end of 1 minute waiting period use 0.1 g spoon (0699) to add two measures
10. At the end of the 12 minute waiting period, immediately mix and insert tube into
11. Press to turn colorimeter off or press
ENTER
to select Testing Menu.
Testing Menu.
sample.
of *Salicylate Ammonia #1 (3978). Cap and mix.
(7457). Cap and mix until dissolved. Wait 1 minute.
of *Salicylate #3 Reagent Powder (7458). Cap and shake vigorously for at least
30 seconds and all solid has dissolved. Wait 12 minutes for maximum color
development.
chamber, close lid and select Scan Sample. Record result.
EXIT
to exit to a previous menu or
make another menu selection.
Test Procedures
CALCULATIONS:
To express results as Unionized Ammonia (NH3):
ppm Unionized Ammonia (NH3) =
ppm Ammonia-Nitrogen (NH3–N) x 1.2
To express results as Ionized Ammonia (NH4):
ppm Ionized Ammonia (NH
+
) =
4
ppm Ammonia-Nitrogen (NH3–N) x 1.3
To determine the percentages of Unionized and Ionized Ammonia-Nitrogen, consult
the Appendix.
NOTE: It is strongly suggested that a reagent blank be determined to account for
any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank.
Then follow the above procedure to perform the test on a distilled or deionized water
sample. This test result is the reagent blank. Subtract the reagent blank from all
subsequent test results of unknown samples. It is necessary to determine the reagent
blank only when a new lot number of reagents is obtained.
Smart3 Test Procedures 09.16 AMMONIA NITROGEN, Low Range
Page 76
PROCEDURE - SALT WATER
1. Press and hold until colorimeter turns on.
2. Press
3. Select All Tests (or another sequence containing 007 Ammonia-N LRS) from
4. Scroll to and select 007 Ammonia-N LRS from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
6. Insert tube into chamber, close lid and select Scan Blank. (See Note.)
7. Remove tube from colorimeter. Use the 1.0 mL plastic pipet (0354) to add 2.0 mL
8. Use the 0.15 g spoon (0727) to add two measures of *Salicylate #2 Reagent
9. At end of 1 minute waiting period use 0.1 g spoon (0699) to add two measures
10. At the end of the 20 minute waiting period, immediately mix and insert tube into
11. Press to turn colorimeter off or press
ENTER
to select Testing Menu.
Testing Menu.
sample.
of *Salicylate Ammonia #1 (3978). Cap and mix.
(7457). Cap and mix until dissolved. Wait 1 minute.
of *Salicylate #3 Reagent Powder (7458). Cap and shake vigorously for at least
30 seconds and all solid has dissolved. Wait 20 minutes for maximum color
development.
chamber, close lid and select Scan Sample. Record result.
EXIT
to exit to a previous menu or
make another menu selection.
CALCULATIONS:
Test Procedures
To express results as Unionized Ammonia (NH3):
ppm Unionized Ammonia (NH3) =
ppm Ammonia-Nitrogen (NH3–N) x 1.2
To express results as Ionized Ammonia (NH4):
ppm Ionized Ammonia (NH
4
+
) =
ppm Ammonia-Nitrogen (NH3–N) x 1.3
To determine the percentages of Unionized and Ionized Ammonia-Nitrogen, consult
the Appendix.
NOTE: It is strongly suggested that a reagent blank be determined to account for
any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank.
Then follow the above procedure to perform the test on a distilled or deionized water
sample. This test result is the reagent blank. Subtract the reagent blank from all
subsequent test results of unknown samples. It is necessary to determine the reagent
blank only when a new lot number of reagents is obtained.
AMMONIA NITROGEN, Low Range Smart3 Test Procedures 09.16
Page 77
AMMONIA NITROGEN - HIGH RANGE
NESSLERIZATION METHOD • CODE 3642-SC
QUANTITY CONTENTS CODE
30 mL Ammonia Nitrogen Reagent #1 V-4797-G
2 x 30 mL*Ammonia Nitrogen Reagent #2*V-4798-G
1 Pipet, 1 mL, plastic 0354
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Ammonia nitrogen is present in various concentrations in many surface and ground
water supplies. Any sudden change in the concentration of ammonia nitrogen in a
water supply is cause for suspicion. A product of microbiological activity, ammonia
nitrogen is sometimes accepted as chemical evidence of pollution when encountered
in natural waters.
Ammonia is rapidly oxidized in natural water systems by special bacterial groups that
produce nitrite and nitrate. This oxidation requires that dissolved oxygen be available
in the water. Ammonia is an additional source of nitrogen as a nutrient which may
contribute to the expanded growth of undesirable algae and other forms of plant
growth that overload the natural system and cause pollution.
Test Procedures
Smart3 Test Procedures 09.16 AMMONIA NITROGEN, High Range
Page 78
APPLICATION:Drinking, surface, and saline waters; domestic and
industrial wastes.
RANGE:0.00–4.00 ppm Ammonia Nitrogen
MDL:0.05 ppm
METHOD:Ammonia forms a colored complex with Nessler’s
Reagent in proportion to the amount of ammonia
present in the sample. Rochelle salt is added to prevent
precipitation of calcium or magnesium in undistilled
samples.
SAMPLE HANDLING &
PRESERVATION:
Ammonia solutions tend to be unstable and should be
analyzed immediately. Sample may be stored for 24
hours at 4°C or 28 days at –20°C.
INTERFERENCES:Sample turbidity and color may interfere. Turbidity may
be removed by a filtration procedure. Color interference
may be eliminated by blanking the instrument with a
sample blank.
Test Procedures
AMMONIA NITROGEN, High Range Smart3 Test Procedures 09.16
Page 79
PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
3. Scroll to and select All Tests (or another sequence containing 008 Ammonia-N HR)
4. Scroll to and select 008 Ammonia-N HR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select Scan Blank. (See Note)
7. Remove tube from colorimeter. Add 8 drops of Ammonia Nitrogen Reagent #1 (V-
8. Use the 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen Reagent #2 (V-
9. At end of the 5 minute waiting period, immediately mix, insert tube into chamber,
10. Press to turn the colorimeter off or press the
CALCULATIONS:
ENTER
to select Testing Menu.
from Testing Menu.
4797). Cap and mix. Wait 1 minute.
4798). Cap and mix. Allow 5 minutes for maximum color development.
close lid and select Scan Sample. Record result.
EXIT
exit to a previous menu
or make another menu selection.
To express results as Unionized Ammonia (NH3):
ppm Unionized Ammonia (NH3) =
ppm Ammonia-Nitrogen (NH3–N) x 1.2
To express results as Ionized Ammonia (NH4):
ppm Ionized Ammonia (NH
4
+
) =
ppm Ammonia-Nitrogen (NH3–N) x 1.3
To determine the percentages of Unionized and Ionized Ammonia-Nitrogen, consult the
Appendix.
NOTE: It is strongly suggested that a reagent blank be determined to account for any
contribution to the test result by the reagent system. To determine the reagent blank,
follow the above test procedure to scan a distilled or deionized water blank. Then follow
the above procedure to perform the test on a distilled or deionized water sample. This
test result is the reagent blank. Subtract the reagent blank from all subsequent test
results of unknown samples. It is necessary to determine the reagent blank only when a
new lot number of reagents is obtained.
Smart3 Test Procedures 09.16 AMMONIA NITROGEN, High Range
Test Procedures
Page 80
Test Procedures
AMMONIA NITROGEN, High Range Smart3 Test Procedures 09.16
Page 81
BARIUM
BARIUM CHLORIDE METHOD • CODE 3638-SC
QUANTITYCONTENTSCODE
50 gBarium Reagent Powder6330-H
1Spoon, 0.5g, plastic0698
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the four
digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a
printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Barium is a naturally occurring metal that is found in rocks and, in trace amounts, in
natural waters. A barium concentration above 2 ppm in drinking water is a classified as
a contaminant. Barium sulfate is used in the medical field as an X-ray radio contrast
agent for imaging the gastrointestinal tract. Barium has many industrial applications.
It is used in the manufacture of paints, brakes, root canal fillings, glass, and motor oil
detergents. Barium nitrate imparts a bright yellow-green color to fireworks and flares.
Barium sulfate is a component of oil well drilling mud which is used to lubricate drill bits.
Barium is also a constituent of some electro ceramics and high temperature yttrium
barium copper oxide (YBCO) superconductors.
Test Procedures
APPLICATION:Industrial
RANGE:0 – 200 ppm barium
MDL:5 ppm
METHOD:Barium is precipitated in an acid medium with sodium
sulfate to form a barium sulfate suspension in
proportion to the amount of barium present.
SAMPLE HANDLING &
PRESERVATION:
INTERFERENCES:Suspended matter and color interference may be
Smart3 Test Procedures 09.16 BARIUM
Barium samples may be preserved by refrigeration at
4°C up to 7 days in glass or plastic containers without
any change in concentration.
removed by a filtration step. Silica in excess of 500
mg/L will interfere. Strontium will interfere. Check for
stray light interference (see page 20).
Page 82
PROCEDURE
11. Press and hold until colorimeter turns on.
12. Press
13. Select All Tests (or another sequence containing 009 Barium) from Testing
14. Scroll to and select 009 Barium from menu.
15. Rinse a tube (0290) with sample water. Fill to the 10 mL line with sample.
16. Insert the tube into chamber, close the lid and select Scan Blank.
17. Remove the tube from the colorimeter.
18. Use the 0.5 g spoon (0698) to add one measure of Barium Reagent Powder (6330).
19. Insert tube into chamber, close lid and select Scan Sample. Record result in ppm
20. Press to turn the colorimeter off or press
NOTE: A while film is deposited on the inside of test tubes as a result of the barium test.
Thoroughly clean and rinse test tubes after each test.
ENTER
to select Testing Menu.
Menu.
Cap and mix by inverting for one minute.
barium.
make another menu selection.
EXIT
to exit to a previous menu or
For the most accurate results, samples and reactions should be at 25 ±4°C.
Test Procedures
BARIUM Smart3 Test Procedures 09.16
Page 83
BIGUANIDE
COLORIMETRIC METHOD • CODE 4044
QUANTITYCONTENTSCODE
2 X 60 mLBiguanide Indicator3994-H
1Pipet, plastic, 1.0 mL0354
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Biguanide is a non-chlorine, non-bromine chemical sanitizer. It is more stable than
chlorine or bromine and has little chemical odor. Biguanide is an effective bacteriacide
but, unlike chlorine and bromine, it does not destroy organic contaminants. Therefore,
hydrogen peroxide is added to biguanide pools on a regular basis to eliminate organic
contaminants. The optimum recommended level of biguanide is 30 to 50 ppm.
APPLICATION:Swimming pools
RANGE:0–70 ppm Biguanide
MDL:2 ppm
METHOD: Biguanide complexes with the proprietary indicator
to produce a colored solution. The color ranges from
yellow through green to blue depending on the biguanide
concentration.
SAMPLE HANDLING &
PRESERVATION:
INTERFERENCES:The only interfering substances that are likely to be
Smart3 Test Procedures 09.16 BIGUANIDE
Samples should be analyzed as soon as possible.
encountered in pool water are oxidized manganese and
oxidizing agents, such as chlorine, bromine and ozone.
Test Procedures
Page 84
PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
3. Select All Tests (or another sequence containing 011 Biguanide) from Testing
4. Scroll to and select 011 Biguanide from menu.
5. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.
6. Insert the tube into chamber, close lid and select Scan Blank.
7. Remove the tube from colorimeter.
8. Use the 1.0 mL pipet (0354) to add 2.0 mL of Biguanide Indicator (3994). Cap and
9. Wait 1 minute.
10. Insert the tube into chamber. Close lid.
11. Select Scan Sample. Record result in ppm Biguanide
12. Press to turn the colorimeter off or press
NOTE: For best possible results, a reagent blank should be determined to account for
any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank. Then
follow the above procedure to perform the test on a distilled or deionized water sample.
This test result is the reagent blank. Subtract the reagent blank from all subsequent
test results of unknown samples. It is necessary to determine the reagent blank only
when a new lot number of reagents is obtained.
Test Procedures
ENTER
to select Testing Menu.
Menu.
invert three times to mix.
make another menu selection.
EXIT
to exit to a previous menu or
BIGUANIDE Smart3 Test Procedures 09.16
Page 85
BORATE - UDV
UNIT DOSE VIALS • CODE 4322-J
QUANTITY CONTENTS CODE
100*Borate UDV, 20 pouches*4322-J
Equipment needed but not supplied:
STANDARD ACCESSORY PACKAGE · CODE 1961
1Package of 3 Vials (empty)0156
1Syringe, 3 mL, plastic1184
1Foil Storage Bag9467
Or:
ADVANCED ACCESSORY PACKAGE · CODE 1962
1Pipettor, 3mL, w/tip30529-BRAND
1Cuvette Rack31695
1Package of 3 Vials (empty)0156
1Foil Storage Bag9467
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the four
digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a
printed copy, contact LaMotte by email, phone or fax.
Test Procedures
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Some swimming pools use a borate buffering system. Borates lower the level of
carbon dioxide in the pool, which slows algae growth. This results in a lower chlorine
requirement. Free chlorine levels in pools with borate systems can be maintained at 1.0
ppm.
Small amounts of boron are necessary for plant growth but large amounts can be toxic.
In humans, boron aids in the uptake of calcium and the production of strong bones.
An excess of boron can affect the central nervous system resulting in a syndrome
known as borism. Some natural waters may contain small amounts of boron. Large
concentrations may be due to industrial effluent entering waterways. Boron compounds
are used in cleaning compounds, paper and paints, fertilizers, glass and ceramics,
fire retardants and the production of alloys. In the atomic energy field, boron is a
component of neutron shields and nuclear reactors.
Smart3 Test Procedures 09.16 BORATE, UDV
Page 86
APPLICATION:swimming pools, surface and saline waters, hydroponic
solutions, industrial waste.
RANGE:0.00 – 80.0 ppm boron
MDL:5 ppm
METHOD:Borate reacts with a selective boron indicator powder to
form a colored complex at pH 7.2 to 7.4 in proportion to
the concentration of boron present.
SAMPLE HANDLING &
PRESERVATION:
Store samples in polyethylene bottles. Do not use borate
detergents or glassware.
INTERFERENCES:Interferences in swimming pool water are unlikely.
Test Procedures
BORATE, UDV Smart3 Test Procedures 09.16
Page 87
PROCEDURE
Use COD/UDV adapter
1. Press until colorimeter turns on.
2. Press
3. Select All Tests (or another sequence containing 013 Borate UDV) from Testing
4. Scroll to and select 013 Borate UDV from menu.
5. Rinse a clean vial (0156) with sample water.
6. Use the syringe (1184) to add 3mL of sample to the vial.
7. Insert the vial into chamber, close the lid and select Scan Blank.
8. Remove the vial from the colorimeter.
9. Use the syringe (1184) to add 3mL of sample to a *Borate UDV vial (4322).
10. Wait 3 minute.
11. Invert the vial three times to mix.
12. Immediately insert tube into chamber, close lid and select Scan Sample. Record
13. Press to turn the colorimeter off or press
NOTE: For best possible results, a reagent blank should be determined to account for
any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank. Then
follow the above procedure to perform the test on a distilled or deionized water sample.
This test result is the reagent blank. Subtract the reagent blank from all subsequent
test results of unknown samples. It is necessary to determine the reagent blank only
when a new lot number of reagents are obtained.
ENTER
to select Testing Menu.
Menu.
NOTE: If powder residue remains in the bottom of the vial after inverting or air
bubbles form, invert once more and tap bottom of vial sharply once or twice to
dislodge powder and bubbles. Mix.
result in ppm borate.
EXIT
to exit to a previous menu or
make another menu selection.
Test Procedures
NOTE: UDVs from opened pouches should be used promptly. Store unused vials from
opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the reagent.
Generally, UDVs stored in the bag should be used within 10 days if the humidity is
less than 50% and within 5 days if humidity is greater than 50%. The Foil Storage Bag
contains a desiccant pack with indicator. When the indicator in the window turns from
blue to pink, the bag should be replaced.
Smart3 Test Procedures 09.16 BORATE, UDV
Page 88
Test Procedures
BORATE, UDV Smart3 Test Procedures 09.16
Page 89
BORON
AZOMETHINE-H METHOD · CODE 4868-01
QUANTITY CONTENTS CODE
120 mL *Boron Buffer *4869-J
10 g *Boron Indicator Powder*4870-D
1 Pipet, plastic, 1.0 mL 0354
1 Spoon, 0.15 g 0727
1 Dark storage chamber, brown 0108
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Small amounts of boron are necessary for plant growth but large amounts can be
toxic. In humans, boron aids in the uptake of calcium and the production of strong
bones. An excess of boron can affect the central nervous system resulting in a
syndrome known as borism. Some natural waters may contain small amounts of
boron. Large concentrations may be due to industrial effluent entering waterways.
Boron compounds are used in cleaning compounds, paper and paints, fertilizers, glass
and ceramics, fire retardants and the production of alloys. In the atomic energy field,
boron is a component of neutron shields and nuclear reactors. Some swimming pools
use boron buffering systems.
Test Procedures
Smart3 Test Procedures 09.16 BORON
Page 90
APPLICATION:Surface and saline waters, hydroponic solutions, industrial
waste, swimming pools.
RANGE:0.00–0.80 ppm Boron
MDL:0.05
METHOD:Azomethine-H and borate form a yellow complex at pH 6 in
proportion to the concentration of boron present.
SAMPLE HANDLING &
PRESERVATION:
Store samples in polyethylene bottles. Do not use borate
detergents or glassware.
INTERFERENCES:Interferences in drinking water are unlikely. Manganese,
zirconium, chromium, titanium, copper, vanadium,
aluminum, beryllium and iron may cause high results.
Test Procedures
BORON Smart3 Test Procedures 09.16
Page 91
PROCEDURE
1. This test requires a Reagent Blank. Rinse a tube (0290) with clear, colorless, boron
free water. Fill to 10 mL line with clear, colorless, boron free water.
2. Use the 1.0 mL pipet (0354) to add 2 mL of *Boron Buffer (4869). Cap and
mix.
3. Use the 0.15 g spoon (0727) to add one level measure of *Boron Indicator Powder
(4870). Press full spoon against side of jar to compress powder. Scrape off excess
powder on inside neck of bottle. Tap excess off spoon handle.
4. Cap and shake vigorously for 30 seconds.
5. Insert the tube into meter chamber. Close lid.
6. Start a timer set for 30 minutes. Do not open the lid during the waiting time. The
reaction is photosensitive.
7. Rinse a clean tube (0290) with Sample Water. Fill to the 10 mL line with sample
water. Repeat steps 2–4.
8. Insert the tube into the Dark Storage Chamber (0108). Close top.
9. Start a second timer set for 30 minutes. Do not open the chamber during the
waiting time. The reaction is photosensitive.
10. When 2 minutes remain on the first timer (Reagent Blank), press and hold ON
button until colorimeter turns on.
11. Press and hold until colorimeter turns on.
12. Press
13. Select All Tests (or another sequence containing 014 Boron) from Testing
14. Scroll to and select 014 Boron from menu.At the end of the Reagent Blank 30
15. Insert the tube into meter chamber, close lid and select Scan Blank.
16. Remove the tube from colorimeter.
17. At the end of the Sample Water 30 minute waiting period, remove Sample Water
18. Insert tube into meter chamber, close lid and select Scan Sample. Record result in
19. Press to turn colorimeter off or press the
ENTER
to select Testing Menu.
Menu.
minute waiting period, remove Reagent Blank tube from meter chamber. Invert
several times to mix.
tube from Dark Storage Chamber. Invert several times to mix.
ppm boron.
EXIT
to exit to a previous menu
or make another menu selection.
Test Procedures
Smart3 Test Procedures 09.16 BORON
Page 92
Test Procedures
BORON Smart3 Test Procedures 09.16
Page 93
BROMINE
LIQUID DPD METHOD · CODE 4859
QUANTITY CONTENTS CODE
30 mL DPD 1A Free Chlorine Reagent P-6740-G
30 mL*DPD 1B Free Chlorine Reagent*P-6741-G
30 mL DPD 3 Total Chlorine Reagent P-6743-G
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the four
digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a
printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Like chlorine,bromine is an effective germicidal agent employed in drinking water
treatment, pool and spa water sanitation, food service sanitation, and other public
health applications.
APPLICATION:Drinking, surface, and saline waters; swimming pool water;
domestic and industrial waters and wastes.
RANGE:0.00–9.00 Bromine
MDL:0.10 ppm
METHOD:In buffered sample bromine reacts with diethyl-p-phenylene
diamine (DPD) to produce a pink-red color in proportion to the
concentration of bromine present.
SAMPLE HANDLING
& PRESERVATION:
INTERFERENCE:The only interfering substance likely to be encountered in
Bromine in aqueous solutions is not stable, and the bromine
content of samples or solutions, particularly weak solutions,
will rapidly decrease. Exposure to sunlight or agitation will
accelerate the reduction of bromine present in such solutions.
For best results start analysis immediately after sampling.
Samples to be analyzed for bromine cannot be preserved or
stored.
water is oxidized manganese. The extent of this interference
can be determined by treating a sample with sodium arsenite
to destroy the bromine present so that the degree of
interference can be estimated.
Test Procedures
Iodine and chlorine can also interfere, but these are not
normally present unless they have been added as sanitizers.
Smart3 Test Procedures 09.16 BROMINE, Liquid, DPD
Page 94
PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
3. Select All Tests (or another sequence containing 015 Bromine L DPD) from Testing
4. Scroll to and select 015 Bromine L DPD from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select Scan Blank.
7. Remove tube from colorimeter. Add 5 drops of DPD 1A Free Chlorine Reagent
8. Insert tube into chamber, close lid and select Scan Sample.
9. Press to turn colorimeter off or press the
NOTE: To test for bromine in the presence of chlorine, add 5 drops of Glycine Solution
(6811-E) (sold separately) before adding the DPD Reagents.
ENTER
to select Testing Menu.
Menu.
(P-6740). Add 5 drops of *DPD 1B Free Chlorine Reagent (P-6741). Cap and mix.
Solution will turn pink if bromine is present.
EXIT
to exit to a previous menu or
make another menu selection.
Test Procedures
BROMINE, Liquid, DPDSmart3 Test Procedures 09.16
Page 95
BROMINE
DPD TABLET METHOD • CODE 3643-SC
QUANTITYCONTENTSCODE
100DPD #1 IG Tablets6903A-J
100DPD #3 IG Tablets6197A-J
15 mLGlycine Solution6811-E
1Tablet Crusher0175
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Like chlorine, bromine is an effective germicidal agent employed in drinking water
treatment, pool and spa water sanitization, food service sanitation, and other public
health applications.
Test Procedures
Smart3 Test Procedures 09.16 BROMINE, Tablet, DPD
Page 96
APPLICATION:Drinking, surface, and saline waters; swimming pool water;
domestic and industrial waters and wastes.
RANGE:0.00–9.00 Bromine
MDL:0.10 ppm
METHOD:In buffered sample bromine reacts with diethyl-p-phenylene
diamine (DPD) to produce a pink-red color in proportion to the
concentration of bromine present.
SAMPLE HANDLING
& PRESERVATION:
Bromine in aqueous solutions is not stable, and the bromine
content of samples or solutions, particularly weak solutions,
will rapidly decrease. Exposure to sunlight or agitation will
accelerate the reduction of bromine present in such solutions.
For best results start analysis immediately after sampling.
Samples to be analyzed for bromine cannot be preserved or
stored.
INTERFERENCE:The only interfering substance likely to be encountered in
water is oxidized manganese. The extent of this interference
can be determined by treating a sample with sodium arsenite
to destroy the bromine present so that the degree of
interference can be estimated.
Iodine and chlorine can also interfere, but these are not
normally present unless they have been added as sanitizers.
Test Procedures
BROMINE, Tablet, DPDSmart3 Test Procedures 09.16
Page 97
PROCEDURE A:
BROMINE (NO CHLORINE)
1. Press and hold until colorimeter turns on.
2. Press
ENTER
to select Testing Menu.
3. Select All Tests (or another sequence containing 016 Bromine T DPD) from Testing Menu.
4. Scroll to and select 016Bromine T DPD from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select Scan Blank.
7. Remove tube from colorimeter. Add one DPD #1 IG Tablet (6903A). Cap tube and
and shake for 10 seconds. Invert slowly 5 times. Solution will turn pink if bromine is
present.
8. Insert tube into chamber, close lid and select Scan Sample.
9. Press to turn colorimeter off or press the
EXIT
to exit to a previous menu or
make another menu selection.
PROCEDURE B:
BROMINE IN THE PRESENCE OF CHLORINE
1. Press and hold until colorimeter turns on.
2. Press
ENTER
to select Testing Menu.
3. Select All Tests (or another sequence containing 016 Bromine T DPD) from Testing Menu.
4. Scroll to and select 016 Bromine-LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber close lid and select Scan Blank.
7. Remove blank from colorimeter. Add 5 drops of Glycine Solution (6811). Cap and
mix.
8. Add one DPD#1 IG Tablet (6903A). Cap tube and shake for 10 seconds. Invert slowly
5 times. Solution will turn pink if bromine is present. Insert tube into chamber, close
lid and select Scan Sample.
9. Press to turn colorimeter off or press the
EXIT
to exit to a previous menu or
make another menu selection.
Smart3 Test Procedures 01.18 BROMINE, Tablet, DPD
Test Procedures
Page 98
PROCEDURE C:
FREE AVAILABLE, TOTAL AVAILABLE & COMBINED CHLORINE IN
THE PRESENCE OF BROMINE
1. Perform the test for Free and Combined Chlorine.
2. Perform the test for bromine in the presence of chlorine.
Calculations:
Residual Bromine (ppm) = Reading BR
Free Chlorine in the Presence of Bromine = Free Chlorine - 0.45 (Reading BR)
Total Chlorine in the Presence of Bromine = Total Chlorine - 0.45 (Reading BR)
Combined Chlorine in the Presence of Bromine = Total Chlorine - Free Chlorine
NOTE: Combined chlorine is not affected by the presence of bromine, so the calculation
is the same as when only chlorine is present.
Test Procedures
Smart3 Test Procedures 09.16 BROMINE, Tablet, DPD
Page 99
BROMINE - UDV
DPD METHOD–UNIT DOSE VIALS · CODE 4311-J
QUANTITY CONTENTS CODE
100 Free Chlorine Unit Dose Vials, 20 pouches 4311-J
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the
contents list or in the test procedures. Omit any letter that follows or precedes the
four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain
a printed copy, contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
Equipment needed but not supplied:
STANDARD ACCESSORY PACKAGE · CODE 1961
1Package of 3 Vials (empty)0156
1Syringe, 3 mL, plastic1184
1Foil Storage Bag9467
Or:
ADVANCED ACCESSORY PACKAGE · CODE 1962
1Pipettor, 3mL, w/tip30529-BRAND
1Cuvette Rack31695
1Package of 3 Vials (empty)0156
1Foil Storage Bag9467
Like chlorine, bromine is an effective germicidal agent employed in drinking water
treatment, pool and spa water sanitization, food service sanitation, and other public
health applications.
Smart3 Test Procedures 11.18 BROMINE, UDV
Test Procedures
Page 100
APPLICATION:Drinking, surface, and saline waters; swimming pool water;
domestic and industrial waters and wastes.
RANGE:0.0–22.0 ppm Bromine
MDL: 0.10 ppm
METHOD:In buffered sample bromine reacts with diethyl-p-phenylene
diamine (DPD) to produce a pink-red color in proportion to the
concentration of bromine present.
SAMPLE HANDLING
& PRESERVATION:
Bromine in aqueous solutions is not stable, and the bromine
content of samples or solutions, particularly weak solutions,
will rapidly decrease. Exposure to sunlight or agitation will
accelerate the reduction of bromine present in such solutions.
For best results start analysis immediately after sampling.
Samples to be analyzed for bromine cannot be preserved or
stored.
INTERFERENCE:The only interfering substance likely to be encountered in
water is oxidized manganese. The extent of this interference
can be determined by treating a sample with sodium arsenite
to destroy the bromine present so that the degree of
interference can be estimated.
Iodine and chlorine can also interfere, but these are not
normally present unless they have been added as sanitizers.
Test Procedures
BROMINE, UDV Smart3 Test Procedures 11.18
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.