Packaging & Delivery ······································································5
General Precautions ········································································5
Safety Precautions ···········································································5
Limits of Liability ·············································································5
Limited Warranty··············································································6
Limitations························································································6
Specifications ···················································································6
Statistical and Technical Definitions
Related to Product Specifications··················································· 7-8
Contents and Accessories································································· 8
EPA Compliance ··············································································8
CE Compliance·················································································9
CHEMICAL TESTING
Water Sampling for Chemical Analysis ··········································· 10-11
Filtration ··························································································11
An Introduction to Colorimetric Analysis ······································ 12
Reagent Blank ·················································································13
Colorimeter Tubes ···········································································13
Meter Care························································································13
Selecting an Appropriate Wavelength ············································· 13-14
Calibration························································································14-16
Calibration Curves ···········································································14-16
Standard Additions ··········································································16-17
Sample Dilution Techniques & Volumetric Measurements ············ 17
Interferences ····················································································18
Stray Light Interference ···································································18
OPERATION OF THE SMART 2 COLORIMETER
Overview ·························································································19
Power Source ····················································································19
Components ····················································································20
Quick Start ······················································································21-22
GENERAL OPERATING PROCEDURES
The Keypad ······················································································23
Sample Holders·················································································23
The Display & the Menus ······························································· 24-25
Looping Menus ·················································································26
TESTING
Testing Menu ··················································································· 27
Sequences of Tests ···········································································28
General Testing Procedures ····························································· 29
Testing With the Pre-Programmed Tests ········································· 29-30
SMART2 COLORIMETER 1.073
T ABLE OF CONTENTS (cont.)
Calibrating LaMotte Pre-Progammed Tests······································ 31-32
Measuring in the %T/ABS Mode ····················································· 33-34
EDITING MENU
Edit a Sequence ···············································································35-36
Adding or Deleting Tests ·································································36-38
Edit User Tests ·················································································39
Naming the Test ··············································································40-41
Selecting the Vial and Wavelength ·················································· 42
Entering a New Calibration ····························································· 43-44
Selecting the Numerical Format of the Result ································ 45
Selecting Units of Concentration····················································· 46
Setting the Clock··············································································47
Turning the Data Logger On and Off ··············································· 48
Factory Setup····················································································49
Setting the Power Saver Function ···················································· 49
Replacing the Battery ······································································· 50
MAINTENANCE
Cleaning ···························································································51
Meter Care························································································51
Meter Disposal··················································································51
Experienced packaging personnel at LaMotte Company assure adequate
protection against normal hazards encountered in transportation of shipments.
After the product leaves the manufacturer, all responsibility for its safe delivery
is assured by the transportation company. Damage claims must be filed
immediately with the transportation company to receive compensation for
damaged goods.
Should it be necessary to return the instrument for repair or servicing, pack
instrument carefully in suitable container with adequate packing material. A
return authorization number must be obtained from LaMotte Company by
calling 1-800-344-3100. Attach a letter with the authorization number to the
shipping carton which describes the kind of trouble experienced. This valuable
information will enable the service department to make the required repairs
more efficiently.
GENERAL PRECAUTIONS
n
Before attempting to set up or operate this instrument it is important to read
the instruction manual. Failure to do so could result in personal injury or
damage to the equipment.
The SMART2 Colorimeter should not be stored or used in a wet or corrosive
environment. Care should be taken to prevent water or reagent chemicals from
wet colorimeter tubes from entering the colorimeter chamber.
NEVER PUT WET TUBES IN COLORIMETER.
SAFETY PRECAUTIONS
n
Read the labels on all LaMotte reagent containers prior to use. Some
containers include precautionary notices and first aid information. Certain
reagents are considered hazardous substances and are designated with a * in the
instruction manual. Material Safety Data Sheets (MSDS) are supplied for these
reagents. Read the accompanying MSDS before using these reagents.
Additional emergency information for all LaMotte reagents is available 24
hours a day from the Poison Control Center listed in the front of the phone
book. Be prepared to supply the name and four-digit LaMotte code number
found on the container label or at the top of the MSDS. LaMotte reagents are
registered with a computerized poison control information system available to
all local poison control centers.
Keep equipment and reagent chemicals out of the reach of young children.
Protect Yourself and Equipment: Use Proper Analytical Techniques
LIMITS OF LIABILITY
n
Under no circumstances shall LaMotte Company be liable for loss of life,
property, profits, or other damages incurred through the use or misuse of its
products.
SMART2 COLORIMETER 1.075
LIMITED WARRANTY
n
This instrument is guaranteed to be free from defects in material and workmanship
for a period of two (2) years from original purchase date. In the event that a defect
is found during the warranty time frame, LaMotte Company agrees that it will be
repaired or replaced without charge except for the transporation costs. This
guarantee does not cover batteries.
This product can not be returned without a return authorization number from
Lamotte Company. For warranty support or a Return Authorization Number,
contact LaMotte Company at 1-800-344-3100 or tech @ lamotte.com.
n
LIMITATIONS
This guarantee is void under the following circumstances:
• Damage due to operator negligence, misuse, accident or improper application.
• Damage or alterations from attempted repairs by an unauthorized
(non-LaMotte) service.
• Damage due to improper power source, AC adapter or battery.
• Damage caused by acts of God or natural disaster.
• Damage occurred while in transit with a shipping carrier.
LaMotte Company will service and repair out-of warranty products at a nominal
charge.
SPECIFICATIONS
n
n
INSTRUMENT TYPE: Colorimeter
ReadoutGraphical 4 line, 16 character per line LCD
Wavelengths430nm, 520 nm, 570 nm, 620 nm
Wavelength Accuracy ±2
Readable ResolutionDetermined by reagent system
Wavelength Bandwidth 10 typical
Photometric Range–2 to + 2AU
Photometric Precision± 0.001AU at 1.0AU
Photometric Accuracy ± 0.005AU at 1.0AU
Sample ChamberAccepts 25 mm diameter flat-bottomed test tubes, 10 mm square cuvettes, 16 mm COD test tubes
Light Sources4LEDs
Detectors4 silicon photodiodes with integrated interference filters
ModesAbsorbance, pre-programmed tests
Pre-Programmed Tests YES, with automatic wavelength selection
User Defined TestsUp to 10 user tests can be input
RS232 Port8 pin mini-DIN, 9600b, 8, 1, n
Power RequirementsBattery Operation: 9 volt alkaline, Line Operation: 110/ AC; 50/60 Hz with adapter, 6V 500 mA DC
Dimensions (LxWxH) 8.5 x 16.2 x 6.7 cm, 3.4 x 6
Weight312 g, 11 oz (meter only)
Data Logger350 test results stored for download to a PC
6SMART2 COLORIMETER 10.07
.4 x 2.6 inches
STATISTICAL AND TECHNICAL DEFINITIONS RELATED
n
TO PRODUCT SPECIFICATIONS
Method Detection Limit (MDL): “The method detection limit (MDL) is
defined as the minimum concentration of a substance that can be measured
and reported with 99% confidence that the analyte concentration is greater
than zero and is determined from analysis of a sample in a given matrix
containing the analyte.”
almost all cases when dealing with a limit of detection or limit of
determination, the primary purpose of determining that limit is to stay away
from it.’”
2
1. CFR 40, part 136, appendix B
2. Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and
L Vanatta, American Laboratory, Sept 2003, P. 31.
Precision: Precision is the numerical agreement between two or more
measurements.
3
The precision can be reported as a range for a measurement
(difference between the min and max). It can also be reported as the standard
deviation or the relative standard deviation. It is a measure of how close
together the measurements are, not how close they are to the correct or true
value. The precision can be very good and the accuracy very bad. This is a useful
measure of the performance of a test method.
3. Skoog, D.A., West, D. M., Fundamental of Analytical Chemistry, 2
Holt Rinehart and Winston, Inc, 1969, p. 26.
Accuracy: Accuracy is the nearness of a measurement to the accepted or true
4
value.
The accuracy can be expressed as a range, about the true value, in
which a measurement occurs (i.e. ±0.5 ppm). It can also be expressed as the %
recovery of a know amount of analyte in a determination of the analyte (i.e.
103.5 %). This is a useful measure and what most customers are interested in when
they want to know about the performance of a test method.
4. Skoog D.A., West D. M., Fundamental of Analytical Chemistry, 2
Holt Rinehart and Winston, Inc, 1969, p. 26.
Resolution: Resolution is the smallest discernible difference between any two
measurements that can be made.
places are displayed. (i.e. 0.01). For titrations and various comparators it is the
smallest interval the device is calibrated or marked to (i.e. 1 drop = 10 ppm,
0.2 ppm for a DRT, or ±half a unit difference for an octaslide or color chart).
Note that the resolution many change with concentration or range. In some
cases the resolution may be less than the smallest interval, if it is possible to
make a reading that falls between calibration marks. This is often done with
various comparators. One caveat is, that resolution has very little relationship to
accuracy or precision. The resolution will always be less than the accuracy or
precision but it is not a statistical measure of how well a method of analysis works.
The resolution can be very very good and the accuracy and precision can be very, very
bad! This is not a useful measure of the performance of a test method.
5. Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and
L Vanatta, American Laboratory, Sept 2003, P. 34.
1
Note that, “As Dr. William Horwitz once stated, ‘In
nd
ed.,
nd
ed.,
5
For meters this is usually how many decimal
SMART2 COLORIMETER 1.077
Sensitivity: Sensitivity is the resolution based on how this term is used in
LaMotte catalogs. This term is not listed in any of the references. Sometimes it
is used for detection limit. It is a confusing term and should be avoided.
Repeatability: Repeatability is the within-run precision.
6
A run is a single
data set, from set up to clean up. Generally, one run occurs on one day.
However, for meter calibrations, a single calibration is considered a single run
or data set, even though it may take 2 or 3 days.
6. Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s Textbook of
Quantitative Chemical Analysis, 5
th
ed., Longman Scientific & Technical,
1989, p. 130.
Reproducibility: Reproducibility is the between-run precision.
7. Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s Textbook of
Quantitative Chemical Analysis, 5
th
ed., Longman Scientific & Technical,
7
1989, p. 130.
CONTENTS AND ACCESSORIES
n
n
CONTENTS
SMART2 Colorimeter
Test Tubes, with Caps
Power Supply, 110/220V
SMART2 Colorimeter Quick Start Guide
SMART2 Colorimeter Manual
n
ACCESSORIES
COD AdapterCode 5-0087
UDV AdapterCode 5-0086
Small Field Carrying CaseCode 1919-GCS150
Large Field Carrying CaseCode 1919-BCS440
SMARTLink2 Program & Interface Cable (3.5 disk) Code 1912-3
SMARTLink2 Program & Interface Cable (CD)Code 1912-CD
EPA COMPLIANCE
n
The SMART2 Colorimeter is an EPA-Accepted instrument. EPA-Accepted
means that the instrument meets the requirements for instrumentation as
found in test procedures that are approved for the National Primary Drinking
Water Regulations (NPDWR) or National Pollutant Discharge Elimination
System (NPDES) compliance monitoring programs. EPA-Accepted
instruments may be used with approved test procedures without additional
approval.
8SMART2 COLORIMETER 1.07
CE COMPLIANCE
n
The SMART2 Colorimeter has earned the European CE Mark of
Compliance for electromagnetic compatibility and safety.
I, the undersigned, hereby declare that the equipment specified above
conforms to the above Directive and Standards.
Signature
1/15/02
Date
SMART2 COLORIMETER 1.079
Scott H. Steffen
Name
VP New Products & Quality
Position
CHEMICAL TESTING
WA TER SAMPLING FOR CHEMICAL ANALYSIS
n
n
Taking Representative Samples
The underlying factor to be considered for any type of water sampling is
whether or not the sample is truly representative of the source. To properly
collect a representative sample:
l
Sample as frequently as possible.
l
Collect a large sample or at least enough to conduct whatever tests are
necessary.
l
Make a composite sample for the same sampling area.
l
Handle the sample in such a way as to prevent deterioration or
contamination before the analysis is performed.
l
Perform analysis for dissolved gases such as dissolved oxygen, carbon
dioxide, and hydrogen sulfide immediately at the site of sampling. These
factors, as well as samples for pH, cannot be stored for later examination.
l
Make a list of conditions or observations which may affect the sample.
Other considerations for taking representative samples are dependent
upon the source of the sample. Taking samples from surface waters
involves different considerations than taking samples from impounded and
sub-surface waters.
n
Sampling of Open Water Systems
Surface waters, such as those found in streams and rivers, are usually well
mixed. The sample should be taken downstream from any tributary, industrial
or sewage pollution source. For comparison purposes samples may be taken
upstream and at the source of the pollution before mixing.
In ponds, lakes, and reservoirs with restricted flow, it is necessary to collect a
number of samples in a cross section of the body of water, and where possible
composite samples should be made to ensure representative samples.
To collect samples from surface waters, select a suitable plastic container with a
tight fitting screw cap. Rinse the container several times with the sample to be
tested, then immerse the container below the surface until it is filled to
overflowing and replace the cap. If the sample is not to be tested immediately,
pour a small part of the sample out and reseal. This will allow for any
expansion. Any condition which might affect the sample should be listed.
Sub-surface sampling is required to obtain a vertical profile of streams, lakes,
ponds, and reservoirs at specific depths. This type of sampling requires more
sophisticated sampling equipment.
For dissolved oxygen studies, or for tests requiring small sample sizes, a Water
Sampler (LaMotte Code 1060) will serve as a subsurface or in-depth sampler.
10SMART2 COLORIMETER 1.07
This weighted device is lowered to the sampling depth and allowed to rest at
this depth for a few minutes. The water percolates into the sample chamber
displacing the air which bubbles to the surface. When the bubbles cease to rise,
the device has flushed itself approximately five times and it may be raised to
the surface for examination. The inner chamber of the sampling device is lifted
out and portions of the water sample are carefully dispensed for subsequent
chemical analysis.
A Snap-Plunger Water Sampler (LaMotte Code 1077) is another “in-depth”
sampling device which is designed to collect large samples which can be used
for a multitude of tests. Basically , this collection apparatus is a hollow cylinder
with a spring loaded plunger attached to each end. The device is cocked above
the surface of the water and lowered to the desired depth. A weighted
messenger is sent down the calibrated line to trip the closing mechanism and
the plungers seal the sample from mixing with intermediate layers as it is
brought to the surface. A special drain outlet is provided to draw off samples for
chemical analysis.
n
Sampling of Closed System
To obtain representative samples from confined water systems, such as pipe
lines, tanks, vats, filters, water softeners, evaporators and condensers, different
considerations are required because of chemical changes which occur between
the inlet and outlet water. One must have a basic understanding of the type of
chemical changes which occur for the type of equipment used. Also,
consideration should be given to the rate of passage and retaining time for the
process water.
Temperature changes play an important part in deciding exactly what test
should be performed. Process water should be allowed to come to room
temperature, 20–25°C, before conducting any tests.
When drawing off samples from an outlet pipe such as a tap, allow sample to
run for several minutes, rinsing the container several times before taking the
final sample. Avoid splashing and introduction of any contaminating material.
FILTRATION
n
When testing natural waters that contain significant turbidity due to suspended
solids and algae, filtration is an option. Reagent systems, whether EPA,
Standard Methods, LaMotte or any others, will generally only determine
dissolved constituents. Both EPA and Standard Methods suggest filtration
through a 0.45 micron filter membrane, to remove turbidity, for the
determination of dissolved constituents.** To test for total constituents,
organically bound and suspended or colloidal materials, a rigorous high
temperature acid digestion is necessary.
**LaMotte offers a filtering apparatus: syringe assembly (Code 1050) and membrane
filters, 0.45 micron, (Code 1103).
SMART2 COLORIMETER 1.0711
AN INTRODUCTION TO COLORIMETRIC ANALYSIS
n
Most test substances in water are colorless and undetectable to the human eye.
To test for their presence we must find a way to “see” them. The SMART2
Colorimeter can be used to measure any test substance that is itself colored or
can be reacted to produce a color. In fact a simple definition of colorimetry is
“the measurement of color” and a colorimetric method is “any technique used
to evaluate an unknown color in reference to known colors”. In a colorimetric
chemical test the intensity of the color from the reaction must be proportional
to the concentration of the substance being tested. Some reactions have
limitations or variances inherent to them that may give misleading results.
Many such interferences are discussed with each particular test instruction. In
the most basic colorimetric method the reacted test sample is visually
compared to a known color standard. However, accurate and reproducible
results are limited by the eyesight of the analyst, inconsistencies in the light
sources, and the fading of color standards.
To avoid these sources of error, a colorimeter can be used to photoelectrically
measure the amount of colored light absorbed by a colored sample in reference
to a colorless sample (blank).
White light is made up of many different colors or wavelengths of light. A
colored sample typically absorbs only one color or one band of wavelengths
from the white light. Only a small difference would be measured between white
light before it passes through a colored sample versus after it passes through a
colored sample. The reason for this is that the one color absorbed by the
sample is only a small portion of the total amount of light passing through the
sample. However, if we could select only that one color or band of wavelengths
of light to which the test sample is most sensitive, we would see a large
difference between the light before it passes through the sample and after it
passes through the sample.
The SMART2 Colorimeter passes one of four colored light beams through one
of four optical filters which transmits only one particular color or band of
wavelengths of light to the photodectector where it is measured. The difference
in the amount of colored light transmitted by a colored sample is a
measurement of the amount of colored light absorbed by the sample. In most
colorimetric tests the amount of colored light absorbed is directly proportional
to the concentration of the test factor producing the color and the path length
through the sample. However, for some tests the amount of colored light
absorbed is inversely proportional to the concentration.
The choice of the correct wavelength for testing is important. It is interesting
to note that the wavelength that gives the most sensitivity (lower detection
limit) for a test factor is the complementary color of the test sample. For
example the Nitrate-Nitrogen test produces a pink color proportional to the
nitrate concentration in the sample (the greater the nitrate concentration, the
darker the pink color). A wavelength in the green region should be selected to
analyze this sample since a pinkish-red solution absorbs mostly green light.
12SMART2 COLORIMETER 1.07
REAGENT BLANK
n
Some tests will provide greater accuracy if a reagent blank is determined to
compensate for any color or turbidity resulting from the reagents
themselves. A reagent blank is performed by running the test procedure on
10 mL of demineralized water . Use sample water to SCAN BLANK.Insert
the reagent blank in the colorimeter chamber and select SCAN SAMPLE.
Note result of reagent blank. Perform the tests on the sample water as
described. Subtract results of reagent blank from all subsequent test results.
NOTE: Some tests require a reagent blank to be used to SCAN BLANK.
COLORIMETER TUBES
n
Colorimeter tubes which have been scratched through excessive use should
be discarded and replaced with new ones. Dirty tubes should be cleaned on
both the inside and outside. Fingerprints on the exterior of the tubes can
cause excessive light scattering and result in errors. Handle the tubes
carefully, making sure the bottom half of the tube is not handled.
LaMotte Company makes every effort to provide high quality colorimeter
tubes. However, wall thicknesses and diameter of tubes may still vary
slightly. This may lead to slight variations in results (e.g. if a tube is turned
while in the sample chamber, the reading will likely change slightly). To
eliminate this error put the tubes into the sample chamber with the same
orientation every time.
The tubes that are included with the colorimeter have an index mark to
facilitate this. If possible, use the same tube to SCAN BLANK and SCANSAMPLE.
METER CARE
n
The optical system of the SMART2 must be kept clean and dry for optimal
performance. Dry the colorimeter tubes before placing them in the
chamber to avoid introducing moisture. For best results store the
instrument in a area that is dry and free from aggressive chemical vapors.
SELECTING AN APPROPRIATE WAVELENG TH
n
The most appropriate wavelength to use when creating a calibration curve
is usually the one which gives the greatest change from the lowest reacted
standard concentration to the highest reacted standard concentration.
However, the absorbance of the highest reacted standard concentration
should never be greater than 2.0 absorbance units. Scan the lowest and
highest reacted standards at different wavelengths using the absorbance
mode to find the wavelength which gives the greatest change in
absorbance without exceeding 2.0 absorbance units. Use this wavelength to
create a calibration curve.
SMART2 COLORIMETER 1.0713
Below is a list of suggested wavelengths for the color of the reacted samples.
Use these as a starting point.
CALIBRATION
n
Sample
Color
Yellow430
Pink520
Red570
Green and Blue620
Wavelength
Range
As with all pre-calibrated meters, it is highly recommended, even if not
required by regulations, that the user periodically verify the performance of the
meter by running standards with a predetermined concentration. Results
outside of specification are an indication that the meter needs to be adjusted.
This can be done following the user calibration described on page 31. If the
user calibration fails to properly adjust the meter then the meter should be
returned to LaMotte Company for recalibration. (See page 5).
CALIBRATION CURVES
n
The Smart2 Colorimeter contains tests for the LaMotte reagent systems (see
Page 49). The first step in using a non-LaMotte reagent system with your
Smart2 Colorimeter is to create a calibration curve for the reagent system. To
create a calibration curve, prepare standard solutions of the test factor and use
the reagent system to test the standard solutions with the Smart2 Colorimeter .
Select a wavelength for the test as described above.
Plot the results (in ABS or %Transmittance) versus concentration to create a
calibration curve. The calibration curve may then be used to identify the
concentration of an unknown sample by testing the unknown, reading
Absorbance or %T, and finding the corresponding concentration from the
curve. The linear range of the reagent system can be determined and this
information can be used to input a User Test into the Smart2 Colorimeter (see
EDIT USER TESTS, page 36).
n
PROCEDURE
1. Prepare 5 or 6 standard solutions of the factor being tested. The
concentration of these standards should be evenly distributed throughout
the range of the reagent system, and should include a 0 ppm standard
(distilled water). For instance, the solutions could measure 0, 10%, 30%,
50%, 70%, and 90% of the system’s maximum range.
2. T urn on the Smart2 Colorimeter. Select the appropriate wavelength from
the absorbance mode. Be sure to select the appropriate wavelength for the
color produced by the reagent system.
14SMART2 COLORIMETER 1.07
3. Use the unreacted 0 ppm standard to standardize the colorimeter by using
it to scan blank.
4. Following the individual reagent system instructions, react each standard
solution beginning with 0 ppm. Continue with standards in increasing
concentration. Record the reading and the standard solution
concentration on a chart. Readings can be recorded as percent
transmittance (%T) or absorbance (A).
5. Plot results on graph paper or computer using any available plotting
program. If results are as %T versus concentration, semilog graph paper
must be used. Plot the standard solution concentrations on the horizontal,
linear axis, and the %T on the vertical, logarithmic axis. If results are as
absorbance versus standard solution concentration, simple linear graph
paper can be used. Plot the standard solution concentration on the
horizontal axis, and the absorbance on the vertical axis.
6. After plotting the results, draw a line, or curve, of best fit through the
plotted points. The best fit may not connect the points. There should be
approximately an equal number of points above the curve as below the
curve. Some reagent systems will produce a straight line, while others
produce a curve. Many computer spreadsheet programs can produce the
curve of best fit by regression analysis of the standard solution data.
NOTE: Only reagent systems which produce a straight line can be used for a
User Test.
A sample of each type of graph appears below:
CALIBRATION CURVE
Absorbance vs. Concentration
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
Absorbance
0.4
0.2
0.0
123456
Concentration in ppm
SMART2 COLORIMETER 1.0715
89107
%T Transmission
CALIBRATION CURVE
%T vs. Concentration
100
10
1
0
123456
Concentration in ppm
891070
n
PREPARING DILUTE STANDARD SOLUTIONS
Standard solutions should be prepared to create a calibration curve. Standard
solutions can be prepared by diluting a known concentrated standard by
specified amounts. A chart or computer spreadsheet can be created to
determine the proper dilutions. Use volumetric flasks and volumetric pipets for
all dilutions.
1. In Column A – Record the maximum concentration of test as determined
by the range and path length.
2. In Column B – Record the percent of the maximum concentration the
standard solution will be.
3. In Column C – Calculate the final concentration of the diluted standard
solutions by multiplying the maximum concentration (In Column A) by
B
the % of maximum concentration divided by 100. (C = A x
100
).
4. In Column D – Record the final volume of the diluted sample (i.e. volume
of volumetric flask).
5. In Column E – Record the concentration of the original standard.
6. In Column F – Calculate the milliliters of original standard required
D
(C x
=F).
E
A sample chart appears below:
Final
B
I 00
DEF=Cx
Concentration
Volume of
Standard
of Original
Standard
Standard
Required
ABC=Ax
Maximum
concentration
of test
10.0 ppm909.0 ppm100 mL1000 ppm0.90 mL
10.0 ppm707.0 ppm100 mL1000 ppm0.70 mL
10.0 ppm505.0 ppm100 mL1000 ppm0.50 mL
10.0 ppm303.0 ppm100 mL1000 ppm0.30 mL
10.0 ppm101.0 ppm100 mL1000 ppm0.10 mL
10.0 ppm00 ppm100 mL1000 ppm0 mL
%of
Maximum
concentration
concentration
of Diluted
Standard
mL of
Original
D
E
STAND ARD ADDITIONS
n
A common method to check the accuracy and precision of a test is by standard
additions. In this method a sample is tested to determine the concentration of
the test substance. A second sample is then “spiked” by the addition of a
known quantity of the test substance. The second sample is then tested. The
determined concentration of the spiked sample should equal the concentration
of the first plus the amount added with the spike. The procedure can be
repeated with larger and larger “spikes.” If the determined concentrations do
not equal the concentration of the sample plus that added with the “spike”,
then an interference may exist.
For example, a 10.0 mL water sample was determined to contain 0.3 ppm iron.
To a second 10.0 mL sample, 0.1 mL of 50 ppm iron standard was added. The
16SMART2 COLORIMETER 1.07
concentration of iron due to the “spike” was (0.10 mL x 50 ppm)/10.0 mL =
0.50 ppm. The concentration of iron determined in the spiked sample should
be 0.3 + 0.5 = 0.8 ppm iron. (Note: any error due to the increased volume from
the “spike” is negligible).
LaMotte offers a line of calibration standards which can be used to generate
calibration curves and perform standard additions.
SAMPLE DILUTION TECHNIQUES
n
& VOLUMETRIC MEASUREMENTS
If a test result using the Smart2 Colorimeter gives an over range message then
the the sample must be diluted. The test should be repeated on the diluted
sample to obtain a reading which is in the concentration range for the test.
(Note: This is not true for colorimetric determination of pH.)
Example:
Measure 5 mL of the water sample into a graduated cylinder. Add
demineralized water until the cylinder is filled to the 10 mL line. The sample
has been diluted by one-half, and the dilution factor is therefore 2. Perform the
test procedure, then multiply the resulting concentration by 2 to obtain the
test result.
The following table gives quick reference guidelines on dilutions of various
proportions. All dilutions are based on a 10 mL volume, so several dilutions
will require small volumes of the water sample. Graduated pipets should be
used for all dilutions.
Size of Sample
10 mL0 mL1
5mL5mL2
2.5 mL7.5 mL4
1mL9mL10
0.5 mL9.5 mL20
Deionized W ater to Bring
Volume to 10 mLMultiplication Factor
If the above glassware is not available, dilutions can be made with the
colorimeter tube. Fill the tube to the 10 mL line with the sample then transfer
it to another container. Add 10 mL volumes of demineralized water to the
container and mix. Transfer back 10 mL of the diluted sample to the tube and
follow the test procedure. Continue diluting and testing until a reading, which
is in the concentration range for the test, is obtained. Be sure to multiply the
concentration found by the dilution factor (the number of total 10 mL volumes
used).
Example:
10 mL of sample is diluted with three 10 mL volumes of demineralized water;
the dilution factor is four.
SMART2 COLORIMETER 1.0717
INTERFERENCES
n
LaMotte reagent systems are designed to minimize most common interferences.
Each individual test instruction discusses interferences unique to that test. Be
aware of possible interferences in the water being tested.
The reagent systems also contain buffers to adjust the water sample to the ideal
pH for the reaction. It is possible that the buffer capacity of the water sample
may exceed the buffer capacity of the reagent system and the ideal pH will not
be obtained. If this is suspected, measure the pH of a reacted distilled water
reagent blank using a pH meter. This is the ideal pH for the test. Measure the
pH of a reacted water sample using the pH meter. If the pH is significantly
different from the ideal value, the pH of the sample should be adjusted before
testing.
Interferences due to high concentration of the substance being tested, can be
overcome by sample dilution (see page 16)
STRAY LIGHT INTERFERENCE
n
When scanning samples in 16 mm tubes, such as COD, the sample chamber lid
can not be closed. The COD adapter minimizes stray light. To further reduce
stray light interference, do not scan sample in direct sunlight.
18SMART2 COLORIMETER 1.07
OPERA TION OF THE
SMART2 COLORIMETER
OVERVIEW
n
The SMART2 Colorimeter is a portable, microprocessor controlled, direct
reading colorimeter. It has a graphical 4 line, 16 character liquid crystal display
for graphical, alphabetical and numerical messages. The operation is controlled
with the keypad through menu driven software in response to selections shown
on the display.
The test library consists of 100 LaMotte tests (not all 100 may be available at
present) and 10 “User Tests”. The LaMotte tests are precalibrated for LaMotte
reagent systems. The colorimeter displays the results of these tests directly in
units of concentration. The 10 “User Tests” may be used to enter additional
calibrations. All of these tests may be arranged in any of 3 sequences. These
sequences can be modified a limitless number of times to meet changing testing
needs.
The optics feature 4 different colored LEDs. Each LED has a corresponding
silicon photodiode with an integrated interference filter. The interference
filters select a narrow band of light from the corresponding LED for the
colorimetric measurements. The microprocessor automatically selects the
correct LED/photodiode combination for a test.`
A RS-232 serial port on the back of the colorimeter, and optional software,
allows the SMAR T2 to be interfaced with an IBM compatible personal
computer for real time data acquisition and data storage. This port also allows
an interface with a RS-232 serial printer.
Due to its portability, alternate power sources, and rugged construction, the
SMART2 Colorimeter is ideal for lab and field use.
POWER SOURCE
n
The SMART2 Colorimeter uses a 500 mA AC adapter. Please refer to the
Parts List for the code number for the correct adapter.
USE OF ANY AC ADAPTER OTHER THAN THE ONE SPECIFIED
FOR USE WITH THE Smart2 COLORIMETER MA Y DAMAGE THE
METER AND WILL VOID THE WARRANTY . Do not use the adapter
sold with the original SMART Colorimeter.
To use the adapter , slide the connector pin from the AC adapter into the small
hole on the left side of the meter. Plug the AC adapter into an appropriate wall
socket or power source.
SMART2 COLORIMETER 1.0719
COMPONENTS
p
n
Figure 1 shows a diagram of the Smart2 Colorimeter and its components.
Top View
SMART2
••••••••••••••••••
*
ENTER
OFFEXIT
ON
Bottom View
Serial
Number
Battery
Compartment
Side Views
Lid
AC
ter Socket
Ada
20SMART2 COLORIMETER 1.07
Figure 1
RS232
Serial Port
QUICK START
n
Some quick instructions to get into testing.
1. Press ON to turn on the SMART2. The
LaMotte logo screen will appear for about 2
seconds and then the Start screen appears. Press
Q/ENTER to start testing.
2. The Main Menu will appear. Press
Q/ENTER to select TESTING MENU.
3. Press Q/ENTER to select All Tests.
4. Press t or s to move the * to the desired
test.
VER 1.0
Smart2
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
ALL TESTS
* 001 Alk - UDV
002 Aluminum
003 Ammonia - NLF
5. Press Q/ENTER to select test.
6. Insert blank, press Q/ENTER to scan blank.
7. The screen will display Blank Done for
about 1 second.
SMART2 COLORIMETER 1.0721
ALL TESTS
* 015 Chlorine
016 Cl F-UDV
017 Cl Liq-DPD
015 Chlorine
* Scan Blank
015 Chlorine
Blank Done
* Scan Blank
8. Insert the reacted sample. Press Q/ENTER to
scan sample. The SMART2 will scan the sample
and display the concentration.
015 Chlorine
* Scan Sample
9. After recording test result, scroll with t or s
and make another selection with Q/ENTER.
Press EXIT to escape to previous menus.
015 Chlorine
1.28 ppm
* Scan Sample
22SMART2 COLORIMETER 1.07
GENERAL OPERATING PROCEDURES
The operation of the SMART2 Colorimeter is controlled by a microprocessor.
The microprocessor is programmed with menu driven software. A menu is a list
of choices. This allows a selection of various tasks for the colorimeter to
perform, such as, scan blank, scan sample, and edit test sequences. The keypad
is used to make menu selections which are viewed in the display. There are
three selections accessible from the MAIN MENU: Testing Menu, EditingMenu and PC Link.
THE KEYP AD
n
The keypad has 6 buttons which are used to perform specific tasks.
ON
tThis button will cause the display to scroll down through a list of
sThis button will cause the display to scroll up in a list of menu
ENTER
Q
EXIT
OFF
SAMPLE HOLDERS
n
The sample chamber is designed for 25 mm round tubes. Additional sample
holders for 16 mm COD tubes and for 1 cm square UDV cuvettes are available
for the SMART2 Colorimeter.
Position the COD adapter in the SMART2 chamber so that the grooves in the
adapter are aligned with the ridges located at the rear of the chamber. The
adapter should be inserted with the small hole, containing the ball plunger, at
the top. The ball plunger can be adjusted with a small screwdriver to control
the tightness of the fit of the tube in the adapter.
This button is used to turn the colorimeter on.
menu choices. It will move through a list viewed in the display. It
will auto scroll when held down.
choices. It will move through a list viewed in the display. It will
auto scroll when held down.
This button is used to select the menu choice adjacent to the “*”in
amenuviewedinthedisplay.
This button is an exit or escape button. When pressed, the display
will exit from the current menu and go to the previous menu.
This button turns the colorimeter off.
SMART2 COLORIMETER 1.0723
THE DISPLAY & THE MENUS
n
The display allows menu selections to be viewed and chosen. These choices
instruct the colorimeter to perform specific tasks. The menus are viewed in the
display using two general formats which are followed from one menu to the
next. Each menu is a list of choices or selections.
There are four lines in the display. The top line in each menu is a title or
pertinent instruction. The top line does not change unless a new menu is
selected. The second and third lines are used in two ways. One way is to display
menu choices. The second way takes advantage of the graphical capabilities of
the display. Both lines are used to display important messages, such as test
results, in a large, easy to read format. The fourth line is used for menu choices.
DISPLAY
TESTING MENU
* FIRST CHOICE
SECOND CHOICE
ANOTHER
AND ANOTHER
AND SO ON
TITLE or INSTRUCTION
MENU CHOICE WINDOW
Think of the menu choices as a vertical list in the display which moves up or
down each time an arrow button is pressed. This list or menu is viewed through
a window, the menu choice window, in the display. The menu choice window
is the lower 2 or 3 lines of the display. Pushing the arrow buttons brings
another portion of the menu into menu choice window. This is referred to as
scrolling through the menu.
TESTING MENU
* FIRST CHOICESECOND CHOICEANOTHER
SECOND CHOICE* ANOTHERAND ANOTHER
ANOTHERAND ANOTHER* AND SO ON
AND ANOTHERAND SO ONLAST CHOICE
AND SO ONLAST CHOICE
LAST CHOICE
t
TESTING MENU
t
TESTING MENU
An asterisk, “*”, will start in the far left position of the top line in the menu
choice window. As the menu is scrolled through, different choices appear next
to the “*”. The “*” in the display corresponds with the Q/ENTER button.
Pushing the Q/ENTER button selects the menu choice which is adjacent to
the “*” in the menu choice window.
24SMART2 COLORIMETER 1.07
The second general format of the display takes advantage of the graphics
capabilities of the display. The top line of the display is still a title line. The
middle two lines of the display are used to display important messages, results
or graphics in a large, easy to read format. The menus work in the same way as
described previously but only one line of the menu is visible at the bottom of
the display.
TESTING MENU
TESTING MENU
t
TESTING MENU
t
Result or MessageResult or MessageResult or Message
* ANOTHER* AND ANOTHER* AND SO ON
AND ANOTHERAND SO ONLAST CHOICE
AND SO ONLAST CHOICE
LAST CHOICE
As described previously, the EXIT button allows an exit or escape from the
current menu and a return to the previous menu. This allows a rapid exit from
an inner menu to the main menu by repeatedly pushing the
Pushing
OFF at any time will turn the colorimeter off.
EXIT button.
SMART2 COLORIMETER 1.0725
LOOPING MENUS
n
Long menus, such as All Tests, incorporate a looping feature which allow
the user to quickly reach the last choice in the menu from the first choice. In a
looping menu the last choices in the menu are above the first choice and
scrolling upward moves through the menu in reverse order. Scrolling downward
moves through the menu from first choice to last but the menu starts over
following the last choice. So all menu choices can be reached by scrolling in
either direction. The diagrams below demonstrate a looping menu.
AND SO ONAND ANOTHERANOTHER
: : :AND SO ONAND ANOTHER
: : :: : :AND SON ON
THIRD TO LAST: : :: : :
SECOND TO LASTTHIRD TO LAST: : :
LAST CHOICESECOND TO LASTTHIRD TO LAST
TESTING MENU
* FIRST CHOICE* LAST CHOICE* SECOND TO LAST
SECOND CHOICEFIRST CHOICELAST CHOICE
ANOTHERSECOND CHOICEFIRST CHOICE
AND ANOTHERANOTHERSECOND CHOICE
AND SO ONAND ANOTHERANOTHER
: : :AND SO ONAND ANOTHER
: : :: : :AND SO ON
LAST CHOICE: : :: : :
TESTING MENU
s
TESTING MENU
s
26SMART2 COLORIMETER 1.07
TESTING
TESTING MENU
n
The Testing Menu is used to run all LaMotte pre-programmed tests, USER
TESTS and Absorbance tests at one of four wavelengths. Testing from any of
three sequences can also be done.
1. Press the ON button to turn on the
SMART2 Colorimeter. The LaMotte logo
will appear for about 2 seconds and the the
Start screen appears. Press the Q/ENTER
button to begin testing.
2. The MAIN MENU will appear. Press the
Q/ENTER button to select Testing Menu.
3. Scroll with the t or s buttons and make
a selection with the Q/ENTER button.All Tests has all the available tests. The
three sequences have selected tests and
Absorbance has %T/ABS tests.
VER 1.0
Smart2
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
Sequence 3
Absorbance
SMART2 COLORIMETER 1.0727
SEQUENCES OF TESTS
n
SEQUENCE 1, SEQUENCE 2,andSEQUENCE 3 are alterable sequences.
They may be edited using the Editing Menu. Any of the LaMotte
pre-programmed tests or User Tests may be placed in these sequences in
whatever testing order that is preferred. Some examples of typical sequences
are given below.
SEQUENCE 1SEQUENCE 2SEQUENCE 3
* 015 Chlorine* 002 Aluminum* 003 Ammonia-N LF
079 Phosphate H035 Cyanide032 Cu-DDC
009 Bromine-LR041 Fluoride064 Nitrate-N L
076 pH TB053 Iron Phen067 Nitrite-N L
061 Moly-HR055 Manganese L074 pH CPR
086 Silica Hi064 Nitrate-N L078 Phosphate L
045 Hydrazine067 Nitrite-N L085 Silica Lo
032 Cu-DDC077 Phenol
051 Iron Bipyr078 Phosphate L
090 Sulfide-LR
These alterable sequences allow a series of tests to be setup that are run
frequently. The order of the individual tests in the sequence is determined by
the user. After running a test, use the * button to select the next test in the
sequence. Continue this pattern until the entire sequence has been completed.
All Tests is a fixed sequence containing the LaMotte pre-programmed tests,
User Tests, and Absorbance tests.
Modification of the alterable sequences is accomplished through the EditingMenu. This menu is explained in greater detail in EDITING MENU (p. 32).
Pressing the EXIT button while in a sequence menu will escape back to the
Testing Menu.
Pressing the OFF button at any time will turn the colorimeter off.
28SMART2 COLORIMETER 1.07
GENERAL TESTING PROCEDURES
n
The following are some step by step examples of how to run tests from the
Testing Menu. These test procedures are designed to be used with LaMotte
SMART Reagent Systems.
LaMotte Company continuously updates the list of pre-programmed tests as the
calibrations become available. Pre-programmed calibrations can be added to
the SMART2 Colorimeter in the field. A Windows-based computer running a
Windows Operating System and an 8 pin mini-DIN/9 pin F D-submin serial
cable (order Code 1771) are required.
Call LaMotte Technical Services at 1-800-344-3100 (410-778-3100 outside
the USA) or email at tech@lamotte.com for a current list of available
calibrations and downloading instructions.
TESTING WITH THE LaMOTTE
n
PRE-PROGRAMMED TESTS
Press ON to turn on the Smart2 Colorimeter.
The LaMotte logo will appear for about 2 seconds
and then the Start screen appears. Press the
Q/ENTER button to start testing.
The MAIN MENU will appear. Press the
Q/ENTER button to select Testing Menu.
Press the Q/ENTER button to select AllTests.
Press the t button to move to the 002
Aluminum to *.
VER 1.0
Smart2
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
ALL TESTS
* 001 Alk - UDV
002 Aluminum
003 Ammonia - NLF
SMART2 COLORIMETER 1.0729
Press the Q/ENTER button to select
002 Aluminum.
ALL TESTS
* 002 Aluminum
003 Ammonia - NLF
004 Ammonia - NLS
The SMART2 Colorimeter is ready to scan at the
correct wavelength. Place the blank in the sample
chamber, close the lid and press the Q/ENTER
button to scan blank.
NOTE: Do not keep the button depressed.
The screen will display Blank Done for about
1 second. Scan Sample will be positioned next
to *.
Place the reacted sample in the chamber, close
the lid and press the Q/ENTER button to scan
sample. The colorimeter will scan the sample and
the results screen will appear.
Record test result. To repeat the test, press the
Q/ENTER button to scan the sample again. The
last blank scanned is used to zero the colorimeter
for repeated scans. A different blank can be used
by pressing the s buttontoscrollbackto ScanBlank and then scanning another blank. Scroll
with the t or s buttons and make another
selection with the Q/ENTER button. The %T or
Absorbance of the last test can be viewed by
choosing %T/Abs. Press the EXIT button to
escape to previous menus.
NOTE: The menus loop in this screen so either
the s or t buttons will lead to the menu
selection needed.
002 Aluminum
* Scan Blank
002 Aluminum
Blank Done
* Scan Blank
002 Aluminum
* Scan Sample
002 Aluminum
0.09 ppm
* Scan Sample
Next Test
Previous Test
%/Abs
Calibrate
Scan Blank
30SMART2 COLORIMETER 1.07
Loading...
+ 228 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.